0% found this document useful (0 votes)
24 views59 pages

Scope 58550 CC 2555 1693194169 - Bagson

BAGSON Calibration Lab Pvt Ltd is accredited under ISO/IEC 17025:2017, with a certificate valid from February 14, 2023, to February 13, 2025. The document details various calibration methods and measurement capabilities for different electrical parameters, including AC and DC current, voltage, resistance, and power factor. It includes specific measurement ranges and associated uncertainties for each calibration method employed.

Uploaded by

Anirban Das
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
24 views59 pages

Scope 58550 CC 2555 1693194169 - Bagson

BAGSON Calibration Lab Pvt Ltd is accredited under ISO/IEC 17025:2017, with a certificate valid from February 14, 2023, to February 13, 2025. The document details various calibration methods and measurement capabilities for different electrical parameters, including AC and DC current, voltage, resistance, and power factor. It includes specific measurement ranges and associated uncertainties for each calibration method employed.

Uploaded by

Anirban Das
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 59

BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ

Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 1 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Permanent Facility
ELECTRO-
Using 6 ½ DMM,
TECHNICAL-
Multiproduct
Alternating
1 AC Current @ 50 Hz Calibrator by 0.5 A to 3 A 0.29%
Current (< 1
Comparison/Direct
GHz)
Method
(Measure)
ELECTRO- Using 6 ½ DMM with
TECHNICAL- Shunt 30A/30mV,
Alternating Multiproduct
2 AC Current @ 50 Hz 10 A to 20 A 0.71 % to 0.72 %
Current (< 1 Calibrator by
GHz) Comparison/Direct
(Measure) Method
ELECTRO-
Using 6 ½ DMM,
TECHNICAL-
Multiproduct
Alternating
3 AC Current @ 50 Hz Calibrator by 100 mA to 0.5 A 0.65 % to 0.28 %
Current (< 1
Comparison/Direct
GHz)
Method
(Measure)
ELECTRO- Using 6 ½ DMM With
TECHNICAL- Shunt 30A/30mV ,
Alternating Multiproduct
4 AC Current @50 Hz 3 A to 10 A 0.29 % to 0.71 %
Current (< 1 Calibrator by
GHz) Comparison/Direct
(Measure) Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 2 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using DMM with HV
Alternating Ac High Voltage @
5 Probe by Direct 1 kV to 25 kV 0.1 kV to 1.7 kV
Current (< 1 50 Hz
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using DMM with HV
Alternating AC High Voltage @
6 Probe by Direct 1 kV to 6 kV 0.1 kV to 0.33 kV
Current (< 1 50 Hz
Method
GHz)
(Measure)
ELECTRO-
Using 6 ½ DMM ,
TECHNICAL-
Multiproduct
Alternating
7 AC Voltage @ 50 Hz Calibrator by 10 mV to 30 mV 0.55%
Current (< 1
Comparison/Direct
GHz)
Method
(Measure)
ELECTRO-
Using 6 ½ DMM,
TECHNICAL-
Multiproduct
Alternating
8 AC Voltage @ 50 Hz Calibrator by 100 mV to 750 V 0.20 % to 0.15 %
Current (< 1
Comparison/Direct
GHz)
Method
(Measure)
ELECTRO-
Using 6 ½ DMM ,
TECHNICAL-
Multiproduct
Alternating
9 AC Voltage @ 50 Hz Calibrator by 30 mV to 100 mV 0.55 % to 0.20 %
Current (< 1
Comparison/Direct
GHz)
Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 3 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (50 Hz -
10 Alternating Calibrator by direct 100 µA to 100 mA 0.27 % to 0.16 %
1kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (50 Hz -
11 Alternating Calibrator by Direct 100 mA to 1 A 0.16 % to 0.09 %
1kHz)
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Current (50 Hz -
12 Alternating Calibrator by direct 30 µA to 100 µA 0.54 % to 0.27 %
1kHz)
Current (< 1 method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multi Product
13 Alternating AC Current @ 50 Hz Calibrator by Direct 1 A to 10 A 0.09 % to 0.15 %
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
14 Alternating AC Current @ 50 Hz Calibrator by Direct 10 A to 20 A 0.15 % to 0.21 %
Current (< 1 Method
GHz) (Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator with 50
15 Alternating AC Current @ 50 Hz 50 A to 1000 A 0.8%
Turns coil by Direct
Current (< 1
Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 4 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO- AC Power @50


TECHNICAL- Hz,7.5 W-12kW (15 Using Multiproduct
16 Alternating V to 600V , 0.5 A to Calibrator by Direct 7.5 W to 12 kW 0.61%
Current (< 1 20 A)0.5 Lead & Lag, Method
GHz) (Source) UPF
ELECTRO-
TECHNICAL- Using Multiproduct
AC Voltage @ 50 Hz
17 Alternating Calibrator by Direct 1 V to 1000 V 0.04%
to 1 kHz
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
AC Voltage @ 50 Hz
18 Alternating Calibrator by Direct 10 mV to 1 V 0.11 % to 0.04 %
to 1 kHz
Current (< 1 Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Decade
Capacitance @ 1
19 Alternating Capacitance Box by 100 pF to 10 µF 2.27%
kHz
Current (< 1 Direct Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Std.
20 Alternating Inductance @ 1 kHz Inductance Box by 100 µH to 10 H 2.27%
Current (< 1 Direct Method
GHz) (Source)
ELECTRO-
TECHNICAL- Using Multiproduct
Power Factor @ 50
21 Alternating Calibrator by Direct 0.2 PF to 1 PF 0.006PF
Hz
Current (< 1 Method
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 5 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO- Using 6 ½ DMM,


TECHNICAL- Multiproduct
22 DIRECT DC Current Calibrator by 0.1 mA to 1 mA 2.36 % to 0.39 %
CURRENT Comparison/Direct
(Measure) Method
ELECTRO- Using 6 ½ DMM,
TECHNICAL- Multiproduct
23 DIRECT DC Current Calibrator by 1 mA to 100 mA 0.39 % to 0.17 %
CURRENT Comparison/Direct
(Measure) Method
ELECTRO- Using 6 ½ DMM ,
TECHNICAL- Multiproduct
24 DIRECT DC Current Calibrator by 100 mA to 3 A 0.17 % to 0.18 %
CURRENT Comparison/Direct
(Measure) Method
ELECTRO- Using 6 ½ DMM &
TECHNICAL- Shunt , Multiproduct
25 DIRECT DC Current Calibrator by 3 A to 20 A 0.18 % to 0.88 %
CURRENT Comparison/Direct
(Measure) Method
ELECTRO-
TECHNICAL- Using DMM with HV
26 DIRECT DC High Voltage Probe by Direct 1 kV to 30 kV 0.11 kV to 1.2 kV
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using DMM with HV
27 DIRECT DC High Voltage Probe by Direct 1 kV to 6 kV 0.11 kV to 0.33 kV
CURRENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 6 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO- Using 6 ½ DMM,


TECHNICAL- Multiproduct
28 DIRECT DC Resistance Calibrator by 1 ohm to 10 ohm 0.49 % to 0.13 %
CURRENT Comparison/Direct
(Measure) Method
ELECTRO- Using 6 ½ DMM ,
TECHNICAL- Multiproduct
1 Mohm to 100
29 DIRECT DC Resistance Calibrator by 0.08 % to 0.92 %
Mohm
CURRENT Comparison/Direct
(Measure) Method
ELECTRO- Using 6 ½ DMM,
TECHNICAL- Multiproduct
30 DIRECT DC Resistance Calibrator by 10 ohm to 1 Mohm 0.13 % to 0.08 %
CURRENT Comparison/Direct
(Measure) Method
ELECTRO- Using 6 ½ DMM,
TECHNICAL- Multiproduct
31 DIRECT DC Voltage Calibrator by 1 mV to 100 mV 0.45 % to 0.04 %
CURRENT Comparison/Direct
(Measure) Method
ELECTRO- Using 6 ½ DMM
TECHNICAL- ,Multiproduct
32 DIRECT DC Voltage Calibrator by 1 V to 1000 V 0.008%
CURRENT Comparison/Direct
(Measure) Method
ELECTRO- Using 6 ½ DMM,
TECHNICAL- Multiproduct
33 DIRECT DC Voltage Calibrator by 100 mV to 1 V 0.04 % to 0.008 %
CURRENT Comparison/Direct
(Measure) Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 7 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Multiproduct
34 DIRECT DC Current Calibrator by Direct 10 A to 20 A 0.15 % to 0.21 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
35 DIRECT DC Current Calibrator by Direct 100 µA to 100 mA 0.27 % to 0.16 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
36 DIRECT DC Current Calibrator by Direct 100 mA to 3 A 0.16 % to 0.08 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
37 DIRECT DC Current Calibrator by Direct 3 A to 10 A 0.08 % to 0.15 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multi Product
30 µA
38 DIRECT DC Current Calibrator by direct 0.52 % to 0.27 %
to 100 µA
CURRENT method
(Source)
ELECTRO-
Using Multiproduct
TECHNICAL-
Calibrator with 50
39 DIRECT DC Current 50 A to 1000 A 0.66 % to 0.7 %
Turns Coil by Direct
CURRENT
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 8 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using Resistance
TECHNICAL-
Box (Discrete
40 DIRECT DC Resistance 1 mohm 0.78%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Resistance
TECHNICAL-
Box(Discrete
41 DIRECT DC Resistance 1 ohm 0.23%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Resistance
TECHNICAL-
Box(Discrete
42 DIRECT DC Resistance 10 mohm 0.25%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Resistance
TECHNICAL-
Box(Discrete
43 DIRECT DC Resistance 100 mohm 0.24%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
44 DIRECT DC Voltage Calibrator by Direct 1 mV to 100 mV 0.17 % to 0.004 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
45 DIRECT DC Voltage Calibrator by Direct 1 V to 1000 V 0.002%
CURRENT Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 9 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Multiproduct
46 DIRECT DC Voltage Calibrator by Direct 100 mV to 1 V 0.004 % to 0.002 %
CURRENT Method
(Source)
ELECTRO-
Using Mega Ohm
TECHNICAL-
Box Sigma (Discrete
47 DIRECT High Resistance 1 Gohm 3.4%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Mega Ohm
TECHNICAL-
Box Sigma (Discrete
48 DIRECT High Resistance 100 Mohm 3.4%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Mega Ohm
TECHNICAL-
Box Sigma (Discrete
49 DIRECT High Resistance 2 GOhm 4.0%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Mega Ohm
TECHNICAL-
Box Sigma (Discrete
50 DIRECT High Resistance 50 Mohm 3.4%
Values)by Direct
CURRENT
Method
(Source)
ELECTRO-
Using Mega Ohm
TECHNICAL-
Box Sigma (Discrete
51 DIRECT High Resistance 500 Mohm 3.4%
Values)by Direct
CURRENT
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 10 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Multiproduct
52 DIRECT Resistance Calibrator by Direct 1 ohm to 100 ohm 0.12 % to 0.004 %
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
1 Mohm to 100
53 DIRECT Resistance Calibrator by Direct 0.007 % to 0.07 %
Mohm
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
100 ohm to 100
54 DIRECT Resistance Calibrator by Direct 0.004%
kohm
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
100 kohm to 1
55 DIRECT Resistance Calibrator by Direct 0.004 % to 0.007 %
Mohm
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multiproduct
100 Mohm to 1000
56 DIRECT Resistance Calibrator by Direct 0.07 % to 1.26 %
Mohm
CURRENT Method
(Source)
ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple R 0°C
57 TEMPERATURE Calibrator by direct 0.47°C
Type to 1750°C
SIMULATION method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 11 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple B 600°C
58 TEMPERATURE Calibrator by direct 0.64°C
Type to 1800°C
SIMULATION method
(Measure)
ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple E- (-)200°C
59 TEMPERATURE Calibrator by direct 0.57°C
Type to 1000°C
SIMULATION method
(Measure)
ELECTRO-
TECHNICAL- Using Multi Product
(-)200°C
60 TEMPERATURE Thermocouple J Type Calibrator by direct 0.31°C
to 1000°C
SIMULATION method
(Measure)
ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple K 0°C
61 TEMPERATURE Calibrator by direct 0.46°C
Type to 1300°C
SIMULATION method
(Measure)
ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple N- 0°C
62 TEMPERATURE Calibrator by direct 0.31°C
Type to 1300°C
SIMULATION method
(Measure)
ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple S 0°C
63 TEMPERATURE Calibrator by direct 0.56°C
Type to 1750°C
SIMULATION method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 12 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple T- (-)200°C
64 TEMPERATURE Calibrator by direct 0.28°C
type to 390°C
SIMULATION method
(Measure)
ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple E 0°C
65 TEMPERATURE Calibrator by direct 0.57°C
Type to 1000°C
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple N 0°C
66 TEMPERATURE Calibrator by direct 0.33°C
Type to 1300°C
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Using Multi Product
(-)200 °C
67 TEMPERATURE RTD - PT 100 type Calibrator by direct 0.27°C
to 800°C
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple B
68 TEMPERATURE Calibrator by direct 600 °C to 1800°C 0.64°C
Type
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Using Multi Product
(-)200°C
69 TEMPERATURE Thermocouple J Type Calibrator by direct 0.31°C
to 1000°C
SIMULATION method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 13 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple K 0°C
70 TEMPERATURE Calibrator by direct 0.46°C
Type to 1300°C
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple R 0°C
71 TEMPERATURE Calibrator by direct 0.65°C
Type to 1750°C
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple S 0°C
72 TEMPERATURE Calibrator by direct 0.54°C
Type to 1750°C
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Using Multi Product
Thermocouple T (-)200 °C
73 TEMPERATURE Calibrator by direct 0.72°C
Type to 390°C
SIMULATION method
(Source)
Using 6½ Digital
ELECTRO-
Multimeter ,
TECHNICAL-
Multiproduct
74 TIME & Frequency 10 Hz to 300 kHz 0.08 % to 0.02 %
Calibrator by
FREQUENCY
Comparison/Direct
(Measure)
Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 14 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using digital timer
75 TIME & Time by comparison 1 s to 10 s 0.022 s to 0.083 s
FREQUENCY method
(Measure)
ELECTRO-
TECHNICAL- Using digital timer
76 TIME & Time by comparison 10 s to 1000 s 0.083 s to 0.82 s
FREQUENCY method:
(Measure)
ELECTRO-
TECHNICAL- Using digital timer
77 TIME & Time by by comparison 1000 s to 10000 s 0.82 s to 22.73 s
FREQUENCY method
(Measure)
ELECTRO-
TECHNICAL- Using digital timer
78 TIME & Time by comparison 10000 s to 86400 s 22.73 s to 53.81 s
FREQUENCY method
(Measure)
ELECTRO-
TECHNICAL- Using Multi Product
79 TIME & Frequency Calibrator by direct 10 Hz to 1 MHz 0.007%
FREQUENCY method
(Source)
Using Digital
MECHANICAL- RPM Indicator/
Tachometer & RPM 1000 rpm to 25000
80 ACCELERATION Tachometer(Non- 0.68%
generator source by rpm
AND SPEED Contact Type)
comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 15 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Using Digital


100 rpm to 1000
81 ACCELERATION Centrifuge Tachometer by 2.99%
rpm
AND SPEED comparison Method
MECHANICAL- Using Digital
1000 rpm to 25000
82 ACCELERATION Centrifuge Tachometer by 0.68 %
rpm
AND SPEED comparison Method
Using Digital
MECHANICAL- RPM Indicator/
Tachometer & RPM 100 rpm to 1000
83 ACCELERATION Tachometer(Non- 2.99 %
generator source by rpm
AND SPEED Contact Type)
comparison Method:
Using Sound Level
MECHANICAL- Sound Level meter
84 Calibrator by 94 dB &114 dB 0.64dB
ACOUSTICS (1 kHz)
comparison method
Using Hydrometer
(Readability: 0.001
MECHANICAL- g/ml) and
Density 0.600 g/ml to 1.800
85 DENSITY AND Appropriate Liquids 0.001g/ml
Hydrometers g/ml
VISCOSITY by Comparison
method as per
based on IS 3104
MECHANICAL-
DIMENSION Using LMM, Master
Thread Ring Gauge
(BASIC Ring & Thread
86 & Thread Check Ring 6 mm to 100 mm 2µm
MEASURING Measuring Probe by
Gauge
INSTRUMENT, Comparison Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 16 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Using Sine bar,
(BASIC Gauge Block & Dial
87 Angle Gauge Upto 60 ° 20 Sec of arc
MEASURING gauge by Indirect
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
Using Master
DIMENSION
cylinder, surface
(BASIC Angle Plate / Box
88 plate and gauge 10 mm to 1000 mm 3µm
MEASURING Angle Plate
blocks by
INSTRUMENT,
comparison method
GAUGE ETC.)
MECHANICAL-
DIMENSION Master cylinder,
(BASIC Angle plate, box surface plate and
89 up to 600 mm 3µm
MEASURING plate Gauge blocks by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Dial Indicator,
(BASIC Bench Centre- Electronic Level &
90 up to 1000 mm 4 µm
MEASURING Flatness Mandrel by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Dial Indicator,
(BASIC Bench Centre-Co - Electronic Level &
91 upto 1000 mm 8µm
MEASURING Axiality Mandrel by
INSTRUMENT, Comparison Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 17 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Using Dial Indicator,
(BASIC Bench Centre- Electronic Level &
92 upto 1000 mm 8µm
MEASURING Parallelism Mandrel by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Bevel Protractor & Using Angle Gauges
(BASIC
93 Combination Set by Comparison 0 ° to 360 ° 4min. of arc
MEASURING
(L.C.: 5' ) Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Master
(BASIC Cylinder, Slip Gauge
94 Block square 10mm to 1000mm 3 µm
MEASURING & Surface Plate by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL- Using Dial
DIMENSION Calibration Tester,
Bore Gauge
(BASIC Dial Gauge,
95 Transmission Upto 2 mm 3.0µm
MEASURING Electronic Probe &
movement only
INSTRUMENT, gauge Block by
GAUGE ETC.) Comparison Method
MECHANICAL-
DIMENSION
Using Thickness
(BASIC Coating Thickness
96 Foils by Comparison up to 700 µm 5µm
MEASURING Gauge
Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 18 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Using Thickness Foil
(BASIC Coating Thickness
97 by Comparison upto 1000 µm 5.6µm
MEASURING Gauge
Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Surface Plate
(BASIC Comparator Stand ( & Dial Gauge with
98 300 mm x 300 mm 2µm
MEASURING Flatness of Table) stand by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Cube mould/ Prism Using Digital caliper 25mm x 25 mm x 25
(BASIC
99 mould/ cylindrical by Comparison mm to 150mm x 150 0.05mm
MEASURING
mould Method mm x 150 mm
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Master
(BASIC Cylindrical Square cylinder & gauge
100 upto 1000mm 5 µm
MEASURING (Squareness Only) block by comparison
INSTRUMENT, method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Master
(BASIC Cylindrical Square cylinder & gauge
101 upto 600mm 3 µm
MEASURING (Squareness Only) block by comparison
INSTRUMENT, method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 19 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Using Master
(BASIC Cylindrical Square cylinder & Gauge
102 upto 600 mm 3µm
MEASURING (Squareness Only) Block by Comparison
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
Depth Gauge ( LC - Checker, Gauge
(BASIC
103 0.01 mm) Vernier / Block & Length Bars 0 to 600 mm 7.2µm
MEASURING
Dial, Digital by Comparison
INSTRUMENT,
Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Dial Gauge
(BASIC Dial Gauge (Plunger) Calibrator and
104 0 to 25 mm 2µm
MEASURING L.C - 0.001mm Gauge Blocks by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Dial Gauge
(BASIC Dial Gauge (Plunger) Calibrator and
105 0 to 50 mm 6.11µm
MEASURING L.C :0.01mm Gauge Blocks by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Dial Gauge
(BASIC Dial Gauge (Lever Calibrator and
106 0 to 50 mm 6.11 µm
MEASURING Type) L.C: 0.01mm Gauge Blocks by
INSTRUMENT, Comparison Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 20 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Using Dial Gauge
(BASIC Dial gauge (Lever) Calibrator and
107 0 to 2mm 2 µm
MEASURING L.C: 0.001mm Gauge Blocks by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Gauge Blocks
Dial Gauge
(BASIC & Electronic Probe
108 Calibrator L.C :0.001 0 to 25 mm 2µm
MEASURING by Comparison
mm
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Dial Thickness Using Gauge blocks
(BASIC
109 Gauge , L.C: by Comparison 0 to 25 mm 2µm
MEASURING
0.001mm Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Length bars
(BASIC Digital Caliper 0.01 mm to 1000
110 by Comparison 10µm
MEASURING L.C:0.01 mm mm
Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using sine bar,
DIMENSION Digital Level/Angle
gauge block, master
(BASIC protector/frame
111 cylinder, surface 0° to 360° 0.22 °
MEASURING bubble level L.C.:
plate by Comparison
INSTRUMENT, 0.1°
Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 21 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Using Master
(BASIC Engineer's Square Cylinder, Slip Gauge
112 upto 1000mm 3 µm
MEASURING (Try Square) & Surface Plate by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Master
(BASIC Engineer's Square ( Cylinder, Slip Gauge
113 upto 600 mm 3.0µm
MEASURING Try Square) & Surface Plate by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Gauge Block &
(BASIC External Micrometer
114 Length Bar by 100 mm to 600 mm 7.0µm
MEASURING L.C: 0.01 mm
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Gauge Block &
(BASIC External Micrometer
115 Length Bar by 0 to 150 mm 0.83µm
MEASURING L.C: 0.001 mm
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Gauge Blocks
(BASIC External Micrometer 600 mm to 1000
116 & Length Bars by 4.0 µm
MEASURING L.C: 0.01 mm mm
Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 22 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Using Gauge Block &
(BASIC External Micrometer
117 Length Bar by Upto 100 mm 3.0µm
MEASURING L.C: 0.01 mm
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Digital
(BASIC
118 Feeler Gauge Micrometer by 0.03 mm to 1 mm 3.40µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
Checker, Gauge
(BASIC Height Gauge L.C:
119 Block ,Length Bars & 0 to 300 mm 8 µm
MEASURING 0.01mm
Lever Dial gauge by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Gauge Block ,
(BASIC Height Gauge L.C: Length Bars & Lever 300 mm to 1000
120 10µm
MEASURING 0.01mm Dial Gauge by mm
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Gauge blocks
(BASIC Indside dial Caliper
121 by Comparison 4 mm to 75 mm 8µm
MEASURING L.C: 0.010mm
Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 23 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
Using Caliper
DIMENSION
Inside Micrometer , Checker, Gauge
(BASIC 300 mm to 1000
122 (Two Jaw & Stick Block Accessories & 10µm
MEASURING mm
type) L.C: 0.01mm Length Bars by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
Inside Micrometer , Checker, Gauge
(BASIC
123 (Two Jaw & Stick Block Accessories & 25 mm to 50 mm 8µm
MEASURING
type) L.C: 0.01mm Length Bars by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
Inside Micrometer , Checker, Gauge
(BASIC
124 (Two Jaw & Stick Block Accessories & 50 mm to 300 mm 8µm
MEASURING
type) L.C: 0.01mm Length Bars by
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
Inside Micrometer, Checker, Gauge
(BASIC
125 (Two Jaw & Stick Block Accessories & 5 mm to 25 mm 3.4µm
MEASURING
type) L.C: 0.01mm Length Bars by
INSTRUMENT,
Comparison Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 24 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Jigs and fixtures,


MECHANICAL- indenter balls,
Using Digimatic
DIMENSION length, inner and
micrometer,
(BASIC outer dia, depth,
126 Digimatic caliper , 4 mm to 300 mm 1 µm to 15
MEASURING Elongation index
dial gauge by
INSTRUMENT, gauge, Flakiness
Comparison Method
GAUGE ETC.) Index gauge, Gauge
and component
MECHANICAL-
Using Master
DIMENSION
Cylinder, Surface
(BASIC
127 Master Cylinder Plate & Gauge Block upto 600 mm 3.0µm
MEASURING
by Comparison
INSTRUMENT,
Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using scale
(BASIC Pie Tape L.C: 0.1 calibration unit with 1 mm to Every 7µm+sqrt(L)µm
128
MEASURING mm DRO by Comparison subsequent m where L in meter
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Length
(BASIC Measuring Machine
129 Pin Gauge Upto 10 mm 1µm
MEASURING by Comparison
INSTRUMENT, Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 25 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Using Profile
(BASIC
130 Pitch gauge Projector by Upto 10 mm 5µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Length
(BASIC Measuring Machine
131 Plain Plug Gauge 0.5 mm to 100 mm 1µm
MEASURING by Comparison
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Length
(BASIC Measuring Machine
132 Plain Ring Gauge 3 mm to 250 mm 2.0µm
MEASURING & Master Ring by
INSTRUMENT, Comparison method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Profile
(BASIC
133 Radius gauge Projector by upto 50 mm 7.7µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
checker, Lever Dial
(BASIC
134 Riser Block gauge & Length bar Upto 100 mm 1µm
MEASURING
by Comparison
INSTRUMENT,
Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 26 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
Using Caliper
DIMENSION
checker, Lever Dial
(BASIC
135 Riser Block Gauge & Length Bar upto 300 mm 2µm
MEASURING
by Comparison
INSTRUMENT,
Method
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
checker, lever Dial
(BASIC
136 Riser Block Gauge & Length Bar upto 600 mm 3.0µm
MEASURING
by Comparison
INSTRUMENT,
Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Digital Vernier
(BASIC
137 Sieves Caliper by 10 mm to 150 mm 30µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Profile
(BASIC
138 Sieves Projector by Upto 10 mm 6.4µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Angle Gauge,
(BASIC Dial gauge & Gauge
139 Sine bar Upto 300 mm 4.77 µm
MEASURING Block by Comparison
INSTRUMENT, Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 27 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Using Gauge Blocks
(BASIC
140 Snap Gauge by Comparison 2 mm to 100 mm 3µm
MEASURING
Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Electronic
Spirit Level ,
(BASIC Level & Titling Table
141 Sensitivity : 0 ' to 1 ' 8µm/m
MEASURING By Comparison
0.02mm/m
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Electronic
Spirit Level
(BASIC Level & Titling Table
142 Sensitivity : 0 ' to 6 ' 6µm/m
MEASURING by Comparison
0.01mm/m
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Steel Scale
(BASIC Steel Scale L.C: 0.5 70+1.72 sqrt ( L) µm
143 Fixture with DRO by 0 to 2000 mm
MEASURING mm , L in meter)
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Scale
(BASIC Steel Tape L.C :1 calibration unit with
144 0 to 1000 mm 7.04 µm
MEASURING mm DRO by Comparison
INSTRUMENT, Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 28 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
Straight Edge
DIMENSION
Thickness upto Using Surface Plate
(BASIC 1x sqrt (L) µm,
145 10mm I-section & Electronic Level by up to 6000 mm
MEASURING where L in m
Thickness 20mm & indirect method
INSTRUMENT,
above (Straightness)
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Electronic
0.2 x sqrt ( (L+
(BASIC Surface Plate Level least count upto size
146 W)/150 ) µm, Where
MEASURING (Flatness) 0.001 mm/m By Grid (6000X4000) mm
L , W are in mm
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Gauge Block &
(BASIC Digital Micrometer,
147 Taper Plug Gauge Upto 100 mm 3µm
MEASURING Standard Wires by
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
Using Floating
DIMENSION
Carriage Micrometer
(BASIC Taper Thread Plug
148 & Thread Measuring Upto 100 mm 3.5µm
MEASURING Gauge (Only PCD)
wires by Comparison
INSTRUMENT,
Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Gauge Block
(BASIC Comparator &
149 Thickness Foil 10 µm to 1000 µm 4µm
MEASURING Gauge Block by
INSTRUMENT, Comparison Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 29 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
Using Floating
DIMENSION
Thread Plug Gauge, Carriage Micrometer
(BASIC
150 wear Check Plugs & Thread Measuring Upto 100 mm 2.70µm
MEASURING
(Only PCD) wires by Comparison
INSTRUMENT,
Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Standard Ring
Three pin
(BASIC Gauges of different
151 micrometer, Bore 6 mm to 150 mm 3 µm
MEASURING sizes.by Comparison
gauge, Hole tester
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Slip Gauge set
(BASIC Ultra Sonic
152 by Comparison 1 mm to 100 mm 49µm
MEASURING Thickness Gauge
Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
Checker, Gauge
(BASIC Vernier Caliper L.C: 300 mm to 1000
153 Block & Length Bars 10µm
MEASURING 0.01mm mm
by Comparison
INSTRUMENT,
Method
GAUGE ETC.)
MECHANICAL-
Using Caliper
DIMENSION
Checker, Gauge
(BASIC Vernier Caliper
154 Block & Length Bars 0 to 300 mm 7.2µm
MEASURING L.C:0.01mm
by Comparison
INSTRUMENT,
Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 30 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
Dial calibration Using Gauge block
DIMENSION
155 tester (L.C: 0.0001 of Grade 00 by 0 to 25 mm 1 µm
(PRECISION
mm ) Comparison Method
INSTRUMENTS)
MECHANICAL- Using Electronic
DIMENSION Electronic Level (L.C: Level, Titling Table &
156 ± 2 mm/m 3µm/m
(PRECISION 0.001mm/m) Electronic probe by
INSTRUMENTS) Comparison Method
Using Gauge Block
MECHANICAL-
Calibrator and
DIMENSION
157 Gauge Block Gauge Block Set 10 mm to 50 mm 0.20µm
(PRECISION
Grade 00 by
INSTRUMENTS)
Comparison Method
Using Gauge Block
MECHANICAL-
Calibrator and
DIMENSION
158 Gauge Block Gauge Block Set 50 mm to 100 mm 0.41µm
(PRECISION
Grade by
INSTRUMENTS)
Comparison Method
Using Gauge Block
MECHANICAL-
Calibrator and
DIMENSION
159 Gauge Block Gauge Block Set Up to 10 mm 0.12µm
(PRECISION
Grade 00 by
INSTRUMENTS)
Comparison Method
Using Caliper
MECHANICAL-
Checker, Gauge
DIMENSION Height Master(LC -
160 Block ,Length Bars & 0 to 600 mm 3.5µm
(PRECISION 0.001mm)
Lever Dial gauge by
INSTRUMENTS)
Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 31 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Using Length Bars,


DIMENSION LVDT Probe &
161 Length Bars 100 mm to 300 mm 1.10µm
(PRECISION Surface Plate by
INSTRUMENTS) Comparison Method
MECHANICAL- Using Length Bars,
DIMENSION LVDT Probe &
162 Length Bars 300 mm to 600 mm 4.3µm
(PRECISION Surface Plate by
INSTRUMENTS) Comparison Method
MECHANICAL- Using Length Bars,
DIMENSION LVDT Probe & 600 mm to 1000
163 Length Bars 5µm
(PRECISION Surface Plate by mm
INSTRUMENTS) Comparison Method
MECHANICAL- Using Length Bars,
DIMENSION LVDT Probe &
164 Length Bars Upto 100 mm 1.3µm
(PRECISION Surface Plate by
INSTRUMENTS) Comparison Method
MECHANICAL- Using Length Bars ,
DIMENSION Linear Height Gauge Lever Dial gauge
165 0 to 1000 mm 4.2µm
(PRECISION ( L.C: 0.001 mm) and Gauge Blocks by
INSTRUMENTS) Comparison Method
MECHANICAL- Using Gauge Blocks
DIMENSION Linear probe by Comparison
166 0 to 2 mm 1.01µm
(PRECISION (Digital), LC: 0.1µm Method on surface
INSTRUMENTS) plate
MECHANICAL-
Profile Projector Using Angle gauge
DIMENSION
167 Angular Scale (LC: by Comparison upto 90° 57.7Arc sec
(PRECISION
1" ) Method
INSTRUMENTS)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 32 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Using Glass Scale,


Profile Projector
DIMENSION Standard wires &
168 Magnification 10x, up to 100 x 0.17%
(PRECISION Vernier Caliper by
20x,100x
INSTRUMENTS) Comparison Method
MECHANICAL-
Profile Projector
DIMENSION Using Glass Scale by
169 linear scale (L.C: upto 300 mm 2.3µm
(PRECISION Comparison Method
0.001 mm)
INSTRUMENTS)
MECHANICAL-
DIMENSION Steel Scale/ tape Using Length bar by
170 upto 1000 mm 6.9µm
(PRECISION Calibrator Comparison Method
INSTRUMENTS)
Using digital single
pan electronic
MECHANICAL-
171 Shore A tester balance of 0.1 g and 0 to 100 Shore A 0.5Shore A
DUROMETER
fixture as per ASTM
2240D
Using digital single
pan electronic
MECHANICAL-
172 Shore D tester balance of 0.1 g and 0 to 100 Shore D 0.5Shore D
DUROMETER
fixture as per ASTM
2240D
Using fixture with
aluminium and SS
MECHANICAL-
Hangers for push
MOBILE FORCE
173 Mobile Force Gauge and pull mode and 5 N to 1000 N 0.3%
MEASURING
F1 class SS weights
SYSTEM
by comparsion
method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 33 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Digital
Digital Vacuum
MECHANICAL- Vacuum Gauge &
Gauge/ Indicator
PRESSURE Vacuum pump, DMM
174 Vacuum Transducer, (-)0.95 bar to 0 0.006bar
INDICATING Comparator
Vacuum Transmitter,
DEVICES Comparison method
Vacuum Switch.
as per DKD- R-6-1
Digital/Dial Pressure
Digital Pressure
Gauge/ Indicator.
MECHANICAL- Gauge &
Pressure
PRESSURE Comparator, DMM
175 Transducer, 0 to 10 bar 0.013bar
INDICATING by Comparison
Pressure
DEVICES method as DKD-
Transmitter,
R-6-1
Pressure Switch.
Hydraulic Pressure -
Digital/Dial Pressure
Using Digital
MECHANICAL- Gauge/ Indicator.
Pressure Gauge &
PRESSURE Pressure
176 Comparator, DMM 10 bar to 700 bar 0.45bar
INDICATING Transducer,
by Comparison as
DEVICES Pressure
per DKD- R-6-1
Transmitter,
Pressure Switch.
MECHANICAL- Torque Wrench Type Using Digital Torque
TORQUE - 1, Class, A,B,C,D,E Calibrator By
177 2 Nm to 20 Nm 0.96%
GENERATING Type - II, Class, Comparison Method
DEVICES A,B,C,D,E,F, & G as per IS/ISO:6789
MECHANICAL- Torque Wrench Type Using Digital Torque
TORQUE - 1, Class, A,B,C,D,E Calibrator By
178 2 Nm to 200 Nm 0.64%
GENERATING Type - II, Class, Comparison Method
DEVICES A,B,C,D,E,F, & G as per IS/ISO:6789

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 34 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Torque Wrench Type Digital Torque


TORQUE - 1, Class, A,B,C,D,E Calibrator By
179 200 Nm to 2000 Nm 0.75%
GENERATING Type - II, Class, Comparison Method
DEVICES A,B,C,D,E,F, & G as per IS/ISO:6789
Using primary
MECHANICAL-
method, calibrated 1
TORQUE Torque Transducer
180 m beam and F1 100 Nm to 200 Nm 0.15Nm
MEASURING calibration
class SS weights as
DEVICES
per BS7882
Using primary
MECHANICAL- method, calibrated 1
TORQUE Torque Transducer metre beam, pulley
181 5 Nm to 100 Nm 0.04%
MEASURING calibration, and F1 class SS and
DEVICES AL weights as per BS
7882
Using primary
Torque transducer,
MECHANICAL- method, calibrated
Torque sensor
TORQUE pulley and F1 class
182 with/without 0.05 Nm to 50 Nm 0.12%
MEASURING AL weights , Digital
indicator, Torque
DEVICES multimeter as per
meter
BS 7882
Using Standard
Weights F1 Class ,
Precision balances,
Measuring Cylinder/
MECHANICAL- double distilled
183 Volumetric Flask/ 0.1 ml to 100 ml 2µl
VOLUME water of known
Graduated Jar/ Cane
density as per
Gravimetric method
based on ISO 4787

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 35 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Standard Weights F1
Class , precision
balances, double
Measuring Cylinder/
MECHANICAL- distilled water of
184 Volumetric Flask/ 1 l to 5 l 8.6ml
VOLUME known density as
Graduated Jar/ Cane
per Gravimetric
method based on
ISO 4787
Using Standard
Weights F1 Class ,
precision balances,
Measuring Cylinder/
MECHANICAL- double distilled
185 Volumetric Flask/ 100 ml to 1 l 0.5ml
VOLUME water of known
Graduated Jar/ Cane
density as per
Gravimetric method
based on ISO 4787
Using Standard
Weights F1 Class ,
precision balances,
Measuring Cylinder/
MECHANICAL- double distilled
186 Volumetric Flask/ 5 l to 20 l 9.04ml
VOLUME water of known
Graduated Jar/ Cane
density as per
Gravimetric method
based on ISO 4787

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 36 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Standard
Weights & Semi-
micro balance and
MECHANICAL- distilled water of
187 Micro Pipette 1 ml to 10 ml 0.8µl
VOLUME known density as
per Gravimetric
method based on
ISO 8655-6
Using Standard
Weights & Semi-
micro balance and
MECHANICAL- distilled water of
188 Micro Pipette 10 µl to 100 µl 0.5µl
VOLUME known density as
per Gravimetric
method based on
ISO 8655-6
Using Standard
Weights & Semi-
micro balance and
MECHANICAL- distilled water of
189 Micro Pipette 100 µl to 1000 µl 0.6 µl
VOLUME known density as
per Gravimetric
method based on
ISO 8655-6

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 37 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Standard
Weights & Semi-
micro balance and
MECHANICAL- distilled water of
190 Pipettes & Burettes 0.1 ml to 5 ml 1µl
VOLUME known density as
per Gravimetric
method based on
ISO 4787
Standard Weights &
Semi- micro balance
and distilled water of
MECHANICAL-
191 Pipettes & Burettes known density as 5 ml to 100 ml 1.3µl
VOLUME
per Gravimetric
method based on
ISO 4787
Using Standard
MECHANICAL- Digital weighing
Weights of Class E1
WEIGHING balance (Class 1 and
192 as per Procedure 1 mg to 20 g 0.08mg
SCALE AND coarser) Readability:
based on OIML R76
BALANCE 0.001 mg
(2006)
Using Standard
MECHANICAL- Digital weighing
Weights of Class F1
WEIGHING balance (Class I and
193 as per Procedure 0 to 80 g 0.08mg
SCALE AND coarser )
based on OIML R76
BALANCE Readability:0.01 mg
(2006)
Using Standard
MECHANICAL- Digital weighing
Weights of Class E1
WEIGHING balance,(Class 1 and
194 as per Procedure 1 mg to 200 g 0.09mg
SCALE AND coarser)
based on OIML R76
BALANCE Readability:0.01 mg
(2006)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 38 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Digital weighing Using Standard


WEIGHING balance,(Class 1 and Weights of F1 Class
195 0 to 200 g 0.13mg
SCALE AND coarser) as per OIML R76
BALANCE Readability:0.1 mg (2006)
MECHANICAL- Using Standard
Weighing Machine
WEIGHING Weights of F1 Class
196 (Class II & Coarser) 0 to 1 kg 2.5mg
SCALE AND Accuracy as per
Readability:1 mg
BALANCE Oiml -R-76-1
MECHANICAL-
Weighing Machine Using Standard
WEIGHING
197 (Class II & Coarser) Weights of F1 Class upto 5 kg 20mg
SCALE AND
Readability:100 mg As per OIML-R-76-1
BALANCE
MECHANICAL- Using Standard
Weighing Machine
WEIGHING Weights F1 Class
198 (Class III & Coarser) upto 50 kg 4.1 g
SCALE AND Accuracy as per
Readability:1 g
BALANCE OIML-R-76-1
MECHANICAL-
Weighing Machine Using Standard
WEIGHING
199 (Class IV & Coarser ) Weight of F1 Class upto 100 kg 2g
SCALE AND
Readability:1g as per OIML-R-76-1
BALANCE
MECHANICAL-
Weighing Machine Using Standard
WEIGHING
200 (Class IV & Coarser) Weight of F1 Class upto 200 kg 26g
SCALE AND
Readability:20g as per OIML-R-76-1
BALANCE

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 39 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- weights (M1 class (readability 0.2 g) by
201 5 kg 0.12g
WEIGHTS and coarser) Substitution/ ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weight( M1 class (readability 0.2 g )
202 1 kg 0.1g
WEIGHTS and coarser) Substitution/ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision
MECHANICAL- Weight( M1 class Balance(readability
203 50 kg 6g
WEIGHTS and coarser) 5 g )Substitution/
ABA Weighing cycle
method as per OIML
R 111

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 40 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Standard
Weights F1 Class &
Precision Balance(
MECHANICAL- Weight( M1 class readability 0.2 g)
204 500 g 0.1g
WEIGHTS and coarser) Substitution/ ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- weights ( F2 class (readability:0.01mg)
205 1 mg 0.02mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights ( F2 class (readability:0.1mg)
206 10 g 0.20mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 41 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights ( F2 class (readability:0.1mg)
207 10 mg 0.02mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights ( F2 class (readability:0.01mg)
208 100 mg 0.05mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights ( F2 class (readability:0.01mg)
209 20 g 0.25mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111:

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 42 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights ( F2 class (readability:0.01mg)
210 20 mg 0.03mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights ( F2 class (readability:0.1mg)
211 200 g 1.0mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111 :
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights ( F2 class (readability:0.01mg)
212 50 g 0.30mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 43 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (F2 class (readability:0.01mg)
213 1g 0.10mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (F2 class (readability:0.01mg)
214 100 g 0.5mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (F2 class (readability:0.01mg)
215 2g 0.12mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111 :

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 44 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (F2 class (readability:0.01mg)
216 2 mg 0.02mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (F2 class (readability:0.01mg)
217 200 mg 0.06mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111:
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (F2 class (readability:0.1mg)
218 5g 0.16mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 45 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (F2 class (readability:0.01mg)
219 5 mg 0.02mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (F2 class (readability:0.01mg)
220 50 mg 0.04mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (F2 class (readability:0.01mg)
221 500 mg 0.08mg
WEIGHTS and coarser) by Substitution/ ABA
Weighing cycle
method as per OIML
R 111

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 46 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (M1 class (readability 5 g) by
222 10 kg 5g
WEIGHTS and coarser) Substitution/ABA
Weighing cycle
method as per OIML
R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (M1 class (readability 0.2mg)
223 2 kg 0.1g
WEIGHTS and coarser) by Substitution/ABA
Weighing cycle
method as per on
OIML R 111
Using Standard
Weights F1 Class &
Precision Balance
MECHANICAL- Weights (M1 class (readability 5 g) by
224 20 kg 5g
WEIGHTS and coarser) Substitution/ABA
Weighing cycle
method as per OIML
R 111
Using light chamber
and calibrated
OPTICAL-
225 Digital Lux meter illumination meter 1 lx to 20000 lx 3.6%
EQUIPMENTS
by comparison
method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 47 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Relative
Humidity Sensor Humidity Sensor
THERMAL-
with Indicator with Indicator, 10 %rh to 95 %rh @
226 SPECIFIC HEAT 1.16%rh
(Dial/Digital)/Thermo Humidity Generator 25ºC
& HUMIDITY
Hygrometer by Comparison
Method
Thermohygrometer, Using Digital
Temperature Temperature and
THERMAL- indicator,Thermomet Humidity sensor
10 ºC to 50
227 SPECIFIC HEAT er/Data logger/ with Indicator, 0.82 ºC
ºC @50%rh
& HUMIDITY Temperature Gauge Temperature
With inbuilt chamber by
Temprature Sensor comparison method:
Temperature
Transmitter, RTD,
Thermocouples with
Using RTD sensor
& without
with Indicator, 6 ½
controller/Data
THERMAL- DMM, Negative
228 Logger/recorder, (-)30°C to 50°C 0.21°C
TEMPERATURE Temperature bath
Temperature Gauge,
by Comparison
Liquid in Glass
method
Thermometer,
Digital Thermometer
with inbuilt sensor

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 48 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Temperature
Transmitter, RTD,
Thermocouples with
& without
Using RTD sensor
controller/Data
THERMAL- with Indicator, 6 ½
229 Logger/recorder, 50°C to 300°C 0.62°C
TEMPERATURE DMM, Oil bath by
Temperature Gauge,
Comparison method
Liquid in Glass
Thermometer,
Digital Thermometer
with inbuilt sensor
Temperature
Transmitter, Using R Type
Thermocouples with Thermocouple With
THERMAL- & without Indicator, 6½ DMM, 300°C
230 1.5°C
TEMPERATURE controller/Data Dry Block Furnace to 600°C
Logger/recorder, by Comparison
Digital Thermometer Method
with inbuilt sensor
Temperature
Transmitter, Using R Type
Thermocouples with Thermocouple With
THERMAL- & without Indicator, 6½ DMM, 600°C
231 1.88 °C
TEMPERATURE controller/Data Dry Block Furnace to 1200°C
Logger/recorder, by Comparison
Digital Thermometer Method
with inbuilt sensor

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 49 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Site Facility
ELECTRO-
TECHNICAL-
Using DMM with HV
Alternating Ac High Voltage @
1 Probe by Direct 1 kV to 25 kV 0.1 kV to 1.7 kV
Current (< 1 50 Hz
Method
GHz)
(Measure)
ELECTRO-
TECHNICAL- Using DMM with HV
2 DIRECT DC High Voltage Probe by Direct 1 kV to 30 kV 0.11 kV to 1.2 kV
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using Universal
(-)200°C
3 TEMPERATURE RTD - PT 100 type Calibrator by direct 0.40°C
to 850°C
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Using Universal
(-)200°C
4 TEMPERATURE Thermocouple J Type Calibrator by direct 0.35°C
to 1050°C
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Using Universal
Thermocouple K 0°C
5 TEMPERATURE Calibrator by direct 0.46 °C
Type to 1300°C
SIMULATION method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 50 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Universal
Thermocouple R 0°C
6 TEMPERATURE Calibrator by direct 0.57°C
Type to 1750°C
SIMULATION method
(Source)
ELECTRO-
TECHNICAL- Using Universal
Thermocouple S 0 °C
7 TEMPERATURE Calibrator by direct 0.57 °C
Type to 1750 °C
SIMULATION method
(Source)
Using Digital
MECHANICAL- RPM Indicator/
Tachometer & RPM 1000 rpm to 25000
8 ACCELERATION Tachometer(Non- 0.68%
generator source by rpm
AND SPEED Contact Type)
comparison Method
MECHANICAL- Using Digital
100 rpm to 1000
9 ACCELERATION Centrifuge Tachometer by 2.99%
rpm
AND SPEED comparison Method
MECHANICAL- Using Digital
1000 rpm to 25000
10 ACCELERATION Centrifuge Tachometer by 0.68 %
rpm
AND SPEED comparison Method
Using Digital
MECHANICAL- RPM Indicator/
Tachometer & RPM 100 rpm to 1000
11 ACCELERATION Tachometer(Non- 2.99 %
generator source by rpm
AND SPEED Contact Type)
comparison Method:
Using Sound Level
MECHANICAL- Sound Level meter
12 Calibrator by 94 dB &114 dB 0.64dB
ACOUSTICS (1 kHz)
comparison method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 51 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
Straight Edge
DIMENSION
Thickness upto Using Surface Plate
(BASIC 1x sqrt (L) µm,
13 10mm I-section & Electronic Level by up to 6000 mm
MEASURING where L in m
Thickness 20mm & indirect method
INSTRUMENT,
above (Straightness)
GAUGE ETC.)
MECHANICAL-
DIMENSION Using Electronic
0.2 x sqrt ( (L+
(BASIC Surface Plate Level least count upto size
14 W)/150 ) µm, Where
MEASURING (Flatness) 0.001 mm/m By Grid (6000X4000) mm
L , W are in mm
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
Dial calibration Using Gauge block
DIMENSION
15 tester (L.C: 0.0001 of Grade 00 by 0 to 25 mm 1 µm
(PRECISION
mm ) Comparison Method
INSTRUMENTS)
Using Caliper
MECHANICAL-
Checker, Gauge
DIMENSION Height Master(LC -
16 Block ,Length Bars & 0 to 600 mm 3.5µm
(PRECISION 0.001mm)
Lever Dial gauge by
INSTRUMENTS)
Comparison Method
MECHANICAL-
Using Length Bars &
DIMENSION Linear Height Gauge
17 Lever Dial Gauge by upto 1000 mm 4.2µm
(PRECISION ( L.C: 0.0001 mm)
Comparison Method
INSTRUMENTS)
MECHANICAL-
Profile Projector Using Angle gauge
DIMENSION
18 Angular Scale (LC: by Comparison upto 90° 57.7Arc sec
(PRECISION
1" ) Method
INSTRUMENTS)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 52 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Using Glass Scale,


Profile Projector
DIMENSION Standard wires &
19 Magnification 10x, up to 100 x 0.17%
(PRECISION Vernier Caliper by
20x,100x
INSTRUMENTS) Comparison Method
MECHANICAL-
Profile Projector
DIMENSION Using Glass Scale by
20 linear scale (L.C: upto 300 mm 2.3µm
(PRECISION Comparison Method
0.001 mm)
INSTRUMENTS)
MECHANICAL-
DIMENSION Steel Scale/ tape Using Length bar by
21 upto 1000 mm 6.9µm
(PRECISION Calibrator Comparison Method
INSTRUMENTS)
MECHANICAL- Using Hardness
HARDNESS Hardness Rockwell Blocks Indirect
22 20 HRC to 70 HRC 0.45 HRC
TESTING Testing machines Method as per IS
MACHINES :1586
Using Digital
Digital Vacuum
MECHANICAL- Vacuum Gauge &
Gauge/ Indicator
PRESSURE Vacuum pump, DMM
23 Vacuum Transducer, (-)0.95 bar to 0 0.006bar
INDICATING Comparator
Vacuum Transmitter,
DEVICES Comparison method
Vacuum Switch.
as per DKD- R-6-1
Digital/Dial Pressure
Digital Pressure
Gauge/ Indicator.
MECHANICAL- Gauge &
Pressure
PRESSURE Comparator, DMM
24 Transducer, 0 to 10 bar 0.013bar
INDICATING by Comparison
Pressure
DEVICES method as DKD-
Transmitter,
R-6-1
Pressure Switch.

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 53 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Hydraulic Pressure -
Digital/Dial Pressure
Using Digital
MECHANICAL- Gauge/ Indicator.
Pressure Gauge &
PRESSURE Pressure
25 Comparator, DMM 10 bar to 700 bar 0.45bar
INDICATING Transducer,
by Comparison as
DEVICES Pressure
per DKD- R-6-1
Transmitter,
Pressure Switch.
MECHANICAL-
UTM, TENSION
Using Dynamometer
CREEP AND Compression Testing
26 (Class - 1) as per 200 kN to 2000 kN 0.9%
TORSION machine
IS:1828
TESTING
MACHINE
MECHANICAL-
UTM, TENSION
Using Proving Ring
CREEP AND Compression Testing
27 (Class 1) as per IS 200 kN to 3000 kN 0.6%
TORSION Machine
1828
TESTING
MACHINE
MECHANICAL-
UTM, TENSION
Using Dynamometer
CREEP AND Compression Testing
28 (Class- 0/1) as per 50 kN to 1000 kN 0.5%
TORSION machine
IS:1828
TESTING
MACHINE

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 54 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
UTM, TENSION
Using Digital load
CREEP AND Tensile/ compression
29 cell class 0/1 as per 0.5 N to 50 N 0.2%
TORSION testing machines
IS 1828
TESTING
MACHINE
MECHANICAL-
UTM, TENSION
Tensile/ Using Digital Load
CREEP AND
30 Compression Testing Cell (class - 0/1) as 50 N to 50 kN 0.7%
TORSION
mchine per IS :1828
TESTING
MACHINE
Using Standard
MECHANICAL- Digital weighing
Weights of Class E1
WEIGHING balance (Class 1 and
31 as per Procedure 1 mg to 20 g 0.08mg
SCALE AND coarser) Readability:
based on OIML R76
BALANCE 0.001 mg
(2006)
Using Standard
MECHANICAL- Digital weighing
Weights of Class F1
WEIGHING balance (Class I and
32 as per Procedure 0 to 80 g 0.08mg
SCALE AND coarser )
based on OIML R76
BALANCE Readability:0.01 mg
(2006)
Using Standard
MECHANICAL- Digital weighing
Weights of Class E1
WEIGHING balance,(Class 1 and
33 as per Procedure 1 mg to 200 g 0.09mg
SCALE AND coarser)
based on OIML R76
BALANCE Readability:0.01 mg
(2006)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 55 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL- Digital weighing Using Standard


WEIGHING balance,(Class 1 and Weights of F1 Class
34 0 to 200 g 0.13mg
SCALE AND coarser) as per OIML R76
BALANCE Readability:0.1 mg (2006)
MECHANICAL- Using Standard
Weighing Machine
WEIGHING Weights of F1 Class
35 (Class II & Coarser) 0 to 1 kg 2.5mg
SCALE AND Accuracy as per
Readability:1 mg
BALANCE Oiml -R-76-1
MECHANICAL-
Weighing Machine Using Standard
WEIGHING
36 (Class II & Coarser) Weights of F1 Class upto 5 kg 20mg
SCALE AND
Readability:100 mg As per OIML-R-76-1
BALANCE
MECHANICAL- Using Standard
Weighing Machine
WEIGHING Weights F1 Class
37 (Class III & Coarser) upto 50 kg 4.1 g
SCALE AND Accuracy as per
Readability:1 g
BALANCE OIML-R-76-1
MECHANICAL-
Weighing Machine Using Standard
WEIGHING
38 (Class IV & Coarser ) Weight of F1 Class upto 100 kg 2g
SCALE AND
Readability:1g as per OIML-R-76-1
BALANCE
MECHANICAL-
Weighing Machine Using Standard
WEIGHING
39 (Class IV & Coarser) Weight of F1 Class upto 200 kg 26g
SCALE AND
Readability:20g as per OIML-R-76-1
BALANCE

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 56 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Humidity Sensor
with Indicator of
Environmental/condi Using Relative
THERMAL-
tioning Humidity Sensor 10 %rh to 95%rh @
40 SPECIFIC HEAT 1.16%rh
Chamber/Humidity with Indicator by 25ºC
& HUMIDITY
chamber For (Non- comparison method
Medical Purpose)
(Single Position)
Deep Freezer,
Freezer,
Incubator(Non-
medical puprose
only),BOD Incubator Using Data Logger
(Non-medical with RTD Sensors
THERMAL- (-)30 ºC
41 puprose only), (Minimum 9 sensor) 1 ºC
TEMPERATURE to 50 ºC
Oven,Water bath, by Multi Position
Temperature method
Furnace, Chamber,
Auto clave (Non-
medical puprose
only)
Incubator ( Non-
Medical Purpose)
,BOD Incubator (
Using Data Logger
Non-Medical
with RTD Sensors
THERMAL- Purpose), Oven,
42 (Minimum 9 50 ºC to 300 ºC 1ºC
TEMPERATURE Water bath,
Sensors) by Multi
Temperature
Position Method
Furnace, Chamber,
Auto clave (Non-
Medical Purpose)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 57 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Oven, Chamber, Using Data Logger


Temperature with K-Type Sensors
THERMAL-
43 Furnace, Industrial ( Minimum 9 300 ºC to 1200 ºC 6.23ºC
TEMPERATURE
Furnace For (Non- Sensors) by Multi
Medical Purpose) Position Method
Temperature
Indicator With
Sensor of Deep
Freezer, Freezer,
Incubator, BOD
Using RTD sensor
THERMAL- Incubator, Oven, (-)80ºC
44 With Indicator By 0.26°C
TEMPERATURE Water bath, to 50ºC
Comparison Method
Temperature
Furnace, Chamber,
Auto clave For ( Non
Medical Purpose) (
Single Position)
Temperature
Indicator With
Using R Type
Sensor of Dry Block
THERMAL- Thermocouple with
45 furnace , Muffle 300°C to 1200°C 1.88°C
TEMPERATURE Indicator by
Furnace For ( Non-
Comparison method
Medical Purpose) (
Single Position)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 58 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Temperature
Indicator With
Sensor of Incubator (
Non-Medical
Purpose), BOD
Incubator ( Non-
Medical Purpose),
Using RTD sensor
THERMAL- Oven, Temperature
46 with Indicator by 50°C to 300°C 0.65°C
TEMPERATURE Furnace,
Comparison method
Temperature
Chamber, Auto
Clave ( Non-Medical
Purpose), Dry Block
Furnace, Muffle
Furnace ( Single
Position )
Temperature
Transmitter, RTD,
Thermocouples with
Using RTD sensor
& without
with Indicator, 6 ½
controller/Data
THERMAL- DMM, Negative
47 Logger/recorder, (-)30°C to 50°C 0.21°C
TEMPERATURE Temperature bath
Temperature Gauge,
by Comparison
Liquid in Glass
method
Thermometer,
Digital Thermometer
with inbuilt sensor

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BAGSON CALIBRATION LAB PVT LTD, B-14 DSIDC COMPLEX, PATPARGANJ
Laboratory Name :
INDUSTRIAL AREA, EAST, DELHI, INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-2555 Page No 59 of 59
Validity 14/02/2023 to 13/02/2025 Last Amended on 24/07/2023

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Temperature
Transmitter, RTD,
Thermocouples with
& without
Using RTD sensor
controller/Data
THERMAL- with Indicator, 6 ½
48 Logger/recorder, 50°C to 300°C 0.62°C
TEMPERATURE DMM, Oil bath by
Temperature Gauge,
Comparison method
Liquid in Glass
Thermometer,
Digital Thermometer
with inbuilt sensor
Temperature
Transmitter, Using R Type
Thermocouples with Thermocouple With
THERMAL- & without Indicator, 6½ DMM, 300°C
49 1.5°C
TEMPERATURE controller/Data Dry Block Furnace to 600°C
Logger/recorder, by Comparison
Digital Thermometer Method
with inbuilt sensor
Temperature
Transmitter, Using R Type
Thermocouples with Thermocouple With
THERMAL- & without Indicator, 6½ DMM, 600°C
50 1.88 °C
TEMPERATURE controller/Data Dry Block Furnace to 1200°C
Logger/recorder, by Comparison
Digital Thermometer Method
with inbuilt sensor
* CMCs represent expanded uncertainties expressed at approximately the 95% level of confidence, using a coverage factor of k = 2.

This is annexure to 'Certificate of Accreditation' and does not require any signature.