ISOLAB Inc Scope 12 02 2015 - 1449083185

ISOLAB Inc. has been accredited for calibration and measurement capabilities under ISO/IEC 17025:2005, valid until December 2, 2018. The certificate details various calibration parameters including electrical current, resistance, voltage, temperature, length, mass, humidity, and frequency, along with their respective ranges and expanded uncertainties. The document outlines the laboratory's capabilities and the equipment used for calibration, ensuring a confidence level of approximately 95% in measurements.

Uploaded by

4821888
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
49 views5 pages

ISOLAB Inc Scope 12 02 2015 - 1449083185

ISOLAB Inc. has been accredited for calibration and measurement capabilities under ISO/IEC 17025:2005, valid until December 2, 2018. The certificate details various calibration parameters including electrical current, resistance, voltage, temperature, length, mass, humidity, and frequency, along with their respective ranges and expanded uncertainties. The document outlines the laboratory's capabilities and the equipment used for calibration, ensuring a confidence level of approximately 95% in measurements.

Uploaded by

4821888
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

Certificate # L2434

Scope of Accreditation
For
ISOLAB Inc
6260 Hawthorne Dr
Windsor, ON N8T 1J9
Pradip Jansari
519-948-8371

In recognition of a successful assessment to ISO/IEC 17025:2005 to the following Calibration and


Measurement Capabilities, accreditation has been granted to ISOLAB Inc. for the following:

Accreditation granted through: December 2, 2018

Calibration
Electrical – Current
Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-)
(1 to 10) mA 0.004 mA
0.006 mA Comparison with Fluke
(10 to 20) mA
8846A/Agilent 34401A
(20 to 100) mA 0.06 mA
DC Current Generate Multimeter
(0.1 to 1) A 0.006 A (Fluke 741B/760A
(1 to 3) A 0.02 A Calibrator sourcing)
(3 to 10) A 0.06 A
(1 to 10) mA 0.007 mA
(10 to 20) mA 0.003 mA
Measurement with
(20 to 100) mA 0.055 A
DC Current Measure Fluke 8846A/Agilent
(0.1 to 1) A 0.0075 A 34401A Multimeter
(1 to 3) A 0.003 A
(3 to 10) A 0.017 A
(0 to 100) µA 0.65 µA Comparison with Fluke
(0.1 to 100) mA 0.6 mA 8846A/Agilent 34401A
AC Current Generate
(0 to 1) A 0.007 A Multimeter
1 kHz
(0 to 3) A 0.02 A (Fluke 741B/760A
(0 to 10) A 0.07 A Calibrator sourcing)
(0 to 100) µA 0.22 µA
(0.1 to 100) mA 0.016 mA Measurement with
AC Current Measure
(0 to 1) A 0.0016 A 8846A/Agilent 34401A
1 kHz
(0 to 3) A 0.005 A Multimeter
(0 to 10) A 0.022 A

Form 28.8—Rev 11 – 3/1/14 Page 1 of 5


Certificate # L2434
Electrical – Resistance
Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-)
(0 to 10 Ω) 0.004 Ω
(1 to 100) Ω 0.015 Ω
Comparison with Fluke
(0.1 to 1) kΩ 0. 16 Ω
8846A/Agilent 34401A
(1 to 10) kΩ 1.5 Ω
Resistance Multimeter
(10 to 100) KΩ 15 Ω
(Fluke 741B/760A
(0.1 to 1) MΩ 0.002 MΩ
Calibrator sourcing)
(1 to 10) MΩ 0.0045 MΩ
(10 to 100) MΩ 0.82 MΩ
RTD Temperature
Simulation – 0.17 °C Fluke 741B/Fluke 743
(-200 to 400) °C
Temperature Recorder/ B/Decade box.
Controller/indicator

Electrical – Voltage
Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-)
(0 to 100) mV 0.007 5 mV Comparison with Fluke
(0.1 to 1) V 0. 35 mV 8846A/Agilent 34401A
DC Voltage Generate (1 to 10) V 0.0046 V Multimeter
(10 to 100) V 0.005 5 V (Fluke 741B/760A
(100 to 1 000) V 0.053 V Calibrator sourcing)
(0 to 100) mV 0.007 5 mV
(0.1 to 1) V 0.35 mV Measurement with
DC Voltage Measure (1 to 10) V 0.004 6 mV Fluke 8846A/Agilent
(10 to 100) V 0.005 3 mV 34401A Multimeter
(100 to 1 000) V 0.053 mV
(1 to 100) mV
10 Hz 0.6 mV
20 kHz 0.6 mV
50 kHz 1.0 mV
100 kHz 1.0 mV
(1 to 1) V
10 Hz 0.006 V
20 kHz 0.006 V
Comparison with Fluke
50 kHz 0.007 V
8846A/Agilent 34401A
100 kHz 0.01 V
Multimeter
(0 to 10) V
AC Voltage Generate (Fluke 741B/760A
10 Hz 0.06 V
Calibrator sourcing)
20 kHz 0.06V
50 kHz 0.07 V
100 kHz 0.1 V
(0 to 100) V
10 Hz 0.7 V
20 kHz 0.7 V
50 kHz 0.8 V
100 kHz 0.25 V

Form 28.8—Rev 11 – 3/1/14 Page 2 of 5


Certificate # L2434
Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-)
(0 to 100) V
10 Hz 0.1 V
20 kHz 0.1 V
50 kHz 0.7 V
100 kHz 1V
(0 to 1 000) V
10 Hz 7V
20 kHz 7V
50 kHz 7V
100 kHz 7V

(1 to 100) mV
10 Hz 0.11 mV
20 kHz 0.11 mV
50 kHz 0.7 mV
100 kHz 0.7 mV
(1 to 1) V
10 Hz 0.001 mV
20 kHz 0.001 mV
50 kHz 0.002 mV
100 kHz 0.007 mV
(0 to 10) V
10 Hz 0.01 V
Measurement with
20 kHz 0.01 V
AC Voltage Measure Fluke 8846A/Agilent
50 kHz 0.02 V
34401A Multimeter
100 kHz 0.07 V
(0 to 100) V
10 Hz 0.1 V
20 kHz 0.1 V
50 kHz 0.2 V
100 kHz 0.2 V
(0 to 1 000) V
10 Hz 0.9 V
20 kHz 0.9 V
50 kHz 0.9 V
100 kHz 1.0 V

TC Temperature
Simulation –
Temperature Recorder/
Controller/indicator

Type J &K (-180 to 1 370) °C 0.35 °C


Type R & S (-50 to 1 760) °C 0.6 °C Fluke 743/741B
Type T (-200 to 400) °C 0.15 °C

Form 28.8—Rev 11 – 3/1/14 Page 3 of 5


Certificate # L2434
Length – Artifacts and Standards 1D
Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-)
(1 to 6) in 75 µin Gage Blocks and Dial
Micrometer End Rods
(7 to 12) in 175 µin Indicator

Length – Hand Tools and Precision Gages 1D


Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-)
(1 to 6) in 100 µin
Calipers
(7 to 12) in 175 µin
Gauge Blocks/
Dial Indicators (0 to 1) in 15 µin
Micrometer Head
(1 to 6) in 100 µin
Micrometer (Outside)
(7 to 12) in 175 µin

Mass – Pressure/Low Vacuum


Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-) 2
(0 to 1) inH2O 0.002 InH2O
0.01 InH2O Ametek PK-II
(1 to 4) In H2O
Pressure (4 to 354) inH2O 0.03% of reading + 0.6R
Ametek HK-1000
(10 to 50) psi 0.03% of reading + 0.6R Ametek T-50
(50 to 10 000) psi 0.03% of reading + 0.6R

Vacuum (0 to 29) inHg 0.04 inHg + 0.06R Ashcroft PTE100

Mass – Torque
Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-) 2
(0.01 to 250) lbf∙in 0.6 % of reading + 0.6R
Torque Tools1 (250 to 1 000) lbf∙in 0.6% of reading + 0.6R Torque Tester/Torque
(0.01 to 250) lbf∙ft 0.15% of reading + 0.6R Arm and Weights
Torque Tools1 (100 to 1 000) lbf∙ft 0.15 % of reading + 0.6R

Thermodynamic – Humidity
Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-)
(10 to 50) %RH 0.10 % RH
Relative Humidity Thunder Scientific
(51 to 95) % RH 0.25 % RH
Model 6500

Form 28.8—Rev 11 – 3/1/14 Page 4 of 5


Certificate # L2434
Thermodynamic – Thermometers and Probes
Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-) 2
Thermocouples
Type J &K (-40 to 400) °C 0.3 °C
Type J&K (401 to 660) C 0.45 °C Temperature
Type R &S (-40 to 400) °C 0.45 °C Bath/Drywell, SPRT
Type R &S (401 to 660) °C 0.6 °C and Fluke 741B/Hart
Type T (-40 to 400) °C 0.3 °C Scientific 1502
RTD and Thermistor (-40 to 400) °C 0.15 °C
Probes (400 to 660) °C 0.5 °C
Temperature Bath,
Liquid in Glass SPRT and Fluke
(-30 to 250) °C 0.5 °C + 0.6R
Thermometer 741B/Hart Scientific
1502

Thermodynamic – Thermodynamic Sources


Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-)
SPRT and Fluke
(-40 to 250) °C 0.10 °C
Temperature Sources 741B/Hart Scientific
(250 to 660) °C 0.5 °C
1502

Time and Frequency – Frequency / Period


Calibration Range Expanded Uncertainty of Remarks
Parameter/Equipment Measurement (+/-)

(1 to 110) Hz 0.011 Hz
Frequency Comparison with Fluke
(110 to 1 100) Hz 0.013 Hz
Generate 8846A/Agilent 34401A
(1.1 to 11) kHz 0.028 Hz
Multimeter
(Fluke 741B/760A
(1 to 110) Hz 0.06 Hz
Calibrator sourcing)
Frequency Measure (110 to 1 100) Hz 0.6 Hz
(1.1 to 11) kHz 0.012 kHz

Calibration and Measurement Capability (CMC) is expressed in terms of the measurement parameter, measurement range,
expanded uncertainty of measurement and remarks. The expanded uncertainty of measurement is expressed as the standard
uncertainty of the measurement multiplied by a coverage factor of 2 (k=2), corresponding to a confidence level of
approximately 95%.

Notes:

1) Laboratory offers calibration services at the laboratory's own facilities and at the client or other agreed upon facilities.
2) L = Length in inches; R = Resolution of the unit under calibration.

Approved by: Date: December 2, 2015


R. Douglas Leonard
Chief Technical Officer
Issued: 12/2/15

Form 28.8—Rev 11 – 3/1/14 Page 5 of 5