GEHC-DICOM-Conformance SenographePristina9 DOC3075145 Rev1
GEHC-DICOM-Conformance SenographePristina9 DOC3075145 Rev1
Publications
Document DOC3075145
Revision 1
© 2024 by GE HealthCare
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
1 Overview
Senographe Pristina is a digital mammography imaging system. It provides DICOM 3.0 support to receive scheduled procedures from RIS and
update status of procedures, to transfer and/or archive images to PACS or Review Workstation, to print images, and to export data on media.
The "Transfer Syntax Set" column lists the sets of Transfer Syntaxes defined in Table 1-2 that are applicable to each SOP Class. The "DIMSE",
"DICOM Web" and "Media Services" columns indicate the roles supported for each SOP Class.
The "Function" columns indicate how the instances are used by the system:
• Create: The system creates instances of the SOP Class. The type of the created SOP Class is indicated by one of the following abbreviations:
Page 2 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
• Display: The system displays the instances of the SOP Class to the user, either by displaying the SOP Instances natively or by applying
instances of another suitable SOP Class to the image instances (e.g., a Presentation State or CAD SR).
• Process: The system processes the instances of the SOP Class to derive some further information that is made available to the user (e.g., a CAD
processing algorithm, or a 3D Rendering).
• Archive: The system stores the instances of the SOP Class and makes them available again.
Page 3 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 1-1 Storage SOP Classes
SOP Classes Transfer Syntax Set DIMSE DICOM Web Media Services Function
SCU SCP UA OS FSC FSU FSR Create Display Process Archive
Media Storage Directory 1.2.840.10008.1.3.10 BD N N N N Y N Y S Y N N
Storage
Breast Tomosynthesis 1.2.840.10008.5.1.4.1.1.13.1.3 LL; L; U Y Y (*) N N Y N Y SE Y N N
Image Storage
CT Image Storage 1.2.840.10008.5.1.4.1.1.2 LL; U Y N N N N N N SE N N N
Digital Mammography X- 1.2.840.10008.5.1.4.1.1.1.2 U Y Y N N Y N Y SE Y N N
Ray Image Storage - For
Presentation
Digital Mammography X- 1.2.840.10008.5.1.4.1.1.1.2.1 U Y Y N N Y N Y SE Y N N
Ray Image Storage - For
Processing
Grayscale Softcopy 1.2.840.10008.5.1.4.1.1.11.1 NI Y Y N N Y N Y S Y N N
Presentation State
Storage
Secondary Capture Image 1.2.840.10008.5.1.4.1.1.7 U Y Y N N Y N Y N Y N N
Storage
X-Ray Radiation Dose SR 1.2.840.10008.5.1.4.1.1.88.67 NI Y Y N N Y N Y See Table 1-3
Storage
(*) As an SCP for Breast Tomosynthesis Image Storage, the system supports only the 3 uncompressed and the JPEG Lossless, Non-Hierarchical, First-Order Prediction (Process 14
[Selection Value 1]) transfer syntaxes.
Page 4 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
• RENDER: The system displays the content of the SR, without using the data for any processing.
• EXTRACT_DATA: The system can extract structured data from the content and use the data for subsequent processing (e.g., reporting).
• OVERLAY: The system uses the information in the SR to display information directly on the images (e.g., Mammography CAD markers).
The "SOP Class UID" column indicates which of the SR Storage SOP Classes are used to encode the information or to store it. If multiple SOP
Classes are supported the "Condition" column describes the conditions for using the different SOP Classes.
Page 5 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
1.2.2 Storage
For details on supported Storage SOP Classes see Section 1.1.
1.2.4 Query/Retrieve
Table 1-6 lists all supported Query/Retrieve SOP Classes.
1.2.5 Printing
Table 1-7 lists all supported Printing SOP Classes.
Page 6 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 1-7 Printing SOP Classes
SOP Classes Transfer Syntax SCU SCP
Basic Grayscale Print 1.2.840.10008.5.1.1.9 Implicit VR Little Endian 1.2.840.10008.1.2 Y N
Management Meta
Explicit VR Little Endian 1.2.840.10008.1.2.1 Y N
Explicit VR Big Endian 1.2.840.10008.1.2.2 Y N
Basic Film Session 1.2.840.10008.5.1.1.1 Implicit VR Little Endian 1.2.840.10008.1.2 Y N
Explicit VR Little Endian 1.2.840.10008.1.2.1 Y N
Explicit VR Big Endian 1.2.840.10008.1.2.2 Y N
Basic Film Box 1.2.840.10008.5.1.1.2 Implicit VR Little Endian 1.2.840.10008.1.2 Y N
Explicit VR Little Endian 1.2.840.10008.1.2.1 Y N
Explicit VR Big Endian 1.2.840.10008.1.2.2 Y N
Basic Grayscale 1.2.840.10008.5.1.1.4 Implicit VR Little Endian 1.2.840.10008.1.2 Y N
Image Box
Explicit VR Little Endian 1.2.840.10008.1.2.1 Y N
Explicit VR Big Endian 1.2.840.10008.1.2.2 Y N
Printer 1.2.840.10008.5.1.1.16 Implicit VR Little Endian 1.2.840.10008.1.2 Y N
Explicit VR Little Endian 1.2.840.10008.1.2.1 Y N
Explicit VR Big Endian 1.2.840.10008.1.2.2 Y N
Page 7 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
2 Table of Contents
Contents
1 Overview ..................................................................................................................................................................................... 2
1.1 Content and Transfer ............................................................................................................................................. 2
1.1.1 Structured Reporting Root Template IDs ........................................................................................................... 5
1.2 DIMSE Services........................................................................................................................................................ 5
1.2.1 Verification ........................................................................................................................................................... 5
1.2.2 Storage................................................................................................................................................................. 6
1.2.3 Workflow Management ....................................................................................................................................... 6
1.2.4 Query/Retrieve..................................................................................................................................................... 6
1.2.5 Printing ................................................................................................................................................................ 6
1.3 DICOM Web Services - N/A ...................................................................................................................................... 7
1.4 Media Services ........................................................................................................................................................ 7
1.5 Real-Time Video Service - N/A ................................................................................................................................ 7
1.6 De-identification Profiles - N/A ............................................................................................................................... 7
1.7 Specific Character Sets ........................................................................................................................................... 8
2 Table of Contents ....................................................................................................................................................................... 8
3 Introduction ............................................................................................................................................................................. 11
3.1 Audience ............................................................................................................................................................... 11
3.2 Remarks ................................................................................................................................................................ 11
3.3 Trademarks ........................................................................................................................................................... 12
3.4 Omissions and Errors............................................................................................................................................ 12
3.5 Copyrights............................................................................................................................................................. 12
3.6 Terms and Definitions .......................................................................................................................................... 12
3.7 Abbreviations........................................................................................................................................................ 14
3.8 References ............................................................................................................................................................ 16
4 Implementation Model............................................................................................................................................................. 17
4.1 Application Entities and Data Flow ...................................................................................................................... 17
4.1.1 Functional Definition of DICOM Service Application Entity .............................................................................. 18
4.1.2 Functional Definition of Media Interchange Application Entity ....................................................................... 19
5 Service and Interoperability Description ................................................................................................................................. 19
5.1 Mapping of Services to Application Entities ........................................................................................................ 19
5.2 DIMSE Services...................................................................................................................................................... 19
Page 8 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
5.2.1 Basic Worklist Management Service ................................................................................................................. 19
5.2.2 Modality Performed Procedure Step Service .................................................................................................... 27
5.2.3 Unified Worklist and Procedure Step Service - N/A .......................................................................................... 31
5.2.4 Instance Availability Notification Service - N/A ................................................................................................ 31
5.2.5 Storage Service .................................................................................................................................................. 31
5.2.6 Storage Commitment Service ........................................................................................................................... 32
5.2.7 Query/Retrieve Service Class ............................................................................................................................ 34
5.2.8 Print Management Service ................................................................................................................................ 37
5.3 DICOM Web Services - N/A .................................................................................................................................... 41
5.4 Media Services ...................................................................................................................................................... 41
5.4.1 File Set Creator (FSC) ......................................................................................................................................... 41
5.4.2 File Set Reader (FSR) ......................................................................................................................................... 41
5.4.3 File Set Updater (FSU) - N/A .............................................................................................................................. 41
5.5 Real-Time Video Services - N/A ............................................................................................................................ 41
5.6 Cross Service Considerations – N/A ..................................................................................................................... 41
5.7 Specific Character Sets ......................................................................................................................................... 41
6 Configuration ........................................................................................................................................................................... 42
6.1 General Configuration Parameters ...................................................................................................................... 42
6.2 Configuration of DIMSE Services .......................................................................................................................... 43
6.2.1 Basic Worklist Management Service Configuration .......................................................................................... 43
6.2.2 Modality Performed Procedure Step Service Configuration ............................................................................ 43
6.2.3 Unified Worklist and Procedure Step Service Configuration - N/A .................................................................. 44
6.2.4 Instance Availability Notification Service Configuration - N/A ......................................................................... 44
6.2.5 Storage Service Configuration .......................................................................................................................... 44
6.2.6 Storage Commitment Service Configuration ................................................................................................... 45
6.2.7 Query/Retrieve Service Configuration .............................................................................................................. 46
6.2.8 Print Management Service Configuration......................................................................................................... 46
6.3 Configuration of DICOM Web Services - N/A ........................................................................................................ 47
6.4 Configuration of Media Storage Service .............................................................................................................. 47
6.5 Configuration of Real-Time Video Service - N/A .................................................................................................. 47
6.6 Configuration of Audit Trail - Syslog .................................................................................................................... 47
7 Network and Media Communication Details ........................................................................................................................... 48
7.1 General .................................................................................................................................................................. 48
7.1.1 General Association Parameters ....................................................................................................................... 48
7.2 Specifications ....................................................................................................................................................... 48
7.2.1 DICOM Service Application Entity ..................................................................................................................... 48
7.3 Status Codes ......................................................................................................................................................... 58
7.3.1 General AE Communication and Failure Behavior and Handling .................................................................... 58
7.3.2 DIMSE Services .................................................................................................................................................. 59
7.3.3 DICOM Web Services - N/A ................................................................................................................................. 72
8 Security ..................................................................................................................................................................................... 73
8.1 Introduction .......................................................................................................................................................... 73
8.2 External Network Requirements .......................................................................................................................... 73
8.3 TCP Port Configuration ........................................................................................................................................ 73
8.4 DICOM Security Profiles Support ......................................................................................................................... 73
Page 9 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
8.4.1 Secure Use and User Identity Profiles ............................................................................................................... 73
8.4.2 Secure Transport Connection Profiles .............................................................................................................. 74
8.4.3 Media Storage Security Profiles ........................................................................................................................ 74
8.4.4 Attribute Confidentiality Profiles - N/A ............................................................................................................. 74
8.4.5 Digital Signature Profiles – N/A ......................................................................................................................... 74
8.4.6 Additional DICOM Security Profiles – N/A ......................................................................................................... 74
8.5 User Identity Negotiation Support – N/A ............................................................................................................. 74
8.6 Web Services Security Features – N/A .................................................................................................................. 75
8.7 Other Security Features ........................................................................................................................................ 75
8.7.1 Media Storage Security – N/A ............................................................................................................................ 75
8.7.2 Network Security – N/A ..................................................................................................................................... 75
8.7.3 Other Security Features – N/A ........................................................................................................................... 75
A Information Object Definitions (IODs) ..................................................................................................................................... 76
A.1 Information Shared Across Multiple IODs ............................................................................................................ 77
A.1.1 Common Modules ............................................................................................................................................. 77
A.1.2 Common Functional Group Macros - N/A ......................................................................................................... 86
A.1.3 Common Private Modules – N/A ....................................................................................................................... 86
A.1.4 Coded Values ..................................................................................................................................................... 86
A.2 Breast Tomosynthesis Image IOD ........................................................................................................................ 87
A.2.1 Breast Tomosynthesis Image IOD Specific Modules ......................................................................................... 89
A.2.2 Breast Tomosynthesis Image IOD Specific Functional Group Macros ............................................................. 97
A.2.3 Breast Tomosynthesis Image IOD Specific Private Modules .......................................................................... 101
A.2.4 Breast Tomosynthesis Image IOD Specific Values and Code Sets ................................................................. 105
A.3 CT Image IOD ...................................................................................................................................................... 105
A.3.1 CT Image IOD Specific Modules ....................................................................................................................... 106
A.3.2 CT Image IOD Specific Functional Group Macros – N/A .................................................................................. 108
A.3.3 CT Image IOD Specific Private Modules .......................................................................................................... 108
A.3.4 CT Image IOD Specific Values and Code Sets – N/A ........................................................................................ 110
A.4 Digital Mammography X-Ray Image IOD ............................................................................................................ 110
A.4.1 Digital Mammography X-Ray Image IOD Specific Modules ............................................................................ 112
A.4.2 Digital Mammography X-Ray Image IOD Specific Functional Group Macros – N/A ....................................... 121
A.4.3 Digital Mammography X-Ray Image IOD Specific Private Modules ................................................................ 122
A.4.4 Digital Mammography X-Ray Image IOD Specific Values and Code Sets ....................................................... 126
A.5 Grayscale Softcopy Presentation State IOD....................................................................................................... 127
A.5.1 Grayscale Softcopy Presentation State IOD Specific Modules ....................................................................... 128
A.5.2 Grayscale Softcopy Presentation State IOD Specific Functional Group Macros – N/A .................................. 131
A.5.3 Grayscale Softcopy Presentation State IOD Specific Private Modules – N/A ................................................. 131
A.5.4 Grayscale Softcopy Presentation State IOD Specific Values and Code Sets – N/A ........................................ 131
A.6 X-Ray Radiation Dose SR IOD ............................................................................................................................. 132
A.6.1 X-Ray Radiation Dose SR IOD Specific Modules .............................................................................................. 132
A.6.2 X-Ray Radiation Dose SR IOD Specific Functional Group Macros – N/A ......................................................... 135
A.6.3 X-Ray Radiation Dose SR IOD Specific Private Modules – N/A ........................................................................ 135
A.6.4 X-Ray Radiation Dose SR IOD Specific Values and Code Sets – N/A ............................................................... 135
A.7 Basic Directory IOD ............................................................................................................................................. 135
B Structured Report Content Encoding .................................................................................................................................... 140
B.1 Projection X-Ray Radiation Dose (TID 10001) .................................................................................................... 140
Page 10 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
B.1.1 Code Sets ......................................................................................................................................................... 145
B.1.2 Measurement Encoding – N/A ......................................................................................................................... 146
C Security Details....................................................................................................................................................................... 147
C.1 External Network Requirement Details ............................................................................................................. 147
C.1.1 Basic Time Synchronization ............................................................................................................................ 147
C.1.2 Basic Network Address Management ............................................................................................................. 147
C.1.3 Application Configuration Management......................................................................................................... 147
C.1.4 DNS Service Discovery – N/A ........................................................................................................................... 147
C.2 DICOM Security Profile Details ........................................................................................................................... 147
C.2.1 Online Electronic Storage Secure Use – N/A ................................................................................................... 147
C.2.2 Audit Trail Messages ........................................................................................................................................ 147
C.2.3 Audit Trail Message Transmission Profile - SYSLOG - TLS .............................................................................. 158
C.2.4 Audit Trail Message Transmission Profile - SYSLOG - UDP ............................................................................. 159
C.2.5 Secure Transport Connection Details – N/A ................................................................................................... 159
C.2.6 Attribute Confidentiality Details - N/A ............................................................................................................ 159
C.2.7 Digital Signature Details – N/A ........................................................................................................................ 159
C.2.8 Additional DICOM Security Profile Details – N/A ............................................................................................. 159
D Mapping of Attributes............................................................................................................................................................. 159
D.1 Mapping Between Modality Worklist Instances and MPPS ............................................................................... 159
E Code Set Usage....................................................................................................................................................................... 161
3 Introduction
3.1 Audience
This document is intended for the audience listed below. It is assumed that the reader has a working knowledge of the DICOM Standard.
The document structure was designed for easier access to relevant information for different user groups:
• Clinical Users, who want to get an overview of the implemented interoperability features of the system can see Section 4 Implementation
Model.
• Personnel involved in Sales can use the information in Section 1 to assess the compatibility between different systems involved in a sales
situation.
• System Integrators can use information in Section 6 during system installation and also information from Section 5 Service and
Interoperability Description for details regarding the implemented services.
• Field Service Engineers can use the details from Section 5 Service and Interoperability Description and from Section 7 Network and Media
Communication Details for troubleshooting.
• Hospital IT staff focusing on security can use the details provided in Section 8 Security regarding implemented Security features.
• Research Personnel may be interested in using information provided in Annex A Information Object Definitions (IODs) or Annex B Structured
Report Content Encoding to get detailed imaging and measurement information.
3.2 Remarks
The scope of this DICOM Conformance Statement is to facilitate integration between Senographe Pristina and other DICOM products. The
Conformance Statement should be read and understood in conjunction with the DICOM Standard [1] . DICOM by itself does not guarantee
interoperability.
Page 11 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
• The Conformance Statement does, however, facilitate a first-level comparison for interoperability between different applications supporting
compatible DICOM functionality.
• This Conformance Statement should not replace validation with other DICOM equipment to ensure proper exchange of intended information.
In fact, it is the user's responsibility to perform the following validation activities:
• The comparison of Conformance Statements from Senographe Pristina and other DICOM conformant equipment is the first step towards
assessing interconnectivity and interoperability between those systems.
• Test procedures should be defined and executed to validate the required level of interoperability with specific DICOM conformant
equipment, as established by the healthcare facility.
Senographe Pristina has participated in an industry-wide testing program sponsored by Integrating the Healthcare Enterprise (IHE). The IHE
Integration Statement of Senographe Pristina together with the IHE Technical Framework may facilitate the process of validation testing.
3.3 Trademarks
All products and their name brands are trademarks of their respective holders.
GE HealthCare personnel, please use the GE HealthCare PQR Process to report all omissions, errors, and defects in this publication.
3.5 Copyrights
GE is a trademark of General Electric Company used under trademark license.
Abstract Syntax The information agreed to be exchanged between applications, generally equivalent to a
Service/Object Pair (SOP) Class. Examples: Verification SOP Class, Modality Worklist
Information Model Find SOP Class, Computed Radiography Image Storage SOP Class.
Application Entity (AE) A representation of the external behavior of an application process in terms of DICOM
Network Services, Web Services and/or media exchange capabilities implemented in one or
more roles. A single device may have multiple Application Entities.
Application Entity Title (AET) The externally known name of an Application Entity, used to identify a DICOM application to
other DICOM applications on the network.
Application Context The specification of the type of communication used between Application Entities. Example:
DICOM network protocol.
Attribute A unit of information in an Information Object Definition; a Data Element identified by a tag.
The information may be a complex data structure (Sequence), itself composed of lower-level
data elements. Examples: Patient ID (0010,0020), Accession Number (0008,0050),
Photometric Interpretation (0028,0004), Procedure Code Sequence (0008,1032).
Page 12 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Data Element A unit of information as defined by a single entry in the data dictionary. An encoded
Information Object Definition (IOD) Attribute that is composed of, at a minimum, three fields:
a Data Element Tag, a Value Length, and a Value Field. For some specific Transfer Syntaxes, a
Data Element also contains a VR Field where the Value Representation of that Data Element is
specified explicitly
Information Object Definition (IOD) The specified set of Attributes that comprise a type of data object; does not represent a
specific instance of the data object, but rather a class of similar data objects that have the
same properties. Examples: MR Image IOD, CT Image IOD, Print Job IOD. The Attributes within
an IOD may be specified as Mandatory (Type 1), Required but possibly unknown (Type 2), or
Optional (Type 3), and there may be conditions associated with the use of an Attribute (Types
1C and 2C).
Media Application Profile The specification of DICOM information objects and encoding exchanged on removable
media (e.g., CDs).
Module A set of Attributes within an Information Object Definition that are logically related to each
other. Example: Patient Module includes Patient's Name, Patient ID, Patient' Birth Date, and
Patient's Sex.
Negotiation First phase of Association establishment that allows Application Entities to agree on the types
of data to be exchanged and how that data will be encoded.
Origin Server Refers to the program that can originate authoritative responses to HTTP requests for a given
Target Resource. The term "server" refers to any implementation that receives a web service
request message from a user agent.
Presentation Context The set of DICOM Network Services used over an Association, as negotiated between
Application Entities; includes Abstract Syntaxes and Transfer Syntaxes.
Private SOP Class A SOP Class that is not defined in the DICOM Standard but is published in an
implementation's Conformance Statement.
Protocol Data Unit (PDU) A packet (piece) of a DICOM message sent across the network. Devices must specify the
maximum size packet they can receive for DICOM messages.
Security Profile A set of mechanisms, such as encryption, user authentication, or digital signatures, used by
an Application Entity to ensure confidentiality, integrity, and/or availability of exchanged
DICOM data.
Service Class Provider (SCP) Role of an Application Entity that provides a DICOM network service; typically, a server that
performs operations requested by another Application Entity (Service Class User). Examples:
Picture Archiving and Communication System (image storage SCP, and image query/retrieve
SCP), Radiology Information System (modality worklist SCP).
Service Class User (SCU) Role of an Application Entity that uses a DICOM Network Service; typically, a client. Examples:
imaging modality (image storage SCU, and modality worklist SCU), imaging workstation
(image query/retrieve SCU).
Service/Object Pair Class (SOP Class) The specification of the network or media transfer (service) of a particular type of data
(object) ; the fundamental unit of a DICOM interoperability specification. Examples:
Ultrasound Image Storage Service, Basic Grayscale Print Management.
Page 13 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Service/Object Pair Instance (SOP Instance) An information object; a specific occurrence of information exchanged in a SOP Class. E.g., a
specific X-ray image.
Specialized SOP Class A SOP Class that is derived from the Standard that is specialized by additional type 1, 1C, 2,
2C, or 3 Attributes, by enumeration of specific permitted Values for Attributes, or by
enumeration of specific permitted Templates. The additional Attributes may either be drawn
from the Data Dictionary in PS3.6 or may be Private Attributes.
Standard SOP Class A SOP Class defined in the Standard, and that is implemented and used without any
modifications.
Standard Extended SOP Class A SOP Class that is defined in the standard, and that is extended by additional type 3
Attributes. The additional Attributes may either be drawn from the DICOM Data Dictionary in
PS3.6 or may be Private Attributes.
Tag A 32-bit identifier for a Data Element, represented as a pair of four-digit hexadecimal
numbers, the "group" and the "element". If the "group" number is odd, the tag is for a private
(manufacturer-specific) data element. Examples: (0010,0020) [Patient ID], (07FE,0010) [Pixel
Data], (0019,0210) [private data element].
Transfer Syntax The encoding used for exchange of DICOM information objects and messages. Examples:
JPEG compressed (images), Little Endian Explicit Value Representation.
TLS-Secured Port TCP port on which an implementation accepts TLS connections to exchange DICOM
information.
Unique Identifier (UID) A globally unique "dotted decimal" string that identifies a specific object or a class of objects;
an ISO-8824 Object Identifier. Examples: Study Instance UID, SOP Class UID, SOP Instance
UID.
User Agent A client in a network protocol used in communications within a client-server distributed
computing system. In particular, the Hypertext Transfer Protocol (HTTP) identifies the client
software originating the request, using a user-agent header, even when the client is not
operated by a user.
Value Representation (VR) The format type of an individual DICOM data element, such as text, an integer, a person's
name, or a code. DICOM information objects can be transmitted with either explicit
identification of the type of each data element (Explicit VR), or without explicit identification
(Implicit VR) ; with Implicit VR, the receiving application must use a DICOM data dictionary to
look up the format of each data element.
The following list includes product specific definitions used throughout this Conformance Statement:
3.7 Abbreviations
Abbreviations that are used in this DICOM Conformance Statement are listed here.
Page 14 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
AE Application Entity
OS Origin Server
Page 15 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
SR Structured Reporting
UA User Agent
UI User Interface
UL Upper Layer
VR Value Representation
3.8 References
[1] National Electrical Manufacturers Association (NEMA), Rosslyn, VA USA. PS3 / ISO 12052 Digital Imaging and Communications in Medicine
(DICOM) Standard. http://www.dicomstandard.org .
[2] Integrating the Healthcare Enterprise (IHE). IHE Radiology Technical Framework.
http://www.ihe.net/Resources/technical_frameworks/#radiology .
Page 16 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
4 Implementation Model
This DICOM Conformance Statement covers all versions 9.x of Senographe Pristina, including its options Senographe Pristina 3D, SenoBright HD,
Pristina Serena, Pristina Serena 3D, and Pristina SerenaBright.
This section describes the organization of the supported Services into Application Entities based on the default configuration of the system. This
may change based on the actual setup at the customer site. See Section 6 for details about the configurability of Services into AEs.
Page 17 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
The user or the system initiates a modality worklist query to the modality worklist SCP with a given set of query parameters. The worklist
query can be manual or automatic. The modality worklist SCP returns responses matching the query parameters. Worklist items from the
returned worklist query responses are presented to the user. The user then chooses the desired worklist item and begins the image
acquisition process.
When the user begins the image acquisition process and generates the first image, the DICOM Service AE sends a N-CREATE message to the
configured MPPS SCP to indicate that the image acquisition process has been started for the requested procedure.
The user can close the acquisition session either by completing the acquisition process or discontinuing the ongoing procedure. On closing
the acquisition session, the DICOM Service AE sends N-SET message to the configured MPPS SCP to indicate the acquisition state of the
requested procedure, with appropriate MPPS status (COMPLETED or DISCONTINUED).
• Push Images
For this operation, the user selects some studies, series or images on the browser and then sends the selected studies, series, or images to
one or several remote DICOM AE. The transfer activity is displayed on a specific icon. The declaration of remote DICOM AE is done through a
configuration tool.
Push of images can also be automatic. This can be configured in Image preference.
• Archive Images
For this operation, the user selects some studies, series or images on the browser and sends them to the wanted remote DICOM AE. This
initiates the DICOM push of the selected exam/series/images to the remote AE. After the successful storage of the selected
exam/series/images on the remote AE, a storage commitment request is sent to the remote DICOM AE. The transfer/storage commit activity
is displayed in the job management queue.
The declaration of remote DICOM AE with archiving support is done through the configuration tool.
Push of images can also be automatic. This can be configured in Image preference. After every successful storage of the automatic pushed
exam/series/images to the remote AE, a storage commitment request is sent to the remote DICOM AE, and the storage commit activity is
displayed in the job management queue.
The DICOM Service AE is indefinitely listening for association requests and no user action is required to receive the Storage Commitment
notification (N-EVENT-REPORT) from the remote AE.
• Query / Retrieve
For this operation, the user queries one or a set of remote DICOM databases to obtain a list of data at Study/Series/Image level by clicking
on the icon that represents the wanted remote DICOM AE.
Once the remote browser has displayed query results, the user can retrieve the study/series/images from the remote DICOM AE.
• Receive Images
When remote DICOM hosts sends DICOM images to DICOM Service AE, images are stored in the local database, the browser displays the
content of the local database.
Page 18 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
• Print
The Film composer allows the user to select printers. Direct print from the viewer is also supported. When user presses the "Print" Button,
the DICOM Service AE tries to establish the association with requested printer and sends the images for printing. It is possible to set the print
in an automatic mode. DBT volume cannot be printed.
• Verify Connectivity
The user can choose to verify a remote Application Entity. Upon user’s request, DICOM Service AE sends out a verification request to a
Remote AE.
The DICOM Service AE also acts as an SCP for incoming Verification requests. No user action is required to process Verification requests from
Remote AEs.
• Generate and write a DICOM File Set (FSC) in a one-shot activity. (SAVE).
• Read a DICOM File Set (FSR) on an Interchange Media (QUERY).
• It can copy SOP instances from the media onto local storage. (RESTORE).
As a Service Class User of the Modality Worklist Information Model - FIND SOP Class, the Senographe Pristina uses the C-FIND-RQ message to
query the SCP. It supports the Query Keys listed in Table 5-2.
• UNIVERSAL: SCU can request that the Attribute be a return Value (universal matching).
In the "Query Value Source" column, the following Values can be used:
• GENERATED: The query Value is generated by the system (e.g., current date as the study date).
• SCANNED: The query Value is read from a barcode scanner or similar device.
• EMPTY: The query Value is sent with a zero-length Value to indicate it is a return key only.
Table 5-2 Supported C-FIND Query Parameters for Modality Worklist - SCU
Attribute Name Tag Matching Type Query Value Value Display Comments
Source on UI
Scheduled Procedure Step
Scheduled Procedure Step (0040,0100) SEQUENCE FIXED N
Sequence
>Scheduled Station AE Title (0040,0001) SINGLE_VALUE; USER; N Matching is supported as follows:
UNIVERSAL EMPTY 1. Worklist AET is supplied for matching (Single
value matching).
2. No AE title is supplied (universal matching),
In case of universal matching, if attribute is absent
or empty in response, the SPS is accepted
>Scheduled Procedure Step (0040,0002) UNIVERSAL; EMPTY; D Matching is supported as one of the following:
Start date SINGLE_VALUE; USER 1. No date is supplied (universal matching).
RANGE 2. One date is supplied for matching (Single value
matching).
3. A start date and an end date are supplied for
matching (Range of date matching).
Page 20 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Attribute Name Tag Matching Type Query Value Value Display Comments
Source on UI
>Scheduled Procedure Step (0040,0003) UNIVERSAL EMPTY D Requested in the MWL queries with zero-length
Start Time value.
If attribute is absent or empty in answer, the SPS
is accepted.
If value is outside of a valid range, the value will
be set to zero length
>Modality (0008,0060) UNIVERSAL; EMPTY; MG N Matching is supported as follows:
SINGLE_VALUE FIXED 1. No modality is supplied (universal matching).
2. Single Value MG is supplied for matching (Single
value matching).
If attribute is absent or empty in answer, the SPS
is accepted.
>Scheduled Performing (0040,0006) UNIVERSAL EMPTY D If value does not meet DICOM encoding
Physician's Name requirements, it will be corrected before usage.
Value is copied into image in attribute Performing
Physician’s Name (0008,1050).
If element is absent or empty in response, the SPS
is accepted, and attribute is not copied into image
(except if edited by the user beforehand).
>Scheduled Procedure Step (0040,0007) UNIVERSAL EMPTY N If value does not meet DICOM encoding
Description requirements, it will be corrected before usage.
If element is absent or empty in answer, the SPS is
accepted, and attribute value is not copied into
images.
Attribute is copied into images into Request
Attribute Sequence (0040,0275).
If Requested Procedure Description (0032,1060)
and Requested Procedure Code Seq. (0032,1064)
code meaning tag values are absent or empty, the
value is also copied to Study description
(0008,1030) in images.
>Scheduled Protocol Code (0040,0008) SEQUENCE FIXED N Sequence is copied into images within sequence
Sequence Request Attribute Sequence (0040,0275)
Page 21 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Attribute Name Tag Matching Type Query Value Value Display Comments
Source on UI
>>Coding Scheme Version (0008,0103) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
>>Code Meaning (0008,0104) UNIVERSAL EMPTY N If answer is received with absent or empty code
value in code sequence, the SPS is accepted.
Sequence will be corrected at image creation.
>Pre-Medication (0040,0012) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
>Scheduled Procedure Step (0040,0009) UNIVERSAL EMPTY N If value does not meet DICOM encoding
ID requirements, SPS is accepted, and value will be
corrected before usage.
Page 22 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Attribute Name Tag Matching Type Query Value Value Display Comments
Source on UI
Requested Procedure Code (0032,1064) UNIVERSAL EMPTY N If sequence is absent or empty in answer, the SPS
Sequence is accepted, and sequence not copied to image.
Page 23 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Attribute Name Tag Matching Type Query Value Value Display Comments
Source on UI
Names of Intended (0040,1010) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Recipients of Results
Imaging Service Request
Accession Number (0008,0050) UNIVERSAL; EMPTY; D Matching is supported as follows:
SINGLE_VALUE USER 1. No value is supplied (universal matching).
2. one value is supplied for matching (Single value
matching).
Page 24 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Attribute Name Tag Matching Type Query Value Value Display Comments
Source on UI
Patient ID (0010,0020) UNIVERSAL; EMPTY; D Matching is supported as follows:
SINGLE VALUE USER No value is supplied (universal matching)
one value is supplied for matching (Single value
matching).
If value does not meet DICOM encoding
requirements, it will be corrected before usage.
If absent or empty in answer, the SPS is accepted,
and attribute will be created empty in images.
Issuer of Patient ID (0010,0021) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Other Patient IDs (0010,1000) UNIVERSAL EMPTY N If value does not meet DICOM encoding
requirements, it will be corrected before usage.
If absent or empty in answer, the SPS is accepted,
and attribute is not copied to images.
Patient Demographics
Patients Birth Date (0010,0030) UNIVERSAL EMPTY D If absent or empty in answer, the SPS is accepted,
and attribute is not copied to images.
If value does not meet DICOM encoding
requirements, it will be corrected before usage.
Patient's Sex (0010,0040) UNIVERSAL EMPTY D If absent or empty in answer, the SPS is accepted,
and attribute is created empty in images.
If value does not meet DICOM encoding
requirements, it will be corrected before usage.
Patient's Weight (0010,1030) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Confidentiality constraint on (0040,3001) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
patient data Description
Patient’s Size (0010,1020) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Patient’s Address (0010,1040) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Patient’s Telephone (0010,2154) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Numbers
Patient Comments (0010,4000) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Patient Medical
Patient State (0038,0500) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Pregnancy Status (0010,21C0) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Medical Alerts (0010,2000) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Allergies (0010,2110) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Special Needs (0038,0050) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Additional Patient History (0010,21B0) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Standard Extended
Study Description (0008,1030) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
NOTE: The study description in image is copied
from Requested Procedure Description.
Page 25 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Attribute Name Tag Matching Type Query Value Value Display Comments
Source on UI
Name of Physician(s) (0008,1060) UNIVERSAL EMPTY N If absent or empty in answer, SPS is accepted.
Reading Study
Operator’s Name (0008,1070) UNIVERSAL EMPTY D If absent or empty in answer, SPS is accepted.
Breast Implant Present (0028,1300) UNIVERSAL EMPTY D If absent or empty in answer, SPS is accepted.
User can cancel the ongoing worklist query by clicking on the Cancel button in the UI. Upon cancel, the Senographe Pristina will send a C-FIND-
CANCEL request to the Modality Worklist SCP.
This implementation can receive multiple C-FIND results over a single association. Only one association is opened at a time.
Each C-FIND response received from the remote AE is parsed to verify the length/type of the items in the response. The DICOM Service AE has a
lenient policy regarding the rejection of the MWL. In case any parameters retrieved from the MWL are invalid, the system will correct
automatically or require/allow the user to enter the value through the UI.
Following acceptance and correction policy is applied on the received worklist items in order to be lenient on DICOM technical violations by
the remote AE:
• Study Instance UID will be locally generated if the Study Instance UID in the worklist item is absent or invalid (for example if it is
more than 64 characters).
• Strings (SH, LO) that are VM=1 and contain \ characters will be processed for HL7v2 escape character conversion:
o \F\ replaced by |
o \S\ replaced by ^
o \E\ replaced by #
o Any other use of \ be replaced by # and the rest of the string will be truncated.
• Strings (CS, SH, LO) that are too long will be truncated. Last character in each truncated value will be set to # (except for CS)
• Replaces individual <CR>, individual <LF>, <CR><LF> combination, and <LF><CR> combination with a canonical NewLine control in
Text (ST, LT) . The NewLine control set is appropriate to the display environment (e.g., a single <LF> character in Unix).
• Text (ST, LT) that is too long will be truncated. Last character in each truncated value will be set to #.
• Names (PN) will be truncated at 64 characters per name component group or after the 5th caret.
• Date/Time (DA, DT, TM) that are Type 2 or 3 and are outside a valid range will be set to zero length.
• CS characters that are lower case will be set to upper case, and – changed to _.
• Any missing Type 2 attributes in the query response will be added in with zero length.
• Any invalid Type 2 attribute after the above processing will be set to zero length.
• Any invalid Type 3 attribute after the above processing will be discarded.
Worklist item that does not comply to this acceptance policy is discarded and the next response (if any) is considered.
Page 26 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
In case of ERROR, persisted worklist item(s) got from previous request to the configured SCP would be shown.
ISO_IR_100 is the Specific Character Set (0008, 0005) value sent during worklist query and while interpreting response.
5.2.1.2 SCP of the Modality Worklist Information Model - FIND SOP Class - N/A
N/A
As a Service Class User of the Modality Performed Procedure Step SOP Class, the Senographe Pristina supports the Attributes listed in Table 5-3
in the N-CREATE-RQ and N-SET-RQ messages, if it creates the message.
Table 5-3 Supported N-CREATE and N-SET Attributes for Modality Performed Procedure Step - SCU
Attribute Name Tag Source Value N-CREATE Value N-SET Comments
Specific Character Set (0008,0005) FIXED ISO_IR 100 ISO_IR 100
Performed Procedure Step Relationship
Scheduled Step Attributes (0040,0270) Not sent
Sequence
>Study Instance UID (0020,000D) MWL; Copied from SPS if received, Not sent
GENERATED else generated.
>Referenced Study (0008,1110) MWL; EMPTY Copied from SPS if received, Not sent
Sequence else sent empty.
>>Referenced SOP Class UID (0008,1150) MWL Copied from SPS if received, Not sent
else not sent.
>>Referenced SOP Instance (0008,1155) MWL Copied from SPS if received, Not sent
UID else not sent.
>Accession Number (0008,0050) MWL; USER; Copied from SPS if received, Not sent
EMPTY or user input if entered, else
sent empty.
>Requested Procedure ID (0040,1001) MWL; EMPTY Copied from SPS if received, Not sent
else sent empty.
Page 27 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Page 28 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Page 29 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Page 30 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
5.2.2.2 SCP of the Modality Performed Procedure Step SOP Class - N/A
N/A
As a Service Class User of the Storage Service Class, the Senographe Pristina uses the C-STORE-RQ message to request storage of DICOM objects
by a remote SCP. See Section 1.1 Content and Transfer in the Overview for the list of supported SOP Classes.
For details regarding the content of SOP Instances that are created by the system, see Annex A, which describes the underlying IOD of the
supported SOP Classes.
Storage requests can be triggered by the user when he first selects the exam/series/image on the local database browser and clicks on the
Remote DICOM AE to which the user desires to send the exam/series/image to.
The user can also trigger storage request of a particular image during the exam session (push-on-demand).
The product can be configured to automatically send acquired exam upon study closure by the user.
The product can be configured to automatically send acquired image upon acquisition of the next image (push-on-the-fly).
All instances that have been created using the same worklist entry are grouped in the same study.
The Senographe Pristina creates one series per mammography view and per type of image.
As a Service Class Provider of the Storage Service Class, the Senographe Pristina receives the C-STORE-RQ message from remote SCUs. See
Section 1.1 Content and Transfer in the Section 1 for the list of supported SOP Classes.
Table 5-4 defines the conformance levels of Senographe Pristina.
Table 5-5 lists any limitations on displaying or processing instances, e.g., display or processing of the respective SOP Instances is prevented by
an unsupported Value for an Attribute or the absence of that Attribute.
The "Effect" column describes what happens if the limitation is encountered. The following Values are used:
Page 31 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
• ND: Display is not possible
Table 5-6 lists the actions performed upon receiving instances from a remote AE and the system behavior when certain conditions are
encountered.
- The stored SOP Instance can be accessed by pushing it from the Senographe Pristina to another device.
- The storage duration is 2 days by default and configurable by the user. After that delay a SOP Instance received from a C-STORE
operation is automatically deleted.
The meaning of each case of an unsuccessful C-STORE response status is described in chapter 7.3.2.5.2 SCP of the Storage SOP Classes - C-
STORE.
As a Service Class User of the Storage Commitment SOP Class, the Senographe Pristina uses the N-ACTION-RQ message to request storage
commitment from a remote SCP. In turn, it receives N-EVENT-REPORT-RQ messages from the SCP indicating success or failure of the request.
Page 32 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
As a Service Class User of the Storage Commitment Push Model SOP Classes the product supports committing all Storage SOP Classes listed in
Section 1.1 Content and Transfer are supported.
The Senographe Pristina doesn’t support the Storage Media File Set ID and UID attributes in the N-ACTION-Request.
N-EVENT-REPORT can be received by the Senographe Pristina on the same Association as the N-ACTION, or on a different one.
If the received N-ACTION Response from the SCP has a success status, the Senographe Pristina waits 10 seconds for an N-EVENT-REPORT on the
same association. This time-out is not configurable. If the N-EVENT-REPORT request is not received in this time, it closes the association and
changes the Job state to ”Waiting” indicating the job is waiting for the response from commitment provider. Then it can receive a N-EVENT-
REPORT from the SCP at any time on a different association.
Up receiving an N-EVENT-REPORT with an Event Type of 1, the “Archived” flag information in the browser for all the successfully archived
exam/series will be updated. The archive status column in the browser will be changed to display “Archived” icon to indicate that the
exam/series has been archived successfully. The job queue entry will be removed.
Table 5-7 lists the behavior of Senographe Pristina for each possible Failure Reason (0008,1197) in the Failed SOP Sequence (0008,1198) upon
receiving an N-EVENT-REPORT request from the SCP with an Event Type ID of 2 (Storage Commitment Request Complete - Failures Exist).
0110 Processing failure: A general failure in processing the operation The system displays the appropriate error message in the job
was encountered. manager user interface and job is paused.
0112 No such object instance: One or more of the elements in the The system displays the appropriate error message in the job
Referenced SOP Instance Sequence was not available. manager user interface and job is paused.
0119 Class / Instance conflict: The SOP Class of an element in the The system displays the appropriate error message in the job
Referenced SOP Instance Sequence did not correspond to the manager user interface and job is paused.
SOP Class registered for this SOP Instance at the SCP.
0122 Referenced SOP Class not supported: Storage Commitment has The system displays the appropriate error message in the job
been requested for a SOP Instance with a SOP Class that is not manager user interface and job is paused.
supported by the SCP.
0131 Duplicate Transaction UID: The Transaction UID of the Storage The system displays the appropriate error message in the job
Commitment Request is already in use. manager user interface and job is paused.
0213 Resource limitation: The SCP does not currently have enough The system displays the appropriate error message in the job
resources to store the requested SOP Instance(s). manager user interface and job is paused.
Jobs paused because of a failure of the storage commitment are not automatically retried. However, the user can manually retry the failed jobs
from the job queue. Such requests will be treated as new requests. This will go through the complete sequence of operations once again.
N/A
Page 33 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
As a Service Class User of the Study Root Q/R - Information Model - FIND SOP Class, the Senographe Pristina uses the C-FIND-RQ message and
supports the Query Keys listed in Table 5-8 for hierarchical queries.
• UNIVERSAL: SCU can request that the Attribute be a return Value (universal matching).
In the "Query Value Source" column the following Values can be used:
• GENERATED: The query Value is generated by the system (e.g., current date as the study date).
• SCANNED: The query Value is read from a barcode scanner or similar device.
• EMPTY: The query Value is sent with a zero-length value to indicate it is a return key only.
Table 5-8 Supported C-FIND Attribute Matching for Study Root Q/R Model -SCU
Attribute Name Tag Matching Type Query Value Source Value Display on UI Comments
Study Level
Study Date (0008,0020) SINGLE_VALUE; USER D
RANGE
Study Time (0008,0030) UNIVERSAL EMPTY N
Accession Number (0008,0050) SINGLE_VALUE; USER N
WILDCARD
Patient's Name (0010,0010) SINGLE_VALUE; USER D
WILDCARD
Patient ID (0010,0020) SINGLE_VALUE; USER D
WILDCARD
Study ID (0020,0010) SINGLE_VALUE; USER N
Page 34 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Attribute Name Tag Matching Type Query Value Source Value Display on UI Comments
WILDCARD
Study Instance UID (0020,000D) UNIVERSAL EMPTY N
Modalities in Study (0008,0061) UNIVERSAL EMPTY D
Station Name (0008,1010) UNIVERSAL EMPTY N
Study Description (0008,1030) UNIVERSAL EMPTY D
Patient’s Birth Date (0010,0030) UNIVERSAL EMPTY N
Patient’s Sex (0010,0040) UNIVERSAL EMPTY N
Patient’s Birth Time (0010,0032) UNIVERSAL EMPTY N
Number of Study Related (0020,1208) UNIVERSAL EMPTY N
Instances
Series Level
Series Number (0008,0011) UNIVERSAL EMPTY D
Modality (0008,0060) UNIVERSAL EMPTY N
Series Instance UID (0020,000E) UNIVERSAL EMPTY N
Series Description (0008,103E) UNIVERSAL EMPTY D
Series Date (0008,0021) UNIVERSAL EMPTY D
Series Time (0008,0031) UNIVERSAL EMPTY N
Series Type (0054,1000) UNIVERSAL EMPTY D
Manufacturer (0008,0070) UNIVERSAL EMPTY N
Number Of Series Related (0020,1209) UNIVERSAL EMPTY N
Instances
Requested Attributes (0040,0275) UNIVERSAL EMPTY N
Sequence
Scheduled Procedure Step (0040,0007) UNIVERSAL EMPTY N
Description
Study Instance UID (0020,000D) UID GENERATED N
Image Type (0008,0008) UNIVERSAL EMPTY N
Instance Level
Instance Number (0020,0013) UNIVERSAL EMPTY D
SOP Instance UID (0008,0018) UNIVERSAL EMPTY N
Image Type (0008,0008) UNIVERSAL EMPTY N
Acquisition Time (0008,0032) UNIVERSAL EMPTY D
Modality (0008,0060) UNIVERSAL EMPTY N
Contrast / Bolus Agent (0018,0010) UNIVERSAL EMPTY N
Slice Thickness (0018,0050) UNIVERSAL EMPTY N
Repetition Time (0018,0080) UNIVERSAL EMPTY N
Echo Time (0018,0081) UNIVERSAL EMPTY N
Page 35 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Attribute Name Tag Matching Type Query Value Source Value Display on UI Comments
Inversion Time (0018,0082) UNIVERSAL EMPTY N
Number Of Averages (0018,0083) UNIVERSAL EMPTY N
Echo Number (0018,0086) UNIVERSAL EMPTY N
Spacing Between Slices (0018,0088) UNIVERSAL EMPTY N
Data Collection Diameter (0018,0090) UNIVERSAL EMPTY N
Trigger Time (0018,1060) UNIVERSAL EMPTY N
Reconstruction Diameter (0018,1100) UNIVERSAL EMPTY N
Gantry / Detector Tilt (0018,1120) UNIVERSAL EMPTY N
Convolution Kernel (0018,1210) UNIVERSAL EMPTY N
Flip Angle (0018,1314) UNIVERSAL EMPTY N
Image Position (Patient) (0020,0032) UNIVERSAL EMPTY N
Image Orientation (0020,0037) UNIVERSAL EMPTY N
(Patient)
Slice Location (0020,1041) UNIVERSAL EMPTY N
Rows (0028,0010) UNIVERSAL EMPTY N
Columns (0028,0011) UNIVERSAL EMPTY N
Pixel Spacing (0028,0030) UNIVERSAL EMPTY N
Units (0054,1001) UNIVERSAL EMPTY N
Reconstruction Method (0054,1103) UNIVERSAL EMPTY N
Image ID (0054,0400) UNIVERSAL EMPTY N
Specific Character Set (0008,0005) UNIVERSAL EMPTY N
Detector Secondary Angle (0018,1531) UNIVERSAL EMPTY D
Study Instance UID (0020,000D) UID GENERATED N
Series Instance UID (0020,000E) UID GENERATED N
ISO_IR_100 is the Specific Character Set (0008, 0005) value sent during query and while interpreting response.
5.2.7.2 SCU of the Patient Root Q/R Information Model - FIND SOP Class - N/A
N/A
5.2.7.3 SCU of the Study Root Q/R Information Model - MOVE SOP Class
When the user initiates a Move operation at any level (Study, Series, Image) the DICOM Service AE will initiate a C-MOVE-RQ request to the
Remote AE with the DICOM Service AE as the Destination AE. The Storage SCP will handle the incoming images.
Each C-MOVE SCU supports an “Association Timer” and an “Operation Inactivity Timer”. These timers are set to 60 and 300 seconds respectively.
The C-MOVE SCU supports only the Baseline behavior and does not negotiate for Extended behavior during association. During the move
operations, Unique Key values for all levels above the Query/Retrieve Level specified will be supplied in the C-MOVE request.
Page 36 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
The DICOM Service AE does not send a C-MOVE request with List of UIDs. If the user chooses multiple exams from the remote browser for
retrieval, it creates separate association and send one C-MOVE request with single exam UID, for each study. The system works the same way
when the user tries to retrieve multiple series.
An individual retrieve-job is created in the job management queue for each exam/series request.
If the user tries to retrieve multiple images within a series, it will create single association and sends multiple C-MOVE requests; one C-MOVE
request per image within the same association.
However, for each user request, a single retrieve-job is created in the job management queue.
If the C-MOVE SCU receives a status different from success (0x0000) or pending (0xFF00) during the association, the DICOM Service AE will release
the association. This information will be logged in the system log files and the network/job manager queue will be updated accordingly.
When a Cancel response (0xFE00) is received, the DICOM Service AE will release the association. This information will be logged in the system log
files and the network queue will be updated accordingly.
5.2.7.4 SCU of the Patient Root Q/R Information Model - MOVE SOP Class - N/A
N/A
5.2.7.5 SCP of the Study Root Q/R Information Model - FIND SOP Class - N/A
N/A
5.2.7.6 SCP of the Patient Root Q/R Information Model - FIND SOP Class - N/A
N/A
5.2.7.7 SCP of the Study Root Q/R Information Model - MOVE SOP Class - N/A
N/A
5.2.7.8 SCP of the Patient Root Q/R - Information Model - MOVE SOP Class - N/A
N/A
The Basic Grayscale Print Management Meta SOP Class is composed of the mandatory SOP Classes listed in Table 5-9.
Table 5-9 Basic Grayscale Print Management Meta SOP Classes - SCU
SOP Class Name SOP Class UID
Basic Film Session 1.2.840.10008.5.1.1.1
Basic Film Box 1.2.840.10008.5.1.1.2
Basic Grayscale Image Box 1.2.840.10008.5.1.1.4
Printer 1.2.840.10008.5.1.1.16
Page 37 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 5-10 Services for the Basic Film Session SOP Class - SCU
DIMSE Service Element Purpose
N-CREATE Create the Film Session
Table 5-11 lists the supported N-CREATE and N-SET Attributes for Basic Film Session:
Table 5-11 Supported N-CREATE and N-SET Attributes for the Basic Film Session SOP Class - SCU
Attribute Name Tag Values Default
Number of Copies (2000,0010) [1…99] 1
Medium Type (2000,0030) BLUE FILM
Film Destination (2000,0040) PROCESSOR
Table 5-12 Supported Services for the Basic Film Box SOP Classes
DIMSE Service Element Purpose
N-CREATE Create the Film Box in a previously created Film Session
N-ACTION Print the Film Box
N-DELETE Delete the Film Box
Table 5-13 lists the supported N-CREATE and N-SET Attributes for Basic Film Box:
Table 5-13 Supported N-CREATE and N-SET Attributes for the Basic Film Box SOP Class - SCU
Attribute Name Tag Values Default
Image Display Format (2010,0010) STANDARD\C,R STANDARD\1,1
Film Orientation (2010,0040) PORTRAIT; LANDSCAPE PORTRAIT
Film Size ID (2010,0050) 14INX17IN; 8INX10IN; 10INX12IN; 10INX14IN; 11INX14IN;
14INX14IN; 24CMX24CM; 24CMX30CM
Magnification Type (2010,0060) REPLICATE; BILINEAR; CUBIC; NONE CUBIC
Smoothing Type (2010,0080) SMOOTH; SHARP SMOOTH
Border Density (2010,0100) BLACK
Trim (2010,0140) YES; NO YES
Configuration Information (2010,0150)
Referenced Film Session Sequence (2010,0500)
>Referenced SOP Class UID (0008,1150) 1.2.840.10008.5.1.1.1
>Referenced SOP Instance UID (0008,1155)
Page 38 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 5-14 Services for the Basic Grayscale Image Box SOP Class
DIMSE Service Element Purpose
N-SET Set Image Attributes for a previously created Film Box
Table 5-15 lists the supported N-SET Attributes for Basic Grayscale Image Box:
Table 5-15 Supported N-SET Attributes for the Basic Grayscale Image Box SOP Class -SCU
Attribute Name Tag Values Default
Image Box Position (2020,0010)
Requested Image Size (2020,0030)
Requested Decimate/Crop Behavior (2020,0040) DECIMATE; CROP; FAIL
Basic Grayscale Image Sequence (2020,0110)
>Samples per Pixel (0028,0002) 1
>Photometric Interpretation (0028,0004) MONOCHROME2
>Rows (0028,0010)
>Columns (0028,0011)
>Bits Allocated (0028,0100) 8; 16
>Bits Stored (0028,0101) 8; 12
>High Bit (0028,0102) 7; 11
>Pixel Representation (0028,0103) 0
>Pixel Data (7FE0,0010)
An N-EVENT-REPORT request can be received by the SCU at any time during an Association.
Table 5-17 summarizes the behavior of the SCU when receiving Event Types within the N-EVENT-REPORT.
Table 5-18 Supported N-GET Attributes for the Printer SOP Class - SCU
Attribute Name Tag Behavior
Printer Status (2110,0010) NORMAL: Status is logged and sends Film
Session N-CREATE request.
WARNING: Status is logged. and sends
Film Session N-CREATE request.
FAILURE: Status is logged. Fails the print
job and displays print failure message on
the Job Manager UI.
Printer Status Info (2110,0020) Logs the message.
Printer Name (2110,0030) Logged if sent by SCP.
Manufacturer (0008,0070) Logged if sent by SCP.
Manufacturer Model Name (0008,1090) Logged if sent by SCP.
Device Serial Number (0018,1000) Logged if sent by SCP.
Software Versions (0018,1020) Logged if sent by SCP.
N-Get request is sent when the user initiates the print activity by clicking on the Print button. Based on the N-Get response, the DICOM
Service AE continues to send the Film Session’s N-CREATE request if status is NORMAL, else pauses the print job and displays the error
message on the Film Job Manager UI.
5.2.8.2 SCU of the Basic Color Print Management Meta SOP Class – N/A
N/A
N/A
N/A
N/A
N/A
5.2.8.7 SCP of the Basic Grayscale Print Management Meta SOP Class - N/A
N/A
5.2.8.8 SCP of the Basic Color Print Management Meta SOP Class - N/A
N/A
Page 40 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
N/A
N/A
N/A
N/A
For a list of supported Media Application Profiles, see Section 1.4 in the Overview.
For a list of supported SOP Classes, see Section 1.1 in the Overview.
For a list of supported SOP Classes, see Section 1.1 in the Overview.
To display or process DICOM Instances contained on the Media, see Section 5.2.5.2.
As a Query SCU, it will accept response items with any value of Specific Character Set. However, it will display in the user interface only
characters specified as within ISO_IR 6 (ASCII) or ISO_IR 100.
Page 41 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
The product user interface will allow the user to enter characters that are within ASCII or ISO_IR 100.
As a Modality Worklist SCU, it only uses ISO_IR 100 Specific Character Set in the worklist query requests.
Generic configuration for Specific Character Sets is covered in Section 6.1. Service specific configuration for Specific Character Sets is addressed
in respective subsections of Section 6.2 or Section 6.3.
6 Configuration
Parameters that can be configured can be accessed through the Service User Interface in the Service Desktop.
Throughout all subsections the following Values can be used in the "Configurable" column:
• FIXED: The parameter is not configurable (it has a fixed Value). The Value is required for the configuration of the remote system.
• N/A: The parameter is not applicable for the local or the remote system.
Page 42 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
In order to identify whether Senographe Pristina is an SCP and / or an SCU, the following applies:
• SCP: The (Secured) Local Called AET and Remote Calling AET parameters are present.
• SCU: The (Secured) Local Calling AET and Remote Called AET parameters are present.
Page 43 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 6-3 MPPS Service Parameters
Local Configuration Parameters - MPPS Service
Parameter Configurable Default Value Comments
Calling AE Title (SCU) SERVICE The Calling AE Title is the one configured
globally for the system and is the same for all
services.
Remote Configuration Parameters - MPPS Service
Parameter Configurable Default Value Comments
Called AE Title (SCP) SERVICE
Port SERVICE
Host SERVICE
Page 44 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Page 45 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Page 46 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Port SERVICE
Host SERVICE
Pixel depth SERVICE 8 Values: 8; 12
Film sizes supported by the Print SCP SERVICE Selection of film sizes that are supported by
the Printer
Default film composer film size SERVICE
Default film composer film SERVICE
orientation
Configuration information SERVICE
Magnification Type SERVICE CUBIC Values: CUBIC; BILINEAR; NONE; REPLICATE
Requested Decimate Crop Behavior SERVICE None Values: None; DECIMATE; CROP; FAIL
Smoothing Factor SERVICE SMOOTH Values: SMOOTH; SHARP
Trim SERVICE YES Values: YES; NO
Page 47 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
7.2 Specifications
7.2.1 DICOM Service Application Entity
7.2.1.1 Sequencing of Real-World Activities for DICOM Service
Following figures describes sequencing of real-world activities during a typical mammography examination with the Senographe Pristina.
Page 48 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Figure 7-1 Sequencing of Real-World Activities for DICOM Service AE – Before starting an exam
The Figure 7-1 describes the sequencing of real-world activities before performing an examination:
- The Senographe Pristina query the worklist from the RIS to get a list of Scheduled Procedure Steps
- The Senographe Pristina can query the PACS to retrieve images of prior study for the patient for which an exam will be performed.
Page 49 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Figure 7-2 Sequencing of Real-World Activities for DICOM Service AE – During and after an exam
The Figure 7-2 describes the sequencing of real-world activities that occurs when an exam is started and until its closure:
- The Senographe Pristina sends a MPPS N-CREATE (in progress) to the PPS Manager when the first image is acquired.
- The Senographe Pristina sends a MPPS N-SET (completed or discontinued) to the PPS Manager when the exam is close.
- At exam closure, or at user request during the exam, the Senographe Pristina pushes the images to a Review System and/or a CAD
(Computer Aided Detection) system using a C-STORE request.
- At exam closure, or at user request during the exam, the Senographe Pristina archive the images to a PACS using a C-STORE request
followed by a Storage Commitment N-ACTION request. Then Senographe Pristina receives the Storage commitment N-EVENT-REPORT
from the PACS.
Page 50 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
The Figure 7-3 describes the sequencing of the real-world activity of Printing. This activity can occur at user request when the Medical
Application is opened, or automatically at exam closure.
N/A
This section details the Association policies of the Application Entity when it is initiating an Association.
If the remote AE does not support the proposed Presentation Context, an appropriate error is logged, and the user is notified.
This implementation can receive multiple C-FIND results over a single association. Only one association is opened at a time.
In case of ERROR, persisted worklist item(s) got from previous request to the configured SCP would be shown.
User can cancel the ongoing worklist query by clicking on the Cancel button in the UI. Upon cancel, the DICOM Service AE will send C-FIND-
CANCEL request to the Modality Worklist SCP.
Extended Negotiation
The Extended Negotiation parameters for all services that are supported by the Application Entity for the Real-World Activity Modality Worklist
Query are described in Table 7-2.
Page 51 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 7-2 Extended Negotiation for Modality Worklist Query of DICOM Service AE - Association Initiation
SOP Class Extended Negotiation Support Requested Value
Modality Worklist
Modality Worklist Information Model - FIND Fuzzy semantic matching of person names N
Timezone query adjustment N
• Start PPS: Initiates a DICOM association to create a DICOM Modality Performed Procedure Step SOP instance in the remote AE via the
N-CREATE service.
• Complete PPS: Initiates a DICOM association to update a DICOM Modality Performed Step instance that is already created with the
remote AE via the N-SET service. The PPS Status is set to ‘COMPLETED’.
• Discontinue PPS: Initiates a DICOM association to update a DICOM Modality Performed Step instance that is already created with the
remote AE via the N-SET service. The PPS Status is set to ‘DISCONTINUED’.
Only one association is opened at a time. The association is open until the response message is returned from the SCP or a network error occurs.
If the remote AE does not support the proposed Presentation context, an appropriate error message is logged.
If multiple exams/series are chosen to be pushed, one association will be established for each of the exam/series. A single association will be
used for multiple C-STORE operations necessary for the images in an individual series for 2D and non-image SOP instances. For 3D SOP
instances, there are multiple separate associations.
This implementation can perform multiple C-STORE operations over a single association.
Upon receiving a C-STORE confirmation containing a successful status, this implementation will perform the next C-STORE operation. The
association will be maintained if possible.
Upon receiving a C-STORE confirmation containing a Refused status, this implementation will consider the current request to be a failure and
will terminate the association.
For all C-STORE statuses other than success or warnings received, this implementation will consider the current request to be a failure and will
terminate the association, except if the C-STORE is invoked from a C-MOVE SCP. In this case it will continue to attempt to send the remaining
images in the request on the same association. The error details will be logged in the log files.
Page 52 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Extended Negotiation
The Extended Negotiation parameters for all services that are supported by the Application Entity for the Real-World Activity Push Images are
described in Table 7-3.
Table 7-3 Extended Negotiation for Push Images of DICOM Service AE - Association Initiation
SOP Class Extended Negotiation Support Requested Value
Storage
Applicable to all Storage SOP Classes listed under Section 5. Level of support N
Level of Digital Signature support N
Element Coercion N
1. If images are successfully sent, then the N-ACTION-RQ request is sent to Storage Commitment Provider in a new association.
(Remote host and storage commitment provider may have different configurations).
2. The system waits for N-ACTION-RSP from Storage Commitment Provider. N-ACTION is on different association than the C-
STORE request.
3. On reception of failure in N-ACTION-RSP, the corresponding job has a failed status in the job management queue.
4. On reception of success, Storage Commitment AE is ready to receive at any time from Storage Commitment Provider the N-
EVENT-REPORT-RQ notification. N-EVENT-REPORT can be in the same association than the N-ACTION or in a new one.
5. On reception of the N-EVENT-REPORT-RQ notification from Storage Commitment Provider, system flags the images in the
database as committed.
6. When all images are flagged, Storage Commitment AE sends a N-EVENT-REPORT-RSP to the Storage Commitment Provider.
The “Query” operation will cause a “Filter Data” menu to appear. The user can enter values for Patient name, Patient Id, Study Id, Accession
number, Study date and Modality. Not entering a value will result in universal match for that field (all possible values). For Study date, the user
selects a date matching type from the "Exam date" pull down menu, where the choices are Today, This Week, 30 days, Date Range. If the user
selects Date Range, the menu will allow entering a start date and an end date. Once the desired parameters are entered, the user chooses
"Search".
The “Query” operation will cause the DICOM Service AE to initiate an association to the selected Remote AE.
The user can also get the detailed content of one or several study displayed by selecting it, the DICOM Service AE to initiate an association to the
selected Remote AE with the Study Instance UID as a matching key (same at series level with Series Instance UID).
Page 53 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
The “Filter” menu contains a “Clear” button. If the user selects it, it will empty all the field in the menu and will cause the DICOM Service AE to
initiate an association to the selected Remote AE, without any matching key.
The user can cancel the current C-FIND request by initiating Query activity to another network node or clicking on the local database. If a query
operation is in progress, the DICOM Service AE will issue a C-FIND-CANCEL message to the remote AE to discontinue the ongoing C-FIND
operation.
If a “Failure” status is returned from the Remote AE, the association is closed, and the operation terminated.
When the user initiates a Retrieve operation at any level (Study, Series, Image) the DICOM Service AE will initiate a C-MOVE-RQ request to the
Remote AE with the DICOM Service AE as the Destination AE. The Storage SCP will handle the incoming images.
The C-MOVE SCU supports only the Baseline behavior and does not negotiate for Extended behavior during association. During the move
operations, Unique Key values for all levels above the Query/Retrieve Level specified will be supplied in the C-MOVE request.
The DICOM Service AE does not send a C-MOVE request with List of UIDs. If the user chooses multiple exams from the remote browser for
retrieval, it creates separate association and send one C-MOVE request with single exam UID, for each study. The system works the same way
when the user tries to retrieve multiple series.
An individual retrieve-job is created in the job management queue for each exam/series request.
If the user tries to retrieve multiple images within a series, it will create single association and sends multiple C-MOVE requests; one C-MOVE
request per image within the same association.
However, for each user request, a single retrieve-job is created in the job management queue.
If the C-MOVE SCU receives a status different from success (0x0000) or pending (0xFF00) during the association, the DICOM Service AE will release
the association. This information will be logged in the system log files and the network/job manager queue will be updated accordingly.
When a Cancel response (0xFE00) is received, the DICOM Service AE will release the association. This information will be logged in the system log
files and the network queue will be updated accordingly.
Extended Negotiation
The Extended Negotiation parameters for all services that are supported by the Application Entity for the Real-World Activity Query/Retrieve are
described in Table 7-4.
Table 7-4 Extended Negotiation for Query/Retrieve of DICOM Service AE - Association Initiation
SOP Class Extended Negotiation Support Requested Value
Query
Applicable to all Query Retrieve - FIND SOP Classes Relational queries N
mentioned under Section 5.
Date-time matching N
Fuzzy semantic matching of person names N
Timezone query adjustment N
Enhanced Multi-Frame Image Conversion N
Retrieve
Applicable to all Query Retrieve - MOVE SOP Classes Relational retrieval N
Page 54 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
N/A
• Autoprint feature.
• Manual print.
• Film composer.
To print manually, the user must select the DICOM printer, from Viewing Applications, and press Print Button.
1. User must select the DICOM printer from the Film Composer Interface.
2. The images to be printed shall be selected into Film Composer slots from Viewing Applications.
3. The user must press the Print Button to print the images.
The DICOM Service AE will start the Print Session. The Print Session involves establishing association with the printer followed by the next
sequence of activities:
a. The DICOM Service AE gets the Printer status using N-GET service. If the Printer returns FAILURE status the print session will be
terminated, and the requester will be notified of the printer status.
b. The film session is created using N-CREATE service. In case of error, the print session will be terminated. The attribute values for the
Film session will be specified with the film session request.
c. The film box is created using N-CREATE service. The print session will be terminated if the printer fails to create the film box. The film
box attribute values will be sent in the film box create request.
d. The image attributes for the images to be printed in this session will be set using the N-SET service. If the printer fails to accommodate
the images in the image set, the print session will be terminated. If the user used “film composer”, they will be as many N-SET box as
images in the film.
e. The film will be printed using the N-ACTION service. Only film box printing is supported. In case of error, the print session will be
terminated.
f. The Film Box instance will be deleted using the N-DELETE service.
g. The SCU does not wait for N-EVENT-REPORT from the Printer after deleting the film box instance. The N-EVENT-REPORT received when
the association was still active is handled, but the data received will be ignored.
h. Finally, the association will be terminated and if all the above operations are successful the requester will be notified of the successful
print session. This status just indicates that the images to be printed have been successfully sent to the printer.
Page 55 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
7.2.1.3.7 Real-World Activity Verify Connectivity
The user can choose to verify a remote Application Entity. Upon user’s request, DICOM Service AE sends out a verification request to a Remote
AE.
The DICOM Service AE provides standard conformance. In case of failure, the verification is not retried.
Each C-ECHO operation supports an “Association Timer”. This timer starts when the association request is sent and stops when the association
is established. Time-out is 20 seconds. If the remote AE does not respond with the association time-out, the verification operation is considered
as failed.
This section details the Association policies of the Application Entity when it is the acceptor of an Association.
Extended Negotiation
The Extended Negotiation parameters for all services that are requested by the Application Entity for the Real-World Activity Receive Images are
described in Table 7-5.
Table 7-5 Extended Negotiation for Receive Images of DICOM Service AE - Association Acceptance
SOP Class Extended Negotiation Support Requested Value
Storage
Applicable to all Storage SOP Classes listed under Section 5. Level of support N
Level of Digital Signature support N
Element Coercion N
This section provides tables that describe the Transfer Syntax preference for different SOP Classes or SOP Class groups when there are multiple
Transfer Syntaxes provided by the Association initiator for Real-World Activity Receive images of DICOM Service AE of the system.
When there are multiple Transfer Syntax provided by the Association initiator, selection of the Transfer Syntax by the DICOM Service AE is not
based on a preference order, but DICOM Service AE will select the first in the presentation context that is supported.
Table 7-6 Transfer Syntax Selection Preference Order - Image SOP Classes for DICOM Service AE
Preference Order Transfer Syntax UID Comments
0 Explicit VR Little-Endian Transfer Syntax 1.2.840.10008.1.2.1 Chosen if it is the first in the Presentation Context
among the supported Transfer Syntaxes.
0 Implicit VR Little-Endian Transfer Syntax 1.2.840.10008.1.2 Chosen if it is the first in the Presentation Context
among the supported Transfer Syntaxes.
0 Explicit VR Big-Endian Transfer Syntax 1.2.840.10008.1.2.2 Chosen if it is the first in the Presentation Context
Page 56 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 7-7 Transfer Syntax Selection Preference Order - Non-Image SOP Classes for DICOM Service AE
Preference Order Transfer Syntax UID Comments
0 Explicit VR Little-Endian Transfer Syntax 1.2.840.10008.1.2.1 Chosen if it is the first in the Presentation Context
among the supported Transfer Syntaxes.
0 Implicit VR little-Endian Transfer Syntax 1.2.840.10008.1.2 Chosen if it is the first in the Presentation Context
among the supported Transfer Syntaxes.
0 Explicit VR Big-Endian Transfer Syntax 1.2.840.10008.1.2.2 Chosen if it is the first in the Presentation Context
among the supported Transfer Syntaxes.
N-EVENT-REPORT can be received by the Senographe Pristina on the same Association as the N-ACTION, or on a different one.
If the received N-ACTION Response from the SCP has a success status, the Senographe Pristina waits 10 seconds for an N-EVENT-REPORT on the
same association. This time-out is not configurable. If the N-EVENT-REPORT request is not received in this time, it closes the association and
changes the Job state to ”Waiting” indicating the job is waiting for the response from commitment provider. Then it can receive a N-EVENT-
REPORT from the SCP at any time on a different association.
This section provides tables that describe the Transfer Syntax preference for different SOP Classes or SOP Class groups when there are multiple
Transfer Syntaxes provided by the Association initiator for Real-World Activity Receive images of DICOM Service AE of the system.
When there are multiple Transfer Syntax provided by the Association initiator, selection of the Transfer Syntax by the DICOM Service AE is not
based on a preference order, but DICOM Service AE will select the first in the presentation context that is supported.
Table 7-8 Transfer Syntax Selection Preference Order - Non-Image SOP Classes for DICOM Service AE
Preference Order Transfer Syntax UID Comments
0 Explicit VR Little-Endian Transfer Syntax 1.2.840.10008.1.2.1 Chosen if it is the first in the Presentation Context
among the supported Transfer Syntaxes.
0 Implicit VR little-Endian Transfer Syntax 1.2.840.10008.1.2 Chosen if it is the first in the Presentation Context
among the supported Transfer Syntaxes.
0 Explicit VR Big-Endian Transfer Syntax 1.2.840.10008.1.2.2 Chosen if it is the first in the Presentation Context
among the supported Transfer Syntaxes.
Page 57 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Role negotiation
The Storage Commitment Provider initiating the association must use the role selection negotiation.
Any remote AE can open an association to the DICOM Service AE for the purpose of verification.
The DICOM Service AE monitors an “Operation Inactivity” timer. The connection with the SCU will be terminated if it is inactive for the time-out
interval. Time-out is 15 seconds and is not configurable.
This section provides tables that describe the Transfer Syntax preference for different SOP Classes or SOP Class groups when there are multiple
Transfer Syntaxes provided by the Association initiator for Real-World Activity Receive images of DICOM Service AE of the system.
When there are multiple Transfer Syntax provided by the Association initiator, selection of the Transfer Syntax by the DICOM Service AE is not
based on a preference order, but DICOM Service AE will select the first in the presentation context that is supported.
Table 7-9 Transfer Syntax Selection Preference Order - Non-Image SOP Classes for DICOM Service AE
Preference Order Transfer Syntax UID Comments
0 Explicit VR Little-Endian Transfer Syntax 1.2.840.10008.1.2.1 Chosen if it is the first in the Presentation Context
among the supported Transfer Syntaxes.
0 Implicit VR little-Endian Transfer Syntax 1.2.840.10008.1.2 Chosen if it is the first in the Presentation Context
among the supported Transfer Syntaxes.
0 Explicit VR Big-Endian Transfer Syntax 1.2.840.10008.1.2.2 Chosen if it is the first in the Presentation Context
among the supported Transfer Syntaxes.
Table 7-12 Status Codes for C-FIND of the Modality Worklist Information Model SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success Matching is complete - No final identifier is supplied 0000 Displays all the worklists items obtained from
the SCP.
Failure Refused: Out of Resources A700 Logs the error and displays all the persisted
worklists.
SOP Class Not Supported 0122 Worklist query operation is considered as failure
and the error message is displayed to the user.
Persisted worklist item(s) got from previous
request to the SCP will be shown.
Error: Data Set does not match SOP Class A900 Logs the error and displays all the persisted
worklists.
Error: Unable to process C000-CFFF Logs the error and displays all the persisted
worklists.
Cancel Matching terminated due to cancel FE00 Closes the association.
Pending Matches are continuing - Current Match is supplied and any FF00 Continues to wait for worklists items from SCP.
Optional Keys were supported in the same manner as
Required Keys.
Matches are continuing - Warning that one or more Optional FF01 Continues to wait for worklists items from SCP.
Keys were not supported for existence for this Identifier
- Other status codes anything else Worklist query operation is considered as failure
and the error message is displayed to the user.
Persisted worklist item(s) got from previous
request to the SCP will be shown.
7.3.2.1.2 SCP of the Modality Worklist Information Model Find SOP Class - C-FIND - N/A
N/A
Page 59 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 7-13 Status Codes for N-CREATE of the Modality Performed Procedure Step SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success Success 0000 The system updates the state and indicates the same on the
user interface.
Failure No Such Attribute 0105 Error message is logged into the system log-file and indicates
the status on the user interface.
Invalid Attribute Value 0106 Error message is logged into the system log-file and indicates
the status on the user interface.
Processing Failure 0110 Error message is logged into the system log-file and indicates
the status on the user interface. Retries the operation after the
configured time is elapsed.
Invalid SOP Instance 0117 Error message is logged into the system log-file and indicates
the status on the user interface.
No Such SOP Class 0118 Error message is logged into the system log-file and indicates
the status on the user interface.
Missing Attribute 0120 Error message is logged into the system log-file and indicates
the status on the user interface.
Duplicate Invocation 0210 Error message is logged into the system log-file and indicates
the status on the user interface.
Unrecognized Operation 0211 Error message is logged into the system log-file and indicates
the status on the user interface.
Mistyped Argument 0212 Error message is logged into the system log-file and indicates
the status on the user interface.
Resource Limitation 0213 Error message is logged into the system log-file and indicates
the status on the user interface. Retries the operation after the
configured time is elapsed.
- Other status codes anything else The operation is deemed ‘Failed’. Detailed message is logged
into the system log-file and indicates the status on the user
interface. Retries the operation after the configured time is
elapsed.
7.3.2.2.2 SCU of the Modality Performed Procedure Step SOP Class - N-SET
Table 7-14 lists the Status Codes that the SCU of the Modality Performed Procedure Step SOP Class supports for the N-SET message and defines
the application behavior when encountering the listed Status Codes.
Table 7-14 Status Codes for N-SET of the Modality Performed Procedure Step SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success Success 0000 The system updates the state and indicates the same on the
user interface.
Page 60 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
7.3.2.2.3 SCP of the Modality Performed Procedure Step SOP Class - N-CREATE - N/A
N/A
7.3.2.2.4 SCP of the Modality Performed Procedure Step SOP Class - N-SET - N/A
N/A
N/A
N/A
Page 61 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 7-15 Status Codes C-STORE for the Storage SOP Classes - SCU
Service Status Further Meaning Status Code Behavior
Success Success 0000 Success status displayed in the Network status and browser job
manager queue. Log files updated.
Warning Coercion of Data Elements B000 Operation considered as successful. Warning is logged in the log files.
Data Set does not match SOP Class B007 Operation considered as successful. Warning is logged in the log files.
Elements Discarded B006 Operation considered as successful. Warning is logged in the log files.
Failure Error: Processing Failure 0110 Operation considered as failed. Error displayed in the Network status
and browser job manager queue. Log-files updated.
Push job is retried
Out of Resources A700-A7FF Operation considered as failed. Error displayed in the Network status
and browser job manager queue. Log-files updated.
Push job is retried
Data Set does not match SOP Class A900-A9FF Operation considered as failed. Error displayed in the Network status
and browser job manager queue. Log-files updated.
Push job is not retried
Cannot Understand C000-CFFF Operation considered as failed. Error displayed in the Network status
and browser job manager queue. Log-files updated.
Push job is retried
- Other status codes anything else Operation considered as failed. Error displayed in the Network status
and browser job manager queue. Log-files updated.
Table 7-16 Status Codes C-STORE of the Storage SOP Classes - SCP
Service Status Further Meaning Status Codes Related Fields Condition (and Comments on Related
fields)
Success Success 0000 N/A DICOM instance stored successfully.
Refused Refused: Out of Resources A700 (0000,0902) Not enough diskspace to store the DICOM
object.
Refused: Out of Resources A711 (0000,0902) Unable to connect to local database for
storage (such as maximum connection limit
reached).
Failure Error: Data Set does not match SOP Class A900 (0000,0902) Storage of the DICOM object failed due to
corrupt/invalid dataset.
Error: Cannot understand C000 (0000,0902) Error while storing DICOM object in the
repository.
Page 62 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 7-17 Status Codes for N-ACTION of the Storage Commitment Push Model SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success 0000 The request for storage comment is considered successfully sent.
System displays “waiting” message in the job manager user interface.
Failure Processing failure 0110 The system displays the appropriate error message in the job
manager user interface.
No such argument 0114 The system displays the appropriate error message in the job
manager user interface.
Invalid argument Value 0115 The system displays the appropriate error message in the job
manager user interface.
Invalid Object instance 0117 The system displays the appropriate error message in the job
manager user interface.
No such SOP Class 0118 The system displays the appropriate error message in the job
manager user interface.
Class-instance conflict 0119 The system displays the appropriate error message in the job
manager user interface.
Unrecognized operation 0211 The system displays the appropriate error message in the job
manager user interface.
Mistyped argument 0212 The system displays the appropriate error message in the job
manager user interface.
Resource limitation 0213 The system logs the appropriate error message and retries the
operation after the configured time is elapsed.
- Other status codes anything else The system displays the appropriate error message in the job
manager user interface.
7.3.2.6.2 SCU of the Storage Commitment Push Model SOP Class - N-EVENT-REPORT
Table 7-18 lists the Status Codes that the SCU of the Storage Commitment Push Model SOP Class supports for the N-EVENT-REPORT message
and defines the application behavior when encountering the listed Status Codes.
Table 7-18 Status Codes for N-EVENT-REPORT for the Storage Commitment Push Model SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success 0000 The storage commitment result received successfully.
Failure Processing failure 0110 Indicates that an internal error occurred while processing.
7.3.2.6.3 SCP of the Storage Commitment Push Model SOP Class - N-ACTION - N/A
N/A
7.3.2.6.4 SCP of the Storage Commitment Push Model SOP Class - N-EVENT-REPORT - N/A
N/A
Page 63 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 7-19 Status Codes C-FIND for Query/Retrieve FIND SOP Classes - SCU
Service Status Further Meaning Status Code Behavior
Success Matching is complete - No final identifier is 0000 Processed and gracefully exits the C-FIND request
supplied processing.
Failure Refused: Out of Resources A700 Browser displays appropriate error to user.
Error: Data Set does not match SOP Class A900 Browser displays appropriate error to user.
Error: Unable to process C000-CFFF Browser displays appropriate error to user.
Cancel Matching terminated due to cancel FE00 1. Considered as invalid status and the operation is
terminated if the query CANCEL was not requested.
2. If the query CANCEL was requested by the SCU, then
system gracefully exits the C-FIND request processing.
Pending Matches are continuing - Current Match is FF00 Processed and the data is displayed in the Browser.
supplied and any Optional Keys were supported
in the same manner as Required Keys.
Matches are continuing - Warning that one or FF01 Processed and the data is displayed in the Browser.
more Optional Keys were not supported for
existence for this Identifier
- Other status codes anything else Operation terminated and the association is closed.
Browser displays appropriate error to user.
Table 7-20 Status Codes C-MOVE for Query/Retrieve MOVE SOP Classes - SCU
Service Status Further Meaning Status Codes Related Fields Behavior
Success Sub-operations Complete - No Failures 0000 (0000,1020) Processed and gracefully exits C-MOVE request
(0000,1021) processing.
(0000,1022)
(0000,1023)
Warning Sub-operations Complete - One or more B000 (0000,1020) The system will display the failure message in
Failures (0000,1022) the job manager user interface.
(0000,1023)
Failed Out of Resources - Unable to calculate A701 (0000,0902) The system will display the failure message in
number of matches job manager user interface and retry the
operation after time is elapsed.
Out of Resources - Unable to perform sub- A702 (0000,1020) The system will display the failure message and
operations (0000,1021) retry the operation after configured time is
(0000,1022) elapsed.
(0000,1023)
Move Destination unknown A801 (0000,0902) The system will display the failure message in
job manager user interface.
Data Set does not match SOP Class A900 (0000,0901) The system will display the failure message in
(0000,0902) job manager user interface.
Page 64 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Cancel Sub-operations terminated due to Cancel FE00 (0000,1020) 1. Considered as invalid status and the
Indication (0000,1021) operation is terminated if the Move CANCEL
(0000,1022) was not requested.
(0000,1023) 2. If the Move CANCEL was requested by the
SCU, then the system gracefully exits the C-
MOVE request processing.
Pending Sub-operations are continuing FF00 (0000,1020) The system processes the information and
(0000,1021) displays the progress in the job manager UI.
(0000,1022)
(0000,1023)
- Other status codes anything else - C-MOVE operation is considered as failure. The
system will display the failure message in the
job manager user interface and log the
information in the log files.
N/A
N/A
Table 7-21 Status Codes for N-CREATE of the Basic Film Session SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success Success 0000 Logs film session instance UID and proceeds for creating film box.
Warning Memory allocation not supported B600 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
Failure No Such Attribute 0105 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
Invalid Attribute Value 0106 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
Duplicate SOP Instance 0111 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
Invalid SOP Instance 0117 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
No Such SOP Class 0118 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
Page 65 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
7.3.2.8.1.2 SCU of the Basic Film Session SOP Class - N-SET – N/A
N/A
7.3.2.8.1.3 SCU of the Basic Film Session SOP Class - N-DELETE – N/A
N/A
7.3.2.8.1.4 SCU of the Basic Film Session SOP Class - N-ACTION – N/A
N/A
7.3.2.8.2.1 SCU of the Basic Film Box Session SOP Class - N-CREATE
Table 7-21 lists the Status Codes that the SCU of the Basic Film Box SOP Class supports for the N-CREATE message and defines the application
behavior when encountering the listed Status Codes.
Table 7-22 Status Codes for N-CREATE of the Basic Film Box SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success Success 0000 Proceed for setting image box.
Warning Requested Min Density or Max Density outside B605 Fails print job and displays print failure message on Job
of printer’s operating range. The printer will use Manager UI. Error details along with the status code is logged.
its respective minimum or maximum density
value instead.
Failure No such attributes 0105 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Invalid Attribute Value 0106 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Processing Failure 0110 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Duplicate SOP Instance 0111 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Invalid SOP Instance 0117 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Page 66 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
7.3.2.8.2.2 SCU of the Basic Film Box Session SOP Class - N-SET -N/A
N/A
7.3.2.8.2.3 SCU of the Basic Film Box Session SOP Class - N-DELETE
Table 7-23 lists the Status Codes that the SCU of the Basic Film Box SOP Class supports for the N-DELETE message and defines the application
behavior when encountering the listed Status Codes.
Table 7-23 Status Codes for N-DELETE of the Basic Film Box SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success Success 0000 Log the success message.
Failure Processing Failure 0110 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
Invalid SOP Instance 0117 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
No Such SOP Class 0118 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
Class Instance Conflict 0119 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
Refused: Not Authorized 0124 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
Duplicate Invocation 0210 Fails print job and displays print failure message on Job Manager UI.
Error details along with the status code is logged.
Page 67 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
7.3.2.8.2.4 SCU of the Basic Film Box Session SOP Class - N-ACTION
Table 7-24 lists the Status Codes that the SCU of the Basic Film Box SOP Class supports for the N-ACTION message and defines the application
behavior when encountering the listed Status Codes.
Table 7-24 Status Codes for N-ACTION of the Basic Film Box SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success Success 0000 Proceed in parsing N-Action response.
Warning Film Box SOP Instance hierarchy does not B603 Fails print job and displays print failure message on Job
contain Image Box SOP Instances (empty page) Manager UI. Error details along with the status code is logged.
Image size is larger than Image Box size. The B604 Fails print job and displays print failure message on Job
image has been demagnified. Manager UI. Error details along with the status code is logged.
Image size is larger than Image Box size. The B609 Fails print job and displays print failure message on Job
image has been cropped to fit. Manager UI. Error details along with the status code is logged.
Image size or Combined Print Image Size is larger B60A Fails print job and displays print failure message on Job
than Image Box size. The image or combined Manager UI. Error details along with the status code is logged.
Print Image has been decimated to fit.
Failure Processing failure 0110 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
No such SOP Instance 0112 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
No Such Argument 0114 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Invalid argument Value 0115 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Invalid SOP Instance 0117 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
No Such SOP Class 0118 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Class-Instance Conflict 0119 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
No Such Action 0123 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Refused: Not Authorized 0124 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Page 68 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
7.3.2.8.3 SCU of the Basic Grayscale Image Box SOP Class - N-SET
Table 7-25 lists the Status Codes that the SCU of the Basic Grayscale Image Box SOP Class supports for the N-SET message and defines the
application behavior when encountering the listed Status Codes.
Table 7-25 Status Codes for N-SET of the Grayscale Image Box SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success Success 0000 Continues setting image box if more images are there, else
sends N-Action request.
Warning Image size is larger than Image Box size. The B604 Fails print job and displays print failure message on Job
image has been demagnified. Manager UI. Error details along with the status code is logged.
Requested Min Density or Max Density outside of B605 Fails print job and displays print failure message on Job
printer's operating range. Manager UI. Error details along with the status code is logged.
Image size is larger than Image Box size. The B609 Fails print job and displays print failure message on Job
image has been cropped to fit. Manager UI. Error details along with the status code is logged.
Image size or Combined Print Image Size is larger B60A Fails print job and displays print failure message on Job
than Image Box size. The image or combined Manager UI. Error details along with the status code is logged.
Print Image has been decimated to fit.
Failure No Such Attribute 0105 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Invalid Attribute Value 0106 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Processing Failure 0110 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Duplicate SOP Instance 0111 Fails print job and displays print failure message on Job
Manager UI. Error details along with the status code is logged.
Page 69 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
7.3.2.8.4 SCU of the Basic Color Image Box SOP Class - N-SET - N/A
N/A
Table 7-26 Status Codes for N-EVENT-REPORT of the Printer SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success Success 0000
Failure Processing Failure 0110
No Such SOP Instance 0112
Page 70 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table 7-27 Status Codes for N-GET of the Printer SOP Class - SCU
Service Status Further Meaning Status Code Behavior
Success Success 0000 Proceed with parsing dataset returned from N-Get response and perform
appropriate operations.
Warning Attribute List Error 0107 Fails print job and displays print failure message on Job Manager UI. Error
details along with the status code is logged.
Failure Processing Failure 0110 Fails print job and displays print failure message on Job Manager UI. Error
details along with the status code is logged.
No Such SOP Instance 0112 Fails print job and displays print failure message on Job Manager UI. Error
details along with the status code is logged.
Invalid SOP Instance 0117 Fails print job and displays print failure message on Job Manager UI. Error
details along with the status code is logged.
No Such SOP Class 0118 Fails print job and displays print failure message on Job Manager UI. Error
details along with the status code is logged.
Class-Instance Conflict 0119 Fails print job and displays print failure message on Job Manager UI. Error
details along with the status code is logged.
Refused: Not Authorized 0124 Fails print job and displays print failure message on Job Manager UI. Error
details along with the status code is logged.
Duplicate Invocation 0210 Fails print job and displays print failure message on Job Manager UI. Error
details along with the status code is logged.
Unrecognized Operation 0211 Fails print job and displays print failure message on Job Manager UI. Error
details along with the status code is logged.
Mistyped Argument 0212 Fails print job and displays print failure message on Job Manager UI. Error
details along with the status code is logged.
Resource Limitation 0213 Fails print job and displays print failure message on Job Manager UI. Error
Page 71 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
7.3.2.8.6 SCU of the Basic Annotation Box SOP Class - N-SET - N/A
N/A
N/A
N/A
7.3.2.8.9 SCU of the Printer Configuration Retrieval SOP Class - N-GET - N/A
N/A
N/A
N/A
7.3.2.8.12 SCP of the Basic Grayscale Image Box SOP Class - N-SET - N/A
N/A
7.3.2.8.13 SCP of the Basic Color Image Box SOP Class - N-SET - N/A
N/A
N/A
7.3.2.8.15 SCP the Basic Annotation Box SOP Class - N-SET - N/A
N/A
N/A
N/A
7.3.2.8.18 SCP of the Printer Configuration Retrieval SOP Class - N-GET - N/A
N/A
Page 72 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
8 Security
8.1 Introduction
The security section describes security features implemented by this product. It includes descriptions of non-DICOM network protocols,
information to configure firewalls and application whitelists, lists of supported DICOM security profiles as well as Web Security features.
Additionally, secured media storage, VPN, etc. are also specified in this security section.
Page 73 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Page 74 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Page 75 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Annexes
A Information Object Definitions (IODs)
This section describes all the SOP Instances natively created by Senographe Pristina , e.g., images created by an acquisition modality or evidence documents created on a review
workstation (i.e., all SOP Classes that are marked in the "Created" column in Table 1-1). Details on Attribute coercion are defined in Section 5.2.5.2.
• QUERY: The Value is determined by performing a query of any of the supported Query/Retrieve Services.
The "Presence" columns reflect the usage of the Module, Functional Group Macro, Attributes, or Value in the Senographe Pristina Implementation and is not necessarily the same
as defined in the DICOM Standard. For the "Presence" column the following Values can be used:
• ALWAYS: the module, functional group macro, Attributes or Value is always present.
• CONDITIONAL: the presence of the module, functional group macro, Attributes or Value is dependent on a condition. The condition must be listed in the "Conditions" column.
• SRC_COPY: The presence of the Attributes and Values depends on the availability of these in the source instances, which are used for copying this information.
Page 76 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>Referenced SOP Class UID (0008,1150) MWL CONDITIONAL ALWAYS If received from RIS
>Referenced SOP Instance (0008,1155) MWL CONDITIONAL ALWAYS If received from RIS
UID
Procedure Code Sequence (0008,1032) MWL CONDITIONAL ALWAYS One or more items may If received from RIS If sequence is received
be present incomplete, it will be
corrected as described
below in sequence
elements
>Code Value (0008,0100) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
copied with code
meaning
>Coding Scheme Designator (0008,0102) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
value is set to
”99LOCAL”.
>Code Meaning (0008,0104) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
value is set to the code
value.
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Series Time (0008,0031) GENERATED ALWAYS ALWAYS Time when generating
new series
Performing Physician's (0008,1050) MWL; USER CONDITIONAL CONDITIONAL If received from RIS or
Name entered by user.
Protocol Name (0018,1030) GENERATED ALWAYS ALWAYS ROUTINE; 3D_ROUTINE;
CESM;
3D_ROUTINE+2D_ROUT
INE; STEREO;
3D_BIOPSY; SAMPLE;
CESM_BIOPSY
Series Description (0008,103E) GENERATED ALWAYS ALWAYS
Operators' Name (0008,1070) MWL; USER CONDITIONAL CONDITIONAL If received from RIS or
entered by user.
Referenced Performed (0008,1111) GENERATED CONDITIONAL ALWAYS Only one item is If a PPS manager has
Procedure Step Sequence present been declared.
>Referenced SOP Class UID (0008,1150) FIXED ALWAYS ALWAYS 1.2.840.10008.3.1.2.3.3
>Referenced SOP Instance (0008,1155) GENERATED ALWAYS ALWAYS
UID
Body Part Examined (0018,0015) FIXED CONDITIONAL ALWAYS BREAST Not sent for specimen
images.
Patient Position (0018,5100) EMPTY CONDITIONAL EMPTY Only in CT images. Sent empty in CT images,
not sent otherwise
Request Attributes (0040,0275) MWL CONDITIONAL ALWAYS One or more items may If received from RIS.
Sequence be present
>Requested Procedure ID (0040,1001) MWL CONDITIONAL ALWAYS If received from RIS.
>Requested Procedure (0032,1060) MWL CONDITIONAL ALWAYS If received from RIS.
Description
>Scheduled Procedure Step (0040,0009) MWL CONDITIONAL ALWAYS If received from RIS.
ID
>Scheduled Procedure Step (0040,0007) MWL CONDITIONAL CONDITIONAL If received from RIS.
Description
Page 79 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>Scheduled Protocol Code (0040,0008) MWL CONDITIONAL ALWAYS One or more items may If received from RIS. If sequence is received
Sequence be present incomplete, it will be
corrected as described
below in sequence
elements
>>Code Value (0008,0100) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
copied with code
meaning
>>Coding Scheme (0008,0102) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
Designator value is set to ”99LOCAL”.
>>Code Meaning (0008,0104) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
value is set to the code
value.
Performed Procedure Step (0040,0253) GENERATED CONDITIONAL ALWAYS If a PPS manager has
ID been declared.
Performed Procedure Step (0040,0244) GENERATED CONDITIONAL ALWAYS If a PPS manager has
Start Date been declared.
Performed Procedure Step (0040,0245) GENERATED CONDITIONAL ALWAYS If a PPS manager has
Start Time been declared.
Performed Procedure Step (0040,0254) CONDITIONAL ALWAYS If a PPS manager has
Description been declared.
Page 80 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Page 81 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Photometric Interpretation (0028,0004) GENERATED ALWAYS ALWAYS MONOCHROME1; MONOCHROME1 in For
MONOCHROME2; Processing images,
MONOCHROME2 in For
Presentation and BTO
images.
Rows (0028,0010) GENERATED ALWAYS ALWAYS
Columns (0028,0011) GENERATED ALWAYS ALWAYS
Bits Allocated (0028,0100) FIXED ALWAYS ALWAYS 16
Bits Stored (0028,0101) GENERATED ALWAYS ALWAYS 14; 12 14 in For Processing
images, 12 in For
Presentation images.
High Bit (0028,0102) GENERATED ALWAYS ALWAYS 13; 11 13 in For Processing
images, 11 in For
Presentation images.
Pixel Representation (0028,0103) FIXED ALWAYS ALWAYS 0000H
Pixel Data (7FE0,0010) GENERATED ALWAYS ALWAYS
Pixel Padding Range Limit (0028,0121) GENERATED CONDITIONAL ALWAYS if For Presentation image.
Page 82 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Content Qualification (0018,9004) FIXED CONDITIONAL ALWAYS PRODUCT Sent only for BTO
Page 83 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Image Laterality (0020,0062) USER ALWAYS ALWAYS R; L; B
Anatomic Region Sequence (0008,2218) FIXED ALWAYS ALWAYS Only one item is
present
>Code Value (0008,0100) FIXED ALWAYS ALWAYS T-04000
>Coding Scheme Designator (0008,0102) FIXED ALWAYS ALWAYS SRT
>Code Meaning (0008,0104) FIXED ALWAYS ALWAYS Breast
Page 84 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Page 85 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Shutter Presentation Value (0018,1622) FIXED ALWAYS ALWAYS 0
Page 86 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Page 87 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table A-18 Breast Tomosynthesis Image IOD
IE Module Name Presence (Module) Condition Reference
Patient Patient ALWAYS Table A-1
Study General Study ALWAYS Table A-2
Patient Study ALWAYS Table A-3
Series General Series ALWAYS Table A-4
Enhanced Mammography Series ALWAYS Table A-20
Frame of Reference Frame of Reference ALWAYS Table A-5
Equipment General Equipment ALWAYS Table A-6
Enhanced General Equipment ALWAYS Table A-7
Image Image Pixel ALWAYS Table A-8
Acquisition Context ALWAYS Table A-9
Multi-frame Functional Groups ALWAYS Table A-21
X-Ray 3D Image ALWAYS Table A-22
Breast Tomosynthesis Contributing Sources ALWAYS Table A-23
Breast Tomosynthesis Acquisition ALWAYS Table A-24
X-Ray 3D Reconstruction ALWAYS Table A-25
Breast View ALWAYS Table A-26
SOP Common ALWAYS Table A-10
Standard Extended Module for Breast Tomosynthesis Image IOD ALWAYS Table A-36
Private Module for Breast Tomosynthesis Image IOD ALWAYS Table A-37
Table A-19 lists the Functional group macros used in Breast Tomosynthesis Image IOD. The usage column defines whether a Macro is used as a shared macro, on a per Frame base
or whether depending on the acquisition context can be used in both contexts. The following values are supported:
• PER_FRAME: The macro is used on a per frame basis, the attributes are included in the Shared Functional Group Sequence (5200,9229)
• SHARED: The macro is shared across all frames; the attributes are included in the Per-Frame Functional Group Sequence (5200,9230)
• CONTEXT_DEPENDENT: depending on the acquisition context the macro can either be used on a per frame basis or be shared across all frames.
Page 88 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table A-19 Functional Group Macros used in Breast Tomosynthesis Image IOD
Functional Group Macro Presence Condition Usage Reference
Pixel Measures ALWAYS SHARED Table A-27
Frame Content ALWAYS PER_FRAME Table A-28
Plane Position (Patient) ALWAYS PER_FRAME Table A-29
Plane Orientation (Patient) ALWAYS SHARED Table A-30
Derivation Image CONDITIONAL Present in slabs and in volumes when SHARED Table A-31
lossy compression is applied
Frame Anatomy ALWAYS SHARED Table A-32
Identity Pixel Value Transformation ALWAYS SHARED Table A-33
Frame VOI LUT With LUT ALWAYS SHARED Table A-34
X-Ray 3D Frame Type ALWAYS PER_FRAME Table A-35
Page 89 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>Scheduled Procedure Step (0040,0009) MWL CONDITIONAL CONDITIONAL If received from RIS.
ID
>Scheduled Procedure Step (0040,0007) MWL CONDITIONAL ALWAYS One or more items may If received from RIS. If sequence is received
Description be present incomplete, it will be
corrected as described
below in sequence
elements
>Scheduled Protocol Code (0040,0008) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
Sequence copied with code
meaning
>>Code Value (0008,0100) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
value is set to ”99LOCAL”.
>>Coding Scheme (0008,0102) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
Designator value is set to the code
value.
Page 90 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table A-22 X-Ray 3D Image Module
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Image Type (0008,0008) GENERATED ALWAYS ALWAYS See Section A.2.4
Pixel Presentation (0008,9205) FIXED ALWAYS ALWAYS MONOCHROME
Volumetric Properties (0008,9206) GENERATED ALWAYS ALWAYS VOLUME; SAMPLED VOLUME for Planes,
SAMPLED for Slabs.
Volume Based Calculation (0008,9207) FIXED ALWAYS ALWAYS TOMOSYNTHESIS
Technique
Bits Allocated (0028,0100) FIXED ALWAYS ALWAYS 16
Bits Stored (0028,0101) FIXED ALWAYS ALWAYS 12
High Bit (0028,0102) FIXED ALWAYS ALWAYS 11
Samples per Pixel (0028,0002) FIXED ALWAYS ALWAYS 1
Photometric Interpretation (0028,0004) FIXED ALWAYS ALWAYS MONOCHROME2
Content Qualification (0018,9004) FIXED ALWAYS ALWAYS PRODUCT
Burned In Annotation (0028,0301) FIXED ALWAYS ALWAYS NO
Lossy Image Compression (0028,2110) GENERATED ALWAYS ALWAYS 00; 01
Lossy Image Compression (0028,2112) GENERATED CONDITIONAL ALWAYS When lossy compression
Ratio applied
Lossy Image Compression (0028,2114) FIXED CONDITIONAL ALWAYS ISO_10918_1 When lossy compression
Method applied
Presentation LUT Shape (2050,0020) FIXED ALWAYS ALWAYS IDENTITY
Page 91 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>>Referenced Series (0008,1115) GENERATED ALWAYS ALWAYS Only one item is
Sequence present
>>>Series Instance UID (0020,000E) SRC_INSTANCE ALWAYS ALWAYS
>>>Series Number (0020,0011) SRC_INSTANCE ALWAYS ALWAYS
>>>Referenced Instance (0008,114A) GENERATED ALWAYS ALWAYS 1 item per projection
Sequence image
>>>>Referenced SOP Class (0008,1150) SRC_INSTANCE ALWAYS ALWAYS
UID
>>>>Referenced SOP (0008,1155) SRC_INSTANCE ALWAYS ALWAYS
Instance UID
>>>>Instance Number (0020,0013) SRC_INSTANCE ALWAYS ALWAYS
>Manufacturer (0008,0070) SRC_INSTANCE ALWAYS ALWAYS
>Manufacturer's Model (0008,1090) SRC_INSTANCE ALWAYS ALWAYS
Name
>Device Serial Number (0018,1000) SRC_INSTANCE ALWAYS ALWAYS
>Software Versions (0018,1020) SRC_INSTANCE ALWAYS ALWAYS
>Acquisition DateTime (0008,002A) SRC_INSTANCE ALWAYS ALWAYS
>Station Name (0008,1010) SRC_INSTANCE ALWAYS ALWAYS
>Operators' Name (0008,1070) SRC_INSTANCE ALWAYS ALWAYS
>Protocol Name (0018,1030) SRC_INSTANCE ALWAYS ALWAYS
>Acquisition Protocol Name (0018,9423) SRC_INSTANCE ALWAYS ALWAYS
>Rows (0028,0010) SRC_INSTANCE ALWAYS ALWAYS
>Columns (0028,0011) SRC_INSTANCE ALWAYS ALWAYS
>Bits Stored (0028,0101) SRC_INSTANCE ALWAYS ALWAYS
>Lossy Image Compression (0028,2110) SRC_INSTANCE ALWAYS ALWAYS
>Detector Type (0018,7004) SRC_INSTANCE ALWAYS ALWAYS
>Detector ID (0018,700A) SRC_INSTANCE ALWAYS ALWAYS
>Date of Last Detector (0018,700C) SRC_INSTANCE ALWAYS ALWAYS
Calibration
>Time of Last Detector (0018,700E) SRC_INSTANCE ALWAYS ALWAYS
Calibration
Page 92 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>Detector Element Spacing (0018,7022) SRC_INSTANCE ALWAYS ALWAYS
Page 93 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>Exposure in mAs (0018,9332) GENERATED ALWAYS ALWAYS Set to the converted total
of the individual
projections’ values
(0018,1153).
>Start Acquisition DateTime (0018,9516) GENERATED ALWAYS ALWAYS
>Primary Positioner Scan (0018,9508) GENERATED ALWAYS ALWAYS
Arc
>Primary Positioner Scan (0018,9510) GENERATED ALWAYS ALWAYS
Start Angle
>Primary Positioner (0018,9514) GENERATED ALWAYS ALWAYS
Increment
>Distance Source to (0018,1110) GENERATED ALWAYS ALWAYS Distance from source to
Detector image plane at 0°
>Distance Source to Patient (0018,1111) GENERATED ALWAYS ALWAYS Distance from source to
breast support at 0°
>Estimated Radiographic (0018,1114) FIXED ALWAYS ALWAYS 1
Magnification Factor
>Anode Target Material (0018,1191) GENERATED ALWAYS ALWAYS
>Body Part Thickness (0018,11A0) GENERATED ALWAYS ALWAYS
>Exposure Control Mode (0018,7060) GENERATED ALWAYS ALWAYS MANUAL; AUTOMATIC
>Exposure Control Mode (0018,7062) GENERATED ALWAYS ALWAYS
Description
>Half Value Layer (0040,0314) GENERATED ALWAYS ALWAYS
>Organ Dose (0040,0316) GENERATED ALWAYS ALWAYS
>Entrance Dose in mGy (0040,8302) GENERATED ALWAYS ALWAYS Total entrance dose of
the acquisition sequence.
>Focal Spot(s) (0018,1190) GENERATED ALWAYS ALWAYS
>Detector Temperature (0018,7001) GENERATED ALWAYS ALWAYS
>Filter Type (0018,1160) GENERATED ALWAYS ALWAYS
>Filter Material (0018,7050) GENERATED ALWAYS ALWAYS
>Compression Force (0018,11A2) GENERATED ALWAYS ALWAYS
>Paddle Description (0018,11A4) GENERATED ALWAYS ALWAYS
Page 94 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>Per Projection Acquisition (0018,9538) GENERATED ALWAYS ALWAYS On item per projection
Sequence image.
>>KVP (0018,0060) GENERATED ALWAYS ALWAYS
>>X-Ray Tube Current in mA (0018,9330) GENERATED ALWAYS ALWAYS Set to conversion from
attribute X-Ray Tube
Current (0018,1151) in
source image.
>>Collimator Shape (0018,1700) GENERATED ALWAYS ALWAYS
>>Collimator Left Vertical (0018,1702) GENERATED ALWAYS ALWAYS
Edge
>>Collimator Right Vertical (0018,1704) GENERATED ALWAYS ALWAYS
Edge
>>Collimator Upper (0018,1706) GENERATED ALWAYS ALWAYS
Horizontal Edge
>>Collimator Lower (0018,1708) GENERATED ALWAYS ALWAYS
Horizontal Edge
>>Positioner Primary Angle (0018,1510) GENERATED ALWAYS ALWAYS
>>Positioner Primary Angle (0018,9559) FIXED ALWAYS ALWAYS CC
Direction
>>Exposure Time in ms (0018,9328) GENERATED ALWAYS ALWAYS Set to the conversion
from attribute Exposure
Time (0018,1150) in
source image.
>>Exposure in mAs (0018,9332) GENERATED ALWAYS ALWAYS Set to the convention
from attribute Exposure
(0018,1152) in source
image.
>>Relative X-Ray Exposure (0018,1405) GENERATED ALWAYS ALWAYS
>>Organ Dose (0040,0316) GENERATED ALWAYS ALWAYS
>>Entrance Dose in mGy (0040,8302) GENERATED ALWAYS ALWAYS
>>Irradiation Event UID (0008,3010) GENERATED ALWAYS ALWAYS
Page 95 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table A-25 X-Ray 3D Reconstruction Module
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
X-Ray 3D Reconstruction (0018,9530) GENERATED ALWAYS ALWAYS One or more items may
Sequence be present
>Reconstruction Description (0018,9531) GENERATED ALWAYS ALWAYS Description of algorithm
used for 3D
reconstruction
>Application Name (0018,9524) GENERATED ALWAYS ALWAYS
>Application Version (0018,9525) GENERATED ALWAYS ALWAYS
>Application Manufacturer (0018,9526) FIXED ALWAYS ALWAYS GE HEALTHCARE
>Algorithm Type (0018,9527) FIXED ALWAYS ALWAYS ITERATIVE
>Acquisition Index (0020,9518) FIXED ALWAYS ALWAYS 1
Page 96 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Page 97 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Table A-29 Plane Position (Patient) Macro
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Plane Position Sequence (0020,9113) GENERATED ALWAYS ALWAYS Only one item is
present
>Image Position (Patient) (0020,0032) GENERATED ALWAYS ALWAYS
Page 98 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>Source Image Sequence (0008,2112) GENERATED ALWAYS ALWAYS One or 2 items may be For Slabs, an item is sent
present with reference to the
corresponding Planes
image.
For Planes and Slabs
when lossy compression
is applied, an item is sent
with reference to the
corresponding
uncompressed image.
>>Referenced SOP Class UID (0008,1150) SRC_INSTANCE ALWAYS ALWAYS
>>Referenced SOP Instance (0008,1155) SRC_INSTANCE ALWAYS ALWAYS
UID
>>Purpose of Reference (0040,A170) GENERATED ALWAYS ALWAYS Only one item is
Code Sequence present.
>>>Code Value (0008,0100) GENERATED ALWAYS ALWAYS 121322; 121320
>>>Coding Scheme (0008,0102) FIXED ALWAYS ALWAYS DCM
Designator
>>>Code Meaning (0008,0104) GENERATED ALWAYS ALWAYS Source image for image
processing operation;
Uncompressed
predecessor
Page 99 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>>Coding Scheme (0008,0102) FIXED ALWAYS ALWAYS SRT
Designator
>>Code Meaning (0008,0104) FIXED ALWAYS ALWAYS BREAST
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>Frame Type (0008,9007) GENERATED ALWAYS ALWAYS See Section A.2.4
>Pixel Presentation (0008,9205) FIXED ALWAYS ALWAYS MONOCHROME
>Volumetric Properties (0008,9206) GENERATED ALWAYS ALWAYS VOLUME; SAMPLED VOLUME for Planes,
SAMPLED for Slabs
>Volume Based Calculation (0008,9207) FIXED ALWAYS ALWAYS TOMOSYNTHESIS
Technique
>Reconstruction Index (0020,9536) FIXED ALWAYS ALWAYS 1
Attribute Name Tag VR VM Identifiable Source Presence of Presence of Value Conditions Comments
Information Attribute Value
>Biopsy device ID (0045,10AF) SH 1 SAFE GENERATED ALWAYS ALWAYS Set to the needle ID
used during the
medical procedure
when target is sent.
X-Ray 3D (0018,9507) SQ 1 SAFE GENERATED ALWAYS ALWAYS
Acquisition
Sequence
>Per Projection (0018,9538) SQ 1-n SAFE GENERATED ALWAYS ALWAYS
Acquisition
Sequence
>>Detector (0018,1531) DS 1 SAFE SRC_INSTANCE ALWAYS ALWAYS
Secondary angle
>>Detector active (0018,7026) DS 1-2 SAFE SRC_INSTANCE ALWAYS ALWAYS
dimension
Attribute Name Tag VR VM Identifiable Source Presence of Presence of Value Conditions Comments
Information Attribute Value
Paddle Properties (0045,10AD) LO 1-n SAFE GENERATED ALWAYS ALWAYS
Biopsy Tool (0045,10B0) SQ 1-n SAFE GENERATED CONDITIONAL ALWAYS Only for Biopsy
Sequence procedure.
>Private Creator (0045,0010) LO 1 SAFE FIXED ALWAYS ALWAYS GEMS_SENO_02
>ID of the Biopsy (0045,10AF) SH 1 SAFE GENERATED ALWAYS ALWAYS
Device
>Biopsy Tool Name (0045,10B1) ST 1 SAFE USER ALWAYS ALWAYS
>Needle Post Fire (0045,10B2) IS 1 SAFE USER ALWAYS ALWAYS Length of the needle
post fire in 1/100 mm
>Needle Notch (0045,10B3) IS 1 SAFE USER ALWAYS ALWAYS Length of the needle
notch in 1/100 mm
>Needle Tip (0045,10B4) IS 1 SAFE USER ALWAYS ALWAYS Length of the needle
tip in 1/100 mm
>Needle Diameter (0045,10B5) IS 1 SAFE USER ALWAYS ALWAYS Needle diameter
value in gauge
>Needle guide head (0045,10B6) IS 1 SAFE USER ALWAYS ALWAYS Needle guide
thickness thickness value in
1/100 mm. Used only
in case of micro
biopsy. Otherwise,
value is set to 0.
>Name of the body (0045,10B7) ST 1 SAFE USER ALWAYS ALWAYS Revolve; ST; Magnum;
Vacora; Eviva; EnCor;
Generic; FNA and
HOOK; Calibration
Tool
>Interface type (0045,10B8) ST 1 SAFE USER ALWAYS ALWAYS H; V
>Body length (0045,10B9) IS 1 SAFE USER ALWAYS ALWAYS Length of the body in
1/100 mm.
Attribute Name Tag VR VM Identifiable Source Presence of Presence of Value Conditions Comments
Information Attribute Value
>Device Type (0045,10BA) ST 1 SAFE USER ALWAYS ALWAYS CB; VAD; CALIB; FNA;
GENERIC
>Coaxial Cannula's (0045,10BB) IS 1 SAFE USER ALWAYS ALWAYS Coaxial cannula's
Length upper length in 1/100
mm
>Needle guide total (0045,10BC) IS 1 SAFE USER ALWAYS ALWAYS Full length of the
length needle in 1/100 mm
>Body Model Image (0045,10BD) ST 1 SAFE GENERATED ALWAYS ALWAYS
Path
> Needle Axis To (0045,10BE) IS 1 SAFE USER CONDITIONAL ALWAYS When Biopsy Tool Distance between
Body Bottom Width is Generic the closest gun part
Distance to breast support in
horizontal position
and the needle in
1/100mm
>Needle Axis To Body (0045,10BF) IS 1 SAFE USER CONDITIONAL ALWAYS When Biopsy Tool Distance between
Top Width Distance is Generic the furthest gun part
to breast support in
horizontal position
and the needle in
1/100mm
>Needle Spacer (0045,10C0) IS 1 SAFE USER ALWAYS ALWAYS Needle spacer in
length 1/100mm
Attribute Name Tag VR VM Identifiable Source Presence of Presence of Value Conditions Comments
Information Attribute Value
Sigmoid Windowing (0045,10C7) DS 2 SAFE GENERATED CONDITIONAL ALWAYS If VOI LUT
parameters Function
(0028,1056) is
LINEAR
A.2.4 Breast Tomosynthesis Image IOD Specific Values and Code Sets
Table A-17 lists Coded Values referenced from the "Value" column of the tables above.
Table A-38 Values and Code Sets for Breast Tomosynthesis Image IOD
Attribute Name Tag Value/Code Condition Comments
Image Type (0008,0008) DERIVED\PRIMARY\TOMOSYNTHESIS\MAXIMUM 3D Routine Slabs image
ORIGINAL\PRIMARY\TOMOSYNTHESIS\NONE 3D Routine Planes image
DERIVED\PRIMARY\TOMOSYNTHESIS\NONE 3D Routine Lossy compressed Planes image
ORIGINAL\PRIMARY\TOMO_SCOUT\NONE 3D Biopsy Planes image
DERIVED\PRIMARY\TOMO_SCOUT\NONE 3D Biopsy Lossy compressed Planes image
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Exposure in microAs (0018,1153) GENERATED ALWAYS ALWAYS Set the total of the
individual projections’
values from projections
tags (0018,1153).
Filter Type (0018,1160) SRC_INSTANCE ALWAYS ALWAYS
Focal Spot(s) (0018,1190) SRC_INSTANCE ALWAYS ALWAYS
Anatomic Region Sequence (0008,2218) FIXED ALWAYS ALWAYS Only one item is
present
>Code Value (0008,0100) FIXED ALWAYS ALWAYS T-04000
>Coding Scheme Designator (0008,0102) FIXED ALWAYS ALWAYS SRT
>Code Meaning (0008,0104) FIXED ALWAYS ALWAYS Breast
View Code Sequence (0054,0220) SRC_INSTANCE ALWAYS ALWAYS Only one item is
present
>Code Value (0008,0100) SRC_INSTANCE ALWAYS ALWAYS
>Coding Scheme Designator (0008,0102) SRC_INSTANCE ALWAYS ALWAYS
>Code Meaning (0008,0104) SRC_INSTANCE ALWAYS ALWAYS
>View Modifier Code (0054,0222) SRC_INSTANCE ALWAYS SRC_COPY Zero or more items may
Sequence be present
>>Code Value (0008,0100) SRC_INSTANCE SRC_COPY ALWAYS
>>Coding Scheme (0008,0102) SRC_INSTANCE SRC_COPY ALWAYS
Designator
>>Code Meaning (0008,0104) SRC_INSTANCE SRC_COPY ALWAYS
Attribute Name Tag VR VM Identifiable Source Presence of Presence of Value Conditions Comments
Information Attribute Value
Filter Material (0018,7050) CS 1 SAFE SRC_INSTANCE ALWAYS ALWAYS
Organ Dose (0040,0316) DS 1 SAFE SRC_INSTANCE ALWAYS ALWAYS
Entrance Dose in (0040,8302) DS 1 SAFE SRC_INSTANCE ALWAYS ALWAYS
mGy
Acquisition Device (0018,1401) LO 1 SAFE GENERATED ALWAYS ALWAYS
Processing Code
Attribute Name Tag VR VM Identifiable Source Presence of Presence of Value Conditions Comments
Information Attribute Value
Private Creator (0045,0010) LO 1 SAFE
Breast ROI X (0045,1060) IS 4 SAFE GENERATED ALWAYS ALWAYS
Breast ROI Y (0045,1061) IS 4 SAFE GENERATED ALWAYS ALWAYS
Radiological (0045,1049) DS 1 SAFE SRC_INSTANCE ALWAYS ALWAYS
Thickness
Reconstruction (0045,10A7) LT 1-n SAFE GENERATED ALWAYS ALWAYS
parameters
Sigmoid Windowing (0045,10C7) DS 2 SAFE GENERATED CONDITIONAL ALWAYS If VOI LUT
parameters Function
(0028,1056) is
LINEAR
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>Scheduled Procedure Step (0040,0009) MWL CONDITIONAL ALWAYS If received from RIS.
ID
>Scheduled Procedure Step (0040,0007) MWL CONDITIONAL CONDITIONAL If received from RIS.
Description
>Scheduled Protocol Code (0040,0008) MWL CONDITIONAL ALWAYS One or more items may If received from RIS. If sequence is received
Sequence be present incomplete, it will be
corrected as described
below in sequence
elements
>>Code Value (0008,0100) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
copied with code
meaning
>>Coding Scheme (0008,0102) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
Designator value is set to ”99LOCAL”.
>>Code Meaning (0008,0104) MWL CONDITIONAL ALWAYS If received from RIS If missing in sequence,
value is set to the code
value.
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Samples per Pixel (0028,0002) FIXED ALWAYS ALWAYS 1
Photometric Interpretation (0028,0004) GENERATED ALWAYS ALWAYS MONOCHROME1; For Processing:
MONOCHROME2 MONOCHROME1.
For Presentation:
MONOCHROME2.
Bits Allocated (0028,0100) FIXED ALWAYS ALWAYS 16
Bits Stored (0028,0101) GENERATED ALWAYS ALWAYS 12; 14 For Processing: 14.
For Presentation: 12.
High Bit (0028,0102) GENERATED ALWAYS ALWAYS 11; 13 For Processing: 13.
For Presentation: 11.
Pixel Representation (0028,0103) FIXED ALWAYS ALWAYS 0000H
Pixel Intensity Relationship (0028,1040) GENERATED ALWAYS ALWAYS LIN; LOG For Processing: LIN.
For Presentation: MG.
Pixel Intensity Relationship (0028,1041) GENERATED ALWAYS ALWAYS +1; -1 For Processing: +1.
Sign For Presentation: -1.
Rescale Intercept (0028,1052) FIXED ALWAYS ALWAYS 0
Rescale Slope (0028,1053) FIXED ALWAYS ALWAYS 1
Rescale Type (0028,1054) FIXED ALWAYS ALWAYS US
Presentation LUT Shape (2050,0020) GENERATED ALWAYS ALWAYS IDENTITY; INVERSE For Processing: INVERSE
For Presentation: -
IDENTITY.
Lossy Image Compression (0028,2110) FIXED ALWAYS ALWAYS 00
Acquisition Device (0018,1400) GENERATED CONDITIONAL ALWAYS Only in For Presentation
Processing Description images
Acquisition Device (0018,1401) GENERATED CONDITIONAL ALWAYS Only in For Presentation
Processing Code images
Patient Orientation (0020,0020) GENERATED CONDITIONAL ALWAYS Not sent for specimen
images
Burned In Annotation (0028,0301) FIXED ALWAYS ALWAYS NO
VOI LUT Sequence (0028,3010) GENERATED CONDITIONAL ALWAYS One to 4 items may be If VOI LUT Function 1 item per Window
present (0028,1056) is SIGMOID Center values (0028,1050)
>LUT Descriptor (0028,3002) GENERATED ALWAYS ALWAYS
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>LUT Explanation (0028,3003) GENERATED ALWAYS ALWAYS NORMAL; SOFTER;
HARDER; USER
>LUT Data (0028,3006) GENERATED ALWAYS ALWAYS
Window Center (0028,1050) GENERATED CONDITIONAL CONDITIONAL If For Presentation image
Window Width (0028,1051) GENERATED CONDITIONAL CONDITIONAL If For Presentation image
Window Center & Width (0028,1055) GENERATED CONDITIONAL ALWAYS NORMAL; SOFTER; If For Presentation image
Explanation HARDER; USER
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Field of View Dimension(s) (0018,1149) GENERATED ALWAYS ALWAYS
Field of View Origin (0018,7030) GENERATED ALWAYS ALWAYS
Field of View Rotation (0018,7032) GENERATED ALWAYS ALWAYS 270; 180; 90; 0
Field of View Horizontal Flip (0018,7034) GENERATED ALWAYS ALWAYS YES; NO
Imager Pixel Spacing (0018,1164) FIXED ALWAYS ALWAYS 0.1\0.1
Pixel Spacing (0028,0030) GENERATED ALWAYS ALWAYS
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>>Code Value (0008,0100) GENERATED ALWAYS ALWAYS See Section A.4.4
>>Coding Scheme (0008,0102) GENERATED ALWAYS ALWAYS See Section A.4.4
Designator
>>Code Meaning (0008,0104) GENERATED ALWAYS ALWAYS See Section A.4.4
Distance Source to Patient (0018,1111) GENERATED CONDITIONAL ALWAYS Not sent for specimen Distance in mm from the
images. source to the breast
support.
Distance Source to Detector (0018,1110) GENERATED ALWAYS ALWAYS Distance in mm from the
source to the image
plane.
Estimated Radiographic (0018,1114) GENERATED ALWAYS ALWAYS Set to SID/(SOD – 20 mm)
Magnification Factor for all images except
breast spacer and
specimen images.
Set to SID/(SOD – 40mm)
for images with breast
spacer.
Set to SID/SOD for
specimen images.
Positioner Type (0018,1508) FIXED ALWAYS ALWAYS MAMMOGRAPHIC
Positioner Primary Angle (0018,1510) GENERATED ALWAYS ALWAYS
Detector Secondary Angle (0018,1531) GENERATED ALWAYS ALWAYS
Body Part Thickness (0018,11A0) GENERATED ALWAYS ALWAYS
Compression Force (0018,11A2) GENERATED ALWAYS ALWAYS
Paddle Description (0018,11A4) GENERATED CONDITIONAL ALWAYS Not sent for specimen
images.
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
X-Ray Tube Current (0018,1151) GENERATED CONDITIONAL ALWAYS Not sent for CESM
Recombined.
Exposure Time (0018,1150) GENERATED CONDITIONAL ALWAYS Not sent for CESM
Recombined.
Exposure (0018,1152) GENERATED CONDITIONAL ALWAYS Not sent for CESM
Recombined.
Exposure in microAs (0018,1153) GENERATED CONDITIONAL ALWAYS Not sent for CESM
Recombined.
Distance Source to Detector (0018,1110) GENERATED ALWAYS ALWAYS Distance in mm from the
source to the image
plane.
Distance Source to Patient (0018,1111) GENERATED CONDITIONAL ALWAYS Not sent for specimen Distance in mm from the
images. source to the breast
support.
Body Part Thickness (0018,11A0) GENERATED CONDITIONAL ALWAYS Not sent for specimen
images.
Relative X-Ray Exposure (0018,1405) GENERATED CONDITIONAL ALWAYS Not sent for CESM
Recombined and
specimen images.
Entrance Dose (0040,0302) GENERATED CONDITIONAL ALWAYS Not sent for CESM
Recombined and
specimen images.
Entrance Dose in mGy (0040,8302) GENERATED CONDITIONAL ALWAYS Not sent for CESM
Recombined and
specimen images.
Distance Source to Entrance (0040,0306) GENERATED ALWAYS ALWAYS Distance in mm from the
source to the
compression paddle.
Value is set to the at 0°
for all TOMO projections.
Comments on Radiation (0040,0310) GENERATED CONDITIONAL ALWAYS Not sent for specimen Breast glandularity in %.
Dose images.
Half Value Layer (0040,0314) GENERATED CONDITIONAL ALWAYS Not sent for CESM
Recombined and
specimen images.
Page 118 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Organ Dose (0040,0316) GENERATED CONDITIONAL ALWAYS Not sent for CESM
Recombined and
specimen images.
Organ Exposed (0040,0318) FIXED ALWAYS ALWAYS BREAST
Anode Target Material (0018,1191) GENERATED ALWAYS ALWAYS MOLYBDENUM;
RHODIUM
Filter Type (0018,1160) FIXED CONDITIONAL ALWAYS STRIP Not sent for CESM
Recombined.
Filter Material (0018,7050) GENERATED CONDITIONAL ALWAYS MOLYBDENUM; SILVER; Not sent for CESM
COPPER Recombined.
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Anatomic Region Sequence (0008,2218) FIXED ALWAYS ALWAYS Only one item is
present
>Code Value (0008,0100) FIXED ALWAYS ALWAYS T-04000
>Coding Scheme Designator (0008,0102) FIXED ALWAYS ALWAYS SRT
>Code Meaning (0008,0104) FIXED ALWAYS ALWAYS Breast
View Code Sequence (0054,0220) GENERATED ALWAYS ALWAYS Only one item is
present
>Code Value (0008,0100) GENERATED ALWAYS ALWAYS See Section A.4.4
>Coding Scheme Designator (0008,0102) GENERATED ALWAYS ALWAYS See Section A.4.4
>Code Meaning (0008,0104) GENERATED ALWAYS ALWAYS See Section A.4.4
>View Modifier Code (0054,0222) GENERATED ALWAYS CONDITIONAL Zero or one items may If view modifier has been
Sequence be present selected by the user.
>>Code Value (0008,0100) GENERATED ALWAYS ALWAYS See Section A.4.4
>>Coding Scheme (0008,0102) GENERATED ALWAYS ALWAYS See Section A.4.4
Designator
>>Code Meaning (0008,0104) GENERATED ALWAYS ALWAYS See Section A.4.4
Biopsy Target Sequence (0018,2041) GENERATED CONDITIONAL ALWAYS One or more items may Only for Biopsy
be present procedure, in “For
Presentation” images,
except specimen images.
>Target UID (0018,2042) GENERATED ALWAYS ALWAYS
>Localizing Cursor Position (0018,2043) GENERATED ALWAYS ALWAYS
>Calculated Target Position (0018,2044) GENERATED ALWAYS ALWAYS
>Displayed Z Value (0018,2046) GENERATED ALWAYS ALWAYS
>Target Label (0018,2045) GENERATED ALWAYS ALWAYS
A.4.2 Digital Mammography X-Ray Image IOD Specific Functional Group Macros – N/A
N/A
Attribute Name Tag VR VM Identifiable Source Presence of Presence of Value Value Conditions Comments
Information Attribute
Breast ROI Y (0045,1061) IS 4 SAFE GENERATED CONDITIONAL ALWAYS If For
Presentation
image.
User Window (0045,1062) IS 1 SAFE GENERATED ALWAYS ALWAYS
Center
User Window (0045,1063) IS 1 SAFE GENERATED ALWAYS ALWAYS
Width
Segmentation (0045,1064) IS 1 SAFE GENERATED CONDITIONAL ALWAYS If For Processing
Threshold image.
Image Crop point (0045,1072) DS 2 SAFE GENERATED ALWAYS ALWAYS
Signal Average (0045,10A0) DS 1 SAFE GENERATED CONDITIONAL ALWAYS Only for CESM
Factor procedure
Organ dose for (0045,10A1) DS 2-n SAFE GENERATED CONDITIONAL ALWAYS Only for CESM
source images procedure
Entrance dose in (0045,10A2) DS 2-n SAFE GENERATED CONDITIONAL ALWAYS Only for CESM
mGy for source procedure
images
X-ray Source (0045,10A3) DS 3 SAFE GENERATED CONDITIONAL ALWAYS Only for Biopsy
Position procedure
Cumulative (0045,10AB) DS 1 SAFE GENERATED CONDITIONAL ALWAYS If DBT projection
organ dose image.
Cumulative (0045,10AC) DS 1 SAFE GENERATED CONDITIONAL ALWAYS If DBT projection
entrance dose image.
Paddle (0045,10AD) LO 1-n SAFE GENERATED CONDITIONAL ALWAYS Not sent for
Properties specimen
images
Biopsy Tool (0045,10B0) SQ 1-n SAFE GENERATED CONDITIONAL ALWAYS Only for Biopsy
Sequence procedure.
>Private Creator (0045,0010) LO 1 SAFE FIXED ALWAYS ALWAYS GEMS_SENO_02
Attribute Name Tag VR VM Identifiable Source Presence of Presence of Value Value Conditions Comments
Information Attribute
>ID of the Biopsy (0045,10AF) SH 1 SAFE GENERATED ALWAYS ALWAYS
device
>Biopsy Device (0045,10B1) ST 1 SAFE USER ALWAYS ALWAYS
Name
>Needle Post Fire (0045,10B2) IS 1 SAFE USER ALWAYS ALWAYS Length of the
needle post fire
in 1/100 mm
>Needle Notch (0045,10B3) IS 1 SAFE USER ALWAYS ALWAYS Length of the
needle notch in
1/100 mm
>Needle Tip (0045,10B4) IS 1 SAFE USER ALWAYS ALWAYS Length of the
needle tip in
1/100 mm
>Needle (0045,10B5) IS 1 SAFE USER ALWAYS ALWAYS Needle
Diameter diameter value
in gauge
>Needle guide (0045,10B6) IS 1 SAFE USER ALWAYS ALWAYS Needle guide
head thickness thickness value
in 1/100 mm.
Used only in
case of micro
biopsy.
Otherwise,
value is set to 0.
>Name of the (0045,10B7) ST 1 SAFE USER ALWAYS ALWAYS Revolve; ST; Magnum;
body Vacora; Eviva; EnCor;
Generic; FNA and
HOOK; Calibration Tool
>Interface type (0045,10B8) ST 1 SAFE USER ALWAYS ALWAYS H; V
>Body length (0045,10B9) IS 1 SAFE USER ALWAYS ALWAYS Length of the
body in 1/100
mm.
Attribute Name Tag VR VM Identifiable Source Presence of Presence of Value Value Conditions Comments
Information Attribute
>Device Type (0045,10BA) ST 1 SAFE USER ALWAYS ALWAYS CB; VAD; CALIB; FNA;
GENERIC
>Coaxial (0045,10BB) IS 1 SAFE USER ALWAYS ALWAYS Coaxial
Cannula's Length cannula's upper
length in 1/100
mm
>Needle guide (0045,10BC) IS 1 SAFE USER ALWAYS ALWAYS Full length of
total length the needle in
1/100 mm
>Body Model (0045,10BD) ST 1 SAFE GENERATED ALWAYS ALWAYS
Image Path
> Needle Axis To (0045,10BE) IS 1 SAFE USER CONDITIONAL ALWAYS When Biopsy Distance
Body Bottom Tool is Generic between the
Width Distance closest gun part
to breast
support in
horizontal
position and the
needle in
1/100mm
>Needle Axis To (0045,10BF) IS 1 SAFE USER CONDITIONAL ALWAYS When Biopsy Distance
Body Top Width Tool is Generic between the
Distance furthest gun
part to breast
support in
horizontal
position and the
needle in
1/100mm
>Needle Spacer (0045,10C0) IS 1 SAFE USER ALWAYS ALWAYS Needle spacer in
length 1/100mm
Attribute Name Tag VR VM Identifiable Source Presence of Presence of Value Value Conditions Comments
Information Attribute
Distance image (0045,10C1) DS 1 SAFE GENERATED CONDITIONAL ALWAYS Only for Biopsy
plan to breast images, except
support specimen.
Breast Support (0045,10C3) LO 2-7 SAFE GENERATED ALWAYS ALWAYS
Properties
IFF inputs (0045,10C4) LO 6 SAFE GENERATED ALWAYS ALWAYS
Sigmoid (0045,10C7) DS 2 SAFE GENERATED CONDITIONAL ALWAYS If VOI LUT
Windowing Function
parameters (0028,1056) is
LINEAR
A.4.4 Digital Mammography X-Ray Image IOD Specific Values and Code Sets
Table A-17 lists Coded Values referenced from the "Value" column of the tables above.
Table A-58 Values and Code Sets for Digital Mammography X-Ray Image IOD
Attribute Name Tag Value/Code Condition Comments
View Code Sequence (0054,0220) (R-10224, SRT, “medio-lateral”) ML view
(R-10226, SRT, “medio-lateral oblique”) MLO view
(R-10228, SRT, “latero-medial”) LM view
(R-10230, SRT, “latero-medial oblique”) LMO view
(R-10242, SRT, “cranio-caudal”) CC view
(R-10244, SRT, “caudo-cranial (from below) ”) FB view
(R-102D0, SRT, “superolateral to inferomedial oblique”) SIO view
(R-40AAA, SRT, “inferomedial to superolateral oblique”) ISO view
(R-1024A, SRT, “cranio-caudal exaggerated laterally”) XCCL view
(R-1024B, SRT, “cranio-caudal exaggerated medially”) XCCM view
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Content Creator's Name (0070,0084) GENERATED ALWAYS ALWAYS
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>>Graphic Data (0070,0022) GENERATED ALWAYS ALWAYS
>>Graphic Type (0070,0023) GENERATED ALWAYS ALWAYS POLYLINE; ELLIPSE
>>Graphic Filled (0070,0024) FIXED CONDITIONAL CONDITIONAL N
A.5.2 Grayscale Softcopy Presentation State IOD Specific Functional Group Macros – N/A
N/A
A.5.3 Grayscale Softcopy Presentation State IOD Specific Private Modules – N/A
N/A
A.5.4 Grayscale Softcopy Presentation State IOD Specific Values and Code Sets – N/A
N/A
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
Referenced Performed (0008,1111) GENERATED ALWAYS CONDITIONAL Zero or one item may If a PPS manager has
Procedure Step Sequence be present been declared.
>Referenced SOP Class UID (0008,1150) FIXED ALWAYS ALWAYS 1.2.840.10008.3.1.2.3.3
>Referenced SOP Instance (0008,1155) GENERATED ALWAYS ALWAYS
UID
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>Requested Procedure (0032,1060) MWL ALWAYS CONDITIONAL If received from RIS
Description
>Requested Procedure Code (0032,1064) MWL ALWAYS CONDITIONAL Zero or one item may If received from RIS
Sequence be present
>>Code Value (0008,0100) MWL ALWAYS ALWAYS
>>Coding Scheme (0008,0102) MWL ALWAYS ALWAYS
Designator
>>Code Meaning (0008,0104) MWL ALWAYS ALWAYS
Performed Procedure Code (0040,A372) EMPTY ALWAYS EMPTY
Sequence
Current Requested (0040,A375) GENERATED ALWAYS ALWAYS One or more items may References to all the
Procedure Evidence be present acquired and stored x-ray
Sequence images of the study.
>Study Instance UID (0020,000D) SRC_INSTANCE ALWAYS ALWAYS
>Referenced Series (0008,1115) SRC_INSTANCE ALWAYS ALWAYS One or more items may
Sequence be present
>>Series Instance UID (0020,000E) SRC_INSTANCE ALWAYS ALWAYS
>>Referenced SOP (0008,1199) GENERATED ALWAYS ALWAYS One or more items may
Sequence be present
>>>Referenced SOP Class (0008,1150) SRC_INSTANCE ALWAYS ALWAYS
UID
>>>Referenced SOP (0008,1155) SRC_INSTANCE ALWAYS ALWAYS
Instance UID
Presence of Presence of
Attribute Name Tag Source Value Conditions Comments
Attribute Value
>Code Meaning (0008,0104) FIXED ALWAYS ALWAYS X-Ray Radiation Dose
Report
Continuity Of Content (0040,A050) FIXED ALWAYS ALWAYS SEPARATE
Content Template Sequence (0040,A504) GENERATED ALWAYS ALWAYS See Annex B for
encoding on supported
TIDs
A.6.2 X-Ray Radiation Dose SR IOD Specific Functional Group Macros – N/A
N/A
A.6.4 X-Ray Radiation Dose SR IOD Specific Values and Code Sets – N/A
N/A
Throughout the tables listed in Annex B the following codes are used for the "Source" and "Presence of Content Item" columns.
• QUERY: The Value is determined by performing a query of any of the supported Query/Retrieve Services.
• ALWAYS: the module, functional group macro, Attributes or Value is always present.
• CONDITIONAL: the presence of the module, functional group macro, Attributes or Value is dependent on a condition. The condition must be listed in the "Comments" column.
• SRC_COPY: The presence of the Attributes and Values depends on the availability of these in the source instances, which are used for copying this information.
NL Rel with Parent VT Concept Name Source Presence of Values TID Comments
Content Item
Data Available")
> CONTAINS CODE (113944, DCM, "X-Ray FIXED ALWAYS (R-0038D, SRT, “Yes”) 10001
Mechanical Data Available")
> CONTAINS CONTAINER (113702, DCM, "Accumulated GENERATED ALWAYS 10002
X-Ray Dose Data")
>> HAS CONCEPT MOD CODE (113764, DCM, "Acquisition FIXED ALWAYS (113622, DCM, "Single Plane") 10002
Plane")
>> CONTAINS NUM (111637, DCM, "Accumulated GENERATED ALWAYS UNITS=(mGy, UCUM, "mGy") 10005
Average Glandular Dose") Value=left breast dose
>>> HAS CONCEPT MOD CODE (272741003, SCT, FIXED ALWAYS (T-04030, SRT, ”Left Breast”) 10005
"Laterality")
>> CONTAINS NUM (111637, DCM, "Accumulated GENERATED ALWAYS UNITS=(mGy, UCUM, "mGy") 10005
Average Glandular Dose") Value=right breast dose
>>> HAS CONCEPT MOD CODE (272741003, SCT, FIXED ALWAYS (T-04020, SRT, ”Right Breast”) 10005
"Laterality")
> CONTAINS CONTAINER (113706, DCM, "Irradiation GENERATED ALWAYS 10003 Repeated for each
Event X-Ray Data") irradiation event
>> HAS CONCEPT MOD CODE (113764, DCM, "Acquisition FIXED ALWAYS (113622, DCM, “Single Plane”) 10003
Plane")
>> CONTAINS UIDREF (113769, DCM, "Irradiation GENERATED ALWAYS 10003
Event UID")
>> CONTAINS DATETIME (111526, DCM, "DateTime GENERATED ALWAYS 10003
Started")
>> CONTAINS CODE (113721, DCM, "Irradiation GENERATED ALWAYS (113611, DCM, “Stationary 10003
Event Type") Acquisition”); (113613, DCM,
“Rotational Acquisition”)
>> CONTAINS TEXT (125203, DCM, "Acquisition GENERATED ALWAYS ROUTINE; 3D_ROUTINE; CESM; 10003
Protocol") 3D_ROUTINE+2D_ROUTINE;
STEREO; 3D_BIOPSY; SAMPLE;
NL Rel with Parent VT Concept Name Source Presence of Values TID Comments
Content Item
CESM_BIOPSY
>> CONTAINS CODE (T-D0005, SRT, "Anatomical FIXED ALWAYS (T-04000, SRT, “Breast”) 10003* Standard Extended
structure")
>>> HAS CONCEPT MOD CODE (G-C171, SRT, "Laterality") GENERATED ALWAYS (G-A100, SRT, “Right”); 10003* Standard Extended
(G-A101, SRT, “Left”);
(G-A102, SRT, “Right and left")
>> CONTAINS CODE (111031, DCM, "Image View") GENERATED ALWAYS See Table B-2 10003
>>> HAS CONCEPT MOD CODE (111032, DCM, "Image View GENERATED CONDITIONAL See Table B-3 10003 If view modifier
Modifier") applied.
>> CONTAINS CODE (123014, DCM, "Target FIXED ALWAYS (T-04000, SRT, “Breast”) 10003
Region")
>> CONTAINS NUM (111634, DCM, "Half Value GENERATED ALWAYS UNITS=(mm, UCUM, "mm") 10003
Layer")
>> CONTAINS NUM (111636, DCM, "Entrance GENERATED ALWAYS UNITS=(mGy, UCUM, "mGy") 10003
Exposure at RP")
>> CONTAINS CODE (113780, DCM, "Reference FIXED ALWAYS (13964, DCM, “At Surface of 10003
Point Definition") Patient”)
>> CONTAINS IMAGE (113795, DCM, "Acquired SRC_INSTANCE ALWAYS Reference (SOP Class and SOP 10003A Row repeated for
Image") instance UID) to the images part of each image to be
this irradiation event. referenced.
>> CONTAINS NUM (111631, DCM, "Average GENERATED ALWAYS UNITS=(mGy, UCUM, "mGy") 10003B
Glandular Dose")
>> CONTAINS NUM (113768, DCM, "Number of GENERATED ALWAYS UNITS=(1, UCUM, "no units") 10003B
Pulses")
>> CONTAINS NUM (113793, DCM, "Pulse Width") GENERATED ALWAYS UNITS=(ms, UCUM, "ms") 10003B Row repeated for
each pulse in the
irradiation event.
>> CONTAINS NUM (113742, DCM, "Irradiation GENERATED ALWAYS UNITS=(s, UCUM, "s") 10003B
Duration")
NL Rel with Parent VT Concept Name Source Presence of Values TID Comments
Content Item
>> CONTAINS NUM (113733, DCM, "KVP") GENERATED ALWAYS UNITS=(kV, UCUM, "kV") 10003B Row repeated for
each pulse in the
irradiation event.
>> CONTAINS NUM (113734, DCM, "X-Ray Tube GENERATED ALWAYS UNITS=(mA, UCUM, "mA") 10003B Row repeated for
Current") each pulse in the
irradiation event.
>> CONTAINS NUM (113824, DCM, "Exposure GENERATED ALWAYS UNITS=(ms, UCUM, "ms") 10003B
Time")
>> CONTAINS NUM (113766, DCM, "Focal Spot GENERATED ALWAYS UNITS=(mm, UCUM, "mm") 10003B
Size")
>> CONTAINS CODE (111632, DCM, "Anode Target GENERATED ALWAYS (C-150F9, SRT, “Molybdenum or 10003B
Material") Molybdenum compound”);
(C-167F9, SRT, “Rhodium or
Rhodium compound”)
>> CONTAINS CONTAINER (113771, DCM, "X-Ray GENERATED ALWAYS 10003B This container and
Filters") its nested items are
repeated for each
pulse of the
irradiation event.
>>> CONTAINS CODE (113772, DCM, "X-Ray Filter FIXED ALWAYS (113650, DCM, “Strip filter”) 10003B
Type")
>>> CONTAINS CODE (113757, DCM, "X-Ray Filter GENERATED ALWAYS (C-150F9, SRT, “Molybdenum or 10003B
Material") Molybdenum compound”);
(C-137F9, SRT, “Silver or Silver
compound”);
(C-127F9, SRT, “Copper or Copper
compound”)
>> CONTAINS NUM (113790, DCM, "Collimated GENERATED ALWAYS UNITS=(m2, UCUM, "m2") 10003B
Field Area")
>> CONTAINS NUM (113788, DCM, "Collimated GENERATED ALWAYS UNITS=(mm, UCUM, "mm") 10003B
Field Height")
NL Rel with Parent VT Concept Name Source Presence of Values TID Comments
Content Item
>> CONTAINS NUM (113789, DCM, "Collimated GENERATED ALWAYS UNITS=(mm, UCUM, "mm") 10003B
Field Width")
>> CONTAINS CODE (111635, DCM, "X-Ray Grid") FIXED ALWAYS (111643, DCM, “Reciprocating grid”) 10003B
>> CONTAINS CODE (113956, DCM, "CR/DR FIXED ALWAYS (113954, DCM, “Upright Stand 10003C
Mechanical Configuration") Mount”)
>> CONTAINS NUM (112011, DCM, "Positioner GENERATED ALWAYS UNITS=(deg, UCUM, "deg") 10003C
Primary Angle")
>> CONTAINS NUM (111633, DCM, "Compression GENERATED ALWAYS UNITS=(mm, UCUM, "mm") 10003C
Thickness")
>> CONTAINS NUM (111647, DCM, "Compression GENERATED ALWAYS UNITS=(N, UCUM, "Newton") 10003C
Force")
>> CONTAINS NUM (113737, DCM, “Distance GENERATED ALWAYS UNITS=(mm, UCUM, "mm") 10003C
Source to Reference Point”)
> CONTAINS CODE (113854, DCM, "Source of FIXED ALWAYS (113856, DCM, “Automated Data 10001
Dose Information") Collection”)
Table B-2 Projection X-Ray Radiation Dose (TID 10001) – Image View
Coding Scheme Designator Code Value Code Meaning
SRT R-10224 medio-lateral
SRT R-10226 medio-lateral oblique
SRT R-10228 latero-medial
SRT R-10230 latero-medial oblique
SRT R-10242 cranio-caudal
SRT R-10244 caudo-cranial (from below)
SRT R-102D0 superolateral to inferomedial oblique
Table B-3 Projection X-Ray Radiation Dose (TID 10001) – Image View Modifier
Coding Scheme Designator Code Value Code Meaning
SRT R-102D2 Cleavage
SRT R-102D1 Axillary Tail
SRT R-102D3 Rolled Lateral
SRT R-102D4 Rolled Medial
SRT R-102CA Rolled Inferior
SRT R-102C9 Rolled Superior
SRT R-102D5 Implant Displaced
SRT R-102D6 Magnification
SRT R-102D7 Spot Compression
SRT R-102C2 Tangential
SRT R-40AB3 Nipple in profile
SRT P2-00161 Anterior compression
SRT R-40ABE Infra-mammary fold
SRT R-40AB2 Axillary tissue
C Security Details
This section provides additional details about security features that are formally described in Section 8.
D Mapping of Attributes
D.1 Mapping Between Modality Worklist Instances and MPPS
Table D-1 describes the mapping of Attributes between Modality Worklist Instances and MPPS messages.
• SCHEDULED: The image acquisition was scheduled at the RIS and procedure details have been communicated in the MWL query)
• UNSCHEDULED: The image acquisition was performed without Modality Worklist information
• APPEND: Instances acquired are added to an existing study after the initial procedure was finalized
In the "Value Source" columns, the following Values are used. The column cell may additionally contain an Attribute Tag if the value is copied
from a different Attribute.
The "Destination" columns either contain TOP, if the Attribute is added to the top level Data Set of the Instance, or contain the Attribute Tag of
the Sequence the Attribute will be added to. The "Comments" column can be used to provide additional information regarding the Values added
to the Instance or MPPS.
Table D-1 Mapping of Attributes from Modality Worklist to Instance and MPPS
Attribute Name in Tag Scenario Image MPPS Comments
Image/MPPS
Value Source Destination Value Source Destination
Performing (0008,1050) SCHEDULED MWL (0040,0006); TOP SRC_INSTANCE (0040,0340) (0040,0006) is
Physician’s Name USER Scheduled Performing
Physician Name
(0040,0340) is
UNSCHEDULED USER TOP SRC_INSTANCE (0040,0340)
Page 159 of 161
Senographe Pristina 9.x
GE HealthCare DICOM CONFORMANCE STATEMENT
DOC3075145 Rev. 1