Technical data
Accessories
METTLER TOLEDO
AM/PM/SM
Balances and scales
A first glance at the M balances and scales
• Animal weighing
0.00
target weight target weight target weight target weight • Data I/O – controlling the balance or scale using a specified
9 command set
Precision, weighing convenience and absolute dependa- • Additive weighing Page 1.2
bility are features you as the owner of a METTLER TOLEDO – Totalization of weight values
balance or scale are accustomed to. • Density determination 1.3
• Differential weighing 1.4
But be honest: – Backweighing
Are you so well acquainted with their capabilities that you – Moisture content and dry weight determination
can also make optimum use of them? • Dispensing – Weighing-in to a target weight – Formula weighing 1.7
– Mixing of substances
The aim of the following section is to provide you with – Net total weighing
information and suggestions on how you can make espe- – Formula weighing by weight and by percent
cially effective use of METTLER TOLEDO technology to • Weight control 1.11
solve your special weighing task. – Simple tolerance check with the METTLER DeltaTrac
– Plus/minus weighing
You will see: – Fill process control, pharmacopeia
Even a METTLER TOLEDO balance or scale can improve its – Statistical quality control
performance! • Calculating with weight data and constants 1.15
• Statistical evaluation 1.16
Together with METTLER TOLEDO software and appropriate • Piece counting 1.17
peripheral units, complete, user-friendly and cost-saving
• Special weighing tasks, freely programmable 1.20
solutions also exist for quite specific application needs.
• Animal weighing – weighing in an unstable environment 1.21
• Weighing under difficult conditions 1.22
– Below-the-balance weighing
– Weighing in a contaminated environment
• Weighing with extremely high accuracy 1.23
1.1
Additive weighing 1.2
Additive weighing
Weigh – add one more part, weigh – add one more part, weigh – ...
You wish to determine individual weights of several components or parts, but without
having to remove the samples already weighed from the weighing pan.
You can achieve this with every M balance or scale by taring after reading off the result.
Particularly well suited to this type of weighing are METTLER DeltaRange balances and
scales, however, as after taring the ten times more accurate fine range is fully accessible
for every individual weighing.
0.0 But what should you do if you also need to know the total of the individual weights exactly?
0
9
Even this poses no problem:
One of the features of the Net total function of the LabPac-M and PharmaPac-M allows the
additive weighing of components and continuously calculates the total of the individual
weights.
Additive weighing can be used to particular effect in weight controls of several identical parts,
e.g. in the case of a pharmacopeia, and is equally well suited to the statistical evaluation
of individual weights and for use in piece counting.
Little acorns – great oaks: the Density Determination Kit from METTLER TOLEDO.
• Simple installation and
• sophisticated design
assure high reproducibility of the density determination of solids by the buoyancy method.
You can also determine the density of liquids by using a displacement body.
Incidentally:
For large bodies or liquid containers, you have the possibility of below-the-balance
weighing. You will find the appropriate information on page 1.22.
7.2
15
9
Automatic density calculation – always the right method!
Whether solid, liquid or porous samples – even more flexibility and convenience is offered
by the Density Determination Kit in combination with the LabWare “Density determination”
software package for the EPSON HX-20:
• uncomplicated user guidance (menu)
• density determination by the buoyancy method or with a pycnometer
• automatic density calculation
• calibration of the pycnometer or displacement body
• statistical evaluation (X, s) of measurement series
7.2
15
• detailed result record with date, time, article, code and method
9
See section 3 for a complete description of the capabilities of the METTLER TOLEDO LabWare
“Density determination”.
Differential weighing
Quality, shelf-life and stability of many products are decisively affected by their moisture
---------------------
DIFF WEIGHING
content.
SAMPLE PREP
--------------------- Are you already using the optimum method to check this important value?
05.05.90 08:16
24.9
96 1 StoT 12.138 g Take a minute to discover what METTLER TOLEDO can offer you here for quality assurance
9
1 StoW 24.996 g
and lab analysis!
---------------------
Backweighing – mere child’s play with every balance or scale of the M series!
The unit percent is built into every AM, PM and SM balance or scale as a basic function.
You set the initial weight of your sample as reference (100 %), process the sample and then
reload on the balance. The result appears directly in percent of the initial weight.
---------------------
DIFF WEIGHING • the individual tare weights and initial weights of 20 samples can be stored
RESULTS
---------------------
05.05.90 08:45 • the results of the backweighing can be displayed in 4 forms
- 47 1 Dif - 47.65 %
.6 5 % 1 Dif -11.911 g
• the samples can be processed and backweighed several times.
1 Abs 52.35 %
1 Abs 13.085 g
In combination with the GA44 Printer, the LabPac-M provides you with accurate records of
--------------------- all intermediate stages and results.
Incidentally:
During sample treatment, you can use the balance for other weighings!
Backweighing in routine operation with even greater flexibility!
Do you need a system that makes an efficient backweighing routine possible in any order
even with a large number of samples?
The METTLER TOLEDO LabWare “Moisture determination” runs on the EPSON HX-20.
See section 3 for a complete description of the capabilities of the LabPac-M and LabWare
“Moisture determination”.
Whether you wish to analyze chemicals or foodstuffs, coffee or wood – the LP16 Infrared
Dryer is equipped to handle all possibilities!
You define various parameters in advance in accordance with the analysis sample, for
instance
• drying time and temperature,
• evaluation program and
• type of record.
Thanks to clear key symbols and convenient user guidance, as easy as winking!
But it is not only in routine operation that you will appreciate the wide variety of possibilities
to configure the LP16 especially for each type of sample!
For routine work in the lab and in production, the simple compact LJ16 instrument is
available.
73 Via the data output built in as standard, the LP16 and LJ16 can be connected to the GA44
.3
5 Printer, which then records all results and settings.
%
Incidentally:
With the Thermometer Set available as an accessory, you can always check the sample
temperature.
The LP16 Application Brochure with its 50 examples provides you with information on the
optimum drying parameters for a wide range of situations.
You will find further details regarding the LP16 and LJ16 in section 4.
Dispensing – Weighing-in to a target weight – Formula weighing
You are looking to considerably ease the load in everyday weighing operations in the lab
or in production.
And in this connection your prime aims are presumably as follows:
• reliable weighing-in or dispensing within a reasonable time,
• dependable compliance with the formulation instructions in weight units or in percent,
• minimal reject quotas,
xfn
• error-free result recording and transfer of the weighing results.
jxfj
xfnsd
xfn jxfjcv
1422
nxy 5 12.12
4577 .1222
yys 7
16:32
253 h
jxfj 31.01
.1985
244
And you envisage that such routine work should progress in a simplified yet accelerated
12:50
h
artike
l
24
Kund
e :1
manner, naturally without meticulousness and precision falling by the wayside. You also
wish to keep an eye on the cost factor.
For example, setting the balance to “fine dispensing” where the display reacts particularly
fast to the weight change and you always have all decimal places in view.
Or the METTLER DeltaTrac, a dynamic graphic indicator with 60 segments. In its basic
function it marks the weighing range already in use. It is thus virtually impossible for you
to exceed the maximum range and no longer be able to read off the weight when mixing
substances or totalizing weight values.
For this you set the balance to %, weigh in the target weight once and set it as reference
(100 %). The DeltaTrac now changes into a plus/minus display with two tolerance vanes
and two pointers.
Each new weighing-in is now referred to the target weight. When the pointers are in the
“6 o'clock” position, the target weight has been reached with an accuracy of ± 0.25 %.
Incidentally:
For particularly precise additions, you have the LV3 Vibrospatula available.
100.
0 %
Mixing and formula weighing made simple!
With the “Net total” function of the LabPac-M or PharmaPac-M, you can
• accurately weigh in additively any number of components from zero,
• enter the target weight of each component,
• weigh in each component to the target weight using the DeltaTrac and
• call up the total of components already added or weighed as a net total.
And all this without any calculation!
Incidentally:
Totalization is also possible with the NetEasy-M software cassette, see GM303 Control Unit
in section 4.
And you wish to have all this on paper? With date, time and identification?
No problem! With the GA44 Printer you obtain all results as well in black and white, e.g. the
target weights and the actual weighed-in amounts of the individual components.
For you have the F03220 Formulation Program for MS-DOS computers available.
• 100 formulations and 200 different components (raw materials) and
• 20 components per formulation with tolerance details
can be stored and recorded in detail in formula weighing.
Incidentally:
Yet another case where you will find the LV3 Vibrospatula extremely useful.
An A4 printer can be used not only to provide result records, but also to print out appropriate
labels for the finished products.
See section 3 for a complete description of the capabilities of the LabPac-M, the Pharma-
Pac-M and FO3220.
Weight control
Built into every AM/PM/SM balance or scale as standard, this graphic indicator marks the
used and available weighing range.
If, on the other hand, the balance is set to the unit “%”, the target weight can be set as
reference (100 %). The DeltaTrac display now becomes a plus/minus display with two
tolerance vanes {{ }} . These mark the fixed tolerance range of ± 2.5 %.
Would you like to select the plus/minus tolerances freely, and perhaps have asymmetrical
tolerances? Or do you perhaps wish to even classify your products as a function of the
result?
There is no escape route – the legislative authorities stipulate exact filling tolerances!
Mean fill
weight
Make therefore a virtue out of necessity:
Economy in operation Look for a control system that, from the managerial economics point of view, can even bring
you a plus instead of the worrying additional costs.
Nominal fill
quantity Hard to believe?
Consumer protection
Take our word for it. You will discover that this need not necessarily be a contradiction in
terms!
On the following pages METTLER TOLEDO introduces you to a range of systems for fill
process control – from the effective, small system up to the complex control system.
Select the optimum performance class for your plant and your requirements!
Balance
sheet §
EC and
§ §
st ds
ar
-
Fill process control in compliance with the law – without an additional computer!
You produce pharmaceutical products, fill solids or liquids or package them. And you are
looking for a small, but powerful system for your checkweighings. For instance to
• check product data by random sampling right next to the filling machine and
• to summarize these samples in a records of totals, or also to
• be able to call up default values of the fill process control (e.g. violations of lower tolerance
limits) at any time.
Then you should take a closer look at the SQC12 StatPac-M1). It has been especially
designed for use as a small system in the field of fill process control and with pharmaco-
peia and needs no additional computer.
Incidentally:
With the appropriate setting, the balance can be used for normal weighings in between
152 sampling operations.
.0
9
In combination with the GM54 Output Module, you can sort your products or control
indicators (lamps etc.) as a function of the result.
See section 3 for a complete description of the capabilities of the SQC12 StatPac-M.
1)
SQC = Statistical Quality Control
In combination with a MS-DOS computer, the SQC33 can manage up to 300 different
articles and each article can be assigned a tolerance system (e.g. EC standards, pharma-
copeia).
And there is no need to worry – the operation is extremely simple thanks to menu guidance!
Incidentally:
Since the sampling operation runs automatically, it is possible to set up the balance and the
computer at separate locations.
Even more powerful features are offered by the METTLER TOLEDO control systems with
several weighing stations, e.g. the SQC52 quality control system.
Ask your METTLER TOLEDO dealer for more information!
497
.9
9
Calculating with weight data and constants
Do you have to convert your weighing results, for instance to determine the weight/m2 of
a sheet of paper or the density of a liquid?
Would you like to prepare a 0.1 M saline solution and weigh out the requisite 5.85 g salt
per liter water only approximately to save time? Yet you subsequently desire reliable
indication of the exact amount of water!
METTLER TOLEDO CalcPac-M, the pocket calculator at the balance, achieves this!
With this you can perform all basic mathematical functions, store weight values, constants
or intermediate results; you can even evaluate measurement series statistically (calculation
of mean value and standard deviation).
Do you weigh precious stones, gold or expensive chemicals and do you then wish to know
the price of the weighed material immediately?
* We recommend the METTLER TOLEDO GoldPac-M, especially for the jewelry and goldsmith
134.9 trades.
9 The selectable weight units are oriented to the special needs of these branches. Thus, you
can weigh precious stones in carats and gold in grams and then at a simple keystroke
effortlessly calculate, for instance, the material value of a piece of jewelry.
See section 3 for a complete description of the capabilities of the CalcPac-M and the
GoldPac-M.
Statistical evaluation
You wish to check the uniformity of tablets, mechanical parts, packages, etc. For this you
must be familiar with the two most important statistical quantities, mean value X and
standard deviation s.
Or you need the two values in order to quantify the scatter of the weighing results within a
series of measurements.
*
s 0.25
9 Just take a look at which METTLER TOLEDO products offer statistics functions for different
weigh-ing tasks – you can then forget paper, pencil and pocket calculator without further
ado!
• With StatEasy-M you can calculate mean value and standard deviation, see GM303
Control Unit in section 4.
• CalcPac-M calculates the total, mean value and standard deviation of the individual
n weights and shows the number of weighings.
1
x= n Σx
i=
i • Mean value, standard deviation and number of samples are available with the Count-
1 Pac-M as functions for quality assessment of the counted parts.
• The “Statistics” function of the PharmaPac-M and ProPac-M calculates X, s and n for
simple quality control in the lab or in production.
• SQC12 StatPac-M and other SQC systems for quality control provide not only X, s and
n n, but also the minimum and maximum value, their difference as well as the number of
s=
1
n- Σ )( x - x
i= 2
i
violations of the lower and upper tolerance limits.
• If you work with the LabWare “Density determination” software package with several
1 1
samples, you can also calculate X and s.
• In backweighings with the LabWare “Moisture determination”, you obtain the mean
value of up to 300 samples.
See section 3 for a complete description of the capabilities of the Pacs and software.
Piece counting
The balance determines the weight of a single part from a reference quantity. The net weight
and the piece weight are then used to calculate the total piece count. This offers enormous
27 stk
advantages compared with traditional counting methods thanks to
• marked time saving,
• greater accuracy and
• outstanding user convenience.
If you wish to count small parts very accurately, you need a high resolution balance. But
you need have no fear of high costs!
Thanks to the recallable fine range of METTLER DeltaRange balances and scales, you can
determine the piece count just as accurately as with a high resolution balance or scale of
the same capacity.
For simple applications, the very thing: a scale of the M series with piece counting built in
as standard.
The reference piece number 10 is programmed in firmware. After the reference has been
set, the sample to be counted is loaded and you can immediately read off the total piece
count on the scale display. That’s all there is to it!
Perhaps you need greater scope in regard to the reference piece number – with the
CountEasy-M software cassette you can expand the possibilities of your balance or scale,
see also GM303 Control Unit in section 4.
Whether you are packaging or portion counting tablets and capsules or counting and
batching piece goods – the solution now goes under the name PharmaPac-M for the lab
and ProPac-M for the production sector.
Both Pacs offer the “Piece counting” function; in addition to the fixed reference piece number
10, you now have available
• a free selection of the reference piece number and
26 stk • the possibility to optimize the reference weight.
This means:
marked easing of the load with one of the most frequent and nerve-racking routine operations
and at the same time greater accuracy in counting.
See section 3 for a complete description of the capabilities of the PharmaPac-M and the
ProPac-M.
Piece counting – for demanding tasks
Whether you are taking inventory or packaging piece goods – for counting tasks in these
areas the solution is called CountPac-M.
In the counting of numerous, extremely light parts, the accuracy of your counting balance
is possibly not sufficient.
Here, the CountPac-M allows a counting system with 2 balances:
With an attached reference balance, the piece weight can be read in by the counting balance
up to two decimal places more accurately. Thus, for example, with an AE, AM or AT balance
as reference balance, up to several million parts weighing 0.1 g can be counted.
Have you still not found a solution for your special application problem in the METTLER
TOLEDO product range?
Then make use of the innumerable possibilities stored by the XPac-M for the solution of
weighing tasks not found in day-to-day operations:
It is freely programmable and thus handles even out-of-the-way problems: For instance,
217.3 • with a program that allows the performance of aerosols to be tested, or
9
• with a program that allows an automatic start in animal weighing and automatic taring
7.215 after removal of the animal, or
9
• with a program that has ready a custom-tailored solution to your special application
program!
Incidentally:
* If you do not wish to write the program yourself, METTLER TOLEDO will be pleased to do it
107.0 for you!
9
See section 3 for a complete description of the capabilities of the XPac-M.
Animal weighing – weighing in an extremely unstable environment
– with the weighing technology from METTLER TOLEDO, you have complete mastery even
of unstable weighing samples!
The cycle time for the determination of a measured value can be adjusted stepwise.
Measurements are made during this time and then their mean value determined.
Weight fluctuations due to movements of the animals are thus irrelevant.
Even severe shocks in the vicinity can no longer influence the weighing results.
Incidentally:
This dynamic weighing is in principle possible with all scales of the M series.
Special animal weighing pans are available as accessories – see the “Accessories” section
for further information.
*
107.0
9
Below-the-balance weighing
Do you wish
• to weigh magnetic weighing samples,
• to weigh under exceptional ambient conditions (e.g. dust, vibrations, extremes of
217
.3
9
temperature) or
• to perform density determinations for large bodies or liquid containers?
Then make use of the possibilities offered by below-the-balance weighing!
For this, the weighing pan or platform is separated from the balance and suspended from
the balance hanger.
With most of the PM and SM balances and scales, the hanger is part of the standard
equipment or available as an accessory. See the table “Standard equipment” in section 5
for more information.
Hazardous substances at the scale location cause problems which require special
solutions:
• in such cases, weighing pan and weighing cell can be separated from the electronics
and the display thereby minimizing disposal problems.
The PM-SE Conversion Kit is available for PM balances up to 6.1 kg. See section 4 “METTLER
TOLEDO Optional accessories” for further information.
23.2
9
Weighing with extremely high accuracy
As a weighing professional
you are responsible for applications that require extremely high accuracy. You thus primarily
demand outstanding reproducibility from your balance.
You work
• with valuable weighing samples or heavy tare vessels – for example in quality control
or in the verification and calibration service.
You are familiar with the requirements needed to obtain optimum weighing results and thus
create the necessary conditions for this regarding the
• surroundings (e.g. vibration-free work room, atmospheric conditions as constant as
possible, avoidance of drafts) and
• operation (e.g. work in a constant time cycle, weighing sample acclimatized and not
500 electrostatically chargeable).
0.0
00
9 In this case you certainly appreciate the particular advantage of the METTLER TOLEDO
PM5003 Comparator: as an unparalleled balance of the METTLER TOLEDO PM range, it
also weighs accurately in the weight class 5 kg down to the milligram range.
The draft shield prevents air currents and helps keep the weighing chamber atmosphere
constant. The circular weighing pan helps ensure exact centering of the weighing sample.
You will find further information in the METTLER TOLEDO booklet “Weighing the Right Way”
(720906).
1.23
1.24
LEERSEITE
2. Result recording and data transfer
The AM/PM/SM balances and scales are fitted with two METTLER TOLEDO offers suggestions for problem solving in the
interface connections “Data I/O” and “GM” as standard. following applications: Page
The “Data I/O” connection can be used for the serial • Attachment of a printer – result recording 2.2
transfer of data to and from the balance or scale (bidirec- • Attachment of a computer – bidirectional communication 2.4
tional). • Hand or foot switch for the print and/or tare command 2.5
This connection facilitates the simple attachment of devices • Attachment of several balances or scales to a central 2.6
– with an RS232C interface data station
– with an active current loop data interface (METTLER • Acquisition and processing of bar-coded information 2.7
TOLEDO CL interface) or • Data transfer to devices with different interfaces 2.8
– with other interfaces via the appropriate METTLER • Control of peripherals by the balance or scale 2.9
TOLEDO interface converters. • Protection of the system against electrical power line 2.10
The connection is point-to-point; only one device can be disturbances
attached. • Overview of the connection possibilities 2.11
2.1
Attachment of a printer – result recording 2.2
The alphanumeric GA44 Thermal Printer for attachment to the CL interface is supplied with
a balance connection cable (1.5 m).
0.00
9
Data transfer is possible when the settings of the balance under “I-FACE” match the factory
settings (see operating instructions of the balance).
You will find further details regarding the GA44 Printer in section 4.
– with printers of other manufacturers including printers with an RS232C interface, e.g.
EPSON FX-80, LX-800
The print command can be triggered from the balance in the same manner as with the GA44
Printer.
For the attachment of a printer, METTLER TOLEDO offers the following cables:
SubD 25-pin, female 1.5 m 33995
SubD 25-pin, male 1.5 m 33640
Cable for EPSON P-40 1.5 m 33688
Extension cable (“MiniMettler” both ends) 2m 216151
Extension cable 5m 216152
Extension cable 10 m 216153
The command set for control of the balance by the computer is described in the operating
instructions “Bidirectional data interface of the PM balances”, see section 6.
DataPac-M
0.00
9 For the attachment of a computer, METTLER TOLEDO offers the following cables:
SubD 25-pin, male (e.g. IBM-XT) 1.5 m 33995
Balance SubD 9-pin, male (e.g. IBM-AT or laptop) 1.5 m 33783
Extension cable (“MiniMettler” both ends) 2m 216151
Extension cable 5m 216152
Extension cable 10 m 216153
Tare command
Attachment of the hand or foot switch to the GM interface of the balance
with the tare cable (length 30 cm) 33872
Hand or foot switch for the print and/or tare command 2.5
Attachment of several balances or scales to a central data station 2.6
CL250 The CL310 5-Channel Line Selector (Multiplexer) allows bidirectional data communication
(full duplex operation) of 5 serial data lines over 1 serial line.
Distance The balance is attached to the CL310 Multiplexer with the 47936 cable.
up to 1 km
CL310 For attachment of the CL310 Multiplexer to a computer with an RS232C interface, the
CL250 Interface Converter is needed.
Data I/O For system expansion, the CL310 Multiplexers can be connected in series if more than 5
instruments have to be attached.
CL310
0.00 9
Data I/O You will find further details regarding CL devices in section 4.
0.00 9
0.00 9
Acquisition and processing of bar-coded information
GA44
Printer A central weighing system for the acquisition of bar codes can comprise the following
units:
• balance (as appropriate, with DataPac-M terminal)
• bar code readers,
e.g. hand scanners (LS7000, LS8100) or wand readers (HP6000 series)
GA50
• central data station (computer)
• GA44 Printer
• GA50 Peripheral Controller
Hand
Wand reader
scanner
The GA50 takes over the editing of the coded information and the data distribution. The
individual units do not need their own data line to the central data station.
GM Data I/O You will find further details regarding the GA50 Peripheral Controller in section 4.
DataPac-M
0.00
For attachment to the GA50 Peripheral Controller, various cables are available from
9
METTLER TOLEDO (see “Cables and connectors” in section 4).
Balance
❏ Matching the data interfaces of the balance to other interfaces used in industry.
❏ Data transfer over long distances (up to 1 km).
Computer
RS232C The METTLER TOLEDO CL interface converters are available for this.
connection
The following interface standards can be matched to the CL interface:
Peripheral,
• BCD parallel CL240 CL-BCD Interface
e.g. CL-RS232C
Interface Adapter • IEEE488 (IEC bus, HP-IB) CL241 CL-IEEE488 Interface
CL cable for
• RS232C CL249 CL-RS232C Adapter
Mettle
r CL 250
communication
over • RS232C (selectable handshake) CL250 CL-RS232C Interface
considerable
distances The CL249 and CL250 Interface Converters are needed when
• distances between 15 m and 1000 m have to be covered. The RS232C interface is not
suitable for such distances. Or if
• CL peripheral units (e.g. CL310) have to be attached to an RS232C interface, e.g. for the
GM attachment of a CL310 Multiplexer to the RS232C interface of a computer.
Data I/O
GM You will find further details regarding the METTLER TOLEDO CL interface converters in section
4, “CL devices”.
0.00
9
To set up the cable connections, the CL extension cables (see section 4, “Cables and
connectors”) are available.
Control of peripherals by the balance or scale
0.00 9 There these signals are converted to switching commands for the attached devices, such
as pumps, motors or signal lamps.
In addition, the outputs of the GM54 Output Module can be controlled by a computer via the
serial interface.
GM54
Output Peripherals can thus be controlled
Module
• as a function of the weighing results or
8 digital outputs for • by the balance via the serial interface of a computer.
control of peripherals
You will find further details regarding GM54 in section 4.
The CL111 Power Line Filter offers excellent protection against power line disturbances of
a moderate to high level. It can be used for loads up to 2 A power consumption.
For protection against extremely high disturbance levels, use of the CL110 Power Line Filter,
which can provide up to 3.15 A current, is advisable.
In addition, it is also possible to switch the attached units on and off centrally.
You will find further details regarding the METTLER TOLEDO power line filter in section 4,
“CL devices”.
Power supply
0.00
9
Overview of the connection possibilities
Printer with
25-pin SubD, f METT
Digital LER
Converto Analog
ter
D/A Converter
33640
I/O
GA44 GA44
Printer extension
47926
I/O data
47936
interface
47936
CL extension
Reference balance
Mettle
r CL 31
0
Mettler
CL 240
GM303
Control Unit
GM extension
33784
Foot switch
I/O
GM
GM extension
33959
Tare cable
33872
Hand switch
GM54
Output Module
LV10 Automatic
Feeder to lamps, motors, pumps, etc.
3. Application Pacs and software • CalcPac-M Page 3.2
– for direct calculations with weight data
• CountPac-M 3.4
The Application Pacs are supplied in two forms: – a match for any piece counting operation
• DataPac-M 3.6
• as a separate entry keypad with cable for AM and – allows dialog with the computer
PM balances, • FlowPac-M 3.8
• as an application terminal together with the display – for mass flow measurement and control
for SM scales. • FO3220 3.10
– the powerful formulation program for your MS-DOS computer
• GoldPac-M 3.12
For high-capacity PM and SM scales or for AM and PM
– for simple price calculation and unit switching in the gold and
balances with the GM303 Control Unit, the following precious stone trades
software cassettes are available for simple applications: • LabPac-M 3.14
– the lab aid in differential weighing and formula weighing
• CountEasy-M • LabWare “Density determination” 3.16
• NetEasy-M – the program for convenient density determination
• StatEasy-M • LabWare “Moisture determination” 3.18
– the program for the simple determination of moisture content and dry weight
Further details can be found in section 4, “GM devices”.
• PharmaPac-M 3.20
– simplifies the production, control and packaging of pharmaceutical products
• ProPac-M 3.22
– supports the checking and packaging of your products
• SQC12 StatPac-M (for one article) 3.24
– controls fill quantities and pharmaceutical products
where tolerances have statutory limits
• SQC33 (for up to 300 articles) 3.26
– the statistics program for MS-DOS computers
• XPac-M 3.28
– programmable for your special requirements
3.1
3.2
CalcPac-M
Print command
F3 I Print
X 1 2 3 -
Clear ∑ Save Weight transfer
F4
∑+ 0 . = + and display
Totalization/statistics • addition or subtraction of individual weight values to or from the current total (∑+, ∑-)
• number of individual weighings n
• mean value X
• standard deviation s
Result recording • printout of values and results with date, time, code and identification number
• complete record of totals
Peripherals • GM units
• GA44 Printer
3.3
3.4
CountPac-M
Totalization/
Portion counting Entry of data and time
Result recording • printout of values and results with date, time, code and two identification numbers
3.5
3.6
DataPac-M
Function keys
Clearing of
Clear wrong entries
A B 7 8 9
C D 4 5 6
E F 1 2 3
Termination o entry
3.7
3.8
Measurement • measurement and recording of mass flows (weight decrease or increase per unit time) for the calibration
or certification of pumps, valves or measuring instruments
• determination of the efficiency of mass metering devices and filters
3.9
3.10
FO3220
****************************************************************************
No. Component Name Component Code Nominal Actual
--------------------------------------------------------------------------
12 Marble powder Batch 56 783.30 g 785.20 g
11 Marble chips Batch 78/9 366.70 g 365.90 g
4 Adhesive A Batch 67-78 29.20 g 29.10 g
8 Color yellow Ye-58.9 120.80 g 119.20 g
3 Additive 3 CD 45 151.70 g 153.20 g
2 Additive 2 YKL 67 498.30 g 493.60 g
16 Water 1050.00 g 1050.00 g
--------------------------------------------------------------------------
Net : 3000.00 g 2996.20 g
3.11
3.12
GoldPac-M
– for simple price calculation and unit switching in the gold and precious stone trades
Result recording • printout of values and results with date, time, code and identification number
• printout on totalization
3.13
3.14
LabPac-M
Formulations according
to percentage composition
ID Date Time
Reset key
Diff. % Net Reset
weighing formula total 7 8 9
Code
Clear and
Clear correction key
Storage
tare Target Comp. 4 5 6
Print command
Comp.
Store (%) Re-zero Print
weight Re-zero 1 2 3
Differential weighing Net total weighing Saving of values (e.g. target weight,
(formulations according set weight of 1st component)
to weight composition) to prevent loss on power outage
Differential weighing • storage of up to 20 tare weights and initial weights
• automatic weight transfer in backweighing in the same order
• result as difference in weight units or in % of the initial weight
• result as absolute value in weight units or in % of the initial weight
• baching of all samples after new treatment repeatable any number of times
• scale available for other weighings during sample processing
% formula • entry of the formulation target weight in the selected weight unit and entry of the components in % of the
target weight or
• automatic calculation of the formulation target weight after entry of the 1st component in % and its initial
weight
• weighing-in of all components from zero with help of the DeltaTrac display
• recall of the formulation target weight, filling of mixtures to target weight
• recall of current net total in weight units or in % of the target weight
Net total • weighing in of any number of components each from zero with the help of the DeltaTrac display
• net total (total of the component weights) recallable at any time
• filling with top-up components (e.g. liquid) from net total up to formulation weight (additive weighing)
• DeltaTrac display as weighing-in aid after entry of the target weight
• totalization of individual weights (without target weight entry)
Result recording • printout of values and results with date, time, code and identification number
3.15
3.16
EPSON HX-20
========================
Density determinat.
07.05.90 17:58
1 Liquids ------------------------
Article : Metal
2 Solids Code : porous
3 Calibration ------------------------
Porous solids
L.density 0.9982 g/ml
Sample 119.86 g
Buoyancy 37.87 g
Density 3.159 g/ml
========================
*** Mettler LabWare ***
Computer • EPSON HX-20 with memory expansion and cassette tape drive
Density determination liquids • by the buoyancy method with the aid of a displacement body
• by use of a pycnometer
Density determination solids • by the buoyancy method for normal or porous solids
• by use of a pycnometer
Result recording • printout of values and results with date, time, article and code by HX-20
3.17
3.18
– the program for the simple determination of moisture content and dry weight
1. -21.47% 2. -21.78%
Mean value -21.62%
========================
*** Mettler LabWare ***
Computer • EPSON HX-20 with memory expansion and cassette tape drive
Calculation • moisture content (difference between residue and initial weight in percent of initial weight, initial weight =
100 %)
• dry weight (residue in percent of initial weight, initial weight = 100 %)
• wet weight (initial weight in percent of residue, residue = 100 %)
• dry weight (difference between residue and initial weight in percent of residue, residue = 100 %)
• summary of batches
• mean value in multiple weighings
• evaluation of up to 300 samples
Result recording • record of initial weights with article name, batch, pan number, tare and initial weight
• backweighing record with article, batch, pan number, initial weight, residue, difference and result in
accordance with the selected method of calculation
3.19
3.20
PharmaPac-M
Print command
Ref. Results Re-zero Print
n pcs. n,x,s
1 2 3
Piece counting Net total weighing Save of value (e.g. piece weight, reference piece number)
to prevent loss on power outage
Display of the piece weight
in piece counting
Piece counting • counting and determination of unknown piece numbers
• with fixed, preset reference piece number Ref. 10 pcs.
• with any selectable reference piece number Ref. n pcs.
• reference optimization, i.e. recalculation of the piece weight from the current values
• piece weight recallable at a keystroke
• switching between piece count and weight
Net total • weighing-in of any number of components each from zero with the aid of the DeltaTrac display
• net total (total of the component weights) recallable at any time
• filling with top-up components (e.g. liquid) from net total up to formulation weight (additive weighing)
• DeltaTrac display as weighing-in aid after entry of the target weight
• totalization of individual weight values (without target weight entry)
Result recording • printout of values and results with date, time, code and identification number
3.21
3.22
ProPac-M
Plus/minus weighing • weighing-in aid with the DeltaTrac display after entry of the target value
• check on adherence to a target weight with selectable plus/minus tolerances
• quality control of finished parts using the weight
• automatic printout of the weight with selectable acceptance criteria
• individual weighing or additive weighing (weighed parts or amounts remain on the scale)
• control of outputs of the GM54 Output Module based on the weight situation in regard to tolerance limits
Result recording • printout of values and results with date, time, code and identification number
3.23
3.24
– controls fill quantities and pharmaceutical products where tolerances have statutory limits
3.25
3.26
F5: Print 'off' F6: Individual 'on' F7: Abort F9: End
Computer • MS-DOS computer with attached A4 or A3 printer
• acquisition and storage of production data of up to 300 articles for different sets of statistics
• computer and weighing station can be separartely installed from each other
Parameters • nominal weight (nominal fill quantity)
• 9 different tolerance systems, e.g. EC, US, Japan, pharmacopeia or 2 free tolerances
• plausibility limits - also asymmetrical - fixed or freely selectable
• individual or mean value tare
• additive weighing possible (parts already weighed remain on the scale)
• 4 different volume units in the appropriate weight unit by entry of the density
• variable sample size (max. 999, with individual tare 99, with pharmacopeia 150)
• selectable feed rate of LV10 as well as automatic discharge at end of sample
• adjustment message for control of the filling machine
• password protection for important functions, e.g. changes to certain data
• classification of the products with the aid of the GM54 Output Module
Record/evaluation • individual values, minimum, maximum and difference into sample record
• sample size, number of weighings
• mean value absolute and in % of the nominal weight
• standard deviation absolute and in % of the mean value
• number of tolerance violations absolute and in %
• graphical representation of the results (histogram)
• printout of an article catalog sorted in accordance with different features
Standard equipment • software on 3.5" diskettes
• operating instructions on diskette
Peripherals • LV10 Automatic Feeder
• GM54 Output Module
• GA44 Printer (with GA50 Peripheral Controller)
• Bar code reader at the weighing station (with GA50 Peripheral Controller)
• connection cables for various computer systems
3.27
3.28
XPac-M
Date Time
Switching back to
F1 F5 ID
Reset normal weighing mode
7 8 9
Code
Clear key
F2 F6
Clear
4 5 6 entry
Print command
F3 F7 EE
Entry of decimal exponent
1 2 3 Print
Acknowledgement and
F4 F8
Enter termination key
0 . - Save register
Result recording • printout of values and results with date, time, code and identification number
Sequence of program development • development of program in XPac-specific language with the editor
• translation of program with the XPac compiler
• transfer of program to cassette
3.29
1.24
LEERSEITE
4. Accessories Page
• Calibration weights 4.2
• Weighing containers 4.3
• METTLER TOLEDO Optional accessories
– Draft shield 4.5
– In-use covers, demo cases 4.6
– Density determination kit, displacement body, weighing tongs, 4.7
triangular holder, anti-theft device
– Printer accessories, fuses, PM-SE Conversion Kit 4.8
• CL devices
– CL240, CL241, CL249, CL250 Interface Converters 4.9
– CL310 5-Channel Line Selector, CL110, CL111 Power Line 4.10
Filters
• GA devices
– GA44 Thermal Printer 4.11
– GA37 D/A Converter, GA50 Peripheral Controller 4.12
• GM devices
– GM303 Control Unit, hand switch, foot switch 4.13
– GM48, GM49, GM53 Auxiliary Displays, GM54 Output Module 4.14
• LP devices
– LP16 and LJ16 Infrared Dryers, Thermometer Set 4.15
• LV devices
– LV3 Vibrospatula, LV10 Automatic Feeder 4.16
• SM terminals and accessories
– Base and application terminals 4.17
– SM-Stand, SM-WallMount Kit, PowerPac-SM, battery charger 4.18
• Cables and connectors 4.19
4.1
Calibration weights 4.2
Calibration weights
Class E2 ≤ NIST Class S (NIST = National Institute of Standards and Technology, Washington, D.C.)
Class F1 ≈ NIST Class M
Weighing containers
Beaker, PVC
1500 cm3 180 mm 150 mm 140 mm 270 g 11284
Animal weighing pan with conical top for PM balances of weighing range ≥ 400 g, stainless steel
1000 cm3 130 mm 220 mm 120 mm < 290 g 47644
Weighing boats
Material Ø Length Tare Number Order Number
U-shaped weighing pans (specially for LP16 and LJ16 Infrared Dryers)
Material Ø Tare Number Order Number
Glass 50 mm 10 g ± 1 mg 81 4506
Glass 70 mm 20 g ± 1 mg 81 2013
Carat pans
Content Material Ø Tare Order Number
Spherical draft shield of glass for analytical balances and carat scales 33520
with weighing pan Ø 80 mm, weighing pan of spherical glass
draft shield Ø 60 mm
Molded in-use cover for AM/PM balances with weighing pan 5 34151
Retainer ring for 34151 41156
Molded in-use cover for PM3000, PM6000, PM6 5 34153
Molded in-use cover for PM11, PM15, PM16, PM30, PM30000, PM34 2 56629
Molded in-use cover for PM11-N, PM15-N, PM16-N, PM30-K,
PM30000-K, PM34-K min. 2 56883
Dust cover for AM/PM balances with shallow glass draft shield
(34300) 33486
Molded in-use cover for SM weighing platforms 5 34216
In-use cover for SM base terminal SM-F or SM-L 5 34222
Molded in-use cover for GM48, GM49 Auxiliary Displays 1 45898
In-use cover for EPSON HX-20 59721
Demo cases
For the density determination of liquids, the displacement body (10 ml)
210260 is also needed.
Weighing tongs
of plastic, with adjustable pitch, opening 65 mm, length 200 mm 210421
Straight tweezers
with fiber tip, length 210 mm 70209
Triangular holder, for test tubes, syringes and containers with round base 210435
Screwdriver 50279
Anti-theft device
Fuses
• distance up to 10 m
23.2
9
• for PM balances up to 6.1 kg
(not possible with PM1200, PM2500, PM5003, PM6100)
CL devices
Interface converters
For the matching of the data transmission between the METTLER TOLEDO CL current loop data interface (20
mA) and BCD, IEEE488 and RS232C Interfaces.
Mettle
r CL 2
40
• CL240 BCD Interface:
Serial-parallel converter (standard N200) for bidirectional transmission via shielded, galvanically separa
ted wire connections.
Mettle
r CL 2
41
• CL241 IEEE488 Interface:
Interface for IEC bus and HP-IB for bidirectional transmission with galvanic separation.
Mett
ler C
L 24
9
• CL249 RS232C Interface:
Interface for bidirectional transmission with galvanic separation.
All METTLER TOLEDO Interface converters have a power supply connection with selectable voltage.
CL devices 4.9
CL devices 4.10
For the suppression of line disturbances and hence an increase in the functional reliability of electronic
equipment.
• CL110:
CL110 very high damping properties with 6 power outputs, 4 of which are filtered.
Load capacity: 3.15 A (filtered), 6.3 A (unfiltered).
On/off switching for all outputs together.
• CL111:
optimum suppression of common-mode interferences with 3 filtered outputs.
Load capacity: 2 A.
CL111
GA devices
GA44 Printer
GA devices 4.11
GA devices 4.12
For the graphical representation of the weight profile as a function of time with the aid of an analog-controlled
recorder (e.g. drying profile with the LP16 Dryer),
for the control of analog-controlled pumps in flow control with FlowPac-M,
for the continuous conversion of results and values from the RS232C serial interface of the balance or a
computer into a corresponding analog voltage signal.
METT
• RS232 input and output
Digita LER
l to A
nalog
Conve
rter • analog voltage output, analog current output
• 16-character, alphanumeric display
• power supply (selectable voltage)
• connection cable for Data I/0 of the scale is enclosed
For simpler and faster operation of built-in applications with AM/PM balances up to 6.1 kg,
for use with the following applications without separate keypad:
- net total (in combination with NetEasy-M)
- statistics (in combination with StatEasy-M)
- piece counting with variable reference (in combination with CountEasy-M)
Foot switch For attachment to the scale, an appropriate adapter cable is required:
Transfer adapter for triggering the print command 47473
Tare adapter for triggering the tare command 33872
GM devices 4.13
GM devices 4.14
Auxiliary displays
For the on/off switching of external loads, e.g. relays, valves, low-power motors, indicator lights.
• suitable for all PM balances with circular weighing pan, particularly for PM balances with a readability of
1 mg
- drying temperature adjustable between 50 °C...160 °C in 5° steps
- drying over a fixed period between 1 and 240 minutes, or
- drying up to a minimal weight decrease of the sample per time interval
- evaluation in % or g, as dry weight, moisture content or as ATRO value
- accuracy dependent on balance and amount of sample used
• power cable and connection cable to balance as well as 80 Al pans included in standard equipment
0.0
0 • data output via CL/RS232C Interface for the automatic recording of the settings, the drying procedure and
9
the final results (e.g. with GA44 Printer or with a recorder via the GA37 D/A Converter)
• power consumption: max. 470 W
LJ 16 Infrared Dryer
The simple compact instrument for routine work in the lab and in production.
LP devices 4.15
LV devices 4.16
LV devices
LV3 Vibrospatula
For the weighing-in of very fine substances with the LV3 Vibrospatula.
• flow rate adjustable
For the automatic feed of small weighing samples (Ø 3 - 33 mm), e.g. of pharmaceutical products or small
mechanical parts.
Base terminals
Application terminals
Terminals with built-in Pac keypad for application-oriented use with ProPac-M, CountPac-M, DataPac-M,
0.0 !
StatPac-M, XPac-M, LabPac-M, PharmaPac-M, CalcPac-M.
OFF
ON
MENU
TARE
SM-AF Application Terminal
- with fluorescent display
SM-Stand 34290
Standard equipment:
Stand complete with base, 2 shims, 2 screws
Standard equipment:
Wall fixture, 2 shims, 2 screws, extension cable (33784)
Key
MM15p 15-pin MiniMettler connector, for Data I/O and GM connection, m = male, 7pf/360 ° 7-pin DIN connector, female, contacts arranged in a circle, screw cap
f = female 2 x Ø 2 mm f 2 banana sockets, Ø 2 mm, hole spacing 8 mm
MMT15p 15-pin MiniMettler T-connector, male/female, for Data I/O and GM connection
SubD25p 25-pin subminiature connector (RS connector), m = male, f = female Transmission/RS handshake: u = unidirectional
SubD9p 9-pin subminiature connector (RS connector), m = male, f = female b = bidirectional
5p/240 ° 5-pin DIN connector, m = male, f = female, contacts arranged over 240 ° T = unidirectional but transfer command
7p/270 ° 7-pin DIN connector, m = male, f = female, contacts arranged over 270 ° All cables are shielded
Cables for serial data transmission (Data I/O connection of the balance)
RS232C for IBM/PC 1.5 m/ 4 MM15pm/Data SubD25pf b/u 33995 for all PCs with subD25pm connector
RS232C for IBM/AT 1.5 m/ 4 MM15pm/Data SubD9pf b/u 33783 for all PCs with subD9pm connector
RS232C for printer 1.5 m/ 4 MM15pm/Data SubD25pm b/u 33640 for all printers with subD25pf connector
CL cable for balance 1.5 m/ 4 MM15pm/Data 5pm/240° b/- 47936 also suitable for GA50
CL extension 0.5 m/ 4 5pf/240° 5pm/240° b/- 42554
CL extension 1.0 m/ 4 5pf/240° 5pm/240° b/- 42560 standard with CL310
CL extension 2.0 m/ 4 5pf/240° 5pm/240° b/- 42555 standard with CL240, CL241
CL extension 5.0 m/ 4 5pf/240° 5pm/240° b/- 42556
CL extension 15.0 m/ 4 5pf/240° 5pm/240° b/- 42557
CL extension 30.0 m/ 4 5pf/240° 5pm/240° b/- 42558
GA44 cable for balance 1.5 m/ 4 MM15pm/Data 7pf/270° T/- 47926 standard with GA44
GA44 extension 0.5 m/ 4 7pf/270° 7pm/270° T/- 42559
GA44 extension 1.5 m/ 4 7pf/270° 7pm/270° T/- 42561 standard with GA44
GA44 extension 5.0 m/ 4 7pf/270° 7pm/270° T/- 42562
GA44 extension 15.0 m/ 4 7pf/270° 7pm/270° T/- 42563
LP16 data cable 0.3 m/ 4 MM15pm/Data 5pm/240° b/- 13969 standard with LP16 and LJ16
Data I/O extension 2.0 m/15 MM15pm/Data MM15pf/Data b/b 216151 for the extension of all connections at the
Data I/O extension 15.0 m/15 MM15pm/Data MM15pf/Data b/b 216152 Data I/O output (CL and RS232C) of
Data I/O extension 10.0 m/15 MM15pm/Data MM15pf/Data b/b 216153 the AM/PM/SM balances and scales
Connection cable 1,5 m/15 MM15pm/Data 7pf/270° b/b 23618 for DL18, DL21, DL25
Designation Length/ Connector 1 Connector 2 Transmission/ Order Number Remarks
No. cores RS Handshake
Cables for the attachment of a reference balance for exact piece counting
AE as ref. for PM/SM 1.5 m/ 2 SubD25pm MMT15p/Data u/- 33956 Data flow SubD25-pin to MMT15p
AM/PM as ref. for PM/SM 1.5 m/ 2 MM15pm/Data MMT15p/Data u/- 33868 Data flow MM15pm to MMT15p
AM/PM/SM as ref. for MMR2.7 m/ 4 MM15pm/Data special/10pm b/- 504123 set with cable and software (ID/NetPac-M)
also for the attachment of AM/PM/SM
balances and scales to an ID terminal
Cable extensions for the connections at the GM output
Transfer cable 0.3 m/ 2 MMT15p/Data 2x Ø 2mm f -/- 47473 button for triggering the print command
Tare cable 0.3 m/ 2 MMT15p/GM 2x Ø 2mm f -/- 33872 button for external taring
LEERSEITE
5. Technical data
Page
• Overview of the AM/PM/SM balance and scale families 5.2
• Technical data for individual models 5.4
• General technical data 5.8
• Standard equipment 5.9
• Weighing ranges in secondary units 5.10
• Decimal places in secondary units 5.11
• Dimension diagrams 5.12
5.1
Overview of the AM/PM/SM balance and scale families 5.2
0.0
g
SM3000 Weighing capacity: 3.1 kg
SM6000 Weighing capacity: 6.1 kg
SM15000 Weighing capacity: 15.0 kg
SM1520 DeltaRange Fine range: 2.0 kg (Coarse range: 15.0 kg)
0.0
g
SM scales are also available in an intrinsically-safe version. Ask your METTLER TOLEDO dealer.
0.0
g
Readability 0.1 mg 0.1 mg 0.001 g 0.001 g 0.001 g 0.001 g 0.001 g 0.01 g 0.01 g
- Fine range (recallable) - - - - - - - 0.001 g 0.001 g
Weighing capacity 51 g 110 g 110 g 210 g 410 g 1200 g 5100 g 410 g 2100 g
- Fine range (recallable) - - - - - - - 80 g 500 g
Taring range (by subtraction) 51 g 110 g 110 g 210 g 410 g 1200 g 5100 g 410 g 2100 g
Reproducibility (s) 0.1 mg 0.1 mg 0.5 mg 0.5 mg 0.001 g 0.001 g 0.0015 g 0.003 g 0.003 g
- Fine range - - - - - - - 0.001 g 0.001 g
Linearity ± 0.2 mg ± 0.2 mg ± 0.002 g ± 0.002 g ± 0.002 g ± 0.002 g ± 0.01 g ± 0.005 g ± 0.005 g
- Fine range - - - - - - - ± 0.002 g ± 0.002 g
Sensitivity drift / °C (10 … 30 °C) 2 x 10-6 2 x 10-6 4 x 10-6 4 x 10-6 3 x 10-6 1.5 x 10-6 2 x 10-6 4 x 10-6 1.5 x 10-6
Stabilization time 1) 2.5/4/6 s 2.5/4/6 s 1.5/2/3 s 1.5/2/3 s 1.5/2/3 s 2.5/4/6 s 9/12/15 s 1.5/2/3 s 2.5/4/6 s
Update speed 0.13 s 0.13 s 0.13 s 0.13 s 0.13 s 0.13 s 0.13 s 0.13 s 0.13 s
Display 2) FD FD FD FD FD LCD LCD FD LCD
Result deviation
in inclined position (1:1000) 0.5 mg 0.5 mg 0.005 g 0.005 g 0.005 g 0.005 g 0.02 g 0.005 g 0.005 g
Weighing pan Ø 80 mm Ø 80 mm Ø 130 mm Ø 130 mm Ø 130 mm Ø 130 mm Ø 130 mm Ø 130 mm Ø 130 mm
Calibration weight 3) 50 g/E2 100 g/E2 100 g/F1 100 g/F1 200 g/F1 1000 g/E2 5000 g/E2 100 g/F1 1000 g/F1
Net weight 6.0 kg 6.6 kg 3.8 kg 3.8 kg 3.8kg 3.8 kg 6.6 kg 3.8 kg 3.8 kg
Balance housing (W x D x H) in mm 194 x 316 x 252 194 x 316 x 337 194 x 316 x 68 194 x 316 x 337 194 x 316 x 68
Power consumption 6 VA
Fusing 63 mA/220 V
125 mA/110 V
PM300 PM600 PM2000 PM4000 PM6100 PM4800 PM3000 PM6000 PM6
DeltaRange
1)
dependent on the setting of the vibration adapter _
2)
FD Fluorescent display, self-luminous
LCD Liquid crystal display, passive
3)
for noncertified version
0.0
PM11-K PM16-K PM30000-K PM34-K DeltaRange PM15-K PM30-K
PM11-N PM16-N PM34-N DeltaRange PM15-N PM30-N
g
1)
dependent on the setting of the vibration adapter _
2)
FD Fluorescent display, self-luminous: SM-F
LCD Liquid crystal display, passive: SM-L
3)
for noncertified version
4)
incl. base terminal
Basic unit of balance/scale, selectable 1) g, kg, lb, oz, ozt, tl, GN, dwt, ct, C.M., k
Switchable 2nd unit 1) g, kg, lb, oz, ozt, tl, GN, dwt, ct, C.M., k
Applications, selectable piece counting, +/- or % weighing, animal weighing
Digital display 7 digits
DeltaTrac 2) 60 segments
Power supply AM and PM balances up to 6.1 kg voltage selectable: 115/230 V, +15/-20 %, 50/60 Hz
Power supply high-capacity PM scales 100/115/200/230 V, +15/-20 %, 50/60 Hz
Power connection SM voltage selectable: 115/230 V, ±15 %, 50/60 Hz
supply via AC adapter: 11 VDC/0.1 A
battery operation: duration 8 h; charge time 16 h
Vibration adapter _ choice of 3 settings, optical display
Weighing process adapter ; choice of 4 settings, optical display
Stability detector Ο choice of 4 settings, optical display
Data interface bidirectional RS232C/CL passive 20 mA
• Baud rate 110...9600 baud
• Parity even, odd, mark, space
• Transmission asynchronous 7-bit ASCII
• Plug-in connection 15-pin MiniMettler socket
• METTLER TOLEDO GM interface 15-pin MiniMettler for connection of peripherals
Admissible ambient conditions
• Temperature 0 °C...40 °C
• Relative humidity 15 %...85 %
• Height below/above sea level -300 m...+600 m
• Vibration 0.3 m/s2
1)
• Degree of protection (following IEC 529) IP 54 see decimal places for secondary units
2)
dynamic graphic indicator and dispensing aid
SM scales are also available in an intrinsically-safe version. Ask your METTLER TOLEDO dealer.
Standard equipment
1)
no hanger possible
2)
AM50: shallow
AM100: deep
PM5003: deep
3)
without calibration weight
AM50 AM100 PM200 PM300 PM400 PM480 PM600 PM1200 PM2000 PM2500 PM5003 PM3000
PM100 PM480 (fine) PM2500 (fine)
Range in g 51 110 210 310 410 80 610 1200 2100 500 5100 3100
lb - 0.243 0.463 0.683 0.904 0.177 1.345 2.645 4.630 1.103 11.243 6.834
oz 1.80 3.88 7.41 10.93 14.49 2.827 21.51 42.33 74.07 17.63 179.90 109.35
ozt 1.640 3.536 6.752 9.97 13.18 2.573 19.61 38.58 67.51 16.07 163.97 99.67
tl 1.360 2.938 5.609 8.28 10.95 2.137 16.29 32.05 56.09 13.35 136.17 82.80
GN 787 1’697 3’240 4’784 6’327 1233 9’413 18’518 32’407 7’713 78’704 47’840
dwt 32.80 70.73 135.0 199.3 263.6 51.44 392.2 771.6 1’350 321.5 3’279 1’993
ct / k / C.M. 255 550 1’050 1’550 2’050 400 3’050 6’000 10’500 2’500 25’500 15’500
PM4000 PM4800 PM6000 PM11-N PM15-N PM30-K PM34-K SM3000 SM6000 SM15000 SM1520
PM4800 (fine) PM6100 PM16-N PM34-K (fine) SM1520 (fine)
PM6 PM30000-K
Range in g 4100 800 6100 11000 16000 32000 4000 3100 6100 15000 2000
lb 9.039 1.764 13.45 24.25 35.27 70.54 8.818 6.834 13.45 33.01 4.41
oz 144.6 28.27 215.1 388.0 564.4 1129 141.1 109.35 215.1 529.1 70.6
ozt 131.8 25.73 196.1 353.6 514.4 1029 128.6 99.67 196.1 482.2 64.30
tl 109.5 21.37 162.9 293.8 427.3 855 106.8 82.80 162.9 400.6 53.42
GN 63’27 12’342 94’137 - - - - - - - -
dwt 2’636 514.4 3’922 7’073 10’288 20’576 2’572 1’993 3’922 9’645 1’285
ct / k / C.M. 20’500 4’000 30’500 55’000 80’000 160’000 20’000 15’500 30’500 75’000 10’000
Decimal places in secondary units
1)
one decimal place less in coarse range
Dimension diagrams
208 24 316
190 169
85
AM50, AM100
95
68
157
194 204 96
PM100, PM200, PM400, PM480, PM1200, PM2500
24 316
169
194
193
185 ø 130
117
95
68
157 204 96
ø 130
PM300
ø 150
PM600, PM2000, PM6100
194 24 316
169
ø 170
PM4000, PM4800
85
68
157 204 96
PM3000, PM6000, PM6
194 24 316
314
183
230
85
68
157 204 96
332
316
85
68
157 204 96
330 360
235 (pan) 355 (pan)
130
78
37 249
13
322
19
350
79 236 35
16
33
8,8,8,8,8,8,8
70
85
331
46
126
20 195 23
80
SM3000, SM6000, SM15000, SM1520 with base terminal and stand
325
0° - 47°
hinged
8,8,8,8,8,
8,8
min. 281 / max. 318
234 45
323
86
20 195 59 80
236 73 293
min. 309 / max. 318 331
366
130 236 35
16
33
8,8,8,8,8,8,8
70
85
331
51
126
20 195 23
80
SM3000, SM6000, SM15000, SM1520 with application terminal and stand
326
0° - 53°
hinged
8,8,8,8,8,
8,8
min. 8 / max. 13 50
234 45
323
86
20 195 59 80
236 73 293
min. 324 / max. 356 331
leerseite
6. Documentation and literature references
For further information, you will find a list of METTLER TOLEDO publications on the following
pages:
Page
• Operating instructions 6.2
• Short-form operating instructions 6.3
• Technical documentation (Engineering support bulletins), 6.4
Technical information bulletins
• Specification bulletins and brochures 6.5
6.1
Documentation and literature references 6.2
Operating instructions
Order No. Order No. Order No. Order No. Order No. Order No. Order No. Order No. Order No.
German English French Spanish Italian Japanese Dutch Arabic Swedish
AM/PM balances (up to 6.1 kg) 702394 702395 702396 702397 702398 702399 703196 703197 704910
High-capacity PM scales 704683 704684 704685 704686 704687 704730 704731 - 704911
SM scales 703875 703876 703877 703878 703879 703880 703881 - -
Bidirectional data interface 702177 702178 702179 701280 702204 - - - -
PM-SE conversion kit 702565 702565 702565 702565 702565 - - - -
CalcPac-M 702558 702559 702560 702561 702562 - - - -
CountEasy-M 704712 704712 704712 704712 704712 - - - -
CountPac-M 702169 702170 702171 702172 702221 - - - -
DataPac-M 702177 702178 702179 702180 702204 - - - -
FlowPac-M 704141 704142 704143 - - - - - -
1)
F03220 704200 704200 704200 - - - - - -
GoldPac-M 702553 702554 702555 702556 702557 - - - -
LabPac-M 702157 702158 702159 702160 702231 703378 - - -
LabWare “Density determination” 702039 702039 702039 - - - - - -
LabWare “Moisture determination” 702040 702040 702040 - - - - - -
NetEasy-M 704713 704713 704713 704713 704713 - - - -
PharmaPac-M 702165 702166 702167 702168 702233 - - - -
ProPac-M 702161 702162 702163 702164 702232 - - - -
SQC12 StatPac-M 702173 702174 702175 702176 702492 - - - -
StatEasy-M 704714 704714 704714 704714 704714 - - - -
XPac-M 703167 703167 703167 - - - - - -
1)
Installation information
Order No. Order No. Order No. Order No. Order No. Order No.
German English French Spanish Italian Japanese
Order No. Order No. Order No. Order No. Order No. Order No. Order No. Order No.
German English French Spanish Italian Japanese Dutch Arabic
AM/PM balances (up to 6.1 kg) 702463 702464 702465 702466 702467 702468 703198 703199
PM…-K 704665 704666 704667 704668 704669 704670 704671 -
PM…-N 704688 704689 704690 704691 704692 704732 704733 -
SM scales 703882 703883 703884 703885 703886 703887 703888 -
DataPac-M - 702446 - - - - - -
GA37 704213 704214 704215 - - - - -
LabPac-M 704693 704694 704695 704696 704697 704698 704699 -
LabWare “Density determination” 702042 702042 702042 - - - - -
Order No. Order No. Order No. Order No. Order No.
German English French Spanish Italian
Order No. Order No. Order No. Order No. Order No.
German English French Spanish Italian
Brochures
Order No. Order No. Order No. Order No. Order No. Order No.
German English French Spanish Italian Japanese
Application – Technical data
– Accessories AM/PM/SM 704700 704701 704702 - 704909 704703
DeltaTrac application brochure 720610 720611 720612 720613 720614 -
Dictionary of Weighing Terms 720113 720114 720115 720711 - -
LP16 application brochure 721065 721066 721067 720608 721068 -
School experiments 720309 720310 720311 - 720814 -
SQC guide 720870 720871 720872 - - -
Weighing the Right Way 720905 720906 720907 720908 720909 720911
Weighing Therms 721158 721159 720115 720711 - -
Fundamentals of Mass Determ. 721074 721118 - - - -
Leerseite
7. Index
Accessories 4.1ff. Control of peripherals 2.9 Filter backweighing 1.4
Additive weighing 1.2 Control of the balance 2.4 Fine dispensing 1.8
Analytical balances AM 5.2 Control unit GM303 4.13 FlowPac-M 2.9, 3.8f.
Animal weighing 1.21 Conversion Kit PM-SE 1.22, 4.8 Foot switch 2.5, 4.13
Animal weighing pan 4.3 CountEasy-M 3.1, 4.13 Formula weighing 1.7ff.
Anti-theft device 4.7 CountPac-M 1.16, 1.19, 3.4f. Formulation program FO3220 1.10, 3.10f.
Attachment of a computer 2.4 D/A Converter GA37 4.12 Freely programmable applications 1.20
Attachment of a printer 2.2f. Data interface 2.4, 6.2 Fuses 4.8, 5.9
Attachment of several balances or scales 2.6 Data transfer 2.1ff. GA devices 2.2, 2.7, 4.11f.
Automatic feeder LV10 4.16 DataPac-M 2.7, 3.6f. GM devices 2.9, 4.13ff.
Auxiliary displays GM48/49/53 4.14 Decimal places 5.11 GoldPac-M 1.15, 3.12f.
Backweighing 1.4f., 1.16 DeltaRange 1.2, 1.17 Hand switch 2.5, 4.13
Bar code identification 2.7 DeltaTrac 1.2, 1.8f., 1.11, 5.8 Hanger 1.22, 5.9
Battery charger 4.18 Demo cases 4.6 In-use covers 4.6, 5.9
Battery set 4.18 Density determination 1.3, 1.22 Infrared dryers 1.6, 4.15
Below-the-balance weighing 1.22 Density determination kit 1.3, 4.7 Interface 2.3ff., 5.8f.
Cables 4.19ff. Differential weighing 1.4ff. Interface converters 2.8, 4.9
CalcPac-M 1.15f., 3.2f. Dimension diagrams 5.12ff. LabPac-M 1.2, 1.4, 1.9, 3.14f.
Calculating with weight data and constants 1.15 Dispensing 1.7ff. LabWare 1.3, 1.5, 1.16, 3.16ff.
Calibration weights 4.2, 5.4ff. Displacement body 1.3, 4.7 Linearity 5.4ff.
CL devices 2.6, 2.8, 4.9ff. Documentation 6.2ff. LP devices 1.6, 4.15
Comparator 1.23, 5.2 Draft shield 4.5, 5.9 LV devices 4.16
Connection possibilities (overview) 2.11 Dynamic weighing 1.21 Mass flow control 3.8
Connectors 4.19ff. Fill process control 1.12ff. Mass flow measurement 3.8
Index 7.1