0% found this document useful (0 votes)
81 views21 pages

AAMI EC53 - 2013-ECG Trunk Cables and Patient 1

The document outlines the AAMI standard ANSI/AAMI EC53:2013 for ECG trunk cables and patient leadwires, emphasizing the importance of safety and performance criteria for medical devices. It highlights the voluntary nature of AAMI standards and the need for professional judgment in their application, while also detailing the standard's objectives and the committee's role in its development. Additionally, it provides guidelines for labeling, construction, and performance requirements to ensure effective and safe use in clinical environments.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
81 views21 pages

AAMI EC53 - 2013-ECG Trunk Cables and Patient 1

The document outlines the AAMI standard ANSI/AAMI EC53:2013 for ECG trunk cables and patient leadwires, emphasizing the importance of safety and performance criteria for medical devices. It highlights the voluntary nature of AAMI standards and the need for professional judgment in their application, while also detailing the standard's objectives and the committee's role in its development. Additionally, it provides guidelines for labeling, construction, and performance requirements to ensure effective and safe use in clinical environments.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 21

American

Draeger Medical Systems Inc


AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

National
Standard

ANSI/AAMI
EC53:2013
ECG trunk cables and patient
leadwires
Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Objectives and uses of AAMI standards and


recommended practices

It is most important that the objectives and potential uses of an Each AAMI standard or recommended practice reflects the
AAMI product standard or recommended practice are clearly collective expertise of a committee of health care professionals and
understood. The objectives of AAMI's technical development industrial representatives, whose work has been reviewed
program derive from AAMI's overall mission: the advancement of nationally (and sometimes internationally). As such, the consensus
medical instrumentation. Essential to such advancement are (1) a recommendations embodied in a standard or recommended practice
continued increase in the safe and effective application of current are intended to respond to clinical needs and, ultimately, to help
technologies to patient care, and (2) the encouragement of new ensure patient safety. A standard or recommended practice is
technologies. It is AAMI's view that standards and recommended limited, however, in the sense that it responds generally to
practices can contribute significantly to the advancement of perceived risks and conditions that may not always be relevant to
medical instrumentation, provided that they are drafted with specific situations. A standard or recommended practice is an
attention to these objectives and provided that arbitrary and important reference in responsible decision-making, but it should
restrictive uses are avoided. never replace responsible decision-making.
A voluntary standard for a medical device recommends to the Despite periodic review and revision (at least once every five
manufacturer the information that should be provided with or on years), a standard or recommended practice is necessarily a static
the product, basic safety and performance criteria that should be document applied to a dynamic technology. Therefore, a standards
considered in qualifying the device for clinical use, and the user must carefully review the reasons why the document was
measurement techniques that can be used to determine whether the initially developed and the specific rationale for each of its
device conforms with the safety and performance criteria and/or to provisions. This review will reveal whether the document remains
compare the performance characteristics of different products. relevant to the specific needs of the user.
Some standards emphasize the information that should be provided Particular care should be taken in applying a product standard
with the device, including performance characteristics, instructions to existing devices and equipment, and in applying a recommended
for use, warnings and precautions, and other data considered practice to current procedures and practices. While observed or
important in ensuring the safe and effective use of the device in the potential risks with existing equipment typically form the basis for
clinical environment. Recommending the disclosure of the safety and performance criteria defined in a standard,
performance characteristics often necessitates the development of professional judgment must be used in applying these criteria to
specialized test methods to facilitate uniformity in reporting; existing equipment. No single source of information will serve to
reaching consensus on these tests can represent a considerable part identify a particular product as "unsafe". A voluntary standard can
of committee work. When a drafting committee determines that be used as one resource, but the ultimate decision as to product
clinical concerns warrant the establishment of minimum safety and safety and efficacy must take into account the specifics of its
performance criteria, referee tests must be provided and the reasons utilization and, of course, cost-benefit considerations. Similarly, a
for establishing the criteria must be documented in the rationale. recommended practice should be analyzed in the context of the
A recommended practice provides guidelines for the use, care, specific needs and resources of the individual institution or firm.
and/or processing of a medical device or system. A recommended Again, the rationale accompanying each AAMI standard and
practice does not address device performance per se, but rather recommended practice is an excellent guide to the reasoning and
procedures and practices that will help ensure that a device is used data underlying its provision.
safely and effectively and that its performance will be maintained. In summary, a standard or recommended practice is truly
Although a device standard is primarily directed to the useful only when it is used in conjunction with other sources of
manufacturer, it may also be of value to the potential purchaser or information and policy guidance and in the context of professional
user of the device as a frame of reference for device evaluation. experience and judgment.
Similarly, even though a recommended practice is usually oriented
towards healthcare professionals, it may be useful to the INTERPRETATIONS OF AAMI STANDARDS
manufacturer in better understanding the environment in which a AND RECOMMENDED PRACTICES
medical device will be used. Also, some recommended practices,
while not addressing device performance criteria, provide Requests for interpretations of AAMI standards and recommended
guidelines to industrial personnel on such subjects as sterilization practices must be made in writing, to the AAMI Vice President,
processing, methods of collecting data to establish safety and Standards Policy and Programs. An official interpretation must be
efficacy, human engineering, and other processing or evaluation approved by letter ballot of the originating committee and
techniques; such guidelines may be useful to health care subsequently reviewed and approved by the AAMI Standards
professionals in understanding industrial practices. Board. The interpretation will become official and representation of
In determining whether an AAMI standard or recommended the Association only upon exhaustion of any appeals and upon
practice is relevant to the specific needs of a potential user of the publication of notice of interpretation in the "Standards Monitor"
document, several important concepts must be recognized: section of the AAMI News. The Association for the Advancement
All AAMI standards and recommended practices are voluntary of Medical Instrumentation disclaims responsibility for any
(unless, of course, they are adopted by government regulatory or characterization or explanation of a standard or recommended
procurement authorities). The application of a standard or practice which has not been developed and communicated in
recommended practice is solely within the discretion and accordance with this procedure and which is not published, by
professional judgment of the user of the document. appropriate notice, as an official interpretation in the AAMI News.
Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

American National Standard ANSI/AAMI EC53:2013


(Revision of ANSI/AAMI EC53:1995 and ANSI/AAMI EC53A:1998)

ECG TRUNK CABLES and PATIENT LEADWIRES

Developed by
Association for the Advancement of Medical Instrumentation

Approved 19 November 2013 by


American National Standards Institute, Inc.

Abstract: The objective of this standard is to allow ECG TRUNK CABLES and PATIENT LEADWIRES to be
interchanged between ECG DEVICES with isolated PATIENT connections by establishing a common
interface between the TRUNK CABLE and the PATIENT LEADWIRE connectors. Performance and safety
criteria for TRUNK CABLES and PATIENT LEADWIRES used with isolated PATIENT connectors are also
specified. This standard’s original scope related to TRUNK CABLES and PATIENT LEADWIRES used
with cardiac monitors. The scope was extended to include PATIENT LEADWIRES used with other
ECG DEVICES including diagnostic electrocardiographs, ambulatory ECG (Holter) recorders/event
recorders and ECG telemetry.

Keywords: electrocardiographic monitoring; cardiac monitoring; cables; patient leadwires


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

AAMI Standard

This Association for the Advancement of Medical Instrumentation (AAMI) standard implies a consensus of those
substantially concerned with its scope and provisions. The existence of an AAMI standard does not in any respect
preclude anyone, whether they have approved the standard or not, from manufacturing, marketing, purchasing, or
using products, processes, or procedures not conforming to the standard. AAMI standards are subject to periodic
review, and users are cautioned to obtain the latest editions.

CAUTION NOTICE: This AAMI standard may be revised or withdrawn at any time. AAMI procedures require that
action be taken to reaffirm, revise, or withdraw this standard no later than five years from the date of publication.
Interested parties may obtain current information on all AAMI standards by calling or writing AAMI, or by visiting the
AAMI website at www.aami.org.

All AAMI standards, recommended practices, technical information reports, and other types of technical documents
developed by AAMI are voluntary, and their application is solely within the discretion and professional judgment of
the user of the document. Occasionally, voluntary technical documents are adopted by government regulatory
agencies or procurement authorities, in which case the adopting agency is responsible for enforcement of its rules
and regulations.

Published by

Association for the Advancement of Medical Instrumentation


4301 N Fairfax Drive, Suite 301
Arlington, VA 22203-1633
www.aami.org

© 2014 by the Association for the Advancement of Medical Instrumentation

All Rights Reserved

This publication is subject to copyright claims of ISO, ANSI, and AAMI. No part of this publication may be reproduced
or distributed in any form, including an electronic retrieval system, without the prior written permission of AAMI. All
requests pertaining to this document should be submitted to AAMI. It is illegal under federal law (17 U.S.C. § 101, et
seq.) to make copies of all or any part of this document (whether internally or externally) without the prior written
permission of the Association for the Advancement of Medical Instrumentation. Violators risk legal action, including
civil and criminal penalties, and damages of $100,000 per offense. For permission regarding the use of all or any
part of this document, complete the reprint request form at www.aami.org or contact AAMI, 4301 N. Fairfax Drive,
Suite 301, Arlington, VA 22203-1633. Phone: +1-703-525-4890; Fax: +1-703-525-1067.

Printed in the United States of America

ISBN 1–57020–510–8
Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Contents

Page
Committee representation.............................................................................................................................................. v
Foreword ....................................................................................................................................................................... vi
1 Scope .......................................................................................................................................................................... 1
2 Normative references .................................................................................................................................................. 1
3 Definitions ................................................................................................................................................................... 1
4 Test methods .............................................................................................................................................................. 2
5 Requirements.............................................................................................................................................................. 3
5.1 *Labeling requirements ..................................................................................................................................... 3
5.1.1 Package labeling .................................................................................................................................... 3
5.1.2 CABLE YOKE labeling ................................................................................................................................ 3
5.1.3 PATIENT LEADWIRE termination labeling.................................................................................................... 3
5.1.4 *Labeling to identify the location of current-limiting devices ................................................................... 3
5.1.5 Optional labeling to identify accessories as not being DEFIBRILLATION-PROOF.......................................... 3
5.2 Construction requirements ................................................................................................................................ 3
5.2.1 *PATIENT LEADWIRE to TRUNK CABLE interconnection ................................................................................ 3
5.2.2 *Current-limiting devices ......................................................................................................................... 5
5.3 Performance requirements — TRUNK CABLES and PATIENT LEADWIRES ............................................................... 5
5.3.1 *Non-DEFIBRILLATION-PROOF TRUNK CABLES and PATIENT LEADWIRES .......................................................... 5
5.3.2 *Cable and leadwire noise ...................................................................................................................... 5
5.3.3 *Flex life of TRUNK CABLE and PATIENT LEADWIRE FLEX RELIEF .................................................................... 6
5.3.4 *Tensile strength of cable connections ................................................................................................... 7
5.3.5 *Number of connector mating/unmating cycles ...................................................................................... 8
5.3.6 *Connector retention force ...................................................................................................................... 8
5.3.7 *Contact resistance ................................................................................................................................ 8
5.3.8 *Leadwire resistance .............................................................................................................................. 8
5.3.9 *Dielectric withstand voltage .................................................................................................................. 9
Annex A ....................................................................................................................................................................... 11
Rationale for the development and provisions of this standard ............................................................................ 11
A.1 Introduction ............................................................................................................................................. 11
A.2 Rationale for specific provisions of this standard .................................................................................... 11

Tables
Table 1—Flex life of TRUNK CABLE and PATIENT LEADWIRE FLEX RELIEF ............................................................................. 6
Table 2—Tensile strength of cable connections in N..................................................................................................... 7
Table 3—Number of connector mating/unmating cycles ............................................................................................... 8
Table 4—Leadwire resistance (Ω) ................................................................................................................................. 9

Figures
Figure 1 - Non-shielded PATIENT LEADWIRE to CABLE YOKE connection ............................................................................ 4
Figure 2 - Shielded PATIENT LEADWIRE to CABLE YOKE connection (equipment side)........................................................ 4
Figure 3 - Shielded PATIENT LEADWIRE to CABLE YOKE connection (PATIENT side)............................................................. 4
Figure 4 – Test setup for cable noise measurement...................................................................................................... 6
Figure 5 – Flex life test setup ......................................................................................................................................... 7
Figure 6 – Wire-to-wire (each pair) dielectric withstand test ........................................................................................ 10
Figure 7 – Wire-to-shield dielectric withstand test ....................................................................................................... 10
Figure 8 – Internal-to-external-conductor dielectric withstand test circuit .................................................................... 10
Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Glossary of equivalent standards

International Standards adopted in the United States may include normative references to other International
Standards. AAMI maintains a current list of each International Standard that has been adopted by AAMI (and ANSI).
Available on the AAMI website at the address below, this list gives the corresponding U.S. designation and level of
equivalency to the International Standard.

www.aami.org/standards/glossary.pdf

iv © 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Committee representation

Association for the Advancement of Medical Instrumentation


Electrocardiograph (ECG) Committee

This standard was developed by the ECG/Electrocardiograph Committee of the Association for the Advancement of
Medical Instrumentation. Committee approval of the standard does not necessarily imply that all committee members
voted for its approval.

At the time this document was published, the AAMI Electrocardiograph Committee had the following members:

Cochairs: Richard A. Sunderland


Ahmet Turkmen
Brian J. Young
Members: Robert William Bain, CBET, Baltimore Medical Engineers & Technician Society
Scott Coggins, Covidien
Prakash C. Deedwania, MD, The VA Medical Center
Laura Dhatt, Physio-Control
Sreeram Dhurjaty, Dhurjaty Electronics Consulting LLC
Richard Diefes, ECRI Institute
Greg Downs, Spacelabs Medical Inc.
Arthur R. Eddy, Jr.
James J. Greco, Medapprove Inc.
Richard Gregg, Philips Electronics North America
Janice M. Jenkins, PhD, University of Michigan College of Engineering
Dongping Lin, PhD
Walter G. Lloyd, Childrens Hospital Boston
Peter W. Macfarlane, Royal Infirmary
Luis A. Melendez, Partners Healthcare
George Moody, Massachusetts Institute of Technology
Cadathur Rajagopalan, PhD SMIEEE, Mindray DS USA Inc.
Linda Ricci, FDA/CDRH
Johann-Jakob Schmid, Schiller AG
Jonathan Steinberg, MD, St Lukes Roosevelt Hospital Center
Richard A. Sunderland, Welch Allyn, Inc.
Ahmet Turkmen, BS MS PhD, University of Wisconsin-Stout
J.S. Wiley, Draeger Medical Systems Inc.
Jeffrey Wiser, 3M Healthcare
Ted Yantsides, Conmed Corp
Brian J. Young, GE Healthcare
Alternates: Mark J. Callahan, Covidien
Kejian Chen, 3M Healthcare
Yu Chen, PhD, Draeger Medical Systems Inc.
Steve Duke, Physio-Control
Charles S. Ho, PhD, FDA/CDRH
Shen Luo, PhD, Mindray Medical
Serkan Sezer, Schiller AG
Donald Stewart, Spacelabs Medical Inc.
John J. Wang, Philips Electronics North America
Yinqi Zhang, Spacelabs Medical Inc.

NOTE—Participation by federal agency representatives in the development of this standard does not constitute
endorsement by the federal government or any of its agencies.

© 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013 v


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Foreword

This standard was developed by the AAMI Electrocardiograph (ECG) Committee. The objective of this standard is to
provide minimum labelling, performance, and safety requirements that will help ensure a reasonable level of clinical
efficacy and PATIENT safety in the use of ECG TRUNK CABLES and PATIENT LEADWIRES on equipment with isolated
PATIENT connections in common ECG applications. The goal of this standard is to promote PATIENT safety by helping
to prevent inadvertent mating of PATIENT LEADWIRES with power mains connectors, and by allowing more rapid
transfer of PATIENTs who require continuous monitoring under emergency conditions.
This standard’s original scope related to TRUNK CABLES and PATIENT LEADWIRES used with cardiac monitors. The scope
was extended to include PATIENT LEADWIRES used with other ECG DEVICES including diagnostic electrocardiographs,
ambulatory ECG (Holter) recorders/event recorders and ECG telemetry.

This is a revision of ANSI/AAMI EC53:1995, ECG cables and leadwires and the amendment, ANSI/AAMI
EC53A:1998. In 1998 and in 2001, two mistakes related to defibrillation withstand (section 5.5.3 and figure 7) were
fixed by issuing an errata.

Continuous monitoring of a PATIENT’s cardiac activity is routine. Monitored PATIENTS are sometimes transferred to
other locations, which use different ECG DEVICES. The committee, therefore, felt it was desirable to require all
TRUNK CABLES and PATIENT LEADWIRES to share a common interface at the CABLE YOKE – PATIENT LEADWIRE CONNECTOR.
The committee also felt that a standardized PATIENT LEADWIRE system would help to reduce confusion, errors, setup
time, and training time. Therefore, in addition to specifying safety and performance criteria, this standard also
establishes a TRUNK CABLE to PATIENT LEADWIRE interface that supports interchangeability. Although using
TRUNK CABLES and PATIENT LEADWIRES from different MANUFACTURERS might affect the quality of an acquired ECG
signal, it is unlikely to otherwise compromise performance, which for cardiac monitors includes detection of life-
threatening events.

As used within the context of this document, “shall” indicates requirements strictly to be followed to conform to the
standard. “Should” indicates that among several possibilities, one is recommended as particularly suitable, without
mentioning or excluding others, or that a certain course of action is preferred but not necessarily required, or that (in
the negative form) a certain possibility or course of action should be avoided but is not prohibited. “May” is used to
indicate that a course of action is permissible within the limits of the recommended practice. “Can” is used as a
statement of possibility and capability. Finally, “must” is used only to describe “unavoidable” situations, including
those mandated by government regulation.The concepts incorporated in this standard should not be considered
inflexible or static. This standard, like any other, must be reviewed and updated periodically to assimilate progressive
technological developments. To remain relevant, it must be modified as technological advances are made and as
new data come to light.

This standard reflects the conscientious efforts of concerned health care professionals and MANUFACTURERS to
develop a standard for those performance levels that can be reasonably achieved at this time.

Suggestions for improving this standard are invited. Comments and suggested revisions should be sent to AAMI,
4301 N. Fairfax Drive, Suite 301, Arlington, VA 22203-1633.

NOTE—This foreword does not contain provisions of the American National Standard, ECG trunk cables and patient
leadwires (ANSI/AAMI EC53:2013), but it does provide important information about the development and intended use
of the document.

vi © 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

American National Standard ANSI/AAMI EC53:2013

ECG TRUNK CABLES and PATIENT LEADWIRES


1 Scope
This standard covers TRUNK CABLES and PATIENT LEADWIRES used to acquire surface electrocardiographic (ECG)
monitoring signals for cardiac monitors/telemetry transmitters (ANSI/AAMI/IEC 60601-2-27), diagnostic
electrocardiographs (ANSI/AAMI/IEC 60601-2-25) and ambulatory ECG recorders/event recorders (ANSI/AAMI/IEC
60601-2-47). In the broadest sense, this standard applies to any ECG DEVICE that uses PATIENT LEADWIRES and
possibly ECG TRUNK CABLES to acquire surface electrocardiographic signals.
This standard covers both disposable and reusable PATIENT LEADWIRES as well as TRUNK CABLES (generally
considered to be reusable). The concept of “disposable” applies to products intended to be used for a single PATIENT
stay, typically up to 7 days (SINGLE-PATIENT USE) as well as to products intended to be used one time, typically a few
hours (SINGLE USE). For terminological consistency, the concept of “reusable” is referred to as MULTI-PATIENT USE.
Some PATIENT LEADWIRES and TRUNK CABLES used for recording applications might not be intended to be
DEFIBRILLATOR-PROOF, but otherwise must meet all the requirements in this standard. Such PATIENT LEADWIRES and
TRUNK CABLES are included within the scope of this standard but must be clearly labelled as not being DEFIBRILLATOR-
PROOF.

The tests in this standard allow MANUFACTURERS to verify compliance of their products to this standard’s
specifications. These tests are not intended to be performed by OPERATORS or RESPONSIBLE ORGANIZATIONS.
ECG TRUNK CABLES and PATIENT LEADWIRES used in applications that require special characteristics, such as a
magnetic resonance imaging (MRI) suite, are excluded from this standard. TRUNK CABLES and PATIENT LEADWIRES
covered by this standard might support physiologic functions in addition to ECG monitoring, such as respiration
monitoring by impedance pneumography. However, the TRUNK CABLE and PATIENT LEADWIRES must meet all of the
requirements of this standard, unless a requirement is specifically excluded.
2 Normative references

The following normative documents contain provisions that, through reference in this text, constitute provisions of
this standard. For dated references, subsequent amendments to, or revisions of, any of these publications do not
apply. However, parties to agreements based on this standard are encouraged to investigate the possibility of
applying the most recent editions of the normative documents indicated below. For undated references, the latest
edition of the normative document referred to applies. AAMI maintains a register of currently valid AAMI technical
documents.
2.1 ANSI/AAMI ES60601-1, Medical electrical equipment—Part 1: General requirements for basic safety and
essential performance. AAMI. 2005.

2.2 ANSI/AAMI/IEC 60601-2-27, Medical electrical equipment — Part 2-27: Particular requirements for basic safety
and essential performance of electrocardiographic monitoring equipment. AAMI. 2011.

2.3 IEC 60601-2-49, Medical electrical equipment — Part 2-49: Particular requirements for the safety of multifunction
patient monitoring equipment. International Electrotechnical Commission. 2011.
3 Definitions

For the purposes of this standard, the following definitions apply.

3.1 CABLE ASSEMBLY: The assembly of individual wires and/or wires bundled into one jacket that connects the PATIENT
to an ECG DEVICE. Typically, this consists of a TRUNK CABLE and a set of PATIENT LEADWIRES. PATIENT LEADWIRES may
be separable from the TRUNK CABLE or may be integrated as a single assembly into the TRUNK CABLE (a one piece
TRUNK CABLE and PATIENT LEADWIRE assembly).

3.2 CABLE YOKE: The end of the TRUNK CABLE into which PATIENT LEADWIRES plug.

© 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013 1


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

NOTE—PATIENT-worn ECG DEVICES such as ambulatory ECG recorders and ECG telemetry transmitters usually incorporate a built-
in CABLE YOKE. In this case, the PATIENT LEADWIRES plug directly into the ECG DEVICE rather than by mating a TRUNK CABLE to the
ECG DEVICE’S INSTRUMENT CONNECTOR RECEPTACLE.

3.3 ECG DEVICE: ME EQUIPMENT that is used to acquire and to display/record surface electrocardiographic signals.

Examples: Cardiac monitor


Ambulatory ECG (Holter) recorder
Cardiac event recorder
ECG telemetry transmitter
Diagnostic electrocardiograph

3.4 FLEX RELIEF: The portion of the TRUNK CABLE or PATIENT LEADWIRE assembly at the junction of the connector to the
assembly that prevents flexure of the assembly from damaging the assembly or the connector.
NOTE—FLEX RELIEF may be part of the strain relief. Strain relief is the part of a connector/cable that supports the connection and
prevents forces that are applied to the cable from being transferred to the contacts and wires within the connector housing.

3.5 INSTRUMENT CONNECTOR: The connector on a TRUNK CABLE that mates to an INSTRUMENT CONNECTOR RECEPTACLE
on an ECG DEVICE.
3.6 INSTRUMENT CONNECTOR RECEPTACLE: The connector on an ECG DEVICE that mates to an INSTRUMENT CONNECTOR.

3.7 MULTI-PATIENT USE: Products specifically not labeled for SINGLE USE or SINGLE-PATIENT USE.

3.8 PATIENT-END TERMINATION: The connector on a PATIENT LEADWIRE that connects to an electrode on the PATIENT.

3.9 PATIENT LEADWIRE CONNECTOR: The connector on a PATIENT LEADWIRE that allows the leadwire to mate to a
CABLE YOKE (or directly to the ECG DEVICE having an integrated CABLE YOKE).

3.10 PATIENT LEADWIRE: A cable that ends with a PATIENT-END TERMINATION and a PATIENT LEADWIRE CONNECTOR.

3.11 SINGLE USE: Products specifically labeled as being for a single use on a single PATIENT.

3.12 SINGLE-PATIENT USE: Products specifically labeled as being for longer term use on a single PATIENT only.

3.13 TRUNK CABLE: The portion of the CABLE ASSEMBLY within which all wires are bundled together into one jacket or
are in some way bound permanently together. Typically, this has a CABLE YOKE at one end with connectors for
PATIENT LEADWIRES, and an INSTRUMENT CONNECTOR at the other end for connecting to the ECG DEVICE.

4 Test methods
NOTE—The test methods specified in Clause 5 are considered to be referee test methods. Other test methods may be used to
prove designs, provided the MANUFACTURER establishes equivalence with the referee tests a priori to ensure comparability of test
results. These test methods do not identify measures that should be taken in order to safely perform each test.

Instrumentation

The following test instruments are required.

—An alternating current (a.c.) current meter capable of measuring 10 microamperes (µA) with ± 1% accuracy;

—An oscilloscope with a differential amplifier having a 3 decibel (dB) bandwidth between 0.1 and 100 hertz (Hz) (6
dB per octave roll off) capable of resolving a 10 microvolt (µV) signal. The input impedance of the differential channel
shall be at least 1 megohm (MΩ). The 3 dB mid-band amplitude accuracy shall be ± 5%;

—A cable flexing apparatus capable of securely clamping the cable to the flexing head and rotating through an arc of
o
± 90 ;
—A pull test apparatus capable of applying an axial force along a cable or connector of at least 135 newton (N);

—A volt/ohm meter with the following minimum specifications:


—d.c. voltage range 10 V (±2%);
—a.c. voltage range (rms) 10 V (±2%);
—d.c. resistance 0.1 Ω to 1 MΩ (±2%).

2 © 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Test circuits

Unless otherwise specified, use resistors with a maximum ± 5% tolerance for frequencies up to 1 megahertz (MHz),
non-polarized capacitors of suitable rating, with a maximum tolerance of ± 10% and inductors with a maximum
tolerance of ± 5%.

Test signals

Unless otherwise specified, input test signals shall be accurate to ± 1% (d.c. voltages) or ± 2 % (a.c. voltages).
5 Requirements
5.1 *Labeling requirements

5.1.1 Package labeling

Each unit package supplied to the clinical OPERATOR shall have the following as part of a package insert or label:

This cable [or leadwire or set of leadwires] meets the requirements of the standard, ECG TRUNK CABLES and
PATIENT LEADWIRES (ANSI/AAMI EC53). This cable [or leadwire or set of leadwires] is intended for SINGLE [or
SINGLE-PATIENT or MULTI-PATIENT] USE.

For TRUNK CABLES and PATIENT LEADWIRES that are not DEFIBRILLATION-PROOF, each unit package supplied to the clinical
OPERATOR shall have the following as part of a package insert or label:

This cable [or leadwire or set of leadwires] is not defibrillator-proof and should not be used in areas where a
PATIENT might be defibrillated.

5.1.2 CABLE YOKE labeling

The CABLE YOKE shall have each PATIENT LEADWIRE position permanently marked (e.g., molded or engraved) with the
appropriate electrode designation and color code per ANSI/AAMI/IEC 60601-2-27 or ANSI/AAMI/IEC 60601-2-47
Labeling of connection polarity is not required, but is customary for products such as ambulatory ECG recorders.

5.1.3 PATIENT LEADWIRE termination labeling


If the entire wire is not appropriately color-coded, the terminations on both ends of the PATIENT LEADWIRE shall be
color-coded per ANSI/AAMI/IEC 60601-2-27or ANSI/AAMI/IEC 60601-2-47. If the PATIENT LEADWIRE CONNECTOR is not
separable from the CABLE YOKE, then only the PATIENT-END TERMINATION shall be color-coded.
The PATIENT-END TERMINATION should include nomenclature that reinforces the color designation (i.e., R, L).

5.1.4 *Labeling to identify the location of current-limiting devices

CABLE ASSEMBLIES or ECG DEVICES that incorporate current-limiting devices (e.g., for defibrillation protection) shall be
permanently marked to indicate that current-limiting devices are present. Also see subclause 5.2.2.

When current-limiting devices are intentionally positioned as close to the patient as possible, the unit package
supplied to the clinical OPERATOR that is associated with the component that contains the current-limiting devices
shall be permanently marked to indicate that current-limiting devices are present.

5.1.5 Optional labeling to identify accessories as not being DEFIBRILLATION-PROOF

TRUNK CABLES and PATIENT LEADWIRES that are not DEFIBRILLATION-PROOF may be permanently marked to indicate that
they are not DEFIBRILLATION-PROOF (e.g., IEC 60601-1’s symbol for DEFIBRILLATION-PROOF covered by a negation cross).
Compliance with all of clause 5.1 is checked by inspection.

5.2 Construction requirements

5.2.1 *PATIENT LEADWIRE to TRUNK CABLE interconnection

Connector stems (see the 4.5 mm dimension in Figures 1 and 2) may include features intended to ensure proper
retention forces for PATIENT LEADWIRE CONNECTORS in the CABLE YOKE provided that interchangeability is not affected.
For CABLE ASSEMBLIES in which non-shielded PATIENT LEADWIRES are separable from the TRUNK CABLE as single
leadwires, the connection between PATIENT LEADWIRE CONNECTORS and the CABLE YOKE should meet the tolerances of
Figure 1 and the CONDUCTIVE CONNECTION requirements of IEC 60601-1 and IEC 60601-2-49.

© 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013 3


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Figure 1 — Non-shielded PATIENT LEADWIRE to CABLE YOKE connection

For CABLE ASSEMBLIES in which PATIENT LEADWIRES are separable from the TRUNK CABLE as shielded leadwires, the
connection between PATIENT LEADWIRE CONNECTORS and CABLE YOKE should meet the requirements of Figures 2 and 3.

Figure 2 — Shielded PATIENT LEADWIRE to CABLE YOKE connection (equipment side)

Figure 3 — Shielded PATIENT LEADWIRE to CABLE YOKE connection (PATIENT side)


NOTE—Figures 2 and 3 illustrate a five-wire CABLE YOKE. Fewer or more PATIENT LEADWIRES may be used as necessary and
PATIENT LEADWIRES may be stacked in multiple rows.

4 © 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

For the shielded interconnection shown in Figure 3:

a) The following lead order should be used in a shielded CABLE YOKE when all PATIENT LEADWIRES are in a single
row, starting with the appropriate PATIENT LEADWIRE at the end marked "*" (as shown in Figure 3, from left to
right): for IEC Code 1 – N (or RF), L, F, R, C1, C2, …; for IEC Code 2 – RL, LA, LL, RA, V1, V2, …). If a
TRUNK CABLE includes PATIENT LEADWIRES other than these, the additional PATIENT LEADWIRES should not
interfere with the specified order, but may otherwise be positioned as desired, provided that the width
spacing is ≥ 0.9 mm and individual shielded patient leadwires can plug in.

b) Other lead orders are permissible if appropriate based upon the INTENDED USE.
Compliance is checked by inspection.

5.2.2 *Current-limiting devices

PATIENT LEADWIRES for all ECG DEVICES, except those intended for use in MRI environments, shall not contain current-
limiting devices unless at least one of the following conditions is met:

1) The PATIENT LEADWIRES are integrated into the TRUNK CABLE and cannot be replaced by the clinical
OPERATOR.

2) The interconnection between the PATIENT LEADWIRE CONNECTOR and the CABLE YOKE/ECG DEVICE is a
proprietary or patented design not used by other MANUFACTURERS.
3) The current-limiting device is located entirely within the PATIENT-END TERMINATION of the PATIENT LEADWIRE.

For 2) and 3) above, the CLINICAL OPERATOR shall be strongly cautioned by prominent labeling and in the instructions
for use to only use the appropriate MANUFACTURER’S PATIENT LEADWIRES with this TRUNK CABLE/ECG DEVICE (using other
PATIENT LEADWIRES could compromise PATIENT safety).

Compliance is checked by inspection.

5.3 Performance requirements — TRUNK CABLES and PATIENT LEADWIRES

All requirements of this standard shall be met while tests are performed in any sequence on the same TRUNK CABLES
and PATIENT LEADWIRES.
5.3.1 *Non-DEFIBRILLATION-PROOF TRUNK CABLES and PATIENT LEADWIRES
TRUNK CABLES and PATIENT LEADWIRES that are explicitly labelled as not being DEFIBRILLATION-PROOF shall not be
treated as DEFIBRILLATION-PROOF APPLIED PARTS nor shall they be subjected to IEC 60601-1’s requirements for
DEFIBRILLATION-PROOF APPLIED PARTS. Subclause 5.1.5 describes suitable labelling requirements.

5.3.2 *Cable and leadwire noise

A 150 centimeter (cm) section of cable material shall not produce noise > 50 µV peak-to-valley.
Compliance is checked using the test setup of Figure 4 with an oscilloscope differential amplifier that has a 3 dB
bandwidth between 0.1 and 100 Hz (6 dB per octave roll off).

Mount a 152 cm length of cable between clamps positioned 122 cm apart. At one end of the cable, connect three
signal wires together through 4.3 (kilohm) kΩ series resistors.
NOTE—Do not connect any other signal wires (such as the extra pair from a five-wire cable) at this end.

Connect two of the signal wires at the other end of the cable to the inputs of the oscilloscope differential amplifier.

a) For wires that are not individually shielded, connect the third signal wire and the overall cable shield to the
oscilloscope ground.
b) For individually shielded wires connect the third signal wire and each wire shield to the oscilloscope ground.

Determine the mass of 0.3 meters (m) of the cable.

Apply a weight of 40 times the identified mass at the center of the cable. Position this weight between the cable
clamps and drop the weight freely.

Verify that the maximum peak-to-valley noise measured on the oscilloscope is ≤ 50 µV.

© 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013 5


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Figure 4 — Test setup for cable noise measurement

5.3.3 *Flex life of TRUNK CABLE and PATIENT LEADWIRE FLEX RELIEF

TRUNK CABLE and PATIENT LEADWIRE interconnections shall withstand flexes of ± 90 degrees as specified in Table 1.
Table 1—Flex life of TRUNK CABLE and PATIENT LEADWIRE FLEX RELIEF

Test location SINGLE USE SINGLE-PATIENT USE MULTI-PATIENT USE


TRUNK CABLE (both ends) 5 10 1000
PATIENT LEADWIRE (both ends) 30 150 500

Compliance is checked using the test setup of Figure 5.

Suspend 0.6 m of cable from the flexing fixture.

Attach a weight to the free end of the cable using a clamp. Ensure that the total mass of the weight and clamp is
0.23 kg (± 5%) if the cable diameter (d) including the jacket is ≤ 3.2 millimeters (mm). If the cable diameter including
the jacket is > 3.2 mm, use the following formula to calculate the total mass of the weight and clamp to use:
2
Total mass = 0.0072 * π * d (± 5%)
Rotate the flexing fixture through the number of flexes specified in Table 1. One flex is defined as rotation from 0 to
90 degrees, back to –90 degrees, and back to 0 degrees.

Failure is defined as any of the following:


a) increase in series resistance of any conductor of more than 50% from original specification;
b) a short between any two conductors;
c) a short between any conductor and shield; or
d) rupture of the specimen jacket (jacket rupture is permitted within 25 mm of the weight attachment position,
but a retest of another specimen is required).

6 © 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Figure 5 — Flex life test setup

5.3.4 *Tensile strength of cable connections

Cable connections shall withstand an axial pull according to the following table.

Table 2—Tensile strength of cable connections in N

Test location SINGLE USE SINGLE-PATIENT USE MULTI-PATIENT USE


TRUNK CABLE to INSTRUMENT CONNECTOR N/A N/A 65
TRUNK CABLE to CABLE YOKE connector N/A N/A 65
TRUNK CABLE material N/A N/A 90
PATIENT LEADWIRE terminations 25 25 50
PATIENT LEADWIRE material per lead 25 25 50

Compliance is checked by comparing the results of the following test to the values in Table 2.

Place one end of the cable sample in the wire holding fixture so that no wire is pinched.

Set the distance between the wire holding fixture and the pull tester to the minimum possible for the wire being
tested. Align the pull tester to exert force axially along the cable sample.

Record the force (N) required to cause the cable to fail between the cable attachment points (a failure at either cable
attachment point does not constitute failure). Cable failure is defined as any of the following:

a) ≥ 50% increase in resistance of any signal wire, including shields;


b) a short between any two conductors; or
c) a short from any signal wire to the shield.

© 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013 7


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

5.3.5 *Number of connector mating/unmating cycles

TRUNK CABLE and PATIENT LEADWIRE connectors shall be capable of being mated and unmated as specified in Table 3.
Following this test, the connections shall meet the requirements of5.3.6 and 5.3.7.

Table 3—Number of connector mating/unmating cycles

Test location Number of cycles


SINGLE USE SINGLE-PATIENT USE MULTI-PATIENT USE
TRUNK CABLE to INSTRUMENT CONNECTOR RECEPTACLE N/A 15 300
TRUNK CABLE YOKE to PATIENT LEADWIRE CONNECTOR N/A — TRUNK CABLE 30 500
5 — Leadwire
PATIENT-END TERMINATION 5 30 500

Compliance is checked by the following test:

Using either a manual or automated device, connect and disconnect each connector termination to/from its intended
mating connector, or equivalent, to ensure acceptable product life. Include the INSTRUMENT CONNECTOR, CABLE YOKE,
PATIENT LEADWIRE CONNECTOR and the PATIENT-END TERMINATION in the test. Complete the number of mating cycles
listed in Table 3 for each connector.

After subjecting the TRUNK CABLE and PATIENT LEADWIRE assemblies to this procedure, verify that these assemblies
comply with 5.3.6 and 5.3.7.

5.3.6 *Connector retention force

The minimum axial force required (per leadwire) to separate the PATIENT LEADWIRE CONNECTOR from the TRUNK CABLE
YOKE shall be ≥ 4.5 N.

The minimum axial force required to separate the INSTRUMENT CONNECTOR from its intended INSTRUMENT CONNECTOR
RECEPTACLEshall be > 30 N.

Compliance is checked by the following test:

Place the intended mating connector, or equivalent, in a holding fixture. Insert the INSTRUMENT CONNECTOR or PATIENT
LEADWIRE CONNECTOR into the mating connector and then attach the pull test device to the cable material
approximately 15 cm from the connector assembly.

Axially pull the cable until the connector assembly disconnects from the mating connector. Record the force in N
required to cause disconnects. Verify that the force meets or exceeds the minimum values specified above.

5.3.7 *Contact resistance

The d.c. resistance of any of the following connections shall not exceed 1.0 Ω when measured between:

a) each PATIENT LEADWIRE to CABLE YOKE connection;


b) each TRUNK CABLE to INSTRUMENT CONNECTOR RECEPTACLE; and
c) each electrode connection.

Compliance is checked by the following test:

Using a digital volt/ohm meter, measure the resistance of the connection after the specified number of mating and
unmating cycles (5.3.5). Before measuring the resistance of the actual connection, short the probes of the meter
together and record the value. Subtract this value from all contact resistance measurements.

5.3.8 *Leadwire resistance

The d.c. resistance of the PATIENT LEADWIRES shall be as specified in Table 4 for the appropriate leadwire material.

8 © 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Table 4—Leadwire resistance (Ω)

Leadwire Resistance (maximum Ω)


length Metallic Composite Organic
(cm) (e.g., copper) (e.g., tinsel) (e.g., carbon)
0–30 1 50 300
31–61 1 50 350
62–91 1 50 400
92–122 1 50 450
123–152 1 50 500
153–183 1 50 550
184–213 1 50 600
214–244 1 50 650

Compliance is checked by the following test:

Using a volt/ohm meter in resistance mode, record the value obtained with the meter’s test leads shorted together.
Connect the test leads across the leadwire material under test and subtract the previously recorded value from the
meter reading to obtain the final resistance. Compare the final resistance to the values for that material in Table 4.
5.3.9 *Dielectric withstand voltage

With the patient leadwires connected to the trunk cable and the trunk cable plugged into its intended receptacle or
equivalent, the assembly shall withstand without breakdown a 1-minute (min) (± 20%) application of 1,000 V root-
mean-square (rms) (± 10%) sine wave at 60 Hz. The voltage shall be applied to all combinations of any two wires,
including the shield.

The assembly shall also withstand, without breakdown, a 1-second (± 10%) application of 5,000 V d.c. (± 10%)
between all wires and shield connected together and any exposed conductive parts.
Breakdown is defined as current flow in excess of 9.25 milliamperes (mA) above the theoretical current flow, given
the voltage and frequency of the stimulus. The current shall be measured at the generator output.

Compliance can be determined by completing the following tests with the patient leadwires connected to the trunk
cable and the trunk cable plugged into its intended receptacle or equivalent.

a) Wire-to-wire test: Use the test circuit shown in Figure 6. Switch positions; wiring must be added to test a
five-wire cable and leadwire assembly.

b) Wire-to-shield test: Patient end connectors are shorted together. Use the test circuit shown in Figure 7.

c) Internal-to-external conductor test: This test is to be performed only if exposed metal components are
present (i.e., a metal nameplate, garment clip, or ungrounded metal connector shell). Use the test circuit
shown in Figure 8.

© 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013 9


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Figure 6 — Wire-to-wire (each pair) dielectric with-


stand test (See 5.3.9)

Figure 7 — Wire-to-shield dielectric withstand test


(See 5.3.9)

Figure 8 — Internal-to-external-conductor dielectric


withstand test circuit. (Connector 1 is made if conductive
materials will be exposed in usage; connection 2 is made if
shield is not available at the patient connection.)

10 © 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

Annex A
(informative)

Rationale for the development and provisions of this standard

A.1 Introduction

There are reports from the early 1990s of PATIENTs being burned or killed when someone plugged detachable
PATIENT LEADWIRES with male-pin connectors into a SUPPLY MAINS connector. These incidents involved non-shielded
PATIENT LEADWIRES or PATIENT LEADWIRES with pre-attached electrodes, but this is also a HAZARD for leadwire sets in
which several PATIENT LEADWIRES are molded together into one block.
These incidents led to mandates to prevent such tragic accidents by eliminating the use of PATIENT LEADWIRES with
exposed male pins. AAMI EC53 was originally written to accelerate the switchover to safer designs. First, EC53
standardized the design of the PATIENT LEADWIRE/TRUNK CABLE interface to support interchangeability in emergency
situations, regardless of brand. (Using one MANUFACTURER’S matched set of PATIENT LEADWIRES and TRUNK CABLES
usually yields optimal performance, though.) Second, EC53 set minimum requirements and defined standardized
test methods to allow OPERATORS (including technical OPERATORS such as biomedical engineers) to correctly select
TRUNK CABLES and PATIENT LEADWIRES.

This revision of EC53 clarifies the earlier revision’s requirements for disposable and reusable products in light of
international use of the terms SINGLE USE and SINGLE-PATIENT USE. Logically, EC53’s disposable requirements were
largely duplicated for SINGLE USE and SINGLE-PATIENT USE. For consistency, reusable became MULTI-PATIENT USE.
A.2 Rationale for specific provisions of this standard

A PATIENT’S stay varies from an hour to more than 6 days. Clinical input before 1995 led to EC53 being based on the
premise that a typical clinical OPERATOR expects MULTI-PATIENT USE TRUNK CABLES to last at least 6 months and MULTI-
PATIENT USE PATIENT LEADWIRES to last at least 3 months. The performance characteristics for MULTI-PATIENT USE
products support these average lives.

This was obviously an attempt to approximate average use lives for many different clinical situations (use lives for a
busy urban emergency room are very different than those for a rural general care ward). The goal is to allow factual
comparison of different products in the same way that automobile fuel economy is compared.

A.5.1 Labeling requirements

The labeling requirements for the CABLE YOKE-PATIENT LEADWIRE interconnection and the PATIENT-END TERMINATION-
ELECTRODE connection ensure that clinical OPERATORS can readily identify which connection is being affected while
connecting electrodes on the PATIENT’S body through PATIENT LEADWIRES to ME EQUIPMENT. Using redundant coding
techniques (e.g., pairing color coding with electrode designations [ANSI/AAMI/IEC 60601-2-27] or connection
polarity [ANSI/AAMI/IEC 60601-2-47]) greatly enhances usability.
To harmonize EC53 with IEC 60601-1, the earlier revision’s requirements for defibrillation withstand were eliminated
(Those requirements fundamentally related to how ECG DEVICES implement defibrillation protection.) Current-limiting
devices may be located between the PATIENT’S body and the ECG DEVICE. As in earlier revisions, this revision of EC53
prohibits placing current-limiting devices within PATIENT LEADWIRES; but now allows exceptions (see 5.2.2).
A.5.1.4 Labeling to identify the location of current-limiting devices

ECG DEVICES, TRUNK CABLES and PATIENT LEADWIRES must be able to withstand repeated defibrillation pulses without
damage. Applying a defibrillation pulse could degrade any current-limiting devices that are present and increases the
resistance of carbon leadwires. Repeatedly applying defibrillation pulses could, therefore, destroy the current-limiting
devices (or leadwires) which, in turn, could damage or destroy the ECG DEVICE or the CABLE ASSEMBLY.
Permanently marking the CABLE ASSEMBLY (typically, the TRUNK CABLE) or ECG DEVICE to indicate that current-limiting
devices are present mitigates a HAZARD related to the use of TRUNK CABLES that are plug-compatible with the
INSTRUMENT CONNECTOR RECEPTACLES of different ECG DEVICES. Simply put, PATIENTS are at RISK if TRUNK CABLES with
and without current-limiting devices are both available because the incorrect TRUNK CABLE might be used with an
ECG DEVICE.

The trend towards hardware miniaturization leads the working group to suspect that future designs might
intentionally position current-limiting devices at the PATIENT (possibly within the electrode itself or where the electrode

© 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013 11


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

connects to the PATIENT-END TERMINATION). In such cases, the unit package (rather than the electrode or
PATIENT LEADWIRE) needs to be labeled to indicate that it contains current-limiting devices.

A.5.2 Construction requirements

A.5.2.1 PATIENT LEADWIRE to TRUNK CABLE interconnection

This standard supports interchangeability between PATIENT LEADWIRES and TRUNK CABLEs. EC53:1995 used the
widely-accepted DIN 42802 connector (Connector, touch proof, for electromedical application, Deutsches Institut für
Normung [DIN], Germany, 1989) as the basis for its model for non-shielded systems. By incorporating the
requirements of DIN 42802 into the configuration of shielded systems, EC53:1995 also standardized the connection
for those systems while allowing non-shielded PATIENT LEADWIRES to connect to a shielded TRUNK CABLE. This latest
revision of EC53 recognizes that current designs largely incorporate touch-proof concepts in unique ways. This
clause’s earlier requirements were, therefore, softened.

Multiple PATIENT LEADWIRES might be configured together as a set. The order of the PATIENT LEADWIRES, the number of
leadwires, and their fixed center distances allow PATIENT LEADWIRES and TRUNK CABLEs from different MANUFACTURERS
to connect properly. The possibility of having TRUNK CABLES with different mating orders is discouraged, but could be
appropriate for specific INTENDED USES.
A.5.2.2 Current-limiting devices

Because a goal of this standard is for any MANUFACTURER’S PATIENT LEADWIRES to connect to any MANUFACTURER’S
CABLE YOKE, specifying that current-limiting devices used for DEFIBRILLATION PROTECTION not be within
PATIENT LEADWIRES helps toensure that PATIENT LEADWIRES work with any ECG DEVICE.

Current-limiting devices might be in either the TRUNK CABLE or the ECG DEVICE itself. Having one or two sets of current-
limiting devices limits how much defibrillation energy is shunted from the PATIENT. Defibrillating a PATIENT when no
current-limiting devices are present is hazardous (the ECG DEVICE absorbs too much energy, possibly compromising
resuscitation and potentially damaging or destroying the ECG DEVICE or CABLE ASSEMBLY).
NOTE—Having two sets of current-limiting devices present (e.g., one in the ECG DEVICE and another in the TRUNK CABLE) might
make an ECG DEVICE’S impedance pneumography-based respiration feature unusable.

The exceptions for integrated CABLE ASSEMBLIES and proprietary/patented connector designs support clinically proven
designs. The exception for intentionally positioning current-limiting devices at the PATIENT supports innovation.
A.5.3 Performance requirements — TRUNK CABLE and PATIENT LEADWIRES

A.5.3.1 Non-DEFIBRILLATION-PROOF TRUNK CABLES and PATIENT LEADWIRES

This subclause allows MANUFACTURERS to explicitly label TRUNK CABLES and PATIENT LEADWIRES appropriately for use in
areas where PATIENTS are unlikely to require defibrillation (e.g., a physician’s office when a PATIENT is connected to a
diagnostic electrocardiograph for a 12-lead ECG report or to an ambulatory ECG recorder for a long-term ECG
recording). Requiring DEFIBRILLATION-PROOF APPLIED PARTS for such applications is an unnecessary expense.
A.5.3.2 Cable and leadwire noise
CABLE ASSEMBLIES flex because of PATIENT movement as well as actions taken by attending OPERATORS. Minimizing
artifact induced by such movements prevents noise from appearing on ECG waveforms and minimizes inaccurate R-
wave detection by ECG devices in general and false alarms by cardiac monitors in particular.

A.5.3.3 Flex life of TRUNK CABLE and PATIENT LEADWIRE FLEX RELIEF
This performance level corresponds to about five hard flexes per day of the INSTRUMENT CONNECTOR to CABLE YOKE or
PATIENT LEADWIRE CONNECTOR to PATIENT-END TERMINATION over the expected life of the cable or leadwire.

A.5.3.4 Tensile strength of cable connections

The tensile test limits (5.3.4) and retention thresholds (5.3.6) are based upon the philosophy that when a PATIENT
moves a PATIENT LEADWIRE by tugging on it, the consequences should occur in the following order:

a) The PATIENT-END TERMINATION disconnects from the electrode.


b) The electrode comes off the PATIENT's skin.
c) THE PATIENT LEADWIRES disconnect from the TRUNK CABLE (at the PATIENT LEADWIRE CONNECTOR-CABLE YOKE
interconnection).
d) The TRUNK CABLE disconnects from the INSTRUMENT CONNECTOR RECEPTACLE;
e) The PATIENT-END TERMINATION is damaged.

12 © 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013


Draeger Medical Systems Inc
AAMI Invoice# 164734 / Order Date: 06/26/2014
Multi-user license through June 2024. Duplication and distribution beyond approved Draeger facilitities is prohibited
Copying, networking and distribution prohibited

f) The PATIENT LEADWIRE CONNECTOR (connection to TRUNK CABLE) is damaged.


g) Either end of the TRUNK CABLE is damaged.
h) The INSTRUMENT CONNECTOR RECEPTACLE is damaged.
i) The ECG DEVICE is ripped from the wall shelf.

A.5.3.5 Number of connector mating/unmating cycles

The performance levels specified in 5.3.5 correspond to two (TRUNK CABLE to INSTRUMENT CONNECTOR RECEPTACLE) or
five (PATIENT LEADWIRE CONNECTOR to CABLE YOKE and PATIENT-END TERMINATION to electrode) mating/unmating cycles
per day.

A.5.3.6 Connector retention force

This standard specifies the retention force for the connection between a PATIENT LEADWIRE and the TRUNK CABLE YOKE
and provides a recommendation for the retention force for the TRUNK CABLE to INSTRUMENT CONNECTOR RECEPTACLE
connection. The latter only applies to ME EQUIPMENT or ME SYSTEMS used with a TRUNK CABLE. Also see A.5.3.4.

A.5.3.7 Contact resistance

Contact resistance refers to components that come into contact with each other; the actual contact resistance should
be ≤ 0.1 Ω. Resistances at this level cannot be measured reliably using a two-input volt/ohm meter, so a four-input
meter is needed. The working group decided that a 1 Ω contact resistance presented no performance limitation,
given the high input impedance of modern ECG DEVICES, and would allow MANUFACTURERS to use simple
instrumentation for the test.

A.5.3.8 PATIENT LEADWIRE resistance


The working group decided to include maximum resistance numbers for alternate materials because leadwire
materials other than copper are used. Input from manufacturers (before 1995) was used to determine the resistance
limits for different leadwire lengths on the basis of the state of the art at that time. Although all leadwires could have
been required to have a resistance less than some maximum value, the working group felt that quantifying realistic
values for different leadwire lengths was important. Mixing leadwires of drastically different resistances could
adversely affect the noise and common-mode performance of ECG DEVICES.
NOTE—This rationale was updated since EC53 was originally published in 1995. The corresponding requirement was not changed.

© 2014 Association for the Advancement of Medical Instrumentation ■ ANSI/AAMI EC53:2013 13

You might also like