0% found this document useful (0 votes)
39 views4 pages

mms6110

The document provides detailed specifications for the MMS 6110 Dual Channel Shaft Vibration Monitor, including its applications, technical data, and environmental conditions. It outlines the power supply requirements, dimensions, and configuration needs for the monitoring system. Additionally, it includes connection diagrams and order numbers for the product and accessories.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
39 views4 pages

mms6110

The document provides detailed specifications for the MMS 6110 Dual Channel Shaft Vibration Monitor, including its applications, technical data, and environmental conditions. It outlines the power supply requirements, dimensions, and configuration needs for the monitoring system. Additionally, it includes connection diagrams and order numbers for the product and accessories.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Machine Monitoring Systems

MMS 6110
Dual Channel Shaft Vibration Monitor

Applications:
Machine Monitoring Systems

Technical data:

Measuring modes:

Programmable measuring parameters:

Limit supervision:
Machine Monitoring Systems

Module and sensor supervision:

Signal outputs at the connecting strip:

Operating elements at the module front:

Power supply: System design:


Redundant supply input via two supply Power consumption: At standalone operation, unlimited
inputs, decoupled via diodes. At least max. 6 W (max. 250 mA at 24 V) number of modules.
one supply input is required for the
supply of the module. Other supply voltages can be realized Max. 31 modules / 62 channels may be
with additional system power supplies. operated at one RS 485 bus
Supply voltage:
18....24....31.2 V DC If more modules / channels are
according to IEC 654-2, class necessary, e.g. with an MMS 6815,
DC4 another RS 485 bus must be installed.

Environmental conditions:
Protection class: Permissible relative humidity: Permissible shock:
Module: IP 00 according to DIN 40050 5....95%, non condensing according to IEC 68-2, part 29
Front plate: IP21 according to DIN peak value of acceleration:
40050 Permissible vibration: 98 m/s2
according to IEC 68-2, part 6 nominal shock duration:
Climate conditions: Vibration amplitude: 16 ms
according to DIN 40040 class KTF 0.15 mm in range 10...55 Hz
operating temperature range: EMC resistance:
0....+65°C Vibration acceleration: according to EN50081-1 / EN50082-2
Temperature range for storage and 16.6 m/s2 in range 55...150Hz
transport:
-30....+85°C
Machine Monitoring Systems

Dimensions:
PCB/EURO card format acc. to
DIN 41494 (100 x 160 mm)
Width: 30,0 mm (6 TE)
Height: 128,4 mm (3 HE)
Length: 160,0 mm
Net weight: app. 320 g
Gross weight: app. 450 g
incl. standard export packing
Packing volume: app. 2,5 dm3
Space requirements:
14 modules (28 channels) fit into each
19“ rack

Fig. 1
Alle Maße in mm
All dimensions in mm

Requirements on configuration PC:


Configuration of modules is made via Processor: Capacity of fixed disk:
the RS 232 interface on the module front 486 DX, 33 MHz min. 5 MB
or via the RS 485 bus by means of a Interfaces: Required working memory:
computer (laptop) with the following one free RS 232 interface (COM 1 min. 620 KB
minimum specifications: or COM 2) with FIFO type 156550 Operating system:
UART MS DOS Version 6.22 or higher or
WIN® 95/98 or NT 4.0

Connection diagram:
+24V/1
+24V/2
0V (24V)
b02 z02 d02
b14 EI1 U- UN+ UB+ A d04
b16 EI2 B z04 RS 485
d18 FX BP b04 BUS
FAKTOR X z22
ALARM STOP ES NGL1 z12 + Static Displacement Voltage CH1
b22 Statische Abstandsspannung Kanal1
EXT. SPERREN GND GND b12 -
d22 d22 KEY (IN) NGL2 d12 Static Displacement Voltage CH2
+ Statische Abstandsspannung Kanal2
MMS
Output CH1 for Indikation or Operation
6110 EO1 d14 Ausg. Kanal1 für Anzeige o. Verknüpfung
+
EO2 d16
+ Output CH2 for Indikation or Operation
Ausg. Kanal2 für Anzeige o. Verknüpfung
B2 AC1 z14 Dynamic Output CH1
+ Dynamisches Ausgangssignal Kanal1
CON 0.. GND b24 -
AC2 z16 + Dynamic Output CH2
out z08 Dynamisches Ausgangssignal Kanal2
SENS1-H I1+ z18 +
z10 SENS1-L I1- b18 Iout CH1 / Iout K1
-

Printed in Germany. Due to continued research and product development


B1 I2+ z20 +
z06 SENS1+ I2- b20 Iout CH2 / Iout K2
-
-24 V

epro reserves the right to change these specifications without notice.


b06 SENS1- SC-A d24 ALERT RELAIS IN QUIESCENT CURRENT
Transducer ALERT RELAIS IN RUHESTROM
Aufnehmer SC-D z24 DANGER RELAIS IN QUIESCENT CURRENT
PR 642./.. DANGER RELAIS IN RUHESTROM
B4 b10 SENC-C A1-C b26 ALERT 1
CON 0.. ALERT 1 A1-E b28
D1-C d26 DANGER 1
out d08 SENS2-H DANGER 1 D1-E d28
d10 SENS2-L A2-C b30 ALERT 2
ALERT 2 A2-E b32
B3
d06 SENS2+ D2-C d30 DANGER 2
-24 V b08 SENS2- DANGER 2 D2-E d32
Transducer
z26 CHANNEL CLEAR 1
Aufnehmer CHANNEL C1-C
z28
PR 642./.. CLEAR 1 C1-E
z30 CHANNEL CLEAR 2
CHANNEL C2-C
z32
CLEAR 2 C2-E
Fig. 2 Connection diagramm
Anschlußschema

Order number:
01/03 Reh

MMS 6110 Dual Channel Shaft Vibration Monitor……………………............................................................. 9100 – 00001


MMS 6910 W Operating accessories ................................................................................................................9510 – 00001
consisting of: operating and installation manual, configuration software and various connection cables

The F48M mating connector has to be ordered separately depending on the intended wiring technology.
6010 – 00001

© epro GmbH
Jöbkesweg 3 D-48599 Gronau Further Information:
Tel. +49 (0) 2562/709-245 Internet: www.epro.de
Fax +49 (0) 2562/709-255 E-Mail: [email protected]