XSA800 Series
Spectrum Analyzer
+ Frequency Range from 9 kHz up to 1.5 GHz
+ -160dBm Displayed Average Noise Level
+ Phase Noise -80dBc/Hz @1Gz and offset at 10KHz
+ Total Amplitude Accuracy <0.7dB
+ 10Hz Minimum Resolution Bandwidth (RBW)
+ EMI pre-compliance test kit, optional EMC test
software
+ Standard tracking generator hardware, can be
remotely upgraded according to needs
+ 9 inches LCD
+Performance Specifications
Model XSA805(TG) XSA810(TG) XSA815(TG)
Frequency
Range 9kHz-500MHz 9kHz-1GHz 9kHz-1.5 GHz
Resolution 1Hz
Frequency span
Range 0 Hz,100 Hz to maximum frequency of device
Accuracy ± span /(swept points -1)
Internal reference
Reference frequency 10 MHz
±[(days from last calibrate × freq aging rate)+ temperature stability + initial
Reference frequency accuracy
accuracy ]
Temperature stability <0.5ppm (15℃ to 35℃)
Aging rate <1ppm/year
Readout
Marker frequency resolution span/(the number of sweep points -1)
±(freq indication × freq reference uncertainty +1%× span +10%× resolution
Uncertainty
bandwidth + Marker Frequency Resolution)
Frequency counter
Resolution 1 Hz,10 Hz,100 Hz,1 kHz
Accuracy ±(marker freq × freq reference uncertainty + counter resolution)
Bandwidth
Resolution bandwidth (-3 dB) 1Hz to 1MHz(in 1 to 10 sequence), 1MHz,3MHz
Resolution filter shape factor <5 nominal(Digital implement, similar to Gauss Pattern)
Accuracy <5% nominal
Video bandwidth (-3 dB) 1Hz to 1MHz
Amplitude Specification
Amplitude and electric level
DANL to +10 dBm, 100KHz to 10MHz,close the preamplifier
Amplitude measurement range
DANL to +20 dBm, 10MHz to 1.5GHz , close the preamplifier
Max input DC voltage 50 VDC
Max. continuous wave RF power +20 dBm (100 mW), attenuation = 40 dB
Max. damage level +30 dBm (1W)
Displayed average noise level(DANL)attenuation = 0 dB, RBW = VBW = 100 Hz, sample detector, trace average ≥ 50,
20℃ to 30℃ , input impendence = 50 Ω
-95 dBm (typical), < -88 dBm ( 9 kHz to 1 MHz)
Preamp off -140 dBm (typical), <-130dBm (1MHz~500MHz)
-138 dBm (typical), <-128 dBm (500MHz~max)
-135 dBm (typical), <-128 dBm ( 9 kHz to 1 MHz)
Preamp on -160 dBm (typical), <-150 dBm (1MHz~500MHz)
-158 dBm (typical), <-148 dBm (500MHz~max)
Phase noise 20 ℃ to 30 ℃,fc=1 GHz
<-80 dBc/Hz @10 kHz offset,
Phase noise <-100 dBc/Hz @100 kHz offset
<-115 dBc/Hz @1 MHz -500MHz offset
Level display range
Log scale coordinate 0.01dB ~255dB
Linear scale coordinate 0 to reference level
level unit dBm, dBuW, dBpW, dBmV, dBuV, W,V
Points 760
Number of traces 5
Positive-peak, negative-peak, sample, normal, RMS, Average, quasi-peak (with EMI
Detectors
option)
Trace functions Clear write, Max Hold, Min Hold, View, Blank, Average, Trace math
Frequency response
20℃ to 30℃,30%~70% relative humidity, 10 dB input attenuation, reference 50 MHz
Preamp off(fc≥9KHz) <0.7dB;
Preamp on(fc≥9KHz) <1.0 dB;
Accuracy
Relative to 10 kHz RBW
RBW Switching Uncertainty
<0.1 dB
20℃ to 30℃, fc=50 MHz,Preamplifier Off, 10dB RF attenuation,input signal 0~40
Input Attenuation Switching
dB
Uncertainty
<0.5 dB
20℃ to 30℃, fc=50 MHz,peak detector, 10 dB RF attenuation,preamplifier off,
Absolute Amplitude Uncertainty input signal level = -10 dBm
< 0.4 dB
Reference Level Range -80 dBm to +30 dBm, in 1 dB step
Reference Level Resolution Log scale 0.01 dB; linear scale 4 digits
95% confidence level, S/N > 20 dB, RBW = VBW = 1 kHz, preamplifier off, attenuation =
Uncertainty 10 dB, -50 dBm < input level ≤ 0 dBm, fc > 10 MHz, 20℃ to 30℃
< 0.7 dB
VSWR input ≥ 10 dB,300 kHz to max;
<1.5,nominal
Distortion and spurious response
fc ≥ 50 MHz,Preamp off, signal input -20 dBm, 10 dB RF attenuation, 20 ℃ to 30 ℃
Second harmonic distortion
> +45 dBm
fc ≥ 50 MHz, two -20 dBm tones at input mixer spaced by 200 kHz, attenuation = 0 dB
Third-order intermodulation
> +10 dBm
fc ≥ 50 MHz, 0 dB RF attenuation,Preamp off , 20 ℃ to 30 ℃
1 dB Gain Compression
>0 dBm, nominal
connect 50 Ω load at input port,0 dB input attenuation,20 ℃ to 30 ℃
Residual response
<-90dBm,nominated
Intermediate frequency <-60 dBc
Referenced to local oscillators, referenced to A/D conversion, referenced to
System related sidebands subharmonic of first LO, referenced to harmonic of first LO
<-60 dBc
-30 dBm signal at input mixer,20 ℃ to 30 ℃
Input related spurious
<-60 dBc
Sweep time and triggering
SPAN ≥ 100 Hz 10ms to 3000s
Sweep time
zero sweep width 10ms to 3000s
SPAN≥ 100 Hz 5%(nominal)
Sweep time uncertainty
zero sweep(Sweep time >1 ms) 5%(nominal)
Mode Continue, single
Trigger Free run, video, external
External trigger level 5 V TTL level
Tracking generator (Only for XSA805-TG, XSA810-TG and XSA815-TG)
Output frequency range 100 kHz to 1.5 GHz(Tracking generator)
Output power level range -40 dBm to 0 dBm
Output power level resolution 1dB
Output flatness ±3dB
Harmonic spurious -30 dBc (Tracking generator output power -10 dBm)
Tracking generator spurious
Non-harmonic spurious -40 dBc (Tracking generator output power -10 dBm)
Tracking source to input
-60 dB (Tracking generator output power 0 dBm)
terminal isolation
Maximum safe reverse level Average total power: +30 dBm,DC : ±50 VDC
Inputs and Outputs
Front panel RF input connector 50 Ω,N-type female
Front panel track generator
50 Ω,N-type female
output
Internal/ External Reference 50 Ω,N-type female
External Trigger Input 50 Ω,N-type female
Communication port USB HOST, USB DEVICE, LAN, earphone port, HDMI
General technical specification
Display TFT LCD,9 inches (1280*800)
Weight (without package) About 3.7 kg
Dimension (W × H × D) 375mm*185mm*120mm
Working temperature 0 ℃ to 40 ℃
Storage temperature -20 °C to +60 °C
Power 100V - 240V 50/60Hz