INTERNATIONAL ISO
STANDARD 9626
Second edition
2016-08-01
Stainless steel needle tubing for the
manufacture of medical devices —
Requirements and test methods
Tubes d'aiguilles en acier inoxydable pour la fabrication de matériel
médical — Exigences et méthodes d'essai
Reference number
180 9626:2016(E)
© 180 20161S0 9626:2016(E)
A COPYRIGHT PROTECTED DOCUMENT
© 150 2016, Published in Switzerland,
All rights reserved. Unless otherwise specified, no part ofthis publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, oF posting on the internet or an intranet, without prior
\written permission. Permission can be requested from either ISO atthe address below or ISO's member body in the country of
the requester:
150 copyright office
Ch. de Blandonnet 8 » CP 401
(CH-1214 Vernier, Geneva, Switzerland
Tel. +4122 74901 11
Fax +41 227490947
[email protected]
wnvrwiso.org
cattle o @ 180 2016 Allrights reserved1S0 9626:2016(E)
Contents
Foreword
Introduction.
1 Scope.
2 Normative references.
3 Terms and definitions
4 Materials
5 Requirements
5.1 General
5.2 Surface finish and visual appearance.
5.3 Cleanliness
5.4 Limits for acidity and alkalinity.
5.5 Size designation.
5.6 Dimensions
5.7 Sample size
5.8 Stiffness
5.9 Resistance to breakage.
5.10 Resistance to corrosion
Annex A (normative) Methods for preparation of extracts.
Annex B (normative) Test method for stiffness of tubing,
Annex C (normative) Test method for resistance of tubing to breakage
Annex D (normative) Test method for resistance to corrosion
Annex E (informative) Rationale with respect to test method for stiffness of tubing
Bibliography
corte mores ouRbAD ODS adlh Fights reserved
Page
=
© @ VVOUNNNNKRN NH
Ree
Bran1S0 9626:2016(E)
Foreword
1S0 (the International Organization for Standardization) isa worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
180 collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
‘The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for Identifying any or all such patent rights, Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see wwiw.iso.ore/patents).
Any trade name used in this document is informal
constitute an endorsement.
nn given for the convenience of users and does not
For an explanation on the meaning of ISO specific terms and expressions related to conformity
assessment, as well as information about ISO's adherence to the WTO principles in the Technical
Barriers to Trade (TBT) see the following URL: Foreword - Supplementary information.
‘The committee responsible for this document is ISO/TC 84, Devices for administration of medicinal
products and catheters.
‘This second edition cancels and replaces the first edition (ISO 9626:1991), which has been technically
revised. Italso incorporates the Amendment ISO 9626:1991/Amd 1:2001
‘The main changes to the previous edition of ISO 9626 introduced by this revision are the following:
a) addition of specifications for stainless steel needle tubing for metric sizes 0,18 mm, 0,2 mm
0,23 mm and 0,25 mm and to reflect the introduction of thinner tubing to allow greater comfort
when injecting, particularly for infants and in paediatric use;
1b) addition of wall thickness designations beyond regular-walled and thin-walled tubing;
)_ addition of minimum inner diameters for additional items where possible;
4d) revision of the means of specifying the steels to be used;
@) revision of the table of tubing dimensions and stiffness parameters.
Annex A, Annex B, Annex C, Annex D and Annex E form an integral part of this International Standard,
caret oni oumeatonte Suendenin @ 180 2016 Allrights reserved1S0 9626:2016(E)
Introduction
Guidance on transition periods for implementing the requirements of this International Standard is
given in ISO/TR 19244,
corte more ouRbiDi ODS aclh Hight reserved v‘orate Cesnnonty SaemetINTERNATIONAL STANDARD ISO 9626:2016(E)
Stainless steel needle tubing for the manufacture of
medical devices — Requirements and test methods
1 Scope
This International Standard applies to rigid stainless steel needle tubing suitable for use in the
manufacture of hypodermic needles and other medical devices primarily for human use.
This International Standard provides requirements and test methods for the tubes manufactured for
needles as component used in medical devices. Additional performance testing on the tube aspect may
be required when the component is incorporated in the ready-to-use device.
This International Standard specifies the dimensions and mechanical properties of steel tubing of
designated metric sizes 3,4 mm (10 Gauge) to 0,18 mm (34 Gauge).
It does not apply to flexible stainless steel! tubing because the mechanical properties differ from those
specified for rigid tubing in this International Standard. However, manufacturers and purchasers
of flexible tubing are encouraged to adopt the dimensional specifications given in this International
Standard.
2 Normative references
‘The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
10 3696, Water for analytical laboratory use — Specification and test methods
10 15510, Stainless steels — Chemical composition
3. Terms and definitions
For the purposes of this document, the following terms and definitions apply.
34
designated metric size
outer diameter designation of the tubing as defined in Table 1
Note 1 to entry: Itis expressed in millimetres.
3.2
gauge
legacy size designation
Note 1 to entry: A particular gauge size corresponds to a designated metric size defining limits for outer
diameters.
33.
wall thickness
material thickness between the inner and outer diameter of the tube
Note 1 to entry: tis expressed as RW=Regular Wall, TW=Thin Wall, ETW= Extra Thin Wall, and UTW=Ultra Thin
Wall as designated in Table 1.
ceo rmraor OoRARA PBA aAAl) Fights reserved 11S0 9626:2016(E)
4 Materials
‘Tubing shall be made of stainless steels listed in ISO 15510. The chosen materials shall be in accordance
with the requirements indicated in this International Standard. Selection of specific stainless steel
material shall be made in consideration with the intended use, e.g. long-term contact with drugs and
should consider biocompatibility requirements.
NOTE Suitable biocompatibility requirements can be found in 10 10993-1.
5 Requirements
5.1 General
For the selection of tubing for a specific application and intended use, a risk based approach shall be
applied.
5.2 Surface finish and visual appearance
‘When examined by normal or corrected vision, the outside surface of the tubing shall be smooth and
free from defects.
Surface finish specifications may be different based on the final function of the medical device; in such
cases, the medical device manufacturer should prepare specific specifications for surface finishing.
When examined by normal or corrected vision, the needle tube shall appear straight and of regular
roundness.
5.3. Cleanliness
When examined by normal or corrected vision, the surfaces of the tubing shall be free from metal soil
and processing agents.
Cleanliness specifications may be different based on the final function of the medical device; in such
cases, the medical device manufacturer should prepare specific specifications for cleanliness.
5.4 Limits for aci
ity and alkalinity
When determined with a laboratory pH meter and using a general purpose electrode, the pH value of
an extract prepared in accordance with Annex A shall be within one pH unit of that of the control fluid.
5.5. Size designation
‘Tubing size shall be designated by the nominal outer diameter, expressed in millimetres (i.e. the
designated metric size), corresponding gauge size (e.g. G31 or 316), and by wall thickness.
EXAMPLE 0,25 mm (31G) ETW.
5.6 Dimensions
‘The dimensions of tubing shall be as given in Table 1.
caret been Oumetonte Suendetin @ 180 2016 Allrights reservedTable 1 — Dimensions of tubing
1S0 9626:2016(E)
Designated metric
op: op: 1D)
aed Gauge sn max | wan iN
‘mn mm mm. mm
RW 0.064
ose 34 0,178 091 Tw 0,091
TW 0,105
RW 0,089
020 3 0,203 0.216 TW. 0.105
ETW 04125
RW 0,089
023 32 0229 0241 tw 0108)
ETW 0425
TW 0146
RW. oat
TW 0.125
02s 31 0254 0.267
ETW 0.146
uTW 0.176
RW 0133
TW. 0,165
030 30 0.298 0,320
FTW 0,190
TW 0240
RW 0133
Tw 0,190
033 29 0324 0351
ETW 0,240
uTw 0.265
RW x
0.36 28 0349 0370 0188
TW 0,190
RW 084
040 ar 0,400 0420
Tw. 0.241
RW 0232
045 26 0440 0470
Tw 0292
RW 0.232
050 25 0,500 0530 Ww 202
RW 0.280
oss m 0550 0,500
TW 0343
RW 0317
060 23 0.600 0.673 Tw. 0.370
FTW 0460
[NOTE 1 RW= Regular Wall TW= Thin Wall, ETW =Extra Thin Wall UTW= Ulera Thin Wal,
INOTE2 Neeile sizes below 0,25 mm, consideration can be made to the measurement uncertainty of
existing measurement equipment.
INOTE3 This International Standard doesnot specify maximum inner diameter.
NOTE 4 0D = outer diameter; ID = inner diameter.
corel wmsiora ORAM PBA SaAlh Fights reserved1S0 9626:2016(E)
Table 1 (continued)
Designated metric Gauge | OMY ODmax Walt IDuiw
RW 0.390
0,70 2 0,698 0,730 TW 0440
eTW, 0,522
RW 0.490
TW 0547
0.80 a 0,800 0,830 rw jo
uTw 0,645
RW 0.560
0.90 20 0,860 0920 iw 8.655
ETW. 0,687
uTw O73
RW 0,648
TW 0.750
110 19 1,030 1,100
ETW. 0,850
uTW 0.891
RW 0,790
1.20 18 1,200 1,300 TW 0910
ETW 1,041
RW 0,950
TW 1,156
1.40 ” 1.400 1.510 aw Tae
uTw 1,276
RW 1,100
1.60 16 1,600 1,690 TW 1,283
ETW. 1,390
RW 1,300
1,80 15 1,750 1,900 TW 1,460
ETW. 1,560
RW 1,500
210 4 1950 2,150 TW 1,600
eTW, 1727
RW 1,700
240 13 2,300 2,500 TW 1386
RW 1,950
2,70 2 2,650 2,850
TW 2,235
[NOTE 1 RW/= Regular Wall TW = Thin Wall, ETW = Extra Thin Wall, UTW= Ultra Thin Wall
NOTE 2. Needle sizes below 0,25 mm, consideration can be made to the measurement uncertainty of
existing measurement equipment.
INOTE 3 This International Standard does not specify maximum inner diameter.
[NOTE 4 OD = outer diameter: ID = inner diameter
corto creneny Sac @ 180 2016 Allrights reserved1S0 9626:2016(E)
Table 1 (continued)
Designated metric ol
0: 0D) it
stee Gauge In MAK | Wan IN
3,00 1 2,950 3,150 Rw 2200
TW 2.464
RW 2,500
340 10 3:300 3500
TW 2,819
INOTET RW = Regular Wall TW
INOTE2 Needle sizes below 0,25 mm, consideration can be made to the measurement uncertainty of|
existing measurement equipment.
fhin Wall, ETW = Extra Thin Wall, UTW = Ultra Thin Wall
NOTE 3 This International Standard does not specify maximum inner diameter.
NOTE 4 0D = outer diameter; ID = inner diameter.
5.7. Sample size
Where sampling is applicable, the sample sizes shall be determined on the basis of risk assessment
principles and be included in the tube specification and based on the intended use.
5.8. Stiffness
When tested in accordance with Annex B, the tubing shall show a deflection not greater than the
relevant value given in Table 2.
For tubes where stiffness test parameters are not defined in this International Standard, the medical
device manufacturer shall prepare specific stiffness requirements based on a risk assessment of the
final intended use of the tube.
Consideration should be given to the final product intended use of the tube to determine if additional
tests are required.
corel wmsiora ORAM PBA SaAlh Fights reserved s:2016(E)
1S0 9626:
‘Urq jo sezis aseqy 10) Son Todor ssoUgTIs AVO@US JOU Seop plepULIG [EUONELIOIL STyy @1OFOIYA O|GETTEAE BIL FIED ON
= © © © © = 9¥O 09 00s zo 09 00s ¥e
c = . c = © S50 os os Wo 005 005 £
. . © © . . so os oor 10 os oor Lt
: © = : . © 390 oor oor BE0 oor oor ve
® e ® ® e . aso oor oe ovo oor os 1
: © 7 7 = = sv ose, ose SEO ose, ose eT
. © © veo ze ose ae oz, sz S70 ove, sz oT
© © © Zao vor os B90 vor ose v0 ot ose vt
. © = © © . 190 a osz 10 vat sz a1
© © © B07 v6 ost 760 £6 ost 120 £6 ost TT
. © © 090 ws sar 9s0 v6 su avo 06 sa 60
© © © 1s0 6 ost so 96 ost seo 96 ost v0
c © © 090 “9 ost zs “9 ost zo 9 ost £0
© . © iso et str evo a a s€0 sF ser v0
: © = . . © 3e0 ¥9 oor ve0 ¥9 ss
® e ® ® e . ovo rs oor ue0 vs so
= © 7 7 = = 150 Te oor sO v0
. © © © . . veo ve se zo Le ¥0
. © @ © © © wo oF oS wo ve os 9E0
aro oT os a0 oT os zo a 0s 00 oT os eo
ozo eT os 90 eT os Tro eT oS Tro eT os e0
wo 1 os 10 v1 os aro vt os ao TT os sco
szo 0 os zo 60 os ozo 60 os ozo 60 os eo
© . . . © © wo v0 os sco 90 os zo
© © . © . . * . . . * . aro
ua TomN uu | FOFN ua | TOFN uu | POTN
uoa2ayep sous | yoruu | wonsayep | au) | osu | uomayep | eae | ro rum | uonseyap | avo) | rosum | azysouneu
wnwixew | Supuag | ueds | wmurxey | suppuog | weds | wmwxew | surpuag | weds | wnwixey | Supuag | ueds pareudisag,
Huyamy pares w-ENIA Fujqm pares uNA-eNxa Bajqm pauea-unyL Huyamy payeareynaoy
189) ssausy]9s 40y suoNpuoD — z a1qeL,
@ 180 2016 Allrights reserved
corer Brsieavomiestnte Seman1S0 9626:2016(E)
5.9 Resistance to breakage
When tested in accordance with Annex C, the tubing shall not show visible breakage when examined by
normal or corrected vision,
NOTE Consideration can be given to the final product’s intended use to determine if additional tests are
warranted.
5.10 Resistance to corrosion
When tested in accordance with Annex D, the immersed half of the tubing shall show no evidence of
corrosion resulting from the test,
NOTE Consideration can be given to the final product's intended use to determine if additional tests are
warranted, eg, long term drug product contact.
cea moran RAR? AAaAAll Fights reserved 21S0 9626:2016(E)
Annex A
(normative)
Methods for preparation of extracts
Al Principle
‘The stainless steel tubing is immersed in water in order to extract soluble components.
A.2 Apparatus and reagents
A.2.1_ Freshly distilled or deionized water, of grade 3 in accordance with ISO 3696.
2.2 Selection of laboratory borosilicate glassware.
A3 Procedure
A.3.1_ Immerse 3 g of tubing in 250 ml of water (4.2.1) in a suitable container made from borosilicate
glass (A.2.2). Ensure that the surface of the stainless tubing, including the inside of the tubing, is in
contact with the water.
Maintain the water at a temperature of (37 +3) °C for (60 + 2) min, Remove the tubing and ensure that
all water from the inside and outside surfaces of the tubing are returned to the container.
A.3.2. Prepare the control fluid by following the procedure given in A.3.1 but omitting the tubing.
cae Roose OumeatoteSareedetin @ 180 2016 Allrights reserved1S0 9626:2016(E)
Annex B
(normative)
Test method for stiffness of tubing
B.1 Principle
A specified force is applied to the centre of the specified length of tubing, which is supported at both
ends, and the amount of deflection measured.
B.2 Apparatus
B.2.1 Stiffness testing apparatus, capable of applying a force of up to 60 N downwards normal to the
tubing with an accuracy of #0,1 N, by means of a plunger having a lower end in the form of a blunt wedge
formed by two plane surfaces inclined at 60° to one another and a cylindrical surface of radius of 1 mm
and length at least 5 mm,
An example ofa suitable apparatus is illustrated in Figure B1.
B.2.2. Equipment, capable of measuring the deflection of the tubing to the nearest 0,01 mm,
— 20°
5mm a , .
'
- ! x
60" ' Ly
\ \
\U/ 4
— pot yA fo.
! ‘
tri
; 3
: 1 2
asi
Key
1 span
2 section ACA
3 seedetail4
4 detail
Figure B.1 — Apparatus for stiffness test
ceo rmraor OoRARA PBA aAAl) Fights reserved 91S0 9626:2016(E)
B.3. Procedure
B.3.1 Place the tubing on the stiffness testing apparatus (B.2.1) and adjust the tubing and the stiffness
testing apparatus so that
a) span isas given in Table 2 for the designated metric size of the tubing,
b) the bottom surface of the plunger is at the centre of the span, and
©) the tubing is normal to the supporting members and the loading plunger, and the centre of the
‘tubing is at the centre of the span.
B32 Apply a downward force given in Table 2 for the designated metric size of the tubing at a speed
between 1 mm/min and 10 mm/min,
B33 Measure and record (3.2.2) to the nearest 0,01 mm the deflection of the tubing at the point of
application of the force.
B.4 Test report
The test report shall contain at least the following information:
a) the identity and designated metric size of the tubing:
b) whether the tubing was of RW, TW, ETW or UTW type;
©) the measured deflection, expressed in millimetres to the nearest 0,01 mm;
4) the date of testing;
©) the type of stainless steel alloy used, chosen according to Clause 4.
caret Ahn oumetone Senden @ 180 2016 Allrights reserved