Test Report

Water Activity Meter


Model EZ-200

Date : 20 September 2024


Company Name :
PT Ultra Sakti
__________________________________________

Address :

Jl. Raya Narogong KM 12


__________________________________________

Cikiwul, Kec. Bantar Gebang


__________________________________________

Bekasi
_________________________________________

Service Signature : Checked by Signature :


Engineer
(Customer)

( M Ridwan ) ( Jeffri )
Date : 20 September 2024 Date : 20 September 2024

1
EZ-200 Basic Function Test Report
Operational Qualification (OQ) Validation Procedures, Criteria and
Result.

Name Serial No.

Water Activity Meter EZ-200 180 105

No Item Description Judgement Note

Visual check Check the appearance and the interior of


1 Pass / Fail Check
of main unit the instrument visually.

(1) Display function test


Pass / Fail Check
Must be able to display each function.

(2) Key pad test


Pass / Fail Check
2 Function Each key must allow input

(3) Buzzer
Pass / Fail Check
Can ‘Beep’ normally

(4) Calender and time shall correctly. Pass / Fail Check


3 Sensor (1) No damage by visually. Pass / Fail
Check
(2) Can detect the temperature in the
chamber. (Temperature difference
between termohygro and main unit Pass / Fail Check
±1ᵒC)
Termohygrometer = S/N TH 19

4 Fan Can operate normal Pass / Fail Check


Confirm by connection main unit and
5 Connection Pass / Fail Check
main unit can operate properly.

2
Measurement Check
Judgement Note
No Standard Suhu Main Nilai Aw
Unit (ᵒC)

C1 0.328 25,0 0,331 Pass / Fail


0,328
± 0.02
0.328 0,333 Pass / Fail
25,0

C2 0.529 25,1 0,533 Pass / Fail


0,529
± 0.02
0.529 25,1 0,533 Pass / Fail

C3 0.753 25,1 0,755 Pass / Fail


0,753
± 0.02
0.753 25,1 0,756 Pass / Fail

C4 0.930 24,9 0,936 Pass / Fail


0,930
± 0.02
0.930 24,9 0,933 Pass / Fail

3
Suhu Selisih Suhu
Termohigrometer Main Unit
Standard Suhu Main Tollerance Judgement
(ᵒC) dengan Suhu
Unit (ᵒC) (ᵒC)
Termohigrometer
S/N = TH 19 (ᵒC)

25,0 25,0 PASS / FAIL


0,0
C1 + 1ᵒC

25,0 25,0 PASS / FAIL


0,0

25,1 25,2 - 0,1 PASS / FAIL


C2 + 1ᵒC

25,1 25,2 - 0,1 PASS / FAIL

25,1 24,9 + 0,2 PASS / FAIL


C3 + 1ᵒC

25,1 25,0 + 0,1 PASS / FAIL

24,9 24,9 0,0 PASS / FAIL


C4 + 1ᵒC

24,9 24,9 0,0 PASS / FAIL