QAQC PRELIM-ZBalbarada
QAQC PRELIM-ZBalbarada
All fields of medicine and hospital department are required to develop and conduct
programs that ensure the quality of patient care and management. Diagnostic imaging
department are leaders in promoting quality patient care.
QUALITY ASSURANCE
Deals with people
QUALITY CONTROL
Deals with instrumentation and equipment
QUALITY ASSURANCE
Health care organization often adopt formal, structured QA models. The joint commission
(TJC) promotes the "The 10 steps monitoring and evaluation process”
1. Assign responsibility
2. Delineate scope of care
3. Identify aspects of care
4. Identify outcomes that affect the aspect of care
5. Establish limit of scope of assessment
6. Collect and organize data
7. Evaluate care when outcomes are reacted
8. Take action to improve care
9. Assess the document actions
10. Communicate information to organization wide quality assurance programs
QUALITY CONTROL
Is more tangible and obvious than QA
Is design to ensure that the radiologist is provided with an optimal image produced
through good equipment performance and resulting in minimal patient radiation exposure
QC begins with xray imaging systems used to produce the image and continues with the
routine evaluation of imaging processing facilities
QC concludes with a dedicated analysis of each image to identify deficiencies and artifacts
and to minimize reexaminations
QC Team
Medical physicist- he/she performs routine activities, monitoring measurement and
observation
QC technologist
Radiologic engr.
Collimation
X-ray must coincide with the light field of the variable aperture light localizing collimator, if
these field are misaligned intended anatomy will be missed and unintended anatomy
irradiated
Misalignment must not exceed 2% of the source to the image receptor distance (SID)
kVp Calibration
Should be evaluated annually or whenever high voltage generator components have
changed significantly
Measured kVp should be within 10% of the indicated kVp
Exposure Linearity
Ability of a radiographic unit to produce a constant radiation output for various
combination of mA and exposure time
Must be within 10% for adjacent mA station
Exposure Reproducibility
Sequential radiation exposures should be reproducible to +- 5%
PROTECTIVE APPAREL
All protective aprons, gloves and gonadal shields should be radiograph or fluoroscope
annually for defects.
If cracks, tears, or holes are evident, the apparel may require replacement.
FILM ILLUMINATORS
Viewbox illumination should be analyzed photometrically on an annual basis.
This is done with an instrument called a photometer.
If the bulbs requires replacement, all bulbs in the illuminators should be replaced and
matched to the type of bulb used in adjacent illuminators.
Intensity should be at least 1500cd/m2 and should not vary by more than +-10% over the
surface of the illuminator
Photometer - measures the light intensity at several areas of the illuminator
PROCESSING CLEANING
The first automatic processor had a dry- to- drop time of 7 minutes, soon this was
shortened to 3 mins by what are known as double capacity processors.
Processing time was reduced further with the fast access system which todays most
popular 90 second processor, processor can handle up to 500 films per hour
Development temperature (95 deg F, 35 deg C) and development immersion time of 22
second
Wash temp should be (87 deg F, 31 deg C)
Cleaning is conducted weekly, records of such cleaning should be maintained
PROCESSOR MAINTENANCE
If equipment is not properly maintained, the processor may fail when least expected
3 Types of Maintenance Program
1. Scheduled maintenance
2. Preventive maintenance
3. Non scheduled maintenance
Scheduled Maintenance
Refers to routine procedures that performed usually weekly or monthly
Include:
Observation of all moving parts for wear
Adjustment of all belts, pulleys and gear
Application of proper lubrication to minimize wear
Preventive Maintenance
Plan program of parts replacement at regular intervals
Requires that a part be replaced before it fails
Nonscheduled Maintenance
The worse kind, a failure in the system that necessities processor is repair