Reduced Graphene Oxide Supported Mxene Based Metal Oxide Ternary Composite Electrodes For Non Enzymatic Glucose Sensor Applications
Reduced Graphene Oxide Supported Mxene Based Metal Oxide Ternary Composite Electrodes For Non Enzymatic Glucose Sensor Applications
com/scientificreports
Diagnosis and monitoring of glucose level in human blood has become a prime necessity to avoid
health risk and to cater this, a sensor’s performance with wide linearity range and high sensitivity
is required. This work reports the use of ternary composite viz. MG–Cu2O (rGO supported MXene
sheet with Cu2O) for non-enzymatic sensing of glucose. It has been prepared by co-precipitation
method and characterized with X-ray powder diffraction, Ultraviolet–visible absorption spectroscopy
(UV–Vis), Raman spectroscopy, Field emission scanning electron microscopy, High resolution
transmission electron microscopy and Selected area diffraction. These analyses show a cubic
structure with spherical shaped Cu2O grown on the MG sheet. Further, the electrocatalytic activity
was carried out with MG–Cu2O sensing element by cyclic voltammetry and chronoamperometry
technique and compared with M–Cu2O (MXene with Cu2O) composite without graphene oxide. Of
these, MG–Cu2O composite was having the high defect density with lower crystalline size of Cu2O,
which might enhance the conductivity thereby increasing the electrocatalytic activity towards the
oxidation of glucose as compared to M–Cu2O. The prepared MG–Cu2O composite shows a sensitivity
of 126.6 µAmM−1 cm−2 with a wide linear range of 0.01to 30 mM, good selectivity, good stability over
30 days and shows a low Relative Standard Deviation (RSD) of 1.7% value towards the sensing of
glucose level in human serum. Thus, the aforementioned finding indicates that the prepared sensing
electrode is a well suitable candidate for the sensing of glucose level for real time applications.
As per the World Health Organization (WHO) survey, a drastic increase in diabetics has made it as the 7th
most deadliest and life-threatening disease in the w orld1,2, which is due to the insufficient amount of insulin in
human body. In order to avoid this, periodic and continuous monitoring of glucose level (in human blood or
tear or saliva or acetone level in breathing air) is necessary and is done by painless method such as non-invasive
or closed loop t echnique3,4. Further economical and efficient sensors are required to be replaced with the cur-
rently available high-cost sensor in the market, which are enzymatic based leading to high cost, poor stability
and reproducibility in the m easurements5. An aforementioned issue with the enzymatic sensor is due to the
denaturation of enzymes with temperature, pH and humidity 6. In order to resolve the pre-existing problems,
huge effort is made by the researchers to develop enzyme free glucose sensors by direct oxidation of glucose mol-
ecule on the surface of the sensing electrode using metal or metal alloys as a sensing element (Pt, Pd, Au, Pt–Au,
Ni–Cu, Pt–Pd and Ni–Cu)7–9. Even though it resolves the enzymatic sensor drawbacks, it still lacks in terms of
production cost, selectivity and slow kinetic mechanism during glucose oxidation10. In order to overcome the
above-mentioned issues, researchers are focusing on various new electrode materials especially metal oxides
1
Centre for Nanotechnology Research, Vellore Institute of Technology, Vellore, Tamil Nadu, India. 2Present address:
Biological and Environmental Sensing Research Unit, King Abdullah Institute for Nanotechnology, King Saud
University, P.O. Box 2455, Riyadh 11451, Saudi Arabia. 3Department of Physics and Astronomy, College of Science,
King Saud University, Riyadh 11451, Saudi Arabia. 4Department of Self‑Development Skills, CFY Deanship, King
Saud University, Riyadh, Saudi Arabia. *email: [email protected]
Vol.:(0123456789)
www.nature.com/scientificreports/
(MO) as a sensing element instead of metal-based electrodes. Various MOs such as CuO, C u2O, Co3O4, NiO, and
Fe2O3 have been investigated as alternative sensing elements11–16. Among the different metal oxides, copper-based
metal oxide shows good catalytic activity towards glucose sensing for non-enzymatic sensor. But a lagging of
sensing parameters such as sensitivity or linearity is observed and to avoid such lagging of sensing parameters,
the active surface area or surface to volume ratio of the sensing elements has to be improved by fine tuning the
size, shape, faces or with the use of composites with carbon-based materials and Cu/Cu2O heterostructures. Such
reported investigations are listed in Table S1 and from the table, it’s clear that these kind of variations improve
the sensitivity but not the linearity range of the sensor, which is also a prime requirement for diabetic patients.
Low conductive nature of single metal oxide may hinder the flow of electrons between the catalyst and
electrode, which will further reduce the sensing performance of non-enzymatic glucose sensors. To cater this,
highly conductive materials in combination with metal oxide as a binary or ternary or multi compound-based
composites have been proposed 17. Various works on metal with MO (Pt/CuO, Co3O4@Pt@MnO2, Au/CuO,
NiO–Au) 18–22, MO with MO (Co3O/NiO, ZnO/NiO, CuO/Cu2O, CuO/ZnO), ternary MO (Mn–Cu–Al), 23–27
carbon materials with metal oxide composites (CuO/rGO, Cu2O/rGO, NiO/rGO, NiCo@f–MWCNT, MWCNT/
Fe3O4)28–32 have been studied as non-enzymatic glucose sensing elements. Among all these, reduced graphene
oxide (rGO) based composites possess fast electron transfer ability during the oxidation of glucose due to its
highly conductive nature, high surface area and good corrosion properties. In addition, composite of rGO with
metal oxides improves the stability of the electrode as compared to other combinations. Zeng et al., prepared
nickel oxide decorated reduced graphene oxide sheet for sensing of glucose molecules, which has shown a higher
response as compared to bare NiO with a linear range of 0.005 to 4.5 mM. rGO sheets in the composite prevent
the growth of clustered nanoparticles thereby enhancing the electron transfer ability between the catalyst and
electrode33. Phetsang et al. prepared copper based rGO composites film for glucose sensing application. The
prepared film showed a sensitivity of 172 µAmM−1 cm−2 with a linear range of 0.1 to 12.5 mM. Recently, Liu
et al. prepared a three-dimensional metal oxide/metal/reduced graphene oxide complex as a sensing element.
The prepared complex showed a higher response as compared to bare C u2O with a linear range of 16.65 mM
of glucose concentration34. Similar process was reported in various literatures in which different combination
of rGO with metal oxides have been studied to improve the sensing performance of non-enzymatic glucose
sensor17,35–38. Although these composites have improved the sensitivity as compared to bare metal oxides, it still
needs a lot of improvement in the linear range with high sensitivity to make it commercially viable.
Recently, a new 2D material called MXene has been explored widely as an alternative to graphene and such
materials are shown to be good for sensing applications typically in the linearity range. Among the different
MXenes, titanium carbide (Ti3C2Tx) has shown to be the best supporting material for electrochemical sensors
due to its biocompatibility, high surface area, high porosity, high conductivity as well as the presence of a large
number of functional groups on the MXene s heet39–41. These properties of T i3C2Tx sheet helps to construct the
best supporting material in the composites as compared to reduced graphene oxide for the preparation of non-
enzymatic biosensors. Rakhi et al. prepared Au/Ti3C2Tx based composite for sensing glucose molecules with a
wide linear range of 0.1 to 18 mM42. Chen et al. prepared Ce-metal organic framework with Ti3C2Tx composite
for the sensing of L-Tryptophan and reported a wide linear range (0.2 to 139 µM) as compared to the carbon-
based composites. As similarly, in our previous study, metal oxide (Cu2O) with Ti3C2Tx composites was shown
to possess a broad linear range (0.01 to 30 mM) of glucose sensing with a good stability43. However, a number of
works have been reported on the synthesis of composites with metal oxide, which suffer in terms of low sensi-
tivity due to restacking or internal sheet aggregation, which in turn limits the access of target molecules during
the electrochemical process. To prevent this, different strategies have been explored like introduction of foreign
atoms or incorporating carbon-based materials in the MXene based composite s tructure44–47.
Based on these concerns, this work is focused on the preparation of C u2O/rGO/Ti3C2Tx composite-based
sensing elements by co-precipitation method. Further, the prepared material was characterized using XRD,
Raman Spectroscopy, FE-SEM, HR-TEM and UV–Visible spectroscopy. The sensing element was fabricated
as electrode probe and further employed to investigate the sensing of glucose using chronoamperometry (CA)
technique.
Experimental methods
Chemicals and reagents. Graphite (325 mesh size), potassium permanganate (KMnO4), copper (II) ace-
tate hydrate (Cu (CH3COO)2·H2O), fructose (C6H12O6), sucrose (C12H22O11), L-ascorbic acid (C6H8O6), sodium
hydroxide (NaOH), sodium chloride (NaCl), potassium chloride (KCl), urea (CH4N2O), uric acid (C5H4N4O3)
and D-glucose (C6H12O6) were purchased from Sigma Aldrich and the MAX phase was purchased from Fors-
man. Hydrofluoric acid (HF), sulfuric acid ( H2SO4), hydrogen peroxide ( H2O2), ethanol ( C2H5OH) and dime-
thyl sulfoxide (DMSO) were purchased from SD Fine Chem Ltd. All the chemicals are of AR grade and utilized
as it is without any further purification.
Synthesis and fabrication of sensing electrode. Preparation of graphene oxide (GO). The prepara-
tion of GO was carried out by modified Hummer’s method as per the previous r eports47,48 with a slight modifica-
tion. In brief, 2 g of graphite powder was added to 100 ml of H2SO4 and stirred for 30 min. The mixed graphitic
solution was then transferred to ice-bath and then 12 g of KMnO4 was added slowly. After 2 h, the solution was
removed from ice bath and stirred at room temperature (RT) for 2 h. After this, 200 ml of ice water was added
gradually and then heated at 50 °C for 2 h. After heating, the solution was kept to cool down to RT and then
600 ml of DI water was added slowly. Finally, 10 ml of H2O2 was added to it and stirred overnight. The obtained
solution was further washed with DI water until the supernatant becomes neutral. The solid was transferred to
petri dish and dried at.
Vol:.(1234567890)
www.nature.com/scientificreports/
Preparation of MXene and MG composites. The conversion of MAX phase to MXene was carried out as similar
to our earlier r eport43 and the preparation of MG composite is given in Fig. 1 (step 1 and 2). The formed MXene
was used for the preparation of MG composites by hydrothermal method. Initially, 100 mg of MXene was added
to 90 ml of DI water and then sonicated for 30 min labeled as solution A. Similarly, 100 mg of GO was added
into 90 ml of DI water and then sonicated for 30 min and named as solution B. The prepared solution B was
added into solution A and sonicated for 1 h. After this, the mixed solution was transferred to 250 ml Teflon lined
stainless steel autoclave and heated at 90 °C for 12 h. After that, the solution was cooled down to RT, the obtained
solution is further centrifuged and washed with DI water several times. Finally, the formed rGO–MXene solid
was dried at 60 °C and named as MG-100. Likewise, different weight ratio of graphene oxide in solution B was
varied with a fixed amount of MXene (100 mg) for the preparation of MG composite and named as MG-0,
MG-25, MG-50 and MG-75 respectively.
Vol.:(0123456789)
www.nature.com/scientificreports/
Preparation of sensing electrode. Cu2O–rGO–MXene composite electrode was fabricated by drop cast-
ing method. Initially, the glassy carbon electrode was cleaned with A
l2O3 powder and then washed with ethanol,
acetone followed by DI water and the washed electrode was further dried under N2 gas. For the deposition of the
prepared composite, a mixture of 1 mg MG–Cu2O, 0.045 ml of DI water and 0.005 ml of 5% nafion solution was
sonicated for 30 min and then 5 µl of mixed solution was dropped on the dried electrode. Further the dropped
electrode was left overnight at room temperature for drying. In the same way, MG composite and M–Cu2O
composites-based electrodes were prepared and then further used for studying the sensing performance.
Vol:.(1234567890)
www.nature.com/scientificreports/
Figure 2. (a) X-ray diffraction patterns of synthesized Cu2O, Ti3C2Tx, MG (50% of GO) and MG–Cu2O
composite (b) Raman spectra for Cu2O, MXene, and their composites (c) Optical absorption spectra of M–
Cu2O, MG–Cu2O and (d) Optical band gap for M–Cu2O, MG-Cu2O.
The corresponding bandgap of material is obtained from the Tauc plot and is given in Figs. 2d, S3d. The
bandgap values are 1.78 eV (MG–Cu2O), 1.87 eV (M–Cu2O) and 1.9 eV (Cu2O) respectively. The above findings
suggest the formation of active heterojunction between the MG and Cu2O interface as well as a strong electron
conduction of graphene in the MG–Cu2O composite58. As per the structural analysis through XRD, Raman and
UV–Visible spectroscopy, a low d-spacing and high defect density of MG-50 composite lead to high conductiv-
ity as compared to MG-0, MG-25, MG-75 and MG-100 respectively. A low value in the crystallite size and a red
shift in the adsorption edge of C
u2O in the composite may lead to more active sites as compared to bare Cu2O.
Morphological study of bare materials and composites. Morphological study of MAX phase,
MXene, MG-50 and MG–Cu2O composites were done with FE-SEM and given in Figs. S5 and 3. After HF etch-
ing of the MAX phases (Fig. S5a,b), a spacing between the layers are observed as similar to the accordion like
structure and is given in Fig. S5c,d. This formation is one more evidence for MXene conversion as supported
with the result of XRD (a shift of (002) peak from 9.54° to ~ 6.4° and disappearance of 38.92° peak). After soni-
cation and hydrothermal treatment of MXene and graphene oxide sheets, the formed reduced graphene oxide
sheets are crevice and cover the MXene sheets 44 (highlighted in green—Fig. 3a,b). It reveals that the two materi-
als are well attached leading to the formation of MG composites59.
For the MG–Cu2O composite (Fig. 3c,d), a non-uniform distribution of aggregated Cu2O nanoparticles
are ornamented on the surfaces of MG-50 composite. These particles are spherical in shape with the size of
500–700 nm. To confirm the role of rGO sheet in the morphological changes, a MG–Cu2O composite is prepared
without rGO sheet (M–Cu2O) and analyzed with FE-SEM. In the absence of reduced graphene oxide sheet, an
octahedron shape of Cu2O (size of 1.5 µm) is formed and highlighted in red circle in Fig. S6a. In contrast, a
spherical shape (Figs. 3d, S6b) of C
u2O (size of 500–700 nm) is formed for MG–Cu2O composite. In the presence
Vol.:(0123456789)
www.nature.com/scientificreports/
Figure 3. FE-SEM images at different magnifications (a, b) MG-50 and (c, d) MG-Cu2O composite.
of rGO sheet, the preclusion of restacking the MXene sheets maintain the larger surface area of sheets as well as
the ample amount of functional groups on the surface and edges of sheet may hinder the nanoparticle agglom-
eration leading to the formation of spherical shape of Cu2O in the MG–Cu2O composite as given in Fig. S6b
(marked in red circle). In order to understand the distribution of C u2O nanoparticles (MG–Cu2O), HR-TEM
was done and given in Fig. 4a,b. A larger number of nanoparticles are embedded on the surface as well as edges
of sheets as observed from Fig. 4b. The average sizes of the distributed nanoparticles are ~ 4.96 ± 1.9 nm and the
respective histogram plot is given in Fig. 4c. Further, the nature of distributed nanoparticles is studied with the
help of SAED pattern for MG–Cu2O composite (Fig. 4d). It shows five different rings along with different spots
indexed with (111), (200), (211), (220) and (311), which are well matching the XRD results.
Electrochemical properties of bare and composite electrodes. In order to find out the highest
electron transfer rate capability of GO loaded MG composites (MXene, MG-25, MG-50, MG-75 and MG-100),
cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) techniques were investigated
in 0.1 M KCl with 5 mM of [Fe(CN)6]4−/3− electrolyte (Fig. S7a–d). As observed from the figure, both anodic
and cathodic peaks are observed in cyclic voltammetry profile of all the materials with the potential difference
(ΔEvp) of 395 mV (MXene), 275 mV (MG-25), 179 mV (MG-50), 233 mV (MG-75) and 250 mV (MG-100)
respectively. Of the various composites, 50 wt% of graphene oxide loaded MG composite (MG-50) shows a lower
ΔEvp as compared to other materials, which indicates the high electron transfer rate as compared to others60.
An increment in graphene oxide loading may reduce the active electrochemical surface area and hinder the flow
of electrons, which is further confirmed with the charge transfer resistance value (Rct) between the modified
electrode and electrolyte for these materials.
For obtaining this, an electrochemical impedance spectral analysis was carried out in the same medium and
given in Fig. S7b,d. For all the wt% of graphene oxide loaded MG composites, the graph shows a semicircle along
with straight-line in the high frequency region. The diameter of the semicircle is equivalent to the charge transfer
resistance value61. The obtained values for Rct are 80 kΩ (MXene), 110 kΩ (MG-25), 0.25 kΩ (MG-50), 1.46 kΩ
(MG-75) and 20 kΩ (MG-100) respectively. A low charge transfer value for MG-50 composite indicates the high
electron transfer rate capability as compared to the other loading, which might be due to its low d-spacing and
highest defect density of the prepared material MG-50.
Further, MG-50 composite has been chosen for the preparation of ternary composites (MG–Cu2O compos-
ites). A cyclic voltammetry and electrochemical impedance spectra analysis of M–Cu2O (Fig. S7e,f), MG-50
and MG–Cu2O composite is given in Fig. 5a,b. The addition of Cu2O catalyst to MG-50 enhances its anodic and
cathodic current as compared to MG-50 alone. A comparative plot of the voltage difference between redox peak
(ΔEvp) and charge transfer resistance (Rct) for the different MG composites, M–Cu2O and MG–Cu2O are given in
Fig. 5c,d. From this plot, a low ΔEvp (121 mV) and Rct (0.1 kΩ) value was observed for MG–Cu2O, which might
be due to its interconnected structure of Cu2O with rGO and MXene sheets as supported with FE-SEM image
Vol:.(1234567890)
www.nature.com/scientificreports/
Figure 4. (a, b) HR-TEM images of MG-Cu2O composite with different magnification (Pink and blue
circle shows the Cu2O on the edges of MXene and rGO sheet, the yellow arrow shows the rGO sheets in the
composites and red arrow indicates the MXene sheet) (c) size distribution of C
u2O nanoparticles on MXene
sheet (d) selected area electron diffraction (SAED) pattern of MG-Cu2O composite.
(Fig. 3c,d). In order to understand the charge transfer mechanism, the diffusion studies was further performed
for MG–Cu2O composite by tracing the peak current response with changing scan rates from 10 to 100 mV/s
(Fig. 6a). A linear response is observed for peak currents with respect to the square root of scan rate, which
rocess62.
indicates that the rate of the reaction is a diffusion-controlled p
Electrocatalytic study of bare and composites electrode. In order to verify the above-mentioned
results, cyclic voltammetry study of MXene, MG-25, MG-50, MG-75 and MG-100 was carried out in 0.1 M
NaOH in the presence of glucose and given in Fig. 6b. A higher current variation is observed for MG-50 com-
posite as compared to others, which implies the faster electron transfer between the electrode and sensing ele-
ment during the oxidation of glucose. Further, the role of rGO and Cu2O in MXene based composite is studied
and given in Fig. 6c. No variation in the current is observed in the absence of glucose analyte for MG-50, M–
Cu2O and MG–Cu2O respectively. With the addition of 5 mM glucose, instantaneous enhancement in the cur-
rent was observed for all the materials, which signifies that all the prepared materials can act as a sensing element
for glucose sensor. Among the composites, a huge enhancement was observed when bare M–Cu2O is replaced
with MG–Cu2O (Fig. 6c). There is a presence of anodic peak in the potential range of 0.4 to 0.8 V, which occurs
due to the oxidation of copper ion in the electrode during the sensing of glucose molecule as explained in Fig. 7.
Briefly, in the presence of glucose, C u+ ions are oxidized to C u2+ with the release of electrons into the solutions
and the formed C u2+ ions are further oxidized to C
u3+ ions. These processes initiate the oxidation of glucose and
simultaneously there is a reduction of Cu3+ to Cu2+ resulting in an enhancement of current. Further, the formed
gluconolactone gets converted to gluconic acid at 0.6 V via the hydrolyzation process63. In the composite, rGO
prevents the stacking of MXene sheet, which improves electron transfer rate with more active sites. This may
enhance the peak and spike current of MG–Cu2O composite, indicating more C u3+ ion formation for MG–Cu2O
Vol.:(0123456789)
www.nature.com/scientificreports/
Figure 5. (a) Cyclic voltammetry comparison of MG-50 and MG-Cu2O composites in 5 mM Fe (CN)3-/4- along
with 0.1 M KCl (b) Nyquist plot of MG-50 and MG-Cu2O composites with the frequency range of 1 MHz to
0.1 Hz (c) 3D bar graph of potential difference between redox peak (ΔEvp) for different materials and (d) 3D bar
graph of charge transfer resistance ( Rct) for different materials.
composite as compared to M–Cu2O. As the concentration of glucose increases, a spike and peak current of
MG–Cu2O modified electrode increases. It shows that the prepared composite is electrochemically active and
responding well during the oxidation of glucose. Therefore, the replacement of M–Cu2O to MG–Cu2O compos-
ite has enhanced the glucose oxidation current beneficiary for sensor applications.
To reinforce the above-mentioned result (CV analysis), a chronoamperometry (CA) analysis of M–Cu2O and
MG–Cu2O composite was carried out in 0.1 M NaOH at different concentrations of glucose (Fig. S8a). As the
concentration of glucose increases, a huge improvement in current for MG–Cu2O (Figs. 6d, S8a) was observed
as compared to M–Cu2O. Therefore, results show that the replacement of M–Cu2O with MG–Cu2O increases
the conductivity of the sensing element and improves the active sites for sensing analyte.
Sensing performance of modified electrode. In order to determine the suitable working potential of
MG–Cu2O, CA study was further carried out with the addition of 0.1 mM glucose in 0.1 M NaOH at various
bias voltages (0.5, 0.6 and 0.7 V). The corresponding current response is plotted with respect to concentration of
glucose and given in Fig. S8b. An increment in the current is observed with an increase in the concentration of
the glucose at all the biasing conditions. Among all the bias, a high change in the current was observed at 0.6 V
and hence this potential was kept constant for further study. The sensitivity of MG–Cu2O modified electrode
was obtained using CA technique with the stepwise addition of glucose at every 50 s (Fig. 8a). A quick change
in current (less than 5 s) was observed with the addition of glucose to the electrolyte. A magnified image at the
lowest concentration of glucose and the respective changes in current is provided in the inset of Fig. 8a. Even
for the lower concentration (10 µM), there is an enhancement in the current. Later, a linear plot of current vs.
concentration of glucose is obtained as given in Fig. 8b. Similarly, M–Cu2O (Fig. 8b) and rGO–Cu2O (Fig. S9a)
composites were studied and the corresponding current response with respect to the change in concentration is
given in Figs. 8c, S9b respectively. MG–Cu2O composite shows a wider linear range as compared to rGO–Cu2O
composites whereas it is similar as M–Cu2O composite. However, a tenfold increment of current is observed by
the replacement of M–Cu2O with MG–Cu2O during the sensing of glucose molecules. It might be due to the
Vol:.(1234567890)
www.nature.com/scientificreports/
Figure 6. (a) Peak current vs. square root of scan rate for MG-Cu2O composite and the inset—cyclic
voltammetry (0.1 M of KCl + 5 mM of Fe (CN)3−/4−solution) at different scan (b) CV response of MG composite
with varying the weight ratio of MXene and Graphene oxide in the presence of 5 mM glucose (in 0.1 M NaOH)
(c) CV curve of MG-50, M-Cu2O and MG-Cu2O composite at 0 and 5 mM glucose (in 0.1 M of NaOH) and (d)
CV of MG-Cu2O composite at different concentrations of glucose in 0.1 M NaOH.
lowest d-spacing and highest defect density of the prepared structure as well as the introduction of rGO, which
avoids the restacking of MXene layers and enhance the active sites as compared to M–Cu2O. Regression equa-
tions for M–Cu2O and rGO–Cu2O are ΔI(µA) = 0.785 [Glucose] mM + 0.3623 with R2 = 0.99, ΔI (µA) = 1.174
[Glucose] mM + 0.5 with R2 = 0.97, respectively.
Moreover, CA response of the MG–Cu2O in glucose environment is divided into two sections of linearity
that is 0.01 to 10 mM and 11 to 30 mM with the slope value of 8.992 and 6.323, respectively. The correspond-
ing regression equations in these two sections are ΔI (µA) = 8.992 [Glucose] mM + 1.73 with R 2 = 0.99 and ΔI
2
(µA) = 6.323 [Glucose] mM + 28.3 with R = 0.99. From these equations, a sensitivity of M–Cu2O, rGO–Cu2O
and MG–Cu2O is calculated and it given in Fig. 8d. A table of comparison of our material with the reported
ones is given in Table 1.
From Fig. 8d and Table 1, it could be understood that the prepared ternary composite (MG–Cu2O) is hav-
ing the enhanced sensitivity with a wide linear range of glucose sensing as compared to other sensing elements
(rGO–Cu2O and M–Cu2O). In these composites, a large surface area, high conductivity (6500 S cm−1) and
hydrophilicity nature of MXene sheet will increase the CuOOH formation in the electrolyte. It might generate
the enediol and lower the energy barrier of electrooxidation reaction with the addition of glucose to the elec-
trolyte. This will increase the number of electrons leading to a wide linear range as per the Yang et al.71. But the
sensitivity of the sensor was low as per the requirement for commercialization. Restacking of MXene sheet may
reduce the rate of electron transfer between the electrode and electrolyte. To prevent this, an incorporation of
rGO sheet (rGO will improve the rate of electron transfer process during the glucose sensor33) to MXene was
carried out. It will maintain the larger surface area of sheets as well as the ample amount of functional groups on
the surface and edges of sheet may hinder the nanoparticle agglomeration leading to the formation of spheri-
cal shape of C u2O in the MG–Cu2O composite, which further enhances the CuOOH formation and increases
the rate of electron transfer between electrode and electrolyte during the oxidation of glucose. In the prepared
ternary composites, MXene improves the linear range by a greater number of CuOOH formation and rGO will
Vol.:(0123456789)
www.nature.com/scientificreports/
Figure 7. A schematic showing the glucose sensing mechanism of MG–Cu2O composite.
improve the rate of electron transfer between the electrode and electrolyte. The combined role of MXene and
rGO has thus promoted the sensitivity of C u2O composite with wide linear range.
Real time performance of MG–Cu2O modified electrode. To understand the real time application of
the prepared sensing electrode, the electrochemical test of the prepared sensing element in the presence of three
different serums was investigated. It was observed that the present sensing electrode shows 7.01 mM concen-
tration of glucose with the addition of serum 1, whereas it was found 7.15 mM concentration of glucose using
a commercial glucometer. A detailed relative standard deviation value between the proposed and commercial
glucometer is tabulated in Table 2. Lower value of deviation suggests that the prepared electrode is well suitable
for real time application with good reliability.
Conclusion
In this report, the role of graphene oxide in MXene–rGO composites is optimized and investigated as a sensing
probe for glucose sensing application. The optimized MG composite is later utilized to prepared the ternary
composite of rGO supported MXene based C u2O composite (MG–Cu2O) by coprecipitation method. Structural
Vol:.(1234567890)
www.nature.com/scientificreports/
Figure 8. (a, b) Current versus time (Chronoamperometry) study of M–Cu2O and MG–Cu2O composite
with the successive addition of glucose at 0.6 V versus Ag/AgCl (inset—zoomed at lower concentrations) (c)
Comparative linear plot of MG–Cu2O and M–Cu2O composite (d) Sensitivity of prepared sensing materials.
Electrode Sensing method Linear range (mM) Detection limit (µM) Sensitivity (µAmM−1 cm−2) Ref
34
Cu2O/Au/GO NE Upto 16.65 0.83 2886
66
CuO/CG-GCE NE 0.1–3.17 0.01 1295
Cu NP on laser induced 67
NE 0.001–6 0.39 495
graphene
68
Ni/rGO/PU NE 0.01–2 1.28 4876
69
Cu-Co/rGO/PGE NE 0.001–4 0.15 240
65
S doped rGO/CuS NE 20.1 0.032 429
18
Pt-CuO-Pt NE 2.2–10 1.42 2921
70
Cu/rGO NE 0.1–12.5 65 172
43
GOx/Au/MX/Nafion E 0.1–18 5.9 4.2
33
NiO/Graphene NE 0.005–4.2 0.1 666.7
71
MXene/NiCo-LDH NE 0.002–4.096 0.53 64.75
44
MXene-Cu2O NE 0.01–30 2.83 11.064
Cu2O/rGO/MXene (MG-
NE 0.01–30 2.1 125.6 This work
Cu2O)
Table 1. A comparison of glucose sensing performance of MG–Cu2O with others glucose sensors. *GCE
glassy carbon electrode, CG carboxylate graphene, PU polyurethane, PCE pencil graphite electrode, LDH
layered double hydroxide, NE non enzymatic, E enzymatic.
Vol.:(0123456789)
www.nature.com/scientificreports/
Figure 9. (a) Sensing analysis of respective elements in human serum for MG–Cu2O (0.5 mM glucose along
with 0.05 mM of other active species) (b) stability study of electrode in electrolyte using CV (inset) and CA (c)
storage stability of modified electrode with 0.1 mM of glucose (d) Sensing response of six different modified
electrodes with MG–Cu2O composite.
analysis confirmed that the prepared composite is having high defect density with lower crystalline size as
compared to M–Cu2O composite. Such composites might lead to enhance the conductivity and amplifies the
signal towards glucose oxidation with MG–Cu2O as compared to M–Cu2O. Therefore, the prepared composite
shows a sensitivity of 125.6 µAmMcm−1 with a broad linear range (0.01 to 30 mM) and further it shows a good
selectivity, stability and reproducibility. Results show that the prepared MG–Cu2O composite is preferable for
analyzing the glucose level in human serum.
Data availability
All data generated or analysed during this study are included in this published article.
References
1. Cho, N. H. et al. IDF diabetes atlas: Global estimates of diabetes prevalence for 2017 and projections for 2045. Diabetes Res. Clin.
Pract. 138, 271–281 (2018).
2. Teymourian, H. et al. Electrochemical glucose sensors in diabetes management: An updated review (2010–2020). Chem. Soc. Rev.
49, 7671–7709 (2020).
Vol:.(1234567890)
www.nature.com/scientificreports/
3. Ma, R. et al. Recent advancements in noninvasive glucose monitoring and closed – Loop management systems for diabetes. J.
Mater. Chem. B10, 5537–5555 (2022).
4. Shokrekhodaei, M. & Quinones, S. Review of non-invasive glucose sensing techniques: Optical, electrical and breath acetone.
Sensors 20, 1251 (2020).
5. Wilson, R. & Turner, A. P. F. Glucose oxidase: An ideal enzyme. Biosens. Bioelectron. 7, 165–185 (1992).
6. Mohapatra, J. et al. Enzymatic and non-enzymatic electrochemical glucose sensor based on carbon nano-onions. Appl. Surf. Sci.
442, 332–341 (2018).
7. Park, S., Boo, H. & Chung, T. D. Electrochemical non-enzymatic glucose sensors. AnalyticaChimicaActa 556, 46–57 (2006).
8. Hwang, D.-W., Lee, S., Seo, M. & Chung, T. D. Recent advances in electrochemical non-enzymatic glucose sensors – A review.
AnalyticaChimicaActa 1033, 1–34 (2018).
9. Park, S., Chung, T. D. & Kim, H. C. Nonenzymatic glucose detection using mesoporous platinum. Anal. Chem. 75, 3046–3049
(2003).
10. Toghill, K. & Compton, R. Electrochemical non-enzymatic glucose sensors: A perspective and an evaluation. Int. J. Electrochem.
Sci. Int. J. 5, 1246–1301 (2010).
11. Dong, Q., Ryu, H. & Lei, Y. Metal oxide based non-enzymatic electrochemical sensors for glucose detection. Electrochim. Acta
370, 137744 (2021).
12. Huang, J. et al. Flexible 3D porous CuO nanowire arrays for enzymeless glucose sensing: In situ engineered versus ex situ piled.
Nanoscale 7, 559–569 (2015).
13. Li, C. et al. An improved sensitivity nonenzymatic glucose biosensor based on a C uxO modified electrode. Biosens. Bioelectron.
26, 903–907 (2010).
14. Xu, J. et al. Co3O4 nanostructures on flexible carbon cloth for crystal plane effect of nonenzymatic electrocatalysis for glucose.
Biosens. Bioelectron. 123, 25–29 (2019).
15. Wang, G. et al. Free-standing nickel oxide nanoflake arrays: Synthesis and application for highly sensitive non-enzymatic glucose
sensors. Nanoscale 4, 3123–3127 (2012).
16. Cao, X. & Wang, N. A novel non-enzymatic glucose sensor modified with Fe2O3 nanowire arrays. Analyst 136, 4241–4246 (2011).
17. Liu, S., Zeng, W., Guo, Q. & Li, Y. Metal oxide-based composite for non-enzymatic glucose sensors. J. Mater. Sci. Mater. Electron.
31, 16111–16136 (2020).
18. Abunahla, H. et al. MOM sense: Metal-oxide-metal elementary glucose sensor. Sci. Rep. 9, 5524 (2019).
19. Xia, H. et al. Hierarchically structured C o3O4@Pt@MnO2 nanowire arrays for high-performance supercapacitors. Sci. Rep. 3, 2978
(2013).
20. Li, Z., Xin, Y., Zhang, Z., Wu, H. & Wang, P. Rational design of binder-free noble metal/metal oxide arrays with nanocauliflower
structure for wide linear range nonenzymatic glucose detection. Sci. Rep. 5, 10617 (2015).
21. Mishra, A. K. et al. Au nanoparticles modified CuO nanowire electrode based non-enzymatic glucose detection with improved
linearity. Sci. Rep. 10, 11451 (2020).
22. Ding, Y., Liu, Y., Parisi, J., Zhang, L. & Lei, Y. A novel NiO - Au hybrid nanobelts based sensor for sensitive and selective glucose
detection. Biosens. Bioelectron. 28, 393–398 (2011).
23. Ramasamy, R. et al. Design and development of C o3O4/NiO composite nanofibers for the application of highly sensitive and selec-
tive non-enzymatic glucose sensors. RSC Adv. 5, 76538–76547 (2015).
24. Liu, Y. et al. Mesoporous ZnO-NiO architectures for use in a high-performance nonenzymatic glucose sensor. MicrochimicaActa
181, 1581–1589 (2014).
25. Zhou, Q.-Q. et al. Controllable synthesis of barnyardgrass-like CuO/Cu2O heterostructure nanowires for highly sensitive non-
enzymatic glucose sensors. J. Mater. Chem. C7, 14874–14880 (2019).
26. Haghparas, Z., Kordrostami, Z., Sorouri, M., Rajabzadeh, M. & Khalifeh, R. Highly sensitive non-enzymatic electrochemical
glucose sensor based on dumbbell-shaped double-shelled hollow nanoporousCuO/ZnO microstructures. Sci. Rep. 11, 344 (2021).
27. Chandrasekaran, N. I. & Matheswaran, M. Unique nonenzymatic glucose sensor using a hollow-shelled triple oxide Mn–Cu–Al
nanocomposite. ACS Omega 5, 23502–23509 (2020).
28. Lohar, G. M., Pore, O. C. & Fulari, A. V. Electrochemical behavior of CuO/rGOnanopellets for flexible supercapacitor, non-
enzymatic glucose, and H2O2 sensing application. Ceram. Int. 47, 16674–16687 (2021).
29. Zhou, D.-L. et al. Facile synthesis of monodisperse porous Cu2O nanospheres on reduced graphene oxide for non-enzymatic
amperometric glucose sensing. Electrochim. Acta 115, 103–108 (2014).
30. Subash, V. S., Alagumalai, K., Chen, S.-M., Shanmugam, R. & Shiuan, H. J. Ultrasonication assisted synthesis of NiO nanoparticles
anchored on graphene oxide: An enzyme-free glucose sensor with ultrahigh sensitivity. New J. Chem. 44, 15071–15080 (2020).
31. Arikan, K., Burhan, H., Bayat, R. & Sen, F. Glucose nano biosensor with non-enzymatic excellent sensitivity prepared with nickel–
cobalt nanocomposites on f-MWCNT. Chemosphere 291, 132720 (2022).
32. Masoomi-Godarzi, S., Khodadadi, A. A., Vesali-Naseh, M. & Mortazavi, Y. Highly stable and selective non-enzymatic glucose
biosensor using carbon nanotubes decorated by F e3O4 nanoparticles. J. Electrochem. Soc. 161, B19–B25 (2013).
33. Zeng, G., Li, W., Ci, S., Jia, J. & Wen, Z. Highly dispersed NiO nanoparticles decorating graphene nanosheets for non-enzymatic
glucose sensor and biofuel cell. Sci. Rep. 6, 36454 (2016).
34. Liu, W., Zhao, X., Dai, Y. & Qi, Y. Preparation of three dimensional C u2O/Au/GO hybrid electrodes and its application as a non-
enzymatic glucose sensor. Microchem. J. 179, 107451 (2022).
35. Gayathri, S., Arunkumar, P., Kim, J. & Han, J. H. Bimetallic layered hydroxide Nitrate@Graphene oxide as an electrocatalyst for
efficient non-enzymatic glucose sensors: tuning sensitivity by hydroxide-regulated M 2(OH)4–n(An–) phases derived from solvent
engineering. ACS Sustain. Chem. Eng. 10, 1689–1701 (2022).
36. Hari Krishna, R. et al. Non-enzymatic catalytic oxidation of glucose and dual mode sensing by fluorescence/electrochemical
methods using MO–GO composites (MO = ZnO, CuO, NiO and Co3O4). Top. Catal. (2022).
37. Güneş, F. et al. Synthesis of hierarchical hetero-composite of graphene foam/α-Fe2O3 nanowires and its application on glucose
biosensors. J. Alloy. Compd. 895, 162688 (2022).
38. Younis, S. et al. Role of functionalized metal oxide–carbon nanocomposites in biomolecule detection. In Metal oxide-carbon hybrid
materials (eds Chaudhry, M. A. et al.) 495–527 (Elsevier, 2022).
39. Hu, T. et al. Anisotropic electronic conduction in stacked two-dimensional titanium carbide. Sci. Rep. 5, 16329 (2015).
40. Zhu, X., Zhang, Y., Liu, M. & Liu, Y. 2D titanium carbide MXenes as emerging optical biosensing platforms. Biosens. Bioelectron.
171, 112730 (2021).
41. Huang, R., Liao, D., Chen, S., Yu, J. & Jiang, X. A strategy for effective electrochemical detection of hydroquinone and catechol:
Decoration of alkalization-intercalated T i3C2 with MOF-derived N-doped porous carbon. Sens. Actuators B Chem. 320, 128386
(2020).
42. Rakhi, R. B., Nayak, P., Xia, C. & Alshareef, H. N. Novel amperometric glucose biosensor based on MXene nanocomposite. Sci.
Rep. 6, 36422 (2016).
43. Gopal, T. S. et al. MXene-based composite electrodes for efficient electrochemical sensing of glucose by non-enzymatic method.
Mater. Today Chem. 24, 100891 (2022).
44. Ni, M. et al. A high-sensitive dopamine electrochemical sensor based on multilayer T i3C2 MXene, graphitized multi-walled carbon
nanotubes and ZnOnanospheres. Microchem. J. 178, 107410 (2022).
Vol.:(0123456789)
www.nature.com/scientificreports/
45. Huang, H. et al. Fabrication of rGO/MXene-Pd/rGO hierarchical framework as high-performance electrochemical sensing platform
for luteolin detection. MicrochimicaActa 189, 59 (2022).
46. Xia, Y., Hu, X., Liu, Y., Zhao, F. & Zeng, B. Molecularly imprinted ratiometric electrochemical sensor based on carbon nanotubes/
cuprous oxide nanoparticles/titanium carbide MXene composite for diethylstilbestrol detection. MicrochimicaActa 189, 137 (2022).
47. Feng, X., Han, G., Cai, J. & Wang, X. Au@Carbon quantum Dots-MXenenanocomposite as an electrochemical sensor for sensitive
detection of nitrite. J. Colloid Interface Sci. 607, 1313–1322 (2022).
48. Hummers, W. S. & Offeman, R. E. Preparation of graphitic oxide. J. Am. Chem. Soc. 80, 1339 (1958).
49. Zhou, T. et al. Super-tough MXene-functionalized graphene sheets. Nat. Commun. 11, 2077 (2020).
50. Asia, S. et al. 2D titanium carbide (MXene) for wireless communication. Sci. Adv. 4, eaau0920 (2022).
51. Sarycheva, A. & Gogotsi, Y. Raman spectroscopy analysis of the structure and surface chemistry of Ti3C2Tx MXene. Chem. Mater.
32, 3480–3488 (2020).
52. Ma, Z., Zhou, X., Deng, W., Lei, D. & Liu, Z. 3D porous MXene (Ti3C2)/reduced graphene oxide hybrid films for advanced lithium
storage. ACS Appl. Mater. Interfaces 10, 3634–3643 (2018).
53. Yan, J. et al. Flexible MXene/graphene films for ultrafast supercapacitors with outstanding volumetric capacitance. Adv. Func.
Mater. 27, 1701264 (2017).
54. Shao, L. et al. MXene/RGO composite aerogels with light and high-strength for supercapacitor electrode materials. Compos. Com‑
mun. 19, 108–113 (2020).
55. Thirumal, V. et al. Heterostructured two dimensional materials of MXene and graphene by hydrothermal method for efficient
hydrogen production and HER activities. Int. J. Hydrog. Energy (2021).
56. Yang, Y.-Y. et al. Terminal groups-dependent near-field enhancement effect of Ti3C2Tx nanosheets. Nanoscale Res. Lett. 16, 60
(2021).
57. Shuck, C. E. et al. Scalable synthesis of T i3C2Tx MXene. Adv. Eng. Mater. 22, 1901241 (2020).
58. Lin, P. et al. Hybrid reduced graphene oxide/TiO2/graphitic carbon nitride composites with improved photocatalytic activity for
organic pollutant degradation. Appl. Phys. A 124, 510 (2018).
59. Liu, C. et al. Creating MXene/reduced graphene oxide hybrid towards highly fire safe thermoplastic polyurethane nanocomposites.
Compos. B Eng. 203, 108486 (2020).
60. Elancheziyan, M., Manoj, D., Saravanakumar, D., Thenmozhi, K. & Senthilkumar, S. Amperometric sensing of catechol using a
glassy carbon electrode modified with ferrocene covalently immobilized on graphene oxide. MicrochimicaActa 184, 2925–2932
(2017).
61. Xu, J., Sun, Y. & Zhang, J. Solvothermal synthesis of F e3O4 nanospheres for high-performance electrochemical non-enzymatic
glucose sensor. Sci. Rep. 10, 16026 (2020).
62. Li, R., Deng, X. & Xia, L. Non-enzymatic sensor for determination of glucose based on PtNi nanoparticles decorated graphene.
Sci. Rep. 10, 16788 (2020).
63. Shabnam, L. et al. Doped graphene/Cu nanocomposite: A high sensitivity non-enzymatic glucose sensor for food. Food Chem.
221, 751–759 (2017).
64. Karikalan, N., Karthik, R., Chen, S.-M., Karuppiah, C. & Elangovan, A. Sonochemical synthesis of sulfur doped reduced graphene
oxide supported CuS nanoparticles for the non-enzymatic glucose sensor applications. Sci. Rep. 7, 2494 (2017).
65. Wu, S., Liu, L., Wang, S., Wang, P. & Guo, H. Copper oxide-carboxylated graphene nanocrystals modified electrode for sensitive
detection of glucose. Mater. Lett. 316, 132011 (2022).
66. Zhang, Y. et al. A flexible non-enzymatic glucose sensor based on coppernanoparticles anchored on laser-induced graphene.
Carbon 156, 506–513 (2020).
67. Guo, S. et al. Three-dimensional flexible polyurethane decorated with Ni and reduced graphene oxide for high-sensitive sensing
of glucose. Mater. Chem. Phys. 278, 125679 (2022).
68. Justice Babu, K., Sheet, S., Lee, Y. S. & Gnanakumar, G. Three-dimensional dendrite Cu–Co/reduced graphene oxide architectures
on a disposable pencil graphite electrode as an electrochemical sensor for nonenzymatic glucose detection. ACS Sustain. Chem.
Eng. 6, 1909–1918 (2018).
69. Phetsang, S. et al. Copper/reduced graphene oxide film modified electrode for non-enzymatic glucose sensing application. Sci.
Rep. 11, 9302 (2021).
70. Li, M. et al. Three-dimensional porous MXene/NiCo-LDH composite for high performance non-enzymatic glucose sensor. Appl.
Surf. Sci. 495, 143554 (2019).
71. Yang, W. L. et al. Improving linear range limitation of non-enzymatic glucose sensor by OH- concentration. Crystals 10, 186 (2020).
72. Liu, H. et al. Glucose-reduced graphene oxide with excellent biocompatibility and photothermal efficiency as well as drug loading.
Nanoscale Res. Lett. 11, 1–10 (2016).
Acknowledgements
The authors extend their appreciation to the Deputyship for Research & Innovation, Ministry of Education in
Saudi Arabia for funding this research work through the project no. (IFKSURG-2-975).
Author contributions
Conceptualization—T.G., A.N.G. Methodology—T.G., A.N.G. Validation—T.G., A.N.G. Formal analysis—
K.E.A., A.K.A., H.A., A.A., M.R.M., S.P. Investigation—A.N.G. Writing—Review and Editing—T.G., A.N.G.
Supervision—T.G., A.N.G. Project administration—K.E.A., A.K.A., H.A., A.A., M.R.M., S.P.
Competing interests
The authors declare no competing interests.
Additional information
Supplementary Information The online version contains supplementary material available at https://doi.org/
10.1038/s41598-022-24700-w.
Correspondence and requests for materials should be addressed to A.N.G.
Reprints and permissions information is available at www.nature.com/reprints.
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and
institutional affiliations.
Vol:.(1234567890)
www.nature.com/scientificreports/
Open Access This article is licensed under a Creative Commons Attribution 4.0 International
License, which permits use, sharing, adaptation, distribution and reproduction in any medium or
format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the
Creative Commons licence, and indicate if changes were made. The images or other third party material in this
article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the
material. If material is not included in the article’s Creative Commons licence and your intended use is not
permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from
the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Vol.:(0123456789)