Calibration Certificate: Lars Enviro PVT LTD Nagpur

This document is a calibration certificate for a pressure gauge issued by Gesa Termómetros, S.L. It details the instrument's specifications, testing methodology, and results, including average differences and uncertainties for various pressure points. The calibration was performed on February 28, 2012, and is valid until February 27, 2013.

Uploaded by

PAWAN
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
37 views2 pages

Calibration Certificate: Lars Enviro PVT LTD Nagpur

This document is a calibration certificate for a pressure gauge issued by Gesa Termómetros, S.L. It details the instrument's specifications, testing methodology, and results, including average differences and uncertainties for various pressure points. The calibration was performed on February 28, 2012, and is valid until February 27, 2013.

Uploaded by

PAWAN
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Gesa Termómetros, S.L.

Polígono Industrial Igeltzera, Pabellón D3


E-48610 Urduliz – Vizcaya – Spain
Telf.: +34-94 676 63 64
Fax.: +34-94 676 63 65
e-mail: [email protected]
http://www.termometros.com
________________________________________________________________________________
CALIBRATION CERTIFICATE
Certificate No :2486/0178/07
CUSTOMER IDENTIFICATION:
Name: Lars Enviro Pvt Ltd Nagpur

IDENTIFICATION OF THE INSTRUMENT TO BE TESTED:


Description: Pressure gauge Bourdon type. Radial connection. Inner elements in
Stainless Steel AISI 316.
Pressure range: 0+25 bar
Subdivision: 0,5 bar
Construction: Case and ring in stainless steel AISI 304 Ø100mm.
Threaded connection ½” BSP according to UNE-EN 10226-1
Glycerine filled
Class 1,0 according to UNE-EN 837-1
Degree of protection IP 65 according to EN 60529 / IEC 529
Dimensions and tolerances according to UNE-EN 22768-1
Inner reference: R936
REFERENCE INSTRUMENT

Description: Digital pressure gauge with radial connection. Class 0,05 according to
UNE-EN 837-1.
ENAC certificate nr. 3913
Uncertainty: U = 8,8 mbar
Inner reference GE-19
The expanded uncertainty U has been obtained by multiplying the standard uncertainty uc by a
coverage factor k=2 according to CEA-ENAC-LC/02 Rev.1 January 98)
Traceability: To international patterns referred to the “Centro Español de Metrología”
– C.E.M. Madrid.

TEST METHODOLOGY
Procedure: Pressure driven by a manual pump. Different pressure points are fixed
on the measuring instrument (inverse calibration) in two series, one
increasing and one decreasing according to the inner procedure Gesa
PC/05/2011
Fluid: Water at ambient temperature.

Page 1
Gesa Termómetros, S.L.
Polígono Industrial Igeltzera, Pabellón D3
E-48610 Urduliz – Vizcaya – Spain
Telf.: +34-94 676 63 64
Fax.: +34-94 676 63 65
e-mail: [email protected]
http://www.termometros.com
________________________________________________________________________________
Certificate Nr G***/201*

TEST RESULTS

Tester (with corrections) Average


Expanded
Instrument diferences
Point Upwards Downwards Upwards Downwards Average Uncertainty
(bar) (Instrument-
1 (bar) 1 (bar) 2 (bar) 2 (bar) (bar) (K=2) (bar)
tester) (bar)
1 5 4,910 4,910 4,910 4,910 4,910 0,090 0,10
2 10 9,906 9,906 9,906 9,906 9,906 0,094 0,11
3 16 15,808 15,808 15,808 15,808 15,808 0,192 0,22
4 18 17,838 17,838 17,838 17,838 17,838 0,162 0,19
5 20 19,784 19,784 19,784 19,784 19,784 0,217 0,25
6 22 21,764 21,764 21,764 21,764 21,764 0,236 0,27
7 25 24,757 24,757 24,757 24,757 24,757 0,243 0,28

The expanded uncertainty U has been obtained by multiplying the standard uncertainty u c by a
coverage factor k=2 . The standard uncertainty uc has been obtained according to document
EAL-R2.

Test conditions:
Room temperature: 22 ± 2ºC
Atmospheric pressure: 1022 mbar
Relative humidity: 75± 10 %

Test date: 28 February 2012


Validity: 27 February 2013