G7115 G7165 DAD MWD UseMa en SD 29000218
G7115 G7165 DAD MWD UseMa en SD 29000218
Wavelength Detector
Agilent InfinityLab LC Series
User Manual
In This Guide...
This manual covers the Agilent 1260 Infinity II Diode Array Detector WR and the
Agilent 1260 Infinity II Multiple Wavelength Detector modules:
• G7115A - 1260 Infinity II DAD WR
• G7165A - 1260 Infinity II MWD
1 Introduction
This chapter gives an introduction to the module.
7 Error Information
This chapter describes the meaning of error messages, and provides information
on probable causes and suggested actions how to recover from error conditions.
9 Maintenance
This chapter describes the maintenance of the detector.
11 Identifying Cables
This chapter provides information on cables used with the Agilent 1200 Infinity
Series modules.
12 Hardware Information
This chapter describes the detector in more detail on hardware and electronics.
13 LAN Configuration
This chapter provides information on connecting the module to the Agilent
ChemStation PC.
14 Appendix
This chapter provides addition information on safety, legal and web.
1 Introduction 9
Introduction to the Detector 10
G7115A Diode Array Detector WR 11
G7165A Multiple Wavelength Detector 13
Optical System 15
Leak and Waste Handling 17
Bio-inert Materials 22
7 Error Information 97
What Are Error Messages 98
General Error Messages 99
Detector Error Messages 107
14 Appendix 264
General Safety Information 265
Waste Electrical and Electronic Equipment (WEEE) Directive 271
Radio Interference 272
Sound Emission 273
UV-Radiation 274
Solvent Information 275
Declaration of Conformity for HOX2 Filter 276
Installation of Stainless Steel Cladded PEEK Capillaries 277
Agilent Technologies on Internet 283
The detector is designed for highest optical performance, GLP compliance and
easy maintenance. It includes the following features:
• 120 Hz data acquisition rate for (ultra-) fast LC applications,
• RFID tags for all flow cells and UV-lamps provides traceable information
about these assemblies,
• long-life deuterium with RFID tag and tungsten lamps for highest intensity and
lowest detection limit over a wavelength range of 190 – 950 nm,
• no loss in sensitivity for up to eight wavelengths simultaneous,
• programmable slit from 1 – 16 nm for complete optimization of sensitivity,
linearity and spectral resolution,
• optional flow-cell cartridges with RFID tag (standard 10 mm13 µL, semi-micro
6 mm5 µL, micro 3 mm2 µL, 80 nL, 500 nL, 10 mm, high pressure
10 mm1.7 µL and prep-cells) are available and can be used depending on the
application needs,
• easy front access to lamps and flow cell for fast replacement, and
• built-in holmium oxide filter for fast wavelength accuracy verification,
• built-in temperature control for improved baseline stability,
• additional diagnostic signals for temperature and lamp voltage monitoring,
For specifications, see “Specifications” on page 29.
Product Description
The 1260 Infinity II DAD WR detector is designed for highest optical performance,
GLP compliance, and easy maintenance. With its 120 Hz data acquisition rate the
detector is perfectly suited for fast LC applications. The long –life deuterium
lamps allow highest intensity and lowest detection limits over a wavelength
range of 190 – 950 nm. The use of RFID tags for all flow cells and UV-lamps
provides traceable information about these assemblies.
The built-in holmium oxide filter features the fast wavelength accuracy
verification, while the built-in temperature controls improves the baseline
stability. Additional diagnostic signals for temperature and lamp voltage
monitoring are available.
Status indicator
Flow cell
Power switch
Leak drain
Features
• Higher sensitivity and selectivity - simultaneous detection of up to eight
compound-specific wavelengths.
• Low detection limits - low noise front-end electronics and the patented flow
cell design delivers very low detection limits thanks to the minimization of
short-term noise (< ± 7 µAU).
• Up to 100 % resolution gain in fast LC - using an 120 Hz data acquisition rate.
• Maximum baseline stability - electronic temperature control (ETC) reduces
baseline drift under fluctuating ambient temperature and humidity conditions.
• Wide linear range - for reliable, simultaneous quantification of primary
compounds, by-products, and impurities.
• Programmable slit (1 – 16 nm) for rapid optimization of sensitivity and
linearity.
• Excellent data traceability - radio frequency identification (RFID) tags on cells
and source lamps improve traceability of data.
• Automatic wavelength verification by built-in holmium oxide filter.
• Nine analytical and preparative flow cells provide you with maximum
application flexibility and choice.
• Extensive diagnostics, error detection and display with Instant Pilot controller
and Lab Advisor software.
Product Description
The 1260 Infinity II Multiple Wavelength Detector is designed for highest optical
performance, GLP compliance, and easy maintenance. With its 120 Hz data
acquisition rate, the detector is perfectly suited for fast LC applications. The
long-life deuterium lamps allows highest intensity and lowest detection limits
over a wavelength range of 190 – 950 nm. The use of RFID tags for all flow cells
and UV-lamps provides traceable information about these assemblies.
The built-in holmium oxide filter features the fast wavelength accuracy
verification, while the built-in temperature controls improves the baseline
stability. Additional diagnostic signals for temperature and lamp voltage
monitoring are available.
Status indicator
Flow cell
Power switch
Leak drain
Features
• Higher sensitivity and selectivity - simultaneous detection of up to eight
compound-specific wavelengths.
• Low detection limits - low noise front-end electronics and the patented flow
cell design delivers very low detection limits thanks to the minimization of
short-term noise (< ± 7 µAU).
• Up to 100 % resolution gain in fast LC - using a 120 Hz data acquisition rate.
• Maximum baseline stability - electronic temperature control (ETC) reduces
baseline drift under fluctuating ambient temperature and humidity conditions.
• Wide linear range - for reliable, simultaneous quantification of primary
compounds, by-products and impurities.
• Programmable slit (1 – 16 nm) for rapid optimization of sensitivity and
linearity.
• Excellent data traceability - radio frequency identification (RFID) tags on cells
and source lamps improve traceability of data.
• Automatic wavelength verification by built-in holmium oxide filter.
• Nine analytical and preparative flow cells provide you with maximum
application flexibility and choice.
• Extensive diagnostics, error detection and display with Instant Pilot controller
and Lab Advisor software.
Optical System
The optical system of the detector is shown in Figure below. Its illumination
source is a combination of a deuterium-arc-discharge lamp for the ultraviolet
(UV) wavelength range and a tungsten lamp for the visible (VIS) and short-wave
near-infrared (SWNIR) wavelength range. The image of the filament of the
tungsten lamp is focused on the discharge aperture of the deuterium lamp by
means of a special rear-access lamp design which allows both light sources to
be optically combined and share a common axis to the source lens. The
achromat (source lens) forms a single, focused beam of light through the flow
cell. Each cell room and lamp are separated by a quartz window which can be
cleaned or replaced. In the spectrograph, light is being dispersed onto the diode
array by a holographic grating. This allows simultaneous access to all
wavelength information.
Coupling lens
Deuterium lamp
Flow cell
Spectro lens
Slit
Grating
Diode array
Figure 3 Optical System of the Detector
Lamps The light source for the UV-wavelength range is a deuterium lamp with a
shine-through aperture. As a result of plasma discharge in low-pressure
deuterium gas, the lamp emits light over the 190 nm to approximately 800 nm
wavelength range. The light source for the visible and SWNIR wavelength range
is a low noise tungsten lamp. This lamp emits light over the wavelength range
470 – 950 nm.
Achromat The achromat receives the light from both lamps and focuses it so that the beam
(Source Lens) passes through the flow cell.
Holmium The holmium oxide filter is electromechanically actuated. During the holmium
Oxide Filter filter test it moves into the light path.
Cell Support The cell support window assembly separates the holmium filter area from the
Window flow cell area.
Flow Cell The optical unit has a flow cell compartment for easy access to flow cells. A
Compartment variety of optional flow cells can be inserted using the same quick, simple
mounting system. The flow cell can be removed to check the optical and
electronic performance of the detector without having influences from the flow
cell.
Spectrograph The spectrograph material is ceramic to reduce thermal effects to a minimum.
The spectrograph consists of the spectrograph lens, the variable entrance slit,
the grating and the photodiode array with front-end electronics. The
spectrograph lens refocuses the light beam after it has passed through the flow
cell. The sampling interval of the diode array is < 1 nm over the wavelength range
190 – 950 nm. Depending on the wavelength this varies from 1.0 to 1.25 diodes
per nanometer (for example a diode every 0.8 to 1 nm).
For a small wavelength range, the small non-linearity could be neglected. With
the wavelength range from 190 – 950 nm a new approach is required to achieve
wavelength accuracy over the full range. Each spectrograph is calibrated
individually. The calibration data is stored in the spectrograph on an EEPROM.
Based on these data, the built-in processors calculate absorbance data with
linear intervals (1.0, 2.0, …) between data points. This results in an excellent
wavelength accuracy and instrument-to-instrument reproducibility.
Variable The micro-slit system makes use of the mechanical properties of silicon
Entrance Slit combined with the precise structuring capabilities of bulk micro-machining. It
System combines the required optical functions — slit and shutter — in a simple and
compact component. The slit width is directly controlled by the micro-processor
of the instrument and can be set as method parameter.
Grating The combination of dispersion and spectral imaging is accomplished by using a
concave holographic grating. The grating separates the light beam into all its
component wavelengths and reflects the light onto the photodiode array.
Diode Array The diode array is a series of 1024 individual photodiodes and control circuits
located on a ceramic carrier. With a wavelength range from 190 – 950 nm the
sampling interval is < 1 nm.
The Agilent InfinityLab LC Series has been designed for safe leak and waste
handling. It is important that all security concepts are understood and
instructions are carefully followed.
The solvent cabinet is designed to store a maximum volume of 8 L solvent. The
maximum volume for an individual bottle stored in the solvent cabinet should not
exceed 2 L. For details, see the usage guideline for the Agilent Infinity II Solvent
Cabinets (a printed copy of the guideline has been shipped with the solvent
cabinet, electronic copies are available on the Internet).
All leak plane outlets are situated in a consistent position so that all Infinity and
Infinity II modules can be stacked on top of each other. Waste tubes are guided
through a channel on the right hand side of the instrument, keeping the front
access clear from tubes.
The leak plane provides leak management by catching all internal liquid leaks,
guiding them to the leak sensor for leak detection, and passing them on to the
next module below, if the leak sensor fails. The leak sensor in the leak plane
stops the running system as soon as the leak detection level is reached.
Solvent and condensate is guided through the waste channel into the waste
container:
• from the detector's flow cell outlet
• from the Multisampler needle wash port
• from the Sample Cooler or Sample Thermostat (condensate)
• from the pump's Seal Wash Sensor (if applicable)
• from the pump's Purge Valve or Multipurpose Valve
The waste tube connected to the leak plane outlet on each of the bottom
instruments guides the solvent to a suitable waste container.
Leak Sensor
Solvent incompatibility
C AU T I O N
The solvent DMF (dimethylformamide) leads to corrosion of the leak sensor.
The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible
with DMF.
Do not use DMF as mobile phase.
Check the leak sensor regularly for corrosion.
Waste Concept
1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use
your own waste solution, make sure that the tubes don't immerse in the liquid.
Bio-inert Materials
For the Bio-inert LC system, Agilent Technologies uses highest quality materials
in the flow path (also referred to as wetted parts), which are widely accepted by
life science scientists, as they are known for optimum inertness to biological
samples and ensure best compatibility with common samples and solvents over
a wide pH range. Explicitly, the complete flow path is free of stainless steel and
free of other alloys containing metals such as iron, nickel, cobalt, chromium,
molybdenum or copper, which can interfere with biological samples. The flow
downstream of the sample introduction contains no metals whatsoever.
Table 1 Used Bio-inert materials
Module Materials
Agilent 1260 Infinity II Bio-inert Analytical Fraction Collector PEEK, ceramic, PTFE
(G5664B)
Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar ( 12 MPa) for MWD/DAD, PEEK, ceramic, sapphire, PTFE
includes 0890-1763 – 0.18 x 1500 mm PEEK capillary and 5063-6591 – PEEK
fittings (G5615-60022)
(for Agilent 1260 Infinity II DAD G7115A, and MWD G7165A)
Bio-inert flow cell, 8 µL, 20 bar (pH 1–12), includes 0890-1763 – 0.18 x 1500 mm PEEK, fused silica, PTFE
PEEK capillary and 5063-6591 – PEEK fittings (G5615-60005)
(for Agilent 1260 Infinity II FLD G7121A/B)
Module Materials
Bio-inert Valve heads (G4235A, G5631A, G5632A, G5639A) PEEK, ceramic (Al2O3 based)
Site Requirements 25
Physical Specifications 28
Performance Specifications 29
Specifications 29
Specification Conditions 35
Site Requirements
A suitable environment is important to ensure optimal performance of the
instrument.
Power Considerations
The module power supply has wide ranging capability. It accepts any line voltage
in the range described in Table 2 on page 28. Consequently there is no voltage
selector in the rear of the module. There are also no externally accessible fuses,
because automatic electronic fuses are implemented in the power supply.
Power Cords
Country-specific power cords are available for the module. The female end of all
power cords is identical. It plugs into the power-input socket at the rear. The male
end of each power cord is different and designed to match the wall socket of a
particular country or region.
Agilent makes sure that your instrument is shipped with the power cord that is
suitable for your particular country or region.
Bench Space
The module dimensions and weight (see Table 2 on page 28) allow you to place
the module on almost any desk or laboratory bench. It needs an additional
2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches)
in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is
designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Agilent recommends that you install the HPLC instrument in the InfinityLab Flex
NOTE Bench rack. This option helps to save bench space as all modules can be placed
into one single stack. It also allows to easily relocate the instrument to another lab.
Environment
Your detector will work within the specifications at ambient temperatures and
relative humidity described in Table 2 on page 28.
ASTM drift tests require a temperature change below 2 °C/hour (3.6 °F/hour)
over one hour period. Our published drift specification (refer also to “Performance
Specifications G7115A” on page 29 and “Performance Specifications
G7165A” on page 32) is based on these conditions. Larger ambient temperature
changes will result in larger drift.
Better drift performance depends on better control of the temperature
fluctuations. To realize the highest performance, minimize the frequency and the
amplitude of the temperature changes to below 1 °C/hour (1.8 °F/hour).
Turbulences around one minute or less can be ignored.
Physical Specifications
Safety standards: IEC, EN, Overvoltage category II, Pollution degree 2 For indoor use only.
CSA, UL
Performance Specifications
Specifications
Performance Specifications G7115A
Table 3 Performance Specifications G7115A
Type Specification
Number of signals 8
Slit width 1, 2, 4 , 8, 16 nm
Diode width ~1 nm
Time programmable Wavelength, polarity, peak width, lamp bandwidth, auto balance,
wavelength range, threshold, spectra storage mode
Type Specification
Flow cells Standard: 13 µL volume, 10 mm cell path length and 120 bar
(1740 psi) pressure maximum
Standard bio-inert: 13 µL volume, 10 mm cell path length and 120 bar
(1740 psi) pressure maximum
Semi-micro: 5 µL volume, 6 mm cell path length and 120 bar
(1740 psi) pressure maximum
Micro: 2 µL volume, 3 mm cell path length, 120 bar (1740 psi)
pressure maximum
Semi-nano: 500 nL volume, 10 mm cell path length and 40 bar
(580 psi) pressure maximum
Nano: 80 nL volume, 6 mm cell path length and 40 bar (580 psi)
pressure maximum
High pressure: 1.7 µL volume, 6 mm cell path length and 400 bar
(5800 psi) pressure maximum
Prep SST: 3 mm cell path length and 120 bar (1740 psi) pressure
maximum
Prep Quartz: 3 mm cell path length and 50 bar (1740 psi) pressure
maximum
Prep Quartz:0.3 mm cell path length and 50 or 20 bar (290 psi)
pressure maximum
Prep Quartz: 0.06 mm cell path length and 50 or 20 bar (290 psi)
pressure maximum
SFC Flow Cell: 13 μL volume, 10 mm cell path length and 400 bar
(5800 psi) pressure maximum
SFC Flow Cell LD: 2 μL volume, 3 mm cell path length and 400 bar
(5800 psi) pressure maximum
Spectral tools Data analysis software for spectra evaluation, including spectral
libraries and peak purity functions
Type Specification
GLP RFID for electronics records of flow cell and UV lamp conditions (path
length, volume, product number, serial number, test passed, usage)
Early maintenance feedback (EMF) for continuous tracking of
instrument usage in terms of lamp burn time with user-setable limits
and feedback messages. Electronic records of maintenance and
errors. Verification of wavelength accuracy with built-in holmium
oxide filter.
Safety and maintenance Extensive diagnostics, error detection and display through Agilent
Instant Pilot and Agilent Lab Advisor software. Leak detection, safe
leak handling, leak output signal for shutdown of pumping system.
Low voltages in major maintenance areas.
Type Specification
Number of signals 8
Slit width 1, 2, 4 , 8, 16 nm
Type Specification
Flow cells Standard: 13 µL volume, 10 mm cell path length and 120 bar
(1740 psi) pressure maximum
Standard bio-inert: 13 µL volume, 10 mm cell path length and 120 bar
(1740 psi) pressure maximum
Semi-micro: 5 µL volume, 6 mm cell path length and 120 bar
(1740 psi) pressure maximum
Micro: 2 µL volume, 3 mm cell path length, 120 bar (1740 psi)
pressure maximum
Semi-nano: 500 nL volume, 10 mm cell path length and 40 bar
(580 psi) pressure maximum
Nano: 80 nL volume, 6 mm cell path length and 40 bar (580 psi)
pressure maximum
High pressure: 1.7 µL volume, 6 mm cell path length and 400 bar
(5800 psi) pressure maximum
Prep SST: 3 mm cell path length and 120 bar (1740 psi) pressure
maximum
Prep Quartz:0.3 mm cell path length and 50 or 20 bar (290 psi)
pressure maximum
Prep Quartz: 0.06 mm cell path length and 50 or 20 bar (290 psi)
pressure maximum
SFC Flow Cell: 13 μL volume, 10 mm cell path length and 400 bar
(5800 psi) pressure maximum
SFC Flow Cell LD: 2 μL volume, 3 mm cell path length and 400 bar
(5800 psi) pressure maximum
GLP RFID for electronics records of flow cell and UV lamp conditions (path
length, volume, product number, serial number, test passed, usage)
Early maintenance feedback (EMF) for continuous tracking of
instrument usage in terms of lamp burn time with user-setable limits
and feedback messages. Electronic records of maintenance and
errors. Verification of wavelength accuracy with built-in holmium
oxide filter.
Type Specification
Safety and maintenance Extensive diagnostics, error detection and display through Agilent
Instant Pilot and Agilent Lab Advisor software. Leak detection, safe
leak handling, leak output signal for shutdown of pumping system.
Low voltages in major maintenance areas. Tracking of flow cells and
lamps with RFID (radio frequency identification) tags.
Specification Conditions
ASTM: “Standard Practice for Variable Wavelength Photometric Detectors Used
in Liquid Chromatography”.
Reference conditions: cell path length 10 mm, wavelength 254 and 750 nm with
reference wavelength 360 nm/100 nm, slit width 4 nm, time constant 2 s (equal
to response time 4 s), flow 1 mL/min LC-grade Methanol.
Linearity: Linearity is measured with caffeine at 265 nm/4 nm with slit width 4 nm
and TC 2 s (or with RT 4 s) with 10 mm pathlength.
For environmental conditions refer to "Environment".
The specifications are based on the standard RFID tag lamp (2140-0820) and
NOTE may be not achieved when other lamp types or aged lamps are used.
Mobile devices used close to the intstrument could affect the detector's short
NOTE term noise level.
ASTM drift tests require a temperature change below 2 °C/hour (3.6 °F/hour)
over one hour period. Our published drift specification is based on these
conditions. Larger ambient temperature changes will result in larger drift. Better
drift performance depends on better control of the temperature fluctuations. To
realize the highest performance, minimize the frequency and the amplitude of the
temperature changes to below 1 °C/hour (1.8 °F/hour). Turbulences around one
minute or less can be ignored.
Performance tests should be done with a completely warmed up optical unit (>
two hours). ASTM measurements require that the detector should be turned on
at least 24 h before start of testing.
Magnets 37
Turn on/off 38
Status Indicators 40
Instrument Configuration 41
Set up the Detector with Agilent Open Lab ChemStation 42
The Detector User Interface 43
Detector Control Settings 45
Method Parameter Settings 46
Agilent Local Control Modules 50
Magnets
Turn on/off
1 2
Power switch: On
Status Indicators
Status indicators
1. Idle
2. Run mode
3. Not-ready. Waiting for a specific pre-run condition to be reached or
completed.
4. Error mode - interrupts the analysis and requires attention (for example, a
leak or defective internal components).
5. Resident mode (blinking) - for example, during update of main firmware.
6. Bootloader mode (fast blinking). Try to re-boot the module or try a
cold-start. Then try a firmware update.
Instrument Configuration
1 Set the switches of the Configuration switch at the rear of the module:
a All switches DOWN: module uses the default IP address 192.168.254.11.
The setup of the detector is shown with the Agilent OpenLab ChemStation
C.01.07 and Driver A.02.14.
This section describes the detector settings only. For information on the Agilent
NOTE OpenLab ChemStation or other 1200 Infinity modules refer to the corresponding
documentation.
After successful load of the OpenLab ChemStation, you should see the module
as an active item in the graphical user interface (GUI).
Within the detector GUI, there are active areas. If you move the
mouse cursor across the icons the cursor will change.
1 Lamp: turn on and off of UV-lamp
2 EMF status
3 Detector status
4 Detector balance status
5 Lamp status (on/off)
6 Lamp information (RFID tag)
7 Flow Cell information (RFID tag)
EMF Status shows Run / Ready / Error state and “Not Ready
text” or “Error text”
• Offline (gray)
• Ok. No Maintenance required (green)
• EMF warning. Maintenance might be required (yellow)
• EMF warning. Maintenance required (red)
Important: The EMF settings can be accessed via Agilent Lab
Advisor. The limit(s) can be changed. Based on the limit, the
User Interface displays the above status.
If the flow cell temperature is critical for your chromatography, you may set the
NOTE Temperature Control to off. This will lower the optical unit and flow cell
temperature by some degree C.
These settings are available via Menu > Instrument > Set up Instrument Method or
via right click into the module’s active area (does not show the Instrument Curves
tab).
For additional help and support: Highlight the desired cell and press F1. A help
NOTE screen will open with additional information and documentation about the topic.
Stoptime/Posttime
The stoptime is the time where either the complete system
stops (As Pump/Injector) or the module (if different from
system stop time). The data collection is stopped at this time.
A posttime period can be used to allow module’s items to
equilibrate (e.g. after gradient change or temperature change).
Analog Output
The range can be set to either 100 mV or 1 V full scale, see
“Detector Control Settings” on page 45.
• Zero Offset: 1 – 99 % in steps of 1 % (5 % equal to 50 mV).
• Attenuation: 0.98 – 2000 mAU at discrete values for either
100 mV or 1 V full scale.
Autobalance
Defines, whether a balance is performed prior to a run and/or
after a run has finished.
Lamp on required for acquisition:
If unchecked, the lamp will be turned off after the analysis has
finished. Note that the lamp on requires at least one hour
warm-up time, see “Warm up of the Detector” on page 89.
Instrument Curves
The detector has several signals (internal temperatures,
voltages of lamps) that can be used for diagnosing problems.
These can be baseline problems deriving from deuterium
lamps wander / drift problems due to temperature changes.
These signals can be used in addition to the normal baseline
signal to determine whether correlation to temperature or
voltage/current of the lamp.
These signals are available via the Agilent ChemStation Online
Plot/Data Signal and/or Agilent Lab Advisor Software.
This chapter provides information on how to set up the module for an analysis
and explains the basic settings.
For details, see the usage guideline for the solvent cabinet. A printed copy of the
NOTE guideline has been shipped with the solvent cabinet, electronic copies are
available in the Agilent Information Center or via the Internet.
Waste Concept
1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use
your own waste solution, make sure that the tubes don't immerse in the liquid.
Setting up an Analysis
When switching between reverse phase and Isopropanol Best solvent to flush air out of the system
normal phase (both times)
After a solvent change Bidistilled water Best solvent to re-dissolve buffer crystals
After the installation of normal phase seals Hexane + 5% Isopropanol Good wetting properties
(P/N 0905-1420)
The pump should never be used for priming empty tubings (never let the pump
NOTE run dry). Use a syringe to draw enough solvent for completely filling the tubings
to the pump inlet before continuing to prime with the pump.
1 Open the purge valve of your pump (by turning it counterclockwise) and set
flow rate to 3 – 5 mL/min.
2 Flush all tubes with at least 30 mL of solvent.
3 Set flow to required value of your application and close the purge valve.
Autosampler
Degasser (optional)
Agilent CDS
Conditions
A single injection of the isocratic test standard is made under the conditions
given in Table 8 on page 56:
Table 8 Conditions
Stoptime 8 min
Temperature Ambient
Injection Volume 1 µL
Typical Chromatogram
A typical chromatogram for this analysis is shown in Figure 9 on page 57. The
exact profile of the chromatogram will depend on the chromatographic
conditions. Variations in solvent quality, column packing, standard concentration
and column temperature will all have a potential effect on peak retention and
response.
7 Select the menu item Run Control > Sample Info and enter information about
this application. Click OK to leave this screen.
Information about using the Data Analysis functions can be obtained from the
NOTE Using your ChemStation manual supplied with your system.
Solvent Information
Disclaimer
Subsequent data was collected from external resources and is meant as a
reference. Agilent cannot guarantee the correctness and completeness of such
information. Data is based on compatibility libraries, which are not specific for
estimating the long-term life time under specific but highly variable conditions of
UHPLC systems, solvents, solvent mixtures and samples. Information can also
not be generalized due to catalytic effects of impurities like metal ions,
complexing agents, oxygen etc. Apart from pure chemical corrosion, other
effects like electro corrosion, electrostatic charging (especially for
non-conductive organic solvents), swelling of polymer parts etc. need to be
considered. Most data available refers to room temperature (typically 20 – 25 °C,
68 – 77 °F). If corrosion is possible, it usually accelerates at higher temperatures.
If in doubt, please consult technical literature on chemical compatibility of
materials.
MP35N
MP35N is a nonmagnetic, nickel-cobalt-chromium-molybdenum alloy
demonstrating excellent corrosion resistance (for example, against nitric and
sulfuric acids, sodium hydroxide, and seawater) over a wide range of
concentrations and temperatures. In addition, this alloy shows exceptional
resistance to high-temperature oxidation. Due to excellent chemical resistance
and toughness, the alloy is used in diverse applications: dental products, medical
devices, nonmagnetic electrical components, chemical and food processing
equipment, marine equipment. Treatment of MP35N alloy samples with 10 %
NaCl in HCl (pH 2.0) does not reveal any detectable corrosion. MP35N also
demonstrates excellent corrosion resistance in a humid environment. Although
the influence of a broad variety of solvents and conditions has been tested, users
should keep in mind that multiple factors can affect corrosion rates, such as
temperature, concentration, pH, impurities, stress, surface finish, and dissimilar
metal contacts.
PEEK
PEEK (Polyether-Ether Ketones) combines excellent properties regarding
biocompatibility, chemical resistance, mechanical and thermal stability. PEEK is
therefore the material of choice for UHPLC and biochemical instrumentation.
It is stable in the specified pH range (for the Bio-Inert LC system: pH 1 – 13, see
bio-inert module manuals for details), and inert to many common solvents.
There is still a number of known incompatibilities with chemicals such as
chloroform, methylene chloride, THF, DMSO, strong acids (nitric acid > 10 %,
sulfuric acid > 10 %, sulfonic acids, trichloroacetic acid), halogens or aqueous
halogen solutions, phenol and derivatives (cresols, salicylic acid, and so on).
When used above room temperature, PEEK is sensitive to bases and various
organic solvents, which can cause it to swell. Under such conditions, normal
PEEK capillaries are sensitive to high pressure. Therefore, Agilent uses stainless
steel cladded PEEK capillaries in bio-inert systems. The use of stainless steel
cladded PEEK capillaries keeps the flow path free of steel and ensures pressure
stability up to 600 bar. If in doubt, consult the available literature about the
chemical compatibility of PEEK.
Polyimide
Agilent uses semi-crystalline polyimide for rotor seals in valves and needle seats
in autosamplers. One supplier of polyimide is DuPont, which brands polyimide as
Vespel, which is also used by Agilent.
Polyimide is stable in a pH range between 1 and 10 and in most organic solvents.
It is incompatible with concentrated mineral acids (e.g. sulphuric acid), glacial
acetic acid, DMSO and THF. It is also degraded by nucleophilic substances like
ammonia (e.g. ammonium salts in basic conditions) or acetates.
Polyethylene (PE)
Agilent uses UHMW (ultra-high molecular weight)-PE/PTFE blends for yellow
piston and wash seals, which are used in 1290 Infinity pumps, 1290 Infinity II
pumps, the G7104C and for normal phase applications in 1260 Infinity pumps.
Polyethylene has a good stability for most common inorganic solvents including
acids and bases in a pH range of 1 to 12.5. It is compatible with many organic
solvents used in chromatographic systems like methanol, acetonitrile and
isopropanol. It has limited stability with aliphatic, aromatic and halogenated
hydrocarbons, THF, phenol and derivatives, concentrated acids and bases. For
normal phase applications, the maximum pressure should be limited to 200 bar.
Tantalum (Ta)
Tantalum is inert to most common HPLC solvents and almost all acids except
fluoric acid and acids with free sulfur trioxide. It can be corroded by strong bases
(e.g. hydroxide solutions > 10 %, diethylamine). It is not recommended for the use
with fluoric acid and fluorides.
Titanium (Ti)
Titanium is highly resistant to oxidizing acids (for example, nitric, perchloric and
hypochlorous acid) over a wide range of concentrations and temperatures. This
is due to a thin oxide layer on the surface, which is stabilized by oxidizing
compounds. Non-oxidizing acids (for example, hydrochloric, sulfuric and
phosphoric acid) can cause slight corrosion, which increases with acid
concentration and temperature. For example, the corrosion rate with 3 % HCl
(about pH 0.1) at room temperature is about 13 μm/year. At room temperature,
titanium is resistant to concentrations of about 5 % sulfuric acid (about pH 0.3).
Addition of nitric acid to hydrochloric or sulfuric acids significantly reduces
corrosion rates. Titanium is sensitive to acidic metal chlorides like FeCl3 or CuCl2.
Titanium is subject to corrosion in anhydrous methanol, which can be avoided by
adding a small amount of water (about 3 %). Slight corrosion is possible with
ammonia > 10 %.
Gold
Gold is inert to all common HPLC solvents, acids and bases within the specified
pH range. It can be corroded by complexing cyanides and concentrated acids like
aqua regia.
Platinum/Iridium
Platinum/Iridium is inert to almost all common acids, bases and solvents. There
are no documented incompatibilities for HPLC applications.
Optimization Overview
Parameter Impact
• Choose flow cell according to used column, see Figure 12 on page 70.
• For flow rates from 0.5 mL/min connect column using the
zero-dead-volume fittings of the detector.
• For small column i.d. (e.g 1 mm) the inlet capillary of the micro flow cell can
be connected directly to the column.
3 Setting the peak width (response time) • peak resolution versus sensitivity
versus disk space
• Sample wavelength:
• Never miss a peak by the use of a browser wavelength like 250 nm with • sensitivity versus selectivity
100 nm bandwidth.
• Select specific wavelength with reduced bandwidth if you need • sensitivity versus linearity
selectivity, e.g. 250,10 nm and 360,100 nm as reference wavelength.
• Set the sample wavelength to a peak or valley in the spectrum to get
best linearity for high concentrations.
• Reference wavelength:
• Select the reference wavelength with broad bandwidth (30...100 nm) • baseline drift due to RI effects.
wavelength range where your analytes have little or no absorbance (e.g.
sample at 254 nm, reference at 320 nm).
Parameter Impact
Typical flow 0.01 ... 0.2 0.2 ... 0.4 0.4 ... 0.4 1 ... 5 0.01 ... 5
rate ml/min ml/min ml/min ml/min ml/min
where
T is the transmission, defined as the quotient of the intensity of the transmitted
light I divided by the intensity of the incident light, I0,
Time (min)
Figure 13 Influence of Cell Path Length on Signal Height
Standard flow cell, 10 mm, 13 µL, 120 bar ( 12 MPa) 9.80±0.07 mm 10/9.8
(G1315-60022)
Semi-micro flow cell, 6 mm, 5 µL, 120 bar ( 12 MPa) 5.80±0.07 mm 6/5.8
(G1315-60025)
Micro flow cell, 3 mm, 2 µL, 120 bar ( 12 MPa) (G1315-60024) 3.00+0.05 mm/-0.07 mm 3/3
Semi-nano flow cell kit, 10 mm, 500 nL, 5 MPa (G1315-68724) 10.00±0.02 mm 10/10
Nano flow cell kit, 6 mm, 80 nL, 5 MPa ( G1315-68716) 6.00±0.02 mm 6/6
Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar ( 12 MPa) for 9.80±0.07 mm 10/9.8
MWD/DAD, includes 0890-1763 – 0.18 x 1500 mm PEEK
capillary and 5063-6591 – PEEK fittings (G5615-60022)
Unfiltered
Response time
0.05 min
Response time
0.1 min
Table 11 on page 74 list the filter choices of the detector. To get optimum results,
set peak width as close as possible to a narrow peak of interest in your
chromatogram. Response time will the be approximately 1/3 of the peak width,
resulting in less than 5 % peak-height reduction and less than 5 % additional peak
dispersion. Decreasing the peak width setting in the detector will result in less
than 5 % gain in peak height but baseline noise will increase by a factor of 1.4 for
a factor of 2 response-time reduction. Increasing peak width (response time) by
factor of two from the recommended setting (over-filtering) will reduce peak
height by about 20 % and reduce baseline noise by a factor of 1.4. This gives you
the best possible signal-to-noise ratio, but may affect peak resolution.
Peak width at half Response [s] Scan data Scan data Scan data
height [min]1 rate[Hz] ≤251 rate[Hz] ≤501 rate[Hz] > 501
pts/scan pts/scan pts/scan
> 0.05 1 5 5 5 5
The detector can compute and store at run time up to 8 signals with these
properties:
• sample wavelength, the center of a wavelength band with the width of sample
bandwidth (BW), and optionally
• reference wavelength, the center of a wavelength band with the width of
reference bandwidth.
The signals comprises a series of data points over time, with the average
absorbance in the sample wavelength band minus the average absorbance of
the reference wavelength band.
Signal A in the detector default method is set to sample 250,100, reference
360,100, that is, the average absorbance from 200 – 300 nm minus the average
absorbance from 300 – 400 nm. As all analytes show higher absorbance at
200 – 300 nm than at 300 – 400 nm, this signal will show you virtually every
compound which can be detected by UV absorbance.
Many compounds show absorbance bands in the spectrum. Figure 15 on
page 76 shows the spectrum of anisic acid as an example.
To optimize for lowest possible detectable concentrations of anisic acid, set the
sample wavelength to the peak of the absorbance band (that is, 252 nm) and the
sample bandwidth to the width of the absorbance band (that is, 30 nm). A
reference of 360,100 is adequate. Anisic acid does not absorb in this range.
If you work with high concentrations, you may get better linearity above 1.5 AU by
setting the sample wavelength to a valley in the spectrum, like 225 nm for anisic
acid.
30 nm
bandwidth
Reference wavelength 360
Wavelength (nm)
Bandwidth
30 nm
12 nm
4 nm
Because the detector averages absorbance values that are calculated for each
wavelength, using a wide bandwidth does not negatively impact linearity.
The use of a reference wavelength is highly recommended to further reduce
baseline drift and wander induced by room temperature fluctuations or refractive
index changes during a gradient.
An example of the reduction of baseline drifts is shown in Figure 17 on page 77
for PTH-amino acids. Without a reference wavelength, the chromatogram drifts
downwards due to refractive index changes induced by the gradient. This is
almost completely eliminated by using a reference wavelength. With this
technique, PTH-amino acids can be quantified in the low picomole range even in
a gradient analysis.
PTH-ASN
PTH-PHE
PTH-ALA
PTH-PRO
PTH-ARG
1pmol each
Wavelength
267 nm
Reference 380nm
Wavelength
267nm
No reference
Time (min)
Grad.: 0.02 m KH2 PO4 /ACN from 12% ACN to 45% ACN in 12 min
Figure 17 Gradient Analysis of PTH-Amino Acids (1 pmol each), with and without Reference
Slit Width
The detector has a variable slit at the entrance of the spectrograph. This is an
effective tool to adapt the detector to changing demand of different analytical
problems.
A narrow slit provides spectral resolution for analytes with very fine structures in
the absorbance spectrum. An example of such a spectrum is benzene. The five
main absorbance bands (fingers) are only 2.5 nm wide and just 6 nm apart from
each other.
16 nm
4 nm
1 nm
A wide slit uses more of the light shining through the flow cell. This gives lower
baseline noise as shown in Figure 19 on page 78.
Slit width 1 nm
Slit width 4 nm
Slit width 16 nm
However, with a wider slit, the spectrograph’s optical resolution (its ability to
distinguish between different wavelengths) diminishes. Any photodiode receives
light within a range of wavelength determined by the slit width. This explains why
the fine spectral structure of benzene disappears when using a 16-nm wide slit.
Furthermore, the absorbance is no longer strictly linear with concentration for
wavelengths at a steep slope of a compound’s spectrum.
Substances with fine structures and steep slopes like benzene are very rare.
In most cases the width of absorbance bands in the spectrum is more like 30 nm
as with anisic acid (Figure 15 on page 76).
In most situations, a slit width of 4 nm will give the best results.
Use a narrow slit (1 or 2 nm) if you want to identify compounds with fine spectral
structures or if you need to quantify at high concentrations (> 1000 mAU) with a
wavelength at the slope of the spectrum. Signals with a wide bandwidth can be
used to reduce baseline noise. Because (digital) bandwidth is computed as
average of absorbance, there is no impact on linearity.
Use a wide (8 or 16 nm) slit when your sample contains very small
concentrations. Always use signals with bandwidth at least as wide as the slit
width.
Optimizing Selectivity
WL3 WL4
With a UV-visible detector based on a diode array and the correct choice of a
reference wavelength setting, quantitative detection is possible. To suppress
caffeine, the reference wavelength must be set to 282 nm. At this wavelength,
caffeine shows exactly the same absorbance as at 222 nm. When the
absorbance values are subtracted from each another, any indication of the
presence of caffeine is eliminated. In the same way, hydrochlorothiazide can be
suppressed if caffeine is to be quantified. In this case the wavelength is set to
204 nm and the reference wavelength to 260 nm. Figure 21 on page 82 shows
the chromatographic results of the peak suppression technique.
The trade-off for this procedure is a loss in sensitivity. The sample signal
decreases by the absorbance at the reference wavelength relative to the signal
wavelength. Sensitivity may be decreased by as much as 10–30 %.
Hydrochlorothiazide Hydrochlorothiazide
and caffeine and caffeine
Wavelength 204 nm Wavelength 222 nm
No reference No reference
Extra-Column Volume
Extra-column volume is a source of peak dispersion that will reduce the
resolution of the separation and so should be minimized. Smaller diameter
columns require proportionally smaller extra-column volumes to keep peak
dispersion at a minimum.
In a liquid chromatograph the extra-column volume will depend on the
connection tubing between the autosampler, column and detector; and on the
volume of the flow cell in the detector. The extra-column volume is minimized
with the Agilent InfinityLab LC Series system due to the narrow-bore (0.12 mm
i.d.) tubing, the low-volume heat exchangers in the column compartment and the
flow cell in the detector.
Flow Cell
The Max-Light cartridge flow cell has a standard 10 mm path length and is
optimized for minimal volume and dispersion ( volume 1.0 µL). It has high light
transmission minimizing noise to reduce noise due to the optofluidic waveguide.
It is suitable for use with a wide range of analytical columns from short
narrow-bore columns to long standard diameter (4.6 mm) columns. Generally the
peak dispersion volume (calculated from peak width x flow rate) should be
greater than about 2 µL for this cell (for example 0.02 min x 200 µL/min = 4 µL).
The Max-Light high sensitivity cell has a path length of 60 mm and this will give
between three and five times increase in signal-to-noise values depending on the
application conditions. The dispersion volume is fractionally increased compared
to the standard cell.
Peak width at half Response [s] Scan data Scan data Scan data
height [min]1 rate[Hz] ≤251 rate[Hz] ≤501 rate[Hz] > 501
pts/scan pts/scan pts/scan
> 0.05 1 5 5 5 5
The maximum spectra scan rate depends on the data points per scan, see
NOTE Table 12 on page 88.
Give the optical unit enough time to warm-up and stabilize (> 60 minutes). The
detector is temperature controlled. After turn-on of the detector, it goes through a
cycle of different states:
• 0 to 0.5 minutes the heater control is OFF and the heater element runs at 0 %
duty cycle.
• 0.5 to 1 minutes the heater control is OFF and the heater element runs at 66%
duty cycle. This first minute is used as self-test of the heater functionality.
• 1 to 30 minutes the heater control is OFF and the heater element runs at 40%
duty cycle.
• After 30 minutes the heater control is ON and is working with optimized
parameters to get the optical unit into the optimal temperature window
stabilized.
This cycle starts
• when the detector is turned off/on
• when the lamp is turned off/on
to ensure that the temperature control operates in a defined control range.
The times to stabilize the baseline may vary from instrument to instrument and
NOTE depends on the environment. The example below was done under stable
environmental conditions.
The figures below show the first two hours of a detector warm-up phase. The
lamp was turned on immediately after turn on of the detector.
Depending on the used interface, the available tests and the screens/reports
NOTE may vary.
Preferred tool should be the Agilent Lab Advisor, see “Agilent Lab Advisor
Software” on page 93.
Agilent Lab Advisor B.02.08 or later is required.
Screenshots used within these procedures are based on the Agilent Lab Advisor
software.
The Agilent Lab Advisor Software (basic license, shipped with an Agilent LC
pump) is a standalone product that can be used with or without a
chromatographic data system. Agilent Lab Advisor helps to manage the lab for
high-quality chromatographic results by providing a detailed system overview of
all connected analytical instruments with instrument status, Early Maintenance
Feedback counters (EMF), instrument configuration information, and diagnostic
tests. With the push of a button, a detailed diagnostic report can be generated.
Upon request, the user can send this report to Agilent for a significantly improved
troubleshooting and repair process.
The Agilent Lab Advisor software is available in two versions:
• Lab Advisor Basic
• Lab Advisor Advanced
Lab Advisor Basic is included with every Agilent 1200 Infinity Series and Agilent
InfinityLab LC Series instrument.
The Lab Advisor Advanced features can be unlocked by purchasing a license key,
and include real-time monitoring of instrument actuals, all various instrument
signals, and state machines. In addition, all diagnostic test results, calibration
results, and acquired signal data can be uploaded to a shared network folder. The
Review Client included in Lab Advisor Advanced allows to load and examine the
uploaded data no matter on which instrument it was generated. This makes Data
Sharing an ideal tool for internal support groups and users who want to track the
instrument history of their analytical systems.
The optional Agilent Maintenance Wizard Add-on provides an easy-to-use,
step-by-step multimedia guide for performing preventive maintenance on Agilent
1200 Infinity LC Series instrument.
The tests and diagnostic features that are provided by the Agilent Lab Advisor
software may differ from the descriptions in this manual. For details, refer to the
Agilent Lab Advisor software help files.
Product Level
Tests
Calibrations
Tools
Product Level
-Configuration
-Analog Output 1 Range Advanced
-Control
-Vis lamp Basic Advanced
-Module Information
Statemachines
-UV Lamp Basic Advanced
Signals
Product Level
EMF Counters
-Accumulated UV Lamp Basic Advanced
On-Time
-Number of UV Lamp Ignitions Basic Advanced
This chapter describes the meaning of error messages, and provides information
on probable causes and suggested actions how to recover from error conditions.
General error messages are generic to all Agilent series HPLC modules and may
show up on other modules as well.
Timeout
Error ID: 0062
The timeout threshold was exceeded.
1 The analysis was completed successfully, Check the logbook for the occurrence and source
and the timeout function switched off the of a not-ready condition. Restart the analysis
module as requested. where required.
2 A not-ready condition was present during a Check the logbook for the occurrence and source
sequence or multiple-injection run for a of a not-ready condition. Restart the analysis
period longer than the timeout threshold. where required.
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals.
A LOW signal input on pin 4 of the remote connector generates the error
message.
1 Leak detected in another module with a CAN Fix the leak in the external instrument before
connection to the system. restarting the module.
2 Leak detected in an external instrument with Fix the leak in the external instrument before
a remote connection to the system. restarting the module.
4 The degasser failed to generate sufficient Check the vacuum degasser for an error
vacuum for solvent degassing. condition. Refer to the Service Manual for the
degasser or the pump that has the degasser
built-in.
Remote Timeout
Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one minute
of starting the analysis. If a not-ready condition is still present on the remote line
after one minute the error message is generated.
1 Not-ready condition in one of the Ensure the instrument showing the not-ready
instruments connected to the remote line. condition is installed correctly, and is set up
correctly for analysis.
3 Defective components in the instrument Check the instrument for defects (refer to the
showing the not-ready condition. instrument’s documentation).
1 CAN cable disconnected. • Ensure all the CAN cables are connected
correctly.
• Ensure all CAN cables are installed correctly.
3 Defective mainboard in another module. Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.
2 Leak sensor incorrectly routed, being Please contact your Agilent service representative.
pinched by a metal component.
3 Power switch assembly defective Please contact your Agilent service representative.
3 Leak sensor incorrectly routed, Please contact your Agilent service representative.
being pinched by a metal
component.
4 Power switch assembly defective Please contact your Agilent service representative.
1 Loose connection between the power switch Please contact your Agilent service
board and the mainboard representative.
2 Loose connection between the power switch Please contact your Agilent service
board and the mainboard representative.
Fan Failed
Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the mainboard to monitor the fan
speed. If the fan speed falls below a certain limit for a certain length of time, the
error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are turned
off to assure that the module does not overheat inside.
Leak
Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted
temperature-compensation sensor) are used by the leak algorithm to determine
whether a leak is present. When a leak occurs, the leak sensor is cooled by the
solvent. This changes the resistance of the leak sensor which is sensed by the
leak sensor circuit on the main board.
Open Cover
Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the foam
is removed, the fan is switched off, and the error message is generated.
1 The top foam was removed during operation. Please contact your Agilent service
representative.
2 Foam not activating the sensor. Please contact your Agilent service
representative.
Cover Violation
Error ID: 7461
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the foam
is removed while the lamps are on (or if an attempt is made to switch on for
example the lamps with the foam removed), the lamps are switched off, and the
error message is generated.
1 The top foam was removed during operation. Please contact your Agilent service
representative.
2 Foam not activating the sensor. Please contact your Agilent service
representative.
ERI Messages
Error ID: 11120 (+5 V) , 11121 (+25 V)
The ERI (Enhanced Remote Interface) provides two error events related to over
current situations on the +5 V and +24 V lines.
UV Lamp Current
Error ID: 7450
The UV lamp current is missing.
The processor continually monitors the anode current drawn by the lamp during
operation. If the anode current falls below the lower current limit, the error
message is generated.
UV Lamp Voltage
Error ID: 7451
The UV lamp anode voltage is missing.
The processor continually monitors the anode voltage across the lamp during
operation. If the anode voltage falls below the lower limit, the error message is
generated.
UV Ignition Failed
Error ID: 7452
The UV lamp failed to ignite.
The processor monitors the UV lamp current during the ignition cycle. If the lamp
current does not rise above the lower limit within 2 – 5 seconds, the error
message is generated.
1 Lamp too hot. Hot gas discharge lamps may Switch off the lamp and allow it to cool down for
not ignite as easily as cold lamps. at least 15 minutes.
UV Heater Current
Error ID: 7453
The UV lamp heater current is missing.
During UV lamp ignition, the processor monitors the heater current. If the current
does not rise above the lower limit within one second, the error message is
generated.
2 Ignition started without the top foam in Please contact your Agilent service
place. representative.
2 Defective or dirty flow cell. Ensure the flow cell is inserted correctly, and is
free from contamination (cell windows, buffers
etc.).
2 Detector is exposed to illegal ambient Verify that the ambient conditions are within the
conditions. allowed range.
1 The temperature sensor is defect. • Replace the cable to the main board.
• Please contact your Agilent service
representative.
2 Detector is exposed to illegal ambient Verify that the ambient conditions are within the
conditions. allowed range.
2 The detector has been repaired. Please contact your Agilent service
representative.
1 Random communication error. • Switch the detector off and on again at the
power switch. If the error reoccurs:
• Please contact your Agilent service
representative.
Introduction 116
Conditions of Detector 120
Failing a Test 121
Self-Test 122
Intensity Test 124
Test Failed 126
Cell Test 127
Test Failed (low ratio value) 128
Filter Test 129
Test Evaluation 130
Holmium Oxide Test 131
Test Evaluation 131
ASTM Drift and Noise Test 134
Test Failed 137
Slit Test 138
Test Failed 140
Wavelength Verification Test 141
Wavelength Calibration 143
Wavelength Recalibration Fails 145
D/A Converter (DAC) Test 146
Test Evaluation 147
Dark-Current Test 148
Test Failed 149
Spectral Scan 150
Other Lab Advisor Functions 151
EMFs - Early Maintenance Feature 151
Introduction
All tests are described based on the Agilent Lab Advisor Software B.02.08. Other
user interfaces may not provide any test or just a few.
For details on the use of the interface refer to the interface documentation.
The Lab Advisor shows the available test under Service & Diagnostics.
Table 14 Available Diagnostic Functions vs. Product Level
Product Level
Tests
Calibrations
Tools
Product Level
Controls
-Configuration
-Control
-Method Parameters
-Module Information
-Special Commands
Product Level
Statemachines
Signals
EMF Counters
Conditions of Detector
The test usually should be performed with a detector turned on for at least one
hour, so that the temperature regulation of the optical unit is working (not active
during the first 30 minutes after turn on). If the detector is on, tests can be
performed usually 10 minutes after the UV-lamp has been turned on.
Failing a Test
If a test fails with the flow cell installed, repeat the test with removed flow cell and
compare. If the test fails also, then start with proposed actions mentioned in the
details of the tests.
Self-Test
The self-test runs a series of individual tests (described on the next pages), and
evaluates the results automatically. The following tests are run:
• Filter Test
• Slit Test
• Dark Current Test
• Intensity Test
• Wavelength Verification Test
• Holmium Oxide Test
• Spectral Flatness Test
• ASTM Noise Test (without testing the Drift)
1 Run the Self Test with Agilent Lab Advisor (for further information see
Online-Help of user interface).
Under the tab Signals you can find the detailed signals from the tests.
Intensity Test
The intensity test measures the intensity of the UV-lamp over the full wavelength
range (190 - 950 nm). Four spectral ranges are used to evaluate the intensity
spectrum. The test is used to determine the performance of the lamp and optics
(see also “Cell Test” on page 127). When the test is started, the 1 nm slit is moved
into the light path automatically. To eliminate effects due to absorbing solvents,
the test should be done with water in the flow cell or with the flow cell removed.
The shape of the intensity spectrum is primarily dependent on the lamp, grating,
and diode array characteristics. Therefore, intensity spectra will differ slightly
between instruments.
1 Run the Intensity Test with Agilent Lab Advisor (for further information see
Online-Help of user interface).
Test Failed
Intensity Test Evaluation
1 Absorbing solvent or air bubble in flow cell. • Ensure the flow cell is filled with water, and
free from air bubbles.
• Repeat test without flow cell and compare
results.
3 Dirty or contaminated flow cell. Run the cell test. If the test fails, flush the flow
cell. See also “Maintenance of Standard,
Semi-Micro or Micro Flow Cell” on page 167.
6 Defect optical unit. If the test fails with flow cell and new UV-lamp,
please contact your Agilent service
representative.
If only one range fails and the application does not require this range, the lamp
NOTE may not be changed.
Cell Test
The cell test measures the intensity of the UV- and tungsten lamps over the full
wavelength range (190 - 950 nm), once with the flow cell installed, and once with
the flow cell removed. The resulting intensity ratio is a measure of the amount of
light absorbed by the flow cell. The test can be used to check for dirty or
contaminated flow cell windows. When the test is started, the 1 nm slit is moved
into the light path automatically.
This test should be performed initially with a new detector/flow cell. The values
should be kept for later reference/comparison.
1 Run the Cell-Test with Agilent Lab Advisor (for further information see
Online-Help of user interface).
Figure 31 Cell Test – Signals (example shows low intensity for flow cell)
1 Absorbing solvent or air bubble in flow cell. Ensure the flow cell is filled with water, and free
from air bubbles.
2 Dirty or contaminated flow cell. Clean the flow cell as described in “Maintenance
of Standard, Semi-Micro or Micro Flow Cell” on
page 167.
Filter Test
The filter test checks the correct operation of the filter assembly. When the test is
started, the holmium oxide filter is moved into position. During filter movement,
the absorbance signal is monitored. As the edge of the filter passes through the
light path, an absorbance maximum is seen. Once the filter is in position, the
absorbance maximum (of holmium oxide) is determined. Finally, the filter is
moved out of the light path. During movement, an additional absorbance
maximum is expected as the edge of the filter passes through the light path. The
test passes successfully, if the two maxima resulting from the edge of the filter
assembly (during filter movement) are seen, and the absorbance maximum of
holmium oxide is within the limits.
Test Evaluation
Filter Test Failed
Test Failed
1 Filter assembly (lever and filter) not installed. Install the filter assembly.
1 Holmium oxide filter not installed. Install the holmium oxide filter.
The holmium oxide test uses characteristic absorbance maxima of the built-in
holmium oxide filter to verify wavelength accuracy (see also “Wavelength
Verification Test” on page 141). When the test is started, the 1-nm slit is moved
into the light path automatically. To eliminate effects due to absorbing solvents,
the test should be done with water in the flow cell or with removed flow cell.
Test Evaluation
Holmium Oxide Test Evaluation
Limits:
The test is evaluated by the instrument, and the measured maxima are displayed
automatically. The test fails if one or more of the maxima lies outside of the
limits (see Figure 33 on page 132).
Test Failed
1 Absorbing solvent or air bubble in flow cell. Ensure the flow cell is filled with water.
3 Dirty or contaminated flow cell. Run the cell test (see “Cell Test” on page 127). If
the test fails, exchange the flow cell windows.
4 Dirty or contaminated optical components Clean optical components with alcohol and
(achromat, windows). lint-free cloth or replace the parts (see “Intensity
Test” on page 124).
The ASTM noise test determines the detector noise over a period of 20 minutes.
The test is done with installed flow cell or flow cell removed.
This test also checks for the drift. It is also part of the Self Test (without checking
for the drift).
If the test is performed with the flow cell removed, the test results are not
influenced by solvent or pump effects.
1 Run the ASTM Drift and Noise Test with Agilent Lab Advisor (for further
information see Online-Help of user interface).
Figure 35 ASTM Drift and Noise Test – Results (with Flow Cell removed)
Figure 36 ASTM Drift and Noise Test – Signal (with Flow Cell removed)
Test Failed
ASTM Noise Test Evaluation
1 Insufficient lamp warm-up time. Allow detector and UV-lamp turned on for at least
2 hours.
2 Absorbing solvent or air bubble in flow cell. Ensure the flow cell is filled with water, and free
from air bubbles.
Slit Test
1 Run the Slit Test with the Agilent Lab Advisor (for further information see
Online-Help of user interface).
Test Failed
Slit Test Evaluation
1 Air bubble in flow cell. Flush the flow cell or remove the flow cell.
2 Old lamp. Run the “Intensity Test”. Exchange the lamp if old
or defective.
The detector uses the alpha (656.1 nm) and beta (486 nm) emission lines of the
UV-lamp for wavelength calibration. The sharp emission lines enable accurate
calibration. When verification is started, the 1-nm slit is moved into the light path
automatically. The test is run with Flow Cell removed or with Flow Cell installed.
If the test is performed with the Flow Cell removed, the test results are not
influenced by solvent or pump effects.
When The detector is calibrated at the factory, and under normal operating conditions should not require
recalibration. However, it is advisable to recalibrate:
• after repair of components in the optical unit,
• after exchange of the optical unit or main board,
• after replacing the Flow Cell or UV-lamp,
• after significant environmental condition changes (temperature, humidity),
• at a regular interval, at least once per year (for example, prior to an Operational
Qualification/Performance Verification procedure), and
• when chromatographic results indicate the detector may require recalibration.
1 Run the Wavelength Verification Test with the Agilent Lab Advisor (for further
information see Online-Help of user interface).
Wavelength Calibration
The detector uses the alpha (656.1 nm) and beta (486 nm) emission lines of the
deuterium lamp for wavelength calibration. The sharp emission lines enable
more accurate calibration than is possible with holmium oxide. When
recalibration is started, the 1 nm slit is moved into the light path automatically.
The gain is set to zero.
On completion of the scan, the alpha- and beta-line deviations (in nm) are
displayed. These values indicate how far the detector calibration deviates from
the actual positions of the alpha and beta emission lines. After calibration, the
deviation is zero.
To eliminate effects due to absorbing solvents, remove the flow cell before
starting the test.
When The detector is calibrated at the factory, and under normal operating conditions should not require
recalibration. However, it is advisable to recalibrate:
• after maintenance (flow cell or UV-lamp),
• after repair of components in the optical unit,
• after exchange of the optical unit or main board,
• after significant environmental condition changes (temperature, humidity),
• at a regular interval, at least once per year (for example, prior to an Operational
Qualification/Performance Verification procedure), and
• when chromatographic results indicate the detector may require recalibration.
If the detector is operated in a lab environment that differs at average from the
NOTE final test environment (25 °C) then the detector should be recalibrated for this
temperature.
If the detector was repaired (opened covers), the wavelength calibration can be
NOTE done 10 minutes after lamp on. A final wavelength calibration should be
repeated after complete warm-up of the detector.
1 Run the Wavelength Calibration with the Agilent Lab Advisor (for further
information see Online-Help of user interface).
1 Absorbing solvent or air bubble in Flow Cell. Repeat calibration with Flow Cell and compare
results.
2 Dirty or contaminated Flow Cell. • Ensure the Flow Cell is filled with water.
• Recalibrate.
4 Dirty or contaminated optical components. Run the Cell Test. If the test fails, flush the flow
cell.
If the test fails with Flow Cell and new UV-lamp, the optical unit must be
NOTE checked/replaced.
The detector provides analog output of chromatographic signals for use with
integrators, chart recorders or data systems. The analog signal is converted from
the digital format by the digital-analog-converter (DAC).
The DAC test is used to verify correct operation of the digital-analog-converter by
applying a digital test signal to the DAC.
The DAC outputs an analog signal of approximately 50 mV (if the zero offset of
the analog output is set to the default value of 5 %) which can be plotted on an
integrator. A continuous square wave with an amplitude of 10 µV and a
frequency of approximately 1 cycle/24 seconds is applied to the signal.
The amplitude of the square wave and the peak-to-peak noise are used to
evaluate the DAC test.
Preparations Lamp must be on for at least 10 minutes. Connect integrator, chart recorder or data system to the
detector analog output.
1 Run the D/A Converter (DAC) Test with the Agilent Lab Advisor (for further
information see Online-Help of user interface).
Test Evaluation
The noise on the step should be less than 3 µV.
Dark-Current Test
The dark-current test measures the leakage current from each diode. The test is
used to check for leaking diodes which may cause non-linearity at specific
wavelengths. During the test, the slit assembly moves to the dark position,
cutting off all light falling onto the diode array. Next, the leakage current from
each diode is measured, and displayed graphically. The leakage current
(represented in counts) for each diode should fall within the limits.
1 Run the Dark Current Test with the recommended user interface (for further
information see Online-Help of user interface).
Test Failed
1 Defective slit assembly (stray light). Run the “Slit Test” on page 138 (part of the
“Self-Test” on page 122).
Spectral Scan
The Spectral Scan tool is available for diode-array and variable wavelength
detectors (DAD/MWD and VWD). It allows you to scan a spectrum over a
specified wavelength range and export the data to a csv (comma-separated
values) file that can be used in other applications (for example, Microsoft Excel).
Introduction to Maintenance
The module is designed for easy maintenance. Maintenance can be done from
the front with module in place in the system.
Module is partially energized when switched off, as long as the power cord
WAR N IN G
is plugged in.
Risk of stroke and other personal injury. Repair work at the module can
lead to personal injuries, e. g. shock hazard, when the module cover is
opened and the instrument is connected to power.
Never perform any adjustment, maintenance or repair of the module
with the top cover removed and with the power cord plugged in.
The security lever at the power input socket prevents that the module
cover is taken off when line power is still connected. Never plug the
power line back in when cover is removed.
Overview of Maintenance
The following pages describe maintenance (simple repairs) of the detector that
can be carried out without opening the main cover.
Table 15 Overview of Maintenance
Deuterium lamp or If noise and/or drift exceeds your application limits or An intensity test should be performed
tungsten lamp lamp does not ignite. after replacement.
exchange
Flow cell exchange If application requires a different flow cell type. A holmium or wavelength calibration
test should be performed after
replacement.
Flow cell parts Cleaning If leaking or if intensity drops due to contaminated flow A pressure tightness test should be done
or exchange cell windows. after repair.
The module case should be kept clean. Cleaning should be done with a soft cloth
slightly dampened with water or a solution of water and mild detergent. Do not
use an excessively damp cloth allowing liquid to drip into the module.
Liquid dripping into the electronic compartment of your module can cause
WAR N IN G
shock hazard and damage the module
Do not use an excessively damp cloth during cleaning.
Drain all solvent lines before opening any connections in the flow path.
The figures shown in this procedure exemplarily show the Infinity II Multisampler
NOTE module.
The principle of how to remove and/or install doors works in the same way for all
Infinity II modules.
1 Press the release buttons and pull the front door out. 2 For the Installation of the front door. Insert the hinges
into their guides and move the door in until the release
buttons click into their final position.
Replace a Lamp
When If noise or drift exceeds application limits or lamp does not ignite
Eye damage may result from directly viewing the light produced by the
deuterium lamp used in this product.
Always turn the deuterium lamp off before removing it.
1 Open the doors. 2 Grab the lamp cover and pull it off (it is fixed by two
magnets in the center of the cover).
3 Unplug the lamp connector as required 4 Unscrew (do not remove) the two lamp screws
. (Pozidriv) as required.
5 Remove the deuterium lamp and place it on a clean 6 Remove the Vis-lamp and place it on a clean place.
place.
NOTE NOTE
Do not touch the glass bulb with your fingers. It may reduce Do not touch the glass bulb with your fingers. It may reduce
the light output. the light output.
7 Insert the deuterium lamp (RFID tag on the top side). 8 Insert the Vis-lamp (flat side to the right).
9 Reconnect the lamp connector as required. 10 Place the lamp cables in the lamp cover.
11 Slide the lamp cover into the top position of the metal 12 Close the doors.
front and press the lamp cover completely in until it
clicks.
1.
2.
Next Steps:
13 Perform a Wavelength Verification Test or a Holmium Oxide Test to check the correct positioning of the lamp
(“Wavelength Verification Test” on page 141, “Holmium Oxide Test” on page 131).
14 Perform an Intensity Test.
When If an application needs a different type of flow cell or the flow cell needs repair.
1 Open the doors. 2 Grab the lamp cover and pull it off (it is fixed by two
magnets in the center of the cover).
3 Open the flow cell door. 4 Disconnect the flow cell inlet capillary (top) from the
union.
5 Disconnect the waste tubing (bottom) from the union. 6 Loosen the thumb screw (1.) and remove the flow cell
outlet capillary (bottom) with the union (2.).
2. 1.
7 Remove the flow cell while pressing the flow cell holder. 8 Insert the flow cell while pressing the flow cell holder.
NOTE
The label attached to the flow cell provides information on
part number, path length, and maximum pressure.
1.
2. 3.
11 Reconnect the waste tubing (bottom) to the union. 12 Close the flow cell door.
Establish a flow and check for leaks.
NOTE
To check for leaks, establish a flow and observe the flow
cell (outside of the cell compartment) and all capillary
connections.
13 Slide the lamp cover into the top position of the metal 14 Close the doors.
front and press the lamp cover completely in until it
clicks.
1.
2.
15 Perform a Wavelength Verification and Calibration or a Holmium Oxide Test to check the correct positioning of the flow
cell.
When If the flow cell needs repair due to leaks or contaminations (reduced light throughput)
The gaskets used in the standard and semi-micro/micro flow cell are different.
NOTE
1 Use a 4 mm hex key to unscrew the window assembly 2 Use a tooth pick to remove the quartz window from the
(1) and remove the gasket (6,7) from the cell body. window assembly.
1 NOTE
7
If the washers fall out of the window assembly, they must
be inserted in the correct order with the PTFE ring to prevent
any leaks from the flow cell window.
6
1
NOTE
Carefully take one of the gaskets (#6 back or # 7 front) and
insert it into the cell body.
8
7
65
43 2
11
9 1
10
NOTE
Gaskets # 6 and #7 have different hole diameters.
5 Correct orientation of spring washers (2) is required. 6 Press the PTFE ring into the window assembly.
1 - Window screw
2 - Spring washers
3 - Compression washer
4 - Window holder
5 - Quartz window
6 - Gasket
1 2 3 4 5 6
7 Press the window assembly onto the new or cleaned 8 Insert the window assembly (1) into the cell body.
quartz window.
1
7
6
1
NOTE
Do not mix the gasket #6 and #7 (different hole diameter).
Next Steps:
9 Using a 4-mm hex key, tighten the window screw hand tight plus a quarter turn.
10 Reconnect the capillaries, see “Remove and Install a Flow Cell” on page 163.
11 Perform a leak test.
12 Insert the flow cell.
13 Replace the front cover
14 Perform a Wavelength Verification Test or a Holmium Oxide Test to check the correct positioning of the lamp
(“Wavelength Verification Test” on page 141, “Holmium Oxide Test” on page 131).
When If the flow cell needs repair due to leaks or contaminations (reduced light throughput)
All descriptions in this procedure are based on the default orientation of the cell
NOTE (as it is manufactured). The heat exchanger/capillary and the cell body can be
fixed mirror symmetrically to have both capillaries routed to the bottom or to the
top (depending on the routing of the capillaries to the column).
1 Use a 4 mm hex key to unscrew the window assembly (1) and remove the
gasket (2) from the cell body.
1
2
2
1
If you want to replace the gasket only, reinsert the window assembly into the
cell body.
2 Use a tooth pick to remove the quartz window from the window assembly.
If the washers fall out of the window assembly, they must be inserted in the
NOTE correct order with the PTFE ring to prevent any leaks from the flow cell window.
6
5
4
2
3 2
9
8
2 Seal ring
3 Quartz window
4 Compression washer
5 Spring washers
6 Window screw
7 Inlet capillary
8 Outlet capillary
All descriptions in this procedure are based on the default orientation of the cell
NOTE (as it is manufactured). The heat exchanger/capillary and the cell body can be
fixed mirror symmetrically to have both capillaries routed to the bottom or to the
top (depending on the routing of the capillaries to the column).
The fittings at the flow cell body are special types for low dead volumes and not
NOTE compatible with other fittings.
When retightening the fittings, make sure that they are carefully tightened
(handtight plus 1/4 turn with a wrench). Otherwise damage of the flow cell body
or blockage may result.
1 Identify the inlet and outlet capillaries. To replace the 2 Remove the outlet capillary.
inlet capillary, continue with step "To replace the inlet
capillary, use a 4 mm wrench for the fitting."
Inlet
Outlet
3 To replace the inlet capillary, use a 4 mm wrench to 4 Unscrew the cell body from the heat exchanger.
open the fitting.
5 Remove the heat exchanger from the clamp unit. 6 Use a small flat screw driver to carefully lift off the I.D.
tag. Shown is the default orientation. See Note at the
beginning of this section.
7 Unscrew the fixing screw and unwrap the inlet capillary 8 Take the new inlet capillary and bend it 90° about 35
from the grove in the flow cell body. mm from its end.
35 cm
90°
9 Bend the capillary again by 90° as shown below. 10 Insert the capillary into the hole between fixing screw
and the inlet fitting.
90°
11 The capillary lays in the grove and should be tied around 12 Insert the fixing screw, so that the capillary cannot leave
the body (in the grove) 5 times. the grove.
13 Carefully insert the I.D. tag into the new heat exchanger. 14 Fix the heat exchanger to the clamp unit.
Shown is the default orientation. See Note at the
beginning of this section.
15 Fix the flow cell body to the heat exchanger. 16 Fix the inlet capillary to the flow cell body handtight first.
Then do a 1/4 turn with a 4 mm wrench.
17 After replacing the outlet capillary, fix it handtight first. 18 Check for a centered holder vs. hole. If required adjust
Then do a 1/4 turn with a 4 mm wrench. with the holder screws.
Next Steps:
19 Reconnect the capillaries, see “Remove and Install a Flow Cell” on page 163.
20 Perform a leak test.
21 Insert the flow cell.
22 Close the doors of the module.
23 Perform a Wavelength Verification Test or a Holmium Oxide Test to check the correct positioning of the lamp
(“Wavelength Verification Test” on page 141, “Holmium Oxide Test” on page 131).
All descriptions in this procedure are based on the default orientation of the cell
NOTE (as it is manufactured). The heat exchanger/capillary and the cell body can be
fixed mirror symmetrically to have both capillaries routed to the bottom or to the
top (depending on the routing of the capillaries to the column).
The fittings at the flow cell body are special types for low dead volumes and not
NOTE compatible with other fittings. When retightening the fittings, make sure that they
are carefully tightened (handtight plus 1/4 turn with a wrench). Otherwise
damage of the flow cell body or blockage may result.
1 Identify the inlet and outlet capillaries. 2 Remove the outlet capillary.
Inlet
Outlet
3 To replace the inlet capillary, use a 4 mm wrench to 4 Unscrew the cell body from the heat exchanger.
open the fitting.
5 Remove the heat exchanger from the clamp unit. 6 Use a small flat screw driver to carefully lift off the I.D.
tag. Shown is the default orientation. See Note at the
beginning of this section.
7 Carefully insert the I.D. tag into the new heat exchanger. 8 Fix the heat exchanger to the clamp unit.
Shown is the default orientation. See Note at the
beginning of this section.
9 Fix the flow cell body to the heat exchanger. 10 Fix the inlet capillary to the flow cell body handtight first.
Then do a 1/4 turn with a 4 mm wrench.
11 After replacing the outlet capillary, fix it handtight first. 12 Check for a centered holder vs. hole. If required adjust
Then do a 1/4 turn with a 4 mm wrench. with the holder screws.
Next Steps:
13 Reconnect the capillaries, see “Remove and Install a Flow Cell” on page 163.
14 Perform a leak test.
15 Insert the flow cell.
16 Close the doors.
17 Perform a Wavelength Verification Test or a Holmium Oxide Test to check the correct positioning of the lamp
(“Wavelength Verification Test” on page 141, “Holmium Oxide Test” on page 131).
For details refer to the technical note that comes with the nano-flow cell kit.
NOTE
The quartz block can be cleaned with alcohol. DO NOT touch the inlet and outlet
NOTE windows at the quartz block.
1 Disconnect the capillaries from the capillary holder and 2 Unscrew the cell body from the holder.
remove the flow cell, see “Remove and Install a Flow
Cell” on page 163.
3 Unscrew the capillaries from the flow cell. DO NOT use 4 Using for example a toothpick, press on the plastic part
the adapter at this time! and slide the quartz body out of the cell housing.
5 The quartz body and the cell seal assembly can be 6 This figure shows the correct holding of the quartz body
separated for cleaning purpose. and the cell seal assembly.
7 Replace the cell seal assembly onto the quartz body. 8 Slide the quartz body completely into the cell body to
Always use a new seal assembly to exclude damage the front stop (use for example a toothpick).
during disassembling.
9 Insert the flow cell capillaries and tighten them 10 Reassemble the flow cell body to the holder.
fingertight. Use the wrench and torque adapter as
described on Figure 46 on page 184 and tighten the
fittings alternately.
Next Steps:
11 Install the flow cell, see “Remove and Install a Flow Cell” on page 163.
12 Perform a leak test with the flow cell outside of the detector.
13 If no leak is observed, install the flow cell and you are ready to work.
14 Make sure that the flow cell assembly is inserted correctly and fits perfectly in the optical unit (especially when PEEK
capillaries are used).
With the instrument accessory kit comes a 4-mm wrench and with the Sealing
NOTE Kit a special adapter. Both together work as a torque wrench with pre-defined
torque (maximum allowed torque for the cell fittings is 0.7 Nm). It can be used to
tight the capillary fittings at the flow cell body. The wrench has to be plugged into
the adapter as shown in Figure 46 on page 184.
Adapter
max
0.7 Nm
1 Unscrew the six screws and remove the flow cell cover. 2 If not already in this position, move the filter up.
3 Release the holder with a screw driver (at the top). 4 Remove the holmium oxide filter.
NOTE NOTE
Do not scratch the holmium oxide filter. The holmium oxide filter can be cleaned with alcohol and a
lint-free cloth.
5 Insert the holmium oxide filter. 6 Replace the flow cell cover and fix the six screws.
7 Perform a holmium oxide test (see “Holmium Oxide 9 Close the doors.
Test” on page 131) to check the proper function of the
holmium oxide filter.
8 Install the flow cell, see “Remove and Install a Flow
Cell” on page 163.
Correcting Leaks
When If a leakage has occurred in the flow cell area or at the heat exchanger or at the capillary connections
Use tissue to dry the leak sensor area and the leak pan.
NOTE
1 Open the doors. 2 Grab the lamp cover and pull it off (it is fixed by two
magnets in the center of the cover).
3 Open the flow cell door. 4 Observe the capillary connections and the flow cell area
for leaks and correct, if required.
NOTE
To check for leaks, establish a flow and observe the flow
cell (outside of the cell compartment) and all capillary
connections.
5 Check the Leak Sensor area for leaks and correct, if 6 Install the flow cell and close the flow cell door.
required.
7 Slide the lamp cover into the top position of the metal 8 Close the doors.
front and press the lamp cover completely in until it
clicks.
1.
2.
9 Perform a Wavelength Verification and Calibration or a Holmium Oxide Test to check the correct positioning of the flow
cell.
1 Open the doors. 2 Press the Leak Adapter down (1.) and remove it
together with the tubing (2.).
1.
2.
3 Install the Leak Adapter by pressing it into the Main 4 Insert the Tubing (approximately 85 mm required for
Cover. replacement) between Leak Adapter outlet and Leak
Panel.
Preparations Read update documentation provided with the Firmware Update Tool.
Compatibility with When using the G7115A or G7165A in a system, all other modules must
1260/1290 Infinity have firmware revision from firmware set 7.00 or above (main and
modules resident).
Compatibility with When using the G7115A or G7165A in a system, all other modules must
1100/1200 Series have firmware revision from firmware set 7.00 or above (main and
modules resident). Otherwise the communication will not work.
• accumulated UV on time (in hours) • nominal path length of the cell (in mm)
The pressure value is always displayed in bar, even if the user interface uses
NOTE other units, e.g. PSI.
4 2140-0820 Longlife Deuterium lamp “C“ (with black cover and RFID tag)
G1315-60022 Standard flow cell, 10 mm, 13 µL, 120 bar (12 MPa)
G5615-60022 Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar (12 MPa) for
MWD/DAD, includes 0890-1763 – 0.18 x 1500 mm PEEK capillary and
5063-6591 – PEEK fittings
G1315-60025 Semi-micro flow cell, 6 mm, 5 µL, 120 bar (12 MPa)
G1315-60024 Micro flow cell, 3 mm, 2 µL, 120 bar (12 MPa)
G1315-60015 High pressure flow cell, 6 mm, 1.7 µL, 400 bar (40 MPa)
G1315-60016 Prep flow cell SST - 3 mm, 120 bar (12 MPa)
3 6
1
8
7
4
2 9
11
10
Figure 47 Overview of maintenance parts
8
7
6
5
4
3
11 2
1
9
10
1 - Window screw
2 - Spring washers
3 - Compression washer
4 - Window holder
5 - Quartz window
6 - Gasket
1 2 3 4 5 6
8
7
6
5
4
3
11 2
1
9
10
1 - Window screw
2 - Spring washers
3 - Compression washer
4 - Window holder
5 - Sapphire window
6 - Gasket
1 2 3 4 5 6
G1315-68713 Cell repair kit semi-micro, includes window screw kit, Gasket Kit BACK,
Gasket Kit FRONT and 4 mm hexagonal wrench
8
7
6
5
4
3
11 2
1
9
10
1 - Window screw
2 - Spring washers
3 - Compression washer
4 - Window holder
5 - Quartz window
6 - Gasket
1 2 3 4 5 6
G1315-68713 Cell repair kit semi-micro, includes window screw kit, Gasket Kit BACK,
Gasket Kit FRONT and 4 mm hexagonal wrench
8
7
6
5
4
3
11 2
1
9
10
1 - Window screw
2 - Spring washers
3 - Compression washer
4 - Window holder
5 - Quartz window
6 - Gasket
1 2 3 4 5 6
79883-68700 High pressure cell repair kit (includes 1 quartz window, 1 compression
washer, 5 spring washers, 2 seal rings)
6
5
4
2
3 2
9
8
For more details on the Preparative Flow Cells refer to the technical note that
NOTE comes with the flow cells.
1
3
5
2 1
6
4 - inlet
4 - outlet
4a
1 - Window screw
2 - Spring washers
3 - Compression washer
4 - Window holder
5 - Quartz window
6 - Gasket
1 2 3 4 5 6
For more details on the Preparative Flow Cells refer to the technical note that
NOTE comes with the flow cells.
The flow cell comes with two tubings 0.8 mm i.d. and one 0.5 mm i.d. so that the
NOTE combination at the flow cell could be either 0.8/0.8 or 0.5/0.8 (inlet/outlet).
Standard is 0.8/0.8. Depending on the system pressure (<30 mL/min) or
bandbroadening, the inlet tubing might be changed to 0.5 mm.
2
3
Semi-nano flow cell kit, 10 mm, 500 nL, 5 completely assembled (includes items 1, 2, 3, 4, 10,
MPa (G1315-68724) 11, 12, 13, 14, 15, and 16)
Nano flow cell kit, 6 mm, 80 nL, 5 MPa ( completely assembled (includes items 1, 2, 3, 4, 10,
G1315-68716) 11, 12, 13, 14, 15, and 16)
Figure 60 on page 213 shows all parts delivered with the nano-flow cell kits.
Generic parts for both nano-flow cells:
2
11
12 3, 4 3, 4 1
13
5
10 7
16
15
8 9
14
17
1 G1315-87333 PEEK coated fused silica capillary Inlet (100 µm) pre-mounted to cell,
includes Inlet capillary, 300 mm long, 100 µm i.d. with pre-fixed
ferrules (#4) and fittings (#3), plus one PEEK Fitting FT (#5)
2 G1315-87338 PEEK coated fused silica capillary Outlet (100 µm) pre-mounted to
cell, includes Outlet capillary, 120 mm long, 100 µm i.d. with pre-fixed
ferrules (#4) and fitting (#3), plus one PEEK Fitting FT (#5)
1 G1315-87323 PEEK coated fused silica capillary Inlet (50 µm) alternative, includes
Inlet capillary, 400 mm long, 50 µm i.d. with pre-fixed ferrules (#4) and
fittings (#3), plus one PEEK Fitting FT (#5)
2 G1315-87328 PEEK coated fused silica capillary Outlet (50 µm), alternative, includes
Outlet capillary, 120 mm long, 50 µm i.d. with pre-fixed ferrules (#4)
and fitting (#3), plus one PEEK Fitting FT (#5)
1 G1315-87323 PEEK coated fused silica capillary Inlet (50 µm) alternative, includes
Inlet capillary, 400 mm long, 50 µm i.d. with pre-fixed ferrules (#4) and
fittings (#3), plus one PEEK Fitting FT (#5)
2 G1315-87328 PEEK coated fused silica capillary Outlet (50 µm), alternative, includes
Outlet capillary, 120 mm long, 50 µm i.d. with pre-fixed ferrules (#4)
and fitting (#3), plus one PEEK Fitting FT (#5)
1 G1315-87313 PEEK coated fused silica capillary Inlet (25 µm) alternative, includes
Inlet capillary, 200 mm long, 25 µm i.d. with pre-fixed ferrules (#4) and
fittings (#3), plus one PEEK Fitting FT (#5)
2 G1315-87318 PEEK coated fused silica capillary Outlet (25 µm) alternative, includes
Outlet capillary, 600 mm long, 25 µm i.d. with pre-fixed ferrules (#4)
and fitting (#3), plus one PEEK Fitting FT (#5)
Accessory Kits
p/n Description
5062-8535 Waste accessory kit
5500-1191 InfinityLab Quick Turn Capillary ST 0.12 mm x 280 mm, long socket
2 1460-1510 Spring
When the filter motor has been removed, the filter lever should not be reused.
NOTE Use always a new filter lever to assure correct fit on the filter motor shaft.
3
4
This chapter provides information on cables used with the Agilent 1200 Infinity
Series modules.
Cable Overview
Never use cables other than the ones supplied by Agilent Technologies to ensure
NOTE proper functionality and compliance with safety or EMC regulations.
Analog cables
p/n Description
35900-60750 Agilent module to 3394/6 integrators
Remote cables
p/n Description
03394-60600 Agilent module to 3396A Series I integrators
BCD cables
p/n Description
03396-60560 Agilent module to 3396 integrators
CAN cables
p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m
LAN cables
p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)
RS-232 cables
p/n Description
RS232-61601 RS-232 cable, 2.5 m
Instrument to PC, 9-to-9 pin (female). This cable has special pin-out,
and is not compatible with connecting printers and plotters. It is also
called "Null Modem Cable" with full handshaking where the wiring is
made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
Analog Cables
1 Not connected
2 Shield Analog -
3 Center Analog +
1 Not connected
2 Black Analog -
3 Red Analog +
Remote Cables
p/n 5061-3378 Pin 35900 A/D Pin Agilent Signal Name Active
module (TTL)
BCD Cables
One end of these cables provides a 15-pin BCD connector to be connected to the
Agilent modules. The other end depends on the instrument to be connected to
p/n G1351-81600 Wire Color Pin Agilent Signal Name BCD Digit
module
Green 1 BCD 5 20
Violet 2 BCD 7 80
Blue 3 BCD 6 40
Yellow 4 BCD 4 10
Black 5 BCD 0 1
Orange 6 BCD 3 8
Red 7 BCD 2 4
Brown 8 BCD 1 2
not connected 14
p/n 03396-60560 Pin 3396 Pin Agilent Signal Name BCD Digit
module
1 1 BCD 5 20
2 2 BCD 7 80
3 3 BCD 6 40
4 4 BCD 4 10
5 5 BCD0 1
6 6 BCD 3 8
7 7 BCD 2 4
8 8 BCD 1 2
9 9 Digital ground
NC 15 +5V Low
CAN/LAN Cables
p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m
LAN Cables
p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)
p/n Description
RS232-61601 RS-232 cable, 2.5 m
Instrument to PC, 9-to-9 pin (female). This cable has special pin-out,
and is not compatible with connecting printers and plotters. It is also
called "Null Modem Cable" with full handshaking where the wiring is
made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
This chapter describes the detector in more detail on hardware and electronics.
Firmware Description
Resident System
This resident section of the firmware is identical for all Agilent
1100/1200/1220/1260/1290 series modules. Its properties are:
• the complete communication capabilities (CAN, LAN, USB and RS- 232)
• memory management
• ability to update the firmware of the 'main system'
Main System
Its properties are:
• the complete communication capabilities (CAN, LAN, USB and RS- 232)
• memory management
• ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are divided
into common functions like
• run synchronization through APG/ERI remote,
• error handling,
• diagnostic functions,
• or module specific functions like
• internal events such as lamp control, filter movements,
• raw data collection and conversion to absorbance.
Firmware Updates
Firmware updates can be done with the Agilent Lab Advisor software with files
on the hard disk (latest version should be used).
Required tools, firmware and documentation are available from the Agilent web:
http://www.agilent.com/en-us/firmwareDownload?whid=69761
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
• PPPP is the product number, for example, 1315B for the G1315B DAD,
• R the firmware revision, for example, A for G1315B or B for the G1315C DAD,
• VVV is the revision number, for example 650 is revision 6.50,
• XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware in
chapter "Maintenance" or use the documentation provided with the Firmware
Update Tools.
Update of main system can be done in the resident system only. Update of the
NOTE resident system can be done in the main system only.
Main and resident firmware must be from the same set.
Main FW update
Resident System Main System
Resident FW Update
The firmware update tools, firmware and documentation are available from the
Agilent web.
• http://www.agilent.com/en-us/firmwareDownload?whid=69761
Electrical Connections
• The CAN bus is a serial bus with high-speed data transfer. The two
connectors for the CAN bus are used for internal module data transfer and
synchronization.
• One analog output provides signals for integrators or data handling systems.
• The ERI/REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
start, stop, common shutdown, prepare, and so on.
• With the appropriate software, the LAN connector may be used to control the
module from a computer through a LAN connection. This connector is
activated and can be configured with the configuration switch.
• With the appropriate software, the USB connector may be used to control the
module from a computer through a USB connection.
• The power input socket accepts a line voltage of 100 – 240 VAC ± 10 % with a
line frequency of 50 or 60 Hz. Maximum power consumption varies by
module. There is no voltage selector on your module because the power
supply has wide-ranging capability. There are no externally accessible fuses
because automatic electronic fuses are implemented in the power supply.
Never use cables other than the ones supplied by Agilent Technologies to ensure
NOTE proper functionality and compliance with safety or EMC regulations.
CCXZZ00000 Format
CC Country of manufacturing
• DE = Germany
• JP = Japan
• CN = China
Interfaces
Pumps
Samplers
Detectors
Fraction Collectors
Others
G7122A No No No Yes No A
Interfaces Overview
CAN
The CAN is inter-module communication interface. It is a 2-wire serial bus
system supporting high speed data communication and real-time requirement.
LAN
The modules have either an interface slot for a LAN card (e.g. Agilent G1369B/C
LAN Interface) or they have an on-board LAN interface (e.g. detectors G1315C/D
DAD and G1365C/D MWD). This interface allows the control of the
module/system via a PC with the appropriate control software. Some modules
have neither on-board LAN nor an interface slot for a LAN card (e.g. G1170A
Valve Drive or G4227A Flexible Cube). These are hosted modules and require a
Host module with firmware B.06.40 or later or with additional G1369C LAN Card.
USB
The USB interface replaces the RS-232 Serial interface in new FUSION generation
modules. For details on USB refer to “USB (Universal Serial Bus)” on page 242.
Remote (ERI)
The ERI (Enhanced Remote Interface) connector may be used in combination
with other analytical instruments from Agilent Technologies if you want to use
features as common shut down, prepare, and so on.
It allows easy connection between single instruments or systems to ensure
coordinated analysis with simple coupling requirements.
The subminiature D connector is used. The module provides one remote
connector which is inputs/outputs (wired- or technique).
All common TTL circuits operate with a 5 V power supply. A TTL signal is defined
NOTE as "low" or L when between 0 V and 0.8 V and "high" or H when between 2.0 V
and 5.0 V (with respect to the ground terminal).
1 START REQUEST (L) Request to start injection cycle (for example, by start key on any
module). Receiver is the autosampler.
2 STOP (L) Request to reach system ready state as soon as possible (for
example, stop run, abort or finish and stop injection). Receiver is any
module performing run-time controlled activities.
3 READY (H) System is ready for next analysis. Receiver is any sequence
controller.
4 POWER ON (H) All modules connected to system are switched on. Receiver is
any module relying on operation of others.
5 Not used
6 SHUT DOWN (L) System has serious problem (for example, leak: stops pump).
Receiver is any module capable to reduce safety risk.
Special Interfaces
There is no special interface for this module.
ERI Description
The ERI interface contains eight individual programmable input/output pins. In
addition, it provides 24 V power and 5 V power and a serial data line to detect and
recognize further add-ons that could be connected to this interface. This way the
interface can support various additional devices like sensors, triggers (in and out)
and small controllers, etc.
ERI
1 IO 1 (START
REQUEST)
2 IO 2 (STOP)
3 IO 3 (READY)
4 IO 4 (POWER ON)
5 IO 5 (NOT USED)
6 IO 6 (SHUT DOWN)
7 IO 7 (START)
8 IO 8 (PREPARE)
9 1 wire DATA
10 DGND
11 +5 V ERI out
12 PGND
13 PGND
IO (Input/Output) Lines
• Eight generic bi-directional channels (input or output).
• Same as the APG Remote.
• Devices like valves, relays, ADCs, DACs, controllers can be
supported/controlled.
The 6-bit configuration switch is located at the rear of the module with FUSION
electronics. Switch settings provide configuration parameters for LAN and
instrument specific initialization procedures.
All modules with FUSION electronics:
• Default is ALL switches DOWN (best settings).
• Default IP address for LAN 192.168.254.11
• For specific LAN modes switches 4-5 must be set as required.
• For boot resident/cold start modes switches 1+2 or 6 must be UP.
Configuration switch
Mode Function/Setting
1 When selecting mode COM, settings are stored to non-volatile memory. When selecting mode TEST, COM settings are taken from
non-volatile memory.
Special Settings
Boot-Resident/Main
Firmware update procedures may require this mode in case of firmware loading
errors (main/resident firmware part).
If you use the following switch settings and power the instrument up again, the
instrument firmware stays in the resident/main mode. In resident mode, it is not
operable as a module. It only uses basic functions of the operating system for
example, for communication. In this mode the main firmware can be loaded
(using update utilities).
Loss of data
C AU T I O N
Forced cold start erases all methods and data stored in the non-volatile
memory. Exceptions are calibration settings, diagnosis and repair log books
which will not be erased.
Save your methods and data before executing a forced cold start.
If you use the following switch settings and power the instrument up again, it will
start as described above.
Instrument Layout
EMF Counters
EMF counters increment with use and can be assigned a maximum limit which
provides visual feedback in the user interface when the limit is exceeded. Some
counters can be reset to zero after the required maintenance procedure.
lamp without RFID tag YES via Lab Advisor or Instant Pilot
This function is only available via Agilent Lab Advisor or Instant Pilot.
NOTE
This chapter is generic and may show figures that differ from your module. The
NOTE functionality is the same.
1 Note the MAC (Media Access Control) address for further reference. The MAC
or hardware address of the LAN interfaces is a world wide unique identifier.
No other network device will have the same hardware address. The MAC
address can be found on a label at the rear of the module underneath the
configuration switch (see Figure 68 on page 251).
Figure 67 MAC-Label
LAN interface
MAC label
Configuration Switches
Configuration switch
The module is shipped with all switches set to OFF, as shown above.
To perform any LAN configuration, SW1 and SW2 must be set to OFF.
NOTE
0 0 0 1 0 0 Use DHCP
Using Stored
When initialization mode Using Stored is selected, the parameters are taken from
the non-volatile memory of the module. The TCP/IP connection will be
established using these parameters. The parameters were configured previously
by one of the described methods.
Non-Volatile
RAM
Active
Parameter
Stored
Parameter
Using Default
When Using Default is selected, the factory default parameters are taken instead.
These parameters enable a TCP/IP connection to the LAN interface without
further configuration, see Table 24 on page 255.
Default Active
Parameter Parameter
Using the default address in your local area network may result in network
NOTE problems. Take care and change it to a valid address immediately.
IP address: 192.168.254.11
Since the default IP address is a so-called local address, it will not be routed by
any network device. Thus, the PC and the module must reside in the same
subnet.
The user may open a Telnet session using the default IP address and change the
parameters stored in the non-volatile memory of the module. He may then close
the session, select the initialization mode Using Stored, power-on again and
establish the TCP/IP connection using the new parameters.
When the module is wired to the PC directly (e.g. using a cross-over cable or a
local hub), separated from the local area network, the user may simply keep the
default parameters to establish the TCP/IP connection.
In the Using Default mode, the parameters stored in the memory of the module
NOTE are not cleared automatically. If not changed by the user, they are still available,
when switching back to the mode Using Stored.
DHCP Active
Server Parameter
1 It may take some time until the DHCP server has updated the DNS server with
NOTE the hostname information.
2 It may be necessary to fully qualify the hostname with the DNS suffix, e.g.
0030d3177321.country.company.com.
3 The DHCP server may reject the hostname proposed by the card and assign a
name following local naming conventions.
Setup (DHCP)
The DHCP functionality is available on all Agilent HPLC modules with on-board LAN Interface or LAN
Interface Card G1369C, and “B”-firmware (B.06.40 or above) or modules with "D"-firmware. All
modules should use latest firmware from the same set.
1 Note the MAC address of the LAN interface (provided with G1369C LAN
Interface Card or mainboard). This MAC address is on a label on the card or at
the rear of the mainboard, for example, 0030d3177321.
On the Local Controller the MAC address can be found under Details in the
LAN section.
2 Set the configuration switch to DHCP either on the G1369C LAN Interface
Card or the mainboard of above mentioned modules.
Table 25 G1369C LAN Interface Card (configuration switch on the card)
SW 4 SW 5 SW 6 SW 7 SW 8 Initialization Mode
Table 26 LC Modules with 8-bit configuration switch (B-firmware) (configuration switch at rear
of the instrument)
SW 6 SW 7 SW 8 Initialization Mode
Manual Configuration
Manual configuration only alters the set of parameters stored in the non-volatile
memory of the module. It never affects the currently active parameters.
Therefore, manual configuration can be done at any time. A power cycle is
mandatory to make the stored parameters become the active parameters, given
that the initialization mode selection switches are allowing it.
TELNET
Session
Non-Volatile
RAM
Stored
Parameter
Control
Module
With Telnet
Whenever a TCP/IP connection to the module is possible (TCP/IP parameters set
by any method), the parameters may be altered by opening a Telnet session.
1 Open the system (DOS) prompt window by clicking on Windows START
button and select “Run...”. Type “cmd” and press OK.
2 Type the following at the system (DOS) prompt:
• c:\>telnet <IP address> or
• c:\>telnet <host name>
where <IP address> may be the assigned address from a Bootp cycle, a
configuration session with the Handheld Controller, or the default IP address
(see “Configuration Switches” on page 253).
When the connection was established successfully, the module responds
with the following:
3 Type
? and press enter to see the available commands.
Value Description
5 Use the “/” and press Enter to list the current settings.
6 Change the IP address (in this example 192.168.254.12) and type “/” to list
current settings.
change of IP setting to
Initialization mode is Using Stored
If the Initialization Mode Switch is changed now to “Using Stored” mode, the
NOTE instrument will take the stored settings when the module is re-booted. In the
example above it would be 192.168.254.12.
5 Press the Edit button (only visible if not in Edit mode), perform the required
changes and press the Done button.
6 Leave the screen by clicking Exit.
Safety Standards
This is a Safety Class I instrument (provided with terminal for protective earthing)
and has been manufactured and tested according to international safety
standards.
General
Do not use this product in any manner not specified by the manufacturer. The
protective features of this product may be impaired if it is used in a manner not
specified in the operation instructions.
In Case of Damage
Damage to the module
WAR N IN G
Personal injury (for example electrical shock, intoxication)
Instruments that appear damaged or defective should be made
inoperative and secured against unintended operation until they can be
repaired by qualified service personnel.
Solvents
Toxic, flammable and hazardous solvents, samples and reagents
WAR N IN G
The handling of solvents, samples and reagents can hold health and safety
risks.
When working with these substances observe appropriate safety
procedures (for example by wearing goggles, safety gloves and
protective clothing) as described in the material handling and safety
data sheet supplied by the vendor, and follow good laboratory practice.
Do not use solvents with an auto-ignition temperature below 200 °C
(392 °F). Do not use solvents with a boiling point below 56 °C (133 °F).
Avoid high vapor concentrations. Keep the solvent temperature at least
40 °C (72 °F) below the boiling point of the solvent used. This includes
the solvent temperature in the sample compartment. For the solvents
methanol and ethanol keep the solvent temperature at least 25 °C
(45 °F) below the boiling point.
Do not operate the instrument in an explosive atmosphere.
Do not use solvents of ignition Class IIC according IEC 60079-20-1 (for
example, carbon disulfide).
Reduce the volume of substances to the minimum required for the
analysis.
Never exceed the maximum permissible volume of solvents (8 L) in the
solvent cabinet. Do not use bottles that exceed the maximum
permissible volume as specified in the usage guideline for solvent
cabinet.
Ground the waste container.
Regularly check the filling level of the waste container. The residual free
volume in the waste container must be large enough to collect the waste
liquid.
To achieve maximal safety, regularly check the tubing for correct
installation.
For details, see the usage guideline for the solvent cabinet. A printed copy of the
NOTE guideline has been shipped with the solvent cabinet, electronic copies are
available in the Agilent Information Center or via the Internet.
Safety Symbols
Table 28 Symbols
The apparatus is marked with this symbol when the user shall refer to the
instruction manual in order to protect risk of harm to the operator and to
protect the apparatus against damage.
The apparatus is marked with this symbol when hot surfaces are available
and the user should not touch it when heated up.
Flammable Material
For Sample Thermostat which uses flammable refrigerant consult Agilent
Information Center / User Manual before attempting to install or service this
equipment. All safety precautions must be followed.
Manufacturing date.
Pacemaker
Magnets could affect the functioning of pacemakers and implanted heart
defibrillators.
A pacemaker could switch into test mode and cause illness. A heart
defibrillator may stop working. If you wear these devices keep at least
55 mm distance to magnets. Warn others who wear these devices from
getting too close to magnets.
Table 28 Symbols
Magnetic field
Magnets produce a far-reaching, strong magnetic field. They could damage
TVs and laptops, computer hard drives, credit and ATM cards, data storage
media, mechanical watches, hearing aids and speakers. Keep magnets at
least 25 mm away from devices and objects that could be damaged by
strong magnetic fields.
A WARNING
WAR N IN G
alerts you to situations that could cause physical injury or death.
Do not proceed beyond a warning until you have fully understood and
met the indicated conditions.
A CAUTION
C AU T I O N
alerts you to situations that could cause loss of data, or damage of equipment.
Do not proceed beyond a caution until you have fully understood and met
the indicated conditions.
This product complies with the European WEEE Directive marking requirements.
The affixed label indicates that you must not discard this electrical/electronic
product in domestic household waste.
Radio Interference
Sound Emission
Sound pressure
Sound pressure Lp <70 db(A) accroding to DIN-EN 27779
Schalldruckpegel
Schalldruckpegel Lp <70 db(A) nach DIN-EN 27779
UV-Radiation
This information is only valid for UV-lamps without cover (e.g. 2140-0590 and
NOTE 2140-0813).
Emissions of ultraviolet radiation (200-315 nm) from this product is limited such
that radiant exposure incident upon the unprotected skin or eye of operator or
service personnel is limited to the following TLVs (Threshold Limit Values)
according to the American Conference of Governmental Industrial Hygienists:
Table 29 UV-Radiation Limits
Typically the radiation values are much smaller than these limits:
Table 30 UV-Radiation Typical Values
Solvent Information
Flow Cell
To protect optimal functionality of your flow-cell:
• Avoid the use of alkaline solutions (pH > 9.5) which can attack quartz and
thus impair the optical properties of the flow cell.
Use of Solvents
Observe the following recommendations on the use of solvents.
• Brown glass ware can avoid growth of algae.
• Avoid the use of the following steel-corrosive solvents:
• solutions of alkali halides and their respective acids (for example, lithium
iodide, potassium chloride, and so on),
• high concentrations of inorganic acids like sulfuric acid and nitric acid,
especially at higher temperatures (if your chromatography method allows,
replace by phosphoric acid or phosphate buffer which are less corrosive
against stainless steel),
• halogenated solvents or mixtures which form radicals and/or acids, for
example:
2CHCl3 + O2→ 2COCl2 + 2HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs
quickly with dried chloroform if the drying process removes the stabilizing
alcohol,
• chromatographic grade ethers, which can contain peroxides (for example,
THF, dioxane, diisopropyl ether) should be filtered through dry aluminium
oxide which adsorbs the peroxides,
• solvents containing strong complexing agents (e.g. EDTA),
• mixtures of carbon tetrachloride with 2-propanol or THF.
• Avoid the use of dimethyl formamide (DMF). Polyvinylidene fluoride (PVDF),
which is used in leak sensors, is not resistant to DMF.
This installation procedure applies for capillaries and corresponding fittings used
NOTE in modules delivered before January 2013.
The 1260 Infinity Bio-inert LC system uses PEEK capillaries that are cladded with
stainless steel. These capillaries combine the high pressure stability of steel with
the inertness of PEEK. They are used in the high pressure flow path after sample
introduction (loop/needle seat capillary) through the thermostatted column
compartment/heat exchangers to the column. Such capillaries need to be
installed carefully in order to keep them tight without damaging them by
over-tightening.
The installation consists of two steps. In the first step, the fitting is installed
finger-tight without using tools. Finger-tight means that the fitting will grip and
hold the capillary. This brings the fitting to the appropriate start position (marked
as 0 ° below) for the second step.
possibly
leaky
recommended
range
possibly
leaky
recommended
range
recommended
range
possibly
leaky
recommended
range
Hard connectors
possibly
possibly leaky
leaky
recommended
range
recommended
range
Soft connectors
possibly
leaky possibly
leaky
recommended recommended
range range
Removing Capillaries
Potential damage of capillaries
C AU T I O N
Do not remove fittings from used capillaries.
To keep the flow path free of stainless steel, the front end of the capillary is made
of PEEK. Under high pressure, or when in contact with some solvents, PEEK can
expand to the shape of the connector where the capillary is installed. If the
capillary is removed, this may become visible as a small step. In such cases, do
not try to pull the fitting from the capillary, as this can destroy the front part of the
capillary. Instead, carefully pull it to the rear. During installation of the capillary,
the fitting will end up in the correct position.
Rear Front
Step
For the latest information on products and services visit our worldwide web site
on the Internet at:
http://www.agilent.com
www.agilent.com
Agilent Technologies Inc. 2016-2021
Edition: 03/2021
Document No: SD-29000218 Rev. E