Site Preparation Drawings

Table of contents
Sheet Type Subject

General
1 C-1 Cover sheet . Table of contents
2 AL-1 Site requirements
3 A-1 Site layout
4 A-2 Building modifications
5 AD-1 Schematic cross section(s)
Building works
6 X-1 Floor provisions (all disciplines)
7 X-2 Ceiling provisions (all disciplines)
8 X-3 Wall provisions (all disciplines)
9 X-4 Wall provisions (all disciplines)
Technical details
10 XD-1 Details 1 (all disciplines)
11 XD-2 Details 2 (all disciplines)
12 XD-3 Details 3 (all disciplines)
13 XD-4 Details 4 (all disciplines)

Typical Site Preparation drawings


14 XD-5 Details 5 (all disciplines)
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

MR5300
Project
Typical configuration shown (not site specific) Revision history
H
G
F

Table of contents
E
D
C

Cover sheet
B

Sheet Subject
A
0 20-Feb-2023 Typical Site Preparation drawings

Important note: Rev Date Drw Chk Proj. phase Description

Important notes

This package provides only indicative information intended for use in very early project phases to provide an
impression of room set ups and required site preparation. Any information,
figure or value shown is not necessarily applicable, accurate and/or up-to-date.
© Koninklijke Philips Electronics N.V. 2020

Your Philips contact can arrange Room Designs and Site preparation specifications for your specific situation.

Quote No
Order No
Drawing
Philips contact Customer ID Technical ID

Philips Healthcare A3 mm

C-1
Sheet 1 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
Drawing alerts: Site requirements
MR5300
Dimensions for building works in the examination room (like RF-cage construction Mains power Mass Heat Noise
1 · Supply configuration: Resp No Description [kg] [W] [dB(A)]
and wall openings) are for indication and bidding only! Refer to the drawing of the RF
cage vendor for actual construction dimensions. For ranges refer to AD-1 sheet. Star, 3 phase+neutral+protective earth (PE).
Examination Area 2000* -
Delta is allowed for the 480V version.
MDU version to be ordered via order questionnaire, A 1.1 Scanner gantry 1.5T 3700 - -
Passive shielding or fencing may be required to contain the 0.5mT field line within the · Nominal Line Voltage: 1.2
2 A Patient table 260 - -
floor and ceiling of the examination room. 220/380V +/-10% 50/60Hz +/-1Hz
230/400V +/-10% 50/60Hz +/-1Hz A 1.3 System filter box 80 - -
480V +/-10% 60Hz +/-1Hz 1.4
A Accessory cabinet (Optional) - - -
· Mains impedance (at nominal voltage):
< 200mOhm Control Area (Basic system excluding options) 300 40
Passive shielding or fencing may be required to contain the 0.5mT field line within the
3 A 2.1 Operating station - - -
back and side wall of the examination room. Electrical power rating (230/400voltage input)
Apparent Power: 60kVA A 2.2 Emergency run down button - - -
Rated current: 90A/phase
A 2.3 Storage Rail - - -
It is recommended to install a small RF feedthrough to avoid air handling issues like Peak current: 400A/phase <5ms (*)
4 2.4
under or overpressure due to air flow unbalance. If not place it could be difficult to Peak current: 500A/phase <1ms (*) A Operator's table (Optional) 100 - -
open or close the RF-door. Furthermore the RF-walls could be damaged. Nominal fuse rating (**): 100A rms
Technical Area (Basic system excluding options) 8000* 75
Typical dimensions 300mm x 150mm (LxH).
To avoid unwanted acoustic noise issues we suggest locating the balance grid in the (*) A 3.1 Mains distribution unit 400V 50Hz 135 - -
RF-ceiling and no acoustic path towards the Operators room. If an UPS for the complete MR system is required, please provide the UPS supplier not
A 3.2 Liquid cooling cabinet 326 - -
only the required kVA but also the required above-mentioned peak current. This can
result in a higher (kVA) specification of the UPS needed. A 3.3 Gradient amplifier cabinet 781 467 - -
A 3.4 Data Acquisition & Control Cabinet 397 - -
(**)
Ferromagnetic beams perpendicular to the Z-axis of the magnet must be located at
5 • For 230V Y-configuration ( = fase to neutral ), type AM or GL according DIN 43620, VDE A 3.5 System Air Cooling Unit (SACU) 25 - 72
least 250mm below the finished floor. All other ferromagnetic beams must be
0636 part 1 and 2a
located at least 600mm below the finished floor. B 3.6 Chilled water connection - - -
Remote service diagnostics B 3.7 Mains switch - - -
To establish this feature, a RJ45 type Ethernet connector must be B 3.8 Earth reference terminal - - -
For the square area of 3000mm x 3000mm symmetrically around magnet isocentre, installed with access to the customer's network (10/100/1000Mbps)
6
ferro-magnetic reinforcement must be: B 3.9 RF-waveguide System Air Cooling Unit (SACU) - - -
- NO less than 50mm below finished floor. Environmental requirements
- NO greater than 25kg/m² average concentration between 50mm and 250mm A 3.10 Air cooled Cryo-Cooler 110 - -
Examination room:
below the finished floor. Reinforcements below 250mm can be ignored. Temperature 18°C-22°C A 3.11 Back up Power Connection Panel (BCP) (WiP) 135 - -
- Evenly distributed. Maximum temperature change 5°C per 10min
General
Relative humidity 40% - 70%, no condensation
Required Air Conditioning Capacity: 2kW B 4.1 RF door - - -

Typical Site Preparation drawings


- Energy dissipated in the examination room must be cooled by hospital air handling system. B 4.2 RF window - - -
- Gradient coil heat dissipation (1 to 15kW) will be removed via liquid cooling of the
gradient coil by Philips to the LCC located in technical room.
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

Technical room:
Temperature 15°C-24°C
Maximum temperature change 5°C per 10min
Maximum Cable lengths: Relative humidity 30% - 70%, no condensation
Required Air Conditioning Capacity: 2kW (8kW if air cooled cryo cooler is active)
Maximum Cable lengths between the Cable in- and Outlets of the Cabinets - Standby heat dissipation: 2kW (8kW if air cooled cryo cooler is active)
- Heat dissipation to Water: 5 - 40kW
From: To: Meter:
LCC (**) Air cooled Cryo-Cooler 0 Control room:
Temperature for human comfort 18°C-24°C
System Filter Box DACC 7

MR5300
Temperature required for X-ray films 15°C-30°C
System Filter Box LCC (**) 7 Temperature MR equipment 10°C-35°C
System Filter Box Gradient Amplifier (**)(***) 7 Maximum temperature change 5°C per 10min

Project
LCC (**) SACU (*) 14 Relative humidity 30% - 70%, no condensation
Required Air Conditioning Capacity: 0.3kW
DACC Gradient Amplifier (**) 7
DACC LCC (**) 7 Responsibilities
Gradient Amplifier (**) LCC (**) 7 A Delivered and installed by Philips
MDU(**) DACC 10 B Delivered and installed by customer/contractor

Equipment legend
Site requirements
MDU(**) Gradient Amplifier (**) 8 C (Pre) Delivered by Philips, installed by customer
MDU(**) LCC (**) 13 D (Pre) Delivered by customer and installed by Philips
E Existing
MDU(**) 50Hz Countries Air cooled Cryo-Cooler 10

Sheet Subject
MDU(**) 60Hz Countries Transformer 7
Transformer Air cooled Cryo-Cooler 7
MDU(**) All other units 10

(*) Location of the SACU in the technical room is not required; Interface with
examination room return air duct outside the RF-enclosure is leading. Keep the * Heat load indicated is peak dissipation for each cabinet measured individually. Peak room
acoustic noise in mind! heat dissipation as indicated in Site Requirements column will be less than the sum of each
individual cabinet in a given room due to the fact that not all cabinets will run peak heat
(**) Cable and hose connections on top of the cabinet only.

0
loads at the same time.
(***) Gradient cables are supplied in one set with a length of 15m. Cable length can be
divided/ cut between length needed inside technical room and length needed inside
© Koninklijke Philips Electronics N.V. 2020

the examination room. Project notes MR


Default length is reduced to reduce costs because in many cases shorter gradient · The feasibility of this project is based on site information (like drawings) provided to

20-Feb-2023

Technical ID
cables are sufficient. 18m length is possible with an extra cable set. Please contact Philips. Philips cannot assume any liability for the accuracy of this information and the

Quote No
Order No
Drawing
SPS to order a duplicate extra set of cables to be able to extend the length to consequences.
maximum 18m.
· The equipment proposed in this room layout produces a strong magnetic field. It is the
customer's responsibility to comply with the local regulation. Normally this field has to be
contained in a controlled area that is only accessible by authorised personnel and screened A3 mm
patients. Consult the Philips contact for details and support.
AL-1
Sheet 2 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

1m

Dimension origin
at finished wall,
near floor level.

Y 2320
4 1

3.8
3.7
0.1mT
4.1
0.3mT 3.5

2800
0.5mT 3.9
1mT
3mT 3.1
Examination area
2.2 3.11
2.2 10mT
Control area 3.3
2.1
3.4

Typical Site Preparation drawings


Z

4.2 3.2
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

1.2
2.3 3.6
1.4
1.1 1.3
3.10
2.4

Technical area

MR5300
Project
3

Site layout
Sheet Subject
Place to be defined locally by customer.

Legend

0
© Koninklijke Philips Electronics N.V. 2020

Walls

20-Feb-2023

Technical ID
New walls Dimensioning legend

Quote No
Order No
Drawing
Existing doors/walls to be removed
XYZ System position dimension
Beams or other building construction elements
(ISO-centre)
A3 1 : 50 mm
(XYZ) Architectural dimension
Columns (materials and strength unknown)
to be checked on site A-1
Preferred service area Dimension from origin Sheet 3 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

1m

Dimension origin
at finished wall,
near floor level.

(3000) (7000) (2000)


Y 2320 (900)

1315
6 5
Opening wall duct

2800
(5150) Opening SACU

(5000)

Typical Site Preparation drawings


Z
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

Opening
System filter box

725

MR5300
Project
Scanner gantry
transfer opening
4200
2500 x 2600

Building modifications
Sheet Subject
Legend

0
© Koninklijke Philips Electronics N.V. 2020

Walls

20-Feb-2023

Technical ID
New walls Dimensioning legend

Quote No
Order No
Drawing
Existing doors/walls to be removed
XYZ System position dimension
Beams or other building construction elements
(ISO-centre)
A3 1 : 50 mm
(XYZ) Architectural dimension
Columns (materials and strength unknown)
to be checked on site A-2
Preferred service area Dimension from origin Sheet 4 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

1m X 2
Finished Wall
RF Wall
Architectural Wall

Suspended ceiling
+10/-25mm maximum/minimum
recess range, for actual dimensions
refer to RF cage vendor drawings!

(3400)
3000
2520
+10 0
-25
Z
1004

50
50
Min. 50
1386

3mT 2
2186

0.5mT
1mT Detail - Floor lowering principle (scale 1:20)

Typical Site Preparation drawings


0.3mT
X
2
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

0.1mT

Schematic cross section A-A

Suspended ceiling

MR5300
Project
(3400)
3000
2520
B

Schematic cross sections


Y

A A 1004

Sheet Subject
1386

3mT
1mT 2
2186

0.5mT

0
0.3mT
© Koninklijke Philips Electronics N.V. 2020

0.1mT

20-Feb-2023

Technical ID
B

Quote No
Order No
Drawing
Situation (scale 1:100) Schematic cross section B-B A3 1 : 50 mm

AD-1
Sheet 5 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

1m

Dimension origin
at finished wall,
near floor level.

(7000)
Y 2320

2800
1030 605
120 710 200
120 300

B A

(5000)
150

485
318
373
Z

485
C C

Typical Site Preparation drawings


60
1 2 A
B

150
150
3 3
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

5
4 4
696 696
663 1710

MR5300
Project
Floor provisions
(all disciplines)
Sheet Subject
Item list:
Resp No Description Qty Detail Text ref

B 1 Floor pad 1 1 XD-1 -


Legend

0
B 2 Floor pad 2 1 XD-1 -
© Koninklijke Philips Electronics N.V. 2020

B 3 Threaded mounting holes M10 2 XD-1 -


Walls
B 4 Suitable anchoring system 2 XD-1 -

20-Feb-2023

Technical ID
New walls Dimensioning legend

Quote No
Order No
B 5 Floor pad 3 4 XD-1 -

Drawing
Existing doors/walls to be removed
Responsibilities
XYZ System position dimension Works symbology
Beams or other building construction elements A Delivered and installed by Philips
(ISO-centre) B Delivered and installed by customer/contractor Construction works A3 1 : 50 mm
(XYZ) Architectural dimension C (Pre) Delivered by Philips, installed by customer Electrical works
Columns (materials and strength unknown)
to be checked on site D (Pre) Delivered by customer and installed by Philips Mechanical works X-1
Preferred service area Dimension from origin E Existing Equipment works Sheet 6 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

1m

Dimension origin
at finished wall,
near floor level.

300 (7000)

150
Y 2320
2
3

300
Wall duct
see X-3

2370

2800
Cable ladder
wall mounted,
see X-3

(5000)
Wall duct
see X-3 Z

Typical Site Preparation drawings


1
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

Cable ladder
wall mounted,
see X-3

350

MR5300
300

Project
Ceiling provisions
(all disciplines)
Sheet Subject
Item list:
Resp No Description Qty Detail Text ref
Legend

0
B 1 500 x 60 Cable ladder - - § 3.3
© Koninklijke Philips Electronics N.V. 2020

B 2 130 x 60 Cable duct above suspended ceiling - - § 3.3


Walls
B 3 Wall opening 1 - -

20-Feb-2023

Technical ID
New walls Dimensioning legend

Quote No
Order No
Drawing
Existing doors/walls to be removed
Responsibilities
XYZ System position dimension Works symbology
Beams or other building construction elements A Delivered and installed by Philips
(ISO-centre) B Delivered and installed by customer/contractor Construction works A3 1 : 50 mm
(XYZ) Architectural dimension C (Pre) Delivered by Philips, installed by customer Electrical works
Columns (materials and strength unknown)
to be checked on site D (Pre) Delivered by customer and installed by Philips Mechanical works X-2
Preferred service area Dimension from origin E Existing Equipment works Sheet 7 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

250 1190
1015
1m
2715 1550 740
1200 1245 300 850 500

15 14
Ceiling duct see X-2
13 16
Ceiling duct see X-2

250
Ceiling ladder
see X-2

850
Suspended ceiling
7 10

100
7 8 7

(3400)

(3400)
9

3050
2

950
1 4

2700
12

2650

2650
2550

2060

2300
560 560
1800

1500
1400
1260
6

995

900
680
5

7 11 4 4
3
845 735
1345

Typical Site Preparation drawings


Elevation A Elevation B
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

Item list:

MR5300
Resp No Description Qty Detail Text ref

Project
B 1 RF window 1 - § 1.6
B 2 RF door 1 - § 1.6
B 3 Additional power outlet (customer) 2 - § 3.2
B 4 Network connection 100Mb, RJ45 2 - § 3.4
B 5 Mains switch 1 - § 3.2

Wall provisions
B Earth reference terminal 1 - § 3.2

(all disciplines)
6
B 7 130 x 60 Wall duct - - § 3.3

Sheet Subject
B 8 500 x 60 Cable ladder - - § 3.3
B 9 300 x 60 Cable ladder - - § 3.3
B
B 10 Ø 19mm Conduit flush mounted - - § 3.3
A 11 Cable storage rail 1 XD-5 -
A
A 12 Emergency run down button 1 - § 3.2

0
B 13 Chilled water connection 1 - -
A System Air Cooling Unit (SACU) 1 XD-2 -
© Koninklijke Philips Electronics N.V. 2020

14
B 15 Opening system filter box 1 XD-4 § 2.4

20-Feb-2023

Technical ID
Quote No
Order No
B 16 Opening SACU 1 - § 4.2

Drawing
Responsibilities
A Delivered and installed by Philips Works symbology
Situation (scale 1:100) B Delivered and installed by customer/contractor
C (Pre) Delivered by Philips, installed by customer
Construction works
Electrical works
A3 1 : 50 mm

D (Pre) Delivered by customer and installed by Philips Mechanical works X-3


E Existing Equipment works Sheet 8 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
0 1 2 3 4 5m

1m

2950 2500

(3400)

(3400)
2600
1

Typical Site Preparation drawings


Elevation C Elevation D
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

MR5300
Project
Wall provisions 2
(all disciplines)
Sheet Subject
D

0
Item list:
© Koninklijke Philips Electronics N.V. 2020

Resp No Description Qty Detail Text ref

20-Feb-2023

Technical ID
Quote No
Order No
B 1 Scanner gantry transfer opening 1 - § 1.7

Drawing
Responsibilities
A Delivered and installed by Philips Works symbology
Situation (scale 1:100) B Delivered and installed by customer/contractor
C (Pre) Delivered by Philips, installed by customer
Construction works
Electrical works
A3 1 : 50 mm

D (Pre) Delivered by customer and installed by Philips Mechanical works X-4


E Existing Equipment works Sheet 9 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
Cable duct cover

Min. 14mm
Antistatic floor covering
Top layer, Epoxy
Wood etc.
Cable duct
Heigth depending RF shield
on top layer type
Building floor

Cross section A-A

Pad 1
Ref. level
+0/-2.5mm
372
Antistatic floor covering
Top layer, Epoxy
Wood etc.
Max. 5mm

RF shield

Typical Site Preparation drawings


Building floor
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

Cross section B-B

Patient table outline Magnet covers outline

Pad 2

MR5300
Pad 1 Reference Pad 3 Pad 3
Ref. level Fixation point Ref. level Ref. level

Project
+0/-2.5mm Patient table +0/-2.5mm +0/-2.5mm
Antistatic floor covering
Top layer, Epoxy
Wood etc.
RF shield
Building floor
Cable duct
RF tight floor fixation Cable duct cover
(responsibility of RF cage supplier)

(all disciplines)
Cross section C-C

Sheet Subject
Details 1
Detail - Scanner gantry/patient table pads (not to scale)

0
© Koninklijke Philips Electronics N.V. 2020

20-Feb-2023

Technical ID
Quote No
Order No
Drawing
A3 mm

XD-1
Sheet 10 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
1 200
506 3 150
124
2

220
275

275
60

35

27
130 122 122 130
18
RF shielding: do not make contact or
penetrate the RF shielding

R7
Detail - System air cooling unit mounting (scale 1:10)

Interface location max. 2m distance from SACU. Third party supplied. 127mm outer
diameter interface. Airflow 400 m³/h Heat load 0.7 kW during scanning To avoid too Rf feedthrough PHILIPS supplied hose through wave guide. Diameter min.160mm. Air in from magnet interface to SACU
much interaction between airflow of air return and SACU return minimum air duct Typical location above suspended ceiling of examination room near hospital air
cross 0.35m² return duct.

Examination area

Return from Technical area


examination
room

Typical Site Preparation drawings


All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

1000

MR5300
Project

(all disciplines)
Sheet Subject
Details 2
2375

Item list:
Resp No Description Qty Detail Text ref

0
B 1 Threaded M6 rods 4 - -
Hospital air conditioning
© Koninklijke Philips Electronics N.V. 2020

A 2 Mounting provision for Air cooling unit 1 - -


B 3 Drilled holes suitable for M6 Threaded rods 4 - -

20-Feb-2023

Technical ID
Quote No
Order No
Drawing
Detail - connection diagram SACU (scale 1:10) Responsibilities
A Delivered and installed by Philips Works symbology
B Delivered and installed by customer/contractor Construction works A3 mm
C (Pre) Delivered by Philips, installed by customer Electrical works
D (Pre) Delivered by customer and installed by Philips Mechanical works XD-2
E Existing Equipment works Sheet 11 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
1000 2200 >1500 1000

Ceiling speaker area

Patient comfort zone


890 no direct lighting
1440 Isocenter

Typical Site Preparation drawings


1300
1250

3000
Removable ceiling grid Z
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

required for servicing


800
720

Z
360

MR5300
Isocenter
No suspended ceiling in this area 2500 1000 750 880

Project
Y
Detail - Ceiling area for lighting and speakers (scale 1:50)
Detail - Suspended ceiling magnet service area
(scale 1:50)

(all disciplines)
Sheet Subject
0Details 3
© Koninklijke Philips Electronics N.V. 2020

20-Feb-2023

Technical ID
Quote No
Order No
Drawing
A3 mm

XD-3
Sheet 12 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
670

960 20 45 45 45 45 45 45 45 45 45 45 45 45 45 20
23 22
385 465

1 2
Aperture in suspended ceiling

205
Cable feedthrough
RF Wall/Feedthrough Frame Aperture

800

605
670
Top view

605

Typical Site Preparation drawings


Underside RF ceiling
960 Detail - Feedthrough System Filter Box (scale 1:10)
Minimum 340

440 520
All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

Suspended ceiling
35 ±5
340

50 50

MR5300
Free space for patient
ventilation air intake

Project
Typical 2520 (see X3 and X4)

and fixation of the


screws on the cover
1260

(all disciplines)
No obstacles
allowed in

Sheet Subject
this area

Details 4
Finished floor level in examination room
Item list:

0
Cross section Resp No Description Qty Detail Text ref
© Koninklijke Philips Electronics N.V. 2020

B 1 Threaded holes 60 - -

Detail - Service area for System Filter Box (scale 1:20) B 2 Suitable fixing system M5 60 - -

20-Feb-2023

Technical ID
Quote No
Order No
Drawing
Responsibilities
A Delivered and installed by Philips Works symbology
B Delivered and installed by customer/contractor Construction works A3 mm
C (Pre) Delivered by Philips, installed by customer Electrical works
D (Pre) Delivered by customer and installed by Philips Mechanical works XD-4
E Existing Equipment works Sheet 13 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED
Table

1200

Ø6mm (8x)

120

Mounted to the hospital wall.

200

Ø6mm (8x)

Mounted to the underside of


the table.

Typical Site Preparation drawings


All rights reserved. Reproduction in whole or in part is prohibited without prior written consent of the copyright holder.

MR5300
Suspended from the table
(Clamps included)

Project

(all disciplines)
Detail - Mounting instruction Storage tray

Sheet Subject
Details 5
(Not to Scale)

0
© Koninklijke Philips Electronics N.V. 2020

20-Feb-2023

Technical ID
Quote No
Order No
Drawing
A3 mm

XD-5
Sheet 14 / 14
THIS SHEET IS PART OF THE DOCUMENT SET LISTED ON SHEET C-1 AND SHOULD NOT BE SEPARATED