PROVEO 8 Surgical Microscope Manual
PROVEO 8 Surgical Microscope Manual
PROVEO 8
User Manual
10 733 910 – Version 03
www.leica-microsystems.com
Thank you for choosing our products. We hope that you will enjoy the quality and
performance of your Leica Microsystems surgical microscope.
Legal disclaimer
All specifications are subject to change without notice.
The information provided by this manual is directly related to the operation of the
equipment. Medical decision remains the responsibility of the clinician.
Leica Microsystems has made every effort to provide a complete and clear user
manual highlighting the key areas of product use. Should additional information
regarding the use of the product be required, please contact your local Leica
representative.
You should never use a medical product of Leica Microsystems without the full
understanding of the use and the performance of the product.
Liability
For our liability, please see our standard sales terms and conditions. Nothing in this
disclaimer will limit any of our liabilities in any way that is not permitted under
applicable law, or exclude any of our liabilities that may not be excluded under
applicable law.
Contents
1 Introduction 3 8.1 Switching the microscope on 39
1.1 About this user manual 3 8.2 Surgeon panel 40
1.2 Symbols in this user manual 3 8.3 Positioning the optics carrier 41
1.3 Optional product features 3 8.4 Adjusting the optics carrier 42
8.5 Integrated slit illumination 44
2 Product identification 3
8.6 Transport position (F42) 44
3 Safety notes 4 8.7 Shutting down the surgical microscope 45
3.1 Intended use 4
9 Control unit with touch panel function 45
3.2 Directions for the person responsible for the
9.1 Menu structure 45
instrument 4
9.2 Selecting users 46
3.3 Directions for the operator of the instrument 5
9.3 Menu – User Settings 47
3.4 Dangers of use 6
3.5 Signs and labels 9 10 Menu 59
10.1 Menu – "Maintenance" 59
4 Design 13
10.2 Menu – "How to..." 60
4.1 F42 floor stand 13
10.3 Menu – "Service" 60
4.2 CT42 telescope mount 14
4.3 PROVEO 8 Optics carrier modules 15 11 Accessories 61
11.1 Devices and accessories manufactured by Leica
5 Functions 16
Microsystems (Schweiz) AG 61
5.1 Balancing system 16
11.2 Accessories list for advanced configurations 62
5.2 Brakes 17
11.3 Drapes 62
5.3 Illumination 19
5.4 Fusion Optics 19 12 Care and maintenance 63
12.1 Maintenance instructions 63
6 Controls 20
12.2 Cleaning the touch panel 63
6.1 PROVEO 8 optics carrier including focus, tilt and XY
12.3 Maintenance 63
with IVC 20
12.4 Changing fuses 64
6.2 PROVEO 8 optics carrier including focus, tilt and XY
12.5 Notes on reprocessing of resterilizable products 64
with IVA 21
6.3 Surgeon panel 22 13 Disposal 67
6.4 Balancing 22
14 What to do if ...? 68
6.5 Control unit 22
14.1 Malfunctions 68
6.6 F42 stand 22
14.2 Malfunctions documentation accessories 68
6.7 F42 Terminals 23
14.2 Error messages on the control unit 69
6.8 CT42 Terminals on the rack controller 24
6.9 CT42 horizontal arm 25 15 Specifications 71
6.10 Footswitch and handles 26 15.1 PROVEO 8 71
15.2 F42 floor stand 72
7 Preparation before surgery 28
15.3 CT42 telescope mount 72
7.1 Transportation (F42) 28
15.4 CT42 rack controller 73
7.2 Installing a binocular tube 29
15.5 Ambient conditions 73
7.3 Setting the binocular tube 30
15.6 Standards fulfilled 73
7.4 Adjusting the eyepiece 30
15.7 Limitations of use (only F42) 73
7.5 Change the 0° assistant to another position 31
15.8 Dimensional drawings 75
7.6 Changing the camera 32
7.7 Balancing and locking the parallelogram 33 16 Manufacturer's declaration of electromagnetic
7.8 Positioning on the operating table 35 compatibility (EMC) 78
7.9 Attaching sterile controls and drape 37 16.1 Table 1 from EN 60601-1-2 78
7.10 Function check 38 16.2 Table 2 from EN 60601-1-2 79
16.3 Table 4 from EN 60601-1-2 80
8 Operation 39
17 Annex 82
17.1 Checklist before the operation 82
The symbols used in this user manual have the following meaning:
3 Safety notes X When instructing new users, do so thoroughly and explain the
meanings of the warning signs and messages.
The PROVEO 8 surgical microscope is state-of-the-art technology. X Allocate responsibilities for commissioning, operation and
Nevertheless, hazards can arise during operation. maintenance. Monitor compliance with this.
X Always follow the instructions in this user manual, and in X The PROVEO 8 is intended for professional use only.
particular the safety notes.
WARNING
3.1 Intended use
Risk of electric shock!
• The PROVEO 8 surgical microscope is an optical instrument for X Connect this equipment only to a supply mains with
improving the visibility of objects through magnification and protective earth.
illumination. It can be applied for observation and
documentation and for human and veterinary medical X Only use the PROVEO 8 surgical microscope if it is free of defects.
treatment. X Inform your Leica representative or Leica Microsystems
• The major field of use is Ophtalmology.
(Schweiz) AG, Medical Division, 9435 Heerbrugg, Switzerland,
• The microscope device is not for use in microsurgery (Neuro,
immediately about any product defect that could potentially
Plastic/Reconstructive, Ear, Nose and Throat surgeries). cause injury or harm.
• The PROVEO 8 surgical microscope may be used only in closed X If you use accessories from other manufacturers with the
rooms and must be placed on a solid floor or mounted to the PROVEO 8 surgical microscope, make sure that these
ceiling. manufacturers confirm that the combination is safe to use.
• It is available on the floor stand F42 and on the telescope mount
Follow the instructions in the user manual for those accessories.
CT42. • Modifications, installations to or service on the PROVEO 8
The F42 is a floor stand for three-dimensional positioning of the surgical microscope may be carried out only by technicians who
PROVEO 8 in the room. are explicitly authorized by Leica to do so.
The CT42 is a ceiling mounted electrical linear actuator for For further details refer to the installation manual.
lifting and lowering the PROVEO 8. It is operated by a hand • Only original Leica replacement parts may be used in servicing
switch. the product.
• The PROVEO 8 surgical microscope is subject to special • After service work or technical modifications, the device must
precautionary measures for electromagnetic compatibility. It be readjusted in accordance with our technical specifications.
must be installed and commissioned in accordance with the • If the instrument is modified or serviced by unauthorized
guidelines and manufacturer's declarations and recommended persons, is improperly maintained (as long as maintenance was
safety distances (according to EMC tables based on not carried out by a qualified trained service engineer), or is
EN60601‑1‑2). handled improperly, Leica Microsystems will not accept any
• Portable and mobile as well as stationary RF communications
liability.
equipment can have a negative effect on the reliability of the • The effect of the surgical microscope on other instruments has
PROVEO 8 surgical microscope's functionality. been tested as specified in EN 60601-1-2. The system passed the
• Always release the brakes to move or relocate the PROVEO 8
emission and immunity test. Comply with the usual pre
surgical microscope. cautionary and safety measures relating to electromagnetic and
• The essential performance of the PROVEO 8 is to provide
other forms of radiation.
illumination and mechanical stability of the optics carrier in any • The electric installation in the building must conform to the
position. national standard, e.g., current-operated ground leakage
protection (fault-current protection) is suggested.
3.2 Directions for the person • Like any other instrument in the operating theater, this system
responsible for the instrument may fail. Leica Microsystems (Schweiz) AG therefore
X Ensure that the PROVEO 8 surgical microscope is used only by recommends that a backup system be kept available during the
persons qualified to do so. operation.
X Ensure that this user manual is always available at the place
where the PROVEO 8 surgical microscope is in use.
X Carry out regular inspections to make certain that the
authorized users are adhering to safety requirements.
Coaxial Red Reflex illumination Decisions about brightness must be made case by case. In any
event, the surgeon must evaluate the risks and benefits of the used
Light setting Recommended maximum exposure time according to 1) [min.]
light intensity. Despite all efforts to minimize the risk of retinal
25% 10
injury by surgical microscopes, damage may still occur.
50 % 4.5
Photochemical retinal damage is a possible complication of the
75% 3 necessity to use bright light to make eye structures visible during
100 % 2 difficult ophthalmological processes.
Retina Protection
14
Function Activated
In addition, the Retina Protection function can be activated during
the surgery to reduce the main light intensity below 10% and
RedReflex below 20%.
If the integrated slit illumination motor fails, the slit • when transporting the microscope
illumination can be manually activated/deactivated using • when re-equipping
the rotary knob (2).
WARNING CAUTION
Risk of injury from parts falling down! Damage of the touch panel!
X Before the operation, make sure that the optical X Operate the touch panel using your fingers only.
components and accessories are sufficiently secured and Never use hard, sharp or pointed objects made out of
cannot move. wood, metal or plastic.
X Never clean the touch panel using cleaners that contain
WARNING abrasive substances. These substances can scratch the
Risk of infection! surface and cause it to be become dull.
X Always use the PROVEO 8 surgical with sterile controls
and knobs. CAUTION
Surgical microscope can move without warning!
WARNING X Always lock the footbrake when you are not moving the
Danger of fatal electrical shock! system.
X Operate the system only with all equipment in its proper
position (all covers fitted, doors closed).
CAUTION
Damage to the PROVEO 8 surgical microscope due to
WARNING uncontrolled tilting!
Danger of fatal electrical shock! X Hold the handle when releasing the brake.
X The PROVEO 8 surgical microscope may be connected to
a grounded socket only (F42).
CAUTION
Damage to the PROVEO 8 surgical microscope during
WARNING transportation!
Motors return to their park position! X Never move the stand in the extended condition.
X Before switching on the microscope, ensure that the X Never roll over cables lying on the floor.
travel paths of XY-unit and focus are free of X Do not transport or store the system in areas with an
obstructions. elevation angle bigger than 10°.
WARNING CAUTION
Light that is too intense can damage the retina! Damage to the housing of the stand or to the touch panel of
X Observe the warning messages in the chapter on "Safety the control unit!
notes". If the optics carrier is moved into the transport position or
from the transport position to the operating position:
X Ensure that the transport lock must be locked.
WARNING
Danger to the patient due to failure of the magnification
motor! CAUTION
If the magnification motor fails, the magnification can be There is a risk of damage to the surgical microscope from
manually adjusted using the rotary knob. uncontrolled tilting!
X Firmly hold the handles before triggering the "All
CAUTION
Risk of infection!
X Leave sufficient space around the stand to ensure that
the sterile drape does not come into contact with non-
sterile components.
CAUTION
Risk of overheating of the system!
Covering the air inlet can result in a controlled shutdown of
the system due to overheating.
X Make sure that there is always some space between the
air inlet and the drape.
CAUTION
Damage of the magnification motor!
X Only adjust the magnification manually if the
magnification motor is defective.
CAUTION
Danger to the patient due to changes in the user settings!
X Never change the configuration settings or edit the user
list during an operation.
CAUTION
Risk of injury!
X Pay special attention to the required safety distances if
you use the Combination Mode function together with
accessories from other manufacturers that can reduce
the working distance to less than 140 mm (non-contact
wide-angle observation systems), since focus together
with Combination Mode is a semi-automated function.
CAUTION
Danger of collision!
The surgical microscope can collide with surrounding
components, the ceiling or lamps.
X Check the danger area before moving the swing arm.
X Carefully move the ceiling mount upwards, and observe
ceiling and lamps.
13
11
12
13
13
14
10
11 12
13
9 8 8
2 15 3 4 5 4 6 6
MD Medical device
2 Fabrication label
7 System weight label
8 Only to be operated
Trained personnel
by trained personnel
Utilisation réservée à
un personnel formé
a Prefix number
b Leica system article no. 9 Equipotential bonding
c Serial number
d Incremental number
starting at 1 for each batch
e JJ = year (2 ditis) 10 Max. load for Optics
max. 8.0 kg (17.64 lb)
f MM = month (2 digits) Carrier
g TT = day (2 digits)
h Production starting date 11 Open
3 Follow the User Manual
12 Closed
4 Grounding reliability can only Grounding label
be achieved when EQUIPMENT is
connected to equivalent
receptacle marked "Hospital
only" or "Hospital Grade".
13 Crushing hazard label
For US and Canada only
La fiabilité de la mise à la
terre n'est assurée que si
l'équipement est connecté à une 14 Monitor weight label
max. 16 kg (35.27 lb)
prise équivalente, marquée
"Hôpital seulement" ou "Qualité
hôpital".
15 UDI Label
5 MET label
Device Identifier (DI)
GS1 Data Matrix Code
Production Identifier (PI)
Serial number
Date of manufacturing
A
B
100 –240 V 50 - 60 Hz
500 VA
7 Equipotential
2 Fabrication label bonding
8 Max. load for optics
Max. 8.0 kg (17.64 lb)
carrier
9 Open
10 Closed
a Prefix number
b Leica system article no.
c Serial number 11 Crushing hazard label
d Incremental number starting at
1 for each batch
e JJ = year (2 digits)
f MM = month (2 digits) 12 Monitor weight label
g TT = day (2 digits) max. 16 kg (35.27 lb)
La fiabilité de la mise à la
terre n'est assurée que si 14 MODEL PROVEO 8 Type label
l'équipement est connecté à une STAND C42 / CT42
prise équivalente, marquée
"Hôpital seulement" ou "Qualité MODULE Swing Arm
hôpital".
15 UDI Label
5 MET label
Device Identifier (DI)
GS1 Data Matrix Code
Production Identifier (PI)
Serial number
Date of manufacturing
4 Design
4.1 F42 floor stand
Balancing the optics carrier at the CT42 telescope mount 5.2 Brakes
The PROVEO 8 surgical microscope has 4 electromagnetic brakes
which stop the movements of the stand, the ceiling mount and the
surgical microscope:
The electromagnetic brakes can be released via handle or nurse
switch.
5.2.2 Selected brakes at the F42 floor stand 5.2.3 Brakes at the CT42 telescope mount
With the function "Selected brakes" the user can release any
individual brake.
1 Rotation of tower
2 Rotation of parallelogram
3 Up/Down movement of parallelogram
4 Rotation of microscope carrier
5.3 Illumination
The illumination of the surgical microscope PROVEO 8 consists of
two LED modules and are located in the optics carrier.
There are two lamps, main lamp and Red Reflex lamp.
6 Controls
6.1 PROVEO 8 optics carrier including focus, tilt and XY with IVC
19
18 17
6.2 PROVEO 8 optics carrier including focus, tilt and XY with IVA
1 XY Coupling 11 0° assistant
2 Rotary knob for changing the 0° assistant position 12 Surgeon panel
3 Fine focus for external camera 13 2× CAN sockets - only for Leica accessories
4 Rotary knob "Magnification" (emergency operation only) 14 C-mount adapter
5 Rotary knob "Slit illumination" (emergency operation only) 15 Cooling slots
6 Slot for filter slide 16 Objective
7 Handle 17 Mounting threads for accessories
8 Rotary knob for Red Reflex illumination diameter 18 Socket for BIOM
9 PROVEO 8 optics carrier
10 Fine focus for assistant
1 Surgeon panel
2 Infrared receiver for remote control (Recorder)
5
3 Touchpanel
1
6.4 Balancing
4
2
3
WARNING
Risk of human errors due to loud disturbing noises during
surgery!
X Don't plug/unplug external devices at the "Line In" input
during surgery.
X Make sure that the volume level on audio device is set to
the minimum before connecting to the "Line In" input.
1 Mains switch
2 Mains fuses
3 Equipotential bonding socket
For connecting the PROVEO 8 to an equipotential bonding
device. This is part of the customer's building installation.
Observe the requirements of EN 60601-1 (§ 8.6.7).
See also User manual Wireless Footswitch, 14 functions. Footswitches and handles can be assigned individually for
each user in the configuration menu.
Focus− Focus+
Magnification− Magnification+
Main light+ Red Reflex+
All Lights On/Off VR Mode On/Off
Main light− Ref Reflex−
Focus− Focus+
Magnification− Magnification+
Handles
All Lights On/Off
Handles
Footswitches and handles can be assigned individually for Footswitches and handles can be assigned individually for
each user in the configuration menu. each user in the configuration menu.
OCT Up OCT Up
OCT Save OCT Change Joystick State OCT Save OCT Change Joystick State
OCT Scan OCT Next Procedure OCT Scan OCT Next Procedure
OCT Optimize Image OCT Auto Locate OCT Optimize Image OCT Auto Locate
OCT Live Mode/Stop OCT Mode On/Off OCT Live Mode/Stop OCT Mode On/Off
Handles Handles
Change View OCT Frame Forward Change View OCT Frame Forward
All Brakes OCT Frame Backward All Brakes OCT Frame Backward
7 Preparation before surgery X Ensure that the PROVEO 8 is in the transport position.
CAUTION
Surgical microscope can move without warning!
X Always lock the footbrake when you are not moving the
system.
CAUTION
Damage to the PROVEO 8 surgical microscope due to
uncontrolled tilting! In case the PROVEO 8 is not in transport position, refer to
X Hold the handle when releasing the brake. section 8.6.
X Press the footbrake padel on the right (1) to unlock the castor
CAUTION wheels lock.
Damage to the PROVEO 8 surgical microscope during X Move the microscope using the handle.
transportation! X Press the footbrake padel on the left (2) to lock the castor
X Never move the stand in the extended condition. wheels lock.
X Never roll over cables lying on the floor.
X Do not transport or store the system in areas with an
elevation angle bigger than 10°.
CAUTION
Damage to the housing of the stand or to the touch panel of
the control unit!
If the optics carrier is moved into the transport position or 2
from the transport position to the operating position:
X Ensure that the transport lock must be locked.
WARNING
Risk of injury from parts falling down!
X Before the operation, make sure that the optical
components and accessories are sufficiently secured and
cannot move.
X Insert the laserfilter IVC (4) into the laserfilter slot of the optics 7.7 Balancing and locking the
carrier.
parallelogram
7.7.1 Balancing the parallelogram
WARNING
Risk of injury due to downward movement of the surgical
microscope!
X Never change the accessories or attempt to rebalance
the microscope while it is above the field of operation.
X After re-equipping, always rebalance the microscope on
the parallelogram.
The plane of filter is sharply imaged in the same plane as the object. WARNING
The GG420 UV protection filter is built-in. In addition, the "GG475 Risk of injury due to downward movement of the surgical
UV protection filter" and the "Protection Filter 5×" are available. microscope!
X Remove the filter cover. X Always lock the parallelogram:
X Push in the filter slide, inclined slightly upwards, until it • when transporting the microscope
engages. • when re-equipping
CAUTION
There is a risk of damage to the surgical microscope from
uncontrolled tilting!
X Firmly hold the handles before triggering the
"All Brakes" function.
X Pull the transport lock (1) and bring it into a horizontal position. 7.7.4 Releasing the brakes
WARNING
Risk of injury from downward movement of surgical
microscope!
X Complete all preparations and adjustments to the stand
before the operation.
X If settings need to be altered during the operation, first
swing the microscope away from the operating field.
X If the microscope needs to be re-equipped, do this
before the operation.
X Before re-equipping, always lock the parallelogram.
X Do not use the handles or remote brake release when
the instrument is in an unbalanced state.
CAUTION
There is a risk of damage to the surgical microscope from Forwards
uncontrolled tilting!
X Firmly hold the handles before triggering the
"All Brakes" function.
X Grip and turn one handle to release the brakes. Backwards
X At the same time, pull the transport lock (1) and bring it into a
vertical position.
Unless they are individually configured for the current user, the
brakes are released by turning the handles as follows:
X Turn backwards and hold: Selected brakes are released
X Turn forwards and hold: All brakes are released
7.8 Positioning on the operating table X Press the footbrake padel on the left (2) to lock the castor
wheels lock.
7.8.1 F42 floor stand
WARNING
Risk of injury due to downward movement of the surgical
microscope!
X Complete all preparations and adjustments to the stand
before the operation.
X Never change the accessories or attempt to rebalance
the microscope while it is above the field of operation. 2
X Before changing accessories, always lock the
parallelogram (see chapter 7.7.2).
X Balance the PROVEO 8 after re-equipping it.
X Do not release the brakes when the instrument is in an
1
unbalanced state.
X Before re-equipping during the operation, first swing
the microscope away from the operating field.
X Position the footswitch below the operating table.
CAUTION X Plug the power cable into the power outlet.
Danger of damage! X Connect the equipotential bonding to the stand.
X Before lifting the optics carrier make sure that the area X Start the system.
above the parallelogram is clear to avoid collisions with X Release the brakes (see chapter 7.7.4) and bring the system to a
OR lamps, ceiling etc. possible position (see picture below).
Maximum swing arm extension is at 135° angle.
X Using the handrail, carefully push the surgical microscope to the
operating table and position it as desired.
• All positions are also possible as the mirror image max. 135°
angle
position.
• The instrument must be positioned such that the range
of movement is large enough for the expected tasks.
7.8.2 CT42 telescope mount Raising and lowering the CT42 telescope mount
The CT42 telescope mount can be raised and lowered electrically.
CT42 telescope mount These functions can be controlled via buttons on the remote control
unit.
7.9 Attaching sterile controls and 7.9.2 Sterile drape for optics carrier
drape • Only use the sterile tested drapes specified in the
WARNING Accessories section.
• Drape only as far as indicated (see picture below).
Risk of infection!
X Always use the PROVEO 8 surgical with sterile controls
and knobs. CAUTION
Risk of infection!
7.9.1 Covers for rotary buttons X Leave sufficient space around the stand to ensure that
the sterile drape does not come into contact with non-
Use the covers also when you use sterile disposable drapes. sterile components.
The controls will be easier to grasp. X Activate the "All Brakes" function on the handle and extend the
X Fit steam-sterilizable covers on the magnification, working parallelogram.
distance and Red Reflex illumination manual override knobs.
Maximum swing arm extension is at 135° angle.
CAUTION
Risk of overheating of the system!
Covering the air inlet (1) can result in a controlled shutdown
of the system due to overheating.
X Make sure that there is always some space between the
air inlet (1) and the drape.
WARNING
8.1 Switching the microscope on Danger of fatal electrical shock!
X The PROVEO 8 surgical microscope may be connected to
WARNING
a grounded socket only (F42).
Danger of fatal electrical shock!
X Operate the system only with all equipment in its proper X Switch on the microscope using the power switch (1) on the
position (all covers fitted, doors closed). stand.
The system is starting an initialization procedure.
WARNING
Motors return to their park position! 1
X Before switching on the microscope, ensure that the
travel paths of XY-unit and focus are free of
obstructions.
Retina
Protect
CAUTION
Damage to the PROVEO 8 surgical microscope due to
uncontrolled movement!
X Hold the handle when releasing the brake.
8.4 Adjusting the optics carrier 8.4.2 Adjusting the Red Reflex illumination
diameter
8.4.1 Adjusting the brightness
You can adjust the Red Reflex illumination diameter using the
rotary knob (1) or the footswitch/handle.
WARNING
X Turn rotary knob (1) and set the Red Reflex illumination
Light that is too intense can damage the retina! diameter as required.
X Observe the warning messages in the chapter on "Safety
notes".
You can make the illumination brighter or darker for the Main Light
and the Red Reflex Light using either the touch panel monitor, a
footswitch, or handle.
On the touch panel monitor in the "Main" menu screen
On the footswitch/handle
Depending on the assignment (see chapters 9.3.4 and 9.3.5), you
can also increase and decrease the brightness of the illumination
use the correspondingly assigned buttons on the footswitch/
handle.
8.4.5 Adjusting the magnification (zoom) Manually adjusting the magnification (zoom)
You can adjust the magnification using a footswitch/handle or the WARNING
"Magnification" adjustment bar on the "Main" menu screen of the
Danger to the patient due to failure of the magnification
control unit.
motor!
On the touch panel monitor in the "Main" menu screen If the magnification motor fails, the magnification can be
manually adjusted using the rotary knob (1).
CAUTION
Damage to the magnification motor!
X Only adjust the magnification manually if the
magnification motor is defective.
WARNING
Danger to the patient due to failure of the integrated slit
illumination!
If the integrated slit illumination motor fails, the slit
illumination can be manually activated/deactivated using
the rotary knob (2).
NOTE
X Make sure that the video monitor does not collide with the
parallelogram of the stand.
1 2 3 4 5 6 7
The quick access line is displayed at the top of each screen to enable
access to the following menus at any time.
1 Main menu
2 Speed menu (see chapter 9.3.3)
3 Menu (see chapter 9.3)
4 DIC menu
5 Warning messages
6 Change Monitor Display View (activated when OCT is enabled in
the accessory list)
7 Display recording in progress (green blinking), no recording (no
color)
9.2.1 Presets
9.2 Selecting users You can find a list of default users preset by Leica for the most
In the "Main" and "Speed" menu screens, the three buttons common types of operation under "Presets".
"Presets", "User List" and "Show Settings" appear in the dynamic
button bar at all times.
"Load" Loads the settings of an existing user from the X Select the user.
user list for modifying. The available functions are displayed in the dynamic button
"Create New User" Opens a new user with "blank" settings. line:
"New (Preset)" Opens the "Preset" screen for selecting a default
user in order to create a new user with the "Cancel" Cancel the action.
settings of the desired preset and to load or "Move" Moves the selected user to another
modify the user's settings. available location of your choosing.
"Edit User List" Allows to rename, to move or to delete users. "Delete" Deletes the selected user.
"Rename" Renames an existing user. The user's
• You can also add a user from the operational menu. settings are not changed.
• If you want to keep the current settings, you can save "Change password" Changes the password.
them by clicking the "Save" button (which appears as
soon as the basic settings of the current users have been CAUTION
changed), either for the current user ("Save") or under a
Danger to the patient due to changes in the user settings!
new username ("Save as New").
X Never change the configuration settings or edit the user
list during an operation.
Editing the user list
Various functions are available in the user list depending on the
situation.
Fast Slitlamp When activated, the slitlamp moves with Configuring individual keys
fast travel speed. X In the left selection field, select the function group with the
Focus Linked to Zoom When activated, the focus speed depends desired functions.
on the magnification: X You can scroll forwards or backwards in the list by clicking the
• low magnifaction – fast focus speed arrowheads.
• high magnification – slow focus speed X Select the desired function.
X Click the caption of the desired key to assign the selected
XY Linked to Zoom When activated, the XY speed depends on
function to it.
the magnification:
• low magnifaction – fast XY speed
• high magnification – slow XY speed Overview of function groups
The possible configuration is divided into the following function
9.3.4 Footswitch assignment groups:
(Footswitch 1 and Footswitch 2) XY
Here the user can configure individual settings for the footswitch. • X+
• X–
• Y+
• Y–
• XY Reverse
Drive
• Magnification +
• Magnification –
• Focus +
• Focus –
• VR Lens Focus +
• VR Lens Focus –
• Tilt +
• Tilt –
• Slitwidth +
• Slitwidth –
• Slitlamp left
The built-in footswitch wireless receiver is Footswitch 1, the
• Slitlamp right
inserted optional footswitch is Footswitch 2.
• Main Inverters In/Out
• All Inverters In/Out
X First select a footswitch 14 or 12 functions. • VR Synchronized Focus+
X Click the left or right arrow to select Normal Mode, VR • VR Synchronized Focus–
Mode, IOL Mode, OCT Mode, or OCT VR Mode.
X Click the button "Cataract", "Retina", "IOL" or "OCT". Light
The default settings are assigned to the footswitch. • Mainlight On/Off
X You can then modify these settings as you like. • Red Reflex On/Off
Clicking the "Clear All" button clears the assignments for all • All Lights On/Off
keys. • Mainlight +
• Mainlight –
• Red Reflex +
• Red Reflex –
• Red Reflex Diameter +
• Red Reflex Diameter –
• Slitlamp In/Out
• Retina Protection On/OFF
The user can set desired values for quick focus and quick tilt
according to his preferences.
X Activate the footswitch button assigned to the quick focus or
quick tilt function by clicking it.
Distance for quick Distance to move upwards from current
focus position after activating the assigned key.
NOTE
X This function can be deactivated by your Leica Microsystems
service technician.
X Click the "OK" button to save the user at the desired location
under the name you have entered.
Select Eyepiece
In this field you can select the type of the eyepieces being used by
the surgeon.
X Scroll forwards or backwards in the list by clicking the
arrowheads.
Select Objective
In this field you can enter the objective currently being used by the
surgeon.
X Scroll forwards or backwards in the list by clicking the
arrowheads. X Press the button for the topic desired.
Detailed information "How to ..." is displayed.
10.2 Menu – "How to..."
X Press the Menu button and select "HOW TO" (1). 10.3 Menu – "Service"
X Press the Menu button and select "SERVICE MENU" (1).
11.1.2 Patient-side
Binocular tube, 30° to 150°, Picture Devices and accessories
T, Typ II L
Objective APO WD175 OCT ready
Objective APO WD200 OCT ready
Objective APO WD225 OCT ready
RUV800 WD175
Binocular tube, 10° to 50°, Typ II,
RUV800 WD200
UltraLow ™III
Slit Illumination
Storage
No special requirements.
Additional information
None
Contact information of manufacturer
Address of local agent
Leica Microsystems (Schweiz) AG verified that the aforementioned
instructions for the preparation of a product are suitable for its
reuse. The processing person is responsible for reprocessing with
the equipment, materials and personnel and for achieving the
desired results in the reprocessing installation. In general, this
requires validations and routine monitoring of the process. Every
deviation from the supplied instructions should also be examined
carefully by the processing person to determine effectiveness and
possible detrimental consequences.
1) This medical device falls within the validated sertility claims of the STERRAD®100S / STERRAD® 100NX™ / STERRAD®50 / STERRAD®200
Systems.Please follow the instructions for use of your STERRAD® System Umser's Guide prior to sterilizing devices in STERRAD® Systems.
13 Disposal
The respective applicable national laws must be observed for
disposal of the products, with the involvement of corresponding
disposal companies. The unit packaging is to be recycled.
14 What to do if ...?
If your instrument has a malfunction that is not described
here, please contact your Leica representative.
14.1 Malfunctions
Malfunction Cause Remedy
The image remains unfocussed. Eyepieces are not mounted correctly. X Screw the eyepieces all the way on.
Diopters not set correctly. X Perform dioptric correction exactly according to the
instructions (see chapter 7.4.1).
Magnification cannot be adjusted Failure of magnification motor. X Press the magnification rotary knob.
electrically. X Set magnification by turning (see chapter 8.4.5).
Parallelogram cannot be moved. Parallelogram locked in position. X Release the locking mechanism (see chapter 7.7.3).
PROVEO 8 is not correctly balanced. Position of accessory was changed after X Balance out PROVEO 8 (see chapter 7.7.1).
balancing.
The image appears shaded through the Accessories not installed properly. X Install the accessories exactly in the holders (see
microscope at the edges and the chapter 7.2).
illumination field is outside the field of
vision.
Functions cannot be activated using the Incorrect assignment entered at control X Change the assignment using the control unit.
foot switch. unit.
Error code Source Visible for user Meaning Possible cause of error
0112h PROVEO 8 Optics yes Main LED overtemperature • Main LED too hot.
• No air flow possible.
• Opitcs Carrier might be covered.
0116h PROVEO 8 Optics yes RedReflex LED overtemperature • RedReflex LED too hot.
• No air flow possible.
• Opitcs Carrier might be covered.
0402h Slitlamp adapter yes Lamp defective Lamp defective
FE01h MDC yes Optics carrier not found • CAN connection loose.
• Cable damaged.
• Controller defective.
FE02h MDC yes XY controller not found • CAN connection loose.
• Cable damaged.
• Controller defective.
FE03h MDC yes F42 controller not found • CAN connection loose.
• Cable damaged.
• Controller defective.
FE04h MDC yes Slitlamp adapter not found • CAN connection loose.
• Cable damaged.
• Controller defective.
• No slitlamp adapter installed (Device list not correctly
configured).
FE06h MDC yes DIC800 not found • CAN connection loose.
• Cable damaged.
• Controller defective.
• No DIC800 installed (Device list not correctly
configured)
FE07h MDC yes HDR not found • RS232 connection loose.
• Cable damaged.
• Controller defective.
• No IVR850 installed (Device list not correctly
configured).
FE08h MDC yes External inverter (SDI) not found • CAN connection loose.
• Cable damaged.
• Controller defective.
• No SDI module installed (Device list not correctly
configured).
Error code Source Visible for user Meaning Possible cause of error
FE09h MDC yes Camera control unit not found • RS232 connection loose.
• Cable damaged.
• Controller defective.
• No CCU module installed (Device list not correctly
configured).
15.1.3 Microscope carrier Integrated Prepared for integration of video camera system
documentation and digital recording system.
Rotation of optics 380° Open architecture
Inclination tilt –15° / +105° Connectors Numerous built-in connectors for Video and control
data transfer. Internal power supply 12 VDC,
XY speed Zoom linked XY speed
19 VDC, 24 VDC and AC terminals
Indicator LED for Video record status
Carrier for monitor 860 mm long and flexible arm with 4 axis for
IR sensor for remote control of the HDR rotation and inclination to carry optional video
monitor.
XY range 62 × 62 mm Max. monitor weight: 15.5 kg
Focus range 75 mm Materials Use of RoHS conform materials
Camera image size with respect to the field of view Surface coating The microscope is coated with a paint which is
system designed to provide an antimicrobial effect on
• Camera for surfaces.
visible light Field of view
Maximum height In park position: 1950 mm
Range Cantilever Max. 1925 mm
Camera image size Load Max. 8.0 kg from microscope dovetail ring interface
Maximum weight Approx. 345 kg without load
Field of view
15.3 CT42 telescope mount
• TLC 12ZWDN-100 (230 V)
15.2 F42 floor stand • TLC 12ZWDN2-100 (120 V)
Electrical data Electrical data
Power connection 600 VA Power connection 100–120 V~ 50-60 Hz 1200 W
100–240 V~ 50-60 Hz 220–230 V~ 50 Hz 890 W
Protection class Class 1 Protection class Class 1
Type Floor stand with 4 electromagnetic brakes. Ceiling attachment Max. distance from concrete ceiling to
intermediate ceiling: 1500 mm
Base 760 × 760 mm with four 360° rotating castors with Attachment to concrete shell construction ceiling:
a diameter of 150 mm each, one parking brake 440 mm hole circle
4 x M12 HSLB M12/15
Balancing Adjustable gas spring via balancing knob
Weight Swing arm 70 kg
Floor stand control The latest electronics control for the continuous
Monitor arm 15.5 kg
unit governing of all motor functions and the light
Ceiling bracket 60 kg
intensity.
Telescope 50 kg
Menu selection based on unique software for user-
specific configuration, with built-in electronic Total weight Approx. 200 kg
auto-diagnosis and user support
Brakes Swing arm:
Control unit stand Open architecture for future software Four electromagnetic brakes, operated by turning
developments the handles, one stop lever for vertical movement
Control elements 14/12 function wireless footswitch and rotary Load Swing arm: Max. 8.0 kg from microscope/dovetail
handles ring interface
max. 1571
(380)
max. 1231
max. 1
50°
°
. 190
max
no
nde
sto
p
max
. 190
°
max.
180°
380x380
Ø400-Ø530
170
40
H* (standard = 1140 / max. 1640)
±190°
500 (stroke)
600
2900* (max 3600)
390**
350**
1600 (recommended)
1720 (max. 2320)
350° ±150°
1420 (max. 1920)
1205**
415 985 90
1490
The PROVEO 8 surgical microscope is intended for operation in an environment as specified below.
The customer or the user of the PROVEO 8 surgical microscope should make sure that it is used in such an environment.
RF emissions in accordance with CISPR 11 Group 1 The PROVEO 8 surgical microscope uses RF energy only for
its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
conducted emissions in accordance with CISPR 11 Class A The PROVEO 8 is suitable for use in establishments other
than domestic and those directly connected to the public
Harmonic emissions according to IEC 61000-3-2 Class A low-voltage power supply network that supplies buildings
used for residential purposes.
Emission of voltage fluctuations/flicker according to IEC 61000-3-3 Complies
The PROVEO 8 surgical microscope is intended for operation in an environment as specified below.
The customer or the user of the PROVEO 8 surgical microscope should make sure that it is used in such an environment.
When the PROVEO 8 surgical microscope is exposed to any of the disturbances below, you might notice one of the following effects:
• flickering/noise on the HD Monitor
• interruptions on the HD Montior
Non of the listed effects above have an impact on the essential performance or safety and effectiveness of the PROVEO 8 surgical microscope. Its no
unacceptable risk for user, patient or environment to be expected.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Discharge of static ± 8 kV contact discharge ± 8 kV contact discharge Floors should be of wood, concrete or ceramic
electricity (ESD) according tile. If floors are covered with synthetic
to IEC 61000-4-2 ± 15 kV air discharge ± 15 kV air discharge material, the relative humidity should be at
least 30 %.
Electrical fast transient/ ± 2 kV for power supply line ± 2 kV for power supply line Mains power quality should be that of a typical
burst immunity according commercial or hospital environment.
to IEC 61000‑4-4 ± 1 kV for input and output ± 1 kV for input and output
lines lines
Surges according to ± 1 kV differential mode ± 1 kV differential mode Mains power quality should be that of a typical
IEC 61000-4-5 commercial or hospital environment.
± 2 kV common mode ± 2 kV common mode
Voltage dips, short <5 % UT (>95 % dip in UT) 70% UT 25/30 cycles Mains power quality should be that of a typical
interruptions and power for ½ cycle commercial or hospital environment.
supply voltage fluctuations 40% UT 10/12 cycles When short interruptions of 5 % UT for
IEC 61000-4-11 40 % UT (60 % dip in UT) 5 seconds occur, the PROVEO 8 surgical
for 5 cycles 40% UT 5/6 cycles microscope will cease operation and restart
automatically. It can be brought back to the
70 % UT (30 % dip in UT) 0% UT 0.5/0.5 cycles state it was before with user intervention.
for 25 cycles If the user of the PROVEO 8 surgical microscope
0% UT 1/1 cycles requires that the instrument remain functional
<5 % UT (>95 % dip in UT) even after power interruptions, it is
for 5 sec 0% UT 250/300 cycles recommended that the PROVEO 8 surgical
microscope be provided with an auxiliary power
source such as an uninterruptible power supply
(UPS) or battery back-up.
The PROVEO 8 surgical microscope is intended for operation in an environment as specified below.
The customer or the user of the PROVEO 8 surgical microscope should make sure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Conducted RF - Disturbance 3/6 Veff 10 Veff d = 0.35 √P for 150 kHz to 80 MHz
variables according to 150 kHz to 80 MHz d = 0.35 √P for 80 MHz to 800 MHz
IEC 61000-4-6 d = 0.7 √P for 800 MHz to 2.7 GHz
Radiated RF - Disturbance 3 V/m 10 V/m Where P is the maximum output power rating of the transmitter in
variables according to 80 MHz to 2.7 GHz 80 MHz to 3 GHz watts (W) according to the transmitter manufacturer and d is the
IEC 61000-4-3 recommended separation distance in meters (m). Field strengths
from fixed RF transmitters, as determined by an electromagnetic site
Proximity fields survey, should be less than the compliance level in each frequency
range. Interference may occur in the vicinity of equipment marked
with the following symbol:
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation amount is affected by absorption and reflection from
structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the PROVEO 8 surgical microscope is used exceeds the applicable RF compliance level above, the PROVEO 8
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the PROVEO 8.
Recommend separation distances between portable and mobile RF telecommunications equipment and the PROVEO 8 surgical
microscope
The PROVEO 8 surgical microscope is intended for operation in an electromagnetic environment in which radiated RF interference is controlled. The
customer or user of the PROVEO 8 surgical microscope can help prevent electromagnetic interference by maintaining the minimum distance between
portable/mobile RF communication equipment (transmitters) and the PROVEO 8 surgical microscope, depending on the output power of the
communication equipment, as stated below.
Rated maximum output 150 kHz up to 80 MHz 80 MHz up to 800 MHz 800 MHz up to 2.5 GHz
power of transmitter in W d = 0.35 √P d = 0.35 √P d = 0.7 √P
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Note 1 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
If use is made of accessories or cables other than those specified in this user manual or approved by the manufacturer of the
PROVEO 8 surgical microscope, this can lead to an increase in electromagnetic radiation or a reduction in EMC.
The PROVEO 8 surgical microscope must not be used directly adjacent other instruments. If it is necessary to operate it in the vicinity
of other instruments, the devices should be monitored to ensure that they function properly in this arrangement.
17 Annex
17.1 Checklist before the operation
Patient ........................................................................................................................................................................................
Surgeon ........................................................................................................................................................................................
Date ........................................................................................................................................................................................
www.leica-microsystems.com