MSLFP06 User Manual
MSLFP06 User Manual
MSLFP06
User Manual
Document Version: A1
Before operating, please read this user manual and pay attention to all safety precautions.
Please ensure that this user's manual is properly maintained so that it can be accessed at
any time ( reserve ).
E
To Customers
To Customers
Congratulations on your purchase of the Fixed Digital Flat Panel (hereinafter referred to as MSLFP06)
which is manufactured by Guangzhou Medsinglong Medical Equipment Co., Ltd.
(Hereinafter referred to as MSL).
At MSL, we strive to not only make the world-class products that deliver the best
value possible to our customers but also offer the highest quality of service and
customer care. Please take time to read through this user guide in order to utilize
the product effectively. We hope you enjoy the experience with MSL MSLFP06
(configuration:MSLFP062 ).
If you have any questions or suggestions, please feel free to contact us.
Use only computers and image display monitors complying with IEC 60601-1 or
IEC 60950-1. For details, consult our sales representative or local MSL dealer.
Use only the dedicated cables. Do not use any cables other than those supplied
with this product.
Request your sales representative or local MSL dealer to install this product。
1
MSLFP06 Digital Flat Panel Detector User Manual
Disclaimer
MSL shall not be liable to the purchaser of this product or third parties for any
damage, losse, or injury incurred by purchaser or third parties as a result of fire,
earthquake, any accident, misuse or abuse of this product.
MSL shall not be liable to any damage, loss, or injury arising from unauthorized
modifications, repairs, or alterations to this product or failure to strictly comply with
MSL’s operating and maintenance instructions.
MSL shall not be liable for any damage or loss arising from the use of any options
or consumable products other than those dedicated as Original MSL Products by
MSL Technology.
Copyright
All rights reserved
Trademarks
The MSL name and MSL logo are registered trademarks of Guangzhou Medsinglong Medical Equipment Co.,
Ltd.
2
To Customers
3
MSLFP06 Digital Flat Panel Detector User Manual
图 含
标 义
Numerical Order
Year
Date
Month
Version
Product Code
Stand-by.
4
To Customers
Protective grounding.
5
MSLFP06 Digital Flat Panel Detector User Manual
Contents
TO CUSTOMERS......................................................................................................................... 1
CONTENTS .................................................................................................................................. 6
1. SAFETY ................................................................................................................................. 8
3. INSTALLATION ...................................................................................................................... 32
4. OPERATION ........................................................................................................................... 57
6
To Customers
7
MSLFP06 Digital Flat Panel Detector User Manual
1. Safety
1.1 Safety precautions .............................................................................. 9
1.2 Notes for Using ................................................................................. 13
8
1. Safety
WARNING
Installation and Do not use or store the equipment near flammable
chemicals such as alcohol, thinner, benzene, etc.
environment of
If chemicals are spilled or evaporate, it may result in fire or
use electric shock through contact with electric parts inside the
equipment. Also, some disinfectants are flammable. Be
sure to take care when using them.
Do not connect the equipment with anything other than
specified.
Doing so may result in fire or electric shock.
All the patients with active implantable medical
devices should be kept away from the equipment.
Power supply Do not operate the equipment using any type of power
supply other than the one indicated on the rating label.
Otherwise, it may result in fire or electric shock.
Do not handle the equipment with wet hands.
You may experience electric shock that could result in
death or serious injury.
Do not place heavy object such as medical equipment
on cables and cords. Do not pull, bend, bundle, or step
on them to prevent their sheath from being damaged,
and do not alter them neither.
Doing so may damage the cords which could result in fire
or electric shock.
Do not supply power to more than one piece of
equipment using the same AC outlet.
Doing so may result in fire or electric shock.
Do not turn ON the system power when condensation
has formed on the equipment.
Doing so may result in fire or electric shock.
Do not connect a multiple portable socket-outlet or
extension cord to the system.
Doing so may result in fire or electric shock.
To avoid the risk of electric shock, this equipment
must only be connected to power supply with
protective earth.
Not doing so may result in fire or electric shock.
9
MSLFP06 Digital Flat Panel Detector User Manual
WARNING
Handling Never disassemble or modify the equipment. No
modification of this equipment is allowed. Parts of the
VENU1717MN that are not serviced or maintained
while in use with the patient.
Doing so may result in fire or electric shock. Also, since the
equipment incorporates parts that may cause electric
shock as well as other hazardous parts, touching them
may cause death or serious injury.
Do not place anything on top of the equipment.
The object may fall and cause an injury. Also, if metal
objects such as needles or clips fall into the equipment, or
if liquid is spilled, it may result in fire or electric shock.
Do not hit or drop the equipment.
The equipment may be damaged if it receives a strong jolt,
which may result in fire or electric shock if the equipment is
used without being repaired.
Do not put the equipment and pointed objects
together.
The equipment may be damaged. If so, the equipment
should be used in bucky.
Have the patient take a fixed posture and do not let the
patient touch parts unnecessarily.
If the patient touches connectors or switches, it may result
in electric shock or malfunction of the equipment.
Maintenance and Please turn OFF the power of the equipment and
unplug the power cord of adaptor before cleaning.
inspection
NEVER use alcohol, ether and other flammable
cleaning agent for safety. NEVER use methanol,
benzene, acid and base because they will erode the
equipment.
DON'T dip the equipment into the liquid.
Please make sure that the equipment's surface & plugs
are dry before turning ON.
Otherwise, it may result in fire or electric shock.
10
1. Safety
CAUTION
Installation and Do not install the equipment in any of the locations
listed below. Doing so may result in failure,
environment of malfunction, equipment falling, fire or injury.
use Close to facilities where water is used
Where it will be exposed to direct sunlight
Close to the air outlet of an air-conditioner or ventilation
equipment
Close to heat source such as a heater
Where the power supply is unstable
In a dusty environment
In a saline or sulfurous environment
Where temperature or humidity is high
Where there is freezing or condensation
In areas prone to vibration
On an incline or in an unstable area
Take care that cables do not become tangled during
use. Also, be careful not to get your feet caught by
cable.
Otherwise, it may cause a malfunction of the equipment or
injury of the user due to tripping over the cable.
1
1
MSLFP06 Digital Flat Panel Detector User Manual
CAUTION
Handling Handle the equipment carefully.
Do not submerge the equipment in water.
The internal image sensor may be damaged if
something hits against it or it is dropped.
CAUTION
Do not close to fire, do not use in high temperature
Do not invert positive and negative pole
Do not contact with metal in case of short circuit
12
1. Safety
Before exposure
Be sure to check the equipment daily and confirm that it works properly.
Sudden heating of the room in cold areas will cause condensation to form on
the equipment. In this case, wait until the condensation evaporates before
performing an exposure. If the equipment is used while condensation is formed
on it, problems may occur in the quality of captured images. When an air-
conditioner is used, be sure to raise/lower the temperature gradually so that a
difference of temperature in the room and equipment does not occur, to
prevent condensation.
The detector should warm up for 15 minutes before exposure or updating the
gain map or defect map.
During exposure
Do not move the power or Ethernet Cables during exposure, or it may cause
image noise or artifacts, even incorrect images.
Do not use the devices near the equipment generating a strong magnetic field.
Otherwise, it may cause image noise, artifacts or even incorrect images.
Wipe it with a cloth slightly damped with a neutral detergent. Do not use
solvents such as alcohol, thinner, benzene, acid and base. Doing so may
damage the surface of the equipment.
Replace Cables
Turn OFF the power of the equipment and unplug the power cord from the AC
outlet before operation. Unplug the Detector cable from the float outlet, or it
may result in fire or electric shock.
Eliminate the static before replacing cable, including operating platform, tools and
operator, or ESD may damage the detector.
1
3
MSLFP06 Digital Flat Panel Detector User Manual
2. General Description
2.1 Scope ............................................................................................... 15
2.2 Model ................................................................................................. 15
2.3 Characteristic ..................................................................................... 16
2.4 Intended use/ essential performance/ application specification ....... 16
2.5 Essential performance ....................................................................... 16
2.6 The relative position between patient and detector ........................... 17
2.7 Product Components ......................................................................... 18
2.8 Optional Product Component ............................................................ 20
2.9 Components Description................................................................... 20
2.10 Product Specification ....................................................................... 24
14
2. General Description
2.1 Scope
This manual contains information about the MSLFP06. Information in the manual,
including the illustrations, is based on prototype. If your configuration does not have any
of these items, information about these items does not apply to your panel.
2.2 Model
Mars □ □ □
15
MSLFP06 Digital Flat Panel Detector User Manual
2.3 Characteristic
Cassette-size
Sync-shot exposure trigger
MSLFP06 Wireless Digital Flat Panel Detector is indicated for digital imaging solution
This panel provides digital X ray imaging for diagnosis of disease, injury, or any
applicable health problem. The image is obtained as the result of X ray passing through
MSL would provide hardware and software support for integration of system.
According to the MSLFP06 series intended use and the result of risk management,
getting imaging and function of data transmission is defined as essential performance.
Getting qualified dark image proves that essential performance does not influence
intended use. Method for getting dark image in detail refers to section “install” and
“operation”
16
2. General Description
PATIENT population:
Nationality: multiple
Intended OPERATOR:
All of use, maintenance and operation steps should be carried out by the operator who
has accepted the professional training offered by the company's customer service staff.
Life-time:
Because of the crosstalk effect of Amorphous silicon flat-panel detector, Pay attention to
the relative position of patient and detector, the recommended position as shown below,
Otherwise, the image is prone to abnormal light lines.
17
MSLFP06 Digital Flat Panel Detector User Manual
MSLFP06 comes with both DC power supply and battery package. Once powered on, it
would build a connection with Workstation through Ethernet cable (only for service) or
Wireless connection.
Item Description
1pcs
MSLFP06 Detector
Main Unit
2 pcs
Battery Battery pack
18
2. General Description
1pcs
Gigabit Ethernet Cable
3m
1 pcs
AC Power Cable
1 pcs
DC Power Cable
3.5 m
1pcs
Gain correction data
CD-Rom Defect correction map
SDK
Manual
19
MSLFP06 Digital Flat Panel Detector User Manual
ExternalSignals Input
20
2. General Description
Control Panel
F Antenna Antenna
2.9.2 Battery
21
MSLFP06 Digital Flat Panel Detector User Manual
A Battery Label /
C Pilot Pin /
22
2. General Description
Major 00-11
Minor 0000-1111
I grid blinking
Battery Insert with capacity ≤30%,charging
II and III grid off
23
MSLFP06 Digital Flat Panel Detector User Manual
2.10.1 Detector
2.10.1.1 Basic
Item Specification
MSLFP06-TSI (CsI)
24
2. General Description
Both AP mode and Client mode Processed Acquisition Time : 7 sec. (including Preview Time)
Weight (with one battery) MSLFP06-PSI: 4.45 kg without battery, 4.67 kg with battery
25
MSLFP06 Digital Flat Panel Detector User Manual
2.10.2 Battery
Item Specifications
Model Battery-KV
1 month-20℃-+50℃
Storage Temperature
3 month -20℃-+40℃
26
2. General Description
6 month -20℃-+20℃
Weight 0.22kg
Item Specifications
Model Charger-KV
Weight 1.26 kg
Item Specifications
27
MSLFP06 Digital Flat Panel Detector User Manual
MSLFP06 does not include Infrared Device. User can choose by themself, however some
basic requirements should be followed.
Item Specifications
MSLFP06 do not include AP Router. Users can choose AP Router as they wish, however
specification below is a requirement.
Item Specifications
Item Description
19: Europe
8: Japan
28
2. General Description
802.11a/g:
Modulation
64 QAM, 16 QAM, QPSK, BPSK
802.11n:
Item Description
Memory 4G DDR3
29
MSLFP06 Digital Flat Panel Detector User Manual
30
2. General Description
Temperature Atmospheric
Temperature Humidity Pressure Change
change Pressure
<10kp/min
Operating 5~30℃ <1k/min 30%~75% RH 700~1060hPa (1kp=1.0197E-
5Pa)
<10kp/min
700~1060hPa
Storage -10~40℃ <1k/min 10%~90% RH (1kp=1.0197E-
5Pa)
The MSLFP06 serial detectors shall operate at an altitude specified not more than 3000m, the environment
is only for detector.
31
MSLFP06 Digital Flat Panel Detector User Manual
3. Installation
3.1 Panel Installation ............................................................................... 33
3.2 Battery Charger Installation ............................................................... 36
3.3 Software Installation.......................................................................... 37
3.4 Panel Infrastructure ........................................................................... 37
32
3. Installation
MSLFP06 can be powered by both battery package and DC power. Once battery package is inserted or
DC power is on, Panel would be activated immediately. If none of battery and DC power is on,
MSLFP06 would power off. Please see below for battery installation.
33
MSLFP06 Digital Flat Panel Detector User Manual
3.1.1.3 Booting Up
On the control panel, user can press power button to power on/off.
34
3. Installation
After booting up, user can check the status LED indicator.
Operating Status
Power Indicator Lighting Status
Battery
Operating DC Input
Capacity
Orange Slow
Power ON ≤10% YES
Blinking
Green Fast
Power OFF >10% YES
Blinking
Green Slow
Power ON >10% YES
Blinking
35
MSLFP06 Digital Flat Panel Detector User Manual
Shut down
OFF wired connection broken and wireless
connection not ready
Panel Initialization
Green blinking
Infrared configuration
Shut down
OFF
Idle
Orange ON Initialization
Operation Figure
36
3. Installation
MSLFP06 supports two connection modes as follows, the IP address and other information mentioned below
is as the example, user should configure the connection with the specific requirement.
37
MSLFP06 Digital Flat Panel Detector User Manual
A
P
2) Wireless AP Mode
A
P
To build connection between workstation and Panel, User should follow steps below.
38
3. Installation
39
MSLFP06 Digital Flat Panel Detector User Manual
Wireless setup
Password: 12345678
40
3. Installation
Password: 12345678
LAN setup
External wireless AP reboot Apply above settings and reboot your wireless router.
41
MSLFP06 Digital Flat Panel Detector User Manual
Configuration of detector
Either Wired Cable or Infrared device can be used to configure detector in wireless client
mode. The wired connection should be used by the service operator only.
a.To start configuration with wired cable. It is necessary to finish 3.4.1.1, then proceed to the
steps below.
Connect panel to Workstation with
Ethernet Cable like 3.4.1
42
3. Installation
Click “Detector”
Click”Read Config”
Choose Client mode
Click “Add”
Type SSID and Password
Click”Apply”
43
MSLFP06 Digital Flat Panel Detector User Manual
Since we have chosen default SSID and password, it would connect to wireless AP immediately
after powered on next time.
44
3. Installation
/
Click “ ” to open wifi
setting
Change SSID and password, do
not select AP mode
Click “Apply” /
/
Click” ”
45
MSLFP06 Digital Flat Panel Detector User Manual
Click”Start”
46
3. Installation
Configuration of detector
Either Wired cable or Infrared device can be used to configure panel wireless AP mode. The wired
connection should be used by the service operator only.
a. To start wired cable configuration, users should finish 3.4.1.1, then proceed to the steps below.
47
MSLFP06 Digital Flat Panel Detector User Manual
Click “Detector”
Select ”wifi”
Choose AP mode
48
3. Installation
49
MSLFP06 Digital Flat Panel Detector User Manual
Since we have chosen default SSID and password, it would connect to wireless AP immediately after
powered on next time.
50
3. Installation
Start IrDARegister.exe
51
MSLFP06 Digital Flat Panel Detector User Manual
52
3. Installation
Click “Apply” /
/
Click” ”
Click”Start”
53
MSLFP06 Digital Flat Panel Detector User Manual
54
3. Installation
55
MSLFP06 Digital Flat Panel Detector User Manual
56
4. Operation
4. Operation
57
MSLFP06 Digital Flat Panel Detector User Manual
MSLFP06 provides SDK for user to integrate panel into their DR system. Additionally, it also
provides an application for demonstration, i.e. iDetector. User can use iDetector to control
panel without DR system.
To Acquire X ray image is the main operation of MSLFP06. Most importantly, panel
should build synchronization with X ray generator.MSLFP06 is born with four ways to
acquire x ray image, that is Software Mode, Inner Mode, Prep Mode and FreeSync Mode.
Software mode is the basic way to acquire x ray image. Please see figure below for
general feature
Workstation(Software) HV
G
Tub
e
Workstation is a host device installed with iDetector and SDK. Chapter 3 has described
58
4. Operation
Click “Prep”
Shoot X ray at any time. However the longer time you wait. The worse
image would be. So please shoot once x ray generator is ready.
To set a clear scenario for programming, see diagram below for details
Detector
2 3 1 1
0 1
Workstation
1 4 9 12
Xray
Generator e
Exposur
5 Anod
e 6 Exposure
Enable
7 X ray
shooting 8 e
Exposur
2. Panel receives “clear” from workstation, starts clearing leakage of panel. Meanwhile,
59
MSLFP06 Digital Flat Panel Detector User Manual
4. Workstation shows “Exposure Enable” on the IDetector’s message bar to tell user
5. User triggers x ray generator to initialize and do anode rotation to prepare for X ray
shooting.
6. X ray generator finishes preparation for X ray shooting and reminds user to shoot.
10. Panel receives “Data Acquisition” from workstation, start data acquisition operation.
11. Panel completes image acquisition and begins to send data to workstation.
12. Workstation receives all image data from panel which are after calibration if Hardware
calibration is on.
Workstation(Inner) HV
G
Tub
e
Workstation is a host PC device installed with IDetector and SDK. Chapter 3 has
described how to establish connection between panels and workstation. In inner mode,
workstation does not control x ray generator. Users would decide when to shoot x ray.
60
4. Operation
Click “Prepare”
Panel preparation, Wait until the warning message change from “Exposure
Prohibit” to “Exposure Enable”
Shoot X ray in exposure window. However the longer time user waits. the
worse image would be. So please shoot once x ray generator is ready.
After shooting, X ray sensor in panel would trigger panel to start image
To set a clear scenario for program, see diagram below for details
2. Panel receives “clear” from Workstation, start clear operation. Meanwhile, panel would send
4. Workstation shows “Exposure Enable” on the iDetector’s message bar to tell user shoot X
ray.
61
MSLFP06 Digital Flat Panel Detector User Manual
5. User triggers x ray generator to initialize and do anode rotation to prepare for X ray shooting
10. Panel completes image acquisition and begins to send data to Workstation.
12. Workstation receives all image data from panel which are after calibration is Hardware
calibration is on.
Syncbox
Workstation(Prep) HV
G
Tub
e
62
4. Operation
4.1.3.2 WorkFlow
Panel preparation, Wait until the warning message change from “Exposure
Prohibit” to “Exposure Enable”
Shoot X ray in exposure window (for click “Acq2” button). Or the exposure
flow will start when use hand-switch.
After shooting, the detector will acquire the image when the setting
“Set_Delay_Time” is reached. The Set_Delay_Time indicates the
exposure window.
Syncbox I II
Xray 5 7 8
4
Exposure Anod Exposure X ray Exposure
Generato Prohibited e
Rotat
Enable shooting Prohibited
r e
1. Workstation receives “I” information from the syncbox and check the state if the exposure
63
MSLFP06 Digital Flat Panel Detector User Manual
2. Panel receives the check information and send the “OK” reply, after that it will wait for the
acq2 command.
3. Workstation receives the “OK” information and replies to the syncbox to open the prepare
5. After preparation of the HVG, the syncbox send the “II” information to the workstation to
start the prepare process of the detector. And the HVG enters the exposure waiting state.
6. Workstation receives the “II” information and sends the “Acq2” command to the panel, after
get the reply of “Exposure Enable” from the detector, it replies the syncbox to start the
exposure.
If Hardware Pre-offset and Hardware calibration is selected, image got is the final image.
If Software Pre-offset and Software Calibration is selected, image got would be raw image,
Workstation(FreeSync) HV
G
Tub
e
64
4. Operation
Workstation is a host PC device installed with iDetector and SDK. Chapter 3 has described
how to establish connection between panel and Workstation. In FreeSync mode, User
doesn’t interact with Workstation. After shooting, images would be shown on screen
immediately.
Shoot X ray
Workstation 2
3 4 5
Panel
1
X ray
X
Generator Anode
Ra
Rotate
y
1. X ray generator is ready for X ray shooting and begins to release X ray.
3. Panel starts uploading Pre-dark image and Light image to Workstation for preview. If
hardware offset is selected, panel would do offset first, and then upload preview image
(2X2 binning).
65
MSLFP06 Digital Flat Panel Detector User Manual
4. Panel starts uploading Post-dark image to Workstation. If hardware offset is chosen, panel
would do correction and calibration first, then upload processed image to Workstation.
Note:
1. Once changing connection from different network card, user must re-connect panel with
different IP address.
2. The rule of Multi-Share control is based on IP address. The second terminal with different IP
address is not allowed to operate panel after the first one connected. If there is no
command transmission between panel and Workstation over 5 minutes, panel releases
access authority.
66
4. Operation
67
MSLFP06 Digital Flat Panel Detector User Manual
The correction and calibration should be performed after installation and it is recommended to
perform the new correction and calibration after any major change on the system settings
68
4. Operation
Before Gain template generating, make sure SID1.2m, no copper is required, the GUI of
the software maybe different with the below figures.
69
MSLFP06 Digital Flat Panel Detector User Manual
On Gain template
generating page, there are
five images that need to be
got
70
4. Operation
71
MSLFP06 Digital Flat Panel Detector User Manual
2. Make sure your x ray dose is right, if your dose is out of the range, IDetector will remind you
to adjust the dose. Then you can click “start creating” and try again.
3. If users operate with two panels, SDK has a probability of quit automatically.
72
4. Operation
“OPEN” provides two features for image check and uploading. Local Image Check, Panel
Image Upload. Local Image Check defines function to check image saved in Workstation.
Panel Image Upload defines function to upload images stored in panel.
73
MSLFP06 Digital Flat Panel Detector User Manual
74
4. Operation
75
MSLFP06 Digital Flat Panel Detector User Manual
76
4. Operation
Panel supports correction and calibration template storage. So template in panel could be
uploaded to Workstation, and template in Workstation could also be downloaded to panel.
77
MSLFP06 Digital Flat Panel Detector User Manual
Panel supports two ways to do correction and calibration. Software Correction and Calibration
defines the scenario that Workstation completes all correction and calibration. If panel
complete all correction and calibration by itself, it is named as Hardware Correction and
Calibration.
78
4. Operation
79
MSLFP06 Digital Flat Panel Detector User Manual
80
4. Operation
Click “ CPU”;
Select “ST STM32L151C8”;
Click “Target->Connect” /
Click “Target-> Erase” /
Click “Target->Program” /
Click “Target->Start Application” /
Note:1. Make sure panel is powered up.
If current MCU version is 2.5.2.*, Please refer to 4.10.2 and 4.10.3 for upgrading.
81
MSLFP06 Digital Flat Panel Detector User Manual
Note:
1. If it is MCU update, choose MCU image file. Otherwise, choose ALL-Image file, Please
make sure update file is selected, if not, panel will be not in use after updating.
2. There is a progress bar for indication. Make sure battery is inserted and battery capacity is
over 25%
3. Please make sure that iDetector shows “ Ready”. It can also be checked by click “Config”
button, there is firmware version.
82
4. Operation
Double-click the left mouse button, the image displayed in center and with maximum size.
Double-click the right mouse button, the window level and width adjusted to
WL:32767/WW:65535.
Lateral-drag the right mouse button to adjust the window width, and vertical-drag the right
F3 Key: Quickly adjust the image window width and window level.
4.10 Software
1 2 3 4 5 6
7
8
1
0
1
1
MSL provides test tools, such as iDetector for testing the basic performance of detector. It can
connect the detector, acquire image, image correct and calibrate.
83
MSLFP06 Digital Flat Panel Detector User Manual
Function description of regions and buttons within the main window as follows:
84
4. Operation
Message Information
If progress bar is Green when shooting X ray, image quality is acceptable, otherwise image
85
MSLFP06 Digital Flat Panel Detector User Manual
Except the following parameters, the value should not be modified for other parameters.
Description Modify
Type number of the detector NO
Product No
86
4. Operation
87
MSLFP06 Digital Flat Panel Detector User Manual
should be “On”
Self Clear Span
Time Should not be modified, and keep the original
YES
value
Hvg Prep On
Reserved YES
Hvg XRay Enable
Reserved YES
Hvg XRay On
Reserved YES
Tube Ready Time
Reserved YES
Image Pkg Gap
Time Reserved YES
Out Mode Cap
Trigger Reserved YES
88
4. Operation
Ftp Download Host FTP download server IP, keep the same as Host IP
YES
IP
Ftp Download Host FTP download server Port, keep the same as Host YES
Port Port
89
MSLFP06 Digital Flat Panel Detector User Manual
Ftp Upload Host IP FTP upoload server IP, keep the same as Host IP YES
Ftp Upload Host FTP upload server Port, keep the same as Host
YES
Port Port
Clr Acq Delay Time Exposure window for Acq2 work-flow of Prep
NO
trigger mode
Description Modify
90
4. Operation
Read Wifi Status Read the wifi status of the current detector /
to write to panel
91
MSLFP06 Digital Flat Panel Detector User Manual
panel
mode
mode
panel in AP mode
tools
92
4. Operation
93
MSLFP06 Digital Flat Panel Detector User Manual
5. Regulatory Information
5.1 Medical equipment safety standards ................................................. 95
5.2 The compliance for each EMISSIONS and IMMUNITY standard or test
specified by IEC60601-1-2 standard ....................................................... 96
5.3 Radio Frequency Compliance Information ...................................... 100
5.4 Battery Safety Standards ................................................................. 101
94
5. Regulatory Information
Type of protection against electrical External electrical power source equipment Class I
shock Equipment (medical approved adaptor)
shock
water
oxide
EN ISO 13485:2012/EN ISO Medical devices --- Quality management systems --- Requirements
13485:2012/AC:2012 for regulatory purposes
IEC 60601-1:2005+ Medical electrical equipment -- Part 1: General requirements for basic
Amendment 1:2012/EN safety and essential performance
60601-1:2006+
Amendment 1:2013
IEC 60601-1- Medical electrical equipment – Part 1-2: General requirements for
IEC 60601-1-3:2008/EN Medical electrical equipment – Part 1-3: Collateral standard: General
60601-1-3:2008 requirements for radiation protection in diagnostic X ray
equipment
IEC 60601-2-54:2015/EN Medical electrical equipment -- Part 2-54: Particular requirements for
95
MSLFP06 Digital Flat Panel Detector User Manual
efficiency
IEC 62366-1:2015/IEC
Medical devices –part 1: Application of usability engineering to
62366:2007/EN
medical devices
62366:2008
ANSI/AAMI ES60601-1:2005+
Medical Electrical Equipment – Part 1: General requirements for
Amendment 1:2012+
safety and essential performance
Amendment 2:2010
CSA CAN/CSA-C22.2 NO. Medical Electrical Equipment – Part 1: General requirements for
5.2 The compliance for each EMISSIONS and IMMUNITY standard or test
specified by IEC60601-1-2 standard
Emission
96
5. Regulatory Information
Voltage fluctuations and flicker IEC 61000-3-3 Professional healthcare facility environment
Compliance
Enclosure Port
ElectrostaticDischarg ±8 kV contact
IEC 61000-4-2
e ±2kV, ±4kV, ±8kV, ±15kV air
3V/m
80% AM at 1kHz
Rated power
30A/m
frequency IEC 61000-4-8
50Hz or 60Hz
magnetic fields
745
780
810 800-960 Pulse modulation 18Hz, 28V/m
870
97
MSLFP06 Digital Flat Panel Detector User Manual
930
1720 1700-1990 Pulse modulation 217Hz, 28V/m
1845
1970
2450 2400-2570 Pulse modulation 217Hz, 28V/m
5240 5100-5800 Pulse modulation 217Hz, 9V/m
5500
5785
Single phase: at 0º
98
5. Regulatory Information
Recommende
Shielded or
Cable d Number Cable classification
Unshielded
cable length
LAN Cable
3m Shielded 1 pcs Signal
(configuration mode)
MSLFP06 requires special precautions regarding EMC and needs to be installed only by
MSL or authorized personnel and put into service according to EMC information
99
MSLFP06 Digital Flat Panel Detector User Manual
MSLFP06 conforms to this EN60601-1-2:2015 standard for both immunity and emissions.
The use of accessories, transmitters and cables other than those specified by this User
Manual, with the exception of accessories and cables sold by MSL of MSLFP06 as
replacement parts for inner components, may result in increased emission or decreased
immunity.
Country Item
U.S.A FCC Part 15.107 Sub part (b) / 15.109(g) Sub part B
The panel has been tested to comply with limits for a Class B digital device,
pursuant to part 15 of FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation.
100
5. Regulatory Information
The panel must accept any interference received, including interference that may
The panel generates, uses, and radiates radio frequency energy and, if not installed
and used in accordance with the instruction, may cause harmful interference to radio
television reception, which can be determined by turning the panel off and on, the
measure.
Standards Description
IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-
acid electrolytes
ST/SG/AC.10/11/Rev.5/Amend.1&Amend.2
101
MSLFP06 Digital Flat Panel Detector User Manual
6. Trouble Shooting
TROUBLE SHOOTING ..................................................................................................... 102
102
6. Trouble Shooting
Please refer to service manual. If the problem persists, turn off the panel and contact MSL
service department ([email protected]). We would provide the best service.
103
MSLFP06 Digital Flat Panel Detector User Manual
7. Service Information
7.1 Product Lifetime............................................................................... 105
7.2 Regular Inspection and Maintenance .............................................. 105
7.3 Repair .............................................................................................. 105
7.4 Replacement Parts Support ............................................................ 105
104
7. Service Information
7.3 Repair
Product Name:
Series Number:
105
MSLFP06 Digital Flat Panel Detector User Manual
Appendix
Appendix A Information of Manufactures .............................................. 107
Appendix B Information of Medical Device Directive European Representative
............................................................................................................... 108
106
Appendix
107
MSLFP06 Digital Flat Panel Detector User Manual
108