Beginners Guide To Ion-Selective Electrode Measurements
Beginners Guide To Ion-Selective Electrode Measurements
Note that 2.303RT/nF is the Slope of the line (from the straight line plot of E versus log(A)
which is the basis of ISE calibration graphs) and this is an important diagnostic
characteristic of the electrode - generally the slope gets lower as the electrode gets old or
contaminated, and the lower the slope the higher the errors on the sample measurements.
For practical use in measuring pH, it is not normally necessary for the operator to construct
a calibration graph and interpolate the results for unknown samples. Most pH electrodes are
connected directly to a special pH meter which performs the calibration automatically. This
determines the slope mathematically and calculates the unknown pH value for immediate
display on the meter.
These basic principles are exactly the same for all ISEs. Thus it would appear that all can
be used as easily and rapidly as the pH electrode: i.e. simply by calibrating the equipment
by measuring two known solutions, then immersing the electrodes in any test solution and
reading the answer directly from a meter. Whilst it is certainly true that some other ions can
be measured in this simple fashion, it is not the case for most. Unfortunately, some ISE
advertising material tends to gloss over this fact and gives the reader a rather rosy view of
the capabilities of this technique. There are several factors which can cause difficulties
when ISE technology is applied to the measurement of other ions. These are listed below
and discussed in more detail in later sections. Nevertheless, it must be stressed here that
as long as these difficulties are recognised and steps are taken to overcome them, then
ISEs can still be a very useful and cost-effective analytical tool.
b) Differences Between pH and Other Ion-Selective Electrodes
i) In contrast to the pH membrane, other ion-selective membranes are not entirely ion-
specific and can permit the passage of some of the other ions which may be present in the
test solution, thus causing the problem of ionic interference.
ii) Most ISEs have a much lower linear range and higher detection limit than the pH
electrode. Many show a curved calibration line in the region 10-5 to 10-7 moles/l and very few
can be used to determine concentrations below 1x10-7 moles/l. Thus, for low concentration
samples, it may be necessary to construct a calibration graph with several points in order to
define the slope more precisely in the non-linear range.
iii) The calculation of ionic concentration is far more dependent on a precise measurement
of the potential difference than is the pH, because the pH depends on the order of
magnitude of the concentration rather than the precise value. For example it would take an
error of more than 5 millivolts to cause a change of 0.1 pH units, but only a 1 millivolt error
will cause at least a 4% error in the calculated concentration of a mono-valent ion and more
than 8% for a di-valent ion. This is because the theoretical value for the slope at 25°C is
59.2 for mono-valent ions and 29.6 for di-valent ions. In practical application, however these
slopes can vary considerably because of variations in temperature, deviations from "ideal"
behaviour, and minor impurities or contamination of the ion-selective membrane, or if
samples are measured near the detection limit of the electrode, in the non-linear range. The
critical factor is not so much the actual value of the slope but that this should be as high as
possible and remain constant over the range of concentrations and the time period required
for the analyses. Thus, when measuring other ion concentrations, it is essential to take extra
precautions to minimise any errors in the measurement of the electrode potential.
iv) For ion concentration measurements, steps must be taken to minimise the effect of the
Ionic Strength of the sample. This is because pH is defined as the log of the Activity of the
ion (which is measured directly by any ISE) but most measurements of other ions require
the actual concentration, which can differ significantly from activity in samples with complex
matrices and high Ionic Strength.
v) It is more usual to plot a calibration graph using the ionic concentration with a logarithmic
scale on the X-axis rather than the pX factor (analogous to pH) on a linear axis.
vi) Some ISEs will only work effectively over a narrow pH range.
Chapter 4.
TYPES OF ION SELECTIVE ELECTRODES
Ion-selective membranes are currently only available for a limited number of commonly
occurring ionic species. Examination of manufacturer’s catalogues reveals that the most
common are:
CATIONS: Ammonium (NH4+), Barium (Ba++), Calcium (Ca++), Cadmium (Cd++), Copper (Cu+
+
), Lead (Pb++), Mercury (Hg++), Potassium (K+), Sodium (Na+), Silver (Ag+).
ANIONS: Bromide (Br-), Chloride (Cl-), Cyanide (CN-), Fluoride (F-), Iodide (I-), Nitrate
(NO3-), Nitrite (NO2-), Perchlorate (ClO4-), Sulphide (S-), Thiocyanate (SCN-).
The manner in which these different membranes select and transport the particular ions is
highly variable and in many cases highly complex. It is far beyond the scope of this work to
explain in detail the exact mechanism for each ion. Moreover, it is not necessary for the
analyst to understand these mechanisms in order to use the electrodes satisfactorily.
Nevertheless, it may be of interest to the general reader to give some indication of these
processes. There are two main types of membrane material, one based on a solid crystal
matrix, either a single crystal or a polycrystalline compressed pellet, and one based on a
plastic or rubber film impregnated with a complex organic molecule which acts as an ion-
carrier. The development of these organic membranes was based on biological research
which revealed that some antibiotics and vitamins can induce cationic permeation through
cell membranes. One example of each membrane type is described below as an illustration
of the range of technologies employed.
b) Crystal-Membrane Electrodes e.g. Fluoride.
The Fluoride electrode is a typical example of the first type. Here the membrane consists of
a single lanthanum fluoride crystal which has been doped with europium fluoride to reduce
the bulk resistivity of the crystal. It is 100% selective for F- ions and is only interfered with by
OH-which reacts with the lanthanum to form lanthanum hydroxide, with the consequent
release of extra F- ions. This interference can be eliminated by adding a pH buffer to the
samples to keep the pH in the range 4 to 8 and hence ensure a low OH- concentration in the
solutions.
c) Impregnated-PVC-Membrane Electrodes e.g. Potassium.
The Potassium electrode was one of the earliest developed and simplest examples of the
second type. The membrane is usually in the form of a thin disc of PVC impregnated with
the macrocyclic antibiotic valinomycin. This compound has a hexagonal ring structure with
an internal cavity which is almost exactly the same size as the diameter of the K+ ion. Thus
it can form complexes with this ion and preferentially conducts it across the membrane.
Unfortunately it is not 100% selective and can also conduct small numbers of sodium and
ammonium ions. Thus these can cause errors in the potassium determination if they are
present in high concentrations. The majority of other ISEs suffer from similar limitations (see
later section on ‘interference’).
Several other complex processes are employed in ion-selective membranes and details of
these can be found in specialist electrochemistry textbooks and in catalogues from
manufacturers of "ionophores".
d) Care and Maintenance of ISEs.
When handling ISEs, care should be taken to avoid damaging the membrane surface. If the
electrodes are in frequent use then they can simply be left hanging in the electrode holder
with the membrane surface open to the air but protected by a clean dry beaker. For
prolonged storage in a cupboard or drawer, the membrane should be protected by covering
with the rubber or plastic cap which is normally provided with the electrode. After extensive
use the membranes may become coated with a deposit or scoured with fine scratches
which may cause a slow or reduced response (low slope) or unstable readings.
Crystal membranes can be regenerated by washing with alcohol and/or gently polishing with
fine emery paper to remove any deposit or discoloration, then thoroughly washing with de-
ionised water to remove any debris. After this, they may require soaking in the concentrated
standard solution for several hours before a stable reading can be re-established. It must be
noted, however, that prolonged immersion of crystal membranes in aqueous solutions will
eventually cause a build up of oxidation products on the membrane surface and thus inhibit
performance and shorten the active life. Conversely, PVC membranes should not even be
touched, let alone polished, and can be often be regenerated by prolonged (several days)
soaking in the standard solution, after removing any deposit with a fine jet of water, or
rinsing in alcohol.
Chapter 5
REFERENCE ELECTRODES
In order to measure the change in potential difference across the ion-selective membrane as the ionic
concentration changes, it is necessary to include in the circuit a stable reference voltage which acts as a half-
cell from which to measure the relative deviations.
a) The Silver / Silver Chloride Single Junction Reference Electrode.
The most common and simplest reference system is the silver / silver chloride single junction reference
electrode. This generally consists of a cylindrical glass tube containing a 4 Molar solution of KCl saturated with
AgCl. The lower end is sealed with a porous ceramic frit which allows the slow passage of the internal filling
solution and forms the liquid junction with the external test solution. Dipping into the filling solution is a silver
wire coated with a layer of silver chloride (it is chloridised) which is joined to a low-noise cable which connects
to the measuring system.
In electrochemical terms, the half-cell can be represented by:
Ag / AgCl (Satd), KCL (Satd)
and the electrode reaction is:
AgCl (s) + e- = Ag (s) + Cl-
The electrode potential for this half-cell is + 0.2046 V relative to the Standard Hydrogen Electrode at 25°C
b) Double Junction Reference Electrodes.
One problem with reference electrodes is that, in order to ensure a stable voltage, it is necessary to maintain a
steady flow of electrolyte through the porous frit. Thus there is a gradual contamination of the test solution with
electrolyte ions. This can cause problems when trying to measure low levels of K, Cl, or Ag, or when using
other ISEs with which these elements may cause interference problems. In order to overcome this difficulty the
double junction reference electrode was developed. In this case the silver / silver chloride cell described above
forms the inner element and this is inserted into an outer tube containing a different electrolyte which is then in
contact with the outer test solution through a second porous frit. The outer filling solution is said to form a "salt
bridge" between the inner reference system and the test solution and is chosen so that it does not contaminate
the test solution with any ions which would effect the analysis.
Commonly used outer filling solutions are:
potassium nitrate - for Br, Cd, Cl, Cu, CN, I, Pb, Hg, Ag, S, SCN.
sodium chloride - for K,
ammonium sulphate - for N03,
magnesium sulphate - for NH4,
Note that double junction reference electrodes are named after their outer filling solutions.
One disadvantage with double junction reference electrodes is that they introduce an extra interface between
two electrolytes and thus give the opportunity for an extra liquid junction potential to develop.
c) Liquid Junction Potentials.
It must be noted that the standard voltage given by a reference electrode is only correct if there is no additional
voltage supplied by a liquid junction potential formed at the porous plug between the filling solution and the
external test solution. Liquid junction potentials can appear whenever two dissimilar electrolytes come into
contact. At this junction, a potential difference will develop as a result of the tendency of the smaller and faster
ions to move across the boundary more quickly than those of lower mobility. These potentials are difficult to
reproduce, tend to be unstable, and are seldom known with any accuracy; so steps must be taken to minimise
them. Using 4 Molar KCL as the inner filling solution has the advantage that the K+and Cl- ions have nearly
equal mobilities and hence form an equi-transferrent solution. Also, in the single junction electrodes, the
electrolyte concentration is much higher than that of the sample solution thus ensuring that the major portion of
the current is carried by these ions. A third factor in minimising the junction potential is the fact that there is a
small but constant flow of electrolyte out from the electrode thus inhibiting any back-diffusion of sample ions -
although this is less important with modern gel electrolytes.
As indicated above, all these problems are doubled when double junction reference electrodes are used and
an additional problem arises in the case of the last three listed above (Sodium Chloride, Ammonium Sulphate,
Magnesium Sulphate) because the filling solutions are not equi-transferrent and hence have a stronger
tendency to form liquid junction potentials. It must be noted here that Nico2000 Ltd have recently introduced a
novel Lithium Acetate reference electrode which overcomes most of these problems and can be used with all
the ELIT range of ISEs. This is because it contains ions which are very nearly equi-tranferrent and which do
not interfere with any of the commonly used ISEs.
It must be noted that the E0 factor in the Nernst equation is the sum of all the liquid junction potentials present
in the system and any variation in this during analyses can be a major source of potential drift and error in
measurements.
d) Combination Electrodes
The majority of pH electrodes are produced in the form of combination electrodes in which the reference
system is housed in the same cylindrical body as the sensor head. This produces a simple, compact unit for
immersing in the test solution and has the added advantage that the two cells are in close proximity (with the
reference cell normally completely surrounding the sensor element) - thus minimising the effect of any stray
electrostatic fields or any inhomogeneity in the test solution. The main disadvantage of this arrangement is the
fact that it is the reference element which is the most likely to cause problems or fail, long before the ISE head
does, but the whole unit has to be replaced when failure does occur.
In contrast to pH electrodes, some ISEs are produced as mono-electrodes for use with separate reference
systems. One reason for this is because ISE membranes have a far lower impedance than pH sensors and
are less susceptible to stray electrostatic fields. Thus it is not necessary to screen the sensor head by
surrounding it with the reference system. More importantly, the membranes and internal construction of ISEs
are generally far more expensive than pH sensors and it is much more cost-effective to have separate units in
which the reference system can be replaced independently from the ISE.
e) Multiple Electrode Heads: Separable Combinations.
A new concept for combination electrodes has recently been introduced. Both the ISEs and the reference
electrodes are made in the form of 8mm diameter tubes fitted with a gold plated plug-in connector. These can
be inserted separately into special multiple electrode heads which are fitted with the cables and connectors for
attaching to the measuring system. The rigid plastic head ensures that the ISE and reference system remain
firmly linked together at a regular distance apart during operation, but either one can easily be replaced in the
event of failure or need to change the analysis. Moreover, the replacement electrodes are relatively
inexpensive compared to conventional electrodes because they do not incorporate the expensive low-noise
cables.
The ELIT Electrode Head System
A practical and cost effective way to combine Ion Selective and Reference Electrodes
ELIT Electrode Heads are manufactured from a robust plastic material and fitted with low noise cables and
connectors which are compatible with any standard mV/pH/ion meter.
The standard version, for use with an ELIT Ion Analyser / Computer Interface, has a BNC plug, but DIN, US or
S7 versions are available if required.
The sockets on the head and the pins on the plug-in electrodes are gold plated to assure good contact.
Advantages of this 'electrode combination' over conventional combination electrodes:
Use of one reference electrode for several ion-selective electrodes.
Replacement of a defective reference system without sacrificing the more expensive ISE.
Expensive low-noise cable and connector are attached to the re-usable head and do not need to be
replaced if the ISE becomes defective.
ISE is less expensive than conventional types with cable & connectors permanently attached.
ISE can be stored dry and the RE wet.
Increased distance between the ISE and the reference system reduces electrical interference and increases
the precision of measurement.
Five types of ELIT Heads are currently
available:
Mono Head for inserting one ELIT ISE or
Redox electrode, for use with a
conventional reference electrode
(or a combination pH electrode).
0.001 0.965 4%
0.0001 0.988 1%
0.001 0.965 4%
0.0001 0.988 1%
0.001 0.965 3%
0.0001 0.988 1%
0.001 0.965 3%
0.0001 0.989 1%
0.0001 0.955 5%
0.0001 0.955 5%
0.0001 0.955 4%
For samples with high ionic strength, there are five possible methods which can be used to avoid the error
introduced by the difference between activity and concentration.
1) Bring the ionic strength to the same level in both the calibrating standard solutions and the samples by
adding a suitable Ionic Strength Adjustment Buffer (ISAB) to both - see below, section c).
2) Dilute the samples to a level where the ionic strength effect is insignificant – but make sure that the
detected ion is still within the linear range of the electrode.
3) For samples with complex but known matrix, make up the standards in a similar solution which does not
contain the detected ion, or any which would interfere with the measurement.
4) Use the Activity Coefficient to calculate the concentration from the activity. As noted above, the activity
coefficient can be calculated for simple solutions with known concentrations of all the ions, but this is not
possible in many practical applications, where the samples may have a complex or unknown matrix.
5) Use the Standard Addition (or Sample Addition) Method where the voltage is measured before and after a
measured small volume of standard (or sample) is added to a larger measured volume of sample (or standard)
and the ionic strength is not altered significantly.
See later, Chapter 10, and www.nico2000.net/datasheets/staddl.html.
c) Ionic Strength Adjustment Buffers.
The most common way of overcoming the effect of the variable ionic strength of the solutions is to attempt to
make them all the same. Theoretically, this can be done by adding, equally to all standards and samples,
another solution of high ionic strength which does not contain the ion to be measured, or any likely interferent.
These solutions are known as Ionic Strength Adjustment Buffers (ISAB). The idea is that they are added in
sufficient quantity to completely swamp the ionic effects of the host solutions and hence give a uniform ionic
strength in all samples and standards. In this case the straight-line calibration curve can be constructed using
concentration units and the unknown concentrations can be read directly from the calibration graph. Note that
there is no need to recalculate the standard concentrations after adding ISAB as long as all standards and
samples are treated in the same way.
It must be noted, however, that most recipes for the addition of ISAB only produce an increase in ionic
strength of 0.1M and thus will only be effective if the IS of the original sample is much lower than 0.1M.
Furthermore, the difference between activity and concentration is relatively small at low Ionic strengths and
can often be ignored in many practical applications – see table above.
Most ISE suppliers also supply ISABs. These vary in composition depending on the detected ion and
occasionally contain other components which actively suppress interfering ions and/or control the pH value. In
many cases the composition of these solutions are carefully guarded ‘trade secrets’ and the reagents are
simply labelled as e.g. "ISAB for nitrate electrode".
d) Potential Drift.
The effect of potential drift can easily be seen if a series of standard solutions are repeatedly measured over a
period of time. The results show that the difference between the voltages measured in the different solutions
(i.e. the electrode slope) remains essentially the same but the actual value generally drifts in the same
direction by several millivolts (ie, not random variation or instability). This is somewhat surprising in view of the
fact that many ISE specifications quote a potential drift of less than 1 mV per day. However, it must be noted
that this figure is an attempt to define the drift which is attributable to the ISE alone - but this is impossible to
measure directly due to variations in other potential differences in other parts of the measuring system. The
most significant of these is the liquid junction potential of the reference electrode.
Although the LJP does not drift significantly once stabilised in a solution at constant temperature, it can
change by several millivolts every time a reference electrode is immersed in a new solution, or even re-
immersed in the same one. Thus the quoted ISE drift factor only applies when the ISE and reference are
constantly immersed in a 1000 ppm solution at constant temperature. Furthermore, constantly removing and
replacing an ISE in different solutions will produce hysteresis (or memory) effects, the extent of which depends
on the relative concentrations of the new and old solutions. Thus if the same solution is re-measured after
measuring a different one it cannot be expected to give exactly the same millivolt reading the second time.
In view of these various sources of potential drift, some manufacturers include a more realistic figure of 3 mV
per 8 hour working day in the electrode specifications.
For most ISE applications these effects are insignificant when compared to the overall reproducibility and
accuracy requirements, but, for the most precise work, steps must be taken to try to minimise these effects -
see Chapter 8.
Chapter 7.
CALIBRATION THEORY
Calibration is carried out by immersing the electrodes in a series of solutions of known concentration and
plotting a graph of the mV reading versus the log of the activity (or the actual activity on a logarithmic X-axis).
This should give a straight line over the whole linear concentration range. However, as noted above, activity is
difficult to determine in complex solutions and it is generally more useful to plot concentration units. In this
case the effect of variable activity coefficients in solutions with high ionic strength can be minimized by adding
Ionic Strength Adjustment Buffer to all standards and samples - but note the limitations to this detailed in the
previous chapter. Nevertheless, it must be noted that if the samples to be measured are likely to have a total
ionic strength of less than about 0.01M for monovalent ions (0.001M for divalent ions) then the activity effect
should be insignificant and it may not be necessary to add ISAB. However, it must be noted that ISAB may be
useful when using a double junction reference electrode with a non-equi-transferrent outer filling solution, in
order to compensate for drift in the liquid junction potential, and in general most ISE systems give a stable
reading more quickly in high ionic strength solutions.
The slope of the calibration graph is the mV response per decade of concentration change. This is typically
around 54 mV/decade for monovalent ions and 27 for divalent ions and will have a negative value for negative
ions - i.e. a higher concentration means more negative ions in solution and therefore a lower voltage.
a) Linear Range.
The linear range of the electrode is defined as that part of the calibration curve through which a linear
regression would demonstrate that the data points do not deviate from linearity by more than 2 mV. For many
electrodes this range can extend from about 0.1 Molar down to 10-6 or even 10-7 Molar.
b) Total Measuring Range.
The total measuring range includes the linear part of the graph as shown below together with a lower curved
portion where the response to varying concentration becomes progressively less as the concentration
reduces. Samples can be measured in this lower range but it must be noted that more closely spaced
calibration points are required in order to define the curve accurately and the percentage error per mV on the
calculated concentration will be progressively higher as the slope reduces.
c) Limit of Detection.
For monovalent ions, the IUPAC definition is: that concentration at which the measured potential differs from
that predicted by the linear regression by more than 18 mV. The practical limit of detection can be calculated
by plotting a calibration graph using several standards at the lower end of the concentration range, and below
it. Say 100, 10, 1, 0.1, 0.05, 0.01 ppm - i.e. at least two to define the linear slope and two to show the position
of the horizontal section below the limit of detection, where the electrode is unresponsive to concentration
change. The limit of detection is then defined by the crossing point of the two straight lines drawn through
these points.
Chapter 8.
CALIBRATION PRACTICE
a) Standard Solutions.
A single standard solution is usually diluted sequentially to produce concentrations of 1000
ppm, 100 ppm, 10 ppm, 1 ppm and even 0.1 ppm for a complete calibration (or this can be
done using decades of Molarity units). However, if the approximate concentration range of
the samples is known then the calibration should be made using standards which closely
bracket this range. Large errors may occur if samples are measured by extrapolation
beyond the range of the calibration. If you are confident that you are working in the linear
range of the electrode then only a two point calibration is necessary. But three or more are
recommended in order to confirm the linearity and to detect any errors in diluting the
standards, or to define the curve in the non-linear range. Since the slope of the calibration
line is temperature dependent, the temperature of the calibrating solutions and the samples
should be the same within a tolerance of ±2°C.
b) Minimising Drift Effects During Calibration
If more than three standards are being measured, it may help to minimise potential drift and
hysteresis during measurement, if the solutions are measured twice, the second time in the
reverse order (i.e. 1, 2, 3, 4, .., etc. then .., 4, 3, 2, 1 - all at regularly spaced intervals) and
the average mV value taken to construct the calibration graph. In this case, having defined
the calibration slope accurately, it may only be necessary to recalibrate using only one
representative standard, nearest to the range of the sample concentrations, and simply
normalise the rest of the points to this new value. For example, if a subsequent reading for
the recalibration standard gives a millivolt reading of 2 mV below the initial (average) value,
then you can either subtract 2 mV from all the other initial calibration data and re-plot the
graph or, probably more easily, add 2 mV to any sample readings taken immediately after
the recalibration and then read the concentrations directly from the original graph.
Alternatively (or possibly additionally) it may help (particularly for the lowest concentration
standard and during sample measurement) to always immerse the electrodes in pure water
for a short time, then drying, before immersing in the next solution. However, it must be
noted that prolonged immersion in pure water may damage the electrodes and they should
only be immersed for as long as it takes for the voltage to fall below that of the lowest
standard (or rise above for negative ions).
Despite these possible improvements to the conventional calibration method, the most
effective way of reducing potential drift is to use the Standard Addition method to measure
two standards (which bracket the known range of the samples) to find the average slope
and then measure the samples by Standard or Sample Addition methods - see Chapter 10.
c) Frequency of Calibration.
For the most precise results it is best to measure samples soon after calibration. Ideally
each sample measurement should be immediately preceded by a calibration (see Chapter
11 for more details of how to achieve the most precise results). This is relatively quick and
easy if only a one-point recalibration is employed as described above. Ultimately the
operator must decide what is the best compromise between the time taken to make frequent
calibrations and the precision requirements, but for many applications it will be sufficient to
make just one calibration graph before taking a series of sample measurements over
several hours.
Chapter 9.
MEASURING PROCEDURES
a) Adding ISAB
As noted above (Chap. 6) ISAB (Ionic Strength Adjustment Buffer) is normally added to
samples and standards in order to ensure that all measured solutions have the same Ionic
Strength and avoid errors due to differences between the measured activity and the actual
concentration. Indeed, many ISE manufacturers advocate always adding ISAB to samples
and standards, irrespective of whether the samples have high ionic strength or not, because
this can help to stabilise the liquid junction potential of the reference electrode and hence
reduce errors in measurement, and reduce the time taken to reach a stable reading - but
this may not be necessary if the reference electrode has an "equi-transferrent" filling
solution (i.e. both ions have the same, or nearly the same, mobility - e.g. Lithium Acetate, or
Potassium Nitrate).
Nevertheless, it must be noted that most electrode systems will give faster stabilisation with
ISAB when measuring low concentration samples in the non-linear range.
For many applications, however, it may be unnecessary (for samples with IS below 0.01M
for monovalent ions and 0.001M for divalent ions), or ineffective (for samples with IS greater
than about 0.1M), or inconvenient, to add ISAB to all the samples and standards.
One simple way to avoid adding ISAB is to dilute the samples to a level where the activity
effect is insignificant. But this requires a knowledge of the Ionic Strength of the samples,
and care must be taken to avoid diluting so much that the measurements would fall within
the non-linear range of the electrode. In some applications, where only the approximate
concentration of the samples are required, or the differences between samples are more
important than the actual concentrations, the effect of the ionic strength can often be
ignored. Alternatively, if the highest possible precision and accuracy is required then using
the Sample Addition or Standard Addition methods may be a better solution than adding
ISAB.
If it is decided that ISAB should be added then the most important factor is that it should be
added equally to standards and samples. Thus if the ISE instructions say, for example, that
ISAB should be added "2%v/v" it must be recognised that this is only an approximate
recommendation, and it will be more convenient to add 2ml of ISAB to 100ml of sample and
standard (or 1ml to 50ml) rather than adding 2ml to 98ml.
b) Taking Readings and Minimising Drift Effects.
There are several schools of thought as to the best way to take measurements. Some
authorities suggest that the solutions should be stirred slowly with a magnetic stirrer at 50 -
100 rpm during immersion of the electrodes - but care must be taken to ensure that there is
no heat exchange between the stirrer and the solution and that the stirrer is always at the
same speed. Big differences in mV readings can occur if the speed is varied. Moreover, the
magnitude of this effect is related to concentration (most variation in dilute solutions) and
hence will affect the slope of the calibration graph. Some prefer to take a reading whilst
stirring, others suggest that it is better to switch off the stirrer and take a reading in a still
solution - but this can greatly increase the time for the measurement because it may take
several minutes for the mV reading to stabilise whilst stirring, and several more to stabilise
after the stirrer is switched off. Alternatively, and more easily, the solution can simply be
swirled manually after immersion of the electrodes (to ensure good, homogeneous contact
between solution and membrane- i.e. no air bubbles) and then left to stand. This avoids the
problems of heat transfer and the inconvenience of adding and removing the magnetic
stirrer every time. It must be noted however that some electrode systems stabilise more
quickly and give more reproducible measurements when stirred whilst others give better
results in still solutions - this is sometimes mentioned in the electrode operating instructions.
The time at which the measurements are taken can also vary depending on the
characteristics of the particular type of electrode system being used and the balance
between time constraints and precision requirements. In some cases it is best to wait for a
stable mV reading (this can take several minutes). In others it is better to take all readings
after a pre-specified time after immersion. Generally, the mV reading changes rapidly in the
first 10 or 20 seconds as the ISE membrane equilibrates with the solution, then more slowly
and exponentially as the reference electrode liquid junction poential stabilises. Always
taking a reading after say 1 or 2 minutes (depending on which electrode system is being
used) should ensure that all are taken in the shallow part of the stabilisation curve where
only small and insignificant changes are occuring. A third alternative is to observe the drift in
reading as the electrodes equilibrate after immersion and then take a reading at the point
where the direction of drift is definitely reversed - i.e. a different electrochemical process
begins to dominate - but this last effect is not common.
If the highest possible precision is required then it is suggested that each operator should
make his own simple experiments by comparing the reproducibility of repeated
measurements on the same sample using each method discussed above, in order to
determine which is most suitable for his particular application. But note that whichever
method is preferred, the same procedure must be used for all standards and samples in
order to ensure that the electrode response will be as uniform as possible.
c) Cleaning Electrodes and Minimising Hysteresis Between Samples.
The electrode tips must be rinsed by spraying with a jet of deionised water and gently
dabbed dry with a low-lint laboratory tissue between measurements. For the most precise
results, it may help to minimise hysteresis effects if the electrodes are soaked in deionised
water for 20 or 30 seconds after rinsing, before every measurement, so that each new
reading is approached from the same direction, (i.e. always from low concentration to high -
as recommended for calibration measurements, to also minimise cross contamination)
rather than just being dependent on the last sample measured
Chapter 10.
METHODS OF ANALYSIS
a) Direct Potentiometry
Direct potentiometry is the simplest and most widely used method of using ISEs as described above in the
Basic Theory and Calibration sections of this work. Simply measure the electrode response in an unknown
solution and read the concentration directly from the calibration graph (either manually or using special
computer graphics and calculations - see later) or from the meter display on a self-calibrating ion meter. A big
advantage of this method is that it can be used to measure large batches of samples covering a wide range of
concentrations very rapidly without having to change range, recalibrate or make any complicated calculations.
Moreover, if ISAB is not being used, it is not necessary to measure the volume of the samples or standards.
Quite acceptable results can be obtained for some elements by simply dangling the electrodes in a river or
pond or effluent outflow without the need to take samples in small beakers.
b) Incremental Methods
There are three main types of incremental methods in general use:
Standard (or Known) Addition,
Sample Addition,
Sample Subtraction.
Standard Addition and Sample Addition Methods.
These methods involve measuring the voltage in a relatively large, accurately measured, volume of sample
(for standard addition) or standard (for sample addition) then adding a much smaller volume of standard (or
sample) and taking a second reading after the voltage has stabilised in the mixture. For full details
see: www.nico2000.net/datasheets/staddl.html
Summary of Advantages over Direct Potentiometry
The electrodes remain immersed throughout the process so that there is little change in the liquid
junction potential of the reference electrode (which can often be changed by several millivolts when
the electrodes are removed from one solution and placed in another) between calibration and sample
measurement - and therefore this source of measurement error is virtually eliminated.
Calibration and sample measurement are both made essentially at the same time and in the same
solution so that ionic strength and temperature differences between standard and sample are not
significant and ISAB is not normally required.
Once the approximate concentration for the samples is known, the calibration (slope) can be "fine
tuned" by analysing a standard with a concentration that lies within the range of the samples (and is at
the same temperature) and then adjusting the slope and re-calculating the results until the standard
gives the correct answer. This "fine tune" procedure is very quick and easy using the ELIT ISE/pH Ion
Analyser Software.
Measuring the slope at or very near to the sample concentration means that these methods can be
used with old or worn electrodes which may not be completely linear over their whole range, as long
as the slope is stable and reproducible over the limited range of the samples.
The most effective way of minimising the variation in LJP is by using Standard Addition or Sample Addition
Techniques (see later - section d). Alternatively, but less effectively, by using reference electrodes with nearly
equi-transferrent filling solutions (in which both ions have the same mobility when diffusing through the
ceramic tip) such as KNO3 or Li Acetate - but this is not always possible (depending on likely interference
effects).
The accuracy of the results is affected by several other factors:
1) The presence of interfering ions.
2) Any difference in ionic strength between the sample and standard solutions.
3) Any difference in temperature between sample and standards - A re-calibration should be made if the
sample temperature changes by more than 1 degree C from the calibration temperature.
4) Any variation in the electrode slope in different parts of the curve. Although the calibration graph may show
a straight line over several decades of concentration with an average slope of say 54.5 ± 2 mV/dec. it is highly
unlikely that this will be exactly the same across the whole of this range. If separate two-point calibrations are
made between two more closely spaced points at different concentration ranges then there may be a variation
of several millivolts between the individual slopes. Thus, if samples are calculated using the overall slope then
they will give results which will differ in concentration from those calculated using the appropriate individual
slope by 4% times the difference in mV between the two slopes.
Therefore, for the most accurate results, it is recommended that the electrode slope is determined using two
standards which closely span the expected range of the samples. It must be noted, however, that it is not
beneficial to have standards too close together because the measured slope is dependent on the difference in
voltages. So, for example, if the difference in mV is 50 then a 1 mV error in measurement will only cause a 2%
error in the slope but if the difference is only 10 mV then the same measurement error will result in a 10% error
in the slope. Thus it is normally recommended that calibration standards are about an order of magnitude
different in concentration and should not be less than 20 mV difference in reading.
Nevertheless, whichever slope is used, the reproducibility of replicate measurements of the same sample
should be the same.
By taking special precautions to overcome drift problems (such as frequent recalibration and ensuring that you
wait for stable readings, or read after a regular time interval), and by adding special ISABs to equalise activity
effects and remove interfering ions, direct potentiometry can give very reasonable results (reproducibility of ± 2
or 3%, one standard deviation, and accurate within these precision limits). Even without taking these
precautions, it is possible to achieve satisfactory reproducibility and accuracy (± 10 to 15%) for many
applications where the highest accuracy is not necessary and ionic strength and interfering ions are not a
problem.
b) Reproducibility Experiments using an Ammonium Electrode.
Some of the suggestions in the foregoing discussion, and the levels of accuracy and precision achievable with
careful work, can be illustrated with the results of some experiments conducted by the author. Reproducibility
tests were carried out using an ‘ELIT’ 8 mm diameter, solid-state ammonium electrode (PVC membrane) with
a lithium acetate double junction reference electrode and pure ammonium solutions (no ISAB). Standard
solutions containing 1 ppm and 10 ppm NH4+ were used for calibration and a 5 ppm solution was used as the
test sample. Measurements were made after immersing the electrodes in approximately 50 mls of solution in a
100 ml beaker, swirling the solution for 5 secs. then leaving to stand for 20 secs. Each millivolt measurement
was the average of ten readings taken at one second intervals. The electrodes were rinsed with a jet of de-
ionised water, then soaked in a beaker of water for twenty seconds, then dabbed dry with a low-lint tissue
between each measurement. The solutions were measured in the sequence 1 ppm, 5 ppm, 10 ppm, and this
pattern was repeated six times. The data were obtained using a meterless PC interface and specially written
software.
For this experiment, the concentration results were calculated with an EXCEL spreadsheet using the Nernst
equation in the standard form for a straight line: y = mx + c.
Where:
y is the measured voltage,
m is the electrode slope
(calculated from the two-point calibration data: (V1-V2)/ ((Log ppm1) - (Log ppm2))),
x is the logarithm of the concentration in the sample,
c, the intercept on the y axis, is Eo.
The experimental data were processed in several different ways:
1) Using only the first measurement of the two standards to define the slope and intercept, six measurements
of the 5 ppm sample, taken over approximately half an hour, gave an average of 4.71 ± 0.14 ppm (±2.96% one
standard deviation). However it was noticeable that successive measurements gave progessively lower values
due to electrode drift after calibration (causing a difference of nearly 8% between the highest and lowest
results).
2) The drift effect was compensated for by recalculating each result using different values for the slope and
intercept as defined by the standards measured immediately adjacent to each sample measurement. This
produced a significant improvement in the reproducibility and only a random variation in the results rather than
a progressive drift downwards. This clearly demonstrates the importance of measuring samples soon after
calibration. The average concentration this time was 4.90 ± 0.06 ppm (±1.20%, 1 S.D.) Although remarkably
precise and very close to the true value, the accuracy of this average is not quite within the precision limits. As
noted above, this can probably be explained by variation in the electrode slope and this suggestion is
supported by examining the individual slope values which can be calculated from the various measurements.
The average value for six determinations of the slope between 1 and 5 ppm was 55.92 ± 0.92 whereas that
between 5 and 10 ppm was 58.21 ± 0.78; i.e. there is a significant difference in slope between the two
adjacent ranges.
3) A third method of calculating these results, using the slope defined by the first calibration for all samples but
a different intercept value as given by each successive two-point calibration, was less satisfactory and gave
4.87 ± 0.12 ppm (± 2.35%) which is only slightly better than the results using only a single calibration at the
beginning. Thus these data would appear to suggest that the effect of electrode drift is more significant in
producing changes in the measured slope between different sample measurements rather than producing a
change in the calculated value for the intercept. This conclusion is also borne out by examining the individual
calibration data. Whereas the average slope between 1 and 10 ppm was 56.74 ± 0.51 (± 0.90%) for six
successive measurements and these showed a gradual drift downwards (57.49, 57.04, 56.84, 56.64, 56.49,
55.98) the associated intercept calculations showed a more random distribution and gave a much more
precise average value of 346.18 ± 0.26 mV (± 0.07%).
c) Reproducibility of Chloride Measurements.
A second experiment using the same techniques as above, but with a chloride (crystal membrane) electrode
and calibration standards of 25 and 250 ppm also yielded very impressive results. Eight measurements of a
100 ppm test solution gave an average of 95.4 ± 0.6 ppm (± 0.63%) when two-point calibrations were made
immediately prior to each sample measurement. Again the relatively poor accuracy is probably due to the
calibration standards being relatively far from the measured sample and thus the calculated slope was
probably not exactly correct for the 100 ppm range.
d) Standard Addition and Sample Addition Methods
These methods can potentially yield even more accurate and precise results than direct potentiometry
because the calibration and sample measurement stages are made essentially at the same time and in the
same solution (so that Ionic Strength and temperature differences are not significant) - and the electrodes
remain immersed throughout the measurements so that hysteresis, memory, and variations in the reference
electrode liquid junction potential are eliminated. They are paticularly useful for samples with high ionic
strength or a complex matrix. See Standard & Sample Addition Methods for a full description and experimental
results for precision and accuracy.
e) Conclusions from the Experimental Data.
These experimental results demonstrate that in order to obtain the best possible accuracy and precision, it is
important to measure samples soon after calibration and to use standard solutions that closely bracket the
expected range of sample concentrations. Furthermore, for direct potentiometry measurements, it is best to
make a full two-point recalibration every time, in order to obtain the most precise value for the slope, rather
than just making a single point recalibration and assuming that the slope is constant. This is not necessary for
Standard and Sample Addition techniques because of the possibility of recalculating the results for a known
standard to "fine tune" the slope measurement in the middle of the concentration range expected for the
samples.
These results show that it is possible to obtain accuracy and precision levels of better than ± 3% fairly easily,
and better than ± 2% by making more frequent calibrations or by using Standard or Sample Addition
techniques (better than ± 1% for some crystal membrane electrodes). Thus it has been shown that, with
careful use and a full consideration of all the limiting factors, ISE technology can be compared favorably with
other analytical techniques which require far more sophisticated and costly equipment.
Nevertheless, it must be stressed that these special considerations are only necessary to achieve the highest
possible precision and accuracy. For many applications, a simple two-point calibration followed by one direct
potentiometry measurement will probably be quite sufficient
Chapter 12.
TYPES OF MEASURING DEVICES AND DATA PROCESSING
a) Analogue Meters
The earliest type of pH meters were simple analogue devices with a resolution of only one
or two millivolts. The original meters were calibrated in millivolts and the corresponding pH
value was read from a calibration graph as described earlier. However it was soon realised
that pH electrodes are reasonably uniform and reproducible instruments with very similar
slopes so that it is not necessary to have a unique calibration graph for each electrode. In
this case the meters can be calibrated directly in pH units by the manufacturer and can
simply be recalibrated each time they are used (to compensate for temperature changes or
slight differences in electrode response) by immersing the electrode in just one pH buffer
solution and adjusting the meter output to give the correct reading. This type of meter is
very simple and quick to use and is perfectly adequate for many pH measurements because
it requires a change of more than 5 mV in order to change the pH value by more than 0.1
pH units.
Nevertheless, as noted earlier, in the direct measurement of ionic concentration, a change
of only 1 mV will represent a change of 4% in the concentration of mono-valent ions and 8%
for di-valent ions. Therefore it is necessary to have a more precise measuring system in
order to determine ionic concentrations and for the most precise pH measurements.
Furthermore, Ion Selective electrodes are more variable in their response and do not have
such uniform slopes so it is more important to make a complete multi-point calibration in
order to determine the slope for each electrode, each time it is used. Thus simple pre-
calibrated analogue meters are not appropriate for ISE measurements and in this case it
was always necessary for the analyst to keep a record of the millivolt readings obtained in
the standard and sample solutions. Then a calibration graph was plotted on semi-log paper,
using concentration units on the logarithmic axis, and the sample results were read directly
from this graph, with varying degrees of accuracy depending on the range of concentrations
plotted and the skill of the operator.
b) Digital Meters
A major advance was made when digital meters were introduced with a resolution of 0.1 or
even 0.01 mV. This enabled the analyst to measure and read the voltage with much greater
accuracy and meant that the stability and reproducibility of the electrode response became
the main limiting factor in determining the accuracy and precision. Nevertheless, although
these meters were also capable of being calibrated in pH units, it was still necessary for the
analyst to keep a log of the millivolt readings for ISE measurements and then plot the
calibration on graph paper or using computer graphics. Sample results were then read from
this graph, or calculated manually or by using computer spreadsheets, as with analogue
meters.
c) Self-Calibrating, Direct-Reading Ion Meters.
The next major advance was when micro-processors were introduced which contained
simple programs to calculate the slope and intercept from the calibration data and then use
these to calculate the sample concentration from the millivolt reading in the sample. Thus
the analyst can simply enter the concentrations of the standards and measure the millivolts,
then immerse the electrodes in the sample and read the sample concentration directly from
the meter. Nevertheless, these meters are often quite confusing to operate, with small
keypads and multi-function switches, and they are not suitable for working in the non-linear
range of the electrodes, or using different slopes for different parts of the calibration range,
or for measuring more than one ion at a time. Furthermore, it is often difficult for the analyst
to assess the quality of the calibration or detect errors in data entry and it is still necessary
for the results to be transferred manually to a permanent record.
Most of these problems of data acquisition, transfer, processing, and storage have now
been overcome with the introduction of meterless computer interfaces and sophisticated
software packages which are the subject of the next section of this guide
Chapter 13.
RECENT DEVELOPMENTS
a) Meterless Electrode-Computer Interfaces
The recent development of computer interfaces for connecting the sensor electrodes directly to a personal
computer, without the need for a meter to measure the voltage, has revolutionised ISE data acquisition and
processing. The earliest versions of this type of measuring system were designed in the form of a flimsy
PCMCIA card which were specifically for use only with a lap-top or palm-top computer and with only one
particular style of electrode with special connecting cables.
Later models are far more versatile and built in the
form of a rugged electronic box connected to the
computer and to the electrodes by standard cables
and connectors. They can be used with any type of
electrode plugged into the interface in the same way
as for any conventional ion/pH meter. The electrode
signals are processed by a low-noise amplifier and
analogue-to-digital converter for direct transfer into
any desk-top or lap-top computer. The user-friendly
software will run on any computer using Windows
operating system and is designed for maximum
efficiency and versatility. Apart from the ease and
speed of operation, the great advantage of this type
of measuring system is that it virtually eliminates any
possibility of operator error in recording and
transfering data. It also facilitates far more
sophisticated data processing, and display and
archiving of results, once the raw data are securely
recorded in the computer.
b) Data Processing Software
Data processing software is now available to carry out
most of the functions previously done by the operator,
and much more. The software basically carries out six
main functions:
1) Hardware Set-up: to configure the system for
whichever sensors and reference electrodes are
being used (e.g. temperature, pH, Redox, or any ISE)
and give a full documentation of electrode types, serial numbers, operator details, date and time etc.
2) Signal measurement: The software interprets the signal from the interface and uses a calibrating factor to
produce an accurate display of the millivolts. The operator can select to take single readings, taking about one
second each, or the average of multiple readings. He can also specify a time delay after immersion of the
electrodes, before taking a reading, if required.
3) Calibration: the only operator entry required is of the concentration (or pH) of the calibration standards. For
ISE measurements, calibration graphs are plotted automatically and instantaneously and can be assessed
with various line-fitting techniques.
4) Sample calculation: the sample results are calculated instantaneously with no possibility of operator error.
All results are saved and displayed in data tables where sample numbers and comments can be added.
Sample results can be calculated using partial slopes if the calibration in the range required is not completely
linear. Concentration results can be reported as ppm or moles/l. Samples can be measured using direct
potentiometry or by Sample Addition and Standard Addition methods.
5) Display, reporting and archiving of results: all data can be stored, printed or imported to other software
packages. Continuous recording of pH measurements can be displayed graphically, in real-time, to facilitate
such activities as pH titrations or process monitoring.
6) Help and Advice for the Operator: the software contains several help menus and drop-down hints and
warnings which guide the operator reliably through the analysis. It also includes an off-line ‘Chemtools’
package which permits the simple calculation of various electrochemical parameters and provides information
on interference factors and activity coefficients. Since the computer interfaces and software were designed to
improve on existing ion meters, the initial concept was based on the most up-to-date ion meters available.
These generally have two inputs, one for pH and one for ISE measurements. They contain microprocessors
for making pH and ISE calibrations and reporting the results as millivolts, pH, or concentration. Some of these
meters also include a temperature sensor. Thus the initial design for the interface was with two channels which
could perform all of these functions, but better.