FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED - CC-3245 - VT 17.05.2023 - As On 14.07.2021

FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED is accredited under ISO/IEC 17025:2017 with certificate number CC-3245, valid from 18/05/2021 to 17/05/2023. The document details various calibration methods and measurement capabilities for mechanical instruments, including tachometers, pressure gauges, and weights, specifying measurement ranges and uncertainties. It serves as an annexure to the Certificate of Accreditation and does not require a signature.

Uploaded by

Anirban Das
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
17 views22 pages

FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED - CC-3245 - VT 17.05.2023 - As On 14.07.2021

FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED is accredited under ISO/IEC 17025:2017 with certificate number CC-3245, valid from 18/05/2021 to 17/05/2023. The document details various calibration methods and measurement capabilities for mechanical instruments, including tachometers, pressure gauges, and weights, specifying measurement ranges and uncertainties. It serves as an annexure to the Certificate of Accreditation and does not require a signature.

Uploaded by

Anirban Das
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 22

FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO.

77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 1 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Permanent Facility
Digital Tachometer
MECHANICAL- by Direct Method
Digital Tachometer 300 RPM to 14000
1 ACCELERATION with Reference to 0.29%rdg
(Non-Contact Type) RPM
AND SPEED SANAS TR 45 (Part
01)
Hydraulic Pressure Pressure Calibrator,
Gauges, Pressure Hydraulic (Pressure
MECHANICAL-
Transmitter, Comparator) & DMM
PRESSURE
2 Pressure Transducer (6.5 digit) by 35 bar to 700 bar 0.18bar
INDICATING
& Pressure comparison Method
DEVICES
Indicators using as per DKD-R-6 (Part
water medium 1)
Pneumatic (Air
Pressure Calibrator ,
Medium) Pressure
MECHANICAL- Pneumatic (Hand
Gauges, Pressure
PRESSURE Pump) & DMM (6.5
3 Transmitter, 1 bar to 20 bar 0.007bar
INDICATING digit) by comparison
Pressure Transducer
DEVICES Method as per DKD-
& Pressure
R-6 (Part 1)
Indicators
Pressure Calibrator,
MECHANICAL- Pneumatic (Hand
PRESSURE Vacuum Gauges & Pump) & DMM (6.5
4 -0.85 bar to 0 bar 0.006bar
INDICATING Transmitters digit) by comparison
DEVICES Method as per DKD-
R-6 (Part 1)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 2 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Electronic
Weighing Balance
with a Readability of
MECHANICAL- VOLUME
5 0.01 mg and >100 µl to 1000 µl 0.35 µl
VOLUME Micropipettes
procedure as per
ISO 8655 (Part 6)
standards
Using Electronic
Weighing Balance
MECHANICAL- VOLUME Pipettes & with a Readability of
6 1 ml to 50 ml 0.05 ml
VOLUME Burettes 0.01 mg and
procedure as per
ISO 4787 standards.
Using Electronic
Weighing Balance
VOLUME Volumetric with a Readability of
MECHANICAL-
7 Flasks & Graduated 0.01 mg / 0.1 mg 1 ml to 100 ml 0.33 ml
VOLUME
measuring Cylinders and procedure as
per ISO 4787
standards.
Using Electronic
Weighing Balance
MECHANICAL- VOLUME with a Readability of
8 10 µl to 100 µl 0.17 µl
VOLUME Micropipettes 0.01 mg. Procedure
based on ISO 8655
(Part 6) standards.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 3 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
9 Balance with a 1g 0.05 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.
Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
10 Balance with a 1 mg 0.02 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.
Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
11 Balance with a 10 g 0.03 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 4 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
12 Balance with a 10 mg 0.03 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.
Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
13 Balance with a 100 g 0.18 mg
WEIGHTS F2 and coarser)
readability of 0.1
mg. Procedure
based on OIML R
111 guidelines.
Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
14 Balance with a 100 mg 0.05 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 5 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
15 Balance with a 2g 0.02 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.
Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
16 Balance with a 2 mg 0.02 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.
Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
17 Balance with a 20 g 0.08 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 6 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
18 Balance with a 20 mg 0.02 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.
Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
19 Balance with a 200 g 0.65 mg
WEIGHTS F2 and coarser)
readability of 0.1
mg. Procedure
based on OIML R
111 guidelines.
Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
20 Balance with a 200 mg 0.06 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 7 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
21 Balance with a 5g 0.11 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.
Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
22 Balance with a 5 mg 0.02 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.
Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
23 Balance with a 50 g 0.08 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 8 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
24 Balance with a 50 mg 0.04 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.
Using E2 Accuracy
Class Standard
Weights and
Electronic Weighing
MECHANICAL- MASS Weights (Class
25 Balance with a 500 mg 0.04 mg
WEIGHTS F2 and coarser)
readability of 0.01
mg. Procedure
based on OIML R
111 guidelines.
MEDICAL
By comparison
DEVICES- 0 cmH2O to 110
Method using Gas
26 DISCHARGE BIPAP - EPAP / IPAP 6.3%rdg
flow Analyzer as per cmH2O
EQUIPMENT/DE
IEC 62353
VICES
MEDICAL
DEVICES- Comparison Method
BP Apparatus / 30 mmHg to 400
27 DISCHARGE by using Vital / 6.9%rdg
Sphygmomanometer mmHg
EQUIPMENT/DE Pressure Simulator
VICES

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 9 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MEDICAL
By comparison
DEVICES-
Infusion Pump - Flow Method using Pump
28 DISCHARGE 10 ml/h to 100 ml/h 7.43%rdg
rate Analyzer as per IEC
EQUIPMENT/DE
62353
VICES
MEDICAL
By comparison
DEVICES-
Infusion Pump - Method using Pump 10 mmHg to 1500
29 DISCHARGE 1.34%rdg
Occlusion Analyzer as per IEC mmHg
EQUIPMENT/DE
62353
VICES
MEDICAL
By comparison
DEVICES-
Infusion Pump - Method using Pump
30 DISCHARGE 1 ml to 50 ml 15.95%rdg
Volume Analyzer as per IEC
EQUIPMENT/DE
62353
VICES
MEDICAL
By comparison
DEVICES-
Method using Gas
31 DISCHARGE Nebulizer - Flow Up to 195 lpm 4.69%rdg
flow Analyzer as per
EQUIPMENT/DE
IEC 62353
VICES
MEDICAL
By comparison
DEVICES-
Syringe Pump - Flow Method using Pump
32 DISCHARGE 10 ml/h to 100 ml/h 7.43%rdg
Rate Analyzer as per IEC
EQUIPMENT/DE
62353
VICES
MEDICAL
By comparison
DEVICES-
Syringe Pump - Method using Pump 10 mmHg to 1500
33 DISCHARGE 1.34%rdg
Occlusion Analyzer as per IEC mmHg
EQUIPMENT/DE
62353
VICES

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 10 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MEDICAL
By comparison
DEVICES-
Syringe Pump - Method using Pump
34 DISCHARGE 1 ml to 50 ml 15.95%rdg
Volume Analyzer and as per
EQUIPMENT/DE
IEC 62353
VICES
MEDICAL Simulation Method -
DEVICES- ECG Machine - Using Vital Sign 12.6 %rdg to 2.4
35 0.5 mV to 5 mV
IMAGING/PLOT Amplitude Simulator as per IEC %rdg
TERS 62353
MEDICAL Simulation Method -
DEVICES- ECG Machine - Heart Using Vital Sign
36 60 bpm to 300 bpm 3.29%rdg
IMAGING/PLOT Rate Simulator as per IEC
TERS 62353
Heart Rate (Patient
MEDICAL Monitor / Multipara
Simulation Method -
DEVICES- Monitor), Pulse Rate
37 Using Vital Sign 60 bpm to 300 bpm 7.93%rdg
MONITORING (NIBP Apparatus) &
Simulator
UNIT Pulse Rate (Pulse
Oximeter)
MEDICAL Simulation Method -
Patient Monitor /
DEVICES- Using Vital Sign
38 Multipara Monitor 60 bpm to 180 bpm 6.11%rdg
MONITORING Simulator as per IEC
(Respiration Rate)
UNIT 62353
Patient Monitor /
MEDICAL Multipara Monitor, Simulation Method -
DEVICES- NIBP Apparatus - Using Vital Sign 0 mmHg to 300
39 9.35%rdg
MONITORING NIBP (Dynamic) Simulator as per IEC mmHg
UNIT Manometer (Systolic 62353
& Diastolic)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 11 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MEDICAL Simulation Method -


DEVICES- SPO2 / Pulse Using Vital Sign
40 55 % to 100 % 2.45%rdg
MONITORING Oximeter Simulator as per IEC
UNIT 62353
MEDICAL
DEVICES- Simulation Method -
Defibrillator - Charge
PATIENT Using Defibrillator
41 Time; Less than 15 sec 2.18% rdg
CONDITIONING Analyzer as per IEC
Synchronization test
/ 62353:
MAINTENANCE
MEDICAL
DEVICES- By Comparison
PATIENT Defibrillator - Energy Method using 20 Joules to 300
42 7.8%rdg
CONDITIONING Output Defibrillator Joules
/ Analyzer
MAINTENANCE
MEDICAL
DEVICES- Simulation Method -
PATIENT Defibrillator - Heart Using Defibrillator
43 30 bpm to 300 bpm 3.42%rdg
CONDITIONING Rate Analyzer as per IEC
/ 62353
MAINTENANCE
MEDICAL
By comparison
DEVICES- Diathermy /
Method using
PATIENT Electrosurgical Unit /
44 Electrosurgical 0 W to 300 W 13.95%rdg
CONDITIONING Cautery Machine -
Analyzer as per IEC
/ Output Power
62353
MAINTENANCE

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 12 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MEDICAL
DEVICES- Comparison Method
PATIENT Tourniquet - - Using Vital Sign 0 mmHg to 400
45 3%rdg
CONDITIONING Pressure Simulator, Stop mmHg
/ Watch as per 62353
MAINTENANCE
MEDICAL
Comparison Method
DEVICES-
- Using Vital Sign
PATIENT Tourniquet - Time
46 Simulator, Stop Up to 30 min 0.19%rdg
CONDITIONING Interval
Watch as per IEC
/
62353
MAINTENANCE
MEDICAL
DEVICES- By comparison
PATIENT Method using Gas
47 Ventilator - I:E Ratio 0.25 s to 5 s 2.31%rdg
CONDITIONING flow Analyzer as per
/ IEC 62353
MAINTENANCE
MEDICAL
DEVICES- By comparison
PATIENT Ventilator - Oxygen Method using Gas
48 21 % to 100 % 2.34%rdg
CONDITIONING Accuracy flow Analyzer as per
/ IEC 62353
MAINTENANCE
MEDICAL
DEVICES- By comparison
PATIENT Ventilator - Pressure Method using Gas 0 cmH2O to 110
49 3.8%rdg
CONDITIONING Accuracy / PEEP flow Analyzer as per cmH2O
/ IEC 62353
MAINTENANCE

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 13 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MEDICAL
DEVICES- By comparison
PATIENT Ventilator - Method using Gas
50 6 bpm to 150 bpm 2.1%rdg
CONDITIONING Respiration Rate flow Analyzer as per
/ IEC 62353
MAINTENANCE
MEDICAL
DEVICES- By comparison
PATIENT Ventilator - Volume Method using Gas
51 Up to 3 l 4.33%rdg
CONDITIONING Accuracy flow Analyzer as per
/ IEC 62353
MAINTENANCE
Using Standard
RH / Temperature
THERMAL- Temp. / RH Indicator
Indicator with 18 %RH to 80
52 SPECIFIC HEAT with sensor by 1.82%RH
Sensor / Thermo %RH @25°C
& HUMIDITY Comparison Method
Hygrometer
and RH chamber
RH / Temperature Using RTD / DMM by
THERMAL-
Indicator with Comparison Method 10 °C to 50 °C @ 50
53 SPECIFIC HEAT 0.7°C
Sensor / Thermo and Low Temp. % RH
& HUMIDITY
Hygrometer Chamber
RTD, Thermocouple Using Standard RTD
THERMAL- with or Without (PT 100) / DMM, Low
54 >25 °C to 50 °C 0.3°C
TEMPERATURE Indicator, Bimetallic Temperature Liquid
Thermometers Bath
RTD, Thermocouple Using Standard RTD
THERMAL- with or Without (PT 100) / DMM, Low
55 -80 °C to 25 °C 0.28°C
TEMPERATURE Indicator, Bimetallic Temperature Liquid
Thermometers Bath

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 14 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

RTD, Thermocouple
with or Without Using Standard RTD
THERMAL-
56 Indicator, Bimetallic (PT 100) / DMM, Dry >50 °C to 400 °C 0.9°C
TEMPERATURE
Thermometers, Block Calibrator
Temperature Switch

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 15 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Site Facility
Digital Tachometer
MECHANICAL- Centrifuge, RPM by Direct Method
>22000 RPM to
1 ACCELERATION Indicators (Non- with Reference to 0.06%rdg
75000 RPM
AND SPEED Contact Type) SANAS TR 45 (Part
01)
Digital Tachometer
MECHANICAL- Centrifuge, RPM by Direct Method
100 RPM to 22000
2 ACCELERATION Indicators (Non- with Reference to 0.64% rdg
RPM
AND SPEED Contact Type) SANAS TR 45 (Part
01)
Hydraulic Pressure Pressure Calibrator,
Gauges, Pressure Hydraulic (Pressure
MECHANICAL-
Transmitter, Comparator) & DMM
PRESSURE
3 Pressure Transducer (6.5 digit) by 35 bar to 700 bar 0.18bar
INDICATING
& Pressure comparison Method
DEVICES
Indicators using as per DKD-R-6 (Part
water medium 1)
Pneumatic (Air
Pressure Calibrator ,
Medium) Pressure
MECHANICAL- Pneumatic (Hand
Gauges, Pressure
PRESSURE Pump) & DMM (6.5
4 Transmitter, 1 bar to 20 bar 0.007bar
INDICATING digit) by comparison
Pressure Transducer
DEVICES Method as per DKD-
& Pressure
R-6 (Part 1)
Indicators

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 16 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Pressure Calibrator,
MECHANICAL- Pneumatic (Hand
PRESSURE Vacuum Gauges & Pump) & DMM (6.5
5 -0.85 bar to 0 bar 0.006bar
INDICATING Transmitters digit) by comparison
DEVICES Method as per DKD-
R-6 (Part 1)
Using M1 Accuracy
Electronic Weighing
MECHANICAL- Class Standard
Balances
WEIGHING Weights and the
6 Readability: 0.1 kg >15 kg to 200 kg 58.0 g
SCALE AND procedure as per
(Class IV and
BALANCE OIML R-47
coarser)
guidelines.
Using E2 Accuracy
MECHANICAL- Electronic Weighing Class Standard
WEIGHING Balances Weights and the
7 1 mg to 200 g 0.15 mg
SCALE AND Readability: 0.1 mg procedure as per
BALANCE (Class I and coarser) OIML R-76 (Part 1)
guidelines.
Using M1 Accuracy
Electronic Weighing
MECHANICAL- Class Standard
Balances
WEIGHING Weights and the
8 Readability: 0.5 g >200 g to 15 kg 0.35 g
SCALE AND procedure as per
(Class III and
BALANCE OIML R-76 (Part 1)
coarser)
guidelines.
MEDICAL
By comparison
DEVICES- 0 cmH2O to 110
Method using Gas
9 DISCHARGE BIPAP - EPAP / IPAP 6.3%rdg
flow Analyzer as per cmH2O
EQUIPMENT/DE
IEC 62353
VICES

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 17 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MEDICAL
DEVICES- Comparison Method
BP Apparatus / 30 mmHg to 400
10 DISCHARGE by using Vital / 6.9%rdg
Sphygmomanometer mmHg
EQUIPMENT/DE Pressure Simulator
VICES
MEDICAL
By comparison
DEVICES-
Infusion Pump - Flow Method using Pump
11 DISCHARGE 10 ml/h to 100 ml/h 7.43%rdg
rate Analyzer as per IEC
EQUIPMENT/DE
62353
VICES
MEDICAL
By comparison
DEVICES-
Infusion Pump - Method using Pump 10 mmHg to 1500
12 DISCHARGE 1.34%rdg
Occlusion Analyzer as per IEC mmHg
EQUIPMENT/DE
62353
VICES
MEDICAL
By comparison
DEVICES-
Infusion Pump - Method using Pump
13 DISCHARGE 1 ml to 50 ml 15.95%rdg
Volume Analyzer as per IEC
EQUIPMENT/DE
62353
VICES
MEDICAL
By comparison
DEVICES-
Method using Gas
14 DISCHARGE Nebulizer - Flow Up to 195 lpm 4.69%rdg
flow Analyzer as per
EQUIPMENT/DE
IEC 62353
VICES
MEDICAL
By comparison
DEVICES-
Syringe Pump - Flow Method using Pump
15 DISCHARGE 10 ml/h to 100 ml/h 7.43%rdg
Rate Analyzer as per IEC
EQUIPMENT/DE
62353
VICES

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 18 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MEDICAL
By comparison
DEVICES-
Syringe Pump - Method using Pump 10 mmHg to 1500
16 DISCHARGE 1.34%rdg
Occlusion Analyzer as per IEC mmHg
EQUIPMENT/DE
62353
VICES
MEDICAL
By comparison
DEVICES-
Syringe Pump - Method using Pump
17 DISCHARGE 1 ml to 50 ml 15.95%rdg
Volume Analyzer and as per
EQUIPMENT/DE
IEC 62353
VICES
MEDICAL Simulation Method -
DEVICES- ECG Machine - Using Vital Sign 12.6 %rdg to 2.4
18 0.5 mV to 5 mV
IMAGING/PLOT Amplitude Simulator as per IEC %rdg
TERS 62353
MEDICAL Simulation Method -
DEVICES- ECG Machine - Heart Using Vital Sign
19 60 bpm to 300 bpm 3.29%rdg
IMAGING/PLOT Rate Simulator as per IEC
TERS 62353
MEDICAL Trans Illuminator
By comparison
DEVICES- Light Source / OT
20 Method using Lux 100 lux to 32000 lux 9.27%rdg
IMAGING/PLOT Light / SPOT Light /
meter
TERS Examination Light
Heart Rate (Patient
MEDICAL Monitor / Multipara
Simulation Method -
DEVICES- Monitor), Pulse Rate
21 Using Vital Sign 60 bpm to 300 bpm 7.93%rdg
MONITORING (NIBP Apparatus) &
Simulator
UNIT Pulse Rate (Pulse
Oximeter)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 19 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MEDICAL Simulation Method -


Patient Monitor /
DEVICES- Using Vital Sign
22 Multipara Monitor 60 bpm to 180 bpm 6.11%rdg
MONITORING Simulator as per IEC
(Respiration Rate)
UNIT 62353
Patient Monitor /
MEDICAL Multipara Monitor, Simulation Method -
DEVICES- NIBP Apparatus - Using Vital Sign 0 mmHg to 300
23 9.35%rdg
MONITORING NIBP (Dynamic) Simulator as per IEC mmHg
UNIT Manometer (Systolic 62353
& Diastolic)
MEDICAL Simulation Method -
DEVICES- SPO2 / Pulse Using Vital Sign
24 55 % to 100 % 2.45%rdg
MONITORING Oximeter Simulator as per IEC
UNIT 62353
MEDICAL
DEVICES- Simulation Method -
Defibrillator - Charge
PATIENT Using Defibrillator
25 Time; Less than 15 sec 2.18% rdg
CONDITIONING Analyzer as per IEC
Synchronization test
/ 62353:
MAINTENANCE
MEDICAL
DEVICES- By Comparison
PATIENT Defibrillator - Energy Method using 20 Joules to 300
26 7.8%rdg
CONDITIONING Output Defibrillator Joules
/ Analyzer
MAINTENANCE

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 20 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MEDICAL
DEVICES- Simulation Method -
PATIENT Defibrillator - Heart Using Defibrillator
27 30 bpm to 300 bpm 3.42%rdg
CONDITIONING Rate Analyzer as per IEC
/ 62353
MAINTENANCE
MEDICAL
By comparison
DEVICES- Diathermy /
Method using
PATIENT Electrosurgical Unit /
28 Electrosurgical 0 W to 300 W 13.95%rdg
CONDITIONING Cautery Machine -
Analyzer as per IEC
/ Output Power
62353
MAINTENANCE
MEDICAL
DEVICES- Comparison Method
PATIENT Tourniquet - - Using Vital Sign 0 mmHg to 400
29 3%rdg
CONDITIONING Pressure Simulator, Stop mmHg
/ Watch as per 62353
MAINTENANCE
MEDICAL
Comparison Method
DEVICES-
- Using Vital Sign
PATIENT Tourniquet - Time
30 Simulator, Stop Up to 30 min 0.19%rdg
CONDITIONING Interval
Watch as per IEC
/
62353
MAINTENANCE
MEDICAL
DEVICES- By comparison
PATIENT Method using Gas
31 Ventilator - I:E Ratio 0.25 s to 5 s 2.31%rdg
CONDITIONING flow Analyzer as per
/ IEC 62353
MAINTENANCE

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 21 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MEDICAL
DEVICES- By comparison
PATIENT Ventilator - Oxygen Method using Gas
32 21 % to 100 % 2.34%rdg
CONDITIONING Accuracy flow Analyzer as per
/ IEC 62353
MAINTENANCE
MEDICAL
DEVICES- By comparison
PATIENT Ventilator - Pressure Method using Gas 0 cmH2O to 110
33 3.8%rdg
CONDITIONING Accuracy / PEEP flow Analyzer as per cmH2O
/ IEC 62353
MAINTENANCE
MEDICAL
DEVICES- By comparison
PATIENT Ventilator - Method using Gas
34 6 bpm to 150 bpm 2.1%rdg
CONDITIONING Respiration Rate flow Analyzer as per
/ IEC 62353
MAINTENANCE
MEDICAL
DEVICES- By comparison
PATIENT Ventilator - Volume Method using Gas
35 Up to 3 l 4.33%rdg
CONDITIONING Accuracy flow Analyzer as per
/ IEC 62353
MAINTENANCE
Using Standard
indicator of RH and Temp. / RH Indicator
THERMAL-
Temperature with sensor by 10 °C to 50 °C @ 50
36 SPECIFIC HEAT 0.7°C
chamber / Climatic Comparison Method % RH
& HUMIDITY
Chamber at single specified
position

This is annexure to 'Certificate of Accreditation' and does not require any signature.
FABER SINDOORI MANAGEMENT SERVICES PRIVATE LIMITED, NO. 77, POTTIPATI
Laboratory Name :
PLAZA, 4TH FLOOR, CHENNAI, TAMIL NADU, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3245 Page No 22 of 22
Validity 18/05/2021 to 17/05/2023 Last Amended on 14/07/2021

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Standard
Temp. / RH Indicator
THERMAL- RH and Temperature
with sensor by 18 %RH to 80
37 SPECIFIC HEAT chamber / Climatic 1.38%RH
Comparison Method %RH @25°C
& HUMIDITY Chamber
at single specified
position
Temperature
Indicator with Using Secondary
Sensor of (Water RTD (PT 100) and
Bath, Refrigerator, RTD Calibrator by
THERMAL-
38 Oven / Incubator, Comparison Method 0 °C to 400 °C 0.57°C
TEMPERATURE
Deep Freezer, for Non-Medical
Temperature Bath, Purpose - Single
Liquid Bath, Dry Position
Bath) with Controller
Temperature
Indicator with Using Secondary
Sensor of (Water RTD (PT 100) and
Bath, Refrigerator, RTD Calibrator by
THERMAL-
39 Oven / Incubator, Comparison Method -80 °C to 0 °C 0.51°C
TEMPERATURE
Deep Freezer, for Non-Medical
Temperature Bath, Purpose - single
Liquid Bath, Dry position
Bath) with Controller
* CMCs represent expanded uncertainties expressed at approximately the 95% level of confidence, using a coverage factor of k = 2.

This is annexure to 'Certificate of Accreditation' and does not require any signature.