DENTAL IMPLANT MOTOR SYSTEM

OPERATION MANUAL
Please read this Operation Manual carefully before
using, and filing for future reference.

Foshan COXO Medical Instrument Co.,Ltd


Address:BLDG 4,District A,Guangdong New Light Source Industrial Base,
Langsha Luocun,Shishan Town,Naihai District,Foshan City,
Guangdong Province,China
Tel:0086-757-81800058
Fax:0086-757-81800058
E-mail:[email protected]
Http://www.coxotec.com

Wellkang Ltd.
Address:Suite B,29 Harley Street,London W1G9QR,
United Kingdom

Ver2.0 Revision Date:2015-09-30


Thank you for purchasing the COXO C-Sailor dental implant motor system .
1. Safety precautions prior to use
We recommend that prior to use, you read this document carefully regarding instructions for using, handling
method, or maintenance check so that you can carry on using the unit in the future. Read Handpiece’s Operation manual carefully before use.
In addition, please keep this operation manual in a place where a user can refer to it at any given time.
INTENDED TO USE

C-Sailor is intended for use in dental oral surgery and surgical procedures by qualified personnel.
◆Classification of equipment
WARNING
• Type of protection against electric shock :
-Class l equipment The system may present a possibility of malfunction when used in the presence of an electromagnetic interference wave. Do not

• Degree of protection against electric shock : install the system in the vicinity of the device which emits magnetic waves. Turn off the power switch of the Control Unit of this
system when an ultrasonic oscillation device or an electrode knife is located in the vicinity is used.
-Type B applied part
• Method of sterilization or disinfection recommended by the manufacture : CAUTION
-See 7. Cleaning, disinfection, packing and sterilization • C-Sailor needs special precautions regarding EMC and needs to be installed and put into service according to the EMC
• Degree of protection against ingress of water as detailed in the current edition of IEC 60529 : information.
-Foot Control: IPX7 (Protected against the effects of 30 minutes’ immersion in water) • Portable and mobile RF communications equipment can affect C-Sailor. Do not use RF equipment outskirts for the product.

• Mode of operation : • The use of accessories, Motors and cables other than those specified, with the exception of Motors and cables sold by the
manufacturer of C-Sailor as replacement parts for internal components, may result in increased EMISSIONS or decreased
-Intermittent operation(ON 2 min/OFF 10 min)
IMMUNITY of the Control Unit.
-The Implanted is designed for intermittent operating mode with an operating time of 2 minutes and an idle time
• C-Sailor should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the
of 10 minutes. If the operating mode specified is observed no overheating of the system and therefore no injury
Control Unit should be observed to verify normal operation in the configuration in which it will be used.
to the patient, user or third persons arises.
• When operating this system always consider the safety of the patient.
-Not to position the ME equipment so that it is difficult to operate the disconnection device. • Read this Operation Manual before use, and fully understand the functions of each part for operation key.
• Do not attempt to disassemble the Control Unit /Foot Control / Micromotor nor temper with the mechanism.
Indications: • Check for vibration, noise and overheating before use and if any abnormalities are found in use, stop using immediately and
• Missing a tooth, multiple teeth, the teeth of all patients. contact your dealer.
• Mandibular alveolar absorb all full mouth, wearing the traditional full mouth dentures is difficult, is not good, • Use an electrical outlet that is grounded.
wear is not strong person. • Do not drop, hit, or subject to excessive shock.
• Wear traditional dentures fixed differential, non-functional mucosa can not endure. • Do not bend the Irrigation Tube while the water pomp is operating. It could cause tube breakage.
• For dentures aesthetics, function special requirements. • Do not use bent, damaged or sub-standard burs. The Shank could bend or brake.

• Remaining teeth missing teeth around the problem, can not do the traditional dentures. • Do not exceed the recommended speed.
• After each Operation, be sure to lubricate and sterilize the handpiece as soon as it's cleaned. Blood coagulation can cause
Contraindications: corrosion and rusting. However, do not lubricate the Micromotor. Oil could generate excessive heat and cause damage.
• Systemic diseases (cancer, cardiovascular diseases serious diseases, the blood system, the immune system • The Control Unit can be cleaned with a moist cloth. Disconnect the power supply before cleaning. The Control Unit and the Foot
The disease, ...). Control cannot be sterilized by any method.
• Ongoing and topical treatment of certain systems (anticoagulant therapy, chemotherapy, radiotherapy, ...). • When the product is very frequently used, consider the maintenance of a small stock of replaceable parts.
• Poor quantity and quality of bone. • Do not disconnect the Motor Cord from the motor.
• Do not wipe with or, clean or immerse in, high acid water or sterilizing solutions.
Cautions for operation • Follow the table below for the motion time of motor and handpiece. Continuous use over a long time may cause the overheating
■Read these safety cautions thoroughly before use and operate the product properly. of the handpiece, which in turn may result in an accident.
■These indicators are to allow you to use the product safely, prevent danger and harm to you and others. • Applied parts for patient and/or operator is Handpiece.
These are classified by degree of danger, damage and seriousness. All indicators concern safety, be sure to
NOTICE
follow them.
• Turn off the Main Power Switch after each use.
Classification Degree of Danger or Damage and Seriousness
• For service requirements and spare parts contact your dealer.
WARNING Explains an instruction where personal injury or physical damage may occur. • The use of COXO genuine pre-sterilized, disposable Irrigation tube Kit is recommended.
CAUTION Explains an instruction where minor to medium injury or physical damage may occur.

NOTICE Explains an instruction that should be observed for safety reasons.


1 2
Temperature Humidity Atmospheric pressure 3.Control Unit with an Irrigation Pump
Use Between 0-40˚C (32-104˚F) Between10-85%RH Between 700-1060hPa

Store Transportation Between -10-55˚C(14-131˚F) Between 10-85%RH Between 500-1060hPa

** No moisture condensation in the Control Unit. Main Power Switch AC Eletrical Cord
Fuse Holder Connetion socket
** Using at outside of these limits may cause malfunction.
Foot Control Cord connection socket

Coolant Solution Hanger


Post Holder
2. Package Contents

Speed Key Torque Key


Program Key
FWD-REV Key
Gear Ratio Key
Coolant Flow key
Memory Key PROGRAM SPEED TORQUE

P
MEMORY GEAR FWD/REV COOLANT

PROGRAM SPEED TORQUE

MEMORY GEAR FWD/REV COOLANT

Micromotor Cord
Connection socket
1 2 3 4 5

Description of Operation
◆Keys on the Control Unit
(1) Program Key
Use to Cycle through available programs. Press [+] key to ascend and [-] key to descend program number. You
will find all program numbers sequentially by pressing keys.
6 7 8 9 (2) Speed Key
Use to set the Micromotor speed. Press [+] key to increase, and [-] key to decrease speed. (When the upper
and lower speed limits are reached an audible alarm 'beep' is sounded).
Item No. Description Quantity (3) Torque Key
Use to set the torque range. Press [+] key to increase and [-] key to decrease torque range. (When the upper and lower torque
1 Control Unit 1
range limits are reached an audible alarm 'beep' is sounded). The torque range setting should be selected according to the
2 Foot Control with Cord(2M) 1 attached handpiece gear ratio.

3 1 (4) Gear Ratio Key


AC Electrical Cord(1.5M)
Use to set gear ratio of the attached handpiece before use. Repeat press this key until the LCD display indicates the correct gear
4 Optic Motor/Non-Optic Motor(with Motor Cord) 1 ratio of the handpiece.
5 Coolant Solution Hanger Post 1 (5) Coolant Flow Key
Use to select 6 levels of coolant solution flow from 0 to 5, repeat press key to cycle through coolant flow levels.
6 Handpiece Stand 1
(6) FWD-REV Key
7 Tube Holder 8 Use to select the direction of rotation. Press this key once to change the rotational direction. (Once in reverse the Control Unit
audibly warns with a 'beep')
8 Spare Fuse 1
(7) Memorize Key
9 Prevent steam bolt 1 Use to memorize the program parameters set by the operator. Press this key once to memorize parameters. When beep sounds,
the new program parameters have been memorized.

3 4
LCD display on the Control Unit Console
4.Foot Control

Forward/Reverse
Coolant Flow Level Speed

rpm Gear Ratio


Program Number GEAR
Torque Foot Control Cord and Plug

PRG(Program) Button
Pedal Link Indicator N.cm Pedal Progress Bar Coolant Solution Flow
Volume Button P Forward/Reverse Button
0 20 40 60 80 100%

Speed Control Pedal

(1) Coolant Flow Level


Displays the selected coolant solution flow level; the selected Flow level is indicated by 1 to 5 levels of
illuminated indicators. No illumination indicates when the Coolant Solution Flow is off.

Position 0 1 2 3 4 5
(1) Coolant Solution Flow Volume Button
Flowrate 0 50mL 60mL 75mL 85mL 100mL
Use to select 6 levels of coolant solution flow from 0 to 5; each level may be increased by one step pressing
(2) Program Number this button. If you press this button in level 5, it returns to level 0.
Displays the selected program number. (2) PRG (Program) Button
(3) Gear Ratio Use to select the desired program number. Program numbers will always ascend each time this button is
Displays the selected gear ratio of the handpiece. pressed and released.
(4) Forward/Reverse Indicator (3) Speed Control Pedal
Displays the selected direction of the Micromotor. Used to start and stop the Micromotor and to control the speed during operation. The Micromotor operational
(5) Speed speed is directly proportional to the position of the foot control, up to the preset maximum.
Displays the selected speed. (4) Forward/Reverse button
Depressing Foot Control:Rotation speed of the motor displayed on the LCD (Assuming the correct gear ratio Used to change the rotational direction of the Micromotor. Push once to change the rotational direction.
is selected)
Without pressing Foot Control : Max Rotation preset setting displayed on the LCD
(6) Torque: Displays the selected torque.
* When using the 1:1 direct drive or Speed increasing handpiece, the torque is not displayed. (This function is
appropriate for 4:1 increase speed handpiece or more)
(7) Pedal Link Indicator
Displayed when the foot is connected, it is not displayed when the pedal is not connected.

CAUTION

The LCD display panel is produced from liquid crystal and should always be treated with care.

5 6
5. Installation After the tubes are correctly positioned, close the Pump Cover Fig. 6
by turning the Pump Cover Lever to the ‘CLOSE’ position

5-1 Connecting the Motor Cord (180 degrees counter-clockwise).(Fig. 6&7)


Fig. 1
RELEASE CLAMP

Plug the Motor cord into the motor interface of the main unit, fasten
the Fixed nut after that. When remove the motor cord, loosen the
Pump Cover Lever
fixed nut, pull out the plug from the interface.(Fig. 1)

Fig. 7

5-2 Connecting the Foot Control Fig. 2


Face the screw on the Foot Control Cord Plug downward then insert
the plug into the Foot Control Cord connector socket on the Control
Unit. Secure the plug by fastening the Lock Nut.(Fig. 2)
5-5 Mounting the Coolant Solution Hanger Post Fig. 8
Mount the Coolant Solution Hanger Post onto the Holder on the
Control Unit(the support bar maximum can hang 2 kg weight.),
5-3 Connecting the AC Electrical Cord Fig. 3 the post will only fit into one position.(Fig. 8)

Correctly align then insert the AC Electrical Cord into the AC Electrical
Cord Connection at the back of the Control Unit.(Fig. 3)

5-6 Insertion of the irrigation Tube Fig. 9


5-4 Installing the Irrigation Tube Fig. 4 Fixed
Position (1) Close the Tube Clamp, between the Irrigation Tube Needle and
Please install the water supply pipe onto the side of the Tube Guide Tube clamp
the Irrigation Pump.(Fig. 9)
main unit. Then please send water supply pipe needle side
(2) Insert the Irrigation Tube Needle into the Bottle Cap.(Fig. 10)
of the clamp ring alignment host back in,it will be another
Fixed ‘Needle’ (3) Open the Tube Cap to supply air into the bottle.
clamp ring alignment water supply pipe guide of the load. Position Direction
(4) Open the Tube Clamp.
(Fig. 4) ‘Control Unit’
Stopper Direction Fig. 10
Bottle Cap
Irrigation
Tube Needle
CAUTION
Make sure that the tube is securely set on the rollers Fig. 5
when closing the Pump Cover. If the tube is not correctly
positioned on the Rollers and the cover is closed, the Rollers
tube could be cut or sheared.(Fig. 5) CAUTION

Do not operate the Irrigation Pump if the tube is bent or the Tube Clamp is in the closed position. This could cause the tube to
burst or slip out of the bottle.

7 8
5-7 Compatibility check of Internal Irrigation Nozzle/Drill
6. Operation
Internal irrigation nozzles accompanied with this product; is not necessarily fitted into all the drills on the
market. Follow the instructions given below for confirmation prior to use.Failure to do so or to fit the internal
6-1 Programming the Micromotor Operation
irrigation nozzle into drills may cause a leakage of saline solution, which will result in problems such as rust or
The Control Unit can memorize 10 programs . Each program includes the following functions which will be
sudden stop of equipment during use.
automatically performed when the appropriate program number is selected.

Instructions:
◇ Gear ratio of contra angle handpieces
(1) Attach a bottle of saline solution to the Control Unit. Fig. 11
◇ Speed
Internal Irrigation Nozzle
(2) Connect the Internal Irrigation Nozzle into the tip of the irrigation tube.
◇ Direction of rotation
(3) Insert the Internal Irrigation Nozzle into the drill from the back.(Fig. 11) Irrigation Tube
Drill ◇ Torque upper limit
(4) Purge at “Maximum” for 5 seconds.
◇ Coolant Solution Flow

Points to be checked : (1) Turn on the power by pushing the Main Power Switch toward [ ]; Power Switch
Symbol Mark
◇ Cleanliness of the saline solution coming out from the drill; if solution is colored there could be rust inside on power up program #1 is displayed by default.
FUNCTION OFF ON
of the drill. If so renew the drill. (2) Select a program number by using either step (a) or step (b):
◇ Water Flow; if the flow is low and/or the flow from the drill is asymmetric, renew the drill. (a) Press the [Program] Key on the Control Panel until the desired program number is displayed.
◇ No water leakage between Internal Irrigation Nozzle and drill. (b) Press the [Program] button on the Foot Control until the desired program number is displayed.
Before use, ensure no water is leaking from the entry point of irrigation nozzle, a broken seal or no seal in (3) Selecting the Gear Ratio of the handpiece relevant to the program; Press the [Gear Ratio] Key to select the
the drill.may be the cause. Replace the drill even if its new, saline solution ingress into handpiece will gear ratio of the handpiece(Gear Ratio will display on the LCD).
cause malfunction. (4) Set the required max operating speed by pressing the [Speed] Key. Each time this Key is pressed display
changes to the next speed level. By pressing this Key for more than 1 second brings the speed quickly to the
CAUTION next level until the speed display reaches its upper or lower limit.
If malfunction such as a leakage of saline solution from the back of Contra Head is detected during use, stop using and perform - When the speed setting reaches the upper or the lower limit, an audible beep is heard and the speed setting
some troubleshooting. cannot be changed any further.
(5) Set the torque upper limit by pressing the [Torque] Key on the Control Panel. Each time this Key is pressed
5-8 Irrigation Nozzle Attachment display changes to the next torque level. By pressing this Key for more than 1 second brings the torque
C6-9/C6-19 has 1 irrigation method available depending on tool and application; External, Internal or both, quickly to the next level until the torque display reaches its upper or lower limit.
For installation detail, refer to the handpiece Operation Manual. - When the torque setting reaches the upper or the lower limit, an audible beep is heard and the torque
setting cannot be changed any further.

5-9 Attaching the Tube Holder (6) Set the rate of the Coolant Solution Flow volume by pressing the [Coolant Flow] Key. The rate of Coolant
Fig. 12
Tube Holder Solution Flow volume has 6 Flow rates (0-5) (0 = no coolant Flow).
Use the Motor Cord as a strain relief for the 'Irrigation Tube'. It is Motor Cord

easier to insert Motor Cord first, then the Irrigation Tube.(Fig. 12) (7) Memorize setting; after completing steps 2 - 6 press and hold [Memorize] key until beep is heard. The beep
Irrigation Tube confirms that the programming is completed. If you hear a short beep when the [Memorize] Key is first
pressed ignore this sound and keep the [Memorize] Key depressed until a long beep is heard.
* Repeat the above steps 2 - 7 to program any one of the 10 available programs.

9 10
6-2 Standard Operation
7. Cleaning, disinfection, packing and sterilization
All standard operational functions can be control at the Foot Control.
(1) Turn on the Main Power Switch: The Control Unit is ready to perform the Program. 7-1 Manual cleaning
(2) Select the desired program number: Step on the Foot Control PRG (Program) Button and the program • Use the softened water(< 38 °C) and brush to clean the 1# samples.
display ascends to the next program number. Pressing the PRG (Program) Button for one second more will a. Surface cleaning
descend the to the next program number. • Use the softened water(< 38 °C) and brush to clean the surface of motor and line until visually clean.
b. Crevices and cavities cleaning
(3) Verify the details of the program on the display.
• Use the softened water(< 38 °C) and brush to brush E type connector screw, the gap part, the connection gap
(4) Operating the Micromotor : Step on the speed control pedal in the middle of the Foot Control the Micromotor
of the motor rear part, pay particular attention to all crevices and cavities. Repeat several times until visually
will start to run. The Coolant Pump will also run (if programmed). Speed increases as the pedal is depressed. clean. Then Use the water absorbent cloth to remove any remnant of liquid.
When the Speed Control Pedal is fully depressed the speed reaches the maximum preset value.
(5) Activation of the torque limiter: During use, when the drilling load reaches the preset torque upper limit, the 7-2 Manual disinfection
integrated torque limiter automatically activates to prevent over torque. When the torque limiter activates, Use KaVo Cavicide disinfectant liquid to disinfect the test samples.
the motor stops after beeping for 1 second. To reactivate the Micromotor, release the speed control pedal
7-3 Packing
and depress it again.
Note:
(6) Stopping the Micromotor: Release the Foot Control Pedal, and the Micromotor will automatically stop.
▶The quality and use of the sterilisation packaging must comply with applicable standards
(7) Reversing the Micromotor rotational direction: To reverse direction of the Micromotor (and bur) simply step
and be suitable for the sterilisation procedure!
on the Foot Control Forward/Reverse button. A warning beep can be heard when the rotational direction is
▶If potentially infectious liquids and particles can contact the products, it is recommended
in reverse mode.
to cover and protect these areas with sterile disposable products.

Seal the handpiece tray and motor cable in a sterilisation pouch.

7-4 Sterilization
• Autoclave sterilization is recommended.
• Autoclave sterilization is required for the first time use and after each patient as noted below.
• Autoclave sterilizatio holding time15 minutes at 121 °C.
• Attach the Autoclave Plug to the Micromotor.(Fig. 13&14)
• Place those in autoclave pouch (not included in the package) and seal it.
The following items can be autoclaved.
• Implant Handpiece
• Micromotor with Motor Cord (Including the Motor Housing)
图13
• Handpiece Stand
• Internal Irrigation Nozzle
• Tube Holder
• Nozzle Holder
• Autoclave Plug
• Calibration Bur
图14

11 12
8-3 Fuse Replacement
8. Care and Maintenance
If the Control Unit does not function, check the fuses ( Fuse Box
Fig. 15
lock located on the rear of the Control Unit ). To access the Fuse,
8-1 Protection Circuit Fuse
use a pointed tool push on the fuse locking latch and the drawer
An electronic circuit breaker automatically functions to protect the Micromotor and the Control Unit if the will spring open.(Fig. 15)
Internal fuse replacement:Tear down the host at the bottom of the
Micromotor is ever overloaded. Power supply to the Micromotor will automatically be terminated an Error code screw with a screwdriver tool to take out the safety device can
will be displayed on the Control Unit. replace the internal fuse.
Fuse Ratings
Fuse box
120V F3AL 250V
◆Resetting the Protection Circuit 230V F2AL 250V

To reset the Protection Circuit, release and then depress the Speed Control Pedal.
8-4 Maintenance of the Control Unit and Foot Control
If blood or saline solution has stained the Control Unit or Foot Control, remove the AC Electrical Cord; wipe
8-2 Malfunction and Remedy
Control Unit or Foot Control with a damp cloth, then with an alcohol-absorbed cloth.
If the device is not working properly, please check the following table before calling our service center.

Malfunction Cause of Malfunction Remedy 8-5 Maintenance of the Handpiece Attachment


• Memory components malfunction Every time after treatment, please put the head of the handpiece into the clean water and rotate for 4-5 seconds,
System malfunction Please contact with the seller
• Memory components damaged to clean the blood and physiological saline. If the attachments on the handpiece is too much, use the dry and soft
• Long time use under overload (over current) rag to wipe them out. Please do not put the whole handpiece into the water, in case of the water flow into the
Over current • Power cord short circuit
• Motor coil short circuit handpiece from behind.
Probably the circuit is bad connected.Please
Over voltage • Main power cord broken connect the motor line properly. If still no
improvement,please contact with the seller.
• Motor sensor malfunction [after use, do not forget the lubrication before sterilizing under the high temperature and high pressure.]
• Motor cord not been connected
• Motor line (signal line) malfunction As shown in Fig. 16, put the nozzle into the handpiece behind, spray for more than 2 seconds.
Motor sensor
malfunction Please contact with the dealers.(in high pressure
• Water flow in to the motor high temperature sterilization,please be sure to Fig. 16
install the anti-steam dossil.
• Due to the long time use under overload, the Please wait until the temperature is cool down
The main unit temperature of the main unit arise. before using.Please place it in the environment
inside overheat • To use the unit under the high temperature that easy to cool down. If still no improvement,
environment (direct sunlight) please contact with the seller.
In a short time repeated operation and stop,
• Abnormal voltage appears the start and stop. running, inhibit suddenly accelerated circuit may
Braking device NOTICE
• Start and stop circuit malfunction be start.Please wait for 1~2 seconds before
malfunction
operation. If still no improvement, please contact When lubricating, please hold the hadnpiece tight, in case of the handpiece fly off due to the force of the spray air.
with the seller. In order to make the oil and gas completely mix, please shake the tank for some time.
Probably the chuck is opened or not fully closed. When lubricating, please be sure to clean up the lubrication oil tank.
• Malfunction of the handpiece
If still no improvement, please contact with the
Motor running • Malfunction of the motor
seller. If the head of the handpiece attach too much blood and dirt, need to be separately cleaned and refueled. Install
malfunction
• When the Torque reaches its limit, the stop Not malfunction. It is for the safety. It can still be the clean nozzle to it, and wash directly from inside(Fig. 17).
process lasts formore than 5 second. used after connect.
If the water supply nozzle is stuck, please use the needle to clean it(Fig. 18).
Please confirm the state of water supply pipe. If it
Water supply pump • Water supply pipe not been fully attached to the
overvoltage right place of the water supply pump
is normal, but still appear the malfunction signal. If there is dirt around the nozzle, please use brush to clean it(Fig. 19).
please contact with the seller.
Water supply pump • Water supply pump malfunction please contact with the seller. Fig. 17 Fig. 18 Fig. 19
over current
• Bad connection Please confirm connection status
Foot control abnormal If the foot control is in normal, but still show the
• Foot control inside malfunction malfunction number, please contact with the seller.

13 14
9. Specifications 11. Disposing Product

Consult with dealer from whom you purchased it about waste disposal.
9-1 Control Unit
Model C-Sailor
Power Supply Voltage AC230V
12. Symbols
Frequency 50Hz
Power Consumption 120VA
Max. Pump Output 100mL / min
The EU directive 93/42/EEC was applied in the design and production of this medical device.
Dimensions W285 x D265 x H155mm

9-2 Micromotor Protected against the effects of 30 minutes’ immersion in water.


7

Speed Range 300-50,000r/min


Dispose of this device and its accessories via methods approved for electronic device and in
Input Voltage DC30V
Dimensions Ø27 x L123mm
compliance with the Directive 2002/96/CE.

9-3 Foot Control


Code length 1.77m Type B applied part See Operation Manual

9-4 Handpiece(Purchased separately)

Model C6-9 C6-19 This way up Used indoor only


Max. Rotation Speed 2,500r/min
Gear Ratio 20:1 Reduction
Spray Type External, Internal* Fragile,handle with care Keep away from rain

Fit ISO 3964 20:1 and IEC 80601-2-60 Standard requirement Implant contra angle
Torque: more than 55 N.cm Foot pedal connector Warning
Do not exceed the rotation speed that bur manufacture recommends. Adjust the rotation speed which bur
manufacture recommend, if you use the allowable rotation speed is less than 120,000rmin-1.
ON(power connection) OFF(power disconnection)
* If you use internal irrigation system drills.

9-5 Infusion Tube


User can purchase separately from hospitals ,pharmacies and W&H Dentalwerk Bürmoos GmbH company. Alternating Current Electric fuse
(REF 04363600) .

9-6 The Cooling Liquid European Union agent CE mark


Physiological saline(Internal and external cooling) (concentration:0.9%; flow velocity:0~100 ML/min)

Date of manufacture Manufacturers

10. Warranty

The warranty of main unit and foot pedals are 12 months,motor is 6 months, other accessories are not included
in the warranty.
If necessary, we can provide circuit diagram, components list,describes, calibration instructions and so on.