0% found this document useful (0 votes)
75 views9 pages

Ultraviolet Absorption Spectrophotometry Guide

UV/VIS BP Volume 5
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
75 views9 pages

Ultraviolet Absorption Spectrophotometry Guide

UV/VIS BP Volume 5
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Status: You are viewing the legally effective version.

Appendix II B. Ultraviolet and Visible


Absorption Spectrophotometry
(Ph. Eur. method 2.2.25)

Determination of absorbance

The absorbance (A) of a solution is defined as the logarithm to base 10 of the reciprocal of the
transmittance (T) for monochromatic radiation:

T = I/I0;

I0 = intensity of incident monochromatic radiation;

I = intensity of transmitted monochromatic radiation.

In the absence of other physico-chemical factors, the absorbance (A) is proportional to the path
length (b) through which the radiation passes and to the concentration (c) of the substance in
solution in accordance with the equation:

∊ = molar absorptivity, if b is expressed in centimetres and c in moles per litre.

The expression representing the specific absorbance of a dissolved substance refers to


the absorbance of a 10 g/L solution in a 1 cm cell and measured at a defined wavelength so that:

Unless otherwise prescribed, measure the absorbance at the prescribed wavelength using a path
length of 1 cm. Unless otherwise prescribed, the measurements are carried out with reference to
the same solvent or the same mixture of solvents. The absorbance of the solvent measured
against air and at the prescribed wavelength shall not exceed 0.4 and is preferably less than 0.2.
Plot the absorption spectrum with absorbance or function of absorbance as ordinate against
wavelength or function of wavelength as abscissa.

Where a monograph gives a single value for the position of an absorption maximum, it is
understood that the value obtained may differ by not more than ± 2 nm.

Apparatus Spectrophotometers suitable for measuring in the ultraviolet and visible range of the
spectrum consist of an optical system capable of producing monochromatic radiation in the range
of 200-800 nm and a device suitable for measuring the absorbance.

Control of wavelengths Verify the wavelength scale using the absorption maxima of holmium
perchlorate solution R, the line of a hydrogen or deuterium discharge lamp or the lines of a
mercury vapour arc shown in Table 2.2.25.-1. The permitted tolerance is ± 1 nm for the ultraviolet
range and ± 3 nm for the visible range. Suitable certified reference materials may also be used.

Control of absorbance Check the absorbance using suitable filters or a solution of potassium
dichromate R at the wavelengths indicated in Table 2.2.25.-2, which gives for each wavelength
the exact value and the permitted limits of the specific absorbance. The table is based on a
tolerance for the absorbance of ± 0.01.

For the control of absorbance, use solutions of potassium dichromate R that has been previously
dried to constant mass at 130 °C. For the control of absorbance at 235 nm, 257 nm, 313 nm and
350 nm, dissolve 57.0-63.0 mg of potassium dichromate R in 0.005 M sulfuric acid and dilute to
1000.0 mL with the same acid. For the control of absorbance at 430 nm, dissolve 57.0-63.0 mg
of potassium dichromate R in 0.005 M sulfuric acid and dilute to 100.0 mL with the same acid.
Suitable certified reference materials may also be used.
Limit of stray light Stray light may be detected at a given wavelength with suitable filters or
solutions: for example, the absorbance of a 12 g/L solution of potassium chloride R in a 1 cm cell
increases steeply between 220 nm and 200 nm and is greater than 2.0 at 198 nm when
compared with water as compensation liquid. Suitable certified reference materials may also be
used.

Resolution (for qualitative analysis) When prescribed in a monograph, measure the resolution
of the apparatus as follows: record the spectrum of a 0.02 per cent V/V solution of toluene R in
hexane R. The minimum ratio of the absorbance at the maximum at 269 nm to that at the
minimum at 266 nm is stated in the monograph. Suitable certified reference materials may also
be used.

Spectral slit-width (for quantitative analysis) To avoid errors due to spectral slit-width, when
using an instrument on which the slit-width is variable at the selected wavelength, the slit-width
must be small compared with the half-width of the absorption band but it must be as large as
possible to obtain a high value of I0. Therefore, a slit-width is chosen such that further reduction
does not result in a change in absorbance reading.

Cells The tolerance on the path length of the cells used is ± 0.005 cm. When filled with the same
solvent, the cells intended to contain the solution to be examined and the compensation liquid
must have the same transmittance. If this is not the case, an appropriate correction must be
applied.

The cells must be cleaned and handled with care.

DERIVATIVE SPECTROPHOTOMETRY
Derivative spectrophotometry involves the transformation of absorption spectra (zero-order) into
first-, second- or higher-order-derivative spectra.
A first-order-derivative spectrum is a plot of the gradient of the absorption curve (rate of change
of the absorbance with wavelength, dA/dλ) against wavelength.

A second-order-derivative spectrum is a plot of the curvature of the absorption spectrum against


wavelength (d2A/dλ2). The second-order-derivative spectrum at any wavelength λ is related to
concentration by the following equation:

c′ = concentration of the absorbing solute, in grams per litre.

Apparatus Use a spectrophotometer complying with the requirements prescribed above and
equipped with an analogue resistance-capacitance differentiation module or a digital
differentiator or other means of producing derivative spectra. Some methods of producing
second-order-derivative spectra produce a wavelength shift relative to the zero-order spectrum
and this is to be taken into account where applicable.

Figure 2.2.25.-1

Resolution power When prescribed in a monograph, record the second-order-derivative


spectrum of a 0.02 per cent V/V solution of toluene R in methanol R, using methanol R as the
compensation liquid. The spectrum shows a small negative extremum located between 2 large
negative extrema at 261 nm and 268 nm, respectively, as shown in Figure 2.2.25.-1. Unless
otherwise prescribed in the monograph, the ratio A/B (see Figure 2.2.25.-1) is not less than 0.2.

Procedure Prepare the solution of the substance to be examined, adjust the various instrument
settings according to the manufacturer’s instructions, and calculate the amount of the substance
to be determined as prescribed in the monograph.

Status: You are not viewing the legally effective [Link] version becomes effective on
01/01/2016.

Appendix II B. Ultraviolet and Visible


Absorption Spectrophotometry
([Link]. method 2.2.25)

Determination of absorbance

The absorbance (A) of a solution is defined as the logarithm to base 10 of the reciprocal of the
transmittance (T) for monochromatic radiation:

T = I/I0;

I0 = intensity of incident monochromatic radiation;

I = intensity of transmitted monochromatic radiation.

In the absence of other physico-chemical factors, the absorbance (A) is proportional to the path
length (b) through which the radiation passes and to the concentration (c) of the substance in
solution in accordance with the equation:
ε = molar absorptivity, if b is expressed in centimetres and c in moles per litre.

The expression representing the specific absorbance of a dissolved substance refers to


the absorbance of a 10 g/L solution in a 1 cm cell and measured at a defined wavelength so that:

Unless otherwise prescribed, measure the absorbance at the prescribed wavelength using a path
length of 1 cm. Unless otherwise prescribed, the measurements are carried out with reference to
the same solvent or the same mixture of solvents. The absorbance of the solvent measured
against air and at the prescribed wavelength shall not exceed 0.4 and is preferably less than 0.2.
Plot the absorption spectrum with absorbance or function of absorbance as ordinate against
wavelength or function of wavelength as abscissa.

Where a monograph gives a single value for the position of an absorption maximum, it is
understood that the value obtained may differ by not more than ± 2 nm.

Apparatus Spectrophotometers suitable for measuring in the ultraviolet and visible range of the
spectrum consist of an optical system capable of producing monochromatic radiation in the range
of 200-800 nm and a device suitable for measuring the absorbance.

Control of wavelengths Verify the wavelength scale using the absorption maxima of holmium
perchlorate solution R, the line of a hydrogen or deuterium discharge lamp or the lines of a
mercury vapour arc shown in Table 2.2.25.-1. The permitted tolerance is ± 1 nm for the ultraviolet
range and ± 3 nm for the visible range. Suitable certified reference materials may also be used.

Control of absorbance Check the absorbance using suitable filters or a solution of potassium
dichromate R at the wavelengths indicated in Table 2.2.25.-2, which gives for each wavelength
the exact value and the permitted limits of the specific absorbance. The table is based on a
tolerance for the absorbance of ± 0.01.

For the control of absorbance, use solutions of potassium dichromate R that has been previously
dried to constant mass at 130 °C. For the control of absorbance at 235 nm, 257 nm, 313 nm and
350 nm, dissolve 57.0-63.0 mg of potassium dichromate R in 0.005 M sulfuric acid and dilute to
1000.0 mL with the same acid. For the control of absorbance at 430 nm, dissolve 57.0-63.0 mg
of potassium dichromate R in 0.005 M sulfuric acid and dilute to 100.0 mL with the same acid.
Suitable certified reference materials may also be used.

Limit of stray light Stray light may be detected at a given wavelength with suitable filters or
solutions: for example, the absorbance of a 12 g/L solution of potassium chloride R in a 1 cm cell
increases steeply between 220 nm and 200 nm and is greater than 2.0 at 198 nm when
compared with water as compensation liquid. Suitable certified reference materials may also be
used.

Resolution (for qualitative analysis) When prescribed in a monograph, measure the resolution
of the apparatus as follows: record the spectrum of a 0.02 per cent V/V solution of toluene R in
hexane R. The minimum ratio of the absorbance at the maximum at 269 nm to that at the
minimum at 266 nm is stated in the monograph. Suitable certified reference materials may also
be used.

Spectral slit-width (for quantitative analysis) To avoid errors due to spectral slit-width, when
using an instrument on which the slit-width is variable at the selected wavelength, the slit-width
must be small compared with the half-width of the absorption band but it must be as large as
possible to obtain a high value of I0. Therefore, a slit-width is chosen such that further reduction
does not result in a change in absorbance reading.

Cells The tolerance on the path length of the cells used is ± 0.005 cm. When filled with the same
solvent, the cells intended to contain the solution to be examined and the compensation liquid
must have the same transmittance. If this is not the case, an appropriate correction must be
applied.

The cells must be cleaned and handled with care.

DERIVATIVE SPECTROPHOTOMETRY
Derivative spectrophotometry involves the transformation of absorption spectra (zero-order) into
first-, second- or higher-order-derivative spectra.

A first-order-derivative spectrum is a plot of the gradient of the absorption curve (rate of change
of the absorbance with wavelength, dA/dλ) against wavelength.

A second-order-derivative spectrum is a plot of the curvature of the absorption spectrum against


wavelength (d2A/dλ2). The second-order-derivative spectrum at any wavelength λ is related to
concentration by the following equation:

c′ = concentration of the absorbing solute, in grams per litre.

Apparatus Use a spectrophotometer complying with the requirements prescribed above and
equipped with an analogue resistance-capacitance differentiation module or a digital
differentiator or other means of producing derivative spectra. Some methods of producing
second-order-derivative spectra produce a wavelength shift relative to the zero-order spectrum
and this is to be taken into account where applicable.
Figure 2.2.25.-1

Resolution power When prescribed in a monograph, record the second-order-derivative


spectrum of a 0.02 per cent V/V solution of toluene R in methanol R, using methanol R as the
compensation liquid. The spectrum shows a small negative extremum located between 2 large
negative extrema at 261 nm and 268 nm, respectively, as shown in Figure 2.2.25.-1. Unless
otherwise prescribed in the monograph, the ratio A/B (see Figure 2.2.25.-1) is not less than 0.2.

Procedure Prepare the solution of the substance to be examined, adjust the various instrument
settings according to the manufacturer’s instructions, and calculate the amount of the substance
to be determined as prescribed in the monograph.

Common questions

Powered by AI

Spectral slit-width plays a crucial role in quantitative analysis using spectrophotometry as it must be small compared to the half-width of the absorption band to avoid errors. A smaller slit-width enhances resolution and accuracy by ensuring that the maximum intensity possible is transmitted through the spectrophotometer, which improves the precision of the absorbance measurements .

In qualitative spectrophotometry, the ratio of absorbance maxima, such as the ratio of absorbance at maximum to minimum (e.g., 269 nm to 266 nm in toluene R and hexane R), is used to assess the resolution power of the spectrophotometer. A specific ratio value, like not less than 0.2, indicates that the instrument can resolve peaks effectively. This resolution is crucial for distinguishing between closely spaced spectral features in complex samples .

To ensure consistent transmittance in cells used for spectrophotometry measurements, the cells intended for containing the solution and compensation liquid must have the same transmittance. If they do not, an appropriate correction must be applied. Cells should be handled with care and properly cleaned to avoid any discrepancies caused by residues or damage .

A wavelength shift in second-order derivative spectra relative to the zero-order spectrum can lead to inaccuracies in identifying specific spectral features or concentrations of analytes. This shift must be accounted for by adjusting the analysis procedure accordingly, as the positions of peaks and troughs will not directly align with those in the zero-order spectrum, potentially leading to incorrect interpretation of results .

Derivative spectrophotometry enhances the analysis of absorption spectra by transforming zero-order absorption spectra into higher-order derivative spectra, such as first- and second-order derivatives. The first-order derivative spectrum plots the gradient of the absorption curve (dA/dλ) against wavelength, while the second-order derivative plots the curvature (d²A/dλ²). These transformations allow for improved resolution and detection of overlapping spectral features, making it easier to analyze complex mixtures .

Stray light in UV-visible spectrophotometry is detected by using suitable filters or solutions. For example, the absorbance of a 12 g/L solution of potassium chloride in a 1 cm cell increases steeply between 220 nm and 200 nm, and it is greater than 2.0 at 198 nm when compared with water. Stray light can significantly affect the accuracy of absorbance readings because it contributes to an unaccounted background signal, thereby leading to erroneous results .

In ultraviolet-visible spectrophotometry, the tolerance on the path length of the cells is ±0.005 cm. Variations beyond this tolerance can affect the accuracy of absorbance measurements because absorbance is directly proportional to the path length through which radiation passes. Ensuring consistent path length is vital for achieving reliable and replicable results .

For wavelength accuracy verification in a spectrophotometer used for ultraviolet and visible spectrophotometry, the wavelength scale should be verified using absorption maxima of holmium perchlorate solution, lines from a hydrogen or deuterium discharge lamp, or mercury vapor arc lines. The permitted tolerance is ±1 nm for the ultraviolet range and ±3 nm for the visible range. Certified reference materials can also be used for this verification .

The fundamental principle behind absorbance in ultraviolet and visible absorption spectrophotometry is that absorbance (A) is defined as the logarithm to base 10 of the reciprocal of the transmittance (T) of monochromatic radiation. Transmittance is determined by the ratio of the intensity of transmitted radiation (I) to the intensity of incident radiation (I0). This principle ensures that absorbance is proportional to the path length of the radiation and the concentration of the substance in solution .

Potassium dichromate is used for the control of absorbance in UV-visible spectrophotometry because it provides consistent and reliable absorbance values at specific wavelengths due to its well-defined spectrum. It serves as a standard reference material that allows for verification of the spectrophotometer's performance, ensuring accurate and precise absorbance measurements .

You might also like