AUTOCLAVE

SCITEK GLOBAL CO.,LTD ClassN / Class B Benchtop Autoclave

Add: No. 7000 Jingshi Road, Jinan City, Shandong Vertical / Horizontal Autoclave
Province, China Hot Air Sterilizer
Tel: +86-531-82893797 H2O2 Plasma Sterilizer
Inquiry: [email protected]
Service: [email protected] Pharmaceutical Field
Complaints: [email protected] Others
Web:www.scitekglobal.com
CONTENT
SCITEK GLOBAL CO.,LTD Class N Benchtop Autoclave.....................................................01

Class B Benchtop Autoclave.....................................................05

Vertical Autoclave..................................................................... 11

Horizontal Autoclave................................................................. 30

Portable Autoclave.................................................................... 43

Glass Bead Sterilizer................................................................ 45

ABOUT
SCITEK GLOBAL IS PROFESSIONAL MANUFACTURE OF LABORATORY INSTRU-
MENTS AND MEDICAL DEVICES, CERTIFIED BY ISO9001, ISO13485 AND ISO14001.
PRODUCTS ARE CERTIFIED BY ETL, CE AND FDA ETC. Infrared Bacti-Cinerator Sterilizer..............................................46

SCITEK
WE LOOK FORWARD TO ESTABLISHING COOPERATION WITH MORE DISTRIBU-
TORS ALL OVER THE WORLD, AND WORKING TOGETHER TO CREATE GREATER
CUSTOMER VALUE. TOGETHER WE WILL BE DIFFERENT.
Infrared Sterilizer....................................................................... 47

Hot Air Sterilizer........................................................................ 48

Cassette Sterilizer..................................................................... 49

Ethylene Oxide Sterilizer.......................................................... 50

H2O2 Plasma Sterilizer.............................................................53

ISO 14001 ISO 9001 ISO 45001 Low-temperature Steam Formaldehyde Sterilizer.................... 57

Pharmaceutical Field

Horizontal Pulse Vacuum sterilizer.......................................... 59


H2O2 serious Hydrogen peroxide steam sterillzer.....................66

Dry heat sterilizer.......................................................................67


CE CE CE CE
General evaporator................................................................... 68
/ /

Class N Benchtop Autoclave Class N Benchtop Autoclave

Application Application

Class N Benchtop Autoclave


Class N Benchtop Autoclave

Used for laboratory sterilization, compact in size, can be placed on a desktop. It has a single program and can be used for sterilizing Used for laboratory sterilization, compact in size, can be placed on a desktop. It has a single program and can be used for sterilizing
items such as culture media, test tubes, and tools. It can also be used for sterilization in hospitals, dental clinics, and other facilities. items such as culture media, test tubes, and tools. It can also be used for sterilization in hospitals, dental clinics, and other facilities.

LED display
screen

Pressure
Gauge

Advantage
●Complies with European Class N standard for sterilization effectiveness.

Advantage ●Microcomputer control system for automatic cycle programming and temperature setting (121 and 134 ).User-friendly
●Steam sterilizer compliant with Europe Class N standard, offering wrapped and unwrapped system options with 121 and 134 interface with touch buttons for easy operation.
sterilization settings. ●Safety interlocking device to prevent opening of the cover during pressurization or high temperature.
●Microprocessor-controlled system for intelligent control and user-friendly interface, suitable for Stomatology, Ophthalmology, Surgery, ●Equipped with a standard test interface for additional functionality.
and Lab departments.
●Quality stainless steel inner tank and sealed cover for durability and airtight operation.
●Chamber constructed with special stainless steel #304, capable of withstanding working pressures from -0.9 bar to 2.3 bar
●Open-type top water tank for convenient water filling.
and a maximum temperature of 136 .
●Automatic protection functions, including over-temperature, over-pressure, low water level, and anti-dry burning.
●Equipped with safety features including a safety valve, manual door lock, error self-test, and overheat protector.
●Buzzer reminder and automatic stop after sterilization.
●Three sterilizing plates included with the bracket for efficient use of space.
●Automatic discharge of cold air and steam exhaust after sterilization.
●Quick sterilization process ranging from 15 to 30 minutes, depending on the selected temperature.
●Built-in steam generator for quick production of saturated steam.

Specification
HDO-45N HDO-70N HDO-140N HDO-230N
Model Specification
Model HDO-45
ST-B16N HDO-70 HDO-140
ST-B18N HDO-230
ST-B24N HDO-45N HDO-70N HDO-140N HDO-230N
Model
Capacity 16L 18L 24L Model HDO-45
ST-B18NI HDO-70 HDO-140 ST-B24NI HDO-230
Sterilization Temperature 134 /121 Capacity 18L 24L
Display accuracy ±0.5 Rated working temperature 134
Recording accuracy ±0.5 Design pressure 0.25Mpa
Sensor accuracy 0.01 Design temperature 139
Tray 3pcs SS trays on SS shelf Rated working pressure 0.22Mpa
Display LED display Temp accuracy ±1
Consumption 1800W 1800W 2000W Consumption 1.5kw 2kw
Power Supply AC220V/110V±10%;50/60Hz Power Supply AC220V/110V±10%;50/60Hz
Chamber Size (mm) Ø230x360 Ø249x355 Ø249x450 Chamber Size (mm) Ø250x370 Ø250x452
External Size(WxDxH)(mm) 550x445x395 550x445x395 640x445x395 External Size(WxDxH)(mm) 540x400x350 640x400x350
Shipping Size(WxDxH)(mm) 670x550x500 670x550x500 740x550x500 Shipping Size(WxDxH)(mm) 540x450x390 640x450x390
N.W./G.W. (kg) 26/32 27/33 33/39 N.W./G.W. (kg) 24/26 27/29

1 2
/ /

Benchtop Class N Autoclave

Application Door-temperature interlock safety: preventing from explosion of liquids if the liquids temperature is not cooled down the safe limit

Benchtop Class N Autoclave


Benchtop Class N Autoclave

Used for laboratory sterilization, compact in size, can be placed on a desktop. It has a single program and can be used for sterilizing (20 lower than the local boiling point).
items such as culture media, test tubes, and tools. It can also be used for sterilization in hospitals, dental clinics, and other facilities. Anti-burning protection automatically cutting off heating with audible and visual alarming when the autoclave is lack of water.

Steam pressure safety valve: automatically works when the chamber pressure exceeds the limit.

Electrical circuit safety device for over-current and shortcut protection.

●Automatic operation of sterilization (heating, sterilization, steam discharge and drying) can be easily set and monitored by

the LCD touch screen.

●Built-in steam generator prevents the media from contamination.

●Buzzer for working errors and the end of the cycle.

●Automatic atmospheric pressure adjustment for the installation and operation in plateau areas.

●Optional printer, real-time record sterilization process, error and alarm information.

Specification
Model
Model ST-B45N
Capacity 45L
External Size(W*D*H) 870x600x520mm
Chamber Size(Ø*H) Φ316x590mm
Chamber Material SUS304
Description
Sterilization Temperature 105~136
Designed Temperature 150
Bechtop Class N Autoclave uses saturated steam as the sterilization medium,exhausts the cold air by gravity. So that the saturated
Max.Working Temperature 136
steam fully penetrates the surface of the items. By utilizing high-temperature saturated steam under high pressure, it effectively kills all Temp.Accuracy 0.1
Temp. Uniformity ≤±0.5
microorganisms and their spores, providing reliable sterilization results. It is considered one of the most effective methods among Designed Pressure 0.3Mpa
Sterilization Time range 0.23Mpa
physical sterilization techniques.
Temp. Controller PID control by microprocessor
Display LCD touch screen
Drying Function YES
Application
Drying Time 15min
It is widely used in fields such as pharmaceuticals, biotechnology, and laboratories for sterilizing items. such as dressing/fabrics, ●Dressings/Fabric program
medical instruments, rubber, and liquids. ●Medical instrument program
Sterilization Program ●Rubber program
●Liquid program
Advantage ●DIY program
Water Injection method Built-in water tank
●Multi Safety Features Consumption 3.6KW
Power Supply 220V 50Hz
Overheat protection automatically cutting off heating with audible and visual alarming when the autoclave is overheated. Standard Sterilization Shelf 1pcs SS shelf
Electric interlock safety: preventing from starting a cycle if door is not properly locked or from door opening if the chamber is pressur- Optional Accessory Printer
ized or the power is disconnected. Net Weight 65kg
Gross Weight 120kg
Shipping Size(W*D*H) 1150×700×700mm

3 4
/ /

Class B Benchtop Autoclave, ST-BB series Class B Benchtop Autoclave

Class B Benchtop Autoclave


Class B Benchtop Autoclave

ST-BB auto door ST-BB manual ST-B8B, ST-B12B ST-B18B, ST-B24B

Advantage
Specification
●Complies with European Class B standard for eective sterilization.
Model
Model ST-B24DB ST-B29DZB
●Microcomputer control wirh LCD display and touch buttons ensures easy monitoring and precise operation.
Capacity 24L 29L
Door Manual Automatic ●Equipped with BD test, HELIX test, and VACUUM test for comprehensive sterilization performance.
Display 5.0 inch Color touch screen
touch screen ●Automatic protection function includes over temperature, over pressure, and low water level protection, as well as anti-dry burning
Chamber Material SUS304 and safety door lock. Safety interlocking device ensures proper operation and prevents accidents.
Sterilization Temperature 105~138
●Three times pre-vacuum enhances sterilization eciency and eectiveness.
Max.Working Temperature 134
Temp.Uniformity ≤±1 ●Buzzer reminder signals the completion of sterilization cycles, and automatic stop ensures safety.
Designed Pressure -0.1~0.3Mpa
Timer Range 0~99min ●Automatic discharge of cold air and steam exhaust after sterilization for ecient and safe operation.
Sterilization Pressure 0.23Mpa
●Built-in steam generator provides quick production of saturated steam for sterilization processes.
Drying Function Yes
Drying Time 2min 20min 25min ●Optional accessory:external or internal printer,usb disk.
Pulse time 1~5
Fast 134
Fabric 134
Packaging Equipment 134 Specification
Enhanced Mode 134
Model
Sterilization Program Rubber program 121 Model ST-B8B ST-B12B ST-B18B ST-B24B
Model
Packaging Equipment 121 Model
Capacity 8L 12L 18L 24L
DIY Program Consumption 2kw 2kw 2kw 2kw
HELIX/B D Test Power Supply 220V110V;50/60Hz
Vacuum Test Design pressure 0.25Mpa
Evaporator New generation evaporator Design temperature 139
Water Injection method Build in water tank Rated working pressure 0.22 Mpa
Standard Sterilization Shelf 3pcs SS shelf Temp accuracy ±1
Consumption 2700w 2800w Rated working temperature 134
Power Supply 220V ±22V 50/60Hz ±1Hz Chamber Size Ø170x320 Ø200x355 Ø250x370 Ø250x452
Chamber Size(ØxH) Ø248x450mm Ø320x362mm External Size(WxDxH)(mm) 580x445x400 600x445x400 640x480x450 690x480x450
External Size(WxDxH) 700x483x455mm 658x638x580mm Shipping Size(WxDxH)(mm) 650x500x450 660x510x460 720x560x510 760x560x510
Net Weight 53kg 95kg N.W./G.W. (kg) 42/45 44/48 48/57 53/60

5 6
/ /

Class B Table Top Autoclave Class B Benchtop Autoclave

Class B Benchtop Autoclave


Class B Benchtop Autoclave

Application Description

Pre-vacuum rapid pressure steam sterilizer is suitable for medical dressings, biological products, instruments and vessels, sterile It is a sterilization equipment that fully complies with CSSD and GMP specifications. The steam sterilizer uses vacuum to form a
clothing, waste, moisture and heat-resistant instruments or items with high sterilization requirements, as well as dentistry, ophthalmolo- negative pressure in the sterilization cabinet and extracts cold air so that the saturated steam can quickly penetrate inside the items for
gy, operating rooms, supply rooms, dialysis Sterilization of packaged or unpackaged solid instruments, Class A cavity instruments sterilization.
(dental handpieces, endoscopes), implantable instruments, dressing fabrics, rubber catheters and other items in medical and health
departments such as rooms and delivery rooms.

Advantage

1.Complies with European Class B standard for effective sterilization.

2.Automatic door switch for convenient and safe operation for ST-PV24B.Automatic door for ST-PV45B.

3.Microprocessor control.

4.LCD touch screen graphical and digital display of all the operating status of each function such as heating, pressure relief,
exhaust, power off.

5.Automatic water filling Automatic hydration.

6.Drying function.

7.Standard accessory: USB Port,Thermal Printer.

Safety protection
ST-PV24B
ST-PV45B Automatic fault diagnosis

Automatic over-temperature protection

Over-pressure protection

Specification
Model
Model
Model ST-PV24B ST-PV45B
Model
Capacity 23L 45L
Chamber Size Φ246X450 Φ320x620mm
External Sizes (WxDxH)(mm) 715X513X425 900x565x600
Working Temperature 5C-40C
Sterilization Temperature 134 /121
Sterilization Pressure 0.25Mpa
Chamber Material 304 corrosion-resistant stainless steel material
Strilization Mode Exposed instruments,Packaged instruments, Fabric hollow, Rubber products, BD testing, Vacuum testing
Drying Mode Powerful vacuum drying, suitable for drying different items; residual moisture in equipment <0.2%, residual
moisture in dressings <1%.
Display LCD touch screen
Capacty for Water Tank 2L 6L
Water consumption 0.5-1L 0.8-1.5L
Displacement 0.5-1L 0.8-1.5L
304 corrosion-resistant LCD touch screen Consumption 1.7KW 3.5KW
stainless steel material Power Supply 220/110V,50/60Hz
Gross weight 49KG 108kg
Shipping Size (W×D×H) 830x600x530mm 1050x700x800mm

7 8
/ /

Class B Benchtop Autoclave, Automatic Door Class B Benchtop Autoclave, Automatic Door

Class B Benchtop Autoclave


Class B Benchtop Autoclave

Advantage
Built-in vacuum pump, with three adjustable pulsating vacuum cycles, achieving a vacuum degree of up to -0.090MPa.

Vacuum drying system, with fabric drying rate less than 1% and instrument drying rate less than 0.1%.

Multi Safety Features

Overheat protection: Automatically cutting off. Audible and visual alarm will be there.

Interlock safety: To prevent a cycle from starting if the door is not locked or stop dooropening when the chamber is pressurized.

Anti-burning protection:Auto shut-off with audible and visual alerts for low water levels in autoclave.

Steam pressure safety valve: Automatically starts when the chamber pressure exceeds the limit. Advantage
Autoclave Functions: automatic operation, touch screen interface, built-in vacuum pump, steam generator, post-vacuum drying, ●Complies with European CLASS B standard for effective sterilization.
positive pulsing, air filter, optional printer for real-time record keeping.
●Automatic door switch for convenient and safe operation.
Automatic door switch for convenient and safe operation.
●Microcomputer control with LED display ensures precise and reliable sterilization.
Specification
●Equipped with BD test, HELIX test, and Vacuum test for comprehensive sterilization performance.
Model
Model ST-B45B
●Standard test interface for convenient testing and validation.
Capacity 45L
External Size(W*D*H) 870x600x520mm ●Automatic protection function includes over temperature, over pressure, low water level, anti dry burning, and safety door lock.
Chamber Size(Ø*H) Φ316x590mm
Chamber Material SUS304 ●Safety interlocking device for secure operation.
Sterilization Temperature 105~136
●Three times pre-vacuum enhances sterilization efficiency.
Designed Temperature 150
Max.Working Temperature 136 ●Buzzer reminder and automatic stop after sterilization cycle completion.
Temp.Accuracy 0.1
Temp. Uniformity ≤±0.5 ●Automatic discharge of cold air and steam exhaust for efficient and safe operation.
Designed Pressure 0.3Mpa
●Built-in steam generator provides quick production of saturated steam.
Sterilization Time range 0.23Mpa
Pressure Resoluiton 0.1KPa ●Built-in printer for record-keeping and documentation.
Ultimate Vacuum -0.096MPa
Temp. Controller PID control by microprocessor Specification
Display LCD touch screen Model
Drying Function YES Model ST-BA18B ST-BA24B ST-BA45B
Drying Time 15min Capacity 18L 24L 45L
Sterilization Program ●Medical instrument program ●Luminal instruments ●Dressings program ●Rubber program Consumption 2.2kw 2.2kw 3.5kw
●Liquid program ●BD Test program ●Vacuum Test progrm ●4 pcs DIY progrm Power Supply 220V110V;50/60Hz
Water Injection method Built-in water tank Design pressure 0.25Mpa
Consumption 3.6KW Design temperature 139
Power Supply 220V 50Hz Rated working pressure 0.22Mpa
Standard Sterilization Shelf 1pcs SS shelf Rated working temperature 134
Standard Accessory Printer Chamber Size Ø250x370 Ø250x470 Ø316x615
Net Weight 65kg External Size(WxDxH)(mm) 505x620x450 505x710x450 657x870x575
Gross Weight 120kg Shipping Size(WxDxH)(mm) 560x670x510 560x760x510 750x980x630
Shipping Size(W*D*H) 1150×700×700mm N.W./G.W. (kg) 53/62 58/65 105/142

9 10
/ /

Vertical Pressure Steam Autoclave

4.Dual air convection rapid cooling fan


Vertical Pressure Steam Autoclave

Vertical Pressure Steam Autoclave


Dual air convection rapid cooling fan, automatically activates after default sterilization for temperature reduction, adjustable cooling lid
opening temperature, lid cannot be opened if temperature is not reached.

5.Multi-stage exhaust

Built-in steam collection box, utilizing multi-stage exhaust, without impacting the surrounding environment.

Safety Features
Microcomputor Control SUS304 Chamber
1.Safety interlocking device:Equipped with a high-temperature and high-pressure interlocking protection system, which automatically
detects and ensures that the chamber cover is closed in place before the work procedure can be started.

1.Anti-dry burning alarm:system automatically stops working when water level is lower than the minimum water level;system directly
cuts off the power supply when the water level is lower than the heating tube

2.Over temperature protection,Over pressure protection

Gasket Door 3.Electrical safety protection:Leakage, over current, and short circuit protection functions

Optional Accessory

1.Large capacity water tank with automatic water replenishment, fully automatic internal circulation.

2.Thermosensitive or needle-type miniature printer or recorder.

3.USB data storage device for recording sterilization cycles and sterilization data.
Vertical Flip Door Cooling Fan
4.Threaded sealing interface for 3Q validation.
Control Feature 5.JUMO mobile probe.
1.Microcomputer fully automated control sterilization process. 6.High-efficiency filtration device.
The status flowchart displays the current sterilization process, allowing for easy visual access to the current running steps at any time.
Specification
2.Multiple Sterilization Modes.

With liquid, solid, agar, instruments, etc.More than five sterilization modes, expandable memory store 80 sterilization programs. Model ST-VL65 ST-VL80 ST-VL90 ST-VL110
Capacity 62L 80L 88L 110L

Mechanical Feature External Size (W×D×H)(mm) 550×640×900 550×640×1150


Weight 72kg 82kg 85kg 95kg
1.Vertical flip cover structure.
Internal Size φ365×484H φ414×484H φ365×734H φ414×734H
Easy to open, and automatic locking upon downward pressure, with lock status determined based on indicator light prompts. Chamber Material SUS304 Stainless steel
Power Supply AC 220V 50Hz 16A
2.Electromagnetic suction type 6/8 claw mechanical interlock.
Consumption 4000W
Electromagnetic suction type 6/8 claw mechanical interlock, with 6/8 claws installed on the top cover. The sealing rubber strip is evenly
Sterilization Temperature/Time 115-138°C Continuously adjustable 0-9999 minute countdown display
stressed, ensuring a secure and reliable seal. The operating surface is clean, providing a smooth loading and unloading experience
Dissolution Temperature/Time 60-100°C continuously adjustable/0-9999 minutes countdown display
without any obstacles.
Holding Temperature/Time Preset Range 45-80°C adjustable/300 minutes
3.Dovetail type sealing groove.
Unlock Temperature 60-95°C can be set (default 80°C)
Dovetail type sealing groove, self-expanding medical grade silicone molded sealing ring, corrosion-resistant and high-temperature Verification Interface Reserved pressure and temperature pipe thread verification interface
resistant, can prevent accidents caused by high-temperature opening operations, providing a safer and more secure seal. Temp. Controller Using PT100AA^ temperature sensor, with a resolution of 0.1
Appointment feature Advanced booking for any time within the next year
Pressure 0.26Mpa
Standard Accessory Mesh basket or tote bag, gloves, and drain hose.

11 12
/ /

Vertical Pressure Steam Autoclave

Application

Vertical Pressure Steam Autoclave


Vertical Pressure Steam Autoclave

ST-VM series Liquid culture media, tip boxes, sterile clothing, exposed instruments, solid metal instruments and other occasions that ●Rapidly Cooling: Standard Configuration: Dual-cycle air cooling can shorten the cooling time. For fragile objects, when combined with
require sterilization. positive pressure, maintaining pressure and quick air cooling can enhance efficiency. The FO value is automatically calculated to
ST-VY series Liquid culture media, tip boxes, sterile clothing, exposed instruments, solid metal instruments, animal cages, bedding, determine if sterilization is effective. Sterilization completion can be determined based on FO value rather than time.
packages and other places that require rapid post-drying.
ST-VY Drying Function

●Disinfection packages with moisture content greater than 3% are referred to as wet packs. Wet packs cannot be used and must be
re-sterilized. The reason is that the moisture in the air forms a wet channel for bacterial contamination, leading to internal infection.

●Replaces the need for an oven, saving space.

●Simplifies complexity, saving time.

●Eliminates secondary contamination during the transfer process.

Specification

ST-VM56 ST-VM80 ST-VM75 ST-VM105 ST-VM120 ST-VM95 ST-VM135 ST-VM150


Model ST-VY56 ST-VY80 ST-VY75 ST-VY105 ST-VY120 ST-VY95 ST-VY135 ST-VY150
Model
Model ST-VM Series: Normal Type / ST-VY Series: Drying Function Type
LED Screen Gasket Door
Capacity 56L 80L 75L 105L 120L 95L 135L 150L
Advantage External Size(WxDxH)(mm) 680x600x895 680x600x1145 680x600x895 680x600x895 680x600x1235 680x600x895 680x600x1145 680x600x1145
Chamber Size(ØxH)(mm) Ø350x645 Ø350x895 Ø400x661 Ø400x911 Ø400x1001 Ø450x677 Ø450x927 Ø450x1017
●Automatically detect the position of the door lock and provide voice-guided instructions for easy operation. Chamber Material S30408
Sterilization Temperature 105~135
●Blue breathing light indicates standby, alert, and power outage statuses.
Designed Temperature 138
●Safety valve calibration prompts, with the ability to send notification messages to a WeChat public account through cloud-based Max.Working Temperature 135
pushing. Dissolution Temperature 60~100
Warming Temperature 45~65
●Dual-hold electromagnetic locks ensure automatic locking under high temperature or pressure. Unlock conditions are met for Unlock temperature 60~95
automatic release, and the lid can be opened directly by lifting the lever. Temp.Accuracy 0.1
Designed Pressure 0.42Mpa 0.42Mpa 0.38Mpa 0.38Mpa 0.38Mpa 0.35Mpa 0.35Mpa 0.35Mpa
●No steam leakage: Internal integration of copper pipe steam condenser converts high-temperature steam into water before discharg-
Max.Sterilization Pressure 0.24Mpa
ing it into the waste steam collection tank.
Sterilization Time range 1~9999min
●Voice prompts and guidance for user-friendly operation. Temp. Controller PID control by microprocessor
Display LED screen
●Optional sterilization report USB export and online storage of sterilization reports. Drying Function 3 Optional Drying Solutions (Select 1): ●Thermal control drying drying ●Positive pressure drying ●Vacuum
drying
●Free computer software provided for generating sterilization reports, allowing data exported from the USB drive to be output as PDF
Drying Time 10~300min (Optional)
reports. Can generate reports in both graphical and data formats.
Sterilization Program ●Liquid program ●Liquid & insulation program ●Bagged waste ●Dissolution & insulation program ●Instrument
●PHM/PHY models are equipped with online sterilization quality control functions such as "FO value calculation, FO value process program ●Self-cleaning program ●Safety valve test program ●Leakage test program ●B-D test program
calculation, FO value aging treatment." Water Injection method Build-in water tankOptional: Water supply pipeline+automatic replenishment, filtered pure water+automatic replenishment,
positive pressure automatic drainage in the chamber, automatic detection and removal of discharged wastewater
●Full automatic water replenishment, built-in current detection module to monitor the operation status of the heater in real-time, Drying Consumption 700W 900W 1200W 1400W 1600W 1200W 1400W 1600W
preventing dry heating online. Consumption 2500W 3800W 2500W 3800W 3800W 3800W 3800W 3800W
Power Supply 220V,50/60Hz
●Standard configuration: Dual over-temperature protection + dual over-pressure protection for the heating system, ensuring enhanced
Standard Accessory ●USB port ●Buli-in water tank ●User management
safety.●
Optional Accessory ●Printer (Impact) ●PTFE filter ●Sample temperature sensor ●Electric drainage ●Water supply pipe ●Bulid-in waste
Expandable + optional: For positive waste sterilization, an optional PTFE filter can be added to prevent exhaust pollution and potential tank ●Positive pressure compensation ●Auto un-lock and door open ●Leakage test module ●Vacuum test module
personnel poisoning. ●Water cooling system
Shipping Size(WxDxH)(mm) 700x780x1070 700x780x1070 700x780x1320 700x780x1070 700x780x1320 700x780x1410 700x780x1320 700x780x1410
●Reservation and preheating: Time reservation for automatic sterilization. Preheating capability prevents condensation due to Gross Weight(KG)
temperature drop after drying. (ST-VM/ST-VY) 116/126 123/133 131/141 126/136 140/150 151/161 144/154 155/165

13 14
/ /

Vertical Pressure Steam Sterillzer

Vertical Pressure Steam Sterillzer


Vertical Pressure Steam Sterillzer

Performance feature

Inner tank made of high-quality stainless steel

Self-expanding seal

Hand wheel translation quick opening structure

Water cutoff protection control

Safety interlocking device.

Buzzer reminds automatic shutdown after sterilization is completed

Knob control, time relay, temperature controller

Standard double-layer stainless steel mesh basket

Over-temperature, over-pressure automatic protection function

Optional drying system, gas collecting bucket, etc.

Automatically discharges cold air and automatically discharges steam after sterilization is completed

5-inch Color Touch Screen Simple operation, safe and reliable, purely mechanical control, low failure rate

Specification

Model ST-VLA50 ST-VLA80 ST-VLA100 ST-VL50 ST-VL80I ST-VL100 ST-VL120


Chamber Size(Ø*H) Ø386*515 Ø386*695 Ø386*875 Ø386*515 Ø386*695 Ø386*875 Ø386*1054
Capacity(L) 60 80 100 60 80 100 120
Consumption 4.6KW 6.1KW 7.3KW 3.6KW 5.1KW 6.1KW 6.1KW
Temperature( 121 -134
Highest working pressure 0.25Mpa
Hand Wheel Structure Basket Basket
Temp.Accuracy 0.5
Design pressure 0.28~0.3Mpa
Description
Ambient temperature 5 ~40
The vertical sterilizer is a safe, reliable and automatically controlled sterilization equipment, consisting of a heating system, a micro- Relative humidity ≤85%
computer control system, an automatic overheating and an overheating and overpressure protection system. The container has the Atmospheric pressure 70kpa~106kpa
advantages of reliable sterilization effect, easy operation, safe use, energy saving, durability and low price. It is suitable for pharma-
Power supply AC 220V ± 22V, 50Hz ± 1Hz
ceutical, medical, scientific research and other industries to sterilize equipment, medical dressings, utensils, liquids, culture media,
waste and other items.

15 16
/ /

Vertical Pressure Steam Autoclave

Vertical Pressure Steam Autoclave


Vertical Pressure Steam Autoclave

Performance feature

●Inner tank made of high-quality stainless steel.

●Self-expanding seal.

●Metal power switch, metal handwheel.

●Built-in water tank, water cutoff protection control.

●Safety interlocking device.

●Buzzer reminds automatic shutdown after sterilization is completed.

●Computer controlled automatic cycle and arbitrarily set sterilization program.

●USB communication socket, built-in mobile probe (for various program modes such as open liquid, culture medium, etc.)

●5-inch color touch screen.

●It can automatically replenish water during the whole operation process and realize real internal circulation of soda and water.

●Double automatic protection function for over-temperature and over-pressure.

●The internal circulation system of steam and water is equipped with an exhaust cooling system, which can avoid the generation of
5-inch Color Touch Screen
aerosols ,does not discharge steam to the outside, and the environment is clean and dry.

●Optional drying system and printer.

●Sensor failure alarm.

●Optional reverse pressure quick cooling function.

●Pressure module overpressure alarm protection.

Specification

Model ST-VPA50 ST-VPA80 ST-VPA100 ST-VP50 ST-VP80 ST-VP100 ST-VP120


Chamber Size(Ø*H) Ø386*515 Ø386*695 Ø386*875 Ø386*515 Ø386*695 Ø386*875 Ø386*1054
Capacity(L) 60 80 100 60 80 100 120
Consumption 4.6KW 6.1KW 7.3KW 3.6KW 5.1KW 6.1KW 6.1VA
Hand Wheel Structure Basket Basket Temperature( 121 ~134
Highest working pressure 0.25Mpa
Temp.Accuracy 0.5
Description Design pressure 0.28~0.3Mpa

The vertical sterilizer is a safe, reliable and automatically controlled sterilization equipment, consisting of a heating system, a micro- Ambient temperature 5 ~40
computer control system, an automatic overheating and an overheating and overpressure protection system. The container has the Relative humidity ≤85%
advantages of reliable sterilization effect, easy operation, safe use, energy saving, durability and low price. It is suitable for pharma- Atmospheric pressure 70kpa~106kpa
ceutical, medical, scientific research and other industries to sterilize equipment, medical dressings, utensils, liquids, culture media,
Power supply AC 220V ± 22V, 50Hz ± 1Hz
waste and other items.
Shipping Size(WxDxH)(mm) 560×640×990 560×640×1072 560×640×1258 560×640×990 560×640×1072 560×640×1258 560×640×1437
Gross Weight(KG) 111 135 155 106 130 150 155

17 18
/ /

Vertical Pressure Steam Autoclave

Vertical Pressure Steam Autoclave


Vertical Pressure Steam Autoclave

Performance feature

●Inner tank made of high-quality stainless steel

●Self-expanding seal

●Metal power switch, metal handwheel

●Water cutoff protection control

●Safety interlocking device.

●Buzzer reminds automatic shutdown after sterilization is completed

●Computer controlled automatic cycle and arbitrarily set sterilization program

●USB communication socket, built-in mobile probe (for various program modes such as open liquid, culture medium, etc.)

●5-inch color touch screen

●Built-in water tank and quick-release filter for easy maintenance and cleaning.

●Double automatic protection function for over-temperature and over-pressure.

5-inch Color Touch Screen ●The internal circulation system of steam and water is equipped with an exhaust cooling system, which can avoid the generation of
aerosols ,does not discharge steam to the outside, and the environment is clean and dry

●Sensor failure alarm

●Optional reverse pressure quick cooling function

●Pressure module overpressure alarm protection.

Specification

Model ST-VGA50 ST-VGA80 ST-VGA100 ST-VG50 ST-VG80 ST-VG100 ST-VG120


Chamber Size(Ø*H) Ø386*515 Ø386*695 Ø386*875 Ø386*515 Ø386*695 Ø386*875 Ø386*1054
Capacity(L) 60 80 100 60 80 100 120
Consumption 4.6KW 6.1KW 7.3KW 3.6KW 5.1KW 6.1KW 6.1KW
Hand Wheel Structure Basket Basket Temperature( 121 -134
Highest working pressure 0.25Mpa
Description Temp.Accuracy 0.5

The vertical sterilizer is a safe, reliable and automatically controlled sterilization equipment, consisting of a heating system, a micro- Design pressure 0.28~0.3Mpa
computer control system, an automatic overheating and an overheating and overpressure protection system. The container has the Ambient temperature 5 ~40
advantages of reliable sterilization effect, easy operation, safe use, energy saving, durability and low price. It is suitable for pharmaceu- Relative humidity ≤85%
tical, medical, scientific research and other industries to sterilize equipment, medical dressings, utensils, liquids, culture media, waste Atmospheric pressure 70kpa~106kpa
and other items.
Power supply AC 220V ± 22V, 50Hz ± 1Hz
Shipping Size(WxDxH)(mm) 560×640×990 560×640×1072 560×640×1258 560×640×990 560×640×1072 560×640×1258 560×640×1437
Gross Weight(KG) 111 135 155 106 130 150 155

19 20
/ /

Vertical Pressure Steam Autoclave


LED Display,No Water Tank

Vertical Pressure Steam Autoclave


Vertical Pressure Steam Autoclave

Description
Application
●Multi Safety Features
The Vertical Autoclave is a small type automatic, high-tem-perature and high-pressure sterilization device that using pressure steam as
the medium. It is suitable for various levels of biological research institutions such as biological laboratories. It is used for sterilizing -Overheat protection automatically cutting off heating with audible and visual indication when the autoclave is overheated.
items such as fabrics, instruments, rubber, liquids, culture medium and waste materials.
-Anti-burning protection automatically cutting off heating with audible and visual indication when the autoclave is lack of water.

-Interlock safety: preventing from starting a cycle if door is not properly closed or from door opening if there is pressure inside cham-
ber.

-Steam pressure safety valve: automatically works when the chamber pressure exceeds the limit.

●Automatic operation of sterilization with PID microprocessor control and LED display.

●Positive pulsing function removes the air efficiently.

●Automatic air discharge control prevents boil over of liquids and explosions of bottles

●With Power-off memory function.

Specification
Model
Model ST-V50E ST-V80E ST-V100E ST-V120E ST-V150E
Capacity 50L 80L 100L 120L 150L
External Size(W*D*H) 476x616x990mm 546x688x1030mm 566x708x1295mm 620x760x1190mm 680x820x1180mm
Chamber Size(Ø*H) Φ316x667mm Φ386x695mm Φ400x800mm Φ450x760mm Φ500x780mm
Pressure gauge Chamber Material SUS304
Sterilization Temperature 105~136
Designed Temperature 150
Max.Working Temperature 136
Dissolution Temperature 40~100
Warming Temperature 40~60
Temp.Accuracy 0.1
Designed Pressure 0.3Mpa
Max.Sterilization Pressure 0.23Mpa
Sterilization Time range 0~99min
Temp. Controller PID control by microprocessor
Display LED display
Drying Function YES
Drying Time 15min
●High temperature instrument program
●Low temperature instrument program
Sterilization Program ●Culture medium sterilization warming program
●Culture medium dissolution warming program
Standard SS Basket Water Injection method Manual
Consumption 3KW 4.5KW 5.4KW 5.4KW 9.3KW
Power Supply 220V 50Hz 220V 50Hz 220V 50Hz 380V 50Hz 380V 50Hz
Standard SS Basket 2pcs 2pcs 2pcs 2pcs 2pcs
Net Weight 55kg 65kg 80kg 165kg 185kg
Gross Weight 90kg 100kg 120kg 205kg 225kg
Shipping Size(W*D*H) 730x700x1250 800x770x1300 800x770x1480 950x900x1280 950x900x1440

21 22
/ /

Vertical Pressure Steam Autoclave


LED Display,Built-in Water Tank

Anti-burning protection automatically cutting off heating with audible and visual indication when the autoclave is lack of water.
Vertical Pressure Steam Autoclave

Vertical Pressure Steam Autoclave


Application
Interlock safety: preventing from starting a cycle if door is not properly closed or from door opening if there is pressure inside
The Vertical Autoclave is a small type automatic, high-temperature and high-pressure sterilization device that using pressure steam as
the medium. It is suitable for various levels of biological research institutions such as biological laboratories. It is used for sterilizing chamber.
items such as fabrics, instruments, rubber, liquids, culture medium and waste materials.
Steam pressure safety valve: automatically works when the chamber pressure exceeds the limit.

●Automatic operation of sterilization with PID microprocessor control and LED display.

●Positive pulsing function removes the air efficiently.

●Automatic air discharge control prevents boil over of liquids and explosions of bottles.

●With Power-off memory function.

●Standard Printer (thermal).

Specification

Model ST-V50 ST-V80 ST-V100 ST-V120 ST-V150


Capacity 50L 80L 100L 120L 150L
External Size(W*D*H) 476x616x990mm 546x688x1030mm 566x708x1295mm 620x760x1190mm 680x820x1180mm
Pressure gauge
Chamber Size(Ø*H) Φ316x667mm Φ386x695mm Φ400x800mm Φ450x760mm Φ500x780mm
Chamber Material SUS304
Sterilization Temperature 105~136
Designed Temperature 150
Max.Working Temperature 136
Dissolution Temperature 40~100
Warming Temperature 40~60
Temp.Accuracy 0.1
Designed Pressure 0.3Mpa
Max.Sterilization Pressure 0.23Mpa
Sterilization Time range 0~99min
Temp. Controller PID control by microprocessor
Display LED display
Drying Function YES
Standard SS Basket
Drying Time 15min
Sterilization Program ●High temperature instrument program
●Low temperature instrument program
●Culture medium sterilization warming program
●Culture medium dissolution warming program
Water Injection method Built-in water tank
Description
Consumption 3KW 4.5KW 5.4KW 5.4KW 9.3KW
●Multi Safety Features Power Supply 220V 50Hz
220V 50Hz 220V 50Hz 380V 50Hz 380V 50Hz
Overheat protection automatically cutting off heating with audible and visual indication when the autoclave is overheated. Standard SS Basket 2pcs 2pcs 2pcs 2pcs 2pcs
Net Weight 68kg 92kg 100kg 175kg 195kg
Gross Weight 103kg 127kg 140kg 215kg 235kg
Shipping Size(W*D*H) 730x820x1250 800x880x1300 800x880x1480 950x1050x1280 950x1050x1440

23 24
/ /

Vertical Pressure Steam Autoclave


LCD Display,Built-in Water Tank

Vertical Pressure Steam Autoclave


Vertical Pressure Steam Autoclave

Description
Application
●Multi Safety Features
Porous load: fabric, sterile clothing, wrapped equipment, filter element, rubber, etc.
-Overheat protection automatically cutting off heating with audible and visual indication when the autoclave is overheated.
Hard load: tools, lumen instruments, glassware, metal containers, etc.
-Anti-burning protection automatically cutting off heating with audible and visual indication when the autoclave is lack of water.
Liquid load: non-sealed liquids such as culture media
-Interlock safety: preventing from starting a cycle if door is not properly closed or from door opening if there is pressure inside chamber.

-Steam pressure safety valve: automatically works when the chamber pressure exceeds the limit.

●Automatic operation of sterilization with PID microprocessor control and LCD display.

●Positive pulsing function removes the air efficiently.

●Liquid program with automatic pressure control prevents boil over of liquids and explosions of bottles or bags.

●With Power-off memory function.

●Internal water tank with steam-water internal circulation.

●Standard printer, real-time record sterilization process,error and alarm information.

Model
Model
Specification
Model
Model ST-V50A ST-V80A ST-V100A ST-V120A ST-V150A
Capacity 50L 80L 100L 120L 150L
External Size(W*D*H) 476x616x990mm 546x688x1030mm 566x708x1295mm 620x760x1190mm 680x820x1180mm
Pressure gauge
Chamber Size(Ø*H) Φ316x667mm Φ386x695mm Φ400x800mm Φ450x760mm Φ500x780mm
Chamber Material SUS304
Sterilization Temperature 105~136
Designed Temperature 150
Max.Working Temperature 136
Dissolution Temperature 40~100
Warming Temperature 40~60
Temp.Accuracy 0.1
Designed Pressure 0.3Mpa
Max.Sterilization Pressure 0.23Mpa
Sterilization Time range 0~99min
Display LCD touch screen
Drying Function YES
Drying Time 15min
●Dressings/Fabric program ●Tools program ●Rubber program ●Culture medium sterilization warming program
Sterilization Program ●Culture medium dissolution warming program ●Liquids in closed containers (glass bottle, plastic bottle and bag)
program ●Waste program
Water Injection method Built-in water tank
Standard SS Basket Consumption 3KW 4.5KW 5.4KW 5.4KW 9.3KW
Power Supply 220V 50Hz 220V 50Hz 220V 50Hz 380V 50Hz 380V 50Hz
Standard SS Basket 2pcs 2pcs 2pcs 2pcs 2pcs
Optional Accessory Printer, USB port
Net Weight 68kg 92kg 100kg 175kg 195kg
Gross Weight 103kg 127kg 140kg 215kg 235kg
Shipping Size(W*D*H) 730x820x1250 800x880x1300 800x880x1480 950x1050x1280 950x1050x1440

25 26
/ /

Vertical Pulse Vacuum Autoclave


Vertical Pulse Vacuum Steam Autoclave

Horizontal Pressure Steam Autoclave


Application Description

Porous load: fabric, sterile clothing, wrapped equipment, filter element, rubber, etc. Automatic operation from sterilization to drying (Pre-vacuum sequence, heating, sterilization, steam discharge and post-vacuum
drying) can be easily set and monitored by the LCD touch screen
Hard load: tools, lumen instruments, glassware, metal containers, etc.
Built-in vacuum pump provides 3-time pulsed vacuum process,extracting the air out of the medical dressing and the Class A hollow
medical instrument efficiently.

●Built-in steam generator prevents the media from contamination.

●Post-vacuum drying function offers efficient drying results.

●Positive pulsing function removes the air efficiently.

●With 0.22µm air filter, blocks all microorganisms from exiting the chamber during the air removal phase.

●Standard built-in printer.Optional real-time record sterilization process, error and alarm information.

Specification

Model ST-V50PV ST-V80PV ST-V100PV ST-V120PV ST-V150PV


Capacity 50L 80L 100L 120L 150L
External Size(W*D*H) 476x636x990mm 636x710x1030mm 630x710x1295mm 620x780x1190mm 680x820x1180mm
Chamber Size(Ø*H) Φ316x667mm Φ386x695mm Φ386x860mm Φ450x760mm Φ500x780mm
Chamber Material SUS304
Pressure gauge Sterilization Temperature 105~136
Designed Temperature 150
Max.Working Temperature 136
Dissolution Temperature 40~100
Warming Temperature 40~60
Temp.Accuracy 0.1
Temp. Uniformity ≤±0.5
Designed Pressure 0.3Mpa
Max.Sterilization Pressure 0.23Mpa
Ultimate Vacuum -0.096MPa
Sterilization Time range 0~99min
Temp. Controller PID control by microprocessor
Display LCD touch screen
Drying Function YES
Drying Time 15min
●Medical instrument program ●Luminal instruments ●Dressings/Fabric program
Sterilization Program ●Rubber program ●Liquid program ●BD Test program ●Vacuum Test progrm
●4 pcs DIY programs
Standard SS Basket Water Injection method Built-in water tank
Consumption 3KW 4.5KW 5.4KW 7.2KW 9.3KW
Power Supply 220V 50Hz 220V 50Hz 220V 50Hz 380V 50Hz 380V 50Hz
Standard SS Basket 2pcs 2pcs 2pcs 2pcs 2pcs
Optional Accessory Printer, USB port
Net Weight 85kg 95kg 112kg 180kg 210kg
Gross Weight 120kg 130kg 152kg 220kg 250kg
Shipping Size(W*D*H) 830x840x1250 880x910x1300 800x880x1550 1030x1050x1300 1030x1050x1450

27 28
/ /

Vertical Pulse Vacuum Steam Autoclave, ST-VPVM series Horizontal Pressure Steam Autoclave

Advantage Application
Vertical Pulse Vacuum Steam Autoclave

Horizontal Pressure Steam Autoclave


●Hand wheel translation and quick opening structure The autoclave is suitable for hospital, scientific research institute and other
units to sterilize asepsis clothes, tools, ulture medium and so on.
●Three times Pre-Vacuum

●High quality stainless steel material Advantage


●Horizontal circular inner chamber and jacket structure for safety, heat preser-
●Microcomputer automatic control, arbitrary setting
vation, and drying function.
●Sterilization parameters
●Control system featuring HMI man-machine interface, touch screen, and
●LED touch screen display microcomputer for intuitive operation.

●Disc type embedded gear and Electric lock door structure for reliable and
●Standard test interface
secure operation.
●Self-expanding seal
●Self-expanding door seals for effective sealing.
●With drying function ●Equipped with a standard equipment verification interface for easy calibration
and validation.
●Safety interlocking device
●Optimal design pipeline system with SUS 304 pipeline and clamp structure for
●With stainless steel bucket or basket
quick assembly.

●FO value and temperature-time double guarantee sterilization effect, with


comprehensive sterilization records.
●With automatic protection function:over temperature protection; over pressure self discharge protection ;low water level protection;
anti dry burning. ●Multiple safety protections including over-temperature and over-pressure
protection.
●Automatic water
●Built-in printer & USB function.
●External steam Generator

●Buzzer reminder after sterilization, automatic stop. Specification


●Automatic discharge of cold air, automatic steam exhaust after sterilization Model
Model ST-H150I ST-H200I ST-H300I ST-H400I
●Internal printer Capacity(L) 150 200 300 400
Chamber Size((Ø*L)mm Ø472x880 Ø472x1150 Ø620x995 Ø620x1320
Chamber Material 304 Stainless Steel
Specification Design pressure -0.1~0.25Mpa
Model Design temperature 139
Model ST-V35PVM ST-V50PVM ST-V75PVM ST-V100PVM Working pressure 0.225Mpa
Capacity 35L 50L 75L 100L Working Temperature 105-136
Consumption 3.5kW 3.5kW 4.5kW 4.5kW Vacuum -0.086Mpa
Power Supply AC 220V 50HZ Temp accuracy ±1
Design pressure 0.25Mpa Temperature Equilibrium <+2.0
Design temperature 139 Compressed air pressure 0.4~0.7Mpa
Rated working pressure 0.22Mpa Steam Pressure 0.3-0.7Mpa
Rated working temperature 134 Drying Function Double door/Single door
Sterilization temperature 116 ~134 Printer YES
Sterilization time setting range 4~120 min Door Built-in Printer
Temp accuracy ±1 USB Port YES
Drying time setting range 0~240min Steam Consumption(KG) ≤12 ≤13 ≤18 ≤22
Chamber Size (mm) Ø350×400 Ø400×505 Ø400×655 Ø450×650 Water Consumption(KG) ≤30 ≤30 ≤40 ≤40
Bucket Size Ø330×320 Ø380×430 Ø380×560 Ø420×540 Consumption(KW) ≤2
Basket Size(optional) Ø320×350 Ø360×280×1 Ø360×280×2 Ø410×300×2 Power Supply Three-phase five-wire AC 380V 50Hz
External Size(W×D×H)(mm) 698×498×940 630×500×1180 725×525×1180 760×560×1260 External Sizes (WxDxH)(mm) 1290x690x1680 1560x690x1680 1430x850x1760 1755x850x1760
Shipping Size (W×D×H)(mm) 790x610x1130 800x600x1230 840x610x1310 870x650x1380 Shipping Size(WxDxH)(mm) 1570x840x1850 1740x840x1850 1610x990x1900 1935x990x1900
N.W./G.W. (kg) 115/130 118/138 124/150 132/160 N.W./G.W. (kg) 400/500 450/550 510/610 650/750

29 30
/ /

Horizontal Pressure Steam Autoclave


Horizontal Pressure Steam Autoclave

Description Advantage

Horizontal Pressure Steam Autoclave


Horizontal Pressure Steam
Autoclave uses saturated steam as ●Multi Safety Features
the sterilization medium,exhausts
the cold air by gravity. So that the -Overheat protection automatically cutting off heating with audible and visual alarming when the autoclave is overheated.
saturated steam fully penetrates
-Interlock safety: preventing from starting a cycle if door is not properly locked or from door opening if the chamber is pressurized.
the surface of the items. By
utilizing high-temperature saturat- -Anti-burning protection automatically cutting off heating with audible and visual indication when the autoclave is lack of water.
ed steam under high pressure, it
effectively kills all microorganisms -Steam pressure safety valve: automatically works when the chamber pressure exceeds the limit.
and their spores, providing reliable
●Electrical circuit safety device for over-current and shortcut protection.
sterilization results. It is considered
one of the most effective methods ●Automatic operation of sterilization (heating, sterilization, steam discharge and drying) can be easily set and monitored by the LCD
among physical sterilization touch screen.
techniques.
●Built-in steam generator prevents the media from contamination.
It is widely used in fields such as
pharmaceuticals, biotechnology, ●Buzzer for working errors and the end of the cycle.
and laboratories for sterilizing
●Automatic atmospheric pressure adjustment for the installation and operation in plateau areas.
items such as dressing/fabrics,
medical instruments, rubber, and ●Standard printer.Optional real-time record sterilization process, error and alarm information.
liquids.
ST-H60/80/100 ST-H135 /150/185/200/300/450

Specification
Model
Model ST-H60 ST-H80 ST-H100 ST-H135 ST-H150 ST-H185 ST-H200 ST-H300 ST-H450
Capacity 60L 80L 100L 135L 150L 185L 200L 300L 450L
External Size(W*D*H) 985x600x1450mm 985x600x1450mm 1200x600x1450mm 1000x710x1550mm 1330x710x1550mm 1250x710x1550mm 1320x710x1550mm 1350x900x1750mm 1750x900x1750mm
Chamber Size(Ø*H) Φ386x500mm Φ386x700mm Φ386x950mm Φ500x700mm Φ500x780mm Φ500x950mm Φ500x1000mm Φ650x950mm Φ650x1360mm
Chamber Material SUS304
Sterilization Temperature 105~136
Designed Temperature 150
Max.Working Temperature 136
Temp.Accuracy 0.1
Temp. Uniformity ≤±0.5
Designed Pressure 0.3Mpa
Sterilization Pressure 0.23Mpa
Temp. Controller PID control by microprocessor
Display LCD touch screen
Drying Function YES
Drying Time 15min
●Dressings/Fabric program
●Medical instrument program
Sterilization Program ●Rubber program
●Liquid program
●DIY program
Consumption 5.7KW 5.7KW 6KW 9.35KW 9.35KW 10.8KW 10.8KW 12KW 20KW
Power Supply 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz
Standard Sterilization Shelf 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs
Standard Accessory Printer
Net Weight 105kg 120kg 140kg 225kg 240kg 255kg 265kg 350kg 450kg
Gross Weight 165kg 185kg 210kg 315kg 340kg 365kg 380kg 470kg 570kg
Shipping Size(W*D*H) 1280×700×1600mm 1280×700×1600mm 1450×700×1600mm 1300×850×1900mm 1400×900×1900mm 1440×930×2020mm 1440×930×2020mm 1650×1100×2120mm 2100×1100×2120mm

31 32
/ /

Horizontal Pressure Steam Sterilizer Horizontal Pressure Steam Sterilizer

Application
Horizontal Pressure Steam Sterilizer

Sterilizer
Autoclave
The horizontal autoclave, which adopt the way of gravity exchange to release the Advantage
cold air from the chamber more completely to ensure the reliable sterilization. The
control system automatically adjust the steam inlet and outlet according to chamber * For 4~6 minutes rapidly sterilization.

Horizontal Pressure Steam


temperature during sterilizing. This unit is an ideal equipment for clinics, scientific
* Digital display of working status, touch type key.
research institutes and other organizations to sterilize surgical instruments, fabric
glassed, and the culture media etc. * With-protection of water lacking.

* Exhaust the h 3 fixed cycles of sterilizing program and adjustable for user.

* The whole process of water adding, temperature rising, sterilizing, drying


steam discharging controlled automatically.

* Steam-water inner circulation system: no steam discharge, and the


environment for sterilizing will be clean and dry.

* Safe cool air automatically.

* Door safety lock system.

Pressure Gage Water Level Meter Chamber * With 2 stainless steel sterilizing plates.

* The chamber of the sterilizer is made of stainless steel.

Advantage * Automatically shut off with beep reminding after sterilization.

* Automatically controlled the sterilizing process, easy to operate. * With drying function.

* With drying function, suitable for the medical dressing drying.


Application
* With over-temperature, over-pressure auto-protect.
The product utilizes saturated steam to rapidly and efficiently sterilize
* The door can not be open until the pressure in the chamber is reduced to 0.027MPa. And the unit can not be started on if the door articles. lt can be used in hospitals, public health centers, health stations,
doesn't close well. clinics in factories and mines, scientific research institutes, etc, to sterilize
medical equipment, surgical dressings, glass utensils, solutions and
* The safety valve will be open automatically when the inner pressure over 0.24MPa and the steam would be exhausted the water substrata. It can also be used as cooker in plateau areas and be used to
tank. make high quality drinking water in industrial and mining enterprises.

* Automatically cut off the power if the water-lack happened, and alarm at the meanwhile.

* The chamber of the sterilizer is made of stainless steel.

Specification
Specification
Model ST-H150YDA ST-H200YDA ST-H280YDA ST-H400YDA ST-H500YDA
Capacity 150L 200L 280L 400L 500L Model ST-H100YDB
Chamber Size(ΦxH) Φ440x1000mm Φ515x1000mm Φ600x1000mm Φ700x1100mm Φ700x1300mm Capacity 100L
Working Pressure 0.22MPa Chamber Size(ΦxH) Φ390x670mm
Max. Working Temperature 134 Working Pressure 0.22MPa
Sterilization Temperature 40~134 Max. Working Temperature 134
Sterilization Time Range 0~60min Sterilization Temperature 105~134
Drying Time Range 0~60min Sterilization Time Range 0~99min
Temp. Uniformity ≤±2 Temp. Uniformity ≤±1
Consumption 9KW 9KW 12KW 18KW 18KW Consumption 9KW
Power Supply 380V, 50Hz Power Supply 220V, 50Hz
External Size(WxDxH) 1400x600x1300mm 1400x670x1650mm 1400x770x1780mm 1430x880x1830 1800x900x1820 External Size(WxDxH) 930x690x1330mm
Net Weight 240kg 260kg 365kg 420kg 470kg Net Weight 245kg
Gross Weight 320kg 350kg 465kg 530kg 580kg Gross Weight 280kg
Shipping Size(WxDxH) 1560x750x1850mm 1560x820x1850mm 1680x920x1950mm 1600x1050x2050mm 1850x1050x2050mm Shipping Size(WxDxH) 1100x720x1420mm

33 34
/ /

Horizontal Pressure Steam Sterilizer


Horizontal Pressure Steam Sterilizer

Sterilizer
Autoclave
Horizontal Pressure Steam
Application

The horizontal autoclave, which adopted the way of gravity exchange to release the cold air from the chamber more completely to
ensure the reliable sterilization. The control system automatically adjust the steam inlet and outlet according to chamber temperature
during sterilizing. This unit is an ideal equipment for clinics, scientific research institutes and other organizations to sterilize surgical
instruments, fabric glassed, and the culture media etc.
Advantage

* Automatically controlled the sterilizing process, easy to operate.

* With drying function, suitable for the medical dressing drying.

* With over-temperature, over-pressure auto-protect.

* The door can not be open until the pressure in the chamber is reduced to 0.027MPa. And the unit can not be started on if the door
doesn't close well.

* The safety valve will be open automatically when the inner pressure over 0.24MPa and the steam would be exhausted the water
tank.

* Automatically cut off the power if the water-lack happened, and alarm at the meanwhile.

* The chamber of the sterilizer is made of stainless steel.

* With printer.
Specification

Model ST-H150YDB ST-H200YDB ST-H280YDB


Capacity 150L 200L 280L
Chamber Size(ΦxH) Φ440x1000mm Φ515x1000mm Φ600x1000mm
Working Pressure 0.22MPa
Max. Working Temperature 134
Sterilization Temperature 40~134
Sterilization Time Range 0~99min
Drying Time Range 0~99min
Temp. Uniformity ≤±2
Consumption 9KW 9KW 12KW
Power Supply 380V, 50Hz
External Size(WxDxH) 1400x650x1820mm 1400x750x1700mm 1520x910x1900mm
Net Weight 340kg 350kg 462kg
Gross Weight 430kg 436kg 570kg
Shipping Size(WxDxH) 1560x860x1820mm 1560x910x1880mm 1680x1060x2050mm

35 36
/ /

Horizontal Pulse Vacuum Autoclave

Advantage
Horizontal Pulse Vacuum Autoclave

Horizontal Pulse Vacuum Autoclave


●Multi Safety Features

Description Overheat protection automatically cutting off heating with audible and visual alarming when the autoclave is overheated.
Horizontal Pulse Vacuum Autoclave uses saturated steam as the steriliza-
Interlock safety: preventing from starting a cycle if door is not properly locked or from door opening if the chamber is pressurized.
tion medium. Through the pulsating vacuum method, the saturated steam
fully penetrates the surface of the items. By utilizing high-temperature Anti-burning protection automatically cutting off heating with audible and visual indication when the autoclave is lack of water.
saturated steam under high pressure, it effectively kills all microorganisms
and their spores, providing reliable sterilization results. It is considered the Steam pressure safety valve: automatically works when the chamber pressure exceeds the limit.
most effective method among physical sterilization techniques.
●Automatic operation from sterilization to drying (Pre-vacuum sequence, heating, sterilization, steam discharge and post-vacuum
drying) can be easily set and monitored by LCD touch screen.

It is widely used in fields such as pharmaceuticals, biotechnology, and ●Built-in vacuum pump provides 3-time pulsed vacuum process,extracting the air out of the medical dressing and the A degree hollow
laboratories for sterilizing items such as dressing/fabrics, medical instru- medical instrument efficiently.
ments, lumen instruments, rubber, and liquids.
●Built-in steam generator prevents the media from contamination.

●Post-vacuum drying function offers efficient drying results.

●Positive pulsing function removes the air efficiently.

●With 0.22µm air filter, blocks all microorganisms from exiting the chamber during the air removal phase.

●Standard printer.Optional real-time record sterilization process, error and alarm information.

Specification
Model
Model ST-H60PV ST-H80PV ST-H100PV ST-H135PV ST-H150PV ST-H185PV ST-H200PV ST-H300PV ST-H450PV
Capacity 60L 80L 100L 135L 150L 185L 200L 300L 450L
External Size(W*D*H) 985x600x1450mm 985x600x1450mm 1200x600x1450mm 1000x710x1550mm 1330x710x1550mm 1250x710x1550mm 1320x710x1550mm 1350x900x1750mm 1750x900x1750mm
Chamber Size(Ø*H) Φ386x500mm Φ386x700mm Φ386x950mm Φ500x700mm Φ500x780mm Φ500x950mm Φ500x1000mm Φ650x950mm Φ650x1360mm
Chamber Material SUS304
Sterilization Temperature 105~136
Designed Temperature 150
Max.Working Temperature 136
Temp.Accuracy 0.1
Temp. Uniformity ≤±0.5
Designed Pressure -0.1/0.3Mpa
Sterilization Pressure 0.23Mpa
Pressure Resolution 0.1KPa
Ultimate Vacuum -0.096MPa
Temp. Controller PID control by microprocessor
Display LCD touch screen
Drying Function YES
Drying Time 15min
Sterilization Program ●Medical instrument program ●Luminal instruments ●Dressings program ●Rubber program
●Liquid program ●BD Test program ●Vacuum Test progrm ●4 pcs DIY progrm
Consumption 5.7KW 56.6KW 6.6KW 12KW 12KW 12KW 12KW 20KW 20KW
Power Supply 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz 380V 50Hz
Standard Sterilization Shelf 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs 1pcs
Standard Accessory Printer
Net Weight 105kg 120kg 140kg 225kg 240kg 255kg 265kg 350kg 480kg
Gross Weight 165kg 185kg 210kg 315kg 340kg 355kg 380kg 470kg 570kg
Shipping Size(W*D*H) 1280×700×1600mm 1280×700×1600mm 1450×700×1600mm 1300×850×1900mm 1400×900×1900mm 1440×930×2020mm 1440×930×2020mm 1650×1100×2120mm 2100×1100×2120mm

37 38
/ /

Horizontal Pulse Vacuum Autoclave

Horizontal Pulse Vacuum Autoclave


Horizontal Pulse Vacuum Autoclave

5.The built-in efficient steam generator can quickly generate constant-pressure steam without contaminating sterilized items.

6.The vacuum drying system ensures fabric drying below 1% and instrument drying below 0.1%.

7.A 0.22μm bacterial air filter is included to ensure that the air introduced after breaking the vacuum is sterile, avoiding recontamina-
tion.

8.An optional mini printer can be selected to record real-time sterilization processes and fault alarm information.

9.Automatic calibration of atmospheric pressure is available to avoid high-altitude reactions.

Safety Function

Over-temperature protection device: When the temperature of the pot exceeds the set temperature, the system automatically cuts off
the heating power and displays an alarm.

Door safety interlock device: The program can only start when the door is closed properly; the door cannot be opened when there is
dangerous pressure in the inner chamber.

Dual overpressure protection: Automatic pressure relief safety valve, when the pressure exceeds the design pressure, the safety valve
opens to release the pressure; automatic overpressure protection control, when the pressure exceeds the set pressure, automatic
steam discharge and pressure relief, and displays an alarm.

Low water level alarm device: When the water level is below the specified level, the heating is automatically cut off and an alarm is
displayed.
ST-H60PVI ST-H80PVI ST-H100PVI ST-H135PVI ST-H185PVI ST-H300PVI
Description

The ST-HPVI series cabinet-type pulsating vacuum sterilizer utilizes saturated steam as the sterilization medium. Through pulsating
vacuum, the saturated steam completely penetrates the surface of the items, using high temperature and pressure to kill all microor-
ganisms and their spores. The sterilization effect is reliable, making it the most effective method in physical sterilization. It is widely
used in pharmaceuticals, biotechnology, laboratories, and other fields to sterilize items such as fabrics, sterile clothing, instruments,
tools, luminal instruments, rubber, liquids, and culture media.
Specification

Model ST-H60PVI ST-H80PVI ST-H100PVI ST-H135PVI ST-H185PVI ST-H300PVI


Advantage Capacity (L) 60 80 100 135 185 300

1. Using microcomputer control technology and touch screen operation, the entire process (pulse vacuum, heating, sterilization, Chamber Size ( Φ ×L) 386×500 386×700 386×950 500×700 500×950 630×1000
exhaust, vacuum drying) runs automatically. External Size (L×W×H) 985×600×1450 985×600×1450 1200×600×1450 1000×710×1550 1250×710×1550 1350×900×1750
Net Weight 105 120 140 225 255 350
2.The touch screen automatically displays the running process status, temperature, pressure, time, and other parameters, as well as
fault alarm information. Power Supply 380V 50Hz
Consumption 5.7Kw 5.7Kw 6Kw 9.35Kw 10.8Kw 12Kw
3.There are four sterilization programs available for instruments, dressings, luminal instruments, rubber, liquids, and custom options,
Design Pressure -0.1/0.3MPa
as well as BD testing and vacuum testing programs for users to choose from.
Ultimate vacuum -0.096MPa
4.The built-in vacuum pump allows for three pulsating vacuums (adjustable), with a vacuum degree of up to -0.090 MPa, which can Sterilizing Temp. 105~136
completely remove air from dressings and Class A luminal instruments.
Max Temp. 136
Chamber Material SUS304 stainless steel
Rated Working Pressure 0.23Mpa
Pressure Display Accuracy 0.1KPa
Temp. Display Accuracy 0.1
Temp. Uniformity ≤±0.5

39 40
/ /

Horizontal Pulse Vacuum Sterilizer


Motorized Vertical Sliding Door

The inflatable pressure-driven sealing ring achieves automatic sealing. The new high-quality tear-resistant silicone rubber material
circular hollow door sealing ring is treated with a special surface coating, which is wear-resistant and smooth, providing better sealing
Autoclave

Autoclave
effect and longer service life. The lifespan can reach up to 1500 cycles.

The main structure is rectangular in section, with a European-style circular reinforcement rib clamp-on flange structure. The sealing
groove is directly welded to the two end circular clamps.
Control Panel
Control system: Siemens programmable logic controller (PLC) and color touch screen. The entire process is automatically controlled
by the program, displaying the running process status, running curve, and parameters such as temperature, pressure, and time.

Intelligent maintenance system: Equipped with the MNcold Internet of Things remote monitoring and maintenance module, it can
Pump connect to the device's information in real-time and remotely from anywhere using a computer, tablet, or smartphone. This includes
Display Screen real-time and historical operating data, fault troubleshooting notifications, and remote software upgrades.

It has sterilization programs for fabrics, instruments, luminal instruments, rubber, liquids, and custom items, as well as testing
programs such as BD testing and vacuum leak testing.

Pulsation modes: It uses negative pressure pulsation, cross-pressure pulsation, and positive pressure pulsation modes, with an air
exclusion rate of over 99.99%. This saves steam consumption and improves steam sterilization and drying efficiency.

Flash steam vacuum drying technology, fabric drying degree less than 0.6%, instrument drying degree less than 0.1%.
Pressure Meter Control Panel Door
Advantage Real-time printing of process data through built-in printer, equipped with paperless recorder for pharmaceutical customers.
The structure of the lift door switch is a motor-driven chain that vertically lifts and slides the door, eliminating the need for manual
operation and saving space. It is a double-door channel type with interlocking doors. It is equipped with pressure safety interlock and Pipeline system: Stainless steel sanitary grade pipeline, clamp connection, all welds are welded with automatic pipe welding machine;
temperature safety interlock for liquid programs. main control valves are imported German angle seat valves and solenoid valves, German SPECK direct-coupled water ring vacuum
pump.

Specification
Model
Model ST-BV150PV ST-BV180PV ST-BV250PV ST-BV300PV ST-BV360PV ST-BV450PV ST-BV600PV ST-BV800PV
Chamber Size(ØxL)mm 450×450×750 450×450×900 550×550×845 550×550×1000 610×610×1000 652×690×1000 652×690×1350 652×690×1800
Capacity 150L 180L 250L 300L 360L 450L 600L 800L
External Size(W*D*H)mm 1045×1120×1800 1185×1120×1800 1200×1290×1850 1350×1290×1850 1350×1350×1850 1350×1400×1950 1700×1400×1950 2150×1400×1950
Steam Consumption(KG/C) 12 12 16 18 20 22 35 52
Water Consumption(KG) 80 90 120 140 160 180 320 400
Power 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ
2KW+(12KW) 2KW+(15KW) 2KW+(20KW) 2KW+(20KW) 2KW+(20KW) 2KW+(30KW) 3KW+(36KW) 3.5KW+(48KW)
Shipping Size Main Unit of Autoclave 1210×1280×2110 1350×1270×2100 1350×1450×2100 1500×1400×2100 1470×1500×2150 1500×1550×2200 1850×1550×2250 2150×1430×2250
(WxDxH)(mm) Single Door Vehicle 1450x700x1100 / 1420×800×1050 / 1600×750×1100 / 2350×1350×1100
Double Door Vehicle 1450x1210x1100 1520×1070×1150 1420×1200×1050 1500×1150×960 / 1450×1150×1050 1900×1150×1050 /
N.W./G.W. (kg) 520/670 560/710 745/895 820/970 870/1020 965/1115 1150/1300 1350/1500
Chamber design pressure -0.1/0.3MPa
Jacket design pressure 0.3MPa
Generator design pressure 1MPa
Design temperaure 150
Maximum operating temperature 139
Maximum operating pressure 0.25MPa
Chamber relief valve 0.28MPa
Jacket relief valve 0.28MPa
Accuracy of temperature 0.1
Accuracy of pressure 0.1KPa
Sterilization temp control accuracy ±1
Temperature uniformity ±1
Negative pressure pulsating vacuum -80 ~0KPa
Across the pressure pulsation -80 ~80KPa
Positive pressure pulse 0 ~ 9 times
Maximum vacuum -96KPa

43 44
/ /

Horizontal Pulse Vacuum Sterilizer


Horizontal Sliding Door

Pulsation modes: Negative pressure pulsation, cross-pressure pulsation, and positive pressure pulsation modes are used, with an air
Horizontal Pulse Vacuum Sterilizer

Horizontal Pulse Vacuum Sterilizer


exclusion rate of over 99.99%, saving steam consumption and improving steam sterilization and drying effects.

Flash steam vacuum drying technology, fabric drying rate is less than 0.6%, instrument drying rate is less than 0.1%.

Real-time printing of process data through the built-in printer, pharmaceutical customers equipped with paperless recorders.

Piping system: Stainless steel sanitary grade piping with clamp connections, all welds are done using automatic pipe welding
machines. The main control valves are imported German angle seat valves and solenoid valves, and the German SPECK direct-cou-
pled water ring vacuum pump is used.
Door
Equipped with a high-efficiency filter for exhaust gas, with a filtration accuracy of 0.22 microns and a sterilization rate of 99.99%. It has
online SIP function to ensure that the discharged gas is sterilized by the high-efficiency filter.

The supply system is equipped with a detection and alarm system: equipped with steam, water source, and power phase sequence
detection and alarm devices. Once the supply system does not meet the equipment operating conditions, it will automatically alarm,
prohibit program startup, and terminate program operation.

It has an automatic drainage temperature regulation system, with a drainage temperature <60 .

Steam Generator

Specification
Model
Model ST-BH120PV ST-BH150PV
Display Screen Electric Box Switch Chamber Size (ØxL)mm 680×1180×1500 680×1180×1800
Capacity 120L 150L
Advantage External Size(W×D×H)(mm) 2007×2110×2113 2357×2110×2113
Weight(kg) 2300 2450
Sliding door switch structure, the door slides horizontally to open and close, driven by a motor chain, no manual operation required. Steam Consumption(KG/C) 600 750
Water Consumption(KG) 65 80
Double-door passage type, equipped with pressure safety interlock and temperature safety interlock for liquid programs, with mutual Power Supply and Consumption 380V,50HZ 3.5KW 72Kw 380V,50HZ 4KW(90KW)
interlocking of the two doors. N.W.(kg) 2300 2450
The inflatable pressure drives the sealing ring to achieve automatic sealing. The new high-quality tear-resistant silicone rubber Chamber design pressure -0.1/0.3MPa
material circular hollow door sealing ring has a special surface coating treatment, which is wear-resistant, smooth, and has better Jecket design pressure 0.3MPa
sealing effect and longer service life, with a lifespan of up to 1500 cycles. Generator design pressure 1.0MPa
Design temperaure 150
The main structure is a rectangular section with a European-style circular reinforcement clamp-on flange structure, and the sealing Maximum operating temperature 139
groove is directly welded to the two end circular clamps. Maximum operating pressure 0.25MPa
Chamber relief valve 0.28MPa
Control system: Siemens programmable logic controller (PLC) and color touch screen, the entire process is automatically controlled by Jacket relief valve 0.28MPa
the program, automatically displaying the operating status, operating curve, and parameters such as temperature, pressure, and time. Accuracy of temperature 0.1
Smart Maintenance System: Equipped with MNcold IoT remote monitoring and maintenance module, it can access device information Accuracy of pressure 0.1KPa
in real time from anywhere through computers, tablets, or smartphones. This includes real-time and historical operational data, Sterilization temperature control accuracy ±1
troubleshooting notifications, and remote software upgrades. Temperature uniformity ±1
Negative pressure pulsating vacuum -80 ~0KPa
It has sterilization programs for fabrics, instruments, luminal instruments, rubber, liquids, waste, customization, as well as testing Across the pressure pulsation -80~80KPa
programs such as BD testing and vacuum leak testing. Positive pressure pulse 0 ~ 9 times
Maximum vacuum -96KPa

45 46
/ /

Horizontal Pulse Vacuum Sterilizer


Motor Hinge Door
Horizontal Pulse Vacuum Sterilizer

Horizontal Pulse Vacuum Sterilizer


Intelligent maintenance system: Equipped with MNcold Internet of Things remote monitoring and maintenance module, it can connect
to computers, tablets, or smartphones to access real-time information of the equipment, including real-time and historical operational
data, fault troubleshooting notifications, and remote software upgrades. It has sterilization programs for fabrics, instruments, lumen
instruments, rubber, liquids, custom programs, as well as testing programs such as BD testing and vacuum leak testing.

Pulsation modes: It adopts negative pressure pulsation, cross-pressure pulsation, and positive pressure pulsation modes, with an air
elimination rate greater than 99.99%. This saves steam consumption and improves steam sterilization and drying efficiency.

Flash steam vacuum drying technology, with fabric drying degree less than 0.6% and instrument drying degree less than 0.1%.

Real-time printing of process data through the built-in printer, equipped with paperless recorders for pharmaceutical customers.

Pipeline system: Stainless steel sanitary-grade pipelines with clamp connections. All welds are done using automatic pipe welding
machines. The main control valves are imported German angle seat valves and solenoid valves, with a German SPECK direct-coupled
water ring vacuum pump.

Supply system detection and alarm system: Equipped with steam, water source, and power phase sequence detection and alarm
devices. If the supply system fails to meet the equipment's operating conditions, it will automatically sound an alarm and prevent
program startup and termination.

It has an automatic drainage temperature regulation system, with a drainage temperature below 60 .

Specification
Model
Model ST-BA650PV ST-BA800PV ST-BA1000PV ST-BA1200PV ST-BA1500PV
Chamber Size(ØxL)mm 610×910×1200 610×910×1500 610×910×1800 680×1180×1500 680×1180×1850
Capacity 650L 800L 1000L 1200L 1500L
External Size(W×D×H)(mm) 1500×1350×1900 1800×1350×1900 2100×1350×1900 1800×1420×2000 2100×1420×2000
Weight(kg) 1250 1350 1550 1650 1850
Steam Consumption(KG/C) 35 47 55 65 80
Water Consumption(KG) 320 400 500 600 750
Power Supply and Consumption 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ
3KW+(36KW) 3.5KW+(48KW) 3.5KW+(54KW) 3.5KW+(72KW) 4KW+(90KW)
N.W. (kg) 1350 1550 1650 1850
Advantage Chamber design pressure 0.3MPa
Generator design pressure 1.0MPa
Design temperaure 150
Sliding door with detachable gear meshing seal structure, driven by motor chain to lift the door up and down.
Maximum operating temperature 139
Double door passage type, equipped with pressure safety interlock and temperature safety interlock for liquid programs, with mutual Maximum operating pressure 0.25MPa
interlocking of the double doors. Chamber relief valve 0.28MPa
Jacket relief valve 0.28MPa
The automatic sealing is achieved by the inflatable pressure-driven sealing ring. The new high-quality tear-resistant silicone rubber Accuracy of pressure 0.1KPa
material circular hollow door sealing ring is coated with a special surface treatment, which is wear-resistant and smooth, providing Sterilization temperature ±1
better sealing effect and longer service life, with a lifespan of up to 1500 cycles. control accuracy
Temperature uniformity ±1
The main body structure is rectangular in section, with a European-style circular reinforcement rib sleeve flangeless structure, and the
Negative pressure -80 ~0KPa
sealing groove is directly welded to the two end circular sleeves.
pulsating vacuum
Control system: Siemens programmable logic controller (PLC) and color touch screen, the entire process is automatically controlled by Across the pressure pulsation -80~80KPa
the program, automatically displaying the operation status, operation curve, and parameters such as temperature, pressure, and time. Positive pressure pulse 0 ~ 9 times
Maximum vacuum -96KPa

47 48
/ /

Portable Autoclave
Autoclave

Autoclave
Digital Display Electric Heating Model
Electric Heating Mode Electric or LPG heating type

Advantage Advantage
●High-quality stainless steel body construction for durability and longevity. ●Microcomputer control with digital display, adjustable sterilization temperature and time

●Self-inflating type seal for effective sealing during operation. ●All Stainless Steel 304 body construction for durability and longevity.

●Electric heating mode for efficient and controlled heating.(For model:ST-P18, ST-P24) ●Overpressure auto-release for enhanced safety.

●Dual heating options: electric and gas.(For model:ST-P18B, ST-P24B) ●Clear and easy-to-read double-scale pressure gauge.

●Automatic release when overpressure occurs, with low water level protection to prevent dry burning. ●Self-inflating seal for secure and reliable operation.

●Double scale indication pressure gauge for accurate monitoring and measurement. ●Simple operation with a focus on safety and reliability.

Specification Specification
Model Model
Model ST-P18 ST-P24 ST-P18B ST-P24B Model ST-P18E ST-P24E
Capacity 18L 24L 18L 24L Capacity 18L 24L
Rated working temperature 126 Rated working temperature 126
Rated working pressure 0.142Mpa Rated working pressure 0.142Mpa
Temp accuracy ±1 Temp accuracy ±1
Consumption 2.0KW Consumption 2.0KW
Power Supply 220V/110V,50Hz/60Hz Power Supply 220V/110V,50Hz/60Hz
Chamber Size(mm) Ø280x295 Ø280x390 Ø280x295 Ø280x390 Chamber Size(mm) Ø280x295 Ø280x390
Shipping Size(W×D×H)(mm) 410x410x560 410x410x680 420x420x430 420x420x540 Shipping Size(W×D×H)(mm) 420x420x560 420x420x680
N.W./G.W. (kg) 16/17 17.5/19 15/16 16/18 N.W./G.W. (kg) 17/18 17.5/19

49 50
/ /

Glass Bead Sterilizer Infrared Bacti-Cinerator Sterilizer

Infrared Bacti-Cinerator Sterilizer


Glass Bead Sterilizer

Model

Description BCS-800A BCS-800B

Glass Bead Sterilizer used for surgical instrument sterilization. Uses high temperature (up to 300°C) with glass beads to eliminate Description
bacteria and microorganisms. Safe and chemical-free sterilization process. Ensures pathogen-free and contaminant-free instruments.
Bacti-Cinerator Sterilizer is a convenient alternative to alcohol lamps for sterilizing inoculating loops and needles. It is commonly used
Compact design saves bench space and can be used in laminar flow hoods.
in biological safety cabinets, exhaust fans, purification tables, and mobile vans for degassing purposes.

Advantage Model
Advantage
●Digital control allows for precise temperature adjustment up to 300°C.
●Safety and convenience: The Bacti-Cinerator Sterilizer offers a safer and more convenient alternative to alcohol lamps.
●Designed for quick sterilization of forceps, scissors, needles, and other small surgical instruments (typically within 10-15 seconds).
●Rapid sterilization: With a highest temperature of 825 , it can sterilize in just 5-7 seconds.
●Safe to use without the need for gases, fumes, open flames, or harsh liquids.
●Smart and lightweight design: It is smartly designed, lightweight, and easy to handle.
●Features a stainless steel inner container with a built-in temperature controller and over-temperature protection for added safety.
●Compatible with anaerobic chambers: The sterilizer can be used effectively in anaerobic chambers.
●Includes a bag of glass beads for efficient and effective sterilization.
●Effective pollution prevention: The use of podzolic organics helps prevent pollution effectively.

●Longer and safer use: Its smart technique ensures longer and safer use of the sterilizer.
Specification
Specification
Model GBS-300L GBS-300H
Temperature range 100-300 Model BCS-800A BCS-800B
Temp accuracy 5 Maximum Temp in Central Zone 825°C ± 25°C 825°C ± 25°C
Display accuracy 1 Maximum Outer Diameter of Ø15mm Ø35mm
Time range 1 min ~ 99h59 min/∞ Sterilization Goods
Heating time 25 min(RT.to300°C) Length of Heating Area 140mm 100mm
Material of lid Stainless steel Heating Time 15mins 15mins
Chamber volume(Φ x depth) 40mmx80mm 40mmx140mm Consumption 150W 250W
Glass beads capacity 150g 300g Power Supply AC110V or 220V, 50/60Hz AC110v or 220V, 50/60Hz
Max.Consumption 120W 250W Fuse 250V 3A 5x20 250V 3A 5x20
Power Supply AC110V or 220V, 50/60Hz Ambient Temperature 5~40 5~40
External Size(W×D×H)(mm) 176x135x190 176x135x250 External Size(W×D×H)(mm) 155x120x180 130x120x200
Shipping Size(W×D×H)(mm) 330x300x300 330x300x300 Shipping Size(W×D×H)(mm) 270x230x200 270x230x220
N.W./G.W. (kg) 2.5/4 3.3/5 N.W./G.W. (kg) 1/1.5 1.3/1.8

51 52
/ /

Infrared Sterilizer Hot Air Sterilizer

Autoclave
Application
Infrared Sterilizer

Hot Air Sterilizer


Dedicated to the rapid sterilization and disinfection of small metal instruments with
high temperature resistance. Such as inoculation loops, inoculation needles,
tweezers, small scissors, etc

Advantage
●Automatic parameter adjustment: Control parameters are automatically adjusted based on ambient temperature and load for optimal
heating and reduced energy consumption.

●High efficiency compressor ensures stable temperature control with low power consumption.

●Large-screen LCD display provides real-time parameter monitoring, with a self-locking function for easy operation.

●Equipped with a high-quality fan and large impeller, the incubator achieves higher ventilation efficiency and improved temperature
uniformity.

●Back suction and bottom vertical air supply system ensures effective circulation within the chamber.
Advantage
●The shelf can be freely adjusted up and down, with a minimum spacing of 30mm, allowing for flexible sample placement.
1.No flammable gas, no naked flame.
●A built-in observation window allows for clear visibility of objects inside the chamber.
2.Infrared Tungsten Halogen Lamp Sterilization,Sterilization temperature reach to 900-1300 in a few seconds, instantaneous heating,
low energy consumption. ●Door handle incorporates an anti-scalding design for easy and safe operation.

3.Infrared sensor control, automatically triggers the start,it can be operated with one hand. ●Sterilizer is equipped with a leakage protection device and a temperature control sound and light alarm system for operator and
equipment safety.
4.Automatic termination after sterilization complete.
Specification
5.No preheating required, low cost of use.
Model ST-H25A ST-H50A ST-H70A ST-H100A ST-H200A
6.Don’t consume oxygen when working, it can be used for microbial experiments in anaerobic chambers. Capacity 20L 50L 70L 100L 200L
Circulation Forced air convection
Specification Range RT+10-250
Fluctuation ±0.5
Model IR-950S Temperature Resolution 0.1
External Size (W*D*H) 160x140x220mm Uniformity 0.25
Sensor PT100
Heat source Infrared-Tungsten Halogen Lamp
Controller PID
Heat temperature 900~1300 Display LCD
Heat zone material Quartz glass Timer(min) 1-9999
Heating zone size Φ15x100mm Material Internal Stainless steel
External Electro-galvanized steel with antimicrobial powder coating
Automatic sterilization shutdown time 7s
Shelves 2
Infrared sensor Yes Test Hole (mm) Optional Ø25, 50, 80
Input voltage 220~240V, 50/60Hz,110V,60Hz Safety Device Leakage protector / Over- temperature alarm
Power 600W Electricity Power Supply 220V/50Hz
Consumption 500W 750W 1050W 1500W 2000W
Protection class IP20
Size(WxDxH) Internal (mm) 300x300x270 420x350x350 400x375x470 550x350x520 600x550x570
Permissible ambient temperature 5~40 , 80%RH External (mm) 590x470x440 710x520x520 540x550x870 840x540x730 890x730x770
Net Weight 2.7kg Shipping Size (WxDxH) (mm) 710x600x550 810x640x600 670x660x1040 970x740x880 1020x890x930
N.W./G.W. (kg) 30/45 42/55 55/75 70/86 86/100
Gross Weight 3.2kg
Optional 1.Independent temperature limiting controller
Shipping Size (W*D*H) 175x165x275mm
2.Intelligent LCD procedure temperature controller(Optional USB,RS485/232 and printer)

53 54
/ /

Cassette Sterilizer Ethylene Oxide Sterilizer

Description Description

Ethylene Oxide Sterilizer


Cassette sterilizers are commonly used in various medical and Ethylene oxide sterilizer is mainly through its non-specific
Cassette Sterilizer

laboratory settings for the sterilization of medical instruments, alkylation with genetic materials such as protein and DNA /
equipment, and materials. They are designed to provide RNA in microorganisms, resulting in the denaturation of
efficient and reliable sterilization by utilizing heat and pressure proteins and genetic materials, and finally the death of microor-
to eliminate bacteria, viruses, and other microorganisms. These ganisms due to blocked metabolism.
sterilizers are widely employed in hospitals, dental clinics,
research laboratories, and other healthcare facilities to ensure
the safety and sterility of medical tools and equipment before Application
their use in patient care or research procedures. Rigid endoscope, hose endoscope, ophthalmic lens,fifiber
optic cable, surgical instruments, metal, glass, silica gel,
Model
rubber, traditional Chinese medicine, western medicine,human
plants, DNA, etc.

Advantage

●The sterilization effect is reliable and rapid, with a non-wrapped sterilization time of 6 minutes, making it suitable for small instruments
in departments such as stomatology and ophthalmology. Model
Advantage
●Easy operation with a display screen showing the system operation. Sterilization procedures can be safely interrupted at any time.
●Humidification: Utilizes steam degradation analysis method with built-in steam generator for comprehensive and uniform humidifica-
●Self-detection system automatically stops and displays an error code on the screen if a problem occurs during operation. tion and heating while minimizing water mist.

●Suitable for sterilizing instruments with cavities and solid instruments, offering the option for non-wrapped sterilization. ●3.5" LCD screen: Equipped with a filter for stability, prolonging the service life and accuracy of the control screen.

●Power-off memory function: Preserves sterilization data and automatically resumes unfinished work procedures in case of accidents
or power loss.
Specification
●Safety protection function: Includes alarm protection for abnormal temperature, ultra-high temperature, abnormal pressure, low
vacuum rate, residual faults, phase sequence errors, and power shortage in the sterilization chamber to prevent further issues.
Model ST-2000C ST-5000C ST-6000C
Sterilization Capacity About 1.8 L About 5.2 L About 5.2 L
Tank Capacity 3.4 L 3.4 L 3.4 L
Maximum Operating Pressure 242 kPa Ethylene Oxide Consumable
Operating Pressure Range 42 kPa ~ 212kPa
Maximum Operating Temperature 138° C Steam EO sterilization pouch roll,Chemistry indicator strip,Biological indicator,Chemistry indicator tape, Chemical indicator label,EO
Operating Temperature Range 115 ° C ~ 135 ° C sterilization gas, Medical non-woven fabric,Sealing machine,EO sterilization gas
Display accuracy 0.1
Ultra-pure water, deionized Ultra-pure water, deionized Ultra-pure water, deionized
Specification
Water water, water,laboratory water, distilled water,laboratory water, distilled
laboratory water. water,purified water. water,purified water. Model ST-EO40
Atmospheric Pressure Range 70.0 kPa to 106.0 kPa Capacity 23L
Ambient Temperature Range + 5 ° C ~ + 40 ° C Product form Desktop
Steam Generator Power 1.2 kVA ± 5% Heating method Electric heating
Fuse F8AL250V phi 5 x 20 mm Working temperature 50 ±3
Safety Valve Setting Pressure 0.25 MPa Selected gas Pure ethylene oxide gas
Equipment Working Medium Water Vapor Sterilization time 0-99 hours adjustable
Working Relative Humidity ≤85% Temperature display 3.5 inch LCD display
Normal Use Place Work room Maximum anti-stress value -140kpa
Run Mode Run Intermittently (≤6 cycles/hour) Power Supply AC220V, 50Hz
Power Supply AC 220V/50 Hz Consumption 3.3KW
Consumption ≤1.3 kVA Size(WxDxH) Internal (mm) (DiameterxDepth): Ø246x450
External Size(W×D×H)(mm) 570x415x170 580x460x190 580x460x190 External (mm) 730x510x410
Shipping Size (W×D×H)(mm) 700x580x580 700x580x580 700x580x580 Shipping Size (WxDxH) (mm) 900x700x600
N.W./G.W. (kg) 35/50 45/60 45/60 N.W./G.W. (kg) 56/76

55 56
/ /

Cassette Sterilizer(three-layer)
Cassette Sterilizer(three-layer)

Cassette Sterilizer(three-layer)
Advantage

The sterilization effect is reliable and rapid, positive pressure pulsating exhaust sterilization, and the entire sterilization process only
takes 7-10 minutes. It is suitable for multi department instrument continuous surgery and emergency sterilization.

7-inch touch interface, graphical operation. The system operation status is displayed on the screen, and the sterilization program can
be safely interrupted at any time.

Compact design, saving space. Each layer is independently controlled.

Multi level permission management ensures the authenticity and effectiveness of sterilization data. Network communication interface,
which can be connected to the traceability system.

Equipped with a built-in self-priming circulating water tank and automatic water inlet function, the program operation is not affected by
the volume of the water tank, and it is equipped with a low water level detector to prevent the evaporator from dry burning.

Built in thermal printer, real-time recording of sterilization process parameters, with the function of recording and storing the last 60
ST-2000TC ST-5000TC ST-8000TC running processes.

Equipped with USB interface for data storage.

Self inspection system. Once a problem occurs during operation, the system automatically terminates operation and displays a fault
code on the screen.

Customize the settings program to meet the sterilization requirements of various large and special instruments.

Specification

Model ST-2000TC ST-5000TC ST-6000TC


External Size(WxDxH)mm 280x600×660
Cassette External Size(WxDxH)mm 196x390×40 196x480×80 196x×196×80
ST-2000TC and Cassette ST-5000TC and Cassette ST-8000TC and Cassette
Cassette Internal Size(WxDxH)mm 180x280×38 180x380×78 180x480×78
Cassette Capacity 1.8Lx3 5.2Lx3 6.0Lx3
Rated Working Temperature 138
Rated Working Pressure 242kPa
Working Temperature Range 115 ~135
Water Tank Capacity 2.5L

Application Power Supply AC 220V ± 22V; 50Hz ± 1Hz;


Working Medium water vapor
ICassette sterilizers are commonly used in various medical and laboratory settings for the sterilization of medical instruments,
Water source supply: ultrapure water, deionized water, laboratory water, distilled water, purified water
equipment, and materials. They are designed to provide efficient and reliable sterilization by utilizing heat and pressure to eliminate
Working mode: Intermittent operation (≤6 cycles/hour)
bacteria, viruses, and other microorganisms. These sterilizers are widely employed in hospitals, dental clinics, research laboratories,
and other healthcare facilities to ensure the safety and sterility of medical tools and equipment before their use in patient care or Atmospheric Pressure Range 70.0kPa~106.0kPa
research procedures. Environmental Temperature Range +5 to+40
Working Relative Humidity ≤ 85%
The three-layer Cassette sterilizers has three times the sterilization capacity, saving space, and each layer is independently
controlled. External Size(WxDxH)mm 280x600×660
Package Size(WxDxH)mm 460x710x810
Net Weight(kg) 60 84 86
Gross Weight 80 104 106

57 58
/ /

Ethylene Oxide Sterilizer


Ethylene Oxide Sterilizer

Ethylene Oxide Sterilizer


Advantage

1. Fully automatic control system, one click start to end, complete the entire process automatically

2. The display screen displays temperature, humidity, pressure, disinfection time, residual gas treatment time, and frequency

3. Built in sensor, automatically detects temperature, humidity, and pressure inside the cavity

4. The 5052 aluminum material sealing door has a pressure resistance of 200Kpa, and there is no leakage of ethylene oxide gas.

5. Residual gas treatment device filter for particle diameter ≥ 0.2 μm The filtration rate of m is not less than 99.5%

Specification

Model ST-EO80 ST-EO120 ST-EO220 ST-EO330 ST-EO120I ST-EO220I


Capacity 67 L 103 L 196 L 294L 126L 192L
Sterilizing Temp. 36-55
Product form Manual door vertical
Heating method Electric heating
Display 5’’ LCD Touch Screen 7’’ LCD Touch Screen
Controller PLC
Sterilizing Time Adjustable 0-99 hour
Sterilizer Pressure -60kpa

ST-EO80/120/220/330 ST-EO120I/220I Sterilization Method Low temperature, 100% Pure EO gas sterilization
Safety Protection Temperature and Pressure protection
Tray 2 stainless net steel
Printer Built-in Thermal mini-printer
USB port Yes
Application Internal Material SS304 Stainless Steel 5052 Aluminium, has better thermal

Materials that can be sterilized include hysteroscopes, laparoscopes, laryngoscopes, probes, rigid endoscopes, flexible endoscopes, conductivity than stainless steel
ophthalmic lenses, fiber optic intracranial sensors, cryotherapy probes, prostatectomy devices, esophageal dilators, electrocautery External Material Carbon steel
equipment, defibrillation motors, laser heads, metals, glass, silicone, rubber, traditional Chinese medicine, Western medicine, human Vacuum system High-performance, low-noise, oil-free vacuum pump
implants, etc
Humidity system Use saturated steam to humidify the sterilization chamber, and the air humidity is more uniform
Timing system EO gas diffusion time and residual EO aeration timing settable
Description Pre-set Standard Sterilization Mode and Residual Washing

Ethylene oxide sterilization is currently the most effective gas sterilization method known. It has the advantages of broad-spectrum Power Supply 220V, 50Hz
sterilization, strong penetration, no damage to items, no pollution to the environment, reliable chemical and biological testing methods, Consumption 2.6KW 3.0KW 3.3KW 3.6KW 3.0KW 3.6KW
and easy storage of sterilized items. It is an essential low-temperature sterilization equipment for hospitals. Internal Size 585*275*422mm 585*375*472mm 705*505*552mm 760*545*712mm 700*450*400mm 820*510*460mm
External Size 795*615*1140mm 795*615*1140mm 968*800*1240mm 968*800*1400mm 976*790*1715mm 976*790*1715mm
Net Weight 124kg 147kg 207kg 227kg 227kg 247kg
Shipping Size (W×D×H) 970*770*1310mm 940*750*1310mm 1120*920*1400mm 1020*930*1530mm 1130*920*186 0mm 1130*920*1860 mm

Gross Weight 150kg 177kg 237kg 277kg 274kg 297kg

59 60
/ /

H2O2 Plasma Sterilizer


Hydrogen Peroxide low temperature Plasma Sterilizer

Description ●Upper and lower layers of aluminum alloy punching plates for loading sterilized items.
Low temperature hydrogen peroxide plasma sterilizers are used for sterilizing heat-sensitive medical devices, dental equipment, ●Electric lifting door with infrared sensor switch and door anti-pinch function.

H2O2 Plasma Sterilizer


H2O2 Plasma Sterilizer

laboratory equipment, pharmaceuticals, veterinary instruments, and beauty industry tools. They provide safe and effective sterilization
while preserving delicate materials. ●Preheating heating time of ≤30 minutes.

●Electric lifting door with infrared sensor switch and door anti-pinch function.

●Preheating heating time of ≤30 minutes.

●Full automatic control system with LCD touch screen display and PLC circuit design.

●Micro-printer for printing and saving data during sterilization.

●Siemens smart series PLC control system.

●7-inch color touch screen display with various status information.

●Storage function for recording sterilization data.

●Dedicated constant temperature incubator for biological indicator monitoring.

●Built-in pressure and temperature sensors for precise control.

●Two program options for circulation and rapid circulation.

●Network connection port available.

●Vacuum system with fast vacuum rate and long service life.

●Oil mist filter and air filter for exhaust filtration.

●Transistor control power supply for stable power output.

●Cartridge filling method for hydrogen peroxide.


Technical Advantage
●Explosion-proof flame retardant heating system.
●Low temperature sterilization at 50 ±5 , preserving delicate items without damage. ●Over-voltage and over-temperature protection functions.
●Green and pollution-free process using hydrogen peroxide solution, no toxic substances or pollution.

●Stable sterilization with balanced temperature and stable hydrogen peroxide concentration. Specification
Model
●Efficient design with rectangular sterilization chamber for increased loading capacity. Model ST-HT100P ST-HT130P ST-HT220P
Capacity 108L 135L 192L
●Safe and stable operation with electric lift door and safety interlock device. Door type Automatic door vertical
●Simple operation with intelligent control processor and one-key start. Heating method Electric heating
Working temperature 40-60
●Configured with high-quality components for reliable performance. Select sterilizer 58%-60% hydrogen peroxide (H202)
Sterilization time 35-55 minutes
●Time and cost-saving with short sterilization cycle and low operating costs. Total Cycle Time Short cycle 30 minutes, long cycle 50 minutes, lumen cycle 60 minutes
Installation environment 10-40
●Equipped with sterile filter for air inlet to prevent secondary pollution.
Temperature display LCD touch screen
●U disk storage function for storing and recording sterilization data. Maximum anti-stress value 10pa
Operation height ≤160cm
Model Advantage Power Supply AC220V, 50Hz
Consumption 3.5kW 4.2kW 4.8kW
●High-quality 5052 aluminum material for sterilization room, with excellent thermal conductivity. Size(W×D×H) Internal (mm) 700x420x360 750x440x400 820x510x 460
External (mm) 1120x790x1770
●ABS plastic disposable outer box material.
Shipping Size (W×D×H) (mm) 1300x950x1950 1300x950x1950 1300x950x1950
●Aluminum alloy electrode mesh and sealing door material. N.W./G.W. (kg) 265/320 272/325 305/360

61 62
/ /

H2O2 Plasma Sterilizer


Hydrogen Peroxide low temperature Plasma Sterilizer

●Electric lifting door with infrared sensor switch and door anti-pinch function.
Hydrogen Peroxide Low Temperature Plasma Sterilizer

Hydrogen Peroxide Low Temperature Plasma Sterilizer


●Preheating heating time of ≤30 minutes.

●Full automatic control system with LCD touch screen display and PLC circuit design.

●Micro-printer for printing and saving data during sterilization.

●Siemens smart series PLC control system.

●7-inch color touch screen display with various status information.

●Storage function for recording sterilization data.

●Dedicated constant temperature incubator for biological indicator monitoring.

●Built-in pressure and temperature sensors for precise control.

ST-HT60P ST-HT100PI/140P/190P ●Two program options for circulation and rapid circulation.
Application ●Network connection port available.
Low temperature hydrogen peroxide plasma sterilizers are used for sterilizing heat-sensitive medical devices, dental equipment, ●Vacuum system with fast vacuum rate and long service life.
laboratory equipment, pharmaceuticals, veterinary instruments, and beauty industry tools. They provide safe and effective sterilization ●Oil mist filter and air filter for exhaust filtration.
while preserving delicate materials. ●Transistor control power supply for stable power output.

●Cartridge filling method for hydrogen peroxide.


Advantage
●Explosion-proof flame retardant heating system.
●Low temperature sterilization at 50 ±5 , preserving delicate items without damage.
●Over-voltage and over-temperature protection functions
●Green and pollution-free process using hydrogen peroxide solution, no toxic substances or pollution.

●Stable sterilization with balanced temperature and stable hydrogen peroxide concentration.
Specification
●Efficient design with rectangular sterilization chamber for increased loading capacity.
Model ST-HT60P ST-HT100PI ST-HT140P ST-HT190P
●Safe and stable operation with electric lift door and safety interlock device.
Chamber Volume 60 L 100L 135L 190L
●Simple operation with intelligent control processor and one-key start.
Construction 5052 Aluminium
●Configured with high-quality components for reliable performance. Processing Temperature 45-55°C
Control System Siemens PLC
●Time and cost-saving with short sterilization cycle and low operating costs.
Display 7 inch Touch Screen
●Equipped with sterile filter for air inlet to prevent secondary pollution. Printer Thermal Printer

●U disk storage function for storing and recording sterilization data. Touch Screen Functions Phase Temperature Date and time Cycle Selected Alert message
Data Storage Memory Card
Description
Total Cycle Time Short cycle 30 minutes, long cycle 50 minutes, lumen cycle 60 minutes
●High-quality 5052 aluminum material for sterilization room, with excellent thermal conductivity. Noise < 60 db
Door Manual Automatic
●ABS plastic disposable outer box material.
Alarm Door Lock Dosing failure Low / High Temp. range Vacuum & Pressure
●Aluminum alloy electrode mesh and sealing door material. Incubator No Biological Incubator

●Upper and lower layers of aluminum alloy punching plates for loading sterilized items. Consumption 2.6KW 3.4KW 3.5KW 4.3KW
Power Supply 220V, 50Hz
●Electric lifting door with infrared sensor switch and door anti-pinch function.
Shipping Size (W×D×H) 890*750*1660mm 1270*920*1940mm 1270*920*1940mm
●Preheating heating time of ≤30 minutes. Gross Weight 230kg 350kg

63 64
/ /

Low-temperature Steam Formaldehyde Sterilizer

9.Printing function: capable of printing program name, sterilization date, sterilization pot number, sterilization start and end time, as
Low-temperature Steam Formaldehyde Sterilizer

Low-temperature Steam Formaldehyde Sterilizer


well as pressure, temperature, stage time and other information of each stage of the sterilization process. Printing records should be
kept for more than 5 years.

10.Storage function: Store sterilization data for no less than 15000 cycles, including program name, sterilization date, sterilization
cycles, sterilization start and end time, as well as pressure, temperature, stage time, and end status of each stage of the sterilization
process.

11.Alarm system: There is a high temperature alarm system.

12.Network connection: Equipped with a network connection port.

13.Vacuum pump: Adopting a German water ring vacuum pump with reliable quality, it not only meets the design pressure require-
ments of the sterilizer, but also accelerates the decomposition and absorption of formaldehyde vapor to a certain extent, with a higher
safety factor.

14.Air filter: filtration accuracy ≤ 0.22 μ M;

15.Pressure sensor: The built-in pressure sensor uses high-precision sensors to automatically detect the pressure inside the cavity,
ensuring that the accuracy of the pressure inside the cavity is controlled within an effective range.

16.Temperature sensor: embedded temperature sensor automatically detects the temperature inside the cavity, ensuring that the
Application accuracy of the temperature inside the cavity is controlled within an effective range.

Mainly used for low-temperature sterilization of heat and moisture resistant items, it can be used in operating rooms, supply rooms, 17.Multiple protection functions: equipped with multiple protection functions such as overvoltage, undervoltage, overtemperature, door
endoscopy rooms, and other treatment rooms in medical institutions. failure, and door safety interlock device

Description Specification

The low-temperature steam formaldehyde sterilizer uses a vacuum system to create a low-pressure environment in the sterilization Model ST-FT100 ST-FT130 ST-FT150
room. In the low-pressure environment, the formaldehyde solution is vaporized to form low-temperature formaldehyde vapor. The
Capacity 100L 130L 150L
formaldehyde vapor is used to act on the items in the sterilization room, causing microbial proteins and genetic substances on the
items to denature and cause microbial death, in order to achieve the purpose of sterilization. Chamber Size(W*D*H) 700*430*360mm 750*450*400mm 820*645*400mm
Consumption 5.3KW
Sterilizer Solution 2% formaldehyde, 3% alcohol, and 95% distilled water solution
Advantage
Sterilizer Solution Packaging Specification 2L
1.Loading method: A movable stainless steel basket with support rods to avoid stacking items. Single cycle dosage 0.6-1.3Kg
Operating temperature 60 and 76
2.Control system: Adopting Siemens PLC control system and friendly human-machine interaction interface, safe and reliable; One click
start to finish, the entire process is automatically completed, displaying information such as temperature, pressure, sterilization time, Sterilization time 2-3.5 hours
etc. The system can be maintained and upgraded. Water consumption 3.95Kg/min
Heating Method Electric heating
3.Heating system: PID temperature controller is used to control the door, inner tank, and steam generator respectively and ensure their
working stability. Anti-pressure 5pa
Operation Height ≤160cm
4.Display function: It can display temperature, pressure, time, operating mode, and alarm information.
Chamber Material 5052 corrosion resistant aluminum alloy, with antioxidant treated
5.Power outage memory function: reduces operating costs and risk of misoperation. Cabinet material ABS
Door panel material 5052 corrosion resistant aluminum alloy, with antioxidant treated, ≥10mm
6.Steam generator: A self-developed fast evaporator that not only improves the vaporization rate of formaldehyde solution, but also
enhances the sterilization effect. Pipe materials SS304 stainless steel
Insulation materials High quality insulation cotton
7.Cleaning and drying function: With the independent addition of automatic cleaning and drying function, it greatly reduces the formal-
Door form Automatic door
dehyde residue in the processed items while ensuring the basic requirement of not getting wet, ensuring that it is below the national
standard value. Display 8 inch Touch Screen
External Size(W×D×H) 1070*790*1770mm
8.Automatic dosing function: Unique formaldehyde solution soft bag filling technology, automatic puncture filling, to avoid manual risks.
Shipping Size (W×D×H) 1200*900*1980mm

65 66
/ /

Horizontal Pulse Vacuum sterilizer


Mechanical door
Horizontal Pulse Vacuum sterilizer

Horizontal Pulse Vacuum sterilizer


Advantage

1. Translation door, with anti-pinch hand function.

2.Double-door channel type, with pressure safety interlock, liquid program with temperature door safety interlock.

Internal Structure Internal Structure 3.Automatic sealing ring, sealing effect is good, the service life is up to 1500 cycles.

4.The main structure is European ring reinforcement jacket structure, programmable controller PLC, color touch screen.

5.With aseptic clothing, utensils, rubber, liquid, medium, waste, custom procedures such as sterilization procedures and BD test,
vacuum leak test and other test procedures.

6.Built-in micro needle printer, equipped with paperless recorder.

7.Intelligent maintenance system: optional remote monitoring and maintenance module; Water saving noise reduction system is
Internal Structure Pressure gauge Pump optional.

Description 8.Pipeline system: sanitary pipeline, clamp connection, automatic pipe welding machine welding, direct-connected water ring vacuum
pump.
ST-BVA series pulsating vacuum sterilizer uses saturated steam for sterilization. The steam thoroughly penetrates into the surface of
articles by means of pulsating vacuum. The high pressure and high temperature saturated steam can effectively kill all microorgan- 9. Automatic drainage device, optional jacket automatic control drainage device, the inner chamber are installed automatic control
isms and their spores, which is the most effective method in physical sterilization. The effect of cold air on temperature is eliminated drainage device, automatic detection jacket and inner chamber temperature pressure automatic drainage, to ensure the saturation of
by pulsating vacuum exhaust, and the articles are sterilized by vacuum dehumidification and jacket drying. The sterilizer is suitable steam.
for pharmaceutical biological engineering, medical health, laboratory animals and other fields, and can be used for sterilizing utensils, 10. With drainage temperature automatic adjustment system, drainage temperature ≤55°C.
aseptic clothes, rubber plugs, aluminum caps, raw materials, filters, media and so on.
11.Air filter: the filtration accuracy is <0.22 microns, the bactericidal rate of 99.97%.

Specification

Model ST-BVA250PV ST-BVA360PV ST-BVA600PV ST-BVA800PV ST-BVA1000PV ST-BVA1200PV ST-BVA1500PV ST-BVA2000PV ST-BVA2500PV ST-BVA3000PV ST-BVA4000PV ST-BVA6000PV
Capacity(L) 255 360 600 800 1000 1200 1500 2000 2500 3000 4000 6000
Interior design pressure -0.1~03MPa
Jacket design pressure 0.3MPa
Design temperature 150
Max working temperature 139
Max working pressure 0.25MPa
Vacuum leakage rate ≤0.13KPa/min
Final vacuum -96KPa
Inner chamber Opening pressure 0.28MPa
Jacket safety valve opening pressure 0.28MPa
Sterilization temperature control precision 0-2
Temperature uniformity ±0.5
Positive pressure pulsation 1~3
Inner tank size (WxHxD)(mm) 550x550x850 610x610x1000 610x910x1200 610x910x1500 610x910x1800 680x1180x1500 680x1180x1850 900x1380x1620 900x1380x2050 900x1380x2400 900x1380x3200 900x1380x4800
Overall dimension (LxWxH)(mm) 1150x1280x1800 1300x1350x1850 1500x1440x1900 1800x1440x1900 2100x1440x1900 1800x1480x2000 2100x1480x2000 1970x1750x2200 2350x1750x2200 2700x1750x2200 3500x1750x2200 5400x1750x2200
Weight(kg) 750 850 1250 1350 1550 1650 1850 2300 2700 3500 3500 6200
Gas consumption(Kg/c) 18 22 35 47 55 65 80 106 132 132 132 132
Water consumption(Kg/c) 120 180 320 400 500 600 750 1000 1250 1250 1250 1250
Power Supply 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ

67 68
/ /

Horizontal Pulse Vacuum sterilizer


Overhead door
Horizontal Pulse Vacuum sterilizer

Horizontal Pulse Vacuum sterilizer


Advantage

1. Translation door, with anti-pinch hand function.

2.Double-door channel type, with pressure safety interlock, liquid program with temperature door safety interlock.

Internal Structure Internal Structure 3.Automatic sealing ring, sealing effect is good, the service life is up to 1500 cycles.

4.The main structure is European ring reinforcement jacket structure, programmable controller PLC, color touch screen.

5.With aseptic clothing, utensils, rubber, liquid, medium, waste, custom procedures such as sterilization procedures and BD test,
vacuum leak test and other test procedures.

6.Built-in micro needle printer, equipped with paperless recorder.

7.Intelligent maintenance system: optional remote monitoring and maintenance module; Water saving noise reduction system is
Internal Structure Pressure gauge Pump optional.

Description 8.Pipeline system: sanitary pipeline, clamp connection, automatic pipe welding machine welding, direct-connected water ring vacuum
pump.
ST-BVB series pulsating vacuum sterilizer uses saturated steam as the medium, and completely kills microorganisms and spores on
the surface of articles by pulsating vacuum. This is the most effective method of physical sterilization, and the sterilization effect is 9.Automatic drainage device, optional jacket automatic control drainage device, the inner chamber are installed automatic control
reliable. The effect of cold air on temperature is eliminated by pulsating vacuum exhaust, and the articles are sterilized by vacuum drainage device, automatic detection jacket and inner chamber temperature pressure automatic drainage, to ensure the saturation of
dehumidification and jacket drying. It can be widely used in pharmaceutical biological engineering, medical health, laboratory animals steam.
and other fields, and is suitable for the sterilization of utensils, aseptic clothes, rubber plugs, aluminum caps, raw materials, filters,
10.With drainage temperature automatic adjustment system, drainage temperature ≤55°C.
media and other items with high sterilization requirements.
11.Air filter: the filtration accuracy is <0.22 microns, the bactericidal rate of 99.97%.

Specification

Model ST-BVB150PV ST-BVB250PV ST-BVB300PV ST-BVB360PV ST-BVB450PV ST-BVB600PV ST-BVB800PV


Capacity (L) 150 250 300 360 450 600 800
Interior design pressure -0.1~03MPa
Jacket design pressure 0.3MPa
Design temperature 150
Max working temperature 139
Max working pressure 0.25MPa
Vacuum leakage rate ≤0.13KPa/min
Final vacuum -96KPa
Inner chamber Opening pressure 0.28MPa
Jacket safety valve opening pressure 0.28MPa
Sterilization temperature control precision 0-2
Temperature uniformity ±0.5
Positive pressure pulsation 1~3
Inner tank size (WxHxD)(mm) 450x450x750 550x550x850 550x550x1000 610x610x1000 650x700x1000 650x700x1320 650x700x1750
Overall dimension (LxWxH)(mm) 1050x1200x1750 1150x1300x1800 1300x1300x1800 1300x1350x1900 1300x1360x1950 1620x1360x1950 2050x1360x1950
Weight(kg) 550 850 850 950 1050 1150 1350
Gas consumption(Kg/c) 12 18 18 22 30 36 45
Power Supply 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ
Consumption 2KW+(12KW) 2KW+(24KW) 2KW+(24KW) 2KW+(30KW) 2KW+(30KW) 2KW+(48KW) 2KW+(48KW)

69 70
/ /

Horizontal Pulse Vacuum sterilizer


Sliding door

Advantage
Horizontal Pulse Vacuum sterilizer

Horizontal Pulse Vacuum sterilizer


1.Translation door, with anti-pinch hand function.

2.Double-door channel type, with pressure safety interlock, liquid program with temperature door safety interlock.

3. Automatic sealing ring, sealing effect is good, the service life is up to 1500 cycles.

Internal Structure Internal Structure 4. The main structure is European ring reinforcement jacket structure, programmable controller PLC, color touch screen.

5.With aseptic clothing, utensils, rubber, liquid, medium, waste, custom procedures such as sterilization procedures and BD test,
vacuum leak test and other test procedures

6. Built-in micro needle printer, equipped with paperless recorder.

7.Intelligent maintenance system: optional remote monitoring and maintenance module; Water saving noise reduction system is
optional
Internal Structure Pressure gauge Pump 8.Pipeline system: sanitary pipeline, clamp connection, automatic pipe welding machine welding, direct-connected water ring vacuum
pump.
Description 9.Automatic drainage device, optional jacket automatic control drainage device, the inner chamber are installed automatic control
ST-BVH series pulsating vacuum sterilizer uses saturated steam as the medium, through pulsating vacuum, efficient killing of microor- drainage device, automatic detection jacket and inner chamber temperature pressure automatic drainage, to ensure the saturation of
ganisms and their spores, is the most effective physical sterilization method. Pulsating vacuum exhaust is adopted to eliminate the steam.
influence of cold air on temperature, and finally sterilization is achieved through vacuum dehumidification combined with jacket drying. 10.With drainage temperature automatic adjustment system, drainage temperature ≤55°C.
Suitable for pharmaceutical biological engineering, medical health, laboratory animals and other fields, the high requirements of
sterilization of equipment, aseptic clothing, rubber stopper, aluminum cap, raw materials, filters, media and other items. 11.Air filter: the filtration accuracy is <0.22 microns, the bactericidal rate of 99.97%.

Specification

Model ST-BVH1200PV ST-BVH1500PV ST-BVH2000PV ST-BVH2500PV ST-BVH3000PV ST-BVH6000PV ST-BVH6700PV ST-BVH13600PV


Capacity(L) 1200 1500 2000 2500 3000 6000 6700 13600
Interior design pressure -0.1~03MPa
Jacket design pressure 0.3MPa
Design temperature 150
Max working temperature 139
Max working pressure 0.25MPa
Vacuum leakage rate ≤0.13KPa/min
Final vacuum -96KPa
Inner chamber Opening pressure 0.28MPa
Jacket safety valve opening pressure 0.28MPa
Sterilization temperature control precision 0-2
Temperature uniformity ±0.5
Positive pressure pulsation 1~3
Inner tank size(WxHxD)(mm) 680x1180x1500 680x1180x1800 900x1380x1620 900x1380x2020 900x1380x2400 900x1380x4800 1250x1380x3910 1250x1380x7850
Overall dimension (LxWxH)(mm) 2007x2710x2173 2357x2110x2713 1950x2750x2200 2300x2750x2200 2700x2750x2200 5400x2750x2200 5400x3200x2200 8500x3200x2200
Weight(kg) 2300 2450 3000 3600 3800 6600 7000 11000
Gas consumption(Kg/c) 65 80 100 120 150 320 350 620
Water consumption(Kg/c) 600 750 900 1100 1300 2400 2600 4800
Power Supply 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ 380V,50HZ
Consumption 3.5kw+(73kw) 3.5kw+(90kw) 4.5kw 6kw 8kw 12kw 12kw 24kw

71 72
/ /

Ethylene oxide sterilizer H2O2 serious Hydrogen peroxide steam sterillzer(Customized type)

H2O2 serious Hydrogen peroxide steam sterillzer


Ethylene oxide sterilizer

Application

It is mainly used for various types of endoscopy, cotton, silicone, rubber, catheter wires and various medical consumables, and will
not corrode or damage the equipment.

Items that cannot be sterilized with ethylene oxide: liquids, aluminum-plastic packaging.

Description

The ST-EO series ethylene oxide sterilizer has the characteristics of sophisticated design and exquisite workmanship. It is mainly
composed of a microcomputer control system, a residual gas treatment device, a sterilization studio and other parts. The principle is
Application
to rely on the ethylene oxide gas to circulate within a certain range. The sterilization effect is achieved under humidity pressure.
Ethylene oxide gas directly destroys microbial proteins, prevents their metabolism, and can kill stubborn Bacillus. Hydrogen peroxide vapor sterilizer (VHP) takes advantage of the fact that the gas state of hydrogen peroxide at room temperature is
more capable of killing spores than the liquid state. It generates free hydroxyl groups and is used to attack cellular components,
including lipids, Protein and DNA tissue, etc., meet the requirements of complete sterilization. It is a specially designed and manufac-
tured special equipment that can be used for sterilization of closed spaces such as isolation type and isolator. It has broad-spectrum
Advantage bactericidal effect and is suitable for sterilizing fungi, bacteria, viruses and spores. Compared with traditional sterilization technology,
1. Ethylene oxide sterilizer is the most effective gas sterilization method currently known. It does no damage to items and the environ- it is pollution-free, has better material compatibility, and is easy to use and verify. It can be widely used in pharmaceutical factories,
ment is not polluted. laboratories and scientific research institutes.

2 Ethylene oxide is a colorless liquid at low temperatures with strong penetrating power. It can sterilize items at low temperatures with
good sterilization effect.
Description
3 The rectangular sterilization chamber design makes it easier to load more instruments at one time and increases the use space of
the sterilization chamber. The hydrogen peroxide steam sterilizer is equipped with a vacuum system and a hydrogen peroxide steam generator. A specially
configured hydrogen peroxide solution is injected into the evaporator, and the hydrogen peroxide is injected into the sterilization
4 The sterilization chamber uses electric heating technology to make the temperature of the sterilization chamber more uniform and the
chamber under low temperature and vacuum conditions. The steam is maintained for a corresponding period of time to allow the
sterilization effect is stable and reliable.
active substances to chemically react with the proteins and nucleic acids in the microorganisms, destroying the microorganisms and
5 It adopts a highly intelligent control processor and can be started with one button. It is easy to operate and can be used immediately disturbing the survival functions of the microorganisms, and then uses H2O2 as the medium to act on the microbial cells to further kill
after plugging in the power supply. the microorganisms. The sterilizing agent generally uses a food-grade hydrogen peroxide solution of a certain concentration, which
has no toxic substance residues, does not require drainage or ventilation, does no harm to the operator's body, and does not pollute
6 The disk capacity is ≥8G and can store and record sterilization data and related information for ≥1 million times. the environment.

Specification

Model ST-EO3000 ST-EO6000 ST-EO10000 ST-EO20000


Chamber Size(W*D*H) 2220*960*1420 3350*1200*1500 5600*1200*1500 5900*1700*2000
External Size(W*D*H) 2400*1705*2200 3550*1900*2260 5850*1900*2260 6200*4300*3000
Consumption 14KVA 29KVA 35KVA 60KVA

73 74
/ /

Dry heat sterilizer General evaporator

Application

suitable for dry heat sterilization on the door Pharmaceutical vials, ampoules, aluminum cap, metal and glass parts Sterile pyrogen
Dry heat sterilizer

General evaporator
removing and drying sterilization of solid materials.

Description

Under the action of the circulating fan, heating tube, and dehumidifying fan, the dry heat sterilizer rapidly heats up. At the same time,
Description
clean and dry air enters the sterilization chamber through a high-efficiency filter, absorbing moisture from the surface of the materials.
A steam generator is a device that converts electrical energy into thermal energy to produce steam. It heats water in a sealed
The absorbed moisture evaporates and is discharged through the circulating channel. The dry air circulates in a directed manner
container using electric heating elements to generate high-pressure steam, which is then supplied to steam-using equipment. The
under the action of the fan. As the water vapor gradually decreases, fresh filtered air is intermittently replenished. The chamber is
evaporator consists of a main body, electrical control system, water inlet pump, electric heating elements, liquid level controller,
maintained at a slightly negative pressure, and the temperature in the sterilization chamber reaches the set value. It is kept at the set
pipeline system, and other components. It has various automatic control functions and advantages such as small size, no pollution,
temperature for insulation circulation, achieving sterilization of the sterilized items.
simple operation, and safety and reliability.

Specification Specification

Model ST-DH360 ST-DH450 ST-DH650 ST-DH800 ST-DH1000 ST-DH1200 ST-DH2000 Model ZFQ-G54 ZFQ-G60 ZFQ-G72 ZFQ-G90 ZFQ-G120
Chamber Material 304/316L Material Q245R/304
Capacity(L) 360 450 650 800 1000 1200 2000 Capacity(L) 28L 28L 42L 48L 49
Power Supply 18kVA 20kVA 21kVA 24kVA 27kVA 27kVA 36kVA Power Supply 54kVA 60kVA 72kVA 90kVA 120kVA
Sterilizer Temperature RT~300 Pressure 0.8
Noise ≤85dB Water supply method Automatic water inlet
Cleanliness Grade CLass 100 Heating method Electric heating
Cooling method Air Cooling/Water Cooling Control method Automatic control

75 76
/ /

Vertical steam sterilizer Multifunctional Self-control Internal Circulation Drying Vertical steam sterilizer Multifunctional Self-control Drying

Advantage Advantage

Vertical Pressure Steam Sterillzer


Vertical Pressure Steam Sterillzer

1. Hand wheel translation and quick opening structure 1. Hand wheel translation and quick opening structure

2. High quality stainless steel material 2. High quality stainless steel material

3. With drying function 3. Microcomputer automatic control, arbitrary setting

4. Microcomputer automatic control, arbitrary setting sterilization parameters.

sterilization parameters 4. Safety interlocking device

5. Safety interlocking device 5. Digital display and Touch buttons

6. Digital display and Touch buttons 6. Standard test interface

7. Standard test interface 7. Self-expanding seal

8. Self-expanding seal 8. With drying function

9. With automatic protection function:over temperatureprotection; 9. With automatic protection function:over temperatureprotec-
over pressure self discharge protection ;low water level protec- tion; over pressure protection; low water level protection; anti
tion; anti dry burning. dry burning.

10.With stainless steel bucket or basket 10. With stainless steel bucket or basket

Buzzer reminder after sterilization, automatic stop Automatic 11. Buzzer reminder after sterilization, automatic stop5?
discharge of cold air, automatic steamexhaust after steriliza-
tion,Internal steam cycle. 12. Automatic discharge of cold air,automatic steam exhaust
after sterilization.
*Optional external printer
*Optional external printer
*Optional automatic water
*Optional automatic water

Specification Specification

Model ST-VX35 ST-VX50 ST-VX75 ST-VX100 Model ST-VG35 ST-VG50I ST-VG75 ST-VG100I
Volume 35L 50L 75L 100L Volume 35L 50L 75L 100L
Power 3.5kW 3.5kW 4.5kW 4.5kW Power 3.5kW 3.5kW 4.5kW 4.5kW
Voltage AC 220V,50HZ Voltage AC 220V,50HZ
Design pressure 0.25MPa Design pressure 0.25MPa
Design temperature 139 Design temperature 139
Rated working pressure 0.22MPa Rated working pressure 0.22MPa
Rated working temperature 134 Rated working temperature 134
Sterilization temperature setting range 116 ~134 Sterilization temperature setting range 116 ~134
Sterilization time setting range 4~120min Sterilization time setting range 4~120min
Drying time setting range 0~240min Drying time setting range 0~240min
Chamber volume Ø350*400 Ø350*525 Ø400*625 Ø450*650 Chamber volume Ø350*400 Ø350*525 Ø400*625 Ø450*650
Bucket dimension Ø330*320 Ø330*460 Ø380*560 Ø420*540 Bucket dimension Ø330*320 Ø330*460 Ø380*560 Ø420*540
Basket dimension Ø350*400 Ø350*525 Ø400*625 Ø450*650 Basket dimension Ø320*350 Ø320*240*2 Ø360*280*2 Ø410*300*2
Outer Dimension 698*498*940 630*500*1180 725*525*1170 788*588*1255 Outer Dimension 480*480*1000 480*480*1120 525*525*1190 550*550*1250
Packing Size(L*W*H 790*610*1130 800*600*1230 840*610*1310 870*650*1380 Packing Size(L*W*H 580*630*1130 580*630*1230 630*680*1290 690*650*1410
Gross weight 125kg 130kg 145kg 156kg Gross weight 79kg 106kg 120kg 133kg
Net weight 110kg 112kg 114kg 129kg Net weight 58kg 80kg 90kg 106kg

77 78
/ /

Vertical Pressure Steam Sterillzer Digital Display Self-control

Advantage

Vertical Pressure Steam Sterillzer


Vertical Pressure Steam Sterillzer

1. Hand wheel translation and quick opening structure

2. High quality stainless steel material

3. Microcomputer automatic control, arbitrary setting sterilization


parameters.

4. Safety interlocking device

5. Digital display and Touch buttons

6. Standard test interface

7. Self-expanding seal

8. With automatic protection function:over temperature protection;


over pressure self discharge protection,low water level protection;
anti dry burning.

9. With stainless steel bucket or basket

10. Buzzer reminder after sterilization, automatic stop

11. Automatic discharge of cold air,automatic steam exhaust after


sterilization.

*Optional automatic water

Specification

Model ST-VB35 ST-VB50 ST-VB75 ST-VB100


Volume 35L 50L 75L 100L
Power 3.5kW 3.5kW 4.5kW 4.5kW
Voltage AC 220V,50HZ
Design pressure 0.25MPa
Design temperature 139
Rated working pressure 0.22MPa
Rated working temperature 134
Sterilization temperature setting range 116 ~134
Sterilization time setting range 4~120min
Drying time setting range 0~240min
Chamber volume Ø350*400 Ø350*525 Ø400*625 Ø450*650
Bucket dimension Ø330*320 Ø330*460 Ø380*560 Ø420*540
Basket dimension Ø320*350 Ø320*240*2 Ø360*280*2 Ø410*300*2
Outer Dimension 480*480*1000 480*480*1120 525*525*1190 550*550*1250
Packing Size(L*W*H 580*630*1130 580*630*1230 630*680*1290 690*650*1410
Gross weight 79kg 106kg 120kg 133kg
Net weight 58kg 80kg 90kg 106kg

79 80
ISO 9001/13485/14001/45001

Pre-vacuum pressure Steam Sterilizer Consumables

Biological indicator Chemical indicating tape


Pre-vacuum pressure Steam Sterilizer Consumables

Pre-vacuum pressure Steam Sterilizer Consumables


Advantage

The color of the indicating ink on the tape changes after disinfection and sterilization, so as to determine whether the item has been
Advantage sterilized, and intuitively and conveniently reflect the sterilization status of the item.

This product is composed of spores of Bacillus stearothermophilus (ATCC7953) and culture medium. according to Method of use
The color change of the biological indicator culture medium reflects the survival of the spores of Bacillus stearothermoris, thereby Cut off an appropriate length of indicator tape and paste it on the seal or surface of the item to be sterilized, and then place it in the
judging sterilizer. After sterilization, the twill on the indicator tape indicates the color change of the mark, which can determine that the steriliza-
tion package has been sterilized.
Check whether pressure steam sterilization is qualified.

Chemical indicator Paper-plastic wrapping

Advantage

This product is specially designed for pressure steam sterilization chemical indicator card, long structure, indicator card surface with Advantage
chemical indicator chemical change, accompanied by a color change, from light yellow to brown or black. The depth of its color is
closely related to the humidity, temperature and duration of the steam. If the indicator card color does not change or the color is lighter This product is a special sealing tape under the pressure steam sterilization method. No special chemical indicator is printed on the
than the standard color, the surface has not reached the specified sterilization temperature and time. surface of the sealing tape, which cannot indicate whether the sterilization package has been sterilized.

81 82
ISO 9001/13485/14001/45001

Pre-vacuum pressure Steam Sterilizer Consumables

BD test standard test package Method of use


Pre-vacuum pressure Steam Sterilizer Consumables

Pre-vacuum pressure Steam Sterilizer Consumables


Attach the chemical indication label to the surface of the package to be sterilized. After sterilization, remove the package and observe
the indicator label: if the color of the chemical indicator block of the indicator label is from yellow to a contrasting color or darker than
the contrasting color, it indicates that the sterilization package has been sterilized.

Medical bag

Advantage

The indication label on the outside of the test kit is also printed with steam sensitive indication ink. After the steam is extinguished, the
yellow ink turns to black, indicating that the test kit has been used. Test for testing the air exclusion effect of a pre-vacuum pressure
steam sterilizer.

Sterilization Chemical instruction card (crawling type)


Advantage

This product is completely made of safe polypropylene fiber, which is a disposable medical products final sterilization packaging
material formed by one layer of antibacterial layer and two layers of anti-stretch layer. It has the characteristics of anti-bacteria,
anti-static, anti-tension, etc., suitable for packaging medical supplies sterilization, can be used for packaging medical equipment for
pressure steam sterilization, ethylene oxide sterilization, low temperature formaldehyde or hydrogen peroxide low temperature plasma
sterilization.
Method of use
Advantage
Put the sterilized items into the roll bag, the length of the sterilized items should be suitable for the sterilization bag sealing, otherwise
This product is specially designed for pressure steam sterilization chemical indicator card, long structure, indicator card surface with the sterilization bag may burst; Abnormity and sharp objects should be placed correctly to ensure safety; Seal the mouth of the
chemical indicator chemical change, accompanied by a color change, from light yellow to brown or black. The depth of its color is sterilization bag with sealing equipment, indicating the sterilization time, use and other information; Items are properly placed in the
closely related to the humidity, temperature and duration of the steam. If the indicator card color does not change or the color is lighter sterilizer to sterilize. When using the instrument, open the sterilizing bag and remove the items in the direction indicated by stripping.
than the standard color, the surface has not reached the specified sterilization temperature and time.
Medical sterilization packaging non-woven fabric
Chemical indictor label

Advantage
Advantage It is suitable for sterilization of packaged medical devices by pressure steam, ethylene oxide, hydrogen peroxide low temperature
plasma and low temperature steam formaldehyde.
This product is a special chemical indication label for pressure steam sterilization. The special chemical indicator color block printed on
the surface of the label, under the action of moisture and heat of saturated steam, the indicator color block has a chemical reaction and Method of use
produces a color change, so as to judge whether the indicated item is pure
The sterilized items are divided into 2 layers by the closed packaging method (repeated folding can form a long curved path to prevent
Sterilized treatment. It is suitable for the monitoring of pressure steam sterilization process under saturated steam conditions of 121 microorganisms from "easily" entering the sterilization package), and then the interface can be pasted with an indication label or
/20min, 134 /4min and 132 /4min. packaging tape.

83 84
ISO 9001/13485/14001/45001

Low Temperature Steam Formaldehyde Sterilizer

Chemical indicator card Method of use


Low Temperature Steam Formaldehyde Sterilizer

Low Temperature Steam Formaldehyde Sterilizer


1. Put the biological indicator into the low-temperature steam formaldehyde sterilization bag and seal it, place the bag in the most
difficult sterilization position in the sterilizer, and run the sterilization procedure.

2. After sterilization, take out the biological indicators in the packaging bag. After confirming that the color of the chemical indicator on
the label on the biological indicator changed from red to yellow, the biological indicator was cooled to room temperature, and the
ampoule inside the biological indicator was broken with a clamp so that the bacteria pieces were fully in contact with the medium.

3. Culture in an incubator at 56 -58 . At the same time, another batch of unsterilized biological indicator with the same batch number
was broken in ampoules and cultured under the same conditions as positive control.

4. After 24 hours of culture, observe the color of the medium to confirm the sterilization effect. Remove the label from the biometric
indicator bottle and store it as a record.

Qualification criteria: After 24 hours of culture, the color of the medium remains blue and purple.

Chemical indicating tape

Advantage

This product is a special chemical indicator card for low temperature steam formaldehyde sterilization. The special chemical indicator
color block printed on the surface of the indicator card is used to conduct biochemical reactions under the conditions of 2% formalde-
hyde content, 60 temperature and 30min action time and 2% formaldehyde content, 78 temperature and 10min action time during
low temperature steam formaldehyde sterilization

Produce a color change.

Method of use

Place the chemical indicator card in the center of the package to be sterilized. After sterilization, open the package and take out the
indicator card for observation: if the color of the chemical indicator block of the indicator card reaches or is lighter than the contrast
color, it indicates that the sterilization package center has been completely sterilized; If the color of the indicated color block does not
change or the color is darker than the contrast color, it indicates that the specified sterilization conditions have not been met. The color
change reaction of the indicator block is irreversible, and the indicator card after the reaction can be stored as a record at room Advantage
temperature
This product is a special chemical indicating tape for low temperature steam formaldehyde sterilization. The special chemical indicator
Biological indicator color block printed on the surface of the indicator label can produce biochemical reactions and color changes under the conditions of
2% formaldehyde content, 60 temperature and 30min action time and 2% formaldehyde content, 78 temperature and 10min action
time during steam formaldehyde sterilization at low temperature.

Method of use

Cut off the appropriate length of chemical indicator tape and apply to the surface of the package to be sterilized. After sterilization,
remove the package and observe the indicator tape: if the chemical indicator color block of the indicator tape changes from red to
green, it indicates that the sterilized package has been sterilized; If the color of the indicated color block does not change or the
discoloration is not obvious, it indicates that the package has not been sterilized or has not met the specified sterilization conditions.
The color change reaction of the indicator block is irreversible, and the indicator tape after the reaction can be stored as a record at
room temperature.

Point of note

1. This product can not be used as a determination of sterilization effect, only indicates whether the item has been sterilized.

2. Do not contact with acidic, alkaline and oxidizing substances, moisture-proof (relative humidity is less than 60%).

3. Avoid light (ultraviolet lamps, fluorescent lamps and sunlight), and store in a dry and ventilated place. Store in light and cool place
(below 25 ).

85 86
ISO 9001/13485/14001/45001

Low Temperature Steam Formaldehyde Sterilizer

Medical bag Chemical indicator label


Low Temperature Steam Formaldehyde Sterilizer

Low Temperature Steam Formaldehyde Sterilizer


Advantage

This product is a special packing bag for low temperature steam formaldehyde sterilization. The packaging bag is made of medical
dialysis paper and high temperature film. Special chemical indicator color blocks are printed on the surface of the medical dialysis
paper of the packaging bag. When sterilized by steam formaldehyde at low temperature, the biochemical reaction and color change
are produced under the conditions of 2% formaldehyde content, 60 , action time 30min and 2% formaldehyde content, 78 , action
time 10min. According to the color change, distinguish whether the item has been sterilized.

Method of use

According to the volume of the items to be sterilized, cut out the appropriate length of the packaging bag, put into the sterilized items,
and then seal the two ends of the sealing machine (sealing temperature is 180'C-200'C), put the packaging bag into the sterilizer for Advantage
sterilization. After sterilization, take out the packaging bag and observe the discoloration of the chemical indicator color block on the This product is a special chemical indication label for low temperature steam formaldehyde sterilization. The special chemical indicator
surface of the medical dialysis paper. If the color of the indicator color block changes (from red to yellow after sterilization by low color block printed on the surface of the indicator label will produce biochemical reactions and color changes when the formaldehyde
temperature steam formaldehyde), it indicates that the packaging bag has been sterilized by low temperature steam formaldehyde. content is 2%, the temperature is 60 , the action time is 30min, and the formaldehyde content is 2%, the temperature is 78 , the
Point of note action time is 10min.

1. The minimum package length used must contain at least one color-changing block. Method of use
2. If the seal is not good or damaged after sterilization, sterilized items should not be used. Attach the chemical indicator label to the surface of the package to be sterilized. After sterilization, remove the package and observe
the indicator label: If the chemical indicator color block of the indicator label changes from red to a contrasting color (yellow) or lighter
3. Do not contact with acidic, alkaline and chlorinated substances, moisture-proof (relative humidity is less than 60%).
than the contrast color, it indicates that the sterilized package has been sterilized; If the color of the indicated color block does not
4. Avoid light (ultraviolet lamps, fluorescent lamps and sunlight), sealed in a dry and ventilated place. 5. One-time use, do not reuse. change or the color is darker than the contrast color (yellow), it indicates that the package has not been sterilized or has not met the
specified conditions for sterilization. The color change reaction of the indicator block is irreversible, and the indicator label after the
reaction can be stored as a record at room temperature

87 88
ISO 9001/13485/14001/45001

Hydrogen Peroxide Low Temperature Plasma Sterilizer Consumables

Microbicide(H2O2) Advantage

Hydrogen Peroxide Low Temperature Plasma Sterilizer Consumables


Hydrogen Peroxide Low Temperature Plasma Sterilizer Consumables

Use when sterilized articles are sterilized by sterilizer.

Method of use

Put the sterilized items into the roll bag, the length of the sterilized items should be suitable for the sterilization bag sealing, otherwise
the sterilization bag may burst; Abnormity and sharp objects should be placed correctly to ensure safety; Seal the mouth of the
sterilization bag with sealing equipment, indicating the sterilization time, use and other information; Items are properly placed in the
Advantage sterilizer to sterilize. When using the instrument, open the sterilizing bag and remove the items in the direction indicated by stripping.

This product uses hydrogen peroxide as the main component of sterilization solution, hydrogen peroxide content is 60%±3%, can kill Chemical indictor label
bacterial spores.
Point of note

1. The sterilizer in the liquid storage tank must be cleaned and replaced regularly.

2. Should be stored in a well-ventilated, dry and cool environment.

3. If the sterilizer comes into contact with eyes and skin, rinse immediately with plenty of water. If the symptoms are serious, seek
medical attention.

4. The sterilizer and other consumables need to be stored separately to prevent the possible impact of the sterilizer on other consum-
ables.
Biological indicator

Advantage

Advantage This product is a special chemical indication label for hydrogen peroxide low temperature plasma sterilization. A special chemical
indicator printed on the surface of the indicator card, under the action of plasma vaporization, the indicator chemical reaction and color
This product uses the thermophilic fatty Bacillus (ATCC7953) as the indicator bacteria, according to the color change after culture to change, in order to determine whether the indicated item has been sterilized.
judge the hydrogen peroxide low temperature plasma sterilization is qualified.
Chemical indicator card
Application

It is used to monitor the sterilization effect of low temperature plasma hydrogen peroxide sterilizer.

Medical bag

89 90
ISO 9001/13485/14001/45001

Hydrogen Peroxide Low Temperature Plasma Sterilizer Consumables

Advantage Seal test card

Hydrogen Peroxide Low Temperature Plasma Sterilizer Consumables


Hydrogen Peroxide Low Temperature Plasma Sterilizer Consumables

Instructions for low temperature plasma sterilization processes. Quickly and intuitively reflect whether the sterilized item has been
sterilized. After use, it can be used as a sterilization record for preservation.
Method of use

1. Place the indicator card and the sterilized object in the packing bag.

2.try to avoid the device completely covered on the card.

3.After sterilization, observe the indicator card and compare the indicator color on the card to judge the sterilization situation.
Method of use

1.Away from acids, alkalis and oxidizing substances exist in the environment.

2. Avoid light and store in a cool place (below 25 ).

3. Take out the required number of roots and tighten the lid. Advantage

Suitable for low temperature paper plastic bag packaging for continuous sealing machine.
Chemical indicating tape
Method of use

Take out the test card, check whether it is complete, make sure that there is no wrinkles and no pollution. After filling in the information
parameters on the test card one by one, place the test card between the transparent surface of the paper bag and the paper surface
into the sealing machine for sealing test. The black side is in contact with the plastic surface. The test temperature is set to the daily
use temperature, and the sealing speed is set according to the actual condition of the device.

Medical sterilization Non-woven fabric

Advantage

This product is a medical sterilization packaging material formed by a layer of melt-blown layer and two layers of anti-stick layer. In the
standard state, the storage time of the items packaged by this product after sterilization can be up to 6 months. It has the characteris-
tics of resistance to bacteria, waterproof, strong permeability, high tensile strength and so on.

Method of use
Advantage
Cut off an appropriate length of indicator tape and paste it on the seal or surface of the item to be sterilized, and then place it in the
The indicating ink on the tape changes color after disinfection and sterilization, and determines whether the item has been sterilized, sterilizer. After sterilization, the twill on the indicator tape indicates the color change of the mark, which can determine that the steriliza-
and intuitively and conveniently reflects the sterilization status of the item. tion package has been sterilized.
Method of use
Application
Cut off an appropriate length of indicator tape and paste it on the seal or surface of the item to be sterilized, and then place it in the
sterilizer. After sterilization, the twill on the indicator tape indicates the color change of the mark, which can determine that the steriliza- It is suitable for sterilization of packaged medical devices by pressure steam, ethylene oxide, hydrogen peroxide low temperature
tion package has been sterilized. plasma and low temperature steam formaldehyde.

91 92
ISO 9001/13485/14001/45001

Peracetic Acid Low Temperature Sterilizer

Chemical indicator 2. After sterilization, take out the biological indicators in the packaging bag. After confirming that the color of the chemical indicator on
Peracetic Acid Low Temperature Sterilizer

Peracetic Acid Low Temperature Sterilizer


the label changed from blue to red, the biological indicator was cooled to room temperature, the indicator cap was pressed down to
close the side air holes of the cap, and the plastic tube of the indicator was squeezed with the special slot on the reader to break the
ampoule inside the plastic tube. After confirming that the bacteria flakes were fully in contact with the medium, the ampoule was
inserted into the quick reader for culture.

3. At the same time, another unsterilized biological indicator with the same batch number was extruded according to the previous
method to break the ampoule, and cultured under the same conditions as the positive control.

4. For the biological indicator for culture detection, the medium ampoule was put into the reader for culture within 2 minutes after
breaking.

5. The detection time of this product in the fast reader is 0.5 hours. If the color change of the recovery medium needs to be observed, it
needs to be cultured in the incubator for 24 hours.

Chemical indicating tape

Advantage

This product is a special chemical indicator card for peracetic acid low temperature sterilization.

There are special chemical indicator color blocks printed on the surface of the indicator card, which undergo chemical reactions and
produce color changes during low temperature sterilization with peracetic acid.

Method of use

Place the chemical indicator card in the center of the package to be sterilized. After sterilization, open the package and take out the
indicator card for observation: if the color of the chemical indicator block of the indicator card reaches or is lighter than the contrast
color, it indicates that the sterilization package center has been completely sterilized; If the color of the indicated color block does not
change or the color is darker than the contrast color, it indicates that the specified sterilization conditions have not been met. The color
change reaction of the indicator block is irreversible, and the indicator card after the reaction can be stored as a record at room
temperature.

Biological indicator

Advantage

This product is a special chemical indicator tape for peracetic acid low temperature sterilization. The special chemical indicator printed
on the surface of the tape will undergo a chemical reaction and produce a color change during low temperature sterilization with
peracetic acid.

Advantage Method of use

It is suitable for monitoring the sterilization effect of peracetic acid low temperature sterilizer. Remove the appropriate length of chemical indicator tape and paste it on the surface of the package to be sterilized. After sterilization,
remove the package and observe the indicator tape: If the chemical indicator color block of the tape changes from blue to a contrasting
Method of use color (red) or lighter than the contrast color, it indicates that the sterilized package has been sterilized: If the color of the indicated color
1, according to the requirements, put this product into the peracetic acid low-temperature sterilization bag, put the bag in the most block does not change or the color is darker than the contrast color (red), it indicates that the package has not been sterilized or has
difficult position in the sterilizer, and run the sterilization procedure. not met the specified conditions for sterilization. The color change reaction of the indicator block is irreversible, and the indicator label
after the reaction can be stored as a record at room temperature.

93 94
ISO 9001/13485/14001/45001

Peracetic Acid Low Temperature Sterilizer

Chemical indicator label


Peracetic Acid Low Temperature Sterilizer

Peracetic Acid Low Temperature Sterilizer


Advantage

This product is a special packaging bag for peracetic acid low temperature sterilization. The bag is made of Tyvek paper and low
temperature film. Special chemical indicator color blocks printed on the Tyvek paper surface of the packaging bag occur chemical
reactions and produce color changes during low temperature sterilization with peracetic acid. According to the color change, distin-
guish whether the item has been sterilized.

Method of use

According to the volume of the items to be sterilized, cut out the appropriate length of the packaging bag, put the sterilized items into
the two ends of the sealing machine to seal (sealing temperature of 120 ), put the packaging bag into the sterilizer for sterilization.
After sterilization, take out the packaging bag and observe the change of the chemical indicator color block on the surface of Tyvek
paper. If the color of the indicator color block changes (from blue to red after low-temperature sterilization with oxyacetic acid), it
indicates that the packaging bag has been low-temperature sterilization with peroxyacetic acid.

Point of use

1. The minimum package length used must contain at least one color-changing block.

2. If the seal is not good or damaged after sterilization, sterilized items should not be used.
Advantage
3. Do not contact with acidic, alkaline and chlorinated substances, moisture-proof (relative humidity is less than 60%).
This product is a special chemical indication label for peracetic acid low temperature sterilization. The special chemical indicator
printed on the surface of the label, which causes a chemical reaction and color change during the low temperature sterilization of 4. Avoid light (ultraviolet lamps, fluorescent lamps and sunlight), sealed in a dry and ventilated place.
peracetic acid, distinguishes whether the item has been sterilized according to the color change. 5. One-time use, do not reuse.
Application:It is suitable for monitoring of low temperature sterilization process of peracetic acid.

Method of use Medical automatic sealing machine


Place the chemical indicator card in the center of the package to be sterilized. After sterilization, open the package and take out the
indicator card for observation: if the color of the chemical indicator block of the indicator card reaches or is lighter than the contrast
color, it indicates that the sterilization package center has been completely sterilized; If the color of the indicated color block does not
change or the color is darker than the contrast color, it indicates that the specified sterilization conditions have not been met. The color
change reaction of the indicator block is irreversible, and the indicator card after the reaction can be stored as a record at room
temperature.

Medical bag

Advantage
The sealing machine can set temperature, display instant temperature, short sealing time, high sealing firmness, can print sealing date,
expiration date, operator number, operation batch and other parameters, continuous work without interruption.

95 96
ISO 9001/13485/14001/45001

Ethylene oxide sterilizer consumables

EO sterilizing solution Chemical indicator


Ethylene oxide sterilizer consumables

Ethylene oxide sterilizer consumables


Advantage

It is suitable for monitoring the sterilization effect and sterilization operating conditions of ethylene oxide.

This product is a special chemical indicator card for ethylene oxide sterilization. The card is printed on the surface
Advantage Special chemical indicator, under the action of ethylene oxide gas, the indicator chemical reaction and produce color
This product is the necessary consumables for ethylene oxide sterilization, the product is strictly prohibited from high temperature, low Color change.
pressure and violent collision, 100% pure gas.
Method of use
Biological indicator
Place the chemical indicator card in the center of the package to be sterilized. When sterilization is complete, open the package and
remove it

Indicator card for observation: if the indicator card chemical indicator color block color from red to a contrasting color or

Darker than the contrast color, indicating that the sterilization package center has been completely sterilized, if the color of the indicat-
ed color block has not changed

Or the discoloration is not obvious, it indicates that the specified sterilization conditions have not been met. Indicates the discoloration
reaction of the color block

Irreversible reaction, the indicator card after the reaction can be stored as a record at room temperature.

Method of use

1. This product is suitable for monitoring the sterilization effect of ethylene oxide.
Advantage
2. Do not contact with acidic, alkaline and oxidizing substances, moisture-proof (relative humidity is less than 60%)
It is suitable for monitoring the sterilization effect of ethylene oxide.
Method of use 3. Avoid light (ultraviolet lamps, fluorescent lamps and sunlight), and store in a dry and ventilated place.

1. Place the biological indicator in the center of the standard test kit (23cmX23cmX 15cm), place the test kit above the exhaust vent in 4. Please use within the validity period.
the sterilizer or at the most difficult position in the sterilizer, and run the sterilization procedure.

2. After sterilization, take out the biological indicators in the test kit. After confirming that the color of the chemical indicator block on the
label of the biological indicator changes from yellow to black, let the biological indicator cool to room temperature, and crack the safety
bottle inside the biological indicator with a clamp so that the bacteria can fully contact the medium.

3. Culture in an incubator at 56 -58 . At the same time, another unsterilized biological indicator with the same batch number was
taken as a positive control and cultured under the same conditions.

4. After 24 hours of culture, observe the color of the medium to confirm the sterilization effect. Remove the label from the biometric
indicator bottle and store it as a record.

97 98
ISO 9001/13485/14001/45001

Ethylene oxide sterilizer consumables

Chemical indictor label


Ethylene oxide sterilizer consumables

Ethylene oxide sterilizer consumables


Advantage

The color of the indicating ink on the tape changes after disinfection and sterilization, so as to determine whether the item has been
sterilized, and intuitively and conveniently reflect the sterilization status of the item.

Method of use

Cut off the appropriate length of chemical indicator tape and apply to the surface of the package to be sterilized. After sterilization,
remove the package and observe the indicator tape: if the chemical indicator color block of the indicator tape changes from
reddish-brown to green, it indicates that the sterilized package has been sterilized; If the color of the indicated color block does not
change or the discoloration is not obvious, it indicates that the package has not been sterilized or has not met the specified sterilization
conditions. The color change reaction of the indicator block is irreversible, and the indicator tape after the reaction can be stored as a
record at room temperature.

Medical bag

Advantage

Used for ethylene oxide sterilization monitoring, used to mark whether the sterilization package has been ethylene oxide

The process.
Method of use

The indicator label is lifted from the backing paper and attached to the package or directly to the sterilized item, and then placed in the
ethylene oxide sterilization chamber. After sterilization, the label changes from red to blue.

Storage method

The storage temperature is 15-25 , the relative humidity is not more than 50%RH, avoid light, and must not coexist with pollution or
toxic chemicals.

Chemical indicating tape

Application

Use when sterilized articles are sterilized by sterilizer.


Method of use

Put the sterilized items into the roll bag, the length of the sterilized items should be suitable for the sterilization bag sealing, otherwise
the sterilization bag may burst; Abnormity and sharp objects should be placed correctly to ensure safety; Seal the mouth of the
sterilization bag with sealing equipment, indicating the sterilization time, use and other information; Items are properly placed in the
sterilizer to sterilize. When using the instrument, open the sterilizing bag and remove the items in the direction indicated by stripping.
Advantage

This product is completely made of safe polypropylene fiber, which is a disposable medical products final sterilization packaging
material formed by one layer of antibacterial layer and two layers of anti-stretch layer. It has the characteristics of anti-bacteria,
Application anti-static, anti-tension, etc., suitable for packaging medical supplies sterilization, can be used for packaging medical equipment for
pressure steam sterilization, ethylene oxide sterilization, low temperature formaldehyde or hydrogen peroxide low temperature plasma
It is suitable for monitoring the sterilization process of ethylene oxide. sterilization

99 100
ISO 9001/13485/14001/45001

Top Speed Bioiogical Indicator Series Products

High speed Bioindicators for pressure steam sterilization (0.5h) Hydrogen peroxide Low temperature plasma Sterilization Bioindicator (0.5h)

Top Speed Bioiogical Indicator Series Products


Top Speed Bioiogical Indicator Series Products

Advantage
Advantage
This product is made up of Bacillus thermophilus spore (ATCC7953) as indicator bacteria and consists of bacteria tablets, recovery
This product is made of Bacillus thermophilus spore (ATCC7953) as indicator bacteria and consists of bacteria tablets, recovery
medium (loaded in ampoules) and plastic shell. By detecting the activity of A-glucosidase, the survival condition of Bacillus adipogenes
medium (packed in ampoules) and plastic shell. By detecting the activity of A-glucosidase, the survival status of the spore of Steato-
thermophilus was reflected, so as to determine whether hydrogen peroxide low temperature plasma sterilization was qualified. The
bacterium thermophilus was reflected, so as to determine whether the pressure steam sterilization was qualified. The biometric
biometric indicator can be used in conjunction with the corresponding biometric reader to read the biological monitoring results of the
indicator can be used in conjunction with the corresponding biometric reader to read the biological monitoring results of the sterilization
sterilization process within 0.5 hours.
process within 0.5 hours.
Method of use
Method of use
1. Put the product into a low-temperature plasma bag according to the requirements, place the bag in the most difficult position of
1. Put the product into the standard test kit or the package to be sterilized according to the requirements, place it in the most difficult
sterilization in the sterilizer, and run the sterilization procedure.
position to sterilize in the pressure steam sterilizer, load it normally, and run the sterilization procedure.
2. After sterilization, take out the biological indicators in the packaging bag. After confirming that the color of the chemical indicator on
2. After sterilization, remove the biological indicators. After confirming that the color of the chemical indicator on the label on the
the label of the biological indicator changes from blue to red, let the biological indicator cool to room temperature, press down the
biological indicator changed from yellow to black, the indicator cap was cooled to room temperature, the indicator cap was pressed
indicator cap to close the side air vents of the cap, and read
down to close the air holes on the side of the cap, the plastic tube of the indicator was squeezed with the special slot on the reader to
make the ampoule inside the plastic tube broken, and the bacteria flakes were fully in contact with the medium, then inserted into the The special slot on the device extruded the indicator plastic tube to break the ampoule inside the plastic tube, and confirmed that the
top speed reader and covered with the reader cap for culture. Wait for the reader to send the result signal corresponding to the culture bacteria pieces were fully in contact with the medium, and inserted into the speed reader for culture.
hole.
3. At the same time, another unsterilized biological indicator with the same batch number was extruded according to the previous
3. Another unsterilized biological indicator with the same batch number was extruded according to the previous method to break the method to break the ampoule, and cultured under the same conditions as the positive control.
ampoule and cultured under the same conditions as the positive control.
4. For the biological indicators for culture detection, the media ampoule was placed into the reader for culture within 2 minutes after
4. For the biological indicators for culture detection, the media ampoule was placed into the reader for culture within 2 minutes after breaking.
breaking.
5. The detection time of this product in the speed reader is 0.5 hours. If the color change of the recovery medium needs to be
5. The detection time of this product in the speed reader is 0.5 hours. observed, it needs to be cultured in the incubator for 24 hours.
Result determination Result determination
Composite: For sterilized indicators, fluorescent negative (green or -) Composite: For sterilized indicators, fluorescent negative (green or -)
Nonconforming: Fluorescent positive (red light or +) for sterilized indicators Nonconforming: Fluorescent positive (red light or +) for sterilized indicators
The above two results are only valid if the positive control has a positive fluorescence result The above two results are only valid if the positive control has a positive fluorescence result.

101 102
ISO 9001/13485/14001/45001

Top Speed Bioiogical Indicator Series Products

High speed biological indicator for low temperature Steam formaldehyde sterilization (0.5h) Peracetic acid low temperature sterilization Bioindicators (0.5h)

Top Speed Bioiogical Indicator Series Products


Top Speed Bioiogical Indicator Series Products

Advantage
Advantage
This product is made up of Bacillus thermophilus spore (ATCC7953) as indicator bacteria and consists of bacteria tablets, recovery
medium (loaded in ampoules) and plastic shell. By detecting the activity of A-glucosidase, the survival condition of Bacillus adipose This product is made up of Bacillus thermophilus spore (ATCC7953) as indicator bacteria and consists of bacteria tablets, recovery
thermophilus spores was reflected, so as to determine whether the low temperature steam formaldehyde sterilization was qualified. medium (loaded in ampoules) and plastic shell. By detecting the activity of A-glucosidase, the survival condition of Bacillus adipose
The biometric monitoring results of the sterilization process can be read within 0.5 hours when the biometric indicator is used together thermophilus spores was reflected, so as to determine whether peracetic acid low temperature sterilization was qualified. The biometric
with the corresponding biometric reader. indicator can be used in conjunction with the corresponding biometric reader to read the biological monitoring results of the sterilization
process within 0.5 hours.
Method of use
Method of use
1. According to the requirements, put the product into the standard test kit or the paper plastic bag to be sterilized, place it in the most
difficult position in the low-temperature steam formaldehyde sterilizer, load it normally, and run the sterilization procedure. 1. Put the product into the peracetic acid low-temperature sterilization bag according to the requirements, place the bag in the most
difficult sterilization position in the sterilizer, and run the sterilization procedure.
2. After sterilization, take out the biological indicators in the packaging bag. After it was confirmed that the color of the chemical
indicator on the label of the biological indicator changed from red to yellow, the indicator cap was left to cool to room temperature, the 2. After sterilization, take out the biological indicators in the packaging bag. After it was confirmed that the color of the chemical
indicator cap was pressed down to close the side air holes of the lid, the plastic tube of the indicator was squeezed with the special indicator on the label changed from blue to red, the biological indicator was cooled to room temperature, the indicator cap was pressed
card slot on the reader to break the ampoule inside the plastic tube, and the bacteria sheets were fully in contact with the medium, down to close the side air holes of the cap, the plastic tube of the indicator was squeezed with the special slot on the reader to break
inserted into the ultra-fast reader and closed the reader cap for culture, etc the ampoule inside the plastic tube, and the bacteria slices were confirmed to be in full contact with the medium and inserted into the
speed reader for culture.
Wait for the reader to send the result signal corresponding to the culture hole.
3. At the same time, another unsterilized biological indicator with the same batch number was extruded according to the previous
3. Another unsterilized biological indicator with the same batch number was extruded according to the previous method to break the method to break the ampoule, and cultured under the same conditions as the positive control.
ampoule and cultured under the same conditions as the positive control.
4. For the biological indicators for culture detection, the media ampoule was placed into the reader for culture within 2 minutes after
4. For the biological indicators for culture detection, the media ampoule was placed into the reader for culture within 2 minutes after breaking.
breaking.
5. The detection time of this product in the speed reader is 0.5 hours. If the color change of the recovery medium needs to be
5. The detection time of this product in the speed reader is 0.5h. observed, it needs to be cultured in the incubator for 24 hours.

Result determination Result determination

Composite: For sterilized indicators, fluorescent negative (or -) Composite: For sterilized indicators, fluorescent negative (green or -)

Nonconforming: Fluorescent positive (or +) for sterilized indicators Nonconforming: Fluorescent positive (red light or +) for sterilized indicators

The above two results are only valid if the positive control has a positive fluorescence result. The above two results are only valid if the positive control has a positive fluorescence result.