INSTALLATION INSTRUCTIONS
1 INTRODUCTION
This manual contains instructions required to install a 3D system on a Planmeca ProMax
2D X-ray unit.
NOTE
Read this manual carefully before installation.
NOTE
Perform the installation in the order given in this manual.
1.3 Sensors
There are two versions of the sensor: 3D s and 3D Classic. The 3D Classic sensor has a
larger active area than the 3D s sensor.
3D_sensor.eps
3Ds_sensor.eps
2 PRE-INSTALLATION REQUIREMENTS
Before starting any installation tasks, check carefully the following pre-installation
requirements.
For more information, see Planmeca ProMax tube head versions and compatibility
information, Technical Bulletin (PX-076B).
2.1.3 Collimator
The existing 2D collimator must be changed to 3D collimator.
NOTE
The PC(s) and other equipment connected to the system must be:
- approved by local authorities
- located outside the patient area
- protectively grounded
• The minimum requirement for the Ethernet cables used in the 3D system network is
Category 6 (CAT6).
• The Romexis PC hardware is not by default included to the Planmeca ProMax 3D s /
3D Classic retrofit kit delivery.
• Each image acquisition workstation (Romexis PC) in the network has to use a
different IP address.
• If connected to any common network, the ProMax network must be protected with a
firewall. The firewall settings must be carefully configured to allow all vital traffic
between X-ray unit, Reco PC and Romexis PC. For more information, refer to the
Romexis Technical manual (10037884). However, all information security issues in
general must be considered together with the clinic’s information security personnel.
• In case of standalone ProMax installation, there is no need for firewall.
ProMax
IP = 10.0.0.1
subnetmask = 255.255.255.0
gateway 0.0.0.0
Sensor head
IP = 10.0.0.2
subnetmask = 255.255.255.0
Romexis PC
Reco PC eth1
eth1 IP = 10.0.0.4
IP = 10.0.0.3 subnetmask = 255.255.255.0
subnetmask = 255.255.255.0 eth0
ProCeph eth0 IP e.g. 192.168.3.xxx
IP = 10.0.0.6 broadcom subnetmask 255.255.252.0
ProTouch
subnetmask = 255.255.255.0
IP = 10.0.0.10
subnetmask = 255.255.255.0
Switch Network
3 INSTALLING 3D HARDWARE
4. From the list on the display that appears select Calibrations (3200).
5. Select Collimator calibration (3260).
7. Touch the Run button to move the collimator to a position where the collimator screws
are visible.
The collimator will now be in a position where the screws that secure the collimator in
place are visible.
8. Switch off the X-ray unit.
NOTE
Removing the covers might slightly differ depending on the X-ray unit model (age), for more
detailed instructions, see the original version of the X-ray unit’s Technical manual.
WARNING
Always switch the X-ray unit off before removing any covers.
2 8x
Digi_C_arm_covers1.eps
2. Unscrew the two screws from the inner side of the sensor head connector back cover
(1) and loosen two from the top (2) with 2.5 mm Allen key.
2
Mid_63_1.eps
1
3. Detach the cover from its position (3).
NOTE
You have to remove the C-arm upper covers before removing the tube head covers.
1. Unscrew the two attachment screws from the top of the front cover (1) and the two
screws from the bottom of the cover (2). The cover can now be removed.
Digi_C_arm_covers6.eps
1
2. Unscrew the two attachment screws from the top of the back cover (1) and the two
screws from the side of the cover (2). The cover can now be removed.
1
Digi_C_arm_covers5.eps
NOTE
You have to remove the C-arm upper covers and sensor head covers before removing the
C-arm inner cover.
1. Unscrew the four screws with a 2.5 mm Allen key (1) and remove the cover (2).
Digi_C_arm_covers4.eps
4x
Remove_tubehead1.eps
5
4
1
C- 21-1
J2 Ar 0-
m 05
5 PC
B
J2
4
1
2
3
ONLY IN TOMOGRAPHIC
10001173 IMAGING MODE
10002151 121-10-08
121-10-08
10002151J1
Tube Head
10002431
J1 J2 J3
J4
J6
10002225
J1
2
J5
J8
J7
121-10-06
10002353 Tube Feedback
Receptor Pan / Primary Collimator
1
121-10-05 PCB
C-Arm Rotation Motor
J9
Receptor Rocation
C-Arm Angle
Sensor
C-Arm PCB
Aux. (J18)
KT60LM06-004
J10
10002180
CAM
3
J11 J20
Laser 121-10-08
J27 00050852
ONLY IN TOMOGRAPHIC CH7 CH8
IMAGING MODE CH4 CH5 CH6 Aux.
J18
J20
Aux.
10003369
J26
J21 J22 J23 J24 J25
J19
1
10002181
4
Y-Collimator Motor
DIMAX2 KT35FM1-010
PAN
10002181
10003699
5 X-Collimator Motor
KT35FM1-010
Primary Collimators
2. Unscrew the two screws (1) which secure the collimator in position and remove the
collimator (3).
The screws should be visible and you should be able to access them. Manually lift up
the lower blade to access the lower screw.
2
Remove_collimator_for_3D_RetroFit.eps
NOTE
If you did not drive the collimator to Show Screws -position as instructed in section 3.2
"Driving collimator to Show Screws -position" on page 7, you have to move the
collimator mechanism manually. To do this, rotate the stepper motor spindles (2)
located behind the collimator blades until you can access the screws.
3. Using a 4 mm Allen key, unscrew the four screws (4) which hold the tube head in
position and remove the tube head (5).
3D collimator
with copper filter
NOTE
If the 3D sensor is equipped with Planmeca ProFace option, do not touch the glass
windows when you hold the sensor. Fingerprints or other stains on the glass surface affects
to the image quality.
C-arm
3D sensor
2. Turn the locking knob over the fastening mechanism. This will secure the sensor head
in position.
CBCT_s2.eps
3. Push in the button of the C-arm electrical connector on the other side to secure the
sensor head in position. This will make the electrical connection between the sensor
head and C-arm.
cbct_s5.eps
NOTE
The front cover located on the inside of the C-arm must have an opening for the cooling
grille (see below). This will ensure that the sensor cooling system functions properly.
Replace the front cover if necessary. The front cover is secured with Velcro fasteners.
C-arm
3Ds_sensor.eps
3D sensor
2. Enter the service menus, touch the service spanner icon on the Main display.
4. From the list on the display that appears select Reset to Factory Defaults (2500).
5. Accept the reset by touching the green check mark button.
Set the X-ray unit type and the 3D sensor type according to the retrofit kit type:
• Type: Depending on the retrofit kit 3D sensor type, select either: ProMax 3D s or
ProMax 3D Classic
• Sensor Type: Depending on the retrofit kit 3D sensor type, select either: 3D s or
3D v2
4. Accept the changes by touching the green check mark button.
The X-ray unit restarts.
5. Enter the technical settings, select Technical.
On these displays (2110 - 2180) you can select the X-ray unit’s programs. These
selections determine which programs are shown on the touch screen when using the
X-ray unit.
In addition, on these displays you can define preset kV and mA values for different
type of exposures, i.e. the selected program, patient size, segmentation and jaw size.
2. Acknowledge the changes by touching the green check mark button.
3. Set the program features, select Program --> 2200 Program Features.
NOTE
DEC is available for panoramic and cephalometric programs. It is not available for
SmartPan programs.
On the Licences display you can change and enable/disable the program licences.
You will need a license code to add a new feature to the X-ray unit. Touch the feature
field to switch this feature on. A number display will appear. Enter the license code to
enable the selected feature.
6. Acknowledge the changes by touching the green check mark button.
7. Switch off the X-ray unit.
NOTE
Make sure that the X-ray unit is switched off.
1. Connect all five tube head cables from the C-arm PCB.
Remove_tubehead1.eps
5
4
1
C- 21-1
J2 Ar 0-
m 05
5 PC
B
J2
4
1
2
3
ONLY IN TOMOGRAPHIC
10001173 IMAGING MODE
10002151 121-10-08
121-10-08
10002151J1
Tube Head
10002431
J1 J2 J3
J4
J6
10002225
J1
2
J5
J8
J7
121-10-06
10002353 Tube Feedback
Receptor Pan / Primary Collimator
1
121-10-05 PCB
C-Arm Rotation Motor
J9
Receptor Rocation
C-Arm Angle
Sensor
C-Arm PCB
Aux. (J18)
KT60LM06-004
J10
10002180
CAM
3
J11 J20
Laser 121-10-08
J27 00050852
ONLY IN TOMOGRAPHIC CH7 CH8
IMAGING MODE CH4 CH5 CH6 Aux.
J18
J20
Aux.
10003369
J26
J21 J22 J23 J24 J25
J19
1
10002181
4
Y-Collimator Motor
DIMAX2 KT35FM1-010
PAN
10002181
10003699
5 X-Collimator Motor
KT35FM1-010
Primary Collimators
NOTE
The minimum requirement for the Ethernet cables used in the 3D system network is
Category 6 (CAT6).
X-ray unit
Reco PC Romexis PC
A
B
Ethernet switch
PoE In 10/100/1000 Base-T Ports (1-8) - Ports are Auto - MDIX
1 2 3 4 5 6 7 8
a. Use an Ethernet cable (A) to connect the X-ray unit to the Ethernet switch.
The Ethernet cable is attached to the left terminal (marked “ETHERNET”) on the
underside of the stationary column top.
CAUTION!
Wrong connection
can cause damage!
Make sure that you
connect the cable to
the connector
marked “ETHERNET”.
exp_sw_input_sw.eps
Reconstruction PC
Terminal for
Ethernet cable
(Reconstruction PC -
Ethernet switch)
c. Use an Ethernet cable (C) to connect the Romexis PC to the Ethernet switch.
4. Switch on the X-ray unit.
5 FINALISING INSTALLATION