006543341
006543341
INSUFFLATOR Highflow 45
2235
GERMANY USA UK
RICHARD WOLF GmbH RICHARD WOLF RICHARD WOLF UK Ltd.
75438 Knittlingen Medical Instruments Corporation Waterside Way
Pforzheimerstr. 32 353 Corporate Woods Parkway Wimbledon
Telephone: +49 70 43 35-0 Vernon Hills, Illinois 60061 SW17 0HB
Telefax: +49 70 43 35-300 Toll Free: 001 (800) 323 - 9653 Telephone: + 44 20 89 44 74 47
MANUFACTURER Phone: 001 (847) 913 - 1113 Telefax: + 44 20 89 44 13 11
[email protected] Fax: 001 (847) 913 - 1488 [email protected]
www.richard-wolf.com [email protected] www.richardwolf.uk.com
www.richardwolfusa.com
BELGIUM / NETHERLANDS FRANCE AUSTRIA
N.V. Endoscopie RICHARD WOLF France S.A.R.L. RICHARD WOLF Austria
RICHARD WOLF Belgium S.A. Rue Daniel Berger Ges.m.b.H.
Industriezone Drongen Z.A.C. La Neuvillette Wilhelminenstraße 93 a
Landegemstraat 6 F-51100 Reims A-1160 Vienna
9031 Gent Drongen Telephone: +33 3 26 87 02 89 Telephone: +43 14 05 51 51
Telephone: +32 92 80 81 00 Telefax: +33 3 26 87 60 33 Telefax: +43 14 05 51 51 45
Telefax: +32 92 82 92 16 [email protected] [email protected]
[email protected] www.richard-wolf.com
www.richard-wolf.be
GA-A 274 0
Contents
1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Indications and field of use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.4 Contraindications and side effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.4.1 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.4.2 Side effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.5 Combinations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.5.1 Instruments for INSUFFLATOR Highflow 45 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.5.2 Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5.3 Requirements for products / components of a combination . . . . . . . . . . . . . . . . . . . . . . . 7
1.6 Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.7 Connection diagram - in dialog mode (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2 Illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1 Front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2 Rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1 Language settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2 CO2 connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.1 Preparation for CO2 operation with gas bottle/cylinder . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.2 Preparation for CO2 operation using centralized gas supply . . . . . . . . . . . . . . . . . . . . . . 15
3.3 Connecting the insufflation tube set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.3.1 Reprocessable insufflation tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.3.2 Disposable insufflations tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4 Highflow insufflation tubes with heating wire . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4.1 Reprocessable highflow insufflation tube with heating wire . . . . . . . . . . . . . . . . . . . . . . . 17
3.4.2 Disposable highflow insufflation tube with heating wire . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.4.3 Disposable tube set for smoke gas evacuation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4 Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.1 Visual check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.2 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.2.1 Device function after switching on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.2.2 Checking the gas supply (status message) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.2.3 Manual checks of the control function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.2.4 Checking the pressure and flow control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5 Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.1 Operating principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.2 Controls and modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.2.1 Operation and display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.2.2 Dispaly of remaining CO2 gas supply in the status display . . . . . . . . . . . . . . . . . . . . . . . 22
5.2.3 Gas consumption in status display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.2.4 Indicator for gas heating (only 2235021, 2235031 / 2235621, 2235631) . . . . . . . . . . . . 23
5.2.5 Footswitch for smoke gas evacuation (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2.6 Video on-screen display with dialog function (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.3 Control sequence and menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
GA-A 274 I
5.3.1 Overview - control sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.3.2 Overview - menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.4 Menu levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.5 Menu control and touchscreen functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.6 Selection of main level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.7 Selection of device modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.8 Selection of Richard Wolf profiles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.8.1 Changing the standard profiles (default profiles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.9 Profiles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.10 Device parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.10.1 Adjusting the device parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.11 Display of actual and nominal values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.11.1 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.11.2 General service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.12 Operation of INSUFFLATOR Highflow 45 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.12.1 Important note for using the reprocessable highflow insufflation tube with heating wire 34
5.12.2 Notes and instruction for using smoke gas evacuation . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.12.3 Creating a pneumoperitoneum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.12.4 Highflow device mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.12.5 Measures after carrying out insufflation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.12.6 Possible alarms during the application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.12.7 Power failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.12.8 Taking out of service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.13 Overview of operator messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.13.1 Structure of operator messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.13.2 Message types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.13.3 Functionality of control elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.13.4 Operator's messages (green) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.13.5 LOW error (turqoise) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.13.6 MEDIUM error (yellow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.13.7 HIGH error (red) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
II GA-A 274
Contents
7.5.3 Leakage test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
8 Technical description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
8.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
8.1.1 Fault in the core system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
8.2 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
8.2.1 Application parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
8.2.2 Pressure and flow values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
8.2.3 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
8.2.4 Optional device variants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
8.3 Operating, storage, transport and shipping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
8.4 Spare parts and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.5 Replacing parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
8.5.1 Device fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
8.5.2 Disposal of product, packaging material and accessories . . . . . . . . . . . . . . . . . . . . . . . . 51
1.1 Symbols
Symbols Designation
Attention, Caution
Potential equalization
Hygiene filter
Gas heating
Alarm
Signal sound
Footswitch
Recycle the product separately. Do not discard together with other waste.
Lot designation
Manufacturer
Manufacturing date
Quantity
Use-by date
1 GA-A 274
Symbols Designation
Do not reuse.
Latex-free
-25_C
+ 50_C
Maximum temperature range for shipment, transport and storage
A Registered Trademark of a Recognized Testing Laboratory, confirm the compliance to the standard
of Medical Electrical Equipment CAN/CSA C22.2 No.60601-1 (c) and ANSI/AAMI ES60601-1 (us)
Identification in conformity with medical devices directive 93/42/EEC only valid if the product and/or
packaging is marked with this symbol. Products of category IIa and above, as well as sterile products
or products with measuring function of category I, are additionally identified with the code no. of the
notified body (0124).
GA-A 274 2
1.3 Indications and field of use
The INSUFFLATOR Highflow 45 (2235) is primarily used in diagnostic
and surgical laparoscopy but can also be used for TEM using the integ
rated smoke gas evacuation (2235011/2235031) and the "Continuous
Flow" (2235104) software module. For TEM applications please follow
the latest instruction manual for the TEM system.
On the basis of the patient’s general condition, the user in charge must
decide whether the planned use is possible or not. For further notes and
instructions please refer to the latest medical literature.
CAUTION!
In case of therapeutic applications an adequate backup device with the
same capabilities must be available should the device fail.
CAUTION!
It is not allowed to use mobile phones or similar devices while the INSUF
FLATOR Highflow 45 is used.
. NOTE!
We recommend reading relevant literature for the planned application.
1.4.1 Contraindications
The use of this device is contraindicated whenever laparoscopy is con
traindicated.
WARNING!
The device has a highflow rate.
The device is not suitable for hysteroscopy.
This device must not be used for distending the cavum uteri.
Contraindications resulting from the patient's general condition and de
scribed in the relevant literature must be observed.
Absolute contraindications:
D Hypovolemic shock represents an absolute contraindication with the
exception of laparoscopic surgeons who have adequate experience in
procedures for blood extraction from the abdomen and for finding and
stopping a hemorrhage site and can execute them safely and so can
achieve hemostasis quicker than by laparotomy.
D In the case of severe inflation of the intestine during intestinal obstruc
tion, increased possibility of adhesions in the pelvis and in the abdomen
as well as in the case of an elevated diaphragm, the required insuf
flation is contraindicated and the space available for a visualization is
severely limited and the probability of injuring the intestine is severely
increased.
D A further contraindication is the existence of a large mass in the pelvis
or in the area of the pelvis and abdomen (which in general extends to
the umbilicus or above). There is an increased risk of injuring this mass,
and there is a high probability that endoscopic visualization is poor and
that the extension and the nature of the mass cannot be determined. A
mass of this size would probably have to be removed by a laparotomy.
3 GA-A 274
D A type 4 cardiac decompensation also represents a contraindication as
the insufflation and its positioning could cause irreversible arrhythmia to
the heart which is filled with congested blood and cause cardiac failure.
Relative contraindications:
D Septic peritonitis
D Pelvic abscess
D Multiple interventions in the abdomen
D Diaphragmatic hernia
D Obesity
D Pregnancy
GA-A 274 4
1.4.2 Side effects
Warnings:
Metabolic acidosis and the resulting cardiac arrhythmia. Intraabdominal
pressures of more than 20 mmHg (mercury column) for an extended
period of time must be avoided.
This may lead to the following complications:
D Reduced breathing with impaired diaphragm excursion
D Reduced venous backflow
D Reduced cardiac output
D Acidosis
Excessive absorption of CO2 is either due to excessive flow or excessive
pressure. The abdomen can be sufficiently distended with pressures
between 15 - 20 mmHg (mercury column). Hardly ever abdominal pres
sures of more than 20 mmHg are required. At these values only limited
absorption should occur. Pressures of more than 20 mmHg (mercury
column) are actually never necessary and will lead to increased and
faster absorption. Adequate artificial respiration will help to avoid prob
lems with regard to CO2.
Z Idiosyncratic reactions. In patients suffering from sickle-cell disease or
pulmonary insufficience the use of these devices can lead to an in
creased risk of a metabolic imbalance due to excessive CO2 absorp
tion.
Z Hypothermia. Highflow rates bear the potential risk of hypothermia.
Precautions:
It is strongly recommended using a hydrophobic bacteria filter to prevent
the transmission of microorganisms between patients.
5 GA-A 274
1.5 Combinations
. IMPORTANT!
In addition to this instruction manual, follow the manuals for the products
used in combination with this product.
WARNING!
Danger of life-threatening gas or air embolism.
Purge the connection tubes/hoses with a sufficient amount of CO2 before
use.
The use of devices using additional gaseous media in combination with
the INSUFFLATOR Highflow 45 is exclusively within the responsibility of
the user.
WARNING!
Danger of intrabdominal excess pressure when using a second gas
source.
When using argon plasma coagulators an additional visual or manual
pressure monitoring device of the user is mandatory because the
pressure monitoring function / pressure relief of the INSUFFLATOR High
flow 45 can be inactive as a result of a kinked insufflation tube or a closed
instrument stopcock.
The preselected gas flow rates of the argon plasma coagulator should not
exceed 2 l/min. Activate the argon plasma coagulator only briefly.
GA-A 274 6
1.5.2 Potential equalization
The potential equalization cable establishes a direct connection between a medical electrical de
vice and an equipotential bonding rail.
It serves to equalize differences in potential between enclosures of electrical equipment and firmly
installed conductive parts in the patient environment.
1.5m
1.5m
1.5m The general requirements depend on whether the products /
components are inside or outside the patient environment.
MP MP
- - Verification of the total patient leakage current
~ ~
MP MP
~ **
MP NMP
**
~ ~
MP NMP
- Verification of leakage currents
a) additional protective earth connection
**
~ - (consult the corresponding manufacturer),
MP or
NMP b) additional isolating transformer for medical applications **
~
MP NMP
~
MP MP / NMP
**
~
7 GA-A 274
. IMPORTANT!
The persons combining products to form a system are responsible for not impairing the system's compliance with
performance and safety requirements, and that the technical data and the intended use are adequately fulfilled.
Possible electromagnetic or other interference that may occur between the product and other products can cause
faults or malfunctions.
When selecting the system components, make sure that they meet the necessary requirements of the medical envi
ronment they are used in, in particular IEC/ EN 60601-1 (3. Edition IEC/EN 60601-1, section 16). In case of doubt
contact the manufacturer(s) of the system components.
Do not touch connectors for electrical connections between various components (such as signal input connectors
and signal output connectors for video signals, data exchange, controls etc.) and the patient at the same time.
GA-A 274 8
Guidelines and manufacturer's declaration - Electromagnetic immunity for products that are not life-supporting
The product is intended for use in the environment specified below. The user must assure that the product is used in such an environment.
Immunity tests IEC 60601 test level Compliance Electromagnetic environment - Guidelines
Portable and mobile RF communications equipment should
be used no closer to any part of the product, including ca
bles, than the recommended separation distance calculated
from equation applicable to the frequency of the transmitter.
REMARKS:
At 80 MHz and 800 MHz the higher frequency range applies. These guidelines may not apply in all situations. The propagation of elec
tromagnetic waves is affected by absorption and reflexion from buildings, objects and people.
1 = The field strength of stationary transmitters (e.g. base station for radio phone, earth to earth radio stations,, amateur radio stations, radio
and television transmitters) cannot be exactly predicted in theory. To assess the EMC environment due to fixed transmitters an electromagnetic
site survey should be conducted. If the measured field strength in the location in which the product is used exceeds the applicable compliance
level above, the product should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be required, such as reorienting or relocating the product.
2 = Over the frequency range between 150 kHz and 80 MHz the field strength should be below 3 V/m.
Recommended separation distances between portable and mobile HF telecommunication devices and devices which
are not life-supporting
The product is intended for use in an electromagnetic environment where HF disturbances are controlled.
The user can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile HF telecommunica
tions equipment and the product.
Separation distance as a function of transmitter frequency (m)
Rated nominal output power of the
transmitter (W) 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
d = 1.2 p P d = 1.2 p P d = 2.3 p P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a nominal output power not listed in the table above, the recommended separation distance (d) in meters (m) can be
determined using the applicable equation (observe frequency). P = nominal power of the transmitter in Watts (W).
REMARKS:
At 80 MHz and 800 MHz the higher frequency range applies. These guidelines may not apply in all situations. The propagation of elec
tromagnetic waves is affected by absorption and reflexion from buildings, objects and people.
9 GA-A 274
1.7 Connection diagram - in dialog mode (option)
1.1
1 2
1.1
Legend
Direct connection
1 CAN BUS connection cable (option)
1.1 CAN BUS termination (option)
2 Video cable to meet the signal requirements of the camera controller used
. IMPORTANT!
The optional dialog function is only possible is only possible with enabled software module 3 of INSUF
FLATOR Highflow 45.
Connection of the CAN BUS connection cable (1) only with interactive camera controller.
If the CAN BUS connection cable is connected, the free CAN BUS connectors of INSUFFLATOR Highflow 45 and
the camera controller must be terminated with CAN BUS terminating resistors (1.1).
GA-A 274 10
2 Illustration
5 4 3 2 1.1 1
Legend
1 Power ON/OFF switch
1.1 Power ON/OFF LED
2 Touchscreen
3 Connector for disposable highflow insufflation tube with heating wire
' only 2235021, 2235031 / 2235621, 2235631
4 Connector for reusable highflow insufflation tube with heating wire
' only 2235021, 2235031 / 2235621, 2235631
5 Connector for insufflation tube
(type CF applied part)
6 Connector for smoke gas evacuation
(type CF applied part)
' only 2235011, 2235031 / 2235611, 2235631
11 GA-A 274
2.2 Rear view
13 14
12
15
11 10 8 7
16
Legend
7 Potential equalization connector
8 Connector for footswitch for smoke gas evacuation
' only 2235011, 2235031 / 2235611, 2235631
9 Service interface
10 Silencer
' only 2235011, 2235031 / 2235611, 2235631
11 Power input connector with fuse holder
12 Fuse plate
13 Device connector for gas supply
14 CAN BUS interface
15 Mounting points for bottle holding bracket
16 Identification plate
GA-A 274 12
3 Setup
WARNING!
The device is not protected against explosions.
Explosion hazard.
Do not operate this device in areas where there is a danger of explosion.
WARNING!
Danger if a power supply without protective earth is used.
Danger of electric shock!
Connect the device only to a power supply with protective earth connec
tor.
. NOTE!
The line / mains voltage must correspond with the voltage indicated on
the identification plate. Connect the device only via the supplied power
cable or a power cable with the same specifications.
Do not block any ventilation slots!
CAUTION!
Danger of faults and malfunctions.
To guarantee the safety of the user, patient and others, use only acces
sories and spare parts as specified by the manufacturer of this product.
Other accessories or spare parts can cause the emission of increased
electromagnetic radiation or reduced immunity against interference.
. IMPORTANT!
Medical electrical equipment is subject to special precautions with regard
to electromagnetic compatibility (EMC).
Make sure you follow the notes on EMC for installation and operation.
Medical electrical devices can be influenced by mobile HF communication
devices.
If it is necessary to stack the devices or place them next to each other
and HF interference is observed, make sure that the devices are ope
rated as specified in the intended use.
WARNING!
Use the device only with the insufflation tubes and filters specified by the
manufacturer. Sterilize the insufflation tube before each use.
Danger of infection due to unsterile accessories and possible malfunction
(loss of gas pressure or gas flow) when using spare parts supplied by
other manufacturers or in the case of wrong assembly.
CAUTION!
Disposable items have been designed for one single use only and are
sterilely packed. Do not use disposable items if:
' the sterile packaging is damaged
' the use-by-date has been exceeded
. NOTE!
For use, place the device on a level surface.
. NOTE!
Remove the protective foil from the touchscreen before the first use.
13 GA-A 274
3.1 Language settings
The language is selected in the Device parameters menu (see section 5).
When switching on the device for the first time (setup) as well as during a
new start after having reset the device to "factory settings", the required
language must be selected.
' Language selection for the “Standard profile”.
' For "User-specific operator profile" it is possible to define a separate
language.
Fig. 1
If a wrong language was selected and it is impossible to select the
language selection menu in this language, proceed as follows:
Z Switch off the device and wait for at least 60 seconds.
Z Switch on the device and press the device name on the touchscreen
until the language selection menu is displayed.
Z Let go of the button and select the required language.
' Language selection for the “Standard profile”.
WARNING!
Make sure that no liquid CO2 enters the INSUFFLATOR Highflow 45 as a
result of incorrect positioning of the gas bottle/cylinder!
This will endanger the patient as a result of possible malfunctions or fail
ure of the INSUFFLATOR Highflow 45 during the surgical intervention.
Operate the gas bottle/cylinder only in vertical position with the valve on
top.
. IMPORTANT!
Observe the manual for the pressure reducing valve.
18
Z Use mounting screws (15) to attach the bottle holding bracket (18) to
the device if required
' Optionally, large gas bottles/cylinders can be fastened to a device
cart.
15
. IMPORTANT!
When using gas bottles with more than 2 l of volume, use a separate
holding bracket (do not attach to the INSUFFLATOR Highflow 45) and
secure the gas bottle/cylinder against falling over.
GA-A 274 14
Z Place the gas bottle (max. 2 l of volume) into the bottle holding bracket
and secure with the belt (18.1).
18.1
20.1 19.1
Z Connect the pressure reducing valve (20) to the gas bottle (19) and
19 tighten the union nut (20.1) with your hand.
20
Z Connect the CO2 pressure tube (21) to the pressure reducing valve
21.1 (20) and tighten the union nut (21.1) with your hand.
Z Open the bottle valve (19.1) by turning counterclockwise.
21
. IMPORTANT!
To connect and disconnect the insufflation tube always hold it by the con
nectors.
15 GA-A 274
3.3.1 Reprocessable insufflation tubes
D 8170.101: for Veress cannula/RIWO-ART trocar sleeve, dia. 10 mm
(B1) + further models with luer lock fitting
D 8170.232: for high-flow trocar sleeves, dia. 10 mm (B1, B2)
Cut off the luer lock fitting on the insufflation tube and
connect the end of the tube onto the connecting piece (x).
B2
(X)
B1
B1
B2 B1
B1 B1
B2 B2
Cut off the luer lock fitting on the insufflation tube and connect
the end of the tube onto the connecting piece (x).
(X)
B B B
GA-A 274 16
3.4 Highflow insufflation tubes with heating wire
Optionally, a reprocessable high-flow insufflation tube with heating wire
as well as a disposable insufflaction tube with heating wire and integrated
hygiene filter can be connected to the device.
The plugs are encoded accordingly and can only be connected to the re
spective socket provided on the device.
For gas heating mode see section 5.
17 GA-A 274
3.4.2 Disposable highflow insufflation tube with heating wire
. IMPORTANT!
The disposable tube set for smoke gas evacuation must only be used for
evacuating smoke gas and not for suctioning out liquids.
GA-A 274 18
4 Checks
. IMPORTANT!
Run through the checks before and after each use.
Do not use the products if they are damaged or incomplete or have loose
parts.
Return damaged products together with any loose parts for repair.
Do not attempt to do any repairs yourself.
19 GA-A 274
4.2.2 Checking the gas supply (status message)
Z The CO2 gas supply is displayed in the status display.
' Check that the gas bottle is open and there is sufficient CO2 supply.
' For displaying the fill levels on the status display see section 5.2.2.
Pressure preselection:
Z Change the nominal pressure using the pressure (+) and (-) buttons.
' When insufflation is on, the preselected pressure (nominal value) is
displayed for approx. 2 s.
GA-A 274 20
Highflow:
Z Nominal pressure to be preselected: 12 mmHg
Flow rate to be preselected: 10 l/min
Z Select highflow.
' First switch on insufflation, then select highflow.
' The current flow rate must agree with the preselected flow rate of 10
l/min.
' Permissible tolerance: ¦ 2.0 l/min
Z Close the insufflation tube with your thumb.
' As soon as the current pressure exceeds the preselected pressure,
the device must release pressure.
Z Open the insufflation tube.
Z Switch back from highflow to primary insufflation.
' After 1.5 s at the latest, a flow rate of 1 l/min ¦0.5 l/min must be dis
played.
Z Switch off insufflation.
21 GA-A 274
5 Use
Display Description
GA-A 274 22
5.2.4 Indicator for gas heating (only 2235021, 2235031 / 2235621, 2235631)
Display Description
23 GA-A 274
5.3 Control sequence and menu structure
GA-A 274 24
5.3.2 Overview - menu structure
User Line/
mains power
ON
Selftest
Mode selection
Standard profile
Profile selection (option)
User-specific profile
Menu level
Select
Device parameters
Display Settings
Audio Settings
Language Settings
Service
25 GA-A 274
5.4 Menu levels
The device menu is subdivided into 4 areas:
D Main level for user
D Menu level for user
D General service area for user
D Device service for authorized persons - with password only
. IMPORTANT!
When controlled via the touchscreen, it is sufficient to touch the monitor
surface slightly.
Do not touch the surface of the touchscreen monitor with sharp,
pointed or contaminated objects as this will lead to reduced image
quality or a damaged surface.
. NOTE!
The illustrations of the user interfaces are only exemplary and vary
depending on the use.
XXXX
XXXX
XXXXXXXX Touchscreen:
The touchscreen monitor is used for information output and the input or
selection of parameters.
XXXXXXXXX - XXXX X
XXXXX
Touchscreen areas:
Area for displays and settings:
When the displayed function is active, the displayed numerical value is
shown in red. The corresponding numerical value is adjusted using the
(+) and (-) buttons or the slide control.
XXXXXXX XXX Status area:
0 100 For approx. 10 s the corresponding status message appears "!" and an
acoustic signal (option) is sounded (e.g. when the lowest or highest
permissible value of the adjustment is reached).
Status area Device parameters:
Indicator for additional parameters, e.g. gas consumption in case of an
Line for main functions insufflator.
Line for main functions:
Calling up the menus, switching on and off device functions, save, delete
etc.
Navigation:
Navigation buttons A = up/down between individual menu items (single
A B steps).
Navigation buttons B = up/down by page or paragraph.
Please note:
The navigation buttons are only active if further menu items can be
called up which are not visible on the first menu page.
GA-A 274 26
Control elements:
Menu Not selected: Button is dark, letters are white (button not depressed).
20
Parameter
designation Unit
0 100 Display and parameter setting:
The parameters are displayed as bars and as numbers.
With the (+) and (-) buttons the value can be increased or decreased
and, optionally changed using the slide control (nominal value).
When you keep the (+) / (-) button depressed, the value changes incre
mentally.
Back
XXXX
The selection on the main level depends on the device version of the
optional software modules and on the device mode. See table in section
8.2.3.
.
27 GA-A 274
5.7 Selection of device modes
In the case of minilaparotomy – primary insufflation via Veress cannula is not applica
Highflow
ble.
Special mode for newborns and babies. Precise pressure and generation of flow in
Baby mode (option)
steps of 0.1 l.
Continuous flow mode with a maximum flow of 8 l/min (only in conjunction with smoke
TEM (option)
gas evacuation).
Continuous flow (option) Continuous flow mode for laparoscopy/thoracic surgery at a maximum flow of 8 l/min.
Profiles Description
The preconfigured Richard Wolf profiles (e.g. urology, gynecology) can be changed or
Richard Wolf profiles deleted as required. After a reset to the "factory setting", the stored profiles are
reinstated.
Resetting the user-specific During a reset to the factory settings, the custom profiles are deleted and the Richard
profiles Wolf profiles are reinstated.
GA-A 274 28
5.9 Profiles
Select
Standardsetting
Standard profile Standardsetting
Standard profile Standard setting
Standard profile
Save profile Delete profile Back Save profile Delete profile Back Save profile Delete profile Back
Q F E C T Z U I O P Ü Q F E C T Z U I O P Ü
A S D F G H Y K L Ö Ä A S D F G H Y K L Ö Ä
Y X C V B N M ,< . - Y X C V B N M ,< . -
1. Dr. Bones
2. TEP hernia
3. New profile
29 GA-A 274
Select
Overview - Changing the settings in a profile
Active
Example: Change display brightness Selection possible
Not active
Select Selection not possible
profile
3. Dr. Muster
Select Display
Device-specific functions Audio
Device parameters Device info
Language
Back Back
Adjust
Display
Adjust brightness
min.
80 max.
brightness
Automatic brightness control
Back
Please note:
After changing a parameter, the setting must be saved
Display
in the "Profile" menu.
Audio
A red profile name and the red "save profile" button
indicate that the changed setting has not yet been Device info
saved. Language
Any settings that haven't been saved are lost when
changing between the profiles or in case of a restart. Dr. Bones
Back
3. Dr. Muster
Dr. Bones
Accept setting
GA-A 274 30
5.10 Device parameters
Display
Volume
XXX
min. 50 max.
Volume adjustment:
In the Audio submenu, the volume of the key tones, acoustic operator's
notes and alarms can be adjusted.
Acoustic operator's notes
To adjust the volume of the audio signals, use the (+) and (-) buttons or
Button sound the slide control.
With the Key tone button or the acoustic operator's notes button acoustic
signals for the operation of the keys or operator's notes can be switched
Back off and on. The alarms cannot be switched off.
Back
31 GA-A 274
British English
US English
Language selection:
German
The "Language" submenu displays the selected language. Here you may
Français select the required language.
The navigation buttons can be used to select other languages.
Back
OK Cancel
Factory setting:
Do you want to reset the device to factory
setting? The Factory settings submenu resets all settings to the delivery state.
Confirming the factory settings with OK resets all settings and deletes all
Profiles (custom settings) and automatically the device is restarted.
After the restart the language selection is displayed. This language
setting defines the language of the "Standard profile".
OK Cancel
GA-A 274 32
5.11.1 Service
. NOTE!
All menu screens in the service area are exclusively in English.
General service
General service:
Service area for the user.
Device service (protected)
Device-related display, resets and settings.
Please note:
To leave the service area the device must be switched off and then on
! Message text ! again.
Device service (protected):
Back
Only for authorized persons with a password.
33 GA-A 274
5.12 Operation of INSUFFLATOR Highflow 45
CAUTION!
Danger of infection due to unsterile accessories.
Sterilize the reprocessable containers and accessories as described in
section 6.
Discard used disposable items after each use as required by the laws
and regulations valid in your country.
Follow the sterilization rules for the different products!
CAUTION!
The products have only limited strength!
Applying excessive force will lead to damage, impair the function and
endanger the patient.
Immediately before and after each use, check the products for damage,
loose parts and completeness.
Make sure that no missing parts remain in the patient.
Do not use the products if they are damaged or incomplete or have loose
parts.
. IMPORTANT!
Before you start laparoscopy observe and carry out the measures de
scribed in section 3 "Setup" and section 4 "Checks".
. IMPORTANT!
Before each use, check the INSUFFLATOR Highflow 45 manually for
proper control functions.
5.12.1 Important note for using the reprocessable highflow insufflation tube with heating wire
The reprocessable highflow insufflation tube with heating wire heats the
insufflation gas to body temperature.
In the case of low gas flow rates (0.1 l/min - 1 l/min) the tube set temper
ature on the outside may exceed 50 °C!
The actual gas flow rate depends on the overpressure generated in the
abdomen. The preset maximum nominal gas flow is only reached as long
as no or only a very slight overpressure is measured in the abdomen.
Otherwise, the gas flow is of secondary importance relative to the pres
sure, i.e. it may be reduced to less than 1 l/min during the operation even
though higher settings were chosen.
To avoid possible injuries of the patient, user or others, make sure that
the following precautions are observed:
D Make sure that the tube set is not connected to the patient for more
than 1 minute.
D Avoid direct contact between the tube set and the patient.
D Before the operation, place the tube set in such a way that it either
does not touch the patient at all or that the contact surface is well insu
lated (several layers of sterile cloth or gauze between the tube set and
the patient).
GA-A 274 34
5.12.2 Notes and instruction for using smoke gas evacuation
. IMPORTANT!
The disposable tube set for smoke gas evacuation must only be used for
evacuating smoke gas and not for suctioning off liquids.
D Smoke gas evacuation can only be activated in the "Highflow" device
mode.
D Smoke gas evacuation is not available in Baby mode (option).
D The smoke gas evacuation pump is only switched on at pressures > 3
mmHg and is switched off automatically at pressures < 3 mmHg.
D Smoke gas evacuation can be activated and deactivated via the touch
screen using the Evacuation button or via the footswitch (option).
WARNING!
Danger of gas embolism.
The "Highflow" device mode should only be used under endoscopic view.
. NOTE!
If insufflation at a highflow rate is used over a prolonged period of time,
the CO2 gas supply may cool down and cause a transitional drop in the
gas pressure of the supply container. At the same time an operator's note
is displayed.
When insufflation is stopped, the original gas pressure will build up again
in the supply tank/cylinder.
35 GA-A 274
5.12.5 Measures after carrying out insufflation
. IMPORTANT!
Close the gas bottle/cylinder valve after the application. Any leaks at
bottle/cylinder valve or on the connection can cause the gas bottle/cy
linder to be completely emptied after some time.
Z Close the gas bottle/cylinder valve.
Z Remove and discard the hygiene filter.
Z Discard or reprocess, respectively, the insufflation tube.
GA-A 274 36
5.13 Overview of operator messages
1 alarm pulse
Informs the user about imminent dangers.
Turqoise LOW error Symbol does not
Error of low priority.
blink.
No button The message window is displayed until the conditions are fulfilled.
Service Once this button is displayed, you can only switch to the "Service" submenu.
The device can no longer be operated.
37 GA-A 274
5.13.4 Operator's messages (green)
Message
Message text Possible cause Corrective action
number
During the runtime a connection
xxx-xxx-005 Connection to core system established to the core system has been ' Status message
established
During the runtime the connection
xxx-xxx-006 Connection to core system disconnected to the core system was inter ' Status message
rupted
Language switchover by the core
xxx-xxx-007 Language switchover by core system successful ' Status message
system
xxx-xxx-008 Upper range limit reached ------- ' Status message
xxx-xxx-009 Lower range limit reached ------- ' Status message
Please wait, the device parameters are being
xxx-xxx-010 ------- ' Status message
reset!
xxx-xxx-011 Memory access, please wait ------- ' Status message
xxx-xxx-012 Please wait, profile is being deleted ------- ' Status message
xxx-xxx-013 Please wait, a new profile is generated ------- ' Status message
Maximum number of profile entries is reached
xxx-xxx-014 To generate a new profile, an existing profile must Maximum profile entries reached ' Carry out message
be deleted.
Invalid profile name
xxx-xxx-015 Please enter profile name again. The name must Incorrect profile name input ' Carry out message
contain at least 1 alphanumeric character.
The profile name entered already exists
xxx-xxx-016 Double profile name ' Carry out message
Please enter a different name
To exit the service menu, a device restart is requi Message when logging into the
xxx-xxx-018 ' Status message
red. service area
xxx-xxx-019 Wrong service password Password spelling error ' Status message
xxx-xxx-020 Please wait ------- ' Status message
xxx-xxx-021 Function cannot be activated ------- ' Status message
xxx-xxx-027 Please restart device A device restart is necessary ' Carry out message
' Open gas bottle/cylinder or
connect full gas bottle/cylinder
xxx-xxx-436 Insufficient gas supply Gas supply interrupted ' If this message occurs
repeatedly although the gas
xxx-xxx-437 Please check the gas supply CO2 pressure sensor defective
bottle/cylinder is full and the
valve is open, contact the ser
vice department
Message
Message text Possible cause Corrective action
number
' Confirm with OK button -
Profile data not correctly initialized
automatic device restart
xxx-xxx-017 Press "OK" to delete the profile data and set to EEPROM memory error
' If fault occurs again, contact
default values.
service department
' Allow the device to cool
down
xxx-xxx-028 Device temperature too high
Device temperature too high ' Carry out a device restart
xxx-xxx-029 Please allow the device to cool down
' If fault occurs again, contact
service department
Device temperature too low ' Wait until device temperature
Device temperature too low
xxx-xxx-030 (outside the permissible operating meets the required operating
Please allow the device to warm up conditions
conditions)
xxx-xxx-031 Device error - Temperature sensor defective Temperature sensor defective ' Contact service department.
GA-A 274 38
Message
Message text Possible cause Corrective action
number
Environment light sensor defective
Environment light sensor defec
xxx-xxx-033 Automatic brightness control of display is swit ' Contact service department.
tive
ched off.
xxx-xxx-034 ' Carry out a device restart
Device error - No connection to core system pos
to Electronics error ' If fault occurs again, contact
sible
xxx-xxx-036 service department
' Carry out recalibration on the
Faulty touchscreen calibration Saved EEPROM values of the service level (general service).
xxx-xxx-037 The default calibration is used, for recalibration touch calibration are not within
the calibration range ' If fault occurs again, contact
switch to service menu.
service department
xxx-xxx-038 ' Carry out a device restart
Device error - Faulty initialization, device is not
to Electronics error ' If fault occurs again, contact
ready
xxx-xxx-042 service department
Software error - Internal data
xxx-xxx-043 Device error - Data transmission ' Contact service department.
transmision
' Carry out a device restart
Internal operating voltages not
xxx-xxx-061 Device error - Device not ready ' If fault occurs again, contact
okay
service department
' Carry out a device restart
Device error - Faulty initialization, device is not Hardware and software versions
xxx-xxx-062 ' If fault occurs again, contact
ready are not compatible
service department
' Carry out a device restart
Device error - No connection to core system pos
xxx-xxx-063
sible
core processor update faulty ' If fault occurs again, contact
service department
' Carry out a device restart
The type number/model number
xxx-xxx-064 Device error - type number ' If fault occurs again, contact
saved in the device is faulty
service department
' Carry out a device restart
Device error - Initialization failed
xxx-xxx-066 Electronics error ' If fault occurs again, contact
The device is not ready
service department
Software error - Internal data
xxx-xxx-067 Device error - Data transmission ' Contact service department.
transmision
xxx-xxx-400 Device error - Hardware safety activated Device error ' Contact service department.
xxx-xxx-403
to Device error - Pneumatic component Electronics error ' Contact service department.
xxx-xxx-404
xxx-xxx-405 Device error - Pneumatic test Pneumatic component defective ' Contact service department.
' Carry out a device restart
xxx-xxx-406 Device error - Hardware incompatible Electronics error ' If fault occurs again, contact
service department
xxx-xxx-407 Device error - Smoke gas evacuation module Electronics error ' Contact service department.
xxx-xxx-408 Device error - Electronics Electronics error ' Contact service department.
xxx-xxx-409
Device error - Pneumatic component Pneumatic component defective ' Contact service department.
xxx-xxx-410
xxx-xxx-411
Device error - Electronics Electronics error ' Contact service department.
xxx-xxx-412
xxx-xxx-413 Device error - Gas heating module Electronics error ' Contact service department.
xxx-xxx-415 Device error - Electronics Electronics error ' Contact service department.
xxx-xxx-428 Device error - Gas heating module Electronics error ' Contact service department.
' When actuating the OK but
ton, the gas heating module is
Device error - Data transmission Software error - Internal data
xxx-xxx-433 deactivated
Gas heating module is deactivated transmision
' If fault occurs again, contact
service department
39 GA-A 274
5.13.6 MEDIUM error (yellow)
Message
Message text Possible cause Corrective action
number
xxx-xxx-032 Device error - Temperature sensor defective Temperature sensor defective ' Contact service department.
Software error - Internal data
xxx-xxx-044 Device error - Data transmission ' Contact service department.
transmision
Software error - Internal data
xxx-xxx-068 Device error - Data transmission ' Contact service department.
transmision
Message
Message text Possible cause Corrective action
number
xxx-xxx-401 Device error - Hardware safety activated Device error ' Contact service department.
Overpressure alarm -
Critical overpressure > 30 mmHg (mercury co ' Decrease intraabdominal
xxx-xxx-402 pressure > 30 mmHg (mercury
lumn) pressure
column) for more than 5 s
xxx-xxx-416 Device error - Electronics Electronics error ' Contact service department.
xxx-xxx-420 Device error - Pneumatic component Electronics error ' Contact service department.
xxx-xxx-423 Device error - Electronics Electronics error ' Contact service department.
xxx-xxx-426 Device error - Software Software error ' Contact service department.
GA-A 274 40
6 Operation within the core system
. IMPORTANT!
Should the core system fail, the device can still be controlled manually via
the controls on the device itself. For a description please refer to the in
struction manual of the respective device.
41 GA-A 274
6.4 Connection to the core system
Touchscreen
monitor
RIWO
CONTROL USB
core.control headset connector
CAN BUS
e.g.
INSUFFLATOR
Highflow 45
CAN BUS
e.g.
HIGHLIGHT X300
light projector
CAN BUS
e.g.
Camera
. IMPORTANT!
The device system must be operated via a "galvanically isolated transformer".
. IMPORTANT!
The last device in the CAN BUS chain must be terminated using the supplied terminating resistor.
GA-A 274 42
7 Reprocessing and maintenance
Clean the device with a soft cloth moistened with surface disinfectant,
alcohol or spirit.
Follow the disinfectant manufacturer's instructions!
. NOTE!
For a description of the surface cleaning and disinfecting agents for
devices and universal device carts / trolleys and utility carts approved by
Richard Wolf, please refer to manual GA-J020 "Reprocessing of
RICHARD WOLF Heat-stable Instruments".
. IMPORTANT!
Make sure that no humidity enters the device. Do not use any cleaning
agents, scouring agents or solvents on this device.
WARNING!
Do not reprocess disposable items!
The service life of products marked as disposable i.e. for single use only,
has been designed for only one use in or on a single patient.
If disposable items are reprocessed to be used again, a deterioration of
the product quality cannot be excluded, which will endanger the patient,
the user and others.
In this case, the manufacturer can no longer guarantee product safety
and performance.
43 GA-A 274
7.2.1 Cleaning
Z Rinse out the tubes with a cleaning gun and dry with compressed air.
7.2.2 Sterilization
Steam sterilization:
Z Steam sterilization using a fractional pre-vacuum procedure at 134 C
(272 F).
7.4 Maintenance
. IMPORTANT!
In the case of inquiries or in your correspondence please always indicate
the product number and the serial number listed on the identification
plate. Further documentation is available from the manufacturer on re
quest.
. IMPORTANT!
To avoid any incidents or damage caused by aging and wear it is neces
sary to service the product and the accessories at adequate intervals.
Depending on the frequency of use, but at least once a year, have an
expert check the functional and operational safety of the equipment.
. IMPORTANT!
The device must not be used if the specified values are not displayed or
functions are not fulfilled.
GA-A 274 44
7.5.2 Checking the pressure and flow control
Primary insufflation:
Z Preselect the following nominal pressure: 12 mmHg (mercury column)
Z Connect the hygiene filter and insufflation tube to the insufflation con
nector.
Z Start insufflation.
' The current flow rate must agree with the preselected flow rate of 1
l/min.
' Permissible tolerance: ¦ 0.5 l/min
Z Close the insufflation tube with your thumb.
' As soon as the current pressure exceeds the preselected pressure,
the device must release pressure.
Z Switch off insufflation and open the insufflation tube.
Highflow:
D Preselect the following nominal pressure: 12 mmHg (mercury column)
D Preselected the following flow rate: 10 l/min
Z Start insufflation, then switch over to Highflow.
' The current flow rate must agree with the preselected flow rate of 10
l/min.
' Permissible tolerance: ¦ 2.0 l/min
Z Close the insufflation tube with your thumb.
' As soon as the current pressure exceeds the preselected pressure,
the device must release pressure.
Z Open the insufflation tube.
Z Switch back from highflow to primary insufflation.
' After 1.5 s at the latest, a flow rate of 1 l/min ¦0.5 l/min must be dis
played.
Z Switch off insufflation.
45 GA-A 274
8 Technical description
8.1 Troubleshooting
. IMPORTANT!
If the faults cannot be corrected with the help of this table, please contact
our service department or return the device for repair.
'Do not attempt to do any repairs yourself!
Device is not functional Power switch is not on 'Actuate the power switch
Power cable is not connected 'Connect the power cable
Fuse in device is defective 'Replace the fuse
No line voltage/mains 'Check in-house power supply
No gas flow Insufflation not activated 'Switch on insufflation
Gas bottle/cylinder empty or not open 'Replace or open gas bottle/cylinder
Preselected pressure is exceeded As a result of manipulations of the
user, the abdominal pressure increa 'Remove cause
ses
Device defective 'Contact the service dept.
Other devices cause a pressure incre 'Remove cause
ase
Pressure won't build up Tube system is leaking 'Check tube and connectors
Flow preselection is to low 'Increase flow preselection
Instrument stopcock is closed 'Open instrument stopcock
Smoke gas evacuation won't start Pressure < 3 mmHg 'Generate pressure > 3 mmHg
Evacuation not actuated 'Actuate evacuation
'Open evacuation tube
'Check connections
'Replace CAN BUS cable
Device is not logged in the core system Data transfer error
'Set to the same system ID
'Contact the service dept.
GA-A 274 46
8.2 Technical data
Voltage Frequency Power Current Fuse
Product no. consumption rating
Vμ Hz VA A A
2235001
100-240 50 / 60 60 0.6 - 0.2 T 3.15 H
2235011
2235021 100-127 /
50 / 60 80 0.8 - 0.3 T 3.15 H
2235031 220-240
2235601 (USA)
100-240 50 / 60 60 0.6 - 0.2 T 3.15 H
2235611 (USA)
2235621 (USA) 100-127 /
50 / 60 80 0.8 - 0.3 T 3.15 H
2235631 (USA) 220-240
47 GA-A 274
8.2.2 Pressure and flow values
Device mode Pressure values min. / max. Flow values min. / max.
Primary insufflation (1l/min)
min.: 3 mmHg (mercury column)
Default min.: 2 l/min
max.: 25 mmHg (mercury column)
max.: 45 l/min
min.: 3 mmHg (mercury column) min: 2 l/min
Highflow
max.: 25 mmHg (mercury column) max.: 45 l/min
min.: 3 mmHg (mercury column) min.: 0,1 l/min
Baby mode (option)
max.: 20 mmHg (mercury column) max.: 2 l/min
Primary insufflation (1l/min)
min.: 3 mmHg (mercury column)
TEM (option) min.: 2 l/min
max.: 25 mmHg (mercury column)
max.: 8 l/min
Primary insufflation (1l/min)
min.: 3 mmHg (mercury column)
Continuous flow (option) min.: 2 l/min
max.: 25 mmHg (mercury column)
max.: 8 l/min
Primary insufflation (1l/min)
min.: 3 mmHg (mercury column)
Gynecology min.: 2 l/min
max.: 25 mmHg (mercury column)
max.: 45 l/min
Primary insufflation (1l/min)
min.: 3 mmHg (mercury column)
Urology min.: 2 l/min
max.: 25 mmHg (mercury column)
max.: 45 l/min
8.2.3 Interfaces
GA-A 274 48
8.2.4 Optional device variants
Device mode
Default
Device mode
Highflow
Device mode
1) 1) 1) 1)
Baby mode *
Device mode
----- 2) ----- 2)
TEM
Device mode
2) 2) 2) 2)
Continuous flow
. NOTE!
To prevent damage during transport or shipment of the products we recommend using the original
packaging material.
49 GA-A 274
8.4 Spare parts and accessories
ISO connector (ISO 5145) DIN connector (DIN EN 738-1) PIN Index connector (DIN EN ISO 407)
The products can be combined as required provided the relevant technical data and intended uses are observed. For
the total overview please refer to the latest catalog sheets and brochures or contact Richard Wolf or your representative.
GA-A 274 50
8.5 Replacing parts
CAUTION!
The specifications of the device fuses must correspond with the fuse ra
tings on the identification plate.
Use only the fuses specified in the spare parts list.
2 2 3
Z Switch off the device and disconnect the power cable from the wall
socket and from the power input connector of the device.
Z Unscrew the fuse holder [1] using a screwdriver.
Z Take the fuses [2] out of the fuse holder [1] and replace.
Z Reinsert the fuse holder and push until it clicks in place [3].
51 GA-A 274