JAN-JUNE 2025 BSCBT 6 SEM V9 BSCBT604A Question Bank B.SC. VI Sem Elective Chemistry BSCBT604A BPCM603A BZBC603A
JAN-JUNE 2025 BSCBT 6 SEM V9 BSCBT604A Question Bank B.SC. VI Sem Elective Chemistry BSCBT604A BPCM603A BZBC603A
DEPARTMENT OF CHEMISTRY
SECTION - A
Unit I
Question 1: What are some key spectroscopic methods covered in the core chemistry syllabus,
and how do they utilize electromagnetic radiation for analytical purposes?
Answer: Several spectroscopic methods are covered in the core chemistry syllabus, including
UV-Visible spectroscopy, infrared spectroscopy (IR), nuclear magnetic resonance spectroscopy
(NMR), and atomic absorption spectroscopy (AAS). These methods utilize electromagnetic
radiation to interact with sample molecules, leading to measurable signals that provide valuable
information about the sample's composition, structure, and concentration.
Question 2: How is analytical data treated in spectroscopy, and why is error analysis important
in the interpretation of spectroscopic results?
Question 3: Discuss the classification of analytical methods, with a focus on the types of
instrumental methods commonly employed. Provide examples of instrumental techniques within
each classification.
Answer: Analytical methods can be classified into two main categories: classical methods and
instrumental methods. Instrumental methods further divide into electrochemical methods (e.g.,
voltammetry), spectroscopic methods (e.g., UV-Visible spectroscopy), chromatographic methods
(e.g., gas chromatography), and mass spectrometric methods (e.g., mass spectrometry). Each
classification offers distinct advantages, and the choice of method depends on the specific
analytical requirements.
Question 4: How does electromagnetic radiation play a central role in various spectroscopic
methods? Briefly explain its interaction with sample molecules and the information it provides.
Question 5: How does the consideration of electromagnetic radiation contribute to the versatility
and selectivity of spectroscopic methods? Provide an example of a spectroscopic technique and
its application in analytical chemistry.
Unit II
Q1. Define Infrared Spectroscopy.
Answer: Infrared (IR) Spectroscopy is a form of spectroscopy that deals with the infrared part of
the electromagnetic spectrum. The infrared region’s rays have a longer wavelength than light but
a lower frequency. Absorption spectroscopy is the foundation of infrared spectroscopy.
As a result, a spectrum of colours ranging from red to violet can be seen, with red
suffering the least variation due to its longest wavelength.
As violet merges into blue, blue into green, and so on, this type of spectrum is known as
a continuous spectrum.
MS can also tell you whether a compound contains particular elements like bromine or chlorine.
These halogens can be easily detected by comparing the intensity ratios of ions with atomic mass
units (m/z) that differ by two.
Nuclear magnetic resonance with respect to hydrogen-1 nuclei within a substance’s molecules is
known as proton nuclear magnetic resonance (proton NMR, hydrogen-1 NMR, or 1H NMR). In
samples containing natural hydrogen (H), the isotope 1H makes up nearly all of the hydrogen
(hydrogen-1; i.e., having a proton for a nucleus).
Answer: The science of spectroscopy is the study of the interaction between matter and radiated
energy. It’s the study of matter’s absorption properties or absorption behaviour when exposed to
electromagnetic radiation. Spectroscopy is a theoretical method of science that does not produce
any results.
Unit III
Question 1: What are the primary factors influencing the separation in chromatography, and
how do these techniques differ in terms of separation mechanisms?
Answer: The primary factors influencing separation in chromatography are volatility, solubility,
interactions with the stationary phase, size, and electrical field. Gas chromatography separates
based on volatility, liquid chromatography on solubility, and electrophoresis on size and charge
interactions.
Question 2: Explain the importance of column technology in chromatography, specifically
discussing the role of packing materials and capillaries.
Answer: Column technology is crucial in chromatography for efficient separation. Packing
materials and capillary dimensions influence separation efficiency. Different packing materials
provide diverse stationary phases, allowing for the separation of compounds based on their
interactions with the stationary phase.
Question 3: How is mass spectrometry used to make gaseous molecules into ions, and what are
the different techniques employed for making liquids and solids into ions?
Answer: Gaseous molecules are made into ions in mass spectrometry through techniques like
electron impact and chemical ionization. For liquids and solids, ionization techniques include
electrospray, electrical discharge, laser desorption, and fast atom bombardment.
Question 4: Discuss the detection methods in chromatography and mass spectrometry. How is
detection used as a means of further analysis, including the use of tags and coupling to other
analytical techniques?
Answer: Detection methods in chromatography include simple (e.g., UV-Vis) and specific (e.g.,
MS) techniques. Detection is used for further analysis by incorporating tags and coupling to
infrared (IR) and mass spectrometry (MS) techniques, providing additional information about the
compounds being analyzed.
Question 5: In electrophoresis, how does it contribute to DNA analysis, and what role does it
play in immunological assays (immunoassays)?
Answer: In DNA analysis, electrophoresis is used to separate DNA fragments based on size.
Capillary electrophoresis offers high-resolution DNA sequencing. In immunoassays,
electrophoresis helps separate and analyze proteins, aiding in the detection and quantification of
specific antibodies or antigens.
Unit IV
Q1. What are the different electroanalytical methods and their application?
Answer: The four main categories are potentiometry (the difference in electrode potentials is
measured), amperometry (electric current is the analytical signal), coulometry (charge passed
during a certain time is recorded), and voltammetry (the cell's current is measured while actively
altering the cell's potential).
Electrosynthesis.
Electrolysis.
Electrochemical Corrosion.
Battery Testing.
Biological Electrochemistry.
Photovoltaics.
Fuel Cells.
Supercapacitors.
Acid-base titration.
Precipitation titration.
Replacement titration.
Complexometric titration.
Answer: pH is defined as the negative log of the hydrogen ion concentration. The range of pH
extends from zero to 14.pH is a logarithm (the negative of the logarithm of H+ activity), and as
such, it has no units.
Unit V
Answer: The principle behind solvent extraction is extremely basic. The goal is to use a liquid
(solvent) to dissolve (solvate) a target molecule or group of compounds (solute) and to wash
them out of the solid plant material. The solvent is then separated from the solute in order to
concentrate the solute.
Answer: Ion exchange chromatography (IEX) separates molecules by their surface charge, a
property that can vary vastly between different proteins. There are two types of IEX, cation
exchange and anion exchange chromatography.
SECTION B
Unit I
Q1. Differentiate between qualitative and quantitative analysis.
Answer: The main distinction between qualitative and quantitative chemistry is that qualitative
chemistry determines the presence or absence of various chemical components in a sample,
whereas quantitative chemistry determines the amount of various chemical components present
in a sample.
= 75 and unknown standard deviation. Using the formula for the t-statistic, the
calculated t equals 2. For a two-sided test at a common level of significance α = 0.05, the critical
values from the t distribution on 24 degrees of freedom are −2.064 and 2.064. The calculated t
does not exceed these values, hence the null hypothesis cannot be rejected with 95 percent
confidence. (The confidence level is 1 − α.)
A second application of the t distribution tests the hypothesis that two independent random
samples have the same mean. The t distribution can also be used to construct confidence
intervals for the true mean of a population (the first application) or for the difference between
two sample means (the second application).
Q4. How To Reduce Errors In Measurement.
Answer: Keeping an eye on the procedure and following the below listed points can help to
reduce the error.
The important thing to know about the Normal Distribution is that the probability of getting a
certain result decreases the farther that result is from the mean. The concept of the normal
distribution will be important when we talk about 1- and 2-tailed tests, confidence levels, and
statistical tests, such as the t-test and F-test.
There are many other types of distributions, but we will only consider the normal distribution
here. This is because we assume that the measurements we perform in this course will be
normally distributed about the mean, and that the random errors will also be normally
distributed. Generally, this is a good assumption, though there are many situations where it does
not apply.
Unit II
Q1: Explain Nuclear Magnetic Resonance (NMR) spectroscopy.
Answer: The study of molecules by recording the interaction of radiofrequency (Rf)
electromagnetic radiations with the nuclei of molecules placed in a strong magnetic field is
known as nuclear magnetic resonance (NMR) spectroscopy.
It’s a study method that takes advantage of the magnetic characteristics of specific atomic nuclei.
The physical and chemical properties of atoms or molecules are determined using NMR
spectroscopy.
It uses the nuclear magnetic resonance phenomena to offer extensive information on a
molecule’s structure, dynamics, reaction state, and chemical environment.
Monitoring the environment and climate change by analysing air, water, and soil quality.
Carbon dating of samples and determining the composition of rock and soil.
Answer: Advantages:
2. Inexpensive.
3. The determination of elements such as alkali and alkaline earth metals is performed
easily with most reliable and convenient methods.
4. Quite quick, convenient, and selective and sensitive to even parts per million (ppm) to
parts per billion (ppb) range.
Disadvantages: Moreover the flame photometer has a wide range of applications in the
analytical chemistry, it possess many disadvantages which are explained below:
1. The concentration of the metal ion in the solution cannot be measured accurately..
4. The information about the molecular structure of the compound present in the sample
solution cannot be determined.
5. The elements such as carbon, hydrogen and halides cannot be detected due to its
non-radiating nature.
Unit III
Question 1: Discuss the principles of gas chromatography (GC) and liquid chromatography
(LC), emphasizing the importance of column technology in separation. How do the separation
mechanisms differ in GC and LC, and how does column packing influence efficiency?
Answer: Gas chromatography (GC) and liquid chromatography (LC) are widely used separation
techniques. GC separates volatile compounds based on their vaporization and partitioning in a
column, while LC separates compounds based on solubility and interactions with the stationary
phase. The choice of column technology, including packing materials and capillaries, is pivotal
in achieving efficient separations. The stationary phase's nature, thickness, and column
dimensions impact resolution and selectivity. In GC, compounds with higher volatility elute
faster, whereas in LC, solubility and interactions play a critical role. The variation in column
packing materials allows tailoring separation conditions for specific analytes, making column
technology a key factor in chromatographic performance.
Answer: Mass spectrometry (MS) is a powerful analytical technique for identifying and
characterizing molecules. Gaseous molecules are ionized through electron impact or chemical
ionization, while liquids and solids undergo ionization methods such as electrospray, electrical
discharge, laser desorption, and fast atom bombardment. The separation of ions is achieved
based on their mass-to-charge ratio (m/z). Different mass analyzers contribute to ion separation.
Magnetic sector analyzers utilize a magnetic field, time-of-flight analyzers rely on ion flight
time, and electric quadrupole analyzers use varying electric fields. Each method offers unique
advantages, and the choice depends on factors such as resolution requirements and the type of
sample being analyzed.
Question 3: Examine the role of detection methods in chromatography and mass spectrometry.
Distinguish between simple and specific detection, and elaborate on how detection is utilized for
further analysis, including the use of tags and coupling to other analytical techniques like
infrared (IR) and mass spectrometry (MS).
Answer: Detection methods play a crucial role in chromatography and mass spectrometry,
providing insights into the identity and concentration of analytes. Simple detection methods,
such as UV-Vis in chromatography, offer general information about the sample. Specific
detection, as seen in mass spectrometry (MS), provides detailed information on individual
compounds. Detection serves as a means for further analysis through the incorporation of tags
and coupling to other techniques. Tags enhance specificity, allowing the detection of specific
functional groups or compounds. Coupling to infrared (IR) and mass spectrometry (MS)
techniques provides additional information about the molecular structure and composition,
contributing to a more comprehensive analysis of complex samples.
Question 5: Explain the factors influencing resolution, time, and multiple separations in mass
spectrometry. Delve into the detection and interpretation processes, emphasizing the linkage to
excitation and its role in analyzing complex samples.
Answer: Resolution, time, and the ability to perform multiple separations are critical aspects of
mass spectrometry (MS). Resolution is influenced by factors such as the type of ionization, mass
analyzer, and magnetic field strength. Time considerations involve the speed of analysis, with
time-of-flight (TOF) analyzers providing rapid results. Multiple separations are achievable by
combining various ionization and separation methods, enhancing the technique's versatility.
Detection in mass spectrometry involves analyzing mass spectra, and interpretation relies on
characteristic peaks corresponding to specific ions. The linkage to excitation refers to the initial
energy input during ionization, influencing the observed mass spectra. Understanding excitation
aids in identifying and characterizing compounds in complex samples, providing valuable
information for detailed molecular analysis and structural elucidation.
Unit IV
Q1. What is Potentiometric Titration? Explain.
Answer: It is the procedure through which the quantity of the given test substance is determined
by the measured addition of titrant until the entire test substance undergoes reaction. After the
titration process, the potential difference between the two electrodes (namely the reference and
indicator electrode) is measured in conditions where a thermodynamic equilibrium is maintained
and the current passing through the electrodes does not disturb this equilibrium.
Potentiometric Titration Principle
Potentiometric titration is a laboratory method to determine the concentration of a given analyte.
It is used in the characterization of acids. In this method, there is no use of a chemical indicator.
Instead, the electric potential across the substance is measured.
Potentiometric Titration Method
Potentiometric Titration is done via the usage of two electrodes – an indicator electrode and a
reference electrode (generally a hydrogen electrode or a silver chloride electrode). One half-cell
is formed with the indicator electrode and the ions of the analyte, which is generally an
electrolyte solution. The other half-cell is formed by the reference electrode.
The overall cell potential can be calculated using the formula given below.
Where the potential drop between the indicator and reference electrodes over the electrolyte
solution is given by Esol.
The overall cell potential, Ecell is calculated in every interval where the titrant is measured and
added. Now, a graph is plotted with the Potential difference on the Y-axis and the volume on the
X-axis as shown below.
It can be observed from the graph that the electric potential of the cell is dependent on the
concentration of ions which are in contact with the indicator electrode. Therefore, the E cell is
measured with each addition of the titrant.
Types of Potentiometric Titration
There are four types of titration that fall under the category of potentiometric titration, namely
acid-base titration, redox titration, complexometric titration, and precipitation titration. A brief
description of each of these types of titration is given below.
Acid-Base Titration: This type of potentiometric titration is used to determine the concentration
of a given acid/base by neutralizing it exactly using a standard solution of base/acid whose
concentration is known.
Redox Titration: This type of potentiometric titration involves an analyte and titrant that
undergo a redox reaction. An example of this type of titration would be the treatment of an
iodine solution with a reducing agent which produces iodide ion (a starch indicator is used to get
the endpoint).
Complexometric Titration: This type of titration can also be referred to as chelatometry. In this
method, a coloured complex is formed, indicating the end point of the titration. This method is
used to determine a mixture of metal ions in a given solution.
Precipitation Titration: This type of titration involves a reaction between the given analyte and
the titrant wherein an insoluble precipitate is formed. The end-point of this titration is noted
when the addition of the titrant no longer forms a precipitate.
Q2. What is pKa? How do we calculate the pKa? Describe the relation between pKa and
pKb.
Answer: The pKa value is the negative base -10 logarithm of the acid dissociation
constant (Ka) of a solution.
The quantitative behavior of acids and bases in solution can be understood only if their pKa
values are known. In particular, the pH of a solution can be predicted when the analytical
concentration and pKa values of all acids and bases are known; conversely, it is possible to
calculate the equilibrium concentration of the acids and bases in solution when the pH is known.
These calculations find application in many different areas of chemistry, biology, medicine, and
geology. For example, many compounds used for medication are weak acids or bases, and a
knowledge of the pKa values.
Definition of pKa: pKa is a number that describes the acidity of a particular molecule. It
measures the strength of an acid by how tightly a proton is held by a Bronsted acid. The lower
the value of pKa, the stronger the acid and the greater its ability to donate its protons. describe
the acidity of a particular molecule. Ka denotes the acid dissociation constant. It measures how
completely an acid dissociates in an aqueous solution. The larger the value of Ka, the stronger
the acid as acid largely dissociates into its ions and has lower pka value. The relationship
between pKa and Ka is described by the following equation:
pKa = -log[Ka]
Acid dissociation constants, or pKa values, are essential for understanding many fundamental
reactions in chemistry. These values reveal the deprotonation state of a molecule in a particular
solvent. There is great interest in using theoretical methods to calculate the pKa values for many
different types of molecules.
Calulation of pKa
Let us consider a weak acid HA which ionises in the aqueous solution as:
o HA(acid) + H2O ⇋ H3O+ (aq) + A– (aq)(conjugate base)
Examples with Phosphoric acid which contain 3 protons that produced dissociation of the first
proton may be denoted as Ka1 and the constants for the dissociation of successive protons as
Ka2, and Ka3.
H3PO4 ⇋ H2PO4– + H+
Ka1 = [H2PO4– ] [H+] / [H3PO4]
pKa1 = – log Ka1 = – log {[H2PO4– ] [H+] / [H3PO4]}
H2PO4– ⇋ HPO42- + H+
Ka2 = [HPO42- ] [H+] / [H2PO4– ]
pKa2 = – log Ka2 = – log {[HPO42- ] [H+] / [H2PO4– ]}
HPO42- ⇋ PO43- + H+
Ka3 = [PO43- ] [H+] / [HPO42- ]
pKa3 = – log Ka3 = – log {[PO43- ] [H+] / [HPO42- ]}
pKa and pH of buffer solution
From the Henderson equation of acidic buffer, the pH of the solution is defined as the
pH = pKa + log { [salt] / [Acid]}
When [salt] / [Acid] = 10 then,
pH = pKa + 1
When [salt] / [Acid] = 1/10 then,
pH = pKa – 1
Note: So weak acid may be used for preparing buffer solutions having pH values lying within
the ranges pKa + 1 and pKa – 1.
The acetic acid has a pK a of about 4.8. It may therefore be used for making buffer solutions with
pH values lying roughly between the range 3.8 to 5.8.
Relation between pKa and pKb
Let us consider a weak acid HA which ionises in the aqueous solution as:
HA + H2O ⇋ H3O+ (aq) + A– (aq)
(acid) (conjugate base)
The dissociation constant of acid is defined by
Ka = [H3O+] [A–] / [HA] ……….. (1)
The conjugate base A- behaves as a weak base in water
A– + H2O ⇋ HA + OH–
For base Kb = [HA] [ OH–] / [A– ] …………. (2)
Multiply equation (i) and (ii)
Ka x Kb = {[H3O+] [A–] / [HA]} x {[HA] [OH–] / [A– ]} = [H3O+] [OH–] = Kw
Ka x Kb = Kw
On taking negative logarithm on both side
– log Ka – log Kb = – log Kw
pKa + pKb = pKw
Q3. How does optical polarimetry work?
Answer: Polarimetry machines are used in chemistry in a variety of ways. Their primary use is
to measure the angle of rotation of an optically active substance using polarized light. The
polarized light will either rotate clockwise or counter-clockwise and the amount it rotates
indicates the angle of rotation. Polarimetry is important in chemistry due to the fact that it allows
one to distinguish between optically active stereoisomers using optical activity as a measuring
point. There are other key tests that are used in chemistry for identification of substances, such as
melting point. However, these tests would prove non conclusive for identification of some
stereoisomers, such as enantiomers, as they will have identical physical properties, such as
melting points and boiling points. The step by step process of rotation of plane-polarized light in
a polarimeter is depicted by the schematic diagram below:
The specific rotation value of a chiral substance is dependent on numerous variables. The main
variables are concentration of the substance and the length of the tube in the polarimeter
machine.
Unit V
Question 1: Explain the principle of NMR spectroscopy and how it provides information about
the molecular structure of compounds.
Answer: Nuclear Magnetic Resonance (NMR) spectroscopy is based on the principle of nuclear
spin. When a sample is exposed to a strong magnetic field and radiofrequency radiation, certain
nuclei with a nonzero magnetic moment absorb energy and undergo resonance. The energy
absorbed is proportional to the strength of the magnetic field, allowing for the determination of
the chemical environment of the nuclei. NMR provides detailed information about molecular
structure, including the types of atoms present, their connectivity, and the local environment,
making it a powerful tool for structural elucidation in organic and biochemistry.
Question 2: Describe the instrumentation used in NMR spectroscopy and the role of magnets
and radiofrequency pulses in the NMR experiment.
Answer: NMR spectroscopy utilizes specialized instruments consisting of powerful magnets and
radiofrequency (RF) pulse systems. The magnets generate a strong and homogeneous magnetic
field, aligning nuclear spins in the sample. RF pulses are applied perpendicular to the magnetic
field, perturbing the spin alignment. When the RF pulse is turned off, the spins relax back to
their equilibrium positions, emitting signals that are detected and used to generate NMR spectra.
The strength and homogeneity of the magnetic field, as well as the precision of the RF pulses,
are critical for obtaining high-quality NMR data.
Question 3: What are the factors affecting chemical shift in NMR spectroscopy, and how do
they influence the positions of signals in the NMR spectrum?
Question 5: Explain the principles of X-ray analysis and electron spectroscopy in surface
analysis. How are these techniques used to investigate the composition and structure of materials
at the atomic level?
Answer: X-ray analysis and electron spectroscopy are powerful techniques for surface analysis.
X-ray analysis, including X-ray diffraction (XRD) and X-ray photoelectron spectroscopy (XPS),
utilizes X-rays to determine crystal structures and elemental compositions. Electron
spectroscopy, such as Auger electron spectroscopy (AES) and scanning electron microscopy
(SEM), uses electrons for surface analysis. These techniques provide information about the
elemental composition, crystal structure, and morphology of materials at the atomic level. XRD
is particularly useful for crystalline materials, while XPS and AES offer insights into elemental
composition and chemical bonding on surfaces. SEM provides high-resolution images, allowing
for detailed morphological analysis of surfaces.
SECTION C
Unit I
Answer: Types of errors: There are three types of errors that are classified based on the source
they arise from; They are:
Gross Errors
Random Errors
Systematic Errors
Gross Errors
This category basically takes into account human oversight and other mistakes while reading,
recording, and readings. The most common human error in measurement falls under this
category of measurement errors. For example, the person taking the reading from the meter of
the instrument may read 23 as 28. Gross errors can be avoided by using two suitable measures,
and they are written below:
Proper care should be taken in reading, recording the data. Also, the calculation of error
should be done accurately.
By increasing the number of experimenters, we can reduce the gross errors. If each
experimenter takes different readings at different points, then by taking the average of
more readings, we can reduce the gross errors
Random Errors
The random errors are those errors, which occur irregularly and hence are random. These can
arise due to random and unpredictable fluctuations in experimental conditions (Example:
unpredictable fluctuations in temperature, voltage supply, mechanical vibrations of experimental
set-ups, etc, errors by the observer taking readings, etc. For example, when the same person
repeats the same observation, he may likely get different readings every time.
This article explored the various types of errors in the measurements we make. These errors are
everywhere in every measurement we make.
Systematic Errors:
Systematic errors can be better understood if we divide them into subgroups; They are:
Environmental Errors
Observational Errors
Instrumental Errors
Environmental Errors: This type of error arises in the measurement due to the effect of the
external conditions on the measurement. The external condition includes temperature, pressure,
and humidity and can also include an external magnetic field. If you measure your temperature
under the armpits and during the measurement, if the electricity goes out and the room gets hot,
it will affect your body temperature, affecting the reading.
Observational Errors: These are the errors that arise due to an individual’s bias, lack of proper
setting of the apparatus, or an individual’s carelessness in taking observations. The measurement
errors also include wrong readings due to Parallax errors.
Instrumental Errors: These errors arise due to faulty construction and calibration of the
measuring instruments. Such errors arise due to the hysteresis of the equipment or due to
friction. Lots of the time, the equipment being used is faulty due to misuse or neglect, which
changes the reading of the equipment. The zero error is a very common type of error. This error
is common in devices like Vernier callipers and screw gauges. The zero error can be either
positive or negative. Sometimes the scale readings are worn off, which can also lead to a bad
reading.
Instrumental error takes place due to :
An inherent constraint of devices
Misuse of Apparatus
Effect of Loading
Their value deviates significantly from the mean. The Q-Test can be used to determine
whether an individual measurement should be rejected or retained. The quantity Q is the absolute
difference between the questioned measurement (xq) and the next closest measurement (xn)
divided by the spread (ω), the difference between the largest and smallest measurement, of the
entire set of data.
(xq − xn)ω
will fall into a range around the mean (x).) If Q is greater than the values listed below for
a particular confidence level, the measurement should be rejected. If Q is less than the
The limiting value is 0.60 for rejecting data at the 99% confidence level for 9 measurements.
Since 0.82 is greater than 0.60, the student should reject the measurement.
* A shortcut can be used to test a suspect measurement: Calculate the mean and average or
standard deviation without using the suspect measurement and reject the suspect
measurement if its deviation from the mean is greater than four times the average or standard
deviation.
Application Comparing the means of two populations. Comparing two population variances.
Basis for
T-test F-test
Comparison
Q4. How do we measure the different types of errors? A length was calculated to be 10.1
feet, but the absolute length was 10.5 feet. Calculate the absolute error.
Heterogeneous material is first divided into relatively homogenous group i.e. material is
first divided into groups possessing similar characteristics. Then from each group
samples are drawn at random it is called as stratified sampling.
E.g.: ores exist as lumps of various sizes; hence lumps of different sizes are grouped on basis of
their sizes then from each group of sizes samples are drawn at random at mixed.
B) Non-random or Systematic Sampling:-
This type of sampling appears to be more scientific method than random sampling,
though not necessary that it will give better sample.
In systematic sampling, sample units are drawn in a definite sequence at equal intervals
from one another.
Purpose of Sampling:-
Judging acceptability :-
A sampling of bulk many times is done to conclude the material from which sample is done to
conclude whether the material from which sample is drawn meets the requirements such as
purchase or sales specifications. If it meets requirements it can be accepted otherwise rejected.
Detection of contaminations :-
The second purpose of sampling is to assure that the material under consideration is free from
contamination or any unwanted material. E.g.: Urea in drinking milk.
Identifying material :-
The third purpose of sampling is to identify an unknown material. A carefully drawn sample can
adequately serve to establish the identity of bulk material under study.
Estimation of material : –
Answer: The principle of flame photometry is based on the fact that the compounds of some
elements can be thermally dissociated in a flame and that some of the atoms produced in this
process can get excited to a higher energy level. A flame photometer has three essential parts.
1. Source of flame:
A burner that provides flame and can be maintained in a constant form and at a constant
temperature.
Helps to transport the homogeneous solution of the substance into the flame at a steady
rate.
3. Optical system (optical filter):
The optical system comprises three parts: convex mirror, lens and filter. The convex
mirror helps to transmit light emitted from the atoms and focus the emissions to the
lens. The convex lens help to focus the light on a point called slit. The reflections from
the mirror pass through the slit and reach the filters. This will isolate the wavelength to
be measured from that of any other extraneous emissions. Hence it acts as interference
type color filters.
4. Photo detector:
Detect the emitted light and measure the intensity of radiation emitted by the flame.
That is, the emitted radiation is converted to an electrical signal with the help of photo
detector. The produced electrical signals are directly proportional to the intensity of
light. A schematic representation of flame photometer is shown in figure 1.
Fig 1: A schematic representation of flame photometer
Mechanism of working:
The working of the flame photometer involves a series of steps which is discussed in the
following sections.
Nebulisation:
The solution of the substance to be analyzed is first aspirated into the burner, which is then
dispersed into the flame as fine spray particles.
2. This solid particles move towards the flame, where the gaseous atoms and ions are
produced.
3. The ions absorb the energy from the flame and excited to high energy levels.
4. When the atoms return to the ground state radiation of the characteristic element is
emitted.
1. Desolvation: The metal particles in the flame are dehydrated by the flame and hence
the solvent is evaporated.
2. Vapourisation: The metal particles in the sample are dehydrated. This also led to the
evaporation of the solvent.
3. Atomization: Reduction of metal ions in the solvent to metal atoms by the flame heat.
4. Excitation: The electrostatic force of attraction between the electrons and nucleus of
the atom helps them to absorb a particular amount of energy. The atoms then jump to
the exited energy state.
5. Emission process: Since the higher energy state is unstable the atoms jump back to
the stable low energy state with the emission of energy in the form of radiation of
characteristic wavelength, which is measured by the photo detector.
The energy level diagram of the sodium atom is shown in figure 3.
The intensity of the light emitted could be described by the Scheibe-Lomakin equation:
Then,
That is the intensity of emitted light is directly related to the concentration of the sample.
The flame emissions of the alkali and alkaline earth metals in terms of the emission
wavelength and the characteristic color produced by each element is shown in table 1
Name of the element Emitted wavelength
Observed colour of the flame
range (nm)
Violet
Red
Orange
Yellow
Applications:
Flame photometer has both quantitative and qualitative applications. Flame photometer with
monochromators emits radiations of characteristic wavelengths which help to detect the
presence of a particular metal in the sample. This help to determine the availability of alkali
and alkaline earth metals which are critical for soil cultivation. In agriculture, the fertilizer
requirement of the soil is analyzed by flame test analysis of the soil. In clinical field, Na+ and
K+ ions in body fluids, muscles and heart can be determined by diluting the blood serum and
aspiration into the flame. Analysis of soft drinks, fruit juices and alcoholic beverages can also
be analyzed by using flame photometry.
It is important to understand the optical layout of both single beam and double beam
systems before you begin to appreciate the advantages of one over the other. So let’s
get into the details and know these systems better.
Understanding Single Beam Spectrophotometer
Single
Beam Schematic Diagram
This spectrophotometer is used to measure the relative intensity of light before and
after inserting a test sample. It functions by either blanking the instrument or
standardizing it with respect to the reference. There are four basic single beam
spectrophotometer components:
Monochromator
Detector
Light source (should be stable over time, low cost, wide wavelength range
brightness, and long service life)
Sample (object)
Working: The light source comprising of a hollow cathode lamp emits sharp atomic
line of the element whose determination is required. The light is modulated (switched
on and off) rapidly by means of a rotating chopper located between the light source
and the flame.
Modulation can also be achieved by pulsing the power (switched on and off rapidly)
to the light source. Modulation serves to differentiate the light coming from the
source lamp from the emission from the flame.
The modulated light is led to the flame where ground state atoms of the element of
interest are present and after absorption is led to the monochromator which isolates
the wavelength of interest which is then led to the detector.
Now that you know what a single beam spectrophotometer is and how it works, let’s
see some of the advantages and disadvantages of this instrument.
Advantages of a Single Beam Spectrophotometer
Cost-Effective: Single beam instruments are less expensive as compared to
the other alternative.
Better Performance: High energy throughput due to the non-splitting of the
source beam results in high sensitivity of detection.
Disadvantages of a Single Beam Spectrophotometer
Instability: This happens due to lack of compensation for disturbances like
electronic circuit fluctuations, voltage fluctuations, mechanical component’s
instability, or drift in the energy of light sources. Such drifts cause abnormal
fluctuations in the results.
Understanding Double Beam Spectrophotometer
Answer: An IR spectrum is essentially a graph plotted with the infrared light absorbed
on the Y-axis against. frequency or wavelength on the X-axis. An illustration
highlighting the different regions that light can be classified into is given below.
IR Spectroscopy detects frequencies of infrared light that are absorbed by a molecule.
Molecules tend to absorb these specific frequencies of light since they correspond to the
frequency of the vibration of bonds in the molecule.
The energy required to excite the bonds belonging to a molecule, and to make them
vibrate with more amplitude, occurs in the Infrared region. A bond will only interact
with the electromagnetic infrared radiation, however, if it is polar.
The presence of separate areas of partial positive and negative charge in a molecule
allows the electric field component of the electromagnetic wave to excite the vibrational
energy of the molecule.
The change in the vibrational energy leads to another corresponding change in the dipole
moment of the given molecule. The intensity of the absorption depends on the polarity of
the bond. Symmetrical non-polar bonds in N≡N and O=O do not absorb radiation, as
they cannot interact with an electric field.
Regions of the Infrared spectrum
Most of the bands that indicate what functional group is present are found in the region
from 4000 cm-1 to 1300 cm-1. Their bands can be identified and used to determine the
functional group of an unknown compound.
Bands that are unique to each molecule, similar to a fingerprint, are found in the
fingerprint region, from 1300 cm-1 to 400 cm-1. These bands are only used to compare the
spectra of one compound to another.
Samples in Infrared Spectroscopy
The samples used in IR spectroscopy can be either in the solid, liquid, or gaseous state.
Solid samples can be prepared by crushing the sample with a mulling agent
which has an oily texture. A thin layer of this mull can now be applied on a salt
plate to be measured.
Liquid samples are generally kept between two salt plates and measured since the
plates are transparent to IR light. Salt plates can be made up of sodium chloride,
calcium fluoride, or even potassium bromide.
Since the concentration of gaseous samples can be in parts per million, the
sample cell must have a relatively long pathlength, i.e. light must travel for a
relatively long distance in the sample cell.
Answer: Gas chromatography (GC) and liquid chromatography (LC) are sophisticated
separation techniques widely employed in analytical chemistry. GC relies on the
vaporization and partitioning of volatile compounds in a column, while LC separates
compounds based on their solubility and interactions with the stationary phase. The
choice of column technology, including packing materials and capillaries, significantly
impacts the efficiency and effectiveness of chromatographic separations. The stationary
phase's nature and thickness, along with column dimensions, play crucial roles in
achieving optimal resolution and selectivity.
In GC, the separation mechanism is primarily based on the volatility of compounds.
Compounds with higher volatility elute faster, resulting in a chromatogram that reflects
their relative boiling points. In contrast, LC separation involves factors such as
solubility, interactions with the stationary phase, and molecular size. Column packing
materials, which can vary in polarity and selectivity, allow tailoring separation
conditions for specific analytes.
The significance of column technology is evident in the chromatographic performance it
affords. Packed columns, consisting of particles coated with a stationary phase, provide
efficient separations for a wide range of compounds. Capillary columns, with their
smaller diameter and increased surface area, offer improved resolution and sensitivity.
The careful selection of column packing materials ensures optimal interactions between
the stationary phase and analytes, contributing to the success of chromatographic
analyses.
Q2. Discuss the basic p rinciple and instrumentation of Mass
Spectrometry.
Basic Principle
A mass spectrometer generates multiple ions from the sample under
investigation, it then separates them according to the ir specific mass-to-
charge ratio (m/z), and then records the relative abundance of each ion
type.
Component/instrumentation
With all the above components, a mass spectrometer should always perform
the following processes:
Process the signals from the detector that are transmitted to the computer
and control the instrument using feedback.
Question 3: Examine the role of detection methods in chromatography and mass
spectrometry. Distinguish between simple and specific detection, and elaborate on how
detection is utilized for further analysis, including the use of tags and coupling to other
analytical techniques like infrared (IR) and mass spectrometry (MS).
Answer: Detection methods in chromatography and mass spectrometry are integral
components that contribute to the identification and quantification of analytes. Simple
detection methods, such as UV-Vis spectroscopy in chromatography, provide general
information about the sample, often based on the absorption of light by compounds.
Specific detection, as seen in mass spectrometry (MS), offers detailed information on
individual compounds through their unique mass spectra.
Detection serves as a means for further analysis by incorporating tags and coupling to
other analytical techniques. Tags, or chemical labels, are introduced to the analytes to
enhance specificity. For instance, in liquid chromatography-mass spectrometry (LC-
MS), specific tags can be attached to compounds, allowing for enhanced detection and
identification. Coupling chromatography or electrophoresis with mass spectrometry
provides additional information about the molecular structure and composition of the
compounds being analyzed.
Infrared (IR) spectroscopy is another powerful technique often coupled with
chromatography or mass spectrometry. It provides information about molecular
vibrations, aiding in the identification of functional groups within compounds. The
combination of chromatography, mass spectrometry, and infrared spectroscopy allows
for a more comprehensive analysis of complex samples, offering insights into both the
identity and structure of analytes.
Question 5: Explain the factors influencing resolution, time, and multiple separations in
mass spectrometry. Delve into the detection and interpretation processes, emphasizing
the linkage to excitation and its role in analyzing complex samples.
Answer: Resolution, time, and the ability to perform multiple separations are critical
considerations in mass spectrometry (MS), influencing the technique's analytical
performance. Resolution is influenced by various factors, including the type of ionization
technique, mass analyzer, and the strength of the magnetic field. Higher resolution
allows for the distinction of closely spaced peaks in a mass spectrum, providing more
detailed information about the components of a sample.
Time considerations in MS involve the speed of analysis. Time-of-flight (TOF)
analyzers, for example, provide rapid results by measuring the time ions take to travel a
specific distance. Faster analysis times are advantageous in high-throughput applications
where efficiency is paramount.
The ability to perform multiple separations is a characteristic feature of advanced MS
setups. By combining different ionization and separation methods, researchers can
achieve comprehensive separations of complex samples. For example, liquid
chromatography-mass spectrometry (LC-MS) combines the high separation efficiency of
liquid chromatography with the sensitivity and specificity of mass spectrometry.
Detection and interpretation processes in MS are essential for extracting meaningful
information from mass spectra. Detection involves the identification of ions based on
their mass-to-charge ratio (m/z). Interpretation includes the analysis of characteristic
peaks and fragmentation patterns in mass spectra. The linkage to excitation refers to the
initial energy input during ionization. Understanding excitation processes is crucial as it
influences the observed mass spectra and aids in the identification and structural
elucidation of compounds in complex samples.
In summary, MS is a powerful analytical technique that continues to evolve, driven by
advancements in resolution, speed, and the ability to perform multiple separations,
making it an indispensable tool in modern analytical chemistry.
Unit IV
Square-wave Voltammetry (SWV) is used for both quantitative chemical analysis and
study of the mechanism, kinetics, and thermodynamics of chemical reactions.
SWV used as an analytical tool offers three major advantages when compared to other
electrochemical techniques.
SWV is very sensitive, often allowing direct analyses at the ppb (parts per
billion) level and even the low ppt (parts per trillion) level when used in a
stripping mode.
SWV requires less time per sweep than older techniques such as differential
pulse polarography. A SWV sweep can often be recorded in less than ten
seconds, in contrast with a differential pulse polarogram that typically requires
more than two minutes for data acquisition.
In Figure, the reduction process occurs from (a) the initial potential to (d) the switching
potential. In this region the potential is scanned negatively to cause a reduction. The
resulting current is called cathodic current (i pc). The corresponding peak potential occurs
at (c), and is called the cathodic peak potential (E pc). The Epc is reached when all of the
substrate at the surface of the electrode has been reduced. After the switching potential
has been reached (d), the potential scans positively from (d) to (g). This results in anodic
current (Ipa) and oxidation to occur. The peak potential at (f) is called the anodic peak
potential (Epa), and is reached when all of the substrate at the surface of the electrode has
been oxidized.
Instrumentation:
A CV system consists of an electrolysis cell, a potentiostat, a current-to-voltage
converter, and a data acquisition system. The electrolysis cell consists of a working
electrode, counter electrode, reference electrode, and electrolytic solution. The working
electrode’s potential is varied linearly with time, while the reference electrode maintains
a constant potential. The counter electrode conducts electricity from the signal source to
the working electrode. The purpose of the electrolytic solution is to provide ions to the
electrodes during oxidation and reduction. A potentiostat is an electronic device which
uses a dc power source to produce a potential which can be maintained and accurately
determined, while allowing small currents to be drawn into the system without changing
the voltage. The current-to-voltage converter measures the resulting current, and the data
acquisition system produces the resulting voltammogram.
Applications: Cyclic Voltammetry can be used to study qualitative information about
electrochemical processes under various conditions, such as the presence of
intermediates in oxidation-reduction reactions, the reversibility of a reaction. CV can
also be used to determine the electron stoichiometry of a system, the diffusion
coefficient of an analyte, and the formal reduction potential, which can be used as an
identification tool. In addition, because concentration is proportional to current in a
reversible, Nernstian system, concentration of an unknown solution can be determined
by generating a calibration curve of current vs. concentration.
Types of Titration
There are many types of titration when considering goals and procedures. However, the
most common types of titration in quantitative chemical analysis are redox titration and
acid-base titration.
1. Acid-base Titrations
2. Redox Titrations.
3. Precipitation Titrations.
4. Complexometric Titrations.
1. Acid-Base Titration
The strength of an acid can be determined using a standard solution of a base. This
process is called acidimetry. In the same way, the strength of a base can be found with
the help of a standard solution of an acid, which is known as alkalimetry. Both titrations
involve in the neutralization reaction of alkali.
HA+BOH→BA+H2O
Or H+ + A– + B+ + OH– → B+ + A– + H2O
Or H+ + OH– → H2O
The acid-base titration is based on the reaction that neutralization is between a base or an
acidic and analyte. In this type, a reagent is mixed with the sample solution until it
reaches the required pH level. This type of titration majorly depends on the track change
in pH or a pH meter.
2. Redox Titrations
Permanganate Titrations
Dichromate Titrations
Iodimetric and Iodometric Titrations
Permanganate Titrations
Further, the solution remains colourless before the endpoint. The potassium
permanganate is used to estimate oxalic acid, ferrous salts, hydrogen peroxide, oxalates
and more. While the solution of potassium permanganate is always standardized before it
is used.
Dichromate Titrations
The solution of potassium dichromate can be directly used for titrations. It is mainly used
for the estimation of ferrous salts and iodides.
l2 + 2e → 2l–……………. (reduction)
2l– + 2e → 2e ……………. (oxidation)
The solution is used as an indicator. Free iodine is used in the iodometric titration, while
in the iodometric titration an oxidation agent is used to react to liberate free iodine.
3. Precipitation Titrations
The titration is based on the insoluble precipitate formation when the two reacting
substances are brought into contact are called precipitation titration. For instance, when
the solution of silver nitrate is used to a solution of ammonium thiocyanate or sodium
chloride, it reacts and forms a white precipitate of silver thiocyanate or silver chloride.
4. Complexometric Titrations
Redox titration is based on an oxidation-reduction reaction between the titrant and the
analyte. It is one of the most common laboratory methods to identify the concentration of
unknown analytes.
In order to evaluate redox titrations, the shape of the corresponding titration curve must
be obtained. In these types of titration, it proves convenient to monitor the reaction
potential instead of monitoring the concentration of a reacting species.
As discussed earlier, redox reactions involve both oxidation and reduction. The key
features of reduction and oxidation are discussed below.
Reduction
Oxidation
Thus, it can be understood that redox titrations involve a transfer of electrons between
the given analyte and the titrant. An example of a redox titration is the treatment of an
iodine solution with a reducing agent. The endpoint of this titration is detected with the
help of a starch indicator.
In the example described above, the diatomic iodine is reduced to iodide ions (I –), and
the iodine solution loses its blue colour. This titration is commonly referred to as
iodometric titration.
In this titration, the analyte is oxalic acid and the titrant is potassium permanganate. The
oxalic acid acts as a reducing agent, and the KMnO 4 acts as an oxidizing agent. Since the
reaction takes place in an acidic medium, the oxidizing power of the permanganate ion is
increased. This acidic medium is created by the addition of dilute sulfuric acid.
MnO−4+8H++5e−→Mn2++4H2O
KMnO4 acts as an indicator of where the permanganate ions are a deep purple colour. In
this redox titration, MnO4– is reduced to colourless manganous ions (Mn2+) in the acidic
medium. The last drop of permanganate gives a light pink colour on reaching the
endpoint. The following chemical equation can represent the reaction that occurs.
Molecular equation
2KMnO4+3H2SO4→K2SO4+2MnSO4+3H2O+5[O]
H2C2O4.2H2O+[O]→2CO2+3[H2O]×5
Complete Reaction
2KMnO4+3H2SO4+5H2C2O4.2H2O→K2SO4+2MnSO4+18H2O+10CO2
Ionic equation
MnO−4+8H++5e−→Mn2++4H2O]×2
C2O2−4→2CO2+2e−]×5
Complete Reaction
2MnO−4+16H++5C2O2−4→2Mn2++8H2O+10CO2
lies far to the right. Carrying out the reaction in a basic buffer solution removes H + as it
is formed, which also favors the formation of the EDTA-metal cation complex reaction
product. For most purposes it can be considered that the formation of the metal cation-
EDTA complex goes to completion, and this is chiefly why EDTA is used in titrations
and standardizations of this type.
Unit V
Principle:
Instrumentation:
NMR spectrometers consist of a powerful magnet, radiofrequency transmitter, and a
detector. The magnet aligns nuclear spins, RF pulses perturb the spin alignment, and the
resulting signals, detected by the instrument, generate the NMR spectrum.
Spin Coupling:
Spin coupling in NMR occurs when adjacent nuclei influence each other's magnetic
environment. This phenomenon, known as coupling or splitting, provides additional
information about the connectivity and structure of a molecule.
Applications:
Potentiometry:
Voltammetry:
Applications:
Applications:
X-ray Analysis:
X-ray analysis includes techniques such as X-ray diffraction (XRD) and X-ray
photoelectron spectroscopy (XPS). XRD is used for determining crystal structures, while
XPS provides information on elemental composition and chemical bonding.
Electron Spectroscopy:
Applications:
X-ray and electron spectroscopy techniques are essential in material science, providing
insights into crystallography, surface composition, and morphology. They are widely
used in fields such as nanotechnology, catalysis, and semiconductor research.
Extra Questions
To inspect cosmetics.
(ii)Retardation factor: The Rf (retardation factor) value is the ratio of the solute’s
distance travelled to the solvent’s distance travelled.
RF Value Explanation
The stationary phase in paper chromatography is water molecules found in the pores of
the filter paper, whereas the moving phase is a solvent such as hexane, toluene, acetone,
or a mixture of solvents such as methanol-water mixture. As the moving phase passes
through the area where the sample has been adsorbed, it dissolves the components more
or less quickly, depending on their solubility, and carries them with it as it moves across
the spot.
It is possible to determine the characteristic rate of movement of each substance on the
chromatography paper as the moving phase moves at a certain temperature and for a
specific solvent. This is represented by the Rf value, which stands for relative front or
retardation factor. Even if the mobile phase (solvent) is the same, various compounds
have varying Rf values. Furthermore, the Rf value of a chemical may vary depending on
the solvent. The following expression can be used to calculate Rf values:
Rf = Distance travelled by the substance from reference line (cm)/Distance travelled by
the solvent front from reference line (cm)
need of the RF value: In chromatography, Rf values are the most basic prerequisite of
the experiment. These numbers indicate whether the analyte (solute) prefers the
stationary or mobile phase. With stationary and mobile phases, R f values are used to
determine polarity, relative masses, and relative solubilities, among other things.
When comparing the results of one chromatogram to those of another, retention factors
come in handy. The retention factor for a given substance should stay constant if the
chromatogram is conducted under the same conditions (same mobile and stationary
phases). This enables the comparison of unknown materials to known materials. They
are not the same compound if the retention factor of an unknown substance differs from
that of a known substance.
Calculation of RF Value
Depending on the nature of the analytes and the stationary phases, a chromatogram must
first be generated with an appropriate solvent (mobile phase). After drying the
chromatogram, the locations (migration values) of the analytes and the solvent front are
measured.
This separation theory was introduced in the year 1940s which was published by
Richard Laurence Millington Synge and Archer Martin. It is also known as
Liquid-liquid chromatography (LLC). Or if gas is the mobile phase it is called Gas-
liquid chromatography (GLC).
The separation of the components from the sample mixture is carried out by the
process of partition of the components between 2 phases. Both phases are in liquid
form. In this process, the immiscible solid surface coated with the liquid surface on the
stationary phase is in the mobile phase. The liquid surface is immobilized by a
stationary phase which results in making its a stationary phase. The mobile phase
moves from the stationary phase and components get separated. The separation
depends on different partition coefficient.
Partition Chromatography Procedure
In the chemical and pharmaceutical sciences, both phases usually are solvents.[2]
Most commonly, one of the solvents is water, while the second is hydrophobic,
such as 1-octanol. Hence the partition coefficient measures how
hydrophilic ("water-loving") or hydrophobic ("water-fearing") a chemical substance
is. Partition coefficients are useful in estimating the distribution of drugs within the
body. Hydrophobic drugs with high octanol-water partition coefficients are mainly
distributed to hydrophobic areas such as lipid bilayers of cells. Conversely,
hydrophilic drugs (low octanol/water partition coefficients) are found primarily in
aqueous regions such as blood serum.
If one of the solvents is a gas and the other a liquid, a gas/liquid partition coefficient
can be determined. For example, the blood/gas partition coefficient of a general
anesthetic measures how easily the anesthetic passes from gas to blood. Partition
coefficients can also be defined when one of the phases is solid, for instance, when
one phase is a molten metal and the second is a solid metal, or when both phases are
solids. The partitioning of a substance into a solid results in a solid solution.
Distribution Ratio
Although the distribution ratio and partition coefficient are often used synonymously,
they are not necessarily so. Solutes may exist in more than one form in any particular
phase, which would mean that the partition coefficient (Kd) and distribution ratio (D)
will have different values. This is an important distinction to make as whilst the
partition coefficient has a fixed value for the partitioning of a solute between two
phases, the distribution ratio changes with differing conditions in the solvent.
1. Cartridge.
2. Columns fashioned like syringe barrels.
3. Disks.
Also, sorbent phases can be purchased, and typical column housings are manufactured of
polypropylene or glass. In order to contain the column with the sorbent phase, porous
frits made of polyethylene, stainless steel, or Teflon can be used [2].
There are some examples for applying SPE as sample preparation step before GC
detection of different analytes in a variety of samples. Some of them are pointed below.
1. Direct extraction, in which the coated fiber is immersed in the aqueous sample.
2. Headspace configuration, for sampling air or the volatiles from the headspace above an
aqueous sample. However, headspace techniques are more applicable to volatile
organics than to the semivolatile organic compounds.
3. Membrane protection configuration, in which the coated fiber is protected with a
membrane, for analyzing the analytes in too much dirty samples.