Quick Start Guide 499aoz Dissolved Ozone Sensor Rosemount en 990774
Quick Start Guide 499aoz Dissolved Ozone Sensor Rosemount en 990774
00825-0100-3492, Rev AB
November 2019
Rosemount™ 499AOZ
Safety information
CAUTION
Sensor/process application compatibility
The wetted sensor materials may not be compatible with process composition and operating
conditions.
Application compatibility is entirely the operator's responsibility.
CAUTION
Equipment damage
Do not exceed pressure and temperature specifications
Pressure: 65 psig (549 kPa abs) max.
Temperature: 32 to 122 °F (0 to 50 °C)
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end
users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your
system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for
all systems used within the facility.
Contents
First steps..................................................................................................................................... 3
Install........................................................................................................................................... 6
Wire........................................................................................................................................... 10
Calibrate.................................................................................................................................... 15
Maintenance.............................................................................................................................. 17
Accessories................................................................................................................................ 20
2 Emerson.com/Rosemount
November 2019 Quick Start Guide
1 First steps
1.1 Unpack and inspect
Procedure
1. Inspect the shipping container. If it is damaged, contact the shipper
immediately for instructions.
2. If there is no apparent damage, unpack the container. Be sure all
items shown on the packing list are present. If items are missing,
notify Emerson immediately.
A. Membrane retainer
B. Membrane assembly
C. O-ring
D. Cathode
E. Electrolyte fill plug (wrap with pipe tape)
F. Pressure equalizing port
G. Sensor cable (integral cable shown)
1.3 Specifications
Table 1-1: Sensor Specifications
Cable length See Ordering information table in the Product Data Sheet for cable
length options.
Cable length (maximum) 300 ft. (91 m), up to 100 ft. (30.5 m) is standard.
(1) Weights and shipping weights are rounded up to the nearest whole pound or 0.5 kg.
2-in. tee 915240-03 PVC and Buna ¾-in. NFPT 120 °F (49 °C) 60 psig (515
N; body is kPa abs)
915240-04 schedule 80 1-in. NFPT
Low flow 24091-00 and Polycarbon- Compression 158 °F (70 °C) 90 psig
cell (1) 24091-01 ate/ fitting for (722 kPa abs)
polyester, ¼-in. O.D.
316 stainless tubing or
steel, and ¼-in. female
silicone national pipe
thread (FNPT)
4 Emerson.com/Rosemount
November 2019 Quick Start Guide
Valved 9390004 for Acrylic, 316 ¼-in. NFPT 150 °F (65 °C) 100 psig (858
rotameter use with low stainless (316 stainless kPa abs)
flow cell steel, and steel)
Flow: 0.5 to 5 Viton
gph (1.4 to 19
L/hr)
(1) Temperature and pressure specifications for the low flow cell exceed the temperature and
pressure specifications for the sensor.
2 Install
Install the sensor in a flowing sample. Keep the sample flow as constant as
possible at a value within the following limits:
6 Emerson.com/Rosemount
November 2019 Quick Start Guide
A. Union coupler
B. 1-in. national pipe thread (NPT), two places
C. Sensor body: Rosemount 499A
D. 1-in. NPT flow cell adapter
E. O-ring 2-222
F. 1½-in. sched 80 CPVC tee body
A. Union coupler
B. Adapter
C. 1-in. NPT (two places)
D. Sensor body: Rosemount 499A
E. O-ring 2-222
F. 2-in. sched 80 PVC tee body
8 Emerson.com/Rosemount
November 2019 Quick Start Guide
A. Inches
B. Millimeters
C. Outlet
D. Inlet
3 Wire
NOTICE
For additional wiring information on this product, including sensor
combinations not shown here, please refer to the Liquid Transmitter Wiring
Diagrams.
3 C Red RTD in
8 E Gray Anode
10 Emerson.com/Rosemount
November 2019 Quick Start Guide
10 F Orange Cathode
5 C Red RTD in
10 N/A N/A pH in
11 N/A N/A -5 V
12 N/A N/A +5 V
13 E Gray Anode
14 F Orange Cathode
12 Emerson.com/Rosemount
November 2019 Quick Start Guide
Note
Connect clear shield wires to solution ground terminal on TB 2. Use wire nut
and pigtail if necessary.
A Orange Cathode
B Gray Anode
C White Return
D White/red Sense
E Red RTD in
1 Cathode
2 N/A
3 RTD sense
4 Anode
5 RTD return
6 RTD in
NOTICE
Use a wire nut and pigtail (included) when connecting several wires to the
same terminal.
14 Emerson.com/Rosemount
November 2019 Quick Start Guide
4 Calibrate
4.1 Zero point calibration
Even in the absence of ozone, the sensor generates a small signal called the
zero current. Failing to correct for the zero current can introduce a bias,
particularly if the ozone concentration is small (<0.4 ppm). Zero the sensor
when it is first placed in service and every time the fill solution is changed.
To zero the sensor:
Procedure
1. Pour a cup of deionized or bottled water.
2. Place the sensor in the water.
3. Wait until the sensor current has reached a stable low value (at least
two hours).
4. Follow the transmitter prompts for zeroing the sensor.
Note
Refer to the manual for the transmitter you are using (Rosemount
56, 1056, 5081, or 1066).
Procedure
1. Place the sensor in the flow cell.
2. Start the sample and reagent flow.
3. Adjust the sample flow to the correct range.
4. Adjust the concentration so that it is near the upper end of the
operating range.
5. Wait for the readings to stabilize.
6. Follow the transmitter prompts to complete the calibration.
Note
Refer to the manual for the transmitter you are using (Rosemount
56, 1056, 5081, or 1066).
The sensitivity should be between 250 and 450 nA/ppm. For more
information, refer to the transmitter manual.
16 Emerson.com/Rosemount
November 2019 Quick Start Guide
5 Maintenance
Periodic maintenance and cleaning are required for best performance of the
sensor. Generally, the membrane and fill solution should be replaced every
four to six months. Sensors installed in harsh or dirty environments require
more frequent maintenance. The optimum maintenance frequency is best
determined by experience.
WARNING
Pressurized spray injury
Before removing the sensor, be absolutely certain that the process pressure
is reduced to 0 psig and the process temperature is lowered to a safe level!
Procedure
1. Unscrew the membrane retainer.
2. Remove the membrane assembly and O-ring.
See Figure 1-1.
3. Hold the sensor over a container with the cathode pointing down.
4. Remove the fill plug.
5. Allow the electrolyte solution to drain out.
6. Inspect the cathode.
a) If it is tarnished, clean it by gently rubbing in the direction of
the existing scratches (do not use a circular motion) with
400-600 grit silicon carbide finishing paper.
b) Rinse thoroughly with water.
10. Hold the sensor at about a 45 degree angle with the cathode end
pointing up.
11. Add electrolyte solution through the fill hole until the liquid
overflows.
12. Tap the sensor near the threads to release trapped air bubbles.
13. Add more electrolyte solution if necessary.
14. Place the fill plug in the electrolyte port and begin screwing it in.
15. After several threads have engaged, rotate the sensor so that the
cathode is pointing up and continue tightening the fill plug.
Do not overtighten.
16. Place a new O-ring in the groove around the cathode post.
17. Cover the holes at the base of the cathode stem with several drops of
electrolyte solution.
18. Insert a small blunt probe, like a toothpick with the end cut off,
through the pressure equalizing port.
See Figure 1-1.
CAUTION
Equipment damage
A sharp probe may puncture the bladder and destroy the sensor.
Do not use a sharp probe.
19. Gently press the probe against the bladder several times to force
liquid through the holes at the base of the cathode stem. Keep
pressing the bladder until no air bubbles can be seen leaving the
holes.
Be sure the holes remain covered with electrolyte solution.
20. Place a drop of electrolyte solution on the cathode; then place the
membrane assembly over the cathode.
21. Screw the membrane retainer in place.
18 Emerson.com/Rosemount
November 2019 Quick Start Guide
6 Accessories
Part number Description
915240-03 2-in. flow through tee with ¾-in. female national pipe
thread (FNPT) connections
20 Emerson.com/Rosemount
November 2019 Quick Start Guide
22 Emerson.com/Rosemount
November 2019 Quick Start Guide