Planmeca

This document provides installation instructions for the Planmeca ProMax 2D X-ray unit retrofit kit to upgrade to a 3D system. It includes pre-installation requirements, detailed steps for hardware installation, configuration of 3D features, and finalizing the installation process. The manual emphasizes the importance of following the instructions in order and outlines necessary components and software updates.

Uploaded by

melquicedec
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
20 views27 pages

Planmeca

This document provides installation instructions for the Planmeca ProMax 2D X-ray unit retrofit kit to upgrade to a 3D system. It includes pre-installation requirements, detailed steps for hardware installation, configuration of 3D features, and finalizing the installation process. The manual emphasizes the importance of following the instructions in order and outlines necessary components and software updates.

Uploaded by

melquicedec
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 27

PLANMECA PROMAX 2D X-RAY UNIT

3D s / 3D CLASSIC RETROFIT KIT

INSTALLATION INSTRUCTIONS

Publication number 10014908 version 9


Copyright Planmeca
Released 26 June 2017
Table of contents
1 INTRODUCTION ................................................................................................................. 3
1.1 Retrofit kit contents ................................................................................................................. 3
1.2 Associated documentation ...................................................................................................... 3
1.3 Sensors .................................................................................................................................. 4
2 PRE-INSTALLATION REQUIREMENTS ............................................................................... 5
2.1 X-ray unit’s hardware requirements ........................................................................................ 5
2.1.1 Graphical user interface ..................................................................................... 5
2.1.2 Tube head types ................................................................................................ 5
2.1.3 Collimator ........................................................................................................... 5
2.2 3D system requirements ......................................................................................................... 5
2.3 3D network architecture requirements .................................................................................... 6
3 INSTALLING 3D HARDWARE ............................................................................................. 7
3.1 Upgrading X-ray unit software ................................................................................................ 7
3.2 Driving collimator to Show Screws -position ........................................................................... 7
3.3 Removing covers .................................................................................................................... 9
3.3.1 Removing C-arm upper covers .......................................................................... 9
3.3.2 Removing sensor holder covers ....................................................................... 10
3.3.3 Removing tube head covers ............................................................................. 11
3.3.4 Removing C-arm inner cover ........................................................................... 12
3.4 Replacing tube head ............................................................................................................. 13
3.5 Attaching 3D sensor ............................................................................................................. 16
4 CONFIGURING 3D FEATURES AND CONNECTING CABLES ........................................... 18
4.1 Restoring factory default settings ......................................................................................... 18
4.2 Setting unit configuration ...................................................................................................... 19
4.3 Enabling 3D features ............................................................................................................ 21
4.4 Connecting cables ................................................................................................................ 23
5 FINALISING INSTALLATION ............................................................................................. 26
5.1 Running filament definition ................................................................................................... 26
5.2 Installing Planmeca Romexis software ................................................................................. 27
5.3 Configuring network settings ................................................................................................ 27
5.4 Checking software versions .................................................................................................. 27
5.5 Adjustments and calibrations ................................................................................................ 27
5.6 Attach removed covers ......................................................................................................... 27

2 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


1 INTRODUCTION

1 INTRODUCTION
This manual contains instructions required to install a 3D system on a Planmeca ProMax
2D X-ray unit.

NOTE
Read this manual carefully before installation.

NOTE
Perform the installation in the order given in this manual.

1.1 Retrofit kit contents


The contents of the retrofit kit depends on the existing 2D system to be updated and the
chosen sensor type.
• sensor 3D s or 3D Classic
• tube head (if needed)
• collimator
• reconstruction PC (Reco PC)
• Ethernet switch
• 3D positioning lights (if needed)
• other installation accessories

1.2 Associated documentation


In addition to these instructions, the following manuals are needed:
• Planmeca ProMax 2D, 3D s and 3D Classic with ProTouch technical manual
(10033257)
• Planmeca Romexis technical manual (10037884)
• Calibration instructions with Planmeca Device Tool (10031558)
• Planmeca ProMax with ProTouch installation manual (10038029)

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 3


1 INTRODUCTION

1.3 Sensors
There are two versions of the sensor: 3D s and 3D Classic. The 3D Classic sensor has a
larger active area than the 3D s sensor.

3D_sensor.eps

3Ds_sensor.eps

3D Classic sensor 3D s sensor


(active area 13 x 13 cm) (active area 8 x 13 cm)

4 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


2 PRE-INSTALLATION REQUIREMENTS

2 PRE-INSTALLATION REQUIREMENTS
Before starting any installation tasks, check carefully the following pre-installation
requirements.

2.1 X-ray unit’s hardware requirements

2.1.1 Graphical user interface


In these instructions it is assumed that the 2D X-ray unit is originally equipped with the
Planmeca ProTouch touch screen, or the ProTouch touch screen has been installed as a
retrofit to the X-ray unit.
If the 2D X-ray unit is not equipped with the ProTouch touch screen, contact Planmeca
After Sales for instructions on how to proceed with the 3D retrofit installation.

2.1.2 Tube head types


The requirement for changing the tube head depends on the type of the tube head
installed to the X-ray unit, see the table below.

Serial number Tube head type and need for changing


RPPxxxxxx 2D tube head
must be changed
TPPxxxxxx 3D fixed tube head
must be changed if rotating tube head is needed
KPPxxxxxx 2D/3D fixed and rotating tube head
no need to change

For more information, see Planmeca ProMax tube head versions and compatibility
information, Technical Bulletin (PX-076B).

2.1.3 Collimator
The existing 2D collimator must be changed to 3D collimator.

2.2 3D system requirements


Check the latest up-to-date system recommendation from the following www-page:
http://www.planmeca.com/System-requirements

NOTE
The PC(s) and other equipment connected to the system must be:
- approved by local authorities
- located outside the patient area
- protectively grounded

2.3 3D network architecture requirements


In the figure below, there is an example of IP network specific settings used in the
Planmeca ProMax 3D s and 3D Classic networks.
Note the following issues:

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 5


2 PRE-INSTALLATION REQUIREMENTS

• The minimum requirement for the Ethernet cables used in the 3D system network is
Category 6 (CAT6).
• The Romexis PC hardware is not by default included to the Planmeca ProMax 3D s /
3D Classic retrofit kit delivery.
• Each image acquisition workstation (Romexis PC) in the network has to use a
different IP address.
• If connected to any common network, the ProMax network must be protected with a
firewall. The firewall settings must be carefully configured to allow all vital traffic
between X-ray unit, Reco PC and Romexis PC. For more information, refer to the
Romexis Technical manual (10037884). However, all information security issues in
general must be considered together with the clinic’s information security personnel.
• In case of standalone ProMax installation, there is no need for firewall.

ProMax
IP = 10.0.0.1
subnetmask = 255.255.255.0
gateway 0.0.0.0
Sensor head
IP = 10.0.0.2
subnetmask = 255.255.255.0

Romexis PC
Reco PC eth1
eth1 IP = 10.0.0.4
IP = 10.0.0.3 subnetmask = 255.255.255.0
subnetmask = 255.255.255.0 eth0
ProCeph eth0 IP e.g. 192.168.3.xxx
IP = 10.0.0.6 broadcom subnetmask 255.255.252.0
ProTouch
subnetmask = 255.255.255.0
IP = 10.0.0.10
subnetmask = 255.255.255.0

Switch Network

6 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


3 INSTALLING 3D HARDWARE

3 INSTALLING 3D HARDWARE

3.1 Upgrading X-ray unit software


Upgrade the X-ray unit software (CPU PCB and ProTouch) to the latest available version.
Update the software with the Planmeca Device Tool, see instruction in the Calibration
instructions with Planmeca Device Tool (10031558).

3.2 Driving collimator to Show Screws -position


Follow the instructions below to drive the collimator to a position where the collimator
screws are visible.
1. Switch on the X-ray unit.
2. To enter the calibration mode, touch the service spanner icon on the Main display.
3. Select Technical from the bottom of the display. The technical setting mode is
password protected and the password is asked when the mode is entered for the first
time after switching the unit on The password is 1701.

4. From the list on the display that appears select Calibrations (3200).
5. Select Collimator calibration (3260).

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 7


3 INSTALLING 3D HARDWARE

6. Select Collimator Adjustment (3264) and select Show Screws option.

7. Touch the Run button to move the collimator to a position where the collimator screws
are visible.
The collimator will now be in a position where the screws that secure the collimator in
place are visible.
8. Switch off the X-ray unit.

8 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


3 INSTALLING 3D HARDWARE

3.3 Removing covers


Follow the instructions below to remove the following X-ray unit covers:
• C-arm upper covers
• sensor holder covers
• tube head covers
• C-arm inner cover

NOTE
Removing the covers might slightly differ depending on the X-ray unit model (age), for more
detailed instructions, see the original version of the X-ray unit’s Technical manual.

WARNING
Always switch the X-ray unit off before removing any covers.

3.3.1 Removing C-arm upper covers


1. Loosen the eight attachment screws with the 2.5 mm Allen key (1).
Do not remove the screws.
2. Slide the tube head cover plate (2) and the receptor cover plate (3) from their
positions as shown below.

2 8x
Digi_C_arm_covers1.eps

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 9


3 INSTALLING 3D HARDWARE

3.3.2 Removing sensor holder covers


1. Remove the sensor head from its holder.
Push in the button of the C-arm electrical connector. The sensor head connector front
cover is attached to the C-arm with mounting springs and can be removed by pulling it
from its position (1).
Mid_62.eps

2. Unscrew the two screws from the inner side of the sensor head connector back cover
(1) and loosen two from the top (2) with 2.5 mm Allen key.

2
Mid_63_1.eps

1
3. Detach the cover from its position (3).

10 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


3 INSTALLING 3D HARDWARE

3.3.3 Removing tube head covers

NOTE
You have to remove the C-arm upper covers before removing the tube head covers.

1. Unscrew the two attachment screws from the top of the front cover (1) and the two
screws from the bottom of the cover (2). The cover can now be removed.

Digi_C_arm_covers6.eps
1

2. Unscrew the two attachment screws from the top of the back cover (1) and the two
screws from the side of the cover (2). The cover can now be removed.

1
Digi_C_arm_covers5.eps

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 11


3 INSTALLING 3D HARDWARE

3.3.4 Removing C-arm inner cover

NOTE
You have to remove the C-arm upper covers and sensor head covers before removing the
C-arm inner cover.

1. Unscrew the four screws with a 2.5 mm Allen key (1) and remove the cover (2).
Digi_C_arm_covers4.eps

4x

12 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


3 INSTALLING 3D HARDWARE

3.4 Replacing tube head


If the old tube head is suitable for the 3D purposes, there is no need to replace the tube
head, see the tube head hardware requirements in section 2.1.2 "Tube head types" on
page 5.
1. Disconnect all five tube head cables from the C-arm PCB.

Remove_tubehead1.eps
5
4
1
C- 21-1
J2 Ar 0-
m 05
5 PC
B

J2
4

1
2
3

ONLY IN TOMOGRAPHIC
10001173 IMAGING MODE
10002151 121-10-08
121-10-08

Top Center Top Left


Laser Laser
J1

10002151J1

Tube Head
10002431
J1 J2 J3
J4
J6
10002225
J1

2
J5
J8
J7
121-10-06
10002353 Tube Feedback
Receptor Pan / Primary Collimator
1

Primary Collimator / Aux. (J20)


2

121-10-05 PCB
C-Arm Rotation Motor

J9
Receptor Rocation

C-Arm Angle
Sensor
C-Arm PCB
Aux. (J18)

KT60LM06-004
J10
10002180
CAM

3
J11 J20

C-Arm Motor 10002151 FBK


Top Right J1

Laser 121-10-08
J27 00050852
ONLY IN TOMOGRAPHIC CH7 CH8
IMAGING MODE CH4 CH5 CH6 Aux.
J18
J20
Aux.
10003369
J26
J21 J22 J23 J24 J25
J19

1
10002181

4
Y-Collimator Motor
DIMAX2 KT35FM1-010
PAN

10002181

10003699
5 X-Collimator Motor
KT35FM1-010

Primary Collimators

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 13


3 INSTALLING 3D HARDWARE

2. Unscrew the two screws (1) which secure the collimator in position and remove the
collimator (3).
The screws should be visible and you should be able to access them. Manually lift up
the lower blade to access the lower screw.

2
Remove_collimator_for_3D_RetroFit.eps

Lift up lower blade


to access lower screw

NOTE
If you did not drive the collimator to Show Screws -position as instructed in section 3.2
"Driving collimator to Show Screws -position" on page 7, you have to move the
collimator mechanism manually. To do this, rotate the stepper motor spindles (2)
located behind the collimator blades until you can access the screws.

3. Using a 4 mm Allen key, unscrew the four screws (4) which hold the tube head in
position and remove the tube head (5).

14 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


3 INSTALLING 3D HARDWARE

4. Install the new 3D s or 3D Classic tube head in reverse order.


Make sure that the tube head is parallel with the C-arm casting back edge.
5. Attach the 3D collimator to the tube head.
a. Manually rotate the stepper motor spindles located behind the collimator blades
until you can access the two screw holes.
b. Attach the screws using a 2.5 mm Allen key.
c. Then route the cables as shown in the figure below, but do not connect the
cables yet.

3D collimator
with copper filter

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 15


3 INSTALLING 3D HARDWARE

3.5 Attaching 3D sensor

NOTE
If the 3D sensor is equipped with Planmeca ProFace option, do not touch the glass
windows when you hold the sensor. Fingerprints or other stains on the glass surface affects
to the image quality.

1. Push the sensor head onto the connector on the C-arm.


CBCT_s3.eps

C-arm

3D sensor

2. Turn the locking knob over the fastening mechanism. This will secure the sensor head
in position.
CBCT_s2.eps

16 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


3 INSTALLING 3D HARDWARE

3. Push in the button of the C-arm electrical connector on the other side to secure the
sensor head in position. This will make the electrical connection between the sensor
head and C-arm.

Button of the C-arm


electrical connector

cbct_s5.eps
NOTE
The front cover located on the inside of the C-arm must have an opening for the cooling
grille (see below). This will ensure that the sensor cooling system functions properly.
Replace the front cover if necessary. The front cover is secured with Velcro fasteners.

C-arm
3Ds_sensor.eps

Front cover with opening


Cooling grille for cooling grille

3D sensor

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 17


4 CONFIGURING 3D FEATURES AND CONNECTING CABLES

4 CONFIGURING 3D FEATURES AND CONNECTING


CABLES

4.1 Restoring factory default settings


1. Switch on the X-ray unit.

2. Enter the service menus, touch the service spanner icon on the Main display.

3. Select Program from the bottom of the display.

4. From the list on the display that appears select Reset to Factory Defaults (2500).
5. Accept the reset by touching the green check mark button.

18 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


4 CONFIGURING 3D FEATURES AND CONNECTING CABLES

4.2 Setting unit configuration


1. Enter the technical settings, select Technical.
2. The technical setting mode is password protected, the password is 1701.

3. Select 3100 Unit Configuration.

Set the X-ray unit type and the 3D sensor type according to the retrofit kit type:
• Type: Depending on the retrofit kit 3D sensor type, select either: ProMax 3D s or
ProMax 3D Classic
• Sensor Type: Depending on the retrofit kit 3D sensor type, select either: 3D s or
3D v2
4. Accept the changes by touching the green check mark button.
The X-ray unit restarts.
5. Enter the technical settings, select Technical.

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 19


4 CONFIGURING 3D FEATURES AND CONNECTING CABLES

6. The technical setting mode is password protected, the password is 1701.

7. Select 3100 Unit Configuration.

Modify the rest of the unit configuration settings:


• Collimator Type: Select: v2.
• Panoramic: Enable/disable panoramic programs, the following options: ON or
OFF.
• Cephalostat: Enable (ON) / disable (OFF) cephalostat.
• Tube Head Type: Defines the tube head type: Fixed or Rotating.
• Cephalostat Type: Defines the used cephalostat type:
- Scanning for Planmeca ProMax cephalostat.
- ProCeph for Planmeca ProCeph cephalostat.
• Logo: Defines the used logo: Planmeca Default, SCS or Planmeca 3D.
8. Accept the changes by touching the green check mark button.

20 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


4 CONFIGURING 3D FEATURES AND CONNECTING CABLES

4.3 Enabling 3D features


1. Activate and configure the needed 3D programs, select Program --> 2100 Programs.

On these displays (2110 - 2180) you can select the X-ray unit’s programs. These
selections determine which programs are shown on the touch screen when using the
X-ray unit.
In addition, on these displays you can define preset kV and mA values for different
type of exposures, i.e. the selected program, patient size, segmentation and jaw size.
2. Acknowledge the changes by touching the green check mark button.
3. Set the program features, select Program --> 2200 Program Features.

Modify the following settings:


• Panoramic System: Dimax or SmartPan.
• Autofocus: ON/OFF
• Fast Forward: ON/OFF
• 2D Image Shift: This setting is not valid for ProMax 3D s or 3D Classic X-ray
units.
• Lateral TMJ Default Angle, PA TMJ Default Angle: Adjust the angle (lat or PA).
• TMJ 2/2 Shift: In double TMJ exposures the target position will be automatically
moved forwards by 10 mm for the second exposure (open view projection). The
setting can be changed by the user.
• 3D Scout: ON/OFF

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 21


4 CONFIGURING 3D FEATURES AND CONNECTING CABLES

• 2D Views for 3D: ON/OFF.


You can set the X-ray unit so that two 2D images (Scout, Lat-PA, Lat or PA) will
be shown on the computer screen before you take a 3D image. This will allow
you to check the position of the target area or, if necessary, confirm the need for
a 3D exposure.
• Pan DEC Density, Ceph DEC Density Lat and Ceph DEC Density PA/AP:
DEC density is expressed in percentage in comparison to DEC calibration value.
The setting can be adjusted between 20% (lower exposure values --> brighter
image) and 200% (higher exposure values --> darker image). The recommended
setting is 100% (default setting).

NOTE
DEC is available for panoramic and cephalometric programs. It is not available for
SmartPan programs.

4. Acknowledge the changes by touching the green check mark button.


5. If additional features are taken into use, activate those licenses. Select 2300
Licences.
Note that, the basic 3D features are available without extra licences.

On the Licences display you can change and enable/disable the program licences.
You will need a license code to add a new feature to the X-ray unit. Touch the feature
field to switch this feature on. A number display will appear. Enter the license code to
enable the selected feature.
6. Acknowledge the changes by touching the green check mark button.
7. Switch off the X-ray unit.

22 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


4 CONFIGURING 3D FEATURES AND CONNECTING CABLES

4.4 Connecting cables

NOTE
Make sure that the X-ray unit is switched off.

1. Connect all five tube head cables from the C-arm PCB.

Remove_tubehead1.eps
5
4
1
C- 21-1
J2 Ar 0-
m 05
5 PC
B

J2
4

1
2
3

ONLY IN TOMOGRAPHIC
10001173 IMAGING MODE
10002151 121-10-08
121-10-08

Top Center Top Left


Laser Laser
J1

10002151J1

Tube Head
10002431
J1 J2 J3
J4
J6
10002225
J1

2
J5
J8
J7
121-10-06
10002353 Tube Feedback
Receptor Pan / Primary Collimator
1

Primary Collimator / Aux. (J20)


2

121-10-05 PCB
C-Arm Rotation Motor

J9
Receptor Rocation

C-Arm Angle
Sensor
C-Arm PCB
Aux. (J18)

KT60LM06-004
J10
10002180
CAM

3
J11 J20

C-Arm Motor 10002151 FBK


Top Right J1

Laser 121-10-08
J27 00050852
ONLY IN TOMOGRAPHIC CH7 CH8
IMAGING MODE CH4 CH5 CH6 Aux.
J18
J20
Aux.
10003369
J26
J21 J22 J23 J24 J25
J19

1
10002181

4
Y-Collimator Motor
DIMAX2 KT35FM1-010
PAN

10002181

10003699
5 X-Collimator Motor
KT35FM1-010

Primary Collimators

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 23


4 CONFIGURING 3D FEATURES AND CONNECTING CABLES

2. Disconnect old Ethernet cable from the X-ray unit.


3. Connect the Ethernet cables between X-ray unit, Ethernet switch, Reco PC and
Romexis PC.

NOTE
The minimum requirement for the Ethernet cables used in the 3D system network is
Category 6 (CAT6).
X-ray unit

Reco PC Romexis PC

A
B

Ethernet switch
PoE In 10/100/1000 Base-T Ports (1-8) - Ports are Auto - MDIX
1 2 3 4 5 6 7 8

a. Use an Ethernet cable (A) to connect the X-ray unit to the Ethernet switch.

The Ethernet cable is attached to the left terminal (marked “ETHERNET”) on the
underside of the stationary column top.

CAUTION!
Wrong connection
can cause damage!
Make sure that you
connect the cable to
the connector
marked “ETHERNET”.
exp_sw_input_sw.eps

24 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


4 CONFIGURING 3D FEATURES AND CONNECTING CABLES

b. Use an Ethernet cable (B) to connect the reconstruction PC to the Ethernet


switch.

Reconstruction PC

Terminal for
Ethernet cable
(Reconstruction PC -
Ethernet switch)

c. Use an Ethernet cable (C) to connect the Romexis PC to the Ethernet switch.
4. Switch on the X-ray unit.

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 25


5 FINALISING INSTALLATION

5 FINALISING INSTALLATION

5.1 Running filament definition


The X-ray tube filament definition must be performed whenever the tube head and/or the
Power supply PCB is replaced.
1. Switch on the X-ray unit.
2. Enter the technical settings, select Technical. The technical setting mode is password
protected, the password is 1701.

3. Select 3200 Calibrations --> 3210 Filament Definition.


4. Press and hold down the exposure switch until the series of exposures ends.

26 Planmeca ProMax 3D s / 3D Classic retrofit kit Installation instructions


5 FINALISING INSTALLATION

5.2 Installing Planmeca Romexis software


Install the Planmeca Romexis software according to the instructions given in the Romexis
technical manual (publication number 10037884).

5.3 Configuring network settings


The default IP addresses of the Planmeca ProMax 3D system is presented in section 2.3
"3D network architecture requirements" on page 5. It is recommended not to change the
default IP adresses.
In some cases, the Reco PC IP address needs to be changed because of the existing
network configuration. If the Reco PC IP address needs to be changed, see instructions
given in the Planmeca ProMax with ProTouch, installation manual (10038029).

5.4 Checking software versions


Check from the Planmeca Romexis bulletin that the installed Planmeca ProMax X-ray unit,
Reconstructions PC (RecoPC) and 3D sensor software versions are on the correct level,
and update if necessary.
Use Planmeca Device Tool to check and update the software versions.

5.5 Adjustments and calibrations


Perform the adjustments and calibrations according to the instructions given in the
Planmeca ProMax, 2D, 3D s and 3D Classic with ProTouch, Technical manual
(10033257) in section C-ARM AND IMAGING ARM, and in the Calibration instructions
with Planmeca Device Tool (10031558).

5.6 Attach removed covers


Attach all removed covers, see instructions (in reverse order) in section 3.3 "Removing
covers" on page 9, and/or refer to the Planmeca ProMax, 2D, 3D s and 3D Classic with
ProTouch, Technical manual (10033257).

Installation instructions Planmeca ProMax 3D s / 3D Classic retrofit kit 27