ABB Ax400 - Series Orp and PH Controller Manual
ABB Ax400 - Series Orp and PH Controller Manual
A B B M E A S U R E M E N T & A N A LY T I C S | U S E R G U I D E | I M/A X 4 P H R E V. P
AX400 series
For more information
pH/Redox (ORP)
analyzers
Further publications are available for free download
from:
www.abb.com/analytical
Contents
1 Introduction ..................................................................... 2 6 Installation .................................................................... 44
1.1 System Description .................................................. 2 6.1 Siting Requirements .............................................. 44
1.2 PID Control .............................................................. 2 6.2 Mounting ............................................................... 45
1.3 AX400 Series Analyzer Options ................................ 2 6.2.1 Wall-/Pipe-mount Analyzers ....................... 45
6.2.2 Panel-mount Analyzers ............................... 46
2 Operation ......................................................................... 3 6.3 Connections, General ............................................ 47
6.3.1 Relay Contact Protection
2.1 Powering Up the Analyzer ........................................ 3
and Interference Suppression ..................... 48
2.2 Displays and Controls .............................................. 3
6.3.2 Cable Entry Knockouts,
2.2.1 Membrane Key Functions ............................. 3
Wall-/Pipe-mount Analyzer ......................... 49
2.3 Operating Page ........................................................ 6
6.4 Wall-/Pipe-mount Analyzer Connections ................ 50
2.3.1 Single Input pH ............................................. 6
6.4.1 Access to Terminals ................................... 50
2.3.2 Dual Input pH ............................................... 7
6.4.2 Connections ............................................... 51
2.3.3 Single Input Redox (ORP) ............................. 8
6.5 Panel-mount Analyzer Connections ....................... 52
2.3.4 Dual Input Redox (ORP) ................................ 9
6.5.1 Access to Terminals ................................... 52
2.3.5 Dual Input pH and Redox (ORP) ................. 10
6.5.2 Connections ............................................... 53
2.3.6 Wash Function ............................................ 11
6.6 pH Sensor Systems Connections ......................... 54
6.6.1 Standard ph Systems Connection – 2867,
3 Operator Views .............................................................. 12 AP100, AP300, 7650/60, TB5, Non-ABB ... 54
3.1 View Set Points ...................................................... 12 6.6.2 Differential ph Systems Connections –
3.2 View Outputs ......................................................... 14 Capable of Providing Sensor Diagnostics
3.3 View Hardware ....................................................... 14 (AP200, TBX5) ........................................... 55
3.4 View Software ........................................................ 15
3.5 View Logbook ........................................................ 16 7 Calibration ..................................................................... 56
3.6 View Clock ............................................................. 18
7.1 Equipment Required .............................................. 56
7.2 Preparation ............................................................ 56
4 Setup ............................................................................. 19 7.3 Factory Settings .................................................... 57
4.1 Sensor Calibration .................................................. 19
4.1.1 Set Buffer Type (pH Only) ........................... 19 8 Simple Fault Finding ..................................................... 62
4.1.2 Set Up User Defined Buffers (pH Only) ........ 21
8.1 Error Messages ..................................................... 62
4.1.3 Adjust Offset (Redox/ORP Only) .................. 22
8.2 Calibration Fail Message or No
4.1.4 Automatic, Single- and
Response to pH/Redox Changes .......................... 62
Two-Point Calibration (pH Only) .................. 23
8.3 Checking the Temperature Input ............................ 63
4.1.5 Manual, Single- and
Two-Point Calibration (pH Only) .................. 25
4.1.6 Grab Calibration (pH Only) .......................... 27 9 Specification ................................................................. 64
IM/AX4PH Rev. P 1
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 1 Introduction
1 Introduction All models incorporate a wash facility for system cleaning; the
Alarm 3 relay can be configured to control the wash system
1.1 System Description either automatically or manually. The relay can be programmed
The AX460 single input and AX466 dual input pH/Redox (ORP) to deliver either a continuous or pulsed signal to control an
analyzers and associated electrode systems have been external power supply to a solenoid or pump and the frequency,
designed for continuous monitoring and control of pH and duration and recovery time for the wash cycle are also
Redox (ORP). The electrode system can be standardized to the programmable. During a wash cycle, the analog output value is
analyzer using the built-in calibration facility and a single point held in its pre-cycle condition.
buffering facility provides easy re-calibration after initial Analyzer operation and programming are performed using five
standardization. tactile membrane keys on the front panel. Programmed
The analyzer is available in wall-/pipe-mount or panel-mount functions are protected from unauthorized alteration by a
versions with either one or two programmable, pH or Redox four-digit security code.
(ORP) input channels, each with its own associated temperature
input channel. When making temperature compensated 1.2 PID Control
measurements, the sample temperature is sensed by a The AX460 single input pH analyzer incorporates Proportional
resistance thermometer (Pt100, Pt1000 or Balco 3K) mounted Integral Derivative (PID) control as standard. Refer to the
in the electrode system. PID Control Supplementary User Guide, IM/AX4PID for a full
description and instructions on how to configure and operate
The analyzer can be configured for, and connected to, either a PID control.
standard pH input (single, high impedance input >1013 ) or
differential pH input (dual, high impedance inputs, both >1013 ). 1.3 AX400 Series Analyzer Options
Differential pH input is designed for use with pH electrode Table 1.1 shows the range of configurations that are possible for
systems that incorporate a solution earth (ground) rod. The the AX400 Series analyzers. The analyzer detects the type of
measuring electrode and reference electrode signals are input board fitted for each input automatically and displays only
measured separately using two, high impedance amplifiers and the operating and programming frames applicable to that input
compared with the solution earth (ground) potential. The board type. If no input board is fitted for a second input
difference between the results is the value used for the pH (Sensor B), Sensor B frames are not displayed.
measurement.
AX411 Dual Input 2-Electrode Conductivity (0 to 10,000 µS/cm) 2-Electrode Conductivity 2-Electrode Conductivity
AX413 Dual Input 2-Electrode Conductivity and 4-Electrode Conductivity 2-Electrode Conductivity 4-Electrode Conductivity
AX416 Dual Input 2-Electrode Conductivity and pH/Redox (ORP) 2-Electrode Conductivity pH/Redox (ORP)
AX418 Dual Input 2-Electrode Conductivity and Dissolved Oxygen 2-Electrode Conductivity Dissolved Oxygen
AX433 Dual Input 4-Electrode Conductivity (0 to 2,000 mS/cm) 4-Electrode Conductivity 4-Electrode Conductivity
AX436 Dual Input 4-Electrode Conductivity and pH/Redox (ORP) 4-Electrode Conductivity pH/Redox (ORP)
AX438 Dual Input 4-Electrode Conductivity and Dissolved Oxygen 4-Electrode Conductivity Dissolved Oxygen
AX455 Dual Input 2-Electrode Conductivity (USP) 2-Electrode Conductivity 2-Electrode Conductivity
AX456 Dual Input 2-Electrode Conductivity (USP) and pH/Redox (ORP) 2-Electrode Conductivity pH/Redox (ORP)
AX468 Dual Input pH/Redox (ORP) and Dissolved Oxygen pH/Redox (ORP) Dissolved Oxygen
2 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
Menu 2
1. Ensure the input sensors are connected correctly.
A Moving Between Menus
2. Switch on the power supply to the analyzer. A start-up
screen is displayed while internal checks are performed,
then the Operating Page (Section 2.3) is displayed as the Advance to
pH or Redox (ORP) monitoring operation starts. next page
Page X
Display Frame 1
Lines
Advance to
0.00 pH Units
next Frame
Lower
Display Line 0.00pH Frame 2
Frame 3
Monitoring pH
Frame 4
Alarm C Moving Between Frames
LEDs
Parameter X
Fig. 2.1 Location of Controls and Displays Y Select
Z
New value is
automatically stored
IM/AX4PH Rev. P 3
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
Section 2.3, Page 6 Section 3.1, Page 12 Section 3.2, Page 14 Section 3.3, Page 14 Section 3.4, Page 15 Section 3.5, Page 16 Section 3.6, Page 18
OPERATING PAGE VIEW SETPOINTS VIEW OUTPUTS VIEW HARDWARE VIEW SOFTWARE VIEW LOGBOOK VIEW CLOCK
A1: Setpoint Analog Output 1 Sensor A Module AX400/2000 Issue Alarms Date 01:01:03
A2: Setpoint Analog Output 2 Sensor B Module Errors Time 12:00
Use the Downscroll
Key to scroll through A3: Setpoint Analog Output 3 Option Board Power
the Frames Cals
A4: Setpoint Analog Output 4
within each Page
A5: Setpoint
A: Electrode B: Electrode
A: Temp. Comp B: Temp. Comp
A: Temp. Sensor B: Temp. Sensor
* A:
A: Enable Cals
Preset Temp.
B: Enable Cals
B: Preset Temp.
*
A: Sample Comp. B: Sample Comp.
A: Sample Coeff. B: Sample Coeff.
A: Set Min Slope B: Set Min Slope
4 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
*
A1: Assign A2: Assign A3: Assign A4: Assign A5: Assign
Use the Downscroll
Key to scroll through A1: Failsafe A2: Failsafe A3: Failsafe Wash Mode A4: Failsafe A5: Failsafe
the Frames A1: Action A2: Action A3: Action Wash Frequency A4: Action A5: Action
within each Page
A1: Setpoint A2: Setpoint A3: Setpoint Wash Duration A4: Setpoint A5: Setpoint
A1: Hysteresis A2: Hysteresis A3: Hysteresis Recovery Period A4: Hysteresis A5: Hysteresis
A1: Delay A2: Delay A3: Delay A4: Delay A5: Delay
* toApplicable
Alarm 3
only
AO1: Zero Value AO2: Zero Value AO3: Zero Value AO4: Zero Value
AO1: Default AO2: Default AO3: Default AO4: Default
AO1: Default Val AO2: Default Val AO3: Default Val AO4: Default Val
Time 12:00
Press To Set Press To Abort
Displayed only if option board fitted and serial communications feature enabled (Section 7.3)
CONFIG.SERIAL
see Supplementary Manual Profibus® Datalink Description (IM/AX4/PBS)
CONFIG. CONTROL Single Input Analyzer Only see Supplementary Manual PID Control (IM/AX4PID)
Test Output 4
To FACTORY SETTINGS
(see Section 7.3, Page 57)
IM/AX4PH Rev. P 5
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
Measured Values
7.00 pH pH.
Monitoring pH
Measured Millivolts
Millivolts.
404 mV
Millivolts
7.00 pH A value between the programmed minimum % slope value (see Set Min Slope – see
Section 5.3, page 30) and 105% is displayed. If the value is outside these limits, check
A: Slope & Check the electrode system.
pH check value (zero point).
Displayed as an additional indication of pH electrode system condition; 7 pH is the
optimum value for glass electrodes and 0 ph for Antimony electrodes.
6 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
7.00 pH Sensor B.
Monitoring pH
Measured Temperature
25.6 Deg.C
Sensor A.
404 mV
Measured Millivolts
Sensor A.
-256 mV Sensor B.
Millivolts
7.00 pH A value between the programmed minimum % slope value (see Set Min Slope – see
Section 5.3, page 30) and 105 % is displayed. If the value is outside these limits, check
A: Slope & Check the electrode system.
pH check value (zero point).
Displayed as an additional indication of pH electrode system condition; 7 ph being the
optimum value for glass electrodes and 0 ph for Antimony electrodes.
100.0 %
% Slope and pH Check Value – Sensor B
% slope value.
7.00 pH pH check value (zero point).
See Sensor A above.
IM/AX4PH Rev. P 7
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
25.6Deg.C Temperature.
Redox (ORP)
Offset – Sensor A
Displays the offset value for the sensor set in A: Adjust Offset – see Section 4.1.3,
A: mV Offset
8 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
-256mV Sensor B.
Note. If Probe Type for both Sensor A and Sensor B is set to ORP (Section 5.3), the
Dual Redox
lower display line shows Dual ORP.
Sample Temperature
25.6Deg.C Sensor A.
24.4Deg.C Sensor B.
Note. The measured temperature is displayed only if Temp. Sensor is not set to None –
Temperature see Section 5.3, page 30.
Offset – Sensor A
Displays the offset value for Sensor A set in A: Adjust Offset – see Section 4.1.3,
0mV page 22.
A: mV Offset
Offset – Sensor B
Displays the offset value for Sensor B set in B: Adjust Offset – see Section 4.1.3,
B: mV Offset
IM/AX4PH Rev. P 9
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
-256 mV Sensor B.
Note. The Probe Type for Sensors A and B can be set to any combination of pH, Redox
pH/Redox (ORP)
or ORP – see Section 5.3, page 30. The display indications change depending on
Probe Type settings, e.g. if Sensor A is set to Redox and Sensor B to pH, the lower
display shows Redox (ORP)/pH.
Measured Temperature
25.6 Deg.C Sensor A.
404 mV
Measured Millivolts
Sensor A.
-256 mV Sensor B.
Millivolts
Offset – Sensor B
Displays the offset value for Sensor B set in B: Adjust Offset – see Section 4.1.3,
page 22.
0 mV
B: mV Offset
100.0 %
% Slope and pH Check Value – Sensor A
% slope value.
7.00 pH A value between the programmed minimum % slope value (see Set Min Slope –
A: Slope & Check Section 5.3) and 105 % is displayed. If the value is outside these limits, check the
electrode system.
pH check value (zero value).
Displayed as an additional indication of pH electrode system condition; 7 ph is the
optimum value for glass electrodes and 0 ph for Antimony electrodes.
VIEW SETPOINTS See Section 3.1, page 12.
SENSOR CAL. See Section 4.1, page 19.
Wash Function A3: Type set to Wash (Section 5.5) – see Section 2.3.6, page 11.
pH/Redox (ORP) A3: Type not set to Wash (Section 5.5) – return to top of page.
10 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
Note. The Wash function is available only if A3: Type is set to Wash – see Section 5.5, page 34.
Wash Function
Manual
On Off – Wash function off. Lower display line of Operating Page shows WASH
----- Off
On
INHIBITED.
– Wash function controlled automatically. Lower display line of Operating
Wash Function Page shows WASH IN PROGRESS.
Manual – Enables wash function to be initiated manually – see below.
Note. Set Wash Function to Off before removing the sensor from the process.
VIEW SETPOINTS See Section 3.1, page 12.
Probe Type set to pH (for either sensor if dual input analyzer) – see Section 5.3,
page 30.
SENSOR CAL. See Section 4.1, page 19.
Probe Type set to Redox or ORP (for both sensors in any combination if dual input
analyzer) – see Section 5.3, page 30.
SENSOR CAL. Enable Cals set to Yes (Section 5.3) – see Section 4.1, page 19.
Security Code Alter Sec. Code not set to zero (Section 5.8) – see Section 5.1, page 28.
CONFIG. DISPLAY Alter Sec. Code set to zero (Section 5.8) – see Section 5.2, page 29.
Redox (ORP)
Wash Function not set to Manual. The display returns to the top of the Operating
Dual Redox
Page.
pH/Redox (ORP)
-----
display line.
Press To Wash
Press To Abort
WASH IN PROGRESS Press the key to initiate the wash cycle. The display returns to the top of the
Operating Page and the lower display line shows WASH IN PROGRESS until the wash
cycle is completed. The Wash Function selection reverts to the one that was set before
Manual was selected.
Monitoring pH
Redox (ORP) Press the key to abort the wash cycle. The display returns to the top of the
Dual Redox Operating Page.
pH/Redox (ORP)
IM/AX4PH Rev. P 11
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
3 Operator Views
3.1 View Set Points
Note. The parameter names and units of measurement displayed in the View Set Points page depend on the Probe Type settings
for Sensors A and B – see Section 5.3, page 30. Those shown below are given as examples only.
-----
together with the name of the parameter it is assigned to.
Alarm assignments, set point values and relay/LED actions are programmable – see
VIEW SETPOINTS Section 5.4, page 33.
8.30pH
A1: Setpoint
35.0Deg.C
A2: Setpoint
6.80pH
A3: Setpoint
A4: Setpoint
A5: Setpoint
VIEW SETPOINTS
VIEW OUTPUTS See Section 3.2, page 14.
SENSOR CAL.
Security Code
See Note on next page.
CONFIG. DISPLAY
12 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
Note. The menu displayed when pressing the key from the Operator View pages depends on analyzer configuration, i.e.:
Single Input Analyzers
SENSOR CAL. Probe Type set to pH
or
Probe Type set to Redox or ORP and Enable Cals set to Yes (Section 5.3) – see Section 4.1, page 19.
Security Code Probe Type set to Redox or ORP and Enable Cals set to No (Section 5.3) and Alter Sec. Code not set to zero
(Section 5.8) – see Section 5.1, page 28.
CONFIG. DISPLAY Probe Type set to Redox or ORP and Enable Cals set to No (Section 5.3) and Alter Sec. Code set to zero
(Section 5.8) – see Section 5.2, page 29.
Dual Input Analyzers
SENSOR CAL. Probe Type for either sensor set to pH
or
Probe Type for both sensors set to Redox or ORP and Enable Cals for either sensor set to Yes (Section 5.3) –
see Section 4.1, page 19.
Security Code Probe Type for both sensors set to Redox or ORP and Enable Cals for both sensors set to No (Section 5.3) and
Alter Sec. Code not set to zero (Section 5.8) – see Section 5.1, page 28.
CONFIG. DISPLAY Probe Type for both sensors set to Redox or ORP and Enable Cals for both sensors set to No (Section 5.3) and
Alter Sec. Code set to zero (Section 5.8) – see Section 5.2, page 29.
IM/AX4PH Rev. P 13
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
----- Note. Analog outputs 3 and 4 available only if option board fitted and analog features
enabled – see Section 7.3, page 57.
VIEW OUTPUTS
50.0% Current output shown as a percentage of full scale for the output range set in CONFIG.
OUPUTS – see Section 5.6, page 38.
Analog Output 1
SENSOR CAL.
Security Code
See Note on Page 13.
CONFIG. DISPLAY
Analog Output 2
Advance to analog output 2 (and outputs 3 and 4 if option board fitted and analog
features enabled – see Section 7.3, page 57).
-----
VIEW HARDWARE
Sensor A Module
Shows the type of input board fitted to the analyzer for the Sensor A input.
----- pH
Sensor A Module
----- pH
Sensor B Module
Option Board
Analog
Note. Displayed only if the option board is fitted.
----- Pb Dp Displays the optional features enabled in the Factory Settings page – see Section 7.3,
Option Board page 57.
Security Code
See Note on Page 13.
CONFIG. DISPLAY
14 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
-----
VIEW SOFTWARE
Issue
Shows the version number of the operating software.
1.00
AX400/2000 Issue
VIEW SOFTWARE VIEW LOGBOOK Option board fitted and analog features enabled (Section 7.3) and Logbook set to On
(Section 5.9) – see Section 3.5, page 16.
Monitoring pH
Redox (ORP)
Dual Redox
Operating Page (option board not fitted) – see Section 2.3, page 6.
pH/Redox (ORP)
SENSOR CAL.
IM/AX4PH Rev. P 15
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
Note. The View Logbook function is available only if the option board is fitted and analog features enabled (see Section 7.3,
page 57) and Logbook is set to On (see Section 5.9, page 41).
The logbook stores data entries for alarm events, sensor errors, power failures and pH
calibration information.
-----
VIEW LOGBOOK
----- Alarms Note. If no entries are stored in the Alarms logbook, the display shows No More
Entries.
VIEW LOGBOOK
1 A1
Alarms
The Alarms logbook contains up to 10 entries (entry 1 is the most recent), each
-----On comprising an alarm number, alarm state (On or Off) and the date/time of the
occurrence.
11:09:02 12:34
VIEW CLOCK Option board fitted and analog features enabled (Section 7.3) – see Section 3.6,
page 18.
SENSOR CAL.
Security Code
See Note on Page 13.
CONFIG. DISPLAY
1 Sen.A Errors
The Errors logbook contains up to 5 entries (entry 1 is the most recent), each
-----Pt100 comprising the sensor letter, error number and the date/time of the occurrence.
11:09:02 12:34
VIEW CLOCK Option board fitted and analog features enabled (Section 7.3) – see Section 3.6,
page 18.
SENSOR CAL.
Security Code
See Note on Page 13.
CONFIG. DISPLAY
16 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
View Logbook
Errors
Alarms Use the and keys to access the Power logbook.
Cals
----- Power Note. If no entries are stored in the Power logbook, the display shows No More Entries.
VIEW LOGBOOK
Power
1 The Power logbook contains up to 2 entries (entry 1 is the most recent), each
-----Off comprising the power state (On or Off) and the date/time of the occurrence.
11:09:02 12:34
VIEW CLOCK
Option board fitted and analog features enabled (Section 7.3) – see Section 3.6,
page 18.
SENSOR CAL.
Security Code
See Note on Page 13.
CONFIG. DISPLAY
2 Advance to entry 2.
Note. If no more entries are stored, the display shows No More Entries.
1 Sen.A
Calibration
The Cals logbook contains up to 5 entries (entry 1 is the most recent), each comprising
-----Passed 2 frames. Frame 1 contains the entry number, sensor letter and the calibration pass/fail
indication.
Calibration
Frame 2 contains the % slope value, the pH check value and the date/time of the
100.0 %
occurrence.
7.00 pH
11:09:02 12:34
VIEW CLOCK Option board fitted and analog features enabled (Section 7.3) – see Section 3.6,
page 18.
SENSOR CAL.
IM/AX4PH Rev. P 17
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
Note. The View Clock function is available only if the option board is fitted and analog features enabled – see Section 7.3,
page 57.
-----
VIEW CLOCK
Date
Shows the current date.
-----
Date 01:02:04
Time
Shows the current time.
-----
Time 12:00
Monitoring pH
Redox (ORP)
Operating Page – see Section 2.3, page 6.
Dual Redox
VIEW CLOCK pH/Redox (ORP)
SENSOR CAL.
18 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 4 Setup
4 Setup
4.1 Sensor Calibration
Note. If Probe Type for either sensor (Sensor A only if single input) is set to Redox or ORP, the sensor can be calibrated only if
Enable Cals. for that sensor is set to Yes – see Section 5.3, page 30.
Sensor Calibration
-----
SENSOR CAL.
Sensor Calibration Security Code
Note. This frame is displayed only if Alter Cal. Code is not set to zero – see Section
Sensor Cal. A
Probe Type set to Redox or ORP (for both sensors if dual input analyzer – see Section
5.3, page 30) – continued on page 22.
-----
Set Auto Buffers
Set Buffer 1 Buffer Type not set to User – continued on next page.
A: Enter Point 1 Buffer Type set to User – see Section 4.1.2, page 21.
IM/AX4PH Rev. P 19
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 4 Setup
Set Buffer 1
Set the pH value of the buffer 1 solution – see Appendix A for pH tables.
----- 4.O1pH
Set Buffer 1
Set Buffer 2
Set the pH value of the buffer 2 solution.
-----9.18pH Note. The solution selected for buffer 2 must be at least 2 ph greater than that
selected for buffer 1, e.g. if buffer 1 is set to 7 pH, buffer 2 must be set to at least 9 pH.
Set Buffer 2
20 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 4 Setup
15.0 Deg.C
Solution A: Enter point 1 (to 5)
Deg.C and Adjust are shown alternately on the upper display line. Using the and
0.00 pH keys, adjust the temperature reading (in 5 º increments) to the first of the
temperatures on the pH/temperature curve.
A: Enter point 1
15.0 Deg.C
3.99 pH
pH and Adjust are shown alternately on the center display line. Using the and
keys, adjust the pH reading (in 0.01 ph increments) to the pH reading that corresponds
A: Enter point 1 to the temperature reading entered above.
Notes.
1) For accurate calibration, it is important to repeat the above for buffer solution A at all
35.0 Deg.C 5 points along the pH/temperature curve.
4.02 pH 2) The displayed temperature value increases automatically by 5 ºC from the value set
for the previous point. The setting may be increased but not decreased.
A: Enter point 5
3.99 pH
B: Enter point 1
35.0 Deg.C Note. For accurate calibration, it is important to repeat the above for buffer solution B
at all 5 points along the pH/temperature curve.
4.02 pH The analyzer calculates the pH/temperature relationship from the data entered.
B: Enter point 5
IM/AX4PH Rev. P 21
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 4 Setup
Calibrate Sensor A
-----
Sensor Cal. A
Sensor Cal. B Sensor B (dual input analyzers only) calibration is identical to Sensor A calibration.
SENSOR CAL. Probe Type for Sensor B (dual input analyzers only) set to Redox or ORP and Enable
Cals. set to No (Section 5.3) – return to top of page.
SECURITY CODE Alter Sec. Code not set to zero (Section 5.8) – see Section 5.1, page 28.
CONFIG. DISPLAY Alter Sec. Code set to zero (Section 5.8) – see Section 5.2, page 29.
A: Adjust Offset Probe Type set to Redox or ORP (Section 5.3) – continued below.
A: Buffer Method Probe Type set to pH (Section 5.3) – see Section 4.1.1, page 19.
407mV keys to adjust the upper display line to the required offset value for the process.
The offset value is adjustable between –240 and +240 mV.
A: Adjust Offset
Probe Type for Sensor B (dual input analyzers only) set to Redox or ORP and Enable
SENSOR CAL.
Cals. set to No (Section 5.3) – return to top of page.
Alter Sec. Code not set to zero (Section 5.8) – see Section 5.1, page 28.
SECURITY CODE
Alter Sec. Code set to zero (Section 5.8) – see Section 5.2, page 29.
CONFIG. DISPLAY
22 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 4 Setup
4.00pH
Calibrate Buffer (Single-Point Calibration) or Calibrate Buffer 1(Two-Point Calibration)
Immerse Sensor A in the buffer solution.
25.0Deg.C
A: Immerse Buf.1
Press the key to initiate calibration.
Note. To abort calibration, press the key again at any time before calibration is
complete – see below.
175mV The center display line shows the measured sensor output in millivolts.
#### 100% #### As calibration proceeds, a progress indicator appears in the lower display line. When
the measured sensor output stabilizes, the lower display line shows
##### 100 % #####.
4.01pH
The display then changes for 2 seconds to show the temperature-corrected buffer
value in the upper display line, then advances automatically to the next frame.
A: Cal. Buffer 1
Abort Calibration
Select Yes or No.
-----Yes
A: Abort Cal.
A: SENSOR CAL. Yes selected – return to the main menu.
A: Immerse Buf.2 No selected – calibration continues.
Calibration
Min. Max. Explanation Action
Message
Calibration
40 to 70 % 105 % The new calibration coefficients are accepted None
Passed
Calibration The new calibration coefficients are ignored and the Check buffer values and repeat buffering.
0% 40 to 70 %
Failed last known valid calibration coefficients are used If the fault persists, replace the electrodes
IM/AX4PH Rev. P 23
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 4 Setup
A: Buffer Method
set to Auto 2-Pt
128mV
The center display line shows the measured sensor output in millivolts.
As calibration proceeds, a progress indicator appears in the lower display line. When
#### 100% #### the measured sensor output stabilizes, the lower display line shows
##### 100 % #####.
4.01pH
The display then changes for 2 seconds to show the temperature-corrected buffer
value in the upper display line, then advances automatically to the next frame.
A: Cal. Buffer 2
A: Buffer Method
set to Auto 1-Pt
Calibration Message
See Table 4.1 for details of calibration messages.
-----Passed
A: Calibration
Sensor Cal. A
pH check value.
Displayed as an additional indication of electrode system condition; 7 ph is the
optimum value for glass electrodes and 0 ph for Antimony electrodes.
Sensor Cal. B Sensor B (dual input analyzers only) calibration is identical to Sensor A calibration.
SENSOR CAL. Probe Type for Sensor B (dual input analyzers only) set to Redox or ORP and Enable
Cals. set to No (Section 5.3) – return to top of page.
SECURITY CODE Alter Sec. Code not set to zero (Section 5.8) – see Section 5.1, page 28.
CONFIG. DISPLAY Alter Sec. Code set to zero (Section 5.8) – see Section 5.2, page 29.
24 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 4 Setup
20.0Deg.C and keys to adjust the displayed temperature value to the required buffer
temperature (–20 to150 ºC or –4 to 302 ºF).
A:Set Buf. Temp. Note. If A: Buffer Method is set to Man 2-Pt, the temperature selected is used for both
buffers.
4.00pH
Calibrate Buffer (Single-Point Calibration) or Calibrate Buffer 1(Two-Point Calibration)
Immerse Sensor A in the buffer solution.
25.0Deg.C pH and Adjust are shown alternately on the upper display line. Use the and
keys to set the displayed pH value to the temperature-corrected pH value of the
A: Immerse Buf.1
chosen solution (see the data sheet provided with the solution).
Press the key to initiate calibration.
Note. To abort calibration, press the key again at any time before calibration is
4.01pH
complete – see below.
Measured pH value from the last successful calibration.
Cal. Accepted is shown for 2 seconds to confirm that the calibration has been
A: Cal. Accepted
accepted, the display then advances automatically to the next frame.
4.01pH The display then changes for 2 seconds to show the temperature-corrected buffer
value in the upper display line, then advances automatically to the next frame.
A: Cal. Buffer 1
Abort Calibration
Select Yes or No.
-----Yes
A: Abort Cal.
A: SENSOR CAL. Yes selected – return to the main menu.
A: Immerse Buf.2 No selected – calibration continues.
IM/AX4PH Rev. P 25
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 4 Setup
A: Buffer Method
set to Man 2-Pt
9.22pH
complete – see previous page.
Measured pH value from the last successful calibration.
105mV Measured sensor output in millivolts.
Press button When the measured sensor output stabilizes, press the key to accept the
when stable calibration.
9.22pH
Cal. Accepted is shown for 2 seconds to confirm that the calibration has been
A: Cal. Accepted
accepted, the display then advances automatically to the next frame.
9.22pH The display then changes for 2 seconds to show the temperature-corrected buffer
value in the upper display line, then advances automatically to the next frame.
A: Cal. Buffer 2
A: Buffer Method
set to Man 1-Pt
Calibration Message
See Table 4.1 for details of calibration messages.
-----Passed
A: Calibration
100.0%
Slope Value
% slope value.
7.00pH A value between the programmed minimum % slope value (see Set Min Slope –
Section 5.3) and 105 % is displayed. If the value is outside these limits, check the
A: Slope & Check electrode system.
pH check value.
Sensor Cal. A
Displayed as an additional indication of electrode system condition; 7 ph is the
optimum value for glass electrodes and 0 ph for Antimony electrodes.
Sensor Cal. B Sensor B calibration (dual input analyzers only) is identical to Sensor A calibration.
SENSOR CAL. Probe Type for Sensor B (dual input analyzers only) set to Redox or ORP and Enable
Cals. set to No (Section 5.3) – return to top of page.
SECURITY CODE Alter Sec. Code not set to zero (Section 5.8) – see Section 5.1, page 28.
CONFIG. DISPLAY Alter Sec. Code set to zero (Section 5.8) – see Section 5.2, page 29.
26 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 4 Setup
-----Grab
A: Buffer Method
9.18pH
Adjust Value
ph and Adjust are shown alternately on the upper display line. The displayed pH value
25.0Deg.C is the reading sampled by the analyzer as this frame is selected and is held until the
display is advanced to the next frame. Use the and keys to adjust the
A: Adjust Value displayed value (in 0.01 pH increments) to match the pH value of the measured grab
sample.
Notes.
If the displayed value is adjusted by more than ±3pH, WARNING - OFFSET is
shown on the lower display line. If the measured grab sample value is correct
and the analyzer reading has not been over-adjusted, clean the electrode, check
the sensor connections and try again.
If the displayed value is adjusted by ±5pH, OUT OF RANGE is shown on the
lower display line, indicating that maximum adjustment has been reached.
Further adjustment is not possible.
100.0%
Slope Value
% slope value.
7.00pH The value generated during the last valid two-point calibration, between the
programmed minimum % slope value (see Set Min Slope – Section 5.3) and 105 %, is
A: Slope & Check displayed.
pH check value.
Sensor Cal. A
The value generated during the last valid two-point calibration, adjusted by the value
applied in Adjust Value (above), is displayed.
Note. The pH check value is reset to the previous, valid check value if a single- or
two-point calibration is carried out after a grab calibration.
Sensor Cal. B Sensor B calibration (dual input analyzers only) is identical to Sensor A calibration.
SENSOR CAL. Probe Type for Sensor B (dual input analyzers only) set to Redox or ORP and Enable
Cals. set to No (Section 5.3) – return to top of page.
SECURITY CODE Alter Sec. Code not set to zero (Section 5.8) – see Section 5.1, page 28.
CONFIG. DISPLAY Alter Sec. Code set to zero (Section 5.8) – see Section 5.2, page 29.
IM/AX4PH Rev. P 27
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
5 Programming
5.1 Security Code
Note. This frame is displayed only if Alter Sec. Code is not set to zero – see Section
5.8, page 41.
0000 Enter the required code number (between 0000 and 19999), to gain access to the
configuration pages. If an incorrect value is entered, access to the configuration pages
SECURITY CODE is prevented and the display reverts to the Operating Page – see Section 2.3, page 6.
28 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
-----
CONFIG. DISPLAY
Set Language
Sets the language to be used on all displays.
-----
Set Language
Language Page
Use the and keys to select the required language.
-----
English
Deutsch
Francais
Espanol
Italiano
Set Language
-----
Set Temp. Units
Temperature Units
Use the and keys to select the sample temperature display units.
Deg. F
----- Deg. C
Off
Temp. Units
-----
Set Backlight
Backlight
Use the and keys to select the required backlight option:
Auto.
----- On Auto. – Backlight comes on at each button press and switches off one
minute after the last button press.
LED Backlight On – Backlight is always on.
CONFIG. DISPLAY Return to main menu.
Set Backlight
CONFIG. SENSORS See Section 5.3, page 30.
IM/AX4PH Rev. P 29
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
-----
CONFIG. SENSORS
Configure Sensor A
-----
Config. Sensor A
Config. Sensor B Sensor B configuration (dual input analyzers only) is identical to Sensor A
configuration.
Probe Type
ORP
Redox Select the required probe type.
----- pH ORP
Redox
–
–
Millivolt display
Millivolt display
A: Probe Type
pH – pH display
Differential Input
Yes Yes – Select if electrode system is equipped with a solution earth (ground)
----- No
No –
rod and sensor diagnostics are required – see Section 5.4, page 33.
Select if electrode system is not equipped with a solution earth
A: Diff. Input
(ground) rod or sensor diagnostics are not required.
Note. Ensure the electrode system is connected correctly for the type of input
selected (Standard or Differential) – see Fig. 6.9 (wall-/pipe-mount analyzers) or
Fig. 6.11 (panel-mount analyzers).
A: Probe Type
set to Redox/ORP
None Temperature Sensor
3K Bal
Pt1000 Select the type of temperature sensor used: Pt100, Pt1000, Balco 3K or None.
----- Pt100
A: Temp. Sensor
Enable Calibration
Select Yes to enable sensor calibration.
Yes
----- No Select No to disable sensor calibration and all associated frames.
A: Enable Cals
30 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
A: Probe Type
set to pH
pH Electrode Type
Temperature Compensation
Manual Select Auto to enable the analyzer to compensate automatically for fluctuations in
A: Temp. Comp.
Manual
Auto
IM/AX4PH Rev. P 31
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
Yes Select Yes to enable compensation for solution temperature effects referenced to
----- No 25 ºC.
A: Sample Comp.
No
Yes
Sample Coefficient
If Sample Comp. is set to Yes, enter the temperature coefficient of the sample, in pH/ºC,
-0.035 pH/°C within the range 0.020 to –0.050 (in –0.001 increments). For ammonia and sodium
hydroxide dosed boilers the value is typically –0.035 (this depends on the individual
A: Sample Coeff. boiler chemistry). The exact value for a particular sample needs to be determined by
laboratory analysis.
60.0%
Set the required pH calibration minimum slope value, in %, within the range 60.0 to
90.0 (in 0.1 increments). The calibration fail limit is set automatically to 20 % below the
minimum slope setting – see Table 4.1.
A: Set Min Slope
Config. Sensor A
Config. Sensor B Sensor B (dual input analyzers only) configuration is identical to Sensor A
configuration.
CONFIG. DIAGS see Section 5.4, page 33
32 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
Note. The Configure Diagnostics function is applicable only if Diff. Input for Sensor A and/or Sensor B is set to Yes – see Section
5.3, page 30.
-----
CONFIG. DIAGS
Configure Sensor A
-----
Config. Sensor A
Config. Sensor B Sensor B configuration (dual input analyzers only) is identical to Sensor A
configuration.
pH Glass Check
Set to On to enable the impedance of the glass electrode circuit to be monitored in
Off
----- On
order to provide low glass impedance and out-of-sample/broken cable warnings.
Note. This frame is displayed only if A: Probe Type is set to pH and A: Electrode is set
A: pH Glass to Glass – see Section 5.3, page 30.
Off Set to On to enable the condition of the reference electrode to be monitored in order to
----- On provide:
– an indication of when sensor cleaning or replacement is required
A: Ref. Checking
– out-of-sample/broken cable warnings
Config. Sensor A Config. Sensor B Sensor B (dual input analyzers only) configuration is identical to Sensor A
configuration.
Off
On CONFIG. ALARMS See section 5.5, page 34.
Reference Alarm
Set the impedance value above which the reference alarm is activated.
50kOhm
A: Ref. Alarm
Config. Sensor A
Config. Sensor B Sensor B (dual input analyzers only) configuration is identical to Sensor A
configuration.
CONFIG. ALARMS See section 5.5, page 34.
IM/AX4PH Rev. P 33
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
-----
CONFIG. ALARMS
Configure Alarm 1
-----
Config. Alarm 1
Config. Alarm 2 Alarms 2 and 3 configuration (and Alarms 4 and 5 if option board fitted and analog
features enabled – see Section 7.3, page 57) is identical to Alarm 1 configuration.
Alarm 3 can also be configured as a Wash alarm if A3: Type is set to Wash – see
following frame.
34 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
Alarm 1 Failsafe
Select Yes to enable failsafe action, otherwise select No.
Yes
----- No See also Fig. 5.2 to Fig. 5.6 (page 37).
A1: Failsafe
Alarm 1 Action
High Select the alarm action required, High or Low.
-----Low See also Fig. 5.2 to Fig. 5.6 (page 37).
A1: Action
A-B
mV Alarm 1 Set Point
Deg.C
Deg.F The Alarm 1 Set Point can be set within the following ranges:
100.0 pH pH
mV
–
–
–2.00 to 16.00 pH
–1200 to 1200 mV
A1: Setpoint
Deg.C – –10.0 to 150.0
Deg.F – –14.0 to 302.0
A-B – 0.00 to 14.00 pH
Set to the value required.
Alarm 1 Hysteresis
A differential set point can be defined between 0 and 5 % of the alarm set point value.
0.0
% Set the required hysteresis in 0.1 % increments.
See also Fig. 5.2 to Fig. 5.6 (page 37).
A1: Hysteresis
Alarm 1 Delay
If an alarm condition occurs, the activation of the relays and LEDs can be delayed for a
0Secs specified time period. If the alarm clears within the period, the alarm is not activated.
A1: Delay Set the required delay, in the range 0 to 60 seconds in 1 second increments.
See also Fig. 5.2 to Fig. 5.6 (page 37).
Config. Alarm 2 Alarms 2 and 3 configuration (and Alarms 4 and 5 if option board fitted and analog
Config. Alarm 1 features enabled – see Section 7.3, page 57) is identical to Alarm 1 configuration.
CONFIG. OUTPUTS See Section 5.6, page 38.
IM/AX4PH Rev. P 35
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
Wash Frequency
Wash Duration
Recovery Period
Set the recovery period required, between 0.5 and 5.0 minutes in 0.5 minute
1.0 Mins increments.
Recovery Period
Config. Alarm 4 Option board fitted and analog features enabled (see Section 7.3, page 57) – Alarm 4
Config. Alarm 3
configuration is identical to Alarm 1.
CONFIG. OUTPUTS Option board not fitted or option board fitted and analog features disabled (see
Section 7.3, page 57) – see Section 5.6, page 38.
Frequency
Continuous
t
1s 1s
Pulsed
t
36 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
Note. The following examples illustrate High Alarm Actions, i.e. the alarm is activated when the process variable exceeds the
defined set point. Low Alarm Actions are the same, except the alarm is activated when the process variable drops below the
defined set point.
Relay Energized,
LED Off Relay Energized, LED On
Relay De-energized,
Relay De-energized, LED On LED Off
Fig. 5.2 High Failsafe Alarm without Hysteresis and Delay Fig. 5.5 High Non-Failsafe Alarm without Delay and Hysteresis
Hysteresis
Delay
Relay De-energized, LED On Relay De-energized, LED On
Fig. 5.3 High Failsafe Alarm with Hysteresis but no Delay Fig. 5.6 High Failsafe Alarm with Delay but no Hysteresis
Process Variable
Hysteresis
Relay Energized,
LED Off
Delay
Relay De-energized, LED On
IM/AX4PH Rev. P 37
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
-----
CONFIG. OUTPUTS
Configure Output 1
-----
Config. Output 1
Config. Output 2
Output 2 configuration (and Outputs 3 and 4 if option board fitted and analog features
enabled – see Section 7.3, page 57) is identical to Output 1 configuration.
A-B Assign
Temp.B
Sen.B Select the sensor and analog output required:
Temp.A
----- Sen.A Sen.A
Sen.B
– pH/Redox (ORP)/mV for the selected sensor.
AO1: Assign
Temp.A
– Temperature for the selected sensor.
Temp.B
A-B – Difference between the Sensor A and Sensor B readings.
Note. Sen.B, Temp.B and A-B are applicable only to dual input analyzers.
Range
4-20mA
0-20mA Select the analog output current range for the selected output.
----- 0-10mA
AO1: Range
7.00pH
Span Value
pH (or mV or Deg.C or Deg.F or A-B) and Adjust are shown alternately on the upper
38 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
Default Output
Hold
On Select the system reaction to failure:
----- Off
mA Hold
On
–
–
Hold the analog output at the value prior to the failure.
Stop on failure. This drives the analog output to the level set in
AO1: Default
the Default Val frame below.
Config. Output 1
Off – Ignore failure and continue operation.
Off
or
Hold
On
Default Value
The level to which the analog output is driven if a failure occurs.
12.00mA Set the value between 0.00 and 22.00 mA
AO1: Default Val
Config. Output 2 Output 2 configuration (and Outputs 3 and 4 if option board fitted and analog features
Config. Output 1 enabled – see Section 7.3, page 57) is identical to Output 1 configuration.
CONFIG. CLOCK Option board fitted and analog features enabled (see Section 7.3, page 57) – see
Section 5.7, page 40.
CONFIG. SERIAL
Option board fitted and Serial Communications feature enabled (see Section 7.3,
page 57) – see Supplementary Manual Profibus® Datalink Description (IM/AX4/PBS).
CONFIG. CONTROL
Single input analyzer and option board not fitted – see PID Control Supplement
(IM/AX4PID).
CONFIG. SECURITY
Dual input analyzer and option board not fitted – see Section 5.8, page 41.
IM/AX4PH Rev. P 39
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
Note. The Configure Clock function is available only if the option board is fitted and analog features enabled – see Section 7.3,
page 57.
-----
CONFIG. CLOCK
Set Clock
Set the system clock.
-----
Set Clock?
CONFIG. CLOCK
Return to main menu.
CONFIG. SERIAL
Option board fitted and Serial Communications feature enabled (see Section 7.3,
page 57) – see Supplementary Manual Profibus® Datalink Description (IM/AX4/PBS).
CONFIG. CONTROL
Single input analyzer and option board not fitted – see PID Control Supplement
(IM/AX4PID).
CONFIG. SECURITY Dual input analyzer and option board not fitted – see Section 5.8, page 41.
Date Format
Select the required date format.
-----
mm:dd:yy
Format dd:mm:yy
Date
Set Set the date in the format selected above.
-----Day Press
Use the
to move between the day, month and year fields.
and keys to adjust each field.
Date 01:01:01
Time
Set
Set the time in the format hh:mm.
----- Hours Press to move between hours and minute fields.
Time 12:00 Use the and keys to adjust each field.
----- Press to Set and Press to Abort are shown alternately on the lower display line.
Press To Set
Press To Abort Press the appropriate key to set the clock or abort the changes.
Set Clock?
40 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
-----
CONFIG. SECURITY
Alter Security Code
Set the security code to a value between 0000 and 19999.
0000
Alter Sec. Code
Note. The Configure Logbook function is available only if the option board is fitted and analog features enabled – see Section
7.3, page 57.
-----
CONFIG. LOGBOOK
Configure Logbook
Off Use the and keys to set the logbook On or Off.
-----On If Off is selected, all data entries in the logbook are cleared.
Logbook
IM/AX4PH Rev. P 41
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
-----
TEST/MAINTENANCE
Test Outputs
Displays the output test details for the analog outputs.
----- Note. Outputs 3 and 4 are available only if the option board is fitted and analog
features enabled – see Section 7.3, page 57.
Test Outputs
Test Output 1 frame only is shown; the format of frames for the remaining outputs is
identical.
4.00mA
Test Output 1
The theoretical output current value.
Maintenance
-----
Maintenance
Hold Outputs
Auto. Enables the relay action and analog outputs to be maintained.
On
----- Off Auto. – Changes in relay action and analog outputs are inhibited during sensor
calibration.
Hold Outputs
On – Changes in relay action and analog outputs are inhibited.
Off – Changes in relay action and analog outputs are not inhibited.
Note. The LEDs flash while the analyzer is in 'Hold' mode.
42 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
Hold Outputs
set to Auto.
1Hrs
Automatic Time
If required, set a time period between 1 and 6 hours, in 30 minute increments, for
30Mins which the outputs are held when Hold Outputs is set to Auto.
At the default setting of Off, changes in relay action and analog outputs are inhibited
Automatic Time
during sensor calibration and released automatically at the end of the procedure.
If a time is set, changes in relay action and analog outputs are inhibited during sensor
calibration, but if the calibration is not completed within the set time, the calibration is
aborted, the display returns to the Operating Page and CAL. ABORTED is displayed.
Maintenance
Load/Save Config Continued below.
FACTORY SETTINGS See Section 7.3, page 57.
Load/Save Configuration
Select whether a configuration is to be loaded or saved.
Yes
----- No Note. If No is selected, pressing the key has no effect.
Load/Save Config
----- Save Factory Config. – resets all the parameters in the Configuration Pages to
the Company Standard.
User Config.
Factory Config. Save User Config. – saves the current configuration into memory.
Load User Config. – reads the saved user configuration into memory.
User Config. and Factory Config. are displayed alternately if a User Configuration has
been saved previously. Use the and keys to make the required selection.
-----
Press To Set. Press to Set and Press to Abort are shown alternately on the lower display line.
Press To Abort
Press the appropriate key to load/save the configuration or abort the changes.
TEST/MAINTENANCE
IM/AX4PH Rev. P 43
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
6 Installation
6.1 Siting Requirements
Note.
Mount in a location free from excessive vibration, and
where the temperature and humidity specification will
not be exceeded.
Mount away from harmful vapors and/or dripping
fluids and ensure that it is suitably protected from
direct sunlight, rain, snow and hail.
Where possible, mount the analyzer at eye level to
allow an unrestricted view of the front panel displays
and controls.
Maximum Distance
Sensor
100m (275 ft)
55°C
(131°F)
Max.
–20°C
(–4°F)
Min.
IP65*
44 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
6.2 Mounting
6.2.1 Wall-/Pipe-mount Analyzers
150 (5.9)
175 (6.9)
192 (7.56)
6)
.2
(0
50
6.
Ø
25
61 (23/8) OD
Position 'U' bolts on pipe 1
Vertical or
Drill suitable Horizontal
2 holes Position plate over 'U' bolts 2 Post
1
Mark fixing centers
(see Fig. 6.2)
3
Secure plate
3
Secure instrument to
wall using 4 Secure transmitter to mounting plate
suitable fixings
A Wall-mounting B Pipe-mounting
IM/AX4PH Rev. P 45
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
Dimensions in mm (in.)
25 92 +0.8
0 (3.62 +0.03
0 )
96 (3.78) 137.50 (5.41) (0.98)
91.60 (3.6)
96 (3.78)
92 +0.8
0
Panel Cut-out (3.62 +0.03 )
0
5.40 (0.2)
Loosen the
retaining screw on 2
each panel clamp
Note. The clamp must fit flat on the analyzer casing. The clamp uses a torque limiter, so it is not possible to over-tighten the
securing screws.
46 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
Warning.
The instrument is not fitted with a switch therefore a disconnecting device such as a switch or circuit breaker conforming to
local safety standards must be fitted to the final installation. It must be fitted in close proximity to the instrument within easy
reach of the operator and must be marked clearly as the disconnection device for the instrument.
Remove all power from supply, relay and any powered control circuits and high common mode voltages before accessing
or making any connections.
The power supply earth (ground) must be connected to reduce the effects of RFI interference and ensure the correct
operation of the power supply interference filter.
The power supply earth (ground) must be connected to the earth (ground) stud on the analyzer case – see Fig. 6.8
(wall-/pipe-mount analyzers) or Fig. 6.10 (panel-mount analyzers).
Use cable appropriate for the load currents. The terminals accept cables from 20 to 14 AWG (0.5 to 2.5mm2) UL Category
AVLV2.
The instrument conforms to Mains Power Input Insulation Category III. All other inputs and outputs conform to Category II.
All connections to secondary circuits must have basic insulation.
After installation, there must be no access to live parts, e.g. terminals.
Terminals for external circuits are for use only with equipment with no accessible live parts.
The relay contacts are voltage-free and must be appropriately connected in series with the power supply and the
alarm/control device which they are to actuate. Ensure that the contact rating is not exceeded. Refer also to Section 6.3.1
for relay contact protection details when the relays are to be used for switching loads.
Do not exceed the maximum load specification for the selected analog output range.
The analog output is isolated, therefore the –ve terminal must be connected to earth (ground) if connecting to the isolated
input of another device.
If the instrument is used in a manner not specified by the Company, the protection provided by the equipment may be
impaired.
All equipment connected to the instrument's terminals must comply with local safety standards (IEC 60950, EN61010-1).
Note.
An earthing (grounding) – stud terminal is fitted to the analyzer case for bus-bar earth (ground) connection – see Fig. 6.8
(wall-/pipe-mount analyzers) or Fig. 6.10 (panel-mount analyzers).
Always route signal output/sensor cell cable leads and mains-carrying/relay cables separately, ideally in earthed (grounded)
metal conduit. Use twisted pair output leads or screened cable with the screen connected to the case earth (ground) stud.
Ensure that the cables enter the analyzer through the glands nearest the appropriate screw terminals and are short and
direct. Do not tuck excess cable into the terminal compartment.
Ensure that the IP65 rating is not compromised when using cable glands, conduit fittings and blanking plugs/bungs (M20
holes). The M20 glands accept cable of between 5 and 9mm (0.2 and 0.35 in.) diameter.
IM/AX4PH Rev. P 47
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
Note. For reliable switching the minimum voltage must be greater than 12V and the minimum current greater than 100mA.
Diode
R C
Load Load
External L N External +
AC Supply DC Supply
A AC Applications B DC Applications
48 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
5 Insert the cable gland into the hole in the analyzer case from the outside.
Tighten the gland to a torque of 3.75 Nm (33 lbf. in.)
Note. The cable glands must be tightened to a torque of 3.75 Nm (33 lbf. in.)
IM/AX4PH Rev. P 49
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
Terminal Block C
(Option Board)
Case Earth
(Ground) Stud Terminal Block A Terminal Block B
Note. When refitting the terminal cover plate, tighten the captive screws to a torque of 0.40 Nm (3.5 lbf. in.)
50 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
6.4.2 Connections
Analog Output 3
Analog Output 4
refer to IM/PROBUS
PROFIBUS DP
connections –
Terminal Block B
Relay 4
Relay 5
Temperature Compensator Temperature Compensator
Not used
Not used
Third Lead
Third Lead
NO
C12 NO
NC
C11 NC
Common
Common
C14 —
C16 —
C
C10 C
C13 +
C15 +
C1
C2
C3
C4
C5
C6
C7
C8
C9
TC
TC
** L
** N
** E
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
B16
B15
B14
B13
B12
B11
B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
*
C
NC
NO
C
NC
NO
C
NC
NO
+
—
+
—
Connect supply earth (ground) to stud on case
Connection Details
Line
Relay 2
Analog Output 1
Analog Output 2
100 to 240 V AC
+
–
12 to 30V DC
Terminal Block A
Power Supplies
Note.
Relay 3 can be configured to control the wash facility – see Section 5.4, page 33.
Tighten the terminal screws to a torque of 0.60 Nm (5.3 lbf. in.).
IM/AX4PH Rev. P 51
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
Terminal Block A
Terminal Block C
(Option Board) Terminal Block B
52 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
6.5.2 Connections
Earth (Ground)
Stud on Case
(see Fig. 6.10)
* 250 mA Type T fues (mains AC) or Refer to Section 6.6 for pH Sensor System
2 A Type T fuse (DC) Connection Details
Note.
Relay 3 can be configured to control the wash facility – see Section 5.4, page 33.
Tighten the terminal screws to a torque of 0.60 Nm (5.3 lbf. in.).
IM/AX4PH Rev. P 53
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
B2 B10 Temperature Compensator (if fitted) 3rd Lead Not Used Red Grey
B7 B15 Screen/Shield (if fitted) Not Used Not Used Not Used
B2 B10 Temperature Compensator (if fitted) 3rd Lead Red Link to White
Connect as per
B3 B11 Temperature Compensator (if fitted) White Red function –
refer to
B4 B12 Not Used Not Used Not Used (non-ABB)
B5 B13 Not Used Not Used Not Used sensor system
manual for cable
B6 B14 Reference Electrode Black Black colors
*See Note 2
Note.
1. If the sensor is fitted with a 2-wire PT100, Pt1000 or 3K Balco temperature compensator, link terminals B9 and B10 (and
B1 and B2 if dual input analyzer).
2. Discard large green wire as not required with this transmitter.
3. Redox systems are not temperature compensated so do not have temperature sensors. To remove temperature error
messages, set Temperature Sensor to NONE. If a temperature sensor is used for a separate temperature display, set
Temperature Sensor to the correct type – see Section 5.3, page 30.
54 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
6.6.2 Differential ph Systems Connections – Capable of Providing Sensor Diagnostics (AP200, TBX5)
When connecting one of these pH systems to the AX400 transmitter, ensure the differential input switch for the relevant sensor is ON.
See Section 5.4, page 33 for sensor diagnostics configuration. If diagnostics are not required leave them switched off.
B2 B10 Temperature Compensator (if fitted) 3rd Lead White Link to White
*In normal operation do not connect the heavy green wire. If noisy readings are obtained, connect the wire to the earth stud.
Note.
1. If the sensor is fitted with a 2-wire PT100, Pt1000 or 3K Balco temperature compensator, link terminals B9 and B10 (and
B1 and B2 if dual input analyzer.).
2. Redox systems are not temperature compensated so do not have temperature sensors. To remove temperature error
messages, set Temperature Sensor to NONE. If a temperature sensor is used for a separate temperature display, set
Temperature Sensor to the correct type – see Section 5.3, page 30.
IM/AX4PH Rev. P 55
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
7 Calibration
Note.
The analyzer is calibrated by the Company prior to dispatch and the Factory Settings pages are protected by an access
code.
Routine recalibration is not necessary – high stability components are used in the analyzer's input circuitry and, once
calibrated, the Analog-to-Digital converter chip self-compensates for zero and span drift. It is therefore unlikely that the
calibration will change over time.
Do Not attempt recalibration without first contacting ABB.
Do Not attempt recalibration unless the input board has been replaced or the Factory Calibration tampered with.
Prior to attempting recalibration, test the analyzer's accuracy using suitably calibrated test equipment – see Section 7.1,
page 56 and see Section 7.2, page 56.
Note. Resistance boxes have an inherent residual resistance that may range from a few m up to 1 . This value must be taken
into account when simulating input levels, as should the overall tolerance of the resistors within the boxes.
7.2 Preparation
1. Switch off the supply and disconnect the conductivity cell(s), temperature compensator(s) and current output(s) from the
analyzer's terminal blocks.
2. Sensor A – Fig. 7.1:
a. Link terminals B9 and B10.
b. Connect the millivolt source to terminals B14 (–ve) and B16 (+ve) to simulate the pH or Redox input. Connect the millivolt
source earth to the Case Earth (Ground) Stud – see Fig. 6.8 (wall-/pipe-mount analyzer) or Fig. 6.10 (panel-mount analyzer).
c. Connect the 0 to 10 k decade resistance box to terminals B9 and B11 to simulate the Pt100/Pt1000/Balco 3K.
Sensor B (dual input analyzers only) – Fig. 7.1:
a. Link terminals B1 and B2.
b. Connect the millivolt source to terminals B6 (–ve) and B8 (+ve) to simulate the pH or Redox input. Connect the millivolt source
earth to the Case Earth (Ground) Stud – see Fig. 6.8 or (wall-/pipe-mount analyzer) or Fig. 6.10 (panel-mount analyzer).
c. Connect the 0 to 10 k decade resistance box to terminals B1 and B3 to simulate the Pt100/Pt1000/Balco 3K.
3. Connect the milliammeter to the analog output terminals.
4. Switch on the supply and allow ten minutes for the circuits to stabilize.
5. Select the FACTORY SETTINGS page and carry out Section 7.3.
Terminal link
Sensor A Terminal Numbers B9 B10 B11 B12 B13 B14 B15 B16
Fig. 7.1 Analyzer Terminal Links and Decade Resistance Box Connections
56 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
Use the Sidescroll Key to scroll through the Pages within each Menu
Cal. Output 2 Cal. Output 3 Cal. Output 4 Option Board Alter Fact.Code
Key
Available only if the Option Board is fitted
and Analog features enabled see page 61
IM/AX4PH Rev. P 57
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
-----
FACTORY SETTINGS
Factory Settings Access Code
Enter the required code number (between 0000 and 19999) to gain access to the
Calibrate Sensor A
Note. The values in the display lines for sensor calibration are shown only as
Cal. Sensor A
Cal. Sensor B Sensor B calibration (dual input analyzers only) is identical to Sensor A calibration.
Monitoring Redox
1000 mV
Millivolt Span
Set the millivolt source to +1000 mV.
1000 Calib The display advances automatically to the next step once a stable and valid value is
recorded.
A:mV Span (+1V)
100.0 Ohms
Temperature Zero (100R)
Set the temperature simulator to 100
100.0 Calib The display advances automatically to the next step once a stable and valid value is
recorded.
A:T.Zero (100R)
58 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
150.0 Ohms
Temperature Span (150R)
Set the temperature simulator to 150
150.0 Calib The display advances automatically to the next step once a stable and valid value is
recorded.
A:T.Span (150R)
1000 Ohms
Temperature Zero (1k0)
Set the temperature simulator to 1000
1000 Calib The display advances automatically to the next step once a stable and valid value is
A:T.Zero (1K0) recorded.
1500
Temperature Span (1k5)
Ohms Set the temperature simulator to 1500
1500 Calib The display advances automatically to the next step once a stable and valid value is
recorded.
A:T.Span (1K5)
2000 Ohms
Temperature Zero (2k0)
Set the temperature simulator to 2000
2000 Calib The display advances automatically to the next step once a stable and valid value is
recorded.
A:T.Zero (2K0)
Abort Calibration
No Select Yes or No
----- Yes
A: Abort Cal.
Yes selected:
A:T.Zero (100R) – before completion of A:mV Span (+1V) frame – calibration advances to A:T.Zero
(100R) and continues.
CAL. SENSOR A – after completion of A:mV Span (+1V) frame – the display returns to the Calibrate
Sensor A page.
No selected – calibration continues from the point at which the key was
pressed.
IM/AX4PH Rev. P 59
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
Calibrate Output 1
Note. When adjusting the 4 and 20 mA outputs, the display reading is unimportant
-----
and is used only to indicate that the output is changing when the and keys
are pressed.
Cal. Output 1
Adjust 4mA
Use the and keys to set the milliammeter reading to 4 mA.
16000 Note. The analog output range selected in Configure Outputs (Section 5.6) does not
affect the reading.
O1: Adjust 4mA
Adjust 20mA
Use the and keys to set the milliammeter reading to 20 mA.
7200 Note. The analog output range selected in Configure Outputs (Section 5.6) does not
affect the reading.
O1: Adjust 20mA
Redox (ORP)
Calibrate Output 2
Note. Output 2 calibration is identical to Output 1 calibration.
-----
Cal. Output 2
Cal. Output 3 Option board fitted and analog features enabled – continued on next page.
Option Board Option board fitted, additional features disabled – continued on next page.
Alter Fact. Code Option board not fitted – continued on next page.
7200
O1: Adjust 20mA
Cal. Output 3 Option board fitted and analog features enabled – continued on next page.
Cal. Output 2
Option Board Option board fitted, additional features disabled – continued on next page.
Alter Fact. Code Option board not fitted – continued on next page.
Monitoring pH
Redox (ORP)
60 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
Option Board
Fitted and Analog
Features Enabled
Calibrate Output 3
Note.
----- Output 3 (and Output 4) calibration is applicable only if the Option board is fitted
and analog features enabled – see below.
Cal. Output 3
Output 3 calibration is identical to Output 2 calibration.
Calibrate Output 4
Note. Output 4 calibration is identical to Output 3 calibration.
-----
Cal. Output 4
Option Board
Fitted, Additional
Features Disabled
Configure Option Board
Analog
Note.
0000
Alter Fact. Code
Monitoring pH
Redox (ORP)
Dual Redox
Operating Page – see Section 2.3, page 6.
pH/Redox (ORP)
IM/AX4PH Rev. P 61
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 8 Simple Fault Finding
Although the calibration has not failed, the Note. A normal laboratory mV source is not suitable
A: CAL LOW SLOPE electrode pair associated with the sensor for use as a pH simulator.
B: CAL LOW SLOPE indicated is becoming fatigued and
replacement is recommended.
Failure to respond to the input indicates a fault with the
The calibration of the sensor indicated has analyzer which must be returned to the Company for
A: PH CAL FAILED failed. Check buffer values and repeat repair. Correct response, but with incorrect readings,
B: PH CAL FAILED buffering. If the fault persists, replace the usually indicates a calibration problem. Recalibrate the
electrodes. analyzer as detailed in Section 7.
Wash Function is set to Off. Set Wash 2. Use the pH simulator to carry out an impedance check on
WASH INHIBITED
Function to On – see Section 2.3.3, page 8. the analyzer, i.e. glass to reference, glass to earth and
A: OUT OF SAMPLE 1. The sensor indicated is not fully immersed reference to earth – refer to simulator manual.
A: BROKEN CABLE in sample. If the analyzer fails this test, check for moisture within the
(alternating display) transmitter and in particular the terminal compartment. It is
2.The cable associated with the sensor
B: OUT OF SAMPLE
indicated may be damaged. vital that all evidence of moisture is removed with the use
B: BROKEN CABLE
(alternating display)
of a hot air drier.
3. Reconnect the electrode cable and connect the simulator
A: BROKEN CABLE The cable associated with the sensor
B: BROKEN CABLE indicated may be damaged.
to the electrode end of the cable. Repeat the procedures
1) and 2) above. If the analyzer fails test 2), check for
A: LOW GLASS IMP.
1. The glass electrode associated with the moisture around the connections and check that the
sensor indicated may be broken. insulation on the inner co-axial conductor is clean and that
A: BROKEN CABLE
(alternating display) 2.The cable associated with the sensor the graphite layer has been removed.
B: LOW GLASS IMP. indicated may be damaged.
B: BROKEN CABLE
3.The connections associated with the
(alternating display)
sensor indicated may be faulty.
62 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 8 Simple Fault Finding
IM/AX4PH Rev. P 63
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 9 Specification
9 Specification
pH/Redox (ORP) – AX460 and AX466 Conductivity – AX416 Only
Inputs Range
One or two* pH or mV inputs and solution earth Programmable 0 to 0.5 to 0 to 10,000 µS/cm
One or two* temperature sensors (with various cell constants)
Enables connection to glass or enamel pH and reference sensors and Minimum span
Redox (ORP) sensors 10 x cell constant
*AX466 only Maximum span
Input resistance 10,000 x cell constant
Glass >1 x 1013 Units of measure
Reference 1 x 10
13 µS/cm, µS/m, mS/cm, mS/m, M-cm and TDS
Range Accuracy
–2 to 16 ph or –1200 to +1200 mV Better than ±0.01 % of span (0 to 100 µS/cm)
64 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 9 Specification
IM/AX4PH Rev. P 65
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 9 Specification
Languages
Languages configurable:
English
French
German
Italian
Spanish
DS/AX4PH–EN Rev. K
66 IM/AX4PH Rev. P
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 Appendix A – Buffer Solutions
The pH value of buffer solutions is influenced considerably by ºC ºF 1.68 pH 4.01 pH 6.86 pH 9.18 pH
temperature variations. Thus, when significant temperature 0 32 1.666 4.003 6.984 9.464
fluctuations occur, it is general practice to correct automatically 5 41 1.668 3.999 6.951 9.395
the measured, prevailing pH to the value that would apply if the
10 50 1.670 3.998 6.923 9.332
solution temperature were 25 ºC (77 ºF), the internationally
accepted standard. 15 59 1.672 3.999 6.900 9.276
Tables A1 to A5 include the pH values for ABB, DIN, Merck, 20 68 1.675 4.002 6.881 9.225
NIST, and US Technical buffer solutions. Standards are for 4, 7 25 77 1.679 4.008 6.865 9.180
and 9 pH values, from 0 to 95 ºC (32 to 203 ºF). 30 86 1.683 4.015 6.853 9.139
IM/AX4PH Rev. P 67
Single and dual input analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 Appendix A – Buffer Solutions
Table A.4 NIST Buffer Solutions Table A.5 US Technical Buffer Solutions
68 IM/AX4PH Rev. P
Acknowledgments
PROFIBUS is a registered trademark of PROFIBUS and
PROFINET International (PI).
Sales Service
—
ABB Limited
Measurement & Analytics
Oldends Lane
Stonehouse
Gloucestershire
GL10 3TA
UK
Tel: +44 (0)1453 826 661
Fax: +44 (0)1453 829 671
Email: [email protected]
ABB Inc.
Measurement & Analytics
125 E County Line Road
Warminster
PA 18974
USA
Tel: +1 215 674 6000
Fax: +1 215 674 7183
abb.com/analytical
—
We reserve the right to make technical changes or modify the contents of this document
12.2018
without prior notice. With regard to purchase orders, the agreed particulars shall prevail.
ABB does not accept any responsibility whatsoever for potential errors or possible lack of
information in this document.
IM/AX4PH Rev. P
We reserve all rights in this document and in the subject matter and illustrations contained
therein. Any reproduction, disclosure to third parties or utilization of its contents – in
whole or in parts – is forbidden without prior written consent of ABB.
© ABB 2018