Hospira GemStar 13086 Infusion Pump - User Manual
Hospira GemStar 13086 Infusion Pump - User Manual
List 13000-04-07/08
List 13100-04-07/08
PLEASE
Read this entire manual before
using the Hospira GemStar Pump
This manual is designed for use by healthcare professionals, caregivers, and patients.
The Hospira Technical Support Operations hotline is available 24 hours a day (in the
USA) to provide consultation and technical assistance regarding the Hospira GemStar
Pump.
Change History
Part Number Description of Change Pages Changed
430-00897-001 Original Issue N/A
(Rev. 9/02)
Aligned Note and Tip icons 21, 61, 69, 88, 97, 100–102,
106–107, 111, 122, 125, 144,
149, 178, 190
Note: Change page manuals assembled by Hospira include a change page identifier
in the part number on the cover page and on the bottom of each changed page. This
change page manual is identified as 430-00897-A02.
Contents
1. Introduction 1
GemStar Pump Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
GemStar Pump Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Therapies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
GemStar Pump Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Menu System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Indications for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Contraindications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2. Basic Operation 17
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Selecting a Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Connecting to the GemStar Docking Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Powering On the Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Accessing the Programming Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Programming the Pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Starting a New Container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
GemStar Pump Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Loading the Cassette. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Releasing the Cassette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Priming a GemStar Pump Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Contents
4. Pain Management 53
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Pain Management Programming. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Bolus Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Loading Dose Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Changing a Pain Management Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Pain Management Programming Worksheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5. Intermittent 61
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Intermittent Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Intermittent Delivery Interruptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Changing an Intermittent Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Intermittent Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
6. Continuous 69
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Continuous Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Rate Titration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Piggybacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Changing a Continuous Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Continuous Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
7. Weight Dosed 77
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Weight Dosed Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Auto KVO. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Rate Titration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
VTBI Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Changing a Weight Dosed Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Weight Dosed Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Contents
8. mL/hr Only 83
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
mL/hr Only Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Auto KVO. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Rate Titration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
VTBI Titration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Piggybacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Changing a mL/hr Only Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
mL/hr Only Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
9. Variable Time 91
Therapy Features and Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Variable Time Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Variable Time Delivery Interruptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Changing a Variable Time Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Variable Time Programming Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Contents
Index 191
Chapter 1
Introduction
Tip: A TIP contains information that makes the pump easier to use.
Note: Examples are provided only to help explain a pump function and may not
be clinically relevant. Graphics are for illustrative purposes only and may not
reflect the actual product.
1. Introduction
1. Introduction
1 DISPLAY • Displays the pump’s status.
2 • Silences an alarm.
3 • Stops infusion.
4 • Starts infusion.
1. Introduction
Bottom/Side View
1. Introduction
Therapies
Seven types of therapy are available for infusion of medications and fluids. The
following table lists possible applications and available features of each therapy.
Therapy Possible
L
WN P
Applications
VO
RAT RA TE
DOSADING
KVO IONA
VTB RATE
/DOPER U
G
BACGGY-
O K
KIN
US
E
I
TIT
OPT
BOL
TIT
AUT
LO
PI
TA
TPN Total Parenteral
Nutrition fluid • •
delivery
Pain Infusion of
Management analgesics or
anesthetics for
patient-
• •
controlled pain
management (PCA)
Intermittent Interval-based
therapies such as •
antibiotics
Continuous Infusion of mg, mcg,
or mL/hr with a
selectable KVO • • •
(Keep Vein Open)
option
Weight Dosed Critical care,
pediatrics
• • •
mL/hr Only Simple rate
programming
• • • •
Variable Time Chronotherapy • •
• = Available
1. Introduction
1. Introduction
Menu System
The GemStar Pump uses a menu system that provides step-by-step instructions
through all of the pump’s functions. Refer to the following table for a list of functions
in each menu.
Use the keypad to select menu items, respond to display messages, and enter numeric
values. For example, press [2] to select PAIN MANAGEMENT from the Programming
Menu.
Tip: A menu item does not have to be on the display to be selected. For example,
when the Programming Menu first displays, only the first four therapies are shown.
Instead of pressing the down arrow to see the other menu items, press [7] to go
directly to Variable Time Programming.
1. Introduction
1. Introduction
Operating Modes
The GemStar Pump has four operating modes:
Mode Description
Help Mode The pump enters the HELP mode when is pressed during
programming or an alarm condition. The HELP mode displays
additional information about the current condition.
Programming Mode The pump enters the PROGRAMMING mode after the pump is
powered on or when is pressed.
Run Mode After is pressed, the pump enters the RUN mode, even
when not infusing.
Stop Mode When programming is complete, the pump enters the STOP
mode until is pressed.
1. Introduction
1. Introduction
1. Introduction
1. Introduction
1. Introduction
When removing the cord from the bottom of the pump, firmly grasp the connector
and pull straight out. DO NOT twist or bend the cord or connector.
To ensure proper pump operation, ALWAYS replace BOTH disposable batteries
with fresh, disposable AA batteries when a change is required.
ALWAYS fully close the battery door when using disposable batteries.
DO NOT use rechargeable batteries in the battery compartment.
Ensure two fresh, disposable AA batteries are installed for backup power when
using other power sources (e.g, docking station, battery pack). The pump will
continue to operate on backup power if the external power source fails.
ALWAYS avoid sources of high-intensity electromagnetic radiation (e.g., radio
transmitters, MRI scanners, microwave ovens, X-ray machines, and CAT
scanners).
Use of radio-frequency emitting devices, such as cellular telephones, 2-way
radios, and Electrical Surgical Devices (ESU), in close proximity to this device
may affect its operation.
Possible explosion hazard exists if the pump is used in the presence of flammable
anesthetics. NEVER use the pump in the presence of flammable or explosive
vapors.
Non-hazardous, low-level electrical potentials are commonly observed when
fluids are administered using infusion devices. These potentials are well within
accepted safety standards, but may create artifacts on voltage-sensing equipment,
such as ECG, EMG, and EEG machines. These artifacts vary at a rate that is
associated with the infusion rate. If the monitoring machine is not operating
correctly or has loose or defective connections to its sensing electrodes, these
artifacts may be accentuated so as to simulate actual physiological signals. To
determine if the abnormality in the monitoring equipment is caused by the
infusion device instead of some other source in the environment, set the infusion
device so that it is temporarily not delivering fluid (if clinically acceptable).
Disappearance of the abnormality indicates that it was probably caused by
electronic noise generated by the infusion device. Proper setup and maintenance
of the monitoring equipment should eliminate the artifact. Refer to the appropriate
monitoring system documentation for setup and maintenance instructions.
1. Introduction
The GemStar pump, when connected to the GemStar AC adaptor, has been tested
and found to comply with EMC/EMI limits in accordance with IEC/EN
60601-1-2 (2001). When connected to other power sources and/or optional system
components, the GemStar system complies with EMC/EMI limits in accordance
with IEC/EN 60601-1-2 (1993). These limits are designed to provide reasonable
protection against harmful interference in a typical medical installation. The
equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful
interference to other devices in the vicinity. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause
harmful interference with other devices, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the interference by
one or more of the following measures:
Reorient or relocate the receiving device
Increase the separation between the equipment
Connect the equipment to an outlet on a circuit different from that to which
the other device is connected
Consult the manufacturer or field service technician for help
The use of portable and mobile RF equipment may have an impact on this and
other pieces of medical equipment
1. Introduction
Some cleaning and sanitizing compounds may slowly degrade components made
from some plastic materials. DO NOT use compounds containing combinations of
isopropyl alcohol and dimethyl benzyl ammonium chloride (i.e., alcohol wipes).
DO NOT sterilize by heat, steam, ethylene oxide (ETO), or radiation. Apply
disinfectants to the outside surface of the pump only. DO NOT use abrasive
cleaners or materials on the pump. Using abrasive cleaners or cleaning solutions
not recommended by Hospira may result in product damage.
Please comply with local disposal and recycling regulations as appropriate for
disposable batteries, rechargeable battery packs, medical electronic components,
and pump sets.
Chapter 2
Basic Operation
The minimum elements required for using the GemStar Pump are:
GemStar Pump
Appropriate power source
Appropriate GemStar Pump Set
Appropriate fluid container
Patient-access device
Overview
To start a therapy:
1 Determine the program settings. Refer to the appropriate therapy chapter for a
worksheet that may be copied and used to record program settings.
2 Select a power source. Refer to Selecting a Power Source on page 18.
3 Prime the pump set. Refer to GemStar Pump Set on page 29.
4 Load the cassette into the pump. Refer to Loading the Cassette on page 32.
5 Power-on the pump and go to the Programming Menu. Refer to Powering On
the Pump on page 24.
6 Select the appropriate optional system components, such as a bolus cord.
7 Select the therapy type and enter a program. Refer to the appropriate therapy
chapter for detailed instructions.
8 Press [START] to begin delivery.
2. Basic Operation
2. Basic Operation
2. Basic Operation
2. Basic Operation
Note: A charged battery pack will last approximately 24 hours when the pump is
infusing at rates less than 125 mL/hr.
2. Basic Operation
CAUTION: When removing the cord from the bottom of the pump, firmly
grasp the connector and pull straight out. DO NOT twist or bend
the cord or connector.
Note: The battery pack light is yellow while the battery pack is charging. When the
battery pack is charged, the battery pack light is green.
2. Basic Operation
Note: The docking station light is yellow while the docking station is charging. When
the docking station is charged, the light is green.
2. Basic Operation
Note: If a cassette is not properly loaded, the pump will alarm after the power-on test is
complete.
1 Press [ON/OFF] to power-on the pump. The pump
GEMSTAR 7
completes a self-test and displays the time and date. No UNIT SELF-TEST
response is required. TIME IS 10:00 AM
FRI, JUN 11, 04
2 The pump checks for available power sources. Respond to the display message as
shown in the following table:.
2. Basic Operation
YES
USING BATTERIES Disposable batteries Press ENTER
to confirm.
only.
ENTER TO CONFIRM If an external power source is in use,
ensure the connections are secure.
YES
USING EXT BATT External batteries Press ENTER
to confirm.
(docking station
ENTER TO CONFIRM battery or battery If an AC mains adaptor is in use,
pack) and disposable ensure the connections are secure.
batteries.
Note: To power-off the pump, press and hold until the display clears. The pump
cannot be powered off during a CHECK CASSETTE alarm (except during
PROGRAMMING mode). Refer to Chapter 12 Troubleshooting for information on
clearing a CHECK CASSETTE alarm.
2. Basic Operation
Tip: To enter a new program when the pump is already powered on:
f 1 Press [STOP] to place the pump in STOP mode.
2 Press [CHANGE] to display the Change Menu.
3 Select NEW PROGRAM. The pump clears the current program and
displays the Programming Menu.
2. Basic Operation
Programming Overview
The pump’s menu system provides step-by-step guidance through all of the
programming steps. Refer to the appropriate therapy chapter for detailed
programming instructions.
Programming worksheets are available at the end of each therapy chapter. Each
worksheet lists the minimum and maximum value (program range) allowed for that
therapy. If a number outside of the program range is entered, the pump displays the
minimum or maximum value allowed.
2. Basic Operation
Programming Tips
The following table explains how to use the keypad during programming:
Key Function
[0] - [9] Press the number keys to select menu items and to enter numeric values.
Note: Some menu items require pressing two number keys to select that items.
For example, when selecting a unit of measure for a weight dosed program, the
menu may display as:
01 mcg/kg/mg
02 mcg/kg/hr
03 mcg/kg/min
04 mg/kg/min
To select mg/kg/min, press [0][4].
Press YES
ENTER to accept entries or to continue when done viewing a display.
2. Basic Operation
False occlusion alarms and fluid delivery inconsistencies may occur when pumping
viscous fluids in microbore tubing at rates greater than 500 mL/hr.
2. Basic Operation
2. Basic Operation
WARNING: ALWAYS prime the pump set to remove air from the
cassette, tubing, and injection sites prior to connecting to
the patient. ALWAYS disconnect the pump set from the
patient prior to priming or purging. Arrange tubing, cords,
and cables to minimize the risk of patient strangulation or
entanglement.
CAUTION: USE ONLY GemStar Pump Sets with the GemStar Pump. Use of
unauthorized sets may result in injury to the patient or damage to
the pump.
To prevent contamination, use aseptic technique with all fluid-path
connections. Remove protective coverings as assembly progresses.
2. Basic Operation
2 Line up the cassette as shown 3 Insert the cassette into the cassette
(with the flow stop button facing pocket by pushing firmly along the
the pump). entire cassette until it is firmly
seated.
2. Basic Operation
2. Basic Operation
2 Hold the container with the 5 While holding the rotated cassette,
administration port toward the squeeze or roll the container to
ceiling and insert the piercing pin. remove air from the container and
cassette.
2. Basic Operation
7 Push down on the flow stop to Note: Ensure the tubing is behind
close the cassette.
the tubing guide (at the distal
end).
2. Basic Operation
Priming Pump Sets WITHOUT a Drip Chamber Using the PURGE Key
Note: The PURGE function is available only if there is a program in the pump.
2. Basic Operation
5 Press [PURGE].
PURGE HELP
2. Basic Operation
1 Push down on the flow stop to 4 Push down on the flow stop to
close the cassette. open (priming position) the
cassette.
2. Basic Operation
2. Basic Operation
Priming Pump Sets WITH a Drip Chamber Using the PURGE Key
WARNING: Failure to use the anti-siphon valve may result in
unrestricted flow.
Note: The Purge function is available only if there is a program in the pump.
1 Push down on the flow stop to 4 Push down on the flow stop to
close the cassette. open (priming position) the
cassette.
2. Basic Operation
PURGE HELP
2. Basic Operation
CAUTION: When using pre-filled vials, the system may underdeliver at low
flow rates.
1 If using a pump set with the 4 Connect the pump set to a Hospira
Y-extension, use the slide clamp to pre-filled vial and injector.
close off the KVO (Keep Vein
Open) side of the tubing.
2. Basic Operation
6 Hold the vial and set upright 9 If the set includes a KVO
(vertically) on a flat surface with Y-extension, close the slide clamp
the pump set extending from the below the pre-filled vial and open
top. the slide clamp on the KVO side.
7 Slowly push down on the injector Prime the KVO Y-extension.
until fluid flows at the distal end of
the tubing, which indicates that all
air is cleared from the vial,
injector, cassette, and tubing.
2. Basic Operation
2. Basic Operation
2. Basic Operation
PURGE HELP
Chapter 3
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
maximum values allowed for this protocol. Use copies of the programming
worksheet to record program values before programming the pump.
TPN Programming
1 Select TPN from the Programming Menu.
2 Select the TPN delivery mode.
3 Enter the total VTBI (Volume To Be Infused).
Note: Time is entered in hours:minutes (00:00). For one hour thirty minutes,
press [1] [3] [0]; not [9] [0].
Note: KVO runs after Taper Down until the total container is
infused.
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
Note: You cannot change a program when the pump is in FULL or CONTAINER
Lock.
To change a program:
1 From the STOP mode, press [CHANGE] and select CHANGE PROGRAM.
2 As each message displays, make the desired changes or press [YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review.
4 Press [START] to begin infusion.
Note: Time is entered in hours:minutes (00:00). For one hour thirty minutes,
press [1] [3] [0]; not [9] [0].
3 Press [YES/ENTER] to return to the RUN mode, then press [START] to begin Auto
Taper Down.
GemStar®
TPN Programming Worksheet
GIVE:
mL 1 - 9600 mL*
Taper Up No
Notes
Chapter 4
Pain Management
6 In Continuous + Bolus protocols, you can enter zero for either the continuous rate
or the bolus dose when entering or changing a program. The continuous rate and
bolus dose cannot be zero at the same time.
6 The default bolus delivery rate is 125 mL/hr.
6 A loading dose may be set during programming, which can be delivered after
programming and priming are complete. The default loading dose rate is 125
mL/hr.
6 Priming using [PURGE] is only allowed before [START] is pressed or after one of
the following events occur:
6 Complete Review of New Program
6 New Container
6 Air-in-Line Alarm
CAUTION: When using pre-filled vials, the system may underdeliver at low
flow rates.
Note: New Container is allowed only when at least 1/2 of the container volume has
d been infused or 1 hour has elapsed since starting the infusion.
Tip: The Programming Worksheet at the end of this chapter lists the minimum and
f maximum values allowed for this protocol. Use copies of the programming
worksheet to record program values before programming the pump.
4. Pain Management
Note: If Continuous + Bolus is selected, the continuous delivery rate or bolus dose
can be set to zero through the CHANGE menu without reprogramming the pump.
4. Pain Management
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
4. Pain Management
Bolus Delivery
Set bolus doses and delivery limits during programming. The default bolus dose
delivery rate in a pain management protocol is 125 mL/hr.
A bolus cord is available for patient bolus requests. Refer to Connecting the Bolus
Cord in Chapter 15 Optional System Components, on page 156, for more
information.
When the bolus dose is complete, the pump does the following:
Records the bolus request and delivery amount to the event and bolus history logs
Adds the delivered boluses to the total bolus amount in the program amounts
Resets the bolus lockout time (if applicable) and ignores other bolus requests until
the lockout time has elapsed
Continues programmed infusion
Note: ALL bolus requests are recorded to the event and bolus history logs.
If the bolus delivery does not start, it could be locked out by one of the following
conditions:
4. Pain Management
Note: Loading dose amounts are included in a 1-hour or 4-hour dose limit.
4. Pain Management
Pressing [STOP]
An alarm condition
To continue a loading dose:
Note: You cannot change a program when the pump is in FULL or CONTAINER
Lock. With a RATE CHANGE Lock, the continuous delivery rate and bolus amount
can be changed within the set ranges.
Although the delivery mode cannot be changed, if Continuous + Bolus is
programmed, the continuous rate or bolus amount can be changed to zero. The
minimum values set with a RATE CHANGE Lock could restrict this function.
4. Pain Management
To change a program:
1 From the STOP mode, press [CHANGE] and select CHANGE PROGRAM.
2 As each message displays, make the desired changes or press [YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review.
4 Press [START] to begin delivery.
4. Pain Management
GemStar®
Pain Management Programming Worksheet
GIVE:
Parameters Selections and/or Values to be Programmable Ranges
Entered
Therapy Pain Management N/A
Delivery Mode Continuous N/A
Bolus Only
Continuous + Bolus
Unit of Concentration: mg/mL 0.1 - 100 mg/mL
Measure
Concentration: mcg/mL 1 - 1000 mcg/mL
mL
Continuous 0.1 - 2500 mg/hr
Rate 1 - 25000 mcg/hr
0.1 - 25.0 mL/hr
Loading Dose No 0.1 - 2500 mg
1 - 25000 mcg
Yes Amount: 0.1 - 25.0 mL
Bolus Dose 0 - 2500 mg
0 - 25000 mcg
0 - 25.0 mL
Bolus Lockout
minutes 5 - 999 min
Optional Dose 4-hour limit Amount: 4X rate - 400 mL, 40000 mg,
Limit 400000 mcg*
Chapter 5
Intermittent
Therapy Features and Specifications
This therapy allows programming of multiple doses at regular intervals.
Optional KVO (Keep Vein Open) can be set during programming.
The pump can be disconnected from the patient between dose deliveries. A call
back alarm can be set to alert the patient or caregiver to reconnect the pump before
the next scheduled delivery begins. The pump must remain powered on for the
call back alarm to function. At the set time before each dose, the pump produces
an audible and visual alarm and flashes START on the display.
Tip: The Programming Worksheet at the end of this chapter lists the
minimum and maximum values allowed for this protocol. Use copies of the
programming worksheet to record program values before programming the
pump.
5. Intermittent
Intermittent Programming
1 Select INTERMITTENT from the Programming Menu.
2 Enter the dose amount in mL.
3 Enter the infusion time for each dose in hours:minutes (00:00).
For example, press [4] [0] if each dose is to deliver for 40 minutes.
4 Enter the frequency of dose deliveries.
For example, press [4] [0] [0] to start a dose every four hours.
5 Enter the total number of doses in the container.
To determine the number of doses in the container, divide the volume in the
container by the dose amount.
6 Enter the KVO rate in mL.
KVO delivers before a delayed start time, between doses, and after the last dose
until the total container is infused.
If no KVO is desired, press [YES/ENTER] while the value displayed is zero.
7 Enter the container size.
The pump automatically calculates the minimum container size required to deliver
the number of programmed doses and KVO between doses.
Note: If KVO is desired before a delayed start, increase the container size to
include the desired KVO volume.
8 Select the air sensitivity, if required.
Note: DO NOT press [YES/ENTER] after selecting the air sensitivity. If
[YES/ENTER] is accidentally pressed and a call back alarm is not desired, press
[BACK-UP] to return to the beginning of step 9 and press [NO].
9 The pump displays SET CALL BACK ALARM FOR NEXT DOSE?
• Press [NO] if you do not want to set the call back alarm.
• Press [YES/ENTER] to set the call back alarm.
Enter the time before the next dose that the alarm should sound. For example,
press [1] [0] to set the alarm to sound 10 minutes before the start of the next
scheduled delivery.
5. Intermittent
Note: When programming the call back alarm, be sure to allow enough time for the
patient or caregiver to reconnect the pump to the patient and to press [START]
BEFORE the scheduled dose start time.
10 The pump displays the current time.
• If the displayed time is correct, press [YES/ENTER].
• To set the clock, press [OPTIONS].
11 The pump displays the current time and START NOW?
• Press [YES/ENTER] to begin delivery when the pump enters the RUN mode.
• Press [NO] to set a delayed start time. The pump displays SET START TIME.
Enter the desired start time and press [YES/ENTER] when done.
Note: The call back alarm works only when the pump is powered on AND in the
STOP mode. DO NOT power-off the pump while it is disconnected from the patient
if you have set the call back alarm.
When the alarm sounds, press [SILENCE] to silence the alarm for three minutes.
Then reconnect the pump to the patient.
5. Intermittent
Dose Interruptions
To understand how interruptions are resolved, compare the normal Intermittent
delivery diagram below for a 1-hour dose every four hours, to the interruption
diagrams shown in the examples:
12 1 2 3 4 5 6 7 8 9 10 11 12
Dose Delivered -
If an Intermittent dose is interrupted and the pump is placed in the RUN mode
during the on or off cycle of the current dose, the dose continues delivery from the
point where it was interrupted and all remaining doses are shifted.
For example, in the following diagram, an intermittent dose was started at 2:00.
The dose was interrupted during the current dose cycle and was started again
before the end of the current dose cycle.
5. Intermittent
12 1 2 3 4 5 6 7 8 9 10 11 12
If an Intermittent dose is interrupted and the pump is placed in the RUN mode
after the off cycle of the current dose, the next dose begins delivery. All remaining
doses are shifted if necessary.
For example, in the following diagram, an intermittent dose was started at 2:00. It
was interrupted during the current delivery cycle at 2:30, but was not started again
until after the current delivery cycle at 6:45:
12 1 2 3 4 5 6 7 8 9 10 11 12
When the dose is complete, the start time for the Next Dose displays. To change
the start time refer to Chapter 10 Options Menu, page 97.
Any missed Intermittent doses, except the first dose, are added to the end of the
delivery schedule.
5. Intermittent
Note: You cannot change a program when the pump is in FULL or CONTAINER
Lock.
To change a program:
1 From the STOP mode, press [CHANGE] and select CHANGE PROGRAM.
2 As each message displays, make the desired changes or press [YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review.
4 Press [START] to begin delivery.
5. Intermittent
GemStar®
Intermittent Programming Worksheet
GIVE:
Parameters Selections and/or Values to be Entered Programmable
Ranges
Therapy Intermittent N/A
Dose Amount
0.1 - 9600 mL
Infusion Time
Per Dose
hrs:min 1 min - 24 hrs
Dose (Infusion time per
Frequency dose + 1 min) to 96 hrs
hrs:min
Number of
Doses Per
Container 1 - 999
KVO No KVO Rate: 0.0
(Keep Vein
Open) Yes KVO Rate: 0.1 -5.0 mL/hr
Container
Size*
0.1 - 9999 mL
Air Sensitivity On Alarms at approx. 0.5 mL of air N/A
2 mL Alarms at approx. 2 mL of air
Off No alarm
Call Back No
Alarm Yes
1 min to time between
Alarm: hrs:min before start doses
Start Time Start immediately after programming
Notes
Chapter 6
Continuous
6. Continuous
Continuous Programming
1 Select CONTINUOUS from the Programming Menu.
2 Select the unit of measure and concentration as required.
3 Enter the rate.
4 Enter the VTBI (Volume to be Infused).
5 Enter the KVO rate, if desired.
KVO runs after the programmed VTBI until the total container is infused. If no
KVO is desired, press [YES/ENTER] while the value displayed is zero.
6 Enter the container size (VTBI + KVO).
VTBI is the default value on the CONTAINER display.
Note: If KVO is programmed, increase the container size to include the KVO
delivery.
To enter the total in mL when programming a concentration, press to move to
the mL line. Enter the mL size when mL flashes on the display.
7 Select the air sensitivity, if required.
8 Review the program. Press to begin review.
9 If using the PURGE key to prime the pump set, press [PURGE] now.
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
6. Continuous
Rate Titration
To titrate the rate while the pump is infusing:
1 In the RUN mode, press the number key for the new rate. Press [YES/ENTER].
For example: if the current rate is 100 mL/hr, press [5] [0] [YES/ENTER] to change
the rate to 50 mL/hr.
Note: You have approximately five seconds after pressing a key to press another
numeric key, the decimal key, or [YES/ENTER]. If no key is pressed within five
seconds, the rate does not change and the delivery continues as previously
programmed.
2 NEW RATE displays with the new value.
• Press [YES/ENTER] within 10 seconds to accept the change and begin delivery
at the new rate.
• Press [NO] to continue delivery at the previously programmed rate.
Piggybacking
Piggybacking with the GemStar Pump requires a GemStar Primary Piggyback Pump
Set with a proximal Y-site backcheck valve and a secondary piggyback set with an
extension hook. The secondary container must be higher than the primary
container. Refer to the instructions included with the piggyback pump set for more
information.
The piggyback rate and piggyback VTBI display until the secondary container has
been delivered. Then the primary infusion continues.
Note: The piggyback infusion amount is NOT added to the primary programmed
infusion amount.
6. Continuous
6. Continuous
Tip: Prepare and set up containers BEFORE programming the pump for piggyback
delivery.
Note: You must know the amount that has already been delivered. When the piggyback
dose amount is changed, the pump begins a new VTBI countdown.
6. Continuous
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the STOP mode, press [CHANGE] and select CHANGE PROGRAM.
2 As each message displays, make the desired changes or press [YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review.
4 Press [START] to begin delivery.
6. Continuous
GemStar®
Continuous Programming Worksheet
GIVE:
Parameter Selections and/or Values to be Programmable Ranges
Entered
Therapy Continuous N/A
Unit of Measure Concentration: mg/mL 0.1 - 1000 mg/mL
Notes
Chapter 7
Weight Dosed
CAUTION: When infusing short-half-life drugs (i.e., those dosed with units of
mcg/kg/min) at very low rates (0.8 mL/hr or less) the rate may be too
slow for that drug. Consider using a lower drug concentration with
these drugs so a pump rate above 0.8 mL/hr can be used.
7. Weight Dosed
[0] - [9] Selects menu items and Scrolls through program review and
enters numeric values. help messages. The up arrow inserts a
decimal point in a numeric value.
Accepts entries or Clears a numeric value (resets to zero)
continues when done during programming.
viewing a display.
Note: If two numbers appear before a menu time (e.g., 02), press both number
keys to select that item.
3 Enter the patient weight, if required.
4 Enter the concentration:
• Enter the value for the unit of measure when it flashes on the display.
• Press [YES/ENTER] to confirm.
Note: The pump automatically displays the mL/hr equivalent even when another
unit of measure is selected.
6 Enter the VTBI.
7 Review the program. Press to begin review.
7. Weight Dosed
8 If using the PURGE key to prime the pump set, press [PURGE] now.
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
Auto KVO
When the programmed VTBI has completed delivery, Auto KVO delivery begins
automatically and the empty container alarm sounds.
Note: Auto KVO delivers even while the empty container alarm sounds. Press
[SILENCE] to quiet the alarm for 2 minutes.
Rate Titration
To titrate the rate while the pump is infusing:
1 In the RUN mode, press the number keys for the new rate. Press [YES/ENTER].
For example: if the current rate is 4.0 mg/kg/hr, press [5] [YES/ENTER] to change
the rate to 5.0 mg/kg/hr.
Press to enter the rate in mL/hr. When using units other than mL/hr, the pump
automatically adjusts the mL/hr value.
Note: You have approximately five seconds after pressing a key to press another
numeric key, the decimal key, or [YES/ENTER]. If no key is pressed within five
seconds, the rate does not change and the delivery continues as previously
programmed.
2 NEW RATE displays with the new value.
• Press [YES/ENTER] within 10 seconds to accept the change and begin delivery
at the new rate.
• Press [NO] to continue delivery at the previously programmed rate.
7. Weight Dosed
VTBI Titration
To titrate the VTBI while the pump is infusing:
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the STOP mode, press [CHANGE] and select CHANGE PROGRAM.
2 As each message displays, make the desired changes or press [YES/ENTER] to
accept the current setting.
3 Review the program. Press to begin review.
4 Press [START] to begin delivery.
7. Weight Dosed
2.0 - 200.0 kg
Concentration (For all units of measure EXCEPT mL/hr) 0.1 - 9999000 mg
(2 entries are mg U mEq 0.1 - 99999999 U
required. Use in in in 0.1 - 9999 mEq
to toggle) mL mL mL 0.1 - 9999 mL
Rate
(see below *)
VTBI
(Volume To Be
Infused) 0.1 - 9999 mL
* Programmable Ranges for Rate (not all units of measure are available in all pump configurations):
0.001 - 833 mg/kg/min 0.1 - 1000 mL/hr 0.001 - 99999999 U/min
0.001 - 500000 mg/kg/hr 0.001 - 999999 mg/hr 0.001 - 99999999 U/hr
0.001 - 16666 mg/min 0.001 - 9999999 mcg/hr 0.001 - 999999999 mU/min
0.001 - 166666 mcg/kg/min 0.001 - 999 grams/hr 0.001 - 999 mEq/hr
0.001 - 9999999 mcg/kg/hr 0.009 - 999 ng/kg/hr
0.002 - 166666 mcg/min 0.001 - 99999999 U/kg/hr
** If two numbers appear before a menu item (e.g., 02), press both number keys to select that item.
CAUTION: When infusing short-half-life drugs (i.e., those dosed with units of mcg/kg/min) at very low
rates (0.8 mL/hr or less) the rate may be too slow for that drug. Consider using a lower drug
concentration so a pump rate above 0.8 mL/hr can be used.
Notes
Chapter 8
mL/hr Only
8. mL/hr Only
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
8. mL/hr Only
Auto KVO
When the programmed VTBI has completed delivery, Auto KVO delivery begins
automatically and the empty container alarm sounds.
Note: Auto KVO delivers even while the empty container alarm sounds. Press
[SILENCE] to quiet the alarm for 2 minutes.
Rate Titration
To titrate the rate while the pump is infusing:
1 In the RUN mode, press the number keys for the new rate. Press [YES/ENTER].
For example, if the current rate is 100 mL/hr, press [5] [0] [YES/ENTER] to change
the rate to 50 mL/hr.
Note: You have approximately five seconds after pressing a key to press
another numeric key, the decimal key, or [YES/ENTER]. If no key is
pressed within five seconds, the rate does not change and the delivery
continues as previously programmed.
2 NEW RATE displays with the new value.
• Press [YES/ENTER] within 10 seconds to accept the change and begin delivery
at the new rate.
• Press [NO] to continue delivery at the previously programmed rate.
8. mL/hr Only
VTBI Titration
To titrate the VTBI while the pump is infusing:
Piggybacking
Piggybacking with the GemStar Pump requires a GemStar Primary Piggyback Pump
Set with a proximal Y-site backcheck valve and a secondary piggyback set with an
extension hook. The secondary container must be higher than the primary
container. Refer to the instructions included with the piggyback pump set for more
information.
Note: The piggyback infusion amount is NOT added to the primary programmed
infusion amount.
8. mL/hr Only
8. mL/hr Only
Tip: Prepare and set up containers BEFORE programming the pump for piggyback
delivery.
Note: You must know the amount that has already been delivered. When the piggyback
dose amount is changed, the pump begins a new VTBI countdown.
8. mL/hr Only
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the STOP mode, press [CHANGE] and select CHANGE PROGRAM.
2 As each message displays, make the desired changes or press [YES/ENTER] to
accept the current setting.
3 Press [START] to begin delivery.
Note: A program review is not required and is available only through the Change or
Options Menu.
8. mL/hr Only
GemStar®
mL/hr Only Programming Worksheet
GIVE:
1 - 7200 mL
Piggyback Infusion 1 min or
Time dose amount / 300 mL/hr to 24
hrs:min hrs or dose amount / 0.1 mL/hr
Chapter 9
Variable Time
9. Variable Time
7 Enter the phase information for each phase. Up to 24 phases can be programmed.
9. Variable Time
WARNING: ALWAYS disconnect the pump set from the patient before
priming or purging.
Note: With variable time programming, the pump keeps track of where it is in the
program each minute of the entire 24-hour period. For example, if Phase 1 is missed,
and the pump is placed in RUN mode in the middle of Phase 2, it starts delivery
according to the program parameters from the middle of Phase 2.
Before the end of the phase. The phase runs to its scheduled
stop time.
After the end of the phase, but before The next phase begins delivery at
the beginning of the next start time. the scheduled start time.
9. Variable Time
Note: You cannot change a program when the pump is in FULL or CONTAINER Lock.
To change a program:
1 From the STOP mode, press [CHANGE] and select CHANGE PROGRAM.
2 As each program parameter displays, make the desired changes or press
[YES/ENTER] to accept the current setting.
3 Review the program. Press to begin review.
4 Press [START] to begin delivery.
Note: The container size cannot be made smaller than the daily total of phases, base
rate, and KVO combined; or the current amount delivered, whichever is greater.
9. Variable Time
9. Variable Time
Chapter 10
Options Menu
Keys Functions
[1] - [8] Press the corresponding number key to select a menu function.
Tip: A menu item does not have to be on the display to be selected. For example, only
items 1 through 4 are shown when the Options Menu is first displayed. To select the
Print Menu without scrolling the display, press [6].
Note: Some menu items may not be available depending on the configuration of the
pump.
Tip: Press and HOLD or to scroll through the history displays faster.
History
VERSION N.NNN
To display the history event log: S/N XXXXXXXX
1 Press [OPTIONS] and select HISTORIES.
TIME IS 09:00 AM
FRI, JUN 11, 04
2 Select HISTORY from the Histories Menu.
NEW CONTAINER
3 Press or to scroll the display. 08:15AM JUN 11
4 Press [BACK-UP] when done to exit OPTIONS.
HISTORY CLEARED
11:00PM JUN 10
Shift Totals EVENT LOG
PM11:18 SET ON
To display the shift totals: AIR SENSITIVITY
1 Press [OPTIONS] and select PM11:18 SET CONT
SIZE 100 mL
HISTORIES.
PM11:18 SET RATE
2 Select SHIFT TOTALS from the Histories Menu. 1.0 mL/hr
PM11:18 SHIFT
3 The shift totals display. Press or to scroll CLEARED
the display. END OF EVENT LOG
Sample History Event Log
Tip: Press [HELP] to display shift totals when the pump is not in PROGRAMMING
mode or an alarm condition.
Volume Totals
To display the volume totals:
1 Press [OPTIONS] and select HISTORIES.
2 Select VOLUME TOTALS from the Histories Menu.
3 Press [BACK-UP] when done to exit OPTIONS.
Configuration
To display the pump configuration:
1 Press [OPTIONS] and select HISTORIES.
2 Select CONFIGURATION from the Histories Menu.
3 Press [BACK-UP] when done to exit OPTIONS.
Clear History
To clear the history event log and bolus history:
1 Press [OPTIONS] and select HISTORIES.
2 Select CLEAR HISTORY from the Histories Menu.
3 The pump display CLEARING HISTORY for 3–4 seconds before exiting
OPTIONS.
Bolus History
To display the bolus history:
1 Press [OPTIONS] and select HISTORIES.
2 Select BOLUS HISTORY from the
Histories Menu.
TIME IS 09:15 AM
3 Press [BACK-UP] when done to exit OPTIONS. FRI, JUN 11, 04
BOL DEL 30
BOL DEM 37
BOL AMT 30.0 mL
06:10A TO 09:14A
Operation Test
The Operation Test checks several pump functions to ensure the GemStar Pump is
working properly. Refer to Chapter 14 Operation Test on page 143 for instructions.
The GemStar Pump allows two lock levels at the same time. Therapy Lock can be
used with either Full Lock, Container Lock, or Rate Change Lock.
When two lock levels are applied, if a function is available under one lock level but
not the other, the more stringent restrictions apply. The table on the following page
lists the functions available for each lock level.
When the pump is locked under a particular level, a small lock symbol ( ) appears
next to the applied lock level as shown above. When the pump is in RUN or STOP
mode, the lock symbol appears in the lower left corner of the display and the lock
level abbreviation appears in the lower right corner.
AIR SENSOR Change the air The pump alarms when it detects an air
sensor sensitivity. bubble of approximately 0.5 mL when ON;
or at approximately 2 mL when 2 mL is
selected. The air alarm does not sound if
OFF is selected.
Air Sensitivity
Occlusion Sensor
1 Press [OPTIONS] and select PUMP SETTINGS.
2 Select DISTAL OCCLUSION from
the Pump Settings Menu. DISTAL OCCLUSION
1 HIGH
3 Select the desired setting. 2 MEDIUM
4 The pump displays MAKE XXXXX DEFAULT? 3 LOW
Sound Level
1 Press [OPTIONS] and select PUMP SETTINGS.
2 Select SOUND LEVEL from the Pump Settings Menu.
3 When the pump displays the Sound Level function, it
sounds intermittent beeps at the current sound level. - +
The blocks on the top line of the display indicate the USE TO ADJUST
current sound level. THE SOUND LEVEL
ENTER WHEN DONE
• Press to increase the sound level.
• Press to decrease the sound level.
• Press [CHANGE] to restore the sound level to the default setting.
4 Press [YES/ENTER] when done.
ENTER TO CONFIRM
Print
WARNING: Ensure a cassette is installed in the pump before
connecting to a printer or computer.
Assign
1 Press [OPTIONS] and select SPEED PROTOCOL.
2 Select ASSIGN from the Speed Protocol menu.
3 Select [1]–[9] to assign to the new protocol.
If the selected speed protocol is already assigned, the pump displays REPLACE
PROTOCOL ASSIGNMENT?
• Press [YES/ENTER] to replace the protocol. The pump displays the new speed
protocol and exits OPTIONS.
• Press [NO] to keep the existing protocol and exit OPTIONS.
Review/Recall
1 Press [OPTIONS] and select SPEED PROTOCOL.
2 Select REVIEW/RECALL from the Speed Protocol menu.
3 Select the desired protocol to review/recall.
4 Review the program. Press to begin review.
5 Press [YES/ENTER]:
• If in the STOP mode, press [1] to recall the protocol; or press [2] to review
another protocol.
• If in the RUN mode, the pump exits OPTIONS.
Delete
1 Press [OPTIONS] and select SPEED PROTOCOL.
2 Select DELETE from the Speed Protocol Menu.
3 Select the desired protocol to delete.
4 Press [YES/ENTER] to confirm deletion or [NO] to exit OPTIONS without deleting.
Equipment Needed
Configuration Settings
Refer to the printer or computer documentation for information on configuration
settings. Configure the printer or computer as follows to use the Data Port:
Data Bits = 8
Stop Bits = 1
Parity = None
Note: When using the Seiko DPU 414 printer, the switch settings MUST be configured
to work with the GemStar Pump.
Note: For maximum battery life, operate the pump on AC mains when connected to a
printer or computer.
1 Install a cassette into the pump.
2 Connect the GemStar serial cable to the printer and to the data port on the bottom
of the pump.
3 Ensure the printer is ONLINE and ready to receive data. Refer to printer
documentation for more information.
4 Ensure paper is loaded in the printer.
5 Select the desired print function from the pump menu.
Connecting to a Computer
Note: For maximum battery life, operate the pump on AC mains when connected to a
printer or computer.
1 Install a cassette into the pump.
2 Connect the GemStar serial cable to the null modem.
3 Connect the null modem to the serial gender changer.
4 Connect the serial gender changer to the serial port of the computer.
5 Connect the GemStar serial cable to the data port on the bottom of the pump.
6 Configure the communication software on the computer. Refer to Configuration
Settings on page 113.
Tip: After the data transfer is complete, use any text editor to view or print the file.
Note: The following procedure is only needed to create a “New Connection” the first
time the pump is connected to the computer.
Notes
Chapter 11
Clinician Instructions
The instructions in this section apply to special features available to clinicians.
Remove this section from the manual to prevent unauthorized access to
these features.
The GemStar pump provides quick access to commonly used functions. Instead of
accessing these functions by navigating through the menu system, you can enter the
key combinations shown in this table:
Unlock Keypad Place the pump in STOP mode and then enter the lock
sequence to unlock a Full Container or Rate Lock. You
cannot use quick access keys to unlock a Therapy Lock.
Displays Shift Totals Place the pump in STOP mode and then press [HELP] to
display Shift Totals.
Display Clock Place the pump in STOP mode, press [YES], and then press
and hold [1]. The clock displays until you release [1].
Program a Loading From the Pain Management STOP MODE, press
Dose [YES/ENTER], and then press [0] to program a Loading
Dose.
Note: When two lock levels are applied, if a function is available under one lock
level but not the other, the more stringent restrictions apply.
7 Enter the lock sequence number. Asterisks display as the numbers are entered.
• For THERAPY Lock, enter [0][0][0][3][1].
• For UNLOCKED, FULL, CONTAINER, and RATE CHANGE lock levels, enter
[1][3][0][0][0].
Shift Totals
1 Press [OPTIONS] and select HISTORIES.
2 Select SHIFT TOTALS from the Histories Menu.
END OF
3 The shift totals display. Press SHIFT TOTALS
and to scroll the display. When done, the CLEAR TOTALS?
pump displays CLEAR TOTALS? YES OR NO. YES OR NO
• Press [YES/ENTER] to clear the shift totals.
• Press [NO] to exit OPTIONS without clearing
the shift totals.
Tip: Press [HELP] to display shift totals when the pump is not in PROGRAMMING
mode or an alarm condition.
Notes
Chapter 12
Troubleshooting
This section contains information on audible and visual alarms that may occur with
the GemStar Pump.
CAUTION: If the pump does not perform as stated in this manual, stop using it
IMMEDIATELY.
For consultation and technical support contact either an authorized Hospira
representative or the Hospira Technical Support Operations hotline (available 24
hours a day).
To return the pump for service, first contact Hospira Customer Support between 7:00
AM and 4:00 PM Pacific time. Have the pump serial number available.
Hospira, Inc.
Technical Support Operations
755 Jarvis Drive
Morgan Hill, CA 95037
Tip: Press [HELP] for additional information during programming and alarm conditions.
12. Troubleshooting
Alarm Messages
Alarm Display Is the Alarm Cause/ Remedy
Message pump Condition
infusing?
Y = Yes A = Audible alarm
N = No V = Visual alarm
AIR IN LINE N A/V The specified volume 1 Press [SILENCE] to quiet
of air has been the alarm for 1 minute.
• Pump automatically detected in the line 2 Press [STOP] to clear the
enters STOP mode. distal to the cassette alarm.
while infusing.
3 Disconnect the patient
from the pump set.
4 Press and hold [PURGE].
Note: Purge rate is 250
mL/hr. The pump purges for
up to 2 minutes each time
you press and hold
[PURGE]. Purge volume
may be limited in a Pain
Management program.
5 Reconnect the pump set to
the patient-access device.
6 Press [START] to begin
infusion.
CHANGE BATTERIES N A/V The pump has 1 Replace both disposable
entered the STOP batteries with fresh AA
mode due to battery batteries and press to
voltage below restart the pump; OR
acceptable levels. connect the pump to the
AC mains adaptor and the
pump will automatically
restart. Replace both
batteries as soon as
possible.
2 After power-on, resume the
current program. Refer to
Accessing the Programming
Menu in Chapter 2 Basic
Operation for more
information.
12. Troubleshooting
N = No V = Visual alarm
CHECK CASSETTE-X N A/V The cassette is 1 Press [SILENCE] to quiet
improperly installed the alarm for 1 minute.
or not installed. 2 Press [STOP] to place the
pump in the STOP mode.
X indicates where to
look for the cause of 3 Reinstall the cassette.
the alarm: CAUTION: Ensure all
four cassette latches
A = Air Sensor are clearly visible after
the cassette is
D = Distal installed.
P = Proximal Note: Ensure the tubing is
behind the tubing guide (at the
distal end). Refer to Loading
the Cassette on page 32.
4 Press [START] to begin
infusion.
Note: The pump cannot be
powered off during a
CHECK CASSETTE alarm
(except during
PROGRAMMING mode).
CALL N A/V The pump has 1 Write down the telephone
1.800.XXX.XXXX detected a condition number and code.
that may require 2 Power-off the pump. If the
CODE: NN/MMM/TTT service. pump does not power-off,
disconnect external power
and remove the batteries.
3 Call the telephone number
shown on the display
message.
12. Troubleshooting
N = No V = Visual alarm
DIST. OCCLUSION N A/V The pump has 1 Press [SILENCE] to quiet
detected sustained the alarm for 1 minute.
distal pressure over 2 Press [STOP] to place the
the user-selected pump in the stop mode.
limit.
3 Check for sources of distal
To avoid a bolus occlusion:
infusion after an Kinked tubing below the
occlusion, complete pump
the following steps: Clamped patient-
access device
1 Press [STOP]. Clogged I.V. filter
2 Close the distal Other obstructions
clamp. 4 If the problem continues,
3 Press the remove and reinstall the
cassette release cassette.
button to remove CAUTION: Ensure all
the cassette from four cassette latches
the pump. are clearly visible after
4 Open the flow the cassette is
stop. installed.
5 Wait 10 seconds. Note: Ensure the tubing is
behind the tubing guide (at the
6 Close the flow
distal end). Refer to Loading
stop.
the Cassette on page 32.
7 Load the
5 Press [START] to begin
cassette into the
infusion.
pump.
8 Open the distal
clamp.
9 Review the
program.
10 Press [START]
to start the
pump.
12. Troubleshooting
N = No V = Visual alarm
LOW BATTERIES Y A/V The pump has 1 Press [SILENCE] to quiet
detected battery the alarm for 15 minutes.
voltage below 2 Press [STOP] to place the
acceptable levels and pump in the STOP mode.
the batteries are the
primary power 3 Connect the pump to AC
source. mains adaptor,
rechargeable battery pack,
Note: The pump OR replace both
allows the user to disposable batteries with
clear the LOW BAT- fresh AA batteries.
TERIES alarm. If 4 Press [START] to begin
the condition has not infusion.
been corrected after
15 minutes, the alarm
continues.
12. Troubleshooting
N = No V = Visual alarm
POWER LOSS N A/V The pump has 1 Connect the pump to AC
detected a power mains adaptor,
failure due to one of rechargeable battery pack,
the following: OR replace both
disposable batteries with
• Batteries have fresh AA batteries.
been removed
while the pump 2 Power-on the pump.
was on battery Review the program if
source. required.
• External power 3 Press [START] to begin
has failed or has infusion.
been
disconnected and
no batteries are
installed.
• Pump has been
powered on with
insufficient power
source.
12. Troubleshooting
N = No V = Visual alarm
PROX. OCCLUSION N A/V The pump has 1 Press [SILENCE] to quiet
detected sustained the alarm for 1 minute.
negative pressure on 2 Press [STOP] to place the
its proximal side. pump in the stop mode.
WARNING: When a
Proximal Occlusion alarm
activates, DO NOT
power-off the pump.
3 Check for sources of
proximal occlusion:
Kinked tubing above
pump
Blockage in the
container and in the
spike
Closed clamps above
pump
Empty container
High torque syringe
Other obstructions
4 Remove obstructions.
5 Press [START] to begin
infusion.
6 If the problem continues,
power-off the pump. Install
a new primed set and begin
infusion.
CAUTION: Ensure all
four cassette latches
are clearly visible after
the cassette is
installed.
Note: Ensure the tubing is
behind the tubing guides (at
the distal end). Refer to
Loading the Cassette on
page 32.
12. Troubleshooting
N = No V = Visual alarm
USING BATTERIES Y V This is a normal When powering on the pump:
response when:
• Press [YES/ENTER] to
• The pump has confirm disposable batteries
been powered on are the only power source.
with only
disposable When switching from AC mains
batteries as a power:
power source; or
• The pump has • Press [SILENCE] to clear
switched to the alarm.
internal power
(disposable
batteries) due to a
loss of external
power or a
reduction in
external power
below acceptable
levels.
USING EXT BATT Y V This is a normal When powering on the pump:
response when:
• Press [YES/ENTER] to
• The pump has confirm external battery
been powered on power source.
with external
batteries as a When switching from AC mains
power source; or power:
• The pump has
switched to • Press [SILENCE] to clear
external battery the alarm.
power due to a
loss of AC mains
power; or
• The pump has
switched to
external battery
power while
running on
internal batteries.
12. Troubleshooting
Alert Messages
Alert Display Message Is the Alar Cause/Condition Remedy
pump m
infusing?
Y = Yes A = Audible alarm
N = No V = Visual alarm
ALMOST EMPTY Y A/V Delivery will complete in • Press [SILENCE] to
30 minutes or less. quiet the alarm for 10
minutes.
For bolus only program, Note: This alert is
delivery will complete available only with
during the next bolus Pain Management
delivery. protocols.
CHECK PRINTER Y A/V • The printer has not 1 Press [SILENCE] to
responded to quiet the alarm for 2
commands from the minutes.
pump.
2 Ensure the printer is
• The printer is not properly connected
properly connected to to the pump by
the pump.
checking the
following:
Printer is powered
on and in ONLINE
mode.
Cable is plugged
in to both the
pump and the
printer.
Cable part
number and
printer settings
are correct.
3 If the problem
continues, select
STOP PRINT from
the Print Menu to
cancel all print jobs.
12. Troubleshooting
N = No V = Visual alarm
EMPTY CONTAINER N A/V The pump has delivered 1 Press [SILENCE] to
the programmed quiet the alarm for 2
container size or the minutes.
pump has delivered the 2 Press [STOP] to
programmed VTBI; AND place the pump in the
Auto KVO is delivering STOP mode and
(if available in current clear the alarm.
program).
3 Use the NEW
CONTAINER
function to repeat the
current program.
Refer to Starting a
New Container in
Chapter 2 Basic
Operation.
END OF INFUSION N A/V The pump has delivered If no KVO is selected:
the entire programmed
• The pump delivers protocol. 1 Press [SILENCE] to
KVO if selected. quiet the alarm for 2
minutes.
2 Press [STOP] to
place the pump in the
STOP mode and
clear the alarm.
If KVO is selected:
12. Troubleshooting
N = No V = Visual alarm
FLASHING DISPLAY N A/V A Non-programming Follow the instructions on
Callback has occurred the display to clear the
• Line four of the because the pump has alarm.
display flashes. been waiting for more
than three minutes for a
keypress on a non-
programming input or a
confirmation display.
NO INTERNAL BATT N A/V No disposable batteries 1 Press [YES/ENTER]
NO POWER LOSS or dead disposable to confirm.
ALARM AVAILABLE batteries; AND external
ENTER TO CONFIRM power. 2 Power-off the pump
and install two fresh,
disposable AA
batteries as a backup
power source.
PROGRAMMING N A/V Programming has not 1 Press any key to
INCOMPLETE been completed and no clear the alarm.
keys have been pressed 2 Complete the
within three minutes. program entries.
START N A/V A Stop Mode Callback 1 Press [SILENCE] to
has occurred because the quiet the alarm for 3
• START flashes on programmed pump has minutes.
the display. Other remained in the STOP
parts of the display 2 Follow the
mode for more than three instructions on the
are not affected.
minutes. display to clear the
alarm.
12. Troubleshooting
Chapter 13
Maintenance
Pump Storage
Store the GemStar Pump in a cool, dry place. Remove the disposable batteries and
the battery pack before storing the pump.
Program and event history are protected in the software memory for at least one year
when power is removed from the pump.
13. Maintenance
Note: NOT all cleaning solutions are disinfectants. Check product labeling.
13. Maintenance
CAUTION: Use care not to damage the silicone seals around the sensor bodies.
1 Remove the cassette, if installed.
D
2 Use a cotton swab, moistened with an
approved cleaning solution, to clean the
sensor faces (A), tubing channel (B),
plunger tip (C), and sensor pins (D).
3 Dry the sensor faces and tubing channel
and ensure the sensor faces are free of
detergent film and/or debris. A
C
Repair
The GemStar Pump has no user-serviceable components, with the exception of
disposable batteries.
If the display clock stops working, contact Hospira Technical Support Operations
to arrange for internal lithium battery replacement.
Homecare Customers: Call your healthcare professional or homecare company
regarding any required service or repairs. DO NOT attempt to repair the pump for
any reason.
The GemStar Pump is covered by a manufacturer’s warranty for one year after
purchase. During this time, opening the pump case for any reason voids this
warranty. Refer to Chapter 18 Warranty/Technical Assistance, page 189 for
details.
Refer all service to Hospira certified technicians only. A GemStar Technical Field
Service Manual is available to Hospira certified technicians.
Notes
Chapter 14
Operation Test
This test is designed to ensure the GemStar Pump is operating properly. Hospira
recommends performing this test a minimum of once every 12 months. Refer to
facility requirements to determine additional testing needs.
The test can be performed when the pump is in the STOP mode and requires
approximately 15 minutes to complete. The pump provides step-by-step instructions
through each section of the test.
Equipment Required
GemStar Pump
Two GemStar Microbore Pump Sets without a filter
Reservoir with at least 50 mL of water
20+ mL or larger graduated cylinder (graduated in 0.1 mL)
Two fresh, disposable AA batteries
Appropriate GemStar AC Mains Adaptor or other Hospira approved external
power source
Test Setup
Note: The pump clears any program and shift totals stored in pump memory before
starting the Operation Test.
1 Connect the external power source and install two fresh, disposable AA batteries.
2 Power-on the pump.
3 Press [OPTIONS].
4 Select HISTORIES from the Options Menu.
5 Select OPERATION TEST from the Histories Menu.
6 Follow the instructions on the display.
Tip: The checklist at the end of this section may be copied to record test results.
Case Inspection
1 Visually inspect the case and keypad for damage. Check the case for missing
screws, or labels.
2 Enter the results:
• If the pump is free of damage and all the screws and labels are in place, press
[YES/ENTER].
• If damage is found on the case or keypad, or any screws or labels are missing,
press [NO].
A
B B
E F C
Power Check
Tip: Opening the battery compartment door will disconnect the batteries enough for
the pump to recognize external power. Then the door can be tightened and testing
can continue. This can save time when performing multiple tests on a group of
pumps.
1 The display indicates that an external power source and batteries are connected.
Press .
2 Remove battery door and press . The display indicates that only external
power is connected. Press .
3 Return the battery door to the closed position.
4 Disconnect the external power source and press .
5 The display indicates that only batteries are connected and displays the test
results. Press .
6 If the pump fails, reconnect the external power source and press to rerun the
test, or press [YES/ENTER] to confirm failure.
Keypad Test
1 Press each key, including the Bolus Button. Press [YES/ENTER] last.
2 Enter the results:
• If each key makes an audible tone, press [YES/ENTER].
• If any key does not make an audible tone, press [NO].
Display Test
1 Press [YES/ENTER] when prompted. The display fills with solid boxes for
approximately 10 seconds. To view the solid boxes for another 10 seconds after
the display times-out, press .
2 Enter the results:
• If the display fills with solid boxes, press [YES/ENTER].
• If the display does not fill with solid boxes, press [NO].
LED Test
1 Verify that both the Power (green) and Alarm (red) LEDs are lit. Refer to the
GemStar Pump Layout in the “Introduction” section for the location of the LEDs.
2 Enter the results:
• If both LEDs are lit, press [YES/ENTER].
• If either of the LEDs does not light, press [NO].
Tip: Use two microbore pump sets to save time during the next four tests. The first
set should be fully primed BEFORE beginning the volume accuracy test. Use a
second, DRY set for the air-in-line test.
When the proximal occlusion test is complete, remove the WET set from the pump
and insert the DRY set. Press [START] to begin the air-in-line test. The WET set can
then be placed in the next unit ready for testing.
1 Clamp the pump set 10 to 12 inches (25.4 to 30.5 cm) below the cassette.
2 Press [START].
3 The test result displays. Release the clamp and press .
• If the test is passed, the pump advances to the next test.
• If the test is failed, clean the sensors and press to rerun the test or press
[YES/ENTER] to confirm failure.
Note: Clean the sensor pins before repeating the distal occlusion test. Refer to
Chapter 13 Maintenance for instructions.
Note: Clean the sensor pins before repeating the proximal occlusion test. Refer to
Chapter 13 Maintenance for instructions.
Air-in-line Test
The air-in-line alarm should occur within approximately 1 minute to pass this test.
Note: Clean the sensor faces before repeating the air-in-line-test. Refer to Chapter 13
Maintenance for instructions.
Note: For maximum battery life, operate the pump on AC mains power when con-
nected to a printer or computer.
1 Connect the serial printer cable to the pump. Refer to Connecting to the Seiko
Printer in Chapter 10 Options Menu, on page 115.
2 Press [YES/ENTER] to print the test results.
3 The pump displays the Programming Menu when the results have finished
transmitting or if [NO] is pressed.
GemStar®
Operation Test Checklist
Pass Fail
Case Inspection
Power Check
Keypad Check
Display Test
LED Test
Air-in-line Test
Serial Number:
Notes
Chapter 15
CAUTION:When removing the cord from the bottom of the pump, firmly grasp
the connector and pull straight out. DO NOT twist or bend the cord
or connector.
Note: For the GemStar Universal Lockbox (List 12499-01) refer to the lockbox for
instructions.
A
CAUTION: Ensure all four
3 Align the pump with the lockbox cassette latches are clearly
visible after the cassette is
(C). installed. Refer to Loading the
4 Slide the pump down into the Cassette on page 32.
lockbox until fully seated.
B
F
E
10 Ensure the tubing will not be
pinched when closing the lockbox.
D
11 Close and lock the lockbox.
12 Remove the key.
150
100
50
0
8 Place the case around the waist with the front “window” facing away from the
body. Snap the buckle together to secure. Adjust the belt for a snug, comfortable
fit.
G
K
H
Note: This carrying case includes an interior support (Hospira 3 Liter Carrier List
13590-01) that may be removed from the case and used separately.
1 Unzip the carrying case and release all Velcro straps and flaps.
2 Place the fluid bag on the carrying case as shown (A).
3 Hook top strap (B) through the hole in the top of the fluid bag and secure.
4 Fold small flaps (C) and (D) over the bag as shown and secure.
5 Secure straps (E) and (F) across the length of the fluid bag and over the small
flaps.
6 Secure the spike connection with strap (G).
7 Fold large flaps (H) and (I) over as shown and secure.
8 Place the pump, with the keypad facing upward, inside the pocket (J) on the flap.
The pump’s display should be visible through the opening on the front of the
pocket.
9 Secure the pump with straps (K) and (L).
10 Secure the tubing with strap (M).
11 Place the secured pump (N) inside the large zipper case or other carrying case.
12 Zip the large case closed.
13 If desired, route the tubing through the loops on the bottom of the carrying case.
Ensure the tubing is not kinked.
14 Adjust the shoulder straps and waist belt for a snug, comfortable fit.
Tip: When not using the waist belt, secure it under the large flap on the back of
the case.
A
B
C
I H
F
E
G
J K
L
M
N
Note: The docking station light is yellow while the docking station is charging. When
the docking station is fully charged, the light is green.
2 Plug the printer cable into the printer port on the rear,
right side of the docking station.
Chapter 16
Pump Sets
CAUTION: USE ONLY GemStar Pump Sets with the GemStar Pump. Use of
unauthorized sets may result in injury to the patient or damage to
the pump.
Contact a Hospira Representative for appropriate pump set configurations.
Fluid path and areas beneath undisturbed protective set covers are
sterile and nonpyrogenic in the intact unit package. STERILE 2
The GemStar Pump Sets, listed in the table on the following page
are available for use with the GemStar Pump.
Note: The disposable pump sets should be changed per the Centers for Disease
Control (CDC) guidelines or health care provider policy. Discard after use.
All GemStar Pump Sets are composed of latex-free components only.
GemStar Pump Sets contain an anti-siphon valve to reduce the potential for gravity
flow at a typical 36-inch head height.
The following sets are a representation of the GemStar® Pump Sets currently
available:
Ambulatory Sets
13263 Latex-Free GemStar® Pump Set and Extension Set with integral
pressure-activated anti-siphon valve, Secure Lock, Nonvented, Microbore
tubing, 96 in (244 cm)
13264 Latex-Free, LifeShield®, GemStar® Pump Set with 0.2 Micron Filter, Secure
Lock, Prepierced reseal injection site, Nonvented, Pre-attached
pressure-activated anti-siphon valve, Microbore tubing, 96 in (244 cm)
13265 Latex-Free, LifeShield®, GemStar® Pump Set with 1.2 Micron filter, Secure lock,
Prepierced reseal injection site, Nonvented, Pre-attached pressure-activated
anti-siphon valve, NonPhthalate (Non-DEHP) microbore tubing, 96 in (244 cm)
13273 Latex-Free GemStar® Pump Set, Secure lock, Non vented, Pre-attached
pressure-activated anti-siphon valve, Microbore tubing, 96 in (244 cm)
General Infusion Sets
13244 Latex-Free, GemStar® HEMA® II, Y-Type Blood Set, 210 Micron Filter, Lower
CLAVE® Connector, Macrobore tubing, 110 in (279 cm)
13267 Latex-Free, LifeShield®, GemStar® Primary Piggyback Pump Set, Secure Lock,
Convertible, Macrobore Tubing Piercing Pin, 2 CLAVE® connectors,
Pre-attached pressure-activated anti-siphon valve, Macrobore tubing, 110 in
(279 cm)
13272 Latex-Free, LifeShield®, GemStar® 150 x 60 micro drip SoluSet®, Option Lock,
Convertible Piercing Pin, 2 CLAVE® connectors, Macrobore tubing, 110 in
(279 cm)
13274 Latex-Free, LifeShield®, GemStar® Piggyback Pump Set, with 0.2 Micron Filter,
Secure Lock, Convertible Piercing Pin, 3 CLAVE® Connectors, Pre-attached
pressure-activated anti-siphon valve, Macrobore tubing, 110 in (279 cm)
Notes
Chapter 17
List 13026-01 Wall-mount AC mains adaptor; 6 ft. (1.8 m) cord molded plug
(USA) complies with UL 544 Input: 110 VAC
Output: 3.0 VDC
List 13072-01 Tabletop AC mains adaptor; molded plug
(Universal) complies with CSA 22.2 Input: 100–240 VAC
No. 601.1 and UL 60601-1 Output: 3.0 VDC
Battery: Two disposable AA alkaline or lithium batteries
Air Sensitivity
ON: Pump alarms at approximately 0.5 mL of air
Pump Self-Tests and Safety Self-test performed when the power switch is activated
Features
Diagnostic routine, including motor speed and air-detection
monitoring, is repeated continuously while the pump is
powered on
Occlusion Information
CAUTION: When using pre-filled vials, the system may underdeliver at low
flow rates.
Note: False occlusion alarms and fluid delivery inconsistencies may occur when
pumping viscous fluids in microbore tubing at rates greater than 500 mL/hr.
Trumpet Curves
The Trumpet Curve Graphs following the Example show representative maximum
and minimum percent flow rate deviation from the programmed rate over time. This
information was developed in accordance with IEC 60601-2-24: 1998, Sub-Clause
50.102. Refer to this standard for detailed information.
How to read a Trumpet Curve Graph (Refer to example on the following page):
The graphs following the Example plot flow rates at 30 second intervals for the first 2
hours and for the 96th hour of delivery. The graph plots mean delivery rate error
(Average of 3 pumps) for the 2nd hour and the 96th hour as a straight line. The graph
also presents maximum and minimum average delivery rate error for this interval
plotted by averaging delivery errors over intervals of 2, 5, 11, 19 and 31 minutes
(“Trumpet Curve”).
Example
From the Trumpet Curve Graph sample that follows, find the 5 minute interval (A) at
the horizontal axis and read the corresponding points (B) and (C) on the vertical axis.
The values are approximately +2.8% and -0.5%.
This means that at the rate of 25 mL/hr the average maximum flow rate fluctuation for
any 5 minute time interval during the 2nd hour of operation was within the limits of
+2.8% and -0.5% from the nominal rate. The average delivery rate error over the
entire 2nd hour was +1.6% (D).
For other time intervals look at other points at the horizontal axis and determine
corresponding limits as above.
A trained professional can use the resulting graphs to select a pump with the
appropriate startup and flow characteristics to suit the clinical application.
Note: As an example of how the trumpet curves can be used, consider the maximum
and minimum deviations at the 5 minute average interval. The upper curve provides
the maximum expected delivery rate error over a 5 minute interval, the lower curve
provides the minimum expected delivery rate error over a 5 minute interval. An
example would be Dopamine administered at 5 µgm/kg/min. At 5 minutes, the average
drug delivery error would be within the range of +2.8% and -0.5% of the expected
nominal rate.
FLOW RATE
0.1 mL, 0mm Hg Delivery Head, Start Up
Rate (mL/hr)
Time (minutes)
— Actual Rate --- Programmed Rate
PERCENT ERROR
0.1 mL, 0 mm Hg Delivery Head, Start Up
Flow Rate Variability (%)
FLOW RATE
0.1 mL, 0mm Hg Delivery Head, Hour 96
Rate (mL/hr)
Time (minutes)
— Actual Rate --- Programmed Rate
PERCENT ERROR
0.1 mL, 0 mm Hg Delivery Head, Hour 96
Flow Rate Variability (%)
FLOW RATE
1 mL, 0mm Hg Delivery Head, Start Up
2.0
1.8
1.6
1.4
1.2
1.0
Rate (mL/hr)
0.8
0.6
0.4
0.2
0.0
-0.2
-0.4
0 10 20 30 40 50 60 70 80 90 100 110 120
Time (minutes)
— Actual Rate --- Programmed Rate
PERCENT ERROR
1 mL, 0 mm Hg Delivery Head, Start Up
30.0%
Flow Rate Variability (%)
20.0%
10.0%
0.0%
-10.0%
-20.0%
-30.0%
0 5 10 15 20 25 30
FLOW RATE
1 mL, 0mm Hg Delivery Head, Hour 96
2,0
1,8
1,6
1,4
1,2
1,0
Rate (mL/hr)
0,8
0,6
0,4
0,2
0,0
-0,2
-0,4
0 10 20 30 40 50
Time (minutes)
— Actual Rate --- Programmed Rate
PERCENT ERROR
1 mL, 0 mm Hg Delivery Head, Hour 96
30.0%
Flow Rate Variability (%)
20.0%
10.0%
0.0%
-10.0%
-20.0%
-30.0%
0 5 10 15 20 25 30
FLOW RATE
25 mL, 0mm Hg Delivery Head, Start Up
50.0
45.0
40.0
35.0
30.0
Rate (mL/hr)
25.0
20.0
15.0
10.0
5.0
0.0
-5.0
-10.0
0 10 20 30 40 50 60 70 80 90 100 110 120
Time (minutes)
— Actual Rate --- Programmed Rate
PERCENT ERROR
25 mL, 0 mm Hg Delivery Head, Start Up
15.0%
Flow Rate Variability (%)
10.0%
5.0%
0.0%
-5.0%
-10.0%
-15.0%
0 5 10 15 20 25 30
FLOW RATE
25 mL, 0mm Hg Delivery Head, 96 Hour
50.0
45.0
40.0
35.0
30.0
Rate (mL/hr)
25.0
20.0
15.0
10.0
5.0
0.0
-5.0
-10.0
0 10 20 30 40 50
Time (minutes)
— Actual Rate --- Programmed Rate
PERCENT ERROR
25 mL, 0 mm Hg Delivery Head, 96 Hour
15.0%
Flow Rate Variability (%)
10.0%
5.0%
0.0%
-5.0%
-10.0%
-15.0%
0 5 10 15 20 25 30
FLOW RATE
25 mL, -100mm Hg Delivery Head, Start Up
50.0
45.0
40.0
35.0
30.0
25.0
Rate (mL/hr)
20.0
15.0
10.0
5.0
0.0
-5.0
-10.0
0 10 20 30 40 50 60 70 80 90 100 110 120
Time (minutes)
— Actual Rate --- Programmed Rate
PERCENT ERROR
25 mL, -100 mm Hg Delivery Head, Start Up
15.0%
Flow Rate Variability (%)
10.0%
5.0%
0.0%
-5.0%
-10.0%
-15.0%
0 5 10 15 20 25 30
FLOW RATE
25 mL, +100mm Hg Delivery Head, Start Up
50.0
45.0
40.0
35.0
30.0
Rate (mL/hr)
25.0
20.0
15.0
10.0
5.0
0.0
-5.0
-10.0
0 10 20 30 40 50 60 70 80 90 100 110 120
Time (minutes)
— Actual Rate --- Programmed Rate
PERCENT ERROR
25 mL, +100 mm Hg Delivery Head, Start Up
15.0%
10.0%
Flow Rate Variability (%)
5.0%
0.0%
-5.0%
-10.0%
-15.0%
0 5 10 15 20 25 30
FLOW RATE
25 mL, 0mm Hg Delivery Head, -20 in H20 Filling Head, Start Up
50.0
45.0
40.0
35.0
30.0
25.0
Rate (mL/hr)
20.0
15.0
10.0
5.0
0.0
-5.0
-10.0
0 10 20 30 40 50 60 70 80 90 100 110 120
Time (minutes)
— Actual Rate --- Programmed Rate
PERCENT ERROR
25 mL, 0 mm Hg Delivery Head, -20 in H20 Filling Head, Start Up
15.0%
Flow Rate Variability (%)
10.0%
5.0%
0.0%
-5.0%
-10.0%
-15.0%
0 5 10 15 20 25 30
Notes
Chapter 18
Warranty/Technical Assistance
The GemStar Pump has been carefully manufactured using high-quality components.
It is warranted to be free from defects in material and workmanship for a period of one
year from the date of purchase under normal use and service. The warranty on the
optional battery pack and other accessories is limited to 90 days.
On return to Hospira, the pump will be repaired or replaced within the terms of this
warranty. Material returned to Hospira must be properly packaged and sent freight
prepaid.
This warranty shall not apply if the pump has been repaired by anyone other than
Hospira qualified service personnel; or altered in any way which, in Hospira’s
judgment, affects its stability or reliability; or if the serial number has been altered,
effaced, or removed.
Note: Outside the U.S, contact your local Hospira sales office.
Index
Index
C programming of, 69
call back alarm, 61 Programming Worksheet, 75
carrying cases rate titration, 71
using the Abbott carrying cases, 159–164 contraindications for use, 11
cassette, 30
cassette pocket, 4 D
cassette release button, 4 Data Port, connecting to, 112–117
cleaning of, 139–141 configuration settings, 113
error message, 129 connecting to a computer, 116
inspection of, 145 connecting to the Seiko printer, 115
loading the cassette, 32 connector, 4
releasing the cassette, 33 disconnecting the pump, 23
warnings and cautions, 13 distal end, 30, 35, 147
Change History, ii distal occlusion, 6, 106
CHANGE key, 3, 28 alarm message, 130
Change Menu, 8 occlusion sensitivity, 173
clinician instructions, 121–125 occlusion sensor setting, 107
adding a second lock level, 123 test of, 148
locking the pump, 122 troubleshooting, 130
program and deliver a clinician activated docking station, 154
loading dose, 124 connecting the bolus cord and printer
unlocking the pump, 124 cable, 166
configuration settings, data port, 113 connecting the pump, 165
connecting the Abbott GemStar AC Mains disconnecting the pump, 166
Adaptor, 20 dose limit, 55
connecting the bolus cord, 156 drip chamber
connecting to a computer, 116 priming pump sets with a drip
See also Data Port chamber, 38
connecting to the Seiko printer, 115 priming pump sets with a drip chamber
See also Data Port using the PURGE Key, 40
continuous delivery
in Pain Management therapy, 53–58 E
in TPN therapy, 47, 50 event log, 6
Continuous therapy, 5, 69–75 history, display, 100
changing a Continuous Program, 74 print, 98
Clinician instructions, 122, 123
features and specifications, 69 F
piggybacking in Continuous flow stop, 30, 33, 34, 36, 38, 40, 42, 45, 174
therapy, 71–74 Full Lock, 104
programmable ranges, 75 lock level functions, 105
Index
Index
Index
Index
START key, 3 U
starting a new container, 29 unlocking the pump, 124
STOP key, 3 UP/DOWN key, 28
STOP mode, 9
V
T Variable Time therapy, 5, 91–96
taper down, 47, 48, 50 changing a Variable Time program, 94
taper up, 47 delivery interruptions, 93
technical assistance, 190 features and specifications, 91
Therapies, 5 programmable ranges, 95
Continuous, 69–75 programming of, 91
Intermittent, 61–67 Programming Worksheet, 95
mL/hr Only, 83–90 Volume To Be Infused. See VTBI
Pain Management, 53–60 VTBI, 70, 78, 84
Total Parenteral Nutrition (TPN), 47–51 piggyback VTBI, 71–73, 86–88
Variable Time, 91–96 VTBI Titration, 80, 86
Weight Dosed, 77
Therapy Lock, 104 W
lock level functions, 105 warnings and cautions, 11
Locking Sequence Instructions, 122 Weight Dosed therapy, 5, 77
unlocking the pump, 124 Auto KVO, 79
Titrate Rates, 5 changing a Weight Dosed program, 80
Titrate VTBI, 5 features and specifications, 77
Total Parenteral Nutrition. See TPN programmable ranges, 81
therapy programming of, 78
TPN therapy, 5, 47–51 programming worksheet, 81
Auto Taper Down, 50 rate titration, 79
changing a TPN program, 49 VTBI Titration, 80
features and specifications, 47
programmable ranges, 51 Y
programming of, 48 YES/ENTER key, 3, 28, 47, 54, 61, 69, 78,
Programming Worksheet, 51 84, 91
troubleshooting, 127–138
alarm messages, 128–134
alert messages, 135
other display messages, 138
CAUTION: Federal (USA) law restricts this device to sale by or on the order of
a physician or other licensed practitioner.
GemStar, LifeShield, HEMA, and CAIR are registered trademarks of Hospira, Inc. Luer Lock is trademark of Hospira,
Inc., Super Edisonite, Seiko, Vesphene II se, Manu-Klenz, Formula C, Teflon, Velcro, CLAVE, and Microsoft,
Windows, and Windows NT are not trademarks of Hospira.
This document and the subject matter disclosed herein are proprietary information. Hospira retains all the exclusive
rights of dissemination, reproduction, manufacture and sale. Any party using this documentation accepts it in
confidence, and agrees not to duplicate it, in whole or in part, nor disclose it to others without the written consent of
Hospira.