0% found this document useful (0 votes)
78 views2 pages

Sensor Compatibility Calibration Chart BR en Us 2306

The document provides a comprehensive selection and warranty chart for various Dräger sensors, detailing their measuring ranges, accuracy, resolution, part numbers, and recommended calibration intervals. It includes specific sensors for different chemicals and gases, along with their respective calibration requirements and warranty information. The chart is intended for users to select appropriate sensors based on their needs and ensure proper maintenance and calibration.

Uploaded by

cataflam80
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
78 views2 pages

Sensor Compatibility Calibration Chart BR en Us 2306

The document provides a comprehensive selection and warranty chart for various Dräger sensors, detailing their measuring ranges, accuracy, resolution, part numbers, and recommended calibration intervals. It includes specific sensors for different chemicals and gases, along with their respective calibration requirements and warranty information. The chart is intended for users to select appropriate sensors based on their needs and ensure proper maintenance and calibration.

Uploaded by

cataflam80
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

DrägerSensor® Selection and Warranty Chart

6/7/2023

Pac 60/6500

X-am 5100
X-am 2500
X-am 2800
X-am 5000
X-am 5600
X-am 5800
X-am 3500
X-am 8000
Pac 8000
Pac 8500

Warranty
Bump Test or
Sensor Measuring Recommended Maximum Calibration manually Calibration with
Description Range Accuracy Resolution Part No. Cal Interval Cal Interval or with CC-Vision X-dock Comments*
XS DrägerSensor®
Hydrazine (N2H4) XS Max 0 - 3.00 ppm
N2H4 (hydrazine) 0 - 3.00 ppm
≤±5% of measured value 0.01 ppm 6809190 X 1-Yr 3 Month 6 Months DraegerService N/A Factory calibration
CH3NH-NH2 (monomethyl hydrazine) 0 - 3.00 ppm
(CH3)2N-NH2 (dimethyl hydrazine) 0 - 3.00 ppm

Hydrogen Flouride/Hyrdogen Chloride (HF/HCl) XS Max 0 - 30.0 ppm


HCl (hydrogen chloride) 0 - 30.0 ppm
HNO3 (nitric acid) 0 - 30.0 ppm
HBr (hydrogen bromide) 0 - 30.0 ppm ≤±15% of measured value 0.1 ppm 6809140 X 1-Yr 12 Month 18 Months 10 or 25 ppm HCl N/A SS Regulator, PTFE-lined hose (short as possible)
POCl3 (phosphorous thrchlorideoxide) 0 - 30.0 ppm
PCl3 (phosphorous trichloride 0 - 30.0 ppm
HF (hydrogen fluoride) 0 - 30.0 ppm

Hydrogen Peroxide (H2O2) XS 0 - 20.0 ppm ≤±10% of measured value 0.1 ppm 6809170 X 1-Yr 3 Months 6 Months 10 ppm SO2 N/A
XXS DrägerSensor®
Amines XXS Std 0 - 100 ppm
CH3NH2 (methylamine) 0 - 100 ppm
(CH3)2NH (dimethylamine) 0 - 100 ppm
CH3)3N (trimethylamine) 0 - 100 ppm ≤±5% of measured value 1 ppm 6812545 X X X X 1-Yr 3 Months 6 Months 50 ppm NH3 50 ppm NH3 Use Tygon tubing and Stainless-Steel Regulator
C2H5NH2 (ethylamine) 0 - 100 ppm
(C2H5)2NH diethylamine) 0 - 100 ppm
(C2H5)3N (triethylamine) 0 - 100 ppm
Ammonia (NH3) XXS 0 - 300 PPM ≤±3% of measured value 1 ppm 6810888 X X X X X 1-Yr 3 Months 6 Months 50 ppm NH3 50 ppm NH3 Use Tygon tubing and Stainless-Steel Regulator
Carbon Dioxide (CO2) XXS 0 - 7.0% ≤±20% of measured value 0.1% Vol. 6810889 X X X X X 15-Mo 3 Months 6 Months 2.5% Vol CO2 2.5% Vol CO2
1
Carbon Monoxide (CO) XXS 0 - 2,000 ppm ≤±2% of measured value 2 ppm 6810882 X X X X 3-Yr 6 Months 12 Months 100 ppm CO 100 ppm CO
Carbon Monoxide Low Conc. (CO) XXS LC1 0 - 2,000 ppm ≤±2% of measured value 1 ppm 6813210 X X X X X X X X 3-Yr 6 Months 12 Months 100 ppm CO 100 ppm CO
Carbon Monoxide High Conc. (CO) XXS HC 0 - 10,000 ppm ≤±2% of measured value 5 ppm 6812010 X X X X 1-Yr 3 Months 6 Months 100 ppm CO 100 ppm CO
100 ppm CO 100 ppm CO
CO Hydrogen Compensated (CO-H2 CP) XXS 0 - 2,000 ppm ≤±2% of measured value 1 ppm 6811950 X X X X X 1-Yr 6 Months 12 Months
500 ppm H2 500 ppm H2
H2S: ≤ ± 5 % of measured
0 - 2,000 ppm CO 1 ppm 100 ppm CO 100 ppm CO
CO-LC/H2S-LC Dual Sensor XXS value, CO: ≤ ± 2 % of 6813280 X X X X X 2-Yr 6 Months 12 Months
0 - 100 ppm H2S 0.1 ppm 25 ppm H2S 25 ppm H2S
measured value
0 - 2,000 ppm CO 2 ppm 100 ppm CO 100 ppm CO
CO/H2S Dual Sensor XXS ≤±2% of measured value 6811410 X X X 2-Yr 6 Months 12 Months
0 - 200 ppm H2S 1 ppm 25 ppm H2S 25 ppm H2S
Chlorine (Cl2) XXS Std 0 - 20 ppm
Cl2 (chlorine) 0 - 20 ppm Use Stainless-Steel Regulator and PTFE-lined Tygon
F2 (fluorine) 0 - 20 ppm ≤±2% of measured value 0.1 ppm 6810890 X X X X X 1-Yr 3 Months 6 Months 5 ppm Cl2 5 ppm Cl2 tubing with 2ea. 8324098 X-dock PTFE Hose Connector
Br2 (bromine) 0 - 20 ppm Sets for connection to X-dock.
ClO2 (chlorine dioxide) 0 - 20 ppm
Hydrogen Cyanide (HCN) XXS 0 - 50 ppm ≤±5% of measured value 0.1 ppm 6810887 X X X X X 2-Yr 3 Months 6 Months 10 ppm HCN 10 ppm HCN
Hydrogen Cyanide (HCN) XXS PC 0 - 50 ppm ≤±5% of measured value 0.5 ppm 6813165 X X X X 1-Yr 3 Months 6 Months 10 ppm HCN 10 ppm HCN
Hydrogen High Conc. (H2) XXS HC 0 - 4% Vol. ≤±2% of measured value 0.1% Vol. 6812025 X X X X 1-Yr 3 Months 6 Months 2% Vol (50%LEL) H2 2% Vol (50%LEL) H2 Check on compatibility with CO or CO-H2 sensors
Hydrogen Low Conc. (H2) XXS 0 - 2,000 ppm ≤±1% of measured value 5 ppm 6812370 X X X X 1-Yr 3 Months 6 Months 500 ppm H2 500 ppm H2
Hydrogen Sulfide (H2S) XXS1 0 - 200 ppm ≤±2% of measured value 1 ppm 6810883 X X 3-Yr 6 Months 12 Months 25 ppm H2S 25 ppm H2S
Hydrogen Sulfide High Conc. (H2S) XXS HC 0 - 1,000 ppm ≤±2% of measured value 2 ppm 6812015 X X X X 1-Yr 3 Months 6 Months 100 ppm H2S 100 ppm H2S
Hydrogen Sulfide Low Conc. (H2S) XXS LC1 0 - 100 ppm ≤±2% of measured value 0.1 ppm 6811525 X X X X X X X X 3-Yr 6 Months 12 Months 25 ppm H2S 25 ppm H2S
Nitrogen Dioxide (NO2) XXS 0 - 50 ppm ≤±2% of measured value 0.1 ppm 6810884 X X X X X X X X 1-Yr 3 Months 6 Months 5 or 10 ppm NO2 5 or 10 ppm NO2
Nitrogen Dioxide Low Conc. (NO2) XXS LC 0 - 50 ppm ≤±3% of measured value 0.02 ppm 6812600 X X X X X 1-Yr 3 Months 6 Months 5 ppm NO2 5 ppm NO2
Nitrogen Oxide (NO) XXS 0 - 200 ppm ≤±3% of measured value 1 ppm 6811545 X X X X X X 1-Yr 3 Months 6 Months 25 ppm NO 25 ppm NO
Odorant (mercaptan) XXS Std 0 - 40 ppm
THT (tetrahydrothiophene) 0 - 40 ppm
(CH3)3CSH (tert.-butyl mercaptane) 0 - 40 ppm
C2H5CH(CH3)SH (sec.-butyl mercaptane) 0 - 40 ppm 10 ppm MeM or
≤±3% of measured value 1 ppm 6812535 X X X X 1-Yr 3 Months 6 Months 10 ppm MeM MeM available on X-dock with ≥02.02.11 firmware
CH3SH (methyl mercaptane) 0 - 40 ppm 10 ppm THT
C2H5SH (ethyl mercaptane) 0 - 40 ppm
(CH3)2S (dimethyl sulphide) 0 - 100 ppm
CH3SSCH3 (dimethyldisulphid) 0 - 40 ppm
Organic Vapor (OV) XXS Std 0 - 200 ppm
C2H4O (ethylene oxide) 0 - 200 ppm
C3H6O (propylene oxide) 0 - 200 ppm
C2H4 (ethene / ethylene) 0 - 100 ppm
C3H6 (propene) 0 - 100 ppm
C2H3Cl (vinyl chloride) 0 - 100 ppm 10 ppm ETO or 10 ppm ETO or
CH3OH (methanol) 0 - 200 ppm
≤±5% of measured value 0.5 ppm 6811530 X X X X X 1-Yr 3 Months 6 Months 100 ppm CO or 100 ppm CO or Calibration gas capability controlled by sensor code
CH2CHCHCH2 (butadiene) 0 - 100 ppm
HCHO (formaldehyde) 0 - 100 ppm 25 ppm Ethylene 25 ppm Ethylene
(H3C)2CHOH (iso-propanol) 0 - 300 ppm
C4H8O (tetrahydrofuran) 0 - 200 ppm
C2H3OCH2Cl (1-chloro-2,3-epoxypropane) 0 - 100 ppm
C6H5CHCH2 (styrene) 0 -100 ppm
H2CC(CH3)COOCH3 (methyl methacrylate) 0 - 100 ppm

Organic Vapor A (OV-A) XXS Std 0 - 200 ppm


C2H4O (ethylene oxide) 0 - 200 ppm
2CCHCN (acrylonitrile) 0 - 100 ppm
(CH3)2CCH2 (isobutene) 0 - 300 ppm 10 ppm ETO or 10 ppm ETO or
CH3COOC2H3 (vinylacetate) 0 - 100 ppm ≤±20% of measured value 1 ppm 6811535 X X X X X 1-Yr 3 Months 6 Months 100 ppm CO or 100 ppm CO or Calibration gas capability controlled by sensor code
C2H5OH (ethanol) 0 - 300 ppm 25 ppm Ethylene 25 ppm Ethylene
CH3CHO (acetaldehyde) 0 - 200 ppm
(C2H5)2O (diethyl ether) 0 - 200 ppm
C2H2 (ethine / acetylene) 0 - 100 ppm
O2: ≤ ± 1 % of measured
0 - 25% Vol O2 0.1%Vol ambient air ambient air
O2/CO-LC Dual Sensor XXS value, CO: ≤ ± 2 % of 6813275 X X X X X 2-Yr 6 Months 12 Months ≤19% O2 required for bump test of O2 on X-dock
0 - 2,000 ppm CO 1 ppm 100 ppm CO 100 ppm CO
measured value
O2: ≤ ± 1 % of measured
0 - 25% Vol O2 0.1%Vol ambient air ambient air
O2/H2S-LC value, H2S: ≤ ± 5 % of 6814137 X X X X X 2-Yr 6 Months 12 Months ≤19% O2 required for bump test of O2 on X-dock
0 - 100 ppm H2S 0.1 ppm 25 ppm H2S 25 ppm H2S
measured value

Draeger, Inc.
7256 S Sam Houston Pkwy W
Suite 100
Houston, TX 77085
(346) 802-6000 06/07/2023
DrägerSensor® Selection and Warranty Chart

6/7/2023

Pac 60/6500

X-am 5100
X-am 2500
X-am 2800
X-am 5000
X-am 5600
X-am 5800
X-am 3500
X-am 8000
Pac 8000
Pac 8500

Warranty
Bump Test or
Sensor Measuring Recommended Maximum Calibration manually Calibration with
Description Range Accuracy Resolution Part No. Cal Interval Cal Interval or with CC-Vision X-dock Comments*
Oxygen (O2) XXS1 0 - 25% Vol. ≤±1% of measured value 0.1% Vol. 6810881 X X X X X X X X 3-Yr 6 Months 12 Months ambient air ambient air ≤19% O2 required for bump test of O2 on X-dock
Oxygen (O2) XXS O2 100 0 - 100% Vol. ≤±1% of measured value 0.5% Vol. 6812385 X X X X 1-Yr 3 Months 6 Months ambient air ambient air Calibrate at altitude where being used
≤19% O2 required for bump test of O2 on X-dock
Oxygen (O2) XXS PR 0 - 30% Vol. ≤±1% of measured value 0.1% Vol. 6800530 X X X X X X X 2-Yr 6 Months 12 Months ambient air ambient air
Order via configurator
Ozone (O3) XXS 0 - 10 ppm ≤±3% of measured value 0.01 ppm 6811540 X X X X X 1-Yr 3 Months 6 Months 5 ppm NO2 5 ppm NO2 N2 required for zero calibration
Phosgene (COCl2) XXS 0 - 10 ppm ≤±5% of measured value 0.01 ppm 6812005 X X X X X 1-Yr 3 Months 6 Months 2 to 3 ppm COCl2 2 to 3 ppm COCl2 Calibration gas not sold by Draeger
Phosphine (PH3) XXS Std 0 - 20 ppm
PH3 (phosphine) 0 - 20 ppm
AsH3 (arsine) 0 - 20 ppm ≤±2% of measured value 0.1 ppm 6810866 X X X X X 1-Yr 3 Months 6 Months 0.5 ppm PH3 0.5 ppm PH3
B2H6 (diborane) 0 - 20 ppm
SiH4 (silane) 0 - 20 ppm
Phosphine High Conc. (PH3) XXS HC 0 - 1,000 ppm ≤±2% of measured value 1 ppm 6812020 X X X X 1-Yr 3 Months 6 Months 5 ppm PH3 5 ppm PH3 Not compatible with CatEx sensor
Sulfur Dioxide (SO2) XXS 0 - 100 ppm ≤±2% of measured value 0.1 ppm 6810885 X X X X X X X X 2-Yr 3 Months 6 Months 5 ppm SO2 5 ppm SO2 Do not use H2S/SO2 mixed gas on the X-dock
®
Cat Ex DrägerSensor
0 - 100% LEL 1% LEL
Cat Ex 125 PR 0 - 5.0% Vol. ≤±1% of LEL 0.1% Vol. 6812950 X X X X 3-Yr 3 Months 6 Months 50% LEL CH4 50% LEL CH4
5 - 100% Vol.2/3 1.0% Vol.
0 - 5.0% Vol. 0.1% Vol. "ch4L" mode in X-am 8000 for 50 to 50,000 ppm range
Cat Ex 125 PR Mining/Gas (MSHA Approved) ≤±1% of LEL 6813080 X X X 3-Yr 3 Months 6 Months 2.5% Vol CH4 2.5% Vol CH4
5 - 100% Vol.2/3 1.0% Vol. for Method 21 type testing
0 - 100% LEL 1% LEL
≤ ± 1 % LEL at
CarEx 125 SR (Shock Resistant) 0 - 5.0% Vol. 0.05% Vol 6851900 X X 3-Yr 3 Months 6 Months 2.5% Vol CH4 2.5% Vol CH4
50 % LEL
0 - 100% Vol.4 0.1% Vol
IR DrägerSensor®
≤±1.5% LEL methane at
0 - 100%LEL 1% LEL 50% LEL CH4 50% LEL CH4
50% LEL X-am 8000 Only - Must be installed in HPP 1 - No PID
Dual IR-Ex/CO2 HC 0 - 100% Vol. 0.1% Vol. 6800276 X 5-Yr 6 Months 12 Months 50% Vol CH4 50% Vol CH4
≤±0.5 Vol.-% CO2 at sensor can be installed at the same time.
0 - 100% Vol. CO2 0.1% Vol. 50% Vol CO2 50% Vol CO2
50 Vol.-%
≤±1.5% LEL methane at
0 - 100%LEL 1% LEL 50% LEL CH4 50% LEL CH4 N2 required for CO2 zero calibration
50% LEL
Dual IR-Ex/CO2 ES (replaces 6811960) 0 - 100% Vol. 0.1% Vol. 6851880 X X 5-Yr 6 Months 12 Months 50% Vol CH4 50% Vol CH4 0 - 9,900 ppm in X-am 5600
≤±0.08 Vol.-% CO2 at 2.5
0 - 5.00% Vol. CO2 0.02% Vol. 2.5% Vol CO2 2.5% Vol CO2 0 - 50,000 ppm range in X-am 8000
Vol.-%
0 - 100% LEL ≤±1.5% LEL methane at 1% LEL
IR-Ex ES (replaces 6812180) 6851881 X X 5-Yr 6 Months 12 Months 50% LEL CH4 50% LEL CH4 Can be configured for 2 ranges/gases via CC-Vision
0 - 100% Vol. 50% LEL 0.1% Vol.
0 - 9,900 ppm ≤±0.08 Vol.-% CO2 at 2.5 50 ppm N2 required for CO2 zero calibration
IR-CO2 ES (replaces 6812190) 6851882 X X 5-Yr 6 Months 12 Months 2.5% Vol CO2 2.5% Vol CO2
0 - 5.00% Vol. Vol.-% 0.02% Vol. 0 - 50,000 ppm range in X-am 8000
PID DrägerSensor®
0-20 ppm - 100 ppb
at 100 ppm isobutylene: >20 ppm - 200 ppb
≤± 2% of measured value >50 ppm - 500 ppb
Use with benzene scrubber tube for benzene specific
PID HC 0.3 - 2000 ppm >100 ppm - 1 ppm 6813475 X 1-Yr 1 Month 12 Months 100 ppm isobutylene 100 ppm isobutylene
readings up to 100 ppm (01/2021).
at zero point: >200 ppm - 2 ppm
≤± 0.3 ppm isobutylene >500 ppm - 5 ppm
>1000 - 10 ppm
at 5 ppm isobutylene:
Requires zero air or N2 for zero calibration.
≤± 2% of measured value 0-2 ppm - 10 ppb
0-10 ppm isobutylene 30 minute warm up time, if not stored on charger.
PID LC >2 ppm - 20 ppb 6813500 X 1-Yr Each use N/A 5 ppm isobutylene 5 ppm isobutylene
0.025-5 ppm benzene Use with benzene scrubber tube for benzene specific
at zero point: >5 ppm - 50 ppb
readings up to 100 ppm (01/2021).
≤± 0.05 ppm isobutylene

* Comments - FKM tubing is recommended for all sensors with the exception of chlorine, nitrogen dioxide and ammonia. For chlorine use the Teflon-lined Tygon tubing and for nitrogen dioxide and ammonia use the Tygon tubing.
1 - From December 1st, 2018, 5 year warranty when sold as installed in an X-am multi-gas monitor. 3 year for replacement sensors.
2- 0-100%Vol range in X-am 5000 only when selected
3- Full Range mode in X-am 8000
4- Full Range mode in X-am 5800

Draeger, Inc.
7256 S Sam Houston Pkwy W
Suite 100
Houston, TX 77085
(346) 802-6000 06/07/2023