2
CONTENTS
PRECAUTIONS..................................................................... 4
INSTRUCTION...................................................................... 6
FEATURES & MAIN FUNCTIONS................................. 6
TECHNICAL SPECIFICATION....................................... 7
DESCRITION OF OVERALL ........................................... 8
THE OPERATIONS............................................................ 10
TEST MODE......................................................................... 12
SET MODE............................................................................ 15
BATTERY OPERATION................................................... 19
REMOTE CONTROL........................................................ 20
ASSEMBLE AND DISASSEMBLY ................................ 21
ERROR MESSAGE & TROUBLE SHOOTING........ 22
3
PRECAUTIONS
4
5
INSTRUCTION
We greatly appreciate your purchase of the NC-I.
These goods perform excellently and exhibit splendid properties through strike tests.
Before using the NC-I, it is recommended that you read this manual carefully so you
may use this device to its full potential.
FEATURES
Appropriate hoist scale for industrial field
Useful Zero, Tare and Hold function
Weight Back-Up
Easy to move
Be able to control with remote controller
MAIN FUNCTIONS
User can change the stable condition and range of digital-filter at user's option
Self Hardware test (internal A/D value, Battery power, key and etc)
6
TECHNICAL SPECIFICATION
Max Capacity 20kg 30kg 50kg 100kg 200kg 500kg
Kilogram
Division 10g 10g 20g 50g 100g 200g
kg/lb Max Cap 200lb(100kg) 400lb(200kg) 1250lb(500kg)
version Division 0.1lb(0.05kg) 0.2lb(0.1kg) 0.5lb(0.2kg)
U.S.A Max Capacity 250lb(125kg) 500lb(250kg) 1000lb(400kg)
version Division 0.1lb(0.05kg) 0.2lb(0.1kg) 0.5lb(0.2kg)
Max Capacity 250lb 500lb 1000lb
Pound
Division 0.1lb 0.2lb 0.5lb
A/D Speed 20 times/second
Tare Subtraction Full Capacity
Display below zero “ ᅳ “ minus sign
Excessive Weight Maximum + 9d (display OUEr )
Weight Display Type LCD (5 digit + Sign, 1 Inch Height)
Character Display ZERO, TARE, BAT, POWER, CHARGE
Keys ON/OFF, ZERO, TARE, kg lb <ᅳ>(HOLD)
Battery 7.2V , 2A (Li - ion)
Use the Backlight Not use the Backlight
Power and Battery
0.45 W 0.17 W
consumption
Approximate 22 Hours Approximate 58 Hours
Operation Temperature -10 ℃ ~ +40 ℃
Rechargeable Adapter 110/220VAC 12V, 1.5AH
Rechargeable Time 4 ~ 5 Hours
Low BAT & Power Off LOW BAT : Below of 6.7V, POWER OFF : 6.4V
WEIGHT Approx. 5.5kg
7
DESCRIPTION OF OVERALL
1. OVERALL VIEW
2. DISPLAY
8
3.LAMP of DISPLAY PART(Front body)
Power ON Being connected the Adapter
LED Charge RED In charge GREEN Charge is completed
Remocon Receiving part from remote controller
Display the battery recharging time (rechargeable battery)-
BAT about 6.7V remain
- Below zero
LCD
ZERO ON when current weight is 0
TARE ON when tare is activated
OZ/lb/kg Weight Unit
4. KEY
KEYS FUNCTIONS
Power is ON while pressing for 1 second
Power is OFF while pressing for 2.0 seconds
Above function is possible while the power switch on the rear part is ON
Please turn off the power switch on rear part when not use for a long time.
It makes the current weight within the zero range to zero
(±2% of maximum capacity)
Pressing ZERO key for 2 second, can enter the TEST Mode
It is used to weigh with a container
If press the tare key at unload state, Tare state is released
Pressing the ON/OFF key while tare key is pressed, can enter the SET Mode
It is used to weigh the moving things
If you press this key once more in the hold mode, hold state is released
(at AVERAGE HOLD, at MANUAL HOLD, at PEAK HOLD)
It is used to change the unit of weight(kg<->lgToggle _ U.S.A. version)
This key is not operated in the area which lb unit is used only
5. REAR PART
Power Switch
POWER
Main Power Switch (If use, you have to turn on this switch)
Charger Adaptor connector
Jack Cover
If you want to recharge the battery, Connect to the adaptor.
9
THE OPERATION
1. POWER ON
■ Press "ON/OFF" key.
2. ZERO OPERATION
■ Press "ZERO" key when the zero point is drifted.
3. TARE OPERATION
■ Load tare on the hook.
■ Press "TARE" key.
"0" is displayed, and the tare is subtracted.
10
■ Remove tare on the hook.
■ Press "TARE" key.
The tare weight is cleared appers in the display window.
11
4. UNIT CONVERSIONS
■ Place an item to be weight on the hook.(KILOGRAM MODE)
■ Remove weight on the hook.
■ Press "kg/lb" key.(POUND MODE)
■ Place an item to be weight on the hook.
TEST MODE
How to enter TEST Mode
■ Pressing the key for 2 seconds in the normal mode, enter the test
mode with display the message Press the key or key
to execute the test mode.(test2~5 is same sequence.)
I f press the key in the test mode, test mode is ended and then enter the normal mode.
(Ref. You can't enter the test mode while in the state of Hold or Tare.)
12
Test menu ( tESt 1~5)
: Display Test
: Load Cell and A/D Conversion Test
: The IC(CPU) Temperature Test
: Battery Test
: Key Test
Available keys
KEYS FUNCTIONS
Exit Test mode and then enter the Normal mode
Use the backlight(only TEST 1)
Move the next menu and then execute TEST Mode
TEST 1
FUNCTION : LCD TEST
KEYS LCD SCREEN DESCRIPTION
Display TEST 1 state
Display all digits on the screen
TEST 2
FUNCTION : Load Cell and A/D Conversion TEST
KEYS LCD SCREEN DESCRIPTION
Display TEST 2 state
Display the digital value
Corresponding to current weight
13
TEST 3
FUNCTION : The IC (CPU) Temperature TEST
KEYS LCD SCREEN DESCRIPTION
Display TEST 3 state
Display the temperature of IC [CPU]
TEST 4
FUNCTION : Battery TEST
KEYS LCD SCREEN DESCRIPTION
Display TEST 4 state
Display the current value of battery
TEST 5
FUNCTION : Key Test
KEYS LCD SCREEN DESCRIPTION
Display TEST 5 state
Pressing the key to test shows
(Power, Zero, Tare, HoLd)
14
SET MODE
How to enter SET Mode
■ Turn on the power while pressing key , after displaying the message
and then enter the SET Mode.
Return to the normal mode if you press the key or key long.
SET Menu (F0 F6)
Menu Contents Menu Contents
F0 Setting the value of Digital-Filter F4 Weight Backup
F1 Setting the stable time of weight F5 Setting the hold type
F2 Setting the stable range of weight F6 Setting the time of Auto Display Off
F3 Setting the Back Light
Available keys
KEYS DESCRIPTION
Menu Cycle ( F0 ~F6F0)
Increase the setting value
Decrease the setting value
Store the current setting value (Not exit)
After storing all setting value and then exit SET Mode
15
F0
FUNCTION : Setting the value of Digital-Filter
KEYS LCD SCREEN Setting value DESCRIPTION
00 00 : Not use Digital-Filter
(Speed of display conversion is fast)
~
~
0.9 09 : Speed of display conversion is slow
F1
FUNCTION : Setting the stable time of weight (step : 0.1 sec.)
KEYS LCD SCREEN Setting value DESCRIPTION
00
000 : Setting the time within 0.5 second
~
09 : Setting the time within 1.4 second
0.9
F2
FUNCTION : Setting the stable range of weight (step : 0.5 division)
KEYS LCD SCREEN Setting value DESCRIPTION
00 00 : Setting the range within 0.5 division
~
0.9 09 : Setting the range within 4.0 division
F3
FUNCTION : Setting the stable range of weight (step : 0.5 division)
KEYS LCD SCREEN Setting value DESCRIPTION
Always Off
00
Always ON
01
Back Light ON whenever weight is loaded
02
Back Light ON Whenever the weight is
03
stable
16
F4
FUNCTIN : Weight Backup
KEYS LCD SCREEN Setting value DESCRIPTION
00 Not use
01 Use (Zero, Tare, Status)
F5
FUNCTION : Hold Type
KEYS LCD SCREEN Setting value DESCRIPTION
00 Average Calculate the average of
Hold moving weight
~
Peak Calculate the peak of
0.9 Hold moving weight
Description of HOLD Function
1. When the F5 is set to 00
Case 1 ( When the hold key is pressed in unloaded state) : Auto-Average Hold Function
If you press the Hold key in unloaded state, after displaying the and
the Hold function Is started.
Just When the weight is changed(above the 1divison) , being displayed
and the changed weight is hold
If you want to release the Hold function, have to press Hold key in unloaded
state certainly
When the Hold function is released, being displayed the message
Case 2 (When the hold key is pressed in loaded state) : Manual Hold Function
When you want to hold the current weight, just press Hold key.
After displaying the and the Hold function Is started.
17
In Hold state
- if the weight is not changed, the current weight is not changed
- if the weight is changed(above the 10% of Max capa), the current weight is changed and
the Hold function is released
- Whenever unloaded states, the Hold function is released automatically
- Also you want to release the Hold function, have to press Hold key in unloaded
state certainly
- If Hold function is released, you can see the massage .
2. When the F5 is set to 01
The maximum value of weight is showed when the Hold key is pressed
If the Hold key is pressed once, the peak Hold function is started
If Hold function is released, you can see the massage .
If the Hold key is pressed again, the Hold function is released
If Hold function is released, you can see the massage .
(Ref> If the weight is 0 lb or 0 kg, this function is not operated.)
F6
FUNCTION : Setting the time of Display off automatically
KEYS LCD SCREEN Setting value DESCRIPTION
00 Not use
01 Display all off in 3 minutes
02 Display all off in 6 minutes
03 Display all off in 9 minutes
Ref. 1. Operates the function of F6 which anything is set when the weight is in zero state.
2. The Display is ON when the weight is not in zero state or when the any key
In the remote controller and front cover is pressed
18
BATTERY OPERATION
1. CHARGE
A. The rechargeable battery inside the NC scale must be charged using the
voltage which don't exceed ±10% of the voltage indicated on the adapter.
B. If the voltage for charging exceed the above limit, the battery would be damage.
C. Check if an output voltage is DC 12V 1.5A.
D. Remove ADAPTOR JACK COVER of REAR PANEL and connect
adaptor with an scale like the following figure.
E. If adaptor is connected with the scale, Red lamp is on(Power) and Charge
lamp is on(Red : charge, Green : complete charge)
F. Time of charge is approximate 4~5 hours.
G. If main power switch is off, charge is not operate.(Red lamp is blinking.)
Do check main power switch(Main power switch is located Rear panel)
2. Discharge
If you use the battery continuously, discharge time is as fallows.
A. with the backlight -> approximate 22 hours
B. without the backlight -> approximate 50 hours
19
3. LOW BATTERY("BAT.") SIGNAL
A. To protect over-discharge of Battery, A. To protect over-discharge of Battery,
BAT LAMP is on when battery has to be charge. (80% of full charged voltage)
B. A scale turns power off automatically below the fixed voltage
(75% of full charged voltage). You can show LobAt message before turn power off.
BAT LAMP is on when battery has to be charge.
(80% of full charged voltage)
C. A scale turns power off automatically below the fixed voltage
(75% of full charged voltage).
You can show LobAt message before turn power off.
REMOTE CONTROL
■ FUNCTION
RED LED Blinking whenever key is pressed
ON/OFF KEY Display ON/OFF(Toggle)
It makes the current weight within the zero range to zero
ZERO KEY ( 2% of maximum capacity)
TARE KEY It is used to weigh with a container
HOLD KEY/ It is used to weigh the moving things
Kg -> lb KEY It is used to change the unit of weight (kg < - > lb Toggle _ U.S.A. version)
※ Remote controller is working within 60 ℃ and distance of 7~10m
20
■ POWER
Voltage 1.5V
Size AA Size(14.5 X 50.5 m/m) - Battery
Note on the Place batteries in correct polarity(+)(-)direction.
replacing
battery Do not use new batteries with the used
ASSEMBLE AND DISASSEMBLY
21
ERROR MESSAGES & TROUBLE SHOOTING
Err 01
■ Cause
The Indicator can't be initialized due to platform shaking
■ Management
Put the platform on a flat place without shaking and power on
If you can see the same massage continuously, call your CAS dealer please.
Err 02
■ Cause
Initial A/D value of zero is too high or low, when the power is ON
Or in the case of NC-I is lying on the flat
■ Management
Check that the load cell is well connected with main unit
Or calibration is done wrongly
Don't lie the NC-1 on the flat
If you can see the same massage continuously, call your CAS dealer please.
Err 09 (Zero Key)
■ Cause
The current weight exceeds zero range (2 % of Maximum Capacity)
■ Management
Unload the scale
If you can see the same massage continuously, call your CAS dealer please.
Err 13 (Zero Range when the Power is ON)
■ Cause
The initial zero range exceeds above 10% of maximum capacity
■ Management
Unload the scale
If you can see the same massage continuously, call your CAS dealer please.
22
BAT or LobAt
■ Cause
Need to recharge of battery
(After display LobAt , the power is off automatically in 2 second)
■ Management
Connect the adaptor with outlet of charge battery of rear part
(Charging time is about 4~5 hours)
If you can see the same massage continuously, call your CAS dealer please.
oUEr
■ Cause
Too heavy object is put on platform so that it may exceed the maximum capacity
■ Management
Never put the weight exceeding the maximum capacity
If you can see the same massage continuously, call your CAS dealer please.
If the load cell is damaged, replace the load cell
* Ref. The function of this manual can be changed without previous warning !!
23
MEMO
24
MEMO
25
MEMO
26
27
28