Keysight 53147 Service Giude
Keysight 53147 Service Giude
Microwave Frequency
Counter/Power Meter/DVM
Assembly Level
Service Guide
Notices U.S. Government Rights Warranty
The Software is “commercial computer THE MATERIAL CONTAINED IN THIS
Copyright Notice software,” as defined by Federal Acqui- DOCUMENT IS PROVIDED “AS IS,”
sition Regulation (“FAR”) 2.101. Pursu- AND IS SUBJECT TO BEING
© Keysight Technologies 2001 - 2017 ant to FAR 12.212 and 27.405-3 and CHANGED, WITHOUT NOTICE, IN
No part of this manual may be repro- Department of Defense FAR Supple- FUTURE EDITIONS. FURTHER, TO THE
duced in any form or by any means ment (“DFARS”) 227.7202, the U.S. MAXIMUM EXTENT PERMITTED BY
(including electronic storage and government acquires commercial com- APPLICABLE LAW, KEYSIGHT DIS-
retrieval or translation into a foreign puter software under the same terms CLAIMS ALL WARRANTIES, EITHER
language) without prior agreement and by which the software is customarily EXPRESS OR IMPLIED, WITH REGARD
written consent from Keysight Technol- provided to the public. Accordingly, TO THIS MANUAL AND ANY INFORMA-
ogies as governed by United States and Keysight provides the Software to U.S. TION CONTAINED HEREIN, INCLUD-
international copyright laws. government customers under its stan- ING BUT NOT LIMITED TO THE
dard commercial license, which is IMPLIED WARRANTIES OF MER-
Manual Part Number embodied in its End User License CHANTABILITY AND FITNESS FOR A
53147-90010 Agreement (EULA), a copy of which can PARTICULAR PURPOSE. KEYSIGHT
be found at http://www.keysight.com/ SHALL NOT BE LIABLE FOR ERRORS
Edition find/sweula. The license set forth in the OR FOR INCIDENTAL OR CONSE-
Edition 3, November 1, 2017 EULA represents the exclusive authority QUENTIAL DAMAGES IN CONNECTION
by which the U.S. government may use, WITH THE FURNISHING, USE, OR
Printed in: modify, distribute, or disclose the Soft- PERFORMANCE OF THIS DOCUMENT
ware. The EULA and the license set OR OF ANY INFORMATION CON-
Printed in Malaysia forth therein, does not require or per- TAINED HEREIN. SHOULD KEYSIGHT
Published by: mit, among other things, that Keysight: AND THE USER HAVE A SEPARATE
(1) Furnish technical information WRITTEN AGREEMENT WITH WAR-
Keysight Technologies related to commercial computer soft- RANTY TERMS COVERING THE MATE-
Bayan Lepas Free Industrial Zone, ware or commercial computer software RIAL IN THIS DOCUMENT THAT
11900 Penang, Malaysia documentation that is not customarily CONFLICT WITH THESE TERMS, THE
provided to the public; or (2) Relinquish WARRANTY TERMS IN THE SEPARATE
Technology Licenses to, or otherwise provide, the govern- AGREEMENT SHALL CONTROL.
The hardware and/or software ment rights in excess of these rights
described in this document are fur- customarily provided to the public to Safety Information
nished under a license and may be use, modify, reproduce, release, per-
used or copied only in accordance with form, display, or disclose commercial CAUTION
the terms of such license. computer software or commercial com-
puter software documentation. No A CAUTION notice denotes a hazard. It
Declaration of Conformity additional government requirements calls attention to an operating proce-
Declarations of Conformity for this beyond those set forth in the EULA dure, practice, or the like that, if not
product and for other Keysight prod- shall apply, except to the extent that correctly performed or adhered to,
ucts may be downloaded from the those terms, rights, or licenses are could result in damage to the product
Web. Go to http://www.keysight.com/ explicitly required from all providers of or loss of important data. Do not pro-
go/conformity. You can then search by commercial computer software pursu-
ceed beyond a CAUTION notice until
product number to find the latest Dec- ant to the FAR and the DFARS and are
set forth specifically in writing else- the indicated conditions are fully
laration of Conformity. understood and met.
where in the EULA. Keysight shall be
under no obligation to update, revise or
otherwise modify the Software. With
respect to any technical data as WARNING
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS A WARNING notice denotes a hazard. It
227.7102, the U.S. government calls attention to an operating proce-
acquires no greater than Limited Rights dure, practice, or the like that, if not
as defined in FAR 27.401 or DFAR correctly performed or adhered to,
227.7103-5 (c), as applicable in any could result in personal injury or death.
technical data. Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
Certification
Keysight Technologies certifies that this product met its published specification at
the time of shipment from the factory. Keysight further certifies that its calibration
measurements are traceable to the United States National Institute of Standards
and Technology (formerly National Bureau of Standards), to the extent allowed by
the Institute’s calibration facility, and to the calibration facilities of other
International Standards Organization members.
Warranty
Keysight warrants Keysight hardware, accessories and supplies against defects in
materials and workmanship for a period of one year from date of shipment. If
Keysight receives notice of such defects during the warranty period, Keysight will,
at its option, either repair or replace products which prove to be defective.
Replacement products may be either new or like-new.
Keysight warrants that Keysight software will not fail to execute its programming
instructions, for the period specified above, due to defects in material and
workmanship when properly installed and used. If Keysight receives notice of such
defects during the warranty period, Keysight will replace software media which
does not execute its programming instructions due to such defects.
Geräuschemission
LpA<47 dB am Arbeits
platz, normaler Betrieb, geprüft nach EN 27779.
Die Angagen beruhen auf Ergebnissen von Typenprüfungen.
Product category:
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
2 Service
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Returning the Instrument to Keysight Technologies for Service . . . . . 57
Providing Repair Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Packing the Instrument in the Original Packaging Materials . . . . . . 58
Packing the Instrument in Commercially Available Packaging
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Calibration Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
10 MHz Timebase Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Power Reference Oscillator Frequency Calibration . . . . . . . . . . . . . 62
Power Reference Oscillator Level Adjustment . . . . . . . . . . . . . . . . . 66
DVM Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Pre-Troubleshooting Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4 Replaceable Parts
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .108
Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .109
How To Order A Part . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .110
Contacting Keysight Technologies . . . . . . . . . . . . . . . . . . . . . . . . .111
Parts Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
Reference Designations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
Cabinet Parts and Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
Accessories and Miscellaneous Items . . . . . . . . . . . . . . . . . . . . . . .112
5 Backdating
6 Specifications
1 Performance Tests -
Verifying Specifications
Introduction 18
Equipment Required 20
Operational Verification 22
Complete Performance Tests 30
Test 1: Channel 1 Frequency Sensitivity 31
Test 2: Channel 2 Frequency Sensitivity 35
Test 3: Power Measurement 39
Test 4: Power Reference Oscillator Frequency 42
Test 5: Power Meter Reference Level 44
Performance Test Record (Page 1 of 3) 51
Performance Test Record (Page 2 of 3) 52
Performance Test Record (Page 3 of 3) 53
17
1 Performance Tests - Verifying Specifications
Introduction
This chapter provides procedures to test the electrical performance of the
Keysight 53147A, 53148A, and 53149A. These procedures are based on the
specifications in Chapter 6, “Specifications”.
Two types of testing are provided:
– Operational Verification
– Complete Performance Tests
This chapter is organized as follows:
– “Introduction” pg. 18
– “Equipment Required” pg. 20
– “Operational Verification” pg. 22
– “Complete Performance Tests” pg. 30
– “Performance Test Record” pg. 51
Operational Verification
Operational Verification is an abbreviated series of tests that you can perform
(instead of performing the Complete Performance Tests) to provide a high degree
of confidence that the instrument is operating properly. Operational Verification is
useful for incoming inspection, routine maintenance, and after instrument repair.
Test Record
The results of the Operational Verification and the Complete Performance Tests
should be recorded on a copy of the Performance Test Record, located at the end
of the “Complete Performance Tests” section in this chapter.
Equipment Required
Table 1-1 lists the test equipment and accessories needed to perform the tests in
this chapter.
Quantity
Instrument Type Required Characteristics Model Recommended[a] Required Use[b]
Synthesized Signal
10 Hz to 20 MHz, +7 dBm to –40 dBm Keysight 3325A/B 2 OV, P, T
Generator
Power Sensor Cable 2.4 mm to 2.4 mm, 5 ft. Keysight 11730A 2[c] P, T
Quantity
Instrument Type Required Characteristics Model Recommended[a] Required Use[b]
BNC (m) to BNC (m) 2 OV, P,
Cables Keysight 10503 series
2.4 mm (m) to 2.4 mm (m) C, T
BNC (f) to Type N (m) Keysight E9635A 1
BNC (m) to SMA (f) Keysight 1250-2015 1
3.5 mm (f) to 2.92 mm (m) — 1
Adapters 2.92 mm (m) to 2.4 mm (m) Keysight 11904A 1 P, C, T
2.92 mm (m) to 2.4 mm (f) Keysight 11904D 1
2.4 mm (m) to 2.4 mm (m) Keysight 11900A 1
2.4 mm (f) to 2.4 mm (f) Keysight 11900B 1
[a] Equivalent equipment can be substituted for all instruments and accessories.
[b] OV = Operational Verification P = Performance Tests C = Calibration T = Troubleshooting
[c] One supplied with Keysight 53147A/48A/49A
Operational Verification
Operational Verification is an abbreviated series of tests that you can perform
(instead of performing the Complete Performance Tests) to provide a high degree
of confidence that the instrument is operating properly. Operational Verification is
useful for incoming inspection, routine maintenance, and after instrument repair.
If you are not familiar with operating the instrument, you should review the
Chapter 1, “Getting Started,” in the Keysight 53147A/148A/149A Operating
Guide. However, the procedures in this chapter are written so that little
experience is necessary. These procedures should be followed in the order in
which they appear.
– Since the instrument’s power supply automatically senses the line voltage,
NOTE there is no AC input-voltage setting.
– It is normal for the fan in the instrument to run when the instrument is in
Standby mode. Power is supplied to the timebase whenever the
power cord is connected to maintain long-term measurement reliability, and
the fan helps to maintain the timebase’s temperature stability.
To ensure that the test results are valid, the instrument and the test equipment
NOTE should be powered on for at least 30 minutes prior to beginning the tests. This
allows the internal temperatures of the equipment and the timebase to stabilize.
10 MHz Test
1 Connect a BNC-to-BNC cable from the Reference 10MHz connector on
the instrument’s rear panel to the CHANNEL 1 input connector on the front
panel.
2 Press and release the Chan Select key.
3 Verify that Ch 1 is displayed in the upper-left corner of the display.
4 Verify that the display reads 10,000,000 Hz.
5 Mark Pass or Fail in the “Performance Test Record” on page 51, “Test 2”.
1 Disconnect all signal cables from the input and Reference 10MHz connectors.
2 Press and release the Shift key, and then press and release the Menu (Reset/
Local) key.
3 Press the up- and/or down-arrow key(s) as many times as necessary until
DO SELF TEST is displayed.
4 Press and release the Enter key.
5 If no errors are detected, SELF TEST OK is briefly displayed when the self test
is completed. If any error messages are displayed, refer to the troubleshooting
section in Chapter 2, “Service”.
6 Mark Pass or Fail in the “Performance Test Record” on page 51, “Test 3”.
– To ensure that the test results are valid, the instrument and the test
NOTE equipment should be powered on for at least 30 minutes prior to beginning
the tests. This allows the internal temperatures of the equipment to stabilize.
– These tests are appropriate for the Keysight 53147A, 53148A, and 53149A.
1 Connect an Keysight 10503 series coaxial cable between the output of the first
Keysight 3325A/B Synthesizer and the Reference 10MHz connector on the
instrument’s rear panel (see Figure 1-1).
2 Connect an Keysight 10100C 50Ω Feedthrough Terminator to the Counter’s
Channel 1 input connector.
3 Connect an Keysight 10503 series coaxial cable between the output of the
second Synthesizer and the feedthrough terminator on the Counter’s
CHANNEL 1 input connector.
To Reference
10 MHz Connector
(Rear Panel)
10 MHZ Output
2 MHz Output
– To ensure that the test results are valid, the instrument and the test
NOTE equipment should be powered on for at least 30 minutes prior to beginning
the tests. This allows the internal temperatures of the equipment to stabilize.
– These tests are appropriate for the Keysight 53147A, 53148A, and 53149A.
1 Connect one end of the Keysight 11730A cable (supplied with instrument)
to the Power Meter INPUT connector on the front panel of the Keysight
53147A, 53148A, or 53149A.
2 Connect the other end of the Keysight 11730A cable to a power sensor head.
3 Press the Display Power key in the Power Meter area of the front panel.
The display should look like this (the reading will vary):
Ch 2
Freq
dB
dBm
Watts
Pwr mW
uW
%
4 If the power sensor head you are using is an Keysight 8481A, 8481D, 8482A,
8485A, or 8487A, select the power sensor head model in the HEAD menu. If
you are not using one of the power sensor heads listed above, press the Cal
Factor key, and enter the calibration factor for the head.
5 Press the Zero key.
The display shows ZEROING for several seconds.
6 Connect the power sensor head to the Power Meter OUTPUT connector.
7 Press the Cal key.
The display briefly shows CALIBRATING.
8 Turn on the power reference output in the instrument’s menu (PWR REF > ON).
9 Verify that the Counter displays a reading of 0 dBm ± .02 dB.
10 Mark Pass or Fail for “Test 5” in the “Performance Test Record” on page 51.
11 Disconnect the test setup.
DVM Test
This test verifies the accuracy of the DVM.
Equipment Required
Keysight 34401A Digital Multimeter
Keysight E3640A DC Power Supply
– To ensure that the test results are valid, the instrument and the test
NOTE equipment should be powered on for at least 30 minutes prior to beginning
the tests. This allows the internal temperatures of the equipment to stabilize.
– These tests are appropriate for the Keysight 53147A, 53148A, and 53149A.
Keysight 34401A
Digital Multimeter Keysight E3640A
Power Supply
5 Observe and record the voltage reading on the Keysight 34401A Multimeter.
6 Disconnect the test leads from the Keysight E3640A DC Power Supply.
Be careful not to disturb the voltage setting.
7 Connect the test leads to the Keysight 53147A/48A/49A (be sure to observe
the correct polarity).
8 Press the Display DVM key on the Keysight 53147A/48A/49A front panel.
9 Connect the test prods to the output terminals on the Keysight E3640A
DC Power Supply (be sure to observe the correct polarity).
10 Observe and record the voltage reading on the Keysight 53147A/48A/49A.
11 Compare the voltage readings you recorded in steps 5 and 10. If the reading
taken with the Keysight 53147A/48A/49A is within ±0.25% ±10 mV of the
reading taken with the Keysight 34401A, mark Pass for “Test 6” in the
“Performance Test Record” on page 51. If the reading taken with the Keysight
53147A/48A/49A is not within ±0.25% ±10 mV of the reading taken with the
Keysight 34401A, mark Fail for “Test 6” in the “Performance Test Record” on
page 51.
This completes the Operational Verification.
To ensure that the test results are valid, the instrument and the test equipment
NOTE should be powered on for at least 30 minutes prior to beginning the tests. This
allows the internal temperatures of the equipment and the timebase to stabilize.
1 Connect an Keysight 10503 Series Coaxial Cable between the reference output
connector on the Keysight 3325A/B Synthesizer and the instrument’s
Reference 10MHz connector.
2 Connect an Keysight 10100C 50Ω Feedthrough Terminator to the Counter’s
CHANNEL 1 connector.
To Reference Output
Output
(Rear
Keysight Panel)
53147A/48A/49A
Channel 1
4 Verify that the Counter is still set to use an external reference signal (refer to
steps 5 through 10 on page 25).
5 Set the output of the Synthesizer to the first frequency specified for “Test 1a”
in the “Performance Test Record” on page 51 (10 Hz).
6 Set the power output of the Synthesizer to the first power value specified for
“Test 1a” in the “Performance Test Record” on page 51 (–14.9 dBm/40 mV).
7 Record the frequency value read on the Counter (±1 count) for “Test 1a” in the
“Performance Test Record” on page 51.
8 Repeat steps 5, 6, and 7 for each of the remaining frequencies specified for
“Test 1a”. Don’t forget to change the power level when performing the 1 KHz
test (–19.2 dBm/25 mV).
9 If all of the frequency values you recorded for “Test 1a” are correct, mark Pass
in the Performance Test Record for “Test 1a”. If any of the frequency readings
you recorded are incorrect, mark Fail in the Performance Test Record.
2 Connect the Keysight 1250-2015 BNC (m) to SMA (f) Adapter to the
Keysight 10100C Feedthrough Terminator on the Counter’s CHANNEL 1
connector.
3 Connect the Keysight 11904A 2.92 mm (m) to 2.4 mm (m) Adapter to the
Keysight 1250-2015 Adapter.
4 Connect the Keysight 8490D 20 dB Attenuator to the Keysight 11904A
Adapter.
5 Connect one of the output connectors on the Keysight 11667C Power Splitter
to the Keysight 8490D Attenuator
6 Connect the Keysight 11900B 2.4 mm (f) to 2.4 mm (f) Adapter to the RF
output of the Synthesizer.
7 Connect a 2.4 mm (m) to 2.4 mm (m) cable between the Adapter on the
Synthesizer’s RF output connector and the input connector on the
Keysight 11667C Power Splitter.
8 Connect the Keysight 8487A/D Power Sensor to the remaining output
connector on the Keysight 11667C Power Splitter.
9 Connect the Keysight 11730A Power Sensor Cable between the Power Sensor
and the sensor connector on the Keysight 437B Power Meter.
10 Verify that the Counter is still set to use an external reference signal (refer to
steps 5 through 10 on page 25).
11 Set the output of the Keysight 83650B Synthesizer to the first frequency
specified for “Test 1b” in the “Performance Test Record” on page 51 (50 MHz).
12 Set the power output of the Synthesizer (read on the Power Meter) to a power
value 20 dBm above the value specified for “Test 1b” in the “Performance Test
Record” on page 51 (+.8 dBm). (The added 20 dBm compensates for the 20 dB
Attenuator.)
13 Record the frequency value read on the Counter (±1 count) for “Test 1b” in the
“Performance Test Record” on page 51.
14 Set the output of the Synthesizer to 125 MHz, and repeat steps 11 through 13.
15 If both of the frequency values you recorded for “Test 1b” are correct, mark
Pass in the Performance Test Record for “Test 1b”. If either of the frequency
readings you recorded are incorrect, mark Fail in the Performance Test Record.
16 Disconnect the Keysight 10100C Feedthrough Terminator, the
Keysight 1250-2015 Adapter, and the Keysight 11904A Adapter from the
Counter’s CHANNEL 1 connector.
– Do not disconnect the rest of the test setup, as these connections are also
NOTE used in the Channel 2 Frequency Sensitivity test.
– Do not turn off the instrument. Turning the instrument off and back on resets
the reference oscillator (REF OSC) selection to internal (INT).
– To ensure that the test results are valid, the instrument and the test
NOTE equipment should be powered on for at least 30 minutes prior to beginning
the tests. This allows the internal temperatures of the equipment to stabilize.
– This test is appropriate for the Keysight 53147A, 53148A, and 53149A.
1 Connect the Keysight 11904D 2.92 mm (m) to 2.4 mm (f) Adapter to the
Counter’s CHANNEL 2 input connector (refer to Figure 1-5 for steps 1 through
4).
2 Connect the Keysight 8490D Attenuator to the Keysight 11904D Adapter,
as shown in Figure 1-5 (the other end of the Attenuator should remain
connected to the Keysight 11667C Power Splitter).
3 Verify that the cable (with Adapter) is still connected between the
RF output of the Keysight 83650B Synthesizer and the Power Splitter
input connector.
4 Verify that the Power Sensor is still connected to the Power Splitter and that
the Power Sensor Cable is still connected between the Power Sensor and the
sensor connector on the Power Meter.
Keysight 83650B Synthesizer
To Reference
Output
(Rear Panel) Keysight 437B Power Meter
HP 8490D 20 dB Attenuator
Channel 2 (50 MHz to 40 GHz tests only)
HP 11904D Adapter
5 Verify that the instrument is still set to use an external reference signal (refer to
steps 5 through 10 on page 25).
6 Set the output of the Keysight 83650B Synthesizer to the first frequency
specified for “Test 2a” in the “Performance Test Record” on page 52 (50 MHz).
7 Set the power output of the Synthesizer (read on the Power Meter) to a power
value 20 dBm above the value specified for “Test 2a” in the “Performance Test
Record” on page 52 (0 dBm). (The added 20 dBm compensates for the 20 dB
Attenuator.)
8 Record the frequency value read on the Counter (±1 count) for “Test 2a” in the
“Performance Test Record” on page 52.
9 Repeat steps 6, through 8 for each of the remaining frequencies specified for
“Test 2a”. Don’t forget to change the power level when performing the
300 MHz and 19 GHz tests (53147A, 53148A, and 53149A) and the 16 GHz
and 19 GHz tests (53149A only).
It may be necessary to repeat Step 5.
NOTE
10 If all of the frequency values you recorded for “Test 2a” are correct, mark Pass
in the Performance Test Record. If any of the frequency readings you recorded
are incorrect, mark Fail in the Performance Test Record.
1 Use the same test setup as described for “Test 2a” on page 35.
2 Set the output of the Keysight 83650B Synthesizer to the first frequency
specified for “Test 2b” in the “Performance Test Record” on page 52 (22 GHz).
3 Set the power output of the Synthesizer (read on the Power Meter) to a power
value 20 dBm above the value specified for “Test 2b” in the “Performance Test
Record” on page 52 (–5 dBm for the Keysight 53148A;
–7 dBm for the Keysight 53149A). (The added 20 dBm compensates for the
20 dB Attenuator.)
4 Record the frequency value read on the Counter (±1 count) for “Test 2b” in the
“Performance Test Record” on page 52.
5 Repeat steps 2, through 4 for each of the remaining frequencies specified for
“Test 2b”.
6 If all of the frequency values you recorded for “Test 2b” are correct, mark Pass
in the Performance Test Record for “Test 2b”. If any of the frequency readings
you recorded are incorrect, mark Fail in the Performance Test Record.
1 Use the same test setup as described for “Test 2a” on page 35.
2 Set the output of the Keysight 83650B Synthesizer to the first frequency
specified for “Test 2c” in the “Performance Test Record” on page 52 (30 GHz).
3 Set the power output of the Synthesizer (read on the Power Meter) to a power
value 20 dBm above the value specified for “Test 2c” in the “Performance Test
Record” on page 52 (–3 dBm). (The added 20 dBm compensates for the 20 dB
Attenuator.)
4 Record the frequency value read on the Counter (±1 count) for “Test 2c” in the
“Performance Test Record” on page 52.
5 Repeat steps 2, through 4 for each of the remaining frequencies specified for
“Test 2c”. Don’t forget to change the power level for the 42, 44, and 46 GHz
tests.
6 If all of the frequency values you recorded for “Test 2c” are correct, mark Pass
in the Performance Test Record for “Test 2c”. If any of the frequency readings
you recorded are incorrect, mark Fail in the Performance Test Record.
– Do not disconnect the Power Sensor or the Power Sensor Cable from the
NOTE power meter or the Power Splitter. These connections are reused in the
following tests.
– Do not turn off the instrument. Turning the instrument off and back on resets
the reference oscillator (REF OSC) selection to internal (INT).
1 Connect the Keysight 11730A Power Sensor Cable between Keysight 11683A
Range Calibrator’s output connector and the Keysight 53147A, 53148A, or
53149A’s Power Meter INPUT connector.
Keysight
53147A/48A/49A Keysight 11683A
Range Calibrator
Output
Power Input
FUNCTION: STANDBY
POLARITY: NORMAL
RANGE: 1 mW
LINE: ON
5 Press the Zero key on the Keysight 53147A, 53148A, or 53149A. When the
reading appears, verify that it is 0.00 ±0.06 µW.
6 Change the Range Calibrator’s FUNCTION switch to CALIBRATE.
7 Press the Cal key on the Keysight 53147A, 53148A, or 53149A. Wait for the
Power Meter to complete calibration.
8 Verify that the Keysight 53147A, 53148A, or 53149A display reads
1.000 ±0.006 mW.
9 Set the Range Calibrator RANGE switch to each of the positions shown in
Table 1-3. For each setting, verify that the Keysight 53147A, 53148A, or
53149A autoranges properly and that the reading displayed is within the limits
shown in the table.
– The lower the setting of the RANGE switch, the longer it takes for
NOTE the reading to stabilize.
– You may have to repeat Step 5 (re-zero the instrument) for each
measurement.
2 Connect an Keysight 10503 series cable between the Keysight E9635A adapter
on the Keysight 53147A/148A/149A’s Power Meter OUTPUT connector and
the CHANNEL 1 input connector on the Keysight 53150A/51A/52A Frequency
Counter.
Keysight E9635A Adapter on
Power Reference Oscillator
Output Connector
Keysight Keysight
53147A/48A/49A 53150A/51A/52A
Microwave Frequency Counter
53150A 20 GHz Counter CHANNEL 1
GHz MHz kHz Hz 10 Hz to 125 MHz
POWER
Ch 12
Rel Freq 1M :
Offset dB Ext Rel Hold
Avg On dBm Rate Rmt SRQ
Watts
Rel Pwr mW
uW
Error
DAMAGE
Standby Offset % Shift
+30 dBm
MODIFY FREQ POWER CHANNEL 2
Freq Pwr Gate Channel 2 50 MHz to 20 GHz
Menu Offset Offset HPIB dBm/ W DAMAGE
Reset/ Rate Avg Resol Chan Display
+27 dBm
Local Select Power
On/Off
50 :
Rel Freq Rel Pwr
Shift Clear +/- Enter Offset Offset
On/Off On/Off
Channel 1
Input
5 Record the power reading shown on the Keysight 53150A/51A/52A for “Test
4” in the “Performance Test Record” on page 51.
6 Mark Pass or Fail for “Test 4” in the Performance Test Record.
– To ensure that the test results are valid, the Keysight 53147A/48A/49A and
NOTE the test equipment should be powered on for at least 30 minutes prior to
beginning the tests. This allows the internal temperatures of the equipment
to stabilize.
– The calibration of the Keysight E4418B Power Meter must be no more than
one level removed from NIST (National Institute of Standards and
Technology). If the E4418B Power Meter’s calibration is more than one level
removed from NIST, the test results will be invalid.
– This test is appropriate for the Keysight 53147A, 53148A, and 53149A.
1 Connect the Keysight 11730A Power Sensor Cable to the Keysight 8481A
Power Sensor (see Figure 1-8).
Figure 1-8 E4418B Power Meter Zeroing and Calibration Test Setup
2 Connect the free end of the Keysight 11730A Power Sensor Cable to the input
connector on the Keysight E4418B Power Meter.
3 Connect the Keysight 8481A Power Sensor to the power reference output
connector of the Keysight E4418B Power Meter.
4 Zero the E4418B Power Meter.
5 Calibrate the E4418B Power Meter.
6 Disconnect the Keysight 8481A Power Sensor from the Keysight E4418B
Power Meter, and immediately connect it to the Power OUTPUT connector on
the Keysight 53147A/48A/49A (see Figure 1-9).
Output connector (with Keysight 8481A
Power Sensor Attached)
Keysight
53147A/48A/49A Keysight E4418B
Power Meter
Input
7 Record the power reading shown on the Keysight E4418B Power Meter for
“Test 5” in the “Performance Test Record” on page 51.
8 Mark Pass or Fail for “Test 5” in the Performance Test Record.
– To ensure that the test results are valid, the instrument and the test
NOTE equipment should be powered on for at least 30 minutes prior to beginning
the tests. This allows the internal temperatures of the equipment to stabilize.
– This test is appropriate for the Keysight 53147A, 53148A, and 53149A.
1 Turn the Keysight 432A Power Meter off. Do not turn it on until instructed to do
so.
2 Set the Keysight 34401A Digital Multimeter to measure resistance, and
connect the DVM test prods between the center pin of the VRF connector on
the Keysight 432A Power Meter rear panel and pin 1 on the thermistor-mount
end of the Keysight 8120-1082 Power Sensor Cable (see Figure 1-10 to locate
pin 1).
2 1
6 5
3 4
3 Round off the resistance reading to two decimal places and keep a record of
this value as R. It is the internal bridge resistance (R) of the Keysight 432A
Power Meter (this should be approximately 200.3 ohms).
4 Connect the Keysight 478A-H75 or 478A-H76 Thermistor Mount to the
Keysight 53147A/148A/149A Power Meter OUTPUT connector, as shown in
Figure 1-11.
Do not use a cable between the Thermistor Mount and the 53147A/148A/149A
NOTE Power Meter OUTPUT connector. Connect the Thermistor Mount directly to the
OUTPUT connector.
Keysight 478A-H75 or
478A-H76 Thermistor Mount
(connected to Power Keysight 432A
Meter OUTPUT connector) Power Meter
Keysight
53147A/48A/49A Keysight 34401A
Digital Multimeter
VRF - +
-
VCOMP
+
Keysight 8120-1082
Power Sensor Cable
5 Connect the Keysight 8120-1082 Power Sensor Cable between the Thermistor
Mount and the input connector on the Keysight 432A Power Meter.
6 Turn on the 432A Power Meter. Wait 30 minutes for the Power Meter and the
Thermistor Mount to stabilize before proceeding to the next step.
7 Set the 432A Power Meter’s RANGE switch to Coarse Zero.
8 Adjust the COARSE ZERO control on the 432A Power Meter’s front panel to
obtain a zero reading on the meter.
9 Set the 432A Power Meter’s RANGE switch to .01 mW / –20 dBm.
10 Press the FINE ZERO switch on the 432A Power Meter’s front panel down to
obtain a zero reading on the meter.
11 Set the 432A Power Meter’s RANGE switch to 1 mW.
12 Set the Digital Multimeter to measure microvolts.
Ensure that the Keysight 34401A Digital Multimeter’s input leads are
CAUTION isolated from chassis ground before performing the next step.
13 Connect the positive test prod from the Keysight 34401A Multimeter to
the VCOMP connector on the Keysight 432A Power Meter rear panel.
The VCOMP and VRF connectors are BNC female. You can use two
NOTE
BNC-to-banana jack adapters and two BNC cables to make it easier to connect
the Multimeter to the Power Meter.
14 Connect the negative test prod from the Keysight 34401A Digital Multimeter to
the VRF connector on the 432A Power Meter’s rear panel.
15 If the reading on the Multimeter is less than 400 microvolts, proceed to step
16. If the reading is 400 microvolts or more, press the FINE ZERO toggle switch
on the 432A Power Meter’s front panel down.
16 Observe the reading on the Multimeter, round off the reading to the nearest
microvolt, and keep a record of this value as V0.
17 Turn on the Keysight 53147A/148A/149A’s internal power reference oscillator:
Press Shift + Menu, use the up and/or down arrow keys to cycle through the
menu to PWR REF, press the right-arrow key, then use the up or down arrow
key to change the setting to ON, and press the Enter key. The Pwr Ref
annunciator should now be activated.
18 Observe the reading on the Multimeter, and keep a record of this value
as V1.
19 Disconnect the Multimeter’s negative test lead from the 432A Power Meter’s
VRF connector, and connect it to the 432A Power Meter’s chassis ground.
20 Observe the reading on the Multimeter, and keep a record of this value
as VCOMP.
21 Calculate the output level of the 53147A/148A/149A’s power reference
oscillator (Prf ) using the following formula:
Where:
R = 200 W
V0 = 170 μV or 170 x 10–6 V
Vcomp = 5.1583 V
Note that V0 is eventually dropped out of the equation in the following steps,
because its value is very small (effectively zero).
= 1.00147 mW
2 Service
Introduction 56
Returning the Instrument to Keysight Technologies for Service 57
Calibration Procedures 59
Pre-Troubleshooting Information 76
Troubleshooting the Instrument 81
55
2 Service
Introduction
This chapter provides service information for your Keysight 53147A/148A/149A. It
is divided into four major sections:
– “Returning the Instrument to Keysight Technologies for Service”
(page 57). This section provides you with step-by-step instructions on how to
return the instrument for service.
– “Calibration Proced ures” (page 59). This section provides step-by-step
procedures for calibrating the Keysight 53147A/148A/149A.
– “Pre-Troubleshooting Information” (page 76). This section provides you with
pertinent information such as safety considerations, recommended test
equipment, repair and after-service considerations, and assembly
identification and location.
– “Troubleshooting the Instrument” (page 81). This section provides you with
troubleshooting procedures that isolate the faulty assembly or module.
Replacement and recalibration of most modules can only be performed at an
authorized Keysight Technologies Service Center.
Calibration Procedures
There are four calibration procedures that can be performed outside of an
Keysight authorized service facility. The following procedures describe how to
calibrate the frequency of the instrument’s timebase, the frequency and output
level of the power reference oscillator, and the DVM.
Equipment Required
Table 2-1 lists the test equipment and accessories needed to perform the
calibration procedures in this chapter.
[a] Equivalent equipment can be substituted for all instruments and accessories.
[b] One supplied with Keysight 53147A/148A/149A
The instrument and the test equipment must be powered on for at least
NOTE 30 minutes prior to beginning this calibration procedure. This allows the internal
temperatures of the equipment to stabilize.
1 Connect an Keysight 10503 series coaxial cable from the output of your
10 MHz frequency standard to the Counter’s CHANNEL 1 input on the
instrument’s front panel.
2 Press the Resol key, and verify that the resolution is set to 1 HZ.
If necessary change the resolution to 1 HZ.
3 If there is a label covering the calibration adjustment access hole,
remove it (see Figure 2-1).
4 Insert a non-metallic, slot-type adjustment tool through the hole near the
bottom of the instrument’s left side and into the slot in the timebase
calibration screw.
5 Turn the calibration screw clockwise to increase the Counter’s reading or
counter-clockwise to decrease the reading until the frequency value displayed
on the Counter’s front panel is exactly 10,000,000 Hz.
– The instrument and the test equipment must be powered on for at least
NOTE 30 minutes prior to adjusting the power reference oscillator output. This
allows the internal temperatures of the equipment to stabilize.
– This procedure specifies the Keysight 53150A for measuring the frequency
output of the 53147A/148A/149A’s power reference oscillator. If the 53147A/
148A/149A’s Frequency Counter is known to be in calibration (see “Test 1:
Channel 1 Frequency Sensitivity” on page 31), you can use it in place of the
53150A in this procedure.
– Adjusting the power reference oscillator frequency may also affect the output
level of the oscillator. Therefore, after you adjust the frequency to 50.0 MHz ±
0.05 MHz, always check the output level and adjust it, if necessary, as
described in “Power Reference Oscillator Level Adjustment” on page 66.
1 Turn off the Keysight 53147A/148A/149A, and disconnect the power cord.
2 Remove the shroud from the Keysight 53147A/148A/149A (see “Removing the
Shroud” on page 96).
3 Position the Keysight 53147A/148A/149A on its right side so that you have
access to the front panel and the left side of the main circuit board (see
Figure 2-2).
Keysight
53147A/48A/49A
Keysight
53150A/51A/52A
Microwave Frequency Counter
53150A 20 GHz Counter CHANNEL 1
GHz MHz kHz Hz 10 Hz to 125 MHz
POWER
Ch 12
Rel Freq 1M :
Offset dB Ext Rel Hold
Avg On dBm Rate Rmt SRQ
Watts
Rel Pwr mW
uW
Error
DAMAGE
Standby Offset % Shift
+30 dBm
MODIFY FREQ POWER CHANNEL 2
Freq Pwr Gate Channel 2 50 MHz to 20 GHz
Menu Offset Offset HPIB dBm/ W DAMAGE
Reset/ Rate Avg Resol Chan Display
+27 dBm
Local Select Power
On/Off
50 :
Rel Freq Rel Pwr
Shift Clear +/- Enter Offset Offset
On/Off On/Off
Channel 1
Input
5 Turn on the Keysight 53147A/148A/149A and then wait for at least 30 minutes
to allow the internal temperature of the instrument to stabilize.
6 Connect an Keysight E9635A adapter to the Power Meter OUTPUT connector as
shown in Figure 2-2.
7 Connect an Keysight 10503 series cable from the Keysight E9635A adapter on
the Keysight 53147A/148A/149A’s Power Meter OUTPUT connector to the
CHANNEL 1 input connector on the Keysight 53150A/51A/52A Frequency
Counter.
8 Set the Keysight 53150A/51A/52A Frequency Counter to measure frequency
on Channel 1.
9 Turn on the Keysight 53147A/148A/149A’s internal power reference oscillator:
Press Shift + Menu, use the up and/or down arrow keys to cycle through the
menu to PWR REF, press the right-arrow key, then use the up or down arrow
key to change the setting to ON, and press the Enter key. The Pwr Ref
annunciator should now be activated.
10 Insert a small, non-conductive, slot-type adjustment tool into the hole in the
side of the power reference oscillator shield (see Figure 2-3). This opening
allows access to L9, the adjustable inductor that is used to adjust the
frequency of the power reference oscillator.
L9
(Frequency Adjustment)
R98
(Power Adjustment)
11 Set the frequency to 50.0 MHz ± 0.05 MHz (49.95 MHz to 50.05 MHz). Turn the
adjustment tool counter-clockwise to increase the frequency or clockwise to
decrease the frequency.
Method 1
Equipment Required
Keysight E4418B Power Meter
Keysight 8481A Power Sensor
Keysight 11730A Power Sensor Cable (supplied with 53147A/148A/149A)
– To ensure that the results are valid, the Keysight 53147A/148A/149A and the
NOTE test equipment should be powered on for at least 30 minutes prior to
beginning the calibration procedure. This allows the internal temperatures of
the equipment to stabilize.
– The calibration of the Keysight E4418B Power Meter must be no more than
one level removed from NIST (National Institute of Standards and
Technology). If the E4418B Power Meter’s calibration is more than one level
removed from NIST, the test results will be invalid.
– This test is appropriate for the Keysight 53147A, 53148A, and 53149A.
1 Connect the Keysight 11730A Power Sensor Cable to the Keysight 8481A
Power Sensor (see Figure 2-4).
Output connector (with Keysight 8481A
Power Sensor Attached)
Keysight E4418B
Power Meter
Input
Figure 2-4 E4418B Power Meter Zeroing and Calibration Test Setup
2 Connect the free end of the Keysight 11730A Power Sensor Cable to the input
connector on the Keysight E4418B Power Meter.
3 Connect the Keysight 8481A Power Sensor to the power reference output
connector of the Keysight E4418B Power Meter.
4 Zero the E4418B Power Meter.
5 Calibrate the E4418B Power Meter.
6 Disconnect the Keysight 8481A Power Sensor from the Keysight E4418B
Power Meter, and immediately connect it to the Power OUTPUT connector on
the Keysight 53147A/148A/149A (see Figure 2-5).
Output connector (with Keysight 8481A
Power Sensor Attached)
Keysight
53147A/48A/49A Keysight E4418B
Power Meter
Input
Method 2
Equipment Required
Keysight 432A Power Meter
Keysight 478A-H75 or 478A-H76 Thermistor Mount
Keysight 34401A Digital Multimeter
Keysight 8120-1082 Power Sensor Cable
To ensure that the results are valid, the Keysight 53147A/148A/149A and the
NOTE test equipment should be powered on for at least 30 minutes prior to beginning
the calibration procedure. This allows the internal temperatures of the
equipment to stabilize.
1 Make sure the Keysight 432A Power Meter is off. Do not turn it on until
instructed to do so.
2 Set the Keysight 34401A Digital Multimeter to measure resistance, and
connect the DVM test prods between the center pin of the VRF connector on
the Keysight 432A Power Meter rear panel and pin 1 on the thermistor-mount
end of the Keysight 8120-1082 Power Sensor Cable (see Figure 2-6 to locate
pin 1).
2 1
6 5
3 4
3 Round off the resistance reading to two decimal places and keep a record of
this value as R. It is the internal bridge resistance (R) of the Keysight 432A
Power Meter (this should be approximately 200.3 ohms).
4 Connect the Keysight 478A-H75 or 478A-H76 Thermistor Mount to the
Keysight 53147A/148A/149A Power Meter OUTPUT connector, as shown in
Figure 2-7.
Do not use a cable between the Thermistor Mount and the 53147A/148A/149A
NOTE Power Meter OUTPUT connector. Connect the Thermistor Mount directly to the
OUTPUT connector.
Keysight 478A-H75 or
478A-H76 Thermistor Mount
(connected to Power Keysight 432A
Meter OUTPUT connector) Power Meter
Keysight
53147A/48A/49A Keysight 34401A
Digital Multimeter
VRF - +
-
VCOMP
+
Keysight 8120-1082
Power Sensor Cable
5 Connect the Keysight 8120-1082 Power Sensor Cable between the Thermistor
Mount and the input connector on the Keysight 432A Power Meter.
6 Turn on the 432A Power Meter. Wait 30 minutes for the Power Meter and the
Thermistor Mount to stabilize before proceeding to the next step.
7 Set the 432A Power Meter’s RANGE switch to Coarse Zero.
8 Adjust the COARSE ZERO control on the 432A Power Meter’s front panel to
obtain a zero reading on the meter.
9 Set the 432A Power Meter’s RANGE switch to .01 mW / –20 dBm.
10 Press the FINE ZERO control on the 432A Power Meter’s front panel down to
obtain a zero reading on the meter.
11 Set the 432A Power Meter’s RANGE switch to 1 mW.
12 Set the Digital Multimeter to measure microvolts.
Ensure that the Keysight 34401A Digital Multimeter’s input leads are
CAUTION isolated from chassis ground before performing the next step.
13 Connect the positive test prod from the Keysight 34401A Multimeter to the
VCOMP connector on the Keysight 432A Power Meter rear panel.
The VCOMP and VRF connectors are BNC female. You can use two
NOTE
BNC-to-banana jack adapters and two BNC cables to make it easier to connect
the Multimeter to the Power Meter.
14 Connect the negative test prod from the Keysight 34401A Multimeter to the
VRF connector on the 432A Power Meter’s rear panel.
15 If the reading on the Multimeter is less than 400 microvolts, proceed to step
16. If the reading is 400 microvolts or more, press the FINE ZERO toggle switch
on the 432A Power Meter’s front panel down.
16 Observe the reading on the Multimeter, round off the reading to the nearest
microvolt, and keep a record of this value as V0.
17 Turn on the Keysight 53147A/148A/149A’s internal power reference oscillator:
Press Shift + Menu, use the up and/or down arrow keys to cycle through the
menu to PWR REF, press the right-arrow key, then use the up or down arrow
key to change the setting to ON, and press the Enter key. The Pwr Ref
annunciator should now be activated.
18 Observe the reading on the Multimeter, and keep a record of this value as V1.
19 Disconnect the Multimeter’s negative test lead from the 432A Power Meter’s
VRF connector, and connect it to the 432A Power Meter’s chassis ground.
20 Observe the reading on the Multimeter, and keep a record of this value
as VCOMP.
Where:
R = 200 W
V0 = 170 μV or 170 x 10–6 V
Vcomp = 5.1583 V
Note that V0 is eventually dropped out of the equation in the following steps,
because its value is very small (effectively zero).
= 1.00147 mW
DVM Calibration
The DVM calibration procedure is semi-automatic in that it is accomplished via
software embedded in the Keysight 53147A/148A/149A. A computer and an
RS-232 serial cable are required to access the calibration software.
Equipment Required
A computer with an RS-232 serial interface and terminal software
Keysight 53150-60215 RS-232 Serial Cable
Keysight 6613C or 6614C Precision DC Power Supply
To ensure that the results are valid, the Keysight 53147A/148A/149A and the
NOTE Keysight 6613C/6614C Precision DC Power Supply should be powered on for at
least 30 minutes prior to beginning the calibration procedure. This allows the
internal temperatures of the equipment to stabilize.
15 When the calibration program redisplays its menu (as shown in Figure 2-8),
type “e” to exit the program.
16 Close the terminal program and disconnect the serial cable from the computer
and the instrument.
Pre-Troubleshooting Information
This section contains the following pertinent troubleshooting information:
– Safety Considerations
– Recommended Test Equipment
– Repair Considerations
– After Service Considerations
– Assembly Identification and Location
Safety Considerations
Although this instrument has been designed in accordance with international
safety standards, this guide contains information, cautions, and warnings which
must be followed to ensure safe operation and to retain the instrument in a safe
condition. Service instructions, and adjustment procedures requiring removal of
the instrument cover, are for use by service-trained personnel only. To avoid
dangerous electric shock, do not perform any servicing or make any adjustments
with the cover removed, unless qualified to do so.
Any maintenance and repair of the opened instrument under voltage should be
avoided as much as possible and, if necessary, should be carried out only by a
skilled person who is aware of the hazards involved (for example, fire and electric
shock).
Repair Considerations
Electrostatic Discharge
Electronic components and assemblies in the Keysight 53147A/148A/149A can
be permanently degraded or damaged by electrostatic discharge. Use the
following precautions when servicing the instrument:
1 ENSURE that static-sensitive devices or assemblies are serviced at static-safe
work stations providing proper grounding for service personnel.
2 ENSURE that static-sensitive devices or assemblies are stored in static
shielding bags or containers.
3 DO NOT wear clothing subject to static-charge buildup, such as wool or
synthetic materials.
4 DO NOT handle components or assemblies in carpeted areas.
5 DO NOT remove an assembly or component from its static shielding protection
until you are ready to install it.
6 AVOID touching component leads. (Handle by packaging only.)
A7 Battery[a] 53150-80010
A6
A7
A5 A8
A4 A9
A11
A3 A12
A13
A10
A2
A1
The DC Power Supply Assembly is fused for safety reasons. It must be assumed
NOTE that, if the fuses are blown, the power supply has probably suffered damage and
has therefore become unreliable. If the power supply fails, it must be replaced,
even if the fuse in the power supply appears to be all that has failed.
Pin 8
Table 2-3 lists A4 Power Supply's input and output voltages, the corresponding
wire color for each input and output, and the tolerances specified for these
voltages.
Self-Test
The instrument performs a Self-Test procedure when it is powered on. The
Self-Test can also be invoked from the instrument’s Menu.
Table 2-5 and Table 2-6 list and describe the messages that are generated by the
instrument during Self-Test or during operation to indicate that a problem has
been detected. These messages are displayed on the instrument’s front-panel
display and are also sent via the RS-232 serial output (in many cases, the exact
message text that is displayed on the front panel is a condensed form of the
message that is sent via RS-232).
3 Replacing Assemblies -
Disassembly and
Reassembly
Introduction 90
Tools Required 91
Do This First 92
Removing the Bumpers 93
Removing the Handle 94
Removing the Bumper Retainers 95
Removing the Shroud 96
Removing the Front Panel Assembly 98
Removing the Rear Panel Assembly 101
Removing the Cooling Fan 103
Removing the Power Supply 104
89
3 Replacing Assemblies - Disassembly and Reassembly
Introduction
This chapter provides instructions for removing major assemblies in the Keysight
53147A/148A/149A. To install an assembly, apply the removal instructions in
reverse.
The following disassembly and repair procedures are provided:
– “Removing the Bumpers” pg. 93
– “Removing the Handle” pg. 94
– “Removing the Bumper Retainers” pg. 95
– “Removing the Shroud” pg. 96
– “Removing the Front Panel Assembly” pg. 98
– “Removing the Rear Panel Assembly” pg. 101
– “Removing the Cooling Fan” pg. 103
– “Removing the Power Supply” pg. 104
Part numbers for all replaceable parts and assemblies are provided in Chapter 4,
“Replaceable Parts”.
Tools Required
The following tools are required for the removal/replacement procedures in this
chapter:
– #1 Phillips screwdriver
– #2 Phillips screwdriver
– TORX® 15 (T15) hand screwdriver
– 1/4-inch open-end or box wrench
– 1/4-inch nut driver
– 5/16-inch nut driver
– 7 mm nut driver
– Needle-nose pliers
Do This First
Perform the following steps before beginning any of the removal and replacement
procedures:
1 If the instrument is on, press and release the POWER button on the
front panel.
2 Disconnect the AC power cord from the AC input socket on the rear panel.
3 If the instrument has the Battery Option (002) installed, remove the battery
sled, and disconnect any DC input cable from the EXT DC connector on the
rear panel.
While most of the hardware used in the instrument is SAE, the six
CAUTION TORX screws that attach the handle mounts to the instrument are metric. To
avoid damage to the threaded holes in the chassis that handle mount screws
fit in, be sure to use the correct screws.
COUNTER
Gate
2 Use the same procedure (as in Step 1) to remove the front bumper.
The bumpers can be used to support and protect the instrument after the shroud
NOTE is removed. Place both bumpers on their inner edges on your work surface, and
then place the instrument on top of them. The rear bumper can also serve as a
holder for the instrument (once the bumper retainers are removed) when you
need to position it upright on either end.
4 Use a T15 TORX screwdriver to remove two of the 8-32 x 1/4-inch handle
mounting screws (remove the two screws that are closest to the back of
the instrument).
5 Repeat Step 4 for the other handle mount.
6 While holding the handle to support it, remove the remaining two screws (one
on each handle mount).
7 Remove the handle.
When the all of the bumper-retainer screws and the handle are removed, the
CAUTION shroud is not attached to the chassis. Exercise caution when moving the
instrument in this condition to prevent the chassis from sliding out of the
shroud. If you do not intend to remove the shroud, insert the bumper retainer
screws in the holes they were removed from (the holes are countersunk to
accept the screws, and the screws can safely be inserted fully into the
holes).
1 Remove all detachable connector fittings and/or adapters from the connectors
on the instrument’s front panel.
2 Remove the front and rear bumpers (see “Removing the Bumpers” on page
93).
The bumpers can be used to support and protect the instrument after the shroud
NOTE is removed. Place both bumpers on their inner edges on your work surface, and
then place the instrument on top of them. The rear bumper can also serve as a
holder for the instrument (once the bumper retainers are removed) when you
need to position it upright on either end.
6 Slide the shroud off the chassis to the rear (see Figure 3-4), and place it aside
(if the shroud does not slide easily off the chassis, press on the rear panel with
one hand while pulling the shroud with the other hand).
Do not attempt to fully loosen the nut at this time. As soon as the nut turns
CAUTION freely, stop turning it, and proceed to the next step.
11 Remove the two 6-32 x ¼ flathead machine screws from the countersunk
holes at the top-front corners of the chassis (see Figure 3-5).
12 Remove the five 4-40 x ¼-inch flathead machine screws that attach the
front-panel assembly to the bottom of the chassis.
13 Remove the two 6-32 x ¼-inch flathead machine screws from the countersunk
holes at each side of the front the end of the chassis.
14 While holding the front panel assembly with one hand, use the fingers of your
other hand to continue to loosen the nut on the Channel 2 input connector.
Gently pull the front panel assembly away from the chassis as you loosen the
nut until the front panel assembly is completely disconnected.
When reattaching the front panel assembly, reverse the procedure above.
CAUTION You must use a microwave torque wrench to tighten the nut that attaches
the Channel 2 input connector to the sampler. Using any other tool to
tighten the coupling nut will damage the connector, the sampler, or both.
Battery
! !
Battery
Rear panel AC POWER
90-130 VAC 50/60/400 Hz
190-230 VAC 50/60 Hz
FUSE
1.5A T
250V
OPTIONS
001 Oven Time Base option Rear panel
screws 002 Battery
screws Battery
Charging
screws
RS-232 GPIB EXT DC
11 to 18 VDC
+
To avoid electrical shock: No user serviceable parts inside. This unit must be earth grounded. Made in U.S.A.
WARNING: Do not remove covers. Refer all servicing to qualified personnel. of domestic and foreign content
5 Remove the nine 4-40 x 3/16-inch self-locking flathead machine screws that
attach the rear panel assembly to the chassis (see Figure 3-6).
The screw in the lower-left corner of the rear panel assembly (as viewed from the
NOTE rear of the Counter) may be covered by the serial-number sticker. If it is,
carefully cut away the lower-left corner of the sticker to expose the screw head.
11 Carefully pull the rear panel assembly away from the back end of the chassis.
Guide the cables that remain attached to the rear panel assembly as you
remove it to prevent them from catching on other cables or other components
of the Counter.
4 While removing the cooling fan from the chassis, guide the cable to be sure it
does not catch on other cables or other components of the instrument.
8 Turn the instrument on, and perform the “Power Supply Check” procedure on
page 81.
9 If necessary, adjust the +5 VDC output of the power supply using the
procedure titled “Adjusting the +5 VDC Output” on page 82.
10 Reassemble the instrument.
4 Replaceable Parts
Introduction 108
Replaceable Parts 109
How To Order A Part 110
Parts Identification 112
107
4 Replaceable Parts
Introduction
This chapter contains parts-ordering information for the Keysight 553147A/148A/
149A Counter. Table 4-1 lists the reference designations used in the parts list and
throughout this guide, Table 4-2 lists all replaceable assemblies and parts.
Figure 4-1 is an exploded view of the Keysight 553147A/148A/149A’s internal
parts, Figure 4-2 is an exploded view of the cabinet and other external parts, and
Figure 4-3 is a top view of the chassis with the shroud removed. These illustrations
are provided to help you locate and identify parts.
Replaceable Parts
Table 4-2 contains a list of replaceable parts that is organized as follows:
1 Electrical assemblies in alphanumeric order by reference designation.
2 Chassis-mounted electrical parts in alphanumeric order by reference
designation.
3 Chassis-mounted mechanical parts in alphanumeric order by reference
designation.
The information given for each part consists of the following:
1 Reference designation.
2 Keysight part number.
3 The quantity (QTY) of the part used in the instrument (field-replaceable items
only).
4 Description of the part.
5 Typical manufacturer's part number for the part (when available).
Parts Identification
To identify the part(s) you need, refer to the exploded views in Figure 4-1 and
Figure 4-2 and the top view of the chassis in Figure 4-3.
When ordering from Keysight Technologies, the important numbers to note from
the Parts List are the Keysight Part Number and the quantity of the part(s) you
need.
Reference Designations
Table 4-1 lists the reference designations used in the parts lists.
H3, H4 A6
A7
H6
H1
H5
H7
A5 H1 A8
H7
H5 A9
A4 H5 H8
H1
H10 H9
A11
H3
H5
H4 H9
H3 H5
A3 A12 H1
H12
H11
A13
A10
H2 H13
W1
MP1 H2
A2 W2
H2
H1
A1
H1
MP6
H14
H14
MP4
H14 H14
MP5
MP4
H14
MP2
MP4
MP3 H14
H14
MP4
MP4
H14
H14
H15
MP7
Figure 4-2 Keysight 53147A/148A/149A Exploded View — Cabinet and
External Parts
W2
A10
W4
W5
P/O A6 W7
W6
W3 W4
W7
W1
W7
W6
W5
A13
W4
W3 W8
W9
W3
W10 W9
Figure 4-3 Keysight 53147A/148A/149A Top View
5 Backdating
Introduction 122
121
5 Backdating
Introduction
As of the current edition of this guide (Edition 3, November 1, 2017), this chapter
does not contain any information for adapting this guide to apply to older
instruments.
As engineering changes are made, newer instruments may have higher serial
prefix numbers than the ones shown on the title page of this guide. These
instruments may be supplied with a Manual Updating Changes package
containing the required information, or they may be supplied with newer editions
of this guide that contain information about adapting that edition of this guide to
older instruments.
If a Manual Updating Changes package is included with your Counter, modify
existing guide information or replace the affected pages as directed in the pages
of the Manual Updating Changes package. For additional information, contact the
nearest Keysight Sales and Support Office.
6 Specifications
For the specifications and characteristics of the 53147A/148A/149A Microwave
Frequency Counter/Power Meter/DVM, refer to the datasheet at
http://literature.cdn.keysight.com/litweb/pdf/5988-0300EN.pdf
123
6 Specifications
125
A Rack Mounting the Instrument
Unlike the hardware used elsewhere in this instrument, all hardware used to
NOTE attach the handle pivots and the rack mounts to the instrument
is metric.
1 Remove the front bumper by lifting the inner edge of the bumper away from
the shroud near one corner at the top and one side and pulling that corner
away from the instrument. Repeat the process with the remaining corners until
the bumper is free of the instrument.
COUNTER
Gate
2 Pull out on both sides of the handle at the handle pivots, and rotate the handle
towards the top of the instrument until the handle is touching the top of the
shroud.
3 Use a T15 Torx™ tool to remove the three screws that attach each of the
handle pivots to the sides of the instrument, and remove the handle.
4 Use a #1 Phillips™ screwdriver to remove the two screws that attach each of
the front bumper retainers to the top and bottom of the instrument (there are
four front bumper retainers—two each on the top and bottom).
6 Insert the eight 3/8-inch x 4-40, pan-head Phillips screws (2) supplied in the
Rack Mount Kit through the holes in the top and bottom of the rack-mount
panel into the threaded holes in the top and bottom of the instrument. Start
the screws, but do not tighten them at this time.
2 1
3, 4
2
3, 4
1
7 Insert the six 10mm long, 4 mm x .7, pan-head Torx screws (1) supplied in the
Rack Mount Kit through the holes in the side flanges of the rack-mount panel
and tighten them fully.
8 Tighten the eight screws you inserted in Step 6.
You can use angle brackets to help support the instrument in the rack and to
NOTE make it easier to mount the instrument.
9 Place the four Nylon washers supplied in the Rack Mount Kit over the four
10-32 screws (3, 4).
10 Inspect the holes in the rack to determine if they are threaded. If the holes are
threaded, skip Step 11.
11 Insert the four sheet-metal U-nuts provided in the Rack Mount Kit in the rack
flange behind the four holes you intend to use to mount the instrument.
12 While holding the instrument in position in the rack, insert one of the 10-32
screws with the Nylon washer on it (3,4) through one of the upper holes in the
rack-mount panel into the threaded hole in the rack (or into the U-nut you
installed in Step 11 if the hole is not threaded).
13 Repeat the procedure in Step 12 for the remaining upper rack-mount hole and
the two lower rack-mount holes.
14 Tighten all four rack-mount screws.
15 Connect the power cord to the Main ~ Power input connector on the
instrument’s rear panel.
16 Connect the other end of the power cord to an appropriate AC power source.
S W
safety, 76 warranty, 77
Self-Test, 83
Serial Number, 110
service, 55
Shroud, 96
specifications, 30
Standby mode, 22
surface-mount components, 77
T
Test 1
Channel 1 Frequency
Sensitivity, 31
*53147-90010*
53147-90010
www.keysight.com