Instruction Manual

Temperature & Humidity Chamber


VS-5TH-150G
VS-5TH-150G Series
Series

※ Please read this manual enough before using.

VISION SCIENTIFIC CO.,LTD


260, Techno-2 Ro,Yuseong-Gu Daejeon-Si Korea
VISION SCIENTIFIC CO.,LTD TEL.+82 32-672-7779 FAX.+82 32-676-3977
Precaution VS-5TH-150G Series

Prohibit Caution Notice

Apart 0.5m on left and right,


apart 0.3m on rear side of unit.

Switch Board No Multi Plugs

Fragile Glass

Do not leave Do not leave


Correct Voltage inframmables flammables
Wear gloves
for safety

Earth Ground
Caution
Low Temperature

220V 1Ph 3 wire Type


Cord & Socket(Option)
Do not dissembly Do not leave high
Do not handle plug without approval humidity and temp.
Keep constant with wet hand
Temperature
* This unit is need Power Cord and Sockets for 220V 1Ph 3 wire Type.
1st Electricity utility must be prepared in advance by user site.
- 2 -
Index
◈ Precaution .................................................................................. 2

◈ Index ............................................................................................ 3

◈ Caution ......................................................................................... 4

◈ Feature .................................................................................. 5

◈ Specification......................................................................................... 6

◈ Installation .................................................................................... 7

◈ Composition.......................................................................................... 9

- Main Body......................................................... 9

- Control Panel .................................................. 10

◈ Operation.......................................................................................... 11

- Operation.......................................... 11

(Refer to Controller setting guide on page 18)

◈ Maintenance ...................................................................................... 12

- Maintenance & Trouble Shooting ....................12

- Diagram ...........................................15

- Spare Parts .......................................16

◈ Warranty............................................................................................ 17

◈ Controller Setting Guide............................................................................ 18

- 3 -
Caution VS-5TH-150G Series

Thank you very much for purchasing our Product.


It is your responsibility to install it in conformance with local electrical codes. For safe operation, pay attention to the alert symbols
through the manual.
Warning
Warning alert you to a possibility of personal injury.
Caution
Caution alert you to a possibility of damage to the equipment.
Note
Notes alert you to pertinent facts and conditions.
Hot Surface
Hot surface alert you possibility of burning injury by hot instrument surface.
Explosive
Explosive alerts you to possibility of explosion by high pressure.
1) Keep the manual near this unit for operator who is end-user(operator).
2) Operator read this manual and is well aware of this unit before using.
3) Do not change Factory parameter and modify this unit, if modify and have trouble,
manufacturer does not have any responsibility.
4) when you install this unit, connect the Ground(earth) connection.
5) the consumer goods and need spare parts, contact to our service team.
6) Do not install below list place.(Avoid below place)
▶ Avoid , has vibration place, ▶ Avoid has corrosion and flammable gas
▶ Environmental install room : Do not over 30℃ ▶ Avoid, has electromagnetic waves place.
(We suggest room temperature at 25±3℃) ▶ Do not install under direct ray of light
▶ Avoid high humidity place.
7) Do not install on slope place.
8) Maintenance particle and dust in the control panel. particle and dust can make short circuit and make fire.
9) Do not change Parameter setting valve on the controller. It can make trouble shooting.
10) Do not modify electric wire network. Prohibit modify because of electric shock. (Be careful electric shock)
11) Do not install and move, when unit is running. Be careful, Hot and Cold .
12) Attention and be careful, motor and turning part.

- 4 -
Feature VS-5TH-150G Series

▶ Feature
VS-5TH Series are suitable for experiments for temperature and humidity test of various samples. Cooling system can be reached to around set
value of Temperature and humidity by PID digital control system quickly with high efficiency. After they are reached on set value, temperature
and humidity deviation is minimized by stable power supply and kept excellent safety for temperature and humidity.

- Programmable Controller with 5.7inch touch screen Interface


- Constant Temperature and Humidity Test from -40 ~ 150’C (Option:r –60~150’C) and Humidity 25~98%
- Temperature, Humidity and other parameters can be recorded using a Recorder
- 120 Patterns / 1200 segments of programming / Real-Time monitoring and date acquisition
- Free PC-software / RS-232 Port as standard offer
- RS-485 Interface port offers connection of 9 chambers to One pc. (Option)

▶ Convenience & Safety


- Efficient operation and maintenance with easy-to-use Temperature & humidity controls
- Anti-fog, heated glass on observation windows can see inside even in low Temperature
- LED lighting lamp included / Built-in Reservoir cassette type water supply system
- Checking of the water level and the amount of water remaining in the tank
- Access port for external probes or wires
- Additional 50mm or bigger diameter cable port can be ordered (Optional)
- Rounded inner chamber corners for easy cleaning and better air flowing
- Vice type door locking system with silicon packing to ensure a perfect seal
- Over-temperature limit / Electrical Leakage Breaker
- Motor / Compressor over current protector
- Compressor High/Low Pressure Limiter

- 5 -
Specification VS-5TH-150G Series

▶ Specification

Model VS-5TH-150G VS-5TH-180G VS-5TH-350G VS-5TH-800G VS-5TH-1000G

Capacity 150 180 336 800 1000

In) WxDxH/mm 500x500x600 600x500x600 600x700x800 1000x800x1000 1000x1000x1000


Dimension
Out)WxDxH/mm 1000x1020x1600 1100x1020x1600 1100x1220x1850 1500x1320x2150 1500x1520x2150

Range -20’C to 150’C

Stability ±0.1’C at Full Range

Temperature Uniformity ±0.5’C at 40’C

Heating Time From 25’C to 100’C Within 30Min

Cooling Time From 25’C to –20’C Within 50Min.

Range 25~ 95% RH at 20~90’C

Humidity Stability ±0.5% at 60%/ 40’C

Uniformity ±3.0% at 60%/ 40’C

Display Programmable Controller with 5.7Inch LCD Color Touch Screen Interface

Controller PC Interface RS-232C Interface with Free PC Software

Programming 120 Patterns / 1200 Segments

Number of
2/8 2/10 2/12
Shelf Standard / Max

Load / Shelf 25Kg

Internal Stainless Steel #304

Material Shelf Stainless Steel #304

External Epoxy Powder Coated Steel

Temperature 18’C to 30’C (Recommend Temp. Range :20’C)


Working
Humidity 30~85%
Condition
Altitude Less Than 2000m

Safety Device Electrical Leakage Breaker, Over Temp. Cut-Off, Water Level Sensor (Low Water Level)

Main Power 230VAC 50Hz, 1 phase 3 wires Option) 380V 50Hz, 3 Phase 4 wires

* Before use, check the electrical specifications and use the correct voltage.
Primary facilities for electrical supply should be ready in advance. (Please contact the user's building facilities team in advance.

- 6 -
Installation VS-5TH-150G Series

▶ Composition in the package


Check these contents before installation.

Main Body 1 Set


Model Number of Standard Available Max. Number(Option)
VS-5TH-150G 2EA 8EA
VS-5TH-180G 2EA 8EA
Shelves
VS-5TH-350G 2EA 10EA
VS-5TH-800G 2EA 12EA
VS-5TH-1000G 2EA 12EA
Manual 1 EA
Power Cable 1 EA

▶ Installation
① Left and Right Space ②Front and Back Space

0.5m

Wall
Wall

0.5m

0.3m

front

Wall

* Make secure minimum 0.3m for front and Back Space and 0.5m for left and side space from wall for maintenance and cooling air & hot air
circulating.

- 7 -
Installation VS-5TH-150G Series

1) Install the unit on horizontal place.


2) Secure space of unit front : In case of front parts of the unit, door is often opened and test sample are often access in the front of the
unit, so care must be taken not to load other units or material in front of the unit.
3) Installation Environment: Installation Environment is recommended Temperature on 25±3℃ and humidity on 30~85%.
In unavoidable circumstances, room temperature should be kept on less 30'C, avoid to high humidity place if can.
If not, motor or other parts can not be worked well.
If unit must be installed on enclosed or small place, you need to install the separated air exhausting system for
exhaust of hot air from the unit.

4) Because dust is attached to samples in places where there is a lot of dust in the surrounding environment, it is difficult to achieve accurate
results, so it is necessary to keep the surrounding environment clean at all times.

5) Avoid direct sunlight because painting of equipment may be discolored if exposed to direct sunlight for a long time.Do not load objects onto
the top of the machine.

6) Make sure that the voltage is 220V single phase (three wire with ground) or three phase (four wire with ground) before connecting power.
(Caution is required for mobile installation.)

7) If the power is three-phase, check the rotation direction of the fan motor around evaporator after connecting the machine .
If fan is rotated conversely, exchange position of R phase and S phase of the power line each other. If they are connected well, heat air of
the unit is out to rear side of the unit. (Occurred Hot air from the unit must be out to rear side)

8) Voluntary Modification of the machine or operation circuit may cause malfunction, so change is prohibited.
(Be careful for Electric shock of 220V)
9) Connecting the power of the device
Connect to the pre-prepared power line (see product specifications for the relevant model rating).

- 8 -
Composition VS-5TH-150G Series

▶ Composition(Main Body
(Main Body)

Description Function

① Door Door

② Window Viewer Window viewer can be checked inside chamber

⑥ ③ Door Knob Opening and Closing knob for Door


② Operation room
④ Operation Room
Ⓐ (Compressor, Heater and others)
⑤ Wheel with Stopper Fix of unit position on the bottom

Ⓡ Touch Screen, Warning light,
Switches for FAN , COOLING and HEATER
⑦ ⑥ Controller
EMERGENCY Button,
④ Power Circuit Breaker


Water tank for humidity control
[Main Body] ⑦ Water Tank
(* Distilled water must be used)

⑧ ⑧ Main Power Cable Main Power Cable


⑨ Water Supply Water supply
⑩ Water Drain Main Water Drain
⑪ Water Over Drain If Water is over on water tank, it drains.
⑨ ⑩ ⑪
Ⓐ Access Port Access Port hole with rubber closer
Ⓛ Warning Light Warning light (Option)

Ⓡ Digital Thermal Recorder Digital Thermal Recorder(Option)

[Rear Side]
- 9 -
Composition VS-5TH-150G Series

▶ Composition(Control Panel)
NO Description Function
ⓐ TOUCH SCREEN All test condition can be set on display.
FAN Switch & Lamp
ⓑ FAN
-Air Circulate Fan in the chamber
ⓒ T-HEATER Temperature Heater Switch & Lamp
ⓓ H-HEATER Humidity Heater Switch & Lamp


Cooler Switch & Lamp
ⓔ COOLER
- Compressor (Cooling System)
Lamp Switch & Lamp
ⓕ LAMP
-When you check the sample in the chamber, on door lamp
EMERGENCY S/W : When there is a
emergency situation, press this
ⓖ EMERGENCY STOP button immediately. If Emergency
situation is over, dial this

ⓑ ⓒ ⓓ ⓗ Power Circuit Breaker


button clockwise.
it supply main power and overcurrent
blocking device

ⓔ ⓕ ▶ Main Controller and Digital Thermal Recorder

We have offered a guide

ⓖ book for Touch screen


and Thermal recorder(Option),
separately.
Refer to Guide book, please.

Model : TEMI 2500 Model : Yokogawa SR-10000 (Option)
* Main Controller and Digital Thermal Recorder can be changed to other equal or better model without notice for better quality.
- 10 -
Operation VS-5TH-150G Series

▶ Operation
When water is filled in water tank, over 1st distilled water is recommended.
If tap water or unclean water is used, unit might not be worked well.
1) Connect Main electric power cable.
2) Fill clean water or over 1st distilled water into the water tank
3) Turn on the main circuit breaker on Control panel.
4) Turn on the switch for ⓑFAN, ⓒT-HEATER, ⓓH-HEATER and ⓔ COOLER on control panel.
5) Set Temperature, humidity and other test condition on touch screen.

Refer to guide book for specification and operation on end of this manual.
6) Touch ‘RUN” button on Controller.
7) Check temperature and humidity after running.
8) if unit has something wrong and trouble, contact to our service team

▶ Caution
- The main voltage must correspond to the voltage given on the name-plate
(AC 220V, 50/60 Hz, Single Phase or 380V 50/60Hz, Three- Phase )
- Install the unit even level with the ground.
- Avoid vibration place.
- Connect and disconnect power, do not touch the cable, hold the plug
- Be careful drop liquid or water into the controller and unit .
- Explosive alerts you to possibility of explosion by high pressure
- Make clean for maintenance
- This equipment can make high temperature and hot,
- Must use protect hand gloves .When you open the door, be careful hot air.
Must Take protector.
- When unit is running, do not open the door
- Make Space around equipment over 50 cm for Left , right, back.
- Do not impact or shock to the equipment.
- Some parts made by ceramic so easy broken.
- Check electric Ampere and connect.

- 11 -
Maintenance VS-5TH-150G Series

▶ Maintenance & Trouble Shooting


Electric shock(220V or 380V) or power failure by short might be occurred when unit is checked.
Be careful electric shock(220V or 380V) or power failure by short might be occurred when unit is checked. If some problem are
occurred on the unit, turn off the main power first. Then, contact to distributors or service team you bought this unit.

1)Trouble case and solution


▶ Excessive temperature/humidity ▶ Poor rising of Temperature or humidity ▶ Poor of WATER TANK level
▶ High pressure and overload of the compressor ▶ Buzzer of End of test
** If an error occurs, an error message will be displayed on the screen.

2) Excessive temperature/humidity (Warning for T-O.T.C & H-O.H.C )


T-OTC and H-OHC are protector for Over Temperature and Humidity.
T-OTC
T-OTC : Over Temperature Controller(OTC) (Setting Value: 170~200’C)
H-OHC : Over Humidity Controller(OHC) (Setting Value : 100%)

- Check 1 : Check the setting value Temp. & Humi. on Controller H-OHC
- Check 2 : T-OTC / H-OHC setting value.
(There are safety device(T-OTC and H-OHC) on right side of panel

If present temperature in chamber is raised than over setting value, T-OTC is worked.
Check about shown error message on control panel. Check T-OTC after opening side case of the unit.
Turn off the main Power. If fan is rotated conversely, temperature might be raised . Contact to distributor or service team you bought this
unit. After problem is solved, set T-OTC again.
(Depends on your setting value, you should set higher 30~50’C on T-OTC and 20~30% H-OHC. These range is up to your test)

If H-OHC is worked, check about water level in water tank. If water supply is not working even though water is enough in water
tank, H-OHC would be worked. Contact to distributor or service team you bought this unit.
- 12 -
Maintenance VS-5TH-150G Series

- Check3 : Check Voltage Terminal and Float Switch. (They are in right side of the unit)

Lamp

Voltage
Terminal
220VAC
When Lamp is on, DC is applied to the input terminals. The voltage terminal output is energized without contact point. Voltage is applied to heater.
When heater output on controller is "0%", Lamp is blink. But, lamp might be turned on when it is break down.
If Heater output on operating mode is "0%", test AC voltage by using tester.

3) Trouble for excessive Temperature and humidity.


Check setting value for temperature and humidity on control panel. Not only check about setting value for temperature and humidity on control
panel but also running button is pressed or not and condition of heater M.C , SSR, safety device(T-OTC, H-OHC).
If Lamp is on when heater output on controller is 100%, test AC voltage by using tester. If AC voltage is 220V, SSR is break down.
If you can not test by yourselves, contact to distributor or service center you bought this unit.

4) High pressure and overload of the compressor(Ref or P/S warning occurrence)


- High pressure on compressor is occurred when motor for compressor condenser is not be worked or Air flow is not
good because a lots of dust are stacked around condenser by unclean. Take off front-bottom cover of the unit.
Clean the dust on condenser every 3~6months.
- After high pressure warning is occurred, high pressure switch is returned after 3~4 hours.
After 3~4hours, running this unit again. If you can not test by yourselves, contact to distributor or service center you bought this unit.
- 13 -
Maintenance VS-5TH-150G Series

5) Buzzer sound
If there are trouble on the unit, power of heater and compressor is off with buzzer sound.
After checking what is the trouble, solve the problem.

6) Humidity error
During the humidity test, the temperature is normally controlled, but sometimes the humidity is shown 100%.
It can be occurred this trouble when water is not enough in the chamber, check if the water is supplied.
If this trouble is happened even though there is water enough in chamber, the part of the humidity sensor inside the chamber
might be problem.

Disposal for unit trouble :

We strictly forbid any repair or exchange of parts by individuals except the repair by authorized engineer.
The product may be damaged or do not function properly. Furthermore, there is a danger of an electric shock or a fire.
If there are some problem on the unit, contact to local agency or us immediately.

1. When you apply after sales service, let us know below point.

- Model name
- Purchase date and year (There are mentioned on barcode of unit)
- Trouble contents in detail with some photos.
- Let us know about organization name, user name, contact number

※ If you have any question, contact to local service team or our service center.
VISION SCIENTIFIC CO.,LTD
Address : 260, Techno-2 Ro, Yuseong-Gu Daejeon-Si, Korea.
Tel : + 82 32 672 7779 Fax: +82 32 676 3977
E-mail: [email protected]

- 14 -
Diagram VS-5TH-150G Series

▶ Diagram

VISION SCIENTIFIC CO LTD

VS-5TH Series

- 15 -
Spare Parts VS-5TH-150G Series

▶ Spare Parts

No. Contents Q’ty

1 PCB for Temperature Sensor 1

2 Touch Screen + Controller Relay Board 1

3 Compressor 1

4 Heater for Temperature in chamber 1

5 Humidifier Heater 1

6 SSR(Solid State Relay) 2

7 Water Solenoid Valve 1

8 Temperature Sensor(PT 100) 1

9 Sensor for Humidity 1

10 LED Converter 30W 1

11 Transformer 1

- Contact to seller on your country or us if you want to exchange or repair the parts separately.
- We strictly forbid any repair or exchange of parts by individuals except the repair by authorized engineer.
- Related Spare parts can be changed depends on situation of parts supply.
-
- 16 -
Warranty

* Thanks to purchase products for vision scientific CO., LTD.


Items Temperature & Humidity Chamber

* The product quality control and strict inspection and testing by


Model VS-5TH-150G Series
product is issued.
Serial No.
* The remaining amount of this product failures occurring within
one year will be a free repair. However, the following cases even Purchase Date.

within the warranty period, we charge the actual cost.


Seller

1. In case Users of the negligence (unreasonable transportation of Name


products, operations) if the failure and damage caused,
Customer
2. This product is non-specialized technicians to repair internal Add
Info.
disagreements is defective
TEL ( ) -
3. Fire, earthquake, flood and other acts of God
or force majeure when the accident occurred as a failure VISION SCIENTIFIC CO.,LTD
4. How to use or using a normal power (voltage) in violation 260, Techno-2Ro, Yuseong-GuDaejeon-Si, Korea.
of the case of failure TEL: +82 32 672 7779 FAX: +82 32 676 3977

VISION SCIENTIFIC CO.,LTD


- 17 -
Controller Setting Guide
Operation Manual (Temperature and Humidity Programmable Controller)

VARIOUS It is temperature and humidity programmable controller which equips with the general control,
heating and cooling function by supporting high definition TFT-LCD touch screen and SD card.
This manual is commonly used for TEMI2000 Series.
Queries related with after sales service for TEMI2000 series
Please inform the TEMI2000 model name, failure condition and
Contents
contact point for queries of after sales service.

T :and
01. Operation 82-32-326-9120
setting ‥‥‥‥‥‥‥‥‥‥‥‥‥ 4 06. Program setting ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 48



1-1. Basic operation flow chart‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 4 6-1. Program pattern setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥49
F : 82-32-326-9119
1-2. Setting button operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 5 6-2. Pattern repetition setting‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥54
1-3. Parameter setting method ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 6 6-3. File editing‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥56
6-4. Time signal operation‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥59
02. Main screen ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 9

6-5. Standby operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥62
3-1. Pattern graph display‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥12
Customer contact for TEMI2000 series 6-6. Experiment name setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥64
03. Setting graph display and save ‥‥‥‥‥‥‥ 12
07. Appointed operation setting ‥‥‥‥‥‥‥‥‥ 66

Quotation request
graph/view‥
Product request


3-2. Presented value (PV) ‥‥‥‥‥‥‥‥‥‥‥‥‥‥15
Specification
3-3. Presented value (PV) request / Data
graph saving request/ Other request
setting‥‥‥‥‥‥‥‥‥‥‥18 08. Screen display setting ‥‥‥‥‥‥‥‥‥‥‥‥ 71


3-4. Memory save setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥20 8-1. Screen display setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥71
■ Internet
www.samwontech.com
4-1. Stationary operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥23 8-2. Touch screen calibration Seting‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥72
8-3. DI error creation history view‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥74
04. Operation state screen setting ‥‥‥‥‥‥‥‥ 23

■ Eoperation‥
4-2. Program mail ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥30 09. Communication error ‥‥‥‥‥‥‥‥‥‥‥‥ 77


[email protected]
4-3. Auto tuning and tuning point‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥40

05. [email protected]
motion setting ‥‥‥‥‥‥‥‥‥‥‥ 42

5-1. Operation method setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥42
5-2. Fuzzy operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥44
5-3. Setting value change rate (SLOPE) operation‥‥‥‥‥‥‥‥‥‥45

6th Edition of TEMI2000 Series IM : MAR. 27. 2020


01. Cautions (Instructions) for safety
Thank you for your choice of our Temperature and Humidity Programmable Controller(TEMI2000 series). This manual describes the method of installation of the product.

Cautions in this instruction manual


■ Please deliver for the end user to possess always and keep it in the place accessible at any time. Symbol marks for safety
■ Use the product after full understanding of this installation manual.
■ This installation manual does not warrant any other things because it is a description of the details for the function. (A) It means the“Handle with care” or
■ A part or whole of this manual shall not be edited or copied randomly. “Cautions”In case of violation of this
■ The descriptions in this manual may be changed randomly without pre notice or warning. point, it may cause the death, severe
■ Even though this manual was made with elaboration, it will be appreciated if you inform to the purchasing point (Dealer shop and etc)
or sales team in our company in case of deficiency, mistake or omission in the contents. injury or the extreme damage on the
product.
Cautions for the safety and modification (Change) of the product ■ Product : It is marked on the
■ Please use this product after full understanding on the safety cautions in this manual for the protection and safety for this product and the system
connected to this system.
points to be acknowledged
■ Our company is not responsible to the damages occurred by using or handling or unattended using not relying on this installation manual. certainly to protect the human
■ Please install at the outside of this product when the additional protection and safety circuit is installed Octarately for the protection and safety for body and device.
this product and the system connected to this system. ■ Instruction manual : It describes
■ The internal modification (Change) and addition to this product are prohibited.
■ Do not disassemble, repair and modify of this product because it becomes the reasons for electric shock, fire and malfunction.
the cautions to prevent the cases
■ In case of changing the part or the consumables of this product, please contact to the sales department of our company. of endangered situation on the
■ Do not contact to the moisture with this product. It may cause the failure on this product. life and body of the user due to
■ Do not apply the strong impact on this product. It may cause the damage and failure on this product. the electric shock and so on.
With regard to the exemption for the responsibility of this product
■ We are not responsible for any warranty on this product besides the defined cases in the quality assurance condition of our company. (B) It means“Ground terminal”
■ We are not responsible for the direct or indirect damages on the user of any third party due to the not expectable defect or the natural disaster in ■ Make the earth with the ground in
use of this product.
case of product installation and
With regard to the quality assurance condition of this product controlling the product.
■ The warranty period shall be one year from the purchasing of this product. Free of charge repair is available only for the cases of out of order
occurred from normal use conditions.
(C) It means the“supplementary
■ The repair due to the out of order occurred after the warranty period shall be repaired at the actual cost according to the defined condition by our
company. explanation”
■ The out of order occurred within the warranty period shall be repaired at the actual cost for the following cases in spite of within the warranty ■ It describes the points to
period. supplement the explanation.
(1) O
 ut of order due to the mistake or fault of the user (4) Out of order due to the random disassemble, change or damage on the product.
(Ex: Initialization by losing the password and etc.) (5) Out of order due to the electric power instability
(2) Out of order due to the natural disaster (6) Others (D) It describes the“references”
(Ex: Fire and flood and etc) ■ It describes the information and
(3) Out of order due to the movement of product after installation. pages of reference to be referred.
■ P lease contact to the purchasing points or sales part of our company when after sales service is necessary because of the failure on the product.

2
01
Part

Operation and setting


1-1 Basic operation flow chart‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 4
1-2 Setting button operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 5
1-3 Setting button operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 6

3
01. Operation and setting
This product is programmable controller designed with dialogue style touch screen easy-to-use for the users.

1-1. Basic operation flow chart


• T he logo screen and the initial screen are displayed sequentially when the electric power is switched “ON”
after installation of the product and it converts to the program stationary screen.
• It takes about 15 seconds for screen loading
• When button is touched at the top of the program stationary screen, it converts to the main screen.
• Refer to [13. System initial setting] in [Installation manual] for change in the initial screen.

Logo screen Initial screen Program stationary screen Main screen

4
1-2. Setting button operation

Button type Button operation


The “Set value” is touched in stationary operation/still screen and it is used for setting the set value wanted by the user.

The “Pattern No.” is touched in program still screen and it is used for setting the pattern number wanted by the user.

It is used for inputting the general numbers and name.


It is used for selection for one out of many types.
It is used for selection for one out of more than 2 parameter setting. (ON/OFF/Inactive state)
It is used for selection of Y/N for the corresponding parameter. (ON/OFF/Inactive state)
It is used for general screen conversion.
It is used for increasing or decreasing of the page within the same screen.
It is used for the page conversion by the decrease and increase in time axis on the same screen.

01. Operation and setting


5
1-3. Parameter setting method
● When is selected in [1-2 Setting button operation], the input key of the setting value is shown as followings and the necessary data can be input.
● When the data out of the setting range is input, error message (“LIMIT ERROR”) is shown on the input data display window with the error sound (“Beep”).

▲ Input key for setting only the numbers

Input key for setting the pattern experiment name and DI error name ▶
Refer to [11. DI function and operation setting] in [Operation manual] for DI error name input key.

▲ Input key for time signal setting ▲ Display when it is out of the setting range

Touch key lock release


● Input OFF (Lock release state) for key block because the set value is not input when “Key block” is “ON.”
● Refer to [4. Operation state screen setting] for details

6
(1) Method for effectiveness of setting button and setting value
● This product is designed as follows when the setting data input button is touched or to check the effectiveness of the input setting data by sound.
● “Beep” : When the basic setting button is touched or the setting data is input normally
● “Beep and beep” : When the input data by the setting data input key is out of the input range.
● Do not press with sharp thing (Pencil and etc) or excessive force on the input key for basic setting button or setting value.
It may cause the mal operation of the device or damage on the touch panel.

(2) Setting value input method


● Every input data used in this product is set by the set data input key, test name input key and time signal input key.
● The input key for set data is appeared when button is touched in [1-2 Setting button operation] and the value to be set can be input.
● Refer to [6-4 Time signal operation] for time signal input.
● Refer to [11. DI function and operation setting] in [Operation manual] for DI error name input key.

▶ Set the value of the input key features and descriptions


① It displays “Parameter.”
1
10 3 ② It displays “Setting range.”
2
b c e 8 4 It displays “setting display window.”
③ • It displays “LIMIT ERROR” when it is out of the setting range.
6 7 9 5 • It displays “INPUT ERROR” when there is an error in setting unit.

01. Operation and setting


d a f ④ It is used to return to original screen after stopping the input.
⑤ It returns to the original screen by saving the input data.
Ex) Set data input method ⑥ It is used for input the decimal point.
Press the set data input button in the corresponding screen ⑦ It is used for input the symbol (+/-).
Press the “ENTER ” key finally after pressing the corresponding number in sequence
⑧ It is used for erasing the input data by one character.
(ⓐ→ⓑ→ⓒ→ⓓ→ⓔ)
⑨ It is used for erasing all input data.
⑩ It displays the already input setting data.

7
02
Part

Main screen ‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 9

8
02. Main screen

[Fig. 2-1] Main screen [Fig. 2-2] Main screen #2


8 7

1 2 3

4 5 6

02. Main screen


No. Instruction Description No. Instruction Description
Moving to the screen to set Y/N for using graph display, Moving to the screen for setting the screen
① GRAPH & RECORD ⑥ DISPLAY SET brightness, back light electricity saving and humidity
graph record, SD card recording
display.
② OPERATION SCREEN Moving to the operation screen It sets Y/N of use buzzer sound.
Moving to the setting screen for additional function and ⑦ BUZZER SOUND The buzzer sound generated in DI error is operated
③ FUNCTION & FIX


operation method when it is set for no use.
④ PROGRAM SET Moving to the program setting menu screen Set the language to use
Moving to the screen for setting current time, ⑧ LANGUAGE SET (English, Korean, Simplified, Traditional, Japanese,
⑤ RESERVE SET
Russian languages available)
programmed operation time.
9
03
Part

Setting graph display and save


3-1 Pattern graph display‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 12
3-2 Presented value (PV) graph view‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 15
3-3 Presented value (PV) graph save setting‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 18
3-4 Memory save setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 20

10
Setting graph display and save
Flow chart

[Fig. 3-2] Graph & Saving screen 1 (Pattern graph display) [Fig. 3-5 Graph & Saving screen 2 (Graph display selection)

[Fig. 3-12] Graph & Saving screen 4 [Fig. 3-9] Graph & Saving screen 3

Part 03
11
03. Setting graph display and save
3-1. Pattern graph display
● It converts to [Fig. 3-2 Graph & Save screen 1 (Pattern graph display)] when [Graph & Save] is selected at the left top in [Fig. 2-1 Main screen]
● This screen displays the temperature and humidity operation pattern and progress time in program operation.
● It is a screen to display the input pattern in [Fig. 6-2 Pattern editing screen].
● Can change the status of Temperature·Humidity Check, and , even in the graph screen.

[Fig. 3-1] Graph & Save screen 1 (Program operation) [Fig. 3-2] Graph & Save screen 1 (Pattern graph display)

7
2 3 1
4 5

12
Display the graph setting menu [Fig. 3-3] Graph & Save screen 1 (Program operation)
① ● D
 isplay the parameters related to the pattern graph operation at the bottom
depending on ON/OFF operation on the graph menu button

The temperature (Red) is disappeared on the graph screen when the checked 2
button is touched in the currently operating screen and temperature (Red) is

displayed in the graph screen when button is touched again.
● , buttons are used even in during operation(RUN) .
1
The humidity (Blue) is disappeared on the graph screen when the checked
button is touched in the currently operating screen and humidity (Blue) is

displayed in the graph screen when button is touched again.
● , buttons are used even in during operation(RUN) . 3

Setting the pattern No. to be displayed 4


● D
 isplay the input key to set the pattern No. when (Pattern No.) is

touched. ① Display the current temperature during operation
● Refer to [Fig. 3-4 Pattern No. input screen] ② Display the current humidity during operation
Setting the time on graph X axis

03. Setting graph display and save


③ Display in green for the operation ended part
● D
 isplay of the input key to set the time on X axis when Display the processing time for the set pattern in
⑤ ④
(Display time) is touched. [6-1 Program pattern setting]
●The time on X axis can be changed during operation
⑥ Moving from current screen to next screen
Change into the Previous/Next stage on the time axis when is touched

on the current page.

13
[Fig. 3-4] Pattern No. input screen

References
▶ It is a screen to input the pattern No. to be displayed in graph.
▶ The pattern No. can be input even during operation.

Parameter Setting range Unit Initial value


PATTERN NO. 1 ~ 40 ABS 1
30 MIN, 1 HOUR, 3 HOURS, 6 HOURS,
VIEW TIME ABS 30 MIN
12 HOURS, 24 HOURS

14
3-2. Presented value (PV) graph view
● It is a screen to display the data recorded in [4-1(4) Stationary operation 3 operation screen] and [4-2(4) Program operation 3 operation screen].
● The date and time saved into the memory are displayed at the top of the screen. [Reference 1]

[Fig. 3-5] Graph&Save setting screen 2 (Graph display is selected)


When Setting copy the recorded PV file in the internal memory to an SD card
3 4 5 6
④ : Send the selected file.
1 : Send the whole file.

7 Set storage media display


⑤ : Memory card file
8
: SD card file
[Reference1]
9 Set graph display direction
10 ⑥ : Vertical (Y-AXIS)
: Horizontal (X-AXIS)
11
⑦ Moving from current screen to next screen
12
When is touched, the files saved into the internal memory is
2 13 ⑧ displayed.
● Moving from current screen to next screen


⑨ The time axis is expanded or reduced.

03. Setting graph display and save


It displays the temperature (Set-data, Indicated-Data) and humidity (Set-data, ⑩ Moving to the beginning and end of the displayed PV graph page
Indicated-Data)
⑪ Moving of graph screen by one page.
① ● When the checked is touched, it is disappeared on the graph screen and

when is touched, it is displayed on the graph screen. Moving the blue line on the graph screen up/down by 1 DOT
● Refer to [Fig. 3-5 and Fig. 3-6 Graph & Save setting screen 2] ⑫ ● When the screen is touched, the indicated values are displayed while the blue


line moves.
② Key lock / SD card / Storage indicator
③ : Display to update the graph screen immediately which is being saved. ⑬ Copy the recorded PV files into the internal memory to SD card.

15
[Fig. 3-6] Graph&Save setting screen 2 (Graph display is not selected) [Fig 3-11] When setting the X-AXIS of Trend direction, Graph Screen

References
▶ I t is a screen for no selection from the temperature
(Set-data, Indicated-Data) and humidity (Set-data, Indicated-Data)
▶ It is a screen to display the saved file into the internal memory.
▶ Refer to [4-1(4) Stationary operation 3 operation screen] and [4-2(4) Program
operation 3 operation screen] for saving into internal memory.

16
[Fig. 3-7] Graph&Save setting screen 2 (Saved folder is displayed) [Fig. 3-8] Graph&Save setting screen 2 (Saved file is displayed)

5
2

① Copy the recorded PV files into the internal memory to SD card.


Move to the start and end in case of data searching stored in the

internal memory

03. Setting graph display and save


When search for the saved file in the internal memory, up / down to

the end 10 to the units.
④ Close the PV file
⑤ Move to the folder

17
3-3. Presented value (PV) graph saving setting
● T his screen is to set the display range and sampling time which are necessary for graph recording in [4-1(4) Stationary operation 3 operation screen] and [4-2(4) Program operation 3
operation screen].
[Fig. 3-9] Graph&Save setting screen 3
Setting the PV graph saving period
● It is not changeable during PV graph saving
① ● Saving about 31 days is possible when sampling time is set in 1 minute in

1 4 saving into internal memory.


Internal memory and Setting Y/N for saving the data into the SD card
● AUTO : Saving the data in synchronized with Operation/Stop automatically
2 ② ● MANUAL : Saving the data by the saving key in the Operation

5 screen 3 manually
Setting the media for data saving
3 ● MEM : Stored in the internal memory
③ ● SD : Stored in the SD card
● BOTH : Simultaneously stored in the internal memory and SD card

④ Setting the display range of Temperature graph


⑤ Setting the display range of Humidity graph

18
Parameter Setting range Unit Initial value
SAMPLING TIME 1 SEC, 2 SEC, 5 SEC, 10 SEC, 20 SEC, 30 SEC, 1 MIN ABS 1 SEC
RECORDING OPER. AUTO, MANUAL ABS AUTO
RECORE MEDIA MEM, SD, BOTH ABS BOTH
TEMP GRAPH DISPLAY HIGH T.EU (-2.50 ~ 102.50%) T.EU T.EU(100.00%)
TEMP GRAPH DISPLAY LOW (Temp graph display lower limit < Temp graph display upper limit) T.EU T.EU(0.00%)
HUMI GRAPH DISPLAY HIGH H.EU (0.0 ~ 100.0%) H.EU H.EU(100.0%)
HUMI GRAPH DISPLAY LOW (Humi graph display lower limit < Humi graph display upper limit) H.EU H.EU(0.0%)

03. Setting graph display and save


[Fig 3-10] Sampling time of Recoding cycle setting screen

19
3-4. Memory save setting
● It is a screen to set the transmitting of pattern and parameter to SD card.
● It is a screen displayed in SD card option only.

[Fig. 3-12] Graph&Save setting screen 4


Setting the items and direction of transmitting in SD card and TEMI2000
● PTN. : Download or upload the set pattern in
[6-1 Program pattern setting]
1 ● PARA. : Download or upload the set parameter
① ● ALL : Download or upload the pattern and parameter
● DOWNLOAD : Transmitting the selected transmitting items out of

the internal data in TEMI2000 to SD card


● UPLOAD : Transmitting the selected transmitting items out of the
saved data in TEMI2000 to SD card
It displays the current capacity of SD card

● It displays when the SD card is inserted only

③ When the data is not in recording to SD card, is activated and


download and upload are possible when is touched by.
3
2

Parameter Setting range Unit Initial value


BACKUP ITEM PTN., PARA., ALL ABS PTN.
DIRECTION DOWNLOAD, UPLOAD ABS DOWNLOAD

20
04
Part

Operation state screen setting


4-1 Stationary operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 23
4-2 Program operation‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 30
4-3 Auto tuning and tuning point‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 40

21
Operation state screen setting
Flow chart

[Fig. 4-1] Stationary operation still screen 1 [Fig. 4-5] Stationary operation still screen 1 [Fig. 4-7] Stationary operation still screen 2 (General)

[Fig. 4-9] Stationary operation still screen 3 [Fig. 4-8] Stationary operation still screen 2 (Heating·Cooling)

Part 04
22
04. Operation state screen setting
4-1. Stationary operation
(1) Stationary operation still screen 1
● When the operation state screen is selected in [Fig. 2-1 Main Screen], it is converted to “Stationary operation still screen 1.”
● Select the operation method with “Stationary” in [5. Operation motion setting]
● When at the right bottom of [Fig. 4-1] Stationary operation still screen 1 is touched by, it converts to [Fig. 4-5] Stationary operation still screen 1.
● If parameter isn't set due to the noise, it is displayed as [Fig. 4-2] Parameter setting error screen due to noise.

In this case, Please use to establish a noise countermeasure about power and communications cable.

04. Operation state screen setting


[Fig. 4-1] Stationary operation still screen 1 [Fig. 4-2] Parameter setting error screen due to noise

23
[Fig. 4-3] Screen of Input for temperature setting about operation [Fig. 4-4] Screen of Input for humidity setting about operation
temperature

Parameter Setting range Unit Initial value


References
Temp setting data(SP) T.EU(0.00 ~ 100.00%) T.EU T.EU(0.00%)
▶ The input for set value for temperature is activated when
Humi setting data(SP) H.EU(0.0 ~ 100.0%) H.EU H.EU(0.0%)
button is touched [Fig. 4-3 creen of Input
※ T.EU : Range of input data for temperature sensor for temperature setting about operation temperature]
※ H.EU : Range of input data for humidity sensor ▶ The input for set value for humidity is activated when
※ Refer to [Engineering units] button is touched. [Fig. 4-4 Screen of Input
for humidity setting about operation]
▶ When the input of set data of temperature and humidity are
completed, operate the stationary operation by selecting .

24
(2) Stationary operation #1 operation screen
● When the “Setting data(SP)” is touched even in operation, the input key setting for operation is activated.
● It is a screen for Measured data(PV) and Set data(SP).

[Fig. 4-5] Stationary operation still screen 1


7 ⑥ It displays the humidity setting data (SP) to be controlled.
1
It displays the current date/time and LCD backlight is off when it is touched.
8 ⑦
2 ● Red LED lamp at the right top is ON when the backlight is OFF in still state.
9 Moving to [Fig. 2-1 Main screen]
3
⑧ ● It displays the key pad to input the password when main button restriction is set.
● Refer to [Fig. 5-2 Screen in restriction setting of main button]

⑨ Moving from current screen to next screen


4 10
Execution or releasing the auto tuning with temperature set value (SP).
5 11 ⑩
● Y/N of the tuning button display is set in [8. PID group] in [Operation manual]
12 Execution or releasing the auto tuning with humidity set value (SP).
6 ⑪
● Y/N of the tuning button display is set in [8. PID group] in [Operation manual]
13
User button
● Y/N for use in [13. System initial setting] in [Operation manual]

04. Operation state screen setting


It displays the current operation status. ● User uses the wanted relay in [10. DO relay setting] in

● The arrow rotates to the clockwisely during operation. ⑫ [Operation manual] when the user button is used.
② It displays the temperature setting data (SP) to be controlled. Ex) It is used for light the chamber.
③ It displays the present temperature value (PV). ● The set relay is operated when the “User” button is touched by
④ It displays the present humidity value (PV). in the stationary and program operation/stop screen.

⑤ Full screen button ⑬ Stationary operation / Stop button

25
References
▶ , : Setting data > Measuring data is displayed in TEMP and HUMI increase.
▶ , : Setting data = Measuring data is displayed in TEMP maintaining.
▶ , : Setting data < Measuring data is displayed in TEMP decrease.

[Fig. 4-6] Stationary operation still screen 1 Full screen

26
(3) Stationary operation #2 operation screen
● It is a screen to display the display lamps for measuring data, setting data and output volume.
● When the “Output volume” is touched in Heating·Cooling, output volume, Heating volume and Cooling volume are displayed in turn.

[Fig. 4-7] Stationary operation #2 operation screen (General) [Fig. 4-8] Stationary operation #2 operation screen (Heating·Cooling)

1
2 2

3 3

4 5

04. Operation state screen setting


It displays the control output volume (MV). The “ON” state is It displays the control output volume (MV) in humidity. When the

displayed in red and “OFF” state is displayed in dark grey. output volume part is touched in controlling the Heating·Cooling,
① ● Setting the state lamp in [13. System initial setting] in [Installation manual] output volume (MV), Heating volume (H.MV) and Cooling volume
● Setting up to 24 for lamp in [13. System initial setting] (C.MV) is displayed in turn.
It displays the control output volume (MV) in temperature. When the It displays the currently applied PID group number.
output volume part is touched in controlling the Heating·Cooling, ④ ● The applied PID group can be checked in

output volume (MV), Heating volume (H.MV) and Cooling volume [8. PID group] in [Installation manual]
(C.MV) is displayed in turn. ⑤ It displays the total process time of stationary operation.
27
(4) Stationary operation #3 operation screen
● The left of the screen is to display the measuring data, setting data of Temperature and Humidity.
● ( ) check box sets Y/N for data display.
● Press at the right middle to save the data of recording.
● Save the important graph files into the SD card.

● Refer to [3-2 Present value (PV) graph view]

● To download to SD card is not allowed while the measurement data, set data and output data of temperature&humidity is saved to internal memory.

[Fig. 4-9] Stationary operation #3 operation screen

2
[Fig. 4-10] X-AXIS setting screen of Stationary operation

It displays the measuring data, setting data of currently operated Temperature Set graph display direction

and Humidity. ③ : Vertical (Y-AXIS)
It displays the capacity of internal memory. : Horizontal (X-AXIS)

● About 90 days of saving is available when the sampling time is set in 1 second.
It is a button to save the measuring data, setting data of currently recorded

Temperature and Humidity into the memory (Internal memory, SD card).

28
(5) Termination screen for operation of stationary time setting
● The stationary operation is terminated while it shows the message, “The time setting operation is terminated.” when the operation is terminated by the elapse of setting time in
[5. Operation motion setting]
● The message is not appeared on the screen when it is forcibly terminated by pressing “Stop” button during operation.

● The message is disappeared by touching the corresponding part when the operation termination message is display in case of operation termination.

(It is same with the program operation termination.)

04. Operation state screen setting


[Fig. 4-11] Termination screen for operation of stationary time setting

29
4-2. Program operation
(1) Program operation still screen 1
● It converts to “Program operation still screen 1” when the operation state screen is selected in [Fig. 2-1 Main screen].
● Select the operation method with “Pattern” in [5. Operation motion setting].
● Refer to [6-1 Program pattern setting] for pattern setting method.

● It converts to [Fig. 4-12 Program operation #1 operation screen] when is touched by on the right bottom in [Fig. 4-14 Program operation #1 still screen]

[Fig. 4-12] Program operation #1 still screen

30
Parameter Setting range Unit Initial value
Pattern number 1~120 ABS 1

Cautions in operation
● It is not operated when the program is not input into the
pattern number on the screen.
● Refer to [6-1 Program pattern setting]

[Fig. 4-13] Screen for pattern number setting


input key to be operated

References
▶ When the button is touched by for inputting

04. Operation state screen setting


the pattern number setting to be operated, it is activated as
shown in [Fig. 4-13] Screen for pattern number setting input
key to be operated.
▶ Execute the program by selecting button when the input
for the pattern number setting to be operated is completed.

31
(2) Program operation #1 operation screen
● The pattern number cannot be set during operation.
● It is a screen for Measured data and Set data.

[Fig. 4-14] Program operation still screen 1


8 ⑦ Operation start time ~ End time display
1
It displays the current date/time and LCD backlight is off when it is touched.
9

● Red LED lamp at the right top is ON when the backlight is OFF in still state.
2
10 Moving to [Fig. 2-1 Main screen]
3 ⑨ ● It displays the key pad to input the password when main button restriction is set.

11 ● Refer to [Fig. 5-2 Screen in restriction setting of main button]

12 ⑩ Moving from current screen to next screen


Maintaining (Hold On) or Release (Hold Off) the currently operating
4 13 ⑪
temperature and humidity set value.
5 14
Terminating the currently processing segment and forced moving to the next

segment.
15
6 Execution or releasing the auto tuning with temperature set value (SP).
16 ⑬
● Y/N of the tuning button display is set in [8. PID group] in [Operation manual]
7
Execution or releasing the auto tuning with humidity set value (SP).

● Y/N of the tuning button display is set in [8. PID group] in [Operation manual]

It displays the current operation status.


① User button
● The arrow rotates to the clockwisely during operation
● Y/N for use in [13. System initial setting] in [Operation manual]
② It displays the temperature setting data (SP) to be controlled. ● User uses the wanted relay in [10. DO relay setting] in [Operation manual]

③ It displays the present temperature value (PV). ⑮ when the user button is used.
Ex) It is used for light the chamber.
④ It displays the present humidity value (PV). ● The set relay is operated when the “User” button is touched by

⑤ Full screen button in the stationary and program operation/stop screen.


⑥ It displays the humidity setting data (SP) to be controlled. ⑯ Program Operation/stop button

32
References
▶ It displays the direction of current pattern processing.
▶ , : Setting data>Measuring data is displayed in temperature increase.
▶ , : Setting data=Measuring data is displayed in temperature maintaining.
▶ , : Setting data<Measuring data is displayed in temperature decrease.

[Fig. 4-15] Program operation still screen 1 Full screen

04. Operation state screen setting


33
(3) Program operation #2 operation screen
● It is a screen to display the display lamps for measuring data, setting data and output volume.
● When the “Output volume” is touched in Heating·Cooling, output volume, Heating volume and Cooling volume are displayed in turn.

[Fig. 4-16] Program operation #2 operation screen (General) [Fig. 4-17] Program operation #2 operation screen (Heating·Cooling)

1
2 2

3 3
4 7
5 8
6 9

34
It displays the control output volume (MV). The “ON” state is It displays the partial repetition state.
displayed in red and “OFF” state is displayed in dark grey. ●The figure in the front of shows

① ● Setting the state lamp in [13. System initial setting] in [Installation manual] the frequency of repetition and the figure at the end shows the set
● Setting up to 24 for lamp in [13. System initial setting] repetition frequency.
It displays the control output volume (MV) in temperature. When the It displays the currently applied PID ground number.
output volume part is touched in controlling the Heating·Cooling, ⑦ ● The applied PID group can be checked in [8. PID group]

output volume (MV), Heating volume (H.MV) and Cooling volume in [Installation manual].
(C.MV) is displayed in turn. It displays the segment process time and setting time of currently
It displays the control output volume (MV) in humidity. When the processing segment.
output volume part is touched in controlling the Heating·Cooling, ⑧ ●The time in the front of

output volume (MV), Heating volume (H.MV) and Cooling volume shows the segment processing time and the time at the end
(C.MV) is displayed in turn. shows the set time in [6-1 Program pattern setting]
It displays the currently operated program pattern number and ⑨ It displays the total process time of program operation.

segment number.
It displays the pattern repetition state.
●The figure in the front in shows

the frequency of repetition and the figure at the end shows the set
repetition frequency.

04. Operation state screen setting


35
(4) Program operation #3 operation screen
● The left of the screen is to display the measuring data, setting data of Temperature and Humidity.
● ( ) check box sets Y/N for data display.
● Press at the right middle to save the data of recording.
● Save the important graph files into the SD card.

● Refer to [3-2 Present value (PV) graph view]

● To download to SD card is not allowed while the measurement data, set data and output data of temperature&humidity is saved to internal memory.

[Fig. 4-18] Program operation #3 operation screen

2
[Fig. 4-19] X-AXIS setting screen of Program operation

It displays the measuring data, setting data of currently operated Temperature Set graph display direction

and Humidity. ③ : Vertical (Y-AXIS)
It displays the capacity of internal memory. : Horizontal (X-AXIS)

● About 90 days of saving is available when the sampling time is set in 1 second.
It is a button to save the measuring data, setting data of currently recorded

Temperature and Humidity into the memory (Internal memory, SD card).

36
(5) Termination screen for operation of program
● The program operation is terminated while it shows the message, “The program operation is terminated.”
when the operation for segment setting range saved into the pattern is terminated.
● The message is not appeared on the screen when it is forcibly terminated by pressing “Stop” button during operation.

● The message is disappeared by touching the corresponding part when the operation termination message is display in case of operation termination.

(It is same with the stationary operation termination.)

04. Operation state screen setting


[Fig. 4-20] Termination screen for operation of program

37
(6) Other operation screen
● It is a screen to display warning in operation screen.

[Fig. 4-21] Screen for warning of shortage of memory capacity [Fig. 4-22] Warning screen for no extra space in memory

References References
▶ It is a screen when internal memory capacity is up to 60.8 megabytes. ▶ Screen for no extra space in intenal memory.

38
[Fig. 4-23] Warning screen for lack of saved file number [Fig. 4-24] Warning screen for full of saved file number

References References
▶ It is a screen when the number of saved files is over 240 in ▶ It is a screen when the number of saved files is fulled
internal memory. up to 256.

04. Operation state screen setting


39
4-3. Auto tuning and tuning point
● The segment is held during program operation and the segment is processed when the auto tuning is terminated.
● Auto tuning is a function to set the optimal PID integer automatically by measuring and calculating the object of control with controller.
● The controller generates the ON/OFF control output during "2.5 periods" during auto tuning and it calculates the P, I, D data automatically

based on the period and oscillation magnitude using the limit cycle to the object to be controlled.
● The Auto-Tuning is available in program.stationary operation.

The PID data calculated from the set value located PID group is saved automatically through Auto-Tuning from the currently set value.

Under tuning PD control

Cautions in operation
Measuring data
(PV)
●  ny change in set value (SP) in auto tuning does not change the tuning point.
A
And the tuning is started with changed set value (SP) for target set value (TSP)
50.2℃ after auto tuning termination.
SP = 50.0℃ ● The auto tuning is stopped in case of "Sensor short" in input during auto tuning.

49.8℃ At this time, the PID data is kept with the previous set value.
● When auto tuning is processed beyond 27 hours, the auto tuning is stopped.

● The PID set value can be changed during auto tuning, but the obtained PID data

from calculating in auto tuning termination is reset with the obtained PID data.
● The PID set value is maintained with previous set value when the auto tuning is

forcibly terminated.
OH
ON ON ON ● For better Auto-Tuning results, need to wait until stabilization by set

Control output value(temperature), then Autotuning the temperature first.


data (MV) And after Autotuning the humidity, will get better P, I and D value(data)
OFF OFF
OL

Auto tuning operation

References
▶ An example of auto tuning depending on the set value.
-O
 peration method : Stationary operation/ - Temperature auto tuning point : 0.10% → EUS 0.10% = 0.2℃ - Output lower limit (OL) : 0.0% / Output upper limit (OH) :
Sensor input : Temperature(PT_1) - Set value (SP) : 50.0℃ - Range : -50.00℃ ~ 150.00℃ 100.0% 50.2℃ 49.8℃ : Auto tuning point
40
05
Part

Operation motion setting


5-1 Operation method setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 42
5-2 Fuzzy operation‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 44
5-3 Setting value change rate (SLOPE) operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 45

41
05. Operation motion setting
It is a screen for general additional functions and additional setting in stationary operation.

5-1. Operation method setting


● It converts to the “Setting screen for operation related motion” when the operation motion setting is selected in [Fig. 2-1 Main screen]

[Fig. 5-1] Setting screen for operation related motion


Setting the recovery motion in black out
● STOP : A motion to return to the operation stop state after power on from

the black out.


1 5 ②
● COLD : An operation from the beginning after power on from the black out.

● HOT : A motion to return to the previous operation state after power on


from the black out.
2
In case that the external disturbances occure, stabilized the control.
● OFF : The indicated value is stabilized depending on time when the
3 6
③ overshoot is made.
● ON : The indicated value is stabilized faster than No operation case, as the
overshoot controlling function is provided.
4 Automatic increase or decreased with the set rate in case of set value change

● It is adopted in stationary operation only.

Total operation in set time and in [4-1(3) Stationary operation #2 operation


7 screen

The operation stops when the process time is coincided with the set time.
● It is adopted in stationary operation only.

Setting with selection either of pattern or stationary operation The key pad to input the password is displayed when the main button is
for operation mode.(It cannot be changed during operation.) touched by in the operation screen for setting the main button restriction
① ⑥
● PROG : Setting in program operation setting.
● FIX ● Refer to [Fig. 5-2 Screen for main button restriction setting]
: Setting in stationary operation

⑦ The parameter setting is impossible when button is touched by


● Screen rolling and key block releasing is possible.

42
Parameter Setting range Unit Initial value
OPERATION MODE PROG, FIX ABS FIX
POWER STOP MODE STOP, COLD, HOT ABS STOP
FUZZY SELECT OFF, ON ABS OFF
TEMP SLOPE T.EUS(0.00 ~ 100.00%) / MIN T.EUS / MIN T.EUS(0.00%) /MIN
HUMI SLOPE H.EUS(0.0 ~ 100.0%) / MIN H.EUS / MIN H.EUS(0.0%) /MIN
TIME OPERATION UNUSE, USE ABS UNUSE
HOUR 0 ~ 9999 HOUR ABS 0
MIN 0 ~ 59 MIN ABS 0
RESTRICT OF MAIN UNUSE, USE ABS UNUSE

Return motion Stationary


Program operation
in black out operation
STOP Program stop Stop
COLD Operation from the first segment Operation
Operation from the segment
HOT Operation
before black out

05. Operation motion setting


[Fig. 5-2] Main button restriction setting screen

43
5-2. Fuzzy operation
● T he overshoot may be taken place in case of severe change in operation or frequent changes in present value (SP).
More effective control can be made when the fussy function is operated at this time.
● Internal operation sequence of fuzzy function: It controls the overshoot by calculating the control output value (MV)

with sub target value (Super SP) instead of present value (PV) from the overshoot control start time.

Now present value(NPV)

Now present value(NPV)


Set value Set value SUPER SP
(SP) (SP) : Sub set value (Super SP)
Starting of internal calculation
SUPER SP
Overshoot control start
Time Time : Sub set value (Super SP)
SUPER SP Overshoot control start Start of application

44
5-3. Setting value change rate (SLOPE) operation
● The set value is changed by fixed changing rate from the now present value (PV) to set value when the set value is changed.

TSP
70.0℃ References
▶ Operation method: Stationary operation
PV SP 40.0℃ ▶ Temperature change rate: 20.0℃/Min
▶ Change [Changed SP(TSP) - PV at the SP changing point] with slope of 20.0℃ per minute:


30.0℃
(70.0-30.0)℃ = Change 40.0℃ with the slope of 20℃
▶ Increase the current set value (SP) from 30.0℃ to 70.0℃ with uniform increasing rate for


2 minutes.
SP change 2minutes.

Change rate (Slope) operation

05. Operation motion setting


45
06
Part

Program setting
6-1 Program pattern setting‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 49
6-2 Pattern repetition setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 54
6-3 File editing‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 56
6-4 Time signal operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 59
6-5 Standby operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 62
6-6 Experiment name setting‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 64

46
Program setting

[Fig. 6-2] Pattern editing screen [Fig. 6-3] Pattern and segment repetition [Fig. 6-4] File editing screen
setting screen

[Fig. 6-7] Time signal setting #1 [Fig. 6-10] Standby operation setting screen [Fig. 6-11] Experiment name setting screen

Part 06
47
06. Program setting
● It converts to [Fig. 6-1 Program setting screen] when the program setting button is touched by in [Fig. 2-1 Main screen].
● It is a screen group to set the parameters related to the program operation.

[Fig. 6-1] Program setting screen


1
① Moving to [Fig. 2-1 Main screen]

② Moving to the pattern editing screen


③ Moving to the screen for repetitive setting of pattern and segment
2 3 4 ④ Moving to the screen for setting pattern copy and deletion
⑤ Moving to the screen for time signal setting
⑥ Moving to the screen for setting the standby screen
⑦ Moving to the screen for experiment name
5 6 7

48
6-1. Program pattern setting
● It is a screen to set the segment depending on the pattern number.
● Refer to [6-4 Time signal operation] for time signal setting.
● Please refer to the [6-5 Standby operation] to find things about the standby operation.

[Fig. 6-2] Pattern editing screen


① Input the pattern number for segment setting
Setting the start condition for program operation.
● NOW PV
1


The set value (SP) is processed to the set time (TM1) from present value (PV)
to the present value (SP1) set in segment 1 (SEG1) regardless of the slope or
2 the start set value (SSP) in starting of the program operation.
(The remained time for operation is calculated by regarding of time passing
9 till the program operation starting through referring the patterns of set
program.)
10 ②
3 ● TEMP SSP


4 In case of starting the program operation, the present set value (SP) is started
5 from temperature SSP and progresses to the set time (TM1) for currently set
11 value 1 (SP1) in segment 1 (SEG1).
6
● HUMI SSP
7


12 It starts from the set SSP for present set value (SP) in starting of program
8 operation and progressed to the set time (TM1) for currently set value 1 (SP1)

06. Program setting


in segment 1 (SEG1).
③ It sets the temperature set value of segment to be operated.
④ It sets the humidity set value of segment to be operated.
⑤ Setting the time of segment to be operated.
It sets the time signal of the segment to be operated.
● 4 time signals can be set for each segment and each time signal

is set by selection from 18 types of time signal.
● Refer to [6-4 Time signal operation]

49
Selection of Y/N for using the Standby of Segment to operate.

● Refer to [6-5 Standby operation]
Select the soak operation of the temperature and humidity segment.

The selected segment operates soak without sloping.
When one of the buttons (Segment 01~99) is touched by
for segment insertion, the selected button (Segment 01~99)

and button are activated and the selected segment can be
inserted when is touched by

When one of the buttons (Segment 01~99) is touched by


for segment deletion, the selected button (Segment 01~99)

and button are activated and the selected segment can be
inserted when is touched by ▲ The input key to set the pattern number is displayed
when the “Pattern number” button is touched by.
⑪ Moving to left/right on the screen by 4 segment units.
⑫ Moving to [Fig. 6-1 Program setting screen] when is touched by

▲ When the “Starting condition” button is touched by, the input key to set the
starting condition is displayed
50
▲ It is a screen set with “TEMP SSP” for starting condition. ▲ , buttons are activated when
button is activated.

06. Program setting


▲ It is a screen set with “HUMI SSP” for starting condition. ▲ The input key to set the set value is displayed when
(TEMP) button is touched by.
51
▲ The input key to set the set value is displayed when ▲ The input key to set the time signal is displayed when
(HUMI) button is touched by. (Time signal) button is touched by.

▲ The input key to set the segment time is displayed when


▲ It is a screen set with T/H soak setting.
(Time) button is touched by.
52
Parameter Setting range Unit Initial value
PATTERN NO. 1~120 ABS 1
START CODE NOW PV, TEMP SSP, HUMI SSP ABS NOW PV
SEGMENT #n TEMP SP T.EU(0.00~100.00%) T.EU T.EU(0.00%)
SEGMENT #n HUMI SP H.EU(0.0~100.0%) H.EU H.EU(0.0%)
SEGMENT #n TIME -00.00.01(OFF) ~ 999.59.59(Hour, Min, Sec) ABS -00.00.01
SEGMENT #n TIME SIGNAL 1 ~ 4 0 ~ 17 ABS 0
SEGMENT #n WAIT USE UNUSE, USE ABS UNUSE
SEGMENT #n T/H SOAK SET UNUSE, USE ABS UNUSE

※ #n : 1 ~ 99

References
▶ Input by pressing for time signal and SEG alarm input.
▶ Select to escape from input screen.
▶ The wanted time signal group can be input by pressing
TS1~TS4 for the set value in [6-4 Time signal operation].

06. Program setting


53
6-2. Pattern repetition setting
● It is a screen to set the function for entire or partial repetition of set pattern.
● The operation method in pattern operation termination can be set.

[Fig. 6-3] Pattern and segment repetition Setting screen


① It sets the pattern number to perform the repetitive operation.
② It sets the repetition operation frequency of the set pattern.
1 7 It sets the pattern number for repetitive operation in termination of set pattern

operation.
④ It sets the segment to start the partial repetitive operation out of the set patterns.

2 It sets the segment to terminate the partial repetitive operation out of the set

8 patterns.
3
It sets the repetition frequency of the partial repetitive operation out of the set

patterns.
When operations to segment repeat, determines the start SP value.
4
⑦ • END SP : From the late operation END SP, SP value proceed.
5 • START SP : From the start SP of segment, the SP value proceed.
6
It decides the next operation when the set pattern operation is finished.
• RESET : The pattern termination signal is generated and the operation
state is in program stop.

• SEG HOLD : It is operated in last operation set value and hold state is
maintained.
• LINK RUN : The pattern set in the connection pattern is operated.

54
Parameter Setting range Unit Initial value
PATTERN NO. 1~120 ABS 1
SEG. REPEAT MODE END SP, START SP ABS END SP
REPEAT COUNT 0 (Indefinite repetition) ~ 999 ABS 1
LINK PATTERN 1~120 ABS 1
PATTERN END MODE RESET, SEG HOLD, LINK RUN ABS RESET
START SEGMENT NO. 1~4 0~99 ABS 0
END SEGMENT NO. 1~4 0~99 ABS 0
REPEAT COUNT NO. 1~4 0~99 ABS 0

Example)
SEG. Repeat Mode : END SP SEG. Repeat Mode : START SP
- From the late operation END SP, SP value proceed. - From the start SP of segment, the SP value proceed.
℃ ℃
30 30

SP SP
20 20

06. Program setting


10 10

0 0 SSP
(PV) (PV)
ESP
-10 -10

1 2 3 4 5 2 3 4 5 6 Seg 1 2 3 4 5 2 3 4 5 6 Seg
SEG. repeat section SEG. repeat section SEG. repeat section SEG. repeat section

55
6-3. File editing
● It is a screen to copy or delete the input segment values in [Fig. 6-1 Program pattern setting].
● The pattern number in operation cannot be deleted.
● The deleted pattern cannot be recovere.

[Fig. 6-4] File editing screen


① It sets the original pattern number to be copied.
It sets the first and last pattern number to be copied.

1 2 ● The first pattern is copied only when the last pattern is “0.”
6 It deletes the first and last pattern number to be copied.

● The first pattern is deleted only when the last pattern is “0.”.
It displays the total patterns set in [Fig. 6-1 Program pattern setting].
3 ④
7 ● The change is impossible as it is only for reading.
4
8 It displays the total segments set in [Fig. 6-1 Program pattern setting].
5 ⑤ ● The change is impossible as it is only for reading.
⑥ It copies the set pattern to the set pattern ②.
⑦ It initializes the set value of the pattern set in ③.
⑧ It initializes the set value of every pattern.

56
[Fig. 6-5] File editing screen #1 [Fig. 6-6] File editing screen #2

References References
▶ The message : “PARAMETER SETTING SET ERROR” ▶ The message : “THE SELECTED PATTERN IS EMPTY”
appeares at the bottom of screen for copying or deleting by wrong input of Appears at the bottom of screen for copying or deleting without saved data

06. Program setting


pattern number. in the pattern number.

57
Parameter Setting range Unit Initial value
PATTERN NO. 1~120 ABS 0
START PATTERN 0~120 ABS 0
COPY
END PATTERN 0~120 ABS 0
COPY UNUSE, USE ABS UNUSE
START PATTERN 0~120 ABS 0
SEL. DEL
END PATTERN 0~120 ABS 0
ALL. DEL UNUSE, USE ABS UNUSE

Message display Explanation


“There is no saved information in the selected pattern.” It is displayed in case of copy when there is nothing saved in pattern number.
“The copy is completed in the selected pattern.” It is displayed in completion of selected pattern copy.
“The deletion is completed in the selected pattern.” It is displayed in completion of selected pattern deletion.
“The copy is completed in every pattern.” It is displayed in completion of every pattern deletion.
“The pattern to be copies is being used.” It is displayed in use of pattern.

58
6-4. Time signal operation
● The time signal operation is classified into ON/OFF operation, time setting operation and the set time signal is used for setting the time signal No.
in segment setting in [6-1 Program pattern setting]
(1) Time signal ON/OFF operation

[Fig. 6-7] Time signal setting #1


The time signal is OFF during corresponding segment operation
① when “0” is selected.
● The change is impossible as it is only for reading.

1
The time signal is ON during corresponding segment operation
② when “1” is selected.
● The change is impossible as it is only for reading.

The time signal is “ON” after setting time is elapsed in delay time from
2 the corresponding segment starting point.
③ ● However, the time signal is not operated when the corresponding

segment time is bigger than delay time.


The time signal is “ON” by the delay time in corresponding segment
is “ON” only for set time in operation time. .
3 5 ④ ● However, the time signal is “ON” for the corresponding segment operation

4 when the (delay time+operation time) is bigger than corresponding


segment time and it is not operated in the next segment.

06. Program setting


⑤ Moving the screen up/down by 6 time signal units.

59
(2) Time signal time setting operation
● The time signal 2~17 (TS2~17) operate depending on delay time and operation time.

[Fig. 6-8] Time signal setting #2 [Fig. 6-9] Time signal setting #3

Parameter Setting range Unit Initial value


DELAY TIME 000.00.00(OFF)~999.59.59 (Hour, Min, Sec) ABS 000.00.00
OPER. TIME 000.00.00(OFF)~999.59.59 (Hour, Min, Sec) ABS 000.00.00

60
(3) Example of operation in time signal input
Setting Time signal operation
Operation time
ON

1. Delay time = 000.00.00 Time signal

OFF
Segment N time≥
Segment (n-1)segment time n segment time (n+1)segment time
Delay time +
Operation time
Operation time ON

Delay time
2. Delay time ≠ 000.00.00 Time signal

OFF
Segment (n-1)segment time n segment time (n+1)segment time
Operation time
ON
Segment N time<
3. Delay time = 000.00.00 Time signal
Delay time +

06. Program setting


Operation time OFF
Segment (n-1)segment time n segment time (n+1)segment time
☞ It does not make
Operation time
influence on the ON
next segment.
Time signal Delay time
4. Delay time ≠ 000.00.00

OFF

Segment (n-1)segment time n segment time (n+1)segment time


61
6-5. Standby operation
● It is a screen to set the temperature&humidity Stand-by Range and Stand-by Time during Program operation.
● The set standby operation here is applied to [Fig. 6-1 Program pattern setting].
● Definition of standby operation

- Coditions for Stand-by Operation Entry (“OR” condition) : In case of no entry of either of temperature or humidity into the set range for standby operation within the set segment time
- Conditions for Stand-by Operation Release (“AND” condition) : In case of entry of temperature and humidity into the set range for standby operation
- The standby time has indefinite value when the standby time is not set (Initial value).

[Fig. 6-10] Standby operation setting screen


① It sets Y/N of standby operation.
② It sets the temperature range to be applied for standby operation.
1 5 ③ It sets the humidity range to be applied for standby operation.
It sets the standby time to be applied when the indicated data for temperature or
humidity does not enter into the standby operation range (TEMP PV or HUMI PV).
④ ● It standbys indefinitely for entry to the standby operation range when the
2
standby operation time is set in “00.00.”
3
It decides either of “Entire” and “Maintain SEG” for standby operation method.
● Entire : The standby operation is applied to the set entire segment in
4 [6-1 Program pattern setting]
⑤ ● Maintain SEG : The standby operation is applied only to the set maintain range
segment in [6-1 Program pattern setting]

62
Parameter Setting range Unit Initial value
WAIT USE UNUSE, USE ABS UNUSE
TEMP ZONE T.EUS(0.00~100.00%) T.EUS T.EUS(0.00%)
HUMI ZONE H.EUS(0.0~100.0%) H.EUS H.EUS(0.0%)
WAIT TIME 00.00~99.59 (UNUSE, USE) ABS 00.00
WAIT USE METHOD ALL, SEG ABS ALL

Standby set time Standby set time

Set value Set value


Standby operation range Standby operation range

Standby operation range Standby operation range


Measuring data
Measuring data
Standby release timing
Standby operation
Process with segment (n+1) Standby release timing
Standby operation
(Time stop)
(Time stop)
Process with segment (n+1)
Segment (n+1) setting time
Segment (n+1)
Segment n Segment(n+1) Segment(n+2) Segment n Segment(n+1) setting time Segment(n+2)

06. Program setting


In case of standby operation release within In case of no entry of the measuring data into standby operation
standby time (Wait time) range within the standby time (Wait time)

References
▶ I t is a graph for interactive relation between standby operation and
standby time.
▶ S tandby operation range : It displays the humidity standby operation range
for temperature standby operation range for temperature and humidity
standby operation range for humidity.
63
6-6. Experiment name setting
● The experiment name can be set for each pattern. Refer to [4-2(2) Program operation #1 operation screen]

[Fig. 6-11] Experiment name setting screen

[Fig. 6-12] Experiment name setting screen

① Input the experiment name of each pattern.


References
② Converting to the next or previous experiment name screen.
▶ The input key to set the experiment name is displayed when
is touched by.

Parameter Setting range Unit Initial value


PATTERN NAME SET 1~120 0~9. A~Z, Special letter (Maximum 24 letters) ABS EXPERIMENT OF PATTERN 1~ 120

64
07
Part

Appointed operation setting


‥‥‥‥‥‥ 66

65
07. Appointed operation setting
● It converts to [Fig. 7-1 Time setting screen] when the appointed operation setting button is touched by in [Fig. 2-1 Main screen].
● It is a screen to set the current time and appointed operation time.

[Fig. 7-1] Time setting screen


It sets the year, month, day and hour.
① ● The current time is not changeable during recording the measured

1 2 data and operating.


② It sets the year, month, day and hour for appointed operation.
The operation is possible in the set appointed time when is
touched by.
3 ③ ● The appointed time is displayed on the operation screen as shown in
[Fig. 7-2 Operation appointed setting screen] when is
touched by.

66
[Fig. 7-2] Operation appointed setting screen (Program operation) [Fig. 7-3] Operation appointed setting screen (Stationary operation)

07. Appointed operation setting


67
Parameter Setting range Unit Initial value
YEAR 2000~2099 ABS -
MONTH 1~12 ABS -
DATE 1~31 ABS -
CURRENT TIME
AM/PM AM, PM ABS -
HOUR 1~12 ABS -
MIN 0~59 ABS -
YEAR 2000~2099 ABS 2012
MONTH 1~12 ABS 1
DATE 1~31 ABS 1
RESERVE TIME
AM/PM AM, PM ABS AM
HOUR 1~12 ABS 12
MIN 0~59 ABS 0
Appointment Click for appointment.

※ AM12:00 : Night 00:00 / PM12:00 : PM 12:00

68
08
Part

Screen display setting


8-1 Setting display setting‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 71
8-2 Touch screen calibration Seting‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 72
8-3 DI error creation history view‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 74

69
Setting display setting
Flow chart

[Fig. 8-1] Screen display setting screen [Fig. 8-8] DI error creation history screen

Part 08
70
08. Screen display setting
8-1. Screen display setting
● It converts to [Fig. 8-1 Screen display setting screen] when the screen display setting button is touched by in [Fig. 2-1 Main screen].

[Fig. 8-1] Screen display setting screen It sets the backlight electricity saving time.
① ● The electricity saving time sets the operation timing of backlight

OFF when there is not key operation.


1 4 It sets the humidity display pattern on operation screen.
In case of the below cases, the humidity data (PV) is displayed.
5
However, except below cases, “---.-“ is displayed.
2 ● Automatic

- In case that temperature set data (SP) in in the range of relative humidity


6 display condition(DRY LIMIT)
3 ②
- In case that temperature indicated data (PV). Is in 0.0℃ to 100.0℃ range
- In case that humidity set data (SP) is not set to 0.0%
● Manual
7
- The humidity is displayed depending on the set value (SP) and indicated


value(PV) as in the automatic mode and the humidity is displayed even when

08. Setting display setting


the humidity set value (SP) is set to 0.0%.
③ The brightness of LCD is controlled by , button.

Total capacity of internal memory, used capacity, total files to be


saved and display of saved files
④ ● Warning : set to the usage of warning (shortage of memory capacity, excess of

saved files) using the button ,
⑤ Moving from current screen to next screen
⑥ Touch screen calibration
⑦ It deletes every file saved in the internal memory.
71
Parameter Setting range Unit Initial value
OFF TIME O~99 MIN ABS 10
HUMIDITY DISPLAY AUTO, MANUAL ABS AUTO
LED BRIGHTNESS 1~8 ABS 8 steps
TOUCH SCREEN CALIBRATION When you calibrate the touch screen, click.

8-2. Touch screen calibration Seting


● Press the red dot at the left / right upper, left / right bottom and center of the touch screen calibration screen, you can calibrate the touch screen.
● Press on the touch screen calibration screen is not touch screen calibration is stored and stops will move to [Figure 8-1 Screen display setting screen]

[Fig.8-2] Touch screen calibration #1 [Fig.8-3] Touch screen calibration #2

72
[Fig.8-4] Touch screen calibration #3 [Fig.8-5] Touch screen calibration #4

08. Setting display setting


[Fig.8-6] Touch screen calibration #5 [Fig.8-7] Touch screen calibration #6
73
8-3. DI error creation history view
● It is a screen to display the type, date and time of error created DI.
● The error history is saved up to 30 cases and the later history is saved after deletion of the saved history.

[Fig. 8-8] DI error creation history screen


It displays the history of DI error creation.
① ● The name set in [11-2 Error name] in [Operation manual] is displayed.
1 ● The change is impossible as it is only for reading.
② It deletes the entire DI error creation.
③ It checks the previous or next error history.

2
Parameter Setting range Unit Initial value
ALL CLR UNUSE, USE ABS UNUSE

74
[Fig. 8-9] Screen for DI error display method by letter [Fig. 8-10] Screen for DI error display method by photo

References
▶ It is a screen in case of DI error creation.
▶ The setting for letter and photo screen setting can be set in [11. DI function and operation setting] in [Installation manual].

08. Setting display setting


▶ It is converted to the operation screen after escaping from the DI error screen when is touched by.
▶ The same DI error creation is neglected for 1 minute when the screen is changed by pressing button after DI creation. (Here, the neglecting means the DI error screen.)
Ex) It neglects even DI1 is created by escaping with“Return” in the stat of DI1 creation and the DI error screen is displayed when DI1 has been created even after 1 minute.
▶ button is to block the alarming sound when DI error is created.
EX) Explanation depending on lamp state
- DI error no creation (“OFF” state) ( letter, photo)
- DI error creation (“ON” state) ( letter, photo)
- Release after DI error creation (“ON” state after “OFF” ) ( letter, photo)

75
09
Part

Communication error ‥‥‥‥‥‥‥ 77

76
09. Communication error

[Fig. 9-1] Screen for DI error display method by letter [Fig. 9-2] Screen for DI error display method by photo

09. Communication error


References
▶ When there is an error between display and control unit The message, The control part is

not connected. is displayed at the bottom of the screen as shown in [Fig. 9-1] Control unit
communication error screen.
▶ When there is an error between control unit and I/O board communication The message,

The I/O board is not connected.” is displayed at the bottom of the screen as shown in
[Fig. 9-2] I/O board communication error screen.
▶ Communication failure : Communication cable defect Communication cable

connection defect

77
Engineering Units - EU, EUS
When the sensor type (IN-T) or the upper limit.lower limit of input range is changed, the parameters expressed in EU( ), EUS( ) are changed in
proportion to current data. (However, the upper and lower range setting data is initialized.)
Download the instruction manual and communication manual from the homepage.
EU( ) : Value of engineering unit depending on the range of instrument
EUS( ) : Value of engineering unit depending on the span of instrument

▶ Range of EU( ) and EUS( )

Range Center point


EU(0 ~ 100%) RL ~ RH | RH - RL |/2 + RL
EU(-100 ~ 100%) -( | RH - RL | + | RL | ) ~ RH RL
EUS(0 ~ 100%) 0 ~ | RH - RL | | RH - RL |/2
EUS(-100 ~ 100%) - | RH - RL | ~ | RH - RL | 0

(Example)
▶ INPUT = PT_1
▶ RANGE = -90.00℃(RL) ~ 200.00℃(RH)

Range Center point


EU(0 ~ 100%) - 90.00 ~ 200.00℃ 55.00℃
EU(-100 ~ 100%) - 380.00 ~ 200.00℃ - 90.00℃
EUS(0 ~ 100%) 0 ~ 290.00℃ 145.00℃
EUS(-100 ~ 100%) - 290.00 ~ 290.00℃ 0.00℃

RL: Lower limit of input range


RL: Upper limit of input range
78