LightSpeed Series Technical Reference Manual - UM - 5478900-2-1EN - 3
LightSpeed Series Technical Reference Manual - UM - 5478900-2-1EN - 3
LightSpeedTM Series
Technical Reference Manual
GE Hangwei Medical Systems does business as GE Healthcare
LightSpeed Plus
LightSpeed Ultra
LightSpeed16
LightSpeedTM Series
Technical Reference Manual, English
original draft in English
5478900-2-1EN
Revision: 3
© 2016 General Electric Company
All rights reserved.
REVISION HISTORY
Chapter 3: Safety
Introduction............................................................................................................................................3-1
What Do I Need to Know About....................................................................................................3-3
Warning Labels and Symbols................................................................................................3-3
Equipment Warning Labels ............................................................................................3-9
General Safety Guidelines .................................................................................................... 3-12
Implantable Device Safety........................................................................................... 3-14
Radiation Safety........................................................................................................................ 3-15
Authorized Users .............................................................................................................. 3-16
General Radiation Safety.............................................................................................. 3-16
Scans Acquired at the Same Tomographic Plane ............................................ 3-17
Geometric Efficiency....................................................................................................... 3-18
CTDIvol .................................................................................................................................. 3-19
Pediatric and Small Patient Imaging....................................................................... 3-19
X-Ray Tubes........................................................................................................................ 3-19
Electrical Safety ................................................................................................... 3-20
Mechanical Safety.................................................................................................................... 3-22
General Mechanical Safety ......................................................................................... 3-22
Short Footprint Mode ..................................................................................................... 3-22
Patient Positioning........................................................................................................... 3-23
Table capacity ................................................................................................................... 3-23
Laser Safety (Reference 21CFR 1040.10 (h))................................................................. 3-27
Reconstructed Image Orientation.................................................................................... 3-27
Data Safety.................................................................................................................................. 3-29
Application Software Safety................................................................................................ 3-32
Application Specific Safety Topics.................................................................................... 3-32
Helical Scanning ............................................................................................................... 3-32
Cardiac Imaging ............................................................................................................... 3-33
Patient Preparation ......................................................................................................... 3-34
Lung Algorithm.................................................................................................................. 3-35
Autoscan .............................................................................................................................. 3-36
Chapter 1
Before You Start
Anyone who operates this system should have received prior training before they attempt to
scan or diagnose patients. This training should include medical and X-Ray education, in
addition to GE applications training. This guide does not provide medical explanations, but it
does suggest potential applications for some of the software features. It describes potential
Safety problems, and how to avoid them.
Everyone who uses this equipment must read and understand all instructions, precautions
and warnings. This manual should be kept near the equipment. Procedures and safety
precautions should be viewed periodically.
This Guide addresses three safety classifications:
DANGER: The most severe label describes conditions or actions which result in a specific
hazard. You will cause severe or fatal personal injury, or substantial property
damage, if you ignore these instructions.
WARNING: This label identifies conditions or actions for which result in a specific hazard.
You may cause severe personal injury, or substantial property damage, if you
ignore these instructions.
CAUTION: This label applies to conditions or actions that have potential hazard. You can
cause minor injury or property damage if you ignore these instructions.
This Manual uses pictures, or icons, to reinforce the printed message. It uses the
corresponding international symbol or icon next to the danger, warning or caution message.
For example, the upright hand with the lightning bolt across it warns of electrical hazards.
Federal law restricts this device to sale by or on the order of a physician.
Do not use the equipment if a known safety problem exists. Call your local service
provider and have the system repaired.
Applications Help
Although we try to make this guide complete and accurate, undocumented changes or
unexpected results do occur.
If you can't find the answer to your application question, you may call the Customer Center.
Use this phone number for non emergency purposes only, because you may not receive an
immediate response.
1-800-682-5327 Select 1 for Applications Answer line.
Select 3 for CT Application assistance.
If your system fails, or you have an emergency, call
GE Cares at 1-800-437-1171.
CAUTION: This system was designed for use by individuals trained in CT system
operation. Study the Safety Tab of this Manual before you scan the first
patient. Use the Index to find the section and page number of an item of
interest. Periodically review the Learning and Reference Guide and the
Technical Reference Manual.
CAUTION: Improper system usage could void your warranty. More importantly, you
could endanger your patients and yourself if you don't follow the correct
procedures.
Chapter 2
X-ray Protection
CAUTION: Improperly used X-Ray equipment may cause injury. Read and understand
the instructions in this book before you attempt to operate this equipment.
If you fail to follow safe X-Ray practices or ignore the advice presented in the
manual, you and your patient risk exposure to hazardous radiation. The
General Electric Company, Healthcare Group, will gladly assist and cooperate
in placing this equipment into use.
Although this equipment incorporates a high degree of protection against X-Ray outside the
useful beam, no practical design can provide complete protection. Nor can any practical
design compel a user to take adequate precautions to prevent the possibility of any person
carelessly, unwisely, or unknowingly exposing themselves or others to radiation.
Everyone having anything to do with X-Ray must receive proper training and become fully
acquainted with the recommendations of the National Council on Radiation Protection and
Measurements, and the International Commission on Radiation Protection.
NCRP reports are available from:
NCRP Publications
7910 Woodmont Avenue
Room 1016
Bethesda, Maryland 20814
CAUTION: Everyone having anything to do with X-Ray must take adequate steps to
insure protection against injury.
All persons authorized to use the equipment must understand the dangers posed by
excessive X-Ray exposure. We sell the equipment with the understanding that the General
Electric Company, Healthcare Group, its agents, and representatives have no responsibility
for injury or damage which may result from exposure to X-Ray.
GE urges you to use protective materials and devices.
Chapter 3
Safety
Introduction
This chapter provides information about safety precautions and procedures. It is important
for you to read and understand the contents of this chapter so the correct precautions and
procedures are followed.
This manual should be kept near the console for easy access.
CAUTION: This system was designed for use by individuals trained in CT system
operation. Study the Safety Tab of this Manual before you scan the first
patient. Use the Index to find the section and page number of an item of
interest. Periodically review the Learning and Reference Guide, Applications
Tips and Workarounds, and the Technical Reference Manual.
If necessary, additional training is available from a GE Applications Specialist. Contact your
institution’s GE sales representative for additional information about further safety and
operational training.
WARNING: Modification of any existing patient data on the system must follow the
guidelines specified in the User Manual. (Reference 21CFR 801.109)
The system is classified as a Class I, IPX0 equipment, not suitable for use in the presence of a
flammable anesthetic mixture with oxygen or nitrous oxide. It is rated for continuous
operation with intermittent loading. No sterilization is applied. The patient table cradle and
cradle accessories are considered Type B applied parts.
The system is intended to be used for head and whole body computed tomography. Refer to
each Technical Reference Manual for the detail information.
United States Federal Regulation 21CFR 801.109
CAUTION: Federal law restricts this device to sale by or on the order of a physician.
CAUTION: Improper system usage could void your warranty. More importantly, you
could endanger your patients and yourself if you don't follow the correct
procedures.
Watch for electromagnetic compatibility from other hardware. Detailed information can be
found in the Electromagnetic Compatibility chapter in the Technical Reference manual.
DANGER: The most severe label describes conditions or actions which result in a specific
hazard. You will cause severe or fatal personal injury, or substantial property
damage if you ignore these instructions.
WARNING: This label identifies conditions or actions which result in a specific hazard.
You will cause severe personal injury, or substantial property damage if you
ignore these instructions.
CAUTION: This label applies to conditions or actions that have potential hazard. You may
cause minor injury or property damage if you ignore these instructions.
This chapter uses the international symbol or icon along with the danger, warning or caution
message.
Table 3-1 IEC Standards
Symbol Description
Alternating current
ON / Power
Input Power
Output Power
Type B Equipment
Symbol Definition
by (Made
Indicates the Manufacturing Location
by)
Pushing prohibited
Model Number
Serial Number
Date of Manufacture
Minimum Filtration
Figure 3-1 The following warning labels are located at the bottom of the gantry cover
(Reference 21CFR 1040.10 (h))
Figure 3-2 Labels on the front of the gantry (Reference 21CFR 1040.10(h)):
WARNING: This x-ray unit may be dangerous to patient and operator unless safe
exposure factors, operating instructions and maintenance schedules are
observed. To be used by authorized personnel only.
Figure 3-7 The following label is located on the operators console for systems
manufactured before June 10, 2006 (Reference 21CFR 1020.30 (j))
WARNING: This x-ray unit may be dangerous to patient and operator unless safe
exposure factors and operating instructions are observed. To be used by
authorized personnel only.
! CAUTION
PINCH POINT.
Keep hands clear
when tilting.
WARNING: Do not place your hands inside the gantry opening when tilting the gantry.
The gantry can pinch or crush your hands!
Figure 3-9 Warning label on gantry front cover, table, operator console, and PDU
! CAUTION
AVOID INJURY.
Read and understand
information in manuals
before operating product.
CAUTION: AVOID INJURY. Read and understand information in manuals before operating
product.
Figure 3-10 Warning label located on PDU
! CAUTION
PDU CAN MOVE AND
DAMAGE CABLES.
Do not lean on or move
when connected to power.
CAUTION: PDU CAN MOVE AND DAMAGE CABLES. Do not lean on or move when
connected to power.
Figure 3-11 Warning label on the table
! CAUTION
AVOID INJURY.
Do Not Exceed Table
>205kg Maximum Capacity
of 205 kg (450 lb).
CAUTION: AVOID INJURY. Do Not Exceed Table Maximum Capacity of 205 kg (450 lb).
CAUTION: Excessive weight can break accessory and cause injury. Do not load more than
34 kg or 75 pounds.
CAUTION: Do not hit the accessory against the gantry. Patient injury or equipment
damage could result.
CAUTION: Do not load more than 4,5 kg or 10 pounds. Verify that extension collar is
securely tightened before use.
CAUTION: Accessory may fall and cause injury if not latched to cradle. Make sure that
accessory is latched to underside of cradle.
CAUTION: Do not load any non GE approved software onto the computer.
DANGER: Make sure all covers are in place before you use the equipment. The covers
protect you and your patient from moving parts or electrical shock. The
covers also protect the equipment.
NOTE: Only qualified Service personnel should service the system with the covers off.
WARNING: This system is intended for use by healthcare professionals only. This system
may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation measures, such as
reorienting or relocating the system or shielding the location.
WARNING: CT Scans may cause interference with implanted or externally worn electronic
medical devices such as pacemakers, defibrillators, neuro stimulators and
drug infusion pumps. The interference could cause operational changes or
malfunction of the electronic medical device.
Radiation Safety
(Reference 21CFR 1020.30 (h) (1) (i))
WARNING: Improperly used X-Ray equipment may cause injury. Read and understand
the instructions in this book before you attempt to operate this equipment.
If you fail to follow safe X-Ray practices or ignore the advice presented in the
manual, you and your patient risk exposure to hazardous radiation.
Authorized Users
This equipment incorporates a high degree of protection against X-Ray radiation outside the
useful beam. But this equipment can not substitute the essential requirement that every
user must take adequate precautions to prevent the possibility of any person carelessly,
unwisely, or unknowingly exposing themselves or others to radiation.
Everyone having anything to do with X-Ray equipment must receive proper training and
become fully acquainted with the recommendations of the National Council on Radiation
Protection and Measurements and the International Commission on Radiation Protection.
NCRP reports are available from:
NCRP Publications
7910 Woodmont Avenue
Room 1016
Bethesda, Maryland 20814
WARNING: Everyone having anything to do with X-Ray equipment must take adequate
steps to insure protection against injury.
All persons authorized to use the equipment must understand the dangers posed by X-Ray
exposure so that they can prevent any injury or damage that may result from such
exposure. GE urges you to use protective materials and devices to prevent any injury or
damage from X-Ray exposure.
WARNING: Never scan a patient with unauthorized personnel in the scan room. Warn
visitors and patients about potential for harm if they fail to follow
instructions.
WARNING: Never calibrate, test the system, or warm the tube with patients or personnel
present in the scan room without adequate radiation safety precautions
being utilized.
– Stay behind a lead screen or lead glass shield during each X-Ray exposure.
– Use technique factors prescribed by the radiologist or diagnostician. Use a dose that
produces the best diagnostic results with the least X-Ray exposure.
– Amber indicator lights on the gantry display panel, and rear of the gantry, illuminate
during X-Ray exposure.
WARNING: This series contains one or more groups with multiple scans at the same
tomographic plane, i.e. same location.
Do you want to continue?
CAUTION: Prolonged exposure to x-ray in one spot may cause reddening or radiation
burns. Users must be aware of the techniques used and exposure time to
insure safe operation.
Geometric Efficiency
IEC60601-1 third edition and related standards, such as IEC600601-2-44 third edition, state
that when a prescribed scanning parameter has the detector coverage with dose efficiency
of less than 70 %, a warning pop-up shall be displayed and confirmed.
A warning message is posted when the Geometric Efficiency in the Z-direction is less than
70%. Geometric Efficiency is a measure of how much of the X-ray beam in the Z-direction is
used by the system.
Figure 3-18 Warning message when dose efficiency less than 70 %.
WARNING: The Geometric Efficiency in the Z direction is less than 70% for image 1-73.
After reading the message, if you wish to continue with the scan, click [Continue].
NOTE: The image number of 1-73 is just for example.
CTDIvol
As you setup the scan parameters from the view/edit screen, the Dose Information area at
the upper right of the scan monitor contains updated dose information. This dose
information is based on a measurement of the CTDI or CT Dose Index, which is the current
standard for CT dosimetry and performance. By using a measurement called CTDIvol, a
single value is provided to estimate the relative dose for an exam.
The CTDIvol is a weighted average measurement in a reference phantom. This dose is
expressed in milliGrays. For additional information on specific CTDIvol doses and their
calculations, refer to your Technical Reference manual.
The DLP or Dose Length Product is the product of the CTDIvol and the scan length for a
group of scans. This number can be summed over the entire exam to give an estimate of the
total dose. The value is expressed in milliGray centimeters.
The Projected Series DLP shows the DLP that would result from scanning the current group
or groups.
The Accumulated Exam DLP displays the total exam DLP up to the current point in time.
Scout dose is not included in the DLP totals since standards for reporting scout dose are not
yet defined. Scout dose is generally a very small part of the exam.
The dose information updates when technique values such as kV, mA, scan time, slice
thickness, and scan field of view are changed.
Dose information is saved as screen save image in Series 999 upon End Exam and Series
997 contains the DICOM Dose Structured Report.
X-Ray Tubes
The system uses cooling and reconstruction algorithms specifically designed for GE X-Ray
tubes.
You risk three dangers when you do not use GE X-Ray tubes.
• A non GE tube could cause destructive component failure if the cooling delays do not
meet its design requirements.
• The images could exhibit reduced performance or artifacts if your x-ray tube fails to
conform with GE tube performance specifications.
• Radiation leakage may exceed GE specifications when a non GE X-Ray tube is installed
in the system.
CAUTION: We cannot guarantee performance or safety if you use a non GE X-Ray tube
because the cooling and reconstruction algorithms depend upon the tube
design. Radiation leakage may exceed GE specifications when a non GE X-Ray
tube is installed in the system.
Electrical Safety
DANGER: ELECTRICAL SHOCK HAZARD. Avoid all contact with any electrical conductor.
Do not remove or open system covers or plugs. Internal circuits use high
voltage capable of causing serious injury.
To guarantee safe, reliable equipment performance, prepare the site
according to GE Medical Systems requirements. This includes making sure
the equipment is connected to a power supply main with a protective earth
ground. If you have any questions about these requirements, contact GE
Medical Systems.
An electrical hazard may exist if any light, monitor or visual indicator stays
on after the system is shut down. To prevent possible injury, turn off the
main power supply wall switch, and contact your service office immediately.
DANGER: ELECTRICAL FIRE. Conductive fluids that seep into the active circuit
components of the system may cause short circuits that can result in
electrical fires. Therefore, do not place any liquid or food on any part of the
system.
To avoid electrical shocks or burns caused by the use of wrong type of fire
extinguisher, make sure that only fire extinguishers approved for use on
electrical fires are used.
CAUTION: Gantry and console rear plug in panel Receptacles are not for general use.
• Surplus length of power cords or other cables from mobile accessory units that may be
used with some patient scanning should be stored in safe and isolated areas, such as
individually in a figure eight at the base of stationary equipment. This discourages signal
interference and protects cables from damage due to traffic.
CAUTION: The outlets are not for General Use. Operator Console outlet has a rating for
2.5A at 120VAC. Gantry outlets have a rating for 2.0A at 120VAC. Accessories
should not exceed above rating.
A CT System combined with GE approved accessories complies with the IEC60601-1
standards related to safety and performance of medical electrical systems. Refer to the
standard for more information.
• Do not connect electric devices to the CT System that are not approved by GE. It may
create increased electrical leakage current and there is possibility of electric shock.
• The GE console monitors, modem, video amp, and media tower are intended to be
powered by the CT System using cables provided. Do not connect these devices to
power sources other than the CT system (for example, wall outlets or other electrical
equipment). It may create increased electrical leakage current and there is possibility of
electric shock.
• Note that some powered equipment may only be connected by a signal cable to GE
equipment (for example, a network hub). A separation device is required for equipment
that is powered by a different power source.
Mechanical Safety
WARNING: Do not use the table base as a foot rest. You could entrap and injure your foot
while lowering the table. Do not place your hands between the table base and
the table side panels.
WARNING: Do not place your hands inside the gantry opening when tilting the gantry.
The gantry can pinch or crush your hands!
WARNING: Be sure that the Gantry will not touch the patient during Remote Tilt
operation. Pinching or crushing may happen if the Gantry touches the patient.
• Avoid any patient contact with the gantry during tilt or cradle movement (manual or
software driven).
WARNING: The IN-limit position of cradle Short Footprint Mode should be set in order not
to pinch a patient’s foot between the cradle edge and the wall of the scan
room.
WARNING: To Prevent Pinching or Crushing of the Patient, Watch the Patient and
Equipment Carefully At All Times During Table Movement. If Unwanted Motion
Occurs or Motion Does Not Stop, Press the Emergency Stop Switches on the
Console or Gantry.
Patient Positioning
CAUTION: If the head is poorly positioned in the head holder and a tilt is used, images
with different CT numbers and intensities may be seen at the edges of two
rotational interfaces. Make sure the patient is properly position up in the head
holder, not positioned such that the head is at the junction of the head holder
attachment to the cradle. If a repeat scan is needed, make sure the locations
with different intensities are in the middle of the beam collimation. Do not
repeat using exactly same prescription.
DANGER: Do Not Place a Patient on the Table (except High Capacity Table) Weighing
More Than the Upper Limit Of 450 Pounds (205kg). This Could Cause the Table
to Fail and the Patient Could Fall.
Table capacity
For HP Table
Up to 400 lb. (180kg) with ± 0.25 mm positional accuracy guaranteed.
400 to 450 lb. (180 to 205 kg) maximum allowed with normal operation and ± 1 mm
positional accuracy.
CAUTION: When using the external laser alignment light for patient positioning
purposes, be aware that the patient's elevation may be slightly lower with the
cradle extended than with the cradle fully retracted. This is because the cradle
may bend slightly under a patient's weight. This difference should be taken
into consideration for applications where patient position information is
critical, such as Treatment Planning. To minimize these affects, after using
the external laser alignment system to position the patient, advance the
patient to the CT scan plane. Turn on the CT alignment lights to determine if
they line up with the markers on the patient. If necessary, compensate for the
bend in the cradle by elevating the table. When the CT alignment lights line
up with the markers, set the landmark for the scan using the Internal laser
alignment light.
CAUTION: When using patient positioning accessories, make sure there are no areas,
which might cause a pinch point or interfere with patient tubing or IV.
CAUTION: Check to make sure the power injector has enough IV tubing to allow free
movement of the cradle. Make sure the unit itself does not interfere with table
travel.
Ensure excess tubing length is secured to the table top. DO NOT loop
additional IV tubing in the patient's fingers.
• Check the length of all patient health lines (IV tubing, oxygen line, etc.) and make sure
they accommodate cradle travel. Position these lines so they cannot catch on anything
within the patient vicinity or between the table and gantry during cradle travel or gantry
tilt.
CAUTION: The patient positioning straps provided with the system do not support the
full weight of the patient. Patient positioning straps should be used to aid in
patient positioning and are not meant to fully restrain the patient.
CAUTION: Care should be taken to ensure the patient positioning straps, patient
clothing, or other material will not be caught during table motion.
• The scannable range is not indicated by the black mark on the table, scannable range is
indicated by the Tilt and Travel limits button on the Gantry Controls.
CAUTION: If the table is lowered with anything in the red X area as indicated above, the
table could be damaged along with the equipment or object under the table.
CAUTION: Physically assist all patients on and off the table and into position on the
cradle.
CAUTION: The foot pedals at the base of the table for loading and unloading patients
are always active. Care should be taken not to activate the foot pedals once
the patient has been positioned on the cradle and an exam started.
• Return the gantry tilt to the 0 degree upright position, latch the cradle, and adjust the
table to a comfortable height for patient loading and unloading.
• Latch the cradle before you load or unload the patient (the Cradle Unlatch indicator
illuminates when the cradle is unlatched).
WARNING: To prevent pinching or crushing of the patient's extremities, keep the patient's
hands and feet away from the edge of the moving table top/cradle and its
surrounding equipment, or between table base and side panels of the table.
(Take special care when positioning physically large patients).
WARNING: To prevent pinching or crushing of the patient watch the patient and
equipment carefully at all times during gantry tilt or table movement. If
unwanted motion occurs or motion does not stop, press the emergency stop
switches on the console or gantry.
WARNING: The head holder may crack, possibly injuring the patient's head or neck, if the
patient tries to brace himself or herself on the head holder during positioning.
The head holder and cradle extender are only designed to support 75 pounds
(34kg). Ask the patient to move up into the head holder or manually help the
patient into position.
CAUTION: The patient head holder or table extender should be adequately secured to
ensure stability. If they not secured properly, degradation of image quality
may result due to introduced motion of the head holder or table extender.
CAUTION: Use of any cradle extension accessories such as the table extension, head
holder, coronal head holder, and phantom holder are not accounted for in the
table gantry interference matrix. Therefore, additional care needs to be taken
to closely monitor any table up/down, in/out or gantry tilt movement to avoid
contact of the extended accessory with the gantry.
NOTE: Collision sensors are placed under the table surfaces to stop downward motion and
minimize the effects of a collision in most cases. Upward motion is still allowed if a
collision sensor has been activated.
• Check the accessory attachment plate fixed to the end of the cradle. Repair or replace if
loose or damaged.
• Use the cradle extender to support the patient's head or feet during a scan.
• To move the patient out of the gantry in an emergency, the cradle can be manually
withdrawn by applying a minimum of 40 lbs (178 N) of force.
CAUTION: Temporal sampling may be degraded due to changes in timing for the table
to move from location to location if proper positioning methods are not
followed. Make sure that the patient is securely positioned on the table and
their arms are not allowed to drag on the table or allow clothing, sheets or
blankets to get caught causing a table move problem.
WARNING: Temporal interval for some of the images exceeds 3.2 seconds. Use of this
data for processing of CT Perfusion maps may contain errors in the functional
information.
CAUTION: The detector and DAS rotate to position the alignment lights over the laser
ports.
- Keep your hands away from the gantry opening.
- Make sure the gantry side covers are in place.
CAUTION: If you plan to reconstruct images, you must first use the files that reside on
the disk. Either reserve the scan files you plan to retrospectively reconstruct,
or reconstruct unsaved scan files before the system overwrites the file with
new scan data. The system refuses to overwrite reserved scan files.
Remember to release the reserved scan files when you finish retrospective
reconstructions.
R L L R
P A
Head First Supine Head First Prone
A P
R L L R
P A
Feet First Supine Feet First Prone
The patient position information stored in the image header correctly reflects the orientation
(RAS) information for the patient. Viewing applications will correctly reflect Right (R), Left (L),
Anterior (A) and Posterior (P) of the patient.
The reconstructed image orientation may differ from preferred anatomical viewing
presentation in which the patient's Right is on the viewers Left and patient's Left is on the
viewers Right. For example when the patient is scanned Head First and Prone the patients’s
Left is on the viewer’s Left and the patient’s Right is on the viewer’s Right. The image
presentation will need to be modified to display preferred anatomical viewing. Some
viewing stations may not have the capability to flip the image presentation, but if the
capability exists, you must use display tools such as Flip to change the presentation of the
image.
Some remote viewing stations may have the capability to set default viewing protocols, this
is another tool that can be used to set an anatomical viewing presentation.
Post processing applications such as Direct MPR, Reformat and Volume Viewer
automatically orient images in anatomical viewing orientation. These applications create
axial images in anatomical viewing presentation. Please see Auto Applications (Option) of
the Learning and Reference Guide for more information. The system also provides the
capability to create Gray Scale Presentation State Objects (GSPS) to flip the image
orientation.
Flip/Rotate in recon can be used to generate images where right/left or anterior/posterior
are flipped or where both R/L and A/P have been flipped to meet desired image display
preference. An Attention pop-up is displayed at Confirm for series where Flip/Rotate in recon
is selected. Attention: This scan prescription utilizes one of the reconstruction based image
flip and/or rotate options. Please ensure that this prescribed image orientation is displayed
appropriately on all remote viewing devices.
Figure 3-21 Flip Warning Message
CAUTION: The scan prescription utilizes one of the reconstruction based image flip
and/or rotate options. Please ensure the prescribed image orientation is
displayed correctly on all remote viewing devices.
Data Safety
To ensure data safety:
– Verify and record the patient's identification before starting a scan.
– Observe and record the patient's orientation, position and anatomical landmarks
before starting a scan. Ensure that the patient is positioned within the scan
parameters.
– Maintain system image quality by performing Daily QA and other maintenance.
Connectivity - Always verify that the data transferred to another system has been correctly
received.
CAUTION: When entering Patient ID information the system may contain multiple
instances of the same Patient ID. Multiple schedule records can be due to
multiple procedures being ordered under separate accession numbers or
New and Completed records in the Patient schedule for the same Patient ID.
When entering the Patient ID verify that the correct Accession number and
Exam Description selected is what is desired. Scanning with an incorrect
accession number may cause problems reconciling exams on a PACS
system. Please see the Schedule Patients chapter for more information.
CAUTION: The system posts a warning message when expected disk space required to
store scan data from the prescribed exam is insufficient.
CAUTION: The system posts a warning message when expected image space required
to store images from prescribed reconstruction is insufficient.
CAUTION: The system posts a warning message when data was interpolated to generate
images.
CAUTION: The system posts a warning message if there is a failure during the archive
of patient data.
CAUTION: The system posts a warning message if there is a failure during the network
of patient image data.
CAUTION: The system posts a warning message when a scan is aborted due to a failure
in the acquisition chain.
CAUTION: The system posts a warning message when the system has low disk space.
This is due to a partition on the system disk getting too full. Removing images
will not help. Contact service to help with recovery. If you reboot the system
and see the message asking if you want to run storelog, select the option to
remove the logs.
CAUTION: The system posts a warning message if patient orientation has been changed
or does not match after start of exam.
CAUTION: The system posts warning message prior to modifying any existing data set
by a software utility.
CAUTION: Do not remove images while scanning. Always remove images when the
system is idle. Removing images while the system is acquiring and
reconstructing data could cause the system to lock up and require a reboot
and/or force the system to go into data base recovery.
CAUTION: Saving images in Interchange (CD/DVD) while scanning may cause long
interscan delays (ISD) to be missed or may cause Auto Voice to fail to play. Do
not copy or restore images using CD-R or DVD-R while scanning.
CAUTION: When comparing GE CT images with other images, consult the DICOM
Conformance Statement for the details on the DICOM Image Position, Frame
of Reference UID and Slice location values stored.
CAUTION: Some annotation values are stored in private DICOM elements. When viewing
images on a remote station these annotation values may not be visible on the
image. Consult the DICOM conformance statement for information on private
DICOM data fields.
CAUTION: If you plan to reconstruct images, you must use files that reside on the disk.
Either reserve the scan files you plan to retrospectively reconstruct, or
reconstruct unsaved scan files before the system overwrites the files with
new scan data. The system refuses to overwrite reserved scan files.
Remember to release the reserved scan files when you finish retrospective
reconstruction.
CAUTION: Do not initiate a QuickSnap if the system is actively collecting data with x-ray
on.
Helical Scanning
WARNING: Helical scanning has the inherent ability to produce artifacts when scanning
highly sloped anatomy (e.g. pediatric or adult heads). Factors which worsen
this effect are: faster table speeds, thicker image thickness, and gantry tilt.
In some cases these artifacts could be mistaken for a hemorrhage near the
cranium, or a thickening of the skull.
To reduce the occurrence of these artifacts you may prescribe slower table
speeds and/or thinner slices (such as 2.5mm) during helical scans near the
vertex of a pediatric or adult head.
WARNING: It has been documented in radiology literature that an artifact may occur in
the chest that bears the double margin of the great vessels, which emulates
a dissection of the vessel during 0.5 - 1.0 second scans. This can occur in axial
or helical scans. If you have scanned axially with a 0.5 - 1.0 second rotation
time and observe this phenomenon, re-scan the area with a 2 second axial
scan to verify if it is artifact or patient pathology. Segment recon mode for
helical and cine acquisitions may be used in Retro recon to also assess if the
areas is artifact or pathology.
Cardiac Imaging
CAUTION: A patient with any of the conditions listed below may require additional
attention. If patients are scanned with these conditions, the software may not
be able to detect the R-Peaks and the images therefore may be produced as
ungated segment images.
- Patients with multiple pre-contractions or extra systole (e.g. PVC, PAC)
- Patients with persistent or extreme arrhythmia
- Patients with bi-ventricular lead (dual chamber) pacemakers
CAUTION: ECG signal clarity and integrity must be confirmed prior to performing
ECG-gated acquisitions. Items which may require adjustment of equipment
settings or positioning, or patient set-up include:
- External Interference
- Atypical Patient ECG (e.g. elevated T-Waves, low ECG amplitude or signal
strength)
- Suboptimal Patient Connection
ECG lead placement should follow recommended guidelines to optimize
results.
If the ECG lead is disconnected during the scan, or the heart rate drops
below 40 BPM, the images will be reconstructed as non-gated segment
images. This is done to avoid inaccuracy of the z-location of images where
necessary.
CAUTION: The heart rate displayed on CT console is a 3-cycle average. You must review
the actual waveform pattern to determine ECG trace clarity, trigger location
and if any cycle to cycle variability or masked arrhythmias may be present in
order to adapt set up and conditions prior to proceeding with the scan
acquisition.
CAUTION: Cardiac helical scan modes of SnapShot Segment, Burst, and Burst Plus are
optimized for specific heart rate ranges. Select the appropriate scan mode
for each patient's heart rate pattern. If the incorrect mode is selected,
temporal resolution may be insufficient and degraded image quality could
result.
Patient Preparation
CAUTION: Ensure the ECG patches are not past expiration date and that the gel on the
pads is still moist for proper conduction of the ECG signal for successful
gating.
Practice hyperventilation breathing instructions with the patient.
• It is important to explain to the patient the events that will occur during the acquisition
of the contrast enhanced cardiac data. Make sure to explain the warm feeling that may
occur during the injection of the contrast material.
• Use hyperventilation technique for all the series in a cardiac exam. Practice the
hyperventilation instructions with the patient prior to scanning.
• During the practice breath hold, make sure to watch the ECG trigger monitor to
determine the average heart rate, minimum heart rate, and ECG pattern during the
breath hold.
• Position the patient's arm over the patient's head so they are comfortable and will not
move during the acquisition of data.
CAUTION: If, during the scan, the heart rate drops significantly lower than the prescribed
heart rate, there is a potential for gaps in the gated image location. To avoid
image location gaps, a non-gated image is reconstructed for the period where
the patient heart rate dropped below the expected or confirmed heart rate
at the start of the exam. A non-gated image may have more motion and may
not be reconstructed at the prescribed phase.
CAUTION: There is a possibility that the ECG signal may not be detected by the scanner
due to improper lead placements, or a lead falling off during the scan. It is
important to place new leads on the patient before the scan. Make sure the
leads are attached properly, and use only GE recommended ECG leads.
It is important to confirm ECG trace clarity before the scan.
CAUTION: The heart rate displayed on the CT console is a 3-cycle average. You must
review the actual waveform pattern to determine ECG trace clarity, trigger
locations, and if any cycle to cycle variability or masked arrhythmias may be
present in order to adapt set up and conditions prior to proceeding with the
scan acquisition.
Lung Algorithm
• The Lung algorithm setting provides edge enhancement between structures with large
density differences, such as calcium and air, resulting in a sharper lung field when
compared to Standard algorithm.
• For best image quality, prescribe a 5 mm scan thickness when you plan to use the Lung
algorithm. If you plan to prescribe a High Resolution Lung study with 3.75, 2.5, or
1.25 mm, use the Bone algorithm.
• The Lung setting enhances the contrast of small objects. For best viewing and film
quality, select a window width of 1000 to 1500 and a window level of -500 to -600.
• The Lung algorithm setting increases the CT number values at the edge of high contrast
objects. If you plan to take CT number measurements of vessels or nodules in the lung,
please check and compare your results with Standard algorithm images. (ROI and
Histogram functions use CT numbers.)
• Remember: The edge enhancement provided by the Lung setting may not be
appropriate in some clinical cases. Please take individual viewing preferences into
account when you choose the Lung setting.
Autoscan
• Press and release Move to Scan on the console to advance the cradle.
• If Autoscan is disabled, Move to Scan must be pressed for every scan before Start Scan
will become ready.
• If you select Auto Scan during one group Rx, it remains ON for every group in that series.
SmartStep/SmartView Safety
The SmartStep option adds several components to the scan room. These are the In-Room
Monitor, Hand Held Control for table movement as well as image review, and the X-Ray
Control Foot pedal.
Each of the SmartStep components is connected to the system by a cable. When using the
system, ensure that the cables cannot catch on anything when the gantry or table is moved.
Table Float
During the scan the Clinician has the option to float the table between scans. When the
Table Float mode is selected, the table is unlatched and can be moved freely by anyone at
the bedside.
WARNING: Unintended table motion may cause a serious injury. Table may be bumped
or jarred during an interventional procedure. Care must be taken when
performing interventional procedures in the float mode. It is the clinician's
responsibility to ensure that they have control of the table when in this mode
of operation. Table must not be left unattended when in the float mode.
Ensure that the table is latched before leaving the table side.
SmartStep/SmartView Scanning
SmartStep scanning allows multiple scans at one location for interventional procedures. The
system allows up to 90 seconds of scanning in one place. After 90 seconds, the operator
must prescribe a new scan to continue. The accumulated scan time from a procedure is
displayed in the In-Room Monitor.
CAUTION: Expose time to the patient can be up to 90 seconds per confirm compared to
60 or 120 (option) seconds for all other scan modes.
CAUTION: Prolonged exposure to x-ray in one spot may cause reddening or radiation
burns. User must be aware of the techniques used and exposure time to insure
safe operation.
CAUTION: The foot pedal is active if the system is in the “Prepped” state. Care should be
taken not to step on the foot pedal and make an unwanted exposure.
CAUTION: The cabling provided for the integrated Hand-Held Controller (HHC) and Foot
Pedal with the SmartView and SmartStep options may present a trip hazard.
Ensure that the cabling can not catch on anything when the gantry or table
is moved and that the cables are out of the way while loading and unloading
the patient.
Clinician's working in the scan room should wear appropriate protective clothing. Lead
aprons, groin and thyroid protection, as well as protective eye wear are available through
the GE Accessories Catalog.
CAUTION: The continuous AutoView layout format should not be used for display of
images during an interventional study because it does not allow for quick
review of images in a free viewport.
WARNING: When scanning for interventional (biopsy) studies the scan mode, image
thickness, number of images per rotation and the display layout used affect
the display of the images. It is recommended to use the Biopsy Mode provided
on the system. If manually prescribing biopsy scans, Axial 1i scan mode or
Helical scan mode with a slice thickness greater than 2.5 mm must be used.
Do not use Cine scan mode for interventional (Biopsy) imaging. Do not use an
Auto view layout with more than one Auto View image viewport.
Refer to "Select a Multiple Image Display” in the Image Display Viewing Area chapter
of the Learning and Reference Guide for more information on how to set up the
desired viewing options.
Choose one of the following for the best auto view layouts.
WARNING: Non GE images acquired can be loaded in Volume Viewer but GE does not
guarantee the quality or reliability of any reconstruction, segmentation or
measurements performed on these images. Non-GE images can easily be
identified by the corresponding image annotation.
Follow the DICOM acquisition parameter guidelines listed in each
application user guide. Consult GE-published DICOM conformance
statement of Volume Viewer which is available on the GE Healthcare website
at
http://www.gehealthcare.com/usen/interoperability/dicom/products/workstati
on_dicom.html
WARNING: Before using any segmentation tool (threshold, scalpel, remove & keep object,
AutoSelect, “floater” filters…) always make sure that it will not remove
pathologies or other essential anatomical structures.
WARNING: When using any Segmentation tools (AutoSelect, threshold, Paint on slice,
Quick Paint…), check contours to check the reliability of the segmentation.
Make sure the contours match the correct segmentation and volumes Check
segmented volumes match contours.
Measurements
WARNING: Do not use 3D or slab views only to perform any measurements (distance,
angle, Region of Interest, Report Cursor, Area, Volume…). Always check
measurement points position and refer to 2D baseline views (acquisition
images or reformatted images of minimal thickness) to confirm
measurements.
CAUTION: The software calculates and displays measurements with a resolution of one
decimal (such as 0.1 mm, 0.1 degree, etc.). You should be aware that the real
measurement accuracy is generally less for a number of different reasons
(image resolution, acquisition conditions…).
Distance, angle and area measurements are valid only if all trace segments
are longer than the inter- slice distance.
CAUTION: When filming or saving images for diagnostic purposes, always make sure
the patient name and geometry information is displayed on all views and
match information on reference view.
CAUTION: When saving images with a new series description, make sure this description
matches the saved images.
Segment Tools
WARNING: Before using any segmentation tool (threshold, scalpel, remove & keep object,
AutoSelect, “floater” filters…) always make sure that it will not remove
pathologies or other essential anatomical structures.
WARNING: When using any Segmentation tools (AutoSelect, threshold, Paint on slice,
Quick Paint…), check contours to check the reliability of the segmentation.
Make sure the contours match the correct segmentation and volumes Check
segmented volumes match contours.
CAUTION: When filming or saving images for diagnostic purposes, always make sure
the patient name and geometry information is displayed on all views and
match information on reference view.
CAUTION: When saving images with a new series description, make sure this description
matches the saved images.
Image reliability
WARNING: The window width and level (W/L) determine how clearly pathologies and
other anatomical structures can be discerned. Incorrect W/L settings may
result in pathologies and other essential anatomical structures not being
displayed correctly. As a single W/L cannot display all features present in an
exam, use several different setting, when necessary to explore all exam data
Volume Rendering
WARNING: When using Volume Rendering, incorrect setting of opacity curve, opacity
threshold, transparency setting when merging VR objects can result in
pathology or essential anatomies not being visible. Always correlate Volume
Rendering images with original images.
Image quality
WARNING: Default Plaque Color Map preset is provided for information. You must check
and adjust Values and segment names.
Accuracy of Measurements
CAUTION: Note that the measurements are accurate only if the trace segments are
longer than the slice interval.
• For measurements of anatomy in the “X” direction that are NOT at ISO (“Y”):
– The measure error using the straight line distance graphic is less than 5 % of the
measured distance plus 2 mm plus 3 % of measured distance per centimeter from
ISO.
• For measurements of anatomy in the “Z” direction:
– Measure error using the straight line distance graphic is less than two times the
image pixel size.
Figure 3-23 Scout Scan Plane
X-Ray Tube
Focal Spot
Y-Axis
X-Axis Scan
ISO Center Plane
Z-Axis
Patient
Table
Measure Angle
Measurement accuracy using the angle graphic is equal to the displayed angle value +/- 10
degrees for an angle measured between segments which are five times larger than the
image pixel size. Accuracy improves as the length of the segments increases.
ROI
Area measurement accuracy using a region of interest graphic (rectangle, smooth curve,
ellipse or free draw) is equal to the displayed area +/- the circumference of the region
multiplied by (image pixel size)2/2. Mean and standard deviation values for the intensity of
the pixels in the region are also affected by this accuracy. If the ROI is rotated, the area
measurement can vary up to 5%. Region of interest statistics are based on the pixels INSIDE
the graphic defining the region.
CAUTION: CT Numbers are NOT absolute; misdiagnosis is possible. System and patient
variables may effect CT Number accuracy. If you rely solely upon CT numbers
without taking the following variables into consideration you could
misdiagnosis an image.
CAUTION: The limiting measurement resolution of the cursor is 1mm, i.e., the distance
less than 1mm but greater than 0.5mm is rounded to 1mm, therefore, the
accuracy of this testing is limited by the cursor measurement capability. This
is especially important for thin slice measurement where the FWHM is close
to 0.625mm. The results for these thin slice images will be not as accurate as
the thick slice ones. This is the limitation by this testing method.
Posture
Correct posture is very important. To ensure correct posture while sitting at your operator
console, follow these basic steps:
1. Face the monitors and keyboard without twisting your body.
2. Sit comfortably erect with the small of your back well supported.
3. Position your forearms parallel to the floor, with your wrists straight.
4. Position the screen so that your eyes are nearly level with the top of the screen.
5. Keep both feet flat on the footrest, with your thighs parallel to the floor.
If you cannot comfortably maintain this position while working at your operator console, you
should make the necessary adjustments to your operator console environment.
Equipment Adjustments
Chair
Adjusting the fit and height of your chair is very important for comfort. Follow these basic
guidelines:
1. Fit the backrest snugly against your back. People with shorter legs might need a back
cushion.
2. Set your chair height to position your forearms parallel with the floor when your hands
are placed on the keyboard. If your feet dangle, you need a footrest.
Keyboard
Keyboard height is also important. When typing:
– Your wrists should be as straight as possible.
– Your forearms should be parallel to the floor.
– Your hands and fingers should float over the keys or mouse.
Screen
• The recommended viewing distance from the screen is 18 - 28 inches (45 - 70
centimeters).
• With your head straight, your eyes should be looking directly at the top of the screen.
• You should look at the screen straight-on, not at an angle from the side, top or bottom.
• Glare from the screen can disrupt your viewing and cause eyestrain. Do not face a
window, and position the screen at right angles to bright light sources.
Comfort
Comfort at your operator console indicates you've set up your work area correctly. However
even a well-designed area needs frequent adjustment, especially for different users. Take
the time when positioning yourself at your operator console to ensure your comfort.
It is also recommended that if you use the operator console for extended periods of use
(several hours at a time), that you take short breaks to get away from your operator console
and perform simple stretching exercises to reduce the chance of fatigue.
Other considerations:
– Stay alert to your patient's condition.
– Use the speakers and microphones on the table, gantry, and console to stay in
constant communication, even while you sit at the console.
– Follow the exam procedures explained in the Chapters 13 and 14 of the Learning and
Reference Manual. Carefully enter patient information and position before
proceeding.
Accessories
WARNING: Do not connect accessories that are not approved as part of the system. Do
not use accessories from other modalities.
WARNING: None of the accessories support the full weight of a patient. If you sit, stand,
or otherwise apply excessive pressure to these devices, they break or come
off the cradle and may cause injury. Note if an accessory breaks, use caution
when picking it up and do not continue to use.
CAUTION: When using patient positioning accessories that are not GE options, make
sure there are no areas that might cause a pinch point or interfere with
patient tubing or IV.
WARNING: Accessories like arm boards and catheter bag holders are not secured to
gantry and may interfere with gantry if not positioned properly.
WARNING: All non medical equipment connected to the USB port of the media tower on
the CT operator console must comply with IEC/EN/UL60950-1 and should be
approved by GE.
CAUTION: Do not use the USB or Ethernet port on the front cover of the CT operator
console, it is intended for service use only.
CAUTION: Using accessories which are not GE approved accessories might affect dose
and image quality.
GE Approved Accessories
Use only GE approved equipment together with this system.
With each use check all accessories for damage and remove them from service if damaged
or cracked.
Table 3-3 Types and Model
Type Manufacturer/Model
IVY 3100 with ethernet
IVY 3100 - A with ethernet
IVY 3100 - B with ethernet
Cardiac Monitor IVY 3150
IVY 3150 - A
IVY 3150 - B
IVY 7800
Varian 1.6
Respiratory Monitor
Varian 1.7
Partial UPS *1 9155-10GE
Honeywell 3800g
Bar Code Reader *1 Hand Held 3800
Opticon 6125
IV Pole Safety
Care should be taken in the amount of weight and ensuring that the pole is tightened prior
to use.
CAUTION: The IV pole may bend when excessive weight is placed on the pole. Ensure no
more than 4.5 kg or 10 lb. is placed on the IV pole.
CAUTION: Ensure that the IV pole extension collar is tightened prior to use to avoid the
pole height to move on it's own.
Figure 3-24 IV Pole Load Limits
CAUTION: Do not load more than 4,5 kg or 10 pounds. Verify that extension collar is
securely tightened before use.
CAUTION: The maximum allowable weight on the table tray is 9kg or 20 lbs
CAUTION: Objects that may be susceptible to tipping should be strapped down with the
Velcro strap provided.
CAUTION: The injector and the system are operated independently after the Start Scan
button is pressed. When you want to stop both the system and the injector,
use the Stop Scan button on the system SCIM and the stop injector function
on the injector.
CAUTION: When you use Xtream Injector with SmartPrep, injection doesn't start at the
beginning of Baseline phase. It starts at Monitor phase. Going to Scan Phase
without Monitor phase, injection will not start.
CAUTION: Due to access limitations on the left side of the gantry, some procedures may
be affected when ancillary equipment is used. Assess the placement of the
equipment needed for the procedure before the placement of the patient on
the table. Access around the left side of the gantry may also be affected.
Emergency Devices
The system has two types of Emergency buttons:
1. Emergency Stop- when pressed, all table and gantry motions are halted, generation of
X-rays is stopped, laser alignment lights are turned off. The system aborts any data
acquisition in progress, and attempts to save all data acquired prior to the abort. Use
the Emergency Stop button for patient related emergencies.
2. System Emergency Off Button- when pressed, the power to all system components is
removed, stopping all table and gantry motion and generation of X-rays. The system
aborts any acquisitions in progress, and data obtained prior to the abort can become
corrupt or lost. Use the System Emergency OFF button for catastrophic emergencies,
such as fire or earthquake.
CAUTION: If you press the Emergency Stop or Emergency OFF buttons during a scan, the
system will abort the data acquisition.
Emergency Stop
NOTE: Every operator should take a few minutes to locate the Emergency Stops on his or
her system before he or she scans the first patient.
GOC4 console
NIO16 console
NIO16 console
Press an Emergency Stop button in the event of a patient related emergency or if the cradle,
table or gantry starts to move unexpectedly.
• Once an Emergency Stop button is pressed, the Reset gantry key, on the gantry control
panel, flashes about once every two seconds.
• Press the Reset gantry key to restore power to the gantry and table.
When Emergency Stop is applied, the moving cradle and tilting gantry may overrun by less
than 10 mm and less than 0.5 degrees respectively.
• During an exam, the system pauses between scans if you Press any button on the
control panel other than the alignment lights. It stops X-Ray if you Press the same
button(s) during a scan.
• Select Resume on the screen to continue the exam.
Emergency Egress
System operation may be stopped due to power failure or a safety event (something coming
into contact with the collision sensors), or the system may be halted by the operator in
response to emergency conditions.
The Cradle unlatch button should only be used in two situations.
1. In Emergency Egress situations.
2. When using the SmartStep scan type.
Environmental Concerns
This symbol indicates that the waste of electrical and electronic equipment
must not be disposed of as unsorted municipal waste and must be collected
separately. Please contact an authorized representative of the manufacturer for
information concerning the decommissioning of your equipment.
This product consists of devices that may contain mercury, which must be recycled or
disposed of in accordance with local, state, or country laws. (Within this system, the
backlight lamps in the monitor display contain mercury.)
The X-Ray Collimator contains the following potentially hazardous materials:
Lead: Lead salts are toxic and their ingestion may cause serious problems. The
manipulation/ handling of lead is subject to regulations.
WARNING: Do not discard the X-Ray Tube Assembly among industrial waste or domestic
garbage.
WARNING: A damaged X-ray Tube Assembly should not be dispatched through the
national postal service.
The X-Ray Tube Assembly contains the following potentially hazardous materials:
Lead: Lead salts are toxic and their ingestion may cause serious problems. The
manipulation/ handling of lead is subject to regulations.
Oil: Univolt 54 and Crosstrans 206 mineral oil are not toxic, but the prevailing
environmental regulations should be observed for their disposal or recuperation. For
example, it is forbidden to dispose of these oils in the wastewater or sewage system
or in the natural environment.
Your local GEMS field service will advise you on the suitable means of disposal of equipment.
The X-Ray Tube Assembly to be discarded should be forwarded to the GEMS Service
network, and it will be disposed of in a GEMS recycling center.
Precautions
Take all the necessary precautions for the personnel handling the recovery or destruction of
X-Ray Tube Assemblies, and in particular against the risks due to lead.
These personnel must be informed of the danger involved and of the necessity to observe
the safety measures.
Chapter 4
Accessing the Learning and
Reference Guide
The Learning and Reference Guide is designed to get you up and running your system. It is
recommended that each system operator review the complete Learning and Reference
Guide.
To start the Learning and Reference Guide:
• Select the [Learning Solutions] icon.
• Select the language you wish to review in.
Chapter 5
Tube Warmup
To maintain image quality and long tube life, complete the Warmup procedure.
• Display the Scan Monitor screen.
• Select Daily Prep and Tube Warmup to display the Warmup screen.
– When you haven't used the X-Ray tube for more than three hours
– Fastcal - Tube warmup runs automatically when Fastcal is selected.
– Before Calcheck
– Before system calibration
GE suggests you warm up the tube after every three hours of non-use.
• The system displays the following message on the Scan Monitor screen:
More than 2 hours have elapsed since the last scan. Tube needs warming.
NOTE: Tube warmup will run in Autoscan mode.
• Read the compatibility Warning.
• Position the gantry to 0 tilt.
• Click [Accept] when you understand the implications.
• Tube warmup runs a set of tube heating scans.
• When finished, the system display returns to the daily prep screen.
A message that tube warm-up has been completed will also be in the system message log.
Chapter 6
Daily Fast Cal Procedure
To maintain image quality, complete the Fast Cal procedure once a day.
1. Display the Scan Monitor screen.
– Clear the gantry opening.
– Raise the table above the patient loading level.
2. Select [Daily Prep] and [Fast Cal] to display the screen.
– Before the start of every scan day
– Before Cal check
– Before system calibration
3. The system automatically selects the Auto Scan function.
– The system automatically selects the following sequence of scans:
a) Mylar Window Check
b) Cold Warmup
c) Warmup 1
d) Warmup 2
e) Collimator Cal
f) FPA Check Scans
g) Clever Gain Calibration
h) Fast Calibration
NOTE: Some of the warmup scans may not run if tube is sufficently warm before fast cal.
– Follow system instructions to initiate the first scan, and the system acquires the rest
of the scan set.
– Remain near the console during auto scan acquisition, so you can stop X-Ray if
someone enters the scanner room.
Chapter 7
Prepare the System
Chapter 8
Check Disk Space
Maintain Image and Scan file space on the disk, because the system refuses to scan when it
runs out of file space.
Chapter 9
Reset the System
Chapter 10
Stop/Start the Operating System
Shutdown System.
To turn off main disconnect control (MDC) or A1, Shutdown the system.
1. On Display Monitor, select the [Shutdown] icon.
A pop-up attention box appears.
If there are open/active patient scan/tube warmup/fastcal or service tool menus, the
attention window below will be seen.
ATTENTION:Gantry tilt is not zero. Move gantry to zero tilt to enable Shutdown of the
system.
Click Confirm in the pop-up and return the gantry to zero tilt. Then start from step 1 to
select the [Shutdown] icon.
3. If shutdown check 1 and 2 are passed, pop-ups below will be displayed sequentially.
NOTE: The Dialog1 and Dialog2 below will be posted for only the system with Performix Plus
Tube. Click Confirm in the pop-ups to continue shutdown process.
Dialog1
WARNING: System is preparing for shutdown. The gantry will rotate immediately after
Confirm is pressed.
Dialog2
4. Press the STOP button on the MDC or A1 connector panel or power off the console with
power switch at the front of the console (if no MDC or A1 Control Panel).
If You Turn OFF the MDC at the End of the Scan Day:
To start the system if main disconnect has been powered off:
1. Press the START button on the main disconnect control (A1) to restore power to the PDU,
console(s) and subsystem electronics.
2. Press the RESET button to turn on the Gantry Control panel to restore power to the
Gantry drives, X-Ray system, and Table drive.
• When this dialog disappears and “System Reset Successful“ message is seen in the
feature status area, the system is ready to use.
– Attention
OC is initializing please wait...
Chapter 11
General Information
This section provides a simple introduction to CT, or Computed Tomography, for people with
no detailed physics or medical diagnostic education.
System components:
• The system components are explained in the Learning and Reference Guide.
GE Healthcare products are designed to provide optimum performance with GE-supplied
parts. Accordingly, GE can make no assurances that Equipment performance will not be
affected by the use of non-GE-supplied parts. In some instances use of non-GE-supplied
parts may affect Equipment performance or functionality.
To enhance user awareness when non-GE-supplied tube are in use, the Equipment has been
designed to recognize GE-supplied tubes and report to the user the presence of a
non-GE-supplied tube. This will permit the user to make any adjustments to Equipment use
that the user deems appropriate. Use of the Equipment with non-GE-supplied parts is
always at the user’s discretion. GE assumes no liability for the use of non-GE-supplied parts
and disclaims any responsibility for any affect such parts may have on Equipment
performance.
Emergency Stop:
• Emergency Stop procedures are described in the CT Safety tab.
CT Description
In conventional radiography, X-Rays pass through the patient to a film, which records
anatomic “shadows.” The X-Rays create a planar image, called a radiograph.
In computed tomography, electronic circuits detect and measure the X-Rays, and send
these measurements to a computer system that converts the information to a pixel value
matrix. These pixel values appear as a two dimensional image on a monitor. Even though
CT creates X-Ray exposures, we normally refer to them as images.
Additional computer software permits you to manipulate, shade, rotate, correlate, and
measure the images to derive even more information. The system also provides the means
to store the images on permanent or temporary media.
Conventional radiography can discern tissue density differences of 5%. CT can distinguish
density differences of 1% or less.
Traditional CT systems collect 1 row of data at a time. The LightSpeed Series CT scanner
system may improve customer productivity and open the door for new applications and
unparalleled scan speeds via a revolutionary 16-row data acquisition system.
CT Operation Theory
The PDU (power distribution unit) distributes system power. Power travels from the PDU to
the gantry. The components that produce X-Rays reside inside the gantry. The generator
produces high voltage to the X-Ray tube. High voltage propels electrons from the X-Ray
tube filament to its anode. Heat and x-radiation result.
The X-Ray tube's heat capacity and dissipation determine the frequency and length of CT
exposures. A Helical and Cine exposure can last up to 120 seconds and Axial exposures last
from 0.5 to 4 seconds.
The scintillator material in the detector absorbs the X-Ray that passes through the patient,
and generates a corresponding level of light. The detector converts the light levels into a
corresponding electric current. The DAS (Data Acquisition System) samples each detector
cell in all 16-row detector about 1000 times per rotation, amplifies and quantifies the
existing current, then sends the resulting data to the IG (Image Generator).
Each complete sample by the DAS is called a view. The recon engine converts all the views
into a single matrix of pixel values, called an image. The display processor takes a copy of
the digital matrix data, and converts it into television shades of gray, and sends the image to
the monitor for display. The OC (operator console) contains the monitor and controls the
computer, X-Ray, and cradle drives.
DICOM Print
The LightSpeed Series can send a camera request to a camera that has DICOM print
capabilities.
X-Ray
The X-Ray tube contains filaments, a cathode and an anode. The filament provides the
electrons that create X-Rays. The X-Ray system generates a current that heats the filament
until electrons start to “boil off” and break away from the filament. We refer to the filament
current as “mA.” Increasing the mA increases the number of electrons that become
available to make X-Ray. Higher concentrations of electrons improve image resolution.
The X-Ray system creates a high voltage, or kV, potential between the cathode and anode.
The negative charge on the cathode repels the electrons that boil off the filament. The
positive are grounded and attracts the negatively charged electrons. The electrons strike
the rotating anode target and displace electrons in the target material. This interaction
creates heat and X-Ray photons. The target rotates to help spread the heat over a larger
area. Increasing the kV increases the electron strike speed, which in turn increases the
intensity or “hardness” of the X-Ray photon beam.
Centered Patient
Gantry Opening
Detector
DAS
Tube Warmup
Warmup provides an automated group of low technique exposures designed to safely bring
the X-Ray tube to operating temperature before you start to scan for the day. Warmups
increase tube life and help produce more consistent, quality images.
System Characteristics
• New Gantry ID with 4/8/16 slice CT scanner and 16-row (4/8 slice) or 24-row (16 slice)
detector.
• LightSpeed Series optional variable rotation scan speeds (0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 2.0,
3.0, 4.0).
• Helical acquisitions at significantly faster table speeds.
• Potential for new applications due to faster coverage and fewer tube cooling delays.
Tube
Performix Ultra, Performix Plus, MAXIRay CT ULTRA
Detector for 16
• 24 rows in Z axis = 24 physically separated cells in Z
(HighSpeed CT/i = single cell).
• Detector cell segregation in Z provides post-patient collimation.
• Control functions:
– Detector FET array and heater control.
– Error detection and reporting.
Patient Scanning
EMI/EMC
All systems built to global regulatory emissions (EMC) and immunity (EMI) compliance
standards to improve reliability, uptime and performance in its intended environment.
Network
Remote Host Parameters
The LightSpeed Series Network function has new enhancements to support DICOM
networking. When adding or updating a remote list, there are some new parameters
needed. All of the following information, except for Comments, needs to be provided in order
to set up a remote host:
• The Host name to be entered is the name of the device. If the device is DICOM, the name
must match exactly to the name given to the device.
• The Network address of the device is provided by the institution's network administrator.
• The Network protocol is DICOM. If the LightSpeed Series will be sending to this device,
the device must be DICOM and the DICOM network protocol must be selected.
• DICOM protocol has a "dialect" which may cause some troubles like disappearances of
some portion of image annotation when connecting to the situation where data transfer
is not confirmed.
• The Port number is unique to the device. If the device is an Advantage Windows
workstation or another LightSpeed system, the number will be 4006.
• The AE Title is unique to the device. If the device is an Advantage Windows workstation
or another GE Healthcare system, the AE Title will be the same as the Host name.
• The Comment field allows you to input a comment.
Network Compatibility
The LightSpeed image format is DICOM. This image format may only be transferred between
systems using a DICOM network protocol. The receiving station must support DICOM receive
for LightSpeed images to be transferred (send or receive) to it.
Use the following table for network compatibility. The table lists the network protocol to use
and the features available for that system. The far left column lists the system the user is at
(from).
Table 11-1 Network Compatibility
From To
LightSpeed/ HiSpeed HiSpeed CT IC HiSpeed 3rd Party
BrightSpeedTM CT/i Advantage NX/i, X/i, or DICOM
QX/i Station
LightSpeed/ DICOM Query DICOM Advantage*** Advantage*** DICOM DICOM*
Send Query Query Query Query Query**
BrightSpeedTM Receive Send Receive Receive Send Send
Receive Receive Receive**
HiSpeed CT/i DICOM DICOM Advantage Advantage DICOM DICOM*
Query Query Query Query Query Query**
Send Send Send Send Send Send
Receive Receive Receive Receive Receive Receive**
HiSpeed DICOM Advantage Advantage Advantage Advantage DICOM
Advantage Send Query Query Query Send Send
Send Send Send
Receive Receive Receive
From To
LightSpeed/ HiSpeed HiSpeed CT IC HiSpeed 3rd Party
BrightSpeedTM CT/i Advantage NX/i, X/i, or DICOM
QX/i Station
CT IC DICOM Advantage Advantage Advantage Advantage DICOM
Send Query Query Query Send Send
Send Send Send
Receive Receive Receive
HiSpeed DICOM DICOM Advantage Advantage DICOM DICOM*
FX/i, DX/I, LX/i Query Query Query Query Query Query**
Send Send Receive Receive Send Send
Receive Receive Receive Receive**
3rd Party DICOM DICOM DICOM DICOM DICOM DICOM*
DICOM Query Query Query Query Query Query**
Send Send Receive Receive Send Send
Station Receive Receive Receive Receive**
* Some 3rd party stations use the ODINA network protocol. In this case use DICOM protocol
and port number 104.
** Query capability is only available only if station is a query retrieve provider.
*** Advantage Net is not available on PC Base Systems.
NOTE: BrightSpeedTM Elite/Edge/Excel, LightSpeed16, Ultra, Plus, QX/i, or HiSpeed QX/i PC
based systems do not support Advantage Network Protocol.
Table 11-2 Advantage Windows Network Compatibility
From
AW 1.X AW 2.X AW 3.X AW4.X LightSpeed or
BrightSpeedTM
LightSpeed or DICOM DICOM DICOM DICOM DICOM
HiSpeed QX/i Send Send Send Query Query
Send Send
Receive Receive
AW 1.X SdC Net SdC Net SdC Net SdC Net DICOM
Query Query Query Query Query
Send Send Send Send Send
Receive Receive Receive Receive Receive
AW 2.X SdC Net SdC Net SdC Net SdC Net DICOM
Query Query Query Query Query
Send Send Send Send Send
Receive Receive Receive Receive Receive
AW 3.X SdC Net V1 SdC Net V2 SdC Net V3 SdC Net DICOM
Query Query Query Query Query
Send Send Send Send Send
Receive Receive Receive Receive Receive
AW 4.X SdC Net SdC Net SdC Net SdC Net DICOM
Query Query Query Query Query
Send Send Send Send Send
Receive Receive Receive Receive Receive
Host
Computer DAS
Keyboard
ISO
Scan Plane
Z
Patient Table
Components
Figure 11-1 X-ray generation and detection components viewed from side of gantry
Anode/Target
Tube
Cathode
Bowtie
Cam Collimator Uncollimated X-Ray Beam
Tungsten Cams
Detector Collimator
16 Individual Lumex Cells
FET Array
Integrated Flex:
16 Columns x
2 Macro Rows=
Z 32 Signals/flex
CAM Collimator
The pre-patient collimation of the X-Ray beam is accomplished via 2 independently
controlled tungsten cams (refer to Illustration 11-2).
Focal Spot
Offset
Cam Offset
Beam Offset
Wide Beam Narrow Beam Narrow Beam
Centered Centered Cams Offset in Z
e.g. 4x5mm Slices e.g. 4x1.25mm Slices e.g. 4x1.25mm Slices
or
8x2.5mm Slices Offset used to compensate for
Z axis beam offset.
16x0.625 16x1.25
8A, 7A, 6A
, 5A, 4A, 3A, 2A, 1A 1B, 2B, 3B, 4B, 5B, 6B, 7B, 8B 8A, 7A, 6A
, 5A, 4A, 3A, 2A, 1A 1B, 2B, 3B, 4B, 5B, 6B, 7B, 8B
Tungsten Cams
Tungsten Cams Patient Table
Patient Table 10mm Collimation
20mm Collimation
8x1.25 8x2.5
4A, 3A, 2A
, 1A 1B, 2B, 3B, 4B 4A, 3A, 2A
, 1A 1B, 2B, 3B, 4B
Collimator Theory
The purpose of tracking is to follow the focal spot so that we can keep the uniform X-ray of
the narrowest possible beam on the detector to reduce dose and still avoid artifacts.
The focal spot moves in Z due to thermal changes in the tube and a mechanical forces
during gantry rotation and tilt angle.
Each cam is basically an independent system.
Scout
Scout imaging is used for anatomical location in conjunction with scan and recon
prescription, to provide an anatomical cross-reference for axial images, and to provide quick
feedback to the user as to the anatomy scanned. Scout supports the following features:
• Less than 10 second reconstruction.
• All kV and mA stations available, dependant on generator and tube limitations.
• 1.25 mm resolution in Z.
• 100 mm/sec table speed.
Pediatric Imaging
Adult techniques and protocols should not be used on pediatric patients (under 2 years of
age.)
Helical Overview
The 24-row detector and 16-row DAS provide the greatest benefits when used in the helical
mode. In the helical mode, data from 16 detector rows is selectively combined and weighted
during reconstruction in order to achieve the optimal balance between image z-axis
resolution, noise, and helical artifacts.
Helical imaging features include:
• All kV and mA stations available, dependant on generator and tube limitations.
• 120 second maximum helical scan time.
• Pitches:
NOTE: Beginning with the LightSpeed Series M3.1 release, the user interface has changed
the scan mode names. These new names indicate pitch as expressed by the ratio of
table travel per rotation to beam collimation. The new scan mode name designation
will be indicated on image annotation.
HQ 1:1
HQ 0.75:1
HS 1.5:1
UQ 0.625:1
UM 0.875:1
UF 1.35:1
US 1.675:1
Calibration Modes
LightSpeed Series calibrations include:
• Z-axis gain calibration - new for LightSpeed Series
• Tube warm-up
• Air calibration*
• Phantom calibration*, upgrades for 50cm SFOV
• Water Cal
• IQ Cal check
* = similar to HiSpeed, upgraded for 16 row data collection
Focal Spot
The X-Ray tube contains a small filament and a large filament. The small filament
concentrates the focal spot size, which improves spatial resolution but cannot tolerate high
technique. The large filament tolerates high technique but loses some of the small
filament’s spatial resolution.
The system automatically selects the small filament:
• When the technique equals 24kW, or less.
The system automatically selects the large filament:
• When the technique exceeds 24kW.
Example: Reduce mA setting from 210mA to 200mA (120kV) to enable Small Filament.
Filament Selection
On traditional single-slice CT systems, such as HiSpeed CT/i, the filament selection effects
the slice thickness for thin slices scanning, with a larger filament producing a larger effective
slice thickness. The LightSpeed Series has eliminated this effect, and true thin slice imaging
can be achieved using both the small and large filament.
Data Collection
The detector and DAS assembly mounts opposite the X-Ray tube on the rotating base. The
X-Ray beam leaves the X-Ray tube, passes through the gantry opening, and enters the
detector. Any material (patient or phantom) positioned within the gantry opening absorbs or
deflects the weaker X-Ray photons. The numbers of photons that enter the detector
depends upon the intensity of the X-Ray beam and the density of the material in the gantry
opening. An increase in density causes a decrease in the number of photons that enter the
detector.
The revolutionary new LightSpeed detector allows 16 rows of data to be collected at a time
for both axial and helical imaging. This allows 4 axial images to be generated in a single
gantry rotation in the axial mode. This allows helical images to be taken at faster speeds,
and with lower power, than single slice scanners.
The DAS measures the detected X-Ray at regular time intervals, called views, and transmits
the information to the image reconstructor for reconstruction into a display image. The
total degrees of gantry rotation and the scan time determine the number of raw views per
image.
For scans greater than 1 second, the raw views are summed together before image
reconstruction. This allows the system to maintain constant image reconstruction times and
spatial resolution/aliasing for all scan speeds. (Note that image reconstruction of more raw
views reduces aliasing at the expense of reconstruction time.)
Example: A 1-second and a 4-second scan both gather data over 360 degrees, so both
scans reconstruct the same number of views per image. A segmented reconstruction uses
data acquired over 235 degrees, so it reconstructs fewer views per image.
Scan Parameters
Reconstruction
The scanner compares the collected data with the calibration data then converts the
detector channel views into a two dimensional matrix. The system converts each matrix
element (pixel) into a CT number. The system displays a scout image as it acquires the data,
but an axial image will be displayed after reconstruction is completed. The axial and scout
images take approximately one second to reconstruct.
Reconstruction Algorithm: Soft, Std, Detail, Bone, Lung, Edge, Bone Plus.
Your choices control the image outcome. Choose parameters to enhance or tailor the
acquisition and processing to the anatomy of interest. Select scan technique and image
parameters that provide optimum resolution.
The system has disk space for 1500 scan files or 2400 (16 slice) scan rotations. The system
stores the most recent scan data in the oldest scan file with an unreserved status. The
system continually overwrites the scan files with data.
CAUTION: If you plan to reconstruct images, you must use files that reside in the disk.
Either reserve the scan files you plan to retrospectively reconstruct, or
reconstruct unsaved scan files before the system overwrites the files with
new scan data. The system refuses to overwrite reserved scan files.
Remember to release the reserved scan files when you finish retrospective
reconstruction.
Selected Slice
Scan Table
Mode Speed 0.625mm 1.25mm 2.5mm 3.75mm 5.0mm 7.5mm 10mm
(mm/rot)
Axial N/A 0.60 1.09 2.34 3.75 4.84 7.35 9.84
Helical 5.625 0.63 1.25 2.50 3.75 5.00
0.562:1 11.25 1.25 2.50 3.75 5.00 7.50 10.00
Helical 9.375 0.85 1.25 2.50 3.75 5.00
0.938:1 18.75 1.6 2.50 3.75 5.00 7.50 10.00
Helical 13.75 0.80 1.25 2.50 3.75 5.00
1.375:1 27.5 1.60 2.50 3.75 5.00 7.50 10.00
Helical 17.5 0.85 1.25 2.50 3.75 5.00
1.75:1 35.0 1.60 2.50 3.75 5.00 7.50 10.00
Selected Slice
Scan Table
Mode Speed 0.625mm 1.25mm 2.5mm 3.75mm 5.0mm 7.5mm 10mm
(mm/rot)
Helical 6.25 1.25 2.50 3.75 5.00 7.50 10.00
0.625:1 12.5 2.50 3.75 5.00 7.50 10.00
Helical 8.75 1.25 2.50 3.75 5.00 7.50 10.00
0.875:1 17.5 2.50 3.75 5.00 7.50 10.00
Helical 13.5 1.60 2.50 3.75 5.00 7.50 10.00
1.35:1 27.0 3.20 3.75 5.00 7.50 10.00
Helical 16.75 1.70 2.50 3.75 5.00 7.50 10.00
1.675:1 33.5 3.20 3.75 5.00 7.50 10.00
Selected Slice
Scan Table
Mode Speed 0.625mm 1.25mm 2.5mm 3.75mm 5.0mm 7.5mm 10mm
(mm/rot)
Helical 5.625 0.75* 1.50 3.00 4.50 6.00
0.562:1 11.25 1.50 3.00 4.50 6.00 9.00 12.00
Helical 9.375 0.92** 1.50 3.00 4.50 6.00
0.938:1 18.75 1.90 3.00 4.50 6.00 9.00 12.00
Helical 13.75 1.0 1.50 3.00 4.50 6.00
1.375:1 27.5 1.90 3.00 4.50 6.00 9.00 12.00
Helical 17.5 1.0 1.50 3.00 4.50 6.00
1.75:1 35.0 1.90 3.00 4.50 6.00 9.00 12.00
Helical 6.25 1.50 3.00 4.50 6.00 9.00 12.00
0.625:1 12.5 3.00 4.50 6.00 9.00 12.00
Helical 8.75 1.50 3.00 4.50 6.00 9.00 12.00
0.875:1 17.5 3.00 4.50 6.00 9.00 12.00
Helical 13.5 2.00 3.00 4.50 6.00 9.00 12.00
1.35:1 27.0 3.00 4.50 6.00 9.00 12.00
Helical 16.75 2.00 3.00 4.50 6.00 9.00 12.00
1.675:1 33.5 3.00 4.50 6.00 9.00 12.00
When a pitch value significantly below 0.125 is chosen, the slice profile may vary
significantly from image to image in a single helical scan. The profile will depend on the
location of the prescribed image with respect to the detector locations. The variation in the
slice profiles can be understood by thinking of an axial scan. In an axial scan, an arbitrary
image location is not permitted. However, an attempt to create an image at an intermediate
location could be performed by using data from adjacent detector rows. This would result in
an increased slice profile. As the pitch becomes much less than 2.0, the helical scan data
becomes more and more like axial scan data.
At a pitch of 1.4, for example, which is the prescribed pitch for a Cardiac Segment scan with
a patient heartrate of 60 bpm and gantry period of 0.5 sec, a 35% variation in the nominal
FWHM is possible.
Calibration Scans
Calibrations scans of air, and uniform objects called phantoms, provide the baseline
information the system needs to produce patient images. The system needs calibration
data for every possible combination of kV, detector row thickness, focal spot size, and scan
field of view.
Warm-up Required
Warm-up the tube:
• Immediately before Cal Check.
• Immediately before Calibration.
• When the tube has cooled to the point that a warm-up is required to ensure optimal
image quality.
Data Storage
NIO16 console
The Console/Computer contains 292 gigabytes (Gb) of magnetic disk that records and
retains 1500 scan files or 9600 (16slice) scan rotations, a Reconstruction Processor that
processes scan data into image data, and a magnetic disk that stores CT specific scan
software.
The computer contains 146 gigabytes (Gb) of system disk that hold about 250,000
uncompressed 5122 image files, along with software.
Despite storage space, the system eventually runs out of disk space. If your facility plans to
preserve image data, you must periodically transfer images and scan information to the
designated archive media.
GOC4 console
The Console/Computer contains 36 gigabytes (Gb) of magnetic disk that records and retains
2900 1-second scan rotation files, a Reconstruction Processor that processes scan data into
image data, and a magnetic disk that stores CT specific scan software.
The computer contains 73 gigabytes (Gb) of system disk that hold about 130,000
uncompressed 5122 image files, along with software.
Despite storage space, the system eventually runs out of disk space. If your facility plans to
preserve image data, you must periodically transfer images and scan information to the
designated archive media.
Image Display
Requested images pass through the IP (image processor) on their way to the screen. The
Image Processor uses a bulk memory to store images selected for Auto View, MID, paging,
magnification, rotation, reformat or 3D (3D optional).
The images appear on the image monitor. The screen contains a display matrix of 1024 x
1024 picture elements, or 1,048,576 pixels. The 1024 display can be further divided into
viewports. The number of viewports displayed determine the number of pixels within a
viewport. Each pixel displays one of the 256 available shades of gray.
The LightSpeed Series system reconstructs axial and continuous images of 5122 pixels.
Images from other scanners may display 64, 128, 320, or 1024 pixel image matrices.
The amount of anatomy represented by each pixel equals the Display Field of View diameter
in mm divided by the matrix width/height.
The system assigns a unique CT number value, originally called a Hounsfield Unit, to each
pixel. The two dimensional pixel represents a three dimensional portion of patient tissue.
The pixel value represents the proportional amount of X-Ray beam that passed through
anatomy and entered the detector.
Gray Scale
The monitor translates the computed pixel value into a shade of gray. Your window Width
and Level choices control which range of CT values receive emphasis. The window Width
assigns the quantity of pixel values to the gray scale. The window Level determines the
center pixel value in the gray scale.
Level
CT Number
Nominal CT numbers range from -1024 to +3071 and extended CT numbers range from
-31743 to +31743. The LightSpeed Series is defaulted to the extended CT number range.
This system references CT number zero to water and CT number -1000 to Air.
Lung and fat have negative pixel values and normally appear black (nominal CT numbers
range). A CT number over 200 represents dense material like contrast agent, calcium, bone,
and normally appears white (nominal CT numbers range).
Inverse Video reverses video white to black, but pixel values remain the same.
CAUTION: CT Numbers are NOT absolute; misdiagnosis is possible. System and patient
variables may effect CT Number accuracy. If you rely solely upon CT numbers
without taking the following variables into consideration you could
misdiagnosis an image.
The following variables effect CT Number accuracy:
• Partial volume effects of anatomy
• Scans acquired with IV or oral contrast agents
• X-Ray tube deterioration
• Improperly calibrated system (poorly centered phantom, used wrong phantom,
replaced current calibration files with extremely old Cal files)
• Beam hardening due to patient anatomy, especially bone.
Pixels
The anatomic image consists of rows and columns of small, square, picture elements called
pixels. The screen displays 1,048,576 pixels in a matrix of 1024 horizontal rows of 1024
pixels. Add number of viewports selected for viewing to determine the number of pixels used
for display in each viewport. The screen pixel size remains the same, but the amount of
anatomy the pixels represent varies with the scan and display field of view (SFOV & DFOV). A
pixel also represents a specific anatomic area. The system identifies each two dimensional
pixel by its location, area and value.
Pixel Coordinates
Describe pixel location two ways.
axial image
RAS Coordinates
These three distances in millimeters appear on the upper left of the viewport on which the
mouse cursor is on, when Continuous Report Cursor is selected.
• The pixel with the R/L and A/P coordinates closest to zero, represents the SFOV center.
The S/I coordinate always equals the table location at isocenter.
axial plane
axial image
• The 50cm FOV has a 25cm radius, so its area equals 1963cm2.
• The ROI or magnification factor determines the anatomic area covered by a magnified
image.
Example: A pixel represents 0.5 by 0.5mm. Magnify pixel size by 2. Each pixel now represents
0.25 by 0.25mm of anatomy.
Pixel area
equals
pixel size
pixel size squared
(width)
in mm
512
Matrix
DFOV in mm
Reformat displays non axial planes created from contiguous pixels extracted from
multiple images. 3D locates similar pixel values within contiguous images, and generates a
mathematical model to produce images that appear three dimensional. BMD samples pixel
values to estimate bone or tissue density.
Reconstruction assigns one value to every image pixel. CT uses pixel values of -1024
to +3071 (nominal range) or -31743 to +31743 (extended range). MR uses pixel values of
+16,000. The screen pixel translates the assigned value into one of the 256 shades of gray.
Vary the gray scale window width and level to select anatomy for display. Window Width
determines the quantity of gray pixel values. Window Level selects the center Window
Width pixel value.
Example: Two windows may contain identical widths of 100 values, but display completely
different anatomy, because one has a level of -100 and the other has a level of 150
Window Width
The system uses 256 gray shades to display 4000 CT pixel values (nominal range). The
Window Width selection determines the number of CT values represented by each shade of
gray. A narrow window assigns fewer pixels to each gray level than a wide window.
Example: WW = 256 System assigns one pixel value to each gray shade WW = 2560 System
assigns ten pixel values to each gray shade.
Enlarge the window to display anatomy that includes air, tissue, and bone, and subtle
differences in the tissue vanish, because the system assigns too many pixel values to each
shade of gray.
Narrow the window, then adjust the level to display only soft tissue. Subtle differences in
tissue density appear, because the system assigns fewer pixel values to each shade of gray.
Adjust the window whenever the display anatomy changes. Large window widths display a
range of anatomic structures and densities, while narrow windows enable subtle density
discrimination within the anatomy of interest.
Window Level
The Level equals the CT number value of pixel in the center of the Window Width range. The
Level value receives the middle shade of gray. The system displays pixel values that fall
between the center and upper window level as gray to off white. It displays pixel values that
fall between the center and lower window values as gray to charcoal. When you change the
level, the window width moves up and down the CT number line. The CT values change with
Window Level, but the Window Width and number of pixels per gray level don’t change.
Inverse Video reverses display conventions. Dense or high numbers are portrayed as black
rather than white.
Chapter 12
Quality Assurance
In order to assure consistent image quality over the system's lifetime, establish and
maintain a regular Quality Assurance (QA) program.
Constancy testing of the system should be performed in accordance with IEC 61223-2-6
or per your facility's specific QA program.
Scan a known material (usually a phantom) under a prescribed set of conditions.
• Compare the results to previous or optimum values.
• Repeat these tests on a regular basis to detect changes in image quality values before
the problem becomes visible.
– If you notice a degradation in image quality, or a change in QA values, you can
schedule a site visit and let the service person or imaging physicist run more detailed
tests.
– Early intervention could prevent a major breakdown.
Quality Assurance begins with baseline performance data acquired during system
installation, or after the repair or replacement of an X-Ray tube, collimator, detector, DAS, or
PDU circuitry.
• Compare subsequent QA results against the baseline.
• You can save baseline images for a visual comparison with your daily QA checks, but the
measurement values provide a more objective way to monitor quality.
NOTE: Copy the QA Data Form found at the end of this section. Use the form to record
baseline data and subsequent QA data.
• Contrast Scale
• High Contrast Spatial Resolution
• Low Contrast Detectability
• Noise and Uniformity
• Slice Thickness
• Laser Light Accuracy
The QA phantom contains three sections, each corresponding to a single scan plane.
• Section 1: Resolution block S0mm scan location
• Section 2: Contrast membrane S40mm scan location
• Section 3: Water bath S60mm scan location
The following illustration contains a list of the sections and corresponding tests.
Location: S0mm
Section 2
Location: S40mm
Section 3
Location: S60mm
4
5
Number Description
1 Circumferential Reference Line
2 Vertical Reference Line
3 Horizontal Reference Line
4 Horizontal Reference Line
5 Phantom Bracket
6 Phantom Holder
NOTE: Reference Lines are etched into the plastic and are unpainted.
• When positioning or removing the QA Phantom from the Phantom Holder, the patient
cradle should be moved out of the gantry fully (home position) and lowered enough so
the user can lift it straight up and off the holder. If the table is not moved to this position,
and the user reaches into the bore to adjust the QA Phantom, stress can cause the
bracket to break.
• If the QA Phantom is broken, cracked or leaking, it should be discarded immediately and
replaced with a new QA Phantom. Care should be taken to avoid any sharp edges which
may develop on the broken pieces, which can potentially cause injury. If the QA
Phantom is leaking water onto the floor, there is a potential slip and fall hazard. If the QA
Phantom is leaking while on or near the scanner, it is possible for water to get under the
table or gantry covers and cause damage.
Prescribe the QA Series for the Resolution, Low Contrastability, and Noise and
Uniformity Tests
1. Click [New Patient] to display the Patient/Exam Parameters screen.
Use the same ID for all related QA tests so you can store the exams together.
2. Enter any additional information in the corresponding data field(s). (Optional)
3. Exam Description: Enter up to 22 characters to describe the test. (Recommended)
4. Select a protocol from the anatomical selector to select a QA protocol. (If available)
If your facility hasn't created a QA protocol, use the following parameters to finish the QA
series prescription:
On the Helical View Edit screen select the following parameters:
Table 12-2 Parameters for QA
Interface Input
Entry Head First
Position Supine
Anatomical Reference QA
Landmark Location 0 on resolution phantom at circumferential line/cross hatch.
Scan Type Helical 0.562:1
Scan Range I0 - S60
Thickness 10
Table Speed 11.25
Recon Interval 10
Tilt 0 degrees
SFOV Small
kV 120
mA 135
Rotation Speed 1 second
Interface Input
DFOV 25cm (phantom diameter:approximately 21.5cm)
Standard and Bone to test high contrast resolution.
Algorithm Standard to test low contrast detectability
Standard for pixel value noise and uniformity
Matrix 512
Contrast None
Special Processing None
For test purposes, the CT values of water and plexiglass in the phantom represent the
standard against which you track the system contrast scale over time.
The test for contrast scale follows:
Figure 12-3 Contrast Scale Phantom Section
1. Select Box ROI to position a 10mm box cursor on the image, as shown in Figure 12-3.
2. For consistency, use the same size cursor and location each time you perform this test.
3. Select Grid to provide a reference.
4. Select Box ROI to position a cursor over the Plexiglass resolution block (refer to Figure
12-3).
5. Record the mean CT number on the QA Data Form.
6. Optional: Record the Standard deviation
7. Select Box ROI to position a cursor over water (refer to Figure 12-3).
8. Record the mean CT number for water on the QA Data form.
9. Optional: Record the Standard deviation
10. Subtract the CT number of water from the CT number of Plexiglass
Record the difference on the QA Data form.
The difference should equal 120 ±12.
The resolution block contains the following bar sizes:1.6mm, 1.3mm, 1.0mm, 0.8mm, 0.6mm,
and 0.5mm.
1. Examine the bar patterns to determine the limiting resolution, defined as the smallest
bar pattern in which you see all five bars.
2. You should see all five 0.6mm bars in images reconstructed with the Bone algorithm
(15cm FOV).
3. Using the standard algorithm, measure the standard deviation of the pixel values in a
single or multiple bar pattern to provide a quantitative method for assessing changes in
system resolution.
4. ROI standard deviation provides a good indicator of system resolution and a sensitive
method to detect changes in system resolution.
The recommended procedure follows:
Figure 12-4 High Contrast Spatial Resolution Section
MTF (optional)
The Modulation Transfer Function (MTF) mathematically quantifies high contrast
resolution.
MTF measures the contrast preserved for a sine wave pattern as a function of frequency.
An MTF curve begins at 1 for zero frequency, and decreases as frequency increases.
Example: An MTF of 1 equals total preservation of contrast
Example: An MTF of 0.5 equals 50% loss of contrast
The limiting resolution equals the frequency at which MTF falls to 0.
An MTF curve is shown in Figure 12-16.
Measure Frequency in line pairs per centimeter.
One line pair per centimeter equals one 5mm Plexiglass bar next to one water filled 5mm
space.
Optional: Consult the publication listed as Reference 1 in section for MTF measurement
instructions (Droege RT, Morin RL. “A Practical Method to Measure the MTF of CT Scanners,”
Medical Physics, Volume 9, No. 5, pp 758-760, 1982).
Subtract B from A.
Count visible holes. Subtract D from C.
Image noise equals the standard deviation of CT numbers within a region of interest (ROI).
Noise results from electronic, mechanical, and mathematical differences in detected X-Ray
energy, electronic outputs, and reconstruction algorithms.
Tube Warmups, up to date calibration files, and daily Fastcals minimize noise and help
provide uniform images.
Refer to Figure 12-6.
1. Select Erase to remove previous ROI data.
2. Select Box ROI to position a 2 cm box ROI over the center of the image.
3. Record the mean CT number and standard deviation on the QA Data Form.
QA data form (Figure ) is at end of this section.
4. Optional: Select Box ROI and move the cursor to the 12 o'clock position.
5. Record the mean CT number and standard deviation on the QA Data Form.
6. Optional: Select Box ROI to move the cursor to the 3 o'clock position.
7. Record the mean CT number and standard deviation on the QA Data Form.
If the Image is reconstructed with Standard algorithm and Small SFOV, the Mean of
Center ROI should equal 0 ± 3.
Standard deviation of the center ROI should less or equal 3.5 (nominal 3.2)
The uniformity difference between the center ROI and the average of the edge ROIs
should be 0 ± 3.
Slice Thickness
Section 1 of the phantom also tests slice thickness.
Both sides of the resolution block contain a pattern of air filled holes designed to
demonstrate slice thickness (refer to Figure 12-7).
Figure 12-7 Slice Thickness Section
Air filled
holes
The resolution block contains holes drilled 1mm apart and positioned to form a line at 45
degrees to the scan plane.
Each visible hole in the image represents 1mm of beam thickness.
The software assigns less negative CT numbers to partial hole images or holes located on
the edge of the slice profile.
Prescribe the QA Series for the Slice Thickness Test - Phantom Section #1
1. Select a protocol from the anatomical selector to select a QA protocol (if available).
2. If your facility hasn't created a QA protocol, use the following parameters to finish the
QA series prescription:
3. On the Axial View Edit screen select the following parameters:
Table 12-3 QA Protocol for Slice Thickness
Interface Input
Entry Head First
Position Supine
Anatomical Reference QA
Landmark Location 0 on resolution phantom at circumferential line/cross hatch.
Scan Type Axial
Scan Range Prescribe 4 Groups
Thickness
Group Scan Range Spacing
(For 16 slices)
1 5mm/4I I7.5 - S7.5 0
2 3.75mm/4I I5.625 - S5.625 0
3 2.5mm/8I I8.75-S8.75 0
4 1.25mm/8l I4.375-S4.375 0
5 0.625mm/16I I4.688-S4.688 0
6 2.5mm/4I I3.75-S3.75 0
Tilt 0 degrees
SFOV Small
kV 120
mA 260
Rotation Speed 1 second
DFOV 25cm (phantom diameter:approximately 21.5cm)
Algorithm Standard
Matrix 512
Contrast None
Special Processing None
You should see one line for each millimeter of scan thickness.
Figure 12-8 represents a 5mm image.
Interface Input
Entry Head First
Position Supine
Type EX for the external alignment light test
Anatomical Reference
Type IN for the internal alignment light test.
Landmark Location 0 on resolution phantom at circumferential line/cross hatch.
Scan Type Axial
Scan Range I3.75 - S3.75
Thickness 2.5 mm/4i
Tilt 0 degrees
SFOV Small
kV 120
mA 260
Rotation Speed 1 second
DFOV 25cm (phantom diameter:approximately 21.5cm)
Algorithm Standard
Matrix 512
Contrast None
Special Processing None
kV 120
mA 260
Small
5.0mm
Tilt 0.0 R0.0
P125.0
W = 250 L = –50
DOSIMETRY
Dosimetry information is provided in terms of the CTDI and CTDIw dose indices. Optionally
CTDIvol and its associated DLP (dose length product) is automatically computed and
displayed on the patient Rx menu to assist in managing patient dose. This section provides a
brief description to help you better understand these dose reporting standards.
General Information
Dose is the amount of energy imparted by the X-ray beam at a given point in an exposed
material (patient tissue, phantom, air, etc. ) and is measured in units of mGy (milli Gray). (The
old unit was the Rad which equals 10 mGy.) Dose is dependent on the energy absorption
factors of the material and on the X-ray exposure. The X-ray exposure is measured in C/kg
(coulombs per kilogram) and is dependent on the technique factors used for the scan. The
dose is generally proportional to the exposure which increases with increasing mA, kV and
scan time. The X-ray exposure to a point occurs from both direct X-ray from the tube and
from scattered X-ray due to adjacent material exposure.
Patient biological risk is related to dose but is also highly dependent on the specific organs
exposed. The effective dose is a way to characterize patient risk. The effective dose is the
sum of the doses weighted in accordance with the specific radio-sensitivity of the particular
organs or tissues exposed. Weighting values are published in ICRP 60 (International
Committee on Radiation Protection, Publication 60). Although we can accurately describe
the X-ray exposure potential to a patient for a CT scan, we can not easily determine the
patient dose or risk in terms of effective dose. This is because each patient is anatomically
unique and the specific details of his or her anatomy along with the source exposure must
be processed using time consuming monte-carlo computer programs to predict how
radiation will be scattered and accumulated within various patient organs.
Since it is not possible to characterize the specific dose given to individual patients, the CT
dose indices are provided to help make relative comparisons. These dose index values can
be used to compare CT systems and to help select appropriate operating conditions for
scanning. However, it is important to recognize that the dose reported by these indices is
inversely proportional to phantoms size (see Figure 12-11). This means that for the same
scan technique (protocol), smaller phantoms (patients) will produce a higher absorbed dose
than larger phantoms (patients) - see “Influence of phantom diameter, kVp and scan mode
upon computed tomography dose index”, Edward L. Nickoloff, Ajoy K. Dutta, and Zheng F. Lu,
Medical Physics 30, 395 (2003). Therefore, it is critical to remember that the body scan
FOV’s uses the 32cm CTDI phantom and all pediatric and head filter uses the 16cm CTDI
phantom for dose reporting purposes (CTDIVOL display on Scan Rx Menu). Table 12-6
indicates the phantom size used for each SFOV.
Figure 12-11 Relationship between dose and phantom size for head and body filters at
120kVp. Similar curves are obtained for the 80, 100, and 140 kVps.
1 2
Table 12-5
Number Description
1 Head Filter
2 Body Filter
3 Relative Dose
4 Phantom Diameter (cm)
Table 12-6
CTDIw is measured using either a 16 cm (for head scanning) or a 32 cm (for body scanning) PMMA
phantom of at least 14 cm in length. The measurements are taken at the center and peripheral (see
Figure 12-12 points A and B). The doses measured at these locations within the PMMA phantom, are
quoted as the dose absorbed in air rather than PMMA (absorption in air is about 11% higher than
absorption in PMMA).
Table 12-7
The dose at the four peripheral locations is very similar due to the geometry of the system
and phantom. CTDI values are reported only for peripheral location “B” as this conservatively
represents all-peripheral dose values.
CTDI100 and CTDIw dose tables and index factors are provided in the following section. To
determine the CTDI100 dose, select the appropriate standard technique dose (small or large
filters) and multiply by the factors for describing the technique used.
(Typical techniques for CTDI100 and CTDIw are the same as for CTDI stated above)
Table of CTDI100 Dose Values (mGy) at Typical Technique (For 4/8 slice)
Head A Head B Body A Body B
46.75 47.71 14.35 29.33
Table of CTDI100, CTDIw, KV, SCAN MODE, and mAs ADJUSTMENT FACTORS
KV Adjustment Factor
KV Head A Head B Body A Body B
80 0.34 0.37 0.27 0.33
100 0.63 0.65 0.57 0.61
120 1 1 1 1
140 1.37 1.36 1.44 1.36
mAs ADJUSTMENT FACTOR = Rx mA * Rx single rotation time in seconds/ 260
CTDI100 Aperture Adjustment Factors for Large Spot (>24kW) (For 16 slice)
Acquisition Mode Head A Head B Body A Body B
16 x 1.25 1.00 1.00 1.00 1.00
16 x 0.625 1.24 1.24 1.25 1.24
8 x 2.50 1.01 1.00 1.00 1.03
8 x 1.25 1.26 1.27 1.26 1.25
4 x 3.75 1.08 1.08 1.08 1.09
4 x 2.50 1.15 1.15 1.16 1.18
4 x 1.25 1.53 1.54 1.54 1.57
4 x 0.625 2.08 2.11 2.14 2.14
2 x 0.625, 1 x 1.25 2.55 2.60 2.58 2.60
CTDI100 Aperture Adjustment Factors for Small Spot (<24kW) (For 16 slices)
Acquisition Mode Head A Head B Body A Body B
16 x 1.25 1.01 1.01 1.01 1.01
16 x 0.625 1.13 1.14 1.13 1.13
8 x 2.50 0.96 0.96 0.97 0.96
8 x 1.25 1.16 1.17 1.16 1.15
4 x 3.75 1.08 1.08 1.08 1.09
4 x 2.50 1.11 1.10 1.11 1.13
4 x 1.25 1.41 1.43 1.46 1.49
4 x 0.625 1.95 1.98 2.00 2.03
2 x 0.625, 1 x 1.25 1.22 1.26 1.22 1.28
CTDI100 Aperture Adjustment Factors for Large Spot (>24kW) (For 4/8 slice)
Acquisition Mode Head A Head B Body A Body B
8 x 2.5 1.00 1.00 1.00 1.00
8 x 1.25 1.17 1.16 1.16 1.14
4 x 5.0 1.00 1.00 1.00 1.00
4 x 3.75 1.02 1.02 1.02 1.02
CTDI100 Aperture Adjustment Factors for Large Spot (>24kW) (For 4/8 slice)
Acquisition Mode Head A Head B Body A Body B
4 x 2.50 1.10 1.09 1.09 1.08
4 x 1.25 1.47 1.46 1.46 1.45
1 x 1.25 3.37 3.37 3.50 3.50
CTDI100 Aperture Adjustment Factors for Small Spot (<24kW) (For 4/8 slices)
Acquisition Mode Head A Head B Body A Body B
8 x 2.5 0.92 0.92 0.91 0.90
8 x 1.25 1.09 1.09 1.09 1.08
4 x 5.0 0.92 0.92 0.91 0.90
4 x 3.75 1.02 1.01 1.01 1.00
4 x 2.50 1.03 1.03 1.02 1.02
4 x 1.25 1.29 1.29 1.29 1.29
2 x 0.63 1.46 1.50 1.47 1.49
1 x 1.25 2.95 2.95 2.95 2.95
Example 1 - The CTDI100 body peripheral dose for a 8.75mm/sec helical scan in 0.875:1 pitch,
8x1.25 slice mode, 250mA, 1.0s gantry speed and 120kV, is determined as follows:
Example 2 - The CTDI100 body center dose for example 1 is determined as follows:
Example 3 - The CTDIw body dose for example 1 and 2 is computed as:
CTDIw = (2/3)*34.94 + (1/3) *17.07 = 28.99 mGy
N T
CTDIvol = d
CTDIw
where N is the number of slices produced in a single axial scan, T is the slice thickness (or
row detection width), and ∆d is the table travel in z-direction between consecutive scans.
b) For helical scanning
CTDIw
CTDIvol = Scan Mode Adjustment Factor (Pitch)
Example 4 - The CTDIvol head dose for a 10mm 1i, 8x1.25 axial scan, at 150mA, 1.0s gantry
speed, 120kV and a 30mm table increment between consecutive scans, is determined by:
Example 5 - The CTDIvol for a cardiac helical scan using the large SFOV, 16x0.625mm slice
mode, with helical pitch 6.0, 250mA, 0.5s gantry speed, and 140kV is determined as follows:
14.09 (body A), 29.07 (body Center, peripheral dose from CTDI100 Table
B)
x 1.44 (body A), 1.36 (body B) 140kV adjustment factor from Table
x 1.25 (body A), 1.24 (body B) 16x0.625 aperture adjustment factor for large spot from Table
(at 120kV, 250mA > 24 kW)
x 125/260 mAs adjustment factor (250mAx 0.5s)/(260mAx1s)
=12.19mGy, 23.57mGy Center and peripheral CTDI100
= (1/3)x12.19 + (2/3)x23.57 CTDIw
= 19.78 mGy
x 2.67 Cardiac helical “pitch factor” = 2.67 from Cardiac Helical Pitch
Factor Table
1.The following article provides an explanation of the relationship between CTDIvol and peak skin
dose for helical and cine-mode exams: J. A. Bauhs, T. J. Vrieze, A. N. Primak, M. R. Bruesewitz, and
C. H. McCollough, 2008, "CT dosimetry: comparison of measurement techniques and devices," Ra-
diographies Vol. 28, pp. 245-253. Available at: http://radiographics.rsna.org/content/28/
I/245.full.pdf+html.
Maximum peak of dose profile (max {Dperipheral (z)}) = Peak dose from Table 12-30* (actual
scan mA/260mA) * gantry rotation speed in seconds.
Computed Tomography Skin Dose Index (CTSD) is an indicator of the peak absorbed dose in
skin tissue located within the area of coverage of the radiation field.
For Axial scanning:
peripheral
NOTE: For axial scanning with a total table travel of less than N x T, this definition may
overestimate the dose.
NOTE: For the selected CT condition of operation, but irrespective of any scanning length
that may be used clinically, the axial-scanning CTSD is an index of dose based on a
convention of 100 mm range of integration along the z-axis. For axial scanning, CTSD
corresponds to the dose averaged over the phantom central section peripheral holes.
For Helical scanning:
peripheral
CT pitch factor
NOTE: CT pitch factor will be a function of time when d or N x T are variable during the
exposure.
NOTE: For helical scanning with a small number of rotations and a table travel per rotation
of less than N x T, this definition may overestimate the dose.
NOTE: For the selected CT condition of operation, but irrespective of any scanning length
that may be used clinically, the helical-scanning CTSD is an index of dose based on
a convention of 100 mm range of integration along the z-axis. For helical scanning,
CTSD corresponds to the dose averaged over the phantom central section peripheral
holes.
For scanning without movement of the patient support:
max{Dperipheral(z)}
NOTE: Includes situations where the patient support may be moved manually, for example,
during an interventional procedure.
For Axial and Helical scanning involving table travel in two derections (“shuttle” mode):
max{Dperipheral(z)}
kV SFOV 2×0.62
16×1.2 16×0.6 4×0.62
8×2.50 8×1.25 4×3.75 4×2.50 4×1.25 5,1×1.2
5 25 5
5
Head 71 58 72 56 66 62 46 34 27
80
Body 134 109 131 108 124 114 86 63 52
kV SFOV 2×0.62
16×1.2 16×0.6 4×0.62
8×2.50 8×1.25 4×3.75 4×2.50 4×1.25 5,1×1.2
5 25 5
5
Head 40 33 41 32 38 35 26 19 15
100
Body 72 59 71 58 67 62 46 34 28
Head 26 21 26 21 24 23 17 12 10
120
Body 44 36 43 35 41 37 28 20 17
Head 19 15 19 15 18 17 12 9 7
140
Body 32 26 31 26 30 27 20 15 12
LightSpeed16 Slice
Scout Dose
Generally, because of short scan times and low mA, the scout dose will be a small part of the
total patient exam dose, additionally a standardized scout dose calculation method has yet
to be developed for CT, therefore scout dose is not currently reported by the system.
GE used a 21.5cm water filled acrylic phantom to measure all Head performance tests.
Testers placed a 30cm wide acrylic ring around the water phantom to measure Body
performance.
Noise
Statistically measure the CT numbers represented by an array of pixels contained in a 2 x
2cm central region of interest (ROI)
Noise equals the standard deviation expressed in Hounsfield units, divided by 1000 to
represent the contrast scale between air and water, then multiplied by 100 to give a value in
percent.
Noise
1020.33 (C) (3) i
The Standard Deviation In the Center scanned with Typical Technique and Standard 512
Recon at 5mm, 4i mode.
Head Body
0.5% 1.3% 260mAs 120 kVp
GE used the Small Body SFOV, rather than the Head SFOV.
Head: Use 25cm FOV
Body: Use 35cm FOV
Both SFOVs use similar resolution algorithms except the Small Body SFOV does not use the
bone/brain interface correction, which the phantom does not need.
The Small Body and Head SFOV produce negligible image performance differences.
For digital form of representative images, can refer to the following steps to find them on
the software UI.
1. On Display Monitor, select [Service] to show the Common Service Desktop. Then, click
the [Install 21CFR Images] under [Image Quality].
Dose Profile
Slice Thickness Profile
3.75mm, 4i mode
Head
Dose Profile
Slice Thickness Profile
Dose Profile
Slice Thickness Profile
Dose Profile
Slice Thickness Profile
Dose Profile
Slice Thickness Profile
3.75mm, 4i mode
Body
Dose Profile
Slice Thickness Profile
Dose Profile
Slice Thickness Profile
Dose Profile
Slice Thickness Profile
0.8
0.6
mtf
0.4
0.2
0.0
0 2 4 6 8 10 12
lp/cm
MTF
With the protocol used to generate the data reported here, expected deviations for values
on the MTF curve: - 10%.
Maximum deviations may reach - 20%.
Sensitivity Profile
With the protocol used to generate the data reported here, the slice sensitivity profiles
(FWHM) may vary ± 10% or 0.5mm whichever is larger.
Helical Scans: The slice sensitivity profile is triangular, and the FWHM may deviate by 20%,
or 1mm, whichever is greater.
With other methods, the maximum deviation may reach 1.5mm for all thicknesses; thin
slices are most affected by these measurement errors.
proves to be within tolerance for a period of 6 months. In this case the dose measurement
may be repeated annually; all other tests may be done quarterly.
In addition, the constancy tests should be repeated:
• Whenever malfunction is suspected;
• Immediately after the CT scanner undergone maintenance that could affect the
performance parameter under test;
• Whenever the constancy test leads to results outside the established criteria, to confirm
the test result.
Radiation Protection
A qualified radiological health physicist should review scan room shielding requirements.
Consider equipment placement, weekly projected workloads, and materials used for
construction of walls, floors, ceiling, doors and windows.
The following illustrations depict measurable radiation levels within the scan room while
scanning a 32cm CTDI body phantom (body) and a 20cm water phantom (head).
Technique:
0.66 1.31 2.63 5.26 5.26 2.63 1.31 0.66 140kV
100mA
1 Sec
4 x 5.00mm
50cm
50cm
Technique:
140kV
100mA
1 Sec
4 x 5.00mm
50cm
50cm
HEAD PHANTOM
ISO-Contour Levels : 0.33, 0.66, 1.31 and 2.63 μGy/Scan
50cm
50cm
HEAD PHANTOM
ISO-Contour Levels : 0.33, 0.66, 1.31 and 2.63 μGy/Scan
Technique:
140kV
100mA
1 Sec
4 x 5.00mm
50cm
50cm
ISO Center
Gantry
Table
Front
Rear
Horizontal
μGy/Hour
ISO Center
Gantry
Table
Front
Rear
Horizontal
μGy/Hour
X-Ray Tube
Focal Spot
541 mm
(ISO Center)
23.89 mm (Z-Axis)
625.8 mm (X-Axis)
water
difference
120+/-12
EXT axial
0 laser
W/L 250/-25 W/L 250/-100
Chapter 13
CT Acceptance Testing
NOTE: Prepare the standard as described below. Use this manual with such a standard.
This section describes the procedures for CT acceptance testing, based on IEC 61223-3-5,
and additional testing required by MHLW PAL EP6 (Japan) and JJG 1026-2007 (China).
• "(Reference IEC 61223-3-5 Clause 5.1)" on page 13-1
• "(Reference IEC 61223-3-5 Clause 5.2)" on page 13-2
• "(Reference IEC 61223-3-5 Clause 5.3)" on page 13-7
• "(Reference IEC 61223-3-5 Clause 5.4)" on page 13-14
• "(Reference IEC 61223-3-5 Clause 5.5)" on page 13-17
• "(Reference IEC 61223-3-5 Clause 5.6)" on page 13-18
• "(Reference IEC 61223-3-5 Optional)" on page 13-21
6. For the external laser light accuracy, move the cradle out and align the wire with the
external laser light field, and use coronal laser light to center the wire in the up/down
direction.
7. Landmark the wire using external landmark.
8. Scan the wire using the protocol as listed in Table 13-1.
Table 13-1 Scan protocols for axial internal/external light accuracy
For 16 Slice
Aperture/
Scan
Scan Slice Scan Range and Recon DFOV
kV mA speed SFOV
mode thickness Orientation kernel (cm)
(s)
(mm)
Internal 10mm/ I4.5 to S4.875
Axial 120 260 1 small Bone 10
Light 16i x 0.625 head first
External 10mm/ I4.5 to S4.875
Axial 120 260 1 small Bone 10
Light 16i x 0.625 head first
For 16 Slice
Scan Aperture/
Scan Scan Range and Recon DFOV
kV mA speed SFOV Slice thickness
mode Orientation kernel (cm)
(s) (mm)
S30 to I30 head
Preview Scout 120 50 NA NA NA NA NA
first
1.25mm0.625
Helic I3.0 to S3.0 head
Tomo 120 200 1 small pitchInterval Bone 10
al first
0.2mm
For 16 Slice
Scan
Scan Aperture/ Slice Scan Range and Recon DFOV
kV mA speed SFOV
mode thickness (mm) Orientation kernel (cm)
(s)
I4.5 to S4.875
Axial 120 260 1 small 10/16i x 0.625 Bone 10
head first
I3.435 to S3.440
Axial 120 160 1 small 1.25/2i x 0.625 Bone 10
head first
Data Evaluation
1. Select the image with the maximum wire CT number for evaluation. (See Figure 13-2).
For internal laser light, confirm that the wire is in the image with image location
between I2 and S2. Record the image location in Table 13-4.
2. For external laser light, follow the similar procedure as in step 1 and record the image
location in Table 13-4.
3. For preview, follow the similar procedure as in step 1 and confirm that the wire location
in the preview image is same as axial image location. Record the distance between
preview location and axial location in Table 13-4.
4. For sagittal laser light, select the image in the middle row (image #8, for example), adjust
the window width and window level, so that the wire is round and clear in the image (for
example, ww =1500, wl =0). Then, place the cursor at the center of the wire. Record the
Left/Right coordinate value in Table 13-4. See Figure 13-3 for details.
5. For coronal laser light accuracy, record the A/P coordinate value in Table 13-4.
6. The specifications for both internal and external light accuracy are +/- 2mm (I2 - S2).
Preview accuracy is +/- 2mm. Sagittal light accuracy is +/- 3mm. (L3 - R3) Coronal light
accuracy is +/- 3mm (A3 - P3).
Table 13-4 Patient Positioning Accuracy Results and Specifications
Figure 13-2 Internal and External Light Accuracy Image. Left image with maximum wire CT
number is used for evaluation and the image location with the brightest wire is I0.31.
Image Location
Brightest Wire
Figure 13-3 The sagittal and coronal light accuracy. Cursor coordinates are the values next
to “R or L” and “A or P”.
Cursor Coordinates
Cursor
Figure 13-4 View from the top of a GE Performance phantom with tungsten wire at 27
degree vs. the scan plane.
Tungsten wire at
27 degree
vs. scan plane
Scan plane
Z‐Axis
The thin tungsten wire has a diameter of 0.05mm, and linear attenuation much higher than
that of aluminum. With a slope of 1:2, the magnification at the scan plane is by 2x, therefore,
a slice thickness of 0.625mm (FWHM) in z will be projected to a length of 1.25mm (FWHM) in
the scan plane.
Catphan 600 is commercially available and has a module CTP404, which has two pairs of
wire ramps with angle of 23 degreevs. the scan plane (see Figure 13-5). For catphan 600, the
magnification factor in the imaging plane is 1/tan(23 degree)= 2.35, slightly greater than the
wire in GE performance phantom.
Testing can be done with either the GE Performance Phantom or the Catphan 600.
Figure 13-5 Catphan 600 module CTP404 has two pairs of wire ramps, one pair parallel vs.
x-axis, the other parallel vs. y-axis. Ramp angle at 23 degree, equal to a slope of 1:2.35.
Test Procedure
1. Center GE Performance phantom at the mark. If Catphan 600 is used, center it at
module CTP404.
2. Scan the phantom using the scan protocols as listed in Table 13-5. This table provides
the protocols for all the apertures available from the system. To evaluate specific
aperture, please select related protocol (s) from Table 13-5. Images with other slice
thickness can be retrospectively reconstructed.
3. Table 13-6 lists all the slice thickness combinations for each aperture. “N/A” means this
slice thickness is not available for the aperture.
4. Due to limited z-axis width of both phantom inserts, only center few rows are useful for
the slice thickness analysis when the phantom is centered in the middle of insert along
z-axis, especially when the aperture is wide (for example, 20mm). To study the slice
thickness for outer images, the phantom may be offset in z-axis by 10mm.
5. To study the slice thickness for the outer images at the tableside (A-side), move the
phantom and center it at S10. To study the slice thickness of the outer images at the
gantry side (B-side), move the phantom to I10.
Aperture
Scan
Scan Scan Range, /slice Recon DFOV
kV mA speed SFOV
mode orientation thickness kernel (cm)
(s)
(mm)
I5 to S5
Axial 120 260 1 Small 20 /10 bone 15
headfirst
I7.5 to S7.5
Axial 120 260 1 Small 20 /5 bone 15
head first
I8.5 to S9
Axial 120 260 1 Small 20 /2.5 bone 15
headfirst
I9.5 to
Axial* 120 260 1 Small 20 /1.25 bone 15
S9.25headfirst
I4 to S3.5
Axial 120 260 1 Small 15 /7.5 bone 15
headfirst
I5.5 to S5.75
Axial 120 260 1 Small 15 /3.75 bone 15
head first
S0 to S0
Axial 120 260 1 Small 10/10 bone 15
headfirst
I2.5 to S2.5
Axial 120 260 1 Small 10/5 bone 15
head first
I4 to S3.5
Axial 120 260 1 Small 10/2.5 bone 15
headfirst
I4.5 to S4.25
Axial 120 260 1 Small 10/1.25 bone 15
head first
I4.875 to S4.5
Axial * 120 260 1 Small 10/0.625 bone 15
head first
S0 to S0
Axial 120 260 1 Small 5/5 bone 15
headfirst
I1.5 to S1.0
Axial * 120 260 1 Small 5/2.5 bone 15
head first
I2.5 to S2.5
Axial * 120 260 1 Small 5/1.25 bone 15
head first
S0 to S0
Axial 120 260 1 Small 1.25/1.25 bone 15
headfirst
I0.325 to S0.3
Axial 120 260 1 Small 1.25/0.625 bone 15
head first
NOTE: *16x0.625mm and 16x1.25 only available for LightSpeed16;
4x1.25mm and 2x2.5mm are only available for retrospective reconstruction on
LightSpeed16.
Data Evaluation
The tomographic section thickness of an axial scan is evaluated by measuring the width of
the wire ramp along x-axis direction and then multiply the measured in-plane width by the
tangent of the ramp angle (vs. the scan plane.) For GE Performance phantom, the tangent of
the ramp angle is 0.5; for Catphan 600, it is 0.42.
1. Use proper size ROI (ROI should be placed within the wire to get accurate CT number) to
measure the CT number of both wire and background.
2. Adjust the window width to 1, and the window level to average of the CT number of the
wire and the background.
3. In window width and level adjusted, measure the width of both top and bottom wire.
4. Take average of the two widths, and multiply the average by 0.5 for GE Performance
phantom, and 0.42 for Catphan 600.
5. For a step-by-step example, see the next section.
6. The deviation of measured slice thickness is specified in Table 13-6.
CAUTION: The limiting measurement resolution of the cursor is 1mm, i.e., the distance
less than 1mm but greater than 0.5mm is rounded to 1mm, therefore, the
accuracy of this testing is limited by the cursor measurement capability. This
is especially important for thin slice measurement where the FWHM is close
to 0.625mm. The results for these thin slice images will be not as accurate as
the thick slice ones. This is the limitation by this testing method.
Table 13-6 Deviation for the slice thickness @ each aperture
Figure 13-7 To measure FWHM from the image, set WW =1, and WL =114. .
Figure 13-8 Measured the width of both top and bottom wires: 4mm, 4mm.
Dose
(Reference IEC 61223-3-5 Clause 5.4)
For details in dosimetry, refer to Chapter 12 of the Technical Reference Manual, “Quality
Assurance”-> “Dosimetry”.
The dose measurement methodology in the Technical Reference Manual follows those
described in IEC 60601-2-44.
In this section, the protocols are proposed for CTDIw, CTDIfree air per IEC 61223-3-5, Clause
5.4 under head and body scan conditions. The specifications for CTDIw and CTDIfree air under
these proposed CT operating conditions are listed.
Apertur
Scan Scan
Scan CTDI Scan e/Slicet Recon DFOV
kV mA speed RangeOrienta SFOV
conditions phantom mode hicknes kernel (cm)
(s) tion
s (mm)
16cm CTDI I5 to S5 head 20mm/
Head Axial 120 260 1 Head stnd 25
phantom first 2ix10
32cm CTDI I5 to S5 head 20mm/
Body Axial 120 260 1 Large stnd 50
phantom first 2ix10
Table 13-8 Expected Head and Body CTDIw and Maximum Variation using scan protocols
defined in Table 13-7
For 16 Slice
CTDIw expressed in mGy: CTDIw expressed in mGy/100 mAs:
Head 47.8mGy+/-40% Head 18.4mGy+/-40%/100mAs
Body 24.2mGy+/-40% Body 9.3mGy+/-40%/100mAs
Scan
Scanning Scan Recon DFOV
kV mA speed Scan Range SFOV Aperture(mm)
Conditions mode kernel (cm)
(s)
I5 to S5
Head Axial 120 260 1 Head 20mm/2ix10 stnd 25
head first
I5 to S5
Body Axial 80 260 1 Large 20mm/2ix10 stnd 50
head first
I5 to S5
Body Axial 100 260 1 Large 20mm/2ix10 stnd 50
head first
I5 to S5
Body Axial 120 260 1 Large 20mm/2ix10 stnd 50
head first
I5 to S5
Body Axial 140 260 1 Large 20mm/2ix10 stnd 50
head first
I 2.5 to S2.5
Body Axial 120 260 1 Large 10mm/2ix5 stnd 50
head first
I4.40 to
Body Axial 120 260 1 S4.35 Large 10mm/8ix1.25 stnd 50
headfirst
I4.7 to
10mm/16ix0.6
Body Axial 120 260 1 S4.675 Large stnd 50
25
head first
I0.325 to
1.25mm/2ix0.6
Body Axial 120 260 1 S0.3 Large stnd 50
25
headfirst
S0 to S0 1.25mm
Body Axial 120 260 1 Large stnd 50
head first /1ix1.25
I5.6 to
Body Axial 120 260 1 S5.65 Large 15mm /4ix3.75 stnd 50
headfirst
Table 13-11 Expected CTDIfree air for scan conditions under Table 13-9, for 4/8 slices
Body
Axial 80 260 Large 20mm/2ix10 22.4 mGy+/- 40%
Body
Axial 100 260 Large 20mm/2ix10 41.9 mGy+/- 40%
Body
Axial 120 260 Large 20mm/2ix10 63.9 mGy+/- 40%
Body
Axial 140 260 Large 20mm/2ix10 88.1 mGy+/- 40%
Body
Axial 120 260 Large 10mm/2ix5 79.9 mGy+/- 40%
Body
Axial 120 260 Large 10mm/8ix1.25 79.9 mGy+/- 40%
Body 10mm/16ix0.6
Axial 120 260 Large 79.9 mGy+/- 40%
25
Body 1.25mm/2ix0.6
Axial 120 260 Large 155.8 mGy+/- 40%
25
Body 1.25mm
Axial 120 260 Large 155.8 mGy+/- 40%
/1ix1.25
Body
Axial 120 260 Large 15mm /4ix3.75 68.2 mGy+/- 40%
Test Equipment
For head scanning, an outsider diameter of 20cm cylindrical water phantom, such as GE
Quality Assurance (QA) phantom should be used for noise, CT number and uniformity
measurement.
For body scanning, an outsider diameter of 30cm cylindrical water phantom should be used.
Test Procedure
The detail test procedures and data evaluation for noise, mean CT number and uniformity
measurements are well described in IEC 61223-3-5 Sections 5.5.3 and 5.5.4.
Aperture/
Scanning Scan
Scan slice Recon DFOV
condition kV mA speed Scan range SFOV
mode thick-ness kernel (cm)
s (s)
(mm)
I5 to S5
Head Axial 120 260 1 Head 20/2ix10 stnd 25
head first
I5 to S5
Body Axial 120 260 1 Large 20/2ix10 stnd 35
head first
Table 13-14 Expected results for head and body scanning conditions in Table 13-13
Spatial Resolution
(Reference IEC 61223-3-5 Clause 5.6)
Test Equipment
GE Performance phantom has a 0.05mm diameter tungsten wire perpendicular to the
imaging plane. This phantom can be used to evaluate the system Modulation Transfer
Function (MTF) combined with an automated software tool implemented in the GE system.
Scan Slice
Scan Scan range Recon DFOV
kV mA speed SFOV thickness
mode /Orientation kernel (cm)
(s) (mm)
I5 to S5 Head
Axial 120 200 1 Small 2ix10 stnd 25
First
I5 to S5 Head
Axial 120 200 1 Small 2ix10 edge 10
First
I5 to S5 Head
Axial 120 200 1 Large 2ix10 stnd 25
First
I5 to S5 Head
Axial 120 200 1 Large 2ix10 edge 10
First
Figure 13-9 ImageAnalysis2 for MTF analysis using wire images from GE Performance
phantom (top) and the MTF results for the images selected from ImageWorks browser.
Test Equipment
Catphan 600 is a commercially available CT phantom. Its solid image uniformity module
CTP486 can be used to evaluate the low contrast resolution (LCD) in a statistical manner (see
Figure 13-10). Module CTP515 can be used to evaluate visual low contrast resolution.
However, the visual low contrast resolution is highly subjective and requires observer’s study
based on a large population to get accurate results, therefore, in this section, a statistical
method is used.
The alternative to Catphan 600 for the statistical low contrast resolution measurement is the
GE Quality Assurance (QA) phantom. The uniform water section of the phantom can be used
to measure the statistical Low Contrast Detectability (or LCD).
Figure 13-10 Module CTP486 is used for Low Contrast Detectability (statistical) evaluation.
4. GE scanner has an automated tool to calculate statistical LCD values. . From “Service”
desktop -> click ”Image Quality” tab ->click ”Image Analysis” button ->in
”ImageAnalysis2” ->”Manual” button-> ”LCD”. (See Figure 13-11). Select the center two
image(s) from the ImageWorks browser. Click “Accept” button. In the LCD popup window
“Input hole diameter (mm)”. Click the “OK” button to use the default value of 3.00 for the
object size (=3mm).
– A result panel will show the calculated LCD values for each individual image and the
average from all images as shown in Figure 13-11. The individual results are listed
under column “ % Contrast @95% CL”. The average LCD result is listed under the
column “ Avg % Cnst@ 95% CL” as shown in Figure 13-11.
Table 13-17 Scan Protocols for Catphan 600
Aperture
Scan Images
Scan Scan / slice Recon DFOV
kV mA speed Pitch SFOV for LCD
mode range thicknes kernel (cm)
(s) analysis
s (mm)
I5 to S5,
20 /2i x Center 2
Axial 120 280 1 head N/A Small stnd 22.7
10 images
first
I5 to S5,
0.625 Center 2
Helical 120 230 1 head Small 20/10 stnd 22.7
:1 images
first
Figure 13-11 The LCD tool (upper half) panel and the result panel, which displays the LCD
results for each individual image and the average of two images.
Chapter 14
X-Ray Tube Specifications
NOTE: Throughout this chapter, Model Numbers may contain “-X” extensions, which
represent all available numeric characters. For example, 2137130-X refers to
2137130, 2137130-2, 2137130-3, etc.
NOTE: * 2120785-5 applies to Tube D3186T and D3189T.
Environmental Specifications
Operating Environment
Maintain an ambient temperature of less than 35C and 30 to 60% (non-condensing)
relative humidity (50% nominal) during operation
80kV 2.9 mm Al
100kV 3.6 mm Al
120kV 4.3 mm Al
140kV 5.0 mm Al
CT Scan Ratings
These ratings apply to a system with computer controlled technique selection, scan mode,
and scan duration.
The system uses a math model to track tube temperature.
This tube cooling algorithm delays the start of a scan, if necessary, to avoid exceeding
temperatures that may damage the tube anode or unit.
Table 14-2 Performix Ultra Target Load in Kilowatts for Selected Scan Technique
NOTE: Small focal spot used for exposures less than or equal to 24kW. Large focal spot used
for exposures over 24kW.
Figure 14-1 Reference Axis for Target Angle and Focal Track
Reference Axis
Focal Track
7 degrees from Ref. Axis
Weight
Without Mounting Plate: 84 kg
With Mounting Plate: 90 kg
X-Ray Tube Assembly Heating and Cooling Curves (Reference IEC 60613
and 21CFR 1020.30 (h) (2) (ii))
The following chart reflects maximum tube performance. The system software ultimately
limits tube operation.
Figure 14-3 Tube Assembly Heating and Cooling Curves
Performix Ultra Assembly Heating and Cooling Curves
Cooldown
Time (seconds)
NOTE: All Assembly Heating and Cooling curves were generated with the X-Ray tube oil
pump and fans operating normally.
Target Material
Tungsten - Rhenium focal track on a molybdenum alloy substrate backed by graphite
Rotor Speed
8400 RPM
Data Required for Driving the Rotating Anode (Reference IEC 60601-2-28)
Acceleration: 12 seconds, 0 to 140Hz
Maximum Current During Boost: 24A, 3-phase motor, 280Hz (to produce 140Hz drive)
Steady State: 7A at 480V, 3-phase motor, 280Hz
Brake Time: 30 seconds, 12A, 3-phase motor, 140Hz to 20Hz
Maximum Anode Heat Capacity (Reference IEC 60613 and 21CFR 1020.30
(h) (2) (ii))
6.3 MHU / 4700 kJ
Figure 14-4 Performix Ultra Anode Heating and Cooling Curves
Performix Ultra Anode Heating and Cooling Curves
24 kW
12 kW
8 kW
4 kW
Kilojoules
2 kW
Cooldown
Time (seconds)
NOTE: The Figure 14-4 chart reflects maximum tube performance. The system software
ultimately limits tube operation.
Lg. Spot
Sm. Spot
450
350
250
mA
150
50
80 kV
100 kV
0 120 kV
0 10 140 kV
20 30 40 50 60 70 80 90 100
Time – Sec. 110 120
400
300
mA
200
100
80
0 100
0
10
120
20
30
kV
40
50
60
140
70
80
90
100
Time – Sec.
110
120
NOTE: Large Focal Spot used for exposures over 24kW. Limits based on initial storage of
383kJ.
300
250
200
mA
150
100
50
80 kV
0 100 kV
0 10 120 kV
20 30 40 50 60 70 140 kV
Time – Sec. 80 90 100
110 120
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
0 .1 .2 .3 .4 .5 .6 .7 .8
SPATIAL FREQUENCY – LINE PAIRS/mm
0.2
0.1
0
0 .2 .4 .6 .8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0
SPATIAL FREQUENCY – LINE PAIRS/mm
1.0
MODULATION TRANSFER FUNCTION
0.9
0.8 MODULATION TRANSFER FUNCTION
WIDTH DIRECTION
0.7 0.9 mm W x 0.7 mm L FOCAL SPOT
0.6 STANDARD MAGNIFICATION 1.3
0.5
0.4
0.3
0.2
0.1
0
0 .2 .4 .6 .8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0
SPATIAL FREQUENCY – LINE PAIRS/mm
MODULATION TRANSFER FUNCTION
1.0
0.9 MODULATION TRANSFER FUNCTION
LENGTH DIRECTION
0.8
0.9 mm W x 0.7 mm L FOCAL SPOT
0.7 STANDARD MAGNIFICATION 1.3
0.6
0.5
0.4
0.3
0.2
0.1
0
0 .2 .4 .6 .8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0
WARNING: Do not discard the X-Ray Tube Assembly among industrial waste or domestic
garbage.
WARNING: A damaged X-Ray Tube Assembly should not be dispatched through the
national postal service.
Your local GE Healthcare field service representative will advise you on the suitable means
of disposal.
Send any X-Ray Tube to be discarded to the GE Healthcare Service network, where it will be
sent to a recycling center for proper disposal.
Hazardous Materials
The X-Ray Tube Assembly contains the following potentially hazardous materials:
Lead: Lead salts are toxic and their ingestion may cause serious problems. The working of
lead is subject to regulations.
Oil: Univolt 54 and Crosstrans 206 mineral oil are not toxic, but the prevailing environmental
regulations should be observed for their disposal or recovery. For example, it is forbidden to
dispose of these oils in the wastewater or sewerage system or in the natural environment.
Precautions
Take all the necessary precautions for the personnel handling the recovery or destruction of
X-Ray Tube Assemblies, and in particular against the risks due to lead. These personnel must
be informed of the danger involved and of the necessity to observe the safety measures.
Environmental Specifications
Non-Operating Environment
(Reference: IEC 60601-2-28)
Maintain a temperature range between -40°C and 70°C (from 10% to 100% relative
humidity) during storage and shipment of the tube assembly. Use GE Healthcare transport
packaging during shipment. Commercial airline shipment is permissible.
Operating Environment
Maintain a CT room ambient temperature range of 18°C - 26°C and 30% to 60%
(non-condensing) relative humidity (50% nominal) during operation. This CT room ambient
temperature range ensures that the temperature inside the gantry covers does not exceed
the tube maximum ambient operating temperature of 45°C. Operating altitude range
(relative to sea level): -150m to 2400m.
System errors may result when attempting to operate the anode rotation circuit if the tube
assembly has been brought into the operating environment shortly after being stored or
transported at temperatures below 18°C. Error-free operation can be ensured by allowing
the oil circulation pump and heat exchanger to be operated for 40 min. prior to energizing
the anode rotation circuit if the storage or transportation ambient was below 18°C (this 40
min. period covers tube assemblies maintained down to the non-operating environment
lower limit of -40°C).
X-ray Characteristics
Total Filtration
(Reference: IEC 60601-1-3:1994, IEC 60601-2-28:1993 and 21 CFR 1020.30(h)(2))
Permanent Filtration
(Reference: IEC 60601-1-3:2008, IEC 60601-2-28:2010 and 21 CFR 1020.30(h)(2))
The Quality Equivalent Filtration of the PerformixTM Plus X-ray Tube Assembly is 4.9 mm Al
minimum at a tube potential in the range of 70-75 kV. This filtration is non-removable from
the assembly in normal use.
• Code of Federal Regulations, Title 21, Part 820 – Quality System Regulation
• Code of Federal Regulations, Title 21, Sub chapter J – Radiological Health
• IEC 60601-1:1988/2005 ; AAMI ES60601-1; CSA C22.2#60601-1
• IEC 60601-1-3:1994/2008
• IEC 60601-2-28:1993/2010
GE Medical Systems is ISO 9001 (2008) and ISO 13485 (2003) certified
Classification
(Reference: IEC 60601-2-28:1993/2010)
The PerformixTM Plus X-Ray Tube Assembly has the following classification:
• Type of protection against electric shock: CLASS 1
Figure 14-10 Reference Axis for target angle and focal track
Focal spot on
target focal track
Cathode structure
Focal track 7°
Reference axis
HV Connection
(Reference: IEC 60601-2-28)
The PerformixTM Plus X-ray Tube is a bipolar x-ray tube assembly. The anode (+) and cathode
(-) high voltage cables are connected to the X-Ray tube assembly with 3-conductor
connectors complying with US Federal Standard design. This x-ray tube assembly is
designed for use with the BrightSpeed (Excel/Edge/Elite)/Lightspeed (16/Plus/Ultra) x-ray
generator (part number 2371333-2 upgraded with 5456518 Rotation Board).
The accessible metal parts of the X-Ray Tube Assembly body and flexible conductive
housing of high-voltage cables must be connected to the conductive enclosure of the
high-voltage generator.
Figure 14-11
The focal spot position is visually indicated on the X-ray Tube housing by the use of three
labels (solid circle with the words “Focal Spot”).
Construction
(Reference: IEC 60601-2-28)
The X-Ray Tube Housing is made of a lead-lined lightweight alloy. It is filled under vacuum
with specially processed insulating oil. A sealed cooling system for the insulating oil is
integral to the x-ray tube assembly. An expansion volume compensates for oil dilation at
permissible temperatures. The mass of the X-Ray Tube Assembly is approximately 91kg.
The mounting mechanism is integral to the tube assembly.
Cooldown
Time (seconds)
Thermal characteristics
(Reference: IEC 60613)
The PerformixTM Plus X-Ray Tube Assembly must always be equipped with a beam-limiting
device in order to meet requirements for the maximum X-Ray beam extent required for its
specified applications.
The beam-limiting devices compatible with PerformixTM Plus tube assembly are:
• Collimator Assembly 5140000-4
Any PerformixTM Plus X-Ray Tube Assembly having beam limiting devices other than those
listed above is obliged to be checked for compliance examination for beam quality and
leakage radiation according to the requirements of IEC Standard 60601-1-3 and 21 CFR
1020.30(k)(m).
Envelope Voltage
(Reference: IEC 60613:2010)
0 kV with respect to earth (ground)
Envelope Current
(Reference: IEC 60613:2010)
3.5% at an x-ray tube voltage of 120kV
Target Material
(Reference: IEC 60601-2-28)
The target material is a Tungsten-Rhenium focal track on a molybdenum alloy substrate
backed by graphite.
Focal Spots
(Reference: IEC 60336)
Reference Axis:
The reference axis for focal spot characteristics is defined to be normal to the mounting
surface of the X-ray Tube housing and passing through the center of the focal spot.
Anode Rotation
(Reference: IEC 60601-2-28)
8400 RPM
Kilojoules
Cooldown
Time (seconds)
Table 14-5 Serial load ratings (10 min inter-exposure delay) for an initial anode heat content
of 383kJ
140kV
120kV
80kV
Tube Current (mA)
100kV
100kV 80kV
Tube Current (mA)
120kV
140kV
Small Filament
Large Filament
WARNING: Do not discard the X-Ray Tube Assembly among industrial waste or domestic
garbage.
WARNING: A damaged X-Ray Tube Assembly should not be dispatched through the
national postal service.
Your local GEHC field service will advise you on the suitable means of disposal. The X-Ray
Tube Assembly to be discarded should be forwarded to the GEHC Service network, and it will
be disposed of in a GEHC recycling center.
Hazardous Materials
The X-Ray Tube Assembly contains the following potentially dangerous materials:
Lead: Lead salts are toxic and their ingestion may cause serious problems. The working of
lead is subject to regulations.
Oil: Univolt 54 and Crosstrans 206 mineral oil are not toxic, but the prevailing environmental
regulations should be observed for their disposal or recuperation. For example, it is
forbidden to dispose of these oils in the wastewater or sewage system or in the natural
environment.
Precautions
Take all the necessary precautions for the personnel handling the recovery or destruction of
X-Ray Tube Assemblies, and in particular against the risks due to lead. These personnel must
be informed of the danger involved and of the necessity to observe the safety measures.
manufacturer
date of manufacture
permanent filtration
serial number
focal spots
(Large) (Small)
Routine Maintenance
Cleaning of the two inlet air filters of the heat exchanger assembly should be performed
annually; refer to the compatible CT system Service Manual for details about cleaning of
these filters.
Manufacturer Address
GE BE PRIVATE LIMITED
60, Export Promotion Industrial Park
Whitefield, Bangalore 560066
India
or
Chapter 15
Regulatory Information
(After 1-May-2012, the above statement is applicable only for the scanner with NIO16
Console model number: 5411378-X).
Authorized representative for Europe/European registered place of business:
GE Medical Systems SCS
Quality Assurance Manager
283 rue de la Minière
78530 BUC France
Tel +33 130704040
• Code of Federal Regulations, Title 21, Part 820 -Quality System Regulation
• Code of Federal Regulations, Title 21, Sub chapter J -Radiological Health
Federal U.S. law restricts this device for sale by or on the order of a physician.
GE Healthcare is ISO 9001 and ISO 13485 certified.
• Applicable standards of Underwriters' Laboratories, Inc. (UL), an independent
testing laboratory
• Applicable standards of the Canadian Standards Association (CSA)
• Applicable standards of the International Electrotechnical Commission (IEC):
The scanner complies with IEC 60601-1: 1988 +A1:1991 +A2:1995 (all configurations),
IEC 60601-1: 2005 (if the scanner manufactured after 30-January-2012 and with NIO16
Console model number: 5411378-X), UL 60601-1: 2003 (all configurations) and
ANSI/AAMI ES60601-1: 2005 +A1: 2009 +A2:2010 (if the scanner manufactured after
30-January-2012 and with NIO16 Console model number: 5411378-X).
The system is classified as a Class I, IPX0 equipment, not suitable for use in the
presence of a flammable anaesthetic mixture with oxygen or nitrous oxide. It is rated
for continuous operation with intermittent loading. No sterilization is applied. The
patient table cradle and cradle accessories are considered Type B applied parts.
The system is intended to be used for head and whole body computed tomography.
The scanner complies with IEC 60601-1-1: 2000.
All portions specified in the section of Patient Environment in pre-installation manual
of the system are suitable for use in the patient environment.
The system should be used only with GE approved equipment.
The scanner complies with IEC 60601-1-2: 2001+A1: 2004 (ED2.1) (all configurations), IEC
60601-1-2: 2007(if the scanner manufactured after 30-January-2012 and with NIO16
Console model number: 5411378-X).
Detailed information concerning Electromagnetic Compatibility can be found in the
following section.
The system complies with the applicable portions of IEC 60601-2-28.
The scanner with radiation protection in accordance with IEC 60601-1-3: 1994 (all
configurations), IEC 60601-1-3: 2008 (if the scanner manufactured after
30-January-2012 and with NIO16 Console model number: 5411378-X).
The scanner complies with the applicable portions of IEC 60601-2-32.
Associated
Patient Table IEC 60601-2-32: 1994
Equipment
LightSpeed Plus
LightSpeed Ultra IEC 60601-2-44: 2001
CT SCANNER
+A1:2002 (all configurations)
LightSpeed16
IEC 60601-2-44: 2009 (if the
LightSpeed Plus scanner manufactured after
CT SCANNER LightSpeed Ultra 30-January-2012 and with
LightSpeed16 NIO16 Console model
number: 5411378-X)
Intended Use
The system is intended to be used for head and whole body computed tomography.
Chapter 16
Electromagnetic Compatibility
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
3.5 3.5 7
d = ------- P d = ------- P d = --- P
Rated Maximum 3 3 3
Output Power (P) of
Separation Distance Separation Distance Separation Distance
Transmitter Watts
meters meters meters
(W)
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
5.2. This product complies the radiated emission as per CISPR11 Group1 Class A standard
limits.
The LightSpeed Series System is predominantly intended for use, in non-domestic
environments, and not directly connected to the Public Mains Network. The LightSpeed
Series System is predominantly intended for use (e.g. in hospitals) with a dedicated
supply system, and with a X-ray shielded room. In case of using in a domestic
environment (e.g. doctors' offices), in order to avoid interferences, it is recommended to
use a separated AC power distribution panel & line, with a X-ray shielded room.
5.3. Subsystem & accessories Power supply distribution
All components, accessories subsystems, systems which are electrically connected to
the LightSpeed Series System, must have all AC power supplied by the same power
distribution panel & line.
5.4. Stacked components & equipment
The LightSpeed Series System should not be used adjacent to or stacked with other
equipment; if adjacent or stacked use is necessary, the LightSpeed Series System should
be observed in order to verify normal operation in the configuration in which it will be
used.
5.5. Low frequency magnetic field
In case of a digital LightSpeed Series System, the Gantry (digital detector) shall be apart
1 meter from the generator cabinet, and 1 meter apart from the analog (CRT) monitors.
These distance specifications will minimize the low frequency magnetic field
interference risk.
5.6. Static magnetic field limits
In order to avoid interference on the LightSpeed Series System, static field limits from
the surrounding environment are specified.
Static field is specified less than <1 Gauss in Examination room, and in the Control Area.
Static field is specified less than <3 Gauss in the Technical Room.
5.7. Electrostatic discharge environment & recommendations
In order to reduce electrostatic discharge interference, install a charge dissipative floor
material to avoid electrostatic charge buildup.
The relative humidity shall be at least 30 percent.
The dissipative material shall be connected to the system ground reference, it
applicable.
Chapter 17
System Specifications
The following describes system components. For details, refer to subsystem specifications.
Table 17-1 Model NumbersTransporter
FDA
Component Model Number Rating Plate Locations
(Y/N)
(Scanner ID)
2378661-X
Gantry Lower, left Gantry base in rear
2383549-X Y
2383550-X
5411378-X (NIO16)
Operator Console Rear of cabinet Y
2377708-X (GOC4)
Right side, low on front leg
Table 2335179-X Y
Bottom on left leg
Generator 2371333-X Front Y
X-ray Tube Unit 2137130-X On Housing center Y
2214768-x Tube at 12 o’clock: on
Collimator Y
5140001-4 collimator front
Back horizontal surface of top
Power Distribution Unit 2326492-3 N
cover
NOTE: "X" is a variable numerical suffix that represents minor variations in design that do
not affect conformance with the requirements of the standard for Medical Electrical
Equipment; IEC60601-1:1988+A1: 1991+A2: 1995 or IEC60601-1: 2005 and identical
and some modified other standards as IEC.
1. Manufacturer
2. Momentary Plus
3. Continuous
1. Manufacturer
2. Momentary Ultra
3. Continuous
(Plates on rear of Gantry)
1. Manufacturer
2. Momentary
3. Continuous
1.Manufacturer
2.Model No.
3.Serial No.
4.Manufactured
COMPLIES WITH RADIATION
PERFORMANCE STANDARDS,21 CFR SUBCHAPTER J.
1.Manufacturer
2.Model
4. Manufactured
1.Manufacturer
GE HANGWEI MEDICAL SYSTEMS CO. LTD
2.Model Number No. 2 YONG CHANG NORTH ROAD
ECONOMIC-TECHNOLOGY DEVELOPMENT ZONE
3.Serial Number BEIJING P.R. CHINA 100176
MADE FOR GE MEDICAL SYSTEMS CO.
4.Line Frequency POWER DISTRIBUTION UNIT
MODEL No. 2326492-3
5.Input Power SERIAL No. ________________
INPUT VOLTAGE: 3~200//240, 380//480V
6.Momentary LINE FREQUENCY: 50/60Hz
INPUT POWER:
7.Continuous MOMENTARY 90KVA@ 0.85PF
CONTINOUS 20KVA
8.Weight WEIGHT: 770LBS. (350KGS)
MFG DATE: _____________
9.Manufacturing
Date PDU Rating Plate
5.Manufactured MANUFACTURED:
SOURCE: 120VAC 50/60Hz 15A
MADE FOR
GE HANGWEI MEDICAL SYSTEM CO., LTD.
NO.2 NORTH YONG CHANG STREET
BEIJING ECONOMIC & TECHNOLOGICAL DEVELOPMENT ZONE
BEIJING, 100176, P. R. CHINA
BY
GE MEDICAL SYSTEMS, LLC
3000 N GRANDVIEW BLVD., WAUKESHA, WI., U.S.A
SA
Supply
Supply Supply
Component Model Name or Number Frequen
Power Voltage
cy
200//240
LightSpeed16 90 kVA VAC
LightSpeed Ultra *75kVA **380//480 **50/60
*System LightSpeed Plus VAC Hz
3 Phase
670 VDC
2378661-X 100 A 440/254
2383549-X 16 A VAC
Gantry
2383550-X 16/30/40 A 208/120 50/60 Hz
VAC
5411378-X (NIO16)
Operator Console 20 A 120 VAC 50/60 Hz
2377708-X (GOC4)
Table 2335179-X 10 A 120 VAC 50/60 Hz
200//240
VAC
Power Distribution **50/60
2326492-3 90 kVA **380//480
Unit Hz
VAC
3 Phase
39.6
Gantry 87.6 74.1 223 188.1 100.1 2800 1270
26.75-34.
75
Keyboard
table 102-12
Operator GOC4 49 40-48 123.8 68-88.3 396 180
26.75-31. 2.8
Console 75
Monitor
table
NIO16 19 26 29 47 65.6 74 159 72
Table & Cradle 24 44 199 61 112 505 750 340
Power Distribution
28 42 22 711 1067 559 770 349
Unit
Total Weight with Operator Console NIO16 5141 2332
Total Weight with Operator Console GOC4 6179 2802
Line pair values decrease with larger focal spot (by 5% with Standard and by 7% with Edge);
limiting resolution is unaffected.
Measurement Basis: Limiting resolution is determined by reconstruction filter cutoff. The
50% and 10% MTF are demonstrated on GE Performance Phantom. MTF is calculated from
a two-dimensional Fourier transform of the point spread function using pixel data around a
0.05mm tungsten wire.
Standard Algorithm
On 8 inch (20 cm) CATPHAN phantom: 5mm @ 0.30% at 13.3 mGy (1.33 Rad).
Suggested Scan Technique:
120 kVp, 60 mAs, 0.5 to 1.0 second gantry rotation, 0.625:1 pitch mode with 10mm nominal
image thickness, 12.5mm table travel/rotation, 25 cm SFOV, 25 cm DFOV, 512 recon, and
standard algorithm.
On 8 inch (20 cm) CATPHAN phantom: 3mm @ 0.30% at 37.2 mGy (3.72 Rad).
Suggested Scan Technique:
120 kVp, 180 mAs, 0.5 to 1.0 second gantry rotation, 0.625:1 pitch mode with 10mm nominal
image thickness, 12.5mm table travel/rotation, 25 cm SFOV, 25 cm DFOV, 512 recon, and
standard algorithm.
ASiR Reconstruction
On 8 inch (20 cm) CATPHAN phantom: 5mm @ 0.30% at 10.0 mGy (1.00 Rad).
Suggested Scan Technique:
120 kVp, 45 mAs, 0.5 to 1.0 second gantry rotation, 0.625:1 pitch mode with 10mm nominal
image thickness, 12.5mm table travel/rotation, 25 cm SFOV, 25 cm DFOV, 512 recon, and
standard algorithm with ASiR Reconstruction (60%).
On 8 inch (20 cm) CATPHAN phantom: 3mm @ 0.30% at 27.9 mGy (2.79 Rad).
Suggested Scan Technique:
120 kVp, 135 mAs, 0.5 to 1.0 second gantry rotation, 0.625:1 pitch mode with 10mm nominal
image thickness, 12.5mm table travel/rotation, 25 cm SFOV, 25 cm DFOV, 512 recon, and
standard algorithm with ASiR Reconstruction (60%).
Test method is as follows:
1. Measure mean CT # values of an array of pixel groups who’s area (pixel group) equals
the size of the detectable object size.
2. Calculate the standard deviation for the means of the pixel groups.
3. Statistically calculate the % contrasted change needed to ensure with 95% confidence
that an object with this contrast could be detected with the above background noise,
and 95% confidence that it’s not detected when not present.
Measurement Basis: Dose is measured on top surface of the phantom using a pencil probe
with a 10 cm chamber length, with the phantom and probe held stationary, and with scan
time equal to time needed to acquire 30 images in helical mode. Dose is average dose per
image for 30 contiguous images (+/- 15% expected deviation).
Standard Algorithm
On 8 inch (20 cm) CATPHAN phantom: 5mm @ 0.30% 13.3 mGy (1.33 Rad).
Suggested Scan Technique:
120 kVp, 100 mAs, 0.5 to 1.0 second gantry rotation, axial acquisition mode with 10 mm
nominal image thickness, 25 cm SFOV, 25 cm DFOV, 512 recon, and standard algorithm.
On 8 inch (20 cm) CATPHAN phantom: 3mm @ 0.30% at 37.2 mGy (3.72 Rad).
Suggested Scan Technique:
120 kVp, 280 mAs, 0.5 to 1.0 second gantry rotation, axial acquisition mode with 10 mm
nominal image thickness, 25 cm SFOV, 25 cm DFOV, 512 recon, and standard algorithm.
ASiR Reconstruction
On 8 inch (20 cm) CATPHAN phantom: 5mm @ 0.30% 10.0 mGy (1.00 Rad).
Suggested Scan Technique:
120 kVp, 75 mAs, 0.5 to 1.0 second gantry rotation, axial acquisition mode with 10 mm
nominal image thickness, 25 cm SFOV, 25 cm DFOV, 512 recon, and standard algorithm with
ASiR Reconstruction (60%).
On 8 inch (20 cm) CATPHAN phantom: 3mm @ 0.30% at 27.9 mGy (2.79 Rad).
Suggested Scan Technique:
120 kVp, 210 mAs, 0.5 to 1.0 second gantry rotation, axial acquisition mode with 10 mm
nominal image thickness, 25 cm SFOV, 25 cm DFOV, 512 recon, and standard algorithm with
ASiR Reconstruction (60%).
Test method is as follows:
1. Measure mean CT# values of an array of pixel groups who’s area (pixel group) equals the
size of the detectable object size.
2. Calculate the standard deviation for the means of the pixel groups.
3. Statistically calculate the % contrasted change needed to ensure with 95% confidence
that an object with this contrast could be detected with the above background noise,
and 95% confidence that it’s not detected when not present.
Measurement Basis: Dose is measured on top surface of the phantom using a pencil probe
with a 10 cm chamber length, with the phantom and probe held stationary. (+/- 15%
expected deviation).
Helical Image Noise (Reference 21CFR 1020.33 (c) (3) (i) and
YY0310)
Standard Algorithm
Noise: ≦ 0.35% (nominal 0.32%) at 28.5 mGy
Suggested Scan Technique:
120 kVp, 160 mAs, 0.5 to 2.0 second gantry rotation, 0.625:1 pitch mode with 10mm nominal
image thickness, 12.5mm table travel/rotation, 25cm SFOV, 25 cm DFOV, 512 recon, and
standard algorithm.
ASiR Reconstruction
Noise: ≦ 0.35% (nominal 0.32%) at 17.1 mGy (ASiR 40%)
Suggested Scan Technique:
120 kVp, 95 mAs, 0.5 to 2.0 second gantry rotation, 0.625:1 pitch mode with 10mm nominal
image thickness, 12.5mm table travel/rotation, 25cm SFOV, 25 cm DFOV, 512 recon, and
standard algorithm with ASiR reconstruction.
Measurement Basis: Noise is demonstrated on 8.5 in GE Quality Assurance phantom
provided with the system using a 25mm x 25mm box ROI.
Head
Standard Algorithm
Noise: ≦ 0.35% (nominal 0.32%) at 29.3 mGy
Suggested Scan Technique:
120kVp, 260 mAs, 0.6 to 4.0 second gantry rotation, 10mm nominal image thickness, 2i
mode reconstruction, 25cm SFOV, 25 cm DFOV, 512 recon, and standard algorithm.
ASiR Reconstruction
Noise: ≦ 0.35% (nominal 0.32%) at 17.6 mGy (ASiR 40%)
Suggested Scan Technique:
120kVp, 155 mAs, 0.6 to 4.0 second gantry rotation, 10mm nominal image thickness, 2i
mode reconstruction, 25cm SFOV, 25 cm DFOV, 512 recon, and standard algorithm with ASiR
recunstruction.
Measurement Basis: Noise is demonstrated on 8.5 in GE Quality Assurance phantom
provided with the system using a 25mm x 25mm box ROI.
CT Number Linearity
Mean CT number: 0+/-3HU
CT Number Uniformity
CT number of Uniformity: 0+/-3HU
Body
Standard Algorithm
Noise: ≦ 1.035% (nominal 0.90%)
Suggested Scan Technique:
120kVp, 260 mAs, 1 second gantry rotation, 10mm nominal image thickness, 2i mode
reconstruction, 50 cm SFOV, 35 cm DFOV, 512 recon, and standard algorithm.
ASiR Reconstruction
Noise: ≦ 1.035% (nominal 0.90%) (ASiR 40%)
Suggested Scan Technique:
120kVp, 155 mAs, 1 second gantry rotation, 10mm nominal image thickness, 2i mode
reconstruction, 50 cm SFOV, 35 cm DFOV, 512 recon, and standard algorithm with ASiR
recunstruction.
Measurement Basis: Noise is demonstrated on 30cm water phantom provided with the
system using a 25mm x 25mm box ROI.
CT Number Linearity
Mean CT number: 0+/-6HU
CT Number Uniformity
CT number of Uniformity: 0+/-8HU
NOTE: The body specification stated below is to comply with China Regulation and meet the
requirement of YY0310-2005 standard.
Noise: ≦ 0.35% (nominal 0.32%) at 29.3 mGy
Uniformity: 0+/-4HU
Mean CT number: 0+/-4HU
Suggested Scan Technique: 120kVp, 260 mA, 1sec, 2i x 10mm, 25cm SFOV, 25 cm
DFOV, 512 recon, and standard algorithm. Dose is dose per image.
Measurement Basis: Noise is demonstrated on 20cm water phantom using a 25mm x
25mm box ROI.
Dose Performance
Table 17-5 Helical Dose (For 16 slice)
Scan Technique:
120kVp, 240 mAs, 0.5 to 1.0 second gantry rotation, 16 x 1.25, pitch 0.9375:1,
18.75mm/rotation, 10mm slice thickness.
Measurement Basis: Helical CTDI, CTDI100, and CTDIw dose calculated from measured axial
CTDI, CTDI100, and CTDIw data and adjusted for 0.9375:1 scan mode and 240 mAs. Dose is
average dose per image for 30 contiguous images (+/- 15% expected deviation).
Table 17-6 Helical Dose (For 4/8 slices)
Scan Technique:
120kVp, 190 mAs, 0.5 to 1.0 second gantry rotation, 0.75:1 acquisition mode with 10mm
nominal image thickness, 15mm table travel/rotation, 512 recon, and standard algorithm.
Measurement Basis: Helical CTDI, CTDI100, and CTDIw dose calculated from measured axial
CTDI, CTDI100, and CTDIw data and adjusted for 0.75:1 scan mode and 190 mAs. Dose is
average dose per image for 30 contiguous images (+/- 15% expected deviation).
Table 17-7 Axial Dose (For 16 slice)
Scan Technique:
120kVp, 260 mAs, 0.5 to 1.0 second gantry rotation, 16 x 1.25, 10mm nominal image
thickness, 2i mode reconstruction. Dose is dose per image.
Table 17-8 Axial Dose (For 4/8 slice)
Scan Technique:
120kVp, 260 mAs, 0.5 to 4.0 second gantry rotation, 10mm nominal image thickness, 2i
mode reconstruction. Dose is dose per image.
Table 17-9 Table of CTDI free air Dose Value (mGy) at Typical Technique (For 4/8 slices)
Head Body
76.8 64.3
Table 17-10 Table of CTDI free air Dose Value (mGy) at KV Adjustment for Body (For 4/8
slices)
80 22.6
100 42.2
140 88.7
Table 17-11 Table of CTDI free air Dose Value (mGy) at Aperture Adjustment for Body (For 4/8
slice)
Table 17-12 Table of CTDI free air Dose Value (mGy) at Typical Technique (For 16 slice)
Head Body
76.3 63.9
Scan Technique:
120kVp, 260 mAs, 0.5 to 4.0 second gantry rotation, 10mm nominal image thickness, 2i
mode reconstruction.
Table 17-13 Table of CTDI free air Dose Value (mGy) at KV Adjustment for Body (For 16
slice)
80 22.4
100 41.9
140 88.1
Scan Technique:
260 mAs, 0.5 to 4.0 second gantry rotation, 10mm nominal image thickness, 2i mode
reconstruction.
Table 17-14 Table of CTDI free air Dose Value (mGy) at Aperture Adjustment for Body (For
16 slice)
0.5mm
An effective 0.5mm voxel size is clearly seen in coronal and sagittal views (2nd smallest
resolution group).
2. Statistic Measurement
Volumetric MTF is computed from the X, Y, and Z axis of a 0.2mm bead using sagittal and
coronal images of the 8¨ Catphan, MTF module with bead source #CTP445.
Scan technique is 120kVp, 100 mAs, 0.5 sec scan rotation, 16 x 0.625mm, 0.562:1 pitch,
5.625mm/rotation, 25 cm SFOV, 10 cm DFOV, Detail Algorithm.
MTF (z-axis) is 12.4 lp/cm @ 10%, 15lp/cm @ 0% which is 0.33mm limiting resolution.
X&Y
MTF (X, Y, & Z) Z
% modulation
lp/cm
Subsystem Specifications
Image Processor
• NVidia Quadro FX1800
• PCI-E
Image Processor
• Nvidia Quadro4 980XGL AGP 8X graphics with 128MB Memory
• Graphics Processor Unit (GPU) Clock 300Mhz
• Graphics Memory Clock 325Mhz
• Dual 350Mhz Video RAMDAC's
or
• NVidia Quadro FX1400 PCI-Express 16X Graphics with 128MB Memory
• Graphics memory bandwidth of 19.2 GB/sec
• Dual 400Mhz Video RAMDAC's
or
• Image Generator consisting of:
– Dual SMP 2.8 GHz Intel Xeon Processors with 1 MB L2 Cache
– Intel Hyper-threading Technology (4 Logical Processors)
– 2 GB DDR 400 ECC Dual Channel Memory Standard
The LightSpeed Series Operator Console user interface features:
• Two, large 20 inch or 21 inch monitors
– Scan/recon monitor mainly for scan and recon control with no image display
– Image monitor mainly for image display, analysis, processing, and management
– Each monitor provides a 1280 x 1024 high resolution, flicker-free display
• Scan control keyboard assembly with intercom speaker, mic and volume controls
• Three button mouse with mouse pad
• BrightBox (trackball assembly)
• Two wide work surfaces
DAS/Detector
16-Row Detector (For 16 slice)
• 16 rows x 888 active patient elements; 15 reference elements
• 70% geometric efficiency
• 99% absorption efficiency
Table
Load Capacity
• 180 kg (400 lbs) with +/- 0.25mm positional accuracy guaranteed
• 205 kg (450 lbs) maximum allowed with normal operation
Cradle Speeds
• 100mm/sec (scout imaging)
• 1.25 - 33.5mm/gantry rotation (helical imaging)
Full Range
SLOW < 120 seconds
Elevation Accuracy
1.5mm
Elevation Range
516mm to 991mm above the floor
Gantry
Tilt Limits
+30° to -30°
Tilt Speed
60 degrees/min nominal
Rotational Freedom
one infinite continuous rotate
X-Ray Tube
Heat Storage
• Anode: 6.3 MHU / 4700 kJ
• Housing: 5.5 MHU / 4100 kJ
Focal Spots
Dual Focal Spots
Anode
• Target angle: 7 degrees
• Up to 53.2 kW maximum radiographic load
Power Supply
5 volt
Maintenance
• Laser alignment lights do not require user maintenance.
• Qualified service personnel must inspect the lights quarterly to assure proper
alignment.
CAUTION: The width of the LIGHT FIELD of Alignment Light is not exceed 3 mm, which is
not represent the actual width of TOMOGRAPHIC SECTION.
3-Phase 50/60 Hz +/- 0.5 Hz (Reference 21CFR 1020.30 (h) (3) (i))
• Phase-to-phase balance within 2% of lowest phase-to-phase voltage.
Maximum Line Current Demand (Reference 21CFR 1020.30 (h) (3) (ii) and (h) (3) (iii))
108A @ 480 V
Maximum line current demand defined at 140 kV and 380 mA.
kV Choices
80, 100 120, 140 kV
Maximum mA
440
Regulation
Recovery within 2 kV for 10% line variation in 50msec
Rise Time
< 2 msec, to attain 75% of selected value
Fall Time
< 10 msec to fall below 75% of selected value
Kilovolts
kV Selections
80, 100, 120, and 140 kV
Milliamperes
mA Selections
10 to 440 mA, by 5 mA increments
• Patient scanning subject to Scan limitations or 53.2 kW maximum
mA Accuracy
Patient scanning selections of 10 to 440 mA
mA Accuracy = +/- (5% + 0.5mA)
Exposure Time
Normal Axial Scan Selections
0.5, 0.6, 0.7, 0.8, 0.9, 1, 2, 3, or 4 sec
Scout
• Scan range: 50 to 1590mm at 100mm/sec
• Exposure time: exposure time increases proportionally with scan distance selection
exposure time = scan range / table speed + 0.08 sec
Helical (Continuous scans with cradle movement)
Up to 120 seconds for a single continuous exposure.
mA
mA: +/- 5%
Exposure Time
Time: +/- 3%
Line Regulation
6% or less.
Measurement Basis
Kilovolts
Precision 10,000:1 voltage dividers built into system reduce generated anode and cathode
voltages.
Resulting low voltage signals provide continuous closed-loop control of the average kV.
Signals are noise filtered and periodically monitored by the computer system.
Operator console displays monitored values during calibration.
Precision 1000:1 voltage divider, model 46-154966G1, (Catalog # C1515A) provides external
feedback during calibration.
• Calibrate Anode and Cathode signals separately.
Milliamperes
Precision shunt resistors, built into the system, measure the tube current component
returned from the secondaries of the
high-voltage transformers.
• The resulting signals provide continuous closed-loop control of the average mA.
• Signals are noise filtered and periodically monitored by the computer system
• Operator console displays the monitored values during calibration.
Check calibration of the shunt resistors and the low voltage mA measuring circuits with a
calibrated digital milliammeter.
Calibrate Anode and Cathode signals separately.
Exposure Time
Traditional Exposure time interval: Duration of time High voltage remains at or above 75% of
selected value.
LightSpeed Series Exposure time interval: Duration of the Expose Command signal within
the Stationary Controller, minus the HV rise time, plus its fall time with respect to the Expose
Command signal.
LightSpeed Series HV components reside on the gantry rotating base.
During stationary scans, use an oscilloscope to measure the HV rise and fall times, with
respect to the Expose Command signal.
Use the oscilloscope to measure the Expose Command signal during stationary scans, to
verify internal time measurements.
Use the internal timer to monitor time during axial/helical scans.
Table 17-15 Symbol and Classification
~
3 335-1 Three-phase alternating current
3N
~ 335-1
Three-phase alternating current
with neutral conductor
Direct Current
Dangerous voltage
Emergency Stop
417-5009 Stand by
Start
Microphone (Mic)
Contrast
Brightness
50 C
System storage prior to installation:
Maintain storage temperature between 0 deg. and +30
deg. C (32 deg. and 86 deg. F)
Humidity 0-70%
Excluding DO NOT store system longer than 90 days
Condensation
System storage prior to installation:
Air Pressure
Maintain Air Pressure between 750 and 1060hPa.
WEEE
Class 1 Equipment
Any permanently installed equipment containing operator or patient accessible surfaces
must provide backup protection against electric shock, in case the basic insulation fails. In
addition to BASIC INSULATION, Class 1 equipment contains a direct connection to a
PROTECTIVE (EARTH) CONDUCTOR which prevents shocks when a person touches a broken
piece of equipment or touches two different equipment surfaces simultaneously.
Type B Equipment
CLASS 1, 2 or 3 EQUIPMENT or EQUIPMENT WITH INTERNAL ELECTRICAL POWER SOURCES
provide an adequate degree of protection against electric shock arising from (allowable)
LEAKAGE CURRENT or a breakdown in the reliability of the protective earth connection.
Ordinary Equipment
Enclosed EQUIPMENT without protection against the ingress of water.
Operation of Equipment
CONTINUOUS OPERATION WITH INTERMITTENT LOADING
Operation in which EQUIPMENT is connected continuously to the SUPPLY MAINS. The stated
permissible loading time is so short that the long term on-load operating temperature is not
attained. The ensuing interval in loading is, however, not sufficiently long for colling down to
the long term no-load operating temperature.
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC MIXTURE
WITH AIR or WITH OXYGEN or NITROGEN OXIDE.
Environmental Specifications
Ratings and duty cycles of all subsystems apply if the site environment complies with the
following.
The specified environment must be constantly maintained, weekends, holidays, and
throughout the night.
Shutdown the CT system whenever the air conditioning fails.
Optional: Turn air conditioner OFF during CT shutdown for repair.
Minimum Allowance
Subsystem
(Watts / BTU/Hr)
Gantry 5500/18766**
Table 200/682
PDU 1500/5118
Operator Console NIO16 1080/3700
Operator Console GOC4 2400/8189
Optional Laser Camera 800/2730
Recommended cooling values should not be used for calculating
system input power requirements.
** Recommended Allowance: 7200/24566
Growth Allowance: 10400/35485
Ambient Temperature
Scan Room
64° - 78° F (18° - 26° C) for patient comfort
Rate of Change
5°F/Hr Max (3°C)
Electromagnetic Interference
Consult GE Healthcare for recommendations when the peak 60 Hz/50 Hz field within the
gantry region exceeds 0.01 gauss peak.
Consider the following when trying to reduce suspected Electromagnetic Interference (EMI):
• The external field strength from a source of magnetic field decreases rapidly with the
distance from the source.
• A bank of three single phase transformers generates a smaller magnetic field (less
external leakage) than a three-phase transformer with an equivalent power rating.
• Large electric motors generate substantial EMI.
• Steel reinforcing in the building structure can act as an
effective conductor of EMI.
• High powered radio signals can affect computers.
• No substitute exists for proper screening of cables and cabinets.
Pollution
Individual components contain filters to optimize environmental conditions.
• Keep air pollution to a minimum.
• Keep the CT suite clean at all times.
• Do not have dust and fume generating work near the system.
• Keep component filters clean and free from obstructions.
Cleaning
The system is not waterproof. It is not designed to protect internal components against
ingress of liquids. Clean external system surfaces (gantry, table, console, accessories) with a
soft cloth dipped in hot water and wrung damp/dry. If necessary, use only mild soap to
remove dirt.
CAUTION: Avoid damage to equipment. Some “spray and wipe” cleaners etch and
permanently cloud clear plastic surfaces. Use only warm water and mild liquid
soap to clean surfaces.
Carpeting
• Install anti-static carpeting - or- treat existing carpets with an anti-static solution.
• Static discharges affect operation and may cause system failures.
Do NOT use steel wool to clean tile floors in scan suite. Fine metal fibers can enter
enclosures and cause internal shorts.
Lighting
Patient Comfort
Use a variable indirect light source between 20 – 100 foot candles in the scan room
Control Room
Select and position subdued light to reduce monitor reflections, and prevent operator eye
strain
Equipment Room
Provide a bright light source for use during maintenance.
Altitude
• Minimum Altitude: 100 feet (30.5 meters) below sea level
• Maximum Altitude: 7,500 feet (2287 m) above sea level
Chapter 18
Planned Maintenance
The following chart gives a description, and frequency of Planned Maintenance (P.M.)
procedures. Please refer to Direction 2141834-100 for the details of each P.M. procedure
and P.M. report charts. LightSpeed Series PMs are based on gantry revolutions.
PMs will be done every 250k gantry revolutions. The average LightSpeed Series scanner
does 1,000,000 revolutions per year, therefore the average scanner has approximately four
Planned Maintenance activities per year.
Table 18-1 LightSpeed Series PM Schedule (based on 1000k revolution cycle/year)
NOTE: Laser Camera PM’s are based on a Quarterly Schedule (2 PM’s per year) vs. PM’s which are
based on Gantry revolutions.
Chapter 19
Product Manufacturer
This chapter lists the manufacturer of the LightSpeedTM Series product. The manufacturer
listed is authorized to CE mark the product listed.
Table 19-1 Product Manufacturer
Manufacturer
Model Name Manufacturer address
Manufacturing Site
West Area of Building No.3, No.1
Yongchang North Road,
GE HANGWEI MEDICAL
SYSTEMS CO., LTD Beijing Economic and
Technological Development Area,
BEIJING 100176 CHINA
LightSpeed Plus GE Healthcare Japan
Corporation (previously 7-127, Asahigaoka 4-chome
named GE Yokogawa Hino-Shi, Tokyo 191-8503, Japan
Medical Systems, LTD).
3000 N. Grandview Blvd.,
GE Medical Systems, LLC
Waukesha, WI 53188, U.S.A.
West Area of Building No.3, No.1
Yongchang North Road,
GE HANGWEI MEDICAL
SYSTEMS CO., LTD Beijing Economic and
Technological Development Area,
BEIJING 100176 CHINA
LightSpeed Ultra GE Healthcare Japan
Corporation (previously 7-127, Asahigaoka 4-chome
named GE Yokogawa Hino-Shi, Tokyo 191-8503, Japan
Medical Systems, LTD).
3000 N. Grandview Blvd.,
GE Medical Systems, LLC
Waukesha, WI 53188, U.S.A.
Chapter 20
Abbreviations
Abbreviations Acronym
AW Advantage Workstation
CT Computed Tomography
CW Clockwise
Abbreviations Acronym
HU Hounsfield Units
HV High Voltage
IV Intra-venous
kV kilo-volts
kW kilo-watts
mA milli-amps (current)
OC Operator’s Console
PM Planned Maintenance
QA Quality Assurance
RF Radio Frequency
Abbreviations Acronym
UL Underwriters’ Laboratories
WL Window Level
WW Window Width
Table A-1 Configured to deliver lower noise in the images at a higher mA value
Table A-2 Configured to deliver average noise in the images at an average mA value
Table A-3 Configured to deliver higher noise in the images at a lower mA value
Chapter Title
CTDIVOL
2
Pediatric and Small Patient Imaging
Chapter Title
2 Radiation Protection
CTDIVOL
3
Pediatric and Small Patient Imaging
11 Scan Parameters
17 ASiR
Chapter Title
Dosimetry
Dose Profile
12
Radiation Protection
Radiation Field
13 Dose
Option Descriptions
The following table describes the function of the system options.
Option Descriptions
Table C-2 Option Descriptions
Operator Instruction
Function Description
Name
An advanced reconstruction technique that enables reduction
Advanced Statistical in image noise and improvement in low contrast detectability
Iterative Reconstruction (LCD) and image quality. ASiR Review provides the ability to
(ASiR) review and assist you in determining the desired level of noise
reduction.
Acquires prospective ECG gating measurements, which
provide information that is valuable for scan timing. Using the
SmartScore Pro
measurements, the system synchronizes the collection of data
with the cardiac cycle.
Provides users the capability to view ECG waveform displayed
ECG Viewer
on the console.
A mode of scanning designed to be used by the Radiologist or
Physician during interventional procedures.Accomplished by
SmartStep
using the integrated Hand Held Controller (HHC) and foot
switch.
Enables Class1 functionality on a CiA425 complaint injector
Xtream Injector allowing only ON/OFF control. Xtream Injector is a start
synchronization of system and injector.
En Enables Class4 functionality on a CiA425 complaint injector
allowing ON/OFF and injector parameters to be set from the
Enhanced Xtream Injector system. Enhanced Xtream Injector is a start synchronization of
the system and injector and injection parameter
communication between the system and injector.
VolumeViewer provides protocols for reformat, 3D rendering
Volume Viewer 3.1
and volume rendering.
G H
Gantry Helical 11-21
Subsystem Specifications 17-21 Helical Overview 11-21
Gantry Coordinate System 11-16 Helical Scan Data Usage 11-27
GE 1-2 Help 1-2
General Information 11-1
Calibration Scans 11-30 I
CT Description 11-1 Image Display 11-31
CT Number 11-32 Image Quality
CT Operation Theory 11-2 Calibration and Scan Image Quality 12-3
Data Collection 11-25 Maintain image quality 5-1, 6-1
Data Storage 11-30 Phantom Image Test and Analysis
DFOV and Pixel Size 11-35 System Performance 12-7
Filament Selection and Scan Thickness QA Schedule
11-24 QA Procedure 12-3
Filament Selection Table 11-24 System Performance
Focal Spot 11-24 System Performance 12-5
Gray Scale 11-31 Image Space 8-1
Image Display 11-31 Image space 8-1
Pixel Coordinates 11-33 Interventional / Biopsy Scanning 3-37
Pixels 11-33
Pixels and CT Numbers 11-36 K
RAS coordinates 11-34 kV
Reconstruction 11-26 Accuracy 17-24
Tube Warmup 11-3 accuracy 17-25
Variables you cannot control 11-33 Measurement Basis 17-26
Window Width 11-37
X-Ray 11-2 L
X-Ray Tube Capacity Learning and Reference Manual 4-1
X-Ray Tube 11-24 Limited Access Room Configuration
General information 3-51
X-Ray 11-2
Generator Specifications 17-24 M
Exposure Time 17-25, 17-27 mA
Kilovolts 17-24, 17-26 Accuracy 17-25
kV, mA and Time Accuracy 17-24 Measurement Basis 17-27
Main Power Supply 17-23 Maintain Image Quality 12-5
Measurement Basis 17-26 Matrix
Measuring tool variance 17-25 Pixel Coordinates 11-33
X
X-Ray
General information 11-2
X-ray Generation and Detection Details 11-16
X-Ray Tube
Filament Selection 11-24
Filament Selection Table 11-24
Focal Spot 11-24
X-Ray Tube Capacity 11-24
X-Ray Tube Specifications 17-22
CT Scan Ratings 14-3
Diagnostic Source Assembly 14-2, 14-18
Performix Ultra Tube Assembly 14-4
Performix Ultra Tube Insert 14-8
Z
Z-Axis Cell Summation 11-18
Imagination at work