2. INTRODUCTION
The precision electronic weighing scale offers a range of capacities from 3 kg to 30kg.
The scales are very simple to use and applicable for general weighing. The user can also use
the parts counting and percent weighing functions for special applications. Special functions
are available for weighing in up to 4 different units of weight. And the scale use grams as the
default unit.
The standard RS-232 interface allows the data to be sent to a computer or printer. (Only for
the scale with RS-232 interface)
The scale is with large easy to read LCD display with backlight, and all the keypads are sealed
membrane switches.
The scale includes audible alarm for pre-set weights, tare and an accumulation facility that
allows the count to be stored and recalled as an accumulated total.
3. COMMON SPECIFICATIONS
-Weighing
-Counting pieces
-Percentage
Functions -Check-weighing
-Check-counting,
-Check-percentage,
-Accumulation(summing)
Set the zero point for all subsequent weighing. The display shows
zero.
Set the limits for check weighing. Allow setting of either the low
limit or the high limit or both.
Secondary function “◄ ”, is to move the active digit to the left when
setting values for parameters or other functions.
Press this key to enter into accumulation function. The display will
show Accumulation times and total weight.
Secondary function (Exit), is to return to normal operation when
the scale is in parameter setting mode.
Press this key to hold the display when an active thing is weighted.
The scale has an automatic re-zeroing function to account for minor drifting or accumulation
of material on the platform.
However you may need to press the “ZERO” key to re-zero the scale if small amounts of
weight are shown when the platform is empty.
Note: Zeroing function is possible to operate only when the scale is stable with “ ”
6.3 Taring
Before Tare function is used, if zeroing the scale is needed, press the “ZERO” key. Then the
indicator of zero will be on.
● Place a container on the platform, a value for its weight will be displayed.
● Press the “TARE/Enter” key to tare the scale. The weight that was displayed is stored as
the tare value and that value is subtracted from display, leaving zero on the display.
The NET” indicator will be on.
● Put a weight on the pan. Only the weight that is added after taring will be displayed. And
the value displayed is the net weight of the product
● Remove the container a negative value will be shown. If the scale was tared just before
removing the container this value is the gross weight of the container plus all products that
was removed.
● Press “TARE/Enter” key again to enter into the normal weighing mode, and the negative
value will disappear.
Note: Taring function is possible to operate only when the scale is stable with “ ”
6.4 Percentage
This mode shows the proportion of standard weight.
A standard weight is selected whatever you want within max capacity of the scale. When you
place a sample on the container, the display shows the percentage against the standard weight.
For example:
● Place a sample(700g) on the Pan.
● Place a 200g weight on the pan, the display will show 28.57% as 200g is 28.57% of 700g.
Note : the scale may jump by large numbers unexpectedly if small weights are used to set
the 100% level. For example if only 23.5g is on a scale with 0.5g increments and the
scale is set to 100%, the display will show 100%, however a small change of weight
will cause the display to jump to 100.13%.
● Place samples (ex. 200g as 10pcs) on the pan. The number of samples should match the
options for parts counting: 10, 20, 50, 100, 200,500, or 1000 pieces
● Press the “COUNT/ ” key to begin. Then the display will show SP 10”.
To change the number of parts is possible as “COUNT/ ” key. It will cycle through the
options : 10, 20, 50, 100, 200, 500,1000
● Press the “TARE/Enter” key when the number matches the number of parts.
● Place any other sample (ex. 500g) on the pan which you want to know how many the
sample is. The display will be show “25 pcs”. It means the sample as 500g is consisted of
25 pieces.
Note : 1) Pressing “%/ ” key is possible to show unit weight, Total weight and the count
(pcs) by turns.
2) Press the “COUNT / ” key to return to normal weighing.
6.6 Check-weighing/counting/percentage
Check-weighing is a procedure to cause an alarm to sound when the weight on the scale
meets or exceeds values stored in memory. The memory holds values for a high limit and a
low limit. Either limit can be used or both can be used. This function is available in weighing,
counting pieces and percentage.
● Press the “CHECK/ ” key. The display will show the current High Limit with the left
most digit flashing and the “HI” symbol on the left of the display.
● After set high limit value by pressing the “TARE/Enter” key, then the display will show
“LO” symbol on the left side.
● Input low limit value (ex. 100g), the way is as same as high limit was set.
● After inputting low limit value, press “TARE/Enter” key to confirm it. Then the scale will
return to weighing mode with the Check-weighing function enabled
When a weight is placed on the scale the arrows will show if the weight is above or below the
limits and the beeper will sound as described below.
● The display will show “OK” and the beeper will sound when the weight is less than the
Low Limit.
● Above the Low Limit the display will show “HI” and the beeper will be off.
HIGH LIMIT SET
● LOW LIMIT is set to zero
● The display will show “LO” and the beeper will be off when the weight is less than the
High Limit.
● Above the High Limit the display will show “OK” and the beeper will be on.
Note :1) When you would like to use check-counting/percentage, sampling should be set
before you set HI/LO limit values
2) The low limit is set greater than the high limit. In this situation, the display will
show E5 and you are supposed to reset the values.
3) The weight must be greater than 20 scale divisions for the check-weighing to
operate.
4) To disable the Check-weighing function enter zero into both limits by pressing
the “CHECK/ ” key to recall the current settings. Pressing “UNIT/Clear” key
when the current limits are shown to clear the settings, then pressing the
“TARE/Enter” key to store the zero values.
3) After showing of “ ×× ” for a few seconds, the total weight is displayed for about 2
seconds. Then the scale returns to normal weighing mode.
4) Remove all weights from the pan in order to zero the scale. Place the second weight on the
pan(ex. 500g)
5) Press “M+/Exit” key to add the second weight, then “ ×× ” is displayed for a few
seconds.
6) After showing of “ ×× ”, the total weight is displayed for about 2 seconds. Then the
scale returns to normal weighing mode.
7) As soon as the total weight is disappeared, the display will show the new weight.
Note : 1) Any of the accumulated items can be cleared by pressing “UNIT/Clear” key.
Please refer to section “ 6.10 SUBTRACTION FUNCTION”.
2) Accumulation memory can be added up to 99 times, or up to the total value
exceeds in maximum capacity.
3) The display weight must be stable( ) for M+ register.
4) Display must return to zero in order to add next item.
5) Accumulation memory is not retained when the scale is turned off.
6) There are two options of RS-232 transmission method in accumulation mode
(Please refer to section 7.5).
7) When the present weight is less than 10d, and press “M+/Exit” key. The scale
can’t store this value in memory but only display total accumulated values and
times.
6.9 Subtraction function
Use “UNIT/Clear” key to subtract accumulation values. This function is available only in
accumulation mode.
● Nothing on the pan.
If you make a mistake or want to cancel the last value, press, “UNIT/Clear” key with
nothing on the pan. Then the last value will be subtracted from total.
Note : The maximum battery life is achieved with the backlight turned off.
Press the “TARE /Enter” key in order to confirm and move to next setting.
Note : 1) When Gross weight is Negative Value, Output is not be transmitted by pressing
“PRINT” key. But in series transmission, negative value can be transmitted
automatically.
Press the “TARE /Enter” key in order to confirm and move to next setting.
Press the “TARE /Enter” key in order to confirm and move to next setting.
Press the “TARE /Enter” key in order to confirm and move to next setting.
7. 6 Zero tracking range
There are 4 options for zero tracking range. You can select its options by pressing “COUNT/
” key.
Press the “TARE /Enter” key in order to confirm and move to next setting.
Options: 0 = 0
1 = 0.3d
2 = 0.6d
3 = 0.9d
(Default value: 1=0.3d)
Press the “TARE /Enter” key in order to confirm and move to next setting.
Note: The smaller number is the shorter time for display stability.
7.8 Selection of the available units
There are 4 kinds of unit: kg, g, lb, oz
1. When the display show “Unit”, press “TARE/Enter” key to enter into the available
units selection setting.
After setting the last unit (“oz”), press “TARE/Enter” key to go back to normal weighing
mode.
NOTE: 1) A unit can not be exchanged in counting mode, or when tare function performed.
2) In accumulation mode or check-weighing setting is available, give a long press of
“UNIT/Clear” key to exchange the unit. However the accumulation value or the
check-weighing setting will be cleared automatically.
8. BATTERY OPERATION
The scale can be operated from the battery if desired. The battery life is approximately 80
hours.
When the battery needs charging a symbol “ ” on the weight display will turn on. The
battery should be charged when the symbol is on. The scale will still operate for about 10
hours after which it will automatically switch off to protect the battery.
To charge the battery, simply attach the power supply module to the scale and plug in. The
scale does not need to be turned on.
There is an LED to indicate the status of battery charging on the right of display. When the
scale is plugged into the mains power the internal battery will be charged.
As the battery is used it may fail to hold a full charge. If the battery life becomes unacceptable
then contact your distributor.
Note: The battery should be recharged every 3 months if the scale is not used for long time.
9. RS-232 OUTPUT
The scale can be ordered with an optional RS-232 output.
Weighing mode:
========WEIGHT========
Weight: 100.0 g
Weight: 100.0 g
Weight: 100.0 g
Weight: -99.9 g
Weight: 100.0 g
Weight: - 13.8 g
Weight: 100.0 g
Weight: 100.0 g
……………………………….
Total: 386.3 g
Count: 4 times .
Counting mode
========COUNT========
Quantity: 20 pcs
Quantity: 23 pcs
Quantity: -23 pcs
Quantity: 23 pcs
Quantity: - 3 pcs
…………………………………..
Total: 40 pcs
Count: 1 times
Note: In accumulation mode, when the total weight/pcs is”0” and the count is not “0”, the count
of results will be “0”. Take weighing mode for instance.
Weighing mode:
========WEIGHT========
Weight: 150 g
Weight: 400 g
Weight: -550g
………………………………….
Total: 0g
Count: 0 times
When total weight/pcs is more than “0” and real count times is “0”, the results of count
times won’t be “0”. For example:
Counting mode:
========WEIGHT========
Quantity: 90 pcs
Quantity: -56 pcs
Quantity: 90 pcs
Quantity: -56 pcs
…………………………………
Total: 68 pcs
Count: 1 times
● ACP is set off, the scale is in accumulation mode, and limit values are not set. Press PRINT
key to output the total accumulated value and times.
Weighing mode:
………………………………..
Total: 386.3g
Count: 4 times
Counting mode:
…………………………………..
Total: 40 pcs
Count: 1 times
● When the accumulation is not used and limit values are not set, press PRINT key to
transmit the value showing in display window.
Weighing mode:
-Weighing Function-
========WEIGHT========
Weight: 90 g
Tare: 20 g
Gross: 110 g
Counting mode:
-Counting Function-
========COUNT========
Weight: 174 g
U/Weight: 9.04 g
Quantity: 19 pcs
Tare: 29 g
Gross: 205 g
Percentage mode:
-Percentage Function-
========PERCENT========
Weight: 176 g
Percent: 95.02 %
Tare: 28 g
Gross: 204 g
● When the Accumulation function is not used and the limit values are set. And transmit by
pressing PRINT key:
Weighing mode:
Counting mode:
Percentage mode:
DEP-50 form (Au on):
The scale transmits data automatically when it is stable. In this mode, press “M+/Exit” key to
accumulate weight / pcs, but it doesn’t transmit total weight / pcs and count by pressing
“PRINT” key.
Weighing mode:
Weight: 90 g
Weight: 120 g
Weight: 90 g
Counting mode:
Quantity: 10 pcs
Quantity: 34 pcs
Quantity: 50 pcs
Percentage mode:
Percent: 21.33 %
Percent: 51.04 %
Percent: 27.62 %
US-Display is Unstable
DATA (8BYTES)
2E (HEX) =”.”(DECIMALPOINT)
UNIT (4BYTE):
g- 20 ( HEX ) ; 67 ( HEX ) ; 20 ( HEX ) ; 20 ( HEX )
ST , GS , + 0.876 g 0D 0A
US , NT , - 1.568 lb 0D 0A
OL , NT , -------- oz 0D 0A
Command(1byte)
Weighing Mode
Char. Hex
0x4C
L (l) Same as CHECK/ key
0x6C
0x43
C (c) Same as COUNT/▲ Key
0x63
0x52
R (r) Same as %/ Key
0x72
0x55
U (u) Same as UNIT/CLEAR Key
0x75
0x4D
M (m) Same as M+/Exit key
0x6D
0x50
P (p) Same as PRINT Key
0x70
0x5A
Z (z) Same as ZERO Key
0x7A
0x54
T (t) Same as TARE/ENTER Key
0x74
0x48
H (h) Same as HOLD Key
0x68
10. AUTO CALIBRATION (can be done in kg/g, Ib)
Give a long press of TARE/Enter key in normal weighing mode without performing Tare
function. When the display is flashing and showing the weight value, put on the mass. While
the scale is stable, the display will be back to zero automatically.
Please take away the mass while scale is counting backward. The auto calibration is finished.
By pressing the “COUNT/ ” key the weight value can be selected (1/3 Full weight, 2/3 Full
weight and 3/3 Full weight).
Operation:
Turning on the scale, press the “TARE+TARE+COUNT+COUNT+COUNT” keys
orderly during self-check. Then, the scale will begin the self-check again. It means that the
scale will come back with a default calibration value.
12. Variables (The prompt character) used in scale
also in label printer
Variable
Effective Qualification Specifications Size
name
Note: 1) ** P variable: The weight accumulation value exceeds 6 digits, but it can
achieve 6-digit display through righting decimal point. (One or two digits behinds
the decimal point won’t display after being rounded)
** B variable: It would be wrong if not considering the decimal point.
2) The variable can’t be less than 0.
Moreover, the negative value will not be transmitted
13. TABLE LIST FOR FULL CAPACITY
Spec
3kg x 0.1g 6kg x 0.2g 15kg x 0.5g 30kg x 1g
unit
If an error message is shown repeat the procedure that caused the message, turning the scale
on, calibration or other functions. If the error message still is shown then contact your dealer
for further support.
Display Indicators Stability, Center of Zero, Gross, Net, Battery status, Hi-Ok-Lo, Units