0% found this document useful (0 votes)
361 views2 pages

SCCPLCC20241223-033-Feeler Gauge-0.025 To 0.635 MM - 26 Blades

This calibration certificate documents the calibration of a feeler gauge for M/s. Vaishnavi Laminates & Wood Works Pvt. Ltd., performed on December 25, 2024, with results indicating the instrument is in acceptable condition. The next calibration is suggested for December 24, 2025, and the certificate includes detailed calibration results for various nominal values and their corresponding observed values. The certificate is valid under specified conditions and includes important notes regarding its use and limitations.

Uploaded by

gupat.zee
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
361 views2 pages

SCCPLCC20241223-033-Feeler Gauge-0.025 To 0.635 MM - 26 Blades

This calibration certificate documents the calibration of a feeler gauge for M/s. Vaishnavi Laminates & Wood Works Pvt. Ltd., performed on December 25, 2024, with results indicating the instrument is in acceptable condition. The next calibration is suggested for December 24, 2025, and the certificate includes detailed calibration results for various nominal values and their corresponding observed values. The certificate is valid under specified conditions and includes important notes regarding its use and limitations.

Uploaded by

gupat.zee
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 2

CALIBRATION CERTIFICATE

Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLCC20241223-033 ULR No.: CC328724000017253F


Group: Dimension (Basic Measuring Instruments, Gauge etc.) Discipline: Mechanical
Customer’s Name & Address Page No.: 1 of 2 Calibration Performed: At Lab
M/s. Vaishnavi Laminates & Wood Works Pvt. Ltd. Date of Receipt: 23-12-2024
Survey No. 211, Ghoghli Tampa Panchmukhi Service Request No.: SCCPLSRN20241223-002
Hanuman Mandir, Klmeshwar Khapari, Nagpur Certificate Issue Date: 26-12-2024
Date of Calibration: 25-12-2024
Next Due Date (Suggested by Customer): 24-12-2025
Equipment Description (UUC) & Identification
Name of the Instrument: Feeler Gauge Make: Vetpex
Model: ------
Range: 0.025 to 0.635 mm Sr. No.: ------
Least Count: ------ I.D. No.: FG-01
Condition of Instrument: Ok Location: ------
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital External 0 To 25 mm & SCCPL SCCPLCC20240113-004 14-01-2025
Micrometer SCCPL/CAL/016
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 20 ± 2 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 005 IS 3179
CALIBRATION RESULTS
Sr. No. Nominal Value of Observed Ref. Standard Error Expanded Uncertainty
UUC (mm) Value (mm) (mm) (±) mm
1 0.038 0.040 -0.002 0.0022
2 0.051 0.052 -0.001 0.0022
3 0.064 0.065 -0.001 0.0022
4 0.076 0.078 -0.002 0.0022
5 0.102 0.104 -0.002 0.0022
6 0.127 0.125 0.002 0.0022
7 0.152 0.151 0.001 0.0022
8 0.178 0.181 -0.003 0.0022
9 0.203 0.208 -0.005 0.0022
10 0.229 0.233 -0.004 0.0022
11 0.254 0.257 -0.003 0.0022
12 0.279 0.279 -0.004 0.0022
13 0.305 0.308 -0.003 0.0022

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLCC20241223-033 ULR No.: CC328724000017253F
Group: Dimension (Basic Measuring Instruments, Gauge etc.) Discipline: Mechanical
Customer’s Name & Address Page No.: 2 of 2 Calibration Performed: At Lab
M/s. Vaishnavi Laminates & Wood Works Pvt. Ltd. Date of Receipt: 23-12-2024
Survey No. 211, Ghoghli Tampa Panchmukhi Service Request No.: SCCPLSRN20241223-002
Hanuman Mandir, Klmeshwar Khapari, Nagpur Certificate Issue Date: 26-12-2024
Date of Calibration: 25-12-2024
Next Due Date (Suggested by Customer): 24-12-2025
CALIBRATION RESULTS
Sr. No. Nominal Value of Observed Ref. Standard Error Expanded Uncertainty
UUC (mm) Value (mm) (mm) (±) mm
14 0.330 0.333 -0.003 0.0022
15 0.356 0.352 0.004 0.0022
16 0.381 0.383 -0.002 0.0022
17 0.406 0.408 -0.002 0.0022
18 0.432 0.435 -0.003 0.0022
19 0.457 0.459 -0.002 0.0022
20 0.483 0.481 0.002 0.0022
21 0.508 0.505 0.003 0.0022
22 0.533 0.539 -0.006 0.0022
23 0.559 0.554 0.005 0.0022
24 0.584 0.586 -0.002 0.0022
25 0.610 0.615 -0.005 0.0022
26 0.635 0.641 -0.006 0.0022
***End of Certificate***

Calibrated By Reviewed/Approved By
Hemraj L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
9. Scan this QR code to download our NABL Accreditation Scope & Certificate.

You might also like