b) Big Bite punch 30 deg right width 3 to3.

5mm working length 11 to 16cm


c) Big Bite punch 30 deg Left width 3 to 3.5mm working length 11 to16cm
d) Big Bite punch up shaft width 3 to 3.5mm working length 11 to 16cm
e) Narrow punch width 3 to3.5mm working length 11 to 16cm
f) Hook scissors straight width 3 to 3.5mm working length 11 to 16cm
g) 3mm probegraduated
h) Grasper pitbull for loose body removal
i) USFDA/CE approved
Indigenous Mobile cart should be provided along with system.
Binocular Microscope

1. Observation method: Bright field & Dark field.


2. Illuminator: Transmitted light illuminator 0.5W LED lamp housing
3. Focusing mechanism: Vertical stage adjustment.
4. Revolving Nosepiece: 4 fixed mounts.
5. Stage: Mechanical Stage.
6. Condenser: Abbe Condenser N.A. 1.25 (with oil immersion)
7. Observation tubes: Binocular, wide field, F.N. 20, Tilt angle 30º
8. Dimensions: 198(W)×238(D)×386(H)mm

Electro- hydraulic table


USE
1.1 Clinical purpose An operating table, sometimes called operating room table, is the table on which the
patient lies during a surgical operation. This surgical equipment is usually found inside the surgery room
of a hospital.
1.2 Used by clinical department/ward : Operation theatre
TECHNICAL
2. TECHNICAL CHARACTERISTICS
2.1 Technical characteristics (specific to this type of device)
 Should be remote controlled with manual over ride, working range from floor level:600 -1050 or more ±10%
 Should be adjustable to all essential positions.
 Should be equipped with movement controls at side of the table.
 4) Should have frame and bottom made of 304 grade Stainless Steel material.
 Should have reinforced five section radiolucent top.
 Height should be adjustable by remote control.
 Should have detachable head rest which can be easily adjustable to any desired position, above or below the
table top.
 Head section raised from the Horizontal:20°-30°
 Durable and leak-proof hydraulic pump.
 Head section lowered from horizontal:28°-30°
 Back section raised from the horizontal:60°-70°
 Trendelenburg:25-30˚
 Reverse Trendelenburg:≥30°
 Leg section lowered from the Horizontal:40°-50°
 Kidney-position should be achievable by breaking the table.
 Table-top should be radio-lucent.
 Should have handset for position selection by in-built stand-by control.
2.2 User’s interface : Manual
2.3 Software and/or standard of communication(where ever required) : NA
3. PHYSICAL CHARACTERISTICS
70 E BID FOR THE PROCUREMENT OF MACHINERY & EQUIPMENT (2024)
3.1 Dimensions (metric) 1910 x 530 mm
3.2 Weight (lbs, kg) : Should be able to bear patient weight
3.5 Mobility, portability Not portable.
4. ENERGY SOURCE (electricity, UP S, solar, gas, water, CO 2 ....)
4.1 Power Requirements Recharging unit: Input voltage- 220V-240V AC, 50Hz
4.2 Battery operated : Yes
4.4 Protection Should have over-charging cut-off with visual symbol.
5. ACCESSORIES, SPARE PARTS, CONSUMABLES
5.1 Accessories (mandatory, standard, optional); Spare parts (main ones); Consumables / reagents (open,
closed system)
 S.S Arm Rest: 2 no.
 Anaesthetic Screen: 1 no.
 Lithotomy Leg Holders with Stirr-Ups:1 set
 Leather Wristlets:1 set
 Padded Leg Rest(Gutter type)
 Anti static mattress
 Side rails
BIDDING / PROCUREMENT TERMS / DONATION REQUIREMENTS
6. ENVIRONMENTAL AND DEPARTMENTAL CONSIDERATIONS
6.1 Atmosphere /Ambiance (air conditioning, humidity, dust ...)
 Operating condition: Capable of operating continuously in ambient temperature of 10 to 40 deg C and relative
humidity of 15 to 90% in ideal circumstances.
 Storage condition: Capable of being stored continuously in ambient temperature of 0 to 50 deg C and relative
humidity of 15 to 90%.
6.2 User’s care, Cleaning, Disinfection & Sterility issues
 Disinfection: Parts of the Device that are designed to come into contact with the patient or the operator
should either be capable of easy disinfection or be protected by a single use/disposable cover.
 Sterilization not required.

7. STANDARDS AND SAFETY


7.1 Certificates (premarket, sanitary, ..); Performance and safety standards (specific to the device type);Local
and/or international
 Should have US FDA/European CE/BIS approved product.
 All mechanical tests.
 Electrical safety conforms to the standards for electrical safety IEC 60601-General requirements( or equivalent
BIS standard) and IEC 60601-2-46 for usability.
 Certified to be compliant with IEC 60601-2-2 Medical Electrical Equipment Part 2-2 :Particular requirements
for the safety of High frequency Surgical Equipment if applicable or equivalent
7.2 Local and/or international Manufacturer / supplier should have ISO 13485 certificate for quality standard.
8. TRAINING AND INSTALLATION
8.1 Pre-installation requirements: nature, values, quality, tolerance
 Availability of 5 amp socket;
 Safety and operation check before handover;
8.2 Requirements for signoff
 Certificate of calibration and inspection from the manufacturer
8.3 Training of staff (medical, paramedical, technicians)
 Training of users on operation and basic maintenance;
 Advanced maintenance tasks required shall be documented
Operation Table Hydraulic Major
USE
1.1 Clinical purpose An operating table, sometimes called operating room table, is the table on which the
patient lies during a surgical operation. This surgical equipment is usually found inside the surgery room
of a hospital.
1.2 Used by clinical department/ward : Operation theatre
2. TECHNICAL CHARACTERISTICS
2.1 Technical characteristics (specific to this type of device)
 Should be a manually controlled operating table, working range from floor level: 800-1040mm.

71 E BID FOR THE PROCUREMENT OF MACHINERY & EQUIPMENT (2024)


 Should be adjustable to all essential positions.
 Should be equipped with movement controls at side of the table.
 Should have Frame and bottom made of Stainless Steel 304 material.
 Should have reinforced three section stainless steel top.
 Height should be adjustable by oil pump, foot step control.
 Should have detachable head rest which can be easily adjustable to any desired position, above or below table
top.
 Trendelenburg: ≥25°-30°
 Reversed Trendelenburg: ≥30°
 Head Section Raised from the Horizontal: ≥20°-30°
 Head Section Lowered from the Horizontal: ≥28°-30°
 Back Section Raised from the Horizontal: ≥60°-70 °
 Leg Section Lowered from the Horizontal: ≥40°-50°
 Kidney Position should be achievable by breaking the table.
2.2 User’s interface : Manual
2.3 Software and/or standard of communication(where ever required) : NA.
3. PHYSICAL CHARACTERISTICS
3.1 Dimensions (metric) Table top dimension (1900 mm x 525 mm) ± 15%
Table elevation: (700mm -1000 mm) ± 10%
3.2 Weight (lbs, kg) Should be able to bear patient having weight upto 160 kg
3.3 Configuration : NA
3.4 Noise (in dBA) : NA
3.5 Heat Dissipation: Should maintain nominal Temp and the heat should be disbursed through an cooling
mechanism.
4. ENERGY SOURCE (electricity, UP S, solar, gas, water, CO 2 ....)
4.1 Power Requirements Recharging unit: NA
4.2 Battery operated : NA
4.3 Tolerance (to variations, shutdowns) : NA
4.4 Power consumption : NA.
5. ACCESSORIES, SPARE PARTS, CONSUMABLES
5.1 Accessories (mandatory, standard, optional); Spare parts (main ones); Consumables / reagents (open,
closed system)
 S. S. Arm Rest 1 No.
 Anaesthetic Screen 1 No.
 Lithotomy Leg Holders with Stirr-Ups 1 Set
 Leather Wristlets 1 Set.
 Padded Leg Rest (Gutter Type)-2 nos.
 Anti static mattress-2 nos.
 Side rails-2 nos
DDING / PROCUREMENT TERMS / DONATION REQUIREMENTS
6. ENVIRONMENTAL AND DEPARTMENTAL CONSIDERATIONS
6.1 Atmosphere /Ambiance (air conditioning, humidity, dust ...)
 Operating condition: Capable of operating continuously in ambient temperature of 10 to 40 deg C and relative
humidity of 15 to 90% in ideal circumstances.
 Storage condition: Capable of being stored continuously in ambient temperature of 0 to 50 deg C and relative
humidity of 15 to 90%.
6.2 User’s care, Cleaning, Disinfection & Sterility issues
 Disinfection: Parts of the Device that are designed to come into contact with the patient or the operator
should either be capable of easy disinfection or be protected by a single use/disposable cover.
 Sterilization not required.

7. STANDARDS AND SAFETY


7.1 Certificates (premarket, sanitary, ..) ; Performance and safety standards (specific to the device type);Local
and/or international
 Should be USFDA/European CE/BIS approved product.
 Electrical safety conforms to the standards for electrical safety IEC 60601-General requirements(or equivalent
BIS standard) and IEC 60601-2-46 for usability.

72 E BID FOR THE PROCUREMENT OF MACHINERY & EQUIPMENT (2024)


 Shall meet internationally recognised IEC 60601-1-2 for Electromagnetic Compatibility(EMC) and
Electromagnetic Interference(EMI)
7.2 Local and/or international Manufacturer / supplier should have ISO certificate for quality standard.
8. TRAINING AND INSTALLATION
8.1 Pre-installation requirements: nature, values, quality, tolerance.
 Safety and operation check before handover;
8.2 Requirements for signoff
 Certificate of calibration and inspection from the manufacturer
8.3 Training of staff (medical, paramedical, technicians)
 Training of users on operation and basic maintenance;
 Advanced maintenance tasks required shall be documented

Orthopaedic Battery Operated Drill System


1. Drill and Reamer Hand piece
2. Headpiece should be compatible with drill reamer and saw attachments.
3. Headpiece should be compatible with battery.
4. Should have dual triggers for forward/reverse mode
5. Drilling speed should be> than 1400 rpm,Reaming speed -0to 400 rpm
6. Should have variable speed control on the hand piece
7. Drill torque of 15-40 in/Ibs
8. Weight of hand piece with battery should be Not more than 1.5 Kg
9. Fully cannulated 4.0 mm hand piece
10. Toolless 360 degree attachments insertion
11. Should be autoclavable

II
 Saggital saw attachment
Should have 11000 cycles per minute
Blade mount should be adjustable to different angles with 360 degree rotation
should be autoclavable
III
Attachments {AO Quick coupling Attachment for K wire/Reamer/ screw driver
1. ¼ inch Jacobs Drill Attachment with key
2. Small AO attachment
3. Reamer Attachment Chuck
4. K Wire Collect Attachment (1mm to 4mm)

IV Battery Charger
 Should have capability to identify the worn out battery
 Should have to charge four batteries at a time without any module or modification need
 Should have an indicator to provide battery status for charging

V Battery Kit
 Li Ion batteries with low internal impedance to deliver higher current than other battery types,
 Should be minimum 9.6 V
 Should have a run time of minimum 14 minutes
 Should include Autoclavable outer housing
 Shield to protect battery from the housing (Funnel)
 Should have sterilization case (indigenous)
Certification
 Should have US FDA certification and should have European CE Certificate

73 E BID FOR THE PROCUREMENT OF MACHINERY & EQUIPMENT (2024)