GSC53N Rel. 1.
00 del 27/07/04
Integrated meter for test verify and Power Quality analysis Pag 1 of 7
1. ELECTRICAL SPECIFICATIONS – VERIFY TESTS
Accuracy is indicated as ± (% readings + no. of digits) at 23°C ± 5°C, con relative humidity <60%UR.
Continuity test on protective and equalizing conductors
Range (Ω) Resolution (Ω) Accuracy (*)
0.01 ÷ 9.99 0.01
±(2.0%rdg + 2dgt)
10.0 ÷ 99.9 0.1
(*) after cable calibration (which eliminates the cable resistance).
Test current: > 200mA DC for R≤5Ω (included calibration)
Resolution on current measurement: 1mA
Open-circuit voltage: 4V ≤ V0 ≤ 24V
Insulation Resistance (DC voltage)
Test voltage(V) Range (MΩ) Resolution (MΩ) Accuracy
0.01 ÷ 9.99 0.01
±(2.0%rdg + 2dgt)
50 10.0 ÷ 49.9 0.1
50.0 ÷ 99.9 0.1 ±(5.0%rdg + 2dgt)
0.01 ÷ 9.99 0.01
±(2.0%rdg + 2dgt)
100 10.0 ÷ 99.9 0.1
100.0 ÷ 199.9 0.1 ±(5.0%rdg + 2dgt)
0.01 ÷ 9.99 0.01
10.0 ÷ 199.9 0.1 ±(2.0%rdg + 2dgt)
250
200 ÷ 249 1
250 ÷ 499 1 ±(5.0%rdg + 2dgt)
0.01 ÷ 9.99 0.01
10.0 ÷ 199.9 0.1 ±(2.0%rdg + 2dgt)
500
200 ÷ 499 1
500 ÷ 999 1 ±(5.0%rdg + 2dgt)
0.01 ÷ 9.99 0.01
10.0 ÷ 199.9 0.1 ±(2.0%rdg + 2dgt)
1000
200 ÷ 999 1
1000 ÷ 1999 1 ±(5.0%rdg + 2dgt)
Open-circuit voltage: <1.3 x nominal test voltage
Short circuit current: <6.0mA at 500V test voltage
ominal test current: >2.2mA on 230kΩ load (500V); >1mA su 1kΩ per Vnom (others)
Measurement limits fitted: 0.05, 0.10, 0.23, 0.25, 0.50, 1.00, 100MΩ
RCDs Tripping time
Range (ms) Resolution (ms) Accuracy
½ I∆N, I∆N 1÷999
2 I∆N 1÷200 general
1÷250 selective 1 ±(2.0%rdg + 2dgt)
5 I∆N RCD 1÷ 50 general
1÷160 selective
Nominal trip-out currents: 10mA, 30mA, 100mA, 300mA, 500mA
RCDs type: AC, A, General and Selective
Phase-PE voltage: 100V ÷ 255V
Frequency: 50Hz ± 0.5Hz
HT ITALIA srl Tel: +39-0546-621002 - Fax: +39-0546-621144
Via Righi 126 - 48018 Faenza (RA)- Italy email:
[email protected] - web: http://www.htitalia.com
GSC53N Rel. 1.00 del 27/07/04
Integrated meter for test verify and Power Quality analysis Pag 2 of 7
Tripping current of RCDs
RCD type I∆N Range I∆N (mA) Resolution (mA) Accuracy I∆N
AC (0.5 ÷ 1.4) I∆N
I∆N ≤ 10mA 0.1 I∆N -0%, +(5.0% I∆N)
A (0.5 ÷ 2.4) I∆N
AC (0.5 ÷ 1.4) I∆N
I∆N > 10mA 0.1 I∆N -0%, +(5.0% I∆N)
A (0.5 ÷ 2.0) I∆N
Contact voltage Ut
Range (V) Resolution (V) Accuracy
0 ÷ 2Utlim 0.1 -0%, +(5.0% rdg + 3dgt)
Utlim (Ul): 25V , 50V
Line Impedance (Phase-Phase, Phase-Neutral)
Range (Ω) Resolution (Ω) Accuracy (*)
0.01 ÷ 9.99 0.01
±(5.0% rdg + 3dgt)
10.0 ÷ 199.9 0.1
(*) 0.1 mΩ on range 0.0 ÷ 199.9 mΩ (with IMP57 optional accessory)
Maximum peak current: 3.65A (at 127V); 6.64A (at 230V); 11.5A (at 400V)
Test voltage: 100÷265V (Phase-Neutral) / 100÷460V (Phase-Phase); 50Hz ± 0.5Hz
Fault Loop Impedance (Phase-Ground)
Range (Ω) Resolution (Ω) Accuracy (*)
0.01 ÷ 19.99 0.01
20.0 ÷ 199.9 0.1 ±(5.0% rdg + 3dgt)
200 ÷ 1999 1
(*) 0.1 mΩ on range 0.0 ÷ 199.9 mΩ (with IMP57 optional accessory)
Maximum peak current: 3.65A (at 127V); 6.64A (at 230V)
Test voltage: 100÷265V (Phase-Ground); 50Hz ± 0.5Hz
Fault Loop Resistance RA without RCDs tripping
Range (Ω) Resolution (Ω) Accuracy
1 ÷ 1999 1 -0%, +(5.0% rdg + 3dgt)
Test current: 0.5 I∆N set on Ut test
15mA on Ra15mA test
Earth Resistance with rods
Range (Ω) Resolution (Ω) Accuracy (*)
0.01 ÷ 19.99 0.01
20.0 ÷ 199.9 0.1 ±(5.0% rdg + 3dgt)
200 ÷ 1999 1
Test current: <10mA – 77.5Hz
Open-circuit voltage: < 20V rms
Earth resistivity
Range ρ (*) Resolution Accuracy (*)
0.06 ÷ 19.99 Ωm 0.01 Ωm
20.0 ÷ 199.9 Ωm 0.1 Ωm
200 ÷ 1999 Ωm 1 Ωm ±(5.0% rdg + 3dgt)
2.00 ÷ 99.99 kΩm 0.01 kΩm
100.0 ÷ 125.5 kΩm 0.1 kΩm
(*) with distance d=10m
Distance range d: 1 ÷ 10m
Test current: <10mA – 77.5Hz
Open-circuit voltage: < 20V rms
HT ITALIA srl Tel: +39-0546-621002 - Fax: +39-0546-621144
Via Righi 126 - 48018 Faenza (RA)- Italy email:
[email protected] - web: http://www.htitalia.com
GSC53N Rel. 1.00 del 27/07/04
Integrated meter for test verify and Power Quality analysis Pag 3 of 7
Voltage (RCD, LOOP, Phase Sequence)
Range (V) Resolution (V) Accuracy
15 ÷ 460 1 ±(3.0% rdg + 2dgt)
Frequency
Range (Hz) Resolution (Hz) Accuracy
47.0 ÷ 63.6 0.1 ±(0.1% rdg + 1dgt)
HT ITALIA srl Tel: +39-0546-621002 - Fax: +39-0546-621144
Via Righi 126 - 48018 Faenza (RA)- Italy email:
[email protected] - web: http://www.htitalia.com
GSC53N Rel. 1.00 del 27/07/04
Integrated meter for test verify and Power Quality analysis Pag 4 of 7
2. ELECTRICAL SPECIFICATIONS – ANALYZER AND AUX
Accuracy is indicated as ± (% readings + no. of digits) at 23°C ± 5°C, con relative humidity <60%UR.
Voltage – Single phase / Three phase systems (Autorange)
Range (V) Resolution (V) Accuracy Input Impedance
15 ÷ 310 0.2 300 kΩ (Phase-Neutral)
±(0.5% rdg + 2dgt)
310 ÷ 600 0.4 300 kΩ (Phase-Phase)
Voltage Anomalies – Single / Three phase systems (Manual range)
Resolution Resolution Accuracy Accuracy
Range (V)
Voltage (V) Time Voltage Time (ref. 50Hz)
15 ÷ 310 0.2
10ms ±(1.0%rdg + 2dgt) ±10ms
30 ÷ 600 0.4
Input Impedance: 300 kΩ (Phase-Neutral and Phase-Phase)
Current by external clamp transducer – STD
Overload
Range (*) Resolution (mV) Accuracy Input Impedance
protection
0.005 ÷ 0.26V 0.1
±(0.5%rdg + 2dgt) 200kΩ 5V
0.26 ÷ 1V 0.4
(*) Example: by using a clamp whose range is 1000A/1V, the instrument measures currents higher than 5A
Current by external clamp transducer – FlexINT (1000A AC range)
Range (A) Voltage input Resolution Accuracy
10.0 ÷ 19.9 950.0µV ÷ 1.691mV ±(4.0% rdg + 8.5µV)
20.0 ÷ 99.9 1.7mV ÷ 8.491mV 8.5µV ±(1.0% rdg + 8.5µV)
100.0 ÷ 999.9 8.5mV ÷ 84.99mV ±(1.0% rdg + 85µV)
1A = 85µV ; Rinput = 400kΩ
Current by external clamp transducer – FlexINT (3000A AC range)
Range (A) Voltage input Resolution Accuracy
30.0 ÷ 999.9 2.55mV ÷ 84.99mV 8.5µV ±(1.0% rdg +17µV)
1000 ÷ 3000 85.0mV ÷ 255mV 85µV ±(0.5% rdg +85µV)
1A = 85µV ; Rinput = 400kΩ
Power factor (Cosϕ) - Single / Three phase systems
Range (cosϕ) Resolution Accuracy (°)
0.20 ÷ 0.50 1.0
0.50 ÷ 0.80 0.01 0.7
0.80 ÷ 1.00 0.6
Leakage curreny (by optional clamp transducer)
Overload
Range (mA)* Resolution (mA) Accuracy Input Impedance
protection
0.5 ÷ 999.9 0.1 ±(5.0% rdg + 2dgt) 200kΩ 5V
(*) While recording the instrument stores only current values > 5mA with 1mA resolution
Maximum stored value is the peak value calculated with response time of 1ms
HT ITALIA srl Tel: +39-0546-621002 - Fax: +39-0546-621144
Via Righi 126 - 48018 Faenza (RA)- Italy email:
[email protected] - web: http://www.htitalia.com
GSC53N Rel. 1.00 del 27/07/04
Integrated meter for test verify and Power Quality analysis Pag 5 of 7
Power – Single / Three phase systems
Measures type Range Resolution Accuracy
100.0 ÷ 999.9W 0.1W
1.000 ÷ 9.999kW 0.001kW
10.00 ÷ 99.99kW 0.01kW
ACTIVE
100.0 ÷ 999.9kW 0.1kW
POWER 1.000 ÷ 9.999MW 0.001MW
10.00 ÷ 99.99MW 0.01MW
100.0 ÷ 999.9MW 0.1MW
100.0 ÷ 999.9VAR 0.1VAR
1.000 ÷ 9.999kVAR 0.001kVAR
10.00 ÷ 99.99kVAR 0.01kVAR
REACTIVE
100.0 ÷ 999.9kVAR 0.1kVAR
POWER 1.000 ÷ 9.999MVAR 0.001MVAR
10.00 ÷ 99.99MVAR 0.01MVAR
100.0 ÷ 999.9MVAR 0.1MVAR
100.0 ÷ 999.9VA 0.1VA
1.000 ÷ 9.999kVA 0.001kVA
10.00 ÷ 99.99kVA 0.01kVA
APPARENT
100.0 ÷ 999.9kVA 0.1kVA ±(1.0% rdg + 2dgt)
POWER 1.000 ÷ 9.999MVA 0.001MVA
10.00 ÷ 99.99MVA 0.01MVA
100.0 ÷ 999.9MVA 0.1MVA
100.0 ÷ 999.9Wh 0.1Wh
1.000 ÷ 9.999kWh 0.001kWh
ACTIVE 10.00 ÷ 99.99kWh 0.01kWh
ENERGY 100.0 ÷ 999.9kWh 0.1kWh
(Class 2 EN61036) 1.000 ÷ 9.999MWh 0.001MWh
10.00 ÷ 99.99MWh 0.01MWh
100.0 ÷ 999.9MWh 0.1MWh
100.0 ÷ 999.9VARh 0.1VARh
1.000 ÷ 9.999kVARh 0.001kVARh
REACTIVE 10.00 ÷ 99.99kVARh 0.01kVARh
ENERGY 100.0 ÷ 999.9kVARh 0.1kVARh
(Class 3 IEC1268) 1.000 ÷ 9.999MVARh 0.001MVARh
10.00 ÷ 99.99MVARh 0.01MVARh
100.0 ÷ 999.9MVARh 0.1MVARh
Harmonics - Single / Three phase systems
Range Maximum resolution Base accuracy
DC ÷ 25a ±(5.0% rdg + 2dgt)
26a ÷ 33a 0.1V / 0.1 A ±(10% rdg + 2dgt)
34a ÷ 49a ±(15% rdg + 2dgt)
Environmental parameters (AUX function)
Range Resolution Accuracy
-20°C ÷ 80°C 0.1 °C
0 ÷ 100% UR 0.1% UR
0.001Lux ÷ 20.00 Lux (*) 0.001 ÷ 0.02 Lux ±(2.0% rdg + 2 dgt)
0.1 Lux ÷ 2000 Lux (*) 0.1 ÷ 2 Lux
1 Lux ÷ 20 kLux (*) 1 ÷ 20 Lux
(*) Accuracy of HT53 luxmeter probe is according to Class AA
HT ITALIA srl Tel: +39-0546-621002 - Fax: +39-0546-621144
Via Righi 126 - 48018 Faenza (RA)- Italy email:
[email protected] - web: http://www.htitalia.com
GSC53N Rel. 1.00 del 27/07/04
Integrated meter for test verify and Power Quality analysis Pag 6 of 7
3. GENERAL SPECIFICATIONS
SINGLE/THREE PHASE RECORDING:
STORED PARAMETERS:
- Phase and delta voltages
- Phase currents, neutral current
- Phase and total three phase Active, Reactive, Apparent power
- Active energy (Class 2 EN61036), Reactive energy (Class 3 IEC1268)
- Phase and total three phase Power factor Cosϕ
- Voltages, currents harmonics (DC,1,2,…49)
- Voltage anomalies (sags, swells, breaks)
- Predefined recordings (EN50160, Voltage anomalies, Harmonics, Start up, Power & Energy)
- Max selectable parameters: 63 or 3 AUX (Environmental and/or leakage)
- Integrated period: 5 ÷ 3600 sec.
- Recording autonomy: > 30 days with integrated period of 15 minutes
- Memory capacity: 2Mbyte
DISPLAY AND MEMORY:
Features: Dot matrix with backlight
Resolution: 128x128 dots
Visible area: 73x73 mm
Memory: 999 measures
POWER SUPPLY:
Batteries: 6 batteries 1.5V type LR6-AA-AM3-MN 1500
External power supply adapter: Code A0050 (AUX e ANALYZER functions only)
MECHANICAL FEATURES:
Dimensions: 225 (W)x165(L)x105(D) mm
Weight (included batteries): about 1.7kg
WORKING ENVIRONMENTAL CONDITIONS:
Reference temperature: 23°C ± 5°C
Working temperature: 0° ÷ 40°C
Allowed relative humidity: < 80% HR
Storage temperature: -10 ÷ 60°C
Storage humidity: < 80% HR
TEST VERIFIES REFERENCE STANDARDS:
Continuity test: IEC 61557
Insulation resistance: IEC 61557
Earth resistance: IEC 61557
Fault Loop Impedance: IEC 61557
RCDs test: IEC 61557
Insulation test on switchboards: EN60439-1
POWER/ENERGY MEASUREMENTS REFERENCE STANDARDS:
Features of voltage supplied by public utilities: EN50160
Active energy static counters for AC current EN61036 (Class 2)
Reactive energy static counters for AC current IEC1268 (Class 3)
NOISE MEASUREMENTS REFERENCE STANDARDS:
Sound measurements (with HT55 probe): EN60651:1994/A1 type 1
EN60804:1994/A2 type 1
HT ITALIA srl Tel: +39-0546-621002 - Fax: +39-0546-621144
Via Righi 126 - 48018 Faenza (RA)- Italy email:
[email protected] - web: http://www.htitalia.com
GSC53N Rel. 1.00 del 27/07/04
Integrated meter for test verify and Power Quality analysis Pag 7 of 7
GENERAL REFERENCE STANDARDS:
Safety of measuring instruments: EN61010-1 + A2(1997)
Product type standard: IEC61557-1, 2, 3, 4, 5, 6
Insulation: class 2 (double insulation)
Pollution degree: 2
Overvoltage category: CAT II 600V CA / 350V CA (to ground)
CAT III 600V CA / 300V CA (to ground)
Use: internal use; max altitude: 2000m
EMC: EN61326-1 (1998) + A1 (1999)
This instrument complies with the European Directive for CE marking.
HT ITALIA srl Tel: +39-0546-621002 - Fax: +39-0546-621144
Via Righi 126 - 48018 Faenza (RA)- Italy email:
[email protected] - web: http://www.htitalia.com