GK GBK GFK Um en

The document is a manual for the GK/GBK/GFK scales by Adam Equipment, detailing setup, operation, calibration, and specifications. It emphasizes environmentally friendly practices and provides instructions for various functions such as zeroing, taring, parts counting, check-weighing, and dynamic weighing. Additionally, it includes information on RS-232 specifications and warranty details.

Uploaded by

pdiogo78
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
49 views23 pages

GK GBK GFK Um en

The document is a manual for the GK/GBK/GFK scales by Adam Equipment, detailing setup, operation, calibration, and specifications. It emphasizes environmentally friendly practices and provides instructions for various functions such as zeroing, taring, parts counting, check-weighing, and dynamic weighing. Additionally, it includes information on RS-232 specifications and warranty details.

Uploaded by

pdiogo78
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 23

Adam Equipment

GK / GBK / GFK SCALE

(P.N. 3056610542, Revision C2, July 2013)

Adam Equipment strives to be more environmentally focused and uses recycled materials and
environmentally friendly packaging where possible. As part of this initiative we have developed a short
manual that uses less paper and ink to describe the main functions of your new Adam indicator/scale. A
complete version is available at www.adamequipment.com. Thank you for your support of Adam Equipment
and we hope that you enjoy your new scale.

© Adam Equipment Company 2013


© Adam Equipment Company 2013
CONTENT

1.0 SETTING UP AND TURNING ON THE SCALES ................................ 3


2.0 OPERATION ......................................................................................... 3
2.1 ZEROING ........................................................................................... 3
2.2 TARING .............................................................................................. 3
2.3 PARTS COUNTING ........................................................................... 4
2.4 CHECK-WEIGHING ........................................................................... 5
2.5 PERCENT WEIGHING ....................................................................... 6
2.6 ANIMAL (Dynamic) WEIGHING ........................................................ 7
2.7 ACCUMULATED TOTAL ................................................................... 8
3.0 RS-232 SPECIFICATION ...................................................................... 9
4.0 CALIBRATION .................................................................................... 10
4.1 CALIBRATION COUNTER FOR APPROVED SCALES ............... 10
5.0 SPECIFICATIONS ............................................................................... 11
6.0 ERROR MESSAGES .......................................................................... 16
7.0 MENU STRUCTURE ........................................................................... 17
8.0 WARRANTY INFORMATION ............................................................. 18

Page1 © Adam Equipment Company 2013


Page2 © Adam Equipment Company 2013
1.0 SETTING UP AND TURNING ON THE SCALES

GK indicator: It must be connected to a load cell platform and calibrated as necessary to match the platform
and user requirements.

GBK/GFK scales: There is a pillar which must be attached to the base frame first using the 4 bolts
supplied. The pillar is secured to the bracket using 2 sets of screws. Then place the platform in the base.
Level the scale by adjusting the four feet. Attach the indicator module to the pillar by sliding it over the
bracket with the flanges engaged in the groves on the base. Attach the cable from the base to the
connector on the rear of the indicator. Attach the power to the indicator.

After completion of setting up, press the [On/Off] key on the rear of the indicator. The software revision
number will be displayed followed by a self-test showing all digits before the zero is displayed along with the
unit of weight that was selected last.

NOTE: The GK and GBK/GFK can be operated from the rechargeable battery. With a single load cell and
backlight disabled the life is approximately 70 hours before needing to be recharged. The battery should be
charged for 12 hours for full capacity.

2.0 OPERATION

2.1 ZEROING
You can press the [Zero] key at any time to set the zero point from which all other weighing and counting is
measured. The scales have an automatic re-zeroing
function to account for minor drifting or accumulation of
material on a connected platform. However you may
need to press [Zero] to re-zero the indicator if small
amount of weight is still shown when the platform is
empty.

2.2 TARING
To determine a weight of sample which is using a container, you need to tare the empty container.
GK/GBK/GFK scales supply two methods of tare: manual tare and preset tare.

Manual Tare

Zero the indicator by pressing [Zero]. The zero indicator will be on. Place a container on the pan.

Press [Tare] when the reading is stable. The weight that was displayed is
stored as the tare value, leaving zero on the display. The stable and Net
indicator will be on.

As a sample is added only the weight of the product will be shown. The
indicator could be tared a second time if another type of product was added
to the first one.

Press [Tare] or [Zero] to remove the tare value and display zero. The Net indicator will disappear.

Page3 © Adam Equipment Company 2013


Preset Tare

When the indicator or scale is at zero with no weight on the platform it is possible to enter a preset tare. First
zeroing the scale, enter a value using the numeric keys. Press [Tare] to tare the indicator. The value that
was entered is stored as the tare value and it is subtracted from the display, leaving a negative number on
the display.

To change the weighing unit press the [Unit] key. The only alternative weighing unit is pounds. OIML
Approved models, GK-M, GBK-M, GFK-M do not allow pounds units

NOTE: This function is not available in GK-M, GBK-M, GFK-M models.

2.3 PARTS COUNTING


The scale can be used to count parts based on the average weight of a sample weighed. If a container is to
be used, place this container on the platform before entering parts counting and press [Tare]. Press [Cnt] to
enter the Parts Counting mode.

The display will show the last sample size used. For example, “10 Pcs”. To
change the sample size, you can press [CE] to clear the last values and then
enter the value 20 using the numeric keypad.

Place the right number of parts on the platform. Then press [Cnt] to determine
an average piece weight. After the sample has been weighed the scale will
count any other parts added by applying the average piece weight to the weight
of the parts to be counted.

During parts counting the display can be changed to show the net weight, unit weight and number of parts by
each time pressing the [Func] key.

To count a different sample quantity, press the [Count] key. And operate as above. To return to weighing,
press [Unit] when “XX pcs” is displayed.

NOTE: If the parts are too light to measure accurately, the count may become faulty. It is suggested that the
samples to be weighed should each weigh more than the resolution of the indicator.

Page4 © Adam Equipment Company 2013


2.4 CHECK-WEIGHING
Check-weighing is a procedure to display an indicator or sound an alarm when the weight on the platform
meets or exceeds the values stored in the memory. The memory holds values for a high limit and a low limit.
Either or both the limits can be set by the user.

The LCD display will indicate whenever the weight is within or exceeds the limits by showing “OK” (mass is
between the limits), “HI”(mass is above the high limit) or “LO”( mass is below the low limit).

The limits can be locked by the manager (see the menu structure section) .A Limit Password must be used
to change the limits or recall other limits from memory.

Setting up the limits

In normal weighing, Press the [Low Limit] key. It will show the current low limit. The “LO” symbol will appear
on the display.

Press the [CE] key to clear the old value and then enter the new low limit using the numeric keys. Then
press [Tare] to accept the value. If you want to reset the value to zero, press [CE] to clear the value. The
limits are displayed in the weighing unit in use.

To set the high limit press [High Limit], the “HI” symbol will be on. Set the high limit in the same way the low
limit was set. Pressing the [Tare] key to enter the value will return the scale to weighing, with the Check-
weighing function enabled. During parts counting and percent weighing, the limits are set in the same way
as above. The limits are displayed in pcs or %.

To disable the check weighing function, enter zero into both the limits as described above. When the current
limits are shown, press [CE] to clear the settings, then press [Tare] to store the zero values.

NOTE: The weight must be greater than 20 scale divisions for check-weighing to operate.

Limits storing and recalling

The indicator can store up to 10 sets of high and low limits in memory along with the weighing units in use
(including pcs and %) as well as settings for the beeper and bar graph. During Check weighing the current
limits can be stored or previously stored units can be recalled.

If you are already in the check weighing mode the display will ask if you wish to
store the current limits by showing “StOrE” or recall another set of limits by
showing “rECALL”. The [Lim] key can be used to toggle between “StOrE” and
“rECALL”.

If you want to store the limits, when “StOrE” is displayed press the [Tare] key.
The display shows “St ”. Enter a number corresponding to the desired memory
location (0 to 9). “St X” will be displayed for 2 seconds indicating the location X
where the current limits, weighing units and settings for the beeper and bar
graph are stored. The indicator will continue to work with the current settings as
active.

Page5 © Adam Equipment Company 2013


If you wish to recall any of the pre-stored limits, press [Tare] when “rECALL” is
displayed. The display shows “rEC”. Enter the number corresponding to the
desired memory location (0 to 9) to be recalled. “rEC X” will be displayed for 2
seconds indicating the values stored in the location “X” is being recalled. The
indicator will change to the recalled limits, weighing units and settings for the
beeper and bar graph.

NOTE: If the recalled limit is for parts counting or percent weighing, the display will show the last sample
value used, ready for a new sample to be counted.

2.5 PERCENT WEIGHING


The scale will use a mass on the platform as the 100% reference weight or input a reference weight using
the keypad.

Steps:

If using a reference weight (or object) as your 100% reference, add the weight to the platform. Press [Func].
The first option is “FUnC 1”, press the [Func] key 3 more times to display “FUnC 4”

Press the [Tare] key. “F4 PCt” will be displayed. Press [Tare] again to enter percent weighing. The scale will
set the sample mass on the platform as 100% reference weight.

NOTE: If there is no reference weight on the pan and percent weighing function is entered, pressing [Tare]
again will return the indicator to normal weighing.

Remove the sample weight. Then any other weight placed on the platform will be displayed as a percentage
of the original sample.

Users can also input the reference weight using the keypad. Make sure the platform is empty and when “F4
PCT” is displayed, enter the weight to be used for the 100% reference, then press [Tare] to accept the
reference weight. The display will show “0.00 %”.

If the indicator shows “x x . x x %”, which is the last weight used as a reference, press [CE] to clear and use
the numeric keypad to enter a new value. Press [Tare] to accept the new reference weight.

Press [Unit] to return to normal weighing.

NOTE: The weight entered must be greater than 50 scale divisions.

The display may jump by large numbers unexpectedly if small weights are used to set as 100% reference.
The indicator checks if the weight is too small and will show Error 7.

Page6 © Adam Equipment Company 2013


2.6 ANIMAL (DYNAMIC) WEIGHING
Steps:

Press [Func]. The first option is “FUnC 1”, press the [Func] key 3 more times to display “Func 4”.

Press the [Tare] key. “F4 PCt” will be displayed. Press the [Func] key to advance to the second function, “F4
AnL”, Animal weighing. And Press [Tare] to enter.

To use the Animal Weighing function it is necessary to set the amount of filtering. More active animals will
require a higher level of filtering to give a stable result. The display will show “Filt x” where x is a value from 1
to 5. To increment the value shown press the [Func] key then press the [Tare] key to accept it.

The display will flash “Ani “ 2 times then show the current weight, 0.00. The scale is now ready to weigh.
Place containers or blankets onto the platform and press the [Tare] key to zero the display. Place the animal
to be weighed on the platform.

When a stable reading is found, the display will show and lock this value, The display will show the “Hold”
symbol .Remove the animal, the display will hold the weight value.

Press the [Unit] key to unlock the display. The display will flash “Ani” twice, and be ready for the next
animal.

Press [zero] key to return to normal weighing.

Page7 © Adam Equipment Company 2013


2.7 ACCUMULATED TOTAL
The indicator can be set to accumulate when a weight is added to the platform automatically or manually by
pressing [Print]. See menu structure section. The accumulation function is available only during weighing. If
at any time the weighing units are changed, the accumulated data will be lost.

Manual Accumulation
When the scale is set to manual accumulation, the weight displayed will be stored in the memory when the
[Print] key is pressed and the weight is stable.

Steps :

Remove the weight and press [Print] when the display is at zero. The display
will show "ACC 1" and then the weight in memory for 2 seconds before returning
to normal. The weight can be output to a printer or PC using the RS-232
interface.

When the indicator is at zero, place a second weight on the platform. When
stable, press [Print] to accumulate the weight. The display will show "ACC 2"
for 2 seconds and then show the new total.

Continue until all weights have been added. This can continue for up to 99 entries until the capacity of
display is exceeded.

To view the total in memory, press the [Print] key when the indicator is at zero. The display will show the
total number of accumulation "ACC xx" and the total weight before returning to zero.

To print the total, press [Print] to recall and then immediately press [Print] the second time to print the
results. To erase the memory, press [Print] to view the total and then immediately press [CE] to clear the
memory.

Automatic Accumulation
When the indicator has been set to Automatic Accumulation the value will be stored in memory automatically.

Place a weight on the platform. The beeper will sound when the display is stable indicating the value is
accepted. Remove the weight. The display will show "ACC 1" and then the total in the memory before it
returns to zero. Adding a 2nd weight will repeat the process.

While the weight is on the platform, press the [Print] key to view the values- first the accumulation number
"ACC x" and then the total will be shown.

Page8 © Adam Equipment Company 2013


3.0 RS-232 SPECIFICATION
The GK indicator and GBK/GFK scale are both supplied with bi-directional RS-232 interface as standard.
The indicator when connected to a printer or computer outputs the weight with the selected weighing unit
through the RS-232 interface.

Default Specifications:

RS-232 output of weighing data

ASCII code

9600 Baud (user selectable)

8 data bits

No Parity

Connector:

9 pin d-sub miniature socket

Pin 3 Output

Pin 2 Input

Pin 5 Signal Ground

Page9 © Adam Equipment Company 2013


4.0 CALIBRATION
The scale can be calibrated using kilogram weights or using pounds weights,(for OIML models, pound is
disabled) depending on the weighing unit selected at the time of calibration.

Steps:

To start the calibration, turn the scale off and switch on again and then press [Tare] during the self-test.
Enter code number 0000 and press [Tare]. This will take you directly to the calibration section or you can get
into the calibration section through the Indicator Settings (“FUnC 3”- see menu structure).

The display will show "UnLoAd"

Remove all weight from the platform and then press the [Tare] key when the display is stable. After the Zero
point is set, the display will show “Ld xx”. Place the suggested calibration mass on the platform. It is best to
use a weight close to the full capacity of the indicator.

If the mass is different from the displayed value, enter the value of the mass in whole numbers. The kg or
the lb symbol will be on to show the active unit. Press the [Tare] key when the stable indicator is on.

When complete, it will display “PASS” and then either display “S8 CAL” (if entered the calibration section
through the Scale Settings) or return to normal weighing (if entered directly). Remove the calibration mass.

If an error message “FAIL H” or “FAIL L” is shown, re-check the calibration and repeat. If the error cannot be
corrected contact your supplier.

4.1 CALIBRATION COUNTER FOR APPROVED SCALES

The approved (GK-M, GBK-M, GK-M Models) scales have the ability to control access to the calibration or
metrology parameters using a passcode to limit access. The requirements for doing this stipulate the code
should be apparent and recorded in a suitable location on the scale.

At power on, the display will show the current software revision number followed by the message of the
Calibration Count “[AL[nt” then a number i.e. “123”. The number is from the counter memory. Then the
Parameter Counter message of “PAr[nt” and probably a different number. The counters cannot be reset to
0, they will increment until the display can no longer hold the values. (1 to 999999). The scale will then
continue to do the display test and go to normal weighing.

If during the time the counting displays are shown, the user presses the [Tare] key, the user will be given a
message to enter the passcode necessary to calibrate the scale, “P - - - - “ Enter the code “P0000” to
Enter calibration or “P1000” to enter the parameters, followed by pressing the [Tare] key.
The Calibration access will allow user calibration and the parameter code will allow access to the following
parameters.

P a g e 10 © Adam Equipment Company 2013


5.0 SPECIFICATIONS
GK Indicator
INPUT SECTION
Load Cells Up to 4 , 350 ohm load cells
Minimum 87 ohms, maximum 1120 ohms
Connection 6 wires
2 excitation, 2 sense, 2 signal
Excitation 5Vdc
Sensitivity 0.15uv/d (GK‐M, 1.5uv/e)
Linearity 0.01% FS
Zero Range 0‐ 10mv
Signal range 0‐40mv
ADC Sensitivity Approximately 0.02 µv/ADC count
DIGITAL SECTION
Maximum Range Typically 1kg – 30000kg
Divisions Up to 30,000, (GK –M, 3000 or less)
Weigh units g, Kg , Lb, ounces, pounds:ounces (GK‐M, kg only)
Stabilisation Time 2 Seconds typical

Operating ‐10°C ‐ 40°C


Temperature 14°F ‐ 104°F
Power supply 230 VAC 50/60 Hz
12V 800ma adaptor for USA versions
Battery Internal rechargeable battery

Calibration Automatic External


Display 6 digits LCD digital displays
with capacity tracker and symbols for units

Indicator Housing ABS Plastic


Overall 260 x 170 x 115mm
Dimensions 10.2” x 6.7” x 4.5”
(wxdxh)
Net Weight 1.8 kg / 4 lb
Applications Weighing and check weighing
Functions Weighing, Check Weighing, Parts counting, check‐counting, , Animal
Weighing, Accumulating memory,
Interface RS‐232 bi‐directional interface
English, German, French, Spanish selectable text

Note: For approved indicators the input specifications is limited to 1.5 µv per division and the
number of divisions is limited to 3000d. Kilograms only.

P a g e 11 © Adam Equipment Company 2013


GFK Models
Model # GFK 75 / GFK 150 / GFK 300 / GFK 600 / GFK 75H / GFK 150H /
GFK 165a GFK 330a GFK 660a GFK 1320a GFK165aH GFK330aH
Maximum Capacity 75kg / 165lb 150kg / 330lb 300kg / 660lb 600kg / 1320lb 75kg / 165lb 150kg / 330lb
Readability 5g / 0.01lb 10g / 0.02lb 20g / 0.05lb 50g / 0.1lb 1g / 0.002lb 2g / 0.005lb
Repeatability (Std Dev) 5g / 0.01lb 10g / 0.02lb 20g / 0.05lb 50g / 0.1lb 2g / 0.004lb 4g / 0.01lb
Linearity + 10g / 0.02lb 20g / 0.04lb 40g / 0.1lb 100g / 0.2lb 3g / 0.006lb 6g / 0.015lb
Units of Measure Kilograms and pounds, grams for GFK75/165a only, XXXa also to have pounds, Ounces, & Pound/Ounces
Stabilization Time 2‐3 Secs
Operating Temperature ‐10°C to +40°C / +14°F to +104°F
Power Supply 230VAC 50/60Hz. in Europe, Asia and South Africa.
12vDC 800mA UL/CSA adapter for USA
Calibration External
Calibration Mass User Selectable
Display Backlit Green display 20mm with capacity tracker
Balance Housing Cast aluminum base, Pantone cool grey painted base, stainless steel grade 304 Top pan, ABS Cool grey indicator housing
Pan Size 400mm x 500mm x 400mm x 500mm x 400mm x 500mm x 600mm x 800mm x 400mm x 500mm x 400mm x 500mm x
65mm/ 15.7” x 19.7” 65mm/ 15.7” x 19.7” 65mm/ 15.7” x 19.7” 65mm/ 65mm/ 15.7” x 65mm/ 15.7” x 19.7”
x 2.6” x 2.6” x 2.6” 23.6” x 31.5” x 2.6” 19.7” x 2.6” x 2.6”
Overal Dimensions 400mm x 620mm x 400mm x 620mm x 400mm x 620mm x 600 mm x 940 mm x 400mm x 620mm x 400mm x 620mm x
(w x d x h) 860mm / 15.7” x 860mm / 15.7” x 860mm / 15.7” x 900 mm 860mm / 15.7” x 860mm / 15.7” x
24.4” x 33.9” 24.4” x 33.9” 24.4” x 33.9” 23.6” x 37” x 35.4” 24.4” x 33.9” 24.4” x 33.9”
Net Weight 12.5kg / 27.6 Lb 12.5kg / 27.6 Lb 12.5kg / 27.6 Lb 25.5 kg / 56.2 Lb 12.5kg / 27.6 Lb 12.5kg / 27.6 Lb

Features Weighing/Counting/Checkweighing with LED lights/Percentage/Hold function/RS232

P a g e 12 © Adam Equipment Company 2013


GBK Models
Model # GBK 8 / GBK 16/ GBK 32 / GBK 60 / GBK 120 /
GBK 16a GBK 35a GBK 70a GBK 130a GBK 260a
Maximum Capacity 8kg/16lb 16kg/35lb 32kg/70lb 60kg/130lb 120kg/260lb
Readability 0.1g/0.0002lb 0.5g/0.001lb 1g/0.002lb 2g/0.005lb 5g/0.01lb
Repeatability (Std Dev) 0.1g/0.0002lb 0.5g/0.001lb 1g/0.002lb 2g/0.005lb 5g/0.01lb
Linearity + 0.2g/0.0004lb 1g/0.002lb 2g/0.004lb 4g/0.01lb 10g/0.02lb
Units of Measure Kilograms and pounds, grams for all except GFK120/260a only, XXXa also to have pounds, Ounces, & Pound/Ounces
Stabilization Time 2‐3 Secs
Operating Temperature ‐10°C to +40°C / +14°F to +104°F
Power Supply 230VAC 50/60Hz. in Europe, Asia and South Africa.
12vDC 800mA UL/CSA adapter for USA
Calibration External
Calibration Mass User Selectable
Display Backlit Green display 20mm with capacity tracker
Balance Housing Cast aluminum base, Pantone cool grey painted base, stainless steel grade 304 Top pan, ABS Cool grey indicator housing
Pan Size 300mm x 400mm x 50mm / 11.8” x 15.7” x 2”
Overal Dimensions
300mmx 520mm x 660mm / 11.8” x 20.5” x 26”
(w x d x h)
Net Weight 7.6kg / 16.8 Lb

Features Weighing/Counting/Checkweighing with LED lights/Percentage/Hold function/RS232

P a g e 13 © Adam Equipment Company 2013


GFK‐M Models
Model # GFK 60M GFK 150M GFK 300M
Maximum Capacity 60kg 150kg 300kg
Readability 0.02kg 0.05kg 0.1kg
Units of Measure Kilograms only
Stabilization Time 2‐3 Secs
Operating Temperature ‐10°C to +40°C / +14°F to +104°F
Power Supply 230VAC 50/60Hz.
Calibration External
Calibration Mass User Selectable
Display Backlit Green display 25mm with capacity tracker
Balance Housing Cast Aluminum base, Pantone cool grey painted base, stainless steel grade 304 Top pan, ABS Cool grey indicator housing
400mm x 500mm
Pan Size
15.7” x 19.7”
Overal Dimensions 400 mm x 620 mm x 860 mm
(w x d x h) 15.7” x 24.4” x 33.9”
Net Weight 12.5kg / 27.6 Lb

Features Weighing/Counting/Checkweighing with LED lights/Percentage/Hold function/RS232

P a g e 14 © Adam Equipment Company 2013


GBK‐M Models
Model # GBK 6M GBK 15M GBK 30M GBK 60M GBK 150M
Maximum Capacity 6kg 15kg 30kg 60kg 150kg
Readability 0.002kg 0.005kg 0.01kg 0.02kg 0.05kg
Units of Measure Kilograms only
Stabilization Time 2‐3 Secs
Operating Temperature ‐10°C to +40°C / +14°F to +104°F
Power Supply 230VAC 50/60Hz.
Calibration External
Calibration Mass User Selectable
Display Backlit Green display 25mm with capacity tracker
Cast Aluminum base, Pantone cool grey painted base, stainless steel grade 304 Top pan, ABS Cool grey indicator
Balance Housing
housing
Pan Size 300mm x 400mm x 50mm / 11.8” x 15.5” x 2”
Overal Dimensions
300mmx 520mm x 660mm / 11.8” x 20.5” x 26”
(w x d x h)
7.6kg / 16.8 Lb
Net Weight

Features Weighing/Counting/Checkweighing with LED lights/Percentage/Hold function/RS232

P a g e 15 © Adam Equipment Company 2013


6.0 ERROR MESSAGES
During the initial power-on testing or during operation, the indicator may show an error message. The
meaning of the error messages is described below.

If an error message is shown, repeat the step that caused the message. If the error message is still shown
then contact your dealer for support.

ERROR DESCRIPTION POSSIBLE CAUSES


CODE

Err 1 Time input Error Invalid time entry such as “268970” for the time
format “H-m-S”.

Err 2 Date input Error 34th day of a month is an invalid entry.

Err 4 Initial Zero is greater than allowed Weight on the platform when turning the
(4% of maximum capacity) when indicator on.
power is turned on or when the
[Zero/Enter] key is pressed. Excessive weight on the platform when zeroing
the indicator.

Platform is not installed.

Improper calibration of the indicator.

Damaged load cell.

Damaged Electronics.

Err 6 A/D count is not correct when turning Load cell is damaged.
the indicator on.
Electronics is damaged.

Err 7 Percent input error Percent function is entered with no reference


mass on the platform.

Err 8 High limit input error Low limit is set first, then the high limit is set
lower than the low limit and high limit not equal
to zero.

Err 9 Low limit input error High limit is set first, then the low limit is set
higher than the high limit and low limit not equal
to zero.

FAIL H or Calibration error Improper calibration (should be within +10% of


the factory calibration). The old calibration data
FAIL L will be retained until the calibration process is
complete.

P a g e 16 © Adam Equipment Company 2013


7.0 MENU STRUCTURE
PARAMETER LAYOUT for GK / GBK / GFK SCALES
Key functions while in this section [Tare] enter a parameter or accept the changes
Press the [Func] key to [Func] move to next parameter or option
enter Functions mode. [Zero] return to previous parameter or return to weighing

FUNC 1 FUNC 2 FUNC 3 FUNC 4


Check weighing parameters RS‐232 Parameters Scale Parameters Scale Parameters

F1 LLk oFF C1 on Prt on S1 Un kg F4 Pct Enter 100%


Limit Lock Enable RS‐232 Prt oFF Units enable lb Percent Weighing reference weight
PSt (pre‐set)
C2 bd 600 S2 bL EL oFF F4 Ani FLt 1 Filter setting
Baud Rate To 19200 Backlight EL on Animal weighing To
F2 Led bAr (Bar type) EL AU (Auto) FLt 5
LED display mA StA (Manual Stable)
Spot (spot type) mA AnY (Manual Any) S3 AoF SLP 0
C3 Prm Au StA (Auto Stable) Set Auto off time SLP 1
SPEA (whole segment) Printing Mode Au End (Auto End) (min.) SLP 5
Ct StA (continuous Stable) SLP 10
Ct AnY(Continuous Any)
F3 bEP bP oFF
S4 dt Set as described in
Beeper Control C4 Aon on Set time and date manual
bP Int (Inside Limits) Enable Accumulation oFF
S5 dIS All
bP otL (Outside Limits) C5 Ln English Display mode StAb (only when
Language for printing French stable)
F4 CPS Enter using numeric German
Check weighing method Spanish S6 Fi SLoW
password
Set Filter nor (normal)
C6 Uid Enter using numeric keys FASt
F5 NCk On User ID
Negative Check
S7 SPS Enter using numeric
weighing Off C7 Sid Enter using numeric keys Scale password keys
Scale ID
C8 LAb LAb On S8 CAL Perform calibration
Lab Off

17 | P a g e © Adam Equipment Company 2013


8.0 WARRANTY INFORMATION
Adam Equipment offers Limited Warranty (Parts and Labour) for components that fail due to defects in
materials or workmanship. Warranty starts from the date of delivery.

During the warranty period, should any repairs be necessary, the purchaser must inform its supplier or
Adam Equipment Company. The company or its authorised Technician reserves the right to repair or
replace the components at any of its workshops depending on the severity of the problem. However,
any freight involved in sending the faulty units or parts to the service centre will be borne by the
purchaser.

The warranty will cease to operate if the equipment is not returned in the original packaging and with
correct documentation for a claim to be processed. All claims are at the sole discretion of Adam
Equipment.

This warranty does not cover equipment where defects or poor performance is due to misuse,
accidental damage, exposure to radioactive or corrosive materials, negligence, faulty installation,
unauthorised modifications or attempted repair or failure to observe the requirements and
recommendations as given in this User Manual. Additionally rechargeable batteries (where supplied)
are not covered under warranty.

Repairs carried out under the warranty does not extend the warranty period. Components removed
during the warranty repairs become the company property.

The statutory right of the purchaser is not affected by this warranty. The terms of this warranty is
governed by the UK law. For complete details on Warranty Information, see the terms and conditions
of sale available on our web‐site.

18 | P a g e © Adam Equipment Company 2013


19 | P a g e © Adam Equipment Company 2013
WEEE 2012/19/EU

This device may not be disposed of in domestic waste. This also applies to countries outside the
EU, per their specific requirements. Disposal of batteries (if fitted) must conform to local laws
and restrictions.
Cet appareil ne peut être éliminé avec les déchets ménagers. L’élimination de la batterie doit
être effectuée conformément aux lois et restrictions locales.
Dieses Gerät nicht mit dem Hausmüll entsorgt.
Dispositivo no puede ser desechado junto con los residuos domésticos
Dispositivo non può essere smaltito nei rifiuti domestici.
FCC / IC CLASS A DIGITAL DEVICE EMC VERIFICATION STATEMENT
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC rules and Canadian ICES‐003/NMB‐003 regulation. These
limits are designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the user
will be required to correct the interference at his own expense.
CALIFORNIA PROPOSITION 65 ‐ MANDATORY STATEMENT
WARNING: This product includes a sealed lead‐acid battery which contains chemicals known to
the State of California to cause cancer and birth defects or other reproductive harm.

Adam Equipment products have been tested with, and are always supplied
with mains power adaptors which meet all legal requirements for the
intended country or region of operation, including electrical safety,
interference and energy efficiency. As we often update adaptor products to
meet changing legislation it is not possible to refer to the exact model in this
manual. Please contact us if you need specifications or safety information for
your particular item. Do not attempt to connect or use an adaptor not
supplied by us.

© Adam Equipment Company 2013


ADAM EQUIPMENT is an ISO 9001:2008 certified global company with more than 40 years’ experience in
the production and sale of electronic weighing equipment.

Adam products are predominantly designed for the Laboratory, Educational, Health and Fitness, Retail and
Industrial Segments. The product range can be described as follows:

‐Analytical and Precision Laboratory Balances


‐Compact and Portable Balances
‐High Capacity Balances
‐Moisture analysers / balances
‐Mechanical Scales
‐Counting Scales
‐Digital Weighing/Check‐weighing Scales
‐High performance Platform Scales
‐Crane scales
‐Mechanical and Digital Electronic Health and Fitness Scales
‐Retail Scales for Price computing
For a complete listing of all Adam products visit our website at
www.adamequipment.com

Adam Equipment Co. Ltd. Adam Equipment Inc. AE Adam GmbH.


Maidstone Road, Kingston 1, Fox Hollow Rd. Instenkamp 4
Milton Keynes Oxford CT D‐24242 Felde
MK10 0BD 06478
UK USA Germany
Phone:+44 (0)1908 274545 Phone: +1 203 790 4774 Phone +49 (0)4340 40300 0
Fax: +44 (0)1908 641339 Fax: +1 203 792 3406 Fax: +49 (0)4340 40300 20
e‐mail: e‐mail: e‐mail:
[email protected] [email protected] [email protected]

Adam Equipment S.A. (Pty) Ltd. Adam Equipment (S.E. ASIA) PTY Adam Equipment (Wuhan) Co. Ltd.
7 Megawatt Road, Ltd A Building East Jianhua
Spartan EXT 22 2/71 Tacoma Circuit Private Industrial Park
Kempton Park, CANNING VALE 6155 Zhuanyang Avenue
Johannesburg, Perth Wuhan Economic & Technological
Republic of South Africa Western Australia Development Zone
430056 Wuhan
P.R.China
Phone +27 (0)11 974 9745 Phone: +61 (0) 8 6461 6236 Phone: + 86 (27) 59420391
Fax: +27 (0)11 392 2587 Fax +61 (0) 8 9456 4462 Fax + 86 (27) 59420388
e‐mail: e‐mail: e‐mail:
[email protected] [email protected] [email protected]

© Copyright by Adam Equipment Co. All rights reserved. No part of this publication may be reprinted or translated in
any form or by any means without the prior permission of Adam Equipment.
Adam Equipment reserves the right to make changes to the technology, features, specifications and design of the
equipment without notice.
All information contained within this publication is to the best of our knowledge timely, complete and accurate when
issued. However, we are not responsible for misinterpretations which may result from the reading of this material.
The latest version of this publication can be found on our Website.
www.adamequipment.com

© Adam Equipment Company 2013