CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLCC20250226-001 ULR No.: CC328725000006131F
Group: Temperature Discipline: Thermal
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Site
M/s. Tabilac Pharmaceuticals Pvt. Ltd. Date of Receipt: 26-02-2025
Plot No-1396, HSIIDC Ind Area, Rai, Sonipat, Haryana-131029 Service Request No.: SCCPLSRN20250226-001
Certificate Issue Date: 28-02-2025
Date of Calibration: 27-02-2025
Next Due Date (Suggested by Customer): 26-02-2026
Equipment Description (UUC) & Identification
Name of the Instrument: Digital Temperature Indicator/ Make: Rising Sun Enterprises
Controller with Sensor (Vertical Autoclave) Model: RSE-401A3
Range: Ambient to 150 ⁰C Sr. No.: 40A13201
Least Count: 0.1 ⁰C I.D. No.: ------
Condition of Instrument: Ok Location: ------
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 High Precision -100 ⁰C to 600 ⁰C & AFSS Technologies CC350824000000159F 05-12-2025
Thermometer with 4 SCCPL/CAL/022A & Private Limited
wire RTD Sensor SCCPL/CAL/022B
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 20 ± 15 ⁰C, Humidity: 30 to 75 %RH SCCPL 03 SOP 026 DKD-R 5-7
CALIBRATION RESULTS
Sr. No. Set Value UUC Value Ref. Standard Error Expanded
(⁰C) (⁰C) Value (⁰C) (⁰C) Uncertainty (± ⁰C)
1 70 70.0 70.095 -0.09 0.57
2 100 100.0 100.154 -0.154 0.57
3 121 121.1 121.377 -0.277 0.69
***End of Certificate***
Calibrated By Reviewed/Approved By
Amit Kumar L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2 .
9. Scan this QR code to download our NABL Accreditation Scope & Certificate.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLCC20250226-002 ULR No.: CC328725000006132F
Group: Temperature Discipline: Thermal
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At site
M/s. Tabilac Pharmaceuticals Pvt. Ltd. Date of Receipt: 26-02-2025
Plot No-1396, HSIIDC Ind Area, Rai, Sonipat, Haryana-131029 Service Request No.: SCCPLSRN20250226-001
Certificate Issue Date: 28-02-2025
Date of Calibration: 27-02-2025
Next Due Date (Suggested by Customer): 26-02-2026
Equipment Description (UUC) & Identification
Name of the Instrument: Digital Temperature Indicator/ Make: Rising Sun Enterprises
Controller with Sensor (Bacteriological Incubator) Model: RSE-428A7
Range: Ambient to 90 ⁰C Sr. No.: 428A72041
Least Count: 0.1 ⁰C I.D. No.: -------
Condition of Instrument: Ok Location: -------
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 High Precision -100 ⁰C to 600 ⁰C & AFSS Technologies CC350824000000159F 05-12-2025
Thermometer with 4 SCCPL/CAL/022A & Private Limited
wire RTD Sensor SCCPL/CAL/022B
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 20 ± 15 ⁰C, Humidity: 30 to 75 %RH SCCPL 03 SOP 026 DKD-R 5-7
CALIBRATION RESULTS
Sr. No. Set Value UUC Value Ref. Standard Error Expanded
(⁰C) (⁰C) Value (⁰C) (⁰C) Uncertainty (± ⁰C)
1 30 30.0 30.075 -0.075 0.38
2 35 35.0 35.143 -0.143 0.38
3 37 37.0 37.269 -0.269 0.38
4 45 45.2 45.544 -0.344 0.38
5 60 60.1 60.532 -0.432 0.57
***End of Certificate***
Calibrated By Reviewed/Approved By
Amit Kumar L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
9. Scan this QR code to download our NABL Accreditation Scope & Certificate.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLCC20250226-003 ULR No.: CC328725000006133F
Group: Temperature Discipline: Thermal
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Site
M/s. Tabilac Pharmaceuticals Pvt. Ltd. Date of Receipt: 26-02-2025
Plot No-1396, HSIIDC Ind Area, Rai, Sonipat, Haryana-131029 Service Request No.: SCCPLSRN20250226-001
Certificate Issue Date: 28-02-2025
Date of Calibration: 27-02-2025
Next Due Date (Suggested by Customer): 26-02-2026
Equipment Description (UUC) & Identification
Name of the Instrument: Digital Temperature Indicator/ Make: Rising Sun Enterprises
Controller with Sensor (BOD Incubator) Model: RSE-405A1
Range: 5 to 60 ⁰C Sr. No.: 405A12062
Least Count: 0.1 ⁰C I.D. No.: -------
Condition of Instrument: Ok Location: ------
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 High Precision -100 ⁰C to 600 ⁰C & AFSS Technologies CC350824000000159F 05-12-2025
Thermometer with 4 SCCPL/CAL/022A & Private Limited
wire RTD Sensor SCCPL/CAL/022B
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 20 ± 15 ⁰C, Humidity: 30 to 75 %RH SCCPL 03 SOP 026 DKD-R 5-7
CALIBRATION RESULTS
Sr. No. Set Value UUC Value Ref. Standard Error Expanded
(⁰C) (⁰C) Value (⁰C) (⁰C) Uncertainty (± ⁰C)
1 25 25.0 25.097 -0.097 0.38
2 27 27.0 27.176 -0.176 0.38
3 30 30.1 30.351 -0.251 0.38
4 35 35.1 35.468 -0.368 0.38
5 40 40.0 40.475 -0.475 0.38
***End of Certificate***
Calibrated By Reviewed/Approved By
Amit Kumar L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising me dia
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.
9. Scan this QR code to download our NABL Accreditation Scope & Certificate.
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLTP20250226-001
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Site
M/s. Tabilac Pharmaceuticals Pvt. Ltd. Date of Receipt: 26-02-2025
Plot No-1396, HSIIDC Ind Area, Rai, Sonipat, Haryana-131029 Service Request No.: SCCPLTPSRN20250226-001
Certificate Issue Date: 28-02-2025
Date of Calibration: 27-02-2025
Next Due Date (Suggested by Customer): 26-02-2026
Equipment Description (UUC) & Identification
Name of the Instrument: Colony Counter Make: Rising Sun Enterprises
Model: RSEL-36
Range: 0 to 9999 Count Sr. No.: 362050
Least Count: 1 Count I.D. No.: -----
Condition of Instrument: Ok Location: -----
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Tachometer 2.5 To 99999 RPM & Asian Technology AT24000000170 25-02-2025
SCCPL/CAL/014
2 Digital Stop Watch 0 to 24 Hrs. & Testncal Laboratory CC405825000001703F 26-01-2026
SCCPL/SW-01
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP MIS 015 -----
CALIBRATION RESULTS
Sr. No. UUC Value Ref. Standard Value Expanded Uncertainty
(Count) (Count) (± )
1 1 1.0
2 10 10.0
3 100 100.0 1%
4 300 300.0
5 1000 1000.0
6 9999 9999.0
***End of Certificate***
Calibrated By Reviewed/Approved By
Amit Kumar L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2 .
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLTP20250226-002
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Site
M/s. Tabilac Pharmaceuticals Pvt. Ltd. Date of Receipt: 26-02-2025
Plot No-1396, HSIIDC Ind Area, Rai, Sonipat, Haryana-131029 Service Request No.: SCCPLTPSRN20250226-001
Certificate Issue Date: 28-02-2025
Date of Calibration: 27-02-2025
Next Due Date (Suggested by Customer): 26-02-2026
Equipment Description (UUC) & Identification
Name of the Instrument: Conductivity meter Make: Rising Sun Enterprises
Model: RSE-215R
Range: 0 to 20.00, 200.0, 2000 µS/cm Sr. No.: 215R2024
Least Count: 0.01, 0.1, 1 µS/cm I.D. No.: ------
Condition of Instrument: Ok Location: -----
Master Equipment Used :
Sr. No. Standard Used Certified By Certificate. No. Valid Up to
1 Standard Conductivity Solution 84 μS/cm APMIL 882485 (Product Code) Sep-2025
2 Standard Conductivity Solution 1413 μS/cm APMIL 890697 (Product Code) Oct-2025
Environmental Conditions Calibration Procedure
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP MIS 014
CALIBRATION RESULTS
Sr. No. Standard Value Observed UUC Value Expanded Uncertainty
(µs/cm) (µs/cm)
1 87.5 87.9
± 0.8 %
2 1420 1425
***End of Certificate***
Calibrated By Reviewed/Approved By
Amit Kumar L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2 .
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLTP20250226-003
Group: Dimension (Basic Measuring Instruments, Gauge etc.) Discipline: Mechanical
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Site
M/s. Tabilac Pharmaceuticals Pvt. Ltd. Date of Receipt: 26-02-2025
Plot No-1396, HSIIDC Ind Area, Rai, Sonipat, Haryana-131029 Service Request No.: SCCPLTPSRN20250226-001
Certificate Issue Date: 28-02-2025
Date of Calibration: 27-02-2025
Next Due Date (Suggested by Customer): 26-02-2026
Equipment Description (UUC) & Identification
Name of the Instrument: Antibiotic Zone Reader Make: Rising Sun Enterprises
Model: RSE-662A1
Range: 0 to 35 mm Sr. No.: 662A12040
Least Count: 0.2 mm I.D. No.: ------
Condition of Instrument: Ok Location: ------
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Slip Gauge Block Set 0.5 mm To 100 mm & Tanson Instrument TI/K/SGS/017/24 28-11-2026
(Grade-0) SCCPL/CAL/001
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 25 ± 5 ⁰C, Humidity: 50 ± 10 %RH SCCPL 03 SOP 015 --------
CALIBRATION RESULTS
Sr. No. Ref. Standard Value Observed UUC Value Error Expanded Uncertainty
(mm) (mm) (mm) (±) mm
1 0 0.0 0.0 0.15
2 10 10.0 0.0 0.15
3 15 15.0 0.0 0.15
4 25 25.0 0.0 0.15
5 30 29.9 -0.1 0.15
6 35 34.9 -0.1 0.15
***End of Certificate***
Calibrated By Reviewed/Approved By
Amit Kumar L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2 .
CALIBRATION CERTIFICATE
Format No.: SCCPL-QF-7.8.1 Certificate No.: SCCPLTP20250226-004
Group: Acceleration & Speed Discipline: Mechanical
Customer’s Name & Address Page No.: 1 of 1 Calibration Performed: At Site
M/s. Tabilac Pharmaceuticals Pvt. Ltd. Date of Receipt: 26-02-2025
Plot No-1396, HSIIDC Ind Area, Rai, Sonipat, Haryana-131029 Service Request No.: SCCPLTPSRN20250226-001
Certificate Issue Date: 28-02-2025
Date of Calibration: 27-02-2025
Next Due Date (Suggested by Customer): 26-02-2026
Equipment Description (UUC) & Identification
Name of the Instrument: Fume Hood Make: Rising Sun Enterprises
Model: RSE-415A11
Range: As per Instrument Sr. No.: 415A112035
Least Count: As per Results I.D. No.: ------
Condition of Instrument: Ok Location: ----
Master Equipment Used
Sr. No. Instrument’s Name Range & I.D. No. Calibration Agency Calibration Certificate No. Valid Up to
1 Digital Anemometer 0 to 30 m/s & SCCPL SCCPLSC20241228-005 29-12-2025
SCCPL/DAM/001
Environmental Conditions Calibration Procedure Calibration Reference Standard
Temperature: 15 to 45 ⁰C, Humidity: 30 to 80 %RH SCCPL 03 SOP 033 ----
CALIBRATION RESULTS
Parameter Observed Ref. Average Value Ref. Requirement Expanded
Standard Value (FPM) Standard Value (FPM) Uncertainty (±)
93
95
Air Velocity (FPM) 90 92.8 60 FPM to 100 FPM 3 % of rdg.
94
92
***End of Certificate***
Calibrated By Reviewed/Approved By
Amit Kumar L.M. Mahato
Calibration Engineer Chief Executive Officer (CEO)
Note: - 1. The calibration results reported in this certificate are valid at the time of and stated condition of measurement.
2. This report is not to be reproducing wholly or in part and cannot be used as evidence in the Court of Law and should not be used in any advertising media
without our special permission in writing.
3. Total liability of our Organization is limited to the invoiced amount.
4. Calibration Certificate without signature are not valid.
5. All our certificates are metrological traceable.
6. The results relate only to the item calibrated.
7. UUC: Unit Under Calibration.
8. Expanded Uncertainty of measurement expressed at approximately the 95 % level of confidence, using a coverage factor of k = 2.