28. HDSD - Tủ Môi Trường Mẫu Kiểm - VS-5TH-150G (2022) (E) - Tieng Anh
28. HDSD - Tủ Môi Trường Mẫu Kiểm - VS-5TH-150G (2022) (E) - Tieng Anh
Fragile Glass
Earth Ground
Caution
Low Temperature
◈ Index ............................................................................................ 3
◈ Caution ......................................................................................... 4
◈ Feature .................................................................................. 5
◈ Specification......................................................................................... 6
◈ Installation .................................................................................... 7
◈ Composition.......................................................................................... 9
- Main Body......................................................... 9
◈ Operation.......................................................................................... 11
- Operation.......................................... 11
◈ Maintenance ...................................................................................... 12
- Diagram ...........................................15
◈ Warranty............................................................................................ 17
- 3 -
Caution VS-5TH-150G Series
- 4 -
Feature VS-5TH-150G Series
▶ Feature
VS-5TH Series are suitable for experiments for temperature and humidity test of various samples. Cooling system can be reached to around set
value of Temperature and humidity by PID digital control system quickly with high efficiency. After they are reached on set value, temperature
and humidity deviation is minimized by stable power supply and kept excellent safety for temperature and humidity.
- 5 -
Specification VS-5TH-150G Series
▶ Specification
Display Programmable Controller with 5.7Inch LCD Color Touch Screen Interface
Number of
2/8 2/10 2/12
Shelf Standard / Max
Safety Device Electrical Leakage Breaker, Over Temp. Cut-Off, Water Level Sensor (Low Water Level)
Main Power 230VAC 50Hz, 1 phase 3 wires Option) 380V 50Hz, 3 Phase 4 wires
* Before use, check the electrical specifications and use the correct voltage.
Primary facilities for electrical supply should be ready in advance. (Please contact the user's building facilities team in advance.
- 6 -
Installation VS-5TH-150G Series
▶ Installation
① Left and Right Space ②Front and Back Space
0.5m
Wall
Wall
0.5m
0.3m
front
Wall
* Make secure minimum 0.3m for front and Back Space and 0.5m for left and side space from wall for maintenance and cooling air & hot air
circulating.
- 7 -
Installation VS-5TH-150G Series
4) Because dust is attached to samples in places where there is a lot of dust in the surrounding environment, it is difficult to achieve accurate
results, so it is necessary to keep the surrounding environment clean at all times.
5) Avoid direct sunlight because painting of equipment may be discolored if exposed to direct sunlight for a long time.Do not load objects onto
the top of the machine.
6) Make sure that the voltage is 220V single phase (three wire with ground) or three phase (four wire with ground) before connecting power.
(Caution is required for mobile installation.)
7) If the power is three-phase, check the rotation direction of the fan motor around evaporator after connecting the machine .
If fan is rotated conversely, exchange position of R phase and S phase of the power line each other. If they are connected well, heat air of
the unit is out to rear side of the unit. (Occurred Hot air from the unit must be out to rear side)
8) Voluntary Modification of the machine or operation circuit may cause malfunction, so change is prohibited.
(Be careful for Electric shock of 220V)
9) Connecting the power of the device
Connect to the pre-prepared power line (see product specifications for the relevant model rating).
- 8 -
Composition VS-5TH-150G Series
▶ Composition(Main Body
(Main Body)
Ⓛ
Description Function
① Door Door
①
② Window Viewer Window viewer can be checked inside chamber
⑤
Water tank for humidity control
[Main Body] ⑦ Water Tank
(* Distilled water must be used)
[Rear Side]
- 9 -
Composition VS-5TH-150G Series
▶ Composition(Control Panel)
NO Description Function
ⓐ TOUCH SCREEN All test condition can be set on display.
FAN Switch & Lamp
ⓑ FAN
-Air Circulate Fan in the chamber
ⓒ T-HEATER Temperature Heater Switch & Lamp
ⓓ H-HEATER Humidity Heater Switch & Lamp
ⓐ
Cooler Switch & Lamp
ⓔ COOLER
- Compressor (Cooling System)
Lamp Switch & Lamp
ⓕ LAMP
-When you check the sample in the chamber, on door lamp
EMERGENCY S/W : When there is a
emergency situation, press this
ⓖ EMERGENCY STOP button immediately. If Emergency
situation is over, dial this
▶ Operation
When water is filled in water tank, over 1st distilled water is recommended.
If tap water or unclean water is used, unit might not be worked well.
1) Connect Main electric power cable.
2) Fill clean water or over 1st distilled water into the water tank
3) Turn on the main circuit breaker on Control panel.
4) Turn on the switch for ⓑFAN, ⓒT-HEATER, ⓓH-HEATER and ⓔ COOLER on control panel.
5) Set Temperature, humidity and other test condition on touch screen.
Refer to guide book for specification and operation on end of this manual.
6) Touch ‘RUN” button on Controller.
7) Check temperature and humidity after running.
8) if unit has something wrong and trouble, contact to our service team
▶ Caution
- The main voltage must correspond to the voltage given on the name-plate
(AC 220V, 50/60 Hz, Single Phase or 380V 50/60Hz, Three- Phase )
- Install the unit even level with the ground.
- Avoid vibration place.
- Connect and disconnect power, do not touch the cable, hold the plug
- Be careful drop liquid or water into the controller and unit .
- Explosive alerts you to possibility of explosion by high pressure
- Make clean for maintenance
- This equipment can make high temperature and hot,
- Must use protect hand gloves .When you open the door, be careful hot air.
Must Take protector.
- When unit is running, do not open the door
- Make Space around equipment over 50 cm for Left , right, back.
- Do not impact or shock to the equipment.
- Some parts made by ceramic so easy broken.
- Check electric Ampere and connect.
- 11 -
Maintenance VS-5TH-150G Series
- Check 1 : Check the setting value Temp. & Humi. on Controller H-OHC
- Check 2 : T-OTC / H-OHC setting value.
(There are safety device(T-OTC and H-OHC) on right side of panel
If present temperature in chamber is raised than over setting value, T-OTC is worked.
Check about shown error message on control panel. Check T-OTC after opening side case of the unit.
Turn off the main Power. If fan is rotated conversely, temperature might be raised . Contact to distributor or service team you bought this
unit. After problem is solved, set T-OTC again.
(Depends on your setting value, you should set higher 30~50’C on T-OTC and 20~30% H-OHC. These range is up to your test)
If H-OHC is worked, check about water level in water tank. If water supply is not working even though water is enough in water
tank, H-OHC would be worked. Contact to distributor or service team you bought this unit.
- 12 -
Maintenance VS-5TH-150G Series
- Check3 : Check Voltage Terminal and Float Switch. (They are in right side of the unit)
Lamp
Voltage
Terminal
220VAC
When Lamp is on, DC is applied to the input terminals. The voltage terminal output is energized without contact point. Voltage is applied to heater.
When heater output on controller is "0%", Lamp is blink. But, lamp might be turned on when it is break down.
If Heater output on operating mode is "0%", test AC voltage by using tester.
5) Buzzer sound
If there are trouble on the unit, power of heater and compressor is off with buzzer sound.
After checking what is the trouble, solve the problem.
6) Humidity error
During the humidity test, the temperature is normally controlled, but sometimes the humidity is shown 100%.
It can be occurred this trouble when water is not enough in the chamber, check if the water is supplied.
If this trouble is happened even though there is water enough in chamber, the part of the humidity sensor inside the chamber
might be problem.
We strictly forbid any repair or exchange of parts by individuals except the repair by authorized engineer.
The product may be damaged or do not function properly. Furthermore, there is a danger of an electric shock or a fire.
If there are some problem on the unit, contact to local agency or us immediately.
1. When you apply after sales service, let us know below point.
- Model name
- Purchase date and year (There are mentioned on barcode of unit)
- Trouble contents in detail with some photos.
- Let us know about organization name, user name, contact number
※ If you have any question, contact to local service team or our service center.
VISION SCIENTIFIC CO.,LTD
Address : 260, Techno-2 Ro, Yuseong-Gu Daejeon-Si, Korea.
Tel : + 82 32 672 7779 Fax: +82 32 676 3977
E-mail: [email protected]
- 14 -
Diagram VS-5TH-150G Series
▶ Diagram
VS-5TH Series
- 15 -
Spare Parts VS-5TH-150G Series
▶ Spare Parts
3 Compressor 1
5 Humidifier Heater 1
11 Transformer 1
- Contact to seller on your country or us if you want to exchange or repair the parts separately.
- We strictly forbid any repair or exchange of parts by individuals except the repair by authorized engineer.
- Related Spare parts can be changed depends on situation of parts supply.
-
- 16 -
Warranty
VARIOUS It is temperature and humidity programmable controller which equips with the general control,
heating and cooling function by supporting high definition TFT-LCD touch screen and SD card.
This manual is commonly used for TEMI2000 Series.
Queries related with after sales service for TEMI2000 series
Please inform the TEMI2000 model name, failure condition and
Contents
contact point for queries of after sales service.
T :and
01. Operation 82-32-326-9120
setting ‥‥‥‥‥‥‥‥‥‥‥‥‥ 4 06. Program setting ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 48
‥
‥
1-1. Basic operation flow chart‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 4 6-1. Program pattern setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥49
F : 82-32-326-9119
1-2. Setting button operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 5 6-2. Pattern repetition setting‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥54
1-3. Parameter setting method ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 6 6-3. File editing‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥56
6-4. Time signal operation‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥59
02. Main screen ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 9
‥
6-5. Standby operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥62
3-1. Pattern graph display‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥12
Customer contact for TEMI2000 series 6-6. Experiment name setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥64
03. Setting graph display and save ‥‥‥‥‥‥‥ 12
07. Appointed operation setting ‥‥‥‥‥‥‥‥‥ 66
‥
Quotation request
graph/view‥
Product request
‥
3-2. Presented value (PV) ‥‥‥‥‥‥‥‥‥‥‥‥‥‥15
Specification
3-3. Presented value (PV) request / Data
graph saving request/ Other request
setting‥‥‥‥‥‥‥‥‥‥‥18 08. Screen display setting ‥‥‥‥‥‥‥‥‥‥‥‥ 71
‥
3-4. Memory save setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥20 8-1. Screen display setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥71
■ Internet
www.samwontech.com
4-1. Stationary operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥23 8-2. Touch screen calibration Seting‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥72
8-3. DI error creation history view‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥74
04. Operation state screen setting ‥‥‥‥‥‥‥‥ 23
‥
■ Eoperation‥
4-2. Program mail ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥30 09. Communication error ‥‥‥‥‥‥‥‥‥‥‥‥ 77
‥
[email protected]
4-3. Auto tuning and tuning point‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥40
05. [email protected]
motion setting ‥‥‥‥‥‥‥‥‥‥‥ 42
‥
5-1. Operation method setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥42
5-2. Fuzzy operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥44
5-3. Setting value change rate (SLOPE) operation‥‥‥‥‥‥‥‥‥‥45
2
01
Part
3
01. Operation and setting
This product is programmable controller designed with dialogue style touch screen easy-to-use for the users.
4
1-2. Setting button operation
The “Pattern No.” is touched in program still screen and it is used for setting the pattern number wanted by the user.
Input key for setting the pattern experiment name and DI error name ▶
Refer to [11. DI function and operation setting] in [Operation manual] for DI error name input key.
▲ Input key for time signal setting ▲ Display when it is out of the setting range
6
(1) Method for effectiveness of setting button and setting value
● This product is designed as follows when the setting data input button is touched or to check the effectiveness of the input setting data by sound.
● “Beep” : When the basic setting button is touched or the setting data is input normally
● “Beep and beep” : When the input data by the setting data input key is out of the input range.
● Do not press with sharp thing (Pencil and etc) or excessive force on the input key for basic setting button or setting value.
It may cause the mal operation of the device or damage on the touch panel.
7
02
Part
8
02. Main screen
1 2 3
4 5 6
operation method when it is set for no use.
④ PROGRAM SET Moving to the program setting menu screen Set the language to use
Moving to the screen for setting current time, ⑧ LANGUAGE SET (English, Korean, Simplified, Traditional, Japanese,
⑤ RESERVE SET
Russian languages available)
programmed operation time.
9
03
Part
10
Setting graph display and save
Flow chart
[Fig. 3-2] Graph & Saving screen 1 (Pattern graph display) [Fig. 3-5 Graph & Saving screen 2 (Graph display selection)
[Fig. 3-12] Graph & Saving screen 4 [Fig. 3-9] Graph & Saving screen 3
Part 03
11
03. Setting graph display and save
3-1. Pattern graph display
● It converts to [Fig. 3-2 Graph & Save screen 1 (Pattern graph display)] when [Graph & Save] is selected at the left top in [Fig. 2-1 Main screen]
● This screen displays the temperature and humidity operation pattern and progress time in program operation.
● It is a screen to display the input pattern in [Fig. 6-2 Pattern editing screen].
● Can change the status of Temperature·Humidity Check, and , even in the graph screen.
[Fig. 3-1] Graph & Save screen 1 (Program operation) [Fig. 3-2] Graph & Save screen 1 (Pattern graph display)
7
2 3 1
4 5
12
Display the graph setting menu [Fig. 3-3] Graph & Save screen 1 (Program operation)
① ● D
isplay the parameters related to the pattern graph operation at the bottom
depending on ON/OFF operation on the graph menu button
The temperature (Red) is disappeared on the graph screen when the checked 2
button is touched in the currently operating screen and temperature (Red) is
②
displayed in the graph screen when button is touched again.
● , buttons are used even in during operation(RUN) .
1
The humidity (Blue) is disappeared on the graph screen when the checked
button is touched in the currently operating screen and humidity (Blue) is
③
displayed in the graph screen when button is touched again.
● , buttons are used even in during operation(RUN) . 3
13
[Fig. 3-4] Pattern No. input screen
References
▶ It is a screen to input the pattern No. to be displayed in graph.
▶ The pattern No. can be input even during operation.
14
3-2. Presented value (PV) graph view
● It is a screen to display the data recorded in [4-1(4) Stationary operation 3 operation screen] and [4-2(4) Program operation 3 operation screen].
● The date and time saved into the memory are displayed at the top of the screen. [Reference 1]
⑨ The time axis is expanded or reduced.
line moves.
② Key lock / SD card / Storage indicator
③ : Display to update the graph screen immediately which is being saved. ⑬ Copy the recorded PV files into the internal memory to SD card.
15
[Fig. 3-6] Graph&Save setting screen 2 (Graph display is not selected) [Fig 3-11] When setting the X-AXIS of Trend direction, Graph Screen
References
▶ I t is a screen for no selection from the temperature
(Set-data, Indicated-Data) and humidity (Set-data, Indicated-Data)
▶ It is a screen to display the saved file into the internal memory.
▶ Refer to [4-1(4) Stationary operation 3 operation screen] and [4-2(4) Program
operation 3 operation screen] for saving into internal memory.
16
[Fig. 3-7] Graph&Save setting screen 2 (Saved folder is displayed) [Fig. 3-8] Graph&Save setting screen 2 (Saved file is displayed)
5
2
17
3-3. Presented value (PV) graph saving setting
● T his screen is to set the display range and sampling time which are necessary for graph recording in [4-1(4) Stationary operation 3 operation screen] and [4-2(4) Program operation 3
operation screen].
[Fig. 3-9] Graph&Save setting screen 3
Setting the PV graph saving period
● It is not changeable during PV graph saving
① ● Saving about 31 days is possible when sampling time is set in 1 minute in
5 screen 3 manually
Setting the media for data saving
3 ● MEM : Stored in the internal memory
③ ● SD : Stored in the SD card
● BOTH : Simultaneously stored in the internal memory and SD card
18
Parameter Setting range Unit Initial value
SAMPLING TIME 1 SEC, 2 SEC, 5 SEC, 10 SEC, 20 SEC, 30 SEC, 1 MIN ABS 1 SEC
RECORDING OPER. AUTO, MANUAL ABS AUTO
RECORE MEDIA MEM, SD, BOTH ABS BOTH
TEMP GRAPH DISPLAY HIGH T.EU (-2.50 ~ 102.50%) T.EU T.EU(100.00%)
TEMP GRAPH DISPLAY LOW (Temp graph display lower limit < Temp graph display upper limit) T.EU T.EU(0.00%)
HUMI GRAPH DISPLAY HIGH H.EU (0.0 ~ 100.0%) H.EU H.EU(100.0%)
HUMI GRAPH DISPLAY LOW (Humi graph display lower limit < Humi graph display upper limit) H.EU H.EU(0.0%)
19
3-4. Memory save setting
● It is a screen to set the transmitting of pattern and parameter to SD card.
● It is a screen displayed in SD card option only.
20
04
Part
21
Operation state screen setting
Flow chart
[Fig. 4-1] Stationary operation still screen 1 [Fig. 4-5] Stationary operation still screen 1 [Fig. 4-7] Stationary operation still screen 2 (General)
[Fig. 4-9] Stationary operation still screen 3 [Fig. 4-8] Stationary operation still screen 2 (Heating·Cooling)
Part 04
22
04. Operation state screen setting
4-1. Stationary operation
(1) Stationary operation still screen 1
● When the operation state screen is selected in [Fig. 2-1 Main Screen], it is converted to “Stationary operation still screen 1.”
● Select the operation method with “Stationary” in [5. Operation motion setting]
● When at the right bottom of [Fig. 4-1] Stationary operation still screen 1 is touched by, it converts to [Fig. 4-5] Stationary operation still screen 1.
● If parameter isn't set due to the noise, it is displayed as [Fig. 4-2] Parameter setting error screen due to noise.
In this case, Please use to establish a noise countermeasure about power and communications cable.
23
[Fig. 4-3] Screen of Input for temperature setting about operation [Fig. 4-4] Screen of Input for humidity setting about operation
temperature
24
(2) Stationary operation #1 operation screen
● When the “Setting data(SP)” is touched even in operation, the input key setting for operation is activated.
● It is a screen for Measured data(PV) and Set data(SP).
25
References
▶ , : Setting data > Measuring data is displayed in TEMP and HUMI increase.
▶ , : Setting data = Measuring data is displayed in TEMP maintaining.
▶ , : Setting data < Measuring data is displayed in TEMP decrease.
26
(3) Stationary operation #2 operation screen
● It is a screen to display the display lamps for measuring data, setting data and output volume.
● When the “Output volume” is touched in Heating·Cooling, output volume, Heating volume and Cooling volume are displayed in turn.
[Fig. 4-7] Stationary operation #2 operation screen (General) [Fig. 4-8] Stationary operation #2 operation screen (Heating·Cooling)
1
2 2
3 3
4 5
● To download to SD card is not allowed while the measurement data, set data and output data of temperature&humidity is saved to internal memory.
2
[Fig. 4-10] X-AXIS setting screen of Stationary operation
It displays the measuring data, setting data of currently operated Temperature Set graph display direction
①
and Humidity. ③ : Vertical (Y-AXIS)
It displays the capacity of internal memory. : Horizontal (X-AXIS)
②
● About 90 days of saving is available when the sampling time is set in 1 second.
It is a button to save the measuring data, setting data of currently recorded
④
Temperature and Humidity into the memory (Internal memory, SD card).
28
(5) Termination screen for operation of stationary time setting
● The stationary operation is terminated while it shows the message, “The time setting operation is terminated.” when the operation is terminated by the elapse of setting time in
[5. Operation motion setting]
● The message is not appeared on the screen when it is forcibly terminated by pressing “Stop” button during operation.
● The message is disappeared by touching the corresponding part when the operation termination message is display in case of operation termination.
29
4-2. Program operation
(1) Program operation still screen 1
● It converts to “Program operation still screen 1” when the operation state screen is selected in [Fig. 2-1 Main screen].
● Select the operation method with “Pattern” in [5. Operation motion setting].
● Refer to [6-1 Program pattern setting] for pattern setting method.
● It converts to [Fig. 4-12 Program operation #1 operation screen] when is touched by on the right bottom in [Fig. 4-14 Program operation #1 still screen]
30
Parameter Setting range Unit Initial value
Pattern number 1~120 ABS 1
Cautions in operation
● It is not operated when the program is not input into the
pattern number on the screen.
● Refer to [6-1 Program pattern setting]
References
▶ When the button is touched by for inputting
31
(2) Program operation #1 operation screen
● The pattern number cannot be set during operation.
● It is a screen for Measured data and Set data.
③ It displays the present temperature value (PV). ⑮ when the user button is used.
Ex) It is used for light the chamber.
④ It displays the present humidity value (PV). ● The set relay is operated when the “User” button is touched by
32
References
▶ It displays the direction of current pattern processing.
▶ , : Setting data>Measuring data is displayed in temperature increase.
▶ , : Setting data=Measuring data is displayed in temperature maintaining.
▶ , : Setting data<Measuring data is displayed in temperature decrease.
[Fig. 4-16] Program operation #2 operation screen (General) [Fig. 4-17] Program operation #2 operation screen (Heating·Cooling)
1
2 2
3 3
4 7
5 8
6 9
34
It displays the control output volume (MV). The “ON” state is It displays the partial repetition state.
displayed in red and “OFF” state is displayed in dark grey. ●The figure in the front of shows
⑥
① ● Setting the state lamp in [13. System initial setting] in [Installation manual] the frequency of repetition and the figure at the end shows the set
● Setting up to 24 for lamp in [13. System initial setting] repetition frequency.
It displays the control output volume (MV) in temperature. When the It displays the currently applied PID ground number.
output volume part is touched in controlling the Heating·Cooling, ⑦ ● The applied PID group can be checked in [8. PID group]
②
output volume (MV), Heating volume (H.MV) and Cooling volume in [Installation manual].
(C.MV) is displayed in turn. It displays the segment process time and setting time of currently
It displays the control output volume (MV) in humidity. When the processing segment.
output volume part is touched in controlling the Heating·Cooling, ⑧ ●The time in the front of
③
output volume (MV), Heating volume (H.MV) and Cooling volume shows the segment processing time and the time at the end
(C.MV) is displayed in turn. shows the set time in [6-1 Program pattern setting]
It displays the currently operated program pattern number and ⑨ It displays the total process time of program operation.
④
segment number.
It displays the pattern repetition state.
●The figure in the front in shows
⑤
the frequency of repetition and the figure at the end shows the set
repetition frequency.
● To download to SD card is not allowed while the measurement data, set data and output data of temperature&humidity is saved to internal memory.
2
[Fig. 4-19] X-AXIS setting screen of Program operation
It displays the measuring data, setting data of currently operated Temperature Set graph display direction
①
and Humidity. ③ : Vertical (Y-AXIS)
It displays the capacity of internal memory. : Horizontal (X-AXIS)
②
● About 90 days of saving is available when the sampling time is set in 1 second.
It is a button to save the measuring data, setting data of currently recorded
④
Temperature and Humidity into the memory (Internal memory, SD card).
36
(5) Termination screen for operation of program
● The program operation is terminated while it shows the message, “The program operation is terminated.”
when the operation for segment setting range saved into the pattern is terminated.
● The message is not appeared on the screen when it is forcibly terminated by pressing “Stop” button during operation.
● The message is disappeared by touching the corresponding part when the operation termination message is display in case of operation termination.
37
(6) Other operation screen
● It is a screen to display warning in operation screen.
[Fig. 4-21] Screen for warning of shortage of memory capacity [Fig. 4-22] Warning screen for no extra space in memory
References References
▶ It is a screen when internal memory capacity is up to 60.8 megabytes. ▶ Screen for no extra space in intenal memory.
38
[Fig. 4-23] Warning screen for lack of saved file number [Fig. 4-24] Warning screen for full of saved file number
References References
▶ It is a screen when the number of saved files is over 240 in ▶ It is a screen when the number of saved files is fulled
internal memory. up to 256.
based on the period and oscillation magnitude using the limit cycle to the object to be controlled.
● The Auto-Tuning is available in program.stationary operation.
The PID data calculated from the set value located PID group is saved automatically through Auto-Tuning from the currently set value.
Cautions in operation
Measuring data
(PV)
● ny change in set value (SP) in auto tuning does not change the tuning point.
A
And the tuning is started with changed set value (SP) for target set value (TSP)
50.2℃ after auto tuning termination.
SP = 50.0℃ ● The auto tuning is stopped in case of "Sensor short" in input during auto tuning.
49.8℃ At this time, the PID data is kept with the previous set value.
● When auto tuning is processed beyond 27 hours, the auto tuning is stopped.
● The PID set value can be changed during auto tuning, but the obtained PID data
from calculating in auto tuning termination is reset with the obtained PID data.
● The PID set value is maintained with previous set value when the auto tuning is
forcibly terminated.
OH
ON ON ON ● For better Auto-Tuning results, need to wait until stabilization by set
References
▶ An example of auto tuning depending on the set value.
-O
peration method : Stationary operation/ - Temperature auto tuning point : 0.10% → EUS 0.10% = 0.2℃ - Output lower limit (OL) : 0.0% / Output upper limit (OH) :
Sensor input : Temperature(PT_1) - Set value (SP) : 50.0℃ - Range : -50.00℃ ~ 150.00℃ 100.0% 50.2℃ 49.8℃ : Auto tuning point
40
05
Part
41
05. Operation motion setting
It is a screen for general additional functions and additional setting in stationary operation.
Setting with selection either of pattern or stationary operation The key pad to input the password is displayed when the main button is
for operation mode.(It cannot be changed during operation.) touched by in the operation screen for setting the main button restriction
① ⑥
● PROG : Setting in program operation setting.
● FIX ● Refer to [Fig. 5-2 Screen for main button restriction setting]
: Setting in stationary operation
42
Parameter Setting range Unit Initial value
OPERATION MODE PROG, FIX ABS FIX
POWER STOP MODE STOP, COLD, HOT ABS STOP
FUZZY SELECT OFF, ON ABS OFF
TEMP SLOPE T.EUS(0.00 ~ 100.00%) / MIN T.EUS / MIN T.EUS(0.00%) /MIN
HUMI SLOPE H.EUS(0.0 ~ 100.0%) / MIN H.EUS / MIN H.EUS(0.0%) /MIN
TIME OPERATION UNUSE, USE ABS UNUSE
HOUR 0 ~ 9999 HOUR ABS 0
MIN 0 ~ 59 MIN ABS 0
RESTRICT OF MAIN UNUSE, USE ABS UNUSE
43
5-2. Fuzzy operation
● T he overshoot may be taken place in case of severe change in operation or frequent changes in present value (SP).
More effective control can be made when the fussy function is operated at this time.
● Internal operation sequence of fuzzy function: It controls the overshoot by calculating the control output value (MV)
with sub target value (Super SP) instead of present value (PV) from the overshoot control start time.
44
5-3. Setting value change rate (SLOPE) operation
● The set value is changed by fixed changing rate from the now present value (PV) to set value when the set value is changed.
TSP
70.0℃ References
▶ Operation method: Stationary operation
PV SP 40.0℃ ▶ Temperature change rate: 20.0℃/Min
▶ Change [Changed SP(TSP) - PV at the SP changing point] with slope of 20.0℃ per minute:
30.0℃
(70.0-30.0)℃ = Change 40.0℃ with the slope of 20℃
▶ Increase the current set value (SP) from 30.0℃ to 70.0℃ with uniform increasing rate for
2 minutes.
SP change 2minutes.
Program setting
6-1 Program pattern setting‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 49
6-2 Pattern repetition setting‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 54
6-3 File editing‥ ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 56
6-4 Time signal operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 59
6-5 Standby operation‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 62
6-6 Experiment name setting‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 64
46
Program setting
[Fig. 6-2] Pattern editing screen [Fig. 6-3] Pattern and segment repetition [Fig. 6-4] File editing screen
setting screen
[Fig. 6-7] Time signal setting #1 [Fig. 6-10] Standby operation setting screen [Fig. 6-11] Experiment name setting screen
Part 06
47
06. Program setting
● It converts to [Fig. 6-1 Program setting screen] when the program setting button is touched by in [Fig. 2-1 Main screen].
● It is a screen group to set the parameters related to the program operation.
48
6-1. Program pattern setting
● It is a screen to set the segment depending on the pattern number.
● Refer to [6-4 Time signal operation] for time signal setting.
● Please refer to the [6-5 Standby operation] to find things about the standby operation.
The set value (SP) is processed to the set time (TM1) from present value (PV)
to the present value (SP1) set in segment 1 (SEG1) regardless of the slope or
2 the start set value (SSP) in starting of the program operation.
(The remained time for operation is calculated by regarding of time passing
9 till the program operation starting through referring the patterns of set
program.)
10 ②
3 ● TEMP SSP
4 In case of starting the program operation, the present set value (SP) is started
5 from temperature SSP and progresses to the set time (TM1) for currently set
11 value 1 (SP1) in segment 1 (SEG1).
6
● HUMI SSP
7
12 It starts from the set SSP for present set value (SP) in starting of program
8 operation and progressed to the set time (TM1) for currently set value 1 (SP1)
49
Selection of Y/N for using the Standby of Segment to operate.
⑦
● Refer to [6-5 Standby operation]
Select the soak operation of the temperature and humidity segment.
⑧
The selected segment operates soak without sloping.
When one of the buttons (Segment 01~99) is touched by
for segment insertion, the selected button (Segment 01~99)
⑨
and button are activated and the selected segment can be
inserted when is touched by
▲ When the “Starting condition” button is touched by, the input key to set the
starting condition is displayed
50
▲ It is a screen set with “TEMP SSP” for starting condition. ▲ , buttons are activated when
button is activated.
※ #n : 1 ~ 99
References
▶ Input by pressing for time signal and SEG alarm input.
▶ Select to escape from input screen.
▶ The wanted time signal group can be input by pressing
TS1~TS4 for the set value in [6-4 Time signal operation].
2 It sets the segment to terminate the partial repetitive operation out of the set
⑤
8 patterns.
3
It sets the repetition frequency of the partial repetitive operation out of the set
⑥
patterns.
When operations to segment repeat, determines the start SP value.
4
⑦ • END SP : From the late operation END SP, SP value proceed.
5 • START SP : From the start SP of segment, the SP value proceed.
6
It decides the next operation when the set pattern operation is finished.
• RESET : The pattern termination signal is generated and the operation
state is in program stop.
⑧
• SEG HOLD : It is operated in last operation set value and hold state is
maintained.
• LINK RUN : The pattern set in the connection pattern is operated.
54
Parameter Setting range Unit Initial value
PATTERN NO. 1~120 ABS 1
SEG. REPEAT MODE END SP, START SP ABS END SP
REPEAT COUNT 0 (Indefinite repetition) ~ 999 ABS 1
LINK PATTERN 1~120 ABS 1
PATTERN END MODE RESET, SEG HOLD, LINK RUN ABS RESET
START SEGMENT NO. 1~4 0~99 ABS 0
END SEGMENT NO. 1~4 0~99 ABS 0
REPEAT COUNT NO. 1~4 0~99 ABS 0
Example)
SEG. Repeat Mode : END SP SEG. Repeat Mode : START SP
- From the late operation END SP, SP value proceed. - From the start SP of segment, the SP value proceed.
℃ ℃
30 30
SP SP
20 20
0 0 SSP
(PV) (PV)
ESP
-10 -10
1 2 3 4 5 2 3 4 5 6 Seg 1 2 3 4 5 2 3 4 5 6 Seg
SEG. repeat section SEG. repeat section SEG. repeat section SEG. repeat section
55
6-3. File editing
● It is a screen to copy or delete the input segment values in [Fig. 6-1 Program pattern setting].
● The pattern number in operation cannot be deleted.
● The deleted pattern cannot be recovere.
56
[Fig. 6-5] File editing screen #1 [Fig. 6-6] File editing screen #2
References References
▶ The message : “PARAMETER SETTING SET ERROR” ▶ The message : “THE SELECTED PATTERN IS EMPTY”
appeares at the bottom of screen for copying or deleting by wrong input of Appears at the bottom of screen for copying or deleting without saved data
57
Parameter Setting range Unit Initial value
PATTERN NO. 1~120 ABS 0
START PATTERN 0~120 ABS 0
COPY
END PATTERN 0~120 ABS 0
COPY UNUSE, USE ABS UNUSE
START PATTERN 0~120 ABS 0
SEL. DEL
END PATTERN 0~120 ABS 0
ALL. DEL UNUSE, USE ABS UNUSE
58
6-4. Time signal operation
● The time signal operation is classified into ON/OFF operation, time setting operation and the set time signal is used for setting the time signal No.
in segment setting in [6-1 Program pattern setting]
(1) Time signal ON/OFF operation
1
The time signal is ON during corresponding segment operation
② when “1” is selected.
● The change is impossible as it is only for reading.
The time signal is “ON” after setting time is elapsed in delay time from
2 the corresponding segment starting point.
③ ● However, the time signal is not operated when the corresponding
59
(2) Time signal time setting operation
● The time signal 2~17 (TS2~17) operate depending on delay time and operation time.
[Fig. 6-8] Time signal setting #2 [Fig. 6-9] Time signal setting #3
60
(3) Example of operation in time signal input
Setting Time signal operation
Operation time
ON
OFF
Segment N time≥
Segment (n-1)segment time n segment time (n+1)segment time
Delay time +
Operation time
Operation time ON
Delay time
2. Delay time ≠ 000.00.00 Time signal
OFF
Segment (n-1)segment time n segment time (n+1)segment time
Operation time
ON
Segment N time<
3. Delay time = 000.00.00 Time signal
Delay time +
OFF
- Coditions for Stand-by Operation Entry (“OR” condition) : In case of no entry of either of temperature or humidity into the set range for standby operation within the set segment time
- Conditions for Stand-by Operation Release (“AND” condition) : In case of entry of temperature and humidity into the set range for standby operation
- The standby time has indefinite value when the standby time is not set (Initial value).
62
Parameter Setting range Unit Initial value
WAIT USE UNUSE, USE ABS UNUSE
TEMP ZONE T.EUS(0.00~100.00%) T.EUS T.EUS(0.00%)
HUMI ZONE H.EUS(0.0~100.0%) H.EUS H.EUS(0.0%)
WAIT TIME 00.00~99.59 (UNUSE, USE) ABS 00.00
WAIT USE METHOD ALL, SEG ABS ALL
References
▶ I t is a graph for interactive relation between standby operation and
standby time.
▶ S tandby operation range : It displays the humidity standby operation range
for temperature standby operation range for temperature and humidity
standby operation range for humidity.
63
6-6. Experiment name setting
● The experiment name can be set for each pattern. Refer to [4-2(2) Program operation #1 operation screen]
64
07
Part
65
07. Appointed operation setting
● It converts to [Fig. 7-1 Time setting screen] when the appointed operation setting button is touched by in [Fig. 2-1 Main screen].
● It is a screen to set the current time and appointed operation time.
66
[Fig. 7-2] Operation appointed setting screen (Program operation) [Fig. 7-3] Operation appointed setting screen (Stationary operation)
68
08
Part
69
Setting display setting
Flow chart
[Fig. 8-1] Screen display setting screen [Fig. 8-8] DI error creation history screen
Part 08
70
08. Screen display setting
8-1. Screen display setting
● It converts to [Fig. 8-1 Screen display setting screen] when the screen display setting button is touched by in [Fig. 2-1 Main screen].
[Fig. 8-1] Screen display setting screen It sets the backlight electricity saving time.
① ● The electricity saving time sets the operation timing of backlight
- In case that temperature set data (SP) in in the range of relative humidity
6 display condition(DRY LIMIT)
3 ②
- In case that temperature indicated data (PV). Is in 0.0℃ to 100.0℃ range
- In case that humidity set data (SP) is not set to 0.0%
● Manual
7
- The humidity is displayed depending on the set value (SP) and indicated
value(PV) as in the automatic mode and the humidity is displayed even when
72
[Fig.8-4] Touch screen calibration #3 [Fig.8-5] Touch screen calibration #4
2
Parameter Setting range Unit Initial value
ALL CLR UNUSE, USE ABS UNUSE
74
[Fig. 8-9] Screen for DI error display method by letter [Fig. 8-10] Screen for DI error display method by photo
References
▶ It is a screen in case of DI error creation.
▶ The setting for letter and photo screen setting can be set in [11. DI function and operation setting] in [Installation manual].
75
09
Part
76
09. Communication error
[Fig. 9-1] Screen for DI error display method by letter [Fig. 9-2] Screen for DI error display method by photo
77
Engineering Units - EU, EUS
When the sensor type (IN-T) or the upper limit.lower limit of input range is changed, the parameters expressed in EU( ), EUS( ) are changed in
proportion to current data. (However, the upper and lower range setting data is initialized.)
Download the instruction manual and communication manual from the homepage.
EU( ) : Value of engineering unit depending on the range of instrument
EUS( ) : Value of engineering unit depending on the span of instrument
(Example)
▶ INPUT = PT_1
▶ RANGE = -90.00℃(RL) ~ 200.00℃(RH)