Implantable NFC Glucose Monitor Design

This document presents a fully implantable, battery-assisted, passive continuous glucose monitor that integrates a glucose sensor interface, NFC wireless communication, and a power management unit. The system utilizes a 13.56 MHz RF signal for energy harvesting and data transmission, implemented in 14nm CMOS technology. The design aims to provide a reliable and non-invasive solution for continuous glucose monitoring, addressing limitations of current methods.

Uploaded by

adilrchanna
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
60 views16 pages

Implantable NFC Glucose Monitor Design

This document presents a fully implantable, battery-assisted, passive continuous glucose monitor that integrates a glucose sensor interface, NFC wireless communication, and a power management unit. The system utilizes a 13.56 MHz RF signal for energy harvesting and data transmission, implemented in 14nm CMOS technology. The design aims to provide a reliable and non-invasive solution for continuous glucose monitoring, addressing limitations of current methods.

Uploaded by

adilrchanna
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 16

HHS Public Access

Author manuscript
IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017
Author Manuscript

July 10.
Published in final edited form as:
IEEE EMBS Int Conf Biomed Health Inform. 2016 February ; 2016: 612–615. doi:10.1109/BHI.
2016.7455973.
A Fully Implantable, NFC Enabled, Continuous Interstitial
Glucose Monitor
Nijad Anabtawi,
Intel Corporation, Chandler, AZ and the Department of Electrical Engineering, Arizona State
University, Tempe, AZ

Sabrina Freeman, and


Department of Biomedical Engineering, Arizona State University, Tempe, AZ
Author Manuscript

Rony Ferzli [Senior Member, IEEE]


Department of Electrical Engineering, Arizona State University, Tempe, AZ

Abstract
This work presents an integrated system-on-chip (SoC) that forms the core of a long-term, fully
implantable, battery assisted, passive continuous glucose monitor. It integrates an amperometric
glucose sensor interface, a near field communication (NFC) wireless front-end and a fully digital
switched mode power management unit for supply regulation and on board battery charging. It
uses 13.56 MHz (ISM) band to harvest energy and backscatter data to an NFC reader. System was
implemented in 14nm CMOS technology and validated with post layout simulations.
Author Manuscript

I. INTRODUCTION
Currently, an estimated 400 million individuals worldwide have been diagnosed with
diabetes. A leading cause of death and disability throughout the world, this disease is
expected to reach epidemic levels in the next few decades [1]. Management of this condition
centers on maintenance of blood glucose levels within a certain range though the frequent
blood sampling and administration of insulin injections. Continuous glucose monitoring
(CGM) can help early diagnosis and effective control of diabetes complications.

Enzyme-based finger-pricking measurement is most widely used. However, blood glucose


levels can change rapidly (0.125 mM/min) leading to missed detection of hypoglycemic and
hyperglycemic events. Moreover, the procedure is invasive and may cause infection. Another
Author Manuscript

measurement approach is the use of noninvasive near-infrared spectroscopy [2]. This method
analyzes the light scattering or transmission spectrum in the fingertip to infer metabolic
concentration, but is susceptible to optical interference and poor specificity and is therefore
not sufficient for clinical use. Electrochemical sensing of glucose in body fluids such as
tears, or sweat offers the most accurate reading as it is immune to optical and other
metabolite interferences and can be used in a noninvasive manner [3]. However, due to the
poor correlation between tear and sweat with blood glucose and the presence of a significant
event propagation delay, these approaches are not clinically useful. A better approach would
be the measurement of glucose in body interstitial fluids found beneath the skin for instance.
Commercially available CGM systems are transdermal and battery powered requiring
Anabtawi et al. Page 2

frequent costly replacements due to limited battery life, whereas implantable passive
Author Manuscript

systems rely on rectified voltage recovered from an RF carrier field to transiently power
sensing and communication, thereby lacking continuous monitoring capabilities in the
absence of the RFID reader. Battery assisted passive sensing offers the best of both active
and passive paradigms. It allows for small footprint fully implantable sensors that can be
powered using conventional NFC fields and have a small on board rechargeable battery that
enables continued sensing.

This work presents an integrated system-on-chip (SoC) that forms the core of a long-term,
fully implantable, battery assisted, passive continuous glucose monitor (Fig. 1). It uses a
13.56 MHz, Industrial Scientific and Medical band (ISM) RF signal to harvest energy and
communicate data to an interrogating reader. Harvested energy is stored in an off chip
printable battery for use in absence of the RF field. The monitor includes an off-chip
amperometric electrochemical glucose sensor, ferrite antenna, printable battery, and SoC.
Author Manuscript

Section II describes the proposed architecture whereas implementation is provided in section


III. Transistor level simulations are presented in section IV, and finally conclusions in
section V.

II. SYSTEM LEVEL ARCHITECTURE


Fig. 2 illustrates the System level architecture of the SoC. Sensor read out and digitization
includes a potentiostat that interfaces directly with the glucose sensor leads and conditions
the readout signal for digitization by the frequency discriminator based analog to digital
converter (ADC). During wireless transmission, two coupled coils are used, the reader’s coil
and the SoC connected ferrite antenna. NFC front end rectifies harvested RF energy
providing an unregulated supply (Vrec) to the power management unit. It is also responsible
Author Manuscript

for communication with the reader via load modulation. Power management unit generates
two regulated, 0.4V (digital) and 0.9V (analog) supplies. When an interrogating field is
present, Vrec is supplied by rectification circuitry. The regulated digital output powers the
digital core and recharges the off-chip battery and on-chip storage capacitor. Once the RF
field is removed, Vrec is supplied by the battery and storage cap. The digital partition
includes control and calibration circuits, decodes incoming commands and manages system
operation.

III. IMPLEMENTATION
A. Amperometric Sensor Front-End and ADC
The amperometric sensor used has an active area of 0.1 mm2 and sensitivity of 0.05 μA·mm
Author Manuscript

−2 · mM−1 for clinically relevant glucose concentrations (0 – 30 mM). It has three


electrodes, working, reference, and counter electrodes (WE, RE, and CE, respectively, Fig.
3). When a constant voltage is applied between (WE) and (RE) (250 mV in this work) by the
potentiostat, a current proportional to interstitial glucose concentration is produced. For RE
to ideally draw zero current and track the applied voltage, the amplifiers A1 and A2 require
large gain. This is achieved using folded cascaded topology with device stacking and rail-to-
rail input/output stages to drive the large double layer capacitor of CE and WE. The output
current is converted to voltage via the I-to-V converter. The front-end can resolve a current

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 3

range of 1 nA to 1μA with an Integrated input referred noise voltage of 20 μVrms in the
Author Manuscript

measurement band of 1 – 25 Hz. Input referred DC offset is 1.6 nV. A 1st order, digital, none
feedback, ΣΔ frequency Discriminator (ΣΔFDC) is used to implement the ADC [4]. It
processes a frequency modulated (FM) input signal to produce a single bit output stream. A
schematic of the ΣΔFDC is shown in Fig. 4. It consists of a voltage controlled oscillator
(VCO), two D flip flops and one XOR gate. Decimation of ΣΔFDC output is implemented
using CIC filter to generate a 10 bit code word.

B. Power Management
Fig. 5 illustrates the proposed power management unit used in this work. It is a switched
mode single inductor, dual output regulator. Two error signals VERR1 and VERR2 are
generated from the regulated output voltages VOUT1 (digital) and VOUT2 (analog)
respectively [5]. The signals VERR1 and VERR2 are compared using the comparator (Comp).
Author Manuscript

The higher of the two signals is then passed through the ‘signal select’ block to the PID
controller. Instead of using a conventional pulse width (PWM) or pulse code (PCM)
modulator to drive the power stage, a 2nd order ΣΔ modulator (Fig. 6) is used to avoid tonal
noise. The comparator’s decision controls the operation of switches S3 and S4. If VERR1 >
VERR2, then S3 is closed and S4 is left open and vice versa. This arrangement allows the
control loop to regulate the two output voltages by tracking the signals VERR1 and VERR2
and adjusting the on/off times for S3 and S4. Digital implementation of the loop controller
allows for low power consumption, ease of programmability, and high tolerance to mismatch
and process variations. Blocks D1 and D2 digitize the output voltages VOUT1 and VOUT2
using 1st order ΣΔFDCs similar to the one used to digitize the sensor’s output. To generate
the error signals VERR1 and VERR2, the digital representation of VOUT1 and VOUT2 is
subtracted from the digitized reference voltages VREF1 and VREF2 respectively. To reduce
Author Manuscript

the switching ripple generated by the regulator and generate a stable analog supply for the
amperometric sensor, an LDO is used in series with the VOUT2. This arrangement allows for
maximum power efficiency and very low supply line ripple needed for accurate reference
generation and sensor readout. Inductor (L) is realized using on chip top metal spiral
windings.

C. NFC Front End


An ASK demodulation circuit (Fig. 7b) composed of a rectifier, diode based envelope
detector, low pass filter and comparator with an average signal driven threshold is used to
read the interrogator’s commands. Fig. 7a shows the clock extraction circuit used to extract a
13.56-MHz clock from the carrier. It is composed of a high bandwidth differential
comparator followed by inverters. Extracted clock is used to generate remaining system
Author Manuscript

clocks in the digital partition. A 40mV minimum antenna output signal is needed to for
successful extraction. Load shift key (LSK) is used to generate remaining system clocks in
the digital partition. A 40mV minimum antenna output signal is needed to for successful
extraction. Load shift key (LSK) is used to transmit data back to the reader by varying the
load impedance of the passive wireless link.

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 4

IV. Simulation Results


Author Manuscript

The SoC was designed in 14nm bulk CMOS process and validated through simulations.
Target specifications for the integrated power management unit are a total output current of 1
mA while charging the battery and operating the sensor and NFC front ends simultaneously,
down to 15 μA while running off battery power alone. Output switching ripple must remain
below 20 mV and 10 μV for digital and analog regulated supplies respectively. Fig. 8
demonstrates load regulation test in presence of the RF field. The two output voltages, Vout1
and Vout2, are regulated to 0.4 V and 0.9 V respectively. At t = 100μs, Total output current
(IL) is ramped up from 15 μA to 28 μA and back down at t = 250μs, both outputs settle back
to nominal values in 12 μs. Fig. 9a demonstrates the measured sensor only transfer
characteristics and simulated post layout VCO sensitivity. The simulated results of a
continuous glucose flow test are shown in Fig. 10. The SoC layout is provided in Fig. 11.
Key metrics are summarized in Table 1.
Author Manuscript

V. Conclusion
This work presented an integrated system-on-chip (SoC) that forms the core of a long-term,
fully implantable, battery assisted, passive continuous glucose monitor. It integrates an
amperometric glucose sensor interface, a near field communication (NFC) wireless front-
end and a fully digital switched mode power management unit for supply regulation and on
board battery charging. It uses 13.56 MHz (ISM) band to harvest energy and backscatter
data to an NFC reader. System was implemented in 14nm CMOS technology and validated
with post layout simulations.

References
Author Manuscript

1. International diabetic federation. 2014 Dec. [Online]. Available:http://www.idf.org/


2. Dehennis A, et al. An NFC enabled CMOS IC for a Wireless, Fully Implantable Glucose Sensor.
Biomedical and Health Informatics, IEEE Journal of. PP(99):1–1.
3. Ricci F, et al. Ex vivo continuous glucose monitoring with microdialysis technique: The example of
glucoday. IEEE Sens J. Jan; 2008 8(1):63–70.
4. Wismar, et al. A 0.2V, 7.5 μW, 20 kHz ΣΔ modulator with 69 dB SNR in 90 nm CMOS. Solid State
Circuits Conference, 2007. ESSCIRC 2007. 33rd European; Sept. 2007;
5. Anabtawi, N., Ferzli, R. A simplified single-inductor dual-output DC-DC buck converter
architecture with a fully digital Σ-Δ based controller. Quality Electronic Design (ISQED), 2015 16th
International Symposium on; 2–4 March 2015; p. 82-85.
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 5
Author Manuscript
Author Manuscript

Fig. 1.
Wireless powered implantable sensor tag for continuous glucose monitoring.
Author Manuscript
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 6
Author Manuscript

Fig. 2.
Author Manuscript

System level architecture for the interstitial glucose monitor SoC.


Author Manuscript
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 7
Author Manuscript

Fig. 3.
Analog front-end for the glucose sensor.
Author Manuscript
Author Manuscript
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 8
Author Manuscript

Fig. 4.
Author Manuscript

ΣΔ based frequency discriminator analog-to-digital converter.


Author Manuscript
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 9
Author Manuscript
Author Manuscript

Fig. 5.
System level architecture for the power management unit.
Author Manuscript
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 10
Author Manuscript

Fig. 6.
Author Manuscript

2nd order digital ΣΔ modulator implementation.


Author Manuscript
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 11
Author Manuscript
Author Manuscript

Fig. 7.
Schematic diagrams of the (a) clock extractor and (b) demodulation circuits.
Author Manuscript
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 12
Author Manuscript
Author Manuscript
Author Manuscript

Fig. 8.
Power regulation during charging operation showing (a) total output current, (b) regulated
digital output (VOUT1) and (c) regulated analog output (VOUT2). Charging engaged at t =
100μs and disengaged at t = 250μs.
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 13
Author Manuscript
Author Manuscript

Fig. 9.
(a) Measured sensor current versus glucose concentration and (b) simulated VCO frequency
versus sensor output current.
Author Manuscript
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 14
Author Manuscript
Author Manuscript

Fig. 10.
Post layout transient simulation response of the continuous glucose monitor running on
battery power (RF rectification disengaged). Glucose concentration 0mM.at t = 0s.
Author Manuscript
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 15
Author Manuscript
Author Manuscript

Fig. 11.
Layout of the glucose monitor SoC showing base layers only. Inductor windings on top
metal layers are not visible.
Author Manuscript
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.
Anabtawi et al. Page 16

Table 1

Key System Metrics


Author Manuscript

Unregulated Supply Voltage Range (V) 1.0 – 3.3

On-Chip Inductor (nH) 20

Efficiency, Rectifier, Regulator (%) 46, 94

Total Power Consumption (W) Charging Enabled: 47m


Charging Disabled: 24μ
- Potentiostat: 2.5μ
- ΣΔFDC: 50n
- NFC Front-End: 12μ
- Digital Controller:110n
- SMPS (Loop Controller): 9μ
Author Manuscript
Author Manuscript
Author Manuscript

IEEE EMBS Int Conf Biomed Health Inform. Author manuscript; available in PMC 2017 July 10.