Adsorber
Adsorber
TECHNICAL INSTRUCTION
F-50H
COMPRESSOR UNIT
E-mail: [email protected]
URL: http://www.shicryogenics.com
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
TABLE OF CONTENTS
TABLE OF CONTENTS
CROSS REFERANCE 1
1 GENERAL INFORMATION 2
1-1 SPECIFICATIONS 2
1-2 CONSTRUCTION 6
1-2-1 CONTROLS AND COUPLINGS 6
1-2-2 GAS AND OIL FLOW IN THE COMPRESSOR UNIT 9
1-2-3 INTERNAL COMPONENTS 11
1-3 ELECTRICAL DESCRIPTION 13
1-3-1 DIAGNOSTIC INTERFACE 13
1-3-2 SAFETY DEVICES 14
2 INSTALLATION 15
2-1 SITE REQUIREMENT 15
2-2 INPUT POWER VOLTAGE SETTING 17
2-3 INPUT POWER CABLE CONNECTION 20
3 MAINTENANCE 22
3-1 PERIODICAL MAINTENANCE 23
3-1-1 REPLACEMENT OF THE COMPRESSOR ADSORBER 24
3-1-2 RESET THE CIRCUIT PROTECTOR(S) 29
APPENDIX 30
ELECTRICAL SCHEMATIC
DRAWINGS
REVISION CONTROL
F-50H
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
CROSS REFERENCE
CROSS REFERENCE
Thoroughly read this manual and following manuals before using this equipment.
F-50H 1
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
1-1 SPECIFICATIONS
1 GENERAL INFORMATION
1-1 SPECIFICATIONS
The specifications of F-50H Helium Compressor Unit are summarized in Table 1.1.
Operating (High Side)** 1.90 - 2.20 MPa --- approx. 1.90 - 2.20 MPa --- approx.
(Reference) (276 - 319 psig) (276 - 319 psig)
Operating (Low Side ) ** 0.40 - 0.60 MPa --- approx. 0.40 - 0.60 MPa --- approx.
(Reference) (58.0 - 87.0 psig) (58.0 - 87.0 psig)
Ambient Operating 5 to 28 deg.C (41 to 82.4 deg.F)
Temperature 28 to 35 deg.C (82.4 to 95 deg.F) with 5% Capacity Loss
Mass 120 kg (264 LBS) --- approx.
Electrical Requirement
Power Line Voltage AC 380, 400, 415 V / 50 Hz, 3 phase (3W+PE)
(+/-10%) AC 460, 480 V / 60Hz, 3 phase (3W+PE)
(Δground, Commercial Power Source)
“WARNING”
Do not use inverter for the main power source.
F-50H 2
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
1-1 SPECIFICATIONS
The typical flow characteristics are shown in Figure 1.1, and cooling water requirement are shown in
Table 1.2.
For Water
11 3.0
Cooling Water Flow Rate
[gallon/min.]
Admissible Range Admissible Range
[liter/min.]
8 2.0
7
6 1.5
5
1.0
4
3 0.5
0 4 8 12 16 20 24 28 32 30 40 50 60 70 80 90
Cooling Water Inlet Temperature [deg.C] Cooling Water Inlet Temperature [deg.F]
10
2.5
9
Admissible Range
[gallon/min.]
Admissible Range
[liter/min.]
8
2.0
7
6 1.5
5
1.0
4
3 0.5
0 4 8 12 16 20 24 28 32 30 40 50 60 70 80 90
Cooling Water Inlet Temperature [deg.C] Cooling Water Inlet Temperature [deg.F]
0.09
Cooling Water Outlet-Inlet [deg.C]
35 50Hz
Pressure Drop [MPa]
0.08 60Hz
30
0.07
25
0.06
0.05 20
0.04 15
0.03
10
0.02
0.01 5
0.00 0
4 5 6 7 8 9 10 4 5 6 7 8 9 10
Water Flow [liter/min] Water Flow [liter/min]
F-50H 3
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
1-1 SPECIFICATIONS
F-50H 4
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
1-2 CONSTRUCTION
485 (19.1″)
450(17.72″)
GAS
DIAGNOSTIC INTERFACE SUPPLY
MAIN POWER
519(20.43″)
HIG H P RES SUR E
GA S BY PA SS
Illu minates ;Close d
OV ER T EMP -ER ROR
GAS
590.5 (23.24″)
[H eliu m D isc har ge]
OV ER T EMP -ER ROR RETURN
[H eliu m A fte r C oole r]
OV ER T EMP -ER ROR
[W ater Ou tle t]
( For
Onl y
Mai nt ena nc e
) SUPPLY
PRESSURE
COLD HE AD R EMO TE DR I V E
D R IV E DRIVE C OM P RE S SO R U NI T
ON
ON LOCAL
COLD HEAD
OFF REMOTE
RETURN
CRYOCOOLER
PRESSURE
WATER HELIUM
OUTLET GAS
CHARGE
WATER
INLET
SU MIT OMO HEA VY IND UST RIES ,LT D.
CRY OGE NICS DE PAR TME NT
2 -1-1 ,YAT O-CH O,NI SHIT OKYO -CIT Y,
TO KYO, JAPA N
(2.82″) 71.5
450(17.72″) 103 485(19.1″)
(4.06″)
588(23.15″)
XRE05B0249BQ
F-50H 5
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
1-2 CONSTRUCTION
1-2 CONSTRUCTION
the outline view for F-50H Compressor Unit is shown in Figure 1.2.
The function of the Compressor Unit is to supply high pressure He gas to the Cold Head and
re-compress the returned He gas from the Cold Head. The Compressor Unit consists of the following
major components: a Compressor Capsule, a Cooling system, Oil separation and injection system, and
Adsorber.
F-50H 6
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
1-2 CONSTRUCTION
To use for the external signal output of condition monitoring for the
compressor unit. The connector to be “D-Sub 15 Pins (Female type)” in
use. Locking Nut is #4-40UNC.
DIAGNOSTIC INTERFACE : Warning; Pay special attention to its wiring when using the external connector on
15
(JR) the Compressor Unit.
Connecting a jumper wire between Pins No.6 - No.8, No.6 - No.13 and
No.6 - No.15 may result in misoperation in some of safety devices in
the equipment, causing electric shock, burn or malfunction.
A connector for the earth wiring, and verify the tight connecting for earth
16 GROUND TERMINAL : (PE)
wiring as well as Input Power Cable.
To use for connecting of input power supply cable.
At a connecting power cable, verify the phase label markings L1, L2 and
17 FIELD TERMINAL : (TB0)
L3. The compressor unit cannot be operated in case of miss-connecting
the power cable.
CIRCUIT PROTECTOR
18 A circuit protector for protection from over-current and short circuit.
(CP1 , CP2)
COLD HEAD CONNECTOR : To use for connecting the Cold Head Cable to supply a Cold Head
19
(JC) driving power.
TERMINAL BLOCK
20 To be used for input voltage adjustment.
(TB1)
F-50H 7
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
1-2 CONSTRUCTION
1
GAS
2
DIAGNOSTIC INTERFACE SU PPLY
MAIN POWER H IG H P R ES SU R E
G A S BY PA S S
I ll um ina te s; Clo se d
O V ER T E MP - ER RO R
[ H el iu m D i sc ha r ge ]
GAS
O V ER T E MP - ER RO R R ETUR N
[ H el iu m A f te r C oo l er ]
O V ER T E MP - ER RO R
[ W at er Ou t le t]
CP1 CP2 F IE LD WI R IN G D IA G RA M
7 (
F or
O nl y
COLD HEAD
DRIVE
M a in t en a n ce
ON
)
REMOTE
DRIVE
LOCAL
DRIVE
ON
SUPPLY
PRESSURE
COMP RES SOR UNI T
3
COLD HEAD
OFF REMOTE
CRYOCOOLER
RETURN
PRESSURE
6 WA TER
WATER
OUTL ET
HELIUM
GAS
CHARGE
4
I NLE T
S UM I TO M O HE A VY I N DU S TR IE S ,L T D.
C RY O GE NI C S D EP AR T ME N T
2- 1 -1 ,Y AT O- C HO ,N IS HI T OK YO -C IT Y ,
T OK YO ,J A PA N
15 DIAGNOSTIC INTERFACE 9
MAIN POWER HIGH PRESSURE
17
GAS BY PASS
Illuminates;Closed
11
OVER TEMP-ERROR
L1 L2 L3 16 14
10
[Helium Discharge]
OVER TEMP-ERROR
[Helium After Cooler]
OVER TEMP-ERROR
[Water Outlet]
11
16 CP1 CP2
ELAPSED TIME
0.63 0.63
0.8 0.8
1 1
CP1 CP2 (
For Maintenance
)
12
18
BM3RHB-0 01 BM3RHB-001 Only
COLD HEAD REMOTE DRIVE
DRIVE DRIVE
ON
ON LOCAL
COLD HEAD
OFF REMOTE
19 380 400 415 460 480 380 400 415 460 480 380 400 415 460 480
13
20 14
XRE05B0239BQ
F-50H 8
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
1-2 CONSTRUCTION
The flow diagram for F-50H Compressor Unit is shown in Figure 1.4.
Internal components diagram and its functions are described in Figure 1.5 and Table 1.4.
1) Low pressure He gas (Pressure.: below 0.80 MPa (116psig)) discharged from a Cold Head can be
led through a HE-GAS RETURN CONNECTOR to the Compressor Unit.
2) The low pressure (Return) He gas can pass through a STORAGE TANK and a FILTER, and flow
into a COMPRESSOR CAPSULE.
3) The low pressure He gas will be compressed and pressurized in the COMPRESSOR CAPSULE,
and the high pressure with high temperature He gas after the compression will be discharged from
the COMPRESSOR CAPSULE outlet.
4) The high pressure with high temperature He gas will be led to a water cooled HE-GAS COOLER
and cooled down in the cooler.
5) The high pressure He gas after cooling will flow into an OIL SEPARATOR to separate an almost all
of lubricating oil mist from the high pressure He gas.
6) The separated lubricating oil can be returned to the COMPRESSOR CAPSULE through a lub oil
return pipings.
7) The high pressure He gas discharged from the OIL SEPARATOR will be led to an ADSORBER.
8) The remained lub oil contents in the high pressure He gas can be adsorbed through an active
charcoal layer to make the high pressure He gas being pure.
9) The pure high pressure He gas can be supplied to the Cold Head through a HE-GAS SUPPLY
CONNECTOR.
F-50H 9
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
1-2 CONSTRUCTION
8 16 24 21 20 6 19 18 12 13 17 11 15 2
PSL
FROM
COLD
HEAD
5
PSH
TS1 4
TS3
TO
COLD
HEAD
TS2 1
10
OIL LINE
23 26 25 7 22 9 14 He-GAS LINE
OUTLET INLET
COOLING WATER
XRE05B0226AQ
F-50H 10
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
1-2 CONSTRUCTION
The parts list and its functions are described in Table 1.4.
The He-gas flow diagram and internal components are shown in Figure 1.4 and Figure 1.5.
F-50H 11
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
1-2 CONSTRUCTION
19 14 18 9 11 21 7 8 23
20
16
24 VIEW A
27 22 10 12 13
1
DIAGNOSTIC INTERFACE
MAIN POWER
16 14
11
HIGH PRESSURE
GAS BY PASS
Illuminates;Closed
OVER TEMP-ERROR
[Helium Discharge]
OVER TEMP-ERROR
[Helium After Cooler]
OVER TEMP-ERROR
[Water Outlet]
2
CP 1 CP 2
15
ELAPSED TIME
( For
Only
Maint enance
)
COLD HEAD REMOTE DRIVE
DRIVE DRIVE
ON
ON LOCAL
COLD HEAD
OFF REMOTE
3
A
4
25 26 5 17
XRE05B0242BQ
F-50H 12
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
This cryocooler includes a high-voltage section. Touching it may result in electric shock.
Handle it with extreme care.
“IMPORTANT”
See the “ELECTRICAL SCHEMATIC” of F-50H Compressor Unit, for detail.
Diagnostic Interface can be used monitoring the status of the Compressor Unit and the remote control
sequences of the Compressor Unit are described in Table 1.5.
The “D-sub” pins indicated in Figure 1.6 on the control panel for the Compressor Unit can be applied to
an initial condition monitoring for a first-aid diagnostics of the Compressor Unit by means of measuring
the each item with a digital Volt/Ohm Meter. The Fault Condition classified the digital meter reading as
referred to the Table 1.5 can be identified simply an actual operation condition of the Compressor Unit
in the field.
15 14 13 12 11 10 9
F-50H 13
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
The safety devices list for Compressor Unit is shown in Table 1.6.
ITEM FUNCTIONS
Setting temperature; 90 deg.C (194 deg.F) ---- approx.
To shut down the Compressor Unit and illuminate the OVER TEMP-ERROR lamp
THERMOSTAT : (TS1) and signal a high temperature alarm to the DIAGNOSTIC INTERFACE, in case of
higher temperature of a compressed He-gas at a compressor outlet than the
setting temperature.
Setting temperature; 60 deg.C (140 deg.F) ---- approx.
To shut down the Compressor Unit and illuminate the OVER TEMP-ERROR lamp
THERMOSTAT : (TS2) and signal a high temperature alarm to the DIAGNOSTIC INTERFACE, in case of
higher temperature of a compressed He-gas at a He-gas cooler outlet than the
setting temperature.
Setting temperature; 60 deg.C (140 deg.F) ---- approx.
To shut down the Compressor Unit and illuminate the OVER TEMP-ERROR lamp
THERMOSTAT : (TS3) and signal a higher temperature alarm to the DIAGNOSTIC INTERFACE, in case
of higher temperature of a water at a cooling water outlet than the setting
temperature.
To stabilize a pressure for even of the He-gas between the Supply and Return
SOLENOID VALVE : (SV1)
piping, at a shut off the Compressor Unit.
Setting pressure;
“Operate” 2.55 MPa ---- approx.
(369 psig)
HIGH PRESSURE SWITCH : “Reset” 2.26 MPa ---- approx.
(PSH) (327 psig)
To adjust a Supply He-gas pressure smoothly by a function of the pressure switch
for Open and/or Shut, in case of higher pressure of the Supply He-gas than the
setting pressure.
Setting Pressure;
“Operate” 0.15 MPa ---- approx.
LOW PRESSURE SWITCH : (22 psig)
(PSL) To shut down the Compressor Unit and signal a Low pressure alarm to the
External Connector, in case of lower pressure of a compressed He-gas caused by
a smaller quantity of He-gas than original filling in the compressor unit.
Setting pressure;
“Operate” 2.61 - 2.75 MPa
(378 - 398 psig)
“Reset” 2.50 MPa ---- minimum
RELIEF VALVE
(362 psig)
To adjust a Supply He-gas pressure smoothly by a function of the Relief Valve for
blowing off the He-gas to the atmosphere, in case of higher pressure of Supply
He-gas than the setting pressure.
Setting current; 16 A
MAIN POWER SWITCH :
To shut down the Compressor Unit, in case of occurring over-current and/or
(QF1)
short-circuit than the setting current.
PHASE FAILURE To avoid starting-up of the Compressor Unit in case of an abnormal operation
PROTECTION CIRCUIT : caused by irregular connecting of Input Power Cable such as failure connecting.
Setting current; 0.63 A (Both CP1 and CP2)
CIRCUIT PROTECTOR
To protect the Compressor Unit from the over-load caused by short-circuit and/or
(CP1, CP2)
any other electrical failure in the DC power or the Solenoid Valve.
F-50H 14
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
2 INSTALLATION
Avoid using the transformer for the main power source of the compressor unit. If the
transformer is installed at the upstream of the unit, lacking phase protection circuit with the
cryocooler occurs in a malfunction. That may result in misoperation or malfunction. When
using the transformer, install the other lacking phase protection device in upstream of the
transformer.
Secure enough space around the compressor unit for heat radiation and maintenance.
Failing to secure enough space may result in misoperation or malfunction.
Use cooling water with appropriate temperature, flow rate and water quality. Using
inappropriate cooling water may result in misoperation or malfunction.
・ An almost level and even area in the field will be selected to install the Compressor Unit.
・ An area to be influenced by splashing water and/or dusts will not be selected to install the
Compressor Unit installation area.
・ A clean environmental condition without dirt and/or free from an exhausted heat will be selected to
install the Compressor Unit in the field.
・ A quality of cooling water will be secured to use for an appropriate coolant for the Compressor Unit.
・ Any heat sensitive object cannot be positioned on surroundings of the Compressor Unit.
F-50H 15
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
(6″)
153
485 (19.1″)
153 153
(6″) (6″)
(11.8″) 300
MAINTENANCE SPACE
250(9.84″)
GAS
DIAGNOSTIC INTERFACE SU PP LY
MAIN POWER
590.5(23.24″)
H IG H P RE SSU RE
GAS B Y P AS S
Illu minat es;C lose d
EL APS ED T IME
( F or
O nl y
M ai nt ena nc e
) SUPPLY
PRESSURE
COLD HEA D R EM OT E DRIVE
DRIVE DRI VE C OM P R E S SO R U N IT
ON
ON LOCAL
COLD HEAD
OFF REMOT E
CRYOCOOLER
RETURN
PRESSURE
WATER HELIUM
OUTLET GAS
CHARGE
WATER
INLET
SU MI TOM O HEA VY IN DUS TR IES ,L TD.
C RYO GE NIC S DEP AR TME NT
2-1 -1, YAT O-C HO, NIS HIT OKY O-C ITY ,
T OKY O,J APA N
XRE05B0228BQ
F-50H 16
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
For F-50H compressor unit (water cooled, high voltage type), pay attention to the setting of
the applicable input supply voltage. The product is shipped with the input supply voltage set
to 480V. Before installing the equipment, be sure to check your supply voltage and change it
to the appropriate setting if necessary. Operating the equipment with your supply voltage
different from the setting of the compressor unit may result in electric shock or malfunction.
Be sure to turn off and Lock Out with OFF position the main power of the customer's power
source before connecting or disconnecting the input power cable to the Compressor Unit, and
then remove the input power cable from the main power. Failing to observe this precaution
may result in electric shock.
Do not change the setting of the dial above the main power switch of the compressor unit
under any circumstances. Failing to observe this precaution may result in electric shock.
Avoid using the transformer for the main power source of the compressor unit. If the
transformer is installed at the upstream of the unit, lacking phase protection circuit with the
cryocooler occurs in a malfunction. That may result in misoperation or malfunction. When
using the transformer, install the other lacking phase protection device in upstream of the
transformer.
The Compressor Unit can be operated on various input power voltages by changing the terminal
wiring in the Compressor Unit. This terminal is located inside of the Control Box.
Initial factory setting is AC480V / 60Hz.
1. Loosen the screws that hold the terminal cover and the remove the cover.
F-50H 17
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
3. Change the wires (1L1, 1L2 and 1L3) to the appropriate setting. Tightening Torque: 1.2 N・m.
Reinstall the plastic cover.
2
1
3
1L
1L
1L
480V AT 60Hz
3
1
1L
1L
1L
460V AT 60Hz
3
2
1L
1L
1L
415V AT 50Hz
3
1L
1L
1L
400V AT 50Hz
3
1L
1L
1L
380V AT 50Hz
XRE05B0245AQ
F-50H 19
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
For F-50H compressor unit (water cooled, high voltage type), pay attention to the setting of
the applicable input supply voltage. The product is shipped with the input supply voltage set
to 480V. Before installing the equipment, be sure to check your supply voltage and change it
to the appropriate setting if necessary. Operating the equipment with your supply voltage
different from the setting of the compressor unit may result in electric shock or malfunction.
Be sure to turn off and Lock Out with OFF position the main power of the customer's power
source before connecting or disconnecting the input power cable to the Compressor Unit, and
then remove the input power cable from the main power. Failing to observe this precaution
may result in electric shock.
Do not change the setting of the dial above the main power switch of the compressor unit
under any circumstances. Failing to observe this precaution may result in electric shock.
Avoid using the transformer for the main power source of the compressor unit. If the
transformer is installed at the upstream of the unit, lacking phase protection circuit with the
cryocooler occurs in a malfunction. That may result in misoperation or malfunction. When
using the transformer, install the other lacking phase protection device in upstream of the
transformer.
“IMPORTANT”
This cryocooler is provided with a phase reverse protection circuit for the input power. If the input
power is connected with reverse phase, the cryocooler does not start.
“IMPORTANT”
See “ELECTRICAL SCHEMATIC” of F-50H Compressor Unit, for detail.
“IMPORTANT”
See “INPUT POWER CABLE ( HIGH VOLTAGE) ” of “APPENDIX” for detail.
Upstream Protection
Use the fuses or circuit breakers as upstream protection of L1, L2, L3. The recommended rating of
the protection is maximum 30A.
F-50H 20
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
See the Table 1.1 for power requirements. The cables are marked with label and connect as follows:
For
Compressor Unit
WIRING DIAGRAM
FIELD
TERMINAL GROUNDING
TERMINAL
For User's L1 L2 L3
Power Source
L1
L2
L3
Ground
1. Loosen the screws that hold the terminal cover and the remove the cover.
F-50H 21
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
3 MAINTENANCE
3 MAINTENANCE
Make sure no power is applied to the compressor unit before starting operation when
connecting or disconnecting the cold head power cable. Failing to observe this precaution
may result in electric shock.
Be sure to turn off and Lock Out with OFF position the main power of the customer's power
source before connecting or disconnecting the input power cable to the Compressor Unit, and
then remove the input power cable from the main power. Failing to observe this precaution
may result in electric shock.
Do not change the setting of the dial above the main power switch of the compressor unit
under any circumstances. Failing to observe this precaution may result in electric shock.
Be sure to turn off and Lock Out with OFF position the customer's main power before
performing maintenance work such as reset of circuit protectors. Failing to observe this
precaution may result in electric shock.
Do not disassemble the equipment for purposes other than maintenance specified in this
service manual under any circumstances. Disassembling the equipment may result in
electric shock, explosion or escape of gas.
The cold head, compressor unit, compressor adsorber and flex lines are pressurized with
helium gas. Purge the helium gas from all pressurized components before disposing. Open
the purging valve gradually or it may result in serious injury.
Install the cryocooler in the ventilated area, otherwise it may result in asphyxiation in case the
helium gas is purged or leaked.
Secure enough space around the compressor unit for heat radiation and maintenance.
Failing to secure enough space may result in misoperation or malfunction.
Use cooling water with appropriate temperature, flow rate and water quality. Using
inappropriate cooling water may result in misoperation or malfunction.
The cold head drive switch provided with the compressor unit is only used for maintenance.
Be sure to turn it OFF in normal operation. Using the compressor unit with the cold head
drive switch turned ON may result in misoperation or malfunction.
F-50H 22
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
F-50H Compressor Unit is to be required the routine maintenance. The basic maintenance work is to
replace the oil mist Adsorber of the Compressor Unit for every 30,000 Hrs. operation as mentioned
Table 3.1.
F-50H 23
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
Do not disassemble the equipment for purposes other than maintenance specified in this
service manual under any circumstances. Disassembling the equipment may result in
electric shock, explosion or escape of gas.
The cold head, compressor unit, compressor adsorber and flex lines are pressurized with
helium gas. Purge the helium gas from all pressurized components before disposing. Open
the purging valve gradually or it may result in serious injury.
Install the cryocooler in the ventilated area, otherwise it may result in asphyxiation in case the
helium gas is purged or leaked.
WARNING The Adsorber mass is about 11.5kg. Be careful of handling so that it may not get hurt when
replace the adsorber.
The Oil Mist Adsorber is required to replace for every 30,000 Hrs operation.
TOOLS REMARK
1 1“ open-end wrench For Aero-quip coupling
2 1-1/8“ Open-end wrench For Aero-quip coupling
3 1-3/16“ Open-end wrench For Aero-quip coupling
4 Snoop liquid For leak check
5 Cotton wipers For leak check
6 13 mm Open-end wrench For fixing nut for Adsorber
7 Screw driver (phillips(+)) For side panel of Compressor Unit.
F-50H 24
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
PREPARATION
1) Shut down the Cryocooler.
2) Disconnect the Input Power Cable from the Compressor Unit.
3) Disconnect the Supply and Return Flex Lines from the Compressor Unit.
3) Remove the Nut secured the Adsorber to Rear Panel. Use two wrenches.
F-50H 25
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
4) Loosen the Nut and Washer secured the Adsorber to the base panel of the Compressor Unit.
F-50H 26
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
Tightening Torque:23N・m
F-50H 27
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
3) Secure the Adsorber to the base panel of the Compressor Unit by tightened Nut.
Tightening Torque:14.5N・m
5) Sprinkle “Liquid Leak Detector ” on the Flex line connecting coupling .in case the bubbling is found ,
tighten the connecting coupling again and re-check the leakage.
Ensure that the pressure gauge indication is specified value for the type of Cold Head. Charge
the helium gas, in case of low pressure indicating.
F-50H 28
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
Be sure to turn off the customer's main power before performing maintenance work such as
replacement of fuses, and remove the input power cable from the compressor unit. Failing to
observe this precaution may result in electric shock.
RESET PROCEDURE
1) Loosen the two screws that hold the circuit protectors cover and the remove the cover.
F-50H 29
TECHNICAL INSTRUCTION Sumitomo Heavy Industries, Ltd.
APPENDIX
APPENDIX
ELECTRICAL SCHEMATIC
DRAWINGS
F-50H 30
INCOMING POWER SUPPLY
QF1
OL protection MAIN DISCONNECTING MEANS
setting:16A CIRCUIT INTERRUPTER WITH
OUTSIDE OPERATING NOB AND
PADLOCK FITTING HOLE AT.
OFF-POSITION
Instantaneous
OC TRIP:208A
KM1 AWG10(5.5mm 2 )
1 2 (BK) 1T1 U
COMPRESSOR MOTOR
OL1
5 6
AWG14(2.0mm 2 ) (FULL LOAD AMPERES 13A,
(BK) AWG10(5.5mm 2 ) LOCKED ROTOR AMPERES 80A)
AWG10(5.5mm 2 )
(BK)
CP1 (G/Y)
0.63A
AWG16(1.3mm 2 )
TR1 0.2kVA (RD) COLDHEAD
2
200V KA12
AWG16(1.3mm ) 1L1A 480 2L1 3L1 2T1 JC A
480 COLD HEAD MOTOR
(BK) 460 2L2 3L2 B
1L1B 2T2 3φ 200V-50/ 60Hz
460 M2
1L1 415 2L3 3L3 2T3 C
1L1C (FULL LOAD AND LOCKED
415
1L1D 400 D ROTOR AMPERES APPROX. 0.5A)
400 CP2
1L1E 380
0.63A AWG16(1.3mm2 )
380 SA2 (G/Y)
480 200 ON COLDHEAD CABLE
1L2A KA12
480 3L1
460 OFF 3L2A
1L2 1L2B CN14 3L1A
460 -1 -3 KM1
1L2C 415 CN3
415 CONTROL PCB -1 A1 A2
1L2D 400 3L1A
400 CN3 AWG18(0.8mm 2 )
1L2E 380 -3 3L2
(RD)
380
CN2
KA10 KA9 -1
200 SV1
1L3A 480
480 CN2
1L3 1L3B 460 -3
AWG18(0.8mm 2 ) C6
460 0.033uF
1L3C 415 (RD) CN1 R
-1 C7 0.033uF
415
400 KA1
AWG18(0.8mm 2) NOTES:
1L3D -2 (RD/RDW TWIST)
400 KA2 1.ABBREVIATION OF WIRING COLOR IS AS FOLLOWS:
380 -4 S
1L3E
380 BK(:BLACK), RD(:RED),BL(:BLUE)
2N -5 KA22 KA2
RDW(:RED WITH WHITE STRIPES)
S1
TB1 W R15 G/Y(:GREEN WITH YELLOW STRIPES)
-7
2
AWG18(0.8mm ) E T 39K 3W
(BK) AWG18(0.8mm 2)
(G/Y) PE
3L1
KA22
3L2B
RF ~~
AWG18(0.8mm ) 2 R22
G PE
+ - (G/Y) 20K
AWG22(0.3mm 2 ) F-50H
24VDC
(BL)
CONTROL SOURCE ELECTRICAL SCHEMATIC 1/2 XRE05W0284BQ
P N
CN4 CN4 D-Sub
P N -2
-1 JR
CN11 CN11
(PSL) CN6
-1 1 2 -2 + - KA3 PRESSURE FAULT -1 15 1
P V3 KA8 CN6 PRESSURE ALARM
CN10 CN10 -2 16 2 WHEN CONTACT OPENS.
(TS1) KA4
-1 3 4 -2 + - OVER TEMP-ERROR SUPPLIED VOLTAGE TO BE 24V DC OR LESS.
[Helium
[Heiium Discharge]
90℃ V4
CN9 CN9 CN6
(TS2) -3 17
-1 5 6 -2 + - KA5 OVER TEMP-ERROR 3
[Helium After Cooler]
60℃ V5 CN6 OVER TEMP-ERROR〔Helium Discharge〕
CN8
(TS3)
CN8 KA13 WHEN CONTACT OPENS.
-1 7 8 -2 + - KA6 OVER TEMP-ERROR -4 18 4 SUPPLIED VOLTAGE TO BE 24V DC OR LESS.
[Water Outlet]
60℃ V6
AWG22(0.3mm 2 )
KA3 + - KA7 CN5 -5 21
LED PCB KA8 5 OVER TEMP-ERROR〔Helium After Cooler〕
CN15 CN15 V18 R1
V7 WHEN CONTACT OPENS.
CN1 CN1 100K
PRESSURE CN6 -5 13 SUPPLIED VOLTAGE TO BE 24V DC OR LESS.
-5 N -5 -1 -1 Q1 6
KA7 KA8
100uF
+ 1K R2 FAULT TIMING
28 + - C1 DRIVE INDICATION
LED1 - 10K R3 CN6 -6 N
ORANGE
7 WHEN 24V DC IS GIVEN BETWEEN JR 6 AND 7 PIN.
V8
CN6 -7 〔USABLE CURRENT max.0.15A〕
KA4 + - KA12 P 8 7 PIN:COMMON LINE OF 24V DC
CN1 CN15 JP1
-2 27 -2 3 2 1 KA13 V12 CN5 -6 22
V19 100K R4 OVER TEMP-ERROR 9
LED2 Q2 [Helium
[Heiium Discharge] KA15
ORANGE KA12 KA13 + 1K R5 CN16 -1 23 CONTROL VOLTAGE +24V DC
+ - 100uF C2 FAULT TIMING 10
- 10K R6 〔USABLE CURRENT max.0.15A〕
V13 CN16 -2 24 OVER TEMP-ERROR〔Water Outlet〕
CN15 KA17
CN1
-3 11 WHEN CONTACT OPENS.
-3 26 KA5 + - KA14
JP2 CN5 -1 19 SUPPLIED VOLTAGE TO BE 24V DC OR LESS.
LED3 3 2 1 KA15 V14 OVER TEMP-ERROR 12
ORANGE V20 R7 CN5
100K [Helium After Cooler] REMOTE DRIVE
Q3 -2 P
KA14 KA15 + 1K R8 FAULT TIMING 13
CN1 CN15 + - 100uF C3 KA11 CN5
-4 25 -4 - 10K REMOTE RESET
R9 + - -3 20 14
V15 WHEN KA11 IS ENERGIZED BY
LED4 PULSED 24V DC FOR 1 SEC.
GREEN KA6 + - KA16 V11
JP3
KA17
OVER TEMP-ERROR
3 2 1
V16 [Water Outlet] CN5 -4 11
V21 100K 15
R10 Q4
+ R11 FAULT TIMING
KA16 KA17
+ - 100uF 1K
12V C4
30K - 10K R12
+ C9 100 R13 ZD1 V17
-100uF AWG22(0.3mm 2 )
R21 + R14 Q5
KA1
22uF
- 20K
N
C5
N V9 NOTE:
CN7 (PSH) CN7
1.WIRING SIZE IS MIN.AWG24 FOR D-Sub.
-1 9 10 -2 KA9
+ -
PRESSURE CONTROL 2.WIRING COLOR OF 24V DC CIRCUIT IS BLUE.
P
SA3 SA1 V10
CN5 CN5 CN12 CN12 KA11 KA1 KA7 KA12 KA14 KA16
-8 LOCAL -7 11 -1 -2
+ - COMPRESSOR DRIVE
12 KA10 COLDHEAD DRIVE
REMOTE C10 R19
0.033μF 4.3K JP1 JP2 JP3
100
ZD2
13 HOU
CN6 CN5 V26 + HM - ELAPSED TIME 3 2 1 3 2 1 3 2 1
-7 -4 R20 + CN13 CN13
Q6
- R18 CN6 -1 -2 CN6 JP JP JP
CN5 C8 20K -5 -6
-2 + -
KA21 KA21 22uF
F-50L,F-50H
F-60L,F-60H
ELECTRICAL SCHEMATIC 2/2 XRE05W0272CQ
29 .3°
(155)
(260)
PROTECTIVE CAP
2PLACES
446.5
WASHER
21
SPRING WASHER
NUT
NOTE
1.CHARGED HELIUM GAS 1.62MPa・・・・Approx
2.MASS 11.5Kg・・・・Approx
XRE05B0246BQ_E
ADSORBER
O-RING
LIQUID TIGHT
LOCK NUT CONNECTOR CONDUIT
φ21.00max
NOTE (1)
Compressor User′s Power
Unit Side Source Side
φ36.78
(140)
L1 L1 L1:BLACK
L2 L2 L2:BLACK
L3 L3 L3:BLACK
GREEN/YELLOW
(GROUND)
12 12
NOTE
(1)MOUNTING HOLE SIZE : MINφ22.00mm.
(2)FOUR CABLES TO BE INSERTED IN THE CONDUIT.
REVISION CONTROL
REVISION CONTROL
Manual
Revision Remarks Date
No.
-A Publication of first edition. preliminary JULY. 28, 2006
-B Silk change. NOV. 3, 2006
Locking nut for D-sub and LED lamp additional.
-C DEC. 27, 2006
(Page5 and 6)
Starting current and Warning addition.
-D Recommended maintenance and Cooling Water JAN. 15, 2007
CD32ZZ-226
Quality correction.
Typo was corrected.
-E FEB. 16 , 2007
Delete Recommended maintenance.
F-50H