Sensor Compatibility Calibration Chart BR en Us 2306
Sensor Compatibility Calibration Chart BR en Us 2306
6/7/2023
Pac 60/6500
X-am 5100
X-am 2500
X-am 2800
X-am 5000
X-am 5600
X-am 5800
X-am 3500
X-am 8000
Pac 8000
Pac 8500
Warranty
Bump Test or
Sensor Measuring Recommended Maximum Calibration manually Calibration with
Description Range Accuracy Resolution Part No. Cal Interval Cal Interval or with CC-Vision X-dock Comments*
XS DrägerSensor®
Hydrazine (N2H4) XS Max 0 - 3.00 ppm
N2H4 (hydrazine) 0 - 3.00 ppm
≤±5% of measured value 0.01 ppm 6809190 X 1-Yr 3 Month 6 Months DraegerService N/A Factory calibration
CH3NH-NH2 (monomethyl hydrazine) 0 - 3.00 ppm
(CH3)2N-NH2 (dimethyl hydrazine) 0 - 3.00 ppm
Hydrogen Peroxide (H2O2) XS 0 - 20.0 ppm ≤±10% of measured value 0.1 ppm 6809170 X 1-Yr 3 Months 6 Months 10 ppm SO2 N/A
XXS DrägerSensor®
Amines XXS Std 0 - 100 ppm
CH3NH2 (methylamine) 0 - 100 ppm
(CH3)2NH (dimethylamine) 0 - 100 ppm
CH3)3N (trimethylamine) 0 - 100 ppm ≤±5% of measured value 1 ppm 6812545 X X X X 1-Yr 3 Months 6 Months 50 ppm NH3 50 ppm NH3 Use Tygon tubing and Stainless-Steel Regulator
C2H5NH2 (ethylamine) 0 - 100 ppm
(C2H5)2NH diethylamine) 0 - 100 ppm
(C2H5)3N (triethylamine) 0 - 100 ppm
Ammonia (NH3) XXS 0 - 300 PPM ≤±3% of measured value 1 ppm 6810888 X X X X X 1-Yr 3 Months 6 Months 50 ppm NH3 50 ppm NH3 Use Tygon tubing and Stainless-Steel Regulator
Carbon Dioxide (CO2) XXS 0 - 7.0% ≤±20% of measured value 0.1% Vol. 6810889 X X X X X 15-Mo 3 Months 6 Months 2.5% Vol CO2 2.5% Vol CO2
1
Carbon Monoxide (CO) XXS 0 - 2,000 ppm ≤±2% of measured value 2 ppm 6810882 X X X X 3-Yr 6 Months 12 Months 100 ppm CO 100 ppm CO
Carbon Monoxide Low Conc. (CO) XXS LC1 0 - 2,000 ppm ≤±2% of measured value 1 ppm 6813210 X X X X X X X X 3-Yr 6 Months 12 Months 100 ppm CO 100 ppm CO
Carbon Monoxide High Conc. (CO) XXS HC 0 - 10,000 ppm ≤±2% of measured value 5 ppm 6812010 X X X X 1-Yr 3 Months 6 Months 100 ppm CO 100 ppm CO
100 ppm CO 100 ppm CO
CO Hydrogen Compensated (CO-H2 CP) XXS 0 - 2,000 ppm ≤±2% of measured value 1 ppm 6811950 X X X X X 1-Yr 6 Months 12 Months
500 ppm H2 500 ppm H2
H2S: ≤ ± 5 % of measured
0 - 2,000 ppm CO 1 ppm 100 ppm CO 100 ppm CO
CO-LC/H2S-LC Dual Sensor XXS value, CO: ≤ ± 2 % of 6813280 X X X X X 2-Yr 6 Months 12 Months
0 - 100 ppm H2S 0.1 ppm 25 ppm H2S 25 ppm H2S
measured value
0 - 2,000 ppm CO 2 ppm 100 ppm CO 100 ppm CO
CO/H2S Dual Sensor XXS ≤±2% of measured value 6811410 X X X 2-Yr 6 Months 12 Months
0 - 200 ppm H2S 1 ppm 25 ppm H2S 25 ppm H2S
Chlorine (Cl2) XXS Std 0 - 20 ppm
Cl2 (chlorine) 0 - 20 ppm Use Stainless-Steel Regulator and PTFE-lined Tygon
F2 (fluorine) 0 - 20 ppm ≤±2% of measured value 0.1 ppm 6810890 X X X X X 1-Yr 3 Months 6 Months 5 ppm Cl2 5 ppm Cl2 tubing with 2ea. 8324098 X-dock PTFE Hose Connector
Br2 (bromine) 0 - 20 ppm Sets for connection to X-dock.
ClO2 (chlorine dioxide) 0 - 20 ppm
Hydrogen Cyanide (HCN) XXS 0 - 50 ppm ≤±5% of measured value 0.1 ppm 6810887 X X X X X 2-Yr 3 Months 6 Months 10 ppm HCN 10 ppm HCN
Hydrogen Cyanide (HCN) XXS PC 0 - 50 ppm ≤±5% of measured value 0.5 ppm 6813165 X X X X 1-Yr 3 Months 6 Months 10 ppm HCN 10 ppm HCN
Hydrogen High Conc. (H2) XXS HC 0 - 4% Vol. ≤±2% of measured value 0.1% Vol. 6812025 X X X X 1-Yr 3 Months 6 Months 2% Vol (50%LEL) H2 2% Vol (50%LEL) H2 Check on compatibility with CO or CO-H2 sensors
Hydrogen Low Conc. (H2) XXS 0 - 2,000 ppm ≤±1% of measured value 5 ppm 6812370 X X X X 1-Yr 3 Months 6 Months 500 ppm H2 500 ppm H2
Hydrogen Sulfide (H2S) XXS1 0 - 200 ppm ≤±2% of measured value 1 ppm 6810883 X X 3-Yr 6 Months 12 Months 25 ppm H2S 25 ppm H2S
Hydrogen Sulfide High Conc. (H2S) XXS HC 0 - 1,000 ppm ≤±2% of measured value 2 ppm 6812015 X X X X 1-Yr 3 Months 6 Months 100 ppm H2S 100 ppm H2S
Hydrogen Sulfide Low Conc. (H2S) XXS LC1 0 - 100 ppm ≤±2% of measured value 0.1 ppm 6811525 X X X X X X X X 3-Yr 6 Months 12 Months 25 ppm H2S 25 ppm H2S
Nitrogen Dioxide (NO2) XXS 0 - 50 ppm ≤±2% of measured value 0.1 ppm 6810884 X X X X X X X X 1-Yr 3 Months 6 Months 5 or 10 ppm NO2 5 or 10 ppm NO2
Nitrogen Dioxide Low Conc. (NO2) XXS LC 0 - 50 ppm ≤±3% of measured value 0.02 ppm 6812600 X X X X X 1-Yr 3 Months 6 Months 5 ppm NO2 5 ppm NO2
Nitrogen Oxide (NO) XXS 0 - 200 ppm ≤±3% of measured value 1 ppm 6811545 X X X X X X 1-Yr 3 Months 6 Months 25 ppm NO 25 ppm NO
Odorant (mercaptan) XXS Std 0 - 40 ppm
THT (tetrahydrothiophene) 0 - 40 ppm
(CH3)3CSH (tert.-butyl mercaptane) 0 - 40 ppm
C2H5CH(CH3)SH (sec.-butyl mercaptane) 0 - 40 ppm 10 ppm MeM or
≤±3% of measured value 1 ppm 6812535 X X X X 1-Yr 3 Months 6 Months 10 ppm MeM MeM available on X-dock with ≥02.02.11 firmware
CH3SH (methyl mercaptane) 0 - 40 ppm 10 ppm THT
C2H5SH (ethyl mercaptane) 0 - 40 ppm
(CH3)2S (dimethyl sulphide) 0 - 100 ppm
CH3SSCH3 (dimethyldisulphid) 0 - 40 ppm
Organic Vapor (OV) XXS Std 0 - 200 ppm
C2H4O (ethylene oxide) 0 - 200 ppm
C3H6O (propylene oxide) 0 - 200 ppm
C2H4 (ethene / ethylene) 0 - 100 ppm
C3H6 (propene) 0 - 100 ppm
C2H3Cl (vinyl chloride) 0 - 100 ppm 10 ppm ETO or 10 ppm ETO or
CH3OH (methanol) 0 - 200 ppm
≤±5% of measured value 0.5 ppm 6811530 X X X X X 1-Yr 3 Months 6 Months 100 ppm CO or 100 ppm CO or Calibration gas capability controlled by sensor code
CH2CHCHCH2 (butadiene) 0 - 100 ppm
HCHO (formaldehyde) 0 - 100 ppm 25 ppm Ethylene 25 ppm Ethylene
(H3C)2CHOH (iso-propanol) 0 - 300 ppm
C4H8O (tetrahydrofuran) 0 - 200 ppm
C2H3OCH2Cl (1-chloro-2,3-epoxypropane) 0 - 100 ppm
C6H5CHCH2 (styrene) 0 -100 ppm
H2CC(CH3)COOCH3 (methyl methacrylate) 0 - 100 ppm
Draeger, Inc.
7256 S Sam Houston Pkwy W
Suite 100
Houston, TX 77085
(346) 802-6000 06/07/2023
DrägerSensor® Selection and Warranty Chart
6/7/2023
Pac 60/6500
X-am 5100
X-am 2500
X-am 2800
X-am 5000
X-am 5600
X-am 5800
X-am 3500
X-am 8000
Pac 8000
Pac 8500
Warranty
Bump Test or
Sensor Measuring Recommended Maximum Calibration manually Calibration with
Description Range Accuracy Resolution Part No. Cal Interval Cal Interval or with CC-Vision X-dock Comments*
Oxygen (O2) XXS1 0 - 25% Vol. ≤±1% of measured value 0.1% Vol. 6810881 X X X X X X X X 3-Yr 6 Months 12 Months ambient air ambient air ≤19% O2 required for bump test of O2 on X-dock
Oxygen (O2) XXS O2 100 0 - 100% Vol. ≤±1% of measured value 0.5% Vol. 6812385 X X X X 1-Yr 3 Months 6 Months ambient air ambient air Calibrate at altitude where being used
≤19% O2 required for bump test of O2 on X-dock
Oxygen (O2) XXS PR 0 - 30% Vol. ≤±1% of measured value 0.1% Vol. 6800530 X X X X X X X 2-Yr 6 Months 12 Months ambient air ambient air
Order via configurator
Ozone (O3) XXS 0 - 10 ppm ≤±3% of measured value 0.01 ppm 6811540 X X X X X 1-Yr 3 Months 6 Months 5 ppm NO2 5 ppm NO2 N2 required for zero calibration
Phosgene (COCl2) XXS 0 - 10 ppm ≤±5% of measured value 0.01 ppm 6812005 X X X X X 1-Yr 3 Months 6 Months 2 to 3 ppm COCl2 2 to 3 ppm COCl2 Calibration gas not sold by Draeger
Phosphine (PH3) XXS Std 0 - 20 ppm
PH3 (phosphine) 0 - 20 ppm
AsH3 (arsine) 0 - 20 ppm ≤±2% of measured value 0.1 ppm 6810866 X X X X X 1-Yr 3 Months 6 Months 0.5 ppm PH3 0.5 ppm PH3
B2H6 (diborane) 0 - 20 ppm
SiH4 (silane) 0 - 20 ppm
Phosphine High Conc. (PH3) XXS HC 0 - 1,000 ppm ≤±2% of measured value 1 ppm 6812020 X X X X 1-Yr 3 Months 6 Months 5 ppm PH3 5 ppm PH3 Not compatible with CatEx sensor
Sulfur Dioxide (SO2) XXS 0 - 100 ppm ≤±2% of measured value 0.1 ppm 6810885 X X X X X X X X 2-Yr 3 Months 6 Months 5 ppm SO2 5 ppm SO2 Do not use H2S/SO2 mixed gas on the X-dock
®
Cat Ex DrägerSensor
0 - 100% LEL 1% LEL
Cat Ex 125 PR 0 - 5.0% Vol. ≤±1% of LEL 0.1% Vol. 6812950 X X X X 3-Yr 3 Months 6 Months 50% LEL CH4 50% LEL CH4
5 - 100% Vol.2/3 1.0% Vol.
0 - 5.0% Vol. 0.1% Vol. "ch4L" mode in X-am 8000 for 50 to 50,000 ppm range
Cat Ex 125 PR Mining/Gas (MSHA Approved) ≤±1% of LEL 6813080 X X X 3-Yr 3 Months 6 Months 2.5% Vol CH4 2.5% Vol CH4
5 - 100% Vol.2/3 1.0% Vol. for Method 21 type testing
0 - 100% LEL 1% LEL
≤ ± 1 % LEL at
CarEx 125 SR (Shock Resistant) 0 - 5.0% Vol. 0.05% Vol 6851900 X X 3-Yr 3 Months 6 Months 2.5% Vol CH4 2.5% Vol CH4
50 % LEL
0 - 100% Vol.4 0.1% Vol
IR DrägerSensor®
≤±1.5% LEL methane at
0 - 100%LEL 1% LEL 50% LEL CH4 50% LEL CH4
50% LEL X-am 8000 Only - Must be installed in HPP 1 - No PID
Dual IR-Ex/CO2 HC 0 - 100% Vol. 0.1% Vol. 6800276 X 5-Yr 6 Months 12 Months 50% Vol CH4 50% Vol CH4
≤±0.5 Vol.-% CO2 at sensor can be installed at the same time.
0 - 100% Vol. CO2 0.1% Vol. 50% Vol CO2 50% Vol CO2
50 Vol.-%
≤±1.5% LEL methane at
0 - 100%LEL 1% LEL 50% LEL CH4 50% LEL CH4 N2 required for CO2 zero calibration
50% LEL
Dual IR-Ex/CO2 ES (replaces 6811960) 0 - 100% Vol. 0.1% Vol. 6851880 X X 5-Yr 6 Months 12 Months 50% Vol CH4 50% Vol CH4 0 - 9,900 ppm in X-am 5600
≤±0.08 Vol.-% CO2 at 2.5
0 - 5.00% Vol. CO2 0.02% Vol. 2.5% Vol CO2 2.5% Vol CO2 0 - 50,000 ppm range in X-am 8000
Vol.-%
0 - 100% LEL ≤±1.5% LEL methane at 1% LEL
IR-Ex ES (replaces 6812180) 6851881 X X 5-Yr 6 Months 12 Months 50% LEL CH4 50% LEL CH4 Can be configured for 2 ranges/gases via CC-Vision
0 - 100% Vol. 50% LEL 0.1% Vol.
0 - 9,900 ppm ≤±0.08 Vol.-% CO2 at 2.5 50 ppm N2 required for CO2 zero calibration
IR-CO2 ES (replaces 6812190) 6851882 X X 5-Yr 6 Months 12 Months 2.5% Vol CO2 2.5% Vol CO2
0 - 5.00% Vol. Vol.-% 0.02% Vol. 0 - 50,000 ppm range in X-am 8000
PID DrägerSensor®
0-20 ppm - 100 ppb
at 100 ppm isobutylene: >20 ppm - 200 ppb
≤± 2% of measured value >50 ppm - 500 ppb
Use with benzene scrubber tube for benzene specific
PID HC 0.3 - 2000 ppm >100 ppm - 1 ppm 6813475 X 1-Yr 1 Month 12 Months 100 ppm isobutylene 100 ppm isobutylene
readings up to 100 ppm (01/2021).
at zero point: >200 ppm - 2 ppm
≤± 0.3 ppm isobutylene >500 ppm - 5 ppm
>1000 - 10 ppm
at 5 ppm isobutylene:
Requires zero air or N2 for zero calibration.
≤± 2% of measured value 0-2 ppm - 10 ppb
0-10 ppm isobutylene 30 minute warm up time, if not stored on charger.
PID LC >2 ppm - 20 ppb 6813500 X 1-Yr Each use N/A 5 ppm isobutylene 5 ppm isobutylene
0.025-5 ppm benzene Use with benzene scrubber tube for benzene specific
at zero point: >5 ppm - 50 ppb
readings up to 100 ppm (01/2021).
≤± 0.05 ppm isobutylene
* Comments - FKM tubing is recommended for all sensors with the exception of chlorine, nitrogen dioxide and ammonia. For chlorine use the Teflon-lined Tygon tubing and for nitrogen dioxide and ammonia use the Tygon tubing.
1 - From December 1st, 2018, 5 year warranty when sold as installed in an X-am multi-gas monitor. 3 year for replacement sensors.
2- 0-100%Vol range in X-am 5000 only when selected
3- Full Range mode in X-am 8000
4- Full Range mode in X-am 5800
Draeger, Inc.
7256 S Sam Houston Pkwy W
Suite 100
Houston, TX 77085
(346) 802-6000 06/07/2023