SOP of Conductmeter ADWA

The document outlines the Standard Operating Procedure (S.O.P.) for the operation, calibration, cleaning, and maintenance of the Conductivity meter at Eva Cosmetics. It details responsibilities, definitions, procedures for verification, operation, cleaning, and maintenance, along with references to relevant forms and standards. The document is intended to ensure proper usage and upkeep of the Conductivity meter to maintain its accuracy and functionality.

Uploaded by

Fady Emile
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
68 views7 pages

SOP of Conductmeter ADWA

The document outlines the Standard Operating Procedure (S.O.P.) for the operation, calibration, cleaning, and maintenance of the Conductivity meter at Eva Cosmetics. It details responsibilities, definitions, procedures for verification, operation, cleaning, and maintenance, along with references to relevant forms and standards. The document is intended to ensure proper usage and upkeep of the Conductivity meter to maintain its accuracy and functionality.

Uploaded by

Fady Emile
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 7

Isuue/ Revision

No.:001/0
Page 1 of 7
Code: QC-WI-048

S.O.P. of Operation ,Calibration, Cleaning and


Maintenance of Conductivity meter Lab

Name Position Sign./Date

Prepared by Fady Ramzey QC Senior Specialist


Reviewed by Mohamed Gamal Q.C Supervisor
Reviewed by Mina Halim QC Manager
Reviewed by Mark Mazloum QA Manager
Approved by Fady Tawadrous SHEQ Manager
Actual
Planned Revision Revision Result
Ser. No. Revision Date
Revision Date Done By
1-
2-
3-

Issue Date 28.03.2022


Valid From 28.03.2022 Document Stamp
Copy No.

Document Revision Table


Issue Date 28.03.2022
EVA COSMETICS. S.O.P. of Operation ,Calibration,
Corp. S.A.E. Cleaning and Maintenance of
QC Valid From 28.03.2022
Conductivity meter
Issue/Revision No.:001/0 Revision Date : Code: QC-WI-048 Page 7 of 7

Revision
Page No. Date Summary Of Revision
No.

Copy No.: Document Stamp


Issue Date 28.03.2022
EVA COSMETICS. S.O.P. of Operation ,Calibration,
Corp. S.A.E. Cleaning and Maintenance of
QC Valid From 28.03.2022
Conductivity meter
Issue/Revision No.:001/0 Revision Date : Code: QC-WI-048 Page 7 of 7

1. 1. OBJECTIVE:
1.1 To provide a defined procedure for calibration, operation, cleaning and maintenance of
Conductivity meter (s).
2. SCOPE:
2.1 To provide a defined procedure for calibration, operation, cleaning and maintenance of
Conductivity meter (s) found in Eva Cosmetics.

Serial Number 11005260014


Manufacturer ADWA
Made in Hungary
Model AD8000

3. RESPONSIBILITY:
3.1. QC lab Analysts: calibrate, operate ,clean and maintain the instrument .
3.2. Q.C Supervisor: supervise analysts in all these functions.
3.3 Q.C Manger: approved processes

4. DEFINITIONS :
4.1. Calibration : the set operations that establish under specified conditions ,the relationship
between values indicated by an instrument and values of reference standard.
4.2. Maintenance: Actions necessary for retaining or restoring a piece of equipment, machine,
or system to the specified operable condition to achieve its maximum useful life.
4.3. Cleaning : is the process of removing unwanted substances, such as dirt, infectious agents,
and other impurities, from an object or environment
4.4. Verification: provision of objective evidence that a given item fulfills specified
requirements

5. FORMS USED:
5.1. Work Card of operation and cleaning Conductivity meter (QC-F-092).
5.2. Format of Daily Verification of Conductivity Meter (QC-F-093)

6. PROCEDURES:
6.1. Verification:
6.1. Place the meter on a flat surface and protect it from intense heat.
6.2. Connect the device to the electricity power supply
6.3. Ensure that the instrument is clean.
6.4. Adjust Temperature of the solution to 25 ⁰C.
6.5. Open the instrument by press <ON/OFF> key.
6.6. Calibrate the conductmeter as follow (Determining the cell constant).
6.7. Immerse the electrode in potassium chloride solution.
6.8. Press <CAL> repeatedly until CELL is displayed.

Copy No.: Document Stamp


Issue Date 28.03.2022
EVA COSMETICS. S.O.P. of Operation ,Calibration,
Corp. S.A.E. Cleaning and Maintenance of
QC Valid From 28.03.2022
Conductivity meter
Issue/Revision No.:001/0 Revision Date : Code: QC-WI-048 Page 7 of 7

6.9.Adjust ( by using ↓ or ↑ buttons )Cell constant to reach conductivity of potassium chloride


solution to 84 µs/cm .

6.10. Press <CFM> to confirm the selection of CELL.


6.11. The meter stores the cell constant automatically. After this the meter switches to the
measuring mode.
6.12. Rinse the measuring electrode with purified water several times.
6.13 it is possible to set the cell constant value directly , without following the calibration
procedure. To set the cell constant enter the setup mode and select the (cell) item
6.14 Record the verification process in Format of Daily Verification of Conductivity Meter (QC-
F-093)

6.3. Operation:
6.3.1. Ensure that the instrument is clean & free from dust.
6.3.2. Ensure that the plug is connected to the mains.
6.3.3. Press power ‘ ON/OFF ‘ switch.
6.3.4. Ensure that the meter is calibrated and verified.
6.3.5. Ensure that the electrode is washed with purified water.
6.3.6. Dry the glass electrode bulb using a nice filter or tissue paper..
6.3.7. Immerse the measuring electrode into the test sample.
6.3.8. press <CFM> key to start the measurement with stability control. The [AR] display
indicator flashes until a stable value is measure then the flashes stop. This is the conductivity
value of the sample.
6.3.9 Remove the measuring electrode from the sample then rinse it with purified water.
6.3.10. Wipe the measuring electrode with a soft tissue.
6.3.11. Hang the electrode free in air in the holder.
6.3.12 Record the readings in the format of Work Card of operation and cleaning Conductivity
meter (QC-F-092).

6.3. Cleaning:
6.3.1. occasionally wipe the outside of the meter with a damp ,lint -free cloth .
6.3.2. disinfect the housing with isopropanol as required (Not use acetone)

Copy No.: Document Stamp


Issue Date 28.03.2022
EVA COSMETICS. S.O.P. of Operation ,Calibration,
Corp. S.A.E. Cleaning and Maintenance of
QC Valid From 28.03.2022
Conductivity meter
Issue/Revision No.:001/0 Revision Date : Code: QC-WI-048 Page 7 of 7

6.4. Maintenance:
(After measurements, rinse the probe with clean water.
If a more through cleaning is required, remove the probe sleeve and clean the probe with a
cloth.)

Process Time period


Cleaning After each single measurement .
Verification Daily
Calibration Annually (externally)
7.REFERENCES:
7.1. Instrumal Manual.
7.2. ISO9001:2015

Copy No.: Document Stamp


Issue Date 28.03.2022
EVA COSMETICS. S.O.P. of Operation ,Calibration,
Corp. S.A.E. Cleaning and Maintenance of
QC Valid From 28.03.2022
Conductivity meter
Issue/Revision No.:001/0 Revision Date : Code: QC-WI-048 Page 7 of 7

5.2. Format of Daily Verification of Conductivity Meter (QC-F-093)

Verification Of Conductivity meter MODEL: S.N. :

Cleaning
Adjust the cell
Date of signature
constant @25 ̊ c
electrode

Copy No.: Document Stamp


Issue Date 28.03.2022
EVA COSMETICS. S.O.P. of Operation ,Calibration,
Corp. S.A.E. Cleaning and Maintenance of
QC Valid From 28.03.2022
Conductivity meter
Issue/Revision No.:001/0 Revision Date : Code: QC-WI-048 Page 7 of 7

.1. Work Card of operation and cleaning Conductivity Meter AD8000 (QC-F-092).

WORK CARD OF CONDUCTIVITY METER MODEL: ADWA S.N. : 11005260014

NAME OF TESTED CONDUCTIVITY


DATE START ON Q.C. NO. END ON CLEANING SIGN. REMARKS
MATERIAL ΜS/CM