Microsoft Word - UVWIN6 SoftWare Operation Manual
Microsoft Word - UVWIN6 SoftWare Operation Manual
Foreword
1. Introduction in general
UVWin6 is the new analysis software of UV‐VIS Spectrophotometers developed by our company
after UVWin5. Its main functions are controlling, measurement, analysis and data processing. Comparing
with its former version, UVWin6 includes more functions, such as spectral bandwidth scanning,
quantitative data upload etc. UVWin6 is designed to be compliant with GLP, including data file encryption,
instrument validation etc. UVWin6 is a universal software, which is suitable to all the UVS products
produced by PG Instruments, especially for the data disposal of T110 and T112.
Chapter Two “Software environment”. It gives you introduction of User interface, Functions Menu,
Configuration, Communication port and other appearance settings.
Chapter Three “Instrument Control”. It tells some methods of controlling UV‐VIS Spectrophotometers
with UVWin6. You’ll learn how to use the offered functions to control instrument, to accomplish
setting instrument and instrument correction.
Chapter Four “Photometric measurement”. It describes how to use Photometry to accomplish your
measurement. You’ll learn the usage relevant to photometric measurement and its parameter
settings.
Chapter Five “Spectrum Scan”. It describes how to use Spectrum Scan to accomplish scan qualitative
analysis of samples. You’ll learn the methods of spectrum scan functions and its parameter settings.
Chapter Six “Quantitative”. It describes how to use Quantitative Measurement to accomplish your
quantitative analysis of samples. You’ll learn the methods of quantitative measurement functions, its
parameter settings, setup and application of calibration curves.
Chapter Seven “Kinetics”. It describes how to use Kinetics to perform measurement of samples.
You’ll learn the methods of Kinetics and its parameter settings.
Chapter Eight “Spectral bandwidth scanning”. It describes how to use Spectral bandwidth scanning to
scan samples. You’ll learn the methods of spectral bandwidth scanning, its parameter settings.
Chapter Nine “DNA Protein Determination”. It gives you introduction of how to use DNA Protein
UVWin6 Software Operation Instruction Manual
Determination to perform analysis on samples. You’ll learn the usage and parameters setting of DNA
Protein Determination.
Chapter Ten “Pesticide Residual Determination”. It describes how to use Pesticide Residual
Determination to perform analysis of samples. You’ll learn the usage and parameters setting of
Pesticide Residual Determination.
Chapter Eleven “Graphic Process”. It gives you introduction of how to process spectra after scanning.
Chapter Twelve “Data Export”. It gives you introduction of how to export results data to other format
files and database.
Contents
Chapter 1 Software Installation and Configuration……………………………………………………1
1.1 Installation of UVWin6.0 Software……………………………………………...……………………1
1.2 Configuration Program………………………………………………………...………………………3
1.3 Chapter Summary………………………………………………………………….…………………5
Chapter 2 Software Environment…………………………………………...………….…………………6
2.1 Startup of UVWin6.0………………………………………………………….…….…………………6
2.2 Getting Familiar With UVWin6.0 Layout……………………………………..…..…………………8
2.3 Setting UVWin6.0 Operation Environment…………………………..…………….…………………11
2.3.1 Environment Options……………………………………………..………….…………………12
2.3.2 Custom Tool………………………………………………………………….…………………14
2.3.3 Communication Port………………………………………………………….…………………15
2.4 Chapter Summary………………………………………….……………………….…………………16
Chapter 3 Instrument Control…………………………………………………….…………………17
3.1 Instrument Performance…………………………………………………………….…………………17
3.2 Energy Setting……………………………………………..……………………….…………………19
3.3 Baseline Correction………………………………………………………..……….…………………20
3.4 Instrument Parameters Correction………………………………….…………….………………… 20
3.5 Wavelength Correction……………………………………………….…………….…………………23
3.6 Goto Wavelength…………………………………………….…………………….…………………23
3.7 Accessories Setting…………………………………………………………….….…………………24
3.8 Multi-Cell Blank Correction……………………………………………………….…………………25
3.9 Instrument Initialization…………………………………………………………….…………………25
3.10 Instrument Validation………………………………………………..…………….…………………26
3.11 Chapter Summary………………………………………………………………….…………………28
Chapter 4 Photometric Measurement…………………………………………….…………………29
4.1 Brief Introduction Of Photometric Measurement…………………….………….…………………29
4.2 Setting of Photometric Measurement Parameters…………………………………….………………29
4.3 Photometric Measurement………………………………………………..……….…………………33
4.4 Saving and Printout ……………………………………………………….……….…………………34
4.5 Chapter Summary…………………………………….…………………………….…………………40
Chapter 5 Spectrum Scan…………………………………………………….…….…………………41
5.1 Brief introduction of Spectrum Scan………………………………………...…….…………………41
5.2 Setting of Spectrum Scan Parameters……………………………...……………….…………………41
5.3 Spectrum Scan………………………………………………….………………….…………………43
5.4 Viewing Spectrum Information…………………………………………………….…………………43
5.5 Saving And Opening Of Spectrum Files…………………………………………...…………………43
5.6 Chapter Summary………………………………………………………………….…………………43
Chapter 6 Quantitative…………………………………………………………….…………………45
6.1 Brief Introduction Of Quantitative…………………………………...…………….…………………45
6.2 Setting Of Quantitative Parameters……………………………………………….…………………45
6.3 Quantitative………………………………………….…………………………….…………………51
6.4 Saving Measurement Files………………………………………….…………….…………………54
UVWin6 Software Operation Instruction Manual
Key points
This chapter, will show the following contents: UVWin6 operating environment, How to install
UVWin6 onto your PC, and how to use Configuration program to setup the software’s external
parameters.
It includes:
Installation of UVWin6 software
Configuration program
Chapter summary
UVWin6 can run on Microsoft Windows98/2000/XP/Win7 Desktop and NoteBook PC, Win7/Win8
Tablet PC. For your PC hardware, the least requirements are over 1.5GHz CPU, over 2GB Memory, screen
resolution over 1280×1024 pixels, and CD‐ROM or DVD‐ROM drive. For the tablet PC, the least
requirements are over 1.5GHz CPU, over 2GB RAM, screen resolution over 1280×800, and HD over 32GB.
Recommended configuration: Pentium1.4G CPU, 2GB Memory, 32M Graph Display Card, 17 inch
Color Display, CD‐ROM or DVD‐ROM. Among the above, Graph Display Card is mainly for 3D Display. So if
you do not have much need for 3D Display, you could reduce the requirement of Display Card. A usual
Display Card is enough. Besides, UVWin6 requires at least 20M hard disk spaces. So you should assure
enough hard disk space for installation of the software.
UVWin6 installation is not complex. What you should do is running Setup.exe in Installation Disk to
enter Installation Program. Please refer to Figure 1‐1.
Follow to the prompt of Installation Program to finish setup. You can choose “all users” or “only for
me” according to your demand when it’s prompted to choose application authority. Please refer to Figure
1‐2. The serial number of software is kept in Installation Pack. Please input the serial number when it is
prompted.
As UVWin6 is universal software for UV‐VIS Spectrophotometers, its installation program may varies
with different instrument models. And the shortcut and folder created by the installation program may
vary respectively.
If you need to remove UVWin6 from your PC, you could run Uninstall in UVWin6 Folder on Program
Group. Windows System will remove UVWin6 from your PC automatically. Please notice that before you
remove the software, you should backup your measurement data in advance so as not to lose them
accidentally.
The default folder of measurement data files is: “C:\Program Files\UV‐VIS\UVWin Software
v6.0.0\Data” .It may vary with different instrument models. So you’d better to find the relevant folder
from UVWin6 shortcut on desktop.
2
UVWin6 Software Operation Instruction Manual
Tips:“Shortcut”
If you do not know the file path that the shortcut links to, you could point at the shortcut and click
right button of mouse. On pop‐up menu, select Attribute to open the Shortcut Attribute Window. In the
Attribute Window, click the Search Object to open the Windows Explorer and go to the file folder that the
shortcut links.
UVWin6 Configuration Program is a tool of setting UVWin6 operation conditions. You could start it by
Selecting [Start]‐> [Program]‐> [UVWin Software v6.0.0]‐> [Configuration Program]. The operation dialog
box of Configuration Program is as shown in Figure 1‐4.
Accessories Tab
Selecting accessories that the instrument uses. The option includes “8‐Cell Holder”, “5‐Cell Holder”,
“Fixed Cell Holder”, “Fixed Flow Cell Holder” and “Integrating Sphere”, as shown in Figure 1‐5. You may
select it according to instrument outfit.
3
UVWin6 Software Operation Instruction Manual
Bandwidth Tab
Setting a slit value for instruments with changeable slit. If your instrument is fixed slit, you can set slit
value of the instrument here. As shown in Figure 1‐6. The value will be kept with your measurement data
files. And it is also used with printout.
Help Tab
In this option, you can get description of how to use UVWin6 configuration program.
After you finish configuration, click on OK to save settings and exit. If you want to quit settings, click
Cancel or close the dialog box directly. If you check the option box of Run UVWin after exit, System will
run UVWin6 automatically after you exit with OK.
4
UVWin6 Software Operation Instruction Manual
Key points
In this chapter, we’ll tell of the following contents: UVWin6 user interface, menu functions, and
environmental parameters setting.
Startup of UVWin6
Getting familiar with UVWin6 Layout
Setting UVWin6 Operation Environment
When you are going to run UVWin6, be sure that connection between UV‐VIS Spectrophotometer
and PC is good and instrument is turned on. Here, you can click UVWin6 shortcut to start it. The startup
window is shown as Figure 2‐1.
If instrument works normally, it shows the instrument initialization window as shown in Figure 2‐2.
(Please refer to 3.1 Instrument Performances for details of initialization)
6
UVWin6 Software Operation Instruction Manual
After initialization, enter into the UVWin6 main user window. If initialization fails, the System will
display message, as shown in F igure 2‐3. Select “Yes” to enter into System, or “No” to exit. Please notice
that when initialization fails, do not enter into the System to operate the instrument as initialization
failures have various causes, it is possible that there are some problems on an instrument
component. If you do some work on defective components, you may make some needless damages. So it is
best to exit the System after initialization fails.
If instrument does not work well, or connection between instrument and PC has a problem, it would
cause the phenomenon of unable to be online. A message about it will display as shown in Figure 2‐4. At
this time, you should check if communication cable is connected correctly or not and if instrument is turned
on or not. Besides, if your settings of communication port have problems, it may cause the above
phenomenon too. In this case, you should select Cancel to exit the program and run UVWin6
Configuration to reset communication port, and startup UVWin6 again (Please refer to 1.2 Configuration
Program for UVWin6 Configuration Application).
Select “Ignore” to enter into System offline. You are unable to operate instrument in this case.
7
UVWin6 Software Operation Instruction Manual
No matter with which method introduced in last section you choose to enter into UVWin6, UVWin6
Main User Interface will open in front of you finally. As Shown in Figure 2‐5.
Toolbar Menubar
Tree Window
Module Screen
Status Bar
Main User Interface adopts a multi‐file tree structure layout. This kind of layout not only has orderly
structure and good looking, but also eases for users to operate. From above figure, you could find that
UVWin6 main user interface include the parts of Menubar, Toolbar, Statusbar, Tree Structure Window,
and Module Window. Following is the introduction of these composing parts.
Menubar
For most of software, Menu is absolutely necessary. It is necessary for UVWin6 too. A majority of
8
UVWin6 Software Operation Instruction Manual
functions in the software are provided in Menubar. Menubar include submenus of File, Edit, Measure,
Graph, Mathematic Calculation, Administrate, Tools, Application, Window, and Help. As shown in Figure
2‐6. You can use menu functions provided in Menubar to accomplish its corresponding setting for
software. In the following chapters, the manual will give introduction of menu functions combined with
its software functions introduction.
Toolbar
The role of Toolbar is listing those commonly used functions together as tool buttons. In this way,
users can achieve menu functions by single press of these tool buttons, so as to dispense with those trivial
menu operations. UVWin6 Toolbar consists of General Toolbar, Configuration Toolbar, Window Toolbar,
Graph Toolbar, Control Toolbar, and Kit Toolbar. Shown as Figure 2‐7. Between each Toolbar, there is a
partition line. When cursor moves over the partition line, cursor changes to the “ ” shape. At this time,
hold pressing left button of mouse and drag the partition line to change its position. As shown in Figure
2‐8.
Status Bar
Status Bar locates at the bottom of whole window, which is used for display of status message in
software. For example, status of command execution, data file type, current username, file call schedule,
9
UVWin6 Software Operation Instruction Manual
etc. In the spectrum window and the time course measurement window, Status Bar also displays actual
coordinate values of current cursor located. As shown in Figure 2‐9. In addition, if your main interface
window is not maximum, there is a “ ” sign at the right lower of Status Bar. You can drag with cursor the
moving sign to adjust the window size.
Workspace Window
Accessories Window
Figure 2‐10 Work Space Window (Upper) and Accessories Window (lower)
10
UVWin6 Software Operation Instruction Manual
Module Window
As a matter of fact, the Module Window is a sub‐window of main interface window. The fact to call it
as Module, is that each window can accomplish different measurement functions. The Module Window
consists of the Photometry Window, the Spectrum Window, the Quantitative Window, the Kinetics
Measurement Window, and the Spectral Bandwidth Scanning Window. As shown in Figure 2‐11. You can
enable each Module Window by using mouse to hit Window Title Bar, or by selecting corresponding
Module Name in “Work Space”.
Tips: “Window”
In Windows, nearly all messages are showed with “Window”. Here, “Window” not only stands for
those windows, which have Title bar and can be dragged and whose sizes can be adjusted, but also stands
for all layout which could display messages. For example, Resource Manager is a standard window. And
Toolbar of Resource Manager is also a window. But the toolbar is only subject to Resource Manager’s
main window. So the toolbar is called one of son‐windows of Resource Manager. And Resource Manager
is called father‐window. In this way, tool button of Toolbar is also a window. Its father‐window is Toolbar.
Resource Manager is a single‐file program. So, its sub‐windows can but lay on settled position, or make
simple moves. But for multi‐file program, it can setup one or several son‐windows as a standard window.
In this way, the son‐windows can be dragged, change their sizes, and hold their own title bar as their
father‐window. Of cause, its active area is within its father window. UVWin6 adopts multi‐file structure.
Each module is set as movable son‐window, which makes its operation more flexible, and convenient.
UVWin6 working environment consists of Environment Options, Toolbar Setting, Window Setting,
Self‐defined Tool Setting, and Communication Port Setting. With these setup functions, it would make
your routine operation easier. And it would also bring certain convenience to your work. Next,it’ll give
you introduction of detailed settings of these functions.
11
UVWin6 Software Operation Instruction Manual
Indication
Prompt before Measurement: When you press Measurement Button, a prompt message will be
displayed to indicate you to put in sample. After you confirm it, measurement continues. Otherwise, it
does not continue to start measurement.
Prompt after Measurement: Similar to Prompt Before Measurement, but a prompt window will be
displayed after measurement to indicate you that measurement is over.
Beep after Measurement: Similar to Prompt after Measurement, but instead of displaying a prompt
window, a beep will be made after measurement to remind you that measurement is over. If your PC is
fitted with audio card, and acoustics, or earphone, is open, sound would be sent out from acoustics or
earphone. If there is not audio card in PC, PC speaker will send out a short beep.
The promote message of displaying kinetics scanning mode: choose this and choose multi cell
scanning as work mode from the kinetics scanning parameters, the promote message will be shown “It
need 2 seconds to switch between every cell holder, so please set Time interval longer. For example, Time
interval should be longer than 4 seconds if choose 2 cell holders, while 16 seconds for 4 cell holders.
Continue or not?”
File Option
Maximum open files: Setting files number that each module can hold simultaneously. For example, in
Spectrum Scan, it means maximum number of spectra displayed at a time.
Auto‐Save Baseline as a file: The function is designed only for Spectrum Scan Module. If you choose
12
UVWin6 Software Operation Instruction Manual
the function, System will save calibration data to a disk file automatically, after each baseline correction
finishes. And System will load the saved baseline data automatically when UVWin6 runs next time. In this
way, you can do spectrum scan directly without doing baseline correction.
For general UV‐VIS spectrophotometer, wide range spectrum scan is time consuming relatively. And
baseline correction takes nearly same time as that of spectrum scan. So, baseline data saving could not
only save your working time and also reduce wear and tear of instrument. It is a double profit. On the
other hand, being affected by working conditions and environmental factors, operating status of UV‐VIS
spectrophotometer may be different to some extent. So, if you do not think scanning result is satisfactory,
you have to do baseline correction again.
For CCD Fast UV Spectrophotometers, the effect of this function does not work much. As fast UV
Spectrophotometer’s CCD detector receives spectra reflected by grating directly, and CCD is very sensitive
to changes of environmental conditions, so instrument must do baseline correction before to start
spectrum scan. And it is a matter of no account that whether to save or not baseline data.
Data Format
Data Format allows you to set related data display format in UVWin6. Mainly they are decimal digits
for display. For example, “0.00” is two decimal digits for display. The data being able to be set its decimal
digits include:
Abs: Absorption
T% : Transmittance
E%: Energy
Conc.: Concentration in Quantitative measurement.
SD: Standard Deviation in Quantitative measurement.
RSD: Relative Standard Deviation in Quantitative measurement.
R2: Standard Linear Relativity.
Default Directory
Setting default directory for files access. That is original directory displayed in dialog box, when you
save, or open a file. “Data” indicates directory for all data files saving. “Export” represents default
directory for exporting files. You can select corresponding directory through pressing “…” buttons.
Menu Setting
For running UVWin6 on PC, please choose [Standard Menu]; For running UVWin6 on Tablet PC,
please choose [Tablet PC Menu].
Web Service
13
UVWin6 Software Operation Instruction Manual
Left window shows all tools list, while right side shows relevant tool info. Tool info includes some info
about program files called by tools.
Program: Executive program called by current tool. It could be executable file (.EXE), or page file
(HTM). You can click on the button right to program edit box to select program file.
Title: Title text showed by the tool.
Directory: The working folder of called program. The default working folder is the directory in
which the program is saved. When you select a program file, System will set the parameter to a
default.
Parameters: Operation parameters of called program. For example, you want to setup a tool to
open the “hh.txt” file in C Disk root directory, with Windows Notepad. You should set Program
as “C:\Windows\Notepad.exe”, and Parameters as “C:\hh.txt”.
Icon file: Icon file is used to provide display icon for Toolbar. Default icon file is program file. You
also are able to click on the button right to icon file edit box to select icon file.
Icon: For an icon file of “ICO” extension name, it only includes one icon. But for a file of “EXE” or
“DLL”, it may include many icons. So, you can select some other icons in the icon file through
this option.
If you want to add a tool, you can do it by editing tool info directly and press Add button. After you
finish some tools info modification, you can press “Replace” to update them. “Delete” button is used for
deleting current tool. After you finish tool configuration, you close the window and buttons of “Tool”
toolbar in main interface are refreshed.
14
UVWin6 Software Operation Instruction Manual
Figure 2‐14 Communication Port Setting Window ‐ Online (Up); Offline (Down)
You can select a port name in pull‐down box of Port, and then press OK button. If you are not sure
which port connecting with instrument, you could click on Auto‐Test button and System will test all ports
automatically. If it finds the port connecting with instrument, it would be online automatically.
Support USB communication, choose the correct port and connect with the PC.
15
UVWin6 Software Operation Instruction Manual
This chapter gives introduction of UVWin6 working interface and environmental setting. You can
setup UVWin6 operation environment by using Environment Setting functions. In addition, you also be
able to make your future operation easier by utilizing Custom Tool to setup Toolbars. After reading of this
chapter, you must have a preliminary knowledge about UVWin6. In the next chapter, it’ll give you
description of other functions.
16
UVWin6 Software Operation Instruction Manual
Key points
In this chapter, we’ll tell of the following contents: Instrument parameters setting, instrument
correction, instrument validation, instrument accessories, and etc.
It includes:
Instrument performance
Energy setting
Baseline correction
Instrument parameters correction
Wavelength correction
Goto wavelength
Accessories setting
Multi‐Cell blank correction
Instrument initialization
Instrument validation
Chapter Summary
For UV‐VIS spectrophotometers, optics system is a core of whole instrument. As today’s instruments
hold powerful automation, so many additional options for optics system setting would make operation
easier and control more flexible. UVWin6 Instrument Performance Option will provide you a function of
setting instrument optics system parameters. Selecting “Instrument Performance” submenu under
“Measure” Menu, or clicking on toolbar button , System will pop up the Instrument Performance Setting
Window. In the window, you can set status of light source, and some characteristics of optics system. As
shown in figure 3‐1.
17
UVWin6 Software Operation Instruction Manual
Lamp Status
Setting on‐off switch of Deuterium lamp and Tungsten lamp. It is on when button is red, while it is off
when button is grey.
Spectral bandwidth
Setting spectral bandwidth of instrument slit. The optional range from 0.1 to 5nm. Not all the model
include 0.1nm bandwidth, choose as the specification of the instrument. For fixed slit instruments, it
needn’t to set the parameter. If you want to change spectral bandwidth of fixed slit instrument, please
refer to 1.2 Configuration Program.
Response time
Setting instrument sampling response time. More time it is set, more accurate datum is sampled.
The parameter is only usable for general UV‐VIS spectrophotometers. For CCD Fast UV
Spectrophotometers, there is not such parameter.
Beam size
Set the size of the beam.
SBA
Set the size of attenuator for sample beam.
18
UVWin6 Software Operation Instruction Manual
RBA
Set the size of attenuator for reference beam.
For UV‐VIS spectrophotometers, photometric mode includes absorption (Abs), and transmittance
(T%) generally. Sometimes energy mode is required to be used for energy distribution analysis of samples.
In this case, photometric mode is required to be set to energy mode. For energy mode, there are some
special settings. After you choose Energy Mode, you can select Energy submenu under Measure Menu, or
click on tool button, to open the energy settings window. As shown in Figure 3‐2.
Lamp
Setting lamp used under energy mode. For general UV‐VIS spectrophotometers, it is set as deuterium
lamp, tungsten lamp or mercury lamp. For the instrument which adopts see‐through type deuterium
lamp, there is no need to set lamp for energy mode.
Gain
Setting a gain value under energy mode. The gain value range varies with different instrument
models. For example, UV‐VIS split beam spectrophotometer has gain value range of 0‐15 with 16 steps.
And UV‐VIS double beam spectrophotometer (PMT) has gain value range of 1‐7 with 7 steps. The role of
gain is to amplify measured signal, so as to perform observation and measurement on spectra of low
energy. But gain amplification would affect data quality, that is, more gain it is, more noise data signal has.
Presenting on spectra is that curve’s serration becomes larger. Measured data are relatively inaccurate.
After energy setting is finished, press OK to update energy parameters of the instrument. If the lamp
source is the first time set to Hg, it will choose the Hg lamp as the source.
19
UVWin6 Software Operation Instruction Manual
Instrument Parameters Correction includes the dark current correction, Wavelength correction with
multi‐points, attenuator correction and spectral bandwidth correction. Open this window by click
[Measure]‐[Instrument Parameters Correction] menu, shown as figure 3‐3.
20
UVWin6 Software Operation Instruction Manual
Spectral Bandwidth Correction is used to correct the accuracy of spectral bandwidth, only used for
continuously selectable slit instruments. It can be used if the error over 20%. Keep the cell holder clear
when doing this operation as figure 3‐4 shown.
Wavelength
Display the current wavelength when doing spectral bandwidth correction.
Half Peak width
Display the half peak width after scanning and calculating.
Spectral Bandwidth
Input the correct value.
Correct
Click this button to start correction. There will be some prompt if success. Please contact with
the Supplier if failed.
Reset
Click this button to restore the default settings.
Notice: Be caution when using this function, all the data store in the chips will be clear and the
spectral bandwidth need to re‐correct.
21
UVWin6 Software Operation Instruction Manual
Attenuator value
Display the attenuator value of sample and reference from 1% to 10%.
Correct
Click this button to do the attenuation correction, and then check the validity, if yes send to the
instrument.
Add
Click this button to add a new line in the wavelength list.
Delete
Click this button to delete a line in the wavelength list.
Hg scan
Click this button to scan the energy of mercury lamp from 250nm‐550nm, the spectrum and
wavelength will be shown in the window.
Advanced Tab
The wavelength status can be set in this tab, as figure 3‐7 shown.
Not correct
Scan without referring the corrected wavelength list
Corrected
Scan according to the corrected wavelength list
Goto wavelength is moving current instrument wavelength to your set position. Select Goto
23
UVWin6 Software Operation Instruction Manual
Wavelength submenu under Measure Menu to open the Goto Wavelength Window as shown in Figure
3‐8. In this window, input wavelength you want to set and press OK button to confirm it.
Accessories setting allows you to setup accessories that instrument uses. At present, the optional
accessories include Fixed Cell Holder, Flow Cell, Integrating Sphere, 5 Cell Holder, and 8 Cell Holder. Select
Accessories submenu under Measure Menu to open the Accessories window as shown in Figure 3‐9. In
this window, you can select a type of accessories and setup sample types accordingly. If you select
multiple cell holder, you can click on Position to set corresponding cells as “ ”, so as to setup it as current
sample cell.
Type
Corresponding to sample type of each cell. If one cell is set as Blank sample, System will move to it to
do correction when baseline correction is going to be done. If one cell is set as Unused, System will never
24
UVWin6 Software Operation Instruction Manual
Sample Name
Setting sample name of corresponding cell. The name will be saved with measurement file and be
able to be printed out.
If the instrument is equipped with multi‐cell holder (5‐cell holder or 8‐cell holder), you could do the
blank correction to every blank sample in the cell holders alone as shown in figure 3‐10.
Select Initialization submenu under Measure Menu to do instrument initialization again, as shown in
Figure 3‐11. In the initialization window, UVWin6’ll initialize some instrument automation mechanisms.
The automation mechanisms may vary with different instrument models. For example, instrument with
variable slit will initialize its slit motor, while instrument with fixed slit will not. UVWin6 initialization
window covers whole screen. Its background shows some info about the software. In the initialization
window, it shows instrument name, instrument serial number, and initialization items. In initialization
procedure, passed items show “ ” sign, while failed items show “ ” sign. Please refer to instrument
service manual for detailed causality of initialization failure and its troubleshooting.
25
UVWin6 Software Operation Instruction Manual
Instrument validation will help you accomplish instrument performance check automatically. Select
Instrument validation submenu under Measure Menu to open the Instrument validation window as
shown in Figure 3‐12.
26
UVWin6 Software Operation Instruction Manual
Instrument Validation includes 6 items: “Wavelength accuracy”, “Photometric accuracy”, “Stray light”,
“Noise”, “Baseline flatness”, and “Stability”. You are able to select and setup each of them.
Wavelength Accuracy
Using characteristic spectral lines of “Holmium filter” or “Deuterium lamp” to calibrate instrument
wavelength accuracy. You can click on Setting button to make small amendment on characteristic
wavelength.
Photometric Accuracy
Using characteristic spectral lines of “Medium density filter” or “Potassium dichromate solution” to
calibrate instrument photometric accuracy. You also can click on Setting button to make small
amendment on characteristic absorption.
Stray light
Using “Nitrous acid sodium” or “Sodium iodide” solutions to calibrate instrument stray light.
Noise Level
Measure instrument noise level in 120 seconds.
Baseline Flatness
Measure instrument baseline flatness
Stability
Measure instrument stability. You can select measurement of one hour or half an hour.
If you do not want to perform some items check, you can undo the check boxes before the items.
After you finish validation setting, press OK button and System will form a validation wizard according to
your settings. You are able to accomplish instrument validation with instruction of the wizard, as shown in
Figure 3‐13.
If your instrument has Multiple cell holder, you can select Auto Validation in the first page of
validation wizard and setup cell positions for relevant samples. Press Next and System will start validation
automatically. If you do not select Auto Validation, you need to continue validation procedure step by step.
In process of validation, System will mark failed items. In the last page of validation wizard, you can click
Print Results Report button to print out validation results. When printout, System will give different marks
to data and results of each check item. Passed items will be signed by System with “ ”. Otherwise, there
is not such sign.
This chapter mainly gives you introduction of how to control instrument through UVWin6 provided
functions. After reading this chapter, you must have some more knowledge of instrument manipulation
with software. Hope you can learn how to use the software through practice on instrument.
28
UVWin6 Software Operation Instruction Manual
Key points
In this chapter, we’ll tell of the following contents: What’s photometric measurement? How to setup
parameters for photometric measurement? How to perform photometric measurement? How to save and
printout measured results?
It includes:
Brief introduction of Photometric measurement
Setting of photometric measurement parameters
Photometric measurement
Saving and printout of measured results
Chapter Summary
“Photometric measurement” is readout of measuring data at the wavelength points you set. In
UVWin6, you can set multiple wavelength points for photometric measurement. Furthermore, you can do
some simple calculation on measured data. Please refer to next section about this.
To enable Photometric Measurement, you should click on the Photometry tab in Work Space. Select
Parameters settings submenu under Measure Menu to activate the Photometric measurement setting
dialog box, as shown in Figure 4‐1. In this dialog box, there are 5 tabs, and you can setup them according
to your various needs.
29
UVWin6 Software Operation Instruction Manual
Measurement Tab
Wavelengths
In measurement tab, you can input in Wavelength Edit Box the wavelengths at which you want to
measure. Then click Add button to add it into its below wavelength list. The number of measuring
wavelengths can be up to 26, while one is least. If you need to delete wavelength points, or clear the
wavelength list, you can press Delete Button, or Clear Button. After you choose a wavelength in
wavelength list, there shows the same wavelength in Wavelength Edit Box. Here, you can amend the
wavelength and press Replace Button to update it correspondingly in Wavelength list.
Repeat Measurement
Photometric measurement enables you to select number of repeat measurements. If you do not
require repeat measurement, select No in Repeat Measurement Option. If manual repeat measurement is
required, select Manual in the same. And then input number of repeat measurements in Repeat Times
Edit Box. Same as that of Manual, Automatic is also performing repeat measurement. And for Automatic,
it accomplishes multiple repeat measurements automatically, without each press of Measure Button.
What Automatic requires specifically is to set a time interval for measurement, that is, waiting time
between every two measurements. This interval could be zero, which means to perform continuous
measurement without pause. If you have chosen Automatic Repeat, you are also able to check the option
box of “Repeat measurement automatically according to cell number”, which will do once measurement
over all samples in cells. In this way, there is no need to set repeat times, which will be forbidden. If you
set cell holder type as fixed cell holder, you cannot check the option box. Anyway, the main purpose of
repeat measurement is in fact for averaging of repeat measurement data. So you can check “Calculate
average value” to enable average calculation. Then System will calculate and display the average
30
UVWin6 Software Operation Instruction Manual
Calculation formula
In “Calculation formula” edit box, you can input required formula for results calculation. In formula,
letters of A, B, C, and D…represent measured data of corresponding wavelengths. For example, you input
two wavelength points in Wavelength list of Measurement Tab, which are 600nm and 500nm. When you
want to calculate the ratio of measured data of these two wavelength points, you can input A/B in
Formula Edit Box of Simple Calculation Tab. Then press Add Button. The default title of calculation formula
is “Result 1”, “Result2”… If you want to specify a title, you can input it, at same time when you input
calculation formula, in Title Edit Box. If you need to amend a formula, select it correspondingly in formula
list, and revise it in formula edit box. Press Revise Button to make it effective. If you want to delete or
clear the content of formula list, press Delete Button or Clear Button. Number of Formula can be up to 10.
Characters
31
UVWin6 Software Operation Instruction Manual
The role of Characters here is for input imitating as keyboard. Press a character button so as to input
the corresponding character equivalently.
Display
The role of Display option here is for selection of different display modes for calculation formula.
Pull‐down box offers two choices, formula and title. Formula here means calculation formula will be
displayed in result table as title. Title here means the default title or user set title will be displayed.
Instrument Tab
The content of Instrument Tab is same as that of Instrument Performance Settings. Please refer to
3.1 Instrument Performance about it.
Accessories Tab
The content of Accessories Tab is same as that of Accessories Settings. Please refer to 3.7 Accessories
Setting about it.
Quality Control is a new function offered by UVWin6. Its role is quality monitoring on measured data.
In case there are abnormal data measured, System will prompt or act according to your preset. Of cause,
judgment way for data can be preset. Shown as in Figure 4‐3.
In the quality control window, you can set Quality Control on or off through “Enable quality control
function” option.
Quality Control List
Set items for quality control. Letters of A, B, C, D… express columns of measuring wavelength points,
result1, result2, result3…, and calculating results. Click at All Button to select all items, while click on
Invert Button invert selection of all items. Click on Clear Button to clear all selections.
Limitation
In limitation box, you can input High Limit and Low Limit for control of selected items. Fail Action
could be set for System to act when data out of limits occur. The optional actions include: Continue –
continue to measure, Stop – stop measurement, and Re‐measure – Measure current sample again. If you
need to make a mark on result out of limits in result table, click on Marking Button to set mark type, as
shown in Figure 4‐4.
Check the Enable Option to open marking function. In Mark box, you can select marking mode for
those data out of limits. “Grid” is marking the grid data located in. The optional marks can be selected
from pull‐down list of Grid. If you choose “Text”, color and font of text can be setup.
Second fail action
Second fail action means that System would take what action when there are successively twice of
out of limits. The optional actions include “Continue” and “Stop”.
The measuring procedure of Photometric Measurement is very simple. Just click on button to
finish a measure. The measured result is displayed in result table. If you want to delete a measured result,
use mouse to click on it and select Delete submenu under Edit Menu to delete it. If you need to recall the
deleted results, you can click right button of mouse in result table. In popup menu, select the Cancel
Deleting Submenu under the Delete Menu to recall the deleted results. If you want to hide the deleted
33
UVWin6 Software Operation Instruction Manual
results, uncheck the Display Deleted Sample under the Delete Menu.
For measured results, they are not only able to be saved to a file, but also be able to be printed out.
After you finish analytic measurement, you can select Save submenu under File Menu, or click on
button. System would pop up the Save File Window, as shown in Figure 4‐5. Input a filename you want to
save to and press Save button to save files to a specified location.
In lower part of the window, it shows current analyzer. Click on Note Button to make a note on
measured results. Shown as Figure 4‐6.
34
UVWin6 Software Operation Instruction Manual
You are able to input measuring sample name in Sample Name edit box, input notes on measured
results in Comments edit box. All info above will be saved to measuring files for future reference.
If you want to print out measured results, you can select Print submenu under File Menu, or click on
button to print them out. Naturally, you must to ensure that your PC has already installed a printer. If
not, System will give an error prompt. An addition, if you want to modify printout format, you can select
Page Settings submenu under File Menu, or click on button to open the Page Setting Window. Shown
as Figure 4‐7.
Format Tab
The contents of Format Tab are mainly setting the format, font, table, and direction of page output.
Shown as 4‐8.
Text
Setting text printing format. The optional formats include Normal, Simple, and Table. “Normal”
means printing out in text, without any decoration. “Simple” is to underline sort headers, to highlight
headers. “Table” means printing out in table. In addition, you are also able to click on Font Button to set
print fonts.
Row Spacing
Setting the interval of each text row. The optional row spacing includes “Minimum”, “0.5 Word”, “1
35
UVWin6 Software Operation Instruction Manual
36
UVWin6 Software Operation Instruction Manual
“Module Name”: To print the name of current measurement module, such as Photometric,
Spectrum Scan, etc.
“Instrument Name/Number”: To print the name and serial number of current instrument, such as
UV‐VIS Spectrophotometer/01‐1901‐01‐0001.
“Analyser”: To print current login username.
“Company”: To print current user’s company name.
“Page Number”: To print current page number.
“Custom”: To print custom text. If you choose this option, bellow the pull‐down menu, there is an
Edit Box, in which you can input custom text.
If you have input relevant contents, you can click on “Font” Button to set its font. In addition, you
also can select Separation Line to draw separation lines among Header, Footer, and Text.
38
UVWin6 Software Operation Instruction Manual
In the print preview window, click on or buttons to zoom the page. Click on button to
print. Click on button to perform page settings.
This chapter mainly gives you introduction of how to setup and perform measurement in
39
UVWin6 Software Operation Instruction Manual
Photometric. Further more, it tells functions of print settings and print preview, which are common for
other operation in the software. Therefore, it will not give details about print operation in the following
chapters.
40
UVWin6 Software Operation Instruction Manual
Key points
In this chapter, we’ll tell of the following contents: What’s Spectrum Scan? How to setup parameters
for Spectrum Scan? How to perform Spectrum Scan? How to view spectrum information? How to save
and open spectrum files?
It includes:
Brief introduction of Spectrum Scan
Setting of Spectrum Scan parameters
Spectrum Scan
Viewing spectrum information
Saving and Opening of spectrum files
Chapter Summary
Spectrum Scan is scanning on a certain wavelength range with a definite wavelength interval. In the
scanning course, it gives readout of measurement data with each wavelength changes and display the
measured data in a graph of two dimensions for further analysis and study. Spectrum scan is mainly used
for qualitative analysis of samples. With its visual graphic display, Spectrum scan is more clear for
operators with samples’ properties. So it is one of the absolutely necessary functions for UV‐VIS
Spectrophotometers.
If you want to perform spectrum scan, first you should set scanning parameters. To enable the
Spectrum Scan Window, you should click on the Spectrum tab in Work Space. Select Parameters settings
submenu under Measure Menu, or click on button, to activate the Spectrum scan setting dialog box,
as shown in Figure 5‐1.
41
UVWin6 Software Operation Instruction Manual
In this dialog box, there are 3 tabs: Measurement, Instrument, and Accessories. Among them,
Measurement Tab is mainly for setting spectrum scanning parameters, while other two tabs are as same
as those of Instrument Performance and Accessories Setting. Please refer to 3.1 Instrument Performance
and 3.7 Accessories Setting for the details. Here’ll not repeat anything more about it. The following will
focus on scanning parameters setting.
Display Range
Setting the Y‐axis range for spectrum scan. You can input relevant values in Maximum and Minimum
Edit boxes.
Scanning Parameters
Setting scanning parameters of wavelength range, wavelength interval, speed, and etc. Wavelength
range consists of Start and End. Speed expresses Scanning speed. Faster scanning speed will give
comparatively less data quality, while slower scanning speed will give comparatively better data quality.
Interval expresses scanning wavelength interval. That is, at an interval of how many nanometer to sample
each data. The optional scanning intervals are 0.1, 0.2, 0.5, 1.0, 2.0, and 5.0nm. The role of Auto Sampling
Interval is selecting a scanning interval for you automatically according to scanning range set by you.
Scanning Mode
42
UVWin6 Software Operation Instruction Manual
Scanning mode is setting repeat scanning mode for spectrum scan. Single Scanning means
performing scan only one times, no repeat. Repeat Scanning means performing multiple repetitive
scanning. Automatic Scanning is performing scanning according to cell number of accessories you select.
If you choose Repeat Scanning, you should set Time Interval and Repeating Times for scanning. If you
choose Automatic Scanning, you need not to set Repeat Times. Only Time Interval is required to be set.
Select Start Submenu under Measure Menu, or click on button, to start spectrum scan. In the
procedure of scanning, System will draw dynamically scanning data and wavelength on the Spectrum Scan
Window in graphic mode. Furthermore, data will be also displayed in Toolbar. If you want to cancel
scanning, click on Stop Button or press ESC Key.
Select Spectrum Information Submenu under Graph Menu, or click on button, to open the
spectrum information window. Shown as Figure 5‐2. In the window, all spectral items will be displayed in
the list. You can click on any of them and its details will be shown in right window. You can also click on
Color button to change the display color of corresponding spectrum.
If you want to save a spectrum, you can select Save Submenu under File Menu, or click on the
43
UVWin6 Software Operation Instruction Manual
button to popup the Save Window. In the window, you should input a saving filename and press OK
Button. When you want to view the saved spectrum files, you can select Open Submenu under File Menu,
or click on the Button to popup the Spectrum File Window, as shown in Figure 5‐3.
In this window, you can select the spectrum that you want to open. The right graph is the preview
window. Click on Attribute Button to see the details of a spectrum file.
This chapter mainly tells you of some contents about spectrum scan. We trust you have got more
knowledge of spectrum scan application. Hope you could be gradually familiar with software usage
through practice of instrument application.
44
UVWin6 Software Operation Instruction Manual
Chapter 6 Quantitative
Key points
In this chapter, we’ll tell of the following contents: What’s quantitative? How to setup parameters for
quantitative? How to perform quantitative? How to save measurement files?
It includes:
Brief introduction of Quantitative
Setting of quantitative parameters
Quantitative
Saving measurement files
Chapter Summary
To enable the Quantitative Window, you should click on the Quantitative tab in Work Space. Select
Parameters settings submenu under Measure Menu, or click on button, to activate the Quantitative
setting dialog box, as shown in Figure 6‐1.
45
UVWin6 Software Operation Instruction Manual
In this dialog box, there are 6 tabs: Measurement, Calibration Curve, Instrument, Accessories, Quality
Control and Simple Calculation. Among them, Instrument and Accessories tabs are as same as those of
Instrument Performance and Accessories Setting. Please refer to 3.1 Instrument Performance and 3.7
Accessories Setting for the details.
Measure Tab
Measure Tab mainly provides setting of quantitative methods, wavelength points, repeat
measurement, and etc. You can select different measurement methods for various determination
requirements. Meanwhile, you can also set sample number and measurement repeat times. Shown as the
above.
Measurement Method
Setting measurement method for quantitative. The optional measurement methods are: single
wavelength, double wavelength, double wavelength with coefficient, three wavelength, first order
derivation, second order derivation, third order derivation, and fourth order derivation. You can select a
method according to various measurement requirements. If you choose the single wavelength method,
you should input measuring wavelength in Main Wavelength Edit Box. If you choose the double
wavelength method, or the three‐wavelength method, you should input baseline wavelengths in Baseline
Wavelength1 and Baseline Wavelength2 Edit Boxes. Please refer to Appendix A Application Methods of
Quantitative for details of Quantitative methods.
Sample Number
Setting Identifier and Number for standards and samples.
Others
46
UVWin6 Software Operation Instruction Manual
In Others Setting, you can set duplicate measurement for samples. Duplicate measurement is
performing multiple measurements on a sample. And then take the average of the multiple measurement
results into content calculation. Therefore, if you check the Duplicate Measurement Option, you are also
required to input the necessary repeat measurement times, so as to achieve the purpose of repeat
measurement, more accurate measured data. But It is sure to take some more time than usual
measurement. Furthermore, if your instrument is equipped Multi‐cell holder, you can choose Cell Holder
Interchange Automatic option. Then, System would move to next cell after it finishes one sample
measurement. What you have to do is placing samples in their orders.
Curve Equation
Setting an equation used for calibration curve. The optional equations have two: C=f(Abs) and
Abs=f(C). The former is taking absorption as variable to compute concentration, while the latter is an
inverse function of the former, mainly used for computing absorption coefficient.
Equation Order
Equation order expresses that of calibration curve, that is, the order of a polynomial. The optional
orders are: First, Second, Third, and Fourth. Among them, first order equation is called linear equation,
while others are called nonlinear equation. As UV‐VIS Spectrophotometers have good photometric
accuracy and photometric repeatability, the measured data in the allowed absorption range are linear
47
UVWin6 Software Operation Instruction Manual
with their concentrations. So linear equation has the wider range of application. You can select different
equations according to your actual situation. Shown as Figure 6‐3 and Figure 6‐4.
Input type
Conc. (Concentration)
Input the concentration of standard sample before measuring it, then measure the Abs reading. The
software will fit the calibration curve synchronously by taking concentration as abscissa and abs as
ordinate. Then measuring the Abs reading of unknown sample and calculate the concentration by the
calibration curve.
Volume
Input the volume and concentration of standard sample before measuring it, to get the mass of
standard sample, then measure the Abs reading. The software will fit the calibration curve synchronously
by taking concentration as abscissa and abs as ordinate. For unknown sample, input the volume, and
measure the Abs reading, then calculate the mass by the calibration curve, so get the concentration
according to the mass and volume.
48
UVWin6 Software Operation Instruction Manual
Blank correction
If “Blank correction” is chosen, the △Abs will be got by sample Abs reading subtract blank reading.
And the ordinate of calibration curve will be △Abs.
Blank setting
Choose the corresponding sample and set as Blank.
Natural logarithm
If “Natural logarithm” is chosen, calculate ㏑ Abs and take as the ordinate of calibration curve.
If “Natural logarithm” and “Blank correction” are chosen together, calculate ㏑△Abs and take as the
ordinate of calibration curve.
Concentration Unit
Setting concentration unit for samples. The default concentration units have: ng/ul, ng/ml, ug/ul,
ug/ml, mg/l, ppb, ppm, and mol/l. If there is not concentration unit that you need, you can input a new
concentration unit manually.
Zero Interception
Zero Interception is to add a standard point of both zero values for concentration and absorption
before the first standard. The zero point would be taken into fitting calculation.
Curve Evaluation
Curve Evaluation is to evaluate curve quality (that is, correlation coefficient). The evaluation result
will be displayed in the Preferences Window of Quantitative Window. If you need to evaluate calibration
curve, you can set Curve Evaluation Option as R2, Correlation Coefficient. Then click on Setting Button to
popup the Curve Evaluation Setting Window, as shown in Figure 6‐5.
In the setting window, you can set Evaluation Value and Positive Deviation as 1.0 and 0.0. As the
correlation of calibration curve is expressed with percentage, so it should take 100% as evaluation value.
Then the correlation will never be larger than 100%, and Positive Deviation will never be generated. So
Positive Deviation is set as 0.0. Negative Deviation means the negative deviation, which is allowed to
generate on the basic of evaluation value. For example, you can set Negative Deviation as 0.001, which
means that your evaluation lower limit is 1.0‐0.001=0.999. Your evaluation range is 1.0‐‐0.999. The values
out of this range are deemed as unqualified data.
Graphic Character is used for display of evaluation result. Deviation Number means number of
displayed characters. For example, you set evaluation value as 1.0, positive deviation as 1.5, negative
deviation as 0.5, graphic character as “*”, and deviation number as 5. Then if you evaluate the value of 0.5,
49
UVWin6 Software Operation Instruction Manual
you’ll get one of “*”, while if you evaluate the value of 1.0, you’ll get five of “*”. It is because that you set
deviation number as 5, which indicates that the difference of positive deviation and evaluation value, and
the difference of negative deviation and evaluation value, are averagely divided into 5 sections and each
section is expressed with one of “*”. Shown as the bellow.
0.5 0.6 0.7 0.8 0.9 1.0 1.1 1.2 1.3 1.4 1.5
* * * * * * * * * *
The evaluation procedure is as follows: first to judge whether the evaluated value, comparing to
evaluation value, is positive deviation, or negative deviation. That is, the one larger than evaluation value
is positive deviation, while the one smaller than evaluation value is negative deviation. After confirming it
is positive, or negative, it starts evaluation. The evaluation process is actually to see how close the
evaluated value is apart from evaluation value. More closer, more “*” it gets. For example, combining
with the above figure, the procedure of evaluating the value of 0.85 is: as judging that 0.85 is positive
deviation, so taking 0.5‐‐1.0 as evaluation range. First, seeing if 0.85 is between 0.5 and 0.6. If not, adding
one of “*”. Then seeing if it is between 0.6 and 0.7. If not, adding another one of “*”. Continue the
procedure in the same. When the judgment goes to the range of 0.8 to 0.9, it ends, because 0.85 is
between 0.8 and 0.9. Thus, there are totally 4 of “*”. Of cause, the whole procedure is accomplished by
System automatically, without any exterior calculation. You need not care the detail. It is well if you
understand the basic function of curve evaluation, that is, more “*”, more better it proves the curve
correlation is. Of cause, these “*” will be displayed in the Preferences Window of Quantitative Window
and can be printed out with calibration curve.
Calibration Method
Calibration method is divided into two types: Concentration Method and Coefficient Method.
Concentration Method requires you to input concentrations of standard solutions. After absorptions of
standards are measured, the measured data, combining with input concentrations, are fitted to compute
factors of equation and correlation. Coefficient Method would require you to input directly the factors of
equation, without measurement of standards. For example, if you set equation order as second, and
select coefficient method, then you need to input three factors of K0, K1, and K2. K0 expresses constant,
and K1 expresses the factor of first order, and K2 expresses the factor of second order.
50
UVWin6 Software Operation Instruction Manual
Calculation formula
In “Calculation formula” edit box, you can input required formula and unit for unknown sample’s
concentration. In formula, Conc. represents concentration of unknown sample. For example, the result
you got take mg/l as unit, but the unit you need is g/l, then you should input Conc/1000 in the
“Calculation formula” edit box, and g/l in the “Result unit” edit box. Then press Add Button. The default
title of calculation formula is “Result 1”, “Result2”… If you want to specify a title, you can input it, at same
time when you input calculation formula, in Title Edit Box. If you need to amend a formula, select it
correspondingly in formula list, and revise it in formula edit box. Press Revise Button to make it effective.
If you want to delete or clear the content of formula list, press Delete Button or Clear Button. Number of
Formula can be up to 10.
If the input type is “Volume”, also you can add formula about “mass”.
Characters
The role of Characters here is for input imitating as keyboard. Press a character button so as to input
the corresponding character equivalently.
Display
The role of Display option here is for selection of different display modes for calculation formula.
Pull‐down box offers two choices, formula and title. Formula here means calculation formula will be
displayed in result table as title. Title here means the default title or user set title will be displayed.
51
UVWin6 Software Operation Instruction Manual
6.3 Quantitative
The Quantitative Measurement Window is divided into 4 parts: Standards Table, Unknown Samples
Table, Calibration Curve, and Preferences. Shown as Figure 6‐7.
Calibration Curve
Standards Table
Preference
Samples Table
Measurement Table
Standards Table or Samples Table is used to display measuring results of standards or samples.
Generally, the table holds 8 rows: Number, ID, Type, Concentration, Absorption, and Absorptions at
Wavelength, Standard Deviation, and RSD. If you choose the double wavelength method, or the three
wavelength method, there will be displayed other wavelength points in the table.
Number: Indicating the number of standards or samples.
ID: The identifier for standards or samples, which can be edited.
Type: For standards, the type is Standard. For unknown samples, the type is Unknown.
Concentration: The concentration values of standards or samples.
Absorption: The final absorption values of standards or samples. The calculation formula is
shown as Table 6‐1.
52
UVWin6 Software Operation Instruction Manual
In the table, “Abs” expresses the final absorption, “W1” expresses first measuring wavelength, also
called main wavelength. “W2” is second measuring wavelength, also called baseline wavelength1. “W3” is
third measuring wavelength, also called baseline wavelength2. “A1” is the absorption of main wavelength.
“A2” is the absorption of baseline wavelength1. “A3” is the absorption of baseline wavelength2. “K”
expresses the coefficient of Double Wavelength Coefficient Method.
Absorption at wavelength: The absorption of standards or samples at corresponding
wavelength points.
Standard Deviation: The Standard Deviation of repeat measurement.
RSD: The RSD of repeat measurement. If repeat measurement is not checked, Standard
Deviation and RSD are both zero.
When performing standards measurement or unknown samples measurement, click on
corresponding measurement table. At this time, the top of table displays Activated, which shows the table
is enabled and all measurements are operated in the table.
Calibration Curve
Calibration Curve is used for graphic display of current calibration curve. If you do not perform
standards measurement, there is no graph displayed.
Preferences
Preference is used for display of current quantitative settings, including curve equation, coefficients,
and etc.
Measurement
If you set calibration method as concentration method in the quantitative settings, before
measurement, you should have to input standards concentrations in the standards measurement table,
and only then you are able to start measurement. There are two ways to input concentrations. First one:
Double click on the blank box of ID Row. Input an ID and press Enter Key. System would add a new
standard column in the table automatically. Then click on the concentration box of the standard and input
its concentration. Second one: Not to input any concentration, but click directly at Start Button. At this
time, System will prompt you input a concentration, as shown in Figure 6‐8.
After input of concentration, click on OK Button. System will add a standard automatically, whose
concentration is what you input, and then start measurement of the standard.
If you set calibration method as coefficient method in the quantitative settings, you need not to
measure standards. You can start measurement of unknown samples directly.
53
UVWin6 Software Operation Instruction Manual
The data of quantitative are sure to be saved to disk. Select Save Submenu under File Menu, or click
on Button to open the File Save Window. In the window, input the filename you want to save to and
click on Save Button. If you want to add notes to the measurement file, click on Note Button to input
comments and press OK Button.
This chapter mainly introduces you the usage of quantitative measurement. For UV‐VIS
Spectrophotometers, the quantitative is comparatively important function. Correct application of
measurement methods in quantitative would be decisive of your analytic work.
54
UVWin6 Software Operation Instruction Manual
Chapter 7 Kinetics
Key points
In this chapter, we’ll tell of the following contents: What’s kinetics? How to setup parameters for
kinetics? How to perform kinetics? How to save?
It includes:
Brief introduction of Kinetics
Setting of Kinetics parameters
Kinetics
Saving time course curves to files
Chapter Summary
Kinetics is a measurement of continuous reading at a certain time interval and graphic display of
measured data. The method is mainly used for observation of sample change trend with time.
To enable the Kinetics Window, you should click on the Kinetics tab in Work Space. Select
Parameters settings submenu under Measure Menu, or click on button, to activate the Kinetics
setting dialog box, as shown in Figure 7‐1.
55
UVWin6 Software Operation Instruction Manual
In this dialog box, there are 3 tabs: Measurement, Instrument, and Accessories. Among them,
Instrument Tab and Accessories Tab are as same as those of Instrument Performance and Accessories
Setting. Please refer to 3.1 Instrument Performance and 3.7 Accessories Setting for the details. The
following will focus on Kinetics parameters setting.
Wavelengths
Setting wavelength points at which Kinetics runs. As Kinetics takes sample readings continuously at
determinate wavelength points, so you should specify a wavelength for data reading.
Display Range
Setting the Y‐axis range for Kinetics. You can set the display range according to the values created by
samples.
Time Option
Time Option allows you to specify for kinetics the parameters of measuring time, sampling number,
and time interval.
Time Unit
Setting the minimum time unit for Kinetics. The optional time units are: Minute and Second. For
example, you specify the time interval as 1, the time unit as minute. It indicates that reading will be taken
56
UVWin6 Software Operation Instruction Manual
one times per minute. If the time unit is set as second, it means that reading will be taken one times per
second.
Display Mode
If you specify the time unit as minute, you are able to select the display mode. That is, during
measurement procedure, if you select Regular, System will display time in regular way of number, like 50,
120, 300, and etc. If you select Clock, System will display time in clock way, like 00:00:50, 00:01:00,
00:05:00, and etc.
Automation Mode
Automation Mode lets System select for you an appropriate set of the sampling number and the
time interval. What you need to specify is only Measuring Time. If you do not check the Automation
Mode, you need to specify the sampling number and the time interval either.
Measuring Time
Measuring Time is the total time over which kinetics runs. The option is only usable when the
Automation Mode is specified.
Sampling Number
Sampling Number is the total number of readings, which Kinetics will take. The sampling number
multiplying by the time interval is equal to the total time over which Kinetics will run.
Time Interval
Setting the time interval for readings. It is also called as Sampling Frequency.
Measurement Mode
In Kinetics, you are also able to perform repeat measurement. You can select the Measurement
Mode Tab to set parameters for repeat measurement. Shown as in Figure 7‐2.
Single Measure
No repeat. That is, the Kinetics run is performed just once.
Repeat Measure
Here is Able to set Repeat Times and Time Interval. System would perform repeat measurements
according to your settings of Repeat Times, and each run is separated with a determinate time, which is
decided by Time Interval. If your instrument equipped with an auto multi‐cell holder, you can also select
Cell Holder Interchange Automatic. In this case, System will switch automatically to next cell and then run
57
UVWin6 Software Operation Instruction Manual
Time Increment
If you select repeat measure or automatic measure, you can also set another option of Time
Increment. The role of the time Increment is adding a time increment to the total measure time.
Start: It expresses the repeat times at which the time increment starts to be effective. That is,
the time increment starts to be added at which times that the repeat measurements have been
performed.
End: It has the same meaning as that of Start. End expresses the repeat times at which the time
increment is to be ineffective.
Increment: It expresses the time being added each time. The time is the percentage relative to
the last measure time.
7.3 Kinetics
If you are going to run kinetics, you can select Start Submenu under Measure Menu, or click on
Button to commence the measurement. During the measure procedure, System would display
results graphically, as shown in Figure 7‐3.
The time curves of Kinetics can be saved as well. Selecting Save Submenu under File Menu, or click
on Button to open File Save Window. You can input the filename you want to save to and press Save
58
UVWin6 Software Operation Instruction Manual
This chapter mainly gives you introduction of some functions of Kinetics. Kinetics is a good method
to observe the samples changes with time. Reasonable application of Kinetics would help your work much.
Furthermore, you can also realize various analysis by combining Kinetics and Graphic Process, which is
going to be introduced in next chapter.
59
UVWin6 Software Operation Instruction Manual
Key points
In this chapter, we will discuss the following content mainly: what’s called spectral bandwidth
scanning, how to set parameters for spectral bandwidth scanning, how to do spectral bandwidth scanning,
how to find the best spectral bandwidth and how to save spectral bandwidth files.
Detailed content includes:
Brief introduction of spectral bandwidth scanning
Spectral bandwidth scanning parameters setting
Spectral bandwidth scanning
Find the best spectral bandwidth
Save spectral bandwidth files
Chapter summary
When analyzing samples by the spectrophotometer, the accuracy of result is mainly affect by the
spectral bandwidth value, choosing the best spectral bandwidth is one of the main factors to obtain the
real absorbance and transmittance. The standard of best spectral bandwidth is to get the maximum
absorbance at where, or the minimum transmittance. Spectral bandwidth scanning is a measure method
that to do the spectrum scanning continuously according to some spectral bandwidth interval, and display
these spectrums together, and find the spectral bandwidth value with maximum absorbance and
minimum transmittance.
If the spectral bandwidth scanning is needed, firstly set the scanning parameters. Click on the
Spectral bandwidth scanning tab in Work Space. Choose [Parameters] from [Measurement], or click
button to open the spectral bandwidth scanning parameters setting, as figure 8‐1 shown.
60
UVWin6 Software Operation Instruction Manual
Measurement tab
Photometric mode
Set photometric mode of spectral bandwidth scanning. You can specify the Photometric mode as one
of Abs. (Absorption), T% (Transmittance. System will perform measurement according to your
selection of photometric mode.
Display range
Setting the Y‐axis range for spectral bandwidth scanning. You can input the [Max.] and [Min.] value
according to the values created by samples.
Scan parameters
Setting scanning parameters of wavelength range, wavelength interval, speed, and etc. The setting is
the same as spectrum scan parameters, please refer to 5.2 Spectrum scan parameters setting for
more information.
Scanning Mode
Accessories tab
Please choose fix cell holder.
61
UVWin6 Software Operation Instruction Manual
Choose [Start] from [Measurement] menu, or click button to start scan. In the procedure of
scanning, System will draw dynamically scanning data and wavelength on the Spectral bandwidth Scan
Window in graphic mode. As figure 8‐2 shown.
The best spectral bandwidth is at where the absorbance is the maximum or transmittance is the
minimum when measure the specified sample at given wavelength. Shown as figure 8‐3, 1.5nm is the best
spectral bandwidth, because the absorbance is as high as 1.093 at the located wavelength.
Choose [Read Spectrum] mode, locate the lines at that wavelength that you want to check the Abs, it
will show you the correspondence spectral bandwidth and Abs value in checking list form. If the list is too
long to display all of them, you can drag the mouse to change the position in the list area to check all the
information. Shown as figure 8‐3.
62
UVWin6 Software Operation Instruction Manual
The Spectral bandwidth scanning files can be saved as well. Selecting Save Submenu under File Menu,
or click on Button to open File Save Window. You can input the filename you want to save to and
press Save button to save it.
8.6 Application
There is a character absorbance peak at 264nm of benzylpenicillin potassium, so the analyzer is care
about the absorbance at this wavelength when measuring it. Here is an example about how to use
spectral bandwidth scanning when analyzing benzylpenicillin potassium.
Instrument: T92+ Double beam UV Spectrophotometer,
Sample: Benzylpenicillin potassium
Spectral bandwidth range: 0.1nm‐5nm take 0.1nm as interval
Here is the operation procedure:
① Power on the instrument
Power on the computer firstly, confirm there is nothing block the light in the sample compartment,
power on the T92+, start the UVWin6 control software, after initialization it will come to the main
interface of software, warm up 60 minutes.
② Reagent preparation
Weight 0.0470g benzylpenicillin potassium and dissolved into 25mL pure water, then the
concentration is 1.88mg/mL.
③ Parameters setting
Choose “Spectral Bandwidth scanning” mode, enter into parameters setting page, set as following:
63
UVWin6 Software Operation Instruction Manual
Figure 8‐1
Measure
Photometric mode Abs
Display range 0.0~1.0
Wavelength range 260nm~270nm
Scanning interval 0.1nm
Scanning speed Slow
Instrument
Spectral bandwidth 0.1nm~5.0nm
Bandwidth interval 0.1nm
④ Sample measurement
Put the sample into Sample cell holder, and the blank into Reference cell holder. Press “Start” button
to scan. Now the measurement interface will prompt scanning and show the current process, also the
spectrum will be shown. Please wait until the scanning finish automatically with patient, it will take long
time. When scanning it’s not allowed to do the other operation except stop scanning, press “Stop” button
if needed, the spectrum that got before stop will be saved.
⑤ Result analysis
Choose [Read Screen] to locate at 264nm, and the Abs value of different bandwidth will be got, and
shown as list from 0.1nm~5.0nm. (Shown as figure 8‐4, the Abs at 0.5nm is 0.8690 which is the maximum,
so 0.5nm is the best spectral bandwidth for sample analysis).
This chapter mainly discusses the spectral bandwidth scanning function of UVWin6. It’s the good
64
UVWin6 Software Operation Instruction Manual
method to find the best spectral bandwidth for different sample, save a lot of time for fast degradation
speed sample especially.
65
UVWin6 Software Operation Instruction Manual
Key points
In this chapter, we will discuss the following content mainly: the use of DNA protein determination
function, how to set measure parameters, how to measure and how to process the measured result.
Detailed content includes:
The use of DNA protein determination function
Start DNA protein determination function
Set measure parameters
Process measurement result
Chapter Summary
Introduction: As the internal structure of DNA and the secret of genetic mechanism are gradually
exposed, especially when people have learnt that genetic code is transcribed and expressed by RNA,
biologists are no longer satisfied with exploring and suggesting the secrets of biological inheritance, but
they are trying to intervene the generic characters of organism at the molecule level. If a generic code
segment from the DNA of an organism is connected to the DNA sequence of another kind of organism,
and after reorganization of DNA, a new generic material will be designed according to human desire, and
even a new organism type will be created. This process is totally different with traditional method of
cultivating organisms to reproduce offspring. According to human demand, “rebuilding” the “gene” of
one organism with the “gene” of another to create a new combination of genes and create a new
organism, the approach is just like engineering design in technological science. Such a biological
technology of creating new organism from reassembling genes according to human desire is called “gene
engineering”, or “genetic engineering”. As an important branch of biological engineering, gene
engineering, together with cell engineering, enzyme engineering, protein engineering and Microbiological
engineering, forms biological engineering.
From the above introduction, it can be easily seen that gene analysis is a complex yet significant job.
Gene engineering plays an important role to all human kind. Study on human genes has become an
important scientific research subject for developed countries. In gene analysis, analysis tools play an
important role. Only with good analysis tools, can one accomplish the analysis successfully. Therefore, in
software UVWin6, DNA/protein determination function is added. Users can measure the density of
DNA/protein, and different analysis methods can be set to satisfy different demands.
If you want to analyze DNA protein, you can choose “DNA protein analysis” submenu under
“application” menu, then the system will pop up a window for DNA protein determination. As shown in
figure 9‐1, the window is divided into two parts. The central part is the display window, showing the
66
UVWin6 Software Operation Instruction Manual
current measured result. The bottom part is a control panel, where you can start functions by clicking
corresponding buttons.
In buttons on the control panel, there is a button named “setting”. You can open parameter setting
window as shown in figure 9‐2 by clicking this button.
67
UVWin6 Software Operation Instruction Manual
Density of protein=Pf2×A2-Pf1×A1
With 320nm correction
(A1=A1-A320, A2=A2-A320)
Density of DNA=Df1×A1-Df2×A2
Density of protein=Pf2×A2-Pf1×A1
The measurement of DNA protein is quite simple. You simply put samples into the sample pool, then
click “measure” button on the control panel, and the measurement is finished. If you want to print the
measurement result, click “print” button on the control panel. If you want to export measurement result,
you can click “export” button.
This chapter mainly discusses the determination method of DNA protein. You can set parameters
according to specific requirements.
68
UVWin6 Software Operation Instruction Manual
Key points
In this chapter, we’ll tell of the following contents: pesticide residual introduction, file operation,
parameters setting, how to do pesticide residual analysis and report printing.
It includes:
Use of Pesticide Residual determination
Start Pesticide Residual determination
File operation
Pesticide measurement
Printing report
Chapter Summary
It’s forbidden to use the pesticide that contains organic phosphor and carbamate for vegetable crop.
The Pesticide Residual determination function of UVWin6 can detect organic phosphor and carbamate
residual of vegetable very quickly, in order to control the vegetable with high pesticide residual coming
into the market, to ensure the safety of our food.
This method is according to the inhibition ratio of enzyme to determine the organic phosphor and
carbamate in vegetable which is based on the National Standard. The principle is: organic phosphor and
carbamate pesticide will inhibit the normal function of cholinesterase, there is some positive relativity
between inhibition ratio and pesticide concentration. Normally enzyme can catalyze the hydrolyzation of
nerve conduction metabolites, and the hydrolysate will react with developer and produce yellow color
matter. So the UV spectrometer can be used to monitor the changing value of Abs according to the time,
and calculate inhibition ratio. Then the inhibition ratio can be used to determine if there is organic
phosphor or carbamate pesticide exists or not. The detection limit of this method is: methamidophos
0.8mg/L, dipterex 0.25mg/L, carbofuran 0.002mg/L.
Start UVWin6 after connecting with instrument, choose “Pesticide Residual Determination V1.0”
from “Application” to enter into the main interface shown as figure 10‐1. On the left is the function
button panel; on the right is sample information list. For the function button panel there are “File”,
“Measurement” and “Print” three area. Next is the introduction of these three function.
69
UVWin6 Software Operation Instruction Manual
Operator
Click this button, then input the name of operator in the textbox on the right as one of the content of
current file.
Open
Open the pesticide residual determination result file (end as “.pmd”). If there is some sample
information in the current list, there will be a prompting information said “Confirm to delete the current
sample”, and it will be overwrite if click OK.
Save
The data will be saved as .pmd file into the hard disk.
button to start measurement, there will be a Timer display on the right of button. Stop measurement to
press “Measure Reference” button. After 3 minutes, the Timer is stop and measurement is finished. If the
difference of Abs between start and end bigger than 0.3, that’s meaning measurement is passed, the
result will be displayed in the list as reference value. If smaller that’s meaning it’s failed, also warning
information will be prompt, please re‐prepare the reference sample.
Sample measurement
Take the reference solution out, and put the sample that prepared according to pesticide residual
determination operation procedure into cell holder in turn, press “Measure Sample” button to start
measurement, and after 3 minutes the result will be displayed in the list which is the same as reference
measurement.
Page setting
Adjust the layout of print page.
Preview
Preview of the report.
Print
Print out the report.
This chapter mainly gives you introduction of Pesticide Residual determination, set the parameters as
your requirement.
71
UVWin6 Software Operation Instruction Manual
Key points
In this chapter, we’ll tell of the following contents: Purpose of graphic process. How to perform
graphic process on spectrum? How to perform simple calculations? How to perform complicated
calculations?
It includes:
Use of graphic process
Analysis of spectra
Creating of 3D spectrum
Display settings
Chapter Summary
With the uninterrupted development of computer technologies, it has melted into many fields.
Manufacturers can write more professional processing software with the aid of PC high operation speed
to make analysis more convenient, more timesaving. With the dozen’s development of UV‐VIS
spectrophotometers, the early instruments performed processing of measured spectra by means of
electric devices and plotting equipments. Today, with fast‐developed computer technologies, we can do
any graphic process with PC’s powerful abilities.
72
UVWin6 Software Operation Instruction Manual
Display Gridding
Setting display or hiding of gridding line. If you check the Display Gridding, you are able to specify the
type of grading line.
Display Scale Line
Setting display or hiding of scale lines of coordinate axis. If you check the Display Scale Line, you are
able to specify whether the scale lines are drawn inside or outside the coordinates frame. The length of
scale lines can also be set.
Automatic Separation
Setting the way separating grading lines. If you check the Automatic Separation, System will
distribute the grading lines automatically. If you want to specify a separation value, you can uncheck the
Automatic Separation and set the Minimum Interval. The Minimum Interval is input as percentage, that is,
the percent of each gridding vs the whole coordinate axis. If you set the Minimum Interval as 10%, System
will draw 9 gridding lines to divide the whole coordinate axis into 10 sections. If you set the Minimum
Interval as 20%, System will draw 4 gridding lines to separate the whole coordinate axis into 5 sections.
Color
Setting the color of gridding lines and scale lines.
Central Color
Setting the color of central graph.
Frame Color
Setting the color of graph frame section, that is, displayed coordinate values and its title.
Preview
If you want to preview what it is like while you is changing settings, you can check Preview. System
will offer live preview on the graph.
73
UVWin6 Software Operation Instruction Manual
Default
Restore the default settings.
Measure Distance
Select the Measure Distance submenu under the Graph menu, or click on Button. The system
cursor will show as . At this time, you can click the mouse left key at any position (P0) on the graph.
Then move the mouse to another position (P1) and click the mouse left key again. System will give you
automatically the calculation result of the line distance, the vertical distance, and the horizontal distance
of the two points of P0 and P1, as shown in Figure 11‐3.
Fit in Window
Select the Fit in Window submenu under the Graph menu, or click on Button. System will
automatically regulate the coordinate range of spectrum graph to bring all spectra displayed normally.
Peak‐Picking
Peak‐Picking detects all peaks and valleys on a spectrum and labels those peaks and valleys
accordant with picking conditions. Opening a scan spectrum and selecting the Peak‐Picking submenu
under the Graph menu, or clicking on the Button, System will automatically detect peaks and valleys
on the spectra and show the results in a table, as shown in Figure 11‐4.
Valley mark
Peak mark
Result list
Settings bar
After peak picking, spectrum window is divided into three parts: Spectrum window, Result list, and
Settings bar. In the Spectrum window, it shows marks of each picked peak and valley. The result list
displays the wavelengths, measured values and comments of each picked peak and valley. You can
improve the effect of peak‐picking through amending parameters in settings bar, as shown in Figure 11‐5.
Threshold: Setting minimum percent limit of height of picked peaks and valleys, which is the
percentage relative to the difference between the maximum value and the minimum value of a spectrum.
Those peaks and valleys of height larger than the percentage are picked out. You can input directly a
threshold in the Threshold Edit Box and press Return, or click on the ▼ button right to the edit box. In the
poll‐down menu, select a relevant threshold.
Display: Setting display only of peaks or valleys, or both of them.
Marks: Setting mark type of picked peaks and valleys as value or number. The mark value expresses
that of peaks or valleys. The number is serial number of peaks or valleys.
Direction: Setting marking direction. The optional direction is Vertical or Horizontal.
Frame: Setting whether marks have a frame or not.
Marking Note: Using the contents in the Comment Column to replace that of marks of peaks and
valleys. You can input the comment message for each of peaks and valleys and then check the Marking
75
UVWin6 Software Operation Instruction Manual
Note option.
Read Spectrum
The function of Read Spectrum is using a vertical line to do scan on a spectrum and then get readout
of data at the cross point of the vertical line with the spectrum. You can select the Read Spectrum
submenu under the Graph menu, or click on the button. System will draw a vertical line on a
spectrum. You can use mouse to drag the line, or use keys of ← and → to move the line, to read out
spectrum data.
Read Screen
Read Screen is getting readout of data at which mouse points at. Select the Read Screen submenu
under the Graph menu, or click on the button. System will change cursor as +. In the meantime, you
can move cursor on spectral graph and System will show the corresponding coordinate values.
Display Cursor
Display Cursor has similar effect as that of Read Spectrum, that is, reading out of coordinate values.
But cursor will be showed as cross cursor. Select the Read Screen submenu under the Graph menu, or
click on the button. System will draw a cross cursor in the center of spectral graph. You can move
mouse to the center of cursor and the cursor will be displayed as . Click mouse left key to drag the
cursor moved. Click again to fix cursor at the appointed position.
Horizontal Stretch, Horizontal Shrink, Vertical Stretch, Vertical Shrink
Stretch and Shrink are zoom on coordinate axes separately. You can select the Scale Zoom submenu
under the Graph menu and choose the zoom action you require in the submenu. Or you can click on
corresponding button on toolbar: ‐‐ Horizontal Stretch, ‐‐ Horizontal Shrink, ‐‐ Vertical
Stretch, and ‐‐ Vertical Shrink.
Curve Display
Curve Display offers you the various options for curve display. The optional display modes include:
line, point, cross point, dot, and triangle point. You can select the Curve Display submenu ender the Graph
menu and choose one of display modes. Or you can click on corresponding button on toolbar: ‐‐ line,
‐‐ point, ‐‐ cross point, ‐‐ dot, and ‐‐ triangle point.
Combination Display
Combination Display allows you to display multiple spectra in a graph. If your graph holds over two
spectra, you can select the Combination submenu under the Graph menu and choose the Build
Combination in the submenu, or click on the button to open the Combination Spectrum Window, as
shown in Figure 11‐6.
In the window, you can select the spectra being combined and press OK Button. If you want to delete
a combination display, you can select the Delete Combination next to the Build Combination, or select the
Delete All Combination.
76
UVWin6 Software Operation Instruction Manual
Basic Calculation
Basic calculation includes Addition, Subtraction, Multiplication, and Division. You can select the
Addition submenu, the Subtraction submenu, the Multiplication submenu, or the Division submenu
under the Mathematic Calculation menu to open setting window, as shown in Figure 11‐7.
77
UVWin6 Software Operation Instruction Manual
Advanced Calculation
Advanced Calculation mainly provides operations of Smoothing, First Order Derivative, Second
Derivative, Third Derivative, and Fourth Derivative. You can select the Smooth submenu or the Differential
submenu under the Mathematic Calculation menu. System will open the Math Calculation Setting
Window, as shown in Figure 11‐8.
Additive Calculation
Additive Calculation mainly provides operations of Logarithm, Reciprocal, and Transform between
Abs and T%. You can choose the Logarithm submenu, the Reciprocal submenu, and the Transform
submenu under the Mathematic Calculation menu. System will open a setting window, as shown in Figure
11‐9.
78
UVWin6 Software Operation Instruction Manual
11.2.4 3D Spectra
3D Spectra is a new adding function in UVWin6. Its role is combining multiple spectra into 3D display.
In this way, it can perform qualitative analysis on samples through graph observation. If you have scanned
several spectra, or have recalled some spectra, you can select the Establish 3D Spectrum submenu under
the Graph menu, or click on the button to open the 3D Spectra Window, as shown in Figure 11‐10.
Menu and Toolbar
Spectra Tree
Graph Display
Cascade Display
79
UVWin6 Software Operation Instruction Manual
The 3D spectra window is divided into 4 parts: Menu and Toolbar, Spectra Tree, Color Cascade, and
Graph Display.
Spectra Tree
Spectra tree is used for displaying the names of current spectra. You can display or hide spectra by
selecting check boxes before spectral headers.
Peak‐Picking
If you want to perform peak‐picking on 3D spectra, select the Peak‐Picking option, and then input a
detection threshold in the Edit Box and press the Pick button. System will perform peaks detection
automatically and display the detection results in the peak‐picking table. Shown as Figure 11‐11. If you
want to close peak‐picking, check the Close Display option. System will close the display of labeling 3D
spectra peaks.
Layering Display
Layering display is separating 3D spectra into several layers by different colors so as to let you
observe the change trend of spectra more clearly. Here 3D facility offers five layers and you can setup
according to your requirement. Shown as 11‐12.
81
UVWin6 Software Operation Instruction Manual
In layers setting window, you can set on/off for each layer. Also you are able to specify value range
and color for each layer. Each layer holds an upper limit, a lower limit, and a color to be set. For example,
the first layer in the above figure holds a lower limit as 0.000, an upper limit as 30.0000, and a color as red.
It expresses that in 3D spectral graph, the data of bigger than 0 and smaller than 30 are considered as the
first layer and are displayed in red. The second layer, which is located above the first layer, has its lower
limit as the upper limit of the first layer and its upper limit is 60.000, its color is yellow. This expresses that
the data of bigger than 30 and smaller than 60 are considered as the second layer and are displayed in
yellow. Accordingly, similar are the rest up to the fifth. If you want to set the layer’s color, click on the
color tab. The bottom color tab expresses that of those data that are smaller than the lower limit of the
first layer, while the top color tab expresses that of those data that are bigger than the limit of the fifth
layer.
Layering not only can be lengthwise, but also transverse. You can choose the Vertical button or the
Horizontal button for the Layering direction. In addition, click on the Reversion button to reverse the
order of color tabs.
This chapter mainly gives you some introduction of graphic process in UVWin6. Graphic process
facility could help you perform some professional analysis. Appropriate application of this facility would
be much of assistance to your work.
82
UVWin6 Software Operation Instruction Manual
Key points
In this chapter, we’ll tell of the following contents: What’s Data Export? What kind of data can be
exported? How to perform data export? How to perform data export? How to handle the exported data
files?
It includes:
Brief introduction of Data Export
Data Exportable
Data Export
Handling exported data files
Chapter Summary
Whatever to perform chemical analysis or to do scientific research, Data are all an important concept.
Only with data, you can do analyzing and studying. Furthermore, sharing data has been becoming a giant
trend of Today’s scientific and technical development. With the uninterrupted development of computer
technologies, data sharing is being realized. For example, a tabular we make with Excel can be insert into
Word for purpose of data sharing. And sometimes we also need to share the measured data or analyzing
data with other editing software. As general analyzing software do not offer such data sharing functions,
it is very difficult to perform such data sharing. UVWin6 software achieves data sharing with editing
software of Excel and Word by its Export function based on Windows OLE, which makes your analysis
more easy and convenient.
Tips: “OLE”
OLE is the abbreviation of Object linking and embedding. It is a standard interface for programming
supplied by Windows Operating System. An application program can use the interface to access to those
applications that offer OLE function. As Word and Excel supply OLE function, so other applications can
operate on them through OLE.
In UVWin6, data able to be exported are mainly measured data. Of cause there are some special data
that are exportable, like instrument information. You can select the last submenu of Information under
the Measure Menu to open the Instrument Info Window, as shown in figure 12‐1. In the window, it shows
the information about current instrument hardware, including: Model, Serial Number, Wavelength Range,
Spectral Bandwidth. In the lower left corner of the window, there is an Export Button. Pressing the button
will export the information in the window into relevant files.
83
UVWin6 Software Operation Instruction Manual
Export Button
Exporting data is easy and simple. You only need to select the Export Data Submenu under the File
Menu. System will open the Data Export Window, as shown in Figure 12‐2. In the window, you can specify
the file type for export through the option of Export Type (please refer to Table 12‐1 for exportable file
types). In the edit box of Export File, input the filename being exported, or click on the “…” button right to
it to specify a file. After setting, click the Export button. System will export all data to the appointed file
according to your settings. If you check the option box of Open Files Automatically After Export, System
will open the exported file automatically after export finishes.
84
UVWin6 Software Operation Instruction Manual
Once data are exported, it would be flexible to handle them. For example, if you need to compose
some papers and you think that the data reports printed out with UVWin6 do not meet your requirement
about their format, you can export the data to a Word file. Then you edit the word file, as you want. In the
same way, for those requirements of complicated calculation, you can export data to an Excel file. Then
you specify calculations on the data.
This chapter mainly gives you introduction of the Export function of UVWin6 and its usage. It is sure
that you have learned about it to a certain extent. And hope the function would afford facilities for your
work.
85
UVWin6 Software Operation Instruction Manual
Chapter 13 Administration
Key points
In this chapter, we’ll tell of the following contents: Why should we have the Administration Facility in
the software? How to setup users and groups? How to set security parameters? How to setup and browse
the operation log file?
It includes:
Brief introduction of Administration Facility
Create a user account
Create a user group
Set security parameters
Setup and browse the operation log file
Chapter Summary
The users of Microsoft Windows should know that, in Windows system, you are able to setup
different users to log on the system. And you can use Administrator to create users with different
privileges. Actually this is an embodiment of Administration Facility. For more details, please refer to the
related books of Windows.
Administration Facility is to make systematic control on software or operation system with some
administrating methods. Its main usage is multiple users management mentioned above. In UVWin6, it
includes Multiple User Management, Security System, Log Management, and etc. All these functions
would be helpful for you.
User Management is a basic administration function in UVWin6. This function will assist you in
creating users and groups and setting relevant parameters. You can select the User Group submenu under
the Administrate menu to open the Management Window. In the window, there are two tabs: Users Tab
and Groups Tab. You can create users and change users’ settings in the User Tab, and create groups and
modify groups’ settings in the Group Tab, as shown in Figure 13‐1.
86
UVWin6 Software Operation Instruction Manual
If you need to create a password for a new user, click on the Amend Password Button. System will
open the Password Window, as shown in Figure 13‐3. What You need to do are only to enter the
password you want in the Password Box and the Confirm Password Box, and then click OK. If you want to
prompt the user to change its password on its next sign in, you can check the option box of User Should
Amend Password On Next Sign‐In. If you want to terminate the user account, check the option box of
Account is terminated.
88
UVWin6 Software Operation Instruction Manual
Additionally, you can also input the relevant information of a new user in the Information Boxes. The
information includes: Full Name, Company, Section, and Description.
After you finish the setup of a new user, click on the Create Button. System will establish a new user
account according to your settings. Then you can log on with this user account by entering the user name
and its password in your next sign‐in.
Rename
The Rename function can change the name of current user. In the user list, specify the user being
renamed and click on the Rename Button. System will prompt you to input a new username. Shown as in
Figure 13‐4. After you enter a new name, click OK to return.
Remove
The Remove Function allows you to delete a specified user. In the user list, specify the user being
deleted and click on the Remove Button. System will prompt you to confirm the deletion. Select OK to
delete the user you specify. If the user you specify is the current log on user, System will prompt you
unable to delete. In this case, you need to use other user to log on to delete this user.
Copy
The Copy Function creates for you a copy of current specified user. Apart from the username, other
settings are the same as the user you specify.
Property
The Property Function allows you to view the user information. Click on the Property Button, and
System will open the User Attribute Window. The window is nearly same as that of creating a user. In the
window, you can modify the relevant information of a user.
89
UVWin6 Software Operation Instruction Manual
For a group, the most important is its privilege. As you may see in the New Group Window, there are
all managing permissions of UVWin6. You can setup a set of permissions for a group according to its
requirements. These privileges will be applicable for each user subject to this group. That is, if a user is
part of the group, when the user logs on, its useable function will be limited within the permissions of the
group. In the group privilege, those functions that are not specified are unavailable.
Rename
The Rename function can change the name of current group. In the group list, specify the group
being renamed and click on the Rename Button. System will prompt you to input a group username. After
you enter a new name, click OK to return.
Remove
The Remove Function allows you to delete a specified group. In the group list, specify the group
90
UVWin6 Software Operation Instruction Manual
being deleted and click on the Remove Button. System will prompt you to confirm the deletion. Select OK
to delete the group you specify. If the group you specify includes the current log on user, System will
prompt you unable to delete. In this case, you need to use other group user to log on to delete this group.
Copy
The Copy function creates for you a copy of current specified group. Apart from the group name,
other settings are the same as the group you specify.
Property
The Property Function allows you to view the group information. Click on the Property Button, and
System will open the Group Attribute Window. The window is nearly same as that of creating a group. In
the window, you can modify the relevant information of a group.
13.3 Security
Security is an important word for all things. Only on the premise of security being guaranteed, works
required can be done. With the development of computer technologies, the security of information
should not have been neglected any more. Therefore UVWin6 software introduces the concept of Security.
You can achieve data protection by security setting. Select the Safety Settings submenu under the
Administrate menu to open the Security Window, as shown in Figure 13‐6.
Never Expired. If you want a user to change his password each 30 days, then you should choose the other
option and enter the corresponding number of days in its edit box.
Password Length
The password length is used for setting the length of a user password. You can specify the maximum
length and the minimum length. In addition, if you do not specify the minimum length, the blank
password allowed is to be selected. In this case, a user can use a blank password, which is no password.
Security
The function of the Auto Login By Administrator is to start UVWin6 software without displaying the
login window. System will use the administrator account to log in. For the single user system, it is more
convenient without entering a password to log in.
If you do not check the option, it means you need to enter an account and its password for each
login. Then you are able to set the times for failed login. That is, System will count your wrong entering of
an account and its password. Once the times of your failed login reach the number you set, System will
exit automatically.
If you wish to lock the UVWin6 system after a period of time being idle, then check the Lock System
Option and specify a time for automatic lock.
File Protection
The function of file protection is to protect those files of measure data. Its role is that the data files
created by each user cannot be read each other. Only its creator or administrator can read and open a
measure data file.
The log function is to record all your operation. You can select the View Log submenu under the
Administrate menu to open the Operation Log Window, as shown in Figure 13‐7.
92
UVWin6 Software Operation Instruction Manual
You are able to select the type of log on the left window, and the details of log are displayed on the
right window. If you want to set the log parameters, you can select the Option submenu under the
Operation menu, System will open the Log Option Window, as shown in Figure 13‐8.
Save in: Setting the path for Log File saving. For example, you can save the log file in other hard disk,
to prevent from data lost.
Maximum Record Number: Setting the maximum record number allowed for Log File to save. When
the record number is reached, System will act according to your choice.
93
UVWin6 Software Operation Instruction Manual
Only the current user or the administrator can release the status of the system lock. If you enter the
wrong password, System is unable to release the lock. If you choose Cancel, System will exit automatically.
This chapter mainly gives you introduction of Administration in UVWin6. It is sure that you have
learned more about it. The Administration is a new function. Hope the function would afford facilities for
your work and make your software system safer.
94
UVWin6 Software Operation Instruction Manual
Chapter 14 Conclusion
Up to now, we have given you all information about UVWin6 functions. It is sure that you have
learned more about the software. It is certain that getting familiar with software needs time. Hope you
could know well it by reading the book and your practices. It would be our pleasure if the software would
provide you more convenience and assistance for your work.
95
Manufacturer: PG Instruments Ltd
Alma Park
Wibtoft
Leicestershire
LE175BH
United Kingdom
Telephone: 0044 1455 220131
Fax: 0044 1455 220025
Email: [email protected]
Web: www.pginstruments.com