Gesa Termómetros, S.L.
Polígono Industrial Igeltzera, Pabellón D3
E-48610 Urduliz – Vizcaya – Spain
Telf.: +34-94 676 63 64
Fax.: +34-94 676 63 65
e-mail:
[email protected] http://www.termometros.com
________________________________________________________________________________
CALIBRATION CERTIFICATE
Certificate No :2486/0178/07
CUSTOMER IDENTIFICATION:
Name: Lars Enviro Pvt Ltd Nagpur
IDENTIFICATION OF THE INSTRUMENT TO BE TESTED:
Description: Pressure gauge Bourdon type. Radial connection. Inner elements in
Stainless Steel AISI 316.
Pressure range: 0+25 bar
Subdivision: 0,5 bar
Construction: Case and ring in stainless steel AISI 304 Ø100mm.
Threaded connection ½” BSP according to UNE-EN 10226-1
Glycerine filled
Class 1,0 according to UNE-EN 837-1
Degree of protection IP 65 according to EN 60529 / IEC 529
Dimensions and tolerances according to UNE-EN 22768-1
Inner reference: R936
REFERENCE INSTRUMENT
Description: Digital pressure gauge with radial connection. Class 0,05 according to
UNE-EN 837-1.
ENAC certificate nr. 3913
Uncertainty: U = 8,8 mbar
Inner reference GE-19
The expanded uncertainty U has been obtained by multiplying the standard uncertainty uc by a
coverage factor k=2 according to CEA-ENAC-LC/02 Rev.1 January 98)
Traceability: To international patterns referred to the “Centro Español de Metrología”
– C.E.M. Madrid.
TEST METHODOLOGY
Procedure: Pressure driven by a manual pump. Different pressure points are fixed
on the measuring instrument (inverse calibration) in two series, one
increasing and one decreasing according to the inner procedure Gesa
PC/05/2011
Fluid: Water at ambient temperature.
Page 1
Gesa Termómetros, S.L.
Polígono Industrial Igeltzera, Pabellón D3
E-48610 Urduliz – Vizcaya – Spain
Telf.: +34-94 676 63 64
Fax.: +34-94 676 63 65
e-mail: [email protected]
http://www.termometros.com
________________________________________________________________________________
Certificate Nr G***/201*
TEST RESULTS
Tester (with corrections) Average
Expanded
Instrument diferences
Point Upwards Downwards Upwards Downwards Average Uncertainty
(bar) (Instrument-
1 (bar) 1 (bar) 2 (bar) 2 (bar) (bar) (K=2) (bar)
tester) (bar)
1 5 4,910 4,910 4,910 4,910 4,910 0,090 0,10
2 10 9,906 9,906 9,906 9,906 9,906 0,094 0,11
3 16 15,808 15,808 15,808 15,808 15,808 0,192 0,22
4 18 17,838 17,838 17,838 17,838 17,838 0,162 0,19
5 20 19,784 19,784 19,784 19,784 19,784 0,217 0,25
6 22 21,764 21,764 21,764 21,764 21,764 0,236 0,27
7 25 24,757 24,757 24,757 24,757 24,757 0,243 0,28
The expanded uncertainty U has been obtained by multiplying the standard uncertainty u c by a
coverage factor k=2 . The standard uncertainty uc has been obtained according to document
EAL-R2.
Test conditions:
Room temperature: 22 ± 2ºC
Atmospheric pressure: 1022 mbar
Relative humidity: 75± 10 %
Test date: 28 February 2012
Validity: 27 February 2013
Calibration service
Page 2