UOP172
UOP172
SCOPE
This method is for determining carbon dioxide, oxygen and carbon monoxide in flue gases. The lower
limit of detection is 0.2 mol-%. If hydrogen sulfide or other acidic compounds are present, the carbon
dioxide determination will require correction. This method is designed as a rapid control room test. UOP
Method 539, a gas chromatographic technique, is the recommended laboratory test, provided the separation
of oxygen and argon is not required.
OUTLINE OF METHOD
A measured volume of gas is admitted into a series of pipets for the successive removal of the individual
components by absorption with selective liquid reagents. Volume measurements are taken with a buret
before and after each absorption. The decrease in volume as the gases are absorbed represents the amount of
the particular component present in the sample. The results are expressed in mole-percent.
APPARATUS
References to catalog numbers and suppliers are included as a convenience to the method user. Other
suppliers may be used.
Beaker, 1000-mL
Flue gas analyzer (see Figure), Master Model 211, Burrell Corporation, Cat. No. 39-106, including the
following replacement parts:
Absorption pipets, Nos. 1, 2, 3 and 4 (regular bubbler type), Cat. No. 40-105-10
UOP Methods are available through ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken PA 19428-2959,
United States. The Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at
[email protected], 610.832.9555 FAX, or 610.832.9585 PHONE.
2 of 6
Tube, gas sampling, A (see Figure), 250-mL, Fisher Scientific, Cat. No. 10-921B (optional, not necessary
if analyses are run at the source)
Tubing, chemical-resistant, black rubber, 4.8-mm (3/16-inch) ID, 2.4-mm (3/32-inch) wall, Fisher
Scientific, Cat. No. 14-158C
Tubing, transparent, amber latex, 4.8-mm (3/16-inch) ID, 1.6-mm (1/16-inch) wall, Fisher Scientific, Cat.
No. 14-178B
All reagents shall conform to the specifications established by the Committee on Analytical Reagents of
the American Chemical Society, when such specifications are available, unless otherwise specified.
References to water shall mean distilled water.
References to catalog numbers and suppliers are included as a convenience to the method user. Other
suppliers may be used.
Ammonium chloride, 99.9% minimum purity, Fisher Scientific, Cat. No. A661
Copper, wire or turnings, 99% minimum purity, Fisher Scientific, Cat. No. C575
Cuprous chloride, 90.0% minimum purity, Fisher Scientific, Cat. No. C457
Potassium hydroxide, pellets, 85.0% minimum purity, Fisher Scientific, Cat. No. P250
Sodium sulfate, anhydrous, 99.5% minimum purity, Fisher Scientific, Cat. No. S421
Sulfuric acid, concentrated, 95.0-98.0% purity, Fisher Scientific, Cat. No. A300
PREPARATION OF REAGENTS
The following directions are for the preparation of the reagents in quantities of one liter each. Mix the
reagents in one-liter beakers and store the surplus in one-liter glass-stoppered bottles. The pipets in which
they are used are referred to in numerical order, 1 being closest to the buret (see Figure).
172-89
3 of 6
Weigh 513 ± 0.01 g potassium hydroxide into 805 ± 5 mL of water. Mix thoroughly. Add 170 ± 5 mL of
this reagent to pipet 1 to absorb carbon dioxide, hydrogen sulfide and acidic gases.
Weigh 891 ± 0.01 g potassium hydroxide into 625 ± 5 mL of water. Mix thoroughly. Add 140 ± 5 mL of
this solution to pipet 2. Weigh 15 ± 0.01 g of pyrogallol into 25 ± 1 mL of water. Mix thoroughly. Add the
entire quantity of the pyrogallol solution to pipet 2 and stopper immediately. This solution is used to absorb
oxygen.
Weigh 250 ± 0.01 g ammonium chloride, 200 ± 0.01 g cuprous chloride and 50 ± 0.01 g copper wire, or
turnings, into a one-liter glass-stoppered bottle. Add 750 ± 5 mL of water. Let stand for at least 24 hours.
Decant 125 ± 5 mL into each of pipets 3 and 4. Add 45 ± 1 mL of ammonium hydroxide and 10 ± 0.01 g of
copper turnings to each pipet. Pipets 3 and 4 are used for the determination of carbon monoxide.
Pipet 5, leveling bottle, buret and manifold: Sodium sulfate retaining solution
Add 30 ± 0.01 g of sulfuric acid to 965 ± 5 mL water in a one-liter glass bottle. Mix thoroughly. Add 142
± 0.01 g of sodium sulfate and a sufficient amount of methyl orange to color the solution.
PROCEDURE
Carefully lubricate all stopcocks with a minimum of stopcock grease. Connect the pipets to the manifold
with 30-mm (1-1/4 inch) lengths of transparent amber latex tubing. At all points where there is glass-to-
glass contact, butt the glass terminals firmly against one another to avoid excessive dead space. Refer to the
manufacturer’s instruction manual for additional information on set-up and operation of this apparatus.
Prepare the various reagents according to directions given under PREPARATION OF REAGENTS. Fill
pipets 1 through 4 using the following procedure: Open the pipet to the atmosphere through the manifold.
Remove the rubber stopper and aspirator bag. Pour the appropriate solution into the rear compartment.
Close the manifold stopcock, inflate the aspirator bag slightly with nitrogen gas and replace the stopper.
A clean buret is essential for accurately reading volumes. Clean by draining the buret and rinsing with
acetone followed by water.
Connect the leveling bottle to the buret with chemical resistant black rubber tubing. Fill the leveling
bottle, buret, pipet 5, and manifold with the sodium sulfate retaining solution. Examine the apparatus to
ensure that it is completely filled and no air bubbles are present. Care must be exercised to avoid bubbles in
the pipets and the manifold throughout the measurement.
Before starting an analysis and when removing the residual gas after absorption, the solution in each pipet
must be brought to a reference mark located on the capillary stem approximately 3 mm below the stopcock.
Open each pipet in sequence to the buret and by carefully adjusting the leveling bottle, bring the solution to
the proper point. The aspirator bags are inflated slightly when the pipets are filled to permit drawing the
solution up into the capillary. Approximately 50 mL of nitrogen gas should be maintained in each pipet
when the apparatus is not in use. This will eliminate the freezing or sticking of the stopcocks due to contact
with strong alkaline solutions.
Connect the sample source to the buret. Flush the lines thoroughly and measure exactly 100 mL of the
gas into the buret. To accomplish this, draw approximately 101-102 mL of sample into the buret, and turn
the buret stopcock to connect the buret to the manifold and flushing pipet. Raise the leveling bottle to
172-89
4 of 6
compress the gas to about the 99-mL mark and pinch the rubber tubing between the thumb and forefinger.
Set the leveling bottle down and permit the gas to expand to exactly 100 mL by releasing the pressure on
the rubber tubing. Turn the buret stopcock for an instant to allow the slight excess of gas to vent to the air.
The gas sample is now ready for analysis.
Raise the buret leveling bottle slightly, thus putting the gas in the buret under slight pressure to avoid any
possibility of pulling solution from the pipet into the manifold. Turn the manifold stopcock to connect to
pipet 1, which contains the 33% potassium hydroxide solution. Continue raising the leveling bottle until the
retaining liquid reaches the top of the buret and all the gas has been passed into the pipet. Lower the
leveling bottle bringing the gas back into the buret. Repeat this procedure 3 or 4 times, or until a constant
volume reading is obtained in the buret. Final residual volumes are read after flushing the gas from the
manifold to the buret with sodium sulfate solution from pipet 5. Record the final gas volume. Subtract this
value from 100 to determine the amount of carbon dioxide absorbed.
This applies only in the absence of hydrogen sulfide. If hydrogen sulfide is expected to be present,
determine its content by UOP Method 9. To obtain percent carbon dioxide, subtract the percent hydrogen
sulfide from the percent of the total gas absorbed in the 33% potassium hydroxide solution.
Pass the residual gas from the carbon dioxide determination through the alkaline pyrogallol solution,
pipet 2, in the same manner as described above. Passes in groups of 4 are to be made until a constant
reading is obtained. If more than 12 passes are necessary to obtain a constant reading, change the solution.
Record the final residual volume and indicate the difference as oxygen.
Transfer the gas to pipet 3, which contains the first cuprous chloride solution. After 10 passes, remove the
gas to the buret, read the volume and pass it to pipet 4, containing fresh cuprous chloride reagent. Make
sufficient passes in groups of 3 until a constant reading is obtained. Record the final gas residue volume and
indicate the total contraction of the gas volume in pipets 3 and 4 as carbon monoxide. It is essential to have
fresh cuprous chloride in the second cuprous chloride pipet to remove traces of carbon monoxide. After 10
mL of carbon monoxide has been absorbed in pipet 4, move this pipet to the 3 position. Replace the solution
in pipet 3 and put it in the 4 position. Make certain that sufficient clean copper wire or turnings are present
in the cuprous chloride solutions at all times.
CALCULATIONS
Calculate the individual mol-% of carbon dioxide, oxygen and carbon monoxide by the following
equation:
100 ( A − B )
Component, mol-% =
C
where:
A = volume of gas taken before absorption
B = volume of gas after absorption
C = volume of the total sample taken for analysis
PRECAUTIONS
Sufficient passes to obtain constant readings, as described in PROCEDURE, are required when
contacting the sample with the reagents. If an excessive number of passes is required, change the reagent
involved.
172-89
5 of 6
Do not permit the retaining solution to enter the absorption pipets and allow no reagent to enter the
manifold. Keep the retaining solution slightly acidic as indicated by its red color.
PRECISION
An estimated standard deviation is not reported since insufficient data are available to permit this
calculation.
The elapsed time and labor requirement, without preparation of solutions, are identical, 0.5 hour.
REFERENCES
SUGGESTED SUPPLIERS
172-89
6 of 6
172-89