Portable PH e
Portable PH e
Notes
1. The certification shall only be performed with the test equipment listed below.
2. For the certification the pH meter must be calibrated.
3. The measured values and limits of permissible error refer to the information specified in the
technical documentation (service manual).
4. No applications are certified.
5. Definition of terms according to ISO (International vocabulary of basic and general terms in
metrology).
Calibration: Determination of relationship between the indicated value and the true value of the
measured quantify(quantities) using certified and traceable reference inspection, measuring and
test equipment.
Adjustment: Operation of bringing a measuring instrument into a state of performance so that
indicated value lies within the specified limits of permissible error.
Maximum permissible errors: Extreme values of an error permitted by specifications for a
measuring instrument.
1. Test equipment
· Certified digital voltmeter with at least 5½ digits (199999 - (-199999)).
· METTLER I/O Tester2 from test set ME-51107158 (Delivery without 9V battery).
· 9 VDC, 500 mA Power Supply for ex. ME-51309528
· Meter Calibration Kit ME-51302410 consisting of the following resistors.:
- Certified resistor for temperature simulation of the NTC at 0 ˚ C (94.980 kΩ)
- Certified resistor for temperature simulation of the NTC at 25 ˚ C (30.000 kΩ)
- Certified resistor for temperature simulation of the NTC at 50 ˚ C (10.969 kΩ)
- Certified resistor for temperature simulation of the NTC at 75 ˚ C (4.528 kΩ)
- Certified resistor for temperature simulation of the NTC at 100 ˚ C (2.070 kΩ)
- Certified resistor for temperature simulation of the Pt1000 at 0 ˚ C (1.0000 kΩ)
- Certified resistor for temperature simulation of the Pt1000 at 25 ˚ C (1.0973 kΩ)
- Certified resistor for temperature simulation of the Pt1000 at 50 ˚ C (1.1940 kΩ)
- Certified resistor for temperature simulation of the Pt1000 at 75 ˚ C (1.2898 kΩ)
- Certified resistor for temperature simulation of the Pt1000 at 100 ˚ C (1.3850 kΩ)
· Electrode cable BNC-S7 1,2 m (ME-52300004)
· Recorder cable, (ME-51302038)
· 2 pcs Banana/Laboratory 4 mm cables
· Portable Adjustment Adapter (ME-51302425)
2.1. General
The pH meter must be switched on for 30 minutes before measurement may be performed. The
setup data and the user method must be noted down.
After the calibration, adjustment and certification process setup data and user methods have to
be re-entered into the pH meter.
2.2. Procedure for the calibration of the sensor inputs and the recorder
output
The I/O Tester2 is used to calibrate the sensor input and recorder output.
The I/O Tester2 should be equipped with a 9 V battery as power source. The “on” switch is
locked.
The following values are calibrated:
0 mV / 180 mV / 500 mV / 1000 mV / 1900 mV / -1900 mV / -1000 mV / -500 mV / -180 mV
This is only performed on the models MP125 and MA130. Connect the digital voltmeter to the
recorder output. The voltage applied with the I/O Tester2 at the sensor input is calibrated at the
recorder output. Maximum permissible error +/- 10 mV.
The certified temperature resistors of Calibration Set ME-51302410 is used to calibrate the ATC
temperature input of the ph meter. The calibration is carried out in the user mode of the pH meter.
The following temperatures are calibrated.:
4. Certification
For the certification of the pH meter, the values obtained under 2.3., 2.4. and 2.5. must be
entered in the certificate, but only if they lie within the specified limits of error. If this is not the
case, the necessary measures has to be taken according to the Service Manual or other valid
procedure.