3.2.ARIETTA 850DI-Đặc Tính Kỹ Thuật - TA

The document outlines the specifications for the ARIETTA 850 DeepInsight Diagnostic Ultrasound System, detailing various scanning methods, operating modes, and imaging capabilities. It includes information on image display modes, beam forming technology, and advanced features such as real-time tissue elastography and panoramic view. Additionally, it specifies the system's dynamic range, processing channels, and Doppler modes for comprehensive diagnostic imaging.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
365 views35 pages

3.2.ARIETTA 850DI-Đặc Tính Kỹ Thuật - TA

The document outlines the specifications for the ARIETTA 850 DeepInsight Diagnostic Ultrasound System, detailing various scanning methods, operating modes, and imaging capabilities. It includes information on image display modes, beam forming technology, and advanced features such as real-time tissue elastography and panoramic view. Additionally, it specifies the system's dynamic range, processing channels, and Doppler modes for comprehensive diagnostic imaging.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 35

SPECIFICATIONS

for
Diagnostic Ultrasound System

ARIETTA 850 DeepInsight

MODEL: ALOKA ARIETTA 850

SPH-ARIETTA850DI-V50-U01
Printed in Japan, 2022-11

1
Basic Specifications ・ B(eFLOW) and Dual
Scanning Method ・ B(Color Flow) and D simultaneous real-time display
・ Electronic Convex (Triplex mode)
・ Electronic Linear ・ B(Power Doppler) and D simultaneous real-time display
・ Electronic Phased Array Sector (Triplex mode)
・ Electronic Radial ・ B(eFLOW) and D simultaneous real-time display(Triplex
mode)
Operating Modes ・ B and B(Color Flow) simultaneous real-time display
・ B-mode (Dual Flow)
・ BiPlane-mode ・ B and B(Power Doppler) simultaneous real-time display
・ M-mode (Dual CF)
・ D: Spectral Doppler mode (PW, CW, HPRF-PW) ・ B and B(eFLOW) simultaneous real-time display(Dual
・ Dual Gate Doppler mode Flow)
・ Color Flow mode ・ Dynamic Slow-motion Display(Real-time image/Slow-
・ Power Doppler mode (Directional Power Doppler) motion image, side by side display)
・ eFLOW mode (Directional eFLOW) ・ Real-time Biplane(Display real time image in 2 cross-
・ Cardiac 3D mode* sections in biplane probe.)
(3D Zoom, Active 3D, Wide Angle 3D) ・ Panoramic View
* Option: SOP-ARIETTA850-129, EU-9207, PEU-LISENDO880 ・ TDI(Tissue Doppler Imaging)
・ Intermittent trigger mode*2
Image Display Modes*1 ・ Monitor mode(Monitor image/CHI mode image, side by
・ B: gray-scale imaging side)*2
・ Dual B ・ RT-3D(4D) mode*3
・ Quad B ・ Detective Flow Imaging(DFI)*4
・M ・ Request function: In multi-mode display, it is possible to
・ B and M select one mode for full screen display.
・D ・ Real-time Tissue Elastography*5
・ B and D ・ 3D Zoom mode(Color)*6
・ B(Color Flow) ・ Active 3D mode(Color)*6
・ B(Power Doppler) ・ Wide Angle 3D mode(Color)*6
・ B(eFLOW) *1 Probe dependent.
・ Dual B(Color Flow) *2 Option: SOP-ARIETTA850-44
・ Quad B(Color Flow) *3 Option: EU-9184, SOP-ARIETTA850-4
・ Dual B(Power Doppler) *4 Option: SOP-ARIETTA850-105
・ Quad B(Power Doppler) *5 Option: SOP-ARIETTA850-43
・ Dual B(eFLOW) *6 Option: SOP-ARIETTA850-129, EU-9207, PEU-LISENDO880
・ Quad B(eFLOW)
・ M(Color Flow) Beam former
・ M(Power Doppler) ・ Transmission
・ M(eFLOW) CPWG(Compound Pulse Wave Generator)
・ B(Color Flow) and M(Color Flow) Programmable waveform transmission
・ B(Power Doppler) and M(Power Doppler) ・ Reception
・ B(eFLOW) and M(eFLOW) Multi processing high-speed digital beam former
・ B(Color Flow) and D 12-bit A/D converter(4096 gray levels)
・ B(Color Flow) and Dual A/D Sampling Frequency: 40 [MHz]
・ B(Power Doppler) and D Parallel processing: Up to 20 directions
・ B(Power Doppler) and Dual ・ Transmission/Reception method of Harmonics:
・ B(eFLOW) and D  FmT: Filter-method Tissue Harmonic Imaging

2
ARIETTA850DI-V50
THI mode image without reducing the frame rate. Minimum display of 5mm
 WbT: Wide-band Tissue Harmonic Imaging ・ Frequency selection:
THI mode image with higher resolution than FmT.  Fundamental:
 HdT: High Definition Tissue Harmonic Max. 5 frequencies
Imaging(HdTHI)  Tissue harmonic:
THI mode image with higher resolution than WbT. Max. 5 frequencies
Reception frequency: 3.2-22.0 MHz* ・ Depth range selections:
*: 22.0 MHz can be achieved under below conditions. 0.75/1.0/1.5/2.0/2.5/3.0/3.5/4.0/4.5/5.0, 1 cm
 Probe: SML44 intervals from 5.0 to 24 cm, and 2 cm intervals from
 Frequency: High 24 cm to 40 cm(probe dependent)
 Frequency Information: Receive ・ Longitudinal and lateral inversion
・ Tissue Adaptive Technology: ・ Rotation by 90 degrees(probe dependent)
 Sound speed adjustment: 26 steps ・ Gain
 Automated sound speed adjustment 0 to 80 dB, 1 dB step, adjustable in real-time and
・ Focusing frozen.
 Transmission  Echo enhancement: 8 steps(Including OFF)
Multi-stage transmission focus of up to 4 stages  Texture: 2 steps(Smooth/Sharp), not available in M-
out of 16 stages (probe dependent) mode.
 Reception: PixelFocus, eFocusing  Smooth/Enhance:
 -8 to -1: Smoothing level
System Dynamic Range  0: OFF
320 dB  1 to 8: Edge enhancement level
 Persistence
System Processing Channels 8 steps(Including OFF), 2 types(Manual/Auto)
7,072,000 Channels  Time Gain Control(TGC): Gain versus depth curve
control
Frame rate 8 slide controls, TGC curve memory function
Max. 882 frames/sec*  TGC Enhancement: ON/OFF possible, working with CF
* It is a measured value with the setting assuming the  Lateral Gain Control(LGC): Gain versus angle curve
maximum value. This value may be exceeded depending on the control.
conditions. 8 sectors(Sector probes only)
 Dynamic range
Volume rate(Mechanical 3D scanning) 40 to 90 dB, adjustable in real-time and frozen.
Max. Up to 41 volumes/sec*  Gray map: 10 types, adjustable in real-time and frozen.
* Depends on probes and various settings  Auto Gain Control(AGC): 8 steps(Including OFF)
 Gamma Curve
Imaging 4 types(Linear, Window, Parabolic, S-curve), control
B-mode points(Rejection, Center, Saturation), changeable in
・ Display Gray Scale: 256 levels real-time and frozen.
・ Scanning area:  Compound
25 to 100%, Continuously variable Possible by selected linear and convex sector probes.
・ Line Density: Up to 8 steps  Trapezoidal Scanning
・ HI Frame Rate Possible by selected linear probes
Improve frame rate by Multi Parallel Processing  Beam Steer
・ Zoom: Possible by selected linear probes
 HI zoom(real-time image):  DeepInsight
Minimum display of 5mm Clarifies the borders of structures and reduces speckle
 PAN zoom(real-time and frozen image): noise and artifacts.

3
ARIETTA850DI-V50
 eFocusing PLUS brightness and high brightness of grayscale.
Synthesizes multiple types of ultrasound beams to OFF+3 steps(Low, Mid, High)
reduce artifacts such as side lobes, multiple echoes, and ・ Free Angular M(FAM): Displays an M mode image on
speckle noise and to improve the spatial resolution and any line of the B mode image.
sensitivity of the image.
 HI REZ: Image filter that maintains the edges of Spectral Doppler Mode
structures while reducing speckle noise. ・ Display: Power spectrum
8steps, ON/OFF possible ・ Real-time Doppler Auto Trace
 HI REZ PLUS: ON/OFF possible ・ Doppler methods:
 Border Clear Filter(BCF): Image filter that enhance the  PW(Pulsed Wave) Doppler
edges of structures.  HPRF(High Pulse Repetition Frequency) PW Doppler
ON/OFF possible, only available when HI REZ is ON  Dual Gate Doppler
・ Acoustic Noise Reduction(ANR): It reduces artifacts.  CW(Continuous Wave) Doppler*
OFF+9 steps, 2 types(A-B)  TD(Tissue Doppler)-PW
・ Near-field Noise Reduction(NNR): It reduces artifacts ・ Reference frequencies (probe dependent):
and noises in the heart chamber or blood vessels. Maximum 3 frequencies
 OFF+3 steps(Low, Mid, High), 5 types(A-E)  PW: 1.5, 1.9, 2.0, 2.1, 2.5, 3.0, 3.3,
 Carving Imaging: OFF+9 types(combination of NNR 3.5, 3.8, 4.0, 4.4, 5.2, 6.0, 6.3 MHz
type(C, D, E) and levels(Low, Mid, High))  CW: 1.7, 1.8, 2.0, 2.1, 2.5, 3.0, 3.3, 3.8, 5.0 MHz
・ Low Echo Reduction: Suppresses the hypoechoic areas. ・ Pulse repetition frequency
0 to 70%, 2% step, adjustable in real-time and frozen  PW: 0.05 to 40 kHz
・ Grayscale Enhancement: Sets the balance between low ・ Analysis rate
brightness and high brightness of grayscale.  CW: 1.1 to 40 kHz
OFF+3 steps(Low, Mid, High) ・ Max. velocity range:
・ Auto-optimizer: Gain, TGC, LGC, sound velocity  PW/HPRF: ±1.26 cm/sec to ±802.08 cm/sec
・ Needle Emphasis: Improves the visibility of the puncture  CW: ±25.07 cm/sec to ±1600 cm/sec
needle echo. ・ Base line shift: Adjustable in real-time and frozen.
・ Wide Scanning: Display image with a wide field of view. ・ Steerable CW: Possible with selected probes.
・ Panoramic View: Display a panoramic image by moving ・ Beam Steer
the probe in a wide range. -30 to 30 degrees(probe dependent), 5 degrees step,
possible with linear probes.
M-mode ・ Spectrum inversion
・ Sweep method: Moving bar ・ Angle correction
・ Sweep speed 0 to 80 degrees, adjustable in real-time and frozen.
7 steps(40.0, 50.0, 66.7, 100, 133.3, 200, 300 Auto angle correction
mm/sec) ・ Sample Volume
・ Gain Size: 0.5 to 20 mm(0.5 mm,1.0 mm step)
B Gain±30dB, working with B Gain, adjustable in real- Display depth of the sample volume
time and frozen. ・ Wall filter: 12 steps, 1/16 of PRF is Max.
・ Dynamic Range ・ Gain
40 to 90 dB, 1 dB step, adjustable in real-time and 0 to 60 dB, ON/OFF possible, adjustable in real-time
frozen. and frozen.
・ Auto Gain Control(AGC): 8 steps(Including OFF) ・ Low Echo Reduction: Suppresses the hypoechoic areas.
・ Acoustic Noise Reduction(ANR): It reduces artifacts. 0 to 30%, 2% step
Working with B mode. ・ Grayscale Enhancement: Sets the balance between low
・ Low Echo Reduction: Suppresses the hypoechoic areas. brightness and high brightness of grayscale.
0 to 70%, 2% step, adjustable in real-time and frozen. OFF+3 levels(Low, Mid, High)
・ Grayscale Enhancement: Sets the balance between low ・ Echo Enh.: OFF+2 levels(Low,High)

4
ARIETTA850DI-V50
・ Dop. Gamma: Changes the Doppler waveform contrast,  Persistence Level: 8 steps(Including OFF)
and the level of gamma adjustment.  Wall Filter: 6 steps
8 levels, adjustable in real-time and frozen.  Wall Motion Reduction(WMR)
・ Auto Optimizer: Automatically adjusts image. OFF + 3 steps(Low, Middle, High), 2 methods
Gain, PRF, baseline, Angle Correction, position of  Wall Motion Reduction(WMR) PLUS: Remove body
sample volume movement due to clutter signals.
・ Audio output: 2 channels OFF + 3 steps(Low, Middle, High)
* Option: EU-9184  Packet Size: 3 levels(Large, Middle, Small)
 Baseline Shift: Up to double velocity(±64 steps)
Color Doppler Mode  Color Map: 15 kinds
・ Flow area: 25 to 100%, 5% step  TGC Enhancement
・ Beam Steer ON/OFF possible, working with B mode
-30 to 30 degrees(probe dependent), 5 degrees step,  Display superimposed color image
possible with linear probes  Accumu. Imaging: ON/OFF possible, in real-time.
・ Line Density  Capture Imaging: ON/OFF possible, in frozen.
8 levels, adjustable independently with B mode. Power Doppler(PD) mode
・ Gain: 128 levels Displays color on the image according to the intensity of
・ Texture: 2 levels(Smooth/Sharp) the color Doppler signal. It is sensitive to slow blood
・ Glossy Level: OFF+4 levels flows.
・ HI Frame Rate: Improve frame rate by Multi Parallel  Display patterns: PD, Directional PD
Processing.  Gradation: 256 levels
・ Auto Optimizer: Automatically adjusts image.  Color Map: 15 kinds
 Gain, position of ROI  Non-display of B/W image: Possible in ROI
 iVascular  Smoothing: 5 steps, adjustable in real-time and frozen.
Automatically identifies the long-axis blood vessels  Wall Filter: 6 steps
in an image, and sets the position of the flow area,  Wall Motion Reduction(WMR)
the position and size of the sample volume, the OFF + 3 steps(Low, Middle, High), 2 methods
blood flow direction, and the angle of the ultrasound  Wall Motion Reduction(WMR) PLUS: Remove body
beam(angle correction). movement due to clutter signals.
Color Flow(CF) mode OFF + 3 steps(Low, Middle, High)
Displays the orientation and velocity in color on the  Carving Imaging: Clarifies the boundaries of blood flows
tomographic image, based on the Doppler signal and changes the level for reducing speckle noise.
obtained from the blood flow. 5 steps
 Display patterns: High Resolution Power Doppler(eFLOW) mode
Velocity(derived from mean Doppler frequency shift), Displays in the Power Doppler mode in high-resolution.
Velocity + variance, Variance, Velocity + intensity,  Display patterns: eFLOW, Directional eFLOW
Velocity + variance + intensity  Pulse repetition frequency: 0.03 to 19.8 kHz
 Max. velocity range  Gradation: 256 levels(±127 levels for directional)
±0.63cm/sec to ±458.33cm/sec  Color Map: 15 kinds
 Reference frequency(probe dependent)  Smoothing: 5 steps, adjustable in real-time and frozen.
1.9, 2.0, 2.1, 2.5, 3.2, 3.3, 3.5, 4.0, 4.4, 5.2, 6.0, 6.3  Wall Filter: 6 steps
MHz  Wall Motion Reduction(WMR)
 Pulse repetition frequency: 0.03 to 19.8 kHz OFF + 3 steps(Low, Middle, High), 2 methods
 Gradation:  Wall Motion Reduction(WMR) PLUS: Remove body
 Velocity: ±127 levels(red and blue) movement due to clutter signals.
 Variance: 64 levels(green) OFF + 3 steps(Low, Middle, High)
 Invert Color Map  Carving Imaging: Clarifies the boundaries of blood flows
 Smoothing: 5 steps, adjustable in real-time and frozen. and changes the level for reducing speckle noise.

5
ARIETTA850DI-V50
5 steps and the points at 1cm on the right and left of the max
IMT, can be automatically detected and averaged.
Cine Memory
・ Cine search and loop display (in B mode): Stress Echo analysis*
ECG time phase display is possible It is an examination in which the heart is subjected to
・ Capacity stress through exercise or drugs, for example, in order to
 B mode: Max. 63,500 frames. check for abnormalities of myocardium.
 M/D mode: Max. Approx. 900 seconds. Stress Echo involves acquiring images before and after
the application of stress, displaying the images alongside
Physiological Signal Display each other, synchronizing the heartbeats and looping the
・ Displayed information: playback. This permits periodic evaluation of images.
ECG, PCG*1, Pulse*2, Breathing waveform Stress echo report can be created using the scoring
・ ECG synchronized display: Available for one phase function.
・ Detect regular pulse from arrhythmia(RRp/RRpp)  Image display modes in which image acquisition is
・ Automated jumping to most optimized R-R Phase possible: B, Each Flow modes
・ Reducing Filter for Tachy or Bradycardia  Methods for acquiring images
・ Display value for each R-R interval time  Temp acquisition: Acquire a preset heart rate. This
・ Display position: method allows you to acquire color images.
Continuously variable(both in B and M modes)  Conti acquisition: Continuous acquisition. This
・ Bar graph display for breathing waveform method cannot acquire color images.
・ 3 type electrocardiogram(I, II, III)  Protocol: Skip Stage/Skip View is possible during exam.
・ Automated detection end diastole and end systole  Exercise stress protocols:
phase.  Exercise Stress Echo
・ Automated split as end diastole(left) and end  Treadmill Exercise
systole(right)  Bicycle Exercise
*1 Option: MA-300  Pharmacological stress protocols:
*2 Option: TY-307A  DSE
 High-Dose DSE
Tissue Doppler Imaging(TDI)  Low-Dose DSE
・ Switching Color Doppler and Tissue Doppler during  Arbutamine
exam is possible.  Dipyridamole
 User's protocol:
Dual Gate Doppler  The user can make a protocol within 8 views X 12
This is a function which displays Doppler Spectrums of stages in 1 exam.
two different sample points simultaneously.  Full disclosure: Max. 2,481 seconds(Depends on
 Supported combinations are PW/PW, TDI/TDI, and probes and various settings)
PW/TDI*. 2,481 seconds can be achieved under below
 Automated sample gate setting for each gate. conditions.
*: It is possible to measure E/e’ immediately(approx. 5 seconds)  Probe: S121(Adult Heart)
by recognition of imaging view and placement of sample volumes  Protocol: Conti. only(1Stage/1View)
automatically in addition to detect stable heart beats with R-R  Frequency: 30Hz
Navigation.  Scan Area: Minimum
Automated Intima-Media Thickness(IMT)  Line Density: Minimum
Measurement  Automatic labeling: Possible(Auto ON/OFF is also
It is possible to automatically extract max, min, average possible)
and SD of IMT by simply setting ROI(region of interest)  Comparing images
on a long-axis view of the vessel.  Shuffle Stage: Displays the same view of each stage
 The thicknesses at 3 points, i.e., the point at max IMT, for comparison.

6
ARIETTA850DI-V50
 Shuffle View: Displays the different views of the
same stage for comparison.
 Compare: Lines up the same view of any stage
left/right for comparison.
 Loop speed: Adjustable
 Scoring: Possible
 Report
 Chart display: Display scoring results in a list.
 Schema(Shuffle View) display: Display scoring
results by a schema(list of the same Stage).
* Option: PEU-LISENDO880

Automated NT Measurement
It is possible to automatically extract the nuchal
translucency(NT) thickness by simply setting ROI(region
of interest). In addition, thicknesses at max NT, Mean NT
can be automatically calculated.

Protocol Assistant
This function ensures that all measurements and image
saving are performed without fail by carrying out
examinations with the protocol that has been registered
in advance.
・ Protocol registration: 128 kinds can be registered.
・ Suspend protocol function
・ Create and edit protocol on the system
・ Register and edit reference image
・ Import/Export protocol

Manual
It is applicable to browse instruction manual on this
system.
・ On-board operation manual

7
ARIETTA850DI-V50
Data Management Measurement data
Image data It is possible to store measurement data in the main
 Format unit hard disk
 Multiple-frame(moving) image
 DICOM(Raw, MJPEG) Patient data
 PC Format(WMV, AVI, MP4)  Displayed information*1
 Single-frame(still) image  Patient information*2
 DICOM(Non-compressed, RLE, RGB(Plane/Pixel), ID(up to 64 characters), Name(up to 64
JPEG) characters, including middle name), Birthdate, Sex,
 PC Format(Tiff, Bmp, JPEG) Age, Height, Weight, Occupation
 Image acquisition mode  Study information
 Real-time multi-frame image acquisition Procedure ID, Accession, Study ID, Study
Raw, Image, RAW and Image at the same time Description, Referring physician, Reporting Phys,
 Post ECG: Max. 10 cardiac cycles(R-R) Sonographer
 Pre ECG: Max. 10 cardiac cycles(R-R) *1: Conforms to DICOM 3.0 standard
 Post Time: Max. 90 seconds *2: Encryption function is available.
 Pre Time: Max. 16 seconds
 Manual: Data storage
 Raw data: Max. 150 seconds  Main unit hard disk*
 Image data: Max. 180 seconds  Capacity: Approx. 1TB
 Cine loop high-speed data transfer(Raw, Image)  USB memory*
It is possible to selectively store data of arbitrary  USB HDD*
section in the Cine Memory.  CD-R
 Simultaneous output to multiple media  DVD-RAM
It is possible to output still image data to multiple  DVD-R
of storage media include network and printers at  Network interface(DICOM format)
the touch of a button. 10 BASE/T or 100 BASE/TX, automatically switched
 Image data management tool *: Encryption function is available.
 Image viewer
 Compatible with DICOM and PC-format images DICOM network communication
 Simultaneous display of stored and real-time ・ Conformity to DICOM service class:
images is possible(Compare mode)  Ultrasound image storage SCU
 Thumbnail display of stored images(1-36 images)  Ultrasound multi-image storage SCU
 Check mark is put on a transferred image  Storage media FSC/FSR
 Image zoom, rotation, inversion  Print management SCU
 Protect stored images  Modality worklist management SCU
 1:1 replay(main unit HDD or DICOM storage  Modality performed procedure step(MPPS) SCU
data) (For details, please refer to the DICOM Conformance
 DVD-RAM Statement issued by FUJIFILM Healthcare
 CD-R Corporation.)
 USB memory ・ Storage
 USB HDD Possible to store patient information directly to DICOM
 Re-storing to media, transfer file server.
 Adjustment is possible on the reconstructed Raw- ・ Print
data image (Gain, Dynamic Range, Gamma Possible to printout images with DICOM compatible
Curve, and Color Map in CF mode) printer directly.
・ Work list management
Retrieval of patient and reservation information from

8
ARIETTA850DI-V50
hospital information system(HIS)
NOTE: The HIS needs to be compatible with DICOM standard
supplement 10. The HIS network and the DICOM network
need to be linked.
・ Router setting
・ Compatible with SR(Structured Report) for OB,
cardiology, vascular and abdominal measurements*
・ Query/Retrieve
・ Compatible with ED(Evidence Documents)
・ Profile
・ Integrated Healthcare Enterprise(IHE)
 SWF(Scheduled Workflow)
 PDI(Portable Data for Imaging)
 ED(Evidence Documents in Radiology/Cardiology
domain)
 Echocardiography Workflow
 CT(Consistent Time)
* Option: SOP-ARIETTA850-21

Security measures
 User authentication function is available.
3 Types of user authority can be set.
It is possible to set whether password is necessary or
not at the start of operation.
 Audit logs
Accesses related to user management and patient
data are recorded as audit log data.
Handling of these logs is limited to users with Level 1
access.

Teaching file
Possible to create teaching file.

9
ARIETTA850DI-V50
Measurements and Analysis  Common bile duct measurement: CBD
Basic measurements  Liver measurement: Liver
B mode  Pancreas measurement: Pancreas
 Distance measurement: Distance, Dist-trace  Pancreatic duct measurement: P-Duct
 Area and Circumference measurement: Area/Circum  Renal measurement: Renal Volume
 Volume measurement: Volume1, Volume2  Spleen measurement: Spleen
 Angle measurement: Angle  Space-occupying lesion measurement: SOL
 Histogram measurement: Histogram  Blood vessel diameter measurement
 Measurement of Congenital Dislocation of the Hip Joint: Aorta Diam, PV Diam
 Hip J Angle  Stenosis percentage measurement
 General purpose index measurement: B.Index %STENO-Diam, %STENO-Area

M mode D mode
 Distance measurement: M.Length  Artery measurement
 Time measurement: Time A-Ao, CA, CHA, SA, SMA, IMA, CIA, HA, Prandial SMA
 Heart rate measurement: Heart Rate  Renal artery measurement*
 Velocity measurement: M.VEL. Renal-A, RA hilum, Seg.A Upp, Seg.A Mid, Seg.A Low
 General purpose index measurement: M.Index  Portal vein measurement: Main PV, Rt.PV, Lt.PV
 Shunt blood vessel measurement
D mode Pre Shunt PV, Prox Shunt, Mid Shunt, Distal Shunt
 Time measurement: Time
 Heart rate measurement: Heart Rate B/D mode
 Blood flow velocity measurement: D.VEL1, D.VEL2  Blood flow volume measurement
 Acceleration (deceleration) measurement: ACCEL FV(Artery) Abdom, FV(Vein) Abdom
 Resistance index measurement: RI *: Auto trace is possible.
 Pressure half-time measurement: P1/2T
 Dop.Caliper measurement: D.Caliper1, D.Caliper2 Cardiology measurements
 Measurement of mean velocity and mean pressure B mode
gradient*1: Mean VEL.  L Cardiology measurements
 Pulsatility index measurement*1: PI M.Simpson*1, GLS(3P)*2, Area-Length*1, Pombo*2,
 Stenosis flow measurement*1: Steno Flow Teichholz*2, Gibson*2, Bullet, Simpson, BP-Ellipse,
 Regurgitant flow measurement*1: Regurg Flow EyeballEF*3
 Dop.Trace measurement*1: D.Trace1, D.Trace2  Valve area measurement: AVA, MVA
 General purpose index measurement*1: D.Index  Aortic valve measurement: AV(2D)
 Pulmonary valve annulus diameter measurement
B/D mode PV Annulus
 Blood flow volume measurement *1: F.Volume, SV/CO  TAVI/TAVR measurement: TAVI/TAVR
 Right ventricular diameter measurement:
CF mode RV Dimension, RV(2D)
 Blood flow volume measurement: Flow Profile*2  Left atrial diameter/aortic diameter measurement
*1: Auto trace is possible. LA/AO
*2 Option: SOP-ARIETTA850-7  Myocardial wall thickness ratio measurement: Ratio
 Left ventricular myocardial mass measurement
Application measurements LV Mass(AL), LV Mass T-E, LV Mass Pedi, LV Mass Pedi
Abdominal measurements Simpson*2
B mode  Inferior vena cava diameter measurement: IVC
 Gallbladder measurement: Gallbladder  Left atrial/right ventricle volume measurement*4
 Gallbladder wall thickness measurement: GB Wall-T LA Vol., RA Vol.

10
ARIETTA850DI-V50
 Right ventricular function measurement*2: FAC Transverse Ao Doppler, Abdominal Ao Doppler
 Blood Flow Measurement of a Defect Hole for Congenital
M mode Heart Disease*2: ASD, VSD, PDA, PFO
 Left ventricular function measurement*5  Coronary measurement*7
Pombo, Teichholz, Gibson prox LAD(Rest), prox LAD(Peak), distal LAD(Rest),
 Mitral valve measurement: Mitral Valve, MV(M) distal LAD(Peak), RCA(Rest), RCA(Peak), LCX(Rest),
 Tricuspid valve measurement: Tricuspid Valve LCX(Peak), Graft(Rest), Graft(Peak),
 Pulmonary valve measurement: Pulmonary Valve Coronary1,2,3(Rest)(Peak), Coronary Stenosis
 Left atrial diameter/aortic diameter measurement*6  Asynchrony measurement
LA/AO AV Async., InterV.Async., TDI PW Time to Onset, TDI
 Inferior vena cava diameter measurement: IVC PW Time to Peak
 Asynchrony measurement: IntralV.Async.
 Measurement of tricuspid annular plane systolic CF mode
excursion(TAPSE):  M TDI measurement
TAPSE M TDI mFS, M TDI WT(LVPW), M TDI MT(IVS)
 Right ventricular measurement: RV(M) *1: Auto or Full Auto measurement is possible.(Option: SOP-
ARIETTA850-74 is necessary for Full Auto measurement.)
D mode *2: Auto measurement is possible.(Option: SOP-ARIETTA850-74
 Left ventricular ejection flow measurement*7 is necessary for Auto measurement.)
LVOT Flow, LVOT *3 Option: SOP-ARIETTA850-58, PEU-LISENDO880
 Aortic stenosis flow measurement*7: AS Flow, AV *4: Auto or Full Auto measurement is possible.(Option: SOP-
 Aortic valve regurgitation flow measurement*7 ARIETTA850-74 is necessary when Auto or Full Auto
AR Flow, AR measurement.)
 Right ventricular ejection flow measurement*7 *5: Caliper Mark Auto Shift is possible.
RVOT Flow *6: Auto measurement is possible.(Option: SOP-ARIETTA850-74
 Pulmonary stenosis flow measurement*7 and PEU-LISENDO880 is necessary when Auto measurement.)
PS Flow, PVA, PV, PA *7: Auto trace is possible.
 Pulmonary valve regurgitant flow measurement*7 *8: Option: PEU-LISENDO880 is necessary for Doppler Trace
PR Flow, PR method.
 Left ventricular inflow measurement*7*8
Trans M Flow, MV, MV Valsalva Vascular measurements
 Mitral annulus diameter measurement: MV Annulus B mode
 Mitral stenosis flow measurement*7: MS Flow  Carotid Artery:
 Mitral regurgitant flow measurement*7: MR Flow  Stenosis measurement
 Tricuspid annulus diameter measurement: TV Annulus Carotid %STENO-D, Carotid %STENO-A
 Tricuspid inflow measurement: TV, TV Valsalva  mean-IMT measurement: mean-IMT, CmeanIMT*1
 Tricuspid stenosis flow measurement*7: TS Flow  Max-IMT measurement
 Tricuspid valve regurgitant flow measurement*7 max-IMT, IMT-Cmax, IMT-Imax, IMT-Bmax
TR Flow, TVA(VTI), TVA(pV)  IMT Automatic measurement:
 Pulmonary vein flow measurement: PV Flow CCA_IMT, ICA_IMT, ECA_IMT, BIF_IMT
 PISA measurement  IMT-C10 measurement: IMT-C10, IMT-C10 Auto*1
MR Vol. PISA, AR Vol. PISA, TR Vol. PISA, PR Vol. PISA  Upper Ext Artery:
 TDI PW measurement*7*8  Stenosis measurement
TDI PW MW, TDI PW1, TDI PW2, M Annulus(m), M Upper %STENO-D, Upper %STENO-A
Annulus(l), M Annulus(s), T Annulus(s), T Annulus(l)  Lower Ext Artery:
 MPI measurement: LV MPI, RV MPI, RV MPI(Tissue)  Stenosis measurement
 Aorta measurement*2 Lower %STENO-D, Lower %STENO-A
Ascending Ao Doppler, Descending Ao Doppler,

11
ARIETTA850DI-V50
D mode M mode
 Carotid artery blood flow measurement  Left ventricular function measurement: LV Function
CCA proximal, CCA mid, CCA distal, ICA, ECA, BIFUR,  Fetal heart rate measurement: FHR, PreHR, PstHR
VERT
 Upper extremity artery blood flow measurement D mode
ScA, AA, BA, DBA, BasA, RA, UA, SPA  Blood flow measurement
 Lower extremity artery blood flow measurement Umbilical Artery, MCA, Uterine Artery, Descending
CIA, EIA, IIA, CFA, DFA, SFA, PopA, PerA, ATA, PTA, Aorta, Renal Artery
DPA  Left(right) ventricular ejection flow measurement:
 Transit time of vessel flow(TVF) measurement*2 LVOT Flow, RVOT Flow
TVF_CFA, TVF_POP, TVF_ATA, TVF_PTA, TVF_PA  Preload index measurement: PLI
 Upper extremity venous blood flow measurement  Ductus venosus measurement: Ductus Venosus
IJV, ScV, CV, AV, BV, DBV, BasV, RV, UV  Fetal heart rate measurement: FHR, PreHR, PstHR
 Lower extremity venous blood flow measurement
CIV, EIV, IIV, CFV, DFV, SFV, GSV, PopV, PerV, LSV, Others
ATV, PTV  Supports multiple pregnancy examination
 Transcranial artery blood flow measurement *1 Auto measurement is possible.
ACA, MCA, PCA, BA, VA, TICA, ACoA, PCoA *2 Caliper Mark Auto Shift is possible.
*1: IMT Auto measurement is possible. *3 Option: SOP-ARIETTA850-72
*2 Option: SOP-ARIETTA850-47, PEU-LISENDO880 *4: in real time
*5 Option: SOP-ARIETTA850-71
Small parts measurements
B mode Gynecology measurements
 Lesion measurement(for breast use): Lesion(Breast) B mode
 Aspect ratio measurement: D/W ratio  Uterus measurement: Uterus
 Papilla-tumor distance measurement: NT dist  Endometrial thickness measurement: Endom-T
 Thyroid gland volume measurement: Thyroid Volume  Cervix measurement: Cervix
 Thyroid isthmus thickness measurement  Ovarian measurement: Ovary
Isthmus Thickness  Follicular measurement: Follicles
 Follicle volume measurement: Follicles Volume
D mode  Bladder measurement: Bladder Volume
 Artery measurement(for mammary blood flow)
 Artery measurement(for thyroid blood flow) D mode
 Uterine artery measurement: Uterine Artery
Obstetric measurements  Ovarian artery measurement: Ovarian Artery
B mode
 Gestational age measurement*1: GA Urological measurements
 Fetal wight measurement: FW B mode
 Ratio measurement: Fetus Ratio  Prostate measurement: PSA Volume, PRS Slice Vol.
 Amniotic fluid index measurement  Seminal vesicle measurement: Seminal Vesicles
AFI*2, AF Pocket, AFV, MVP  Bladder measurement: Bladder Volume
 Cardiothoracic ratio measurement: CTAR, CTR  Testicular measurement: Testis Volume
 Cervical length measurement: Cervix  Renal measurement: Renal Volume
 Nuchal translucency measurement: Auto NT*1  Cortical measurement: Cortex Thickness
 Automatic measurement of fetal heart rate*1*3  Adrenal measurement: Adrenal
Auto FHR, Auto FHR+*4
 Automatic measurement of fetal fractional D mode
shortening*1*5: Auto FS  Arterial blood flow measurement: Renal Artery

12
ARIETTA850DI-V50
Report functions
 Measurement report
 Abdominal measurement report
 Cardiac function measurement report
 Vascular measurement report
 Small parts measurement report
 Obstetric measurement report
 Gynecology measurement report
 Urological measurement report
 Recall past measurement reports.
 Plot examination data history on the report.
 Direct printout of each report with an optional PC
printer.
 Output of measurement values in CSV file is possible.

Others
 Playback measurement
 Assign measuring functions to the alphabet keys on the
keyboard.
 User's Calculation: Combine basic measurement
functions such as distance, area and flow velocity to
create index calculation formulas for measurement
packages.
It is possible to create measurement packages that
contain a total of 30 calculations per application.
 Reserved Word: Measurement parameter names for
application measurements, or user-registered
measurement names that can be used in User's
Calculation.
60 reserved words can be registered for each
application.
 Font size of measurement result
3 kinds(x1, x1.2, x1.4)

13
ARIETTA850DI-V50
Optional Functions  Global LV twist and LV torsion calculation
PC printer  Automated workflow allows analysis can be started
It is possible to printout Abdomen, cardiology, vascular, without user interaction.
small parts, OB/GYN, and urology report screens including  Displacement and strain analysis
ultrasound images directly with an external PC printer.  Longitudinal, circumferential and radial component
analysis
Cardiac 3D(4D)*  Principal tangential strain and 3D displacement
Cardiac 3D(4D) function provides the capability to display analysis
Cardiology 3D image and to display dual image(reference  Segmental peak, time to peak and global average
image and biplane image) simultaneously by using TTE values in result table
2D array probe.  Segmental parameter over time chart display
・ Scanning rate: Up to 999 volumes/sec  Rotation and twist analysis
・ Biplane Mode(Color)  Twist, Rotation Base, Rotation Apex
・ 3D Zoom Mode(Color)  Segmental parameter over time chart display
・ Active 3D Mode(Color) * Option: SOP-ARIETTA850-112
・ Wide Angle 3D Mode(Color) This option is manufactured by TOMTEC Imaging System GmbH
・ Live 1,2,3,4,5,6 beat 3D volume imaging
・ Live 1,2,3,4,5,6,7 beat 3D color volume imaging 4D RV-Function*
・ It is possible to display 3 arbitrary sections  Semi-Automated quantification of the Right Ventricle
simultaneously.  Fast and easy Workflow
・ MPR(Multi-planar Reconstruction)  Automatic contour proposal - no laborious contour
・ Auto Rotation drawing
・ 2 click cropping operation  RV surface model
・ 3D Cropping  4D Review and Analysis
・ Display Layout(Single, Dual, Quad, MSI)  Real-time interactive 4D model(Beutel) display of
・ Multi slice Image ventricle
・ Function to change 3D Imaging Parameter  Display of 2D ultrasound slices(SAX and LAX) with
・ Change MPR View to operate by drag Beutel contours as overlay
・ Density(Both Lateral and Elevation)  Graph displays for global volumes over time
・ Performance mode(Function to control the frame rate  Display of result window with derived parameters
and volume rate to adjust the temperature at the probe  Measurements
tip.)  Global volumes over time
・ Change Plane Angle to observe arbitrary cross-section.  Calculation of EDV, ESV, EF, SV
* Option: SOP-ARIETTA850-129, EU-9207, PEU-LISENDO880  Right ventricular longitudinal strain(RVLS)
 free wall
iEF*  septum
iEF calculates Biplane EF with full automated ED/ES frame  Additional 2D measurements
detecting and full automated A2C and A4C contour  TAPSE
tracing of biplane images acquired from a transthoracic  FAC
2D matrix probe.  RVDd base(RVD1)
* Option: SOP-ARIETTA850-120, SOP-ARIETTA850-74, SOP-  RVDd mid(RVD2)
ARIETTA850-129, EU-9207, PEU-LISENDO880  RVLd(RVD3)
* Option: SOP-ARIETTA850-113
4D LV-Analysis* This option is manufactured by TOMTEC Imaging System GmbH
Additional features to 4D LV-Function
 Polar plot with parametric display for all parameters. 4D MV-Assessment*
 Dynamic(amplitude) and static(time to peak) parametric  Semi-Automated Annulus and Leaflet Detection
display for all parameters.  Fast and easy Workflow

14
ARIETTA850DI-V50
 Semi automatic annulus detection  Spatio-Temporal Image Correlation(STIC): Captures
 Automatic and dynamic MV model generation via volume data in B mode or Color Doppler mode via a
object segmentation slow scan.
 Automatic measurements for annulus, leaflets and  Rendering Mode
coaptation  Surface Sharp: Displays images that mix Surface(3D
 Morphological and functional analysis views images with a smooth texture) and Sharp(3D
 Analysis screen for 4D presentation of tissue and images with a detailed texture) effects.
annular structures  Surface MIP: Displays a mixed image composed of a
 Transparence rendering to display all objects within surface image and MIP(maximum intensity
tissue projection: A 3D image at its highest echo intensity
 Definition and display of surgical and ventricular view as seen from the viewpoint) image.
orientations  Surface X-ray: Displays a mixed image composed of
 Display of B/W tissue and Doppler color information a surface image and an X-ray(a 3D image averaged
in all views echo data seen from the viewpoint) image.
 Adjustable cut plane(D’Art)  4Dshading: Displays realistic 3D images with light
 Color texture on leaflet models that represents their and shadow that look like endoscopic images.
topology  4Dshading Flow: Displays realistic color Doppler 3D
 4D Reconstruction Mode images with shadows, similar to optical images.
 Real-time rendering  4Dtranslucence: Displays 3D images in which the
 Interactive 4D cube for data presentation and surface contour of the target has been enhanced.
manipulation  3D Image Display Mode
 4D rendering of B/W tissue and color Doppler data  3D/4D: Displays cross-sectional and 3D images.
 3D visualization of automatic measurements  Free Axis of MPR(FMPR): Standard reference points
 Hybrid 4D display of measurements and grayscale can be moved, and the cross-sectional image
data displayed.
 Interactive display settings for optimum display  Multi Slice Imaging(MSI): Displays multiple cross-
 Separate settings for threshold and transparency sections reconstructed from volume data.
 Generic manual measurements  Volume Slice Imaging(VSI): Displays 3D images of
 Further general features volume data of arbitrary thickness at a specific axis.
 On-screen help  Curved MPR(CMPR): Displays cross sections on an
 Planimetry layout for individual MPR adjustment arbitrary curve or straight line from the volume data.
 Carpentier segmentation labels on leaflet surface  Measurements
 Measurements  Cross-sectional images: Distance, circumference,
 Manual measurements such as distance, angle, and area, etc. can be measured in the same way as
curve and area with general B mode images. (except for STIC)
 Static and dynamic annular measurements  3D Image: Measurements are possible in CMPR
 Static and dynamic leaflet measurements mode(distance, circumference) and VSI
 Static and dynamic coaptation measurements mode(distance).
* Option: SOP-ARIETTA850-114  Auto Clipper: This function makes automatic
This option is manufactured by TOMTEC Imaging System GmbH adjustments to achieve an appropriate Clipper
configuration to display a fetal in 3D image.
RT-3D(4D)*  3D image quality adjustment: Lower Threshold, 3D HI
 Method for acquiring volume data REZ, Inversion Mode
 4D(real-time 3D): Captures volume data in a * Option: EU-9184, SOP-ARIETTA850-4
continuous B mode scan.
 HI Definition 3D: Captures volume data in B mode. Real-time Tissue Elastography*1
 Color 3D: Captures volume data in color Doppler It is a function to visualize the stiffness of a tissue in real-
mode. time. The strain generated in a tissue on applying

15
ARIETTA850DI-V50
pressure is represented by colors. Measures the propagation velocity of shear waves in
(It is available to change Elastography color map) the target region.
 Elasto mode  ROI setting function
Overlapping display of Elastography on B-mode A trackball can be used to change the position of the
image. ROI you select to measure.
 Dual Elasto mode  Automatic recording of still images
Dual image display of Elasto mode, possible to select The instrument can be set up to automatically send
different color map on each side. results after measurements.
 Real-time biplane  Integration with measurement functions
Elasto mode overlapping display of Elastography on B- Measurement data is saved for each selected
mode image of either cross-section, in the function to measurement. You can modify a measurement during
display B-mode image with the one of different cross- an examination.
section simultaneously.  Data saved to CSV files
 Strain Ratio measurement Measurement data is saved in a CSV file.
Calculation of ratio of strain between arbitrary 2  Attenuation measurement(iATT)*2
regions. Simultaneously with the measurement of Vs, the
 Assist Strain Ratio ultrasound attenuation(ATT) is measured near the
Clicking center of the tumor automatically sets the center of the ROI.
measurement ROI to the tumor and the fatty layer  Combi-Elasto is possible.
(the mammary gland region and lesion are targeted). *1 Option: SOP-ARIETTA850-151
 Strain Graph *2: It is only available with C251 and C252 probe.
The time variation of distorted average value is
displayed on a graph in real time. Shear Wave Elastography*
 Auto Select Frame It is a function for measuring the propagation velocity of
The frame under stable pressure is chosen shear waves and the stiffness, and creating images of the
automatically. 2D distribution of propagation velocity. The propagation
 Strain Histogram Measurement*2 velocity of shear waves varies according to tissue
Displays a histogram which shows the values of hardness.
relative strain within an ROI and calculates the  Shear Wave Elastography(SWE) mode
quantitative characteristics of the elasticity image and You can superimpose shear wave propagation
Liver Fibrosis Index. velocities in color on B mode images. The color SWE
 Compatible to RAW data image appears inside the SWE ROI.
 Guide of displacement direction  ROI setting function
Display the direction of displacement of the liver This function lets you use the trackball to adjust the
caused by the heart beat in real time. position of the measurement ROI.
 HI Strain  SWE measurement
Performs the strain calculations based on multiple You can measure the propagation velocity of shear
frames, and displays the most stable elasticity image. waves within a single or multiple arbitrary regions.
*1 Option: SOP-ARIETTA850-43 Measurement starts when the image is frozen.
*2 Option: SOP-ARIETTA850-60 * Option: SOP-ARIETTA850-151

Shear Wave Measurement*1 Combi-Elasto*


It is a function for measuring shear wave propagation The Shear Wave Measurement screen and the Real-time
velocity(Vs), interquartile range of the Vs group(IQR), Vs Tissue Elastography screen are displayed at the same
effective rate(VsN) and elastic modulus(E). The time in the dual-screen view. While viewing an elasticity
propagation velocity of shear waves varies with tissue image, you can perform shear wave measurement to
hardness. measure the velocity, hardness, and attenuation value of
 Shear Wave Measurement(SWM) mode shear waves.

16
ARIETTA850DI-V50
* Option: SOP-ARIETTA850-43, SOP-ARIETTA850-60, SOP- image section.
ARIETTA850-151  Last registration
It is possible to return just before the condition of
Automated FHR Measurement* registration.
Set measurement ROI on fetal heart plane in B mode  Reset function
image and measures heart rate automatically. It is possible to return initial condition of registration.
* Option: SOP-ARIETTA850-72.  Built-in magnetic sensor*2
 Displaying fused virtual images*3
Automated FS Measurement* By fusing a parts image or color image of the
It is Possible to automatically extract the fetus left SYNAPSE VINCENT analysis image to the CT or MRI
ventricle inside diameter shortening rate(FS). base image, you can use the fused image as a virtual
* Option: SOP-ARIETTA850-71 image in an RVS examination.
 3D Sim-Navigator*4
Real-time Virtual Sonography*1 3D Sim-Navigator is an auxiliary function provided by
This function is able to synchronize the volume data the Real-time Virtual Sonography(RVS) software.
which was previously acquired by CT, MRI, PET, or US  Displays the puncture guideline on virtual images
and in real-time US image.  Navigator view(3-screen)
The magnetic position sensor is installed on the probe to  Displays C-plane image
determine the position and angle of the probe.  Registration of needle marks
As a result, the display of cross sections that match  Displays distance between multiple needle marks
ultrasound images is enabled, allowing you to use the  E-field Simulator *5
cross sections as a reference for your ultrasound A color map superimposed on the CT image simulates
diagnosis. the distribution of electric current from the given
The same section of ultrasound image and CT/MR/PET location of the multiple electrodes.
image or US volume data is displayed as real time by  Needle Tracking*6
acquiring the position information on the probe from a Tracks and displays the needle tip location in real time
magnetic sensor and reconstructing MPR(Multi Planer during RFA procedures.
Reconstruction) image from CT/MR/PET image and US  Body Motion Tracking*7
volume data. The omniTRAX Active Patient Tracker (manufactured
In various diagnosis areas(liver, breast and prostate) by CIVCO) provides automatic image registration of
examination is possible as it supports convex type, linear fused images using real time ultrasound with
type, intracavital probe. previously acquired CT; with the synchronized status
Moreover, it can use together with Real-time Tissue adjusted when small patient movements occur.
Elastography. *1 Option: EU-9185B or later, EZU-RVF1B, PM-AR850-H004,
 Monitor Display: SOP-ARIETTA850-62
 Dual: Virtual / US *2 Option: PM-AR850-H006(Probe dependent)
 Quad: Virtual / US / Virtual / Virtual *3 Option: SOP-ARIETTA850-97
 Overlay: Virtual + US *4 Option: SOP-ARIETTA850-75
 Magnetic field quality display *5 Option: SOP-ARIETTA850-96
 3D body mark display: ON/OFF *6 Option: SOP-ARIETTA850-84, EU-9197
 3D body mark size: 3 kinds *7 Option: SOP-ARIETTA850-85, EU-9197
 Marking function
 Multi-volume function 2D Tissue Tracking(2DTT)*1
Max. 4 CT/MR/PET/US images for retrieving, saving It is an application for evaluating the thickness and
and the display switching movements of the myocardium.
 US-US RVS function  Analysis
US volume data generated beforehand can be  Free use
displayed as Virtual image in the same to US real-time Analyzes the changes in distance between any two

17
ARIETTA850DI-V50
points and angular variation from the initial time Tissue Doppler Imaging(TDI) analysis*
phase. It performs velocity analysis of the data acquired in the
 SAX TDI mode.
Analyzes differences in thickness between various  B-mode
regions of the myocardium, as well as angular  Temporal VP: Displays a graph of the motion velocity
variation with the center of gravity point, based on distribution in the area specified in the myocardium.
left ventricular short axis images.  Regional VP: Displays a graph of the motion velocity
 Apex-S*2 distribution on a trace line arbitrarily specified in the
Calculates the Longitudinal Strain and Transverse myocardium.
Strain of each region of the left ventricle based on  Strain Rate: Displays a graph of the Strain Rate time
the apex image. The global strain of the variations in a graph.
endocardium trace is also calculated.  Strain: Displays a graph of the Strain time variations
 Apex-V*2 in a graph.
Calculates the volume, ejection fraction, center of  Myocardial Thickness: The changes in myocardial
gravity and velocity of volume change of the right thickness in the time direction of the designated
and left ventricles and atrium based on the apex endocardium and epicardium movement within the
image. myocardium are automatically traced and displayed
 View in a graph. The Strain in that area can also be
 SAX displayed.
SAX, Basal SAX, Mid SAX, Apical SAX  Output as a CSV file: Possible
 APEX-S, -V * Option: SOP-ARIETTA850-13
4ch, 4ch Inv, 2ch, 2ch Inv, 3ch, 3ch Inv
 Graph Type Contrast Harmonic Imaging(CHI)*
 Line Graph: Displays analysis results for each ROI It is a function that displays images more visibly by dying
chronologically in line graphs. them with an ultrasound contrast agent.
 Color Graph: Displays a list the analysis results for Settings ideal for Low, Mid, and High acoustic pressure
each ROI into the colors corresponding to the color are selectable for each contrast agents.
bar.  CHI Method for CHI mode:
 Line & Color Graph: Displays Line Graph and Color  Wide-band Contrast Harmonic Imaging (WbC): The
Graph up and down. Pulse Inversion method enables retrieval of echoes
 Bull’s eye: When the view is assigned, the analysis from the ultrasound contrast agent across a broad
results for each ROI are displayed with color and spectrum and display them in CHI mode at a high
numerical values in a Bull’s-eye image. level of sensitivity.
 Measurements  Tissue Reduction Contrast Harmonic Imaging (TrC):
 Point to Point: Measures the difference between two The Amplitude Modulation method makes it possible
points on a graph. to reduce signals from tissue to get clearer CHI
 Time to Peak: Measures various values of any three mode images.
time phases and the difference between them.  CHI Mode image:
 3 point: Measures various values of any three time  CHI-B/W: Displays signals from the ultrasound
phases and the difference between them. contrast agent as black-and-white images.
 Distance: Measures the distance between any two  CHI-Color: Displays signals from the ultrasound
points on an image. contrast agent as color images.
*1 Option: SOP-ARIETTA850-49  CHI-eFLOW: Displays signals from the ultrasound
*2: It is possible to perform fully automatic measurements using contrast agent on power Doppler images.
2D tracking method such as Global Longitudinal Strain or SAX  Tx Mode:
radial strain and ejection fraction measurements by a single click  Monitor Mode: Displays the monitor image and CHI
operation in combination with multiple settings. mode image side by side.
 Flash: This mode sends high acoustic pressure

18
ARIETTA850DI-V50
transmissions at set intervals to destroy the time it takes to reach -6 dB or -3 dB from the
ultrasound contrast agent. maximum value of the graph, and measures the
 One-shot: This mode sends and receives a single time width between the intersection points on either
shot of an ultrasound signal at high acoustic pressure side of the maximum value.
in response to a manual operation.  Distance: Measures the distance between any two
 Frame Rate Limit points on an image.
Limits the frame rate to prevent ultrasound waves  Inflow Time Mapping(ITM) Measurement: Measures
from breaking down the contrast agent. the number and proportion of color pixels at each
 Overlay display time interval in an ITM image.
Body motion compensation by Motion Compensated  Area Under the Curve(AUC) Measurement: Measures
Mode is possible. the area under the curve for a section on the graph.
* Option: SOP-ARIETTA850-44  Output as a CSV file: Possible
* Option: SOP-ARIETTA850-44
Contrast Harmonic Imaging(CHI) analysis*
It is a function that lets you display the data collected in EyeballEF*
CHI mode as a Time Intensity Curve or analyze it in the Selection of apical four-chamber view(A4C) and apical
form of Subtraction images. two-chamber view(A2C) and tracing endocardium surface
 Analysis automatically, and measure heart volume, ejection
・ Time Intensity Curve(TIC) fraction(EF), and Global Longitudinal Strain(GLS) using M.
Displaying a graph of the temporal changes in the Simpson method.
average(or median, maximum, or mode) intensity  Revising Items:
of the region of interest(ROI).  Specifying a cross section
 Types of graph:  Setting heartbeat analysis
 By ROI: Displays graphs for all frames in  Tracing an end-diastolic image
chronological order.  Specifying an end systolic frame
 By Frame Number: Displays a graph for each  Tracing an end-diastolic image
frame number in a unit. * Option: SOP-ARIETTA850-58, PEU-LISENDO880
 Inflow Time Mapping(ITM)
This function focuses on the difference between eTRACKING*
ultrasound contrast agent inflow times and It is a function to calculate arteriosclerosis indices such as
colorizes changes in brightness across the entire the pressure-distortion elastic constant(Ep) and the
range of the moving image. stiffness parameter(β), which indicate the elastic
 Subtraction characteristics of the blood vessel based on the blood
Display subtraction images that represent the vessel diameter and blood pressure values of the carotid
difference between frames. artery.
 Measurement  Stiffness parameter(β) analysis results:
 Point to Point: Measures the difference between two  beta: Stiffness parameter
points on a graph.  Ep: Pressure-distortion elastic constant
 SD: Measures the analysis data for the specified time  AC: Arterial compliance
phase(Time, Max, Average, Medium, Mode, or SD).  AI: Augmentation Index
 Wash-in: Measures the rise time of the graph from  PWVbeta: Regional pulse wave velocity
the minimum and maximum values.  DATmax: Maximum Distension Acceleration Time
 Wash-out: Measures the fall time of the graph from  Output as a CSV file: Possible
the minimum and maximum values.  Blood pressure value is necessary for this examination.
 Fitting: Displays an approximated curve based on * Option: SOP-ARIETTA850-11, PEU-LISENDO880
the start and end points, and measures the degree
of increase in the signal strength. Flow Mediated Dilatation(FMD) analysis*
 FWHM(Full Width at Half Maximum): Calculates the It is used for the functional assessment of vessels before

19
ARIETTA850DI-V50
organic changes, such as vessel wall thickening and
plaque formation, are manifested.
In an FMD test, blood flow volume is increased
noninvasively, and changes in blood flow velocity and
vessel diameter are recorded precisely for long periods of
time.
Echo Tracking technology is used to capture and record
changes in the state of the vessel at three stages such as
baseline, during a vascularization and dilatation after the
cuff deflation, to assess FMD.
 Output as a CSV file: Possible
 Blood pressure value is necessary for this examination.
* Option: SOP-ARIETTA850-16(includes eTRACKING software),
PEU-LISENDO880

Wave Intensity(WI)*
Wave Intensity (WI) has been proposed as an indicator to
allow simple determination, based on the forms of the
blood pressure waveforms and velocity waveforms at any
point in the arterial system, of which is predominant, the
forward-traveling wave heading from the heart to the
periphery, or the backward-traveling wave returning from
the periphery to the heart.
 Wave Intensity analysis:
 Wave Intensity: Wave Intensity index value
 Arterial Stiffness: Index values for assessing the
hardness of the blood vessel
 Diameter: Distension waveform
 Flow Velocity: Blood flow velocity measurement
value
 Output as a CSV file: Possible
 Blood pressure value is necessary for this examination.
* Option: SOP-ARIETTA850-34 (includes eTRACKING software),
PEU-LISENDO880

Detective Flow Imaging(DFI)*


Removes motion artifacts from the feature amount of
movement obtained by analyzing the received signals in
the ROI, and displays images in colors that vary according
to the signal intensity. Blood flow is displayed at a high
frame rate and is little affected by body movement.
* Option: SOP-ARIETTA850-105

20
ARIETTA850DI-V50
General Specifications Resolution: WXGA++(1600x900)
Acoustic Power Output
 0 to 100%  DVI-D digital
Resolution: WXGA++(1600x900)
Preset Function  Digital video with HDMI connector*
 100 kinds(Max. 10 kinds per each probe) Resolution: FHD(1920x1080), XGA(1024x768),
 Preset contents storable in USB memory VGA(640x480)
 Quick Scanning Selector(Q.S.S.) * Option: EU-9205
Image modifying parameters (e.g. Gain, frequency, ・ Analog Video Input/Output
depth) of your choice can be registered. (Up to 4 sets Input
per preset) ・ Y/C: 1 channel
These parameters can immediately be registered and Output
selected by touch panel during examinations.  Color composite (BNC): 1 channel
 Preset is booted up in conjunction with ID information  Y/C: 1 channel
(BodyParts or etc.) ・ Network: LAN(Wired, Wireless)
・ Others
Characters and graphic displays ・ Audio (L/R): 2 channels (Output 1, Input 1)
 Character input area
ID, name, age, sex, retained text Storage
(Can be corrected after exam.) ・ SSD
・ Input is possible with virtual keyboard on LCD panel ・ HDD
・ Automatic Annotation Labeling
800 words(User registration is possible. 10 classes.) Viewing Monitor
・ Body mark: ・ 22 inch Organic EL monitor(OLED)
 38 kinds are available per each region. ・ Resolution: FHD(1920 x 1080)
 6 regions+1 user is able to register. ・ Tilt and swivel are possible.
 Probe mark: 4 kinds ・ Height adjustment and swivel together with operation
 Display position: changeable panel.
 Fetus mark
Rotatable(Only single horizontal fetus marks) Safety Regulation
・ Assist line display  IEC 60601-1 Ed.3.0: 2005, IEC 60601-1 Ed.3.1: 2012
 Class I, Type BF
Menu control
・ 10.4-inch color TFT LCD touch panel Environmental Requirements
 In Operation
Active Probe Ports  Temperature: +10 to +40 degrees C
・ For electronic scanning probes: 6 (4 active, 2 parking)  Relative Humidity: 30 to 75%(non condensing)
・ For independent probes*: 1  Atmospheric pressure: 700 to 1060 hPa
* Option: EU-9187B  Altitude: Up to 3000 m
 In Storage/Transportation
Input/Output Signals  Temperature: -10 to +50 degrees C
・ Data Input/Output (0 to +50 degrees C for mechanical probes)
・ USB2.0: 5 channels  Relative Humidity: 10 to 90%(non condensing)
(Main unit 2+ Operation Panel 3)  Atmospheric pressure: 700 to 1060 hPa
・ USB3.0: 1 channel
・ Digital Video Input/Output Power Requirement
Input  100 to 120/ 200 to 240 V ±10%, 50 or 60 Hz,
・ DVI-D digital: Max. 1300 VA(with optional recorders connected)

21
ARIETTA850DI-V50
Max. 900 VA(main unit only)
 Shut down tool
Hibernation*
*: The condition is saved even if the power is unplugged.

Dimensions
 55 cm(W) × 90 cm(D) ×122 to 169.5 cm(H)

Weight
 Main unit only: 145 kg±10%
 Main unit and all optional equipment: 163 kg±10%

22
ARIETTA850DI-V50
System Configuration

Continue to the next page

23
ARIETTA850DI-V50
24
ARIETTA850DI-V50
OPTIONAL PROBES
Electronic convex sector probes
Application Frequency Scanning angle
Model Optional accessories
(description) range (MHz) (degrees)
Abdominal C22K 6.0-1.0 82 Puncture Adapter
MP-2781
MP-2781-5
MP-2781-25
Bracket
614-108, 614-109(CIVCO)*1*5
Abdominal C22P 6.0-1.0 74 Needle Guide Bracket
EZU-PA7C2*1
MP-2824
Magnetic Position Sensor Attachment(Standard)
RV-012*2
Abdominal C23 6.0-1.0 110* Verza Bracket
644-095, 644-096(CIVCO)*1*5
BX2 Needle Guide
644-094(CIVCO)*1*5
Magnetic Position Sensor Attachment
RV-017*2
Waterproof Case
WP-001
Abdominal C23RV 6.0-1.0 110* Verza Bracket
644-095, 644-096(CIVCO)*1*5
BX2 Needle Guide
644-094(CIVCO)*1*5
Magnetic Position Sensor Attachment
RV-017*3
Waterproof Case
WP-001
Abdominal C251 5.0-1.0 70 Bracket
644-082, 644-083(CIVCO)*1*5
Magnetic Position Sensor Attachment
RV-004*2
Abdominal C252 6.0-1.0 150** Bracket
644-082, 644-083(CIVCO)*1*5
Magnetic Position Sensor Attachment
RV-004*2
Abdominal C25P*4 5.0-1.0 70 Biopsy Attachment
EZU-PA7B1-1
EZU-PA7B1-2
EZU-PA7B1-3
EZU-PA7B1-4
EZU-PA7B1-C
Magnetic Position Sensor Attachment
RV-005*2
Abdominal C35 8.0–2.0 110* Bracket
644-082, 644-083(CIVCO)*1*5
Magnetic Position Sensor Attachment
RV-004*2
Transvaginal・ C41B 10.0-2.0 200 Puncture Guide Tube
Transrectal MP-2445
Magnetic Position Sensor Attachment
RV-013*2
Waterproof Case
WP-001

25
ARIETTA850DI-V50
Transrectal C41RP 9.0–2.0 180 Puncture Guide Tube
MP-2452-G18(Standard:2pcs)
MP-2452-G21/-G16/-G14(Option)
Waterproof Case
WP-001
Transvaginal C41V1 10.0-2.0 200 Puncture Adapter
EZU-PA7V
Magnetic Position Sensor Attachment
RV-002*2
Abdominal C42 8.0-4.0 80 Needle Guide Bracket
EZU-PA532
Bracket
644-077, 644-078(CIVCO)*1*5
Magnetic Position Sensor Attachment
RV-006*2
RV-007(for EZU-PA532)*2
Intraoperative C42K 10.0-4.0 65 Puncture Adapter
MP-2458
Bracket
614-068, 614-100(CIVCO)*1*5
Intraoperative C42T 10.0-3.0 65 Magnetic Position Sensor Attachment
RV-003*2
Waterproof Case
WP-001
Abdominal C421 12.0-3.0 100* Verza Bracket
644-098, 644-099(CIVCO)*1*5
Magnetic Position Sensor Attachment
RV-017*2
Waterproof Case
WP-001
*When Wide Scanning is ON.
**When Wide Scanning PLUS is ON.

Electronic linear probes


Application Frequency Scanning width
Model Optional accessories
(description) range (MHz) (mm)
Intraoperative L31KP 9.0-2.0 6 Puncture Adapter
MP-2450(Standard)
Waterproof case
WP-001
Small Organ L34 7.0-3.0 38 Coupler
EZU-PA3C1H
Bracket
644-079, 644-080(CIVCO)*1*5
Small Organ L35 9.0-2.0 45 Magnetic Position Sensor Attachment
RV-017*2
CIVCO Verza Needle Guide
610-1500-24(CIVCO)*5
No Bracket for puncture
(Verza Needle Guide can be attached directly)
Intraoperative L43K 12.0-2.0 26 Waterproof Case
WP-001
Peripheral Vessel L441 12.0-2.0 38 Bracket
644-075, 644-076(CIVCO)*1*5
Intraoperative L44K 14.0-2.0 42 Waterproof Case
WP-001
Intraoperative L46K1 14.0-2.0 63 Waterproof Case
WP-001
Intraoperative L51K 15.0-3.0 13 Waterproof Case
WP-001
26
ARIETTA850DI-V50
Intraoperative L53K 15.0-3.0 25 Waterproof Case
WP-001
Intraoperative L44LA 13.0-2.0 36 -
Intraoperative L44LA1 13.0-2.0 38 Waterproof Case
WP-001
Small Organ L55 13.0-5.0 50 Needle Guide Bracket
EZU-PA7L2*1
Elastography Stabilizer
EL-001
Magnetic Position Sensor Attachment
RV-008*2
Small Organ L64 18.0-5.0 38 Needle Guide Bracket
EZU-PA7L3*1
Elastography Stabilizer
EL-002
Magnetic Position Sensor Attachment
RV-009*2
Acoustic Coupler Attachment
EZU-TEATC2
Acoustic Coupler
EZU-TECPL1(for RTE)
SF-001

Electronic phased array sector probes


Frequency
Application Scanning angle*
Model range Optional accessories
(description) (degrees)
(MHz)
Cardiac Adult S121 5.0-1.0 120 -
Cardiac Pediatric S31 9.0-2.0 100 -
Neonatal S42 14.0-3.0 100
-
Cardiology
Cardiac Adult, S3ESL1 9.0-2.0 100 Waterproof Case
TEE WP-001
Cardiac, TEE S3ESEL 8.0-2.0 100 -
Cardiac, TEE S3ESCLS 8.0-2.0 100 -
*When Wide Scanning is ON.

CMUT Probes (linear probes)


Frequency Scanning width
Application Model Optional accessories
range (MHz) (mm)
Small Organ SML44 22.0-2.0 38 Magnetic Position Sensor Attachment
RV-014*2

RT-3D (4D) Probes*


* SOP-ARIETTA850-4 and EU-9184 are necessary.
Frequency Scanning angle
Application Model Optional accessories
range (MHz) (degrees)
Fetal VC34 7.0-2.0 70 -
Fetal VC35 8.0-2.0 72 -
Transvaginal VC41V 8.0-2.0 145 -

27
ARIETTA850DI-V50
Bi-plane probe
Application Frequency range Scanning angle
Model Optional accessories
(description) (MHz) (degrees)
Transrectal CC41R 8.0-4.0 100/120 Puncture Adapter
EZU-PA5V
Puncture Guide Fixture
EZU-PA3U
Magnetic Position Sensor Attachment
RV-010*2
Waterproof Case
WP-001
Transrectal CC41R1 10.0-2.0 180/180 Puncture Adapter
EZU-PA5V
Magnetic Position Sensor Attachment
RV-015*2
Waterproof Case
WP-001
Transrectal C41L47RP Convex 8.0-4.0 200 Puncture Guide Fixture
Linear 10.0-5.0 64mm EZU-PA3U
Magnetic Position Sensor Attachment
RV-011*2
Transrectal CL4416R Convex 10.0-2.0 180 Puncture Guide Fixture
Linear 14.0-2.0 63mm BA-001
Magnetic Position Sensor Attachment
RV-011*2
Waterproof Case
WP-001
Transrectal CL4416R1 Convex 10.0-2.0 200* Puncture Adapter
Linear 14.0-2.0 63mm BA-001
Magnetic Position Sensor Attachment
RV-021*2
Waterproof case
WP-001
*When Wide Scanning is ON.

Matrix 3D(4D) Probes*


*SOP-ARIETTA850-129, EU-9207 and PEU-LISENDO880 are necessary
Frequency range Scanning angle
Application (description) Model Optional accessories
(MHz) (degrees)
Cardiac Adult, TTE MXS1 5.0-1.0 90 -

Independent CW Doppler Probes*


* Independent probe connection unit EU-9184 and EU-9187B are necessary.
Frequency Optional accessories
Application Model
range (MHz)
Cardiac Adult UST-2265-2 2.0 -
Cardiac Adult UST-2266-5 5.0 -

28
ARIETTA850DI-V50
Electrical Radial Probes
Frequency Scanning angle
Application Model Optional accessories
range (MHz) (degrees)
Transrectal R41R 10.0-5.0 360 Waterproof Case
WP-001

*1: Needle Guide Kit(CIVCO) is necessary.


*2: Necessary to perform RVS.
*3: Necessary to perform RVS with an external magnetic sensor.
*4: One of puncture adapter is necessary to use needle.
*5: CIVCO products are manufactured and sold by CIVCO Medical Solutions. It does not suggest that CIVCO
products describe in this specification are commercially available in all countries and regions.

29
ARIETTA850DI-V50
Probes and available functions
Basic Functions

Compound
eFocusing PLUS
DeepInsight
Trapezoid
Wide Scanning
Wide Scanning PLUS
B steer
eFocusing
Acoustic Noise Reduction

Near-field Noise Reduction

Real-time Biplane
OMNI mode
FAM
TGC Enh(B)
TGC Enh(Color)
TDI mode
HI Frame(B)
HI Frame(Color)
Puncture Guide Line
Needle Emphasis
Brachy Grid Display
CW mode
Assist Line
THI(FmT)
THI(WbT)
THI(HdT)
Dual Gate Doppler
STIC
Stress echo
C22K          

C22P           

C23              

C23RV              

C251               

C252                    

C25P           

C35                   

C41B          

C41RP         

C41V1            

C42            

C42K         

C42T       

C421             

CC41R         

CC41R1          

R41R   

L31KP          
 
L34               
 
L35                 
 
L43K          
 
L441                
 
L44K          
 
L46K1          
 
L51K          
 
L53K           

30
ARIETTA850DI-V50
Compound
eFocusing PLUS
DeepInsight
Trapezoid
Wide Scanning
Wide Scanning PLUS
B steer
eFocusing
Acoustic Noise Reduction

Near-field Noise Reduction

Real-time Biplane
OMNI mode
FAM
TGC(B)
TGC(Color)
TDI mode
Hi Frame(B)
Hi Frame(Color)
Puncture Guide Line
Needle Emphasis
Brachy Grid Display
CW mode
Assist Line
THI(FmT)
THI(WbT)
THI(HdT)
Dual Gate Doppler
STIC
Stress echo
 
L44LA          

L44LA1    
 
L55               
 
L64                
 
SML44             

S121               

S31          

S42          

S3ESL1        

S3ESEL        

S3ESCLS        

VC34             

VC35             

VC41V         

MXS1            

C41L47RP(CV)        
 
C41L47RP(LN)           

CL4416R(CV)        
 
CL4416R(LN)           
  
CL4416R1(CV)            
  
CL4416R1(LN)          

UST-2265-2 

UST-2266-5 

31
ARIETTA850DI-V50
Optional Functions

CHI(Low)

CHI(Mid)

CHI(High)

Panoramic

RTE

SWM

iATT

SWE

RVS

Needle Tracking

Cardiac 3D

Real time 3D

eTRACKING

FMD

Wave Intensity

CHI-eFLOW

DFI mode
C22K   

C22P     

C23       

C23RV       

C251           

C252            

C25P     

C35       

C41B     

C41RP

C41V1     

C42   

C42K 

C42T     

C421        

CC41R     

CC41R1  

R41R 

L31KP 

L34        

L35         

L43K    

L441         

L44K    

L46K1    

L51K    

L53K 

32
ARIETTA850DI-V50
CHI(Low)

CHI(Mid)

CHI(High)

Panoramic

RTE

SWM

iATT

SWE

RVS

Needle Tracking

Cardiac 3D

Real time 3D

eTRACKING

FMD

Wave Intensity

CHI-eFLOW

DFI mode
L44LA    

L44LA1

L55          

L64         

SML44       

S121   

S31

S42

S3ESL1

S3ESEL

S3ESCLS

VC34 

VC35 

VC41V  

MXS1 

C41L47RP(CV)  

C41L47RP(LN)   

CL4416R(CV)  

CL4416R(LN)   

CL4416R1(CV)  

CL4416R1(LN)   

UST-2265-2

UST-2266-5

33
ARIETTA850DI-V50
34
ARIETTA850DI-V50
・ The specifications are subject to change without notice.
・ The standard components and optional items differ depending on the country.
Not all products are marketed in all countries.
Please contact your local FUJIFILM Healthcare Corporation distributors for details.
・ ARIETTA, DeepInsight, HdTHI, eFocusing, HI REZ, Carving Imaging, Real-time Tissue Elastography, Real-
time Virtual Sonography, 4Dshading and LISENDO are registered trademarks or trademarks of FUJIFILM
Healthcare Corporation in Japan and other countries.
・ ALOKA is a registered trademark or trademark of Nippon RayTech Co., Ltd. in Japan and other countries.
・ McAfee is a registered trademark or trademark of McAfee LLC in the United States and other countries.
・ DICOM is a registered trademark of the National Electrical Manufacturers Association in the United States
for its standards publications relating to digital communications of medical information.
・ Excel is a registered trademark or trademark of Microsoft Corporation in the United States and other
countries.
・ SYNAPSE and VINCENT are registered trademarks or trademarks of FUJIFILM Corporation in Japan and
other countries.Ueno
・ TOMTEC, LV-Function, Beutel and D’Art are registered trademarks or trademarks of TOMTEC Imaging
Systems GmbH in Germany and other countries.
・ This document is applicable to ARIETTA 850

9-7-3, Akasaka, Minato-ku, Tokyo 107-0052, Japan

35
ARIETTA850DI-V50