XN-L Series Hematology Analyzer Guide
XN-L Series Hematology Analyzer Guide
XN-L series
XN-550/XN-450
/XN-350
Troubleshooting
Read this manual when you encounter a problem, and to perform instrument maintenance.
The explanations in this manual assume that you have already read "General Information".
Revision History
07/2014
Initial issue
Software version: 00-00
10/2014
Software version: 00-04
12/2014
Software version: Ver. 3
06/2015
Software version: Ver. 3
08/2015
Software version: Ver. 4
03/2016
Software version: Ver. 5
06/2016
Software version: Ver. 5
02/2017
Software version: Ver. 5
05/2017
Software version: Ver. 5
01/2018
Software version: Ver. 5
08/2018
Software version: Ver. 5
06/2019
Software version: Ver. 5
04/2020
Software version: Ver. 5
Updated manuals: Basic Operation, General Infomation
3/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
4
Revision History
4/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Table of Contents
Table of Contents
Revision History 3
Chapter 1 Troubleshooting 7
1.1 Error message list (in alphabetical order) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 [Help] dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.3 Causes of errors and remedial actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Chapter 2 Maintenance 69
2.1 List of maintenance tasks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
2.2 Displaying maintenance screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
2.3 Maintenance screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
2.4 Executing routine cleaning (XN-550) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
2.5 Executing routine cleaning (XN-450) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
2.6 Executing routine cleaning (XN-350) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
2.7 Replenishing the reagent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
2.8 Replacing the waste container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
2.9 Testing instrument operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
2.10 Checking the cycle count (counter) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
2.11 Checking the instrument status (sensor) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
2.12 Replacing the reagent. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
2.13 Replacing the piercer (XN-550/XN-450) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
2.14 Replacing the air pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
2.15 Replacing the fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
2.16 Cleaning the aspiration unit tray (XN-550/XN-450) . . . . . . . . . . . . . . . . . . . . . 94
2.17 Cleaning the aspiration unit tray (XN-350) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
2.18 Adjusting the monitor image quality (XN-550) . . . . . . . . . . . . . . . . . . . . . . . . . 98
2.19 Maintenance and inspection checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
5/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Table of Contents
Index 137
6/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Chapter 1 Troubleshooting
Note:
In the situations below, remove the sample tubes from the instrument and turn OFF the power.
• An error message marked with ☎ (phone sign) appears in the error message list.
• An error persists after the indicated action is taken.
Instrument repair or replacement of parts is necessary. Contact your Sysmex service representative.
7/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
☎ Internal Error 32
Invalid analysis item is specified 31
Invalid analysis item is specified (sampler analysis) 31
L L-J Control Error 32
☎ Laser life 31
8/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
9/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
10/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
[Error Message Displays the current errors. If multiple errors exist, errors that have higher priority are
List] displayed at the top.
[Action] Displays the troubleshooting actions for the selected error. Depending on the type of
error, this field may be blank.
[Reset Alarm] Touch to stop the alarm.
[Instruction manual] Touch to display the section of the manual that explains the selected error. This
button cannot be selected if there are no relevant sections.
[Execute]/[Accept] Depending on the error type, either the [Execute] button or the [Accept] button
appears.
Touch [Execute] to execute the action indicated in the [Action] field.
Touch [Accept] to clear the error. Execute the action indicated in the [Action] field.
[Close] Touch to close the [Help] dialog box.
Note:
• After you have closed the [Help] dialog box by touching [Close], you can re-display the dialog box by
touching [Error] in the control menu.
• Even if the [Help] dialog box does not open, you can stop the alarm by touching any part of the screen.
11/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Risk of infection
When taking remedial action to handle an error, wear adequate personal protective equipment, such as
protective gloves, a protective mask, protective eyewear, and a lab coat. Wash your hands after completing
the task.
There is a risk of infection.
The air pump has malfunctioned. Touch [Execute] in the [Help] dialog box.
An air pump test is executed.
The 0.06 MPa relief valve has malfunctioned. Touch [Execute] in the [Help] dialog box.
An air pump test is executed.
12/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Case 2: Drain the pneumatic trap chamber. For details, see the
The pneumatic trap chamber is full of water. following.
13/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
4 Turn the pneumatic trap chamber in the 6 Remove the float, and insert into the
direction of the arrow to remove. pneumatic trap chamber.
Insert the float into the pneumatic trap chamber in the
same orientation as when you removed it.
Float
Pneumatic
trap
chamber
Loosen
Example: XN-550
10 Press the power switch.
1
5 Dispose of the water in the chamber.
The pressure sensor has malfunctioned. Remove the sampler adapter (XN-550 only) and sample tubes
from the instrument, and then touch [Execute] in the [Help]
dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn OFF the main power switch. The instrument requires
servicing. Contact your Sysmex service representative.
The vacuum sensor has malfunctioned. Remove the sampler adapter (XN-550 only) and sample tubes
from the instrument, and then touch [Execute] in the [Help]
dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn OFF the main power switch. The instrument requires
servicing. Contact your Sysmex service representative.
14/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
The temperature of the unit has fallen out of the monitored Touch [Execute] in the [Help] dialog box.
range. While checking the [Sensor Display] dialog box that appears,
wait for the temperature to return within the monitored range.
(➤P.83 "Chapter 2: 2.11 Checking the instrument status
(sensor)")
If the error has not cleared after 30 minutes, contact your
Sysmex service representative.
The ambient temperature of the instrument has fallen out of Touch [Execute] in the [Help] dialog box.
the usable range (7 to 40°C). Check [Environment Temp.] in the [Sensor Display] dialog box
that appears and adjust the room temperature.
(➤P.83 "Chapter 2: 2.11 Checking the instrument status
(sensor)")
The thermistor in the unit has malfunctioned, or there is a Remove the sampler adapter (XN-550 only) and sample tubes
circuit break. from the instrument, and then touch [Execute] in the [Help]
dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn OFF the main power switch. The instrument requires
servicing. Contact your Sysmex service representative.
The temperature of the unit is not stabilizing. Remove the sampler adapter (XN-550 only) and sample tubes
from the instrument, and then touch [Execute] in the [Help]
dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn OFF the main power switch. The instrument requires
servicing. Contact your Sysmex service representative.
15/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
16/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
The reagent has run out. Replace the reagent. For details, see the following.
17/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
3 Pull out the dye holder. 5 Slowly remove the dye cover, taking
care that the dye does not drip.
Dye holder
Dye cover
Caution!
Dye may drip from the nozzle on the dye cover.
If dye drips onto the instrument, immediately wipe it off
using a moistened cloth. Otherwise, the coated surface
of the instrument may become discolored.
Example: XN-550
Fluorocell RET
Holder
New reagent cartridge
18/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
8 Place the dye cover. 9 Place the holder in the dye holder.
Information
Be sure to push the dye holder all the way in.
• The tubing connected to the reagent container is clogged. Check the condition of the tubes, and then touch [Execute] in
• There is a foreign object pressing on the tubing connected the [Help] dialog box.
to the reagent container, or there is a kink in the tubing. Reagent replacement takes place.
• The dye cover opened. Check the condition of the dye cover, and then touch
• Air bubbles have formed in the tubing connected to the [Execute] in the [Help] dialog box.
cartridge. Reagent replacement takes place.
19/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
The drain tubing is clogged. Touch [Execute] in the [Help] dialog box.
The waste chamber is drained.
If the error does not clear, rinse the waste chamber. For
details, see the following.
XN-550
(➤P.20 "Rinsing the waste chamber (XN-550)")
XN-450
(➤P.21 "Rinsing the waste chamber (XN-450)")
XN-350
(➤P.22 "Rinsing the waste chamber (XN-350)")
CELLCLEAN
2 Touch the [Rinse Instrument] icon. AUTO
20/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
4 If the sample tube holder is not ejected, 8 Wait until cleaning finishes.
press the sample tube holder open/close When the process finishes, the dialog box closes.
switch.
The sample tube holder is ejected.
21/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
3 Touch [Rinse Waste Fluid Chamber]. 5 Insert the pipette all the way to the
The [Rinse Waste Fluid Chamber] dialog box appears. bottom of the CELLCLEAN AUTO, and
press the start switch.
The drain tubing is clogged. Touch [Execute] in the [Help] dialog box.
The sample is drained from the reaction chamber.
22/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
The drain tubing is clogged. Touch [Execute] in the [Help] dialog box.
The sample is drained from the reaction chamber.
The drain tubing is clogged. Touch [Execute] in the [Help] dialog box.
The sample is drained from the reaction chamber.
Rinse cup pinch valve move error, Waste chamber1 pinch valve move error,
Waste chamber2 pinch valve move error
The pinch valve has malfunctioned. Touch [Execute] in the [Help] dialog box.
A pinch valve test is executed.
If the error does not clear, the device requires servicing.
Contact your Sysmex service representative.
The waste container is full. Replace the waste container, and then touch [Accept] in the
[Help] dialog box. To replace the waste container, see the
following.
23/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Caution!
Install the waste container below the bottom of the
instrument.
There is a water leak inside the instrument. Touch [Execute] in the [Help] dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn OFF the main power switch, and wipe off the water when
the instrument is wet. The instrument requires servicing.
Contact your Sysmex service representative.
There is a water leak inside the instrument. Touch [Execute] in the [Help] dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn OFF the main power switch. The instrument requires
servicing. Contact your Sysmex service representative.
The water leak sensor has malfunctioned. Remove the sampler adapter (XN-550 only) and sample tubes
from the instrument, and then touch [Execute] in the [Help]
dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn OFF the main power switch. The instrument requires
servicing. Contact your Sysmex service representative.
CELLPACK DST has run out. Check the RU-20. If being prepared, wait briefly.
If the error does not clear, the replacement of CELLPACK DST is
required. For details, see "RU-20 Instructions for Use".
(➤RU-20 Instructions for Use, "Chapter 6: 6.3 Replacing the
reagent")
24/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
The whole blood aspiration motor has malfunctioned. Touch [Execute] in the [Help] dialog box.
A whole blood aspiration motor test is executed.
If the error does not clear, the instrument requires servicing.
Contact your Sysmex service representative.
The sheath motor has malfunctioned. Touch [Execute] in the [Help] dialog box.
A sheath motor test is executed.
If the error does not clear, the instrument requires servicing.
Contact your Sysmex service representative.
The aspiration unit up-down motor has malfunctioned. Touch [Execute] in the [Help] dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn the main power switch OFF and then ON.
If the error does not clear, the instrument requires servicing.
Contact your Sysmex service representative.
The aspiration unit front-back motor has malfunctioned. Touch [Execute] in the [Help] dialog box.
An aspiration unit motor test is executed.
If the error does not clear, the instrument requires servicing.
Contact your Sysmex service representative.
The sample tube holder motor has malfunctioned. Remove the sample tubes, and then touch [Execute] in the
[Help] dialog box.
A sample tube holder motor test is executed.
If the error does not clear, the instrument requires servicing.
Contact your Sysmex service representative.
25/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
26/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
CELLCLEAN
AUTO
Wait until this process finishes. When the process Regular sample
finishes, the dialog box closes and a background tube holder
check starts.
27/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
6 Press the start switch. 4 If the sample tube holder is not ejected,
press the sample tube holder open/close
switch.
The sample tube holder is ejected.
28/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Keep the opener attached, and remove immediately 7 Wait until cleaning finishes.
before you use CELLCLEAN AUTO.
When the process finishes, the dialog box closes, and
the instrument power automatically turns OFF.
The aspiration sensor has malfunctioned. Touch [Accept] in the [Help] dialog box.
The instrument requires servicing. Contact your Sysmex
service representative.
29/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
A sample tube was still in the sample tube holder when Remove the sample tube, and then touch [Accept] in the
analysis started. The hand did not work properly during [Help] dialog box.
sampler analysis.
Both a regular sample tube and a micro collection tube were Remove the sample tube that will not be analyzed, and then
set for manual analysis. touch [Accept] in the [Help] dialog box.
The set amount of time (60 seconds) elapsed after the host Pull out the sampler adapter holder and remix the sample
computer was queried for analysis information during sampler tube in the starting tube position (adapter number – test
analysis. tube position) shown in the instrument status on the left side
of the control menu. Replace the sample tube in its original
position, push in the sampler adapter holder, and touch
[Accept] in the [Help] dialog box.
No sample number was specified at the time of manual Touch [Accept] in the [Help] dialog box.
analysis. Enter the sample number and then execute the analysis.
The barcode label is dirty. Touch [Accept] in the [Help] dialog box. Check the position
The print condition of the barcode label is poor. and orientation of the barcode label, and if the barcode label is
The position or orientation of the barcode label is poor. dirty.
No analysis item was specified at the time of manual analysis. Touch [Accept] in the [Help] dialog box.
Specify the analysis item and then execute analysis.
30/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
An invalid analysis item is specified. Touch [Accept] in the [Help] dialog box.
Change the analysis item and then execute analysis.
During sampler analysis, a sample had an analysis parameter Touch [Accept] in the [Help] dialog box. The error sample is
that cannot be analyzed. skipped and sampler analysis continues. When sampler
analysis finishes, change the analysis parameter of the
skipped sample and then analyze the sample.
It is time to replace the laser. Laser replacement is required. Contact your Sysmex service
representative.
• The FCM cover opened. Remove the sampler adapter (XN-550 only) and sample tubes
• The FCM cover sensor has malfunctioned. from the instrument, and then touch [Execute] in the [Help]
dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn OFF the main power switch. The instrument requires
servicing. Contact your Sysmex service representative.
There was a communication error with the barcode reader. Remove the sampler adapter (XN-550 only) and sample tubes
from the instrument, and then touch [Execute] in the [Help]
dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn the main power switch OFF and then ON.
If the error persists, contact your Sysmex service
representative.
31/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
There was a communication error with the RFID reader/writer. Remove the sampler adapter (XN-550 only) and sample tubes
from the instrument, and then touch [Execute] in the [Help]
dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn the main power switch OFF and then ON.
If the error persists, contact your Sysmex service
representative.
Internal Error
An error occurred in the operation of the program. Remove the sampler adapter (XN-550 only) and sample tubes
from the instrument, and then touch [Execute] in the [Help]
dialog box.
Instrument shutdown is executed. When shutdown is finished,
turn OFF the main power switch. The instrument requires
servicing. Contact your Sysmex service representative.
A QC error (value out of the control limit range) occurred. Touch [Accept] in the [Help] dialog box.
Check the analysis results in the [QC Chart] screen, and
perform calibration if necessary. For details on checking the
QC results, see "Basic Operation".
(➤Basic Operation, "Chapter 3: 3.7 Checking analysis results")
The control blood has expired. Touch [Accept] in the [Help] dialog box.
Use new control blood to execute QC analysis.
When using a new lot, the lot information must be registered in
a new QC file. For details on registering lot information, see
"Basic Operation".
(➤Basic Operation, "Chapter 3: 3.5 Registering lots")
Control blood with an unregistered lot number was used. Touch [Accept] in the [Help] dialog box.
Register the control blood lot information. For details on
registering lot information, see "Basic Operation".
(➤Basic Operation, "Chapter 3: 3.5 Registering lots")
32/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
QC not executed.
A numerical value for a parameter in the scattergram is Touch [Accept] in the [Help] dialog box.
outside the specified range. Check the scattergram in the [Data Browser] screen. For
details on the [Data Browser] screen, see "Basic Operation".
(➤Basic Operation, "Chapter 6: Checking Detailed Analysis
Information (Data Browser)")
The analysis mode is not compatible with the control blood Touch [Accept] in the [Help] dialog box.
type. Change the analysis mode and then analyze.
Air bubbles have formed in the detector. Touch [Accept] in the [Help] dialog box.
The detector is clogged. Execute auto rinse. For details, see the following.
The detector is dirty.
33/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Draining the sample from the RBC Removing the RBC detector clog
isolation chamber
1 Touch the [Maintenance] icon in the
1 Touch the [Maintenance] icon in the [Menu] screen.
[Menu] screen. The [Maintenance] menu screen appears.
The [Maintenance] menu screen appears.
2 Touch the [Rinse Instrument] icon.
2 Touch the [Rinse Instrument] icon. The [Rinse Instrument] dialog box appears.
The [Rinse Instrument] dialog box appears.
3 Touch [Remove RBC Detector Clog].
3 Touch [Drain RBC Isolation Chamber]. Clog removal starts, and the [Remove RBC Detector
Clog] dialog box appears.
Draining starts, and the [Drain RBC Isolation
Chamber] dialog box appears.
34/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Warning!
Never touch the detector when the power of the
instrument is turned ON.
An electrical shock could occur.
Caution!
• Be sure to use CELLCLEAN AUTO only.
• When closing the detector cover, take care not to
kink the tube. It may lead to incorrect analysis.
• When rinsing the detector aperture, use the
supplied unclogging brush and lightly tap on the
detector aperture.
Excessive force will damage the detector aperture.
Opener
Caution!
• Always wear adequate personal protective Right cover
equipment when opening CELLCLEAN AUTO.
• Press down slowly so that the fluid does not spill.
• Store opened CELLCLEAN AUTO standing on the Example: XN-550
rack with the opener attached. If CELLCLEAN
AUTO is tilted, the fluid may leak even with the
opener attached.
3 Touch [OK].
The instrument power automatically turns OFF.
35/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
5 Loosen the screw that fastens the 7 Turn the lid of the detector chamber in
detector cover. the direction of the arrow and pull out.
Detector
cover
Detector aperture
Example: XN-550
Detector cover
Unclogging brush
Note:
If fluid spills, wipe off the spilled fluid with a piece of
tissue paper.
36/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
9 Insert the lid of the detector chamber 10 Install the detector cover, and fasten the
straight in, and turn in the direction of screw with the flathead screwdriver.
the arrow to lock.
Insert the lid of the detector chamber all the way in, 11 Close the right cover and lock.
and turn to the position shown.
Information
Wash the brush and opener well and store in a clean
state.
There is a risk of instrument malfunctioning if there
are small particles or other contaminants on the brush
or opener.
Caution!
When the lid of the detector chamber is not properly
attached, correct results are not obtained.
The instrument may be damaged if leakage occurs.
37/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
CELLCLEAN
AUTO 2 Touch the [Rinse Instrument] icon.
The [Rinse Instrument] dialog box appears.
CELLCLEAN
AUTO
Regular
sample tube
holder
38/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Opener
Caution!
• Always wear adequate personal protective
equipment when opening CELLCLEAN AUTO.
• Press down slowly so that the fluid does not spill.
• Store opened CELLCLEAN AUTO standing on the
rack with the opener attached. If CELLCLEAN
AUTO is tilted, the fluid may leak even with the
opener attached.
39/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
5 Insert the pipette all the way to the 6 Remove CELLCLEAN AUTO from the
bottom of the CELLCLEAN AUTO, and pipette.
press the start switch. Remove CELLCLEAN AUTO without bending the
pipette.
Aspiration starts.
During aspiration, the analysis status indicator LED
blinks green and the beeper sounds repeatedly.
When aspiration finishes, the analysis status indicator
LED turns off and the beeping stops. Cleaning starts.
3 Touch [OK].
The instrument power automatically turns OFF.
40/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
3 Turn OFF the main power switch. 5 Loosen the screw that fastens the
detector cover.
Detector
cover
Example: XN-550
Detector cover
Right cover
Example: XN-550
41/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
7 Turn the lid of the detector chamber in 9 Insert the lid of the detector chamber
the direction of the arrow and pull out. straight in, and turn in the direction of
the arrow to lock.
Insert the lid of the detector chamber all the way in,
and turn to the position shown.
Lid of the detector
chamber
Caution!
When the lid of the detector chamber is not properly
attached, correct results are not obtained.
The instrument may be damaged if leakage occurs.
Detector aperture
11 Close the right cover and lock.
Information
Wash the brush and opener well and store in a clean
state.
Unclogging brush There is a risk of instrument malfunctioning if there
are small particles or other contaminants on the brush
or opener.
Note:
If fluid spills, wipe off the spilled fluid with a piece of
tissue paper.
HGB error
The HGB blank value or the HGB sample value has fallen out Touch [Accept] in the [Help] dialog box.
of the specified range. Re-analyze.
42/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Case 2: Make sure there are no sources of noise near the instrument.
Because of external noise, the particle count has exceeded Move any noise sources away from the instrument and re-
the limit of the display range. analyze.
43/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
CELLCLEAN
AUTO 2 Touch the [Rinse Instrument] icon.
The [Rinse Instrument] dialog box appears.
CELLCLEAN
AUTO
Regular
sample tube
holder
44/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Opener
Aspiration starts.
During aspiration, the analysis status indicator LED
blinks green and the beeper sounds repeatedly.
[Rinse Flowcell] dialog box When aspiration finishes, the analysis status indicator
LED turns off and the beeping stops. Cleaning starts.
45/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Data Errors
An analysis value was outside the set upper/lower limit range. Touch [Accept] in the [Help] dialog box. Check the analysis
data and review the set upper and lower limit values.
When body fluid analysis was executed, analysis data with Remove the sample tubes from the instrument, and then
high values that may affect the next analysis results were touch [Execute] in the [Help] dialog box.
obtained. A background check is executed.
46/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
The piercer (XN-550/XN-450), pipette (XN-350), or whole Touch [Accept] in the [Help] dialog box. Execute auto rinse.
blood aspiration line tubing is clogged. For details, see the following.
(➤P.47 "Executing auto rinse")
XN-550
(➤P.47 "Executing routine cleaning (XN-550)")
XN-450
(➤P.48 "Executing routine cleaning (XN-450)")
XN-350
(➤P.49 "Executing routine cleaning (XN-350)")
2 Touch [Yes].
Auto rinse starts and the [Auto Rinse] dialog box
appears.
[Auto Rinse] dialog box 4 Open the sampler cover (manual unit).
Wait until this process finishes. When the process
finishes, the dialog box closes and a background
check starts.
47/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
5 Place CELLCLEAN AUTO in the sample 2 Touch the [Rinse Instrument] icon.
tube holder. The [Rinse Instrument] dialog box appears.
Place in the regular sample tube holder.
3 Touch [Routine Cleaning].
CELLCLEAN The [Routine Cleaning] dialog box appears.
AUTO
Regular sample
tube holder
48/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Opener
Aspiration starts.
During aspiration, the analysis status indicator LED
blinks green and the beeper sounds repeatedly.
[Routine Cleaning] dialog box When aspiration finishes, the analysis status indicator
LED turns off and the beeping stops. Cleaning starts.
49/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
The FCM detector cover is open. Contact your Sysmex service representative.
The dye holder has been pulled out. Push the dye holder into the instrument.
The right cover opened while the instrument was in operation. Close the right cover and then touch [Execute] in the [Help]
dialog box.
When shutdown is finished, turn OFF the main power switch.
Turn ON the main power switch again and press the power
switch.
The sample tube holder opened while the instrument was in Touch [Execute] in the [Help] dialog box.
operation. When shutdown is finished, turn OFF the main power switch.
Turn ON the main power switch again and press the power
switch.
50/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
It is time to replace the piercer. Replace the piercer (XN-550/XN-450). Contact your Sysmex
service representative.
It is time to execute routine cleaning. Touch [Accept] in the [Help] dialog box.
(The cumulative analysis count since the previous routine Execute routine cleaning. For details, see the following.
cleaning was performed has exceeded 1,000, or 1 week has XN-550
elapsed.) (➤P.51 "Executing routine cleaning (XN-550)")
XN-450
(➤P.52 "Executing routine cleaning (XN-450)")
XN-350
(➤P.53 "Executing routine cleaning (XN-350)")
Routine cleaning must be executed. Touch [Execute] in the [Help] dialog box.
(The cumulative analysis count since the previous routine The [Routine Cleaning] dialog box appears. Execute routine
cleaning was performed has exceeded 1,400.) cleaning. For details, see the following.
XN-550
(➤P.51 "Executing routine cleaning (XN-550)")
XN-450
(➤P.52 "Executing routine cleaning (XN-450)")
XN-350
(➤P.53 "Executing routine cleaning (XN-350)")
51/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
CELLCLEAN
AUTO
1 Touch the [Maintenance] icon in the
[Menu] screen.
The [Maintenance] menu screen appears.
Regular sample
tube holder 2 Touch the [Rinse Instrument] icon.
The [Rinse Instrument] dialog box appears.
52/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
CELLCLEAN
AUTO
Regular
sample tube
holder
Opener
53/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
5 Insert the pipette all the way to the 6 Remove CELLCLEAN AUTO from the
bottom of the CELLCLEAN AUTO, and pipette.
press the start switch. Remove CELLCLEAN AUTO without bending the
pipette.
Aspiration starts.
During aspiration, the analysis status indicator LED
blinks green and the beeper sounds repeatedly.
When aspiration finishes, the analysis status indicator
LED turns off and the beeping stops. Cleaning starts.
It is time to replace the air pump. Touch [Accept] in the [Help] dialog box.
Replace the air pump. For details, see the following.
Replacing the air pump 3 Touch the [Menu] button on the toolbar.
The [Menu] screen appears.
54/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
2 Turn OFF the main power switch, and 4 Remove the screws (x2) with a Phillips
unplug the power cable. head screwdriver, and remove the upper
cover.
Upper cover
Main power
ON switch
OFF
Plug
Screws (x2)
Example: XN-550
Right cover
Example: XN-550
55/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
5 Loosen the screws (x2) with a Phillips 7 Remove the connectors and tubes (x2),
head screwdriver. and remove the air pump unit.
Screws (x2) Tube (P1)
Connector
Tube (V1)
Information
Connect the tube (P1) to the upper side.
56/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
The reagent has expired. Touch [Execute] in the [Help] dialog box.
The [Reagent Replacement] dialog box appears. Replace the
reagent. For details, see the following.
Replacing the diluent and hemolytic When the reagent code is entered,
the reagent in the dialog box shows
agent [Received]. The lot number and
expiration date show the
1 Prepare a new reagent. information of the new reagent.
Make sure the reagent has not expired.
Caution! Note:
• Place the reagent container at a level no more than
When inputting manually, touch the name of the
1 meter above or below the bottom of the analyzer.
reagent to be replaced in the [Reagent Replacement]
Do not place reagents on top of the instrument.
dialog box. The dialog box below appears.
• The new reagent must be allowed to sit for at least
24 hours at room temperature (15 to 35°C) before
use.
• If reagent spills, immediately wipe up the spill
using a moistened cloth.
3 Input the reagent code (barcode) of the Touch [Replace the reagent.] to select the checkbox,
new reagent. enter the [Reagent Code :], and touch [OK].
Input by reading the barcode
Scan the reagent code (barcode) on the outer box of
the new reagent with the hand-held barcode reader. 4 Remove the cap from the new reagent
The reagent code is as shown below.
container.
Make sure the barcode scanning surface is flat before
scanning.
5 Remove the cap from the old reagent
container.
Note:
In case the reagent outer box label shows a "XN
Reagent Code" or "Reagent Code 2" barcode, Example: Lysercell WDF (2 L)
please scan this barcode.
57/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
6 Pull the spout set straight up and out. 7 Insert the spout set straight into the new
reagent container and close the cap.
8 Touch [Execute].
Reagent replacement starts. Wait until this process
finishes. When replacement finishes, the [Reagent
Replacement] dialog box closes.
The reagent has expired. Replace the reagent. For details, see the following.
58/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
3 Pull out the dye holder. 5 Slowly remove the dye cover, taking
care that the dye does not drip.
Dye holder
Dye cover
Caution!
Dye may drip from the nozzle on the dye cover.
If dye drips onto the instrument, immediately wipe it off
using a moistened cloth. Otherwise, the coated surface
of the instrument may become discolored.
Example: XN-550
Fluorocell RET
Holder
New reagent cartridge
59/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
8 Place the dye cover. 9 Place the holder in the dye holder.
Information
Be sure to push the dye holder all the way in.
Press Start SW
This appears when the instrument is in the sleep state. Press the start switch. The fluid path is rinsed and the
instrument wakes up from the sleep state.
On the XN-550, when the instrument is in sampler analysis
mode, it can also be woken from the sleep state by pressing
the sampler start/stop switch.
An incorrect dye type has been installed. Remove the incorrect reagent, wipe the nozzle of the dye
holder, and then install the correct reagent.
Dispose of the incorrect reagent.
The dye is not installed. Install the reagent. For details, see the following.
(➤P.61 "Replacing the dye")
60/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Fluorocell WDF has already been used, Fluorocell RET has already been used
A used dye has been installed. Replace the reagent. For details, see the following.
Fluorocell RET
Holder
61/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
5 Slowly remove the dye cover, taking 8 Place the dye cover.
care that the dye does not drip.
Dye cover
Dye cover
62/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Cannot recognize Fluorocell WDF information, Cannot recognize Fluorocell RET information
The RFID tag on the reagent cartridge is damaged. Touch [Execute] in the [Help] dialog box.
The [Reagent Replacement] dialog box appears. Touch the
error dye and register the reagent information in the dialog
box that appears. For details, see the following.
Note:
[Accept] is enabled 5 seconds after the dialog box
2 Touch the dye that indicates appears. If the lot number is invalid and appears in
[Exchange]. red, [Accept] will not be enabled.
63/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
Fluorocell WDF RFID tag error, Fluorocell RET RFID tag error
Unable to write the information on the RFID tag of the reagent Replace the reagent (a reagent with a normal RFID tag). For
cartridge. details, see the following.
Fluorocell RET
Holder
64/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
5 Slowly remove the dye cover, taking 8 Place the dye cover.
care that the dye does not drip.
Dye cover
Dye cover
65/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
The sampler cover (front) is open. Close the sampler cover (front).
The sampler cover (front) opened while the sampler analysis Remove the sample tubes from the instrument. When the
was in operation. hand is gripping a sample tube, open the sampler cover (front
or manual unit) and remove the sample tubes. Close the
sampler cover (front and manual unit) and then touch
[Execute] in the [Help] dialog box. The sampler test is
executed.
If an abnormality occurs during the operation test, the sensor
that detects the cover has failed. Contact your Sysmex service
representative.
Sampler analysis cannot be performed, but manual analysis is
possible.
The sample cover (manual unit) is open. Close the sample cover (manual unit).
The sampler cover (manual unit) opened while the sampler Remove the sample tubes from the instrument. When the
analysis was in operation. hand is gripping a sample tube, open the sampler cover (front
or manual unit) and remove the sample tubes. Close the
sampler cover (front and manual unit) and then touch
[Execute] in the [Help] dialog box. The sampler test is
executed.
If an abnormality occurs during the operation test, the sensor
that detects the cover has failed. Contact your Sysmex service
representative.
Sampler analysis cannot be performed, but manual analysis is
possible.
The sampler adapter holder (right)/(left) is open. Close the sampler adapter holder (right)/(left).
66/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
The sampler adapter holder (right)/(left) opened while the Remove the sample tubes from the instrument. When the
sampler analysis was in operation. hand is gripping a sample tube, open the sampler cover (front
or manual unit) and remove the sample tubes. Close the
sampler cover (front and manual unit) and then touch
[Execute] in the [Help] dialog box. The sampler test is
executed.
If an abnormality occurs during the operation test, the sensor
that detects the cover has failed. Contact your Sysmex service
representative.
Sampler analysis cannot be performed, but manual analysis is
possible.
Sampler adapter is not placed. Place the sampler adapter and press the sampler start/stop
switch.
An unregistered sampler adapter was placed. Place an appropriate sampler adapter, and then execute
The sampler adapter is damaged. sampler analysis.
If the error persists, execute manual analysis, or execute
sampler analysis using another sampler adapter.
The hand did not work properly. Remove the sample tubes from the instrument. When the hand
is gripping a sample tube, open the sampler cover (front or
manual unit) and remove the sample tubes. Close the sampler
cover (front and manual unit) and then touch [Execute] in the
[Help] dialog box. The sampler test is executed.
If an abnormality occurs during the operation test, the hand
has failed. Contact your Sysmex service representative.
Sampler analysis cannot be performed, but manual analysis is
possible. Before executing manual analysis, check if the hand
is touching the sample tube holder. If so, move the hand.
67/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 1 Troubleshooting
68/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
Chapter 2 Maintenance
This chapter describes the procedures for maintenance, replacing and replenishing supplies, and checking
instrument operation.
Risk of infection
While working, always wear adequate personal protective equipment, such as protective gloves, a
protective mask, protective eyewear, and a lab coat. Wash your hands after completing the task.
There is a risk of infection.
69/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
70/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
[Operation Opens the [Operation [Whole Blood Touch to start the whole blood
Confirmation] Confirmation] dialog box. Select Aspiration Motor aspiration motor test.
the unit you want to check. Test] (➤P.79 "2.9.1 Motor test")
(➤P.71 "● [Operation
[Sheath Motor Touch to start the sheath motor
Confirmation] dialog box")
Test] test.
[Exchange] Opens the [Exchange] dialog (➤P.79 "2.9.1 Motor test")
box. Select the item you want to
[Aspiration Unit Touch to start the aspiration unit
replace.
Motor Test] motor test.
(➤P.72 "● [Exchange] dialog
(➤P.79 "2.9.1 Motor test")
box")
[Air Pump Test] Touch to start the air pump test.
[Rinse Opens the [Rinse Instrument]
(➤P.79 "2.9.2 Air pump test")
Instrument] dialog box. Select the unit to be
rinsed and the rinsing method. [Pinch Valve Touch to start the pinch valve test.
(➤P.72 "● [Rinse Instrument] Test] (➤P.80 "2.9.3 Pinch valve test")
dialog box") [Tube Holder Touch to start the sample tube
[Status Opens the [Status Confirmation] Motor Test]*1 holder motor test.
Confirmation] dialog box. Select the (➤P.79 "2.9.1 Motor test")
information (analysis or cycle [Barcode Use to test the barcode reader in
count/unit status) you want to Reading Test]*2 the instrument.
check. (➤P.80 "2.9.4 Barcode reader
(➤P.73 "● [Status Confirmation] reading test (XN-550)")
dialog box")
[Sampler Touch to start the sampler test.
[Back] Touch to return to the [Menu] Operation Test]*2 (➤P.81 "2.9.5 Sampler operation
screen. test (XN-550)")
[Close] The dialog box closes.
71/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
[Exchange] dialog box [Drain RBC Touch to drain the sample that has
Isolation collected in the RBC isolation
[Reagent Touch to display the [Reagent Chamber] chamber.
Replacement] Replacement] dialog box. Use this (➤P.34 "Chapter 1: Draining the
dialog box to replace a reagent. sample from the RBC isolation
(➤P.84 "2.12 Replacing the chamber")
reagent") [Remove RBC Touch to remove a clog from the
[Close] The dialog box closes. Detector Clog] RBC detector.
(➤P.34 "Chapter 1: Removing the
RBC detector clog")
● [Rinse Instrument] dialog box
[Routine Touch to execute routine cleaning.
Cleaning] XN-550
(➤P.74 "2.4 Executing routine
cleaning (XN-550)")
XN-450
(➤P.75 "2.5 Executing routine
cleaning (XN-450)")
XN-350
(➤P.76 "2.6 Executing routine
cleaning (XN-350)")
[Reagent Use to replace the reagent in the
Replenishment] fluid line.
[Rinse Instrument] dialog box (➤P.77 "2.7 Replenishing the
reagent")
[Drain Waste Touch to drain the waste fluid that [Close] The dialog box closes.
Fluid Chamber] has collected in the waste fluid
chamber.
[Rinse Waste Touch to rinse the waste fluid
Fluid Chamber] chamber.
XN-550
(➤P.20 "Chapter 1: Rinsing the
waste chamber (XN-550)")
XN-450
(➤P.21 "Chapter 1: Rinsing the
waste chamber (XN-450)")
XN-350
(➤P.22 "Chapter 1: Rinsing the
waste chamber (XN-350)")
[Remove Touch to remove bubbles from the
Flowcell Air flowcell.
Bubbles] (➤P.46 "Chapter 1: Removing air
bubbles from the flowcell")
72/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
73/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
1 Touch the [Maintenance] icon in the 5 Place CELLCLEAN AUTO in the sample
[Menu] screen. tube holder.
The [Maintenance] menu screen appears. Place in the regular sample tube holder.
74/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
1 Touch the [Maintenance] icon in the 5 Place CELLCLEAN AUTO in the sample
[Menu] screen. tube holder.
The [Maintenance] menu screen appears. Place in the regular sample tube holder.
Regular
sample tube
holder
75/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
1 Touch the [Maintenance] icon in the 5 Insert the pipette all the way to the
[Menu] screen. bottom of the CELLCLEAN AUTO, and
The [Maintenance] menu screen appears. press the start switch.
Aspiration starts.
During aspiration, the analysis status indicator LED
blinks green and the beeper sounds repeatedly.
When aspiration finishes, the analysis status indicator
LED turns off and the beeping stops. Cleaning starts.
[Routine Cleaning] dialog box
Caution!
• Always wear adequate personal protective
equipment when opening CELLCLEAN AUTO.
• Press down slowly so that the fluid does not spill.
• Store opened CELLCLEAN AUTO standing on the
rack with the opener attached. If CELLCLEAN
AUTO is tilted, the fluid may leak even with the
opener attached.
76/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
6 Touch [OK].
Reagent replenishing starts.
Wait until this process finishes. When the process
finishes, the dialog box closes.
77/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
Waste container
78/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
Name Operation
Whole blood Aspirates and discharges
1 Touch the [Maintenance] icon in the
aspiration motor samples. [Menu] screen.
The [Maintenance] menu screen appears.
Sheath motor Feeds reaction sample to the
FCM detector or RBC/PLT
detector. 2 Touch the [Operation Confirmation]
Aspiration unit motor Moves the piercer (XN-550/ icon.
XN-450) or pipette (XN-350) The [Operation Confirmation] dialog box appears.
that aspirates the sample.
Sample tube holder Ejects and retracts the 3 Touch [Air Pump Test].
motor* sample tube holder.
The operation test starts, and the [Air Pump Test]
* XN-550/XN-450 only.
dialog box appears.
79/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
Reading result
[Barcode Reading Test] dialog box
80/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
10 Touch [Cancel].
The dialog box closes.
81/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
[Pump] tab
1 Touch the [Maintenance] icon in the
[Menu] screen.
The [Maintenance] menu screen appears.
3 Touch [Counter].
The [Counter] dialog box appears.
[Unit] tab
[Measurement] tab
4 Touch [Close].
The dialog box closes.
82/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
3 Touch [Sensor].
The [Sensor Display] dialog box appears.
Sensor status display area
83/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
2.12.1 [Reagent Replacement] Lot No. Displays the lot number of the
dialog box reagent.
Reagent state Displays the remaining number
The [Reagent Replacement] dialog box allows you to of tests that can be executed
check how much reagent remains and replace reagents. with the reagent.
For CELLPACK DCL, the
remaining volume appears.
1 Touch the [Maintenance] icon in the The remaining number of tests is
[Menu] screen. an approximate value. The value
The [Maintenance] menu screen appears. will vary depending on
conditions of use. The value is
2 Touch the [Exchange] icon. not displayed if the reagent has
not been registered. When the
The [Exchange] dialog box appears.
reagent runs low, the
background becomes yellow.
3 Touch [Reagent Replacement]. During diluent or hemolytic
agent replacement, progress is
The [Reagent Replacement] dialog box appears.
indicated as "0 to 100%".
Reagent name Displays the reagent name.
Remaining The remaining volume of the
volume graph reagent is shown as a graph.
This is not displayed if the
reagent has not been registered,
or if the reagent has run out.
Note:
The [Reagent Replacement] dialog box can also be
displayed by touching the remaining reagent display
in the control menu.
In addition, touching [Execute] in the [Help] dialog box
when a reagent replacement error message appears
also displays the [Reagent Replacement] dialog box.
In this case, only the reagent for which the error
[Reagent Replacement] dialog box occurred can be replaced.
Lot No.
Remaining volume
graph
Reagent state
Reagent name
84/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
Caution!
• Place the reagent container at a level no more than
1 meter above or below the bottom of the analyzer.
Do not place reagents on top of the instrument.
• The new reagent must be allowed to sit for at least
24 hours at room temperature (15 to 35°C)* before
use.
Touch [Replace the reagent.] to select the checkbox,
• If reagent spills, immediately wipe up the spill
enter the [Reagent Code :], and touch [OK].
using a moistened cloth.
* Use CELLPACK DST at 15 to 30°C.
Note:
In case the reagent outer box label shows a "XN
Reagent Code" or "Reagent Code 2" barcode,
please scan this barcode.
85/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
7 Insert the spout set straight into the new 3 Pull out the dye holder.
reagent container and close the cap.
Dye holder
8 Touch [Execute].
Reagent replacement starts. Wait until this process
finishes. When replacement finishes, the [Reagent
Replacement] dialog box closes.
Caution!
While performing replacement, always wear
adequate personal protective equipment.
Example: XN-550
Fluorocell RET
86/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
5 Slowly remove the dye cover, taking 8 Place the dye cover.
care that the dye does not drip.
Dye cover
Dye cover
87/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
88/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
Caution, Hot!
The surface of the air pump is hot. After turning OFF the main power, make sure that the air pump has
cooled sufficiently before replacing it.
The air pump must be replaced once every 30,000 analyses, or once every 2 years. Continuing to use the old air
pump may result in breakage or other problems.
When [Replace air pump.] appears, promptly replace the air pump.
A replacement air pump is not included with the instrument. Contact your local Sysmex representative to purchase a
replacement air pump.
1 Shut down the instrument. 2 Turn OFF the main power switch, and
unplug the power cable.
1 Make sure the instrument is in the
ready state.
If the instrument status at the bottom left of the
screen is not green, wait until it becomes green.
5 Touch [OK].
The instrument power automatically turns OFF.
89/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
3 Open the right cover. 5 Loosen the screws (x2) with a Phillips
Release the lock by turning to the left with a flathead head screwdriver.
screwdriver, and open the right cover.
Screws (x2)
Right cover
Example: XN-550
Upper cover
Example: XN-550
Caution!
Take care not to drop the screws.
It may be very difficult to remove a screw that falls into
the instrument.
Screws (x2)
Example: XN-550
90/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
Tube (P1)
Connector
Tube (V1)
Information
Connect the tube (P1) to the upper side.
91/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
Warning!
• Always unplug the power cable before replacing a fuse.
An electrical shock could occur.
• Use only a fuse of the specified type and rating.
Doing so could cause a smoke emission.
1 Turn OFF the main power switch, and 2 Remove the old fuse.
unplug the power cable.
1 Lift the tab on the fuse holder with a
flathead screwdriver and pull out the
fuse holder.
Fuse holder
OFF
Plug
Fuse
Fuse holder
92/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
93/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
If the aspiration unit tray becomes dirty, clean the aspiration unit tray.
Risk of infection
Always wear adequate personal protective equipment, such as protective gloves, a protective mask,
protective eyewear, and a lab coat when cleaning the aspiration unit tray. Wash your hands after
completing the task.
There is a risk of infection if the tray is contaminated with blood or other material.
5 Touch [OK].
The instrument power automatically turns OFF.
Example: XN-550
94/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
4 Remove the screw (x1) with a Phillips 7 Make sure no dirt remains, and wipe off
head screwdriver. the water.
Screw
8 Install the aspiration unit tray in the
instrument, and fasten with the screw
(x1) with a Phillips head screwdriver.
See the illustrations of steps 4 and 5.
Example: XN-550
Caution!
Take care not to drop the screws you removed.
It may be very difficult to remove a screw that falls
into the instrument.
95/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
If the aspiration unit tray becomes dirty, clean the aspiration unit tray.
Risk of infection
Always wear adequate personal protective equipment, such as protective gloves, a protective mask,
protective eyewear, and a lab coat when cleaning the aspiration unit tray. Wash your hands after
completing the task.
There is a risk of infection if the tray is contaminated with blood or other material.
4 Touch [OK].
The instrument power automatically turns OFF.
Right cover
96/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
Information
Install so that the protrusion on the aspiration unit
tray is on the right side.
97/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
2.18.1 Displaying the OSD screen 2.18.2 Closing the OSD screen
Page 2
3 Change the setting.
• To increase a value, press the button.
• To decrease a value, press the button.
98/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 2 Maintenance
Setting item/
Pages Description Setting range Default setting
function
1 Contrast Adjustment of contrast of all RGB colors 0 to 255 55
H Position Horizontal position adjustment 28 to 228 128
V Position Vertical position adjustment 105 to 151* 128
HTotal (Clock) Horizontal total count setting 64 to 192 128
Phase Phase setting (flickering adjustment) 0 to 63 0
Contrast Red Red contrast adjustment (digital setting) 0 to 255 50
Contrast Green Green contrast adjustment (digital setting) 0 to 255 50
Contrast Blue Blue contrast adjustment (digital setting) 0 to 255 50
Backlight Brightness adjustment 0 to 31 31
AutoAdjust Auto position adjustment -
AutoColor Auto color adjustment -
Default Initial setting -
Save Data Data save -
Cancel Cancel or OSD screen closing -
Option Display of page 2 -
2 Red Gain Red ADC gain adjustment (analog setting) 0 to 255 128
Green Gain Green ADC gain adjustment (analog setting) 0 to 255 128
Blue Gain Blue ADC gain adjustment (analog setting) 0 to 255 128
Red Offset Red ADC offset adjustment (analog setting) 0 to 255 128
Green Offset Green ADC offset adjustment (analog setting) 0 to 255 128
Blue Offset Blue ADC offset adjustment (analog setting) 0 to 255 128
Filter Space filter setting Non Non
Gamma Gamma value setting 0.1 to 3.0 1.0
OSD HPosition OSD horizontal position adjustment 3 to 255 128
OSD VPosition OSD vertical position adjustment 41 to 211 128
OSD Effect OSD transparency effect setting 0 to 7 0
OSD Timeout Timeout setting for lights-out of OSD Non, 5 to 10, 15, 10
20, 30, 40, 50, 60
Default Initial setting -
Save Data Data save -
Cancel Display of page 1 -
* Value when SVGA 60 Hz is input. The range varies depending on the input resolution and vertical frequency.
99/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Example:
100/138
2.19
Chapter 2
Signed
Routine cleaning
Maintenance tasks once every 2 years Reagents and supply parts replacement
Maintenance task M/D Signed M/D Signed Maintenance task M/D Signed M/D Signed
* We recommend that our customers prepare a checklist that suits their operating environment.
Chapter 3 Calibration
Calibrator calibration
A special calibrator is used to calibrate the instrument.
Manually analyze the same calibrator 11 times, and check the repeatability and accuracy of the analysis
parameters.
At the same time, the compensation rate can be updated.
Precision check
Check the instrument's repeatability using a normal sample.
Note:
• Before performing calibration, check the level of the reagent connected to the instrument.
If the reagent runs out, calibration will stop.
• The following sample numbers are automatically assigned by the analyzer.
• Calibrator calibration: CAL-CAL-01 to CAL-CAL-11
• Precision check: PRE-CHK-01 to PRE-CHK-11
101/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
● Calibrator calibration
XN CAL: use for WBC, RBC, HGB, HCT, PLT, and RET calibration.
● Precision check
For a precision check, use fresh normal blood that meets the following requirements.
• Blood of a healthy person who is not taking any medicine
• Blood with an appropriate amount of anticoagulant added
• Whole blood volume of 2.0 mL or more per sample
Information
Control blood is not suitable for calibrator calibration. Control blood is intended for quality control, not for
calibration.
102/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
1 Make sure the instrument is in the ready 3 Touch the [Calibrator Calibration] icon.
state. The [Calibrator Calibration] (analysis) dialog box
If the instrument status at the bottom left of the appears.
screen is not green, wait until it becomes green.
Calibration parameters
103/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
Note: Information
• Calibrator calibration is performed using manual If an error occurs during analysis, stop calibrator
analysis. calibration. After clearing the error, repeat the
• The discrete parameters to be analyzed are procedure from the beginning.
specified automatically, and cannot be changed.
• If you have the RET channel analysis license:
CBC+DIFF+RET
• If you do not have the RET channel analysis XN-450:
license:
CBC+DIFF
1 If the sample tube holder is not ejected,
press the sample tube holder open/
XN-550: close switch.
The sample tube holder is ejected.
1 Open the sampler cover (manual unit).
2 Place the vial in the sample tube holder.
Place in the regular sample tube holder.
Vial
Regular sample
tube holder
104/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
Regular sample
tube holder
105/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
106/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
107/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
10 Touch [Next].
The following dialog box appears.
The parameters that appear depend on your system
configuration.
Calibrator parameter
checkboxes Modify checkbox
11 Touch [OK].
The compensation rate is updated.
108/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
1 Make sure the instrument is in the ready 3 Touch the [Precision Check] icon.
state. The [Precision Check] (analysis) dialog box appears.
If the instrument status at the bottom left of the Precision check parameters
screen is not green, wait until it becomes green.
109/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
XN-550:
1 If the sample tube holder is not ejected,
press the sample tube holder open/
1 Open the sampler cover (manual unit). close switch.
The sample tube holder is ejected.
2 Place the vial in the sample tube holder.
Place in the regular sample tube holder.
Vial
Regular sample
tube holder
110/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
Regular sample
tube holder
111/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
8 Touch [OK].
The following dialog box appears.
9 Touch [Yes].
The result of the precision check is recorded.
112/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
Note:
Up to 20 logs of each type are saved. Once 20 logs are saved, each new log automatically deletes the
oldest log.
113/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
USB memory
stick
USB port
114/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
Note:
If the number of logs exceeds 20 during restoring,
the oldest log is automatically deleted.
115/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
2 Insert the USB memory stick that 6 Remove the USB memory stick.
contains the backup log file.
Insert the USB memory stick into a free USB port on
3.4.6 Deleting a calibration log/
the back of the instrument.
A USB memory stick with a password lock function
precision check log
cannot be used.
Follow the steps below to delete a log.
USB memory
stick 2 Select the log you want to delete.
USB port
5 Touch [OK].
The restore starts and the [Waiting to complete
execution] dialog box appears.
The [Waiting to complete execution] dialog box closes
when the restore is complete.
116/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
Toolbar
Log list
[Output] Touch to open a submenu that allows you to output the selected log.
(➤P.113 "3.4.2 Printing the calibration log/precision check log on a printer")
117/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
[File] Touch to open a submenu that allows you to save and restore the log selected in
the list.
[Output in CSV Format]: Use to save a log in CSV format.
(➤P.114 "3.4.3 Saving the calibration log/precision check
log in CSV format")
[Backup]: Use to back up a log.
(➤P.115 "3.4.4 Saving the calibration log/precision check
log (backup)")
[Restore]: Use to restore the backed up data.
(➤P.115 "3.4.5 Restoring a saved calibration log/
precision check log (restore)")
[Delete] Use to delete the log selected in the list.
(➤P.116 "3.4.6 Deleting a calibration log/precision check log")
Log list Displays the calibration logs in the list.
[Calibration Date] Displays the date and time when the calibration was performed.
[Logon Name] Displays the name of the user who was logged on when the calibration was
performed.
[Material] Displays the name of the calibrator.
[Lot No.] Displays the lot number of the calibrator.
Compensation rate Displays the compensation rate of each analysis parameter.
When a new compensation rate has been manually input for a parameter, [*]
appears in front of the compensation rate.
118/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
Toolbar
Log list
[Output] Touch to open a submenu that allows you to output the selected log.
(➤P.113 "3.4.2 Printing the calibration log/precision check log on a printer")
119/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 3 Calibration
[File] Touch to open a submenu that allows you to save and restore the log selected in
the list.
[Output in CSV Format]: Use to save a log in CSV format.
(➤P.114 "3.4.3 Saving the calibration log/precision check
log in CSV format")
[Backup]: Use to back up a log.
(➤P.115 "3.4.4 Saving the calibration log/precision check
log (backup)")
[Restore]: Use to restore the backed up data.
(➤P.115 "3.4.5 Restoring a saved calibration log/
precision check log (restore)")
[Delete] Use to delete the log selected in the list.
(➤P.116 "3.4.6 Deleting a calibration log/precision check log")
Log list Displays the precision check logs.
[Execution Date] Displays the date and time when the precision check was performed.
[Logon Name] Displays the name of the user who was logged on at the time of the precision
check.
Analysis results Displays the CV (%) of each precision check parameter.
120/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 4 Checking Logs
4.1 Logs
You can check the following logs that are saved on the instrument.
• [Audit Log]: Instrument operation log
• [Error Log]: Errors that occurred, and information on when the errors occurred and when they
were cleared
• [Maintenance Log]: Maintenance tasks executed, and information at the time of execution
• [Reagent Replacement Log]: Reagents replaced, and information entered at the time of replacement
Note:
• Up to 5,000 logs of each type are saved. Once 5,000 logs are saved, each new log automatically deletes
the oldest log.
• The calibration logs (calibrator calibration and precision check) cannot be viewed from the [History] menu
screen. For details on calibration logs, see Chapter 3.
(➤P.113 "Chapter 3: 3.4 Checking the calibration log/precision check log")
121/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 4 Checking Logs
5 Touch [OK].
The comment is added.
122/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 4 Checking Logs
Example: XN-XXX_00-01_REAGENTLOG.csv
123/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 4 Checking Logs
Toolbar
Details button
Log list
Details table
* When the Details button is touched, the details appear below the log list.
124/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 4 Checking Logs
*1 A deletion is not recorded when an entry is automatically deleted because the maximum number of
registered entries was exceeded.
*2 A log is displayed for each rule type.
125/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 4 Checking Logs
* When the Details button is touched, the details appear below the log list.
* When the Details button is touched, the details appear below the log list.
A maintenance log is registered when one of the maintenance tasks below is performed.
The maintenance tasks and properties that are displayed are shown below.
Maintenance
Maintenance tasks
properties
[Auto Rinse] [As needed]
[Routine Cleaning] [Daily]
(Including routine cleaning at shutdown)
[Drain Waste Fluid Chamber] [As needed]
[Rinse Waste Fluid Chamber] [As needed]
[Remove Flowcell Air Bubbles] [As needed]
[Rinse Flowcell] [As needed]
[Drain Reaction Chamber] [As needed]
[Drain RBC Isolation Chamber] [As needed]
[Remove RBC Detector Clog] [As needed]
[Reagent Replenishment] [As needed]
[Reset Air Pump Count] [Part Replacement]
126/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 4 Checking Logs
* When the Details button is touched, the details appear below the log list.
127/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 4 Checking Logs
128/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 5 IP Messages
Chapter 5 IP Messages
Caution!
• A Positive or Error judgment indicates the possibility of an abnormality. It is not a diagnosis of the patient.
If a Positive or Error judgment occurs, check the data and repeat the analysis, or examine carefully in
accordance with the protocol of your laboratory.
• IP messages are only intended for use in the clinical laboratory and are not for patient diagnosis. IP
messages provide notification of the possibility of a specific sample abnormality based on examination of
the analysis data.
Positive/Negative judgment
129/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 5 IP Messages
● Message types
There are 2 types of IP messages that may be displayed for WBC, RBC/RET, and PLT: abnormal messages and
suspect messages.
● Positive/Negative judgment
[Positive] Indicates that a blood cell analysis value or cell morphology exceeds the preset
criteria for the IP message (abnormal sample).
Displayed on a red background.
A Positive judgment is classified into the 3 types shown below. Touch [Positive]
to display a dialog box.
[Diff. Abnormal]: Indicates an abnormal blood cell differentiation value.
[Morph. Abnormal]: Indicates an abnormal cell morphology.
[Count Abnormal]: Indicates an abnormal blood cell count.
[Negative] Indicates that there were no analysis errors or abnormalities, and that there is no
IP message (normal sample).
Displayed on a green background.
Note:
Only Positive judgment is performed for analysis in [Pre-Dilution] mode and [Body Fluid] mode. Negative
judgment is not performed.
130/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 5 IP Messages
With respect to the following IP messages, when a sample judgment is Positive, the analysis results are regarded as
having low reliability due to the abnormality, and " * " is displayed to the right of the data or "- - - -" which is a data
mask indicating non-analyzable is displayed.
WBC IP messages
WBC-BF
AS-LYMP#*2 TC-BF#
NEUT# LYMPH# MONO# EO# BASO# IG# AS-LYMP%*2 PMN#
WBC
NEUT% LYMPH% MONO% EO% BASO% IG% RE-LYMP#*2 PMN%
RE-LYMP%*2 MN#
MN%
WBC Abn Scattergram
Lymph, Mono
Neut, Eo
Lymph, Neut
Neut, Mono
Lymph, Baso
Mono, Eo
Mono, Baso
Neut, Baso
Debris, Neut
Debris, Baso
Debris, Lymph
Debris, WBC-BF*1
5 DIFF data
calculation ---- ---- ---- ---- ---- ---- ----
impossible
Abnormal IG
clustering
HF-BF high value
Left Shift?
Blasts/Abn Lympho? ----
Atypical Lympho?
NRBC?
*1 Does not appear when the instrument does not have the body fluid analysis function.
*2 The availability of analysis functions depends on your systems configuration.
131/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 5 IP Messages
RBC/RET IP messages
RET#
RBC
RET%
RET#
HGB IRF
HCT
MCH RDW-SD RDW-CV LFR
MCV
MCHC MFR
MCH
HFR
MCHC
RET-He
RBC Abn Distribution
MP-Flag ---- ----
Abnormal RDW-SD ----
Other abnormal
distributions
Dimorphic Population ---- ----
*1
RET Abn Scattergram
RBC Agglutination?
Turbidity/HGB Interf?
Iron Deficiency?
HGB Defect?
Fragment?
iRBC?*2
*1 Does not appear when the instrument does not have the RET analysis function.
*2 Only appears when the iRBC flag license is registered.
PLT IP messages
PDW
MPV
PLT
P-LCR
PCT
PLT Abn Distribution
Abnormal PDW ----
Other abnormal
distributions
PLT Abn Scattergram*
Foreign particles in PLT
area (serious effect)
Foreign particles in PLT
area
PLT Clumps?
* Does not appear when the instrument does not have the RET analysis function.
132/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 5 IP Messages
● WBC IP messages
Messages Meaning Judgment method/equation
Abnormal messages
WBC Abn Scattergram Abnormal WBC scattergram Judged from clustering in the WDF scattergram.
For body fluid analysis, judged from clustering in
the WDF scattergram and the HF-BF value.
Neutropenia Low neutrophil count NEUT# < 1.00 x 103/μL
or NEUT% < 0.0%
Neutrophilia High neutrophil count NEUT# > 11.00 x 103/μL
or NEUT% > 100.0%
Lymphopenia Low lymphocyte count LYMPH# < 0.80 x 103/μL
or LYMPH% < 0.0%
Lymphocytosis High lymphocyte count LYMPH# > 4.00 x 103/μL
or LYMPH% > 100.0%
Monocytosis High monocyte count MONO# > 1.00 x 103/μL
or MONO% > 100.0%
Eosinophilia High eosinophil count EO# > 0.70 x 103/μL
or EO% > 100.0%
Basophilia High basophil count BASO# > 0.20 x 103/μL
or BASO% > 100.0%
Leukocytopenia Low leukocyte count WBC < 2.50 x 103/μL
Leukocytosis High leukocyte count WBC > 18.00 x 103/μL
IG Present Increased immature IG# > 0.10 x 103/μL
granulocyte or IG% > 100.0%
Suspect messages
Left Shift? Possibility of left shift Judged from the distribution of the upper right
area of NEUT in the WDF scattergram.
Blasts/Abn Lympho? Possibility that blasts are Judged from the presence of Blasts/Abn Lympho
present/Possibility of on the WDF scattergram.
abnormal lymphocytes
Atypical Lympho? Possibility of atypical Judged from the distribution of the upper area of
lymphocytes LYMPH in the WDF scattergram.
NRBC? Possibility of nucleated red Judged from the presence of NRBC in the WDF
blood cells scattergram.
133/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 5 IP Messages
● RBC/RET IP messages
Messages Meaning Judgment method/equation
Abnormal messages
RBC Abn Distribution Abnormal RBC distribution Judged from the red blood cell distribution.
Dimorphic Population Double-peak RBC distribution Judged from the shape of the distribution peak
and the gap between the high and low points.
RET Abn Scattergram*1 Abnormal RET scattergram Judged from clustering in the RET scattergram.
Reticulocytosis*1 Reticulocytosis RET% > 5.0%
or RET# > 0.2000 x 106/μL
Anisocytosis Anisocytosis RDW-SD > 65.0 fL
or RDW-CV > 20.0%
Microcytosis Microcytosis MCV < 70.0 fL
Macrocytosis Macrocytosis MCV > 110.0 fL
Hypochromia Hypochromia MCHC < 29.0 g/dL
Anemia Anemia HGB < 10.0 g/dL
Erythrocytosis Erythrocytosis RBC > 6.50 x 106/μL
Suspect messages
RBC Agglutination? Possibility of RBC agglutination Judged from the red blood cell distribution and
hemoglobin related parameters.
Turbidity/HGB Interf? Possibility of effect on HGB Judged from the red blood cell distribution and
by chylemia hemoglobin related parameters.
Iron Deficiency? Possibility of iron deficiency Judged from the red blood cell distribution and
hemoglobin related parameters.
HGB Defect? Possibility of HGB Judged from the red blood cell distribution
abnormality related parameters.
Fragments? Possibility of fragmented red Judged from the red blood cell distribution,
blood cells platelet distribution, and RET scattergram.
iRBC?*2 Possibility of erythrocyte Judged from various analysis data and
inclusions scattergrams.
*1 Does not appear when the instrument does not have the RET analysis function.
*2 Only appears when the iRBC flag license is registered.
● PLT IP messages
Messages Meaning Judgment method/equation
Abnormal messages
PLT Abn Distribution Abnormal PLT distribution Judged from the platelet distribution.
Thrombocytopenia Thrombocytopenia PLT# < 60 x 103/μL
Thrombocytosis Thrombocytosis PLT# > 600 x 103/μL
PLT Abn Scattergram* Abnormal platelet Judged from clustering in the RET (PLT-O)
scattergram scattergram.
Suspect messages
PLT Clumps? Possibility of PLT clumps Judged from the presence of PLT clumps in the
WDF scattergram.
* Does not appear when the instrument does not have the RET analysis function.
134/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
Chapter 5 IP Messages
Blank data
Blank data is data that meets all of the following conditions:
• WBC < 1.00 x 103/μL
• RBC < 0.30 x 106/μL
• HGB < 1.0 g/dL
• PLT < 20 x 103/μL
● Judgment methods
WBC < 0.50 x 103/μL Judgment for a WBC suspect message ([Left Shift?]) is not performed.
(In [Pre-Dilution] mode, when WBC < 0.20 x 103/μL.)
RBC < 0.50 x 106/μL IP message judgment for RBC other than [RBC Abn Distribution] is not performed.
This is displayed as [RBC Abn Distribution], even if analysis of RBC was not
specified.
• If an error or other condition prevents an analysis parameter necessary for judgment from being calculated
("- - - -" or "+ + + +" is displayed), judgments including that parameter will not be performed.
• Parameters for which the user has not specified that analysis be performed (blank " ") are not used for
judgment.
135/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
BF [Body
Positive/Negative WB [Whole Blood]/LW [Low WBC] PD [Pre-Dilution]
Flag Detection Fluid]
*4
*3
*2
*1
Messages judgment flag
No.*1 channel CBC+DIFF CBC+DIFF
136/138
category CBC CBC+DIFF CBC+RET CBC CBC+DIFF CBC+DIFF
+RET +RET
Chapter 5
WBC
Abnormal messages
Leukocytopenia 9 Count WDF ✓ ✓ ✓ ✓ ✓ ✓ ✓ -
Leukocytosis A Count WDF ✓ ✓ ✓ ✓ ✓ ✓ ✓ -
IG Present F Morph+Count WDF - ✓ - ✓ - ✓ ✓ -
Left Shift? 3 Morph WDF - ✓ - ✓ - ✓ ✓ -
Blasts/Abn Lympho? 7 Morph WDF - ✓ - ✓ - ✓ ✓ -
Atypical Lympho? 4 Morph WDF - ✓ - ✓ - ✓ ✓ -
Suspect
messages
NRBC? 5 Morph WDF - ✓ - ✓ - ✓ ✓ -
RBC Abn Distribution 1 Morph RBC ✓ ✓ ✓ ✓ - - - -
Dimorphic Population 2 Morph RBC ✓ ✓ ✓ ✓ - - - -
● Table of IP message judgments
Abnormal messages
Hypochromia 6 Morph RBC+HGB ✓ ✓ ✓ ✓ ✓ ✓ ✓ -
Anemia 7 Count HGB ✓ ✓ ✓ ✓ ✓ ✓ ✓ -
RBC/RET
In the [Sample Explorer] screen, IP messages are shown by flag number.
Erythrocytosis 8 Count RBC ✓ ✓ ✓ ✓ ✓ ✓ ✓ -
RBC Agglutination? 1 Count RBC+HGB ✓ ✓ ✓ ✓ - - - -
Turbidity/HGB Interf? 2 Count RBC+HGB - - - -
Does not appear when the instrument does not have the RET analysis function.
✓ ✓ ✓ ✓
Iron Deficiency? 3 Morph RBC+HGB ✓ ✓ ✓ ✓ - - - -
HGB Defect? 4 Morph RBC ✓ ✓ ✓ ✓ - - - -
Fragments? 5 Morph RBC, PLT, RET ✓ ✓ ✓ ✓ - - - -
Suspect messages
iRBC?*3 6 Morph WDF, PLT, RET ✓*4 ✓*4 ✓*4 ✓ ✓*4 ✓*4 ✓ -
PLT Abn Distribution 1 Morph PLT ✓ ✓ ✓ ✓ - - - -
Thrombocytopenia 2 Count PLT, RET ✓ ✓ ✓ ✓ ✓ ✓ ✓ -
Thrombocytosis 3 Count PLT, RET ✓ ✓ ✓ ✓ ✓ ✓ ✓ -
Abnormal
messages
PLT Abn Scattergram*2 4 Count RET - - ✓ ✓ - - - -
PLT
PLT Clumps? 1 Count WDF ✓ ✓ ✓ ✓ - - - -
Suspect
messages
Index
A E
Auto rinse..............................................27, 33, 47 Error messages................................................... 7
C H
Calibration........................................................101 [Help] dialog box ............................................... 11
Main functions .............................................102
Overview......................................................101 I
Standards ....................................................101 IP messages
Calibration logs ................................................113 Details table ................................................ 136
[Calibration History] screen .........................117 Displayed items .......................................... 129
Backup.........................................................115 Judgment conditions................................... 135
Check ..........................................................113 Judgment methods ..................................... 135
Delete ..........................................................116 Meaning ...................................................... 133
Output..........................................................113 Types .......................................................... 133
Output in CSV format ..................................114
Restore ........................................................115 L
Calibrator .........................................................102 List of maintenance tasks.................................. 69
Calibrator calibration ........................................103 Logs ................................................................ 121
Calibrators and samples to be used ............102 [Audit Log] screen....................................... 124
Overview......................................................101 [Error Log] screen ....................................... 126
Screens .......................................................117 [Maintenance Log] screen .......................... 126
Causes of errors and remedial actions ..............12 [Reagent Replacement Log] screen ........... 127
Errors related to the analysis.........................33 Check.......................................................... 121
Errors related to the blood aspiration ............26 Comments .................................................. 121
Errors related to the covers ...........................50 Output ......................................................... 122
Errors related to the dye holder .....................60 Output in CSV format.................................. 122
Errors related to the hand and sample tube
holders..............................................67 M
Errors related to the laser ..............................31 Maintenance and inspection checklist............. 100
Errors related to the motors...........................25 Maintenance screens ........................................ 70
Errors related to the orders............................30 [Exchange] dialog box .................................. 72
Errors related to the pressure........................12 [Maintenance] menu screen ......................... 71
Errors related to the quality control (QC).......32 [Operation Confirmation] dialog box ............. 71
Errors related to the reagents and [Rinse Instrument] dialog box ....................... 72
chambers..........................................16 [Status Confirmation] dialog box................... 73
Errors related to the sample numbers ...........30 Monitor
Errors related to the sample tube holder .......30 Changing settings ......................................... 98
Errors related to the sampler analysis Closing the OSD screen ............................... 98
operation...........................................66 Displaying the OSD screen........................... 98
Errors related to the system ..........................31 Image quality adjustment.............................. 98
Errors related to the temperature ..................15 Setting items in the OSD screen................... 99
Errors related to the user maintenance and
warnings ...........................................51 O
Cleaning the aspiration unit tray (XN-350).........96 Operation tests.................................................. 79
Cleaning the aspiration unit tray Air pump ....................................................... 79
(XN-550/XN-450) ............................................94 Aspiration unit motor..................................... 79
Counter ..............................................................82 Barcode reader (XN-550) ............................. 80
Motor............................................................. 79
D Pinch valve ................................................... 80
Draining the pneumatic trap chamber................13 Sampler ........................................................ 81
Draining the sample from the RBC isolation Sheath motor ................................................ 79
chamber ..........................................................34 Whole blood aspiration motor ....................... 79
137/138
XN-L series XN-550/XN-450/XN-350 Troubleshooting
138
Index
P
Precision check ............................................... 109
Calibrators and samples to be used ........... 102
Overview..................................................... 101
Screens....................................................... 119
Precision check logs ....................................... 113
[Precision Check History] screen................ 119
Backup........................................................ 115
Check.......................................................... 113
Delete ......................................................... 116
Output ......................................................... 113
Output in CSV format.................................. 114
Restore ....................................................... 115
R
Reagent replacement........................................ 84
Dialog box..................................................... 84
Diluent and hemolytic agent .............16, 57, 85
Dye ............................................ 17, 58, 61, 86
Reagent replenishment ..................................... 77
Registering reagent information ........................ 63
Removing air bubbles from the flowcell ............ 46
Removing the RBC detector clog...................... 34
Replacing the air pump ............................... 54, 89
Replacing the fuse ............................................ 92
Replacing the piercer
(XN-550/XN-450) ........................................... 88
Replacing the waste container .................... 23, 78
Rinsing the flowcell (XN-350)...................... 39, 45
Rinsing the flowcell (XN-450)...................... 38, 44
Rinsing the flowcell (XN-550)...................... 37, 43
Rinsing the RBC detector aperture ............. 35, 40
Rinsing the waste chamber (XN-350) ............... 22
Rinsing the waste chamber (XN-450) ............... 21
Rinsing the waste chamber (XN-550) ............... 20
Routine cleaning (XN-350)............. 29, 49, 53, 76
Routine cleaning (XN-450)............. 28, 48, 52, 75
Routine cleaning (XN-550)............. 27, 47, 51, 74
S
Sensor............................................................... 83
X
XN CAL ........................................................... 102
138/138