Jedi Error Codes
Jedi Error Codes
3-6 DIAGNOSTICS
This chapter describes diagnostics tools based on error codes and specific
diagnostics.
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS
WARNING OFF. APPLY LOCK OUT-TAG OUT PROCEDURE FOR YOUR OWN SAFETY
WHEN MANIPULATING INSIDE THE EQUIPMENT IS REQUIRED.
3-6-1 Detailed Error Description & Troubleshooting Guide
The first part details errors, while the warning errors are listed in a second part of
this chapter.
The table below provides guidelines to troubleshoot Generator and Collimator
Jedi Error Codes
problems based on error code.
For each code, there is:
-Associated message and additional explanation related to the error
occurrence.
-List potential cause, in the order of expected probability.
Codes are sorted by ascending order both for simplified code and error code.
Refer to the theory of operation for error code structure for Generator error code.
Information about associated data structure is located at the end of each error
code subset whenever it applies.
Whenever wiring, cabling, LED check is mentioned in the recommended actions,
refer to the central listing.
Errors
Tube spits detection Errors (Code 30)
7-18
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 20 DIRECTION 2273022-100
7-19
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 20 DIRECTION 2273022-100
component failure:
0001H=RAM
0002H=RAM stack
0200H=EPLD
8000H=program checksum
7-20
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 20 DIRECTION 2273022-100
other errors:
rotation state:
0=inverter OFF
1=acceleration 0 to low speed
2=acceleration 0 to high speed
3=acceleration low speed to high
4=low speed run
5=high speed run
6=high speed to low speed brake
7=brake reverse
8=brake DC
7-22
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 20 DIRECTION 2273022-100
7-23
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 20 DIRECTION 2273022-100
7-24
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 20 DIRECTION 2273022-100
7-25
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 20 DIRECTION 2273022-100
7-26
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 20 DIRECTION 2273022-100
7-27
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 20 DIRECTION 2273022-100
7-28
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 20 DIRECTION 2273022-100
Not Proteus
AEC communication error 1. AEC board
100-1301 KV Control has detected a 2. I/F board
communication problem 3. kV control
with AEC board. 4. Wiring
7-29
PROTEUS XR/a
GE MEDICAL SYSTEMS Service Manual
REV 20 DIRECTION 2273022-100
7-30