AB-280_3
AB-280_3
Summary
Standards
In principle the gas extraction or oven technique can be used
with all types of samples which release the contained water Standards (with different water contents) suitable for the oven
when the samples are heated. The oven technique is technique are commercially available.
essential whenever direct Karl Fischer titration is impossible
because the sample contains interfering components or, due
Carrier gas
to its consistency, is difficult to place in the titration vessel.
The carrier gas transports the released water into the titration
This Application Bulletin shows examples from food,
vessel. In principle it is not important what gas is used.
pharmaceutical, plastics and petrochemical industry to
However, due to oxidation reactions with oxygen, the thermal
describe automated water content determination using the
stability of organic substances at higher temperatures is often
oven technique in combination with coulometric KF titration.
poor, when using dry air as carrier gas. Inert gases, usually
nitrogen, offer more flexibility and reliability.
Instruments For solids with a water content below 1% we recommend a
874 Oven Sample Processor gas flow rate of 50 mL/min. Gas flow rates of up to 100
mL/min can be used for solids with higher water contents.
KF Titrator (coulometric)
Higher flow rates do usually not lead to shorter determination
times.
Electrodes Measurements in liquid samples (oil, fuel, solvents,…) should
Double Pt wire electrode (for coulometry) be carried out using gas flow rates of 80 mL/min or higher.
Generator electrode with or without diaphragm (for Furthermore it is recommended that the inlet needle is
coulometry)* immersed in the sample and the gas is allowed to flow
* For measurements of samples with absolute water contents
through the sample.
< 100 µg, the generator electrode with diaphragm is recommended.
General
Reagents The combination of the oven technique with coulometric Karl
For the coulometric technique, special reagents suitable for Fischer titration is ideal for samples with low water content.
the oven technique are available. Foodstuff, pharmaceutical products, plastics or mineral oil
products can be analyzed fully automated and accurately.
Due to the constant gas flow during measurements, methanol
contained in the reagents evaporates. With the generator In accordance with the gas extraction principle the water is
electrode without diaphragm a decreasing methanol content driven out of the heated sample by a stream of dry carrier gas
can lead to false high results, with recoveries of up to 110%. and transferred to the titration vessel, where the water
content is determined.
Page 1 of 6
Application Bulletin 280/3 e
Automated water content determination with the 874 OSP
Page 2 of 6
Application Bulletin 280/3 e
Automated water content determination with the 874 OSP
Fig. 3: Sodium tartrate dihydrate before (left) and after (right) Fig. 5: Temperature gradient (0.07 °C/min) of sodium tartrate
running a temperature gradient. dihydrate between 170 and 220 °C.
The temperature curve can be used to determine the optimal Please note that the time to record a temperature gradient
oven temperature for extracting the water from the sample. increases if the heating rate is reduced. The temperature
This temperature should be high enough for the water to be gradient in figure 4 took more than 24 hours to finish.
extracted completely without any decomposition of the
sample. The determination time should be kept as short as
possible. Therefore the oven temperature should be chosen Analysis
as high as possible, but approximately 20 °C below the start Each analysis consists of the following steps:
of decomposition. This procedure works for many different • Conditioning of the complete titration system (titration
samples and can be used for a fast assessment of the oven cell and tubing)
temperature.
Page 3 of 6
Application Bulletin 280/3 e
Automated water content determination with the 874 OSP
water contentabsolute =
water contentsample + blank value + drift x determination time
Page 4 of 6
Application Bulletin 280/3 e
Automated water content determination with the 874 OSP
System preparation means that the whole system is adjusted Stop criterion rel. drift
to the selected conditions. An empty sample vial is treated in Relative stop drift 5 µg/min
exactly the same way as the following samples, but the value
Conditioning Start drift 10 µg/min
is not taken into account. We recommend that this step –
parameters
which is at the same time used for checking that the analysis
Stabilizing time 60 s
system is working properly – is carried out before every new
* 400 mA are used with the generator electrode without diaphragm.
sample series.
The setting auto is chosen for the generator electrode with
Apart from the water in the sample, the sample vial also diaphragm.
contains atmospheric humidity; this makes a blank value
determination necessary. A three-fold determination of the
Troubleshooting
blank value is normally sufficient. The mean value is stored
as a Common Variable and taken into account in the Procedure for poor precision (reproducibility):
calculation of the water content (subtracted). If large sample • Optimize the titration and control parameters.
sizes are used, it might be necessary to correct the blank • Check whether the sample vials are tightly sealed.
value (see KF Application note AN-K-048).
• Drift too high: Switch off the gas flow.
The system preparation and the blank value determination If drift value decreases: Check the needle system,
must be carried out using the same method parameters as transfer tube.
for the analysis of the samples. In order to do this, an empty If drift value does not change: check titration cell,
system preparation vial and three empty blank value vials are septum and/or seals leak, molecular sieve exhausted,
placed on the rack of the 874 Oven Sample Processor and poorly conditioned reagent, ensure thorough mixing.
analyzed before the samples.
• Clean electrodes according to the electrode leaflet.
End of titration • Check the needle system and possibly clean it with
The titration and the gas extraction of the sample is stopped water and methanol. Dry the components afterwards.
as soon as the drift value (amount of water per time) falls • Check the transfer tube and clean it with water and
below a predefined value. Usually the parameter “relative methanol to remove condensates and dirt.
drift” is used to stop a determination. The stop drift is
• KF reagent contaminated/exhausted: change the
calculated by adding the drift at the start of the determination
solution, possibly use a different batch.
and the value entered for the “relative drift”. The endpoint is
reached if the actual drift is smaller than the sum of the two • Check electrical contacts; defective contacts can lead
mentioned values. The higher the chosen value for the to an unstable measuring signal.
“relative drift” the sooner the determination is stopped and the • Check if chosen oven temperature is suitable for
more water remains in the sample. The “relative drift” value sample analysis.
should not be higher than 5 µg/min. If high accuracy is • Balance: too inexact, drafts, temperature equilibrium
required and for small water contents the “relative drift” not reached, sample weight not optimal/too low.
should be reduced (e.g. to 2 µg/min).
• Possibly carry out a qualification of the analysis system.
Please contact Metrohm Service for further information.
Parameters
Table 1: Recommended parameters for the oven technique in
Example applications
combination with coulometry.
The following table 2 provides an overview of samples
Parameter Setting
analyzed by the oven technique. In each case the Karl
General parameters I(pol) 10 µA Fischer water determination was carried out in combination
Generator current 400 mA or with a Coulometer using N2 as a carrier gas (flow rate
auto* 40 mL/min).
Control parameters EP at 50 mV
Dynamics 70 mV
Max. rate Maximum
µg/min
Min. rate 15 µg/min
Page 5 of 6
Application Bulletin 280/3 e
Automated water content determination with the 874 OSP
Table 2: Selection of samples which can be analyzed using the Denture cleaner, 70 0.2–1.5 3.8%
oven technique
effervescent
Sample Temp. Sample Water content tablets4
[°C] weight [g] Drugs 140 0.04 6.7%
Foodstuff Lyophilizate 150 0.01 5.0%
Lyophilizate 120 0.06 1.4% Others
Aromas1 100 0.03–0.08 2.5-5.4% Sodium tartrate 160 0.02–0.08 15.5%
Maltodextrin 100 0.03–0.08 4.7-8.8% dihydrate
Lactose 155 0.06 5.2% Potassium 220 0.03 5.6%
monohydrate citrate
Skimmed milk 90 0.06 4.4% monohydrate
powder2 Polyammonium 220 0.03–0.3 1.0-85.0%
Whole milk 90 0.06 3.9% compounds
powder2 Emulsified fat 220 0.03–0.08 85.0%
Sweet whey 90 0.06 4.9% compound
powder2 Formamidosulfo 220 0.2–0.3 < 1.0%
Glucose 90 0.06 8.9% nic acid
monohydrate2 Pigment 100 0.03–0.3 7.9%
Maltose 120 0.06 5.9% Polyol ether 150 0.6–1.3 0.2%
monohydrate3 Dibutene 100– 0.03 250 µg/g
Garlic powder 110 0.2 3.5% 140
Mineral mixtures 110 0.3 6.0-7.0% Lithium cobaltite 100 0.45–1.0 64 µg/g
Plastics Building rubble
Polypropylene 170 3.0 380 µg/g Surface water 50–60 0.3 0.6%
Polyethylene 115 3.0 40 µg/g Bound water 85–140 0.3 1.0%
Olefins 180 3.0 100 µg/g
1
With the 874 Oven Sample Processor the dissolution of the
Polyamide 180 0.3 0.8% sample in KF reagent, which is required in a direct Karl Fischer
Polyoxy- 170 0.3 0.1% titration and which frequently cannot be carried out completely , is
methylene no longer necessary
2
(POM) With the addition of 4 mL methanol as extraction agent (the water
content of the methanol must be determined and included in the
Polystyrene 120 0.05…0.2 200–500 µg/g
blank value).
Refinery products 3
With the addition of 4 mL 1,5-pentandiol, extraction time 600 s
Transformer oil 150 3.0 80 µg/g (the water content of the 1,5-pentandiol must also be determined
Mineral oil 120 1–3 10–100 µg/g and included in the blank value).
4
Above 70 °C the contained carbonate decomposes.
Insulating oil 140 3 5 µg/g
Crude oil 140 2 500-1200
µg/g Reference
Additive 120 0.01…0.03 4.4% • Metrohm Monograph 8.026.5013 - Water Determination
Antimony 50–130 3.0 700 µg/g by Karl Fischer Titration
dialkyl- • AN-K-048 – Sample preparation with the oven
thiocarbamate in technique – relative blank (available on
crude oil www.metrohm.com)
Pharmaceutical products
Collagens 160 0.07–0.4 10.6-9.2%
Author
Competence Center Titration
Metrohm International Headquarters
Page 6 of 6