ISO 10993-12-2021
ISO 10993-12-2021
Fifth edition
2021-01
Reference number
ISO 10993-12:2021(E)
© ISO 2021
Copyrighted material licensed to University of Toronto by Techstreet LLC, subscriptions.techstreet.com, downloaded on 2021-04-28 19:40:25 +0000 by
No further reproduction or distribution is permitted.
ISO 10993-12:2021(E)
Contents Page
Foreword......................................................................................................................................................................................................................................... iv
Introduction...................................................................................................................................................................................................................................v
1 Scope.................................................................................................................................................................................................................................. 1
2 Normative references....................................................................................................................................................................................... 1
3 Terms and definitions...................................................................................................................................................................................... 1
4 General requirements...................................................................................................................................................................................... 3
5 Reference materials (RMs)......................................................................................................................................................................... 4
5.1 General............................................................................................................................................................................................................ 4
5.2 Certification of RMs for biological safety testing........................................................................................................ 4
6 Use of RMs as experimental controls............................................................................................................................................... 4
7 Test sample selection........................................................................................................................................................................................ 5
8 Test sample and RM preparation......................................................................................................................................................... 5
9 Selection of representative portions from a medical device................................................................................... 6
10 Preparation of extracts of samples..................................................................................................................................................... 6
10.1 General............................................................................................................................................................................................................ 6
10.2 Containers for extraction................................................................................................................................................................ 6
10.3 Extraction conditions and methods....................................................................................................................................... 7
10.4 Extraction conditions for materials that polymerize in situ ....................................................................... 10
11 Records......................................................................................................................................................................................................................... 10
Annex A (informative) Experimental controls.........................................................................................................................................11
Annex B (informative) General principles on, and practices of, test sample preparation and
sample selection.................................................................................................................................................................................................13
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT) see www.iso.org/
iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 194, Biological and clinical evaluation of
medical devices, in collaboration with the European Committee for Standardization (CEN) Technical
Committee CEN/TC 206, Biological and clinical evaluation of medical devices, in accordance with the
Introduction
It is important that sample preparation methods be appropriate for both the biological evaluation
methods and the materials being evaluated. Each biological test method requires the selection of
materials, extraction solvents and conditions.
This document is based on existing national and international standards and regulations, wherever
possible.
1 Scope
This document specifies requirements and gives guidance on the procedures in the preparation of
samples and the selection of reference materials for medical device testing primarily in biological test
systems primarily in accordance with one or more parts of the ISO 10993 series.
Specifically, this document addresses the following:
— test sample selection;
— selection of representative portions from a medical device;
— test sample preparation;
— experimental controls;
— selection of, and requirements for, reference materials;
— preparation of extracts.
This document is not applicable to live cells but can be relevant to the material or medical device
components of combination products containing live cells.
2 Normative references
There are no normative references in this document.
3.2
CRM
certified reference material
reference material (RM) characterized by a metrologically valid procedure for one or more specified
properties, accompanied by an RM certificate that provides the value of the specified property, its
associated uncertainty, and a statement of metrological traceability
Note 1 to entry: The concept of value includes a nominal property or a qualitative attribute such as identity or
sequence. Uncertainties for such attributes may be expressed as probabilities or levels of confidence.
Note 2 to entry: Metrologically valid procedures for the production and certification of RMs are given in, among
others, ISO 17034 and ISO Guide 35.
3.4
exhaustive extraction
extraction conducted until the amount of extractable material in a subsequent extraction is less than
10 % by gravimetric analysis (or that achieved by other means) of that detected in the initial extraction
Note 1 to entry: As it is not possible to demonstrate the exhaustive nature of residual recovery, the definition of
3.5
experimental control
substance with well-characterized responses, which is used in a specific test system to assist in
evaluating if the test system has responded in a reproducible and appropriate manner
3.6
extract
liquid that results from extraction of the test sample or control
3.7
extractable substance
substance that can be released from a medical device or material using either extraction solvents
or extraction conditions, or both, that are expected to be at least as aggressive as the conditions of
clinical use
3.8
homogeneity
consistency of a material's chemical and physical compositions, and uniformity in response to a
biological endpoint
Note 1 to entry: A reference material is said to be homogeneous if the biological response in a specific test is
found to lie within the specified uncertainty limits of the test, irrespective of the batch or lot of material from
which the test sample is extracted.
3.9
leachable substance
substance that can be released from a medical device or material during clinical use
3.10
negative control
well-characterized material and/or substance, which, when tested by a specific procedure, demonstrates
the suitability of the procedure to yield a reproducible, appropriately negative, non-reactive or minimal
response in the test system
Note 1 to entry: In practice, negative controls are reference materials but can include blanks and extraction
vehicles/solvents.
3.11
positive control
well-characterized material and/or substance, which, when evaluated by a specific test method,
demonstrates the suitability of the test system to yield a reproducible, appropriately positive or reactive
response in the test system
3.12
reference material
RM
material, sufficiently homogeneous and stable with respect to one or more specified properties, which
has been established to be fit for its intended use in a measurement process
Note 1 to entry: RM is a generic term.
Note 2 to entry: Properties can be quantitative or qualitative, e.g. identity of substances or species.
Note 3 to entry: Uses may include the calibration of a measurement system, assessment of a measurement
procedure, assigning values to other materials, and quality control.
Note 4 to entry: ISO/IEC Guide 99:2007 has an analogous definition (5.13), but restricts the term “measurement”
to apply to quantitative values. However, Note 3 of ISO/IEC Guide 99:2007, 5.13 (VIM), specifically includes
qualitative properties, called “nominal properties”.
Note 5 to entry: The laboratory is to demonstrate that the simulated-use extraction is carried out under
[SOURCE: ISO Guide 30:2015, 2.1.1 — Note 5 to entry has been added.]
3.13
stability
characteristic of a material, when stored under specified conditions, to maintain a specified property
value within specified limits for a specified period of time
Note 1 to entry: See also the IUPAC Compendium of Analytical Nomenclature[5].
3.14
test sample
medical device, component or material (or a representative sample thereof, manufactured and
processed by equivalent methods), or an extract or portion thereof that is subjected to biological
evaluation testing
4 General requirements
When identifying hazards and estimating risk in relation to medical devices, hazards that arise from
changes in the manufacturing process, or insufficient control of the manufacturing process, shall be
considered in the design and preparation of test samples, as described in ISO 14971. Particular attention
shall be given to material additives, unintentional base material impurities and manufacturing process
residues, e.g. trace elements and cleaning and disinfection agents.
The ISO 10993 series describes many different biological assay systems. Therefore, the individual parts
shall be consulted to ascertain whether these are appropriate for specific test systems.
Experimental controls shall be used in biological evaluations carried out in order to either validate a
test procedure or compare the results between materials, or both. Depending on the specifications of a
particular test, either negative controls, blanks or positive controls, or all three, shall be used.
NOTE The same type of control can be applicable to different tests and can allow cross-reference to other
established materials and test methods. Additional guidance on the selection of experimental controls is given in
Annex A. Use of positive controls for in vivo testing might be affected by animal welfare regulations.
5.1 General
RMs are established by individual laboratories. The extent of chemical, physical and biological
characterization is determined by the individual laboratory. Commercially available articles may be
used as RMs.
NOTE See also ISO Guide 35.
CRMs are selected for their high purity, critical characteristics, suitability for the intended purpose and
general availability. The critical chemical, physical and biological characteristics shall be determined
by collaborative testing in three or more laboratories and made available to the investigator by the
distributor.
It is desirable for users to obtain a commitment from suppliers of RMs or CRMs stating that these
materials will be available to the user for at least five years. A second but less desirable option is for the
source of the RM or CRM to publish an “open formulation” for the material, i.e. publication of the source
materials and details of the processing needed to ensure uniform batches of the RM.
Suppliers of RMs shall certify the materials. The supplier determines the extent of chemical and physical
characterization that is performed. The individual laboratories that use the RM shall identify the
biological characterization necessary to qualify an RM for a specific test or procedure. Commercially
available materials may be used as RMs, provided they are certified and qualified.
Certification of a RM is a procedure that establishes the numerical or qualitative value of the biological
response of the material under the specified test conditions. This process serves to validate the testing
of the material for that particular response and results in the issuance of a certificate. The biological
response of the material shall be established through interlaboratory tests.
example, delayed-type hypersensitivity, shall not be used as an RM for another, for example, cytotoxicity,
without additional validation.
NOTE The use of an RM will facilitate the comparability of the response between laboratories and help
assess reproducibility of the test performance within individual laboratories. For comparison of the biological
response, it is desirable to use RMs having a range of responses, e.g. minimum, intermediate or severe.
RMs used as experimental controls shall meet the established quality assurance procedures of the
manufacturer and test laboratory. They shall be identified in relation to source, manufacturer, grade
and type. RMs are processed as described in Clause 8.
When RMs are used as experimental controls, they shall be in the same material class as the test
sample, i.e. polymer, ceramic, metal, colloid. However, pure chemicals may be used as experimental
controls for mechanistically-based test procedures, for example, genotoxicity and immune delayed-
type hypersensitivity assays.
For absorbable materials that could potentially have toxic degradants and residuals, testing of
intermediate products should be considered.
The same test sample selection procedure applies when an extract is required.
— Test samples from sterilized medical devices and RMs shall be handled aseptically, if appropriate to
the test procedure.
— Test samples which are clean, sterile and disinfected, shall be processed by the method recommended
by the manufacturer and handled aseptically, if appropriate to the test procedure.
— The influence of the cleaning process and cleaning agent shall be considered in the selection and
handling of the test sample.
Test samples from medical devices not required to be sterile in use shall be used as supplied and
handled aseptically throughout the test sample preparation. If sterile test samples are required for a
test procedure, e.g. for cytotoxicity testing, the effect of the sterilization or resterilization process on
the test sample and RM shall be considered.
When test samples and RMs need to be cut into pieces, as described in 10.3.3, the influence of previously
unexposed surfaces, e.g. lumens or cut surfaces shall be considered. Tools used for cutting medical
devices into representative portions for testing shall be cleaned between uses to prevent contamination.
Furthermore, care shall be taken that the tool itself doesn’t contaminate the device.
9.1 If a medical device cannot be tested as a whole, each individual material in the final product that is
required to be tested shall be represented proportionally in the test sample.
— The test sample of the medical devices with surface coatings shall include both the coating material
and the substrate, even if the substrate has no tissue contact.
— The test sample shall include a representative portion of the joint or seal, or both, if adhesives,
radiofrequency (RF) seals or solvent seals are used in the manufacture of a portion of the medical
device which comes into contact with patients.
9.3 When different materials are present in a single medical device, the potential for synergies and
interactions shall be considered in the choice of test sample.
9.4 The test sample shall be chosen to maximize the exposure of the test system to the components of
a medical device that are known to have potential for a biological response.
9.5 Non-patient contacting portions of the medical device should, if possible, be excluded either
physically from test sample extracts or by exclusion of the surface area in the calculation of the extraction
ratio. When this is not possible, the extraction ratio shall be justified. Ensure that all contacting portions
are covered by the selected extraction vehicle volume.
Clinician and user surface contact with materials other than those in common use in consumer products
with a similar nature of contact, should be considered [see ISO 10993-1:2018, 5.2.2, a)].
9.6 Medical device components with different type or duration of tissue contact might need to be
extracted and tested separately.
10.1 General
If extracts of the medical device are required for a test procedure, the extraction vehicles and conditions
of extraction used shall be appropriate to the nature and use of the final product and to the purpose of
the test, for example, hazard identification, risk estimation or risk assessment. The physico-chemical
properties of the medical device materials, leachable substances or residues shall be considered when
choosing the extraction conditions (see ISO 10993-18 and ISO/TS 10993-19). For more information on
sample preparation for testing of nanomaterials or nanostructured materials, see ISO/TR 10993-22.
NOTE For additional guidance on the extraction of samples, see Annex C.
10.3.1 Extraction conditions are based on common practice and are justified on the basis of providing
a standardized approach that is, in many ways, an appropriate exaggeration of product use. Extraction
shall be conducted under one of the following conditions (see also C.5):
10.3.2 For materials that are intended to dissolve or absorb under conditions of use, the selection of
extraction conditions described in 10.3 might need to consider the thermal properties (e.g. glass transition
temperature of polymers) of device materials of construction and the relevant clinical use conditions. For
these materials, the extracts prepared based on 10.3 may have changes either in osmolarities or in the
pH that may not be appropriate for the test system to be dosed. Any adjustment applied to the extracts
prior to biocompatibility testing should be justified.
NOTE For more information on sample preparation for testing of absorbable medical devices, see
ISO 10993-3, ISO 10993-6, ISO 10993-13, ISO 10993-14, ISO 10993-15, ISO 10993-18 and ISO/TS 37137-1.
Perform extraction using the appropriate extraction vehicle and the conditions of time and temperature
to simulate exaggerated exposure wherever possible. Complete dissolution using the extraction vehicles
and conditions recommended by this document can be appropriate, if justified; however, caution should
be taken since complete device dissolution can create challenges for subsequent biological testing
(e.g. difficulty in dosing animals with neat test extract if viscosity is increased, difficulty in interpreting
in vitro cell-based test failure data in case of increased osmolality or pH change). For chemical
characterization and hazard assessment of potential intermediate degradants that cannot otherwise
be evaluated under these testing conditions, see ISO 10993-17 and ISO 10993-18.
10.3.3 The standard surface area can be used to determine the volume of extraction vehicle needed.
This area includes the combined area of all tissue contacting surfaces of the sample and ignores the
contribution of indeterminate surface irregularities. When the surface area cannot be determined due to
configuration of the sample, a mass/volume of extracting fluid shall be used. See Table 1.
Other surface-area-to-volume extraction ratios, for example, those related to evaluation of porous
materials can be used if they simulate the conditions during clinical use or result in a measure of the
hazard potential (ISO/TS 10993-19 describes the tests for the morphological characterization of porous
materials.)
Materials may be cut into small pieces before the extraction to enhance submersion in the extract media,
except when otherwise inappropriate (see 10.3.4). For example, for polymers, pieces approximately
10 mm × 50 mm or 5 mm × 25 mm are appropriate.
10.3.4 Elastomers, coated materials, composites, laminates shall be tested intact whenever possible
because of potential differences in extraction characteristics between the intact and cut surfaces. For
materials such as elastomers and latex, it may be appropriate to use an extraction ratio of 1,25 cm2/ml, if
justified.
10.3.5 Extraction using both polar and non-polar extraction vehicles shall be performed. In some device
specific circumstances, it may be appropriate to extract in only one extraction vehicle, either polar or
non-polar. Where extraction is in one vehicle only, rationale shall be provided. The following are examples
of extraction vehicles:
EXAMPLE 1 For polar extraction vehicle are water, physiological saline, culture media without serum.
EXAMPLE 2 For non-polar extraction vehicle are freshly refined vegetable oil (e.g. cottonseed or sesame oil) of
the quality defined in various pharmacopoeias.
NOTE 2 Other extraction vehicles appropriate to the nature and use of the medical device or to the methods
for hazard identification can also be used if their effects on the material and the biological system are known
(see Annex D).
NOTE 3 An example of a device only requiring a single (polar) extract is a syringe pre-filled only with saline.
The use of a culture medium with serum is preferred for extraction in testing for cytotoxicity because
of its ability to support cellular growth as well as extract both polar and non-polar substances
10.3.6 Extractions should be performed with continuous, mechanical agitation or circulation. When
extraction under static conditions or intermittent agitation is considered to be appropriate, the method
shall be justified, specified and reported. Care should be taken to not to damage the sample or the
container.
10.3.7 Liquid extracts should, if possible, be used immediately after preparation to prevent sorption on
to the extraction container or other changes in composition. If an extract is stored for longer than 24 h
(e.g. refrigerated at 2 °C to 8 °C) then the stability and the homogeneity of the extract under the storage
conditions shall be verified.
10.3.9 The extract shall not routinely be processed by filtration, centrifugation or other methods
to remove suspended particulates. However, if such processing is necessary, the rationale shall be
documented.
10.3.10 For materials or medical devices not expected to dissolve or absorb under conditions
of use, any solvents used in the extraction of a polymeric material or medical device shall not cause
dissolution of the polymer formulation. No softening or deformation of the polymeric material shall occur
in the presence of the volatile solvent. Selected solvents should not compromise (e.g. severe swelling,
particulate generation and degradation) the medical material or devices. The solvent shall be removed
(prior to use in a bioassay) to the extent that any residues do not adversely affect the biological assay
(e.g. cause protein denaturation or skin irritation). For materials or medical devices expected to dissolve
or absorb under conditions of use, see 10.3.2, 10.3.11, and C.7.
10.3.11 For solution and soluble materials, the standard extraction methods used for insoluble
materials might be inappropriate. The following guidance should be considered in addition to information
contained in Table 1.
a) Factors such as test system compatibility, route of administration and extent of dissolution or
degradation should be considered in the final preparation for testing. Use an appropriate vehicle
and conditions to simulate exaggerated exposure wherever possible. A pre-test can help to
determine appropriate conditions.
b) If the material completely dissolves, in a vehicle or diluent that is compatible with the material
and the test system, the resulting solution can be evaluated neat, provided the solution properties
are also compatible with the test system, e.g. pH, osmolarity, solute concentrations. If the resulting
solution contains all the constituents of the material, a second vehicle would not be needed.
c) If the material is an aqueous solution and used in this form, it shall be tested directly and not extracted,
provided the solution properties are compatible with the test system [see also a) and b) above].
d) OECD Guidelines for the Testing of Chemicals[16], or similar chemical testing standards, can be
used as guidance in determining maximum concentrations of test substances used for specific test
methods.
10.3.12 Where fluids circulate through the medical device under normal conditions of use, e.g.
extra-corporeal devices, extraction via re-circulation can be used. When possible, one or more of the
conditions shall be exaggerated, for example, temperature, time, volume, flow rate. The rationale for the
extraction chosen shall be reported.
11 Records
Records of the sample and its preparation shall include, but not be limited to:
— Type and, if known, composition of material, source of material, medical device, medical device
portion or component (a written description, drawing, photograph or other methods can achieve all
or part of this requirement):
— lot or batch number, where appropriate;
— description of processing, cleaning or sterilization treatments, if appropriate;
— extraction techniques, as appropriate, including documentation of extraction vehicle, extraction
ratios, conditions for extraction, means of agitation, as well as any deviations from the conditions
specified in this document, such as filtration of the extract or extraction media. The condition
of the test extract (e.g. color, clarity, presence of any particulates) shall also be described, and
photographed if applicable.
— Documents shall be provided (e.g. schematic or photo) of the medical device components that are
sampled, and those that are not sampled.
Annex A
(informative)
Experimental controls
A.1 The materials listed in the following paragraphs might meet the criteria for an appropriate
experimental control in selected tests. It is the responsibility of the investigator to make the appropriate
choices (see Table A.1).
A.2 Materials that have been used as negative controls or RMs are, for example, high-density
polyethylene1) low-density polyethylene2), silica-free polydimethylsiloxane3),4), polypropylene5),
aluminium oxide ceramic rods, stainless steel and commercially pure (cp) titanium alloys.
1) High-density polyethylene (negative control plastic RS), HDPE film, RM-C and HDPE rod, RM-E (are) an
example(s) of a suitable product(s) available commercially. This information is given for the convenience of users of
this document and does not constitute an endorsement by ISO of these products.
2) PE 140 tubing: RAUMEDIC AG, is an example of a suitable product available commercially. This information is
given for the convenience of users of this document and does not constitute an endorsement by ISO of this product.
3) Biomaterials Program, Devices and Technology Branch, National Heart, Lung and Blood Institute, NIH Building,
7550 Wisconsin Ave., Bethesda, MD 20892, USA.
4) SIK 8363 tubing: RAUMEDIC AG is an example of a suitable product available commercially. This information is
given for the convenience of users of this document and does not constitute an endorsement by ISO of this product.
5) PP 146 tubing: RAUMEDIC AG is an example of a suitable product available commercially. This information is
given for the convenience of users of this document and does not constitute an endorsement by ISO of this product.
A.3 Materials that have been used as positive controls are, for example, polyvinylchloride-containing
organotin additives6), segmented polyurethane rod7) or films containing zinc diethyldithiocarbamate
(SPU-ZDEC)8) or dibutyldithiocarbamate (SPU-ZDBC), and PVC with a known irritant substance (Genapol
X-080)9), as well as certain latex formulations, solutions of zinc salts, and copper. Substances that have
been used as positive controls for extract samples are dilutions of phenol and water.
6) Positive Control Material, code 499-300-000-000: Portex Limited is an example of a suitable product available
commercially. This information is given for the convenience of users of this document and does not constitute an
endorsement by ISO of this product.
7) Polyurethane rod — ZDEC: RM-F, is an example of a suitable product available commercially. This information
is given for the convenience of users of this document and does not constitute an endorsement by ISO of this product.
8) Polyurethane film — ZDEC: RM-A and Polyurethane film — ZDBC (SPU-ZDBC): RM-B are examples of suitable
products available commercially. This information is given for the convenience of users of this document and does
not constitute an endorsement by ISO of these products.
9) Genapol X-080 — Y-3 Pellet and Genapol X-080 — Y-4 Pellet are examples of suitable products available
commercially. This information is given for the convenience of users of this document and does not constitute an
endorsement by ISO of these products.
Annex B
(informative)
The material used in the biological assay should be representative of the composition and surface
characteristics of the final product and of the processes used in its manufacture (see Clause 7).
Documentation of the composition of plastic and rubber materials should include identification of
the resin, polymer and any additives. The formulation description should specify the history of the
material, for example, information on the thermal processing and whether it is virgin or reground and,
if reground, the specification for the maximum allowable regrind.
Materials that may be re-sterilized by the same or alternative methods should be tested after treatment
by the multiple sterilizations.
For example, a material that is sterilized by radiation and re-sterilized by ethylene oxide should be
tested in the final product after radiation and ethylene oxide sterilization.
If a “worst case” exposure can be identified with appropriate justification, testing may be performed
after exposure to this treatment.
Ideally, all biological tests which use a material cut from a medical device, the medical device
component itself as the test material, or an extract prepared from either, should be performed with the
surface of the material exposed to the test system’s biological environment. An alternative method to
cutting the surface is the fabrication of miniatures of the medical device, using the same processes (e.g.
extrusion, dipping), temperatures, time, atmosphere, release agents, and processes such as annealing,
Extraction conditions for implant materials that can cause particle generation in vivo during clinical use
should be considered in the design of tests on the materials. The effect of extraction procedures should
be considered when designing tests for materials where particulates are generated by the extraction
conditions.
The amount of material, and surface area thereof, should be appropriate to the biological and physical
constraints of the test system. In practice, the use of a standard sample size for a specific assay is
recommended.
Users of this document are directed to the discussion of “proper use” and “misuse” of CRMs in the
introduction to ISO Guide 33. This discussion points out areas of both potential under- and over-
utilization of RMs and CRMs. Users of this document should also note that the use of calibration
materials to evaluate the biological response of materials under investigation within a single laboratory
is acceptable.
Annex C
(informative)
WARNING — Application of the test methods of this document to medical device materials
comprising proteins shall be made with great care to ensure that the extraction procedure has
not altered the biological properties of the materials being extracted.
C.1 Extraction of a medical device is carried out to provide a suitable test sample for the biological
evaluation testing.
If extracts of the medical device are prepared, the extraction vehicle and conditions of extraction used
should be appropriate to the nature and use of the final product, as well as to the predictability (e.g. test
purpose, rationale, sensitivity) of the test method. Extraction conditions and application of the extract
to test systems should therefore ideally reflect not only actual conditions of use of the products but also
the purpose and predictability of the tests.
Under normal conditions of use where fluids circulate through the medical device (e.g. extra-corporeal
devices), if a vertical standard exists, it should be consulted for the appropriate extraction techniques.
Biological tests are carried out to identify hazards and estimate the risks of the hazards occurring in
exaggerated use or in actual conditions of use, or both.
C.2 This document assumes that the amount of extractable substance(s) is (are) related to the period
of extraction, the temperature, the ratio of the surface area of the material to the volume of the extraction
vehicle, and the nature of the extraction vehicle.
C.4 Extraction temperatures can vary for the different materials to be tested. Extraction should not cause
significant degradation of the material, unless the material is intended to dissolve or be resorbed during
use (see 10.3.2). The extraction temperature is dependent upon the physico-chemical characteristics of
the medical device material(s). The extraction temperature chosen for polymers, for example, should be
below the glass transition temperature. If the glass transition temperature is below the use temperature,
the extraction temperature should be below the melting temperature. Recommended conditions are
given in 10.3.1.
For polymeric absorbable materials, extraction above in vivo temperatures that are near or above the
glass transition temperature may lead to changes in the polymer properties (e.g. degradation) that
are not representative of clinical conditions and should be avoided. For absorbable metals, elevated
extraction temperatures can introduce new and potentially unrepresentative corrosion mechanisms.
Thus, for most absorbable polymers and metals, the standard extraction temperatures listed in 10.3.1
may not be applicable. When evaluating absorbable devices, extraction of partially degraded materials
and their related intermediate degradation products can be considered.
C.5 The ratio of the surface area of the medical device to the volume of extraction vehicle or solvent
should be sufficient to:
— attain the maximum amount of extractable substance(s) in an appropriate dosage volume for
biological testing (i.e. dosage volume within physiological limits);
— demonstrate the hazard potential of using the medical device in humans;
— cover the material in the solvent volume.
In practice, the use of a standard area and solvent volume is recommended (described in 10.3.3) in
lieu of device-specific parameters. Some test methods require concentration of extracts to increase the
C.8 A standard prescription cannot be constructed to address the specialized needs of preparing
extracts of polymerized in situ products. The individual components, time to polymerization, intended
use and the extraction vehicles should be taken into consideration when developing a relevant extract.
Language should include the recommendation that the polymerization kinetics be used in the design of
the correct methodology to develop a relevant extract for testing. The uncured components should be
considered when selecting an appropriate solvent for extracting the sample.
Annex D
(informative)
D.1 General
Polymeric materials often contain a small amount of low-molecular-weight chemical substances
(LMWCs) such as catalysts, processing aids, or other additives[19], residual monomers, or oligomers.
A major toxicological concern during the biological evaluation of polymeric materials is the toxicity
of any leachables that can migrate from the polymer to the human body during use. This concept is
derived from the agreement of the OECD polymer group regarding health concerns and the exemption
of polymers from testing (see Reference [15]). The report pointed out the following four parameters,
which are important for judging the health risk of polymers:
— number-average molecular weight of the polymer;
— content of low-molecular-weight chemical species;
— presence of reactive functional groups (see Reference [17]);
— presence of bioavailable metals.
NOTE LMWCs are defined as low-molecular-weight chemical substances with a molecular weight not
exceeding 1 000 Da.
solvents for exhaustive extraction of polymeric medical devices be chosen on a case-by-case basis
(see Reference [25]).
The following are important points to be considered for exhaustive extraction of polymeric medical
devices for biological testing.
— For biocompatibility evaluation of medical devices, the extraction of the medical device in both
polar and non-polar solvents should be considered for hazard identification purposes. The selected
solvent(s) should not compromise (e.g. severe swelling, particulate generation and degradation) the
medical devices. Also, a rationale should be provided for the chosen extraction temperature.
— For exhaustive extractions, the duration of the extraction cannot be prescribed in advance but
can be established in the following manner. A series of successive extractions is carried out by
extracting the test sample in the selected solvent for a fixed period (e.g. 24 h), then replacing the
extract with fresh solvent and extracting the test sample for another fixed period. This process is
repeated until the level of the residue for nth successive extraction is one tenth (0,1) of the level in
the first extraction, so the extraction can be deemed complete or exhaustive. Gravimetric analysis
or other analytical methods can be used to determine if the exhaustive endpoint is achieved.
— The extract residue is obtained by evaporating solvents or another drying process following
exhaustive extraction. Data should be provided to demonstrate that the evaporation or drying
process and the subsequent reconstitution step (as described below) do not cause loss of either
volatiles or semi-volatile compounds, or both, from the extract.
— The reconstitution of the extract residue after exhaustive extraction is a critical step of sample
preparation for biological testing. Typically, the extraction solvents for exhaustive extraction are
different from the solvents or media used for biocompatibility testing. If the exhaustive extraction is
conducted in organic solvents, the obtained extractable/leachable of the polymeric medical device
or samples are hydrophobic in nature. The extract residue after solvent evaporation usually has
miscibility issues in the cell culture medium and might not be completely soluble in the solvent used
for biocompatibility testing. Consequently, the test system dosed with such reconstituted solution
of the extract residue can be exposed to a significantly lower amount of extractables/leachables. For
Bibliography
[1] ISO Guide 30:2015, Reference materials — Selected terms and definitions
[2] ISO Guide 31, Reference materials — Contents of certificates, labels and accompanying
documentation
[3] ISO Guide 33, Reference materials — Good practice in using reference materials
[4] ISO Guide 35, Reference materials — Guidance for characterization and assessment of homogeneity
and stability
[5] ISO/IEC Guide 99:2007, International vocabulary of metrology — Basic and general concepts and
associated terms (VIM)
[6] ISO 10993 (all parts), Biological evaluation of medical devices
[7] ISO 14971, Medical devices — Application of risk management to medical devices
[8] ISO 17034, General requirements for the competence of reference material producers
[9] ISO/TS 37137-1, Biological evaluation of medical devices — Part 1: Guidance for absorbable
implants
[10] United States Pharmacopeia/National Formulary, ˂88> Biological Reactivity Tests, In Vivo
[11] The Japan Society for Analytical Chemistry Research Committee of Polymer Analysis, Polymer
Analysis Handbook, pp. 549-558, Kinokuniya-Shoten, Tokyo, 1995 (ISBN 4-314-10110-5 C3043)
[12] MHLW Notification by Director, OMDE, Yakushokuki-hatsu 0301 No.20, March 1, 2012. Basic
Principles of Biological Safety Evaluation Required for Application for Approval to Market
[23] Fuchs O., Solvents and Non-solvents for Polymers in Polymer Handbook (third edition), edited
by Brandrup, J. and Immergut, E.H., VII/379-VII/407, Wiley Interscience, 1989
[24] Matsuoka A., Haishima Y., Hasegawa C., Matsuda Y., Tsuchiya T., Organic-solvent extraction
of model biomaterials for use in the in vitro chromosome aberration test. J. Biomed. Mater. Res.,
Part A, 86, 2008, pp. 13–22
[25] Nakamura et al. Cutaneous and Ocular Toxicology, 22(3), 2003, pp. 169–185
[26] Oba T., Tsuji K., Nakamura A., Shintani H., Mizumachi S., Kikuchi H., Kaniwa M.A., Kojima
S., Kanohta K., Kawasaki Y., Furuya T., Matsumoto K., Tobe M. Artificial Organs, 8(4), 1984,
pp. 429–435
[27] Reid R.C., Sidman K.R., Schwope A.D., Till D.E. Ind. Eng. Chem. Prod. Res. Dev., 19(4), 1980,
pp. 580–587
[28] Reid R.C., Schwope A.D., Sidman K.R., Modeling the migration of additives from polymer films
to foods and food simulating liquids, MIT Industrial Liaison Program Report 1-14-84, Directory
of Current Research: 3.04.077
[29] Shintani H., Nakamura A. J. Biomed. Mater. Res., 25, 1991, pp. 1275–1286
[30] Tsuji K., Mizumachi S., Iida K., Oba T. Kobunshi Ronbunshu, 34(4), 1977, pp. 287–290
[31] Uphill P.F., Christopher D.H., Developing a Positive Control for Cytotoxicity Testing of Medical
Device Materials, Medical Device Technology, Nov./Dec. 1990, pp. 24–27
[32] Vondracek P., Dolezel B., Biostability of Medical Elastomers: A Review, Biomater., 5,
1984, p. 209
[33] Vogel A., Vogel's Textbook of Practical Organic Chemistry: Experimental Techniques (fifth
edition), Chapter 2, Revised by Furniss, B.A. et al., John Wiley & Sons, Inc., New York, 1989