Horizontal Pulse Vacuum Autoclave
Advantage
Horizontal Pulse Vacuum Autoclave
Horizontal Pulse Vacuum Autoclave
●Multi Safety Features
Description Overheat protectionautomatically cutting off heating with audible and visual alarming when the autoclave is overheated.
Horizontal Pulse Vacuum Autoclave uses saturated steam as the steriliza-
Interlock safety: preventing from starting a cycle if door is not properly locked or from door opening if the chamber is pressurized.
tion medium. Through the pulsating vacuum method, the saturated steam
fully penetrates the surface of the items. By utilizing high-temperature Anti-burning protectionautomatically cutting off heating with audible and visual indication when the autoclave is lack of water.
saturated steam under high pressure, it effectively kills all microorganisms
and their spores, providing reliable sterilization results. It is considered the Steam pressure safety valve: automatically works when the chamber pressure exceeds the limit.
most effective method among physical sterilization techniques.
●Automatic operation from sterilization to drying (Pre-vacuum sequence, heating, sterilization, steam discharge and post-vacuum
drying) can be easily set and monitored by LCD touch screen.
It is widely used in fields such as pharmaceuticals, biotechnology, and ●Built-in vacuum pump provides 3-time pulsed vacuum process,extracting the air out of the medical dressing and the A degree hollow
laboratories for sterilizing items such as dressing/fabrics, medical instru- medical instrument efficiently.
ments, lumen instruments, rubber, and liquids.
●Built-in steam generator prevents the media from contamination.
●Post-vacuum drying function offers efficient drying results.
●Positive pulsing function removes the air efficiently.
●With 0.22µm air filter, blocks all microorganisms from exiting the chamber during the air removal phase.
●Standard printer.Optional real-time record sterilization process, error and alarm information.
Specification
Model
Model ST-H60PV ST-H80PV ST-H100PV ST-H135PV ST-H150PV ST-H185PV ST-H200PV ST-H300PV ST-H450PV
Capacity 60L 80L 100L 135L 150L 185L 200L 300L 450L
External Size(W*D*H) 985x600x1450mm 985x600x1450mm 1200x600x1450mm 1000x710x1550mm 1330x710x1550mm 1250x710x1550mm 1320x710x1550mm 1350x900x1750mm 1750x900x1750mm
Chamber Size(Ø*H) Φ386x500mm Φ386x700mm Φ386x950mm Φ500x700mm Φ500x780mm Φ500x950mm Φ500x1000mm Φ650x950mm Φ650x1360mm
Chamber Material SUS304
Sterilization Temperature 105~136
Designed Temperature 150
Max.Working Temperature 136
Temp.Accuracy 0.1
Temp. Uniformity ≤±0.5
Designed Pressure -0.1/0.3Mpa
Sterilization Pressure 0.23Mpa
Pressure Resoluiton 0.1KPa
Ultimate Vacuum -0.096MPa
Temp. Controller PID control by microprocessor
Display LCD touch screen
Drying Function YES
Drying Time 15min
Sterilization Program ●Medical instrument program ●Luminal instruments ●Dressings program ●Rubber program
●Liquid program ●BD Test program ●Vacuum Test progrm ●4 pcs DIY progrm
Consumption 5.7KW 56.6KW 6.6KW 12KW 12KW 12KW 12KW 20KW 20KW
Power Supply 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz
Standard Sterilization Shelf 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs
Standard Accessory Printer
Net Weight 105kg 120kg 140kg 225kg 240kg 255kg 265kg 350kg 480kg
Gross Weight 165kg 185kg 210kg 315kg 340kg 355kg 380kg 470kg 570kg
Shipping Size(W*D*H) 1280×700×1600mm 1280×700×1600mm 1450×700×1600mm 1300×850×1900mm 1400×900×1900mm 1440×930×2020mm 1440×930×2020mm 1650×1100×2120mm 2100×1100×2120mm
33 34